Revision: 12/2021 Copyright 2016 2021 Campbell Scientific, Inc.

Table of contents 1. Introduction 1

2. Precautions 1

3. Initial inspection 2

4. QuickStart 2

5. Overview 6

6. Specifications 7 6.1 SGB3 3-line surge protector 7

7. Installation 8 7.1 Siting 8 7.2 Mounting 8 7.3 Wiring 9 7.4 Programming 10

7.4.1 SDI12Recorder() instruction 10 7.4.2 Slow sequence program instructions 10

8. Operation 11 8.1 Sensor measurements 11

8.1.1 SDI-12 addressing 12 8.1.2 Metadata 14

8.2 Long cables 14 8.3 Power conservation 15

8. Maintenance, calibration, and troubleshooting 15

Appendix A. Importing Short Cut code into CRBasic Editor 17

Appendix B. SDI-12 sensor support 18 B.1 SDI-12 command basics 18

B.1.1 Acknowledge active command (a!) 19 B.1.2 Send identification command (al!) 19 B.1.3 Start verification command (aV!) 20 B.1.4 Address query command (?!) 20

Table of Contents - i

B.1.5 Change address command (aAb!) 20 B.1.6 Start measurement commands (aM!) 20 B.1.7 Stopping a measurement command 21 B.1.8 Send data command (aD0! aD9!) 21 B.1.9 Continuous measurement command (aR0! aR9!) 22

B.2 SDI-12 transparent mode 22 B.2.1 Changing an SDI-12 address 22

B.3 References 25

Table of Contents - ii

1. Introduction The CS225 Temperature String uses SDI-12 digital technology for simple integration and reliability. The CS225 consists of an arrangement of temperature sensors mounted in rugged steel reinforced cable. Temperature points are over molded to provide long-term connection in all mediums. Each CS225 is manufactured to the clients specific requirements and includes the SGB3 to provide electrical surge protection.

NOTE: This manual provides information only for CRBasic data loggers. For retired Edlog data logger support, see an older manual at www.campbellsci.com/old-manuals .

2. Precautions l READ AND UNDERSTAND the Safety section at the back of this manual. l Although the CS225 is designed to be a rugged and reliable device for field use, care

should be taken when handling or moving it to avoid damage. l There are no user-serviceable parts and disassembling the device will void the warranty. l The black outer jacket of the cable is Santoprene rubber. This compound was chosen for

its resistance to temperature extremes, moisture, and UV degradation. However, this jacket will support combustion in air. It is rated as slow burning when tested according to U.L. 94 H.B. and will pass FMVSS302. Local fire codes may preclude its use inside buildings.

l CAUTIONDo not make tight bends near the temperature sensor(s). Doing so has the potential to damage the sensor assembly. The minimum bend radius at any sensor location is 15 cm (6 in). See following image.

CS225 Temperature String 1

3. Initial inspection l Upon receipt of the CS225, inspect the packaging and contents for damage. File damage

claims with the shipping company. l The model number and cable length are printed on a label at the connection end of the

cable. Check this information against the shipping documents to ensure the expected product and cable length were received.

4. QuickStart A video that describes data logger programming using Short Cut is available at: www.campbellsci.com/videos/cr1000x-data logger-getting-started-program-part-3 . Short Cut is an easy way to program your data logger to measure the sensor and assign data logger wiring terminals. Short Cut is available as a download on www.campbellsci.com . It is included in installations of LoggerNet, RTDAQ, and PC400.

The following procedure also shows using Short Cut to program the CS225.

1. Open Short Cut and click Create New Program.

2. Double-click the data logger model.

CS225 Temperature String 2

3. In the Available Sensors and Devices box, type CS225 or locate the sensor in the Sensors > Temperature folder. Double-click CS225 SDI-12 Temperature Profiler. Type the correct Starting SDI-12 Address for the first sensor (SDI-12 addressing (p. 12)). Type the Number of Result Sensors.

CS225 Temperature String 3

4. Click the Wiring tab to see how the sensor is to be wired to the data logger. Click OK after wiring the sensor.

5. Repeat steps 3 and 4 for other sensors.

CS225 Temperature String 4

6. In Output Setup, type the scan rate, meaningful table names, and Data Output Storage Interval.

7. Select the measurement and its associated output option.

CS225 Temperature String 5

8. Click Finish and save the program. Send the program to the data logger if the data logger is connected to the computer.

9. If the sensor is connected to the data logger, check the output of the sensor in the data display in LoggerNet, RTDAQ, or PC400 to make sure it is making reasonable measurements.

