USER MANUAL

FLIR MODEL DM166 Thermal Imaging Digital MultiMeter with IGMTM

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 2

Table of Contents

1. ADVISORIES 4

1.1 Copyright 4

1.2 Quality Assurance 4

1.3 Documentation 4

1.4 Disposal of Electronic Waste 4 2. SAFETY 5 3. INTRODUCTION 7

3.1 Key Features 7 4. METER DESCRIPTION AND REFERENCE GUIDE 8

4.1 Front and Back Meter Descriptions 8

4.2 Function Switch Positions 9

4.3 Function Buttons 10 4.3.1 MODE/Hz Button Sequence of Operations 10

4.4 Display Icons and Status Indicators 11 5. METER POWER 12

5.1 Powering the Meter 12

5.2 Intelligent Auto Power OFF (APO) 12 6. MULTIMETER OPERATION 13

6.1 Auto/Manual Range Modes 13

6.2 Probe Connection Alert 13

6.3 Out of Range Warning (OL) 14

6.4 Data (Display) Hold 14

6.5 MIN-MAX-AVG Mode 14

6.6 Relative Mode 14

6.7 Test Lead Holder Accessory 14

6.8 Voltage (AC/DC), VFD, and Frequency (Hz) Measurements 15

6.9 Non-Contact Voltage Detector 16

6.10 Resistance Measurements 17

6.11 Continuity Test 18

6.12 Diode Test 18

6.13 Capacitance Measurements 19

6.14 Type K Temperature Measurements 20

6.15 Current and Frequency Measurements (A, mA, A) 20 6.15.1 Test Lead Current Measurements (A, mA, and A) 20

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 3

6.15.2 FLEX Clamp Adaptor Current and Frequency Measurements 22 7. IGMTM THERMAL IMAGER OPERATION 23

7.1 IGMTM (Infrared Guided Measurements) Basics 23 7.1.1 Thermal Image Display Description (refer to Fig. 7-1 for the list below) 23 7.1.2 Thermal Imager Operation 23

7.2 Using the Multimeter in the IGMTM mode 25

7.3 Emissivity Factors for Common Materials 25

7.4 Infrared Energy and Thermal Imaging Overview 26 8. MAINTENANCE 27

8.1 Cleaning and Storage 27

8.2 Battery Replacement 27

8.3 Fuse Replacement 27

8.4 Disposal of Electronic Waste 27 9. SPECIFICATIONS 28

9.1 General specifications 28

9.2 Thermal Imaging Specifications 30

9.3 Electrical Specifications 30 10. TECHNICAL SUPPORT 34 11. WARRANTIES 35

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 4

1. Advisories

1.1 Copyright 2017, FLIR Systems, Inc. All rights reserved worldwide. No parts of the software including source code may be reproduced, transmitted, transcribed or translated into any language or computer language in any form or by any means, electronic, magnetic, optical, manual or otherwise, without the prior written permission of FLIR Systems. The documentation must not, in whole or part, be copied, photocopied, reproduced, translated or transmitted to any electronic medium or machine readable form without prior consent, in writing, from FLIR Systems. Names and marks appearing on the products herein are either registered trademarks or trademarks of FLIR Systems and/or its subsidiaries. All other trademarks, trade names or company names referenced herein are used for identification only and are the property of their respective owners.

1.2 Quality Assurance The Quality Management System under which these products are developed and manufactured has been certified in accordance with the ISO 9001 standard. FLIR Systems is committed to a policy of continuous development; therefore, we reserve the right to make changes and improvements on any of the products without prior notice.

1.3 Documentation To access the latest manuals and notifications, go to the Download tab at: http://support.flir.com. It only takes a few minutes to register online. In the download area you will also find the latest releases of manuals for our other products, as well as manuals for our historical and obsolete products.

1.4 Disposal of Electronic Waste As with most electronic products, this equipment must be disposed of in an environmentally friendly way, and in accordance with existing regulations for electronic waste. Please contact your FLIR Systems representative for more details.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 5

2. Safety

Safety Notes Before operating the device, you must read, understand, and follow all instructions,

dangers, warnings, cautions, and notes.

FLIR Systems reserves the right to discontinue models, parts or accessories, and other items, or to change specifications at any time without prior notice.

Remove the batteries if the device is not to be used for an extended period of time.

Warning Statements

Do not operate the device if you do not have the correct knowledge. Incorrect operation of the device can cause damage, shock, injury or death to persons.

Do not start a measuring procedure before you have set the function switch to the correct position. Failure to do so can cause damage to the instrument and can cause injury to persons.

Do not change to the resistance mode when measuring voltage. This can cause damage to the instrument and can cause injury to persons.

Do not measure the current on a circuit when the voltage increases to more than 1000 V. This can cause damage to the instrument and can cause injury to persons.

You must disconnect the test leads from the circuit under test before you change the range. Failure to observe this warning can damage the instrument and cause bodily injury.

Do not replace the batteries before you remove the test leads. This can cause damage to the instrument and can cause injury to persons.

Do not use the device if the test leads and/or the device show signs of damage. Injury to persons can occur.

Be careful performing measurements if the voltages are > 25 VAC rms or 35 VDC. There is a risk of shock from these voltages. Injury to persons can occur.

Do not do diode, resistance or continuity tests before you have removed the power from capacitors and other devices under test. Injury to persons can occur.

Be careful when performing voltage checks on electrical outlets. These checks are difficult because of the uncertainty of the connection to the recessed electrical contacts. You must not rely solely on this this device when determining if the terminals are not live. There is a risk of electrical shock. Injury to person can occur.

Do not touch expired/damaged batteries without gloves. Injury to persons can occur.

Do not cause a short circuit of the batteries. This can cause damage to the instrument and can cause injury to persons.

Do not put the batteries into a fire. Injury to persons can occur.

Use extreme caution when the laser pointer is on.

Do not point the beam toward anyone's eye or allow the beam to strike the eye from a reflective surface.

Do not use the laser near explosive gases or in other potentially explosive areas.

Refer to the CAUTION statement label (shown below) for critical safety information.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 6

Note: The manufacturers address label is located inside the battery compartment.

Cautions Do not use the device in a manner not specified by the manufacturer. This can cause damage to the protection provided.

This symbol, adjacent to another symbol or terminal, indicates that the user must refer to the user manual for further information.

This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Double insulation.

