Flir Extech DCP42 Power Supply User Guide PDF

1 of 19
1 of 19

Summary of Content for Flir Extech DCP42 Power Supply User Guide PDF

USER GUIDE 

Model DCP42  160W Laboratory Grade Constant Power,  Switching Mode Power Supply 

DCP42enGB_V1.5   12/14 2

Table of Contents 

1.  INTRODUCTION AND FEATURES 

2.  SAFETY 

3.  POWER SUPPLY DESCRIPTION 

4.  CONTROL MODES 

4.1 Normal Mode 

4.2 Preset Mode 

4.3 Set Mode  4.3.1 Setting the V/t value  7  4.3.2 Activating the Remote Control Mode  8  4.3.3 Deactivating the Remote Control Mode  8  4.3.4 Factory Default Reset  8 

5.  OPERATION 

5.1 Powering the Supply 

5.2 Switching the Power Supply ON and SelfTest Information 

5.3 Using the Control knobs 

5.4 Connecting the Power Supply and running a Test 

5.5 Keypad Lockout function 

5.6 Manual Output ON/OFF 

5.7 Ammeter/Wattmeter Selection 

6.  SUPPLEMENTAL FUNCTIONS  10 

DCP42enGB_V1.5   12/14 3

6.1 V/t Function  10  6.1.2 Setting the DC generator voltage level  10  6.1.2 Setting the t  10  6.1.3 Generating RampUp or RampDown DC output cycles  10 

6.2 Function A/B/C Waveform Generator  11  6.2.1 Set FUNC (A/B/C):  11  6.2.2 Generating Wave Form  11 

6.3 Upper Voltage and Current Limits  12 

6.4 Remote Sensing  12  6.4.1 Connection  12  6.4.2 Disconnection  12 

6.5 Remote Control Operation  13  6.5.1 REMOTE CONTROL  METHOD 1: Using two external DC voltage sources  13  6.5.2 REMOTE CONTROL  METHOD 2: Using two 0 to 5K Ohm variable resistors 13  6.5.3 Remote Control Output ON/OFF  14 

7.  PC SOFTWARE PROGRAM  14 

8.  AUTOMATIC OUTPUT STEP PROGRAMS  14 

8.1 External Timed Program Window  15 

8.2 Output ON/OFF function  15 

8.3 Automatic Timed Output Operation  15 

9.  APPENDIX  ADDITIONAL OPERATIONAL EXAMPLES  16 

10.  SPECIFICATIONS  17 

11.  TROUBLESHOOTING  19 

DCP42enGB_V1.5   12/14 4

1. Introduction and Features  Thank you for selecting the Extech DCP42 Switching Mode Power Supply.  This newly redesigned  laboratory grade power supply differs from a conventional power supply in that it can calculate  and change the voltage and current limit points according to the available maximum power.  Therefore the maximum limits of the voltage and current are changeable according to the rated  power. A conventional power supply has only a set maximum voltage and current such that all of  the operating V & A must fall within these limits. The DCP42 offers cost saving and space saving  advantages as one constantpower power supply can do the work of several conventional power  supplies.    The DCP42 is highly efficient, incorporates upgraded SMPS circuitry with small form factor, benefits  from an automatic crossover for CC and CV, has three (3) Voltage/Current presets, and can be  remote controlled. The DCP42 is perfect for solving a variety of loading conditions and applications.  Features include the following:   

Dual action (Course/Fine) tuning provides smooth, accurate, and quick setting of  Voltage/Current 

Setting, changing, and verifying the current limit level is convenient and can be  performed without output pole spiking 

The Remote Control feature permits output ON/OFF and Voltage/Current adjustments  without the need to touch the front panel of the power supply 

The USB port offers PC connectivity for programming and running ramp/soak cycles with  20 programmable sets of Voltage/Current and varying test durations (up to 999 cycles) 

The DC ramp and waveform generator can either be panel set or set by the supplied  software which previews the final waveform. 

Applications and industries where the DCP42 can be employed include laboratory,  telecommunications, production testing, field testing, DC network powering, and more 

Three (3) user programmable Voltage/Current settings offer quick recall of frequently  used test settings.   

This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable  service. Please visit our website (www.extech.com) to check for the latest version of this User  Guide, Product Updates, and Customer Support. 

