229 Heat Dissipation Matric Water Potential Sensor

Revision: 5/09

C o p y r i g h t 2 0 0 6 - 2 0 0 9 C a m p b e l l S c i e n t i f i c , I n c .

Warranty and Assistance The 229 HEAT DISSIPATION MATRIC WATER POTENTIAL SENSOR is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC, INC.'s option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CAMPBELL SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such products by surface carrier prepaid. This warranty shall not apply to any CAMPBELL SCIENTIFIC, INC. products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CAMPBELL SCIENTIFIC, INC. is not liable for special, indirect, incidental, or consequential damages.

Products may not be returned without prior authorization. The following contact information is for US and International customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com to determine which Campbell Scientific company serves your country.

To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. CAMPBELL SCIENTIFIC's shipping address is:

CAMPBELL SCIENTIFIC, INC. RMA#_____ 815 West 1800 North Logan, Utah 84321-1784

For all returns, the customer must fill out a Declaration of Hazardous Material and Decontamination form and comply with the requirements specified in it. The form is available from our website at www.campbellsci.com/repair. A completed form must be either emailed to shanna@campbellsci.com or faxed to 435-750-9579. Campbell Scientific will not process any returns until we receive this form. If the form is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customers expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.

CAMPBELL SCIENTIFIC, INC. does not accept collect calls.

229 Sensor Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat bookmarks tab for links to specific sections.

1. General Description ..................................................... 1 1.1 Compatibility ............................................................................................ 2 1.2 Measurement Principle ............................................................................. 2

2. Specifications .............................................................. 4

3. Installation .................................................................... 4 3.1 Orientation ................................................................................................ 4 3.2 Contact ...................................................................................................... 4 3.3 Equilibration and Saturation of the Sensor Before Installation ................ 5

4. Wiring ............................................................................ 5

5. Example Programs ...................................................... 6 5.1 Choosing a Reference for the Thermocouple Readings ........................... 6 5.2 Adjusting for Thermal Properties of Sensor During Early Heating Times7 5.3 Datalogger Program Structure and Multiplexers ...................................... 7 5.4 Temperature Correction ............................................................................ 8 5.5 Example #1 CR1000 with CE4 and Four 229s .................................... 9 5.6 Example #2 CR1000 with AM16/32-series Multiplexer,

CE4 and Sixteen 229 Sensors with Temperature Correction ............. 11 5.7 Example #3 CR10X with 229 Sensor ................................................ 13 5.8 Example #4 CR10X with AM16/32-series, CE4, and

Sixteen 229 Sensors ........................................................................... 16

6. Calibration .................................................................. 19 6.1 General.................................................................................................... 19 6.2 Normalized Temperature Change and Correction for Soil Temperature ...... 20

6.2.1 Normalized Temperature Change ................................................. 20 6.2.2 Correction for Soil Temperature ................................................... 21

6.3 Using Pressurized Extraction Methods ................................................... 24 6.4 General Description of Calibration/Measurement Process using

Pressure Plate Extractor ..................................................................... 24 6.4.1 Wiring for Calibration using Pressure Plate Extractor .................. 25

7. Maintenance ............................................................... 26

8. Troubleshooting ........................................................ 27

i

229 Sensor Table of Contents

9. References .................................................................. 28

List of Figures 1-1. 229 Heat Dissipation Matric Water Potential Sensor and Hypodermic

Assembly ............................................................................................. 1 1-2. CE4 and CE8 Current Excitation Modules ............................................. 2 1-3. Typical Temperature Response of 229 Sensor in Silt Loam Soil ........... 3 4-1. Schematic of Connections for Measurement of a 229 Sensor ................ 6 6-1. Data Points and Regression for Typical Calibration............................. 20 6-2. Measurement error for range of soil temperatures and wide range of

matric potential .................................................................................. 22 6-3. Measurement error for range of soil temperatures and wetter range of

matric potential .................................................................................. 23 6-4. Datalogger and Peripheral Connections for 229 Calibration ................ 26

List of Tables 4-1. 229 Sensor and CE4/CE8 Wiring ........................................................... 5 5-1. Wiring for Four 229s with CR1000 and CE4 ......................................... 9 5-2. 229 Sensor and CE4 Wiring with CR1000 and AM16/32-series ......... 11 5-3. 229 Sensor and CE4/CE8 Wiring with CR10XTCR ............................ 13 5-4. 229 Sensor and CE4/CE8 Wiring with AM16/32 Multiplexer ............. 16

ii

229 Heat Dissipation Matric Water Potential Sensor 1. General Description

The 229 Heat Dissipation Matric Water Potential Sensor uses a heat dissipation method to indirectly measure soil water matric potential. The active part of the 229 Soil Water Potential Sensor is a cylindrically-shaped porous ceramic body. A heating element which has the same length as the ceramic body is positioned at the center of the cylinder. A thermocouple is located at mid-length of the ceramic and heating element. The position of the heating element and the thermocouple is maintained by placing both inside a hypodermic needle. This also protects the delicate wires. The volume inside the needle which is not occupied by wiring is filled with epoxy.

FIGURE 1-1. A 229 Heat Dissipation Matric Water Potential Sensor is shown at the top. The hypodermic assembly (without epoxy and ceramic) is shown just below. Cutaway view shows longitudinal section of the needle with heater and thermocouple junction.

The ceramic cylinder has a diameter of 1.5 cm and a length of 3.2 cm. Three copper wires and one constantan wire, contained in a shielded, burial-grade sheath provide a path for connection to measuring instrumentation. An epoxy section which is the same diameter as the ceramic matrix gives strain relief to the cable.

The 229 is used to measure soil water matric potential in the range -10 kPa to -2500 kPa. The method relies on hydraulic continuity between the soil and the sensor ceramic for water exchange. The variability in heat transfer properties

1

229 Heat Dissipation Matric Water Potential Sensor

among sensors makes individual calibration by the user a requirement. See Section 6 for calibration information.

Use of the 229 sensor requires a constant current source. Campbell Scientific offers the CE4 and CE8 current excitation modules (Figure 1-2), which have respectively four and eight regulated outputs of 50 milliamp 0.25 milliamp. All of the outputs of the excitation module are switched on or off simultaneously by setting a single datalogger control port to its high or low state.

The L option on the model 229 Heat Dissipation Matric Water Potential Sensor (229-L) indicates that the cable length is user specified. This manual refers to the sensor as the 229.

