User Guide

Precision Milliohm Meters Models 380460 and 380462

380460/380462-en-GB_v2.5 8/15 2

8

1

2

3

4

67

5

Introduction Congratulations on your purchase of the Extech Milliohm Meter Model 380460 (110V) or Model 380462 (220V). The milliohm meters are enclosed in rugged carrying cases for durability. Measurements are made with Kelvin test leads (supplied) for optimal accuracy in five (5) selectable ranges. Careful use of this meter will provide years of reliable service.

Meter Description

1. LCD Display

2. Zero adjust knob

3. Range Dial

4. Power switch

5. 200m select switch

6. AC power cord

7. Kelvin lead to meter connections

8. Carrying case

380460/380462-en-GB_v2.5 8/15 3

Operation

Basic Principles of 4-wire Measurements

For each range, the meter provides a specific amount of test current (refer to General Specifications) which flows from the HI to the LO meter terminal and therefore from the HI to the LO clip lead (Refer to the diagram). This is the current that ultimately passes through the device under test (RX in diagram below).

Once the test current is applied to the device under test, clip leads Rx1 & Rx2 measure the voltage drop across the device under test. The following equations detail how the meter accomplishes its measurement tasks.

Measurement Equations

Vx = Is x Rx; Where Vx is the voltage (measured by the meter) across the device under test; Is is the test current; Rx is the resistance of the device under test. From Vx = Is x Rx, the meter moves to the next step which is: Rx = Vx/Is. With this equation the meter determines the resistance of the device under test. Note that the measured resistance (Rx1 & Rx2) is not affected by stray resistance since the test current is supplied directly to the device under test. This is the advantage of the 4-wire Kelvin lead configuration over 2-wire methods which introduce errors into low resistance measurements.

International Symbols

LO HIRx1 Rx2

White

bands

Rx Device under test

To Meter

380460/380462-en-GB_v2.5 8/15 4

Powering the Meter

Connect the meter to the correct power source (110 or 220VAC); the Model 380460 uses 110VAC and the Model 380462 uses 220VAC (note that these meters are NOT interchangeable). Do not apply voltage to the meter input measurement terminals. Meter damage may result.

Measurement Procedure

1. Rotate the Range select switch to the desired range.

2. When using the 200m range, put the 200m select switch to the 200m position; for all other measurement ranges put this switch in the opposite position

3. Perform a Zero Adjustment per the following:

Short the Kelvin Clips as shown in the diagram below.

(Make sure that the clips are clean)

Rotate the Zero Adjust Knob until the meter indicates zero units.

4. Either attach the test leads onto a component or attach the leads between two points on a circuit.

5. Observe the meter reading.

380460/380462-en-GB_v2.5 8/15 5

Specifications

General Specifications

Circuit Custom one-chip LSI microprocessor circuit

Display 18 mm (0.7") LCD with 2000 display counts

Measurement terminals 4-Terminal Kelvin type

Measurement ranges Five (5) ranges (200m, 2, 20, 200, 2000 ohms)

Zero Adjust 50 count adjustment

Sampling Time Approximately 0.4 seconds

Over range indication Indication of "1 "

Operating Temperature 0 C to 50 C (32 F to 122 F)

Operating Humidity 80% Relative Humidity max.

Power Supply 110V (380460) or 220V (380462) 15%, 50/60Hz

Power Consumption Less than 2 VA

Weight 694g (1.53 lbs.)

Dimensions 167 x 122 x 85 mm (6.6 x 4.8 x 3.4") with cover

Accessories Power Cable and 4-wire Kelvin clip leads(380465)

Range Specifications

Range

Resolution

Test

Current

Accuracy

Open Circuit

Voltage

200 m 0.1m 100mA 0.75% + 4 digits 4.4V

2 1m 10mA 0.75% + 2 digits 3.6V

20 10m 10mA 0.75% + 2 digits 3.6V

200 0.1 1mA 0.75% + 2 digits 3.5V

2000 1 1mA 0.75% + 2 digits 3.5V

Note:

Manualsnet FAQs

If you want to find out how the Extech Flir works, you can view and download the Flir Extech 380462 Tester User Guide on the Manualsnet website.

Yes, we have the User Guide for Flir Extech as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Guide should include all the details that are needed to use a Flir Extech. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir Extech 380462 Tester User Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir Extech 380462 Tester User Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir Extech 380462 Tester User Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir Extech 380462 Tester User Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir Extech 380462 Tester User Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.