PWC100 Present Weather Sensor Calibrator

Revision: 3/12

C o p y r i g h t 2 0 0 8 - 2 0 1 2 C a m p b e l l S c i e n t i f i c , I n c .

Warranty PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (Campbell) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless otherwise specified in the corresponding Campbell pricelist or product manual. Products not manufactured, but that are re-sold by Campbell, are warranted only to the limits extended by the original manufacturer. Batteries, fine-wire thermocouples, desiccant, and other consumables have no warranty. Campbell's obligation under this warranty is limited to repairing or replacing (at Campbell's option) defective products, which shall be the sole and exclusive remedy under this warranty. The customer shall assume all costs of removing, reinstalling, and shipping defective products to Campbell. Campbell will return such products by surface carrier prepaid within the continental United States of America. To all other locations, Campbell will return such products best way CIP (Port of Entry) INCOTERM 2010, prepaid. This warranty shall not apply to any products which have been subjected to modification, misuse, neglect, improper service, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied. The warranty for installation services performed by Campbell such as programming to customer specifications, electrical connections to products manufactured by Campbell, and product specific training, is part of Campbells product warranty. CAMPBELL EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Campbell is not liable for any special, indirect, incidental, and/or consequential damages.

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com to determine which Campbell Scientific company serves your country.

To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 227-9000. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. Campbell Scientific's shipping address is:

CAMPBELL SCIENTIFIC, INC. RMA#_____ 815 West 1800 North Logan, Utah 84321-1784

For all returns, the customer must fill out a "Statement of Product Cleanliness and Decontamination" form and comply with the requirements specified in it. The form is available from our web site at www.campbellsci.com/repair. A completed form must be either emailed to repair@campbellsci.com or faxed to (435) 227-9106. Campbell Scientific is unable to process any returns until we receive this form. If the form is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customer's expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.

PWC100 Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to specific sections.

1. Introduction..................................................................1

2. Operation and Use .......................................................1 2.1 Operating Guidelines ................................................................................1 2.2 Initial Configuration of the PWS100 for use of the Calibrator.................2

3. PWS Calibration...........................................................2 3.1 Installing and Connecting up the Calibrator .............................................3 3.2 Starting the Tests ......................................................................................4

4. Cleaning Guide ..........................................................10

5. Calibrator Safety Guidelines.....................................10

6. Electrical Parameters ................................................11 6.1 PWC100 Voltages and Currents.............................................................11 6.2 Earth, 0 V and Transient Protection .......................................................11

Appendix

A. Assembly Guidelines .............................................. A-1 A.1 Assembly Guidelines .......................................................................... A-1 A.2 Connector Pin Outs............................................................................. A-2

Figures 2-1. PWS100 calibration disk constants .........................................................2 3-1. The PWS100 enclosure showing the LEMO connector used by the

calibrator ..............................................................................................3 A-1. Attaching the velocity disk ................................................................ A-1 A-2. Attaching the visibility disk............................................................... A-1 A-3. Attachment of the velocity calibrator ................................................ A-2

i

PWC100 Present Weather Sensor Calibrator 1. Introduction

The PWC100 is a customer field calibration validation device for the PWS100.

The PWC100 calibrator system simulates particle and visibility data for the PWS100 allowing verification of the calibration constants held within the system.

The following features are checked by the calibrator:

Particle velocity

Particle size detection

Laser beam uniformity

Visibility range

Component aging

The PWC100 will give a go/no go indication of the PWS100 calibration to show whether the instrument is within normal operating limits. If required, the PWC100 can be used to partially recalibrate the PWS100. That method is not described in this manual. Please contact Campbell Scientific for more details or refer to the PWS100 manual.

2. Operation and Use 2.1 Operating Guidelines

Please ensure that there are no obstructions between the PWS sensor heads before the tests are started.

Ambient temperature should be 20C 10C.

The test should not be performed in the following conditions:

Do not perform the calibration when it is raining or during any other precipitation event, such as snow, hail or drizzle.

Do not perform the calibration when the local visibility is below approximately 10,000 meters.

Do not perform the calibration if there is a strong wind that could blow debris into the sensing volume of the PWS.