FIGURE 1-2. CE4 and CE8 Current Excitation Modules

1.1 Compatibility Compatible dataloggers include our 21X, CR7, CR10(X), CR23X, CR800, CR850, CR1000, and CR3000. The 229 is not compatible with our CR200- series, CR500, or CR510 dataloggers. The 229 can be connected with a multiplexer. Compatible multiplexers include our AM16/32, AM16/32A, and AM16/32B.

When using multiplexers, the user should be aware that switching currents of greater than 30 mA will degrade the contact surfaces of the mechanical relays. This degradation will adversely affect the suitability of these relays to multiplex low voltage signals. Although a relay used in this manner no longer qualifies for low voltage measurements, it continues to be useful for switching currents in excess of 30 mA. Therefore, the user is advised to record which multiplexer channels are used to multiplex the 50 mA excitation for the 229-L sensors in order to avoid using those channels for low voltage measurements in future applications.

1.2 Measurement Principle Movement of water between the 229 ceramic matrix and the surrounding soil occurs when a water potential gradient exists. When the water potential of the soil surrounding a 229 sensor changes, a water flux with the ceramic matrix will occur. The time required for hydraulic equilibration of the water in the soil and ceramic depends on both the magnitude of the water potential gradient

NOTE

2

229 Heat Dissipation Matric Water Potential Sensor

and the hydraulic conductivity. Typically this equilibration time is on the order of minutes or tens of minutes.

A change in the water potential and water content of the ceramic matrix causes a corresponding change in the thermal conductivity of the ceramic/water complex. As the water content in the ceramic increases, the thermal conductivity of the complex also increases. At very low water contents, the ceramic material controls the thermal conductivity. As water content in the ceramic increases, water films are established between the solid particles, resulting in a rapid increase in thermal conductivity. As the pores in the ceramic continue to fill, the thermal conductivity becomes increasingly controlled by the continuous water and the increase in thermal conductivity of the ceramic/water complex approaches a constant value.

When a constant power is dissipated from the line heat source, the temperature increase near the heat source will depend on the thermal conductivity of the ceramic/water complex surrounding the heater. A temperature increase is caused by heat that is not dissipated. As the water content and thermal conductivity of the ceramic increases, the temperature increase as measured by the thermocouple will be reduced because conduction of the thermal energy from the heat source is greater. A drier sensor will have a lower thermal conductivity, so the thermal energy will not dissipate as quickly and the temperature rise will be greater. When 50 milliamps is passed through the heating element for 30 seconds, the temperature increase ranges from approximately 0.7C under wet conditions to 3.0C when dry. Figure 1-3 presents a typical temperature response in a silt loam.

0.5

1

1.5

2

2.5

3

0 5 10 15 20 25 30 heating time (s)

te m

pe ra

tu re

in cr

ea se

(C )

200 kPa 100 kPa 50 kPa 10 kPa

FIGURE 1-3. Typical Temperature Response of 229 Sensor in Silt Loam Soil

3

229 Heat Dissipation Matric Water Potential Sensor

2. Specifications 229

Measurement range: -10 to -2500 kPa

Measurement time: 30 seconds typical

Thermocouple type: copper / constantan (type T)

Dimensions: 1.5 cm (0.6) diameter 3.2 cm (1.3) length of ceramic cylinder 6.0 cm (2.4) length of entire sensor

Weight: 10 g (0.35 oz) plus 23 g/m (0.25 oz/ft) of cable

Heater resistance: 34 ohms plus cable resistance

Resolution: ~1 kPa at matric potentials greater than -100 kPa

CE4/CE8

Output: 50 mA 0.25 mA per channel, regulated

Output channels: CE4: 4 CE8: 8

Current drain(while active): 25 mA + 50 mA * no. of 229s connected to the CE4 or CE8 output channels.

Dimensions: CE4: 11.5 cm (4.5) x 5.4 cm (2.1) x 2.7 cm (1.1) CE8: 16.5 cm (6.5) x 5.4 cm (2.1) x 2.7 cm (1.1)

Weight: CE4: 131 g (4.6 oz) CE8: 184 g (6.5 oz)

3. Installation 3.1 Orientation

For best measurement results, the 229 should be installed horizontally at the desired depth of the soil. This will reduce distortion of typical vertical water flux.

3.2 Contact Good contact must exist between the ceramic matrix and the soil since the measurement relies on water flux between the two. Adequate contact will result if the sensor is planted in a manner similar to that used for seedlings. Sufficient contact in coarse texture soils such as medium and coarse sand can be obtained by surrounding the ceramic portion with a slurry of fine silica (sometimes referred to as silica flour).

4

229 Heat Dissipation Matric Water Potential Sensor

3.3 Equilibration and Saturation of the Sensor Before Installation

The smaller the difference in water potential between the 229 ceramic and the surrounding soil, the sooner equilibrium will be reached. Filling the ceramic pores with liquid water will optimize the hydraulic conductivity between the ceramic and soil.

Simple immersion of the sensors in water can leave some entrapped air in the pores. Complete saturation can be closely approached if (1) deaerated water is used, and (2) saturation occurs in a vacuum. Soaking the ceramic in free water for 12 hours followed by soaking under a vacuum of 71 kPa (0.7 atm) atmosphere for 1 hour results in complete saturation of the sensor.

4. Wiring Table 4-1 shows wiring information for the 229 sensor and CE4 or CE8 excitation module when connecting sensors directly to the datalogger without the use of a multiplexer. Figure 4-1 shows a simple schematic of these connections.

See the multiplexer program in Section 5 for details on multiplexer wiring.

TABLE 4-1. 229 Sensor and CE4/CE8 Wiring

229 Wire Color

Function

CR10(X), CR23X, CR800, CR850, CR1000, CR3000

CE4/CE8

Blue Thermocouple High High side of differential channel

Red Thermocouple Low Low side of differential channel

Green Heater High Current excitation channel

Black Heater Low Clear Shield G

CE4/CE8 Power 12V +12V

CE4/CE8 Ground G CE4/CE8 Enable Control Port CTRL

5

229 Heat Dissipation Matric Water Potential Sensor

FIGURE 4-1. Schematic of Connections for Measurement of a 229 Sensor

5. Example Programs 5.1 Choosing a Reference for the Thermocouple Readings

A fundamental thermocouple circuit uses two thermocouple junctions with one pair of common-alloy leads tied together and the other pair connected to a voltage readout device. One of the junctions is the reference junction and is generally held at a known temperature. The temperature at the other junction can be determined by knowing the voltage potential difference between the junctions and the reference temperature.

