PROFESSIONAL TORQUE TESTER

PRO-TEST 40, 400 & 1500ER

OPERATORS HANDBOOK (PART No 34237) Issue 5 (ENGLISH)

FOR USE WITH PRO-TESTS FITTED WITH VERSION 37701.101, 37701.102 & 37701.201 SOFTWARE

MODEL NUMBERS COVERED BY THIS MANUAL

43180 PRO-TEST 40 TRANSDUCER 43181 PRO-TEST 400 TRANSDUCER 43189 PRO-TEST 1500ER TRANSDUCER 43184 PRO-TEST DISPLAY MODULE

NORBAR TORQUE TOOLS LTD, Beaumont Road, Banbury, Oxfordshire, OX16 7XJ, UNITED KINGDOM Tel : + 44 (0) 1295 270333, Fax : + 44 (0) 1295 753643

CONTENTS PAGE

Introduction 1 Assembly Diagram 1 Operating Instructions 2 Set Up Menus and Options 4 RS-232-C Serial Data Output Interface 6 Print Inhibit Controller Option 7 Specifications 8 Trouble Shooting 10

PRO-TEST OPERATORS HANDBOOK PAGE 1 OF 10 ISSUE 5 AUGUST 1999

INTRODUCTION The Professional Torque Tester (Pro-Test) has been designed as an accurate, cost effective and easy to operate instrument for testing and calibrating all types of torque wrenches and torque limiting screw drivers.

Pro-Test transducers are mounted in aluminium cast housings along with all the necessary electronics to produce an output for the display module and RS-232-C interface. The housings can be mounted for either horizontal, clockwise, or anti-clockwise torque testing.

The function of the front panel buttons is identified on the display to facilitate quick and easy selection of the operating mode most suitable for the tool to be tested, i.e. track, peak memory with manual reset, peak memory with auto reset, first peak memory with manual reset, or first peak memory with auto reset. These buttons are also used to drive the Set-up Menus if changes to the operational parameters of the Pro-Test are required. These include, language of operation, measurement units to be enabled/disabled, RS-232-C communication parameters, auto reset hold time, first peak memory sensitivity and modes of operation to be enabled/disabled. The Zero button is only active in the track mode.

Pro-Test transducers are calibrated in the clockwise direction only as standard, the anti-clockwise direction of operation will be inhibited unless the transducer has been additionally calibrated in this direction. The direction of use is identified on the display and RS232C output

Optional extras include a Mounting Bracket, 1 sq x 36mm Hex Socket, Print Inhibit Controller, Data Printer and 12 Volt DC power supply for use in vehicles.

ASSEMBLY DIAGRAM

70

2 HOLES M10

2 HOLES 10

43180 PRO-TEST 40 43181 PRO-TEST 400 43189 PRO-TEST 1500ER

29067 3/8 TO 10mm SOCKET 29134 1/2 TO 22mm SOCKET 29189 3/4 TO 22mm SOCKET 29179 3/4 TO 36mm SOCKET 29190 1 TO 36mm SOCKET

62198 (OPTIONAL)

38796

38793

43184

25429

PRO-TEST OPERATORS HANDBOOK PAGE 2 OF 10 ISSUE 5 AUGUST 1999

OPERATING INSTRUCTIONS

1. Securely mount the Pro-Test transducer in the plane of operation required to a bench top, wall or the additional mounting bracket (part number 62198) with the 2 fixing bolts supplied.

IMPORTANT ! The un-bolted end of the unit, must always be in contact with a supporting surface.

MOUNTING POSITION TO ENABLE CLOCKWISE TESTING. (bolts to be hand tightened with allen key provided)

MOUNTING POSITION TO ENABLE ANTI-CLOCKWISE TESTING. (bolts to be hand tightened with allen key provided)

MOUNTING POSITION TO ENABLE CLOCKWISE AND ANTI-CLOCKWISE TESTING

IMPORTANT ! Bolts must be torque tightened to 50 N.m, and replaced when removed.

