Contents

Flir DM62 Multimeter User Manual PDF

1 of 32
1 of 32

Summary of Content for Flir DM62 Multimeter User Manual PDF

USER MANUAL

True RMS Digital MultiMeter Series Models DM62, DM64, and DM66

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 2

Table of Contents

1. ADVISORIES 4

1.1 Copyright 4

1.2 Quality Assurance 4

1.3 Documentation 4

1.4 Disposal of Electronic Waste 4 2. SAFETY 5 3. INTRODUCTION 7 4. METER AND DISPLAY DESCRIPTIONS 8

4.1 Front and Back Meter Descriptions 8

4.2 Rotary Function Selector 9

4.3 Control Buttons 9

4.4 Display Description 10 5. BASIC OPERATION AND FEATURES 11

5.1 Powering the Meter 11

5.2 Intelligent Auto Power OFF (APO) 11

5.3 LCD Backlight 11

5.4 Auto/Manual Range Mode 11

5.5 True RMS Response 11

5.6 Crest Factor 11

5.7 Out-of-range Warning (O.L) 11

5.8 Data Hold 12

5.9 Relative Zero Mode 12

5.10 VFD Mode (Low Pass Filter) Models DM64/DM66 only 12

5.11 MAX-MIN-AVG Record Mode 12

5.12 Test Lead Input Warning 12

5.13 Auto Voltage AC/DC Detection in Lo Z mode (DM64/DM66 only) 12 6. MEASUREMENTS 13

6.1 AC/DC Voltage Measurements (VFD and Frequency on DM64/DM66) 13

6.2 Low Impedance (Lo Z) Auto-Voltage Measurements (DM64/66 only) 14

6.3 Line Frequency Measurements (DM64/DM66 only) 15

6.4 Resistance and Continuity Measurements 16

6.5 Capacitance (DM64/DM66) and Diode Measurements 17

6.6 AC/DC mV and Temperature (DM64/DM66) Measurements 19

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 3

6.7 A, mA, A Current Measurements with Frequency (DM64/DM66) 21

6.8 NCV (Non-Contact Voltage) Detection 22 7. MAINTENANCE 23

7.1 Cleaning and Storage 23

7.2 Battery and Fuse Replacement 23

7.3 Disposal of Electronic Waste 23 8. SPECIFICATIONS 24

8.1 General specifications 24

8.2 Electrical Specifications 25 9. TECHNICAL SUPPORT 29 10. THREE-YEAR LIMITED WARRANTY 30

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 4

1. Advisories

1.1 Copyright 2018, FLIR Systems, Inc. All rights reserved worldwide. No parts of the software including source code may be reproduced, transmitted, transcribed or translated into any language or computer language in any form or by any means, electronic, magnetic, optical, manual or otherwise, without the prior written permission of FLIR Systems. The documentation must not, in whole or part, be copied, photocopied, reproduced, translated or transmitted to any electronic medium or machine readable form without prior consent, in writing, from FLIR Systems. Names and marks appearing on the products herein are either registered trademarks or trademarks of FLIR Systems and/or its subsidiaries. All other trademarks, trade names or company names referenced herein are used for identification only and are the property of their respective owners.

1.2 Quality Assurance The Quality Management System under which these products are developed and manufactured has been certified in accordance with the ISO 9001 standard. FLIR Systems is committed to a policy of continuous development; therefore, we reserve the right to make changes and improvements on any of the products without prior notice.

1.3 Documentation To access the latest manuals and notifications, go to the Download tab at: http://support.flir.com. It only takes a few minutes to register online. In the download area you will also find the latest releases of manuals for our other products, as well as manuals for our historical and obsolete products.

1.4 Disposal of Electronic Waste As with most electronic products, this equipment must be disposed of in an environmentally friendly way, and in accordance with existing regulations for electronic waste. Please contact your FLIR Systems representative for more details.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 5

2. Safety

Safety Notes This manual contains information and warnings that must be followed for operating the meter safely and maintaining the meter in a safe operating condition. If the meter is used in a manner not specified by the manufacturer, the protection provided by the meter may be impaired.

Terms in this manual WARNING Identifies conditions and actions that could result in serious injury or even

death to the user. CAUTION Identifies conditions and actions that could cause damage or malfunction in

the instrument.

WARNING To reduce the risk of fire or electric shock, do not expose this product to rain or moisture. The meter is intended only for indoor use. Keep your hands/fingers behind the hand/finger barriers (of the meter and the test probe assembly, where applicable) that indicate the limits of safe access of the hand-held parts during measurements. Inspect lead wires, connectors, and probes for damaged insulation or exposed metal before using the meter. If any defects are found, replace them immediately. Only use the probe assembly provided with the meter or a UL Listed Probe Assembly to the same meter ratings or better. IEC 61010-031 requires exposed conductive test probe tips to be 4mm for CAT III & CAT IV ratings. Refer to the category markings on your probe assemblies as well as on the add-on accessories (like detachable Caps or Alligator Clips), if any, for applicable rating changes. Observe proper safety precautions when working with voltages above 30 Vrms, 42.4 Vpeak or 60 VDC. These voltage levels pose a potential shock hazard to the user. Before and after hazardous voltage measurements, check the voltage function on a known source such as line voltage to determine proper meter functioning.