5. Overview The CS225 Temperature String uses digital sensor technology allowing for a simple three-wire integration. The CS225 consists of an arrangement of over molded temperature points mounted in a rugged steel reinforced cable. Each CS225 is manufactured to the clients specific requirements.

The CS225 is suited to a wide variety of applications and environments that require temperature profiling. The completely sealed cable assembly permits the CS225 to be buried, submerged, or integrated directly into structures. Examples of applications include temperature profiling in boreholes, soils, water, and frost and permafrost monitoring.

The SGB3 provides surge protection for the CS225 Temperature String. The SGB3 case can be mounted to the backplate of a Campbell Scientific enclosure.

Features:

l Accurate and stable measurements l Each sensor is individually addressed and referenced to its depth l Low power consumption l Digital SDI-12 output l Compatible with the following data loggers: CR6, CR3000, CR1000X, CR800 Series,

CR300 Series, CR1000

CS225 Temperature String 6

6. Specifications Operating range: 55 to 85 C

Accuracy

Typical: 0.2 C over 40 to 85 C,

Worst case: 0.4 C over 40 to 85 C; 0.5 C over 55 to 40 C (includes lifetime drift of sensor)

Resolution: 0.0078 C

Output protocol: SDI-12 1.3

Measurement update interval: 1 s (automatic), occurs in quiescent mode

Warm-up time: 10 s

Maximum sensors per probe: 36

Supply voltage: 9 to 28 VDC

Current Consumption

Quiescent: # sensors 1.0 mA (max)

Active (during SDI-12 communications): 20 mA + (# sensors 1.0 mA)

Temperature point diameter: 2.22 cm (0.88 in)

Maximum cable length: 152 m (500 ft)

Minimum sensor spacing: 5 cm (1.97 in)

EU Declaration of Conformity: View at www.campbellsci.com/cs225-l

6.1 SGB3 3-line surge protector Operating range: 55 to 85 C

Maximum voltage: 28 VDC / 20 VAC (L1, L2, L3 with respect to G terminals)

Maximum current: 2 A per terminal, 4 A total (requires both ground terminals for return current)

Maximum rated surge: 1200 A (8/20 s)

CS225 Temperature String 7

7. Installation If you are programming your data logger with Short Cut, skip Wiring (p. 9) and Programming (p. 10). Short Cut does this work for you. See QuickStart (p. 2) for a Short Cut tutorial.

7.1 Siting Install the CS225 directly in the medium that is to be measured. The types of medium that can be measured are varied, including soils, roadbeds, and water. To make the most representative measurement, it is important that consistent contact be made between the temperature profiler and the medium.

The CS225 consists of the lead cable and the sensor array. The lead cable length is the length between the data logger and sensor array. The sensor array length is the length of cable that incorporates all temperature sensors in their specified configuration. The installation position of the string and its measurement points are referenced from the first sensor position at the end of the sensor array. This information is needed for the sensor configuration process.

7.2 Mounting Orient and secure the CS225 in the measurement medium. Keep materials removed during installation and use that material as backfill.

If a burial depth is required for each temperature sensor, you need to know the length of the sensor array, the measurement from end metadata of each sensor in the array, and any offset used during installation of the string. You will also need to calculate the depth of each temperature sensor.

NOTE: The furthest temperature point from the data logger (or the temperature point at the end of the string) is addressed starting at 1. Each temperature point going up the string towards the data logger changes by one address number/letter.

Orient the lead cable of the CS225 towards the data logger to avoid loops or strain on the cable. Also use a suitable trench or conduit to protect the signal and power cable from damage.

CS225 Temperature String 8

The SGB3 is required to protect against electrical surges (FIGURE 7-1 (p. 9)). The CS225 is shipped with the SGB3, a 2-ft cable for connecting the SGB3 to the data logger, and two pan Phillips screws and two grommets for mounting the SGB3 to an enclosure backplate.

FIGURE 7-1. SGB3 3-line surge protector

7.3 Wiring Table 7-1 (p. 9) provides the connections for the SGB3 surge protector, CS225, and Campbell Scientific data loggers.

Table 7-1: Wire color, function, and connections to the SGB3 and data logger

CS225 wire color Function SGB3 connection terminal

Data logger connection terminal1

Red Power L1 12V

Green SDI-12 Signal L2 C or U2 terminal configured for SDI-12

Black Power Ground G G

Clear Shield G G 1 The SGB3 connects to the data logger using the 2 ft cable shipped with the CS225.

2 U terminals are automatically configured by the measurement instruction.

If multiple SDI-12 sensors are connected to a data logger, Campbell Scientific recommends using separate terminals when possible. For the CR6 and CR1000X, triggering conflicts may occur when a companion terminal is used for a triggering instruction such as TimerInput(), PulseCount(), or WaitDigTrig(). For example, if the CS225 is connected to C3 on a CR1000X, C4 cannot be used in the TimerInput(), PulseCount(), or WaitDigTrig() instructions.