2. Locate the 2 holes in the bottom of the Pro-Test Display Unit on to the bolt heads protruding from the transducer housing. Fit connecting lead between Display Unit and Transducer.

Vertical Testing

(bench mounted)

Vertical Testing

(wall mounted)

Vertical Testing

(bench mounted)

Vertical Testing

(wall mounted)

Horizontal Testing

(bench mounted)

Vertical Testing

(wall mounted)

PRO-TEST OPERATORS HANDBOOK PAGE 3 OF 10 ISSUE 5 AUGUST 1999

3. Plug Power Supply into back of transducer housing then plug mains lead into Power Supply. If a mains plug is not fitted, wire as follows :

BROWN-LIVE BLUE-NEUTRAL GREEN / YELLOW-EARTH

WARNING! It is important that live, neutral and earth are all connected between the Pro- Test and mains power supply. If no earth is available (2 wire mains power supply) it is recommended that a separate earth is connected between the transducer housing (the bottom right hand fixing screw in the corner of the back panel is ideal) and a suitable earth.

4. If power supply has an ON/OFF fitted, switch to ON.

5. Wait 2 seconds for initialisation and allow 5 minutes for the instrument to warm up and stabilise, then press ZERO.

6. Select required units of measurement. Pressing the SELECT UNITS button once will step onto the next available unit of measurement.

7. Select mode of measurement required. Pressing the MODE button once will step onto the next available.

BENDING BEAM WRENCHES, DIAL TYPE WRENCHES & TORQUE SCREWDRIVERS:- ________

INSTRUMENT MODE : Peak memory with manual reset or Peak memory with auto reset

PRESET (CLICK) WRENCHES & TORQUE SCREWDRIVERS:- _____________________________

INSTRUMENT MODE : First peak memory with manual reset or First peak memory with auto reset

8. Fit adapter supplied to hex drive of transducer, then tool to be tested to square of adapter.

9. Operate tool to be tested.

10. If the Pro-Test is loaded in a MANUAL RESET mode, press RESET to clear the display after the torque has been removed. If an AUTO RESET mode is used, the display will clear automatically after the torque has been removed.

NOTE:- If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment could be impaired.

PRO-TEST OPERATORS HANDBOOK PAGE 4 OF 10 ISSUE 5 AUGUST 1999

SET UP MENUS AND OPTIONS

ZERO

To enter SET UP mode, press SELECT UNITS & PRINT buttons simultaneously, then release upon which the SOFTWARE VERSION NUMBER will be displayed for 2 seconds.

ZERO

Repeatedly press the button until desired language of operation flashes, then press EXIT.

ZERO

To enable/disable a unit of measurement repeatedly press the button until the desired unit flashes then press / X ( = enabled, X = disabled). When all selections have been made, press EXIT.

MODE> TRACK PRINT

SELECT UNITS

LANGUAGE MENU

MEASUREMENT UNITS MENU

EXIT

EXIT

ENGLISH DANSK FRANCAIS NEDERLANDS DEUTSCH SUOMI ITALIANO NORSK ESPANOL SVENSKA

PORTUGUES

N.m X lbf.in dN.m X ozf.in cN.m kgf.m X lbf.ft X kgf.cm

/ X

PRO-TEST OPERATORS HANDBOOK PAGE 5 OF 10 ISSUE 5 AUGUST 1999

ZERO

Select desired parameter by repeatedly pressing until text flashes. To change the setting, press the bottom right hand button, this button changes its function as each parameter is selected. When all selections have been made, press EXIT.

ZERO

Select desired functional parameter by repeatedly pressing until text flashes. To change the setting, press the bottom right hand button, this button changes its function as each parameter is selected. When all selections have been made, press EXIT.

ZERO

To enable/disable a mode of measurement repeatedly press until the desired mode flashes, then press / X. There are 4 modes of measurement of which 2 will be displayed at any one time. When all selections have been made, press EXIT, this will take the Pro-Test out of SET UP mode.

