CS500 Temperature and Relative Humidity Probe

Revision: 7/04

C o p y r i g h t ( c ) 1 9 9 5 - 2 0 0 4 C a m p b e l l S c i e n t i f i c , I n c .

Warranty and Assistance The CS500 TEMPERATURE AND RELATIVE HUMIDITY PROBE is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC, INC.'s option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CAMPBELL SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such products by surface carrier prepaid. This warranty shall not apply to any CAMPBELL SCIENTIFIC, INC. products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CAMPBELL SCIENTIFIC, INC. is not liable for special, indirect, incidental, or consequential damages.

Products may not be returned without prior authorization. The following contact information is for US and International customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com to determine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. CAMPBELL SCIENTIFIC's shipping address is:

CAMPBELL SCIENTIFIC, INC. RMA#_____ 815 West 1800 North Logan, Utah 84321-1784

CAMPBELL SCIENTIFIC, INC. does not accept collect calls.

i

CS500 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat bookmarks tab for links to specific sections.

1. General ........................................................................1

2. Specifications .............................................................1 2.1 Temperature Sensor ..................................................................................1 2.2 Relative Humidity Sensor .........................................................................2

3. Installation...................................................................2

4. Wiring ..........................................................................6

5. Example Programs .....................................................6 5.1 Example for CR1000 ................................................................................7 5.2 Example for CR10X .................................................................................7

6. Long Lead Lengths .....................................................8

7. Absolute Humidity ......................................................9

8. Maintenance ..............................................................11

9. References ................................................................11

Figures 1. CS500 and 41301 Radiation Shield on a CM6/CM10 Tripod Mast ...........3 2. CS500 and 41303 Radiation Shield ............................................................4 3. CS500 and 41003 Radiation Shield on a CM6/CM10 Tripod Mast ...........4 4. Radiation Shield, CS500, and 41381 Adapter ............................................5 5. CS500 Wiring .............................................................................................5

Tables 1. Datalogger Connections..............................................................................6 2. Calibration for Temperature .......................................................................6 3. Calibration for Relative Humidity...............................................................6 4. Wiring for CR1000 and CR10X Examples.................................................7 5. CR10(X) Wiring for Example 1..................................................................9

This is a blank page.

1

CS500 Temperature and Relative Humidity Probe

1. General The CS500 Temperature and Relative Humidity probe contains a Platinum Resistance Temperature detector (PRT) and a Vaisala INTERCAP capacitive relative humidity sensor.

The -L option on the model CS500 Temperature and Relative Humidity probe (CS500-L) indicates that the cable length is user specified. This manual refers to the sensor as the CS500.

2. Specifications Operating Temperature: -40C to +60C

Storage Temperature: -40C to +80C

Probe Length: 6.8 cm (2.66 in.)

Probe Body Diameter: 1.2 cm (0.47 in.)

Filter: 0.2 m Teflon membrane

Filter Diameter: 1.2 cm (0.47 in.)

Housing Material: ABS Plastic

Power Consumption: <2 mA

Supply Voltage: 7 to 28 VDC

Settling Time after power is switched on: 1 second

2.1 Temperature Sensor Sensor: 1000 PRT, DIN 43760B

Temperature Measurement Range: -40C to +60C

Temperature Output Signal range: 0 to 1.0 VDC

CS500 Temperature and Relative Humidity Probe

2

Temperature Accuracy:

0

0.4

0.8

1.2

1.6

-0.4

-0.8

-1.2

-1.6 -20-40 0

Temperature ( C)o

( C

) Er

ro r

o

20 40 60

2.2 Relative Humidity Sensor Sensor: INTERCAP

Relative Humidity Measurement Range: 0 to 100% non-condensing

RH Output Signal Range: 0 to 1.0 VDC

Accuracy at 20C unspecified (0 to 10% Relative Humidity) 3% RH (10 to 90% Relative Humidity) 6% RH (90 to 100% Relative Humidity)

Temperature Dependence of Relative Humidity Measurement:

0 0.5 1.0 1.5

-0.5 -1.0 -1.5

10 20 30 40 50 60 70 80 90 100

RH (%)

( % R

H /

)

Te m

pe ra

tu re

D ep

en de

nc e

60 Co

-40 Co

Co

Typical Long Term Stability: Better than 1% RH per year

Response Time (at 20C, 90% response to a steep change in humidity): 15 seconds with membrane filter

3. Installation The CS500 must be housed inside a solar radiation shield when used in the field. The 41303 6-Plate Radiation Shield (Figure 1) mounts to a CM6/CM10 tripod or UT10 tower. The CS500 is held within the 41301 by a mounting clamp (Figure 2).