CAUTION

Disconnect the test leads from the test points before changing functions.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 6

International Electrical Symbols

Marking of Electrical and Electronic Equipment (EEE). Do not dispose of this product as unsorted municipal waste. Contact a qualified recycler

Caution! Refer to the explanation in this Manual

Caution! Possibility of electric shock

Earth (Ground)

Meter protected throughout by Double Insulation or Reinforced insulation

Fuse

Direct Current (DC)

Alternating Current (AC)

3 Three-phase Alternating Current

Measurement Categories

Measurement Category IV is applicable to test and measuring circuits connected at the source of the buildings low-voltage MAINS installation. Examples are measurements on devices installed before the main fuse or circuit breaker in the building installation. Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the buildings low-voltage MAINS installation. Examples are measurements on distribution boards (including secondary meters), circuit-breakers, wiring, including cables, bus-bars, junction boxes, switches, and socket-outlets in the fixed installation, and equipment for industrial use and some other equipment such as stationary motors with permanent connection to the fixed installation.

CENELEC DIRECTIVES The instruments conform to CENELEC Low-voltage directive 2014/35/EC, Electromagnetic compatibility directive 2014/30/EU and RoHS directive 2011/65/EU.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 7

3. Introduction Thank you for selecting the FLIR DM6x Digital Multimeter. This meter can measure up to 600V and 10A AC/DC. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.

Features

6000 Count (3-5/6 digits) backlit LCD with 5 reading per second update rate

Auto Power OFF

Measures ACV, DCV, Resistance, Continuity, Capacitance*, Diode, DCA, ACA, Temperature*, line Frequency*, non-contact Voltage (NCV)

True RMS responding

Auto AC/DC detection for voltage measurements*

Lo Z (low impedance) for ghost voltage protection*

Auto ranging

VFD feature (low pass filter for AC voltage measurements)*

MIN-MAX-AVG Reading Memory

Relative Zero mode

Equipped with test leads, batteries, printed Quick Start (full User Manual available on FLIR website http://support.flir.com), and type K thermocouple*

Optional accessories: Banana plug to type-K socket plug adaptor*, magnetic hanger, and carrying pouch

*Applies to models DM64 and DM66 only

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 8

4. Meter and Display Descriptions

4.1 Front and Back Meter Descriptions

1. Non-Contact Voltage Detector

2. LCD

3. Control buttons

4. Rotary Selector Switch

5. A Current input terminal

6. uA, mA Current input terminal

7. COM (negative -) input terminal

8. Positive (+) input terminal for all inputs except Current

9. Battery compartment lock and fuse access

10. Test Lead holders

11. Battery compartment and fuse access

12. Tilt stand

Fig. 4-1 Front and Back Views (DM66)

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 9

4.2 Rotary Function Selector

Lo Z (low impedance) mode with Auto ACV/DCV detection (DM64/DM6)

Meter power off

AC/DC Voltage and VFD (low pass filter) modes (VFD available in models DM64/DM66 only)

Resistance and Continuity measurements

Temperature (DM64 and DM66 only) and AC/DC milli-volt measurements

Capacitance (DM64 and DM66 only) and Diode measurements

AC/DC Amp and milli-amp current measurements

AC/DC micro-amp current measurements

4.3 Control Buttons

Long press to switch LCD Backlight ON/OFF

Mode select button. For rotary switch positions with more than one function, use this button to select the desired function

Short press to select MANUAL mode and to change ranges, long press to return to AUTO. Also used to select sensitivity in NCV mode and trigger level for line frequency measurements (DM64/DM66)

Short press to access MIN-MAX-AVG Record mode. Short press to step through MIN-MAX-AVG readings. Long press to exit

Long press the Relative/Zero button to switch the mode ON/OFF. When switched on, the displayed reading is stored as reference/offset value

Data Hold button. Short press to freeze/unfreeze displayed reading

Frequency button (DM64/DM66). Short press to view frequency for AC measurements

Short press the Non-Contact Voltage (NCV) button to switch ON/OFF. Use RANGE to select high (H) 24~1000V or low (L) 100~1000V sensitivity

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 10

4.4 Display Description

Fig. 4-2 LCD Display Icons

1. Automatic Range mode

2. Data Hold

3. Maximum, Minimum, and Average Reading Record modes

4. High Sensitivity NCV mode

5. Low Sensitivity NCV mode

6. VFD (low pass filtering) mode*

7. Continuity test

8. Relative/Zero function

9. Units of measure

10. Diode test

11. Main display digits

12. Battery status

13. Lo Z (low impedance) mode*

14. Minus symbol

15. AC and DC symbols

16. Automatic ACV/DCV detection in Lo Z mode*

*Models DM64/DM66 only

5. Basic Operation and Features

5.1 Powering the Meter 1. Set the function switch to any position to switch on the meter. 2. If the battery status indicator appears low or if the meter does not power on, replace the two (2) AAA batteries. See Section 7.2, Battery Replacement.