RS 232 C PARAMETERS MENU

FUNCTIONAL PARAMETERS MENU

MODE MENU

EXIT

EXIT

EXIT

PARITY OFF DATA BITS 8 STOP BITS 2 LEADING CHARACTER BLANK OUTPUT UNITS YES

AUTO RESET HOLD TIME 3 SECONDS FIRST PEAK SENSITIVITY HIGH

MODE> PEAK MEMORY & MANUAL RESET

MODE> PEAK MEMORY & AUTO RESET

EVEN/ODD/OFF

1/2/3/4

/ X

PRO-TEST OPERATORS HANDBOOK PAGE 6 OF 10 ISSUE 5 AUGUST 1999

RS-232-C SERIAL DATA OUTPUT INTERFACE

Output of both measured value and units of measurement (as shown on the display) are in a familiar serial data format for communication with computers, printers, etc.

Data is output on the RS-232-C interface automatically when the memory auto-reset mode timer operates, when the 'MEMORY RESET' button is pressed, or by pressing the PRINT button when in track mode. Output can be requested externally when the 'request to send pin is taken HIGH.

RS-232-C INTERFACE SPECIFICATIONS :- _____________________________________________

Data rate fixed at 1200 baud.

Control Word Options - Parity = odd, even or off. - data bits = 7 or 8. - stop bits = 1 or 2.

All Options are menu selectable (see page 4) and are initially factory set to 8 data bits, 2 stop bits, and no parity

Transmitted data voltage levels +9 volts to -9 volts.

RS-232-C OUTPUT PIN CONNECTIONS :- ______________________________________________

PIN No FUNCTION 1 Arm LED for print inhibit option.

2 Received data (from P.C)

3 Transmitted data (from Pro-Test)

4 Mode switch for print inhibit option.

5 Signal ground 0V.

6 Arm switch for print inhibit option.

7 Request to send to Pro-Test (Low to High transition)

8 + 5 volts (from Pro-Test) WARNING: Not for powering external devices.

9 Calibration Mode Enable input (Active Low)

RS-232-C DATA OUTPUT EXAMPLES :- _______________________________________________

Code : DP=Decimal Point. CR=Carriage Return. SP=SPace.

1. Using the Pro-Test 400 set to the standard data string. Reading 368.4 N m (clockwise).

3 6 8 DP 4 SP N DP m CR

2. Using the Pro-Test 400 set to the standard data string. Reading 45.2 N m (anti-clockwise).

- 4 5 DP 2 SP N DP m CR

3. Using the Pro-Test 1500ER set to the standard data string. Reading 1108 lbf ft (clockwise).

1 1 0 8 SP l b f DP f t CR

PRO-TEST OPERATORS HANDBOOK PAGE 7 OF 10 ISSUE 5 AUGUST 1999

MITUTOYO DATA PROCESSORS :- ___________________________________________________

The instrument can be configured to communicate to Mitutoyo DP3DX, DP7, QM1000 and QM5000 families of data processors.

For DP3DX, DP7, QM1000 and QM5000 families, the units of measurement must be inhibited.

For DP3DX and DP7 families, a '+' character must be added to the start of the data stream This is in addition to the units of measurement being inhibited.

See SET-UP MENU section.

WIRING THE PRO-TEST TO DATA PROCESSORS:-

PRO-TEST RS-232-C OUTPUT

MITUTOYO RS-232-C INPUT

CONNECTOR COVER

CABLE SCREEN

PIN 3 PIN 3 PIN 5 PIN 7

For DP3DX & DP7 link pins 1 to 5 and 4 to 8 on Mitutoyo RS-232-C input. For QM1000 & QM5000 link pins 4 to 5 and 6 to 8 on Mitutoyo RS-232-C input.

For more information please consult your Mitutoyo data processor manual. MITUTOYO is a registered trade mark of Mitutoyo (UK) Ltd.