UL listing is not an indication or a verification of the accuracy of the meter

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 7

3. Introduction Thank you for selecting the FLIR DM166 True RMS Digital MultiMeter with IGMTM (Infrared Guided Measurement) Thermal Imaging. The DM166 can measure voltage up to 600V AC/DC and includes a VFD function (low pass filter). The thermal imager measures surface temperature, offers adjustable emissivity, and features a laser pointer and cross hairs for precise targeting. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.

3.1 Key Features

6000 count 2.4 digital TFT display, 320x240 pixels.

IGMTM Infrared imager with selectable color palette, laser pointer, cross hairs targeting, and adjustable emissivity measures surface temperature

Thermal imager (80x60 pixels) offers a 30:1 distance-to-spot ratio and a 50o x 38.6o field-of- view

DMM measures AC/DC Voltage (V, mV), AC/DC Current (A, mA, A), Frequency, Resistance, Continuity, Diode, Capacitance, and Type-K Temperature

Built-in non-contact voltage detector (NCV)

Flex Clamp adaptor direct input

Automatic and Manual ranging

Input over-voltage warning (OL)

Input warning when test leads are incorrectly connected to meter

MIN-MAX-AVG memory

Relative offset mode

Variable-frequency drive VFD feature (low-pass filter)

Data (Display) Hold

Intelligent Auto Power OFF (APO), customizable

Safety Category Rating: CAT IV-300V, CAT III-600V.

Equipped with batteries, test leads, pouch, and Quick Start booklet.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 8

4. Meter Description and Reference Guide

4.1 Front and Back Meter Descriptions

Fig. 4-1 Front View

1. NCV sensor

2. LCD display

3. MODE/Hz button

4. RANGE button

5. Palette/Emissivity button

6. Function Switch

7. Positive (+) Probe Input Jack for A/mA (Current)

8. Positive (+) Probe Input Jack for Amps (Current)

9. COM (-) Probe Input Jack

10. Positive (+) Probe Input Jack for all inputs except Amps, mA, and A

11. Relative (REL) button

12. Laser pointer ON/OFF button

13. MIN-MAX/APO button

14. HOLD/Temperature units button

15. Thermal Image mode button

1. Laser lens

2. Thermal imaging lens

3. Test Lead holder mounts

4. Tripod mount

5. Tilt stand

6. Battery/Fuse compartment

7. Battery/Fuse door lock

Fig. 4-2 Rear View

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 9

4.2 Function Switch Positions

Detect AC voltage through the non-contact sensor at the top of the meter

FLEX Direct: Auxiliary channel for use with optional Flexible Current clamp or standard clamp adaptors when > 600A measurements are required. In this mode, the meter will display true rms ACA measurements from the connected device. Long press MODE to see the Frequency (Hz).

Meter is switched OFF and in full power-saving mode.

Measure AC voltage (V) through the probe inputs. Use the MODE button to select the VFD (low pass filter) function

Measure DC voltage (V) through the probe inputs.

Measure low voltage (mV) through the probe inputs. Use the MODE button to select AC/DC milli-volts. Measure temperature through the probe inputs using a thermocouple adaptor. Use the MODE button to select temperature.

Measure resistance, continuity (use MODE button to toggle function).

Measure capacitance, or diode through the probe inputs. Use the MODE button to select the desired function.

Measure A current through the probe inputs. Use the MODE button to select AC or DC.

Measure current through the probe inputs (A or mA). Use the MODE button to select AC or DC.

Fig. 4-3 Rotary Function Switch

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 10

4.3 Function Buttons

Short press to toggle VFD on/off in AC Voltage mode or to toggle functions at a two-function switch position. Long press to view Frequency (Hz) in AC modes. Short press to step through AC/DC mV, and temperature when the function switch is set to the mV/Temp position. See Section 4.3.1, MODE Button Sequence of Operations.

From Auto range mode, short press to select Manual range mode. In Manual mode, short press to change range; long press to return to Auto range.

In thermal image mode, short press to select a color palette. Long press to open the emissivity adjustment menu, then short press to select emissivity value and long press to exit.

Short press to enable/disable Relative mode.

Short press to activate the Thermal Imager. Short press again to clear image of all text/icons. Short press a third time to exit the Thermal Imager mode.

Press and hold to activate laser pointer. Release button to deactivate.

Short press to switch ON/OFF the MIN-MAX-AVG display. Long press to open the Auto Power OFF (APO) settings menu, then short press to select APO time and long press to exit.

Short press freezes/unfreezes display (DMM or thermal image mode). Long press toggles oC and oF temperature units in Thermal Image mode only.

4.3.1 MODE/Hz Button Sequence of Operations

Measurement MODE/Hz button sequence of operations

ACA < > Frequency (long button press)

ACV < > Frequency (long press)

AC mV > DC mV > C > F (short press) AC mV < > Frequency (long press)

Resistance < > Continuity (short press)

Capacitance < > Diode (short press)

ACA < > DCA (short press)

AC < > DC (short press); ACA/ACmA < > Frequency (long press)

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 11

4.4 Display Icons and Status Indicators

For the non-contact voltage detector, the display bars and beeper represent the strength of the sensed voltage. Number of bars and beeper rate increase relative to voltage strength.

Lo Low sensitivity mode for non-contact voltage detector (80~1000V range). Use the RANGE button to toggle Hi/Lo setting.

Hi High sensitivity mode for non-contact voltage detector (20~1000V range). Use the RANGE button to toggle Hi/Lo setting.

MAX reading displayed (available in DMM and Thermal Image modes)

MIN reading displayed (available in DMM and Thermal Image modes)

AVG reading displayed (available in DMM and Thermal Image modes)

Auto range mode

VFD (low pass filter)

Data (display) HOLD for DMM and Thermal Image mode

Relative mode (available in DMM and Thermal Image mode)

Emissivity setting

Battery status

AC current or voltage

DC current or voltage

Flex Clamp or standard clamp adaptor direct input

Continuity mode

Resistance mode

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 12

Diode test mode

Capacitance mode

External Type K Thermocouple probe mode

oC/oF Temperature units

5. Meter Power

5.1 Powering the Meter 1. Set the function switch to any position to switch on the meter.

2. If the battery indicator shows that the battery voltage is low, if a Low Battery warning screen appears, or if the meter does not power on, replace the batteries in the rear compartment. See Section 8.2, Battery Replacement. If using the Model TA04 charging system, please recharge the battery pack.