CS225 Temperature String 9

7.4 Programming Short Cut is the best source for up-to-date programming code for Campbell Scientific data loggers. If your data acquisition requirements are simple, you can probably create and maintain a data logger program exclusively with Short Cut. If your data acquisition needs are more complex, the files that Short Cut creates are a great source for programming code to start a new program or add to an existing custom program.

NOTE: Short Cut cannot edit programs after they are imported and edited in CRBasic Editor.

A Short Cut tutorial is available in QuickStart (p. 2). If you wish to import Short Cut code into CRBasic Editor to create or add to a customized program, follow the procedure in Importing Short Cut code into CRBasic Editor (p. 17). Programming basics for CRBasic data loggers are provided in the following section. Downloadable example programs are available at www.campbellsci.com/downloads/cs225-example-programs .

7.4.1 SDI12Recorder() instruction The SDI12Recorder() instruction is used to measure the temperature sensors in the CS225. This instruction sends a request to the sensor to make a measurement and then retrieves the measurement from the sensor. See Sensor measurements (p. 11) for more information.

For most data loggers, the SDI12Recorder() instruction has the following syntax:

SDI12Recorder(Destination, SDIPort, SDIAddress, SDICommand, Multiplier, Offset, FillNAN, WaitonTimeout)

Each temperature sensor in the CS225 needs a unique SDI-12 address. For the SDIAddress, alphabetical characters need to be enclosed in quotes (for example, A). Also enclose the SDICommand in quotes as shown. The Destination parameter must be an array. The required number of values in the array depends on the command (Table 8-1 (p. 11)).

FillNAN and WaitonTimeout are optional parameters (refer to CRBasic Help for more information).

7.4.2 Slow sequence program instructions Use the slow sequence program instructions when the CS225 measurements will exceed the program scan interval of the additional instruments included in the station. For example, if a CS225 consists of 17 or more temperature sensors, the time required to poll all sensors and receive data can be greater than 5 seconds based on the 300 ms execution time for the aR0! command. For more details on the use of the slow sequence program instructions, reference the

CS225 Temperature String 10

related LoggerNet Help, data logger manual, or the slow sequence program at www.campbellsci.com/downloads/cs225-example-programs .

8. Operation 8.1 Sensor measurements The CS225 responds to the SDI-12 commands shown in Table 8-1 (p. 11). When power is supplied to the CS225, the internal electronics continuously measure temperature at a rate of approximately once per second. Every output measurement (aR0! or aM0!) obtained from the sensor is a running average of 10 consecutive readings. The accuracy specification is based on an average of 10 consecutive readings. Therefore, after initial power up, a delay of 10 s is recommended to obtain the best accuracy.

As the sensor is obtaining a measurement every second, Campbell Scientific recommends using the continuous measurement command (aR0!) to obtain the temperature readings. Using the aR0! commands reduces the time taken in comparison to the aM0! to obtain a reading via the SDI-12 protocol.

Outputs of both lifetime and user resettable minimum and maximum temperatures are also available during powered operation from each temperature point in the CS225. The user- resettable minimum and maximum temperatures can be used to monitor specific seasons or periods of measure, without having to review the entire data set. The lifetime minimum and maximum temperatures are used for maintenance and warranty records. The lifetime and user- resettable minimum and maximum temperature values are single 1-second readings.

Table 8-1: SDI-12 commands

SDI-12 command Variable name Description

aR0! Temperature value Temperature floating point (C)

aR1! Serial number, location number, depth

value (in cm) Serial number, location number, depth value (in cm)

aR2! Read user resettable min temperature Min. temperature floating point (C)

aR3! Read user resettable max temperature Max. temperature floating point (C)

aR4! Read lifetime min temperature Min. temperature floating point (C)

CS225 Temperature String 11

Table 8-1: SDI-12 commands

SDI-12 command Variable name Description

aR5! Read lifetime max temperature Max. temperature floating point (C)

aR6! Read and reset user resettable min

temperature

Min. temperature floating point (C). This value constitutes the minimum of all 1-second measurements taken since the previous aR6! command.

aR7! Read and reset user resettable max

temperature

Max. temperature floating point (C). This value constitutes the maximum of all 1-second measurements taken since the previous aR7! command.

aV! Verification command S1 = BootRom Signature

S2 = Firmware Signature

aAb! Change Address command

Valid addresses in sequence are:

19 / AZ / az (no Address 0)

Sending a broadcast message with the address change { can correct units that have conflicting addresses.

aI! SDI-12 Identification command X13CAMPBELLCS225 1.0 SN:XXXXX

8.1.1 SDI-12 addressing Each temperature sensor has a different default SDI-12 address. The starting address is 1 and coincides with the bottom temperature point. The last address coincides with the top sensor, which is closest to the data logger.

