Agilent InfinityLab LC Series Multiwash Upgrade Kit Installation Note

General Information The multiwash upgrade kit is designed for low carryover and can be installed in any Agilent InfinityLab LC Series Multisampler. With the multiwash upgrade kit you can minimize carryover for critical samples with the following measures: Wash the outer needle surface with the choice of three different wash solvents, or Back flush the needle seat and the rotor seal with the appropriate solvents.

This integrated feature reduces the carryover to less than 9 ppm.

Figure 1 Valve in bypass, needle seat backflush (Multiwash)

NOTE The Multiwash Option is incompatible with multidraw.

Wash Pump

Solvent Sampling loop

Metering

Needle seat

Wash port

Multisampler

P1 P2

1

2

3 4 5

6

Pump

device

Column

Waste

2

General Information

Figure 2 Capillary connections (Multiwash)

Solvents on cabinet

to washport

Sample loop flex

to column (port 6)

Plastic blank nut

from pump (port 1)

to metering device (port 2)

(port 4)

(port 3)

to washport/ needle seat

(port 5) to needle seat

to flush pump

Injection valve

Metering device

SSV/piezo pump

Flush pump

1

2

3

4 5

6

3

Delivery Checklist

Delivery Checklist Multiwash upgrade kit (G7167-68712) contains:

Item # p/n Description

1 1 G4267-68708 Drain Management Kit

2 1 G4267-60049 Flush head, 500 L

3 1 5500-1167 Capillary ST 0.17 mm x 250 mm SL-SL

4 2 G4220-60007 Bottle Head Assembly

5 1 G4267-60008 Hydraulic Multi Wash this is an internal partnumber and not orderable

6 1 G4267-60081 Tubing-Kit-Sampler-Multi-Wash

1

2

3

4 5

6

4

Installing the Multiwash Infinity II Upgrade Kit Remove the Hydraulic Box

Installing the Multiwash Infinity II Upgrade Kit The installation of the upgrade kit requires the following procedures: Remove the Hydraulic Box on page 4, Install the Hydraulic Box on page 8, and Configuration of the Multiwash Option on page 12.

Remove the Hydraulic Box

Figure 3 Variants of the hydraulic box

Standard Dual Needle Multiwash

Tools required p/n Description

Hexagonal key, 2.5 mm Hexagonal key, 4 mm

8710-0510 Open-end wrench 1/4 5/16 inch

NOTE For changing the metering devices in the hydraulic boxes of the multisampler, the sampler must be in the service position.

5

Installing the Multiwash Infinity II Upgrade Kit Remove the Hydraulic Box

1 In the Instant Pilot start the maintenance mode and select Change Piston / Change Seal. Wait until the metering drive is in the rear position. OR In the Agilent Lab Advisor software select Service & Diagnostics in the system screen (Tools) Maintenance Positions> Change Piston / Change Seal, click and wait until the needle assembly is in maintenance position.

2 Switch off the module.

3 Remove the capillaries, the tubings, and the leak sensor which are connected to the hydraulic box.

6

Installing the Multiwash Infinity II Upgrade Kit Remove the Hydraulic Box

4 Remove the injection valve and the analytical head from the standard hydraulic box. For further info check the service manual.

5 Remove the tubing holder insert at the right side of the multisampler.

6 Remove the transport protection screw.

7

Installing the Multiwash Infinity II Upgrade Kit Remove the Hydraulic Box

7 Lift the clip and pull the standard hydraulic box out of the mainframe.

8

Installing the Multiwash Infinity II Upgrade Kit Install the Hydraulic Box

Install the Hydraulic Box 1 Check which version of the drain management kit is currently installed in the sampler,

especially the wash port (see figure below). If the old version with metal connector is installed, upgrade the complete drain

management kit. For further information, check the service manual. If the new one is installed, no further change is necessary, proceed to the next step.

2 Install the flush head. 3 Install the injection valve and the analytical head from the standard hydraulic box into the

multiwash hydraulic box.

9

Installing the Multiwash Infinity II Upgrade Kit Install the Hydraulic Box

4 Slide in the whole new multiwash hydraulic box.

5 Install the transport protection screw.

6 Install the tubing holder insert.

NOTE The hydraulic box must be installed correctly. Push from the front against the hydraulic box and check if the lever is locked in place.

