User Manual 

Heavy Duty  Differential Pressure Manometer 

Model SDL720 

Additional User Manual Translations available at www.extech.com 

2 SDL720-en-GB_V1.0 12/16

Introduction  Congratulations on your purchase of the Extech SDL720 Pressure Manometer. This device  measures gauge and differential pressure in the range of 29 psi. Supported measurement units  are bar, psi, Kg/cm2, mm Hg, inch Hg, meters of H2O, inches of H2O, atmosphere, hPA, and kPA.  Logged data readings are stored on an SD card for transfer to a PC. This meter is shipped fully  tested and calibrated and, with proper use, will provide years of reliable service. Please visit our  website (www.extech.com) to check for the latest version of this User Manual, other  Translations of this manual, and Customer Support. 

Safety 

International Safety Symbols 

This symbol, adjacent to another symbol or terminal, indicates the user must refer to the  manual for further information.   

Meter Description 

1. P1 Input (positive +) 

2. P2 Input (negative ) 

3. LCD Display 

4. HOLD and Backlight   key 

5. MAXMIN key 

6. SET and Clock  key 

7. RS232 output jack 

8. Reset button 

9. Power Adaptor jack 

10. SD card slot 

11. ENTER and LOG key 

12. Down arrow  key   (Press & key at same time to zero meter) 

13. Up arrow  / UNIT key  (Press & key at same time to zero meter) 

14. Power ONOFF   key 

Notes: Items 7, 8, and 9 are located behind the snapoff compartment cover on meters right side.

Battery compartment, tilt stand, and tripod mount are located on the rear of the instrument. 

1

11 12 13

14

2

3

4 5 6

7 8 9

10

3 SDL720-en-GB_V1.0 12/16

Getting Started 

Power ONOFF 

Power the meter by pressing and holding the power button    for at least 1.5 seconds. 

Press and hold the power button for at least 1.5 seconds to power OFF the meter. 

This meter is powered by six (6) 1.5VDC AA batteries or by optional AC adaptor. If the  meter will not switch ON please check that fresh batteries are installed in the rear battery  compartment or, in the case of the AC adaptor, check that the adaptor is connected  correctly to the meter and to an AC source. 

Display Backlight 

To turn the display backlight ON or OFF, press and hold the backlight   button for at least 1.5  seconds. The meter will beep when switching the backlight ON or OFF unless the beeper is  disabled. 

Units of Measure 

The currently selected unit of measure is shown below the measurement value on the meters  LCD. To change the unit of measure, press and hold the UNIT button until the desired unit of  measure appears and then release the UNIT button. The meter begins scrolling through the  available units of measure (see table below) after the UNIT button has been depressed for at  least 1.5 seconds. 

UNIT  DISPLAY INDICATOR 

psi  PSI 

Inch Hg  In Hg 

Inch H2O  In H2O 

hPA  hPA 

kPA  _PA 

Bar  bAr 

Kg/cm2  _g C2 

mm Hg  Hg 

Meters H2O  t H2O 

Atmospheres  AtP 

ZERO Adjustment 

To null, or zero, the display for a connected sensor, press and hold the up and down arrow  keys for at least 1.5 seconds. The meter will show CAL 0 on the bottom left of the screen. 

Data Hold 

To freeze a displayed reading on the LCD, momentarily press the HOLD button (the HOLD icon  will appear above the reading). To exit HOLD, press the HOLD button again. 

4 SDL720-en-GB_V1.0 12/16

Setup Mode 

Basic settings at a glance  To view the current configuration of the meter with regard to time, date, and datalogging  sampling rate press the SET button momentarily. The meter will now display the configuration in  quick succession. If the information is missed on the first try, simply press the SET button again  until all of the information is noted. 

Accessing the Setup mode  1. Press and hold the SET button for at least 1.5 seconds to access the Setup menu. 

2. Press the SET button momentarily to step through the available parameters. The parameter  type is shown on the bottom of the LCD and the current selection for that type is shown  above it. 

3. When a parameter is displayed that is to be changed, use the arrow keys to change the  setting. Press the ENTER button to confirm a change. 

4. Press and hold the SET button for at least 1.5 seconds to exit the Setup mode. Note that the  meter automatically switches out of the Setup mode if no key is pressed within 7 seconds.  

