Agilent InfinityLab LC Series Sampler User Manual PDF
Summary of Content for Agilent InfinityLab LC Series Sampler User Manual PDF
Vialsamplers Agilent InfinityLab LC Series
User Manual
Agilent InfinityLab LC Series Vialsamplers User Manual
Notices Document Information Document No: SD-29000242 Rev. E Part No: G7129-90010 Rev. E EDITION 04/2020
Copyright Agilent Technologies, Inc. 2015-2020
No part of this manual may be repro- duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written con- sent from Agilent Technologies, Inc. as governed by United States and interna- tional copyright laws.
Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn
Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future edi- tions. Further, to the maximum extent permitted by applicable law, Agilent dis- claims all warranties, either express or implied, with regard to this manual and any information contained herein, includ- ing but not limited to the implied warran- ties of merchantability and fitness for a particular purpose. Agilent shall not be lia- ble for errors or for incidental or conse- quential damages in connection with the furnishing, use, or performance of this document or of any information con- tained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend U.S. Government Restricted Rights. Soft- ware and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Soft- ware and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Depart- ment of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commer- cial Computer Software or Computer Software Documentation).
Safety Notices
CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARN- ING notice until the indicated conditions are fully understood and met.
Agilent InfinityLab LC Series Vialsamplers User Manual
In This Book:
This manual covers the following modules: Agilent 1260 Infinity II Vialsampler (G7129A) Agilent 1290 Infinity II Vialsampler (G7129B) Agilent 1260 Infinity II Vialsampler (G7129C)
1 Introduction
This chapter gives an introduction to the module and instrument overview.
2 Site Requirements and Specifications
This chapter provides information on environmental requirements, physical and performance specifications.
3 Using the Module
This chapter provides information on how to use the module.
4 Preparing the Module
This chapter explains the operational parameters of the module.
5 Optimizing Performance
This chapter provides information on how to optimize the module.
6 Troubleshooting and Diagnostics
This chapter gives an overview of the maintenance, troubleshooting, and diagnostic features available for the Agilent InfinityLab Series Vialsampler.
7 Error Information
This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.
Agilent InfinityLab LC Series Vialsamplers User Manual 3
8 Maintenance
This chapter describes the maintenance of the module.
9 Parts and Materials for Maintenance
This chapter provides information on parts for maintenance.
10 Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.
11 Hardware Information
This chapter describes the detector in more detail on hardware and electronics.
12 LAN Configuration
This chapter provides information on connecting the module to the Agilent ChemStation PC.
13 Appendix
This chapter provides additional information on safety, legal and web.
Agilent InfinityLab LC Series Vialsamplers User Manual 4
Contents
1 Introduction 9
Product Description (G7129A) 10 Features (G7129A) 11 Product Description (G7129B) 12 Features (G7129B) 13 Product Description (G7129C) 14 Features (G7129C) 15 Overview of the Module 16 Operating Principle 18 Leak and Waste Handling 27
2 Site Requirements and Specifications 33
Site Requirements 34 Physical Specifications 38 Performance Specifications 39 Specifications of the Sample Cooler 47 Specifications of the Sample Thermostat 49 Specifications of the Integrated Column Compartment 51
3 Using the Module 53
Magnets 55 Turn on/off 56 Status Indicators 58 Vial Drawers and Trays 59 Choice of Vials and Caps 70 Install the Optional Integrated Column Compartment 74 Using the Optional Integrated Column Compartment 96 Install the Optional Sample Cooler/Sample Thermostat 116 Using the Optional Sample Cooler/Sample Thermostat 123 Transporting the Sampler 132 Agilent Local Control Modules 137
Agilent InfinityLab LC Series Vialsamplers User Manual 5
4 Preparing the Module 139
Leak and Waste Handling 140 Preparing the Module 141 Solvent Information 142 Capillary Color Coding Guide 149 Installing Capillaries 150 Flow Connections to the Vialsampler 152 Setting up the Vialsampler 154
5 Optimizing Performance 166
Optimization for Lowest Carryover 167 Fast Injection Cycle and Low Delay Volume 171 Precise Injection Volume 173 Choice of Rotor Seal 175
6 Troubleshooting and Diagnostics 176
Status Indicators 177 Overview of Tests and Tools 178 Maintenance and Troubleshooting Tools 179 Diagnostic Tests 187 Agilent Lab Advisor Software 189
7 Error Information 190
What are Error Messages 192 General Error Messages 193 Vialsampler Error Messages 201 Sample Cooler/Sample Thermostat Error Messages 213 Integrated Column Compartment (ICC) Heater Error Messages 219
Agilent InfinityLab LC Series Vialsamplers User Manual 6
8 Maintenance 222
Introduction to Maintenance 223 Warnings and Cautions 224 Overview of Maintenance 226 Cleaning the Module 227 Remove and Install Doors 228 Exchange the Needle Assembly 229 Exchange the Needle Seat Assembly 235 Exchange the Sample Loop Assembly 241 Exchange the Rotor Seal 245 Exchange the Metering Seal and Piston 250 Replace the Analytical Heads 255 Exchange the Gripper Arm 258 Replace the Peristaltic Pump Cartridge 261 Exchange the Wash Port Assembly 263 Replace the Module Firmware 269 Replace the Sample Cooler/Sample Thermostat 270
9 Parts and Materials for Maintenance 275
Standard Parts For Maintenance 276 Accessory Kits 279 Vial Drawers and Trays 283 Multidraw Kits 286 Sample Thermostat Upgrade 287 Integrated Column Compartment 288 Column ID Upgrade Kit 289 Parts for 900 L Injection Upgrade 290 Cabinet and Door Kits 291 Analytical Head Assembly (100 L) 295 Analytical-Head Assembly (900 L) 296 Analytical Head Assembly (40 L) 297 2 Position/6 Port Injection Valve, 600 bar 298 2 Position/6 Port Injection Valve, 800 bar 299 2 Position/6 Port Injection Valve, 1300 bar 300
Agilent InfinityLab LC Series Vialsamplers User Manual 7
10 Identifying Cables 301
Cable Overview 302 Analog Cables 304 Remote Cables 306 CAN/LAN Cables 310 RS-232 Cables 311 USB 312
11 Hardware Information 313
Firmware Description 314 Electrical Connections 317 Interfaces 319 Setting the 6-bit Configuration Switch 326 Instrument Layout 330 Early Maintenance Feedback (EMF) 331
12 LAN Configuration 332
What You Have to Do First 333 TCP/IP parameter configuration 334 Configuration Switches 335 Initialization Mode Selection 336 Dynamic Host Configuration Protocol (DHCP) 338 Manual Configuration 341 PC and Agilent ChemStation Setup 345
13 Appendix 354
General Safety Information 355 Waste Electrical and Electronic Equipment (WEEE) Directive 361 Refrigerant 362 Radio Interference 365 Sound Emission 366 Solvent Information 367 Agilent Technologies on Internet 368
Agilent InfinityLab LC Series Vialsamplers User Manual 8
1 Introduction
Product Description (G7129A) 10 Features (G7129A) 11 Product Description (G7129B) 12 Features (G7129B) 13 Product Description (G7129C) 14 Features (G7129C) 15 Overview of the Module 16 Operating Principle 18 Sampling Sequence 18 Needle Parkstation 23 Hydraulic Box 24 Transport Assembly 26
Leak and Waste Handling 27 Leak Sensor 31 Waste Guidance 31 Waste Concept 32
This chapter gives an introduction to the module and instrument overview.
Agilent InfinityLab LC Series Vialsamplers User Manual 9
1 Introduction Product Description (G7129A)
Product Description (G7129A)
The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the reliability and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 C as well as a sample thermostat for stable temperatures from 4 C to 40 C, all within one module.
Figure 1 Overview of the Vialsampler
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch
Agilent InfinityLab LC Series Vialsamplers User Manual 10
1 Introduction Features (G7129A)
Features (G7129A)
High capacity up to 132 vials (2 mL) or up to 36 vials (6 mL). Reliable injections 0.1 to 100 L injections for up to 600 bar (G7129A) and
up to 800 bar (G7129C). Easy volume extensions for injection volumes up to 1800 L for applications
ranging from microbore to semipreparative chromatography. Lowest carryover with an in needle flush port included, for rinsing the outside
of the needle. Efficient temperature control with an integrated column compartment as
option or upgrade available. Which holds two columns up to 30 cm length, and provides heating capacity from 5 C above ambient up to 80 C for reproducible chromatography data at optimized resolution.
Integrated sample thermostat available as option or upgrade, providing cooling and heating in the range from 4 C - 40 C.
Low internal volume for minimum contribution to a system's total internal volume, which can be even further reduced using "bypass" mode.
Increased productivity with overlapped injections. Customizable injection program available for customizing advanced
injections as well as for sample preparation steps upfront injection.
Agilent InfinityLab LC Series Vialsamplers User Manual 11
1 Introduction Product Description (G7129B)
Product Description (G7129B)
The Agilent 1290 Infinity II Vialsampler (G7129B) is an autosampler designed for UHPLC applications up to 1300 bar. It provides the reliability, safety, and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 C as well as a sample thermostat for stable temperatures from 4 C to 40 C, all within one module.
Figure 2 Overview of the Vialsampler
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch
Agilent InfinityLab LC Series Vialsamplers User Manual 12
1 Introduction Features (G7129B)
Features (G7129B)
Accurate and precise injections - within a wide and flexible range of volumes High capacity - up to 132 vials (2 mL) or up to 36 vials (6 mL). Easy volume extensions - for injection volumes up to 1500 L for applications
ranging from microbore to semipreparative chromatography. Lowest carryover - with an in needle flush port included, for rinsing the outside
of the needle. Efficient temperature control - with an integrated column compartment as
option or upgrade available. Which holds two columns up to 30 cm length, and provides heating capacity from 5 C above ambient up to 80 C for reproducible chromatography data at optimized resolution.
Integrated sample thermostat - available as option or upgrade, providing cooling and heating in the range from 4 40 C.
Low internal volume - for minimum contribution to a system's total internal volume, which can be even further reduced using bypass mode.
Increased productivity - with overlapped injections. Customizable Injection program - available for customizing advanced
injections as well as for sample preparation steps upfront injection.
Agilent InfinityLab LC Series Vialsamplers User Manual 13
1 Introduction Product Description (G7129C)
Product Description (G7129C)
The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the reliability and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 C as well as a sample thermostat for stable temperatures from 4 C to 40 C, all within one module.
Figure 3 Overview of the Vialsampler
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch
Agilent InfinityLab LC Series Vialsamplers User Manual 14
1 Introduction Features (G7129C)
Features (G7129C)
High capacity up to 132 vials (2 mL) or up to 36 vials (6 mL). Reliable injections 0.1 to 100 L injections for up to 600 bar (G7129A) and
up to 800 bar (G7129C). Easy volume extensions for injection volumes up to 1800 L for applications
ranging from microbore to semipreparative chromatography. Lowest carryover with an in needle flush port included, for rinsing the outside
of the needle. Efficient temperature control with an integrated column compartment as
option or upgrade available. Which holds two columns up to 30 cm length, and provides heating capacity from 5 C above ambient up to 80 C for reproducible chromatography data at optimized resolution.
Integrated sample thermostat available as option or upgrade, providing cooling and heating in the range from 4 C - 40 C.
Low internal volume for minimum contribution to a system's total internal volume, which can be even further reduced using "bypass" mode.
Increased productivity with overlapped injections. Customizable injection program available for customizing advanced
injections as well as for sample preparation steps upfront injection.
Agilent InfinityLab LC Series Vialsamplers User Manual 15
1 Introduction Overview of the Module
Overview of the Module
The Agilent InfinityLab LC Series Vialsampler is designed for use with other modules of the Agilent InfinityLab LC Series, 1200 Series, and 1100 Series LC, as well as with other LC systems with adequate remote-control capabilities. The Vialsampler can be controlled from any computer with a suitable chromatography data system (for example, Agilent OpenLab CDS, MassHunter, etc.) or via an Agilent local control module (InfinityLab Companion or 1200 Infinity Series Instant Pilot).
The 1290 Infinity II Vialsampler (G7129B) can be operated at up to 1300 bar. The two 1260 Infinity II Vialsampler variants (G7129A and G7129C) are limited to a maximum system pressure of 600 and 800 bar, respectively.
The Vialsampler can accommodate two vial drawers, providing space for up to 132 samples. The two cartesian vial drawers allow for the use of either 66 2 mL or 18 6 mL vials. A pair of classic vial drawers with a numerical vial assignment system is also available, each capable of hosting 50 2 mL vials. The classic vial drawers are designed to ease method transfer from older Agilent samplers. The Vialsampler can also be equipped with an external tray, which can be beneficial for automation. The Agilent WalkUp solution fully supports the external tray.
The Vialsampler features a robotic gripper arm for the transport of vials, driven by four stepper motors to ensure the highest precision and flexibility for the transport movement.
The default configuration of the analytical head and sample loop allows an injection range from 0.1 20 L for the 1290 Infinity II Vialsampler and 0.1 100 L for the 1260 Infinity II Vialsamplers. An analytical head with extended volume is also available for injection volumes from 0.1 900 L, and can be operated at up to 400 bar. With the Multidraw Kit, the maximum injection volume can even reach 1800 L.
The 2-position/6-port injection valve is driven by a high-speed, hybrid stepper motor, ensuring smooth switching between the mainpass and bypass positions. Thanks to the flow-through design of the Vialsampler, the eluent continuously rinses the component parts of the sample introduction system. This, together with the automated needle wash function for rinsing the outer surface of the needle, ensures the lowest level of carryover for each analysis. For advanced delay volume reduction, the injection valve can be programmed to automatically switch back to the bypass position when the injection is completed.
Agilent InfinityLab LC Series Vialsamplers User Manual 16
1 Introduction Overview of the Module
The Agilent InfinityLab Integrated Column Compartment, an optional upgrade for the Vialsampler, eliminates the need for having a standalone column thermostat in the LC stack. It has the heating capability of reaching temperatures up to 80 C and can hold a total of two columns with maximum lengths of 30 cm.
For applications requiring control over the vial temperature, the Vialsampler can be equipped with the Agilent InfinityLab Sample Thermostat (G7167-60101). It features a vapor-compression refrigeration system and an electric heater, allowing the Vialsampler to reach vial temperatures down to 4 C and up to 40 C.
Agilent InfinityLab LC Series Vialsamplers User Manual 17
1 Introduction Operating Principle
Operating Principle
Sampling Sequence The Vialsampler processor continuously monitors the movements of the vialsampler components during the sampling sequence. The processor defines specific time windows and mechanical ranges for each movement. If a specific step of the sampling sequence cant be completed successfully, an error message is generated.
During the sampling sequence, the solvent bypasses the vialsampler via the injection valve. The gripper arm selects the sample vial, either from a static sample rack, or from external vial positions. The gripper arm places the sample vial below the injection needle. The required volume of sample is drawn into the sample loop by the metering device. Sample is applied to the column when the injection valve returns to the mainpass (main path) position at the end of the sampling sequence. The sampling sequence occurs in the following order: 1 The injection valve switches to the bypass position. 2 The piston of the metering device moves to the initialization position. 3 The gripper arm moves from the home position, and selects the vial. At the
same time, the needle lifts out of the seat.
