Model 092 Barometric Pressure Sensor

Revision: 10/16

C o p y r i g h t 2 0 1 2 - 2 0 1 6 C a m p b e l l S c i e n t i f i c , I n c .

Limited Warranty Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service for twelve months from the date of shipment unless otherwise specified in the corresponding product manual. (Product manuals are available for review online at www.campbellsci.com.) Products not manufactured by CSI, but that are resold by CSI, are warranted only to the limits extended by the original manufacturer. Batteries, fine-wire thermocouples, desiccant, and other consumables have no warranty. CSIs obligation under this warranty is limited to repairing or replacing (at CSIs option) defective Products, which shall be the sole and exclusive remedy under this warranty. The Customer assumes all costs of removing, reinstalling, and shipping defective Products to CSI. CSI will return such Products by surface carrier prepaid within the continental United States of America. To all other locations, CSI will return such Products best way CIP (port of entry) per Incoterms 2010. This warranty shall not apply to any Products which have been subjected to modification, misuse, neglect, improper service, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied. The warranty for installation services performed by CSI such as programming to customer specifications, electrical connections to Products manufactured by CSI, and Product specific training, is part of CSI's product warranty. CSI EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CSI hereby disclaims, to the fullest extent allowed by applicable law, any and all warranties and conditions with respect to the Products, whether express, implied or statutory, other than those expressly provided herein.

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com to determine which Campbell Scientific company serves your country.

To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 227-9000. Please write the issued RMA number clearly on the outside of the shipping container. Campbell Scientifics shipping address is:

CAMPBELL SCIENTIFIC, INC. RMA#_____ 815 West 1800 North Logan, Utah 84321-1784

For all returns, the customer must fill out a Statement of Product Cleanliness and Decontamination form and comply with the requirements specified in it. The form is available from our website at www.campbellsci.com/repair. A completed form must be either emailed to repair@campbellsci.com or faxed to (435) 227-9106. Campbell Scientific is unable to process any returns until we receive this form. If the form is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customers expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.

Safety DANGER MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC. FAILURE TO PROPERLY AND COMPLETELY ASSEMBLE, INSTALL, OPERATE, USE, AND MAINTAIN TRIPODS, TOWERS, AND ATTACHMENTS, AND FAILURE TO HEED WARNINGS, INCREASES THE RISK OF DEATH, ACCIDENT, SERIOUS INJURY, PROPERTY DAMAGE, AND PRODUCT FAILURE. TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS. CHECK WITH YOUR ORGANIZATION'S SAFETY COORDINATOR (OR POLICY) FOR PROCEDURES AND REQUIRED PROTECTIVE EQUIPMENT PRIOR TO PERFORMING ANY WORK.

Use tripods, towers, and attachments to tripods and towers only for purposes for which they are designed. Do not exceed design limits. Be familiar and comply with all instructions provided in product manuals. Manuals are available at www.campbellsci.com or by telephoning (435) 227-9000 (USA). You are responsible for conformance with governing codes and regulations, including safety regulations, and the integrity and location of structures or land to which towers, tripods, and any attachments are attached. Installation sites should be evaluated and approved by a qualified engineer. If questions or concerns arise regarding installation, use, or maintenance of tripods, towers, attachments, or electrical connections, consult with a licensed and qualified engineer or electrician.

General Prior to performing site or installation work, obtain required approvals and permits. Comply

with all governing structure-height regulations, such as those of the FAA in the USA. Use only qualified personnel for installation, use, and maintenance of tripods and towers, and

any attachments to tripods and towers. The use of licensed and qualified contractors is highly recommended.

Read all applicable instructions carefully and understand procedures thoroughly before beginning work.

Wear a hardhat and eye protection, and take other appropriate safety precautions while working on or around tripods and towers.

Do not climb tripods or towers at any time, and prohibit climbing by other persons. Take reasonable precautions to secure tripod and tower sites from trespassers.

Use only manufacturer recommended parts, materials, and tools.

Utility and Electrical You can be killed or sustain serious bodily injury if the tripod, tower, or attachments you are

installing, constructing, using, or maintaining, or a tool, stake, or anchor, come in contact with overhead or underground utility lines.

Maintain a distance of at least one-and-one-half times structure height, 20 feet, or the distance required by applicable law, whichever is greater, between overhead utility lines and the structure (tripod, tower, attachments, or tools).

Prior to performing site or installation work, inform all utility companies and have all underground utilities marked.