The PWC100 is not suitable for long term outdoor use and is only designed to be exposed for the duration of the calibration test. Long term exposure to outdoor environmental conditions may cause damage to the unit.

1

PWC100 Present Weather Sensor Calibrator

Care must be taken not to drop or scratch either the velocity or visibility units as this may cause errors that could result in tests failing when they shouldnt.

2.2 Initial Configuration of the PWS100 for use of the Calibrator The visibility calibration disks provided as part of the calibration kit are individually calibrated and have a number printed on them to indicate the measured signal in a reference instrument. Each PWS100 that is to be used with the calibrator has to have that value entered into its configuration so that it is referenced to the properties of that disk. Once set, the PWS100 will retain the calibrator disk information unless the sensor is reset to its factory defaults. The test procedure outlined in the rest of the manual can then be followed repeatedly without changing the settings in the sensors.

For information on how to set calibration disk constants, please refer to Section 7.4.3 of the PWS100 manual. Figure 2-1 provides the screen in which the values are shown. Please note you only need to change the constants; do not attempt to adjust the calibration of the PWS100 if carrying out the standard test procedure.

FIGURE 2-1. PWS100 calibration disk constants

For normal operation it is only necessary to set the visibility constant, as printed on the disk. It is also useful to set the calibrator serial number for reference purposes. The velocity and size constants can be left at the factory defaults of 5.000 and 2.386 as these are fixed values for a standard calibrator.

3. PWS Calibration The calibration check procedure can be broken up into two separate sections. The first section comprises of visibility tests and the second velocity tests. If only a visibility test is required, then the tests can be finished after the first stage and the velocity tests need not be run.

It may be desirable to carry out a check of the sensor before and after cleaning the sensors lenses to determine if the previous readings of the sensor were adversely affected by dirt. This would be at the customers discretion.

2

PWC100 Present Weather Sensor Calibrator

3.1 Installing and Connecting up the Calibrator Remove the existing plastic blanking screws from their positions on the

PWS center block.

Fix the PWC100 visibility calibrator unit to the center position on the PWS using two of the metal countersunk fixing screws provided.

(Connector for the calibrator)

FIGURE 3-1. The PWS100 enclosure showing the LEMO connector used by the calibrator

3

PWC100 Present Weather Sensor Calibrator

3.2 Starting the Tests All PWC100 tests can be performed without the need to access the PWS with a PC. Only once the tests are finished and if changes to the calibration constants are needed, will a PC be required to change the system configuration.

If at any point the LEDs behave unexpectedly, check the status table at the end of this section.

Step 1: Performing the visibility calibration

Check that the PWC100 visibility calibrator unit is located correctly and securely (refer to Appendix A for additional details).

Connect the LEMO connector coming from the velocity unit to the LEMO socket on the PWS (as detailed in Section 3.1). This will let the PWS know you want to start the tests. Press the RESET/STOP button on the PWC100. Within 10 seconds, all three LEDs should light up for approximately 3 seconds. This indicates that the calibration unit has been detected correctly by the PWS100 and that the LEDs are functioning correctly.

LED L and R should now be flashing slowly and all other LEDs should be off.

Once you are ready, press START.

The test is now automated. LED L and R will stay illuminated until this stage of the test is complete. This test will take approximately two minutes.

Upon successful completion of the test, LED L and LED C will be flashing slowly, indicating that the next test is ready to be started.

4

PWC100 Present Weather Sensor Calibrator

Step 2: Performing the visibility maximum limit check

Remove the PWC100 visibility calibration disk and place it back into its box.

Place blanking plug into upper sensor head as shown in diagram above.

At this point, LED L and LED C should be flashing slowly.

Once you are ready, press START.

LED L and LED C will be illuminated for the duration of the test. This test will take approximately two minutes.

Upon successful completion of the test, LED C will be flashing slowly. This indicates the end of the visibility test.

Remember to remove the blanking plug from the PWS before continuing. (This is important!)

At this point, steps three through five can be ignored if only the visibility calibrations are required.

To view the visibility test results alone, press the RESET/DONE button. This will jump the unit to the end of the tests and will display the results in the normal way detailed at the end of the calibration section (this is the only stage in the tests where pressing the RESET/DONE button will not restart the tests).

Please remember to read the final section relating to tidying up and ensuring all screws and fasteners are returned to their original positions.