4 The gripper moves into the needle station and stops in the draw position.1
5 The needle lowers into the vial. 6 The metering device draws the defined sample volume. 7 The needle lifts out of the vial.
8 The gripper arm moves out slightly and stops in the wash position1. 9 The needle moves downwards and dips into the wash well of the wash port.
Simultaneously the peristaltic pump delivers the flush solvent.1
10 The needle moves back.1
11 The gripper arm moves out of the needle station and the wash port snaps back in position.
1 only if automated needle wash is selected. If this feature is disabled, the gripper arm positions the sample vial directly below the needle (Step 4) and lowers the needle into the vial.
Agilent InfinityLab LC Series Vialsamplers User Manual 18
1 Introduction Operating Principle
12 The gripper arm replaces the vial, and returns to the home position. Simultaneously, the needle lowers into the seat.
13 The injection valve switches to the mainpass (main path) position.
Injection Sequence
Before the start of the injection sequence, and during an analysis, the injection valve is in the mainpass (main path) position (see Figure 4 on page 19). In this position, the mobile phase flows through the autosamplers metering device, sample loop, and needle. This ensures that all parts in contact with sample are flushed during the run, thus minimizing carryover.
Figure 4 Mainpass (main path) Position - standard position during runs and when the sampler is idle
Vial Wash port
Needle seat
Sampling loop
Solvent
Wash pump
Metering device
Agilent InfinityLab LC Series Vialsamplers User Manual 19
1 Introduction Operating Principle
When the sample sequence begins, the valve unit switches to the bypass position (see Figure 5 on page 20). Solvent from the pump enters the valve unit at port 1, and flows directly to the column through port 6.
Figure 5 Valve in bypass - needle in vial, metering device aspirates sample volume
Vial Wash port
Needle seat
Sampling loop
Solvent
Wash pump
Metering device
Agilent InfinityLab LC Series Vialsamplers User Manual 20
1 Introduction Operating Principle
Then the vial is positioned below the needle. The needle moves down into the vial, the metering unit draws the required sample volume into the loop, and the needle is raised. In the next step, the needle is washed (see Figure 6 on page 21).
Figure 6 Outer face of needle getting washed in wash port
Vial Wash port
Needle seat
Sampling loop
Solvent
Wash pump
Metering device
Agilent InfinityLab LC Series Vialsamplers User Manual 21
1 Introduction Operating Principle
When the metering unit has drawn the required volume of sample into the sample loop, the vial is replaced in the sample tray. The wash port flips into the origin position, the needle is lowered into the needle seat, and the injection valve switches back to mainpass (main path) position, flushing the sample onto the column (see Figure 7 on page 22).
Figure 7 Valve switches to mainpass (main path) - sample is transferred towards the LC column
Vial Wash port
Needle seat
Sampling loop
Solvent
Wash pump
Metering device
Agilent InfinityLab LC Series Vialsamplers User Manual 22
1 Introduction Operating Principle
Needle Parkstation The needle parkstation comprises two main assemblies: needle drive and wash port.
Figure 8 Needle Station
Needle-Drive
The needle movement is driven by a stepper motor connected to the spindle assembly by a toothed belt. The circular motion of the motor is converted to linear motion by the drive nut on the spindle assembly. The upper and lower needle positions are detected by reflection sensors on the needle station board, while the needle-in-vial position is determined by counting the motor steps from the upper needle-sensor position.
Agilent InfinityLab LC Series Vialsamplers User Manual 23
1 Introduction Operating Principle
Hydraulic Box The hydraulic box comprises two main assemblies: metering device, and injection valve.
Figure 9 Hydraulic Unit
NOTE The replacement hydraulic box excludes the injection valve and metering head assemblies.
Agilent InfinityLab LC Series Vialsamplers User Manual 24
1 Introduction Operating Principle
Analytical Head
The analytical head is driven by the stepper motor that is connected to the drive shaft by a toothed belt. The drive nut on the spindle converts the circular movement of the spindle to linear motion. The drive nut pushes the sapphire piston against the tension of the spring into the analytical head. The base of the piston sits on the large bearing of the drive nut, which ensures the piston is always centered. A ceramic ring guides the movement of the piston in the analytical head. The home position of the piston is sensed by an optical sensor on the hydraulic unit board while the sample volume is determined by counting the number of steps from the home position. The backward movement of the piston (driven by the spring) draws sample from the vial.
Injection Valve
The two-position 6-port injection valve is driven by a stepper motor. Only five of the six ports are used (port 3 is not used). A lever/slider mechanism transfers the movement of the stepper motor to the injection valve. Two microswitches monitor switching of the valve (bypass and mainpass (main path) end positions). No valve adjustments are required after replacing internal components.
Agilent InfinityLab LC Series Vialsamplers User Manual 25
1 Introduction Operating Principle
Transport Assembly The transport unit comprises an X-axis slide (left-right motion), a Z-axis arm (up-down motion), and a gripper assembly (rotation and vial-gripping).
Figure 10 Transport Assembly
The transport assembly uses four stepper motors driven in closed-loop mode for accurate positioning of the gripper assembly. The rotational movement of the motors is converted to linear motion (X- and Z-axes) by toothed belts connected to the drive spindles. The rotation (theta axes) of the gripper assembly is transferred from the motor by a toothed belt and series of gears. The opening and closing of the gripper fingers are driven by a stepper motor linked by a toothed belt to the planetary gearing inside the gripper assembly.
The stepper motor positions are determined by the optical encoders mounted onto the stepper-motor housing. The encoders monitor the position of the motors continually, and correct for position errors automatically (e.g. if the gripper is accidentally moved out of position when loading vials into the vial tray). The initialization positions of the moving components are sensed by reflection sensors mounted on the flex board. These positions are used by the processor to calculate the actual motor position. An additional six reflection sensors for tray recognition are mounted on the flex board at the front of the assembly.
X motor
Theta motor
Gripper motor
Gripper arm
Z motor
Flex board
Theta axis
X axis
Z axis
Agilent InfinityLab LC Series Vialsamplers User Manual 26
1 Introduction Leak and Waste Handling
Leak and Waste Handling
The Agilent InfinityLab LC Series has been designed for safe leak and waste handling. It is important that all security concepts are understood and instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The maximum volume for an individual bottle stored in the solvent cabinet should not exceed 2 L. For details, see the usage guideline for the Agilent Infinity II Solvent Cabinets (a printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and Infinity II modules can be stacked on top of each other. Waste tubes are guided through a channel on the right hand side of the instrument, keeping the front access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks, guiding them to the leak sensor for leak detection, and passing them on to the next module below, if the leak sensor fails. The leak sensor in the leak plane stops the running system as soon as the leak detection level is reached. Solvent and condensate is guided through the waste channel into the waste container: from the detector's flow cell outlet from the Multisampler needle wash port from the Sample Cooler or Sample Thermostat (condensate) from the pump's Seal Wash Sensor (if applicable) from the pump's Purge Valve or Multipurpose Valve
Agilent InfinityLab LC Series Vialsamplers User Manual 27
1 Introduction Leak and Waste Handling
Figure 11 Infinity II Leak Waste Concept (Flex Bench installation)
Agilent InfinityLab LC Series Vialsamplers User Manual 28
1 Introduction Leak and Waste Handling
Figure 12 Infinity II Single Stack Leak Waste Concept (bench installation)
Agilent InfinityLab LC Series Vialsamplers User Manual 29
1 Introduction Leak and Waste Handling
Figure 13 Infinity II Two Stack Leak Waste Concept (bench installation)
The waste tube connected to the leak plane outlet on each of the bottom instruments guides the solvent to a suitable waste container.
Agilent InfinityLab LC Series Vialsamplers User Manual 30
1 Introduction Leak and Waste Handling
Leak Sensor
Waste Guidance
CAUTION Solvent incompatibility
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
Do not use DMF as mobile phase.
Check the leak sensor regularly for corrosion.
NOTE The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf respectively.
NOTE The waste drainage must go straight into the waste containers. The waste flow must not be restricted at bends or joints.
Agilent InfinityLab LC Series Vialsamplers User Manual 31
1 Introduction Leak and Waste Handling
Waste Concept 1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.
Agilent InfinityLab LC Series Vialsamplers User Manual 32
2 Site Requirements and Specifications
Site Requirements 34 Physical Specifications 38 Performance Specifications 39 Performance Specifications (G7129A) 39 Performance Specifications (G7129B) 42 Performance Specifications (G7129C) 45
Specifications of the Sample Cooler 47 Specifications of the Sample Thermostat 49 Specifications of the Integrated Column Compartment 51
This chapter provides information on environmental requirements, physical and performance specifications.
Agilent InfinityLab LC Series Vialsamplers User Manual 33
2 Site Requirements and Specifications Site Requirements
Site Requirements
Site Requirements A suitable environment is important to ensure optimum performance of the instrument.
Power Considerations The module power supply has wide ranging capability. It accepts any line voltage in the range described in Table 1 on page 38. Consequently there is no voltage selector in the rear of the module. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.
WARNING Hazard of electrical shock or damage of your instrumentation can result, if the devices are connected to a line voltage higher than specified.
Connect your instrument to the specified line voltage only.
WARNING Electrical shock hazard The module is partially energized when switched off, as long as the power cord is plugged in. The cover protects users from personal injuries, for example electrical shock.
Do not open the cover.
Do not operate the instrument and disconnect the power cable in case the cover has any signs of damage.
Contact Agilent for support and request an instrument repair service.
WARNING Inaccessible power plug. In case of emergency it must be possible to disconnect the instrument from the power line at any time.
Make sure the power connector of the instrument can be easily reached and unplugged.
Provide sufficient space behind the power socket of the instrument to unplug the cable.
Agilent InfinityLab LC Series Vialsamplers User Manual 34
2 Site Requirements and Specifications Site Requirements
Power Cords Country-specific power cords are available for the module. The female end of all power cords is identical. It plugs into the power-input socket at the rear. The male end of each power cord is different and designed to match the wall socket of a particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is suitable for your particular country or region.
WARNING Unintended use of power cords
Using power cords for unintended purposes can lead to personal injury or damage of electronic equipment.
Never use a power cord other than the one that Agilent shipped with this instrument.
Never use the power cords that Agilent Technologies supplies with this instrument for any other equipment.
Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
WARNING Absence of ground connection
The absence of ground connection can lead to electric shock or short circuit.
Never operate your instrumentation from a power outlet that has no ground connection.
WARNING Electrical shock hazard
Solvents may damage electrical cables.
Prevent electrical cables from getting in contact with solvents.
Exchange electrical cables after contact with solvents.
Agilent InfinityLab LC Series Vialsamplers User Manual 35
2 Site Requirements and Specifications Site Requirements
Room Size and Ventilation
Bench Space The module dimensions and weight (see Table 1 on page 38) allow you to place the module on almost any desk or laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position, especially if a Sample Cooler or Sample Thermostat is installed. Use a bubble level to check the leveling of the sampler.
WARNING Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample Thermostat and laboratory.
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m3 (1 m3 for every 8 g of R600a refrigerant inside of the Sample Thermostat).
Ensure adequate ventilation: typical air exchange of 25 m3/h per m2 of laboratory floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do not block the openings on the circumference of the Sample Thermostat.
NOTE Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench rack. This option helps to save bench space as all modules can be placed into one single stack. It also allows to easily relocate the instrument to another laboratory.
Agilent InfinityLab LC Series Vialsamplers User Manual 36
2 Site Requirements and Specifications Site Requirements
Condensation
WARNING Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
CAUTION Condensation within the module
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.
Agilent InfinityLab LC Series Vialsamplers User Manual 37
2 Site Requirements and Specifications Physical Specifications
Physical Specifications
Table 1 Physical Specifications G7129A, G7129B, G7129C
Type Specification Comments
Weight 19 kg (41.9 lbs) w/o sample thermostat
Dimensions (height width depth)
320 x 396 x 468 mm (12.8 x 15.6 x 18.4 inches)
Line voltage 100 240 V~, 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, 5 %
Power consumption 350 VA, 350 W, 1195 BTU/h
Ambient operating tempera- ture
4 - 40 C (39 - 104 F); without sample ther- mostat up to 55 C (131 F)
Ambient non-operating tem- perature
-40 70 C (-40 158 F)
Humidity < 95 % r.h. at 40 C (104 F)1
1 If a sample thermostat is included the upper value for humidity can be reduced. Please check your lab conditions to stay beyond dew point values for noncondensing operation.
Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11
Permitted solvents Boiling point 56 C (133 F). Auto-ignition temperature 200 C (394 F).
Agilent InfinityLab LC Series Vialsamplers User Manual 38
2 Site Requirements and Specifications Performance Specifications
Performance Specifications
Performance Specifications (G7129A) Table 2 Performance Specifications G7129A
Type Specification Comments Method/Conditions
Injection range 0.1 100 L in 0.1 L incre- ments with 100 L up to 60 MPa 0.1 900 L in 0.1 L incre- ments with 900 L up to 40 MPa
Up to 1800 L with multiple draw (hardware modification required) Requires 900 L analytical head
Injection precision <0.25 % RSD of peak areas from 5 L to 100 L
Measured caffeine
Pressure range 0 60 MPa (0 600 bar, 0 8702 psi) 0 40 MPa (0 400 bar, 0 5801 psi)
for 900 L Analytical Head
Sample viscosity range
0.2 5.0 cP
Sample capacity 132 x 2 mL vial (two trays default) 100 x 2 mL vial (two classic trays optional) 36 x 6 mL vials (two trays optional)
Agilent InfinityLab LC Series Vialsamplers User Manual 39
2 Site Requirements and Specifications Performance Specifications
Carryover <0.004 % (40 ppm) with needle wash
Using the following conditions: Column: Agilent Pursuit XRs
3 C18, 2.0 x 50 mm Mobile Phase:
A: 0.1 % TFA in water B: 0.1 % TFA in acetoni-
trile Isocratic : %B = 40 % Flow rate: 0.5 mL/min Temperature: Ambient Wavelength: 257/4 nm Injection volume: 1 L Sample: 1200 ng/L Chlor-
hexidine for UV (dissolved with mobile phase A), 1 L injected and measured with Agilent UV detector
Wash solution: H2O with 0.1 % TFA (3 s)
Injection cycle time 18 s for draw speed 200 L/min Ejection speed: 200 L/min Injection volume: 1 L
Minimum sample vol- ume
1 L from 5 L sample in 100 L microvial, or 1 L from 10 L sample in 300 L microvial.
Needle height offset has to be adapted to ensure that needle doesnt touch vial bottom. Default needle height = 0 equates to 2 mm above the vial bottom.
Instrument Control LC & CE Drivers A.02.12 or above Instrument Control Framework (ICF) A.02.03 or above Instant Pilot (G4208A) with firm- ware B.02.19 or above Lab Advisor B.02.07 or above
For details about supported software versions refer to the compatibility matrix of your ver- sion of the LC & CE Drivers
Communication Controller Area Network (CAN),Local Area Network (LAN) ERI: ready, start, stop and shut-down signals
Table 2 Performance Specifications G7129A
Type Specification Comments Method/Conditions
Agilent InfinityLab LC Series Vialsamplers User Manual 40
2 Site Requirements and Specifications Performance Specifications
Maintenance and safety-related features
Extensive diagnostics, error detection and display with Agi- lent Lab Advisor software Leak detection, safe leak han- dling, leak output signal for shutdown of pumping system, and low voltages in major main- tenance areas
GLP features Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feed- back messages. Electronic records of maintenance and errors.