If multiple SDI-12 sensors are connected to the data logger, Campbell Scientific recommends using separate terminals when possible. However, multiple SDI-12 sensors or multiple CS225 sensors can connect to the same data logger control or U terminal if they have different SDI-12 addresses. If changing the SDI-12 addresses, Campbell Scientific recommends starting the readdressing process with the largest temperature sensor address to avoid duplicate addresses. Table 8-2 (p. 13) provides the SDI-12 addresses and positions.

CS225 Temperature String 12

NOTE: Use the default SDI-12 addresses when possible. Inadvertently giving multiple sensors the same SDI-12 address will prevent the sensors from communicating. Use the aA{! command, where a is the affected address to reset the affected sensors to their factory configured address value.

Table 8-2: SDI-12 addresses and positions

Numeric set Uppercase set Lowercase set

1 / 1 A / 10 a / 36

2 / 2 B / 11 b / 37

3 / 3 C / 12 c / 38

4 / 4 D / 13 d / 39

5 / 5 E / 14 e / 40

6 / 6 F / 15 f / 41

7 / 7 G / 16 g / 42

8 / 8 H / 17 h / 43

9 / 9 I / 18 i / 44

J / 19 j / 45

K / 20 k / 46

L / 21 l / 47

M / 22 m / 48

N / 23 n / 49

O / 24 o / 50

P / 25 p / 51

Q / 26 q / 52

R / 27 r / 53

S / 28 s / 54

T / 29 t / 55

U / 30 u / 56

V / 31 v / 57

CS225 Temperature String 13

Table 8-2: SDI-12 addresses and positions

Numeric set Uppercase set Lowercase set

W / 32 w / 58

X / 33 x / 59

Y / 34 y / 60

Z / 35 a / 61

{ reset to factory address

8.1.2 Metadata Every temperature point in a probe includes the following metadata, which can be retrieved using the aR1! SDI-12 command. This information can be used to identify details of the temperature probe and its individual temperature points.

Table 8-3: Metadata details

Name Value range Description

Serial Number 0 to 65534 The serial number that is unique to each probe unit.

Location Number 1 to 255

Each temperature sensor within a probe is assigned its own unique location number which by default is in relation to the SDI-12 address (see Table 8-1 (p. 11)).

Depth Value 0 to 65535 cm

The distance values in centimeters (cm). These are intended to reflect the profile distances. Typically, the top most temperature point would be designated as 0 cm and if the next temperature point below was 10 cm away, then its distance value would be 10 cm. Users may designate other starting values. These can be configured at the time of ordering.

8.2 Long cables As the measurement data is transferred between the temperature profiler and data logger digitally, there are no offset errors incurred with increasing cable length as seen with analog sensors. However, with long enough cable lengths, the digital communications will break down, resulting in either no response from the sensor or corrupted readings. The original SDI-12 standard specifies the maximum total cable length for the cable as being 61 m (200 ft), but we are able to exceed this limit by:

CS225 Temperature String 14

l Using low capacitance, low resistance, screened cable l Ensuring that the power ground cable has low resistance and is connected to the same

ground reference as the data logger control or U terminals

8.3 Power conservation The CS225 draws less than 1 mA of current per sensor between polling sessions from its 12 V supply. In many applications, this is minimal compared to overall system power use, so the sensor can be permanently powered to avoid the warm-up period.

In low power applications, you can switch the power on a minimum of 10 s (allowing for the warm-up period) before polling the CS225. This switching can be achieved in different ways depending on the type and model of your data logger. If available, the switched 12 V output of the data logger can be used.

8. Maintenance, calibration, and troubleshooting The CS225 probe requires no maintenance nor calibration. The following are symptoms and possible corrections.

Symptom: 9999 or NAN for temperature

1. Verify the green wire is connected to the control terminal specified by the SDI-12 measurement instruction.

2. Verify the red power wire is connected to a 12V terminal; check the voltage with a digital volt meter. If a switched 12V terminal is used, temporarily connect the red wire to a 12V terminal (non-switched) for test purposes.

Symptom: Sensor wont respond to command

1. Expected address not used or has been changed.

a. Confirm all addresses in use with the aI! command to determine the individual addresses of each temperature sensor.