5. The available Setup parameters are listed below. Additional detailed information is  provided below this list:  

dAtE  Set the clock (Year/Month/Date; Hours/Minutes/Seconds)  SPt  Set the datalogger sampling rate (Hours/Minutes/Seconds)  PoFF  Automatic poweroff management (Enable or disable the autopower off 

function)  bEEP  Set the beeper sound ON/OFF  dEC  Set the numerical format; USA (decimal: 20.00) or European (comma: 20,00)  Sd F  Format the SD memory card 

Setting the Clock Time   1. Access the dAtE parameter. 

2. Use the arrow keys to change a value 

3. Use the ENTER button to step through the selections 

4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode  (or simply wait 7 seconds for the meter to automatically switch to the normal operating  mode). 

5. The clock will keep accurate time even when the meter is switched off. However, if the  battery expires the clock will have to be reset after fresh batteries are installed. 

 Setting the Datalogger Sampling Time (Rate)   1. Access the SPt parameter. 

2. The sampling rate can be set from 0 seconds (manual log mode) up to 8 hours, 59 minutes,  and 59 seconds. (Data lose may occur when logging at a 1 second rate) 

3. Use the ENTER button to move through the Hours, Minutes, and Seconds digit groups and  use the arrow keys to change the digit values. 

4. Press the ENTER button to confirm the entry. 

5. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode  (or simply wait 7 seconds for the meter to automatically switch to the normal operating  mode). 

5 SDL720-en-GB_V1.0 12/16

 Enabling/Disabling the Auto Power OFF Feature   1. Access the PoFF parameter. 

2. Use the arrow buttons to select ON or OFF. With the Auto Power OFF feature enabled, the  meter will automatically switch OFF after 10 minutes of inactivity. 

3. Press ENTER to confirm setting. 

4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode  (or simply wait 7 seconds for the meter to automatically switch to the normal operating  mode). 

 Set the Beeper Sound ON or OFF   1. Access the bEEP parameter. 

2. Use the arrow buttons to select ON or OFF.   

3. Press ENTER to confirm setting. 

4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode  (or simply wait 7 seconds for the meter to automatically switch to the normal operating  mode). 

Numerical Format (comma or decimal)   European and USA numerical formats differ. The meter defaults to USA mode where a decimal  point is used to separate units from tenths, i.e. 20.00; The European format uses a comma, i.e.  20,00 to separate units from tenths. To change this setting: 

1. Access the dEC parameter. 

2. Use the arrow buttons to select USA or EUro.   

3. Press ENTER to confirm setting. 

4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode  (or simply wait 7 seconds for the meter to automatically switch to the normal operating  mode). 

SD Card FORMATTING   1. Access the Sd F parameter. 

2. Use the arrow buttons to select YES to format the card (select NO to abort). Note that all  data on the card will be lost if formatting is attempted. 

3. Press ENTER to confirm selection. 

4. Press ENTER again to reconfirm. 

5. The meter will automatically return to the normal operating mode when formatting is  complete. If not, press and hold the SET button for at least 1.5 seconds to exit to the normal  operation mode. 

System Reset  If the meters keys become inoperable or if the display freezes the Reset button can be used to  reset the instrument. 

Use a paper clip or similar item to momentarily press the reset button located on the lower  right side of the instrument under the snapoff compartment cover. 

After pressing the Reset button, switch the instrument ON by pressing and holding the  POWER key for at least 1.5 seconds. If using the power adaptor unplug the adaptor and then  plug it back in again to power the meter. 

6 SDL720-en-GB_V1.0 12/16

MaxMin Reading Record  

For a given measurement session, this meter can record the highest (MAX) and the lowest (MIN)  readings for later recall. 

1. Press the MAXMIN button momentarily to access this mode of operation (REC icon  appears). 

2. The meter is now recording the MAX and MIN readings. 

3. Press the MAXMIN button again to view the current MAX readings (MAX icon appears).  The readings on the display are now the highest readings encountered since the REC icon  was switched on (when the MAXMIN button was first pressed). 

4. Press the MAXMIN button again to view the current MIN readings (MIN icon appears). The  readings on the display are now the lowest readings encountered since the REC icon was  switched on (when the MAXMIN button was first pressed). 

5. To exit the MAXMIN mode, press and hold the MAXMIN button for at least 1.5 seconds.  The meter will beep, the RECMAXMIN icons will switch off, the MAXMIN memory will  clear, and the meter will return to the normal operating mode.  