Housing All materials recyclable.
Table 2 Performance Specifications G7129A
Type Specification Comments Method/Conditions
Agilent InfinityLab LC Series Vialsamplers User Manual 41
2 Site Requirements and Specifications Performance Specifications
Performance Specifications (G7129B) Table 3 Performance Specifications G7129B
Type Specification Comments Method/Conditions
Injection range 0.1 20 L in 0.1 L increments (default) 0.1 40 L in 0.1 L increments (if 40 L loop is installed) 0.1 120 L in 0.1 L incre- ments with 1290 Infinity large volume injection kit (hardware modification required) 0.1 100 L in 0.1 L (if 100 L loop and 100 L head is installed)
Up to 1500 L (with 1400 L multi-draw kit and 100 L ana- lytical head)
up to 130 MPa (1300 bar, 18854 psi)
up to 60 MPa (600 bar, 8702 psi)
Injection precision <0.25 % RSD of peak areas from 5 L to 100 L
Measured caffeine
Pressure range Up to 130 MPa (1300 bar, 18854 psi)
Sample viscosity range
0.2 5.0 cP
Sample capacity 132 x 2 mL vial (two trays default) 100 x 2 mL vial (two classic trays optional) 36 x 6 mL vials (two trays optional)
Agilent InfinityLab LC Series Vialsamplers User Manual 42
2 Site Requirements and Specifications Performance Specifications
Carryover <0.004 % (40 ppm) with needle wash
Using the following conditions: Column: Agilent Pursuit XRs
3 C18, 2.0 x 50 mm Mobile Phase:
A: 0.1 % TFA in water B: 0.1 % TFA in acetoni-
trile Isocratic : %B = 40 % Flow rate: 0.5 mL/min Temperature: Ambient Wavelength: 257/4 nm Injection volume: 1 L Sample: 1200 ng/L Chlor-
hexidine for UV (dissolved with mobile phase A), 1 L injected and measured with Agilent UV detector
Wash solution: H2O with 0.1 % TFA (3 s)
Injection cycle time 18 s for draw speed 200 L/min Ejection speed: 200 L/min Injection volume: 1 L
Minimum sample vol- ume
1 L from 5 L sample in 100 L microvial, or 1 L from 10 L sample in 300 L microvial.
Needle height offset has to be adapted to ensure that needle doesnt touch vial bottom. Default needle height = 0 equates to 2 mm above the vial bottom.
Instrument control LC & CE Drivers A.02.12 or above Instrument Control Framework (ICF) A.02.03 or above Instant Pilot (G4208A) with firm- ware B.02.19 or above Lab Advisor B.02.07 or above
For details about supported software versions refer to the compatibility matrix of your ver- sion of the LC & CE Drivers
Communication Controller Area Network (CAN), Local Area Network (LAN), ERI: ready, start, stop and shut-down signals
Table 3 Performance Specifications G7129B
Type Specification Comments Method/Conditions
Agilent InfinityLab LC Series Vialsamplers User Manual 43
2 Site Requirements and Specifications Performance Specifications
Maintenance and safety-related features
Extensive diagnostics, error detection and display with Agi- lent Lab Advisor software Leak detection, safe leak han- dling, leak output signal for shutdown of pumping system, and low voltages in major main- tenance areas
GLP features Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feed- back messages. Electronic records of maintenance and errors.
Housing All materials recyclable.
Table 3 Performance Specifications G7129B
Type Specification Comments Method/Conditions
Agilent InfinityLab LC Series Vialsamplers User Manual 44
2 Site Requirements and Specifications Performance Specifications
Performance Specifications (G7129C) Table 4 Performance Specifications G7129C
Type Specification Comments Method/Conditions
Injection range 0.1 100 L in 0.1 L incre- ments with 100 L up to 80 MPa
Up to 1800 L with multiple draw (hardware modification required)
Injection precision <0.25 % RSD of peak areas from 5 L to 100 L
Measured caffeine
Pressure range 0 80 MPa (0 800 bar, 0 11603 psi)
Sample viscosity range
0.2 5.0 cP
Sample capacity 132 x 2 mL vial (two trays default) 100 x 2 mL vial (two classic trays optional) 36 x 6 mL vials (two trays optional)
Carryover <0.004 % (40 ppm) with needle wash
Using the following conditions: Column: Agilent Pursuit XRs
3 C18, 2.0 x 50 mm Mobile Phase:
A: 0.1 % TFA in water B: 0.1 % TFA in acetoni-
trile Isocratic : %B = 40 % Flow rate: 0.5 mL/min Temperature: Ambient Wavelength: 257/4 nm Injection volume: 1 L Sample: 1200 ng/L Chlor-
hexidine for UV (dissolved with mobile phase A), 1 L injected and measured with Agilent UV detector
Wash solution: H2O with 0.1 % TFA (3 s)
Injection cycle time 18 s for draw speed 200 L/min Ejection speed: 200 L/min Injection volume: 1 L
Agilent InfinityLab LC Series Vialsamplers User Manual 45
2 Site Requirements and Specifications Performance Specifications
Minimum sample vol- ume
1 L from 5 L sample in 100 L microvial, or 1 L from 10 L sample in 300 L microvial.
Needle height offset has to be adapted to ensure that needle doesnt touch vial bottom. Default needle height = 0 equates to 2 mm above the vial bottom.
Instrument Control LC & CE Drivers A.02.17 or above Instrument Control Framework (ICF) A.02.05 or above Instant Pilot (G4208A) with firm- ware B.02.22 or above InfinityLab LC Companion (G7108A) Lab Advisor B.02.10 or above
For details about supported software versions refer to the compatibility matrix of your ver- sion of the LC & CE Drivers
Communication Controller Area Network (CAN), Local Area Network (LAN), ERI: ready, start, stop and shut-down signals
Maintenance and safety-related features
Extensive diagnostics, error detection and display with Agi- lent Lab Advisor software Leak detection, safe leak han- dling, leak output signal for shutdown of pumping system, and low voltages in major main- tenance areas
GLP features Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feed- back messages. Electronic records of maintenance and errors.
Housing All materials recyclable.
Table 4 Performance Specifications G7129C
Type Specification Comments Method/Conditions
Agilent InfinityLab LC Series Vialsamplers User Manual 46
2 Site Requirements and Specifications Specifications of the Sample Cooler
Specifications of the Sample Cooler
The Agilent Infinity II Sample Cooler is a vapor-compression refrigeration system that uses a fluorinated greenhouse gas (HCF-134a) as the refrigerant. For information on carbon dioxide equivalency (CDE) and global warming potential (GWP), see the instrument label.
Table 5 Physical Specification of the Sample Cooler
Type Specification Comment
Weight < 6 kg (< 13.2 lbs)
Dimensions (height width depth) 205 x 340 x 370 mm (8.1 x 13.4 x 14.6 inches)
Refrigerant gas HFC-134a (0.042 kg) Ozone depletion poten- tial (ODP) = 0
Supply voltage 24 VDC
Current 10 A max.
Ambient operating temperature 4 40 C (39 104 F)
Ambient non-operating temperature -40 70 C (-40 158 F)
Humidity < 95 % r.h. at 40 C (104 F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN, CSA, UL
Installation category II, Pollution degree 2
For indoor use only
ISM Classification ISM Group 1 Class B According to CISPR 11
Agilent InfinityLab LC Series Vialsamplers User Manual 47
2 Site Requirements and Specifications Specifications of the Sample Cooler
CAUTION General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
The disposal or scrapping of the Sample Cooler or the Sample Thermostat must be carried out by a qualified disposal company.
All media must be disposed of in accordance with national and local regulations.
Please contact your local Agilent Service Center in regard to safe environmental disposal of the appliance or check www.agilent.com for more info.
Table 6 Performance Specifications of the Sample Cooler
Type Specifications
Operating principle High performance, low-energy consumption micro-compressor based cooler with ozone-friendly HFC-134a coolant (42 g), user-upgradable.
Temperature range from 4 C to 5 C below ambient
Temperature settable from 4 40 C in 1 increments
Temperature accuracy (<25 C, <50 % r.H.) 2 C to 6 C at a setpoint of 4 C
NOTE The Agilent Infinity II Sample Cooler is not available for trade sales anymore and has been replaced by the Agilent InfinityLab Sample Thermostat.
Agilent InfinityLab LC Series Vialsamplers User Manual 48
2 Site Requirements and Specifications Specifications of the Sample Thermostat
Specifications of the Sample Thermostat
The Agilent InfinityLab Sample Thermostat is the combination of an electric heater and a vapor-compression refrigeration system. It uses isobutane as a non-Freon refrigerant, which is harmless to the environment and does not affect the ozone layer and global warming but is combustible. Please adhere to the warnings listed in the manual.
Table 7 Physical Specifications of the Sample Thermostat
Type Specification Comments
Weight <6 kg (< 13.2 lbs)
Dimensions (height x width x depth)
205 x 340 x 370 mm (8.1 x 13.4 x 14.6 inches)
Refrigerant gas R600a (0.030 kg) Ozone depletion potential (ODP) =0 Global warming potential (GWP) =3
Supply voltage 24VDC
Current 10 A max.
Ambient operating temperature
4 40 C (39 104 F)
Ambient non-operat- ing temperature
-40 70 C (-40 158 F)
Humidity < 95 % r.h. at 40 C (104 F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN, CSA, UL
Installation category II, Pollution degree 2
For indoor use only
ISM Classification ISM Group 1 Class B According to CISPR 11
Agilent InfinityLab LC Series Vialsamplers User Manual 49
2 Site Requirements and Specifications Specifications of the Sample Thermostat
CAUTION General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
The disposal or scrapping of the Sample Cooler or the Sample Thermostat must be carried out by a qualified disposal company.
All media must be disposed of in accordance with national and local regulations.
Please contact your local Agilent Service Center in regard to safe environmental disposal of the appliance or check www.agilent.com for more info.
Table 8 Performance Specifications for the Sample Thermostat
Type Specifications
Operating principle High performance, low-energy consumption micro-compressor based cooler with natural R600a coolant (Butane 30 g), user-upgradable
Temperature range from 4 40 C
Temperature settable from 4 40 C in 1 increments
Temperature accuracy (<25 C, <50 % r.H.)
2 C to 6 C at a setpoint of 4 C
NOTE Minimum firmware revision for the Sample Thermostat is D.07.22.
Minimum LC driver revision for the Sample Thermostat is A.02.14.
Agilent InfinityLab LC Series Vialsamplers User Manual 50
2 Site Requirements and Specifications Specifications of the Integrated Column Compartment
Specifications of the Integrated Column Compartment
Physical Specifications Agilent InfinityLab LC Series Integrated Column Compartment (G7130A) Table 9 Physical Specifications (G7130A)
Type Specification Comment
Weight 1.8 kg (4.0 lbs)
Dimensions (height x width x depth)
86.5 x 396.0 x 106.5 mm (3.4 x 15.6 x 4.2 inches)
Maximum outside
Supply Voltage 24 V DC
Power consumption 110 W
Ambient operating temperature 4 55 C (39 131 F)
Ambient non-operating tem- perature
-40 70 C (-40 158 F)
Humidity < 95 % r.h. at 40 C (104 F)1
1 If a sample cooler is included the upper value for humidity can be reduced. Please check your lab condi- tions to stay beyond dew point values for noncondensing operation.
Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards IEC, EN, CSA, UL
Installation category II, Pollution degree 2
For indoor use only
ISM Classification ISM Group 1 Class B According to CISPR 11
Agilent InfinityLab LC Series Vialsamplers User Manual 51
2 Site Requirements and Specifications Specifications of the Integrated Column Compartment
Performance Specifications Agilent InfinityLab LC Series Integrated Column Compartment (G7130A) Table 10 Performance Specifications G7130A
Type Specification Comments
Temperature range 5 C above ambient to 80 C
Column capacity 2 columns up to 30 cm and 4.6 mm ID
Temperature stability 0.10 C at sensor
Temperature accuracy 0.8 K (0.5 K with calibration)
Warm up time 20 40 C in 5 min at sensor
Agilent InfinityLab LC Series Vialsamplers User Manual 52
3 Using the Module
Magnets 55 Turn on/off 56 Status Indicators 58 Vial Drawers and Trays 59 List of Drawers and Trays 59 Exchange Drawers 62 Insert a Vial Into the Sampler 65 Install the External Tray 67
Choice of Vials and Caps 70 Compatible Vials and Caps for the 2 mL Vial Drawers 70 Compatible Vials and Caps for the 6 mL Vial Drawer 73
Install the Optional Integrated Column Compartment 74 Unpacking the Unit 74 Install the Integrated Column Compartment 76 Install a Column in the ICC 82 Remove a Column from the ICC 86 Install the Column Shelf 88 Install the Column ID Upgrade Kit 92 Connect a Column Identification Tag to the Tag Reader 94
Using the Optional Integrated Column Compartment 96 Dashboard 96 Control Interface 97 Control 98 Method Parameters 99 Online Signals 101 Column Assignment 102 Column Tag Information Table 103 Using Column Identification Tags 104
Install the Optional Sample Cooler/Sample Thermostat 116 Unpacking the Unit 116 Install the Sample Cooler/Sample Thermostat 117
Using the Optional Sample Cooler/Sample Thermostat 123 Dashboard 123 Control Interface 124 Control 125 Temperature Mode 127 Online Signals 128
Agilent InfinityLab LC Series Vialsamplers User Manual 53
3 Using the Module Specifications of the Integrated Column Compartment
Reporting Sample Temperature 129 Operation Information 130
Transporting the Sampler 132 Prepare a Sampler Without Cooler/Thermostat for Transportation 132 Prepare a Sampler with Cooler/Thermostat for Transportation 133 Install the Transport Protection Foam 135
Agilent Local Control Modules 137
This chapter provides information on how to use the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 54
3 Using the Module Magnets
Magnets
1 Magnets in doors of pumps, autosamplers, detectors, and fraction collectors.
Agilent InfinityLab LC Series Vialsamplers User Manual 55
3 Using the Module Turn on/off
Turn on/off
This procedure exemplarily shows an arbitrary LC stack configuration.
1 2
Power switch: On
Agilent InfinityLab LC Series Vialsamplers User Manual 56
3 Using the Module Turn on/off
3 Turn instrument On/Off with the control software. 4
Power switch: Off
5
Agilent InfinityLab LC Series Vialsamplers User Manual 57
3 Using the Module Status Indicators
Status Indicators
This procedure exemplarily shows an arbitrary LC stack configuration. 1 The module status indicator indicates one of six possible module conditions:
Status indicators 1. Idle 2. Run mode 3. Not-ready. Waiting for a specific pre-run condition to be reached or completed. 4. Error mode - interrupts the analysis and requires attention (for example a leak or defective internal components). 5. Resident mode (blinking) - for example during update of main firmware. 6. Bootloader mode (fast blinking). Try to re-boot the module or try a cold-start. Then try a firmware update.
Agilent InfinityLab LC Series Vialsamplers User Manual 58
3 Using the Module Vial Drawers and Trays
Vial Drawers and Trays
List of Drawers and Trays Supported drawers for the module:
Drawer Combinations
Drawers can be installed in any combination enabling both 2 mL- and 6 mL-vials to be used simultaneously. The only exception is the usage of the classical drawer option (100 x 2 mL). This option can't combine with the other drawers.
