CPEC306 Closed-Path Eddy-Covariance System

Do not connect or disconnect the EC155 gas analyzer head or the CSAT3A son- ic anemometer head from the EC100 electronics while the EC100 is powered. Doing so can result in unpredictable performance of the system or damage to the instrument head.

Grounding electrical components in the measurement system is critical. Prop- er earth (chassis) grounding will ensure maximum electrostatic discharge (ESD) protection and higher measurement accuracy.

Use care when connecting and disconnecting tube fittings to avoid introduc- ing dust or other contaminants.

Do not overtighten the tube fittings. Consult the manual for information on proper connection.

The CPEC306 power source should be designed thoughtfully to ensure unin- terrupted power. If needed, contact Campbell Scientific for assistance.

Retain all spare caps and plugs as these are required when shipping or storing the CPEC306 system.

IMPORTANT NOTE: This Quick Deploy Guide is meant to be a general reference to give the installer an overview of the steps required to make this system operational. The Owners Manual is the definitive source for detailed installation instructions and information.

1 Caution!!

CSAT3A Sonic Anemometer Head

Using the CR1000KD keypad to configure settings and zero/span on a deployed system

The following tools are required to install the CPEC306 system in the field. Addi- tional tools may be required for a user-supplied tripod or tower.

Attach the horizontal crossarm pole to the desired height on the tripod.

3 Physical Deployment

1. Set up the tripod and crossarm pole.

Secure the tripod to the ground.

QUICK DEPLOY GUIDE

Document Part Number: 34223 Revision Date: September 2018

CM210 crossarm- to-pole bracket

crossarm

When not using datalogger support software such as Loggernet, turn on the +12 Vdc power supply and use the CR1000KD keypad to configure the settings and zero/span. 1. Press Enter to activate the display. Press

Enter again to display the System Control menu.

2. On the System Control menu, select Site Var Settings to customize site specific variables.

3. Enter site-specific variables. Press Esc when complete to return to the main menu.

4. On the System Control menu, select Const Table to modify sensor information.

5. Add and remove sensors by selecting -1 for true and 0 for false. Once the changes are completed, select Apply and Restart at the bottom of the screen. Select Yes to save the changes. The device will then restart.

Site Var Settings: Meas height :2.00000 Pck Surf typ : GRASS Canopy hght :0.50000 d, 0 = auto :0.00000 Z0,0 = auto :0:00000 GPS height :2.00000 Bulk density :1300.00

NMBR_HFP : 4 HFP_SNSTVT_1 : 62.0000 HFP_SNSTVT_2 : 62.0000 HFP_SNSTVT_3 : 62.0000 HFP_SNSTVT_4 : 62.0000 CAL_INTV : 1440 Apply and Restart

2 Required Gear

1. 9/16-in, open-end wrench 2. 1/2-in, open-end wrench 3. 11/16-in, open-end wrench 4. Adjustable wrench 5. Small, flat-tip screwdriver

6. Large, flat-tip screwdriver 7. Sledgehammer (to drive ground-

ing rod into the ground) 8. 3/16-in hex-key wrench

Sample Intake Assembly with EC155 Gas Analyzer

EC100 Electronics

CPEC306 Enclosure

CPEC306 System Diagram

4. Connect the EC155 and CSAT3A to the electronics.

5. Ground the tripod and the enclosures.

6. Connect the system plumbing.

7. Wiring.

Connect the SDM from the main enclosure and EC100 power cables to the EC100 electronics.

Draw the EC100 power cable and EC100 SDM cable through the feedthrough at the bottom of the CPEC306 enclosure and connect them to the DINrail.

8. Datalogger

Insert a MicroSD card into the datalogger and connect power.

Turn on the +12 Vdc power supply and use either LoggerNet, PC200W, or PC400 on a laptop to configure settings and zero/span. If using the CR1000KD keypad, use the instruc- tions on the front page of this quick deploy guide.

1. Connect to the datalogger

2. Connect and select the Const_Table within the EasyFluxTM DL program.

3. Confirm all the sensors used at the site are set to -1. To change a value in this table, right-click on the current value and select View/Modify from the pop-up menu. Press Apply once the new value has been entered. Once all of the sensors and constant settings are correct, scroll to the bottom and set ApplyAndRestart to true.

4. Review the Public table and confirm that site specific variables are set appropriately. Note: Setting these variables does not require an ApplyAndRestart.

2. Setup and mount sensors

Mount the EC155 bypass tube.

Connect the CM250 mount to the crossarm and then connect the sensors mounted on the Mounting Platform, as shown. Use the bubble level on the CSAT3A to level the platform.

3. Mount the enclosures. Mount the enclosures on the legs of the tripod as shown.

EC155/CSAT3A mounting platform

CM250 leveling mount

CSAT3A Sonic Anemometer head

Sample Intake Assembly with EC155 Gas Analyzer

crossarm

EC155 analyzer cable

CSAT3A cable

EC155 sample cell cable

Ground rod

Ground lug

CPEC306 enclosure

Power cable to +12Vdc power supply (Do not apply power until the CPEC306 is fully wired.)

Power cable to EC100

Ground cable to +12Vdc power supply (off)

Gr

Manualsnet FAQs

If you want to find out how the CPEC306 Campbell Scientific works, you can view and download the Campbell Scientific CPEC306 System Quick Deploy Guide on the Manualsnet website.

Yes, we have the Quick Deploy Guide for Campbell Scientific CPEC306 as well as other Campbell Scientific manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Quick Deploy Guide should include all the details that are needed to use a Campbell Scientific CPEC306. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Campbell Scientific CPEC306 System Quick Deploy Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Campbell Scientific CPEC306 System Quick Deploy Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Campbell Scientific CPEC306 System Quick Deploy Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Campbell Scientific CPEC306 System Quick Deploy Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Campbell Scientific CPEC306 System Quick Deploy Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.