5.2 Intelligent Auto Power OFF (APO) The meter enters sleep mode after a programmable period of inactivity, to customize this setting:

1. Long press the APO button to open the APO menu 2. Short press the APO button to step to the desired APO time or to OFF 3. Long press the APO button to exit 4. The last APO selection saved becomes the default APO time 5. The DM166 has an intelligent APO feature where the meter will not power

down if any of the following conditions exist:

Rotary switch or push button is operated

Significant measurement > 8.5% of ranges

Non-OL readings for Resistance, Continuity or Diode function

Non-zero readings for Hz function

Electric field signal present for NCV function

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 13

6. MultiMeter Operation

Caution: Before operating the device, you must read, understand, and follow all instructions, safety warnings, cautions, and notes.

Caution: When the meter is not in use, the function switch should be set to the OFF position.

Caution: When connecting the probe leads to the device under test, connect the COM (negative) lead before connecting the positive lead. When removing the probe leads, remove the positive lead before removing the COM (negative) lead.

6.1 Auto/Manual Range Modes In Auto range mode , the meter automatically selects the most appropriate measurement scale. In Manual range mode, the user can adjust the range (scale). Auto range mode is the default mode of operation.

1. To use Manual range, short press the RANGE button to enter the manual range mode, and short press again until the desired range appears.

2. To return to the Auto range mode, long press the RANGE button until the Auto

Range indicator appears.

6.2 Probe Connection Alert When the probe leads are not plugged into the correct jacks for the measurement selected by the function switch, a display alert is shown and the beeper sounds. The warning shown in Figure 6-1, below left, appears when a test lead is plugged into the A or A/mA input terminals with the rotary switch set to any function other than A, mA, or A. The warning shown in Figure 6-1, below right, appears when the test lead is plugged into the A input terminal with the rotary switch set to the A terminal. In these cases, switch the meter off and correctly connect the test leads before attempting to make measurements.

Fig. 6-1 Probe Connection Alert Screens

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 14

6.3 Out of Range Warning (OL) If the input is outside the selected range in Manual mode, or if the signal has exceeded the maximum input in Auto range mode, OL appears.

6.4 Data (Display) Hold In Hold mode, the displayed reading freezes (DMM mode) or the thermal image freezes

(Thermal Image mode). To enter/exit Hold mode, short press the button. The

indicator appears in the Hold mode.

6.5 MIN-MAX-AVG Mode Short press the MIN/MAX button to start recording/viewing the lowest , highest ,

and average readings. Each time a memory display is updated the meter beeps. Short

press MIN/MAX to return to the normal operating display.

6.6 Relative Mode In Relative mode, you can compare measurements to a stored reference. For example,

if the stored reference value is 10VDC and you take a measurement of 50VDC, the

meter will display 40VDC (50V actual minus the 10V reference.).

Take a measurement and short press the REL button to store the reading (stored

reading appears on the lower right corner of the display).

Now, subsequent measurements will display relative to the stored reference.

Short press REL to return to the normal operating mode.

6.7 Test Lead Holder Accessory Connect the test lead holder accessory (optional) to the two slots on the back of the meter (item 3 in Fig. 4-2) and to the tripod mount (item 4 in Fig. 4-2). The test lead holder is ergonomic and protects the thermal imaging and laser pointer lenses.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 15

6.8 Voltage (AC/DC), VFD, and Frequency (Hz) Measurements

Fig. 6-2(a) AC Voltage/HZ Measurements 1. Set the function switch to one of the following positions:

(VDC) or (VAC) for high voltage measurements. Short press MODE to switch to VFD mode for VAC measurements. Press again to exit.

(milli-volts) for low voltage measurements. Short press MODE to select AC or DC for milli-volt measurements.

2. The indicator appears for AC measurements. The indicator appears for DC measurements.

3. Insert the black probe lead into the negative COM terminal and the red probe lead into the positive terminal.

4. Connect the probe leads in parallel to the part under test.

5. Read the measured voltage value on the display. The meter defaults to Auto Range mode. To use Manual Range mode: short press the Range button or refer to Section 6.1, Auto/Manual Range Modes.

6. If OL appears, the signal being measured is out of the range of the meters capability. In this case, stop the test immediately and review the application.

7. Short press the HOLD button to freeze/unfreeze the displayed reading.

8. To view the Frequency (Hz) of the measured AC voltage, long press the Hz button. Long press again to return to the voltage measurement.

9. For additional functionality, refer to Section 6.5, MIN- MAX-AVG Mode and Section 6.6, Relative Mode.

Fig. 6-2(b) DC Voltage Measurements

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 16

6.9 Non-Contact Voltage Detector Warning: For safety, always test the voltage detector on a known live circuit before using the voltage detector on unknown voltage potentials.

1. Set switch to the NCV position. See Figure 6-3.

2. Be sure to remove the test leads from the meter. 3. Short press the RANGE button to toggle the Hi (20~1000V) and Lo (80~1000V)

sensitivity modes. 4. Position the top of the meter close to a source of voltage. 5. When the meter detects a voltage, it beeps and displays a blue bar animation. The

tone rate and the bar number will increase relative to the detected voltage.

Fig. 6-3 Non-Contact Voltage Detector

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 17

6.10 Resistance Measurements Warning: Do not test resistance/continuity before removing power from capacitors and other devices under test during a measurement. Injury to persons can occur.

1. Refer to Fig. 6-4. Set the function switch to the resistance position. 2. Short press MODE to step to the resistance mode. 3. Insert the black probe lead into the negative COM terminal and red probe lead

into positive terminal. 4. Touch the tips of the probe across the circuit or component under test. 5. Read the resistance value on the display. The meter defaults to Auto Range

mode. To use Manual mode short press the RANGE button or refer to Section 6.1, Auto/Manual Range Modes.

6. If OL appears, the signal is out of range. In this case, stop the test immediately and review the application.

7. Short press the HOLD button to freeze/unfreeze the displayed reading. 8. For additional functions, see Section 6.5, MIN-MAX-AVG Mode and Section 6.6,

Relative Mode.

Fig. 6-4 Resistance and Continuity Measurements

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 18

6.11 Continuity Test Warning: Do not perform continuity tests before removing the power from capacitors and other devices under test during a measurement. Injury to persons can occur.

1. Refer to Fig. 6-4. Set the function switch to the continuity position.

2. Short press the MODE button to select continuity if necessary . 3. Insert the black probe lead into the negative COM terminal and the red probe

lead into the positive terminal. 4. Touch the tips of the probe across the circuit or component under test.

If the resistance is < 30 the meter beeps

If the resistance is > 480 the meter will not beep

If the resistance is > 30 but < 480 the beep stops at an unspecified point

6.12 Diode Test Warning: Do not perform diode tests before removing the power to the diode or other devices under test during a measurement. Injury to persons can occur.