5.2 Intelligent Auto Power OFF (APO) The meter will intelligently avoid entering the APO mode under normal measurement conditions. The APO feature turns the meter off automatically, in order to extend battery life, after approximately 32 minutes if none of the conditions below are met:

1) Rotary switch or push button operation

2) Significant measurement value (reading above 8.5% of the range)

3) Non-OL readings for Resistance, Continuity or Diode function

4) Non-zero readings for Hz function

5) Electric field signal present for EF NCV function

To wake the meter from APO, short press the MODE button, or turn the rotary switch OFF and then back on. Always turn the rotary switch to the OFF position when the meter is not in use.

5.3 LCD Backlight Long press the backlight button to toggle the LCD backlight ON and OFF. The backlight will automatically switch OFF after 10 minutes to conserve battery energy.

5.4 Auto/Manual Range Mode For most auto-ranging functions, short press the RANGE button to select manual- ranging override. The meter will remain in the range it was in and the auto range display icon will switch off. Short press the button again to select the next range. Long press the button to resume auto-ranging. Note: Manual-ranging is not available for Auto-V, Capacitance & Hz functions.

5.5 True RMS Response True RMS (Root Mean Square) allows the meter to respond accurately to the effective RMS value, regardless of the waveform shape (square, saw-tooth, triangle, pulse, spike, as well as distorted waveforms with the presence of harmonics).

5.6 Crest Factor Crest Factor is the ratio of the Crest (instantaneous peak) value to the True RMS value. A pure sine wave has a Crest Factor of 1.414. A badly distorted sinusoidal waveform normally has a much higher Crest Factor.

5.7 Out-of-range Warning (O.L) If the input is over/under the full-scale range in Manual range mode, or if the signal has exceeded the maximum/minimum input in Auto range mode, O.L is displayed.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 12

5.8 Data Hold In Data Hold mode, the displayed reading is held. To enter/exit Data Hold mode, short

press the button. In Hold mode, the indicator is displayed.

5.9 Relative Zero Mode Relative Zero allows the user to offset measurements by the amount determined by a stored reference value. Almost any displayed reading can be set as the relative reference value including MAX/MIN/AVG readings. Long press the button to toggle Relative Zero mode ON and OFF. When switched ON, the meter stores the displayed reading as the offset/reference value, the zero (0) symbol appears on the upper right, and the meter will display the measurement value minus the stored reference.

5.10 VFD Mode (Low Pass Filter) Models DM64/DM66 only The VFD (variable-frequency drive) utility eliminates high-frequency noise from AC voltage measurements. This is accomplished using a low-pass filter. To activate, short press the MODE button until the VFD symbol appears.

5.11 MAX-MIN-AVG Record Mode Short press REC to activate the recording mode, three arrow icons ( [MAX], [MIN], and [AVG]) will appear indicating that the meter is now recording. Short press the REC button to step through the MAX-MIN-AVG readings. The meter beeps when new MAX or MIN readings are detected. Long press REC to exit the recording mode. Auto-ranging is active and Auto-Power-Off is disabled in this mode.

5.12 Test Lead Input Warning The meter beeps (at a fast rate) as well as displays InEr to warn the user that an

improper connection to the A, mA, or A input jack is being made when another function, especially a voltage function, is selected.

5.13 Auto Voltage AC/DC Detection in Lo Z mode (DM64/DM66 only) Auto Voltage Lo Z mode automatically selects DCV or ACV, based on input levels detected through the test leads. This mode uses a Lo Z low ramp-up impedance (approx. 2.1k for low voltages) to drain ghost voltages. Refer to Section 6.2 Low Impedance Auto-Voltage Measurements for complete details.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 13

6. Measurements

Caution: Before and after hazardous voltage measurements, test the voltage function on a known source such as line voltage to determine proper meter functioning.

Caution: Before operating the device, you must read, understand, and follow all instructions, dangers, warnings, cautions, and notes.

Caution: When the meter is not in use, the function switch should be set to the OFF position.

Caution: When connecting the probe leads to the device under test, connect the COM (negative) lead before connecting the positive lead. When removing the probe leads, remove the positive lead before removing the COM (negative) lead.

6.1 AC/DC Voltage Measurements (VFD and Frequency on DM64/DM66)

Move the rotary switch to the position and short press the MODE button to step through the following modes: DCV, ACV, and ACV with VFD*. The last selection will be saved as the power-up default. For AC, short press the Hz button to view the frequency* and press again to return to the voltage measurement (refer to Section 6-3 for more on Frequency measurement features). Refer to Fig. 6-1 for function switch position, test lead connection, push-button operation, and display example. *VFD and Frequency available on models DM64/DM66 only

Figure 6-1 ACV, DCV, and VFD application diagram

Notes: VFD-ACV and the associated Hz measurement are equipped with a digital low- pass filter, and are capable of handling VFD (Variable Frequency Drives) signals for fundamental V & Hz readings. It also improves ACV and Hz reading stability when being used in noisy electrical environments.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 14

6.2 Low Impedance (Lo Z) Auto-Voltage Measurements (DM64/66 only) Auto Voltage Lo Z mode automatically selects DCV or ACV, based on input levels detected through the test leads. This mode uses a Lo Z low ramp-up impedance (approx. 2.1k for low voltages) to drain ghost voltages*. Refer to Fig. 6-2 for function switch position, test lead connection, and display example.

Figure 6-2 LoZ application diagram

With no input, the meter displays - - - - when it is ready for testing.