A Campbell Scientific datalogger can read a single thermocouple junction directly because the temperature at the wiring panel is measured with a thermistor and this temperature is converted to a voltage which is then used as a thermocouple reference. A thermocouple circuit voltage potential is affected by the temperature of all dissimilar metal junctions.

When using a multiplexer with the 229 sensor, the temperature of the multiplexer can be used as the reference temperature if a thermistor probe such as the 107 is taped to the multiplexer panel near the 229 wires. Alternately, a CR10XTCR can be used to get an accurate reading of the CR10X wiring panel temperature and type T thermocouple wire (copper-constantan) can be used for the signal wires between the differential voltage channel on the datalogger and the appropriate common channels on the multiplexer (see program example #2 below). The CR23X, CR800, CR850, CR1000, and CR3000 can use their own internal panel temperature measurement instead of the CR10XTCR and type T thermocouple wire to the multiplexer common channels as previously noted. The use of insulation or an enclosure to keep the multiplexer and temperature sensor at the same temperature will improve measurement quality.

6

229 Heat Dissipation Matric Water Potential Sensor

5.2 Adjusting for Thermal Properties of Sensor During Early Heating Times

The discussion presented at the beginning of the calibration section (Section 6) describes how thermal properties can vary from sensor-to-sensor. The thermal properties of the needle casing, wiring, and the amount of contact area between the needle and the ceramic have a slight effect on the temperature response. Most of the nonideal behavior of the sensor is manifest in the first second of heating. The measurement is improved if the temperature after 1 s is subtracted from some final temperature. A typical T would be T(30 s) - T(1 s).

5.3 Datalogger Program Structure and Multiplexers The sequence of datalogger instructions for a 229 measurement is as follows:

1) Measure sensor temperature prior to heating.

2) Set a control port high to enable constant current excitation module and being heating.

3) Wait for one second of heating and measure sensor temperature.

4) Wait for 29 more seconds of heating and measure sensor temperature.

5) Set control port low to disable the constant current excitation module and end heating.

6) Calculate temperature rise by subtracting T(1 s) from T(30 s).

Since all of the output channels of the CE4 or CE8 are activated when the control terminal is set high, power will be applied to all of the 229 sensors connected to the current source. Inaccurate measurements can result if the temperature of multiple sensors is simply read sequentially. The inaccuracy can occur because a finite amount of time is required to execute each of the temperature measurement instructions.

For example, a CR10X making multiple differential thermocouple readings with 60 Hz rejection takes 34.9 ms to read one thermocouple, and 30.9 ms more for each additional thermocouple. In a configuration where six 229 sensors are connected to a CE8 with their thermocouple wires connected sequentially to the CR10X wiring panel, the sixth 229 sensor will heat for 154.5 ms longer than the first sensor each time its temperature is measured. The amount of time between temperature measurement of the first sensor and the last sensor can be as long as 0.5 seconds under some measurement configurations.

The error caused by this difference in heating times can be minimized if the sensors are connected to the constant current excitation module and datalogger during calibration in exactly the same order they will be wired during field deployment. The difference in heating times can be eliminated altogether by heating the sensors one at a time through a multiplexer such as the AM16/32B.

7

229 Heat Dissipation Matric Water Potential Sensor

The AM16/32B multiplexer in 4x16 mode provides a convenient method to measure up to sixteen 229 sensors. Since four lines are switched at once, both the thermocouple and the heating element leads for each sensor can be connected to a multiplexer channel. A measurement sequence is executed on each sensor. See program example #2 below for instructions on wiring and programming multiple 229 sensors on a multiplexer.

When using multiplexers, the user should be aware that switching currents of greater than 30 mA will degrade the contact surfaces of the mechanical relays. This degradation will adversely affect the suitability of these relays to multiplex low voltage signals. Although a relay used in this manner no longer qualifies for low voltage measurements, it continues to be useful for switching currents in excess of 30 mA. Therefore, the user is advised to record which multiplexer channels are used to multiplex the 50 mA excitation for the 229-L sensors in order to avoid using those channels for low voltage measurements in future applications.

5.4 Temperature Correction The rise in temperature over the 30 second heating time will be affected by the initial temperature of the 229 sensor. The measurement can be corrected for temperature if it can first be normalized. See Section 6.2.1 for more information on normalizing the sensor output.

Correct for temperature in the datalogger program following these steps outlined in Flint, et al. (2002):

1. Determine normalized dimensionless temperature rise, Tnorm as described in Section 6.2.1 equation 3

wetdry

dry norm TT

TT T

=

2. Determine where Tnorm falls on the dimensionless slope function, s*

0013.0011.00203.00747.00559.00133.0* 2345 ++++= normnormnormnormnorm TTTTTs

3. Determine corrected dimensionless temperature rise, Tnormcorr

( )calsoilnormnormcorr TTsTT = *

4. Determine corrected temperature rise, Tcorr

( )wetdrynormcorrdcorr TTTTT =

The corrected temperature rise can now be used to calculate soil water matric potential. Example 4 below shows how to make the temperature correction with a CR1000.

NOTE

8

229 Heat Dissipation Matric Water Potential Sensor

5.5 Example #1 CR1000 with CE4 and Four 229s Table 5-1 shows wiring information for reading four 229 sensors with a CR1000 datalogger and CE4 current excitation module.