1. Set the function switch to the diode position. Short press the MODE button to select the diode test function if necessary.

2. Insert the black probe lead into the negative COM terminal and the red probe lead into the positive terminal.

3. Touch probe tips across the diode or semiconductor junction under test in one polarity (direction) and then in the opposite polarity as shown in Fig. 6-5.

4. If the reading is between 0.400 and 0.800V in one direction and OL (overload) in the opposite direction, the component is good. If the measurement is 0V in both directions (shorted) or OL in both directions (open), the component is bad.

Fig. 6-5 Diode Test

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 19

6.13 Capacitance Measurements Warning: Do not perform capacitance tests before removing power to the capacitor or other devices under test during a measurement. Injury to persons can occur.

1. Set the function switch to the capacitance position. 2. Short press the MODE button to select the capacitance measurement if

necessary. The F (Farad) unit of measure appears. 3. Insert the black probe lead into the negative COM terminal and the red probe

lead into the positive terminal. 4. Touch the tips of the probe across the part under test. 5. Read the capacitance value on the display. Short press the HOLD button to

freeze/unfreeze the displayed reading. 6. For additional functionality, refer to Section 6.5, MIN-MAX-AVG Mode and

Section 6.6, Relative Mode.

Fig. 6-6 Capacitance Measurements

Note: For very large capacitance values, it may take several minutes for the measurement to settle and the final reading to stabilize.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 20

6.14 Type K Temperature Measurements 1. Set the function to the Temperature position. 2. Short presses of the MODE button step to the F or C temperature mode. 3. While observing polarity, insert the

thermocouple adapter into the negative COM terminal and the positive terminal.

4. Touch the tip of the thermocouple to a part under test or hold the thermocouple in the surrounding air. Wait until the reading stabilizes before moving the thermocouple and recording the reading.

5. Read the temperature value on the display. Short press the HOLD button to freeze/unfreeze the displayed reading.

6. To avoid electrical shock, disconnect the thermocouple adapter before turning the function switch to another position.

Fig. 6-7 Temperature Measurements

6.15 Current and Frequency Measurements (A, mA, A) For test lead current measurements, disconnect the part under test and connect the test leads in series with the part, see Figure 6-8.

Fig. 6-8 Disconnected component

6.15.1 Test Lead Current Measurements (A, mA, and A) 1. For test-lead measurements: A (amperes) and mA (milli-amperes) set the

function switch to the position. For A (microampere) measurements, set

the function switch to the position. 2. Insert the black probe lead into the negative COM terminal and the red probe

lead into one of the following positive terminals:

A for high current measurements.

mA for lower current measurements.

A for micro-amp measurements

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 21

If the test lead warning appears on the display and the meter beeps, turn meter power OFF and check the test lead connections to the meter. Do not attempt to use the meter when the alert is showing.

3. Short press the MODE button to select AC or DC measurements.

The indicator appears for AC measurements.

The indicator appears for DC measurements. 4. Connect the probe leads in series with the part or circuit under test. Fig. 6-9

shows a typical setup example. 5. Read the current on the display. If OL appears, the signal being measured is out

of the range of the meters capability. In this case, stop the test immediately and review the application.

6. Short press the HOLD button to freeze/unfreeze the displayed reading. 7. To see the frequency (Hz) of measured AC current, long press the Hz button;

long press again to exit. 8. For additional functionality, refer to Section 6.5, MIN-MAX-AVG Mode and

Section 6.6, Relative Mode.

Fig. 6-9 Current Measurement example

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 22

6.15.2 FLEX Clamp Adaptor Current and Frequency Measurements Connect a FLIR Flex Clamp Adaptor (Model TA72 or TA74, for example) or other clamp adaptor to the DM166 to display current measurements as detailed below:

1. Turn the function dial to the position.

2. Connect a Clamp adaptor as shown in Fig. 6-10.

3. Set the Range of the Flex Clamp Adaptor to match the range of the DM166.

4. Short press the RANGE button to select the range of the DM166 (1, 10, or 100 mv/A, as shown in the upper left corner of the DM166 display).

5. Operate the Flex Clamp meter (or other clamp adaptor) per the instructions provided with the Flex Clamp meter or other adaptor.

6. Read the current measured by the Flex Clamp on the DM166 LCD. If OL appears, the signal being measured is out of the range of the meters capability. In this case, stop the test immediately and review the application.

7. Short press the HOLD button to freeze/unfreeze the displayed reading.

8. To see the frequency value for the measured current, long press the Hz button. Long press the button to return to the current measurement display.

Fig. 6-10 FLEX Clamp Application

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 23

7. IGMTM Thermal Imager Operation

7.1 IGMTM (Infrared Guided Measurements) Basics

In the Thermal Imaging mode, the user can measure a targeted surfaces temperature. The DM166 accomplishes this by detecting the energy emitted by the surface under test. The DM166 sees a thermal image of an area under test in the same manner as with a dedicated thermal imaging device. See Section 7.4, Infrared Energy and Thermal Imaging Overview for in-depth information. The DM166 laser pointer and cross hairs assist in targeting.

Press the IGM button to open the Thermal Imager. In Fig 7-1, the meter is set to color

palette IRON. Short press the Palette/Emissivity button to select other colors.

Long press the Emissivity button to open the emissivity-setting menu. Press the Emissivity button to scroll to the desired setting, press and hold the emissivity button to exit. See Section 7.3, Emissivity Factors for Common Materials

7.1.1 Thermal Image Display Description (refer to Fig. 7-1 for the list below)

1. Surface temperature measurement represents the temperature of the spot. Dashes display while the temperature reading stabilizes

2. DMM measurement data

3. Cross hairs for spot targeting

4. Thermal image frame

5. Icon area

Fig. 7-1 IGM Display Example

7.1.2 Thermal Imager Operation

1. Set the function switch to any position.

2. Short press the IGM button to switch the Thermal Imager ON. Short press again to clear image of all text and icons. Short press again to exit the imager mode.

3. With the Imager on, point the lens (back of meter) at the target.

4. The display will show the temperature measurement in the upper left hand corner for the targeted area.

5. Use the laser pointer and cross hairs for targeting. Press and hold the laser button to use the pointer; release the button to switch OFF the laser.