DCP42enGB_V1.5   12/14 5

2. Safety  This manual contains important safety and operation instructions for correct use of the power  supply. Read through the manual and pay special attention to the markings and labels of this unit  and equipment to be connected. 

Pay special attention to these two types of notices used in this manual: 

WARNING : Failure to observe this warning may cause injury to persons and damage to power  supply or connected equipment. 

CAUTION : Failure to observe this warning may result in damage to equipment and  improper functioning of the power supply. 

WARNING  

Do not use this power supply near water 

Do not operate or touch this power supply with wet hands 

Do not open the casing of the power supply when it is connected to ac mains 

Refer all servicing to qualified service personnel only 

Before replacing the AC fuse, determine the cause and clear up the cause first 

Replace the AC fuse with the same type and rating as the original fuse 

The maximum output voltage of the DCP42 exceeds 60VDC, avoid touching the metal 

contact parts at the output terminals 

CAUTION  

Use a grounded 3 pin AC source 

This unit is intended for indoor use only 

Do not operate or place this unit in a humid, dusty location 

Do not operate this unit in direct sunlight or near any heat source 

Before plugging into local AC mains, check the rating label at the back of the unit 

Do not block any ventilation openings of the unit 

This unit must be used within the specified rating; regular excessive continuous loading may  cause damage to the power supply 

The gauge size of the input power cable must be at least 0.75mm 2 and the total length of 

power cable must not exceed 3m 

Input Fuse Recommended: T3AL250V (3A TimeLag) 

ENVIRONMENTAL CONDITIONS FOR OPERATION 

1080% R.H. (Relative Humidity) 

Altitude up to 2000m (6561 ft.)  

Installation category: CAT 2 

Pollution degree: 2 

Mains supply voltage fluctuation up to 10% of normal voltage   

DCP42enGB_V1.5   12/14 6

3. Power Supply Description   

(1) LED panel meter display with CC/CV/W Indictor 

(2) Remote Control Indicator (Remote Control/ USB mode) 

(3) Output Voltage Control Knob 

(4) Output Current Control Knob 

(5) Power ON/OFF Switch 

(6) 6 button Keypad (with illumination): 

 Preset Keyd 1, 2 & 3 

 Menu Key (Shift Key) 

 Lock/ Unlock Key (A/W meter Key) 

 Output ON/OFF (V/t Key) 

(7) Output Safety Jacks (Rated 10A)

(8) USB Remote Control port (For access to computer to run cyclical operation with programmable  voltage, current, period time and cycle) 

(9) Remote Sensing Terminal 

Warning! Shorting the remote sensing terminal or connecting the sensing terminal with the  polarity reversed will damage the power supply 

 (10) Remote Control Terminal 

(11) AC Inlet socket (with fuse holder) 

DCP42enGB_V1.5   12/14 7

4. Control Modes  There are four control modes for the power supply: NORMAL, PRESET, SET, and REMOTE CONTROL  modes. The power supply is factory preset to Normal Mode with maximum current level CC. 

4.1 Normal Mode  This is the factory default mode; the power supply output Voltage and Current are controlled by  the dual action gain knobs.  1. Push the knobs to toggle the coarse and fine tuning, notice the subtle changes in 

brightness of the related LED.  2. Adjust the knobs to the desired values using coarse and then fine tuning.  3. To check the preset current level, turn the Current Knob lightly in any direction.  4. The display will resume its normal brightness after a few seconds to reaffirm the 

adjustment. 

4.2 Preset Mode  1. There are three preset outputs P1/ P2/ P3 available.  2. Press the Preset button (1, 2, or 3) to enter the preset mode (the buttons light will 

switch ON).  3. The preset values are factory set as shown in the table below and are retained even after 

the unit has been powered down.  4. The user can change these output settings using the voltage and current knobs. 

Recall No.  Output Voltage    Output Current 

P1  5V  Maximum 

P2  13.8V  Maximum 

P3    40V    Maximum 

4.3 Set Mode  Enter the Set Mode by pushing and holding the MENU button for 3 seconds. In the Set Mode the  user can adjust the V/t value, access the remote Control mode, or perform a Factory Reset  recall function. Note: If no selection is made after 10 seconds the power supply defaults to its  prior settings. 