When a signal above the voltage threshold of 8.0V DC or 5.5V AC (up to the rated 600V) is present, the meter displays the voltage value in DC or AC, whichever is larger in peak magnitude. Attempting to measure Lo Z voltages below these limits will result in the same readings as for the normal voltage measurement mode.

Notes: *Ghost-voltages are unwanted stray signals, coupled from adjacent live signals, which can confuse common Multimeter voltage measurements. This feature is helpful when distinguishing between hot and open wires (to ground) in electrical installations, for example. Only HOLD, EF (NCV) & Backlight push-button features are available in Auto Voltage mode.

WARNING: Lo Z Auto Voltage mode input impedance increases abruptly from the initial 2.1k to a few hundred ks on high voltage hard signals. Lo Z appears on the LCD as a reminder. Peak initial load current, while probing 1000VAC for example, can be as high as 673mA (1000V x 1.414 / 2.1k), decreasing abruptly to approx. 2.4mA (1000V x 1.414 / 580k) within a fraction of a second.Do not use this feature on circuits that could be damaged by such low input impedance. Instead, use the high input impedance voltage modes to minimize loading on such circuits (rotary

switch set to the position).

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 15

6.3 Line Frequency Measurements (DM64/DM66 only) Short press the Hz button to toggle the frequency function ON and OFF. Refer to Fig. 6-3 for function switch position, test lead connection, push-button operation, and display example. A voltage application is shown in Fig. 6-3, but line frequency can also be viewed in AC Current applications; however, the input sensitivity adjustment, as detailed below, applies to AC Voltage applications only.

Figure 6-3 Line Frequency Measurements (Hz) for ACV/VFD

Note on Input Sensitivity for voltage applications:

Input sensitivity varies automatically with the range selected. The 6V range has the highest

sensitivity and the 1000V range has the lowest. In the DCV, ACV or VFD-ACV voltage functions, the

trigger voltage will be displayed immediately before starting the Hz measurement. While displayed,

short press the RANGE button to manually select another trigger voltage range (voltage modes

only). First measure the signal voltage (or current) level and then activate the Hz function in that

range to find the optimum trigger level. If the Hz reading becomes unstable, select a lower

sensitivity to avoid electrical noise. If the reading shows zero, select a higher sensitivity.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 16

6.4 Resistance and Continuity Measurements

Warning: Do not perform resistance or continuity tests before removing power from capacitors and other devices under test during a measurement. Injury to persons can occur.

Short press the MODE button to toggle Resistance and Continuity functions. The last selection will be saved as the power-up default. Refer to the Fig. 6-4 for function switch position, test lead connection, push-button operation, and display examples.

Figure 6-4 Resistance and Continuity Measurements

For convenience, the Continuity function offers a continuous beep tone together with a flashing LCD backlight indicating a complete wire. This audio-visual indication is especially handy in noisy work environments. If the resistance is < 30 the meter beeps. If the resistance is > 480 the meter will not beep. If the resistance is > 30 but < 480 the beeping will stop at an unspecified point. CAUTION Be sure to remove power from any circuit or device before measuring resistance and continuity. Measuring resistance and continuity on a live circuit will produce false results and may damage the instrument.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 17

6.5 Capacitance (DM64/DM66) and Diode Measurements

Warning: Do not perform capacitance or diode tests before removing power to the devices under test. Injury to persons can occur.

Short press the MODE button to toggle the Capacitance and Diode functions. The last selection will be saved as the power-up default. Refer to the Fig. 6-5 for function switch position, test lead connection, push-button operation, and display example for Capacitance testing. Refer to Fig 6-6 for diode testing.

Figure 6-5 Capacitance Testing

CAUTION Discharge capacitors before making any measurements. Large value capacitors should be discharged through an appropriate resistance load. Large value capacitors will require ample time before a stable reading is displayed.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 18

In the Diode mode, the normal forward voltage drop (forward biased) for a good silicon diode is between 0.400V to 0.900V (Fig 6-6, left). A reading higher than that indicates a leaky diode (defective). A zero reading indicates a shorted diode (defective). An OL display indicates an open diode (defective). Reverse the test lead connection (reverse biased) across the diode (Fig 6-6, right). The digital display shows OL if the diode is good. Any other readings indicate that the diode is resistive or shorted (defective).

Figure 6-6 Diode testing (forward bias on left; reverse bias on right)

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 19

6.6 AC/DC mV and Temperature (DM64/DM66) Measurements Short press the MODE button to step through mV AC/DC and Temperature oC/oF functions (temperature available only on DM64/DM66). The last selection will be saved as the power-up default. Refer to the Fig. 6-7 for function switch position, test lead connection, push-button operation, and display example for mV measurements. Short press the Hz button to switch the Frequency measurement mode on/off in AC mV mode. Refer to Fig 6-8 for Temperature measurements.

Figure 6-7 AC/DC mV measurements

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 20

Insert the banana plug type-K temperature bead probe observing correct polarity. You can also use a plug adapter (optional) with banana plug to type-K socket to adapt other standard type-K mini plug temperature probes (see the drawing on the right of Fig 6-8, for adaptor illustration).