CS225 Temperature String 15

2. Expected sensor address has been to match another sensor address already in use.

a. When readdressing temperature sensors, you must avoid giving multiple sensors the same address. If this does occur, you will no longer be able to communicate with these sensors. Use the aA{! command, where a is the affected address. This will reset the affected sensors to their factory configured address value. For example, if two sensors have address 2, sending 2A{! will reset both sensors to the original addresses.

CS225 Temperature String 16

Appendix A. Importing Short Cut code into CRBasic Editor Short Cut creates a .DEF file that contains wiring information and a program file that can be imported into the CRBasic Editor. By default, these files reside in the C:\campbellsci\SCWin folder.

Import Short Cut program file and wiring information into CRBasic Editor:

1. Create the Short Cut program. After saving the Short Cut program, click the Advanced tab then the CRBasic Editor button. A program file with a generic name will open in CRBasic. Provide a meaningful name and save the CRBasic program. This program can now be edited for additional refinement.

NOTE: Once the file is edited with CRBasic Editor, Short Cut can no longer be used to edit the program it created.

2. To add the Short Cut wiring information into the new CRBasic program, open the .DEF file located in the C:\campbellsci\SCWin folder, and copy the wiring information, which is at the beginning of the .DEF file.

3. Go into the CRBasic program and paste the wiring information into it.

4. In the CRBasic program, highlight the wiring information, right-click, and select Comment Block. This adds an apostrophe (') to the beginning of each of the highlighted lines, which instructs the data logger compiler to ignore those lines when compiling. The Comment Block feature is demonstrated at about 5:10 in the CRBasic | Features video .

CS225 Temperature String 17

Appendix B. SDI-12 sensor support SDI-12, Serial Data Interface at 1200 baud, is a protocol developed to simplify sensor and data logger compatibility. Only three wires are necessary serial data, ground, and 12 V. With unique addresses, multiple SDI-12 sensors can connect to a single SDI-12 terminal on a Campbell Scientific data logger.

This appendix discusses the structure of SDI-12 commands and the process of querying SDI-12 sensors. For more detailed information, refer to version 1.4 of the SDI-12 protocol, available at www.sdi-12.org .

For additional information, refer to the SDI-12 Sensors | Transparent Mode and SDI-12 Sensors | Watch or Sniffer Mode videos.

B.1 SDI-12 command basics SDI-12 commands have three components:

l Sensor address (a) a single character and the first character of the command. Use the default address of zero (0) unless multiple sensors are connected to the same port.

l Command body an upper case letter (the command), optionally followed by one or more alphanumeric qualifiers.

l Command termination (!) an exclamation mark.

An active sensor responds to each command. Responses have several standard forms and always terminate with (carriage return and line feed). Standard SDI-12 commands are listed in Table B-1 (p. 18).

Table B-1: Campbell Scientific sensor SDI-12 command and response set

Name Command Response1

Acknowledge Active a! a

Send Identification aI! allccccccccmmmmmmvvvxxx...xx

Start Verification aV! atttn

CS225 Temperature String 18

Table B-1: Campbell Scientific sensor SDI-12 command and response set

Name Command Response1

Address Query ?! a

Change Address aAb! b

Start Measurement aM! aM1!...aM9!

atttn

Send Data aD0!...aD9! a or

a 1 Information on each of these commands is given in the following sections.

B.1.1 Acknowledge active command (a!) The Acknowledge Active command (a!) is used to test a sensor on the SDI-12 bus. An active sensor responds with its address.

B.1.2 Send identification command (al!) Sensor identifiers are requested by issuing command aI!. The reply is defined by the sensor manufacturer but usually includes the sensor address, SDI-12 version, manufacturers name, and sensor model information. Serial number or other sensor specific information may also be included.

aI! allccccccccmmmmmmvvvxxx...xx

a Sensor SDI-12 address

ll SDI-12 version number (indicates compatibility)

cccccccc 8-character vendor identification

mmmmmm 6 characters specifying the sensor model

vvv 3 characters specifying the sensor version (operating system)

xxxxx Up to 13 optional characters used for a serial number or other specific sensor information that is not relevant for operation of the data logger

Terminates the response Source: SDI-12: A Serial-Digital Interface Standard for Microprocessor-Based Sensors (see References).

CS225 Temperature String 19

B.1.3 Start verification command (aV!) The response to a Start Verification command can include hardware diagnostics, but like the aI! command, the response is not standardized.

Command: aV!

Response: atttn

a = sensor address

ttt = time, in seconds, until verification information is available

n = the number of values to be returned when one or more subsequent D! commands are issued

B.1.4 Address query command (?!) Command ?! requests the address of the connected sensor. The sensor replies to the query with the address, a. This command should only be used with one sensor on the SDI-12 bus at a time.