4.3.1 Setting the V/t value  a. Press and hold the MENU button for 3 seconds to access the Set Mode.  b. Turn the Voltage control knob until the panel meter displays dudt Set as shown below and  then press the Voltage control knob to enter the V/t setting mode. 

  c. Adjust the Voltage and Current control knobs as desired to set the V/t value. Further details  are provided in the dedicated section for V/t mode later in this guide.  d. When finished, press the Voltage control knob to confirm and return to Set menu  e. Press the Menu key to exit the Set menu  f. To start the V/t function, press the SHIFT button and then press the V/t button to run,  g. To stop the V/t function, press the SHIFT button and then press the V/t button again. 

DCP42enGB_V1.5   12/14 8

4.3.2 Activating the Remote Control Mode  a. Press and hold the MENU button for 3 seconds.  b. Turn the Voltage control knob until the panel meter displays rC Set as shown below and  then press the Voltage control knob to confirm entry into the remote control mode. 

  c. Turn the Current control knob to select remote YES or NO.  d. Press the Voltage control knob to confirm and to return to the Set menu.  e. Press the Menu button to exit the Set menu.  f. Details of the remote control mode are located in the dedicate Remote Control section later in  this guide.  Note: After activating the Remote Control Mode, all function keys located on the front panel will  be locked (the Lock/Unlock button light switches ON). 

4.3.3 Deactivating the Remote Control Mode  a. Press the Lock/Unlock button to unlock all of the function keys  b. Repeat steps a to c in the Activating the Remote Control Mode section above and set 

Remote Control to NO and then press the voltage control knob. 

4.3.4 Factory Default Reset  a. Press and hold the MENU button for 3 seconds.  b. Turn the Voltage control knob until the panel meter displays FACS SEt as shown below and  then press the Voltage control knob to confirm entry into the Factory Reset mode. 

  c. Turn the Current control knob to select Yes or No.  d. Press the Voltage control knob to confirm and to return to the Set menu.  e. Press the Menu button to exit the Set menu.  f. The Power Supply will reboot and revert to its Factory Default settings. 

DCP42enGB_V1.5   12/14 9

5. Operation  Note: The DCP42 Total Output Voltage Range is 42VDC and the Total Rated Current is 0 to 10ADC  

5.1 Powering the Supply  Check the rating label of the power supply and ensure that it complies with the AC mains voltage  that is to be used. Connect the power supply to the AC Mains using the provided power cord.  

5.2 Switching the Power Supply ON and SelfTest Information  The power supply will perform a series of selftests when it is switched ON. The LED and other  indicators on the front panel will illuminate in sequence. When the cooling fan is being checked,  a high speed wind sound can be heard. After the selftests, the CVV and A LED indicators are lit  and the unit displays voltage and 0.0 current. To find the CC current level, turn the current  control knob one click in either direction. The current display returns to 0.0 after a few seconds.   

5.3 Using the Control knobs  The rotary encoder control knobs have fine and coarse tuning with clicking movement. Push  the knobs to toggle between coarse and fine tuning, notice the subtle changes in brightness of  the related LED. Adjust the knobs to the desired values by using coarse and then fine tuning.  The display will resume its normal brightness after few seconds to confirm an adjustment. 

5.4 Connecting the Power Supply and running a Test  1. Connect the equipment under test to the power supply. Red (+) is connected to the positive 

polarity input of the equipment and Black () is connected to the negative polarity input of  the equipment. 

2. Switch ON the power supply first; the panel meter & green CV Indicator should illuminate.  3. Next, switch ON the equipment under test; the panel meter & green CV Indicator should 

remain lit in green.  4. Testing can now begin.  5. When finished, switch OFF the equipment under test first and then switch off the power 

supply. 

5.5 Keypad Lockout function  1. Press LOCK/UNLOCK to lock or unlock all of the function buttons and control knobs on the 

front panel.  2. The LOCK/UNLOCK button light switches ON when the pushbuttons are locked. 

5.6 Manual Output ON/OFF  1. Press OUTPUT ON/OFF button to manually set the output ON or OFF.   2. The OUTPUT ON/OFF button light switches ON when the output is ON. 

5.7 Ammeter/Wattmeter Selection  1. With the output ON and all presets OFF, press SHIFT and then press PRESET.  2. Press the SHIFT button again and then press the LOCK/UNLOCK button to change from 

Ammeter mode to Wattmeter mode.  3. To change back to Ammeter mode from Wattmeter mode, press the SHIFT button and then 

press the LOCK/UNLOCK button. 