Figure 6-8 Temperature measurements (optional adaptor shown on right)

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 21

6.7 A, mA, A Current Measurements with Frequency (DM64/DM66)

WARNING Do not measure current on a circuit when the voltage increases to more than

600V. This can cause damage to the instrument and can cause injury to persons.

Short press the MODE button to toggle DC and AC modes. The last selection will be saved as the power-up default. Refer to the Fig. 6-9 for function switch position, test lead connection, push-button operation, and display examples. Short press the Hz button to view the frequency measurement in AC current modes. Frequency is available only on the DM64/DM66 models.

Figure 6-9 Current measurements (A on left, mA in the center, and A, on right)

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 22

6.8 NCV (Non-Contact Voltage) Detection

Short press the button to switch ON/OFF the non-contact voltage (NCV) detector.

The meter displays H or L (for High 24~1000V or Low 100~1000V sensitivity) when it is ready. Short press the RANGE button to change the sensitivity. Place the top left of the meter near a conductor to test. The detected voltage strength is indicated as a series of display dashes and a variable beeper tone.Refer to the Fig. 6-10 for function switch position, push-button operation, and display example.

Figure 6-10 Non-Contact Voltage Detection

An antenna is located along the top-left end of the meter which detects the electric fields that surround energized, live conductors. It is ideal for tracing live wiring connections, locating wiring breakages and distinguishing between live and earth connections.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 23

7. Maintenance

7.1 Cleaning and Storage Wipe the housing with a damp cloth as needed. Please do not use abrasives or solvents to clean the meter. If the meter is not to be used for an extended period, remove the batteries and store them separately.

7.2 Battery and Fuse Replacement WARNING: To avoid electrical shock, disconnect the meter from any connected circuits, remove the test leads from the meter terminals, and set the function switch to the OFF position before attempting to replace the batteries or fuses.

1. To replace the batteries, unlock the battery compartment

by rotating the lock with a screwdriver or other tool (the lock is item 1 in the accompanying diagram).

2. Remove the battery compartment cover. 3. Replace the two (2) 1.5V AAA batteries, observing correct

polarity. 4. Replace the battery cover and re-lock it before operating

the meter. 5. To replace the fuses, lift the tilt stand and locate the two

screw holes (item 2 in the diagram). 6. The screws are protected by rubber grommets which must

be removed before the screws can be accessed. 7. Remove the two screws and remove one additional screw

located just behind the battery compartment lock to access the fuse area.

8. Replace Fuse F1 for A/mA current input with a 0.4A/600V DC/AC, IR 30kA F fuse or better; Dimension: 6 x 32 mm

9. Replace Fuse F2 for A current input with a 11A/600V DC/AC, IR 20kA F fuse or better; Dimension: 10 x 38mm

10. Secure the meter before operating.

7.3 Disposal of Electronic Waste As with most electronic products, this equipment must be disposed of in an environmentally friendly way, and in accordance with existing regulations for electronic waste. Please contact your FLIR Systems representative for more details.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 24

8. Specifications

8.1 General specifications

Display: 3-5/6 digits 6,000 counts

Update Rate: 5 per second nominal

Operating Temperature: 14oF ~ 122oF (-10oC ~ 50oC)

Relative Humidity: Maximum relative humidity 80% for temperature up to 87.8oF (31oC) decreasing linearly to 50% relative humidity at 122oF (50oC)

Altitude: Operating below 6562 (2000m)

Storage Temperature: -4oF ~ 140oF (-20oC ~ 60oC), < 80% R.H. (with battery removed)

Temperature Coefficient: Nominal 0.15 x (specified accuracy)/ oC @ (14oF ~ 64.6oF [- 10oC ~ 18oC]] or 82.4oF ~ 122oF [28oC ~ 50oC]), or as otherwise specified

Sensing: True RMS response

Ingress Protection: IP40

Drop-proof: to 6.5 (2m)

Pollution Degree: 2

Safety: Certified per IEC/UL/EN61010-1 Ed. 3.0, IEC/UL/EN61010-2-030 Ed. 1.0, IEC/UL/EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and the corresponding CAN/CSA-C22.2 regulations to Measurement Categories:

CAT III 600V and CAT IV 300V AC & DC

Transient Protection: 6.0kV (1.2/50s surge)

E.M.C.: Meets EN61326-1:2013

In an RF field of 3V/m:

Temperature function is not specified

Ohm function: Total Accuracy = Specified Accuracy + 15 digits

Other functions: Total Accuracy = Specified Accuracy

Performance above 3V/m is not specified

Overload Protection:

A & mA: 0.4A/600V DC/AC rms, IR 30kA, F fuse or better

A: 11A/1000V DC/AC rms, IR 20kA, F fuse or better

V & Auto Voltage modes: 600V DC/AC rms

mV, Ohm & others: 600V DC/AC rms

Low Battery warning: Below approx. 2.5V

Power Supply: 1.5V AAA Size battery x 2

Power Consumption (typical): 3.2mA

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 25

APO Consumption (typical): 10A

APO Timing: Powers off after 30 minutes of inactivity

Dimension: 6.3 x 3.1 x 2.0 (161 x 80 x 50mm) L x W x H

Weight: Approx. 11.8 oz. (334 g)