The 41003 10-Plate Radiation Shield (Figure 3) mounts to a CM6/CM10 tripod. The UT12VA 12-Plate Radiation Shield mounts to a UT10 or UT30 tower with the UT018 horizontal mounting arm.

CS500 Temperature and Relative Humidity Probe

3

The CS500 is held in place, within the 41003 or UT12VA Radiation Shield, via an adapter, Model 41381. The 41381 adapter is threaded onto the bottom of the CS500 (Figure 4). The 41004 12-Plate Radiation Shield, used with 207 probes, can be converted to a 41002 with P/N 6638.

The black outer jacket of the cable is Santoprene rubber. This compound was chosen for its resistance to temperature extremes, moisture, and UV degradation. However, this jacket will support combustion in air. It is rated as slow burning when tested according to U.L. 94 H.B. and will pass FMVSS302. Local fire codes may preclude its use inside buildings.

FIGURE 1. CS500 and 41301 Radiation Shield on a CM6/CM10 Tripod Mast

NOTE

CS500 Temperature and Relative Humidity Probe

4

FIGURE 2. CS500 and 41303 Radiation Shield

FIGURE 3. CS500 and 41003 Radiation Shield on a CM6/CM10 Tripod Mast

Mounting Clamp

CS500 Temperature and Relative Humidity Probe

5

FIGURE 4. Radiation Shield, CS500, and 41381 Adapter

FIGURE 5. CS500 Wiring

41003 10-Plate Radiation Shield

Lock nut tightens against adapter

CS500

41381 Adapter

Black Temperature Signal

Brown Relative Humidity Signal

Green Signal & Power Reference

Red Power

Clear Shield

Tripod Mast

Internal threads secure the CS500 to adapter

CS500 Temperature and Relative Humidity Probe

6

TABLE 1. Datalogger Connections

Description Color CR23X/CR1000 CR10(X), CR510 21X, CR7 Temperature Black Single-Ended Input Single-Ended Input Single-Ended Input

Relative Humidity Brown Single-Ended Input Single-Ended Input Single-Ended Input Signal & Power

Reference Green G G

Power Red 12 V 12 V 12 V Shield Clear G

4. Wiring Connections to Campbell Scientific dataloggers are given in Table 1. The probe is measured by two single-ended analog input channels, one for temperature and one for relative humidity.

Always connect the Green lead to the datalogger first, followed by the Black, Brown, and Clear leads. Connect the Red (Power) lead last.

5. Example Programs This section is for users who write their own datalogger programs. A datalogger program to measure this sensor can be created using Campbell Scientifics Short Cut Program Builder Software. You do not need to read this section to use Short Cut.

The temperature and relative humidity signals from the CS500 are measured using two single-ended analog measurements (Instruction 1).

The probe output scale is 0 to 1000 millivolts for the temperature range of - 40C to +60C and for the relative humidity range of 0 to 100%. Tables 2 and 3 provide calibration information for temperature and relative humidity.

TABLE 2. Calibration for Temperature

Units Multiplier (degrees mV-1)

Offset (degrees)

Celsius 0.1 -40 Fahrenheit 0.18 -40

TABLE 3. Calibration for Relative Humidity

Units Multiplier (% mV-1)

Offset (%)

Percent 0.1 0 Fraction 0.001 0

CAUTION

CS500 Temperature and Relative Humidity Probe

7

TABLE 4. Wiring for CR1000 and CR10X Examples

Description Color CR1000 CR10(X) Temperature Black SE 1 SE 3 (2H)

Relative Humidity Brown SE 2 SE 4 (2L) Signal & Power Reference Green G G

Power Red 12 V 12 V Shield Clear G

5.1 Example for CR1000

'CR1000 'Created by SCWIN (2.1)

Public AirTC Public RH

DataTable(Table1,True,-1) DataInterval(0,60,Min,0) Average(1,AirTC,IEEE4,0) Sample(1,RH,IEEE4)

EndTable

BeginProg Scan(5,Sec,1,0)

'CS500 Temperature & Relative Humidity Sensor measurements AirTC and RH: VoltSE(AirTC,1,mV2500,1,0,0,_60Hz,0.1,-40.0) VoltSE(RH,1,mV2500,2,0,0,_60Hz,0.1,0) If (RH>100) And (RH<108) Then RH=100 CallTable(Table1)

NextScan EndProg

5.2 Example for CR10X

;Measure the CS500 temperature. ; 01: Volt (SE) (P1)

1: 1 Reps 2: 5 2500 mV Slow Range ;CR500 (2500 mV); CR23X (1000 mV); 21X,

CR7 (5000 mV) 3: 3 SE Channel ;Black wire (SE 3), Green wire (G) 4: 1 Loc [ T_C ] 5: .1 Mult ;See Table 2 for alternate multipliers 6: -40 Offset ;See Table 2 for alternate offsets