Comply with all electrical codes. Electrical equipment and related grounding devices should be installed by a licensed and qualified electrician.

Elevated Work and Weather Exercise extreme caution when performing elevated work. Use appropriate equipment and safety practices. During installation and maintenance, keep tower and tripod sites clear of un-trained or non-

essential personnel. Take precautions to prevent elevated tools and objects from dropping. Do not perform any work in inclement weather, including wind, rain, snow, lightning, etc.

Maintenance Periodically (at least yearly) check for wear and damage, including corrosion, stress cracks,

frayed cables, loose cable clamps, cable tightness, etc. and take necessary corrective actions. Periodically (at least yearly) check electrical ground connections.

WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL SCIENTIFIC PRODUCTS, THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER INSTALLATION, USE, OR MAINTENANCE OF TRIPODS, TOWERS, OR ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC.

i

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to specific sections.

1. Introduction ................................................................ 1

2. Precautions ................................................................ 1

3. Initial Inspection ......................................................... 1

4. QuickStart ................................................................... 1

5. Overview ..................................................................... 4

6. Specifications ............................................................. 5

7. Installation .................................................................. 5

7.1 Mounting .............................................................................................. 5 7.2 Datalogger Wiring................................................................................ 6 7.3 Programming ........................................................................................ 7

7.3.1 CRBasic Instruction ...................................................................... 7

8. Operations .................................................................. 8

8.1 Conversion Factors .............................................................................. 8 8.2 Correcting Pressure to Sea Level ......................................................... 8

9. Maintenance ............................................................... 9

9.1 Cleaning ............................................................................................... 9 9.2 Calibration............................................................................................ 9

Appendices

A. Importing Short Cut Code Into CRBasic Editor ... A-1

B. Example Program ................................................... B-1

C. Analog Output Configurations .............................. C-1

D. Internal Cable Connections ................................... D-1

Table of Contents

ii

E. CE Certification ...................................................... E-1

Figures 7-1. 092 installed on a vertical mast ........................................................... 6 D-1. Analog wiring schematic ................................................................. D-1

Tables 7-1. Wire Color, Wire Function, and Datalogger Connection .................... 7 8-1. Conversion Factors for Alternative Pressure Units ............................. 8 C-1. Dip Switches SW1: Analog Output Voltage Settings ..................... C-1 C-2. Dip Switches SW2: Analog Output Pressure Range Settings ......... C-2

CRBasic Example B-1. CR1000 Program Measuring the 092 .............................................. B-1

1

Model 092 Barometric Pressure Sensor

1. Introduction The 092 Barometric Pressure Sensor measures ambient atmospheric pressure and is configured at Campbell Scientific for analog output. It is recommended for wind-farm power-performance research.

This manual provides information only for CRBasic dataloggers. It is also compatible with our retired Edlog dataloggers. For Edlog datalogger support, see an older manual at www.campbellsci.com/old-manuals.

2. Precautions READ AND UNDERSTAND the Safety section at the front of this

manual.

Install the sensor such that it faces north in the northern hemisphere, or south in the southern hemisphere. This orientation will shield the polycarbonate sensor housing from most direct sunlight.

The pressure sensor element is light sensitive. Only operate the 092 when the polycarbonate case is closed.

Pressure conditions in the range of 10 to 10,000 millibars will not harm the sensor, but the measurement accuracy is not guaranteed beyond the range of 600 to 1100 millibars.

3. Initial Inspection Upon receipt of the 092, inspect the packaging and contents for damage.

File damage claims with the shipping company, but contact Campbell Scientific to arrange for repair or replacement.

The model number and cable length are printed on a label at the connection end of the cable. Check this information against the shipping documents to ensure the expected product and cable length are received.

4. QuickStart Short Cut is an easy way to program your datalogger to measure the 092 and assign datalogger wiring terminals. Short Cut is available as a download on www.campbellsci.com and the ResourceDVD. It is included in installations of LoggerNet, PC200W, PC400, or RTDAQ.

Use the following procedure to get started.

NOTE

Model 092 Barometric Pressure Sensor

2

1. Open Short Cut. Click New Program.

2. Select Datalogger Model and Scan Interval (default of 5 seconds is OK for most applications). Click Next.

Model 092 Barometric Pressure Sensor

3

3. Under the Available Sensors and Devices list, select Sensors | Meteorological | Barometric Pressure folder. Select 092 Barometric

Pressure Sensor. Click to move the selection to the Selected device window. Enter the Sea Level Elevation Correction. The default units for the sea level elevation correction is meters; this can be changed by clicking on the Elevation Correction Units box and selecting Feet. Defaults for the barometric pressure measurement is mmHg. This can be changed by clicking the Barometric Pressure box and selecting different units.