Special Features: Auto Voltage (Lo Z) (DM64/DM66); VFD; Audio-visual Continuity; Auto-ranging MAX/MIN/AVG Record; Backlit LCD; Auto-ranging Relative Zero mode;

Display Hold; EF-Detection (NCV); Test lead warning on A-mA/A terminals

Accessories: Test lead pair; 2 x 1.5V AAA batteries installed; Quick Start manual; Banana plug type-K thermocouple (DM64/DM66 only)

Optional Purchase Accessories: Banana plug to type-K socket plug adaptor (DM64/DM66 only); Magnetic hanger; Soft carrying pouch

8.2 Electrical Specifications Accuracy is (% reading + number of digits (dgt)) at 73.4 F 9 F (23C 5 C), <80% RH.

ACV & ACA accuracies are specified from 1 % to 100 % of range or as otherwise specified. Maximum Crest Factor <2:1 at full scale & <4:1 at half scale, and with frequency components fall within the meter specified frequency bandwidth for non- sinusoidal waveforms

AC Voltage

RANGE Accuracy

50Hz ~ 60Hz

6.000V 1), 60.00V, 600.0V 0.7% + 3d

45Hz ~ 440Hz

6.000V 1), 60.00V, 600.0V 2.0% + 3d

Input Impedance: 10M, 54pF nominal 1) <5d non-zero residue may appear when backlight is on, which will not affect the

specified measuring range and accuracy ACmV

RANGE Accuracy

10Hz ~ 500Hz

60.00mV 1) 2), 600.0mV 3) 1.0% + 3d

500Hz ~ 800Hz

60.00mV 1) 2), 600.0mV 3) 2.0% + 3d

Input Impedance: 10M, 54pF nominal 1) <5d non-zero residue may appear when backlight is on, which will not affect the

specified measuring range and accuracy 2) Signal peak absolute values, including DC bias, less than 130mVpeak 3) Signal peak absolute values, including DC bias, less than 1300mVpeak

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 26

VFD_ACV (with Low Pass Filter) DM64/DM66 only

RANGE Accuracy 1)

10Hz ~ 100Hz (fundamental)

600.0V 1.0% + 3d

100Hz ~ 400Hz (fundamental)

600.0V 10% + 3d 2) 1) Not specified for fundamental frequency > 400Hz 2) Accuracy linearly decreases from 1% + 3d @100Hz to 10% + 3d @400Hz

Auto Voltage ACV (DM64/DM66)

RANGE Accuracy 1)

45Hz ~ 440Hz

600.0V 2.0% + 3d 1) Not specified at <1VAC Threshold: > 1VAC nominal Approximate input impedance (//164pF) for reference:

At direct input 50Vac (typical) from quiescence:

>8M @ < 5.6Vac

22k @ 7Vac

12k @ 8Vac

2.6k @ 50Vac At direct input >>50V (typical) from quiescence:

Initial impedance is approximately 2.1k. Impedance increases abruptly within a fraction of a second as display voltage (hard signal) is much higher than 50V (typical). Final impedances vs display voltages typically are:

12k @100V

100k @300V

240k @600V

DC Voltage

RANGE Accuracy

60.00mV, 600.0mV, 6.000V 0.3% + 2d

60.00V 0.4% + 2d

600.0V 0.2% + 2d

Input Impedance: 10M, 54pF nominal

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 27

Auto Voltage DCV (DM64/DM66)

RANGE Accuracy 1)

600.0V 2.0% + 3d 1) Not specified at <1VDC Threshold: > +1.0VDC or < -1.0VDC nominal Approximate input impedance (//164pF) for reference:

At direct input 50Vdc (typical) from quiescence:

>8M @ < 8Vdc (Protection clamping threshold)

25k @ 9Vdc

13k @ 10Vdc

2.6k @ 50Vdc At direct input >>50V (typical) from quiescence:

Initial impedance is approximately 2.1k. Impedance increases abruptly within a fraction of a second as display voltage (hard signal) is much higher than 50V (typical). Actual impedances vs display voltages typically are:

12k @100V

100k @300V

240k @600V

Resistance (Ohms)

RANGE 1) Accuracy

600.0, 6.000k 0.3% + 3d

60.00k, 600.0k 0.5% + 3d

6.000M 2), 60.00M 3) 0.9% + 2d 4) 1) Open Circuit Voltage: 1.6VDC typical 2) Constant Test Current: 0.2A Typical 3) Constant Test Current: 0.02A Typical 4) 5%+20d @ >30M Continuity Tester

Continuity Threshold: Between 30 and 480 Continuity ON Response Time: <15ms Audible Indication: Beeper sound Visible Indication: LCD Backlight

Capacitance (DM64/DM66)

RANGE Accuracy

20.00nF, 200.0nF 1.5% + 8d

2000nF, 20.00F, 200.0F, 2000F 1.5% + 2d

10.00mF 4.5% + 10d

Accuracies with film capacitor or better

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 28

Diode Tester

RANGE Accuracy

3.000V 0.9% + 2d

Test Current: 0.3mA typical

Open Circuit Voltage: < 3.2VDC typical DC Current

RANGE Accuracy Burden Voltage

600.0A, 6000A 1.0% + 3d 0.1mV/A

60.00mA, 600.0mA 0.7% + 3d

1.9mV/mA

6.000A, 10.00A 1) 0.04V/A 1)10A continuous, >10A to 20A for 30 seconds max with 5 minutes cool down interval AC Current

