Model 42270 Temperature & Humidity Datalogger

Model 42260 Temperature Datalogger

Introduction

Congratulations on your purchase of the Extech Datalogging Module. The following models are covered in this manual.

Model 42270: Temperature and Humidity Datalogging Module Model 42260: Temperature-only Datalogging Module

The Dataloggers can be used to monitor the temperature of greenhouses, warehouses, food transports, aircraft cabins, refrigerated trucks, containers, museums, and HVAC equipment. The Model 42270 records Relative Humidity readings in addition to temperature readings. Instructions on how to program the dataloggers are not included this manual. Programming and data retrieval instructions are included in the 42275, 42265, 42276 and 42266 kits.

Description

1. Mounting hole

2. Status LEDs (Record and Alarm)

3. LCD display

Displays

LCD Display

REC: Displayed while logging readings. HI and LOW: Displayed when the High or Low Alarm limit is exceeded. RH%: Relative Humidity (42270 only) C or F: Temperature units. COMM: Appears when Datalogger is communicating with a PC.

Recording Status Indication

Both the RED status LED and the display indicator REC will flash every 5 seconds if the sampling rate is set to 5 seconds or higher. If the sampling rate is set lower than 5 seconds, the indicators will flash every 1, 2, 3, or 4 seconds as programmed.

ALARM Status Indication

Both the YELLOW status LED and the display indicator ALM will flash when the recorded value is higher than the user programmed High Alarm value or lower than the Low Alarm value. The Alarm status LED and the ALM display indicator flash every 5 seconds if the sampling rate is programmed for 5 seconds or higher. If the sampling rate is set lower than 5 seconds, the indicators will flash every 1, 2, 3, or 4 seconds as programmed.

Low Battery Indication

When the lithium battery voltage nears the critical operating level, the LCD displays LO. To replace the battery, refer to the Battery Replacement section of this manual.

Specifications

Front panel status LEDs Two (2): RECORD and ALARM Temperature range -40 to 85C (-40 to 185F) Temperature resolution 0.1 up to 99.9; 1 from 100 to 185 Temperature accuracy 0.6C (1.2F) from -20 to 50C (-4 to 122F)

1.2C (2.4F) all other ranges Humidity range & accuracy 0.0 to 99.9% Relative Humidity; Accuracy: 3% Datalogger Memory 16,000 temperature readings (8,000 temperature and

8,000 humidity readings on the 42270) Sampling rate Programmable from 1 to 7200 seconds (2 hours) Power supply One 3.6V AA lithium battery Battery Life 1 year: 5 second sample rate in sleep mode

3.8 months: 24 hour sample rate in non-sleep mode Low battery Indicator Display shows LO Dimensions 76.2 x 57.2 x 22.3mm (3 x 2.3 x 0.88")

The datalogger display will not switch on until activated by software or the printer

programmer.

THESE DATALOGGERS CANNOT BE USED WITHOUT THE SOFTWARE FROM KIT 42275 (42265) OR THE PRINTER

PROGRAMMER 42276 (42266).

Battery Replacement

When the low battery icon (LO) appears on the LCD display, replace the Lithium CR2 battery.

1. Open Datalogger housing by first removing the four (4) rear Phillips screws.

2. Carefully pry off the rear of the Datalogger.

3. The cylindrical lithium battery is located on the bottom of the PCB board.

4. Replace the battery, observing polarity.

5. Assemble the Datalogger housing.

You, as the end user, are legally bound (Battery ordinance) to return all used batteries and accumulators; disposal in the household garbage is prohibited! You can hand over your used ba

Manualsnet FAQs

If you want to find out how the Extech Flir works, you can view and download the Flir Extech 42260 Datalogger User's Guide on the Manualsnet website.

Yes, we have the User's Guide for Flir Extech as well as other Flir manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User's Guide should include all the details that are needed to use a Flir Extech. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Flir Extech 42260 Datalogger User's Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Flir Extech 42260 Datalogger User's Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Flir Extech 42260 Datalogger User's Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Flir Extech 42260 Datalogger User's Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Flir Extech 42260 Datalogger User's Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.