DCP42enGB_V1.5   12/14 10

6. Supplemental Functions 

6.1 V/t Function  There are 3 DC voltage generator settings (A, B, and C).  Preset 1 = A  Preset 2 = B  Preset 3 = C  ab (from voltage level A to voltage level B).  ab (time in seconds from voltage level A to level B, this transit time is adjustable from 0 to 20  seconds) 

6.1.2 Setting the DC generator voltage level  There are 3 generator voltage levels presets (A, B, and C).  Press the Preset button 1, 2 and 3 and adjust the voltage to the desired value.   

6.1.2 Setting the t  There are 6 sets of available, turning the Voltage control knob cycles through the following  sequence: AB BA AC CA BC CB  Using the AB setting as an example, turn the Voltage control knob until Ab appears on the  voltmeter display and then turn the Current control knob to set to desired transit time (5  seconds for example).   

6.1.3 Generating RampUp or RampDown DC output cycles  Ramp up generator example: Set A = 5V; B = 10V; ab = 5 seconds  1. Connect a suitable load to the output and turn ON the output by pressing the OUTPUT 

ON/OFF button; note the output LED indicator.  2. Push Preset 1 and note the output voltage of 5V. Press Preset 2 and note that the 

output voltage rises to 10V in 5 seconds showing a ramp up voltage (the output will  remain at 10V). 

3. Pressing Preset 1 will now generate a ramp down voltage from 10V to 5V in 5 seconds  and will remain at the 5V level. 

Note: The final output voltage level will be the last voltage level generated. 

DCP42enGB_V1.5   12/14 11

6.2 Function A/B/C Waveform Generator  Preset 1 = A  Preset 2 = B  Preset 3 = C  FUNC (A/B/C) sets the duration (from 0 to 600 seconds) that the voltage generator remains ON  at the specified output voltage level before moving to another voltage level. 

6.2.1 Set FUNC (A/B/C):  1. Enter the V/t setting first (Refer to the Set Mode section 4.3.)  2. Turn the Voltage control knob to select the desired function and then turn the Current 

control knob to set the time period (0 to 600 seconds) for the chosen voltage level.  3. Pressing the Current control knob toggles coarse and fine tuning of the time period.  4. To generate waveforms, the Func A/B/C button is used in conjunction with V/t button (the 

time period from one voltage level to another voltage level)  5. Using the above functions simultaneously, 3 reversible pairs (AB & BA, BC & CB, AC & CA) can 

be used to generate waveforms.    Example 1: Pulse waveform with 3sec at 10V and 1sec at 5V.  Set A (Preset 1) = 5V, set B (Preset 2) = 10V  Set Func A = 1 second by turning the Current control knob  Func B = 3 seconds using the same procedure as above  Set t ab = 0 , Set t ba = 0 

6.2.2 Generating Wave Form  To generate the waveform push SHIFT and then press the V/t button The waveform will repeat cyclically until the SHIFT button and then the V/t button is pressed. 

  Example 2: Triangle waveform  Set A = 5V, B = 10 V  Set ab = 3 seconds, Set t ba = 3 seconds  Set Func. A = 3 seconds, Set Func. B = 3 seconds 

  Note: The waveform generator can be operated via the remote programming software (with  waveform preview and graphical datalogging output presentation features). Details are provided  in the section 4 titled PC Software Program. Please verify that the Preset keys are both lit and that one is flashing. The flashing light will jump  from one to the other, A to B, then B to A. The waveform will repeat cyclically until SHIFT and   then V/t are pressed again. Examples 3, 4, and 5 are provided in Appendix A later in this  guide. 

DCP42enGB_V1.5   12/14 12

6.3 Upper Voltage and Current Limits  When the voltage or current at the output terminal exceeds the maximum allowable limits, the  output will automatically switch OFF. This additional protection feature is necessary for a power  supply with such a wide range of voltage and current capabilities.  To set the UVL:   

1. To set the UVL: Press the SHIFT button and then press the Voltage control knob. The  voltmeter will show the current limit setting and the Ammeter will show Suul. 