Numbering of Vial Positions
The standard 2*66 vial drawers have 132 vial positions from P1-A1-P2-F11. However, when using two drawers, the numbering convention is slightly different. The vial positions of the right-hand drawer begin at position P2-A1 as follows:
Left-hand Drawer for 66 x 2 mL Vials: P1-A1 to P1-F11
Left-hand Drawer for 18 x 6 mL Vials: P1-A1 to P1-C6
Right-hand Drawer for 66 x 2 mL Vials: P2-A1 to P2-F11
Right-hand Drawer for 18 x 6 mL Vials: P2-A1 to P2-C6
Drawer for 50 x 2 mL Vials Classic Left: Vial 1-50
p/n Description
G7129-60010 Drawer for 66 x 2 mL Vials
G7129-60110 Drawer for 18 x 6 mL Vials
G7129-68210 Classic Vial Drawer Kit
G7129-60210 Classic Drawer for 50 x 2 mL Vials, Left
G7129-60220 Classic Drawer for 50 x 2 mL Vials, Right
G7129-60000 External Tray for 5 x 2 mL Vials
NOTE Install all drawers for best cooling performance.
Agilent InfinityLab LC Series Vialsamplers User Manual 59
3 Using the Module Vial Drawers and Trays
Drawer for 50 x 2 mL Vials Classic Right: Vial 51-100
External Tray 5-position: 201 205 Position
(The disposal tube is installed into the external tray by turning and pushing it into the backside of the hole position, No. 206)
Figure 14 Numbering of drawer position (left-hand Drawer for 66 x 2 mL Vials)
Figure 15 Numbering of drawer position (right-hand Drawer for 66 x 2 mL Vials)
Figure 16 Numbering of drawer position (left-hand Drawer for 18 x 6 mL Vials)
Figure 17 Numbering of drawer position (right-hand Drawer for 18 x 6 mL Vials)
P1-A1
P1-F11
P2-A1
P2-F11
P1-A1
P1-C6
P2-A1
P2-C6
Agilent InfinityLab LC Series Vialsamplers User Manual 60
3 Using the Module Vial Drawers and Trays
Figure 18 Numbering of drawer position (Drawer for 50 x 2 mL Vials Classic)
Figure 19 Numbering of tray position (External tray)
1
50
51
100
201 202 203
204 205 206
(Waste Disposal)
Agilent InfinityLab LC Series Vialsamplers User Manual 61
3 Using the Module Vial Drawers and Trays
Exchange Drawers
NOTE Do not operate the sampler without drawers installed.
NOTE Install all drawers for best cooling performance.
NOTE Do not mix standard and classic drawers.
Install classic drawer 1-50 to the left, classic drawer 51-100 to the right side.
Agilent InfinityLab LC Series Vialsamplers User Manual 62
3 Using the Module Vial Drawers and Trays
1 Open the doors of the module. 2 Remove the drawer.
a Pull the drawer out.
b Lift the front of the drawer.
c Lift the drawer out.
d Remove the drawer.
Agilent InfinityLab LC Series Vialsamplers User Manual 63
3 Using the Module Vial Drawers and Trays
3 Install the drawer.
a Insert the back of the drawer.
b Align the drawer.
c Push in the drawer.
4 Close the doors.
Agilent InfinityLab LC Series Vialsamplers User Manual 64
3 Using the Module Vial Drawers and Trays
Insert a Vial Into the Sampler
NOTE Attempting to insert a new vial into the sampler while the gripper arm is moving might lead to aborting the ongoing analysis.
1 Open the doors of the module.
NOTE For accessing the left drawer, it is sufficient to open the left door only.
2 Pull out the vial drawer into which you want to insert the vial.
3 Insert the vial into an appropriate position of the drawer. 4 Push the vial drawer back into place.
Agilent InfinityLab LC Series Vialsamplers User Manual 65
3 Using the Module Vial Drawers and Trays
5 Close the doors.
Agilent InfinityLab LC Series Vialsamplers User Manual 66
3 Using the Module Vial Drawers and Trays
Install the External Tray
Tools required Description Flathead screwdriver
Parts required p/n Description G7129-60000 External Tray for 5 x 2 mL Vials G1313-27302 Disposal tube
Preparations Turn off the sampler Open the doors Remove the samples
NOTE Keep foam and plastic cover in a safe place.
For best temperature performance, and if the external tray is not in use, it is best to cover the opening for the external tray with the original parts.
Agilent InfinityLab LC Series Vialsamplers User Manual 67
3 Using the Module Vial Drawers and Trays
1 Remove the foam and the plastic cover.
a Lift up the front part of the foam with a flathead screw driver.
b Remove the foam.
c Push out the plastic cover.
2 Install the external tray and the disposal tube.
a Mount the external tray in the mounting holes on the left side of the sampler and ensure that it is pushed in all the way.
b Install the disposal tube.
Agilent InfinityLab LC Series Vialsamplers User Manual 68
3 Using the Module Vial Drawers and Trays
3 Close the doors. 4 Configure the External Tray in the online view of the Chromatographic Data System: right-click on the sam- pler dashboard and select Modify> External Tray. In the dialog, select the External Tray installed check box.
Agilent InfinityLab LC Series Vialsamplers User Manual 69
3 Using the Module Choice of Vials and Caps
Choice of Vials and Caps
Compatible Vials and Caps for the 2 mL Vial Drawers
List of Compatible Vials and Caps
For reliable operation vials used with the autosampler must not have tapered shoulders or caps that are wider than the body of the vial. The vials in Crimp Top Vials on page 70, Snap Top Vials on page 71 and Screw Top Vials on page 71 and caps in Crimp Caps on page 72, Snap Caps on page 72 and Screw Caps on page 72 (shown with their Part numbers) have been successfully tested using a minimum of 15,000 injections with the autosampler.
Crimp Top Vials
p/n Description
5181-3375 Crimp Top Vial, 2 mL, clear glass, 100/Pack
5183-4491 Crimp Top Vial, 2 mL, clear glass, 1000/Pack
5182-0543 Crimp Top Vial, 2 mL, clear glass, write-on spot, 100/Pack
5183-4492 Crimp Top Vial, 2 mL, clear glass, write-on spot, 1000/Pack
5183-4494 Crimp Top Vial, 2 mL, clear glass, write-on spot, 100/Pack (silanized)
5181-3376 Crimp Top Vial, 2 mL, amber glass, write-on spot, 100/Pack
5183-4493 Crimp Top Vial, 2 mL, amber glass, write-on spot, 1000/Pack
5183-4495 Crimp Top Vial, 2 mL, amber glass, write-on spot, 100/Pack (silanized)
5182-0567 Crimp Top Vial, 1 mL, polypropylene, wide opening, 100/Pack
5183-4496 Crimp Top Vial, 1 mL, polypropylene, wide opening, 100/Pack (silanized)
9301-0978 Crimp top vial, 250 L, polypropylene, wide opening, 1000/Pack
Agilent InfinityLab LC Series Vialsamplers User Manual 70
3 Using the Module Choice of Vials and Caps
Snap Top Vials
Screw Top Vials
p/n Description
5182-0544 Snap Top Vial, 2 mL, clear glass, 100/Pack
5183-4504 Snap Top Vial, 2 mL, clear glass, 1000/Pack
5183-4507 Snap Top Vial, 2 mL, clear glass, 100/Pack (silanized)
5182-0546 Snap Top Vial, 2 mL, clear glass, write-on spot, 100/Pack
5183-4505 Snap Top Vial, 2 mL, clear glass, write-on spot, 1000/Pack
5183-4508 Snap Top Vial, 2 mL, clear glass, write-on spot, 100/Pack (silanized)
5182-0545 Snap Top Vial, 2 mL, amber glass, write-on spot, 100/Pack
5183-4506 Snap Top Vial, 2 mL, amber glass, write-on spot, 1000/Pack
5183-4509 Snap Top Vial, 2 mL, amber glass, write-on spot, 100/Pack (silanized)
p/n Description
5182-0714 Screw Cap Vials, 2 mL, clear glass, 100/Pack
5183-2067 Screw Top Vial, 2 mL, clear glass, 1000/Pack
5183-2070 Screw Top Vial, 2 mL, clear glass, 100/Pack (silanized)
5182-0715 Screw Top Vial, 2 mL, clear glass, write-on spot, 100/Pack
5183-2068 Screw Top Vial, 2 mL, clear glass, write-on spot, 1000/Pack
5183-2071 Screw Top Vial, 2 mL, clear glass, write-on spot, 100/Pack (silanized)
5182-0716 Screw Cap Vial, 2 mL, amber glass, write-on spot, 100/Pack
5183-2069 Screw Top Vial, 2 mL, amber glass, write-on spot, 1000/Pack
5183-2072 Screw Top Vial, 2 mL, amber glass, write-on spot, 100/Pack (silanized)
Agilent InfinityLab LC Series Vialsamplers User Manual 71
3 Using the Module Choice of Vials and Caps
Crimp Caps
Snap Caps
Screw Caps
p/n Description
5181-1210 Crimp Cap, silver aluminum, septum (clear PTFE/red rubber), 100/Pack
5183-4498 Crimp Cap, silver aluminum, septum (clear PTFE/red rubber), 1000/Pack
5181-1215 Crimp Cap, blue aluminum, septum (clear PTFE/red rubber), 100/Pack
5181-1216 Crimp Cap, green aluminum, septum (clear PTFE/red rubber), 100/Pack
5181-1217 Crimp Cap, red aluminum, septum (clear PTFE/red rubber), 100/Pack
p/n Description
5182-0550 Snap Cap, clear polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-3458 Snap Cap, blue polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-3457 Snap Cap, green polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-3459 Snap Cap, red polypropylene, septum (clear PTFE/red rubber), 100/Pack
p/n Description
5182-0717 Screw Cap, blue polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-0718 Screw Cap, green polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-0719 Screw Cap, red polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-0720 Screw Cap, blue polypropylene, septum (clear PTFE/silicone), 100/Pack
5182-0721 Screw Cap, green polypropylene, septum (clear PTFE/silicone), 100/Pack
5182-0722 Screw Cap, red polypropylene, septum (clear PTFE/silicone), 100/Pack
Agilent InfinityLab LC Series Vialsamplers User Manual 72
3 Using the Module Choice of Vials and Caps
Compatible Vials and Caps for the 6 mL Vial Drawer
Screw Top Vials and Caps
Crimp Top Vials and Caps
NOTE For reliable operation of the sampler, the combined height of the vial and cap should never exceed 40 mm.
p/n Description
9301-1377 Screw Top Vial, 6 mL, clear glass, flat bottom, 100/Pack
5188-5369 Screw Top Vial, 5 mL, clear glass, high recovery, 100/Pack
9301-1379 Screw Cap, 16 mm, w/o septum, for 6 mL vials, 100/Pack
9301-1378 Septum, PTFE/silicone, for 16 mm caps, 100/Pack
5188-2758 Septum, preslit, PTFE/silicone, for 16 mm caps, 100/Pack
p/n Description
9301-1419 Crimp Top Vial, 6 mL, clear glass, flat bottom, 100/Pack
9301-1425 Crimp Cap, silver aluminum, septum (PTFE/silicone), 100/Pack
Agilent InfinityLab LC Series Vialsamplers User Manual 73
3 Using the Module Install the Optional Integrated Column Compartment
Install the Optional Integrated Column Compartment
Unpacking the Unit
Damaged Packaging
Damaged Packaging
If the delivery packaging shows signs of external damage, please call your Agilent Technologies sales and service office immediately. Inform your service representative that the instrument may have been damaged during shipment.
Condensation
CAUTION "Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged.
Notify your Agilent sales and service office about the damage.
An Agilent service representative will inspect the instrument at your site and initiate appropriate actions.
CAUTION Condensation within the module
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.
Agilent InfinityLab LC Series Vialsamplers User Manual 74
3 Using the Module Install the Optional Integrated Column Compartment
Delivery Checklist
Ensure all parts and materials have been delivered with your module. The delivery checklist is shown below. For parts identification please check the illustrated parts breakdown in Integrated Column Compartment on page 288. Please report any missing or damaged parts to your local Agilent Technologies sales and service office.
Table 11 Delivery checklist for the Integrated Column Compartment (ICC)
Description Quantity
Integrated Column Compartment (G7130-60030 or G7130-60060) 1
Accessory Kit for ICC 1
Agilent InfinityLab LC Series Vialsamplers User Manual 75
3 Using the Module Install the Optional Integrated Column Compartment
Install the Integrated Column Compartment
For more information on recommended capillary types refer to Capillary Connections on page 278.
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch
Parts required # p/n Description 1 G7130-60030 Integrated Column Compartment, 3 L Heater
OR 1 G7130-60060 Integrated Column Compartment, 6 L Heater 1 Accessory Kit for ICC
(includes capillaries and column holder clips)
Preparations Switch off the pump. Switch off the sampler. Remove the column. Open the doors of the sampler.
WARNING Fire or explosion due to missing leak drainage
Solvents may run into the module if column shelf or Integrated Column Compartment are not installed.
Install either column shelf or Integrated Column Compartment before starting the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 76
3 Using the Module Install the Optional Integrated Column Compartment
1 Remove the column shelf.
a Press in the release buttons on the left and right side.
b Pull out the column shelf.
The column shelf is removed.
2 Install the Integrated Column Compartment.
a Push in the Integrated Column Compartment.
b Ensure that the release buttons are clicked back in their original position.
The Integrated Column Compartment is installed.
Click
Agilent InfinityLab LC Series Vialsamplers User Manual 77
3 Using the Module Install the Optional Integrated Column Compartment
3 Push the latch of the Integrated Column Compartment to the left.
4 Open the flap door.
5 Connect the injector valve (port 6) of the Vialsampler to the inlet port of the ICC heat exchanger with a suitable capillary (see Capillary Connections on page 278 for recommendations).
6 Connect a suitable capillary to the respective outlet port of the ICC heat exchanger (see Capillary Connections on page 278 for recommendations).
Agilent InfinityLab LC Series Vialsamplers User Manual 78
3 Using the Module Install the Optional Integrated Column Compartment
7 Tighten the capillary fitting to make the connection pressure tight.
NOTE Do not overtighten the fitting (maximum torque 2.5 Nm).
NOTE The use of capillaries with SL or SX fittings are recom- mended for a better handling.
8 Mount the column holders on the heat exchanger.
max. 2.5 Nm
Agilent InfinityLab LC Series Vialsamplers User Manual 79
3 Using the Module Install the Optional Integrated Column Compartment
9 Connect the column to the free end of the capillary coming from the ICC heat exchanger. For best perfor- mance, use capillaries with the InfinityLab Quick Con- nect Fitting.
a Turn the column onto the Quick Connect Fitting until the first resistance is observed.
b Close the lever to establish a pressure tight connec- tion.
10 Connect a suitable capillary to the outlet port of the col- umn. For best performance, use capillaries with the InfinityLab Quick Turn Fitting.
a Connect the capillary with the Quick Connect Fitting and tighten it until finger-tightness.
The fitting should be pressure tight up to 600 bar.
b If needed, tighten the fitting further with the help of two suitable wrenches.
