Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide PDF

1 of 3
1 of 3

Summary of Content for Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide PDF

Revision: 08/2021

www.campbellsci.com/barovue10

1. Introduction This Quick Deploy Guide provides an overview of the steps used to install and measure the BaroVUE 10 using SDI-12. The product manual is the definitive source for detailed installation instructions and information. The manual is available at: www.campbellsci.com/barovue10 .

2. Program sensor 1. Connect the data logger to the computer.

2. Open Short Cut and click Create New Program.

3. Double-click the data logger model.

4. In the Available Sensors and Devices box, type BaroVUE10 or find the sensor in the Sensors > Meteorological > Barometric Pressure folder, and double-click BaroVUE10 (SDI-12) Barometric Pressure Sensor. SDI-12 Address defaults to 0. Enter the correct SDI-12 Address for the sensor if it has been changed from the factory default value. Type the elevation in meters. Choose whether to measure the sensor every hour or every scan. The Barometric pressure and Temperature default to hPa and deg C, respectively. These can be changed by clicking the Barometric pressure or Temperature box and selecting another option.

5. Click the Wiring tab to see how the sensor is to be wired to the data logger. The default control terminal for SDI-12 is C1. To change to another terminal, click the terminal name and select another terminal. Click OK after wiring the sensor.

6. In Output Setup, type the scan rate, a Table Name, and Data Output Storage Interval. Click Next.

1

7. Select the output options.

8. Click Finish and save the program. Send the program to the data logger.

3. Calibration sheet

NOTE: The calibration sheet supplied with the BaroVUE 10 applies to the sensor card (BVC10). The serial number on the outside of the unit refers to the serial number of the BaroVUE 10 Main Board. To confirm you have the correct calibration sheet, open the device and check the serial number on the BVC10 sensor card.

4. Test sensor Verify sensor operation in the lab or office before installing in the field using the Device Configuration Utility software available as a download on www.campbellsci.com . It is included in installations of LoggerNet, RTDAQ, and PC400 software.

1. Open the main case by unclipping the latch and opening the main door.

FIGURE 4-1. BaroVUE 10 with main door open

2. Open Device Configuration Utility.

3. Type BaroVUE 10 in the Device Type box and click BaroVUE 10.

2

4. If this is the first time connecting the BaroVUE 10 to the computer, click Install USBDriver before connecting the cable to the computer.

5. Use the USB cable to connect the BaroVUE 10 USB port to a computer USB port. See FIGURE 4-1.

6. Select the Communication Port in the left panel. BaroVUE10 will appear in the selection dialog.

NOTE: It may take a few seconds for the Communication Port to become available for use after physically connecting the BaroVUE 10 to the computer. If you don't see a BaroVUE 10 device listed, check that the USB cable is correctly inserted and that the USB driver was installed in step 4.

7. Click Connect to access the Settings Editor.

8. Click the Live Data tab to view real-time measurements. The Quality Metric Sample value should be greater than 5.

WARNING: Remove the USB cable before closing the door.

5. Wire functions and connections Connect the wires in the order shown in the following table.

Table 5-1: Wire color, function, and data logger connection

Wire color Wire function Data logger connection

White SDI-12 signal C, SDI-12, or U configured for SDI-121

Red Power 12V, Battery+

Black Power ground G

Blue RS-232 TX not connected

Yellow RS-232 RX not connected 1U and C terminals are automatically configured by the measurement instruction.

6. Mounting

CAUTION: To prevent condensation, install the sensor in an environmentally protected enclosure, complete with desiccant. Change the desiccant at regular intervals.

The BaroVUE

Manualsnet FAQs

If you want to find out how the BaroVue 10 Campbell Scientific works, you can view and download the Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide on the Manualsnet website.

Yes, we have the Quick Deploy Guide for Campbell Scientific BaroVue 10 as well as other Campbell Scientific manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Quick Deploy Guide should include all the details that are needed to use a Campbell Scientific BaroVue 10. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Campbell Scientific BaroVue 10 Barometer Quick Deploy Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.