B.1.5 Change address command (aAb!) Multiple SDI-12 sensors can connect to a single SDI-12 terminal on a data logger. Each device on a single terminal must have a unique address.

A sensor address is changed with command aAb!, where a is the current address and b is the new address. For example, to change an address from 0 to 2, the command is 0A2!. The sensor responds with the new address b, which in this case is 2.

NOTE: Only one sensor should be connected to a particular terminal at a time when changing addresses.

B.1.6 Start measurement commands (aM!) A measurement is initiated with the M! command. The response to each command has the form atttn , where

a = sensor address

ttt = time, in seconds, until measurement data is available. When the data is ready, the sensor notifies the data logger, and the data logger begins issuing D commands.

n = the number of values returned when one or more subsequent D commands are issued. For the aM! command, n is an integer from 0 to 9.

CS225 Temperature String 20

When the aM! is issued, the data logger pauses its operation and waits until either it receives the data from the sensor or the time, ttt, expires. Depending on the scan interval of the data logger program and the response time of the sensor, this may cause skipped scans to occur. In this case make sure your scan interval is longer than the longest measurement time (ttt).

Table B-2: Example aM! sequence

0M! The data logger makes a request to sensor 0 to start a measurement.

00352 Sensor 0 immediately indicates that it will return two values within the next 35 seconds.

0 Within 35 seconds, sensor 0 indicates that it has completed the measurement by sending a service request to the data logger.

0D0! The data logger immediately issues the first D command to collect data from the sensor.

0+.859+3.54 The sensor immediately responds with the sensor address and the two values.

B.1.7 Stopping a measurement command A measurement command (M!) is stopped if it detects a break signal before the measurement is complete. A break signal is sent by the data logger before most commands.

A concurrent measurement command (C!) is aborted when another valid command is sent to the sensor before the measurement time has elapsed.

B.1.8 Send data command (aD0! aD9!) The Send Data command requests data from the sensor. It is issued automatically with every type of measurement command (aM!, aMC!, aC!, aCC!). When the measurement command is aM! or aMC!, the data logger issues the aD0! command once a service request has been received from the sensor or the reported time has expired. When the data logger is issuing concurrent commands (aC! or aCC!), the Send Data command is issued after the required time has elapsed (no service request will be sent by the sensor). In transparent mode (see SDI-12 transparent mode (p. 22) ), the user asserts this command to obtain data.

Depending on the type of data returned and the number of values a sensor returns, the data logger may need to issue aD0! up to aD9! to retrieve all data. A sensor may return up to 35 characters of data in response to a D command that follows an M! or MC! command. A sensor may return up to 75 characters of data in response to a D command that follows a C! or CC! command. Data values are separated by plus or minus signs.

Command: aD0! (aD1! aD9!)

CS225 Temperature String 21

Response: a or a

where:

a = the sensor address

= values returned with a polarity sign (+ or )

= terminates the response

= 16-bit CRC code appended if data was requested with aMC! or aCC!.

B.1.9 Continuous measurement command (aR0! aR9!) Sensors that are able to continuously monitor the phenomena to be measured can be read directly with the R commands (R0! R9!). The response to the R commands mirrors the Send Data command (aD0!). A maximum of 75 characters can be returned in the part of the response to the R command.

B.2 SDI-12 transparent mode System operators can manually interrogate and enter settings in probes using transparent mode. Transparent mode is useful in troubleshooting SDI-12 systems because it allows direct communication with probes. Data logger security may need to be unlocked before activating the transparent mode.

Transparent mode is entered while the computer is communicating with the data logger through a terminal emulator program. It is accessed through Campbell Scientific data logger support software or other terminal emulator programs. Data logger keyboards and displays cannot be used.

The terminal emulator is accessed by navigating to the Tools list in PC400 or the Datalogger list in the Connect screen of LoggerNet.

Watch videos/sdi12-sensors-transparent-mode from our website.

Data loggers from other manufacturers will also have a transparent mode. Refer to those manuals on how to use their transparent mode.

The following examples show how to enter transparent mode and change the SDI-12 address of an SDI-12 sensor. The steps shown in Changing an SDI-12 address (p. 22) are used with most Campbell Scientific data loggers.

B.2.1 Changing an SDI-12 address This example was done with a CR1000X, but the steps are only slightly different for CR6, CR3000, CR800-series, CR300-series, CR1000 data loggers.

CS225 Temperature String 22

1. Connect an SDI-12 sensor to the CR1000X.

2. Open Device Configuration utility.

3. Under Device Type, type the data logger model and double-click on the model type. This example uses a CR1000X directly connected to the computer USB port.