5

PWC100 Present Weather Sensor Calibrator

Step 3: Position 1: Center position

Attaching the velocity calibration unit

Fix the PWC100 velocity calibrator unit to the center position on the PWS using two of the countersunk fixing screws provided

Check that the velocity calibrator is located correctly and securely.

LED C should now be flashing slowly and all other LEDs should be off.

Once you are ready, press START.

The test is now automated and the disk should start rotating. LED C will stay illuminated until this stage of the test is complete.

Wait for the calibration disk to stop rotating. This test will take approximately two and a half minutes.

Upon successful completion of the test, LED L will be flashing slowly and LED C will be turned off.

6

PWC100 Present Weather Sensor Calibrator

Step 4: Position 2: Left

Move the calibrator into the left hand position by unscrewing the calibrator mounting screws slightly and sliding the calibrator fully to the left. Then retighten the screws once in position.

Visually check that the calibrator is located correctly and securely.

LED L should now be flashing slowly and all other LEDs should be off.

Once you are ready, press START.

The test is now automated and the disk should start rotating. LED L will stay illuminated until this stage of the test is complete.

Wait for the calibration disk to stop rotating. This test will take approximately two and a half minutes.

Upon successful completion of the test, LED R will be flashing slowly and LED L will be turned off.

7

PWC100 Present Weather Sensor Calibrator

Step 5: Position 3: Right

Move the calibrator into the right-hand position by unscrewing the calibrator mounting screws slightly and sliding the calibrator fully to the right. Then retighten the screws once in position.

Visually check that the calibrator is located correctly and securely.

LED R should now be flashing slowly and all other LEDs should be off.

Once you are ready, press START.

The test is now automated and the disk should start rotating. LED R will stay illuminated until this stage of the test is complete.

Wait for the calibration disk to stop rotating. This test will take approximately two and a half minutes.

Final calibration results

The calibration verification has now been completed. All three of the LEDs should be solidly illuminated indicating that the PWS100 is still within its calibration limits.

If all three LEDs are flashing, then the calibration was found to be outside of desired levels and further action needs to be taken.

If any of the calibration checks have failed (all three light flashing after test), then the tests should be re-run after ensuring the disk is clean and the area between the heads is clear of any obstructions. This should remove any chance that the calibrator or another external factor was the cause.

If the PWC100 still indicates a problem, please contact Campbell Scientific for further advice.

8

PWC100 Present Weather Sensor Calibrator

Step 6: Final stage

Remember to reinstall the plastic blanking screws back into the PWS100 before leaving the site.

If a CS215-PWS or other piece of equipment was unplugged in order to perform the tests, then ensure it is reconnected.

Remember to check that the blanking plug has not been left in the PWS.

Place all calibration equipment carefully back into its correct box being careful not to scratch the visibility disk.

Final calibration result table

Result/ LED Status

LED L LED C LED R Description

Test passed ON ON ON If all tests have passed, all the LEDs will be illuminated at the same time

Calibration required

Flashing Flashing Flashing Calibration values are suspect. Do not use the calibration values created from this test

Fault states during tests

If during the calibration cycle, LED C and R start flashing together, then there has been an error either with the communications to the PWS100 or that the disk has stalled or is unable to attain the required speed for calibration.

State LED L LED C LED R Description

Disk stalled - Flashing Flashing PWS calibrator problem

9

PWC100 Present Weather Sensor Calibrator

Quick look LED status table

Ready/complete state Running state

Test state Left Center Right Left Center Right

Visibility test Flash Flash ON ON

Visibility maximum range test

Flash Flash ON ON

Velocity test position 1 (Center)

Flash ON

Velocity test position 2 (Left)

Flash ON

Velocity test position 3 (Right)

Flash ON

Final result states

Calibration passed ON ON ON

Calibration required Flash Flash Flash

Error conditions

Fault occurrence during tests

Flash Flash

4. Cleaning Guide It is important to ensure that the visibility and velocity disks are clean prior to use. Finger prints and small quantities of dust should not impair performance significantly, but it is highly recommended that the disks are kept clean and care is taken to ensure the surfaces are not scratched, cracked or damaged. A cleaning agent that does not leave a residue such as isopropyl alcohol or a product specifically designed to clean optical equipment should be used along with a lint free cloth.