6. In the Thermal imaging mode, the meter continues to operate normally as a Multimeter; the electrical symbols appear on the left side of the display.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 24

7. The Distance to Spot ratio for the imager is 30:1 meaning that the measurement spot is 30 times smaller than the distance the meter is from the spot (at a distance of 30, the meter sees a target spot of 1). See Fig. 7-2.

8. The thermal imagers FOV (Field of View) is 50 degrees (top view) and 38.6 degrees (side view); see Fig. 7-3 (a) and (b).

Fig. 7-2 Distance-to-Spot ratio 30:1

Fig. 7-3 (a) Field of View side view Fig. 7-3 (b) Field of View - top view

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 25

7.2 Using the Multimeter in the IGMTM mode Use the Multimeter as described in Section 6, Multimeter Operation while the thermal imager is in use. Multimeter readings, icons, and operational modes Relative and MIN-MAX-AVG appear superimposed on the thermal image.

7.3 Emissivity Factors for Common Materials

Material Emissivity Material Emissivity

Asphalt 0.90 to 0.98 Cloth (black) 0.98

Concrete 0.94 Skin (human) 0.98

Cement 0.96 Leather 0.75 to 0.80

Sand 0.90 Charcoal (powder) 0.96

Soil 0.92 to 0.96 Lacquer 0.80 to 0.95

Water 0.92 to 0.96 Lacquer (matt) 0.97

Ice 0.96 to 0.98 Rubber (black) 0.94

Snow 0.83 Plastic 0.85 to 0.95

Glass 0.90 to 0.95 Timber 0.90

Ceramic 0.90 to 0.94 Paper 0.70 to 0.94

Marble 0.94 Chromium Oxides 0.81

Plaster 0.80 to 0.90 Copper Oxides 0.78

Mortar 0.89 to 0.91 Iron Oxides 0.78 to 0.82

Brick 0.93 to 0.96 Textiles 0.90

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 26

7.4 Infrared Energy and Thermal Imaging Overview A thermal imager generates an image based on temperature differences. In a thermal image, the hottest item in the scene appears as white and the coldest item as black. All other items are represented as a gray scale value between white and black. The DM166 also offers color images to simulate hot (lighter colors) and cold (darker colors) temperatures.

It may take some time to get used to the thermal imagery. Having a basic understanding of the differences between thermal and daylight cameras can help with getting the best performance from the DM166.

One difference between thermal and daylight cameras has to do with where the energy comes from to create an image. When viewing an image with an ordinary camera, there has to be some source of visible light (something hot, such as the sun or other lighting) that reflects off the objects in the scene to the camera. The same is true with human eyesight; the vast majority of what people see is based on reflected light energy. On the other hand, the thermal imager detects energy that is directly radiated from objects in the scene.

This is why hot objects such as parts on engines and exhaust pipes appear white, while the sky, puddles of water and other cold objects appear dark (or cool). Scenes with familiar objects will be easy to interpret with some experience.

Infrared energy is part of a complete range of radiation called the electromagnetic spectrum. The electromagnetic spectrum includes gamma rays, X-rays, ultraviolet, visible, infrared, microwaves (RADAR), and radio waves. The only difference is their wavelength or frequency. All of these forms of radiation travel at the speed of light. Infrared radiation lies between the visible and RADAR portions of the electromagnetic spectrum.

The primary source of infrared radiation is heat or thermal radiation. Any object that has a temperature radiates in the infrared portion of the electromagnetic spectrum. Even objects that are very cold, such as an ice cube, emit infrared. When an object is not quite hot enough to radiate visible light, it will emit most of its energy in the infrared. For example, hot charcoal may not give off light, but it does emit infrared radiation, which we feel as heat. The warmer the object, the more infrared radiation it emits.

Infrared imaging devices produce an image of invisible infrared or heat radiation that is unseen by the human eye. There are no colors or shades of gray in infrared, only varying intensities of radiated energy. The infrared imager converts this energy into an image that we can interpret.

The FLIR Infrared Training center offers training (including online training) and certification in all aspects of thermography: http://www.infraredtraining.com/.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 27

8. Maintenance

8.1 Cleaning and Storage Wipe the housing with a damp cloth as needed. Use a high quality lens wipe to remove dirt or smudges from the meter lenses and display window. Please do not use abrasives or solvents to clean the meter housing, lenses, or display window. If the meter is to be stored for an extended period, remove the batteries and store them separately.

8.2 Battery Replacement The Battery symbol flashes with no bars when the battery voltage has reached a critical level. The meter displays readings within specifications while the low battery indicator is on and powers off before it displays an out of tolerance reading.

WARNING: To avoid electrical shock, disconnect the meter from any connected circuits, remove the test leads from the meter terminals, and set the function switch to the OFF position before attempting to replace the batteries.

1. Remove the battery cover (attached to tilt stand) by first unlocking it. To do so, use a flat-head screwdriver to move the lock/unlock screw to the unlocked position.

2. Remove the battery cover and replace the three-1.5V AA batteries observing correct polarity.

3. If using the Model TA04 rechargeable lithium-polymer battery system, please recharge the rechargeable battery.

4. Secure the battery compartment cover before use.

8.3 Fuse Replacement Access the two fuses by first unlocking the battery/fuse compartment cover (attached to the tilt stand). To do so, use a flat-head screwdriver to move the locking/unlocking screw to the unlocked position. Then remove the two small Phillips-heads screws to open the fuse compartment. The fuse ratings are:

FS1: 11A, 1KV FAST

FS2: 400mA, 1KV FAST

8.4 Disposal of Electronic Waste As with most electronic products, this equipment must be disposed of in an environmentally friendly way, and in accordance with existing regulations for electronic waste. Please contact your FLIR Systems representative for more details.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 28

9. Specifications

9.1 General specifications

Display: 3-5/6 digits 6,000 counts

Update Rate: Five (5) per second nominal

Operating Temperature: 14 ~ 122oF (-10oC ~ 50oC)

Relative Humidity: Maximum relative humidity 80% for temperature up to 87.8oF (31oC) decreasing linearly to 50% relative humidity at 122oF (50 oC)

Altitude: Operating below 6560 ft (2000m)

Storage Temperature: -4F ~ 140F (-20oC ~ 60oC), < 80% R.H. (with battery removed)

Temperature Coefficient: Nominal 0.15 x (specified accuracy)/ oC @ 14oF ~ 64.4oF (-10oC ~ 18oC) or 82.4oF ~ 122oF (28oC ~ 50 oC), or as otherwise specified

