Agilent Microwave Plasma Atomic Emission Spectrometers Models 4100, 4200, and 4210

Long Term Shut Down and Start up Instructions

Shut down the system

To shut down the system: 1 Rinse the spray chamber by aspirating water (or the appropriate solvent for your application)

for a few minutes.

NOTE When running organic samples, it is recommended that the spray chamber be cleaned and dried thoroughly between analyses.

2 Remove solution tubing from the solution. 3 Pump all solution from the sample line, nebulizer and spray chamber with the peristaltic pump

and then continue to run for another 30 seconds to ensure that the sample introduction system is solution free.

4 Choose Plasma Off from the Plasma button drop-down arrow to extinguish the plasma. The peristaltic pump stops automatically when the plasma is extinguished.

NOTE The green Plasma enable button, located on the front of the MP-AES instrument, is intended to be used only in an emergency. It is not intended to be used every time you want to extinguish the plasma. If it is used to extinguish the plasma, you will need to reset the Plasma enable button to the ON position before the plasma can be re-ignited.

5 Loosen the pressure on the peristaltic pump tubing by releasing the pressure bars and lifting the tubing out of the grooves in the tube retainer to increase the pump tubing lifetime. To do this: a Push up the pressure bar tensioners (1 in Figure 1).

This releases the tensioner from the pressure bar. b Swing the pressure bar backwards (2 in Figure 1). c Lift the tubing out of the grooves in the tube retainer (3 in Figure 1).

6 Dispose of all pump tubes. 7 Remove all hazardous materials and samples from in and around the equipment and store

appropriately. Special precautions are required for HF applications systems.

Agilent Microwave Plasma Atomic Emission Spectrometers Long Term Shut Down and Start up Instructions

2

WARNING

Warning Hazardous Materials Perform the appropriate decontamination procedure if hazardous material has been spilled on or inside the instrument.

8 Clean any sample spills. 9 Remove sample introduction components (plasma torch, spray chamber and nebulizer) and

store in original packaging. 10 Turn off power to the instrument at the instrument ON/OFF switch. 11 Close all open applications and shut down the computer. 12 Shut off all gas supplies at the regulators/isolation valves and gas cylinders where applicable. 13 Disconnect the instrument from the mains power. 14 Turn off the laboratory exhaust system. 15 Shutdown and disconnect any accessories according to the instructions provided in their

manual.

1 Pressure bar tensioner 2. Pressure bar 3. Tube retainer

Figure 1 Peristaltic pump components

NOTE The torch should be cleaned of any deposits, dirt or residue. For further information about cleaning the torch see the MP-AES Users Guide or search for Agilent MP-AES Easy Fit Torch Maintenance on www.agilent.com

Start up the system

To start up the system: 1 Check that the exhaust and intake lines are secured to the MP-AES instrument. 2 Ensure the gas lines are connected to the MP-AES instrument and that the gas supplies are

turned on and set to the correct pressures. 3 Check that the USB and power cables are plugged in. 4 Check that the pre-optics window is clean and correctly installed and that the interlock is

engaged. 5 Check that the torch is clean and in good condition. 6 Insert the torch and completely close the torch handle. 7 Insert the nebulizer into the spray chamber. 8 Fit the spray chamber socket to the ball joint on the base of the torch and secure using the

torch clamp.

2 3

1

Agilent Microwave Plasma Atomic Emission Spectrometers Long Term Shut Down and Start up Instructions

3

9 Check that all tubing on the spray chamber, nebulizer and peristaltic pump are correctly connected.

10 Switch on the computer, monitor and printer. 11 Switch on the laboratory exhaust system.

12 Turn on the MP-AES instrument. 13 Ensure that the Plasma Enable Switch is in the Enable state (pushed in).

NOTE Wait 1-2 hours for the instrument to reach operating temperature.

14 Double-click the MP Expert desktop icon to start the MP Expert software.

Refer to the Agilent Microwave Plasma Atomic Emission Spectrometers Users Guide and MP Expert Help for more information on preparing your MP-AES instrument for analysis.

WARNING

Hot surfaces noxious fumes non-ionizing radiation

Burn danger, inhalation hazard. The plasma emits heats, ozone and fumes, which can be hazardous. Always switch on the exhaust system before lighting the plasma.

Agilent Microwave Plasma Atomic Emission Spectrometers Long Term Shut Down and Start up Instructions

<

Manualsnet FAQs

If you want to find out how the 4100 Microwave Plasma Atomic Emission Agilent works, you can view and download the Agilent 4100 Microwave Plasma Atomic Emission Spectroscope Instructions on the Manualsnet website.

Yes, we have the Instructions for Agilent 4100 Microwave Plasma Atomic Emission as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Instructions should include all the details that are needed to use a Agilent 4100 Microwave Plasma Atomic Emission. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent 4100 Microwave Plasma Atomic Emission Spectroscope Instructions is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent 4100 Microwave Plasma Atomic Emission Spectroscope Instructions consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent 4100 Microwave Plasma Atomic Emission Spectroscope Instructions free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent 4100 Microwave Plasma Atomic Emission Spectroscope Instructions, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent 4100 Microwave Plasma Atomic Emission Spectroscope Instructions as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.