4. Select the correct Communication Port and click Connect.

5. Click the Terminal tab.

6. Select All Caps Mode.

CS225 Temperature String 23

7. Press Enter until the data logger responds with the data logger (CR1000X>) prompt.

8. Type SDI12 and press Enter.

9. At the Select SDI12 Port prompt, type the number corresponding to the control port where the sensor is connected and press Enter. In this example the sensor is connected to C3. The response Entering SDI12 Terminal indicates that the sensor is ready to accept SDI-12 commands.

10. To query the sensor for its current SDI-12 address, type ?! and press Enter. The sensor responds with its SDI-12 address. If no characters are typed within 60 seconds, the mode is exited. In that case, simply type SDI12 again, press Enter, and type the correct control port number when prompted.

11. To change the SDI-12 address, type aAb!, where a is the current address from the previous step and b is the new address. Press Enter. The sensor changes its address and responds with the new address. In the following example, the sensor address is changed from 0 to B.

12. To exit SDI-12 transparent mode, click Close Terminal.

NOTE: The transparent mode for the CR6, CR3000, CR800-series, CR300-series, and CR1000 data loggers is similar to that shown for the CR1000X.

CS225 Temperature String 24

B.3 References SDI-12 Support Group. SDI-12: A Serial-Digital Interface Standard for Microprocessor-Based Sensors Version 1.4. River Heights, UT: SDI-12 Support Group, 2017. https://sdi- 12.org/specification .

CS225 Temperature String 25

Limited warranty Products manufactured by Campbell Scientific are warranted by Campbell Scientific to be free from defects in materials and workmanship under normal use and service for twelve months from the date of shipment unless otherwise specified on the corresponding product webpage. See Product Details on the Ordering Information pages at www.campbellsci.com . Other manufacturer's products, that are resold by Campbell Scientific, are warranted only to the limits extended by the original manufacturer.

Refer to www.campbellsci.com/terms#warranty for more information.

CAMPBELL SCIENTIFIC EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Campbell Scientific hereby disclaims, to the fullest extent allowed by applicable law, any and all warranties and conditions with respect to the Products, whether express, implied or statutory, other than those expressly provided herein.

Assistance Products may not be returned without prior authorization.

Products shipped to Campbell Scientific require a Returned Materials Authorization (RMA) or Repair Reference number and must be clean and uncontaminated by harmful substances, such as hazardous materials, chemicals, insects, and pests. Please complete the required forms prior to shipping equipment.

Campbell Scientific regional offices handle repairs for customers within their territories. Please see the back page for the Global Sales and Support Network or visit www.campbellsci.com/contact to determine which Campbell Scientific office serves your country.

To obtain a Returned Materials Authorization or Repair Reference number, contact your CAMPBELL SCIENTIFIC regional office. Please write the issued number clearly on the outside of the shipping container and ship as directed.

For all returns, the customer must provide a Statement of Product Cleanliness and Decontamination or Declaration of Hazardous Material and Decontamination form and comply with the requirements specified in it. The form is available from your CAMPBELL SCIENTIFIC regional office. Campbell Scientific is unable to process any returns until we receive this statement. If the statement is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customers expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.

Safety DANGER MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC. FAILURE TO PROPERLY AND COMPLETELY ASSEMBLE, INSTALL, OPERATE, USE, AND MAINTAIN TRIPODS, TOWERS, AND ATTACHMENTS, AND FAILURE TO HEED WARNINGS, INCREASES THE RISK OF DEATH, ACCIDENT, SERIOUS INJURY, PROPERTY DAMAGE, AND PRODUCT FAILURE. TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS. CHECK WITH YOUR ORGANIZATION'S SAFETY COORDINATOR (OR POLICY) FOR PROCEDURES AND REQUIRED PROTECTIVE EQUIPMENT PRIOR TO PERFORMING ANY WORK. Use tripods, towers, and attachments to tripods and towers only for purposes for which they are designed. Do not exceed design limits. Be familiar and comply with all instructions provided in product manuals. Manuals are available at www.campbellsci.com. You are responsible for conformance with governing codes and regulations, including safety regulations, and the integrity and location of structures or land to which towers, tripods, and any attachments are attached. Installation sites should be evaluated and approved by a qualified engineer. If questions or concerns arise regarding installation, use, or maintenance of tripods, towers, attachments, or electrical connections, consult with a licensed and qualified engineer or electrician. General

l Protect from over-voltage. l Protect electrical equipment from water. l Protect from electrostatic discharge (ESD). l Protect from lightning. l Prior to performing site or installation work, obtain required approvals and permits. Comply with all governing structure-height regulations. l Use only qualified personnel for installation, use, and maintenance of tripods and towers, and any attachments to tripods and towers. The

use of licensed and qualified contractors is highly recommended. l Read all applicable instructions carefully and understand procedures thoroughly before beginning work. l Wear a hardhat and eye protection, and take other appropriate safety precautions while working on or around tripods and towers. l Do not climb tripods or towers at any time, and prohibit climbing by other persons. Take reasonable precautions to secure tripod and tower

sites from trespassers. l Use only manufacturer recommended parts, materials, and tools.