2. Turn the Voltage control knob to the desired limit value. Press SHIFT to confirm and to  return to normal operation. 

3. To set the UCL, press SHIFT and then press the Current control knob. The Ammeter will  show the current limit setting and the Voltmeter will show SUCL. 

4. Turn the Current control knob to the desired limit value. Press SHIFT to confirm and to  return to normal operation. 

Note: The UVL and UCL settings default to the maximum output value. 

6.4 Remote Sensing  Note: Observe warnings displayed by the Power Supply, an incorrect disconnection sequence  will damage the Power Supply  Warnings: Never short the Remote Sensing Terminal and always disconnect the Remote Sensing  Terminal first. 

6.4.1 Connection  1. First complete the power connections between power supply and equipment.  2. Check and ensure that the power connections are secure.  3. Make connections between Remote Sensing and equipment. 

Warning!: Never short the Remote Sensing Terminal and never connect the Remote Sensing  Terminal in reverse polarity The diagram below shows the connections for Remote Sensing, Power output, and Equipment.  The remote sensing wire should be AT LEAST 22AWG wire size. 

6.4.2 Disconnection  Warning! Incorrect disconnect sequences will damage the power supply  1. First disconnect the remote sensing connections.  2. Then disconnect the power connections between the power supply and equipment. 

DCP42enGB_V1.5   12/14 13

6.5 Remote Control Operation  There are two methods for remotely controlling current and voltage adjustments. Current must  be controlled for both methods otherwise the unit will default to CC mode.   

6.5.1 REMOTE CONTROL  METHOD 1: Using two external DC voltage sources   

Remote Socket Pin Assignment for external variable voltage source 

PIN  FUNCTIONS  REMARKS 

1  Internal DC +5V  Lower than 50mA 

2  Voltage adjust  0 to 5V 

3  Current adjust  0 to 5V 

4  Ground   

5  Output OFF  Short to ground 

6  N.A.   

7  N.A.   

8  N.A.   

  Check the output voltage range of the power supply by varying the external voltage source.  Short circuit the main output with 14AWG wire to check the CC setting on the display while  varying the external voltage source.     

6.5.2 REMOTE CONTROL  METHOD 2: Using two 0 to 5K Ohm variable resistors 

Remote Socket Pin Assignment for external variable resistor 

PIN  FUNCTIONS  REMARKS 

1  Internal DC +5V  One end of resistor 

2  Voltage adjust  Variable leg of resistor 

3  Current adjust  Variable leg of resistor 

4  Ground  Other end of resistor 

5  Output OFF  Short to ground 

6  N.A.   

7  N.A.   

8  N.A.   

  Check the output voltage range of the power supply by adjusting the 5Kohm variable resistor.  Short circuit the main output with 14AWG wire to check the CC setting on the display while  adjusting the variable resistor. 

DCP42enGB_V1.5   12/14 14

6.5.3 Remote Control Output ON/OFF  Remote output ON/OFF control can be activated in any of the modes Normal, Preset, Set, and  Remote. Review the conditions below:   

By default, Pin 5 is open and the output is ON.  Shorting Pin 5 to Pin 4 (ground) switches the output OFF.  When the output is OFF, the C.V. & C.C. LEDs will flash. The present output voltage and  current setting will be shown on the panel meter. 

The user can adjust the output using the voltage & current control knob to the desired  value when the output is OFF. 

  Remark: Use only the 8pin remote plug provided and connect with 22AWG wires. See  diagram below.   

Note: Pin numbers are marked on the black area of the socket

7. PC Software Program  This meter has the capability to connect to and communicate with a PC. To install and use the  software, please refer to the instructions provided on the supplied CDROM and/or the instructions  provided in the HELP Utility within the software program.      

Check the software download page of the website www.extech.com for the latest version of the PC  software and its operating system compatibility. 

8. Automatic Output Step Programs  To use the Automatic Output Step utility the user must first become familiar with the PC  interface and the supplied software. To install and use the software, please refer to the  instructions provided on the supplied CDROM and/or the instructions provided in the HELP  Utility within the software program. Once familiarity with the software is achieved refer to the  instructions below:    The unit can be externally programmed via a PC to run 20 steps, each step with a preset  voltage/current limiting value and a preset time period of 1 second to 99 minutes. The timed  program can be set to run one cycle or up to an infinite number of cycles. 