Agilent InfinityLab LC Series Vialsamplers User Manual 80
3 Using the Module Install the Optional Integrated Column Compartment
11 Mount the column in the column holder clip(s). 12 Route the capillary coming from the column through the opening between the flap door and the ICC body towards the detector module.
NOTE Alternatively, the capillary can also be routed through the incisions at the right side of the ICC dedicated for this pur- pose.
13 Close the flap door until the latch is locked. 14 Close the doors.
15 Turn on the sampler and the pump.
Click
Agilent InfinityLab LC Series Vialsamplers User Manual 81
3 Using the Module Install the Optional Integrated Column Compartment
Install a Column in the ICC
For more information on recommended capillary types, refer to Capillary Connections on page 278.
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch
Parts required p/n Description Column (up to 30 cm length)
G7116-68003 Column Holder Lamella, 2/pk
Preparations Switch off the pump Switch off the sampler Open the doors of the sampler
NOTE A maximum of two columns can be installed.
1 Push the latch of the Integrated Column Compartment to the left.
2 Open the flap door.
Agilent InfinityLab LC Series Vialsamplers User Manual 82
3 Using the Module Install the Optional Integrated Column Compartment
3 Connect the injector valve (port 6) of the Vialsampler to the inlet port of the ICC heat exchanger with a suitable capillary (see Capillary Connections on page 278 for recommendations).
4 Connect a suitable capillary to the respective outlet port of the ICC heat exchanger (see Capillary Connections on page 278 for recommendations).
5 Tighten the capillary fitting to make the connection pressure tight.
NOTE Do not overtighten the fitting (maximum torque 2.5 Nm).
NOTE The use of capillaries with SL or SX fittings are recom- mended for a better handling.
6 Mount the column holders on the heat exchanger.
max. 2.5 Nm
Agilent InfinityLab LC Series Vialsamplers User Manual 83
3 Using the Module Install the Optional Integrated Column Compartment
7 Connect the column to the free end of the capillary coming from the ICC heat exchanger. For best perfor- mance, use capillaries with the InfinityLab Quick Con- nect Fitting.
a Turn the column onto the Quick Connect Fitting until the first resistance is observed.
b Close the lever to establish a pressure tight connec- tion.
8 Connect a suitable capillary to the outlet port of the col- umn. For best performance, use capillaries with the InfinityLab Quick Turn Fitting.
a Connect the capillary with the Quick Connect Fitting and tighten it until finger-tightness.
The fitting should be pressure tight up to 600 bar.
b If needed, tighten the fitting further with the help of two suitable wrenches.
Agilent InfinityLab LC Series Vialsamplers User Manual 84
3 Using the Module Install the Optional Integrated Column Compartment
9 Mount the column in the column holder clip(s). 10 Route the capillary coming from the column through the opening between the flap door and the ICC body towards the detector module.
NOTE Alternatively, the capillary can also be routed through the incisions at the right side of the ICC dedicated for this pur- pose.
11 Close the flap door until the latch is locked. 12 Close the doors.
Next Steps:
13 If applicable, connect the column identification tag to the tag reader (see Connect a Column Identification Tag to the Tag Reader on page 94).
14 Configure the connected column in the CDS (see Column Assignment on page 102).
Click
Agilent InfinityLab LC Series Vialsamplers User Manual 85
3 Using the Module Install the Optional Integrated Column Compartment
Remove a Column from the ICC
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch
Preparations Switch off the pump Switch off the sampler Open the doors of the sampler
1 Push the latch of the Integrated Column Compartment to the left.
2 Open the flap door.
3 Remove the column from the column holder clips. 4 Loosen the Quick Turn Fitting either with your fingers or with the help of two wrenches.
Agilent InfinityLab LC Series Vialsamplers User Manual 86
3 Using the Module Install the Optional Integrated Column Compartment
5 Remove the capillary from the outlet port of the column. 6 Lift up the lever to release the tension on the Quick Con- nect Fitting.
7 Unscrew and remove the Quick Connect Fitting. 8 Remove the column from the ICC.
9 Close the flap door until the latch is locked.
Click
Agilent InfinityLab LC Series Vialsamplers User Manual 87
3 Using the Module Install the Optional Integrated Column Compartment
Install the Column Shelf
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch
Parts required p/n Description G7129-60023 Column Shelf
Preparations Switch off the pump. Switch off the sampler. Open the doors of the sampler. Open the ICC flap door.
WARNING Fire or explosion due to missing leak drainage
Solvents may run into the module if column shelf or Integrated Column Compartment are not installed.
Install either column shelf or Integrated Column Compartment before starting the module.
1 Remove the capillary connecting the injection valve and the ICC heat exchanger.
2 Remove the column from the column holder clips.
Agilent InfinityLab LC Series Vialsamplers User Manual 88
3 Using the Module Install the Optional Integrated Column Compartment
3 Loosen the Quick Turn Fitting either with your fingers or with the help of two wrenches.
4 Remove the capillary from the outlet port of the column.
5 Lift up the lever to release the tension on the Quick Con- nect Fitting.
6 Unscrew and remove the Quick Connect Fitting.
7 Remove the column from the ICC. 8 Close the flap door until the latch is locked.
Click
Agilent InfinityLab LC Series Vialsamplers User Manual 89
3 Using the Module Install the Optional Integrated Column Compartment
9 Remove the Integrated Column Compartment.
a Press the release buttons on the left and right side.
b Pull out the Integrated Column Compartment.
The Integrated Column Comparment is removed.
10 Install the column shelf
a Push in the column shelf.
b Ensure that the release buttons are clicked back in their original position.
The column shelf is installed.
Click
Agilent InfinityLab LC Series Vialsamplers User Manual 90
3 Using the Module Install the Optional Integrated Column Compartment
11 Close the doors. 12 Turn on the sampler and the pump.
Agilent InfinityLab LC Series Vialsamplers User Manual 91
3 Using the Module Install the Optional Integrated Column Compartment
Install the Column ID Upgrade Kit
Parts required # p/n Description 1 G4751A Column ID Upgrade Kit for the Integrated Column Compartment
(includes the tag reader, 2 column identification tags and a tech- nical note)
OR 1 5067-6153 Column ID Tag Reader 1 5067-5917 InfinityLab Column Identification Tag
Preparations Power off the instrument. Install the Integrated Column Compartment (see Install the Integrated Column
Compartment on page 76).
NOTE To use the G4751A Column ID Upgrade Kit, the Agilent Integrated Column Compartment must be installed in the Vialsampler.
1 Place the two mounting clips onto the left side of the Vialsampler. Make sure of the correct orientation.
2 Mount the tag reader to the mounting clips.
3 Connect the cable of the tag reader to the respective socket on the back of the Vialsampler.
4 Fix the cable to the back of the sampler with the cable clips provided with the tag reader.
Agilent InfinityLab LC Series Vialsamplers User Manual 92
3 Using the Module Install the Optional Integrated Column Compartment
Next Steps:
5 Power on the instrument.
6 Place a column with an identification tag in the ICC and plug the tag into the tag reader (see Connect a Column Identi- fication Tag to the Tag Reader on page 94).
Agilent InfinityLab LC Series Vialsamplers User Manual 93
3 Using the Module Install the Optional Integrated Column Compartment
Connect a Column Identification Tag to the Tag Reader
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch
Parts required p/n Description 5067-5917 InfinityLab Column Identification Tag
Column (up to 30 cm length)
Preparations Install the Column ID Tag Reader (see Install the Column ID Upgrade Kit on page 92). Remove the column if installed (see Remove a Column from the ICC on page 86).
1 If the column is without a preinstalled identification tag, fit a new tag onto it by slipping the lanyard loop over one end of the column and pulling the cord tight through the plastic holder.
NOTE Once the lanyard is fixed, the tag can no longer be removed from the column.
2 Install the column in the ICC (see Install a Column in the ICC on page 82).
Agilent InfinityLab LC Series Vialsamplers User Manual 94
3 Using the Module Install the Optional Integrated Column Compartment
3 Route the cord to the left and plug the column identifi- cation tag into the corresponding socket of the tag reader.
4 Close the flap door until the latch is locked.
5 Close the doors. 6 Configure the connected column with identification tag in the CDS (see Column Assignment on page 102 and Using Column Identification Tags on page 104).
Click
Agilent InfinityLab LC Series Vialsamplers User Manual 95
3 Using the Module Using the Optional Integrated Column Compartment
Using the Optional Integrated Column Compartment
The following section describes how to operate the Agilent Integrated Column Compartment.
Dashboard When the Integrated Column Compartment (ICC) is installed in a Vialsampler and configured in the chromatography data system (CDS), a new tile appears on the dashboard, comprising a standalone graphical user interface (GUI) for the ICC.
1
2
3
4
5
6
7
8
1 ICC: Turn On/Off
2 ICC: Heater Status (On/Off)
3 ICC: Actual Temperature
4 ICC: Set Temperature
5 ICC: Column Information
6 ICC: Status Indicator
7 ICC: Early Maintenance Feedback (EMF) Status
8 ICC: Column Identification Tag Recognised
Agilent InfinityLab LC Series Vialsamplers User Manual 96
3 Using the Module Using the Optional Integrated Column Compartment
Control Interface Right-clicking on the sampler GUI opens the control interface, where control and method parameters can be edited, column assignment modified, and special commands executed.
Agilent InfinityLab LC Series Vialsamplers User Manual 97
3 Using the Module Using the Optional Integrated Column Compartment
Control The Control dialog box enables the user to change some control-specific settings for the ICC: Thermostat:
On: The ICC turns on and the system starts to regulate the temperature of the column oven towards the setpoint.
Off: The ICC turns off. At Power On:
Turn on thermostat: The ICC turns on automatically upon powering on the sampler.
Automatic Turn On: Turn on at: The ICC turns on automatically at the specified time and date.
Agilent InfinityLab LC Series Vialsamplers User Manual 98
3 Using the Module Using the Optional Integrated Column Compartment
Method Parameters The Method dialog box enables the user to change the method parameters relevant for the ICC: Temperature:
Not Controlled: The temperature of the column oven stays uncontrolled. At specified temperature (C): The temperature of the column oven is
regulated to stay at the setpoint temperature specified.
Unchanged: The temperature setting stays unchanged, meaning that the setting currently stored in the system is used by the method.
Valve Position/Column: This section provides information about the location of the column that is being used for the given method. Enforce column for run: Marking this checkbox restricts the execution of
the method to a specific column type, identified by its product number. Stoptime:
As Pump/Injector: The ICC uses the same stoptime setting as specified for the pump or the sampler.
After specified time (min): The stop time defines a time limit for the analysis, after which all method parameters are reset to initial values.
Posttime: Off: No post-run period is applied. For specified time (min): After the analysis, the system stays in the
post-run state for the specified time. Using a post-time period can be beneficial when the column needs to be re-equilibrated before starting the next analysis (for example, gradient elutions).
NOTE Valid inputs are 10 to 80 C in increments of 0.1 C.
NOTE For Stoptime and Posttime, valid inputs are 0.01 to 99999 minutes in increments of 0.1 min.
Agilent InfinityLab LC Series Vialsamplers User Manual 99
3 Using the Module Using the Optional Integrated Column Compartment
Enable Analysis: With any temperature: The analysis starts regardless of the actual column
oven temperature. When temperature is within: The analysis starts only when the actual
column oven temperature is within the specified range of the setpoint temperature.
Timetable: This feature enables the user to use programmed changes to control the oven temperature during the curse of the analysis.
Agilent InfinityLab LC Series Vialsamplers User Manual 100
3 Using the Module Using the Optional Integrated Column Compartment
Online Signals In the Online Signals tab of the CDS, the actual temperature of the column oven can be configured and plotted together with the other instrument actuals. This enables the user to have a better overview of how the temperature changes over time.
Agilent InfinityLab LC Series Vialsamplers User Manual 101
3 Using the Module Using the Optional Integrated Column Compartment
Column Assignment Clicking on Column Assignment in the control interface opens a dialog box, where the column information can be edited: Plumbing: The user can specify whether the column is located in the Upper or
Lower position of the ICC heat exchanger, or indicate that no column is present in the ICC at all (None).
Visualization: This section gives a visual representation of the column configuration in accordance with the information in the Plumbing section.
Column Tag Information: This section holds information on the physical properties of the columns installed in the ICC and their usage (see Column Tag Information Table on page 103 for more information). The fields of the table can be edited and the information therein used for reporting purposes even if no column identification tags are present. This is often referred to as soft tagging. When the Column ID Upgrade Kit is installed and a column identification tag is present, the content of the table corresponds to the information stored on the tag. This is referred to as hard tagging.
NOTE By default, the Column Tag Information table is truncated and only columns up to and including Injections are displayed. Clicking on >> in the top right corner displays the whole table.
Agilent InfinityLab LC Series Vialsamplers User Manual 102
3 Using the Module Using the Optional Integrated Column Compartment
Column Tag Information Table The Column Tag Information table is part of the Column Assignment dialog box, which can be prompted by clicking on the corresponding button of the control interface.
Agilent InfinityLab LC Series Vialsamplers User Manual 103
3 Using the Module Using the Optional Integrated Column Compartment
Using Column Identification Tags The Agilent InfinityLab Column Identification Tag is designed to enable the automatical detection of the columns installed in the column thermostat and help the user to track the usage of the columns in the laboratory. The column identification tag contains information about not only the physical properties of the column but also about its usage. Two types of information are stored: Static fields contain information on the physical characteristics of the column
such as the length, internal diameter and particle size. Typically, this information stays the same for the lifetime of the column.
Dynamic fields contain information on the usage of the column, such as the number of injections and maximum measured temperature. The information stored in the dynamic fields will be automatically updated each time a new analysis is performed with the column. These fields are read-only from the perspective of the user, meaning that their content cannot be manually edited or erased.
For pre-labeled Agilent columns, the static fields are pre-populated with information and set to read-only for compliance reasons, except for the Void Volume and Comment fields. The dynamic fields are blank by default and will be automatically edited by the system upon using the column.
On the contrary, custom column identification tags enable the user to edit the content for all static fields, as long as the tag is not sealed. This can be beneficial when using non-Agilent columns. After entering all the necessary information, the custom tag can be sealed to prevent any further modification of the content in the static fields. The content of the dynamic fields cannot be manually edited or erased, not even for the custom tags.
NOTE The information in the dynamic fields cannot be erased by any means.
NOTE When the column identification tag is sealed, the static fields will become write-protected and only the dynamic fields will remain available for updating. The only exceptions from this are the Void Volume and Commentstatic fields.
Agilent InfinityLab LC Series Vialsamplers User Manual 104
3 Using the Module Using the Optional Integrated Column Compartment
There are two ways to edit the content of custom column identification tags: One way is to type in all the necessary information manually, using the
Column Assignment tab of the control interface (see Column Assignment on page 102 for more information).
The other one is to import the data from a database (an existing catalog or inventory, or the Agilent columns guide). This option is available only for the OpenLab CDS ChemStation (C.02.07 SR2 or higher).
When plugging a new tag into the tag reader, the content that is stored on it will be automatically imported into the Edit Columns tab (only for OpenLab CDS ChemStation, see Using Column ID Tags with ChemStation on page 105 for more information) and the Column Assignment tab of the control interface (see Column Assignment on page 102 for more information).