Sensing: True RMS sensing

Pollution Degree: 2

Safety: Certified per IEC/UL/EN61010-1 Ed. 3.0, IEC/UL/EN61010-2-030 Ed. 1.0, IEC/UL/EN61010-2- 033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and the corresponding CAN/CSA-C22.2 regulations to Measurement Categories:

CAT III 600V and CAT IV 300V AC & DC

Transient Protection: 6.0kV (1.2/50s surge)

E.M.C.: Meets EN61326-1:2013; In an RF field of 3V/m:

Temperature function not specified

Ohm function: Total Accuracy = Specified Accuracy + 15 digits

Other functions: Total Accuracy = Specified Accuracy

Performance above 3V/m not specified

Overload Protection:

A & mA: 0.4A/1000V DC/AC rms, IR 30kA, F fuse or better

A: 11A/1000V DC/AC rms, IR 20kA, F fuse or better

V & Auto V: 1100V DC/AC rms

mV, Ohm, & others: 1000V DC/AC rms

Drop protection: 9.8(3m)

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 29

Power Supply: 3 x 1.5 V AA alkaline or Li/FeS2 lithium batteries or optional Model TA04 lithium polymer rechargeable battery system

Power Requirements for thermal imager:

Battery Type: Alkaline AA Battery x 3

Battery Life: approximately 12 hours

Battery Type: Energizer L91 Lithium (Li/FeS2) AA Battery x 3

Battery Life: approximately 22 hours

Optional Rechargeable Battery: Li-Polymer; FLIR PN: TA04-KIT

Battery Life: approximately 22 hours

Power Consumption (typical): 160mA

APO Consumption (typical): 200A

APO Timing: Five (5) minutes (default), 10 minutes, 20 minutes & OFF selectable

APO counter reset for DMM mode:

Rotary switch or push button operations

Significant measuring readings of above 8.5% of ranges

Non-OL readings for Resistance, Continuity or Diode function

Non-zero readings for Hz function

Electric field signal present for EF function

APO counter reset for Thermal Imager mode:

Rotary switch or push button operations

Accessories: Test lead pair, Batteries, Quick Start manual, Pouch

Calibration Cycle: One year

Weight: 15.1 oz (428.3g)

Dimensions: (L x W x H) 7.5 x 3.4 x 1.9 in. (190 x 86.4 x 48.3mm)

CAT Application Field

I Circuits not connected to mains.

II Circuits directly connected to a low-voltage installation.

III Building installation.

IV Source of the low-voltage installation.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 30

9.2 Thermal Imaging Specifications

IR Temperature Range 14 ~ 302F (-10 ~150C)

IR Temperature Resolution 0.1C/F

Image Sensitivity < or equal to 150mK (0.15C)

IR Temperature Accuracy 9F (5C) or 5% (14~302F [-10 to 150C]) whichever is greater

Emissivity 0.95 maximum (4 presets and a fine tuning feature)

Distance to Spot ratio 30:1

Response time 150ms

Spectral Response 8~14um

Scanning type Continuous

Repeatability 0.5%

Image Detector Lepton

Display resolution 80 x 60 pixels

Field of View (FOV) 38.6o x 50o

Color Palettes Selectable: Iron, Rainbow, and Greyscale

Laser type Class 1

Laser power < 0.4mW

9.3 Electrical Specifications *Accuracy is given as (% of reading + counts of least significant digit) or as otherwise specified at 73.4F 9F (23C 5C), with relative humidity < 80%; AC Voltage & Current accuracies are specified from 1 % to 100 % of range, or as otherwise specified.

Maximum Crest Factor is <2:1 at full scale & <4:1 at half scale, and with frequency components within the specified frequency bandwidth for non-sinusoidal waveforms

Other AC Function notes:

ACV and ACA are ac coupled, true RMS

For all AC functions, the LCD displays 0 counts when the reading < 10 counts

Temperature Coefficient: Nominal 0.15 x (specified accuracy)/ oC @ 14oF ~ 64.4oF (-10oC ~ 18oC) or 82.4oF ~ 122oF (28oC ~ 50 oC), or as otherwise specified

AC Voltage*

RANGE and RESOLUTION Accuracy

50Hz ~ 60Hz

6.000V, 60.00V, 600.0V 0.7% + 3d

45Hz ~ 440Hz

6.000V, 60.00V, 600.0V 2.0% + 3d

Input Impedance: 10M, 54pF nominal

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 31

VFD_ACV* (with Low Pass Filter)

RANGE and RESOLUTION Accuracy 1)

10Hz ~ 100Hz (fundamental)

600.0V, 1.0% + 3d

100Hz ~ 400Hz (fundamental)

600.0V, 10% + 3d 2) 1) Not specified for fundamental frequency > 400Hz 2) Accuracy linearly decreases from 1% + 3d @100Hz to 10% + 3d @400Hz

ACmV*

RANGE and RESOLUTION Accuracy

50Hz ~ 60Hz

60.00mV 1) 2), 600.0mV 3) 1.0% + 3d

10Hz ~ 500Hz

60.00mV 1) 2), 600.0mV 3) 2.0% + 3d

Input Impedance: 10M, 54pF nominal 1) Non-zero residual readings 5d may appear as a zero volt input (short-circuit) when battery indication is

25% 2) Signal peak absolute values, including DC bias: < 130mVpeak 3) Signal peak absolute values, including DC bias: < 1300mVpeak

DC Voltage

RANGE and RESOLUTION Accuracy

60.00mV, 600.0mV, 0.3% + 2d

6.000V, 600.0V 0.4% + 2d

60.00V 0.5% + 2d

Input Impedance: 10M, 54pF nominal

Ohm

RANGE and RESOLUTION 1) Accuracy

600.0 0.3% + 3d

6.000k,60.00k, 600.0k 0.5% + 3d

6.000M 2) 0.9% + 2d

60.00M 3) 4) 5) 1.5% + 2d 1) Open Circuit Voltage: 1.6VDC typical 2) Constant Test Current: 0.1A Typical 3) Constant Test Current: 0.01A Typical 4) 5%+20d @ >30M, 5) Unspecified @ ambient > 40oC

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 32

Audible Continuity Tester

Continuity Threshold: Between 30 and 480

Continuity ON Response Time: <15ms

Capacitance

RANGE and RESOLUTION Accuracy

20.00nF, 200.0nF 1.5% + 8d

2000nF, 20.00F, 200.0F 1.5% + 2d

2000F 2.0% + 2d

10.00mF 5.0% + 10d

Accuracies stated for film capacitors (or better)