Utility and Electrical l You can be killed or sustain serious bodily injury if the tripod, tower, or attachments you are installing, constructing, using, or maintaining, or

a tool, stake, or anchor, come in contact with overhead or underground utility lines. l Maintain a distance of at least one-and-one-half times structure height, 6 meters (20 feet), or the distance required by applicable law,

whichever is greater, between overhead utility lines and the structure (tripod, tower, attachments, or tools). l Prior to performing site or installation work, inform all utility companies and have all underground utilities marked. l Comply with all electrical codes. Electrical equipment and related grounding devices should be installed by a licensed and qualified

electrician. l Only use power sources approved for use in the country of installation to power Campbell Scientific devices.

Elevated Work and Weather l Exercise extreme caution when performing elevated work. l Use appropriate equipment and safety practices. l During installation and maintenance, keep tower and tripod sites clear of un-trained or non-essential personnel. Take precautions to prevent

elevated tools and objects from dropping. l Do not perform any work in inclement weather, including wind, rain, snow, lightning, etc.

Maintenance l Periodically (at least yearly) check for wear and damage, including corrosion, stress cracks, frayed cables, loose cable clamps, cable tightness,

etc. and take necessary corrective actions. l Periodically (at least yearly) check electrical ground connections.

Internal Battery l Be aware of fire, explosion, and severe-burn hazards. l Misuse or improper installation of the internal lithium battery can cause severe injury. l Do not recharge, disassemble, heat above 100 C (212 F), solder directly to the cell, incinerate, or expose contents to water. Dispose of spent

batteries properly. WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL SCIENTIFIC PRODUCTS, THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER INSTALLATION, USE, OR MAINTENANCE OF TRIPODS, TOWERS, OR ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC.

Australia Location: Phone: Email: Website:

Garbutt, QLD Australia 61.7.4401.7700 info@campbellsci.com.au www.campbellsci.com.au

Brazil Location: Phone: Email: Website:

So Paulo, SP Brazil 11.3732.3399 vendas@campbellsci.com.br www.campbellsci.com.br

Canada Location: Phone: Email: Website:

Edmonton, AB Canada 780.454.2505 dataloggers@campbellsci.ca www.campbellsci.ca

China Location: Phone: Email: Website:

Beijing, P. R. China 86.10.6561.0080 info@campbellsci.com.cn www.campbellsci.com.cn

Costa Rica Location: Phone: Email: Website:

San Pedro, Costa Rica 506.2280.1564 info@campbellsci.cc www.campbellsci.cc

France Location: Phone: Email: Website:

Vincennes, France 0033.0.1.56.45.15.20 info@campbellsci.fr www.campbellsci.fr

Germany Location: Phone: Email: Website:

Bremen, Germany 49.0.421.460974.0 info@campbellsci.de www.campbellsci.de

India Location: Phone: Email: Website:

New Delhi, DL India 91.11.46500481.482 info@campbellsci.in www.campbellsci.in

South Africa Location: Phone: Email: Website:

Stellenbosch, South Africa 27.21.8809960 sales@campbellsci.co.za www.campbellsci.co.za

Spain Location: Phone: Email: Website:

Barcelona, Spain 34.93.2323938 info@campbellsci.es www.campbellsci.es

Thailand Location: Phone: Email: Website:

Bangkok, Thailand 66.2.719.3399 info@campbellsci.asia www.campbellsci.asia

UK Location: Phone: Email: Website:

Shepshed, Loughborough, UK 44.0.1509.601141 sales@campbellsci.co.uk www.campbellsci.co.uk

US

Manualsnet FAQs

If you want to find out how the CS225 Campbell Scientific works, you can view and download the Campbell Scientific CS225 Sensor Product Manual on the Manualsnet website.

Yes, we have the Product Manual for Campbell Scientific CS225 as well as other Campbell Scientific manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Product Manual should include all the details that are needed to use a Campbell Scientific CS225. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Campbell Scientific CS225 Sensor Product Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Campbell Scientific CS225 Sensor Product Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Campbell Scientific CS225 Sensor Product Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Campbell Scientific CS225 Sensor Product Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Campbell Scientific CS225 Sensor Product Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.