TABLE 5-1. Wiring for Four 229s with CR1000 and CE4

229 CR1000 229 CE4

229 #1Blue 1H 229 #1 Green Channel 1

229 #1Red 1L 229 #1 Black 229 #2Blue 2H 229 #2 Green Channel 2

229 #2Red 2L 229 #2 Black 229 #3Blue 3H 229 #4 Green Channel 3

229 #3Red 3L 229 #3 Black 229 #4Blue 4H 229 #4 Green Channel 4

229 #4Red 4L 229 #4 Black 229 #1 Clear 229 #2 Clear 229 #3 Clear 229 #4 Clear 12V +12V

G C1 CTRL

9

229 Heat Dissipation Matric Water Potential Sensor

'CR1000 SequentialMode Const Num229 = 4 'Enter number of 229 sensors to measure Dim LoopCount Public RefTemp_C, StartTemp_C(Num229), Temp_1sec_C(Num229) Public Temp_30sec_C(Num229), DeltaT_C(Num229) Public Flag(1) as Boolean Units StartTemp_C()=Deg C Units DeltaT_C()=Deg C DataTable(Matric,Flag(1),-1) Sample(Num229,StartTemp_C(),FP2) Sample(Num229,DeltaT_C(),FP2) EndTable BeginProg Scan(30,Sec,1,0) PanelTemp (RefTemp_C,250) If IfTime (0,240,Min) Then Flag(1)=True 'Every 4 hours set Flag(1) high If Flag(1) = True Then 'Flag(1) true triggers 229 readings 'Measure starting temperature before heating TCDiff(StartTemp_C(),Num229,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0) PortSet (1,1 ) 'Set C1 high to activate CE4 Delay (0,1,Sec) 'Wait 1 second 'Measure temperature after 1 second of heating TCDiff(Temp_1sec_C(),Num229,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0) 'Measure temperature after 30 second of heating Delay (0,29,Sec) 'Wait 29 seconds more for total of 30 seconds heating TCDiff(Temp_30sec_C(),Num229,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0) PortSet (1,0 ) 'Set C1 low to deactivate CE4 For LoopCount=1 to Num229 'Calculate temperature rise DeltaT_C(LoopCount)=Temp_30sec_C(LoopCount)-Temp_1sec_C(LoopCount) 'LoopCount=LoopCount+1 Next LoopCount EndIf 'Ends Flag(1) true condition CallTable(Matric) 'Call Data Tables and Store Data Flag(1)=False 'Set Flag 1 false to disable 229 measurements NextScan EndProg

10

229 Heat Dissipation Matric Water Potential Sensor

5.6 Example #2 CR1000 with AM16/32-series Multiplexer, CE4 and Sixteen 229 Sensors with Temperature Correction

Table 5-2 shows wiring information for connecting multiple 229 sensors and CE4 excitation module to an AM16/32 multiplexer and CR1000 datalogger. See Figure 6-4 for a schematic of this wiring configuration.

TABLE 5-2. 229 Sensor and CE4 Wiring with CR1000 and AM16/32-series

229

107

Function

CR1000

CE4

Multiplexer (4x16 mode)

Blue 229 Thermocouple High ODD H (1H, 3H, etc)

Red 229 Thermocouple Low ODD L (1L, 3L, etc)

Green 229 Heater High EVEN H (2H, 4H, etc)

Black 229 Heater Low EVEN L (2L, 4L, etc)

Clear 229 Shield G CE4 Power 12V +12V

CE4 Ground G

CE4 Enable C3 CTRL

1H COM ODD H

1L COM ODD L

CE4 current excitation channel Channel 1 COM EVEN H

CE4 Ground COM EVEN L

AM16 Power 12V 12V

AM16/32 Ground G GND

AM16/32 Enable C1 RES

AM16/32 Advance C2 CLK

Red 107 Signal SE 3

Black 107 Excitation EX1

Purple 107 Signal Ground

Clear 107 Shield G

11

229 Heat Dissipation Matric Water Potential Sensor

'CR1000 SequentialMode Const Num229 = 16 'Enter number of 229 sensors to measure Const read229 = 60 'Enter Number of minutes between 229-L readings Const CalTemp = 20 'Enter calibration temperature (deg C) Dim i, dTdry(Num229), dTwet(Num229) Dim Tstar, Tstarcorr, DeltaTcorr, s Public RefTemp_C, StartTemp_C(Num229), Temp_1sec_C(Num229) Public Temp_30sec_C(Num229), DeltaT_C(Num229), dTcorr(Num229) Public Flag(1) as Boolean Units StartTemp_C()=Deg C Units DeltaT_C()=Deg C Units dTcorr() = Deg C DataTable(Matric,Flag(1),-1) Sample(Num229,StartTemp_C(),FP2) Sample(Num229,DeltaT_C(),FP2) Sample(Num229,dTcorr(),FP2) EndTable Sub TempCorr 'Subroutine to temperature correct DeltaT_C Tstar=(dTdry(i)-DeltaT_C(i))/(dTdry(i)-dTwet(i)) s=-0.0133*Tstar^5+0.0559*Tstar^4-0.0747*Tstar^3+0.0203*Tstar^2+0.011*Tstar+0.0013 Tstarcorr=Tstar-s*(StartTemp_C(i)-CalTemp) DeltaTcorr=dTdry(i)-Tstarcorr*(dTdry(i)-dTwet(i)) EndSub BeginProg 'Enter dTdry and dTwet values obtained for each probe (these are examples): dTdry(1)= 3.421: dTdry(2)=3.417: dTdry(3)=3.433: dTdry(4)=3.418 dTdry(5)= 3.412: dTdry(6)=3.407: dTdry(7)=3.422: dTdry(8)=3.428 dTdry(9)= 3.399: dTdry(10)=3.377: dTdry(11)=3.405: dTdry(12)=3.406 dTdry(13)=3.422: dTdry(14)=3.431: dTdry(15)=3.423: dTdry(16)=3.408 dTwet(1)= 0.752: dTwet(2)=0.695: dTwet(3)=0.731: dTwet(4)=0.724 dTwet(5)= 0.709: dTwet(6)=0.752: dTwet(7)=0.739: dTwet(8)=0.737 dTwet(9)= 0.723: dTwet(10)=0.754: dTwet(11)=0.691: dTwet(12)=0.760 dTwet(13)=0.722: dTwet(14)=0.745: dTwet(15)=0.739: dTwet(16)=0.748 Scan(30,Sec,1,0) Therm107(RefTemp_C,1,3,1,0,_60Hz,1.0,0.0) 'Reference temperature measurement If IfTime (0,read229,Min) Then Flag(1)=True 'Set Flag(1) high based on time If Flag(1) = True Then 'Flag(1) triggers 229 readings PortSet(1,1) 'Set C1 High to turn on multiplexer For i=1 to Num229 PulsePort(2,10000) 'Switch to next multiplexer channel 'Measure starting temperature before heating TCDiff(StartTemp_C(i),1,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0) PortSet (3,1 ) 'Set C3 high to activate CE4 Delay (0,1,Sec) 'Wait 1 second 'Measure temperature after 1 second of heating TCDiff(Temp_1sec_C(i),1,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0) Delay (0,29,Sec) 'Wait 29 seconds more

12

229 Heat Dissipation Matric Water Potential Sensor

'Measure temperature after 30 second of heating TCDiff(Temp_30sec_C(i),1,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0) PortSet (3,0 ) 'Set C3 low to deactivate CE4 DeltaT_C(i)=Temp_30sec_C(i)-Temp_1sec_C(i)'Calculate temperature rise Call TempCorr 'Call temperature correction subroutine dTcorr(i)=DeltaTcorr Next i EndIf 'Ends Flag(1) high condition PortSet(1,0) 'Turn multiplexer Off CallTable(Matric) 'Call Data Tables and Store Data Flag(1)=False 'Disable 229 measurements NextScan EndProg

5.7 Example #3 CR10X with 229 Sensor Table 5-3 shows wiring information for reading a single sensor with a CR10X datalogger, CE4 current excitation module, and CR10XTCR thermocouple reference temperature sensor.

