Flir Extech HD750 Manometer User Guide PDF

1 of 8
1 of 8

Summary of Content for Flir Extech HD750 Manometer User Guide PDF

USER GUIDE 

Heavy Duty  Differential Pressure Manometer 

Model HD750 

HD750-en-GB_V1.8 4/15 2

Introduction  Congratulations on your purchase of the Extech Manometer Model HD750. This device measures  gauge and differential pressure in the range of 5.000 psi. The HD750 offers eleven (11) units of  measure (user selectable by front panel pushbutton) and has a differential input that uses  convenient quickdisconnect fittings. Advanced features include DATA HOLD, MINMAXAVG  RECORD mode, ZERO/OFFSET, AUTO POWER OFF, and USB PC interface. This meter is shipped fully  tested and calibrated and, with proper use, will provide years of reliable service. Please visit our  website (www.extech.com) to check for the latest version of this User Guide, Product Updates, and  Customer Support.   

Meter Description 

1. P1 input   

2. AC adaptor jack 

3. P2Input  

4. LCD Display  Indicates the measurement data, unit of measure,  elapsed timer, and function symbols as described in this manual 

5. USB PC interface jack (3.5mm)  For use with data acquisition  software 

6. UNIT button  Press to select the unit of measure 

7. DIF button Press to display differential pressure. Also used to  OFFSET the displayed readings. 

8. RECORD button Press to access the MIN/MAX/AVG recording  mode 

9. HOLD / ZERO button Press to freeze the displayed reading. Also  used to zero the display (press and hold until display shows all  zeroes). 

10. BACKLIGHT  button  Press to switch the display backlight ON.  The backlight will automatically turn OFF after 40 seconds. 

11. POWER button  Press to turn the meter ON or OFF 

12. Battery compartment (on rear) 

HD750-en-GB_V1.8 4/15 3

Display 

1. USB data output active 

2. Differential Pressure Mode 

3. Data Hold Mode 

4. Primary display 

5. Record Mode 

6. MAX/MIN/AVG indicators 

7. Low Battery Indicator 

8. Pressure unit of measure 

9. Elapsed Timer 

HD750-en-GB_V1.8 4/15 4

Basic Operation 

Meter Power 

The HD750 uses a 9V battery for power. If the meter does not switch on when the POWER  button is pressed or if the display shows BAT,replace the battery. 

When the POWER buttons is pressed, the meter performs a short selftest and reverts to the  settings in place when the meter was last turned off. 

Automatic Poweroff Feature 

In order to conserve battery life, the meter has an automatic power off feature (after 20  minutes). To defeat this feature, press and hold the HOLD button while turning the meter on.  Release the HOLD button when n appears on the display. The meter will now remain on until  the user turns it off or until the battery weakens. Note that the Auto Power OFF feature will be  reactivated the next time the meter is turned off and then on. 

Unit of Measure 

Press the UNIT button to select the unit of measure. This meter offers eleven (11) units of  measure as listed below. 

Unit of Measure  Display icon 

Pounds per square inch  psi 

Inches of Mercury  inHg 

Millibar  mbar  

Millimeters of Mercury  mmHg 

KiloPascals  kPa 

Kilograms per centimeter squared  Kgcm2 

Inches of water  inH2O 

Feet of water  ftH2O 

Centimeters of water  cmH2O 

Bars  Bar 

Ounces per square inch  ozin2 

Fittings Connection  Connect tubing to the P1, P2 (or both) input port(s). If both inputs are used (differential mode),  the meter displays a positive pressure reading if the P1 pressure is greater than P2 and a  negative reading if P2 is greater than P1. 

HD750-en-GB_V1.8 4/15 5

Zero Adjust and Offset function  The unit must be kept in the upright position when ZEROING or when taking measurements. To  ZERO the meter: Before use and without fittings attached to the meter, press and hold the HOLD  button for 3 seconds or until the display shows all zeroes. 

To OFFSET meter readings: Press the DIF button while taking a measurement. Subsequent  readings will represent the difference between the current measurement and the measurement  that was on the display when the DIF button was pressed. 

MINMAXAVG Record Mode with Elapsed Timer 

The RECORD function tracks the maximum (MAX), minimum (MIN), and average (AVG) readings  with a time stamp. The elapsed timer is displayed on the lower righthand corner of the LCD and  informs the user of the time a MAX or MIN reading was recorded. 

1. Press the RECORD button once. The REC indicator will appear on the display and the  displayed elapsed timer will start (HH:MM:SS format [hours:minutes:seconds]). 