Using Column ID Tags with ChemStation
1 In the Method and Run Control view of the Agilent OpenLAB CDS ChemStation Edition, select Columns from the Instrument menu. The Edit Columns dialog box is displayed. If the Edit Columns table is empty, go straight to Step 3.
Figure 20 The ChemStation Edit Columns table
2 Select a line in the table that contains column information as close as possible to the column you are adding. The selected line acts as a template for the new column.
NOTE It may take up to several seconds for the column tag to be read and the tables to be updated.
NOTE Column catalogs are available only in OpenLab CDS ChemStation Edition C.01.07 SR2 and above with drivers A.02.14 and above. Click the Catalog button to display a dialog box that allows you to choose how to load the catalog into the table.
Agilent InfinityLab LC Series Vialsamplers User Manual 105
3 Using the Module Using the Optional Integrated Column Compartment
3 Click the Insert button to insert a line above the currently selected line, or the Append button to add a line to the end of the table. The new line contains a copy of the information in the template line.
4 Click the Edit button to display a dialog box that allows you to edit the column-specific information such as Serial Number, Batch Number and Description.
5 Add or edit the other column-specific information (for example, maximum pressure, maximum temperature, length, diameter, particle size) in the Edit Columns table.
6 If the column is installed and will be used in the Multicolumn Thermostat, select YES in the Installed column.
7 Click the Plumbing button. The Column Assignment dialog box is displayed.
Figure 21 Column Assignment dialog box
Agilent InfinityLab LC Series Vialsamplers User Manual 106
3 Using the Module Using the Optional Integrated Column Compartment
The Column Assignment dialog box allows you to specify and review detailed information about the columns attached to each position in the column compartment. The information in the Column Assignment dialog box is in three sections: 1 The Plumbing section contains a table where you can specify the plumbing
assignment for each valve position. 2 The Visualization section gives a visual representation of the Multicolumn
Thermostat configuration. 3 The Column Tag Information table shows the information stored on the
column tag(s) of the installed column(s). For more details, see The Column Tag Information Table on page 113.
8 Click in the Import column of an empty line in the Column Tag Information table. The list of columns from the ChemStation's Edit Columns table is displayed.
Figure 22 The list of columns from the ChemStation's Edit Columns table
To reduce the list to only those columns that are marked as Installed, mark the Only show installed Columns check box.
Agilent InfinityLab LC Series Vialsamplers User Manual 107
3 Using the Module Using the Optional Integrated Column Compartment
9 Select the column information to import from the list of columns and click OK. The column information is imported into the Column Tag Information table.
10 Click the >> button at the top right of the Column Tag Information table to show the hidden table columns. Use the horizontal scroll bar to access the columns at the right of the table.
11 If all the information for the column is correct, click the Ok/Write Tag button to write the information to the column ID tag. The information in the ChemStation's Edit Columns table is also updated.
12 You can edit the information on the column ID tag using the ChemStation Edit Columns table. When you have finished editing the information, repeat steps 7 to 9 to update the information in the tag.
NOTE It may take several seconds before the information appears in the Column Tag Information table.
NOTE The characters semicolon (;), single quote (') and double quote (") are invalid for the Column Tag Information table. If these characters are used in any field of the table, an error is displayed, and OK/Write Tag button is disabled. You must delete all invalid characters before you can write the data to the tag.
The Description field is limited to 32 characters in the Column Tag Information table.
NOTE At this stage, you can assign a color to the column using the drop-down list in the Color Code column; this information is displayed in the Visualization panel, but is not written to the tag.
Agilent InfinityLab LC Series Vialsamplers User Manual 108
3 Using the Module Using the Optional Integrated Column Compartment
13 When you are sure that you will not make any more changes to the information in the tag, you can irrevocably seal the tag to set all static fields to read-only. Right-click in the appropriate line in the Column Tag Information table and select Seal Column Tag from the context menu.
The sealed column is shown in the Column Tag Information table with the icon at the beginning of the row. In the Edit Columns table of the ChemStation, it is shown with Sealed in the Tag column.
Figure 23 The Column Tag Information table context menu
NOTE Once a column ID tag has been sealed, the static fields cannot be edited.
Until the column ID tag has been sealed, you can delete all information in static fields on the tag using the Clear Column Tag Information command from the context menu.
Agilent InfinityLab LC Series Vialsamplers User Manual 109
3 Using the Module Using the Optional Integrated Column Compartment
The Plumbing button of the ChemStation Edit Columns table displays the Column Assignment dialog box, which can also be displayed by selecting Column Assignment from the context menu of the column compartment Dashboard panel in the ChemStation's Method and Control view.
Figure 24 Column Assignment dialog box overview
The Column Assignment dialog box has three sections that give you information about the column: The Plumbing section shows the valve connections to the positions in the
column compartment.
NOTE Make the connections to give the shortest distances between the valve ports and the columns, and use a logical order (left column 1 to port 1-1, left column 2 to port 2-2 and so on). Avoid leaving unused ports between used ones.
Agilent InfinityLab LC Series Vialsamplers User Manual 110
3 Using the Module Using the Optional Integrated Column Compartment
The Visualization section gives a visual representation of the configuration of the columns in the column compartment; the columns are color coded. Place the mouse cursor over a column to display a tooltip of the
column information from the column ID tag. Click on a column to highlight the column information in the Column
Tag Information table. The Column Tag Information shows the information in the column ID tags
for all columns in the configuration, including their location in the column compartment and their color codes.
The column compartment panel of the Dashboard in the ChemStation's Method and Control view also allows you a quick view of the column ID tag information.
Figure 25 The dashboard panel of the column compartment
Place the mouse cursor over the column ID tag icon ( ); the tooltip shows the information currently stored on the column ID tag. Note that the column ID tag icon changes according to its state as described in The Column Tag Information Table on page 113.
Agilent InfinityLab LC Series Vialsamplers User Manual 111
3 Using the Module Using the Optional Integrated Column Compartment
Availability of Column Information
The ChemStation Columns Table
The Edit Columns dialog box, which you access using the Columns command of the Instrument menu in Method and Run Control view, shows the detailed information about the columns attached to each position in the device. The table contains all the column-specific information stored in the column ID tags (see Column Tag Information on page 114) plus the following possible additional columns:
The table includes not only the columns that are installed in the column thermostat device(s), but also the inventory of other available columns, for example, columns that have been used in the past. The ChemStation also provides a catalog of column types, which you can load into the Edit Columns table to act as templates for other columns.
Installed YES when the column is installed in a device. no when the column is not installed in a device.
Location Shows the location in the device to which the plumbing of the valve position leads.
Dev. Serial# Present only for a valve thermostat cluster. Shows the serial number of the device that contains the column.
Tag Shows if the column has column ID tag (Used), if it sealed (Sealed) or the column has no tag (empty).
Agilent InfinityLab LC Series Vialsamplers User Manual 112
3 Using the Module Using the Optional Integrated Column Compartment
The Column Tag Information Table
The Column Assignment dialog box, which you access using the Column Assignment command of the Dashboard context menu, shows the detailed information about the columns attached to each position in the device. The dialog box includes the Column Tag Information table which contains all the column-specific information stored on the column ID tags (see Column Tag Information on page 114) plus the following possible additional columns:
By default, only the Column Tag Information table columns up to and including the Injection column are displayed. Click the >> button at the top right of the table to show the full table.
Tag Status shows the status of the column ID tag: Empty: The position is empty or has a column without a column
ID tag.
: A column with a column ID tag is installed at this position.
: A column with a sealed column ID tag is installed at this position.
: A column with a pre-labeled column ID tag supplied by Agilent Technologies is installed at this position.
Location shows the location in the device to which the plumbing of the valve position leads. For a valve thermostat cluster, the Column Host (the device where the column is installed) is also shown.
Color Code shows the color representing the column currently occupying the valve position.
Import present only when the CDS is an Agilent OpenLAB CDS ChemSta- tion Edition. Click the icon to display a dialog box listing all the columns entered in the ChemStation's Edit Columns table, from which you can select appropriate information to import.
Agilent InfinityLab LC Series Vialsamplers User Manual 113
3 Using the Module Using the Optional Integrated Column Compartment
Column Tag Information
The column ID tag contains the following information:
Field Description Type Read/Write permission pre-labeled tag
Read/Write permission custom tag
Description A description of the column. Static Read Write
Length The length of the column in mm. Static Read Write
Diameter The internal diameter of the column in mm. Static Read Write
Particle Size The particle size of the column packing mate- rial in m.
Static Read Write
Maximum Pressure The maximum pressure supported by the col- umn.
Static Read Write
Number of Injections The number of injections that have been made on the column.
Dynamic Read Read
Product Number The product number of the column. Static Read Write
Maximum Measured Temperature
The highest temperature (in C) experienced by the column to date.
Dynamic Read Read
Maximum Tempera- ture
The safe maximum operating temperature of the column (in C).
Static Read Write
Minimum pH The minimum pH supported by the column. Static Read Write
Maximum pH The maximum pH supported by the column. Static Read Write
Void Volume (mL) The void volume of the column and fittings. Static Write Write
First Injection The date and time of the first injection onto the column.
Dynamic Read Read
Recent Injection The date and time of the most recent injection onto the column.
Dynamic Read Read
Manufacturing Date The date of manufacture of the column. Static Read Write
Agilent Column Whether or not the column was supplied by Agilent Technologies.
Static Read Write
Serial Number The serial number of the column. Static Read Write
Batch Number The batch number of the column. Static Read Write
Agilent InfinityLab LC Series Vialsamplers User Manual 114
3 Using the Module Using the Optional Integrated Column Compartment
Tag Sealed Whether or not all static fields except Comment and Void Volume are set irrevocably to read-only.
Static Read Write
Comment A user-generated comment about the column. Static Write Write
Field Description Type Read/Write permission pre-labeled tag
Read/Write permission custom tag
Agilent InfinityLab LC Series Vialsamplers User Manual 115
3 Using the Module Install the Optional Sample Cooler/Sample Thermostat
Install the Optional Sample Cooler/Sample Thermostat
Unpacking the Unit
Damaged Packaging
If the delivery packaging shows signs of external damage, please call your Agilent Technologies sales and service office immediately. Inform your service representative that the instrument may have been damaged during shipment.
Delivery Checklist
Ensure that all parts and materials have been delivered with your module. The delivery checklist is shown below. For parts identification, please check the illustrated parts breakdown in Sample Thermostat Upgrade on page 287. Please report any missing or damaged parts to your local Agilent Technologies sales and service office.
CAUTION "Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged.
Notify your Agilent sales and service office about the damage.
An Agilent service representative will inspect the instrument at your site and initiate appropriate actions.
Table 12 Delivery checklist for the Sample Thermostat
Description Quantity
Sample Thermostat (G7167-60101) 1
Condensate Drainage Kit (5067-6208) 1
Declaration of Conformity 1
Customer Letter 1
NOTE The Agilent Infinity II Sample Cooler is not available for trade sales anymore and has been replaced by the Agilent InfinityLab Sample Thermostat.
Agilent InfinityLab LC Series Vialsamplers User Manual 116
3 Using the Module Install the Optional Sample Cooler/Sample Thermostat
Install the Sample Cooler/Sample Thermostat
Tools required p/n Description 8710-0899 Screwdriver Pozidrive Shaft
(for the Sample Cooler) 5182-3466 Torx screwdriver T10
(for the Sample Thermostat) OR 5023-3089 Torx key set
Parts required # p/n Description 1 Sampler 1 G7167-60005 Sample Cooler
OR 1 G7167-60101 Sample Thermostat 1 Power cord 1 5067-6208 Condensate Drainage Kit
Preparations Sampler is installed in the stack.
WARNING Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample Thermostat and laboratory.
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m3 (1 m3 for every 8 g of R600a refrigerant inside of the Sample Thermostat).
Ensure adequate ventilation: typical air exchange of 25 m3/h per m2 of laboratory floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do not block the openings on the circumference of the Sample Thermostat.
WARNING Flammable refrigerant used
When handling, installing and operating the Sample Thermostat, care should be taken to avoid damage to the refrigerant tubing or any part of the Sample Thermostat.
Agilent InfinityLab LC Series Vialsamplers User Manual 117
3 Using the Module Install the Optional Sample Cooler/Sample Thermostat
CAUTION Condensate inside the Sample Cooler/Sample Thermostat
Damage to the electronics of the module
After installation of the Sample Cooler/Sample Thermostat, wait at least 30 min before switching on the module.
Make sure there is no condensate inside the module.
WARNING In the event of a damage
Keep open fire or sources of ignition away from the device.
Ventilate the room for several minutes.
Do not use the Sample Thermostat any more.
NOTE Do not open the Sample Thermostat. There are no serviceable parts inside.
NOTE If the Sample Cooler/Sample Thermostat is disconnected from the power supply, wait for at least five minutes before replugging and switching on the compressor again.
NOTE Even under average humidity conditions, a significant amount of condensed water gathers every day. A suitable container must be provided and emptied regularly to avoid overflow.
NOTE For best performance of the Sample Cooler/Sample Thermostat, all drawers must be installed in the sampler. For the Multisampler, use dummy drawers (G4267-60024) if no full hotel configuration is needed.
NOTE For the Sample Cooler installation in a sampler, the serial number of the Sample Cooler must be DEBAT02001 or higher.
NOTE Depending on the ambient conditions in the lab, the amount of condensate can vary from 200 mL to 2 L per day. Do not fill waste containers for the condensate to the top. Regularly empty the waste container.
Agilent InfinityLab LC Series Vialsamplers User Manual 118
3 Using the Module Install the Optional Sample Cooler/Sample Thermostat
1 Ensure that the power switch on the front of the module is OFF (switch stands out).
2 Disconnect the power cable from the sampler.
3 Loosen the four screws on the rear of the module. 4 Remove the sheet metal back cover of the sampler.
Agilent InfinityLab LC Series Vialsamplers User Manual 119
3 Using the Module Install the Optional Sample Cooler/Sample Thermostat
5 Slide the Sample Cooler/Sample Thermostat halfway into the sampler. WARNING
Module is partially energized when switched off, as long as the power cord is plugged in. Repair work at the module can lead to personal injuries, e.g. shock hazard, when the cover is opened and the mod- ule is connected to power.
Make sure that it is always possible to access the power plug.
Do not use the Sample Cooler/Sample Thermostat if it is not operating correctly or has been damaged. Dis- connect it from the power supply and call your local service center.
Remove the power cable from the module before open- ing the cover.
Do not connect the power cable to the module while the covers are removed.
If the Sample Cooler/Sample Thermostat is discon- nected from the power supply, you should wait for at least five minutes before switching on the compressor.
CAUTION Damaged electronics
To avoid damages of the electronics of the module make sure the power cords are unplugged before dis- connecting or reconnecting the sampler to the Sample Cooler/Sample Thermostat cables.
6 Connect the power cable and the data cable to the Sam- ple Cooler/Sample Thermostat.
Agilent InfinityLab LC Series Vialsamplers User Manual 120
3 Using the Module Install the Optional Sample Cooler/Sample Thermostat
CAUTION Damage to the cables
Do not bend or pinch the cables.
Make sure that the Sample Cooler/Sample Thermostat fits perfectly in the sampler.