TABLE 5-3. 229 Sensor and CE4/CE8 Wiring with CR10XTCR

229 Function CR10(X) CE4 CR10XTCR

Blue 229 Thermocouple High 1H

Red 229 Thermocouple Low 1L

Green 229 Heater High Channel 1

Black 229 Heater Low

Clear 229 Shield G

CE4/CE8 Power 12V +12V

CE4/CE8 Ground G

CE4/CE8 Enable C1 CTRL

CR10XTCR Signal 2H Red

CR10XTCR Signal Return AG Clear

CR10XTCR Excitation E3 Black

13

229 Heat Dissipation Matric Water Potential Sensor

;{CR10X} ;Program to read 1 229-L sensor ;Reading 1 sensor takes 30 seconds *Table 1 Program 01: 60 Execution Interval (seconds) 1: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 11 Set Flag 1 High 2: If Flag/Port (P91) 1: 11 Do if Flag 1 is High 2: 30 Then Do 3: Temp (107) (P11) ;Measure reference temperature 1: 1 Reps 2: 3 SE Channel 3: 3 Excite all reps w/E3 4: 1 Loc [ Ref_Temp ] 5: 1.0 Multiplier 6: 0.0 Offset 4: Thermocouple Temp (DIFF) (P14) ;Measure initial sensor temperature 1: 1 Reps 2: 1 2.5 mV Slow Range 3: 1 DIFF Channel 4: 1 Type T (Copper-Constantan) 5: 1 Ref Temp (Deg. C) Loc [ Ref_Temp ] 6: 2 Loc [ Tinit_1 ] 7: 1.0 Mult 8: 0.0 Offset 5: Do (P86) ;Turn on CE8 1: 41 Set Port 1 High 6: Excitation with Delay (P22) ;Wait one second before taking first reading 1: 1 Ex Channel 2: 0000 Delay W/Ex (0.01 sec units) 3: 100 Delay After Ex (0.01 sec units) 4: 0 mV Excitation 7: Thermocouple Temp (DIFF) (P14) ;Take 1 second temperature reading 1: 1 Reps 2: 1 2.5 mV Slow Range 3: 4 DIFF Channel 4: 1 Type T (Copper-Constantan) 5: 1 Ref Temp (Deg. C) Loc [ ref_temp ] 6: 3 Loc [ T1sec_1 ] 7: 1.0 Mult 8: 0.0 Offset

14

229 Heat Dissipation Matric Water Potential Sensor

8: Excitation with Delay (P22) ;Wait 29 more seconds for next reading 1: 1 Ex Channel 2: 0 Delay W/Ex (0.01 sec units) 3: 2900 Delay After Ex (0.01 sec units) 4: 0000 mV Excitation 9: Thermocouple Temp (DIFF) (P14) ;Take 30 second temperature reading 1: 1 Reps 2: 1 2.5 mV Slow Range 3: 4 DIFF Channel 4: 1 Type T (Copper-Constantan) 5: 1 Ref Temp (Deg. C) Loc [ ref_temp ] 6: 4 Loc [ T30sec_1 ] 7: 1.0 Mult 8: 0.0 Offset 10: Do (P86) ;turn off CE8 1: 52 Set Port 2 Low 11: Z=X-Y (P35) ;Calculate delta T 1: 4 X Loc [ T30sec_1 ] 2: 3 Y Loc [ T1sec_1 ] 3: 5 Z Loc [ deltaT_1 ] 12: Do (P86) ;Save readings to final storage 1: 10 Set Output Flag High (Flag 0) 13: Set Active Storage Area (P80) 1: 1 Final Storage Area 1 2: 229 Array ID 14: Real Time (P77) 1: 1220 Year,Day,Hour/Minute (midnight = 2400) 15: Sample (P70) 1: 1 Reps 2: 2 Loc [ Tinit_1 ] 16: Sample (P70) 1: 1 Reps 2: 5 Loc [ deltaT_1 ] 17: Do (P86) ;Disable flag 1 1: 21 Set Flag 1 Low 18: End (P95)

15

229 Heat Dissipation Matric Water Potential Sensor

5.8 Example #4 CR10X with AM16/32-series, CE4, and Sixteen 229 Sensors

Table 5-4 shows wiring information for connecting multiple 229 sensors and CE4 or CE8 excitation module to an AM16/32-series multiplexer and CR10X datalogger. A CR10XTCR or 107 probe should be used for the reference temperature measurement as described at the beginning of this section. See Figure 6-4 for a schematic of this wiring configuration.

TABLE 5-4. 229 Sensor and CE4/CE8 Wiring with AM16/32 Multiplexer

229

107/CR10XTCR

Function

CR10(X)

CE4/CE8

AM16/32 (4x16 mode)

Blue 229 Thermocouple High ODD H (1H, 3H, etc)

Red 229 Thermocouple Low ODD L (1L, 3L, etc)

Green 229 Heater High EVEN H (2H, 4H, etc)

Black 229 Heater Low EVEN L (2L, 4L, etc)

Clear 229 Shield G CE4/CE8 Power 12V +12V

CE4/CE8 Ground G

CE4/CE8 Enable C1 CTRL

1H COM ODD H*

1L COM ODD L*

CE4/CE8 current excitation channel

Channel 1 COM EVEN H

CE4/CE8 ground COM EVEN L

AM16/32 Power 12V 12V

AM16/32 Ground G GND

AM16/32 Enable C2 RES

AM16/32 Advance C3 CLK

Red 107 Signal SE5

Black 107 Excitation E1

Purple (Clear for CR10XTCR)

107 Signal Ground AG

Clear (107 only) 107 Shield G

* Run copper wire from COM ODD H and COM ODD L to 1H/1L if using a 107 probe taped to the AM16/32 as the reference temperature sensor. If using a CR10XTCR as the reference temperature sensor, run copper wire from COM ODD H to 1H and run constantan wire from COM ODD L to 1L.