10

Installing the Multiwash Infinity II Upgrade Kit Install the Hydraulic Box

7 Reinstall the capillaries, the tubing and the leak sensor which are connected to the hydraulic box.

Figure 4 Capillary connections (Multiwash)

Solvents on cabinet

to washport

Sample loop flex

to column (port 6)

Plastic blank nut

from pump (port 1)

to metering device (port 2)

(port 4)

(port 3)

to washport/ needle seat

(port 5) to needle seat

to flush pump

Injection valve

Metering device

SSV/piezo pump

Flush pump

1

2

3

4 5

6

11

Installing the Multiwash Infinity II Upgrade Kit Install the Hydraulic Box

8 Switch on the instrument. 9 Prime the tubing with appropriate wash solvent and run Prime to Seat and prime to Wash port

for at least 30 s.

10 Perform a pressure test in LabAdvisor.

12

Installing the Multiwash Infinity II Upgrade Kit Configuration of the Multiwash Option

Configuration of the Multiwash Option The following options for the configuration of the Multiwash Option are available: Configure Multiwash in OpenLab CDS C.01.06 and above on page 12 Configure Multi-wash in Control Module (G4208A) on page 16

Configure Multiwash in OpenLab CDS C.01.06 and above The multiwash mode is automatically enabled and the multiwash icon is shown in the graphical user interface.

Prerequisites For the OpenLab CDS LC driver A.02.11 or higher is required.

1 Right click on the dashboard and select Method.

13

Installing the Multiwash Infinity II Upgrade Kit Configuration of the Multiwash Option

2 Select Multi-wash.

14

Installing the Multiwash Infinity II Upgrade Kit Configuration of the Multiwash Option

3 Specify the duration of your solvent S1, S2, and S3 in the Multi-wash table.

NOTE It takes approximately 30 s to fully exchange one solvent for another in the flush port, and 18 s in the seat. Also, it is recommended to use the autoclean function (right-click the dashboard and select Auto-clean) to flush the multisampler regularly with all installed solvents.

NOTE If Seat Back Flush and Needle Wash are both activated at the same time, they must use the same solvent.

NOTE It is recommended to always activate Start Cond. as the last step of the washing procedure. The flow path is then filled with the appropriate starting solvent conditions for the next sample.

NOTE It is not recommended to use buffers and solvents with high salt content as the last step of the washing procedure. This might cause formation of salt crystals on the needle seat.

15

Installing the Multiwash Infinity II Upgrade Kit Configuration of the Multiwash Option

NOTE Due to the flow path of the Multiwash hydraulic box (see Figure 1 on page 1), when the metering device moves to home position before an injection, a droplet of clean mobile phase will come out from the needle tip and might fall into the seat. This is an expected behavior, independent of the Injection Path Cleaning settings, and does not indicate leakage.

To avoid salt build-up on the needle and the seat when using buffers and solvents with high salt content as mobile phase, follow these recommendations:

Purge the Multisampler daily with water for 15 min.

Visually inspect the needle and seat.

Installing the Multiwash Infinity II Upgrade Kit Configuration of the Multiwash Option

*G7167-90210* *G7167-90210* Manual Part Number: G7167-90210 Rev. D Document Number: SD-29000208 Rev. D

Edition: 07/2020 Printed in Germany

Agilent Technologies, Inc 2016-2020

Agilent Technologies, Inc Hewlett-Packard-Strasse 8

76337 Waldbronn Germany

Configure Multi-wash in Control Module (G4208A) Prerequisites For the Instant Pilot G4208A Firmware B.02.19 or higher is required.

1 Press Control and select Prime/Auto clean the tubing with appropriate wash solv

Manualsnet FAQs

If you want to find out how the InfinityLab LC Series Agilent works, you can view and download the Agilent InfinityLab LC Series Upgrade Kit Installation Note on the Manualsnet website.

Yes, we have the Installation Note for Agilent InfinityLab LC Series as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Installation Note should include all the details that are needed to use a Agilent InfinityLab LC Series. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent InfinityLab LC Series Upgrade Kit Installation Note is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent InfinityLab LC Series Upgrade Kit Installation Note consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent InfinityLab LC Series Upgrade Kit Installation Note free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent InfinityLab LC Series Upgrade Kit Installation Note, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent InfinityLab LC Series Upgrade Kit Installation Note as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.