PRINT INHIBIT CONTROLLER OPTION PART NUMBER 60167

INTRODUCTION :- _________________________________________________________________

The print inhibit controller is a remote, hand held device for controlling the RS-232-C output from the Pro-Test. This option can be retro fitted to any Pro-Test transducer, model numbers 43180-43183.

This option can control unwanted RS-232-C data input to printers, calibration and data collection systems.

OPERATION :-_____________________________________________________________________

Position of ' PRINT NORMALLY/ PRINT WHEN ARMED switch :-

SWITCH POSITION

STATE OF INDICATOR

ACTION

'PRINT NORMALLY'

UNARMED The RS-232-C output acts as normal. Every time an output is requested it will be issued.

'PRINT WHEN ARMED

UNARMED The RS-232-C output is inhibited, so will not function.

PRINT WHEN ARMED

ARMED (Having pressed button on controller)

When the ' ARM ' button is pressed, the arm led will light so indicating that the next request to send data will be acted upon.

Upon the next data output taking place, the arm led will turn off and the RS-232-C output will again be inhibited.

To obtain further data output either press the ARM button or switch to PRINT NORMALLY.

PRO-TEST OPERATORS HANDBOOK PAGE 8 OF 10 ISSUE 5 AUGUST 1999

SPECIFICATIONS

PRO - TEST TRANSDUCER SPECIFICATION :-__________________________________________

PRO-TEST 40 PRO-TEST 400 PRO-TEST 1500ER

RANGE OF OPERATION

0 to 40 N.m 0 to 400 N.m 0 to 1500 N.m

CALIBRATION RANGE

5% to 100% of full scale. 2 to 40 N.m.

5% to 100% of full scale. 20 to 400 N.m.

2% to 100% of full scale. 30 to 1500 N.m.

TRANSDUCER OVERLOAD CAPACITY

150% of Pro-Test calibration range

(60 N.m).

150% of Pro-Test calibration range

(600 N.m).

150% of Pro-Test calibration range

(2250 N.m).

DIMENSIONS Height x Width x Depth (mm)

106 x 106 x 185. 106 x 106 x 185. 106 x 106 x 185.

DRIVE ADAPTERS

3/8 sq to 10 mm hex socket

1/2 sq to 22 mm hex socket

3/4 sq to 22 mm hex socket

3/4 sq to 36 mm hex socket

WEIGHT 3.9 kg (8.6 lb). 4.0 kg (8.9 lb). 4.9 kg (10.8 lb).

PRO-TEST DISPLAY UNIT SPECIFICATION :-___________________________________________

READOUT 240 x 64 pixel dot matrix display giving direction of torque applied, 4 full digits, direct readout of engineering units, stop loading indicator and front panel button identification in all modes of operation and set up menus.

DISPLAY UPDATE RATE 2 per second (2Hz).

RESOLUTION 1 Digit in 9999.

WEIGHT 2.4 kg (5.3 lb)

DIMENSIONS 128 mm high x 185 mm wide x 102 mm deep.

PRO-TEST OPERATORS HANDBOOK PAGE 9 OF 10 ISSUE 5 AUGUST 1999

SYSTEM SPECIFICATIONS :- ________________________________________________________

ACCURACY +/-1% of reading over calibration range.

ZERO SUPPRESSION TRACK None.

ALL MEMORY MODES Suppressed from 0 to approximately 0.5% of transducer calibration range.

UNITS OF MEASUREMENT PRO-TEST 40 N.m, dN.m, cN.m, lbf.ft, lbf.in, ozf.in, kgf.m, kgf.cm. (Menu Selectable).

PRO-TEST 400 N.m, dN.m, lbf.ft, lbf.in, kgf.m, kgf.cm. (Menu Selectable).

PRO-TEST 1500ER N.m, lbf.ft, lbf.in, kgf.m. (Menu Selectable).

MEMORY AUTO-RESET TRIGGER 2.5%(High), 5%(Med), or 10%(Low) of reading (Menu Selectable).

MEMORY AUTO RESET HOLD TIMES 1, 2, 3 or 4 seconds (Menu Selectable).