5. Calibrator Safety Guidelines During the velocity tests, the calibration disk will be rotated at a moderate speed.

It is recommended that the user visually checks the disk for any cracks or distortions on its surface. If the disk is cracked or distorted in any way, the tests should not be performed as this may cause an incorrect result or even damage to the calibration unit. Campbell Scientific should be contacted for advice about any damage that may have occurred to the disk.

When performing the tests, the operator should stand more than two meters from the calibrator. This prevents the operator from interfering with the results by causing reflections, and keeps the operator away from the rotating disk.

10

PWC100 Present Weather Sensor Calibrator

It is also recommended that the operator stands behind the laser hood while the PWS is active.

6. Electrical Parameters 6.1 PWC100 Voltages and Currents

Nominal Max Notes

General details

Standby current 29 mA 30 mA No tests running

Operating current 125 mA 150 mA Disk rotating at test speed

Spin up current 500 mA 700 mA Disk spinning up to speed

+ Volts +12 V +2 8V Main supply voltage

6.2 Earth, 0 V and Transient Protection The user must ensure that the ground lug on the PWS is connected to the earth lug on the logger or another suitable termination point. The PWC100 is connected to this earth via the umbilical cable. This connection acts as the discharge path for electrical transients such as static and surge. Failure to correctly terminate this connection could lead to damage to the unit or faulty operation.

The PWC100 is protected against external electrostatic discharges from user contact but not lightning related surges caused by nearby strikes. Therefore, do not leave the PWC100 connected to the PWS100 for prolonged periods in field conditions.

11

PWC100 Present Weather Sensor Calibrator

12

Appendix A. Assembly Guidelines

A.1 Assembly Guidelines The PWC100 visibility and velocity calibration disks can easily be removed for cleaning or repair if needed.

FIGURE A-1. Attaching the velocity disk

FIGURE A-2. Attaching the visibility disk

A-1

Appendix A. Assembly Guidelines

FIGURE A-3. Attachment of the velocity calibrator

A.2 Connector Pin Outs Connector Front View

4: Data 1: 0 V

2: + Volts3: 0 V

A-2

Campbell Scientific Companies

Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES

www.campbellsci.com info@campbellsci.com

Campbell Scientific Africa Pty. Ltd. (CSAf) PO Box 2450

Somerset West 7129 SOUTH AFRICA

www.csafrica.co.za cleroux@csafrica.co.za

Campbell Scientific Australia Pty. Ltd. (CSA) PO Box 8108

Garbutt Post Shop QLD 4814 AUSTRALIA

www.campbellsci.com.au info@campbellsci.com.au

Campbell Scientific do Brazil Ltda. (CSB) Rua Luisa Crapsi Orsi, 15 Butant

CEP: 005543-000 So Paulo SP BRAZIL www.campbellsci.com.br suporte@campbellsci.com.br

Campbell Scientific Canada Corp. (CSC)

11564 - 149th Street NW Edmonton, Alberta T5M 1W7

CANADA www.campbellsci.ca dataloggers@campbellsci.ca

Campbell Scientific Centro Caribe S.A. (CSCC)

300 N Cementerio, Edificio Breller Santo Domingo, Heredia 40305

COSTA RICA www.campbellsci.cc info@campbellsci.cc

Campbell Scientific Ltd. (CSL)

Campbell Park 80 Hathern Road

Shepshed, Loughborough LE12 9GX UNITED KINGDOM

www.campbellsci.co.uk sales@campbellsci.co.uk

Cam

Manualsnet FAQs

If you want to find out how the PWC100 Campbell Scientific works, you can view and download the Campbell Scientific PWC100 Calibrator Instruction Manual on the Manualsnet website.

Yes, we have the Instruction Manual for Campbell Scientific PWC100 as well as other Campbell Scientific manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Instruction Manual should include all the details that are needed to use a Campbell Scientific PWC100. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Campbell Scientific PWC100 Calibrator Instruction Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Campbell Scientific PWC100 Calibrator Instruction Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Campbell Scientific PWC100 Calibrator Instruction Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Campbell Scientific PWC100 Calibrator Instruction Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Campbell Scientific PWC100 Calibrator Instruction Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.