4. After selecting the sensor, click Wiring Diagram to see how the sensor is to be wired to the datalogger. The wiring diagram can be printed now or after more sensors are added.

Model 092 Barometric Pressure Sensor

4

5. Select any other sensors you have, then finish the remaining Short Cut steps to complete the program. The remaining steps are outlined in Short Cut Help, which is accessed by clicking on Help | Contents | Programming Steps.

6. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the PC to datalogger connection is active, you can click Finish in Short Cut and you will be prompted to send the program just created to the datalogger.

7. If the sensor is connected to the datalogger, as shown in the wiring diagram in step 4, check the output of the sensor in the datalogger support software data display to make sure it is making reasonable measurements.

5. Overview The 092 ships from Campbell Scientific pre-wired for analog output with the following settings:

Analog Output Voltage: 0 to 5 Vdc Range: 600 to 1100 mbar

Refer to Appendix C, Analog Output Configurations (p. C-1), for available analog output options.

The 092 uses an on-board microcontroller and piezoresistive pressure sensor module. This module contains an analog-to-digital converter, a temperature sensor, and non-volatile memory for storage of factory-determined calibration coefficients. The microcontroller polls the sensor module once per second. Measurements are temperature corrected, the calibration coefficients applied, and the processed pressure measurement stored for output.

The pressure sensor element is light sensitive. Do not operate the 092 without the top cover of the polycarbonate case securely in place.

All wiring terminals on the 092 are protected from static-surge damage by transzorbs and current-limiting resistors.

Although the 092 is configured at Campbell Scientific for analog output, digital output options including RS-232 and SDI-12 are available. Contact Campbell Scientific for more information.

NOTE

Model 092 Barometric Pressure Sensor

5

6. Specifications Features:

Configured by Campbell Scientific

Configurable settings:

o measurement range

o data output option / communication option

o analog output range

Compatible with Campbell Scientific CRBasic dataloggers: CR200(X) series, CR300 series, CR6 series, CR800 series, CR1000, CR3000, CR5000, and CR9000(X)

Measurement Range: 600 to 1100 mbar Resolution: 0.1 mbar (.003 Hg) Accuracy: 0.35 mbar at 25 C 0.75 mbar at 0 to 55 C 1.5 mbar at 40 C Long-term stability: 1 mbar in 12 months

Temperature Operating range: 40 to 55 C Compensated range: 40 to 55 C

Power requirement Source: 6 to 16 Vdc Load: 10 mA at 12 Vdc

Output: Analog, 0 to 5 Vdc

Enclosure: Polycarbonate

Size: 120 x 80 x 55 mm (4.7 x 3.1 x 2.2 in)

CE certification: See Appendix E, CE Certification (p. E-1)

Calibration: Factory-calibrated against a NIST traceable anemometer

7. Installation If you are programming your datalogger with Short Cut, skip Section 7.2, Datalogger Wiring (p. 6), and Section 7.3, Programming (p. 7). Short Cut does this work for you. See Section 4, QuickStart (p. 1), for a Short Cut tutorial.

7.1 Mounting The 092 is factory-mounted to a solar shield. The solar shield mounts to a vertical pipe with a diameter of up to 2-inch IPS. Install the sensor such that it

Model 092 Barometric Pressure Sensor

6

faces north in the northern hemisphere, or south in the southern hemisphere. This orientation will shield the polycarbonate sensor housing from most direct sunlight.

To install the 092:

1. The 092 enclosure is pre-supplied with desiccant and does not need to be opened.

2. Attach the 092 to the tripod or tower leg using the supplied U-bolts. FIGURE 7-1 shows the 092 installed on a vertical mast.

FIGURE 7-1. 092 installed on a vertical mast

7.2 Datalogger Wiring Datalogger wiring is provided in TABLE 7-1 and a wiring schematic is provided in Appendix D, Internal Cable Connections (p. D-1).