CS500 Temperature and Relative Humidity Probe

8

;Measure the CS500 relative humidity. ; 02: Volt (SE) (P1)

1: 1 Reps 2: 5 2500 mV Slow Range ;CR500 (2500 mV); CR23X (1000 mV); 21X,

CR7 (5000 mV) 3: 4 SE Channel ;Brown wire (SE 4), Green wire (G) 4: 3 Loc [ RH_pct ] 5: .1 Mult ;See Table 3 for alternate multipliers 6: 0 Offset

;Limit the maximum relative humidity to 100%. ; 03: If (X<=>F) (P89)

1: 3 X Loc [ RH_pct ] 2: 3 >= 3: 100 F 4: 30 Then Do

04: Z=F (P30) 1: 100 F 2: 0 Exponent of 10 3: 3 Z Loc [ RH_pct ]

05: End (P95)

6. Long Lead Lengths Long lead lengths cause errors in the measured temperature and relative humidity. The approximate error in temperature and relative humidity is 0.35C and 0.35% per 100 feet of cable length, respectively.

When long lead lengths are required and the above errors in temperature and relative humidity are unacceptable, use the HMP45C temperature and humidity probe.

Understanding the following details are not required for the general operation of the CS500 with Campbell Scientifics dataloggers. The signal reference and the power ground (black) are the same lead in the CS500. When the CS500 temperature and relative humidity are measured, both the signal reference and power ground are connected to ground at the datalogger. The signal reference/power ground lead serves as the return path for 12 V. There will be a voltage drop along this lead because the wire itself has resistance. The CS500 draws approximately 2 mA when it is powered. The wire used in the CS500 (P/N 9720) has resistance of 17.5 /1000 feet. Using Ohms law, the voltage drop (Vd), along the signal reference/power ground lead, is given by Eq. (1).

V I Rd = = =

mA ft

mV ft 2 17 5 1000 35 1000

. (1)

CS500 Temperature and Relative Humidity Probe

9

This voltage drop will raise the apparent temperature and relative humidity because the difference between the signal and signal reference, at the datalogger, has increased by Vd. The approximate error in temperature and relative humidity is 0.35C and 0.35% per 100 feet of cable length, respectively.

7. Absolute Humidity The CS500 measures the relative humidity. Relative humidity is defined by the equation below:

RH e e

100 s

= (2)

where RH is the relative humidity, e is the vapor pressure in kPa , and es is the saturation vapor pressure in kPa. The vapor pressure, e, is an absolute measure of the amount of water vapor in the air and is related to the dew point temperature. The saturation vapor pressure is the maximum amount of water vapor that air can hold at a given air temperature. The relationship between dew point and vapor pressure, and air temperature and saturation vapor pressure are given by Goff and Gratch (1946), Lowe (1977), and Weiss (1977).

When the air temperature increases, so does the saturation vapor pressure. Conversely, a decrease in air temperature causes a corresponding decrease in saturation vapor pressure. It follows then from Eq. (2) that a change in air temperature will change the relative humidity, without causing a change in absolute humidity.

For example, for an air temperature of 20C and a vapor pressure of 1.17 kPa, the saturation vapor pressure is 2.34 kPa and the relative humidity is 50%. If the air temperature is increased by 5C and no moisture is added or removed from the air, the saturation vapor pressure increases to 3.17 kPa and the relative humidity decreases to 36.9%. After the increase in air temperature, the air can hold more water vapor. However, the actual amount of water vapor in the air has not changed. Thus, the amount of water vapor in the air, relative to saturation, has decreased.

Because of the inverse relationship between relative humidity and air temperature, finding the mean relative humidity is meaningless. A more useful quantity is the mean vapor pressure. The mean vapor pressure can be computed on-line by the datalogger (Example 1).

TABLE 5. CR10(X) Wiring for Example 1

Description Color CR10(X) Temperature Black SE 3 (2H)

Relative Humidity Brown SE 4 (2L) Signal & Power Reference Green G

Power Red 12 V Shield Clear G

CS500 Temperature and Relative Humidity Probe

10

Example 1. Sample CR10(X) Program that Computes Vapor Pressure and Saturation Vapor Pressure

;Measure the CS500 temperature. ; 01: Volt (SE) (P1)

1: 1 Reps 2: 5 2500 mV Slow Range ;CR500 (2500 mV); CR23X (1000 mV); 21X,

CR7 (5000 mV) 3: 3 SE Channel ;Black wire (SE 3), Green wire (G) 4: 1 Loc [ T_C ] 5: .1 Mult ;See Table 2 for alternate multipliers 6: -40 Offset ;See Table 2 for alternate offsets