Diode Tester

RANGE and RESOLUTION Accuracy

3.000V 0.9% + 3d

Test Current: 0.3mA typical Open Circuit Voltage: < 3.2VDC typical

DC Current

RANGE and RESOLUTION Accuracy Burden Voltage

600.0A 1), 6000A 1) 1.0% + 3d 0.1mV/A

60.00mA 1), 600.0mA 1) 2) 0.7% + 3d

1.9mV/mA

6.000A 3) 4), 10.00A 3) 4) 5) 0.04V/A

1) A/mA DC accuracy is affected by extreme internal meter temperatures. For rated accuracies, allow linear proportional cool down intervals from 6 to 20 minutes after measuring currents from 3 to 10A continuously via A-input. 2) <400mA continuous; >400mA for <20 minutes ON per >5 minutes OFF 3) 10A continuous up to ambient 95F (35C); <15 minutes on per >5 minutes off @ 95~122F (35C ~ 50C) 4) >10A to 20A for <30 seconds ON per >5 minutes OFF

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 33

AC Current*

RANGE and RESOLUTION Accuracy Burden Voltage

45HZ ~ 440HZ

600.0A 1) 2), 6000A 2) 1.5% + 3d 0.1mV/A

60.00mA 1) 2), 600.0 mA 2) 3) 1.0% + 3d

1.9mV/mA

6.000A 1) 4) 5), 10.00A 4) 5) 6) 0.04V/A

1) Non-zero residual readings 5d may appear as a zero ampere input when battery indication is 25%

2) A/mA DC accuracy is affected by extreme interior temperature of the meter. For rated accuracies, allow linear proportional cool down intervals from 6 to 20 minutes after measuring currents from 3 to 10A continuously via A-input. 3) <400mA continuous; >400mA for <20 minutes ON per >5 minutes OFF 4) 10A continuous up to ambient 35oC; <15 minutes ON per >5 minutes OFF @ 95~122F (35C ~ 50C) 5) >10A to 20A for <30 seconds ON per >5 minutes OFF 6) Unspecified @ <0.5A

Temperature

RANGE and RESOLUTION Accuracy 1) 2) 3)

-40.0 oC ~ 0.0oC 1% + 1.5 oC

0.0 oC ~ 100.0oC 1% + 1 oC

100.0 oC ~ 400.0oC

-40.0 oF ~ 32.0 oF 1% + 3 oF

32.0 oF ~212.0 oF 1% + 2 oF

212.0oF ~ 752.0 oF

1) Accuracies assume meter interior has the same temperature as the ambient (isothermal stage) for correct junction voltage compensation. Allow enough time to reach the isothermal stage for a significant change of ambient temperature. It can take up to an hour for changes > 9F (5C). 2) Type-K thermocouple range & accuracy not included 3) For isothermal stage, allow linear proportional cool down intervals from 9 to 30 minutes after measuring currents from 3 to 10A continuously via A-input.

Line Frequency

Function Sensitivity (Sine RMS) Range

60mV, 600mV 50mV 10Hz - 50kHz

6V 5V 10Hz - 50kHz

60V 10V 10Hz - 50kHz

600V 50V 10Hz - 1kHz

600V VFD 50V 10Hz-400Hz

600A, 6000A 500A 10Hz - 5kHz

60mA, 600mA 50mA 10Hz - 5kHz

6A, 10A 8A 50Hz - 1kHz

Accuracy: 0.03% + 2d

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 34

FLEX*

RANGE and RESOLUTION Accuracy

50Hz ~ 400Hz

30.00A, 300.0A, 3000A 2.0% + 3d

Input Impedance: 10M, 54pF nominal

Non-Contact Voltage EF-Detection

Bar-Graph Indication EF-H (Hi Sensitivity) EF-L (Lo Sensitivity)

DMM mode IGM mode Typical Voltage (Tolerance)

-- - 20V (10V ~ 30V) 80V (40V ~ 150V)

---- - - 40V (20V ~ 60V) 160V (80V ~ 300V)

------ - - - 80V (40V ~ 150V) 320V (160V ~ 450V)

-------- - - - - 160V (80V ~ 300V) 640V (320V ~ 900V)

--------- - - - - - 320V (>320V) 1000V (>950V)

Indication: Bar-graph segments & audible beep tones proportional to the field strength

Detection Frequency: 50/60Hz

Detection Antenna: Top of meter

10. Technical Support Main Website http://www.flir.com/test

Technical Support Website http://support.flir.com

Technical support Email TMSupport@flir.com

Service/Repair Support Email Repair@flir.com

Support Telephone number +1 855-499-3662 option 3 (toll-free)