Model 092 Barometric Pressure Sensor

7

TABLE 7-1. Wire Color, Wire Function, and Datalogger Connection

Wire Color

Wire Function Datalogger Connection Terminal

White Signal U configured for single-ended analog input1, SE (single-ended, analog input)

Green Signal Reference AG or (analog ground)

Red Power 12V

Black Ground G

Clear Shield AG or (analog ground) 1U channels are automatically configured by the measurement instruction.

7.3 Programming Short Cut is the best source for up-to-date datalogger programming code. Programming code is needed when:

Creating a program for a new datalogger installation Adding sensors to an existing datalogger program

If your data acquisition requirements are simple, you can probably create and maintain a datalogger program exclusively with Short Cut. If your data acquisition needs are more complex, the files that Short Cut creates are a great source for programming code to start a new program or add to an existing custom program.

Short Cut cannot edit programs after they are imported and edited in CRBasic Editor.

A Short Cut tutorial is available in Section 4, QuickStart (p. 1). If you wish to import Short Cut code into CRBasic Editor to create or add to a customized program, follow the procedure in Appendix A, Importing Short Cut Code Into CRBasic Editor (p. A-1). Programming basics for CRBasic dataloggers are in the following section. A complete program example for a select CRBasic datalogger can be found in Appendix B, Example Program (p. B-1).

7.3.1 CRBasic Instruction

The VoltSE() measurement instruction typically programs the datalogger to measure the 092.

VoltSE( Dest, Reps, Range, SEChan, MeasOff, SettlingTime, Integration, Multiplier, Offset )

For the multiplier, enter 0.1. For the offset, enter 600. The offset will need to be adjusted if the barometer is not at sea level (see Section 8.2, Correcting Pressure to Sea Level (p. 8)). If barometric pressure units other than mbar or hPA are desired, see Section 8.1, Conversion Factors (p. 8).

NOTE

Model 092 Barometric Pressure Sensor

8

Atmospheric pressure changes little with time. In most weather station applications, measuring the barometer pressure once an hour is adequate. See Appendix B, Example Program (p. B-1), for more information.

8. Operations 8.1 Conversion Factors

To report pressure in different units, multiply the measured pressure by the appropriate conversion factor. This is done by including an expression in the CRBasic program. See TABLE 8-1 below for conversion factors.

TABLE 8-1. Conversion Factors for Alternative Pressure Units

To Find Multiply By

hPa 1.0

kPa 0.1

mm of Hg 0.75006

in of Hg 0.02953

Psi 0.0145

Atm 0.00099

Torr 0.75006

8.2 Correcting Pressure to Sea Level The weather service, most airports, radio stations, and television stations adjust the atmospheric pressure to a common reference (sea level). Equation 1 can be used to find the difference in pressure between the sea level and the site. That value (dP) is then added to the offset in the measurement instruction. U. S. Standard Atmosphere and dry air were assumed when Equation 1 was derived (Wallace, J. M. and P. V. Hobbes, 1977: Atmospheric Science: An Introductory Survey, Academic Press, pp. 59-61).

dP E

=

1013 25 1 1

44307 69231

5.25328 .

. (1)

The value dP is in millibars and the site elevation, E, is in meters. Add dP value to the offset in the measurement instruction.

Use Equation 2 to convert feet to meters.

E m E ft

ft m ( )

( ) .

= 3 281

(2)

The corrections involved can be significant. For example, at 1000 mb and 20 C, barometric pressure will decrease by 1.1 mb for every 10 m increase in altitude.

Model 092 Barometric Pressure Sensor

9

9. Maintenance 9.1 Cleaning

Periodic cleaning of the sintered filter at the base of the unit may be required in dusty conditions. Remove the filter and clean with distilled water.

9.2 Calibration Calibration is performed against a NIST traceable standard. Calibration coefficients are stored in non-volatile memory at the factory. No user calibration procedure is available. The sensor must be returned to Met One Instruments when calibration is desired. Yearly calibrations are recommended.

For calibration, contact:

Met One Instruments 1600 Washington Blvd. Grants Pass, OR 97526 Phone: (541) 471-7111 Fax: (541) 471-7116

Met One Instruments 3206 Main Street Suite 106 Rowlett, TX 75088 Phone: (972) 412-4747 Fax: (972) 412-4716

Model 092 Barometric Pressure Sensor

10

A-1

Appendix A. Importing Short Cut Code Into CRBasic Editor

This tutorial shows:

How to import a Short Cut program into a program editor for additional refinement

How to import a wiring diagram from Short Cut into the comments of a custom program

Short Cut creates files, which can be imported into CRBasic Editor. Assuming defaults were used when Short Cut was installed, these files reside in the C:\campbellsci\SCWin folder:

.DEF (wiring and memory usage information) .CR2 (CR200(X)-series datalogger code) .CR300 (CR300-series datalogger code) .CR6 (CR6-series datalogger code) .CR8 (CR800-series datalogger code) .CR1 (CR1000 datalogger code) .CR3 (CR3000 datalogger code) .CR5 (CR5000 datalogger code). .CR9 (CR9000(X) datalogger code)

Use the following procedure to import Short Cut code and wiring diagram into CRBasic Editor.