16

229 Heat Dissipation Matric Water Potential Sensor

;{CR10X} ;Program to read 16 229-L sensors using 1 AM16/32 multiplexer ;and 1 CE4 or CE8 constant current interface ;Manually set Flag 1 high to force readings *Table 1 Program 01: 30 Execution Interval (seconds) 1: Batt Voltage (P10) 1: 1 Loc [ Batt_Volt ] 2: If time is (P92) ;Set Flag 1 high each hour 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 11 Set Flag 1 High 3: If Flag/Port (P91) 1: 11 Do if Flag 1 is High 2: 30 Then Do 4: Do (P86) ;Turn on AM16/32 1: 4 Set Port 2 High 5: Beginning of Loop (P87) ;Loop of 16 for 16 sensors on AM16/32 1: 0 Delay 2: 16 Loop Count 6: Do (P86) ;Advance to next multiplexer channel 1: 73 Pulse Port 3 7: Temp (107) (P11) ;Measure reference temperature 1: 1 Reps ;This is the same instruction 2: 5 SE Channel ;for both the CR10X and 107 3: 1 Excite all reps w/E1 4: 2 Loc [ Tref_C ] 5: 1.0 Mult 6: 0.0 Offset 8: Thermocouple Temp (DIFF) (P14) ;Read initial temperature 1: 1 Reps 2: 21 10 mV, 60 Hz Reject, Slow Range 3: 1 DIFF Channel 4: 1 Type T (Copper-Constantan) 5: 2 Ref Temp (Deg. C) Loc [ Tref_C ] 6: 3 -- Loc [ Tinit_1 ] 7: 1.0 Mult 8: 0.0 Offset 9: Do (P86) ;turn on CE8 1: 41 Set Port 1 High 10: Excitation with Delay (P22) ;Delay of 1 second 1: 1 Ex Channel 2: 0000 Delay W/Ex (0.01 sec units) 3: 100 Delay After Ex (0.01 sec units) 4: 0 mV Excitation

17

229 Heat Dissipation Matric Water Potential Sensor

11: Thermocouple Temp (DIFF) (P14) ;Read thermocouple after 1 second of heating 1: 1 Reps 2: 21 10 mV, 60 Hz Reject, Slow Range 3: 1 DIFF Channel 4: 1 Type T (Copper-Constantan) 5: 2 Ref Temp (Deg. C) Loc [ Tref_C ] 6: 19 -- Loc [ T1s_1 ] 7: 1.0 Mult 8: 0.0 Offset 12: Excitation with Delay (P22) ;delay 29 more seconds 1: 1 Ex Channel 2: 0 Delay W/Ex (0.01 sec units) 3: 2900 Delay After Ex (0.01 sec units) 4: 0000 mV Excitation 13: Thermocouple Temp (DIFF) (P14) ;read temperature after 30 seconds heating 1: 1 Reps 2: 21 10 mV, 60 Hz Reject, Slow Range 3: 1 DIFF Channel 4: 1 Type T (Copper-Constantan) 5: 2 Ref Temp (Deg. C) Loc [ Tref_C ] 6: 35 -- Loc [ T30s_1 ] 7: 1.0 Mult 8: 0.0 Offset 14: Do (P86) ;turn off CE8 1: 51 Set Port 1 Low 15: Z=X-Y (P35) ;Calculate temperature rise 1: 35 -- X Loc [ T30s_1 ] 2: 19 -- Y Loc [ T1s_1 ] 3: 5 -- Z Loc [ dT_1 ] 16: End (P95) ;end of loop 17: Do (P86) ;Turn off AM16/32 1: 52 Set Port 2 Low 18: End (P95) ;End of then do statement 19: If Flag/Port (P91) 1: 11 Do if Flag 1 is High 2: 10 Set Output Flag High (Flag 0) 20: Set Active Storage Area (P80) 1: 1 Final Storage Area 1 2: 229 Array ID 21: Real Time (P77) 1: 1220 Year,Day,Hour/Minute (midnight = 2400)

18

229 Heat Dissipation Matric Water Potential Sensor

22: Sample (P70) ;Sample 16 initial soil temperature readings 1: 16 Reps 2: 3 Loc [ Tinit_1 ] 23: Sample (P70) ;Sample 16 delta T readings 1: 16 Reps 2: 51 Loc [ dT_1 ] 24: Do (P86) 1: 21 Set Flag 1 Low

6. Calibration 6.1 General

The heat transfer properties of a 229 sensor depend both on the thermal properties of the various sensor materials and on the interfaces between the different materials. Heat transfer between the stainless steel needle containing the heating element and thermocouple and the ceramic material depends on the density of points-of-contact between two different materials. Heat transfer also depends on the arrangement of the wires in the hypodermic needle and the amount of contact between the needle and the ceramic. The uncontrollable variability in heat transfer properties warrants individual calibration of the 229 sensors.

The calibration used to relate temperature increase and the soil water potential is strictly empirical, and the functional expression of the relationship can take several forms. The most commonly used function is:

)*exp( += T [1]

with the soil water potential, exp the exponential function, T the temperature increase during the chosen heating period of time, the slope and the intercept.

The relationship between the natural logarithm of soil water tension and the temperature increase is linear which simplifies derivation of the calibration function.

+= T*|)ln(| [2]

(Soil water potential is a negative value and becomes more negative as soil dries.) Figure 6-1 is a typical calibration and the data set is easily described with linear regression. A power function works well in applications when calibration is needed for -500 kPa matric potential -10 kPa. A power function calibration has the form || = a*(T)b with the multiplier, a, and exponent, b, as fitted parameters.