DCP42enGB_V1.5   12/14 15

8.1 External Timed Program Window 

  Clear Table: Delete all data in the Display Table and prepare for new data entry.  Run (Stop): To run and stop the Timed Program.  Running Cycle: Enter the number of desired running cycles here.

  The numerical range for cycling is 0999. Note that the maximum cycles can be set to infinite  when 0cycles is entered. The External Timed Program allows the user to set the output either  ON or OFF by selecting the boxes in the last column.   

8.2 Output ON/OFF function  Output ON Ticked: Output is ON for that step 

Output ON Unticked: Output is OFF for that step.   

8.3 Automatic Timed Output Operation  1. Clear old data in the table by clicking [Clear Table]. 

2. Enter data in the table using 'UpDownLeftRight' keys on PC keyboard for new locations. 

3. Data that exceed the rated voltage and current will not be accepted. 

4. Voltages that exceed the UVL (Upper Voltage Limit) will not be accepted. 

5. If retrieved or entered, data that exceed the preset Upper or Lower Limit setting for voltage /  current / time periods will switch to red in color and will not be accepted. 

6. When the running time period of any step is set to zero minutes and zero seconds, this step  becomes the terminating step and the cycle will end at this step. In the above example there are  4 steps each with a 2 second time period, if step 3 is set to zero minutes & seconds, the program  will only cycle step 1 and 2 and will not go to step 4. 

7. Enter the number of desired running cycles. 

8. Click [Run] to run the External Time Program. The External Timed Program feature allows the  user to set the output either ON or OFF by selecting the boxes in the last column. 

DCP42enGB_V1.5   12/14 16

9. Appendix  Additional Operational Examples  Examples of Func A/B/C and Wave Form Generator Features    These examples are continued from Examples 1 and 2 in Section 6.2 earlier in this guide.  Example 3  Irregular waveform  Set A= 5V, B= 10 V  Set ab = 1 second, Set t ba = 2 seconds  Set Func. A = 3 seconds, Set Func.B = 3 seconds 

Example 4  Irregular waveform  Set A= 5V, B= 10 V  Set ab = 2 second, Set t ba = 2 seconds  Set Func. A = 2 seconds, Set Func.B = 6 seconds 

Example 5  Irregular waveform  Set A= 5V, B= 10 V  Set ab = 2 second, Set t ba = 0 seconds  Set Func. A = 4 seconds, Set Func.B = 3 seconds 

  Example 6  Irregular waveform  Set A = 5V, B = 10 V  Set ab = 4 second, Set t ba = 4 seconds  Set Func. A = 2 seconds, Set Func .B = 2 seconds 

DCP42enGB_V1.5   12/14 17

10. Specifications   Output 

Variable Output Voltage (Limited to O/P 160W) 0  42VDC  

Variable Output Current (Limited to O/P 160W) 0  10A 

Nominal Output Power (42V / 3.8A) 160W  Preset Outputs 

Preset 1 (P1) Voltage: 5V 0.2V, Current: 10A 0.2A   

Preset 2 (P2) Voltage: 13.8V 0.2V, Current: 10A 0.2A   

Preset 3 (P3) Voltage: 20V 0.2V, Current: 4.0A 0.2A    Voltage Regulation 

Load (0100% Rated Current) 50mV 

Line (90264VAC Variation) 10mV  Current Regulation 

Load (1090% Rated Voltage) 50mA 

Line (90264VAC Variation) 10mA  Ripple & Noise 

Ripple & Noise (RMS) Voltage 8mVrms   

Ripple & Noise (peakpeak) Voltage 80mVpp   

Current Ripple & Noise (peakpeak) 50mA    Meter Type & Accuracy 

Voltage Meter 4 Digit LED Display (0.1% +5 counts) 

Current Meter 4 Digit LED Display (0.1% +5 counts)  Resolution 

Output Voltage Resolution 0.02V 

Output Current Resolution 0.01A  Setting Accuracy 

Output Voltage (0.2% +0.05) 