Fittings Connection  Connect tubing to the P1 (+), P2 (), or both input port(s). If both inputs are used (differential  mode), the meter displays a positive pressure reading if the P1 (+) pressure is greater than P2 ()  and a negative reading if P2 () is greater than P1 (+). 

Datalogging 

Types of Data Recording 

Manual Datalogging: Manually log up to 99 readings onto an SD card via pushbutton press. 

Automatic Datalogging: Automatically log data onto an SD memory card where the number  of data points is virtually limited only by the card size. Readings are logged at a rate specified  by the user. 

SD Card Information 

Insert an SD card (from 1G size up to 16G) into the SD card slot at the bottom of the meter.  The card must be inserted with the front of the card (label side) facing toward the rear of the  meter. 

If the SD card is being used for the first time it is recommended that the card be formatted  and the loggers clock set to allow for accurate date/time stamping during datalogging  sessions. Refer to the Setup Mode section for SD card formatting and time/date setting  instructions. 

European and USA numerical formats differ. The data on the SD card can be formatted for  either format. The meter defaults to USA mode where a decimal point is used to separate  units from tenths, i.e. 20.00. The European format uses a comma, i.e. 20,00. To change this  setting, refer to the Setup Mode section. 

7 SDL720-en-GB_V1.0 12/16

Manual Datalogging  In the manual mode the user presses the LOG button to manually log a reading onto the SD card. 

1. Set the sampling rate to 0 seconds as described in the Setup Mode section. 

2. Press and hold the LOG button for at least 1.5 seconds and the DATALOGGER icon will  appear on the LCD; the lower portion of the display will show pn (n = memory position  number 199). Note that if PSI is set as the unit of measure it appears as P51 (where a 5 is  used as an S) in the same area of the LCD where memory locations are shown. This can be  disorienting at first. 

3. Momentarily press the LOG button to store a reading. The DATALOGGER icon will flash and  the beeper will sound (if set to ON) each time a data point is stored. 

4. Use the  and  buttons to select one of the 99 data memory positions in which to record.   

5. To exit the manual datalogging mode, press and hold the LOG button for at least 1.5  seconds. The DATALOGGER icon will switch off. 

Automatic Datalogging 

In automatic datalogging mode the meter takes and stores a reading at a userspecified sampling  rate onto an SD memory card. The meter defaults to a sampling rate of two seconds. To change  the sampling rate, refer to the Setup Mode section (the sampling rate cannot be 0 for  automatic datalogging): 

1. Select the sampling rate in the Setup Mode to a value other than zero. 

2. Press and hold the LOG button for at least 1.5 seconds. The meter will flash the  DATALOGGER icon at the selected sampling rate indicating that readings are now being  automatically recorded to the SD card. 

3. If a card is not inserted or if the card is defective, the meter will display EMPTY and exit the  DATALOGGER mode. In this case, switch the meter OFF and try again with a valid SD card. 

4. Pause the datalogger by pressing the LOG button momentarily. The DATALOGGER icon will  stop flashing and the sample rate will display for a short time. To resume logging simply  press the LOG button again momentarily. 

5. To terminate the datalogging session press and hold the LOG button for at least 1.5  seconds. 

6. When an SD card is used for the first time a folder is created on the card and named  PMA01. Up to 99 spreadsheet documents (each with 30,000 readings) can be stored in this  folder. 

7. When datalogging begins a new spreadsheet document named PMA01001.xls is created on  the SD card in the PMA01 folder. The data recorded will be placed in the PMA01001.xls  document until 30,000 readings are reached. 

8. If the measurement session exceeds 30,000 readings, a new document will be created  (PMA01002.xls) where another 30,000 readings can be stored. This method continues for  up to 99 documents, after which another folder is created (PMA02) where another 99  spreadsheet documents can be stored. This process continues in this same fashion with  folders PMA03 through PMA10 (last allowable folder). 

8 SDL720-en-GB_V1.0 12/16

SD Data Card to PC Data Transfer   1. Complete a datalogging session as detailed above in the previous sections. Hint: For the 

first few tests, simply record a small amount of test data. This is to ensure that the  datalogging process is well understood before committing to critical, large scale  datalogging. 

2. With the meter switched OFF, remove the SD Card. 

3. Plug the SD Card directly into a PC SD card reader. If the PC does not have an SD card slot,  use an SD card adaptor (available at most outlets where computer accessories are sold). 

4. Power the PC and run a spreadsheet software program. Open the saved documents in the  spreadsheet software program (see example spreadsheet data screen below). 