7 Slide the Sample Cooler/Sample Thermostat all the way into the sampler.
8 Fix the Sample Cooler/Sample Thermostat with the four screws.
9 Use a bubble level to check the leveling of the sampler.
NOTE To ensure adequate drainage for condensate, the module should be operated in a proper horizontal position.
10 Adhere a tubing holder clamp to the side of the cooler/thermostat where the drain pipe is situated. Ensure a distance of 60 mm from the bottom edge.
6 0
m m
Agilent InfinityLab LC Series Vialsamplers User Manual 121
3 Using the Module Install the Optional Sample Cooler/Sample Thermostat
11 Mount the condensate tubing assembly with the y-con- nector on the drainpipe and fix the venting tube in the tubing holder clamp.
NOTE The condensate handling system should be installed in a way that it allows a continuous slope for the drained liquid. Horizontal or uphill sections may hinder the drainage.
12 Ensure that the tubing runs straight into the waste con- tainer without any bends or joints and that it is free end does not immerse in the liquid.
NOTE For more information, refer to Leak and Waste Handling on page 27.
CAUTION Damage to the Sample Cooler/Sample Thermostat
Wait at least 30 min before switching on the compres- sor of the cooler/thermostat.
This allows the refrigerant and system lubrication to reach equilibrium.
13 Connect the power cable to the power connector at the rear of the module.
14 Configure the Sample Cooler/Sample Thermostat in the CDS.
60 mm
1 00
m m
Agilent InfinityLab LC Series Vialsamplers User Manual 122
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Using the Optional Sample Cooler/Sample Thermostat
The following section describes how to operate the Agilent Infinity II Sample Cooler and the Agilent InfinityLab Sample Thermostat using the Multisampler as an example for the hosting sampler. The operation principle is the same for any other Infinity II sampler type.
Dashboard The status indicator of the Sample Cooler/Sample Thermostat is incorporated in the graphical user interface (GUI) of the hosting sampler, which appears automatically when the unit is configured in the chromatography data system (CDS). When the cooler/thermostat is turned on, the set temperature and the actual temperature are also displayed.
1
2
3
4
5 6
7
8 9
10 11
1 Sampler: Turn On/Off 2 Sampler: Analytical head configuration 3 Sampler: Injection volume 4 Sampler: Hotel configuration 5 Sampler: Status indicator 6 Sampler: Early Maintenance Feedback (EMF) status 7 Sampler: Multiwash installed 8 Cooler/Thermostat: Status indicator (Off) 9 Cooler/Thermostat: Status indicator (On) 10 Cooler/Thermostat: Set temperature 11 Cooler/Thermostat: Actual temperature
Agilent InfinityLab LC Series Vialsamplers User Manual 123
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Control Interface Right-clicking the sampler GUI will prompt the control interface, where control and method parameters can be edited, configuration modified, and special commands executed.
NOTE The actual temperature may deviate from the set temperature by up to 3 C, depending on the temperature setting and ambient conditions.
NOTE If the actual temperature differs by more than 2 C from the set temperature, a yellow highlight is visible around the temperature reading. This, however, will not prevent the system from starting a new analysis, unless the Enable Analysis> Temperature within +/- 2 C function is selected.
Agilent InfinityLab LC Series Vialsamplers User Manual 124
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Control With the Sample Cooler/Sample Thermostat installed, the Control dialog box of the hosting Infinity II sampler will include the following cooler/thermostat-specific control options: At Power On:
Turn On Thermostat: The cooler/thermostat turns on automatically upon powering on the sampler.
Thermostat: On: The cooler/thermostat turns on and the system starts to regulate the
temperature inside the sample space towards the setpoint.
Off: The cooler/thermostat turns off. Enable Analysis
With any temperature: The analysis starts regardless of the actual temperature inside the sampler.
Temperature within +/- 2 C: The analysis starts only when the actual temperature is within the 2 C range of the setpoint temperature.
NOTE For the Sample Cooler, the set temperature must be at least 5 C below ambient for proper temperature control.
NOTE The Enable Analysis control setting is available since LC & CE drivers A.02.19.
NOTE The Temperature within +/- 2 C option is only available for the Sample Thermostat.
Agilent InfinityLab LC Series Vialsamplers User Manual 125
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Agilent InfinityLab LC Series Vialsamplers User Manual 126
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Temperature Mode Selecting Modify> Temperature Mode in the Control Interface will prompt a dialog box, where the temperature control mode can be switched between being a method parameter or a system (control) setting: Constant Temperature Mode: The temperature control mode is defined as a
system (control) setting, meaning that the temperature setting is independent of the method parameters. The temperature stays constant for all methods within a given sequence. This control mode is the default option and recommended for most applications.
Variable Temperature Mode: The temperature control mode is defined as a method parameter, meaning that the temperature setting is part of the method parameters. The temperature can change from method to method within a given sequence. This control mode is not recommended for most analytical workflows but might be used for some special applications, such as degradation studies.
Before using the Variable Temperature Mode setting, here are some hints and tips to consider: Changing the temperature setting from one method to another will affect all
samples inside the sampler. Depending on the extent of the temperature change, it could take up to a
couple of hours until the sample temperature stabilizes at the new setpoint (for example, from 4 to 40 C or vice versa).
It might be beneficial to use the Temperature within +/- 2 C function; otherwise, the next run will start without waiting for the new setpoint being reached.
NOTE For modifying the temperature mode, LC & CE drivers A.02.12 or higher are required. If the system is run on an earlier driver version, the temperature mode is defined as a system setting.
Agilent InfinityLab LC Series Vialsamplers User Manual 127
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Online Signals In the Online Signals tab of the CDS, the actual temperature of the sample space can be configured and plotted together with the other instrument actuals. This enables the user to have a better overview of how the temperature changes over time.
Agilent InfinityLab LC Series Vialsamplers User Manual 128
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Reporting Sample Temperature The actual and setpoint temperature can be included in the analysis report. For this, the Samples> Advanced Run Information field must be included in the report template.
Figure 26 Finding the Advanced Run Information setting in OpenLab CDS 2.4.
Figure 27 Reporting actual and setpoint temperature using theAdvanced Run Information setting.
NOTE For OpenLab CDS ChemStation, this option is only available in Intelligent Reporting.
Agilent InfinityLab LC Series Vialsamplers User Manual 129
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Operation Information
Reaching Setpoint Temperature
Depending on the ambient conditions and the sampler configuration (for example, hotel configuration for the Multisampler), reaching the setpoint temperature can take from 30 min up to a couple of hours.
Condensate Formation
Operating the cooler/thermostat at temperatures below ambient results in condensate formation. This condensed water is collected in the base plate of the cooler/thermostat and drained through the drainpipe at the back of the unit. The container for condensate collection should be regularly emptied to ensure the proper functioning of the system.
Dew Formation
Setting the cooler/thermostat from a lower to a higher temperature setpoint, or just simply turning it off, can result in dew formation on the internal surfaces of the sampler. This is normal and should cease after a couple of hours at the most.
NOTE Reaching the 4 C setpoint from an ambient temperature of 22 C takes about 45 min for the Vialsampler (G7129A/B/C or G7157A), as well as for the Multisampler (G7167A/B, G5668A, or G4767A) with a single 2H drawer installed.
NOTE This relatively slow ramping down of the temperature is necessary to avoid ice formation.
NOTE For the best performance of the Sample Cooler/Sample Thermostat, all drawers must be installed in the sampler. For the Multisampler, use dummy drawers if no full hotel configuration is needed.
NOTE If the container is overfilled or the condensate tubing is blocked, the condensate sensor is triggered, rendering the HPLC system to enter the error state (see Sample temperature control switched off due to condensate on page 213).
NOTE Depending on the ambient conditions in the lab, the amount of condensate can vary from 200 mL to 2 L per day. Waste containers for the condensate should not be filled to the top. The waste container must be emptied regularly.
Agilent InfinityLab LC Series Vialsamplers User Manual 130
3 Using the Module Using the Optional Sample Cooler/Sample Thermostat
Frequent Door/Drawer Opening
Opening the door(s) and/or the sample drawers frequently can compromise the temperature stability, as fresh warm and humid air will enter each time. In a highly humid environment, this could also lead to the formation of significant amounts of condensate on the internal surfaces of the sampler.
Ice Formation
The Sample Cooler/Sample Thermostat was designed to operate without the risk of icing. In an unlikely event of ice formation, turn off the cooler/thermostat and wait until it defrosts.
Shutting Down When the Sample Cooler/Sample Thermostat needs to be turned off for the night or a longer period, the following best practices are recommended: Remove all sample containers and/or vials from the sampler. Let the system reach the ambient temperature. Opening the door(s) of the
sampler facilitates this process. Remove any condensate that might appear on the sample drawers or the
internal surfaces of the sampler. Make sure that all condensate is removed from the cooler/thermostat.
NOTE Do not use mechanical devices or other means to accelerate the defrosting process.
NOTE Gently tapping on the sides of the sampler facilitates the condensate removal. Tilting the module towards its right back corner is not recommended as it can damage the internal parts.
Agilent InfinityLab LC Series Vialsamplers User Manual 131
3 Using the Module Transporting the Sampler
Transporting the Sampler
Prepare a Sampler Without Cooler/Thermostat for Transportation
1 Remove all vials from the vial drawers. 2 Move the transport assembly to the park position using Instant Pilot or Lab
Advisor, see Park Arm on page 186. 3 Turn off the sampler. 4 Install the Transport Protection Foam, see Install the Transport Protection
Foam on page 135.
When The module needs to be transported or relocated.
WARNING Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
CAUTION Unsecured transportation
Mechanical damage
Secure the transport assembly before transporting the sampler.
Agilent InfinityLab LC Series Vialsamplers User Manual 132
3 Using the Module Transporting the Sampler
Prepare a Sampler with Cooler/Thermostat for Transportation
When The module needs to be transported or relocated.
WARNING Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
CAUTION Condensate inside the cooler or thermostat
Damage to the electronics
Unplug the power cords.
Drain off all condensate before dismounting the sample cooler or thermostat.
Make sure that there is no condensate left.
CAUTION Unsecured transportation
Mechanical damage
Secure the transport assembly before transporting the sampler.
NOTE Moving the sampler with the Sample Cooler/Sample Thermostat installed is possible for short distances (for example, from one workbench to another). For longer transportation, remove the cooler/thermostat from the sampler and handle the units separately.
Agilent InfinityLab LC Series Vialsamplers User Manual 133
3 Using the Module Transporting the Sampler
1 Turn off the Sample Cooler/Sample Thermostat. 2 Remove condensate.
a Place a suitable container underneath the outlet pipe. b Remove the drainage tube. c Gently tap the sides of the sampler several times to facilitate the drainage
of the condensate from the system.
3 Remove all vials from the vial drawers. 4 Move the transport assembly to the park position using Instant Pilot or Lab
Advisor, see Park Arm on page 186. 5 Turn off the sampler. 6 Install the Transport Protection Foam, see Install the Transport Protection
Foam on page 135. 7 Remove the Sample Cooler/Sample Thermostat from the sampler if needed,
see Replace the Sample Cooler/Sample Thermostat on page 270.
NOTE Do not tilt the module to avoid damage to the internal parts.
Agilent InfinityLab LC Series Vialsamplers User Manual 134
3 Using the Module Transporting the Sampler
Install the Transport Protection Foam
When To secure the transport arm before transporting or shipping the sampler.
Parts required p/n Description G7129-40050 Transport Protection Foam
Preparations All vials are removed from the vial drawers.
1 Open the doors of the module. 2 Move the gripper arm into the park position (see Park Arm on page 186) and turn off the sampler.
3 Place the Transport Protection Foam into the sampler. 4 Push the Transport Foam in until the endpoint.
Agilent InfinityLab LC Series Vialsamplers User Manual 135
3 Using the Module Transporting the Sampler
5 Ensure that the foam is snapped behind the upper part of the metal frame.
6 Close the doors.
Agilent InfinityLab LC Series Vialsamplers User Manual 136
3 Using the Module Agilent Local Control Modules
Agilent Local Control Modules
Agilent InfinityLab Companion G7108AA
The Agilent InfinityLab Companion gives you complete control, system monitoring, signal plotting and diagnostic capabilities for a wide range of LC system modules.
The instrument control solution is available as full package including all hardware and accessories, but can also be used on your own mobile devices like tablets, mobile phones and other electronic equipment.
Combining the conveniences of the Agilent Instant Pilot features with state of the art mobile technology, the Agilent InfinityLab Companion gives you maximum flexibility and ease of use to control and monitor your LC system modules. Features: Complete local control and monitoring of Agilent Infinity II Prime LC Modules Excellent usability and ease of use through a user interface specifically
tailored for mobile devices - simple, intuitive touch-enabled and visual controllable.
High flexibility through a modern Bring your own device approach. Connection between LC module and mobile device either wireless via WLAN or wired over USB cable (with full package).
Convenient, ergonomic operation either handheld or attached to a module at the stack with newly developed, secure tablet holder (included in the full package).
Preconfigured tablet with all required software already installed (included in the full package).
Centerpiece of the solution is a USB dongle that activates the complete intelligence of the InfinityLab Companion on the instrument stack.
The InfinityLab Companion provides: fast and direct control in front of the instrument a clear overview of the system status control functionalities access to method parameters and sequences a logbook showing events from the modules diagnose tests
Agilent InfinityLab LC Series Vialsamplers User Manual 137
3 Using the Module Agilent Local Control Modules
Agilent 1200 Infinity Series Instant Pilot G4208A
The Agilent 1200 Infinity Series Instant Pilot controller gives you complete control, system monitoring, signal plotting, and diagnostic capabilities for a virtually unlimited number of LC system modules. It is connected to the LC system with a CAN cable for power supply and communication. Features: Complete local control and monitoring of an Agilent 1200 Series, 1260 Infinity,
and 1290 Infinity system or a single module from a single point. However, not for Agilent 1220 Compact LC.
Mixed system configurations supported, for example, 1200 Series, 1200 Series SL-, and 1100 Series.
Excellent readability and usability by large colored display with background light, high resolution, and contrast.
Convenient, ergonomic operation either handheld or at the stack with newly developed, secure attachment.
Handheld or attached to a module in a stack to facilitate operator preferences.
The 1200 Infinity Series Instant Pilot provides: Easy automation recalibration intervals and multimethod sequences satisfy
the most stringent automation routines. Transfer and archiving of methods, sequences, and logbooks via standard
USB memory sticks. Factory installed software flat dialog structure, user configurable interface,
enhanced sequence engine, for example with wait for baseline stabilization, diagnosis with passed/failed.
GLP System logbook and module logbooks record errors, unusual events, and maintenance activities for GLP traceability.
Agilent InfinityLab LC Series Vialsamplers User Manual 138
4 Preparing the Module
Leak and Waste Handling 140 Preparing the Module 141 Solvent Information 142 Capillary Color Coding Guide 149 Installing Capillaries 150 Flow Connections to the Vialsampler 152 Setting up the Vialsampler 154 Control Settings 158 Method Parameter Settings 160 Injector Programm 164 Module Configuration View 165
This chapter explains the operational parameters of the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 139
4 Preparing the Module Leak and Waste Handling
Leak and Waste Handling
For details on correct installation, see separate installation documentation.
WARNING Toxic, flammable and hazardous solvents, samples and reagents The handling of solvents, samples and reagents can hold health and safety risks. When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
Do not use solvents with an auto-ignition temperature below 200 C (392 F). Do not use solvents with a boiling point below 56 C (133 F).