Output Current (0.2% +0.05)  Output Rise & Fall Time 

Output TON (50% Load Current) 80ms   

Output TON (100% Load Current) 1200ms   

Output TOFF (50% Load Current) 90ms   

Output TOFF (100% Load Current) 50ms   

Power 

Operation Input Voltage Range 90 264VAC 

Operation Input Voltage Frequency Range 4565Hz~  No Load Input Current 

(230Vac / 100Vac): 220mA  Full Load Input Current 

(230Vac / 100Vac): 1.0A / 2.3A  Power Consumption 

(230Vac / 100Vac): 190W / 200W @ 42V / 3.8A    Efficiency 

(230Vac / 100Vac): 86% / 84% @ 42V / 3.8A   

DCP42enGB_V1.5   12/14 18

Switching Frequency  4555 KHz~ 

Tracking Over Voltage Protections 

O/P 010V: Set Voltage+ (1.0 0.3V) 

O/P 1042V: 105125% of set voltage  Power Factor 

Control Power factor correction: >0.91 at optimal load  Cooling Method 

Natural Convection  Leakage Current 

2mA  Protections 

Adjustable upper voltage limit, Adjustable upper current limit, Short Circuit, Overload,  Over Temperature , Tracking OVP 

Input Fuse  T3.15AL250V 

Safety & EMC  Safety Standard CE: EN 61010  Withstanding Voltage I/PO/P: 3.0KVac, I/PF/G: 1.5KVac, O/PF/G: 0.5KVac  Insulation Resistance I/PO/P, I/PF/G, O/PF/G : 100Mohm @500Vdc  EMC CE: EN55011  Harmonics Current Emission EN 6100032  Voltage Fluctuations & Flicker EN 6100033  EMC Immunity EN 6100061 

Environmental  Working Temperature 0 to +40C (32 to +104F)  Working Humidity 1080% RH Noncondensing  Storage Temperature and Humidity 15 to +70C (5 to 158F) 1085% RH Noncondensing  Pollution Degree Class 2 

General  Case Material Polycarbonate Front Panel and ElectroGalvanized Steel Enclosure  Case Protection IP20  Connection Dia. 4.0mm Safety Jack x3 (Output Positive, Negative and Function Ground)  Dimensions (WxHxD) 200 x 90 x 250 mm (7.9 x 3.6 x 10 inch)  Weight 2.5 kg (5.5 lb) 

DCP42enGB_V1.5   12/14 19

11. Troubleshooting  OUP: Over Voltage Protection  This unit has a builtin tracking over voltage protection feature. In the event of output voltage  becoming greater than the set value (see specified range from specifications table), protection  will be triggered and the output power will be cut off and OUP warning appears as below. 

  To reset the warning, switch off the unit and remove all loading.  Switch the unit back on again and it should resume normal operation.  If this problem persists, please contact the customer care department or the point of purchase. 

OTP: Over Temperature Protection  There is a thermo sensor inside the unit to monitor and prevent the unit from unduly heating  internally. When an OTP error occurs, there is no output and the following warning will appear  on the LED display. 

  When this warning appears, switch off the unit and remove all loading.  Check the load and output setting. Allow the unit to cool for at least 30 minutes.  Check to see if any ventilation holes are blocked, also check that there is sufficient clearance  around the power supply.  Listen carefully for the short wind sound emanating from the cooling fan when the unit is  switched ON again. If the wind sound is not detected, the fan may be faulty; do not use the  power supply in this case, contact the customer care department or the agent from the point of  purchase. 

OCP: Over Current Protection  Normally the overload protection is sustained by the CC constant current mode. When the CC  mode fails and goes undetected, serious damage to the test piece or load may occur. The OCP is  used to minimize the extent of damage to the loads in the event of power supply failure.  Switch off the power supply as soon as this warning is seen (as shown below). 

  To reset this warning, remove all loading and press the OUTPUT ON/OFF button.  Switch the unit back on again and double check with caution.  If this problem persists, please contact the customer care department or consult with the agent  at the point of purchase. 

Copyright  2014 FLIR Systems, Inc. 

All rights reserved including the right of reproduction in whole or in part in any form 

w

Manualsnet FAQs

If you want to find out how the Extech Flir works, you can view and download the Flir Extech DCP42 Power Supply User Guide on the Manualsnet website.

Yes, we have the User Guide for Flir Extech as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Guide should include all the details that are needed to use a Flir Extech. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir Extech DCP42 Power Supply User Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir Extech DCP42 Power Supply User Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir Extech DCP42 Power Supply User Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir Extech DCP42 Power Supply User Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir Extech DCP42 Power Supply User Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.