Spreadsheet data example 

AC Power Adaptor 

This meter is normally powered by six (6) 1.5V AA batteries. An optional 9V power adaptor is  available. When the adaptor is used, the meter is permanently powered and the power button  will be disabled. 

9 SDL720-en-GB_V1.0 12/16

Hose Couplings 

This meter is supplied with a pair of hose couplings.  They fit 4.0mm (0.157) tubing. 

See diagram on how to connect these couplings to the hose and the meter. 

10 SDL720-en-GB_V1.0 12/16

Battery Replacement and Disposal  When the low battery icon   appears on the LCD, the batteries must be replaced. Several 

hours of accurate readings are still possible in this condition; however, batteries should be  replaced as soon as possible: 

Remove the two (2) Phillips screws from the rear of the meter (directly above the top of the  tilt stand). 

Remove and safely place the battery compartment and screws where they will not be  damaged or lost. 

Replace the six (6) 1.5V AA batteries observing correct polarity. 

Replace the battery compartment cover with the two (2) Phillips screws. 

   Never dispose of used batteries or rechargeable batteries in household waste.  As consumers, users are legally required to take used batteries to appropriate  collection sites, the retail store where the batteries were purchased, or wherever  batteries are sold. 

Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end oflife devices to a designated collection point for the disposal of electrical and electronic  equipment.     Other Battery Safety Reminders  o Never dispose of batteries in a fire. Batteries may explode or leak.

o Never mix battery types. Always install new batteries of the same type.

11 SDL720-en-GB_V1.0 12/16

Specifications 

General Specifications 

Display  Backlit LCD; LCD size: 51 x 37mm (2 x 1.5) 

Status indicators  Overrange audible beep and low battery display icon   

Measurement Units  mBar, psi, Kg/cm2, hPA, mm Hg, inch Hg, meters H2O, kPA, inches  H2O, ATP 

Accuracy (Meter)   (2%FS) at ambient temperature 23C ( 5C) 

Input circuit  Differential inputs (P1 and P2) 

Input Ports  Two metal 0.197(5.0mm), barbed for 0.157" (4.0mm) ID tubing 

Hose couplings  Replacement pair of hose couplings (407915) 

Sensor Builtin piezoelectric sensors 

Datalogger Sampling Rate  AUTO LOGGING: From 1 second up to 8 hours 59 min 59  sec. (Data lose may occur when logging at a 1 second rate) 

  MANUAL LOGGING: Set the sampling rate to 0 seconds.  Select 1 to 99 locations. 

Data error number   0.1% number of total saved data 

Memory Card  SD memory card; 1G to 16GB size 

Data Hold  Freeze the displayed reading 

Memory Recall  Record and Recall the Maximum and Minimum readings 

Display update rate  Approx. 1 second 

Operating Temperature  0 to 50C (32 to 122F) 

Operating Humidity  85% R.H. max. 

Auto Power OFF   After 10 minutes of inactivity (can be disabled) 

Power Supply  Six (6) 1.5 VDC batteries (optional 9V AC adaptor) 

Power Consumption  Normal operation (backlight and datalogger OFF): approx.  7mAdc 

  With backlight OFF and datalogging ON: approx. 25mAdc 

  With backlight ON add approx. 10mAdc 

Weight  265g (0.59 lbs.) 

Dimensions  190 x 68 x 45mm (7.5 x 2.7 x 1.8) 

12 SDL720-en-GB_V1.0 12/16

Range/Resolution Specifications 

Range/Units (Max.)  Resolution 

 2000 mbar  1 

 29 psi  0.01 

 2.040 Kg/cm2  0.001 

 1500 mm Hg  1 

 59.05 inches Hg  0.02 

 200 kPa  0.1 

 2000 hPA  1 

 802 inches H2O  0.5 

 20.40 m H2O  0.01 

 1.974 ATP  0.001 

Conversions 

Useful conversion Factors 

From  To  Multiplier 

in of H2O  in of Hg  0

Manualsnet FAQs

If you want to find out how the Extech Flir works, you can view and download the Flir Extech SDL720 Manometer User Manual on the Manualsnet website.

Yes, we have the User Manual for Flir Extech as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Manual should include all the details that are needed to use a Flir Extech. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir Extech SDL720 Manometer User Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir Extech SDL720 Manometer User Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir Extech SDL720 Manometer User Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir Extech SDL720 Manometer User Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir Extech SDL720 Manometer User Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.