OPERATING TEMP RANGE +5C to +40C.

STORAGE TEMP RANGE -20C to +70C.

MAXIMUM OPERATING HUMIDITY 85% Relative Humidity @30C.

POWER REQUIREMENTS 90 to 264 Volts AC at 50-60 Hz.

POWER CONSUMPTION 12.0 W - maximum.

POWER CABLE 2 meters (6 ft 6 ins) long minimum.

POWER PLUG FUSE (if fitted) 1 Amp

CASE MATERIALS / FINISH Display and Transducer housings engineered in aluminium castings. Finished in powder coated paint.

ENVIRONMENT Indoor use within a light industrial environment.

ELECTROMAGNETIC COMPATIBILITY In conformance with EN 50081-1 : 1992 (EMC) DIRECTIVE & EN 50082-1 : 1992.

LOW VOLTAGE DIRECTIVE In conformance with EN 61010-1 : 1993. To environmental conditions Pollution Degree 2 & Installation Category (Overvoltage Category) II.

TIME/DATE COMPLIANCE This equipment does not utilise time or date functions and thus will not be affected by the issues of date compliance in the future.

Due to continuous improvement all specifications are subject to change without prior notice.

PRO-TEST OPERATORS HANDBOOK PAGE 10 OF 10 ISSUE 5 AUGUST 1999

TROUBLE SHOOTING

1. No display when power is switched on.

a) Check the Power Supply Module is correctly plugged into the back of the transducer housing.

b) Check fuse in mains power plug.

c) Check the lead is fitted between the display module and transducer.

2. Readings in First Peak mode are inconsistent.

Change First Peak sensitivity to be less sensitive, see SET-UP MENU section (page 4).

3. Required units of measurement are not selectable.

a) Conversion from transducer standard units to required units could not be displayed within available digits on display, thus conversion is not allowed.

b) Units required have not been enabled. See SET-UP MENU section (page 4).

4. RS-232-C data output not communicating with other equipment.

a) Check that control word on the Pro-Test and the equipment receiving data match. See page 6.

b) Check that the baud rate is set to 1200 baud on the equipment receiving data.

c) Check that the connecting lead is wired correctly at both ends, see page 6.

d) Check if equipment receiving data requires the units of measurement inhibited or a leading

character. This is applicable when interfacing to Mitutoyo equipment, see page 7.

5. Display shows zero and does not alter when torque applied.

Switch power off or disconnect at the mains power socket, wait 2 seconds, then power up the Pro-Test again.

PRO-TEST CALIBRATION AND REPAIR :-______________________________________________

To maintain the specified accuracy it is recommended that the Pro-Test transducers are recalibrated at least once per year.

Recalibration and repair should be carried out at Norbar or by a Norbar approved agent, where all the facilities to ensure the instrument is functioning at maximum accuracy are available.

Alternatively, procedures for calibrating the Pro-Test to the specified accuracy can be found in the Professional Torque Tester (Pro-Test) service manual Part No. 34240.

CLEANING :- ______________________________________________________________________

Do not use abrasives or solvent based cleaners. We recommend a propriety brand of foam based fabric / vinyl cleaner. Use a sof

Manualsnet FAQs

If you want to find out how the 34237 Norbar works, you can view and download the Norbar PRO-TEST 400 34237 Tester Operator's Handbook on the Manualsnet website.

Yes, we have the Operators Handbook for Norbar 34237 as well as other Norbar manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Operators Handbook should include all the details that are needed to use a Norbar 34237. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Norbar PRO-TEST 400 34237 Tester Operator's Handbook is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Norbar PRO-TEST 400 34237 Tester Operator's Handbook consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Norbar PRO-TEST 400 34237 Tester Operator's Handbook free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Norbar PRO-TEST 400 34237 Tester Operator's Handbook, simply download the document to your computer. Once downloaded, open the PDF file and print the Norbar PRO-TEST 400 34237 Tester Operator's Handbook as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.