RANGE Accuracy Burden Voltage

50HZ ~ 400HZ

600.0A, 6000A 1.5% + 3d 0.1mV/A

60.00mA, 600.0mA 1.0% + 3d

1.9mV/mA

6.000A 1), 10.00A 2) 0.04V/A 1) <5d non-zero residue may appear when backlight is on, which will not affect the

specified measuring range and accuracy 2)10A continuous, >10A to 20A for 30 seconds max with 5 minutes cool down interval Temperature (DM64/DM66)

RANGE Accuracy 1) 2)

-40.0 oC ~ 99.9oC 1% + 1 oC

100 oC ~ 400oC

-40.0 oF ~ 99.9 oF 1% + 2 oF

100 oF ~ 752 oF 1) Accuracies assume meter interior and the ambient air have reached the same temperature for a correct junction voltage compensation. Allow enough settling time for a significant change of ambient temperature. It can take up to an hour for changes > 5oC (>9oF). 2) Type-K thermocouple range & accuracy not included

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 29

Line Frequency (DM64/DM66)

Function Sensitivity (Sine RMS) Range

60mV, 600mV 50mV

10.00Hz ~ 50.00kHz 6V 5V

60V 10V

600V 50V 10.00Hz ~ 1.00kHz

VFD 600V 50V

600A, 6000A 500A 10.00Hz ~ 5.00kHz

60mA, 600mA 50mA

6A, 10A 8A 50.00Hz ~ 1.00kHz

Accuracy: 0.03% + 2d NCV Non-Contact Voltage Detection Indication: Display dashes and audible beep tones proportional to a conductors field strength. Sensitivity: H display for High sensitivity (24~1000V) and L display for Low sensitivity (100~1000V), use the RANGE button to change the sensitivity. Detection Frequency: 50/60Hz Detection Antenna: Top-left of the meter

9. Technical Support

Main Website http://www.flir.com/test

Technical Support Website http://support.flir.com

Technical support Email TMSupport@flir.com

Service/Repair Support Email Repair@flir.com

Support Telephone number +1 855-499-3662 option 3 (toll-free)

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 30

10. Three-Year Limited Warranty Congratulations! You (the Purchaser) are now the owner of a world-class FLIR Test and Measurement product. A qualifying FLIR Test and Measurement product (the Product) purchased either directly from FLIR Commercial Systems Inc. and affiliates (FLIR) or from an authorized FLIR distributor that Purchaser registers on-line with FLIR is eligible for coverage under FLIRs 3-Year Limited Warranty subject to the terms and conditions in this document. This warranty only applies to purchases of Qualifying Products (see below) purchased after April 2013 and only to the original Purchaser of the Product. PLEASE READ THIS DOCUMENT CAREFULLY; IT CONTAINS IMPORTANT INFORMATION ABOUT THE PRODUCTS THAT QUALIFY FOR COVERAGE UNDER THE 3-YEAR LIMITED WARRANTY, PURCHASERS OBLIGATIONS, HOW TO ACTIVATE THE WARRANTY, WARRANTY COVERAGE, AND OTHER IMPORTANT TERMS, CONDITIONS, EXCLUSIONS AND DISCLAIMERS. 1. PRODUCT REGISTRATION. To qualify for FLIRs 3-Year Limited Warranty, the Purchaser must fully register the Product directly with FLIR on-line at www.flir.com WITHIN Sixty (60) DAYS of the date the Product was purchased by the first retail customer (the Purchase Date). PRODUCTS THAT ARE NOT REGISTERED ON-LINE WITHIN Sixty (60) DAYS OF THE PURCHASE DATE OR PRODUCTS WHICH DO NOT QUALIFY FOR THE 3-YEAR WARRANTY WILL HAVE A LIMITED ONE YEAR WARRANTY FROM THE DATE OF PURCHASE. 2. QUALIFYING PRODUCTS. Upon registration, a list of Test and Measurement Products that qualify for coverage under FLIRs 3-Year Limited Warranty can be found at www.flir.com/testwarranty 3. THE WARRANTY PERIOD. The Product is warranted for a period of three (3) Years from the Purchase Date (the Warranty Period). Any Product that is repaired or replaced under warranty is covered under this 3-Year Limited Warranty for one hundred eighty days (180) days from the date of return shipment by FLIR or for the remaining duration of the applicable Warranty Period, whichever is longer. 4. LIMITED WARRANTY. In accordance with the terms and conditions of this 3-Year Limited Warranty, and except as excluded or disclaimed in this document, FLIR warrants, from the Purchase Date, that all fully registered Products will conform to FLIRs published Product specifications and be free from defects in materials and workmanship during the applicable Warranty Period. PURCHASERS SOLE AND EXCLUSIVE REMEDY UNDER THIS WARRANTY, AT FLIRS SOLE DISCRETION, IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PRODUCTS IN A MANNER, AND BY A SERVICE CENTER, AUTHORIZED BY FLIR. IF THIS REMEDY IS ADJUDICATED TO BE INSUFFICIENT, FLIR SHALL REFUND PURCHASERS PAID PURCHASE PRICE AND HAVE NO OTHER OBLIGATION OR LIABILITY TO BUYER WHATSOEVER. 5. WARRANTY EXCLUSIONS AND DISCLAIMERS. FLIR MAKES NO OTHER WARRANTY OF ANY KIND WITH RESPECT TO THE PRODUCTS. ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE (EVEN IF PURCHASER HAS NOTIFIED FLIR OF ITS INTENDED USE FOR THE PRODUCTS), AND NON-INFRINGEMENT ARE EXPRESSLY EXCLUDED FROM THIS AGREEMENT. THIS WARRANTY EXPRESSLY EXCLUDES ROUTINE PRODUCT MAINTENANCE, AND SOFTWARE UPDATES, AND REPLACEMENT OF FUSES OR DISPOSABLE BATTERIES. FLIR FURTHER EXPRESSLY