A variety of calibration methods are suitable. The sole requirement is that the water potential of the medium surrounding the sensor must be known. Either the applied potential can be controlled at a specified value or the water potential can be independently measured. Hanging water columns and pressure plate extractors are typically used. Several data values which

19

229 Heat Dissipation Matric Water Potential Sensor

correspond to the water potential expected during sensor use should be included in the calibration

2

2.5

3

3.5

4

4.5

5

5.5

1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6

Temperature rise,T(30s) - T(1s), deg-C

ln (|m

at ric

w at

er p

ot en

tia l|)

(k Pa

)

FIGURE 6-1. Data Points (x) and Regression for Typical Calibration

6.2 Normalized Temperature Change and Correction for Soil Temperature

6.2.1 Normalized Temperature Change The effect of sensor-to-sensor variability can be reduced by using normalized temperature increase, normT , described by

wetdry

dry norm TT

TT T

= [3]

where dryT is the change in temperature during measurement when the 229

sensor is dry, wetT is the change in temperature during measurement when

the 229 sensor is fully saturated and T is the change in temperature during the measurement. The range of normT will be 0 to 1 with normT equal 0 for dry soil and 1 for saturated soil.

The wetT value requires full saturation of the ceramic. The ceramic portion of the sensor must be immersed in water while under vacuum to remove all air from the ceramic. Measurements at the factory show that complete saturation will occur after 3-4 hours at 0.7 atmospheres (70 kPa) vacuum. However, the arrangement of the sensors in the water during vacuum extraction may hamper movement of air from the ceramic. When tapping or jarring the container shows no release of bubbles, the ceramic is saturated.

20

229 Heat Dissipation Matric Water Potential Sensor

The dryT value requires that the ceramic be as dry as possible. Sensors can be dried with desiccant or in an oven at temperature no greater than 60 C. Temperatures greater than 60 C may damage the sensor cable.

Reece (1996) suggested that inverse thermal conductivity can also be used as a normalization technique but work by Campbell Scientific has not shown significant advantage for this method over normalization as described by equation [3].

A calibration equation using normT and having a form similar to equation [2] is

+= normT*|)ln(| . [4]

The slope of equation [2] will be positive while the slope of equation [4] will be negative.

6.2.2 Correction for Soil Temperature The heat dissipation method of matric potential measurement is sensitive to temperature and correction for the temperature dependence may be necessary to maintain accuracy of the measurement. If the soil temperature when the matric potential measurements are made is close to the temperature at the time of sensor calibration there is no need for correction.

The 229 measurement method uses heat transfer away from a heated line source and the heat transfer depends on the thermal conductivity of the ceramic. The thermal conductivity of the ceramic depends on the combination of the conductivities of water, vapor and solid constituents. The vapor component has a strong temperature dependence and consequently imparts sensitivity of the measurement to temperature. The sensitivity is related to the difference between the temperature of the sensor at time of measurement (soil temperature) and the temperature of the sensor during calibration. Figures 6-2 and 6-3 show the response of the matric potential measurement for a range of temperatures when the calibration temperature is 20 C.

21

229 Heat Dissipation Matric Water Potential Sensor

0 500 1000 1500 2000 600

400

200

0

200

400

600

10 degrees C 16 degrees C 18 degrees C 22 degrees C 24 degrees C 30 degrees C

matric potential (-kPa)

er ro

r (-

kP a)

FIGURE 6-2. Measurement error for range of soil temperatures and wide range of matric potential.

22

229 Heat Dissipation Matric Water Potential Sensor

0 100 200 300 400 500 200

100

0

100

200

10 degrees C 16 degrees C 18 degrees C 22 degrees C 24 degrees C 30 degrees C

matric potential (-kPa)

er ro

r (-

kP a)

FIGURE 6-3. Measurement error for range of soil temperatures and wetter range of matric potential.

A temperature correction for the difference in temperature at time of calibration and time of measurement is provided in the work of Flint et al., 2002. To implement the correction, normalized temperature must be used for the calibration variable. Normalized temperature is as defined in equation [3].

The correction procedure is an iterative method. Examples of implementing the iterative routine with dataloggers are given in programming examples 5 and 6. The general sequence for using the temperature correction method is:

1. Determine dryT and wetT prior to calibration and with sensor

conditions as described in 6.2.1. The recommended T is the sensor temperature after 30 seconds of heating minus the sensor temperature after 1 second of heating.

23

229 Heat Dissipation Matric Water Potential Sensor

2. With the sensor in place, use the T from the in situ measurement

along with the dryT , wetT values for the particular sensor to

calculate normT .

3. Implement the iterative temperature correction as presented in datalogger example program #4 to obtain a corrected normT .

4. Use the corrected normT in the calibration equation, e.g. equation [4].

6.3 Using Pressurized Extraction Methods Pressurized extraction methods use a porous material (typically ceramic or stainless steel) to separate the pressurized soil sample from atmospheric pressure. One side of the porous material is in contact with the soil sample and the other side is at atmospheric pressure. A simple configuration is a cylinder which has the porous material as a bottom and a solid cap at the top which provides a pressure-tight seal after the soil sample has been placed on the porous bottom. This allows pressurization of the soil sample which will force water from the soil. The air entry pressure of the porous material is dependent on the effective pore diameter and must be greater than the maximum pressure applied during calibration. When a specified pressure is applied, all soil water at water potentials greater than -1*(applied pressure) will move through the soil sample and through the porous bottom. The time required for the soil water to leave the sample system depends on the pressure gradient and the hydraulic conductivity of the soil and the porous bottom plate.

Equilibration of water potential throughout the system must be attained or the accuracy of the water potential measurement using the derived calibration will be reduced. The simplest way to confirm equilibration is to repeat the measurements at a given applied pressure until readings do not change.

6.4 General Description of Calibration/Measurement Process using Pressure Plate Extractor

A pressure plate extractor consists of a high-pressure vessel, a porous plate and tubing to remove soil water from the soil sample. A ring with diameter slightly less than the diameter of the ceramic plate can be used to hold the soil in which the 229 sensors are buried. Smaller rings can also be used. The porous plate, soil and 229 sensors must be thoroughly saturated at the beginning of the calibration routine. Complete saturation of the ceramic plate and 229 sensors is better achieved by applying vacuum. The number of sensors that can be calibrated in a single pressure vessel will depend on the 229 cable lengths because of the limited space in the vessel.

The general calibration sequence is:

1. The extractor plate, soil and sensors are saturated and positioned in the pressure vessel.

2. The lowest calibration pressure chosen by the user (>10 kPa) is applied and the soil solution, which is held by the soil at an energy level less than that applied, is allowed to leave the pressurized system.