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 35

11. Warranties 11.1 FLIR Test & Measurement Imaging Product 10-year/10-year Warranty

Congratulations! You (the Purchaser) are now the owner of a world-class FLIR Imaging Test and Measurement product. A qualifying FLIR Imaging Test and Measurement product (the Product) purchased either directly from FLIR Commercial Systems Inc. and affiliates (FLIR) or from an authorized FLIR distributor that Purchaser registers on-line with FLIR is eligible for coverage under FLIRs industry-leading 10-10 Limited Warranty, subject to the terms and conditions in this document. This warranty only applies to purchases of Qualifying Products (see below) purchased after September 2015 and only to the original Purchaser of the Product. PLEASE READ THIS DOCUMENT CAREFULLY; IT CONTAINS IMPORTANT INFORMATION ABOUT THE PRODUCTS THAT QUALIFY FOR COVERAGE UNDER THE 10-10 LIMITED WARRANTY, PURCHASERS OBLIGATIONS, HOW TO ACTIVATE THE WARRANTY, WARRANTY COVERAGE, AND OTHER IMPORTANT TERMS, CONDITIONS, EXCLUSIONS AND DISCLAIMERS. 1. PRODUCT REGISTRATION. To qualify for FLIRs 10-10 Limited Warranty, the Purchaser must fully register the Product directly with FLIR on-line at www.flir.com WITHIN Sixty (60) DAYS of the date the Product was purchased by the first retail customer (the Purchase Date). PRODUCTS THAT ARE NOT REGISTERED ON-LINE WITHIN Sixty (60) DAYS OF THE PURCHASE DATE OR PRODUCTS WHICH DO NOT QUALIFY FOR THE 10-10 WARRANTY WILL HAVE A LIMITED ONE YEAR WARRANTY FROM THE DATE OF PURCHASE. 2. QUALIFYING PRODUCTS. Upon registration, a list of thermal Imaging Test and Measurement Products that qualify for coverage under FLIRs 10-10 Warranty can be found at www.flir.com/testwarranty 3. WARRANTY PERIODS. The 10-10 Limited Warranty has two separate periods of warranty coverage (the Warranty Period), depending on the Imaging Test and Measurement Product part: Product components are warranted for a period of ten (10) years from the Purchase Date; Thermal imaging sensor is warranted for a period of ten (10) years from the Purchase Date. Any Product that is repaired or replaced under warranty is covered under this 10-10 Limited Warranty for one hundred eighty days (180) days from the date of return shipment by FLIR or for the remaining duration of the applicable Warranty Period, whichever is longer. 4. LIMITED WARRANTY. In accordance with the terms and conditions of this 10-10 Limited Warranty, and except as excluded or disclaimed in this document, FLIR warrants, from the Purchase Date, that all fully registered Products will conform to FLIRs published Product specifications and be free from defects in materials and workmanship during the applicable Warranty Period. PURCHASERS SOLE AND EXCLUSIVE REMEDY UNDER THIS WARRANTY, AT FLIRS SOLE DISCRETION, IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PRODUCTS IN A MANNER, AND BY A SERVICE CENTER, AUTHORIZED BY FLIR. IF THIS REMEDY IS ADJUDICATED TO BE INSUFFICIENT, FLIR SHALL REFUND PURCHASERS PAID PURCHASE PRICE AND HAVE NO OTHER OBLIGATION OR LIABILITY TO BUYER WHATSOEVER. 5. WARRANTY EXCLUSIONS AND DISCLAIMERS. FLIR MAKES NO OTHER WARRANTY OF ANY KIND WITH RESPECT TO THE PRODUCTS. ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE (EVEN IF PURCHASER HAS NOTIFIED FLIR OF ITS INTENDED USE FOR THE PRODUCTS), AND NON-INFRINGEMENT ARE EXPRESSLY EXCLUDED FROM THIS AGREEMENT. THIS WARRANTY EXPRESSLY EXCLUDES ROUTINE PRODUCT MAINTENANCE, AND SOFTWARE UPDATES. FLIR FURTHER EXPRESSLY DISCLAIMS ANY WARRANTY COVERAGE WHERE THE ALLEGED NONCONFORMITY IS DUE TO NORMAL WEAR AND TEAR OTHER THAN SENSORS, ALTERATION, MODIFICATION, REPAIR, ATTEMPTED REPAIR, IMPROPER USE, IMPROPER MAINTENANCE, NEGLECT, ABUSE, IMPROPER STORAGE, FAILURE TO FOLLOW ANY PRODUCT INSTRUCTIONS, DAMAGE (WHETHER CAUSED BY ACCIDENT OR OTHERWISE), OR ANY OTHER IMPROPER CARE OR HANDING OF THE PRODUCTS CAUSED BY ANYONE OTHER THAN FLIR OR FLIRS EXPRESSLY AUTHORIZED DESIGNEE.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 36

THIS DOCUMENT CONTAINS THE ENTIRE WARRANTY AGREEMENT BETWEEN PURCHASER AND FLIR AND SUPERSEDES ALL PRIOR WARRANTY NEGOTIATIONS, AGREEMENTS, PROMISES AND UNDERSTANDINGS BETWEEN PURCHASER AND FLIR. THIS WARRANTY MAY NOT BE ALTERED WITHOUT THE EXPRESS WRITTEN CONSENT OF FLIR. 6. WARRANTY RETURN, REPAIR AND REPLACEMENT. To be eligible for warranty repair or replacement, Purchaser must notify FLIR within thirty (30) days of discovering of any apparent defect in materials or workmanship. Before Purchaser may return a Product for warranty service or repair, Purchaser must first obtain a returned material authorization (RMA) number from FLIR. To obtain the RMA number Owner must provide an original proof of purchase. For additional information, to notify FLIR of an apparent defect in materials or workmanship, or to request an RMA number, visit www.flir.com. Purchaser is solely responsible for complying with all RMA instructions provided by FLIR including but not limited to adequately packaging the Product for shipment to FLIR and for all packaging and shipping costs. FLIR will pay for returning to Purchaser any Product that FLIR repairs or replaces under warranty. FLIR reserves the right to determine, in its sole discretion, whether a returned Product is covered under warranty. If FLIR determines that any returned Product is not covered under warranty or is otherwise excluded from warranty coverage, FLIR may charge Purchaser a reasonable handling fee and return the Product to Purchaser, at Purchasers expense, or offer Purchaser the option of handling the Product as a non-warranty return. FLIR shall not be responsible for any data, images or other information that may be stored on the returned Product which was not included in the Product at the time of purchase. It is Purchasers responsibility to save any and all data prior to returning the Product for warranty service. 7. NON-WARRANTY RETURN. Purchase may request that FLIR evaluate and service or repair a Product not covered under warranty, which FLIR may agree to do in its sole discretion. Before Purchaser returns a Product for non-warranty evaluation and repair, Purchaser must contact FLIR by visiting www.flir.com to request an evaluation and obtain an RMA. Purchaser is solely responsible for complying with all RMA instructions provided by FLIR including but not limited to adequately packaging the Product for shipment to FLIR and for all packaging and shipping costs. Upon receipt of an authorized non-warranty return, FLIR will evaluate the Product and contact Purchaser regarding the feasibility of and the costs and fees associated with Purchasers request. Purchaser shall be responsible for the reasonable cost of FLIRs evaluation, for the cost of any repairs or services authorized by Purchaser, and for the cost of repackaging and returning the Product to Purchaser. Any non-warranty repair of a Product is warranted for one hundred eighty days (180) days from the date of return shipment by FLIR to be free from defects in materials and workmanship only, subject to all of the limitations, exclusions and disclaimers in this document.

FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA 37

Corporate Headquarters FLIR Systems, Inc. 2770 SW Parkway Avenue Wilsonville, OR 97070 USA Telephone: +1 503-498-3547 Customer Support Technical Support Website http://support.flir.com Technical Support Email TMSupport@flir.com Service and Repair Email Repair@flir.com Customer Support T

Manualsnet FAQs

If you want to find out how the DM166 Flir works, you can view and download the Flir DM166 Multimeter User Manual on the Manualsnet website.

Yes, we have the User Manual for Flir DM166 as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Manual should include all the details that are needed to use a Flir DM166. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir DM166 Multimeter User Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir DM166 Multimeter User Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir DM166 Multimeter User Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir DM166 Multimeter User Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir DM166 Multimeter User Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.