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 31

DISCLAIMS ANY WARRANTY COVERAGE WHERE THE ALLEGED NONCONFORMITY IS DUE TO NORMAL WEAR AND TEAR, ALTERATION, MODIFICATION, REPAIR, ATTEMPTED REPAIR, IMPROPER USE, IMPROPER MAINTENANCE, NEGLECT, ABUSE, IMPROPER STORAGE, FAILURE TO FOLLOW ANY PRODUCT INSTRUCTIONS, DAMAGE (WHETHER CAUSED BY ACCIDENT OR OTHERWISE), OR ANY OTHER IMPROPER CARE OR HANDING OF THE PRODUCTS CAUSED BY ANYONE OTHER THAN FLIR OR FLIRS EXPRESSLY AUTHORIZED DESIGNEE. THIS DOCUMENT CONTAINS THE ENTIRE WARRANTY AGREEMENT BETWEEN PURCHASER AND FLIR AND SUPERSEDES ALL PRIOR WARRANTY NEGOTIATIONS, AGREEMENTS, PROMISES AND UNDERSTANDINGS BETWEEN PURCHASER AND FLIR. THIS WARRANTY MAY NOT BE ALTERED WITHOUT THE EXPRESS WRITTEN CONSENT OF FLIR. 6. WARRANTY RETURN, REPAIR AND REPLACEMENT. To be eligible for warranty repair or replacement, Purchaser must notify FLIR within thirty (30) days of discovering of any apparent defect in materials or workmanship. Before Purchaser may return a Product for warranty service or repair, Purchaser must first obtain a returned material authorization (RMA) number from FLIR. To obtain the RMA number Owner must provide an original proof of purchase. For additional information, to notify FLIR of an apparent defect in materials or workmanship, or to request an RMA number, visit www.flir.com. Purchaser is solely responsible for complying with all RMA instructions provided by FLIR including but not limited to adequately packaging the Product for shipment to FLIR and for all packaging and shipping costs. FLIR will pay for returning to Purchaser any Product that FLIR repairs or replaces under warranty. FLIR reserves the right to determine, in its sole discretion, whether a returned Product is covered under warranty. If FLIR determines that any returned Product is not covered under warranty or is otherwise excluded from warranty coverage, FLIR may charge Purchaser a reasonable handling fee and return the Product to Purchaser, at Purchasers expense, or offer Purchaser the option of handling the Product as a non-warranty return. FLIR shall not be responsible for any data, images or other information that may be stored on the returned Product which was not included in the Product at the time of purchase. It is Purchasers responsibility to save any and all data prior to returning the Product for warranty service. 7. NON-WARRANTY RETURN. Purchase may request that FLIR evaluate and service or repair a Product not covered under warranty, which FLIR may agree to do in its sole discretion. Before Purchaser returns a Product for non-warranty evaluation and repair, Purchaser must contact FLIR by visiting www.flir.com to request an evaluation and obtain an RMA. Purchaser is solely responsible for complying with all RMA instructions provided by FLIR including but not limited to adequately packaging the Product for shipment to FLIR and for all packaging and shipping costs. Upon receipt of an authorized non-warranty return, FLIR will evaluate the Product and contact Purchaser regarding the feasibility of and the costs and fees associated with Purchasers request. Purchaser shall be responsible for the reasonable cost of FLIRs evaluation, for the cost of any repairs or services authorized by Purchaser, and for the cost of repackaging and returning the Product to Purchaser. Any non-warranty repair of a Product is warranted for one hundred eighty days (180) days from the date of return shipment by FLIR to be free from defects in materials and workmanship only, subject to all of the limitations, exclusions and disclaimers in this document.

FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AB 32

Corporate Headquarters FLIR Systems, Inc. 2770 SW Parkway Avenue Wilsonville, OR 97070 USA Telephone: +1 503-498-3547 Customer Support Technical Support Website http://support.fli

Manualsnet FAQs

If you want to find out how the DM62 Flir works, you can view and download the Flir DM62 Multimeter User Manual on the Manualsnet website.

Yes, we have the User Manual for Flir DM62 as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Manual should include all the details that are needed to use a Flir DM62. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir DM62 Multimeter User Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir DM62 Multimeter User Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir DM62 Multimeter User Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir DM62 Multimeter User Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir DM62 Multimeter User Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.