2. Press the RECORD button again to view the highest (MAX) reading recorded since the  RECORD button was first pressed. The MAX indicator and maximum reading will appear on  the display. The elapsed timer will show the time of the MAX event. 

3. Press the RECORD button again, the MIN indicator and the minimum value will appear on  the display. The elapsed timer will show the time of the MIN event. 

4. Press the RECORD button again, the AVG indicator and the average value will appear on the  display. The elapsed timer will continue to run. 

5. To return to normal operation, press and hold the RECORD button until the display  indicators REC, MAX, etc. switch off. 

Note that in the RECORD mode, the backlight and the power buttons are the only available  functions. 

USB PC Interface 

The meter has a builtin USB port for use with the Manometer Data Acquisition software. The  software allows the user to view, save, export, and print readings from the HD750. Refer to the  documentation included with the software for complete details. 

Displayed Error Codes 

The following codes are displayed when the meter encounters an error. 

ERR 1: Pressure input exceeds specified limits 

ERR 2: Pressure input is too low to read 

ERR 3: The Differential Pressure input exceeds specified limits 

ERR 4: The Differential Pressure input is too low to read 

HD750-en-GB_V1.8 4/15 6

Specifications 

General Specifications 

Display  Four digit LCD with function indicators 

Units of Measure  BAR, mBAR, psi, kg/cm2, mm Hg, inches Hg, kPa, feet H2O, inches  H2O, cm H2O, and oz in

Accuracy  0.3%FS  (@ 25C)

Repeatability  0.2% to 0.5% FS

Linearity/Hysteresis  0.3%FS

Combined Accuracy  1.1%FS 

Maximum Pressure  10psi  

Response time  0.5s typical 

Input circuit  Differential inputs (P1 and P2) 

Input Ports  Two metal 4.8mm, barbed for 1/8" (3.17mm) ID tubing 

Data Hold  Freezes displayed reading 

Sensor  Builtin piezoelectric sensors 

Zero / Offset function  Pushbutton activation 

Sample rate  0.5 sec. (approx.) 

Compatibility  Air or noncorrosive and nonionized gas (not intended for use with  liquids) 

Auto Power OFF  Meter turns off after approx. 20 minutes to conserve energy (can  be deactivated) 

PC Interface  Builtin USB serial interface 

Storage conditions  0 to 60 oC (14 to 140oF); < 80% RH 

Operating conditions  0 to 50 oC (32 to 122oF); < 80% RH 

Battery power  9V alkaline battery (BAT low battery indicator) 

Battery life  125 hours typical 

Dimensions  210 x 75 x 50mm (8.2 x 2.9 x 1.9) 

Weight  280g (9.8oz) 

HD750-en-GB_V1.8 4/15 7

Range Specifications 

Range/Units (Max.)  Resolution 

 0.344bar  0.001 

 344.7mbar  0.1 

 5.000psi  0.001 

 0.351Kg/cm2  0.001 

 258.5 mm Hg  0.1 

 10.18 inches Hg  0.01 

 34.47kPa  0.01 

 11.53feet H2O  0.01 

 138.3inches H2O  0.1 

 351.5cm H2O  0.1 

 80.00oz in2  0.01 

Warning: Pressure inputs of 10 psi or higher will damage the instrument. Do not apply pressure higher than the maximum pressure listed in the specification table

HD750-en-GB_V1.8 4/15 8

Battery Replacement 

When the low battery indicator BAT appears on the display or if the meter does not switch on  when the POWER button is pressed, it is necessary to replace the battery. 

To replace the battery: 

1. Open the battery compartment by sliding the meters rear panel in the direction of the  embossed arrow. 

2. Replace the 9V alkaline battery 

3. Replace the battery compartment cover. 

  Never dispose of used batteries or rechargeable batteries in household waste.  As consumers, users are legally required to take used batteries to appropriate  collection sites, the retail store where the batteries were purchased, or wherever  batteries are sold. 

Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end oflife devices to a designated collection point for the disposal of electrical and electronic  equipment.     Other Battery Safety Reminders  o Never dispose of batteries in a fire. Batteries may explode or leak.

o Never mix battery types. Always install new batteries of the same type.

Conversions 

Useful conversion Factors 

From  To  Multiplier 

in of H2O  in of Hg

Manualsnet FAQs

If you want to find out how the Extech Flir works, you can view and download the Flir Extech HD750 Manometer User Guide on the Manualsnet website.

Yes, we have the User Guide for Flir Extech as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Guide should include all the details that are needed to use a Flir Extech. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir Extech HD750 Manometer User Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir Extech HD750 Manometer User Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir Extech HD750 Manometer User Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir Extech HD750 Manometer User Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir Extech HD750 Manometer User Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.