24

229 Heat Dissipation Matric Water Potential Sensor

3. Measurements of sensor temperature response are made periodically to determine if equilibration is attained. This will require depressurization of the pressure vessel if a pressure-tight feedthrough is not used. Prior to depressurization, it is important that the effluent hose be blocked by clamping or other method to prevent solution from re-entering the soil and sensors.

4. When equilibration is attained, the effluent hose is blocked, the vessel is depressurized and 229 measurements are recorded.

5. The next calibration pressure is then applied and the process repeated.

Pressure-tight bulkhead connectors are available for some pressure vessels. Determining whether equilibrium has been reached is simplified when using a feedthrough connector since the pressure vessel doesnt have to be depressurized and opened for each reading. A temporary connector can be used to disconnect the datalogger from the other components installed in the pressure vessel.

Pressurized readings can be used for determining equilibrium but should not be used for calibration data. Thermal properties are affected by pressure and calibration data should be collected at the same pressure the sensors will be usedin most cases this is atmospheric pressure.

6.4.1 Wiring for Calibration using Pressure Plate Extractor The wiring arrangement of Figure 6-4 depicts a datalogger with an AM16/32 multiplexer, a CE4 current source and a thermistor being used in a typical 229 calibration arrangement. This is similar to the wiring for program example #2 and #4 (see Section 5). All components except the datalogger can be placed inside a 5 bar pressure vessel. If this method is used, the electronic components in the pressure vessel should be protected against moisture damage. Place the components in a container such as plastic bag with desiccant.

25

229 Heat Dissipation Matric Water Potential Sensor

FIGURE 6-4. Datalogger and Peripheral Connections for 229 Calibration

7. Maintenance The 229 does not require maintenance after it is installed in the soil. The datalogger, current excitation module, and multiplexer, if used, should be kept in a weatherproof enclosure. Periodic replacement of the desiccant in the enclosure is required to keep the electronics dry and free of corrosion.

26

229 Heat Dissipation Matric Water Potential Sensor

8. Troubleshooting Symptom Possible Cause Action

Temperature reading is offscale (-6999 or NAN)

Thermocouple wire not connected to correct datalogger channel

Check program to see which differential input channel 229 should be connected to and verify that it has a good connection to that channel

Break in thermocouple wire Use ohm-meter to measure resistance between blue and red wires. The reading in ohms should be approximately the cable length in feet. An open circuit indicates a break in the wire

Multiplexer not operating properly

Make sure that multiplexer has 12V between 12V and GND terminals.

Check for a good electrical connection on the wires that connect RES and CLK to datalogger control ports.

Check for a good electrical connection on the wires going from the common channels to the datalogger and the current excitation module.

Check program to make sure that the control port connected to RES is being set high and the control port connected to CLK is being pulsed.

DeltaT reading close to zero

Heater wire broken or not properly connected

Check resistance between terminal screws for green and black wires. It should read 34 - 40 ohms.

Current excitation module not turning on

Check for 12V between +12V and ground terminals

Check for good electrical connection on wire connecting CTRL with datalogger control port

Check program to make sure that control port connected to CTRL is being set high

Temperature decreases during heating

Thermocouple wires reversed

Make sure blue wire is on the high side of the differential input channel and red is on the low side

Readings for first sensor on multiplexer are all right, but all others read zero

All readings are being written to the same input location (Edlog dataloggers)

Check program to make sure measurement instructions in the multiplexer loop are indexed (-- next to the input location number. Press F4 or C to toggle the --)

27

229 Heat Dissipation Matric Water Potential Sensor

9. References Flint, A. L., G. S. Campbell, K. M. Ellett, and C. Calissendorff. 2002. Calibration and Temperature Correction of Heat Dissipation Matric Potential Sensors. Soil Sci. Soc. Am. J. 66:14391445.

Reece, C.F. 1996. Evaluation of a line heat dissipation sensor for measuring soil matric potential. Soil Sci. Soc. Am. J. 60:10221028.

28

Campbell Scientific Companies

Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES

www.campbellsci.com info@campbellsci.com

Campbell Scientific Africa Pty. Ltd. (CSAf) PO Box 2450

Somerset West 7129 SOUTH AFRICA

www.csafrica.co.za cleroux@csafrica.co.za

Campbell Scientific Australia Pty. Ltd. (CSA) PO Box 8108

Garbutt Post Shop QLD 4814 AUSTRALIA

www.campbellsci.com.au info@campbellsci.com.au

Campbell Scientific (Beijing) Co., Ltd. 8B16, Floor 8 Tower B, Hanwei Plaza

7 Guanghua Road Chaoyang, Beijing 100004

P.R. CHINA www.campbellsci.com info@campbellsci.com.cn

Campbell Scientific do Brasil Ltda. (CSB) Rua Apinags, nbr. 2018 Perdizes CEP: 01258-00 So Paulo SP

BRASIL www.campbellsci.com.br vendas@campbellsci.com.br

Campbell Scientific Canada Corp. (CSC) 14532 131 Avenue NW Edmonton AB T5L 4X4

CANADA www.campbellsci.ca dataloggers@campbellsci.ca

Campbell Scientific Centro Caribe S.A. (CSCC) 300 N Cementerio, Edificio Breller

Santo Domingo, Heredia 40305 COSTA RICA

www.campbellsci.cc info@campbellsci.cc

Campbell Scientific Ltd. (CSL) Campbell Park

80 Hathern Road Shepshed, Loughborough LE12 9GX

UNITED KINGDOM www.campbellsci.co.uk sales@campbellsci.co.uk

Campbell Scientific Ltd. (CSL France) 3 Avenue de la Division Leclerc

92160 ANTONY FRANCE

www.campbellsci.fr info@campbellsci.fr

Campbell Scientific Ltd. (CSL Germany) Fahrenheitstrae 13

28359 Bremen GERMANY

www.campbellsci.de info@campbellsci.de </

Manualsnet FAQs

If you want to find out how the 229 Campbell Scientific works, you can view and download the Campbell Scientific 229 Sensor Instruction Manual on the Manualsnet website.

Yes, we have the Instruction Manual for Campbell Scientific 229 as well as other Campbell Scientific manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Instruction Manual should include all the details that are needed to use a Campbell Scientific 229. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Campbell Scientific 229 Sensor Instruction Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Campbell Scientific 229 Sensor Instruction Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Campbell Scientific 229 Sensor Instruction Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Campbell Scientific 229 Sensor Instruction Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Campbell Scientific 229 Sensor Instruction Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.