1. Create the Short Cut program following the procedure in Section 4, QuickStart (p. 1). Finish the program and exit Short Cut. Make note of the file name used when saving the Short Cut program.

2. Open CRBasic Editor.

3. Click File | Open. Assuming the default paths were used when Short Cut was installed, navigate to C:\CampbellSci\SCWin folder. The file of interest has the .CR2, .CR300, .CR6, .CR8, .CR1, .CR3, .CR5, or .CR9 extension. Select the file and click Open.

4. Immediately save the file in a folder different from C:\Campbellsci\SCWin, or save the file with a different file name.

Once the file is edited with CRBasic Editor, Short Cut can no longer be used to edit the datalogger program. Change the name of the program file or move it, or Short Cut may overwrite it next time it is used.

5. The program can now be edited, saved, and sent to the datalogger.

6. Import wiring information to the program by opening the associated .DEF file. Copy and paste the section beginning with heading -Wiring for CRXXX into the CRBasic program, usually at the head of the file. After pasting, edit the information such that an apostrophe (') begins each line. This character instructs the datalogger compiler to ignore the line when compiling.

NOTE

B-1

Appendix B. Example Program The following CR1000 programming example can be used, in whole or in part, if the datalogger is being programmed with CRBasic Editor. Programming is similar with the CR800 and CR3000.

CRBasic Example B-1. CR1000 Program Measuring the 092

'CR1000 with 092 Barometric Pressure Sensor '-Wiring- '1H: White 'Ground: Green 'Ground: Clear 'G: Black '12V: Red 'Declare Variables and Units Public BattV Public PTemp_C Public BP_mmHg Units BattV=Volts Units PTemp_C=Deg C Units BP_mmHg=mmHg 'Define Data Tables DataTable(Table1,True,-1) DataInterval(0,60,Min,10) Average(1,BP_mmHg,FP2,False) EndTable DataTable(Table2,True,-1) DataInterval(0,1440,Min,10) Minimum(1,BattV,FP2,False,False) EndTable 'Main Program BeginProg 'Main Scan Scan(5,Sec,1,0) 'Default Datalogger Battery Voltage measurement 'BattV' Battery(BattV) 'Default Wiring Panel Temperature measurement 'PTemp_C' PanelTemp(PTemp_C,_60Hz) '092 Barometric Pressure Sensor measurement 'BP_mmHg' VoltSE(BP_mmHg,1,mV5000,1,1,0,_60Hz,0.1,600) BP_mmHg=BP_mmHg*0.75006 'Call Data Tables and Store Data CallTable(Table1) CallTable(Table2) NextScan EndProg

C-1

Appendix C. Analog Output Configurations

The 092 ships from Campbell Scientific pre-wired and pre-configured for analog output with the following settings:

Analog Output Voltage: 0 to 5 Vdc Range: 600 to 1100 mbar

Output voltage and measurement ranges are set by switches. Dip switch settings are listed in TABLE C-1 and TABLE C-2. Highlighted settings are those set at the factory by Campbell Scientific.

TABLE C-1. Dip Switches SW1: Analog Output Voltage Settings

Range SW1-1 SW1-2

0 to 1 Vdc On On

0 to 2 Vdc On Off

0 to 2.5 Vdc Off On

0 to 5 Vdc* Off* Off*

*Factory configuration at Campbell Scientific.