;Measure the CS500 relative humidity. ; 02: Volt (SE) (P1)

1: 1 Reps 2: 5 2500 mV Slow Range ;CR500 (2500 mV); CR23X (1000 mV); 21X,

CR7 (5000 mV) 3: 4 SE Channel ;Brown wire (SE 4), Green wire (G) 4: 2 Loc [ RH_frac ] 5: .001 Mult ;See Table 3 for alternate multipliers 6: 0 Offset

;Limit the maximum value of relative humidity ;to 1 (expressed as a fraction). ; 03: If (X<=>F) (P89)

1: 2 X Loc [ RH_frac ] 2: 3 >= 3: 1 F 4: 30 Then Do

04: Z=F (P30) 1: 1 F 2: 0 Exponent of 10 3: 2 Z Loc [ RH_frac ]

05: End (P95)

;Compute the saturation vapor pressure in kPa. ;The temperature must be in degrees Celsius. ; 06: Saturation Vapor Pressure (P56)

1: 1 Temperature Loc [ T_C ] 2: 3 Loc [ e_sat ]

CS500 Temperature and Relative Humidity Probe

11

;Compute the vapor pressure in kPa. ;Relative humidity must be a fraction. ; 07: Z=X*Y (P36)

1: 3 X Loc [ e_sat ] 2: 2 Y Loc [ RH_frac ] 3: 4 Z Loc [ e ]

8. Maintenance The CS500 Probe requires minimal maintenance. Check monthly to make sure the radiation shield is free from debris. The white screen at the tip of the probe should also be checked for contaminants.

When installed in close proximity to the ocean or other bodies of salt water (e.g., Great Salt Lake), a coating of salt (mostly NaCl) may build up on the radiation shield, sensor, filter and even the chip. NaCl has an affinity for water. The humidity over a saturated NaCl solution is 75%. A buildup of salt on the filter or chip will delay or destroy the response to atmospheric humidity.

The filter can be rinsed gently in distilled water. If necessary, the chip can be removed and rinsed as well. Do not scratch the chip while cleaning.

The offset and gain on the CS500 electronics can not be adjusted as part of a recalibration. Replace the RH chip as needed.

9. References Goff, J. A. and S. Gratch, 1946: Low-pressure properties of water from -160

to 212F, Trans. Amer. Soc. Heat. Vent. Eng., 51, 125-164.

Lowe, P. R., 1977: An approximating polynomial for the computation of saturation vapor pressure, J. Appl. Meteor., 16, 100-103.

Weiss, A., 1977: Algorithms for the calculation of moist air properties on a hand calculator, Amer. Soc. Ag. Eng., 20, 1133-1136.

CS500 Temperature and Relative Humidity Probe

12

This is a blank page.

This is a blank page.

Campbell Scientific Companies

Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES

www.campbellsci.com info@campbellsci.com

Campbell Scientific Africa Pty. Ltd. (CSAf) PO Box 2450

Somerset West 7129 SOUTH AFRICA

www.csafrica.co.za sales@csafrica.co.za

Campbell Scientific Australia Pty. Ltd. (CSA) PO Box 444

Thuringowa Central QLD 4812 AUSTRALIA www.campbellsci.com.au info@campbellsci.com.au

Campbell Scientific do Brazil Ltda. (CSB) Rua Luisa Crapsi Orsi, 15 Butant

CEP: 005543-000 So Paulo SP BRAZIL www.campbellsci.com.br

suporte@campbellsci.com.br

Campbell Scientific Canada Corp. (CSC) 11564 - 149th Street NW

Edmonton, Alberta T5M 1W7 CANADA

www.campbellsci.ca dataloggers@campbellsci.ca

Campbell Scientific Ltd. (CSL) Campbell Park

80 Hathern Road Shepshed, Loughborough LE12 9GX

UNITED KINGDOM www.campbellsci.co.uk sales@campbellsci.co.uk

Campbell Sc

Manualsnet FAQs

If you want to find out how the CS500-L Campbell Scientific works, you can view and download the Campbell Scientific CS500-L Probe Instruction Manual on the Manualsnet website.

Yes, we have the Instruction Manual for Campbell Scientific CS500-L as well as other Campbell Scientific manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Instruction Manual should include all the details that are needed to use a Campbell Scientific CS500-L. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Campbell Scientific CS500-L Probe Instruction Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Campbell Scientific CS500-L Probe Instruction Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Campbell Scientific CS500-L Probe Instruction Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Campbell Scientific CS500-L Probe Instruction Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Campbell Scientific CS500-L Probe Instruction Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.