Installation of the Infinity II Cooler/Thermostat Condensate

Drainage Tubing Kit Technical Note

This Technical Note describes the use and installation of the Infinity II Cooler/Thermostat Condensate Drainage Tubing Kit.

Delivery Checklist Condensate Drainage Kit (5067-6208) contains:

Figure 1 Content of Condensate Drainage Kit (5067-6208)

Item # Description

1 1 Condensate Collector

2 1 Single Drain Connector

3 2 Washer, ST, ID/OD 3.2/9 mm

4 1 Tubing, silicon rubber 2.16 m, ID/OD 6/9 mm (pre-cut)

5 1 Tubing connector, Y-shaped, ID 6.4 mm (with a 60 and 100 mm premounted tubing piece)

6 1 Tubing connector, 90 , ID 6.4 mm

7 1 Tubing holder clip, 19 x 19 x 8 mm

1

2

3

4

5

6

7

Installation of the Condensate Drainage Kit

Installation of the Condensate Drainage Kit

Setup With the Condensate Collector Funnel for Bench Installations

1 Attach the condensate tube to the outlet port of the condensate collector funnel.

Tools required p/n Description

8710-1622 Torx screwdriver T15

OR 5023-3089 Torx key set

Parts required # Description

1 Condensate Collector

1 Single Drain Connector

1 Washer, ST, ID/OD 3.2/9 mm

1 Tubing, silicon rubber 2.16 m, ID/OD 6/9 mm (pre-cut)

1 Tubing connector, 90 , ID 6.4 mm

NOTE The setup with the condensate collector funnel is suitable for bench installations only.

NOTE The condensate collector funnel is not suitable for Flex Bench systems or when the sampler is the lowermost module in a bench installation.

2

Installation of the Condensate Drainage Kit

2 Mount the drain connector on the condensate drainage outlet tube. Ensure the correct orientation of the spout.

3 Remove the screw situated above the condensate drainage outlet tube.

3

Installation of the Condensate Drainage Kit

4 Place the washer over the thread of the screw (1). Screw the screw and washer halfway into the hole in the back of the cooler/thermostat (2).

5 Position the condensate collector funnel underneath the condensate drainage outlet tube (1) and fix it to the back of the cooler/thermostat by tightening the screw (2). Ensure correct orientation and avoid over-tightening the screw.

1.

2.

1.

2.

4

Installation of the Condensate Drainage Kit

6 Shorten the condensate tube so that it runs straight into the waste container without any unnecessary detour (1). If needed, use the 90 tubing connector provided in the kit to eliminate uphill sections, which might occur at the edge of the bench (2). Agilent recommends the use of a separate canister for condensate collection to avoid drainage problems (3).

1

3

2

5

Installation of the Condensate Drainage Kit

7 Ensure that the tubing runs straight into the waste canister without any bends or joints and it is not hindered by any mechanical obstacle. Agilent recommends using a 6 L waste canister equipped with a suitable InfinityLab Stay Safe cap for optimal condensate handling. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.

NOTE Depending on the ambient conditions in the lab, the amount of condensate can vary from 200 mL to 2 L per day. Do not fill the waste container for the condensate to the top. Regularly empty the waste container.

6

Installation of the Condensate Drainage Kit

Setup with the Y-piece Tubing Connector for Bench Installations and Flex Bench Systems

1 Attach the condensate tube to the free end of the Y-piece tubing connector.

Parts required Description

Tubing connector, Y-shaped, ID 6.4 mm (with a 60 and 100 mm premounted tubing piece)

Tubing holder clip, 19 x 19 x 8 mm

Tubing, silicon rubber 2.16 m, ID/OD 6/9 mm (pre-cut)

Tubing connector, 90 , ID 6.4 mm

NOTE The setup with the Y-piece tubing connector is equally suitable for bench installations and Flex Bench systems.

NOTE Agilent Infinity II autosamplers with a cooler/thermostat installed should not be used as the lowermost module in an HPLC stack to ensure adequate condensate drainage.

7

Installation of the Condensate Drainage Kit

2 Adhere a tubing holder clip to the side of the cooler/thermostat where the condensate drainage outlet tube is situated. Ensure a distance of 20 mm from the bottom edge.

3 Mount the shortest tube (60 mm) of the condensate tubing with the Y-piece on the condensate drainage outlet tube and fix the venting tube (100 mm) in the tubing holder clip.

2 0

m m

1 00

m m

60 m

m

8

Installation of the Condensate Drainage Kit

4 Ensure that the condensate handling system is installed in a way that allows a continuous slope for the drained liquid. Horizontal or uphill sections may hinder the drainage.

9

Installation of the Condensate Drainage Kit

5 Shorten the condensate tubing so that it runs straight into the waste container without any unnecessary detour (1). If needed, use the 90 tubing connector provided in the kit to eliminate up-hill sections, which might occur at the edge of the bench (2). Agilent recommends the use of a separate canister for condensate collection to avoid drainage problems (3).

a Routing the condensate tubing for bench installations.

1

3

2

10

Installation of the Condensate Drainage Kit

b Routing the condensate tubing for a Flex Bench system.

1

3

11

6 Ensure that the tubing runs straight into the waste canister without any bends or joints and it is not hindered by any mechanical obstacle. Agilent recommends using a 6 L waste canister equipped with a suitable InfinityLab Stay Safe cap for optimal condensate handling. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.

NOTE Depending on the ambient conditions in the lab, the amount of condensate can vary from 200 mL to 2 L per day. Do not fill the waste container for the condensate t

Manualsnet FAQs

If you want to find out how the Infinity II Agilent works, you can view and download the Agilent Infinity II Tubing Kit Technical Note on the Manualsnet website.

Yes, we have the Technical Note for Agilent Infinity II as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Technical Note should include all the details that are needed to use a Agilent Infinity II. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent Infinity II Tubing Kit Technical Note is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent Infinity II Tubing Kit Technical Note consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent Infinity II Tubing Kit Technical Note free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent Infinity II Tubing Kit Technical Note, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent Infinity II Tubing Kit Technical Note as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.