Appendix C. Analog Output Configurations

C-2

TABLE C-2. Dip Switches SW2: Analog Output Pressure Range Settings

RANGE SELECTION SWITCH SETTINGS

mbar mm Hg in. Hg

LOWER UPPER LOWER UPPER LOWER UPPER SW2-1 SW2-2 SW2-3 SW2-4

600 800 450 600 17.72 23.62 ON ON ON ON

600 900 450 675 17.72 26.58 OFF ON ON ON

600 1000 450 750 17.72 29.53 ON OFF ON ON

6001 11001 4501 8251 17.721 32.481 OFF1 OFF1 ON1 ON1

700 800 525 600 20.67 23.62 ON ON OFF ON

700 900 525 675 20.67 26.58 OFF ON OFF ON

700 1000 525 750 20.67 29.53 ON OFF OFF ON

700 1100 525 825 20.67 32.48 OFF OFF OFF ON

677.1 1083.6 508.0 812.8 20.00 32.00 ON ON ON OFF

800 900 600 675 23.62 26.58 OFF ON ON OFF

800 1000 600 750 23.62 29.53 ON OFF ON OFF

800 1100 600 825 23.62 32.48 OFF OFF ON OFF 2 2 2 2 2 2 ON2 ON2 OFF2 OFF2 2 2 2 2 2 2 OFF2 ON2 OFF2 OFF2

900 1000 675 750 26.58 29.53 ON OFF OFF OFF

900 1100 675 825 26.58 32.48 OFF OFF OFF OFF 1Factory configuration at Campbell Scientific 2Settings with no pressure values are invalid settings. If the switches are set to either of these combinations, the analog output will default to the full range of 600 to 1100 mbar.

D-1

Appendix D. Internal Cable Connections

The 092 ships from Campbell Scientific pre-wired for analog output. Internal cable connections are shown in FIGURE D-1.

FIGURE D-1. Analog wiring schematic

E-1

Appendix E. CE Certification

Campbell Scientific Companies

Campbell Scientific, Inc. 815 West 1800 North Logan, Utah 84321 UNITED STATES

www.campbellsci.com info@campbellsci.com

Campbell Scientific Africa Pty. Ltd. PO Box 2450

Somerset West 7129 SOUTH AFRICA

www.campbellsci.co.za cleroux@csafrica.co.za

Campbell Scientific Southeast Asia Co., Ltd. 877/22 Nirvana@Work, Rama 9 Road

Suan Luang Subdistrict, Suan Luang District Bangkok 10250

THAILAND www.campbellsci.asia info@campbellsci.asia

Campbell Scientific Australia Pty. Ltd. PO Box 8108

Garbutt Post Shop QLD 4814 AUSTRALIA

www.campbellsci.com.au info@campbellsci.com.au

Campbell Scientific (Beijing) Co., Ltd. 8B16, Floor 8 Tower B, Hanwei Plaza

7 Guanghua Road Chaoyang, Beijing 100004

P.R. CHINA www.campbellsci.com info@campbellsci.com.cn

Campbell Scientific do Brasil Ltda. Rua Apinags, nbr. 2018 Perdizes CEP: 01258-00 So Paulo SP

BRASIL www.campbellsci.com.br vendas@campbellsci.com.br

Campbell Scientific Canada Corp. 14532 131 Avenue NW Edmonton AB T5L 4X4

CANADA www.campbellsci.ca dataloggers@campbellsci.ca

Campbell Scientific Centro Caribe S.A. 300 N Cementerio, Edificio Breller

Santo Domingo, Heredia 40305 COSTA RICA

www.campbellsci.cc info@campbellsci.cc

Campbell Scientific Ltd. Campbell Park

80 Hathern Road Shepshed, Loughborough LE12 9GX

UNITED KINGDOM www.campbellsci.co.uk sales@campbellsci.co.uk

Campbell Scientific Ltd. 3 Avenue de la Division Leclerc

92160 ANTONY FRANCE

www.campbellsci.fr info@campbellsci.fr

Campbell Scientific Ltd. Fahrenheitstrae 13

28359 Bremen GERMANY

www.campbellsci.de info@campbellsci.de

Campbell Scientific Spain, S. L. Avda. Pompeu Fabra 7-9, local 1

08024 Barcelona SPAIN

www.campbellsci.es info@campbellsc

Manualsnet FAQs

If you want to find out how the 92 Campbell Scientific works, you can view and download the Campbell Scientific 92 Sensor Instruction Manual on the Manualsnet website.

Yes, we have the Instruction Manual for Campbell Scientific 92 as well as other Campbell Scientific manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Instruction Manual should include all the details that are needed to use a Campbell Scientific 92. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Campbell Scientific 92 Sensor Instruction Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Campbell Scientific 92 Sensor Instruction Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Campbell Scientific 92 Sensor Instruction Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Campbell Scientific 92 Sensor Instruction Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Campbell Scientific 92 Sensor Instruction Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.