User's Guide

Humidity + Temperature + Pressure Datalogger

Model RHT50

RHT50-en-GB_V2.4 4/15 2

Introduction

Congratulations on your purchase of the Relative Humidity / Temperature / Pressure Datalogger. This datalogger can measure and store up to 10,000 readings. The user can easily configure the logging rate, high/low alarm and start-mode, and download the stored data by plugging the module into a PC's USB port and running the supplied PC software. Data can then be graphed, printed and exported to other applications. The data logger is supplied with a long-life lithium battery, which can typically allow logging for one year. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.

Features Memory 10,000 temperature, humidity and pressure readings

Selectable data sampling rate: 1 minute to 18 hours

Manual and automatic start

Status Indication via Red/Yellow LED and Green LED

USB Interface for Set-up and Data Download

User-Programmable Alarm Thresholds

Long battery life

Description 1. USB connector

2. Alarm LED (red/yellow)

3. Record LED (green)

4. Start button

5. Battery cover latch

6. Tripod mount

7. RH, Temperature, Pressure sensors

8. Mounting bracket

RHT50-en-GB_V2.4 4/15 3

Operation

INSTALLING THE GRAPHING SOFTWARE

Install the supplied WindowsTM PC Datalogger Software by placing the supplied program disk in the PCs CD-ROM drive. If the installation program does not automatically open and provide on-screen prompts, open and browse the CD-ROM drive, then find and double-click on the SETUP.EXE file included on the program disk. Follow the on-screen prompts to complete the installation. Before running the graphing software, please connect the RHT50 to the PC and install the USB Driver software as described in the next section.

INSTALLING THE USB DRIVER

Plug the RHT50 into an available USB port on the PC. If this is the first time that the RHT50 has been connected to the particular port on the PC, the following screen will appear: Make sure the program disk is still in the PC CD-ROM drive. Select the recommended installation and follow the on-screen prompts to complete the driver installation. If this is a repeat connection of the RHT50 to the USB port, the driver installation process will be automatically skipped. Refer to the HELP utility in the software program for detailed instructions on using the software to download and process recorded data. The LED status guide provided below is offered for easy reference.

RHT50-en-GB_V2.4 4/15 4

LED STATUS GUIDE

LEDs Record Alarm

Meaning

No LED Logging not started

No LED Logging stopped *

No LED Battery discharged **

Green

Green flash (single): Logging indication *****

Green flashes (five) Reset

Green flashes (six) Start Logging

Red

Red double flash every 30 sec. Low battery ***

Red single flash Logging, alarm event ****

Red flashes (six) Stop logging *

Yellow

Yellow flash every 20 seconds Memory full *

* Connect the datalogger to the PC and download data ** Replace the battery, data will not be lost *** Logging will stop, data will remain in memory **** Occurs if an alarm is exceeded ***** Flash cycle time is set from the software

RHT50-en-GB_V2.4 4/15 5

Specifications

Range Accuracy

Relative Humidity 0 to 20 and 80 to 100% 5.0%

20 to 80% 3.5%

Temperature -40 to 32oF and 104oF to 158oF -40 to 0oC and +40 to +70oC

3.6oF/2oC

32oF to104oF/0oC to +40oC 1.8oF/1oC

Pressure 950 to 1050hPa

(kPa, Bar, PSI)

Accuracy: 0.25%FSO @ (25oC) Repeatability: 0.25%Max+0.3%FSO) Linearity/Hysteresis: 0.24%FSO Temperature coefficient: 0.1%FSO

Notes: FSO: Full Scale Output 1psi = 68.947 mbar = 0.068947 bar 1 psi = 6894.7 Pa = 68.947 hPa = 6.8947 kPa

Resolution 0.1%RH, 0.1oC/F, 0.1hPa

Data storage 10,000 measurements

Logging rate Selectable sampling interval: From 1 minute up to 18 hours

Operating temperature -35 to 80oC (-31 to 176oF)

Battery type 3.6V Lithium (1/2 AA) (SAFT LS14250, Xeno XL050F or equivalent)

Battery life 1 year (typ.) depending on logging rate, ambient temperature, & use of Alarm LEDs

Dimensions/Weight 101x25x23mm (4x1x.9) / 172g (6oz)

RHT50-en-GB_V2.4 4/15 6

Battery Replacement It is highly recommended that the battery be replaced every 12 months, or prior to logging critical data.

The RHT50 does not lose its stored readings when the battery weakens or when the battery is replaced; the data logging process will however be stopped and cannot be re-started until the battery has been replaced and the logged data has been downloaded to PC.

Use only 3.6V lithium batteries. Before replacing the battery, remove the datalogger from the PC. Follow diagrammatic steps below.

NOTE: Leaving the RHT50 plugged into the PC USB port longer than necessary will cause some of the battery capacity to be lost.

1. Unscrew the protective sensor cap counter-clockwise. 2. Remove the protective cap and safely set it aside. 3. While inserting a small flat blade screwdriver into the slot and applying moderate pressure to the

locking tab, pull the front (USB side) of the RHT50 outward. 4. Completely disassemble the meter to access the battery compartment. 5. Insert battery. 6. Re-assemble the meter. 7. Replace the cap. 8. Hand-tighten the cap in a clockwise motion. Do not over tighten.

1 2

3

6

7 8

5

4

WARNING: Handle lithium batteries carefully, observe warnings on battery casing. Dispose of in accordance with local regulations.

You, as the end user, are legally bound (EU Battery ordinance) to return all used batteries, disposal in the household garbage is prohibited! You can hand

Manualsnet FAQs

If you want to find out how the Extech Flir works, you can view and download the Flir Extech RHT50 Datalogger User's Guide on the Manualsnet website.

Yes, we have the User's Guide for Flir Extech as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User's Guide should include all the details that are needed to use a Flir Extech. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir Extech RHT50 Datalogger User's Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir Extech RHT50 Datalogger User's Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir Extech RHT50 Datalogger User's Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir Extech RHT50 Datalogger User's Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir Extech RHT50 Datalogger User's Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.