Avoid high vapor concentrations. Keep the solvent temperature at least 40 K below the boiling point of the solvent used. This includes the solvent temperature in the sample compartment. For the solvents methanol and ethanol keep the solvent temperature at least 25 K below the boiling point.
Do not operate the instrument in an explosive atmosphere. Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for
example, carbon disulfide). Reduce the volume of substances to the minimum required for the
analysis. Never exceed the maximum permissible volume of solvents (8 L) in the
solvent cabinet. Do not use bottles that exceed the maximum permissible volume as specified in the usage guideline for solvent cabinet.
Ground the waste container. Regularly check the filling level of the waste container. The residual free
volume in the waste container must be large enough to collect the waste liquid.
To achieve maximal safety, regularly check the tubing for correct installation.
NOTE For details, see the usage guideline for the solvent cabinet. A printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available in the Agilent Information Center or via the Internet.
Agilent InfinityLab LC Series Vialsamplers User Manual 140
4 Preparing the Module Preparing the Module
Preparing the Module
For best performance of the module When using the module in a system with a vacuum degassing unit, shortly
degas your samples before using them in the module. Filter samples before use in a 1200 Infinity Series system. Use High pressure
filter kit (5067-4638) for inline filtering. When using buffer solutions, flush the system with water before switching it
off. Check the module plungers for scratches, grooves and dents when changing
the piston seal. Damaged plungers cause micro leaks and will decrease the lifetime of the seal.
Solvent Information - Observe recommendations on the use of solvents, see Solvent Information on page 142.
Priming and Purging the System - When the solvents have been exchanged or the system has been turned off for a certain time (for example, overnight) oxygen will re-diffuse into the solvent channel. Therefore priming and purging of the system is required before starting an application.
Table 13 Choice of Priming Solvents for Different Purposes
Activity Solvent Comments
After an installation Isopropanol Best solvent to flush air out of the system
When switching between reverse phase and normal phase (both times)
Isopropanol Best solvent to flush air out of the system
After an installation Ethanol or methanol Alternative to isopropanol (sec- ond choice) if no isopropanol is available
To clean the system when using buffers
Bidistilled water Best solvent to re-dissolve buf- fer crystals
After a solvent change Bidistilled water Best solvent to re-dissolve buf- fer crystals
Agilent InfinityLab LC Series Vialsamplers User Manual 141
4 Preparing the Module Solvent Information
Solvent Information
Observe the following recommendations on the use of solvents. Follow the recommendations for avoiding the growth of algae, see the pump
manuals. Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.22 m filters. Avoid or minimize the use of solvents that may corrode parts in the flow path.
Consider specifications for the pH range given for different materials such as flow cells, valve materials etc. and recommendations in subsequent sections.
Recommended Wash Solvents water ethanol methanol water/acid (especially for basic compounds) water/base (especially for acidic compounds) water/acetonitrile
NOTE For different wash solvents as mentioned above, verify that the wash solvent is suitable for the silicone wash tubing.
Agilent InfinityLab LC Series Vialsamplers User Manual 142
4 Preparing the Module Solvent Information
Solvent Compatibility of Tubings for Peristaltic Pumps The table shows the chemical resistance properties of Silicone and PharMed tubing to different needle wash solvents:
Table 14 Solvent Compatibility of Silicone and PharMed Tubing
Silicone PharMed
Acids weak medium strong
good unsatisfactory not recommended
very good good not recommended
Alkaline solution weak medium strong
good unsatisfactory not recommended
very good very good good
Hydrocarbons aliphatic aromatizised halogenated
not recommended not recommended not recommended
Agilent InfinityLab LC Series Vialsamplers User Manual 143
4 Preparing the Module Solvent Information
General Information about Solvent/Material Compatibility Materials in the flow path are carefully selected based on Agilents experiences in developing highest quality instruments for HPLC analysis over several decades. These materials exhibit excellent robustness under typical HPLC conditions. For any special condition, please consult the material information section or contact Agilent.
Disclaimer
Subsequent data was collected from external resources and is meant as a reference. Agilent cannot guarantee the correctness and completeness of such information. Data is based on compatibility libraries, which are not specific for estimating the long-term life time under specific but highly variable conditions of UHPLC systems, solvents, solvent mixtures and samples. Information can also not be generalized due to catalytic effects of impurities like metal ions, complexing agents, oxygen etc. Apart from pure chemical corrosion, other effects like electro corrosion, electrostatic charging (especially for non-conductive organic solvents), swelling of polymer parts etc. need to be considered. Most data available refers to room temperature (typically 20 25 C, 68 77 F). If corrosion is possible, it usually accelerates at higher temperatures. If in doubt, please consult technical literature on chemical compatibility of materials.
PEEK
PEEK (Polyether-Ether Ketones) combines excellent properties regarding biocompatibility, chemical resistance, mechanical and thermal stability. PEEK is therefore the material of choice for UHPLC and biochemical instrumentation.
It is stable in the specified pH range (for the Bio-inert LC system: pH 1 13, see bio-inert module manuals for details), and inert to many common solvents.
There is still a number of known incompatibilities with chemicals such as chloroform, methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %, sulfuric acid > 10 %, sulfonic acids, trichloroacetic acid), halogens or aqueous halogen solutions, phenol and derivatives (cresols, salicylic acid, and so on).
When used above room temperature, PEEK is sensitive to bases and various organic solvents, which can cause it to swell. Under such conditions, normal PEEK capillaries are very sensitive to high pressure. Therefore, Agilent uses stainless steel cladded PEEK capillaries in bio-inert systems. The use of stainless steel cladded PEEK capillaries keeps the flow path free of steel and ensures pressure stability to at least 600 bar. If in doubt, consult the available literature about the chemical compatibility of PEEK.
Agilent InfinityLab LC Series Vialsamplers User Manual 144
4 Preparing the Module Solvent Information
Polyimide
Agilent uses semi-crystalline polyimide for rotor seals in valves and needle seats in autosamplers. One supplier of polyimide is DuPont, which brands polyimide as Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic solvents. It is incompatible with concentrated mineral acids (e.g. sulphuric acid), glacial acetic acid, DMSO and THF. It is also degraded by nucleophilic substances like ammonia (e.g. ammonium salts in basic conditions) or acetates.
Polyethylene (PE)
Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents including acids and bases in a pH range of 1 to 12.5. It is compatible with many organic solvents used in chromatographic systems like methanol, acetonitrile and isopropanol. It has limited stability with aliphatic, aromatic and halogenated hydrocarbons, THF, phenol and derivatives, concentrated acids and bases. For normal phase applications, the maximum pressure should be limited to 200 bar.
Tantalum (Ta)
Tantalum is inert to most common HPLC solvents and almost all acids except fluoric acid and acids with free sulfur trioxide. It can be corroded by strong bases (e.g. hydroxide solutions > 10 %, diethylamine). It is not recommended for the use with fluoric acid and fluorides.
Stainless Steel (SST) Stainless steel is inert against many common solvents. It is stable in the presence of acids and bases in a pH range of 1 to 12.5. It can be corroded by acids below pH 2.3. It can also corrode in following solvents: Solutions of alkali halides, their respective acids (for example, lithium iodide,
potassium chloride, and so on) and aqueous solutions of halogens. High concentrations of inorganic acids like nitric acid, sulfuric acid and
organic solvents especially at higher temperatures (replace, if your chromatography method allows, by phosphoric acid or phosphate buffer which are less corrosive against stainless steel).
Agilent InfinityLab LC Series Vialsamplers User Manual 145
4 Preparing the Module Solvent Information
Halogenated solvents or mixtures which form radicals and/or acids, for example: 2 CHCl3 + O2 2 COCl2 + 2 HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol.
Chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, diisopropylether). Such ethers should be filtered through dry aluminium oxide which adsorbs the peroxides.
Solutions of organic acids (acetic acid, formic acid, and so on) in organic solvents. For example, a 1 % solution of acetic acid in methanol will attack steel.
Solutions containing strong complexing agents (for example, EDTA, ethylene diamine tetra-acetic acid).
Mixtures of carbon tetrachloride with isopropanol or THF.
Titanium (Ti)
Titanium is highly resistant to oxidizing acids (for example, nitric, perchloric and hypochlorous acid) over a wide range of concentrations and temperatures. This is due to a thin oxide layer on the surface, which is stabilized by oxidizing compounds. Non-oxidizing acids (for example, hydrochloric, sulfuric and phosphoric acid) can cause slight corrosion, which increases with acid concentration and temperature. For example, the corrosion rate with 3 % HCl (about pH 0.1) at room temperature is about 13 m/year. At room temperature, titanium is resistant to concentrations of about 5 % sulfuric acid (about pH 0.3). Addition of nitric acid to hydrochloric or sulfuric acids significantly reduces corrosion rates. Titanium is sensitive to acidic metal chlorides like FeCl3 or CuCl2. Titanium is subject to corrosion in anhydrous methanol, which can be avoided by adding a small amount of water (about 3 %). Slight corrosion is possible with ammonia > 10 %.
Diamond-Like Carbon (DLC)
Diamond-Like Carbon is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Agilent InfinityLab LC Series Vialsamplers User Manual 146
4 Preparing the Module Solvent Information
Fused silica and Quartz (SiO2)
Fused silica is used in Max Light Cartridges. Quartz is used for classical flow cell windows. It is inert against all common solvents and acids except hydrofluoric acid and acidic solvents containing fluorides. It is corroded by strong bases and should not be used above pH 12 at room temperature. The corrosion of flow cell windows can negatively affect measurement results. For a pH greater than 12, the use of flow cells with sapphire windows is recommended.
Gold
Gold is inert to all common HPLC solvents, acids and bases within the specified pH range. It can be corroded by complexing cyanides and concentrated acids like aqua regia.
Zirconium Oxide (ZrO2)
Zirconium Oxide is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Fluorinated polymers (PTFE, PFA, FEP, FFKM, PVDF)
Fluorinated polymers like PTFE (polytetrafluorethylene), PFA (perfluoroalkoxy), and FEP (fluorinated ethylene propylene) are inert to almost all common acids, bases, and solvents. FFKM is perfluorinated rubber, which is also resistant to most chemicals. As an elastomer, it may swell in some organic solvents like halogenated hydrocarbons.
TFE/PDD copolymer tubings, which are used in all Agilent degassers except 1322A/G7122A, are not compatible with fluorinated solvents like Freon, Fluorinert, or Vertrel. They have limited life time in the presence of Hexafluoroisopropanol (HFIP). To ensure the longest possible life with HFIP, it is best to dedicate a particular chamber to this solvent, not to switch solvents, and not to let dry out the chamber. For optimizing the life of the pressure sensor, do not leave HFIP in the chamber when the unit is off.
The tubing of the leak sensor is made of PVDF (polyvinylidene fluoride), which is incompatible with the solvent DMF (dimethyl formamide).
Agilent InfinityLab LC Series Vialsamplers User Manual 147
4 Preparing the Module Solvent Information
Sapphire, Ruby and Al2O3-based ceramics
Sapphire, ruby and ceramics based on aluminum oxide Al2O3 are inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Agilent InfinityLab LC Series Vialsamplers User Manual 148
4 Preparing the Module Capillary Color Coding Guide
Capillary Color Coding Guide
Figure 28 Syntax for capillary description
Agilent InfinityLab LC Series Vialsamplers User Manual 149
4 Preparing the Module Installing Capillaries
Installing Capillaries
For correct installation of capillary connections of the sampler it's important to choose the correct fittings, see Capillary Color Coding Guide on page 149.
The capillaries mentioned above are examples only.
Parts required p/n Description 5067-4650 Capillary ST 0.12 mm x 150 mm SL/SX 5067-4651 Capillary ST 0.12 mm x 280 mm SL/SX 5067-4720 Capillary ST 0.17 mm x 150 mm SL/SX 5067-4722 Capillary ST 0.17 mm x 280 mm SL/SX 5065-4454 Fitting screw long
10/pk Quantity depends on configuration of the module (number of connections to the multisampler).
1 Select a nut that is long enough for the fitting you'll be using.
2 Slide the nut over the end of the tubing or capillary.
Agilent InfinityLab LC Series Vialsamplers User Manual 150
4 Preparing the Module Installing Capillaries
3 Carefully slide the ferrule components on after the nut and then finger-tighten the assembly while ensuring that the tubing is completely seated in the bottom of the end fitting.
4 Use a column or injection valve to gently tighten the fit- ting which forces the ferrule to seat onto the tubing or capillary.
NOTE Don't overtighten. Overtightening will shorten the lifetime of the fitting.
5 Loosen the nut and verify that the ferrule is correctly positioned on the tubing or capillary.
NOTE The first time that the swagelock fitting is used on a column or an injection valve, the position of the ferrule is permanently set. If changing from a column or an injection valve to another, the fitting may leak or decrease the quality of the separation by contributing to band broadening.
Agilent InfinityLab LC Series Vialsamplers User Manual 151
4 Preparing the Module Flow Connections to the Vialsampler
Flow Connections to the Vialsampler
1 Place a needle wash solvent reservoir into the solvent cabinet. 2 Connect the Needle Wash Bottle Head Assembly to the solvent reservoir and
close the bottle. 3 Guide the tube of the Needle Wash Bottle Head Assembly through the cover
opening and connect it to the peristaltic pump. 4 Route the drainage of the washport outlet to the waste container. 5 Prime the wash tubing. 6 Install the capillary from the pump outlet into the port 1 of the injection valve.
Preparations Module is installed in the system.
Use an appropriate solvent based on the sample and mobile phase chemistries. The composition of the wash solvent should be the most solubilizing compatible solvent (your stron- gest diluent). Selecting the wash solvent is part of the method development. A mixture of 50 % up to 100 % organic solvent in distilled water is a good choice for many applica- tions.
NOTE The silicone waste drainage is suitable for most common wash solvents. For critical wash solvents, it may be necessary to replace the silicone tube with a PE tube (Tubing, PE, 1.5 m (5042-9974)).
Agilent InfinityLab LC Series Vialsamplers User Manual 152
4 Preparing the Module Flow Connections to the Vialsampler
7 Install the capillary from port 6 of the injection valve to the ICC or MCT The correct plumbing is shown in the figure below and is attached to the module.
8 Prime and purge the complete sampler.
Waste
Column
Pump
Agilent InfinityLab LC Series Vialsamplers User Manual 153
4 Preparing the Module Setting up the Vialsampler
Setting up the Vialsampler
The setup of the Vialsampler is shown with the Agilent OpenLab ChemStation C.01.07. Depending on the controller (e.g. Agilent Local Controller, OpenLab EZChrom, Masshunter) the screens look different.
Figure 29 ChemStation Method and Run Control
After successful load of the OpenLab ChemStation, you should see the module as an active item in the graphical user interface (GUI).
NOTE This section describes the Vialsampler settings only. For information on the software or other modules refer to the corresponding documentation.
Agilent InfinityLab LC Series Vialsamplers User Manual 154
4 Preparing the Module Setting up the Vialsampler
Table 15 The Vialsampler User Interface
Within the sampler user interface, there are active areas. If you move the mouse cursor across the icons (tray, EMF button), the cursor will change and you may click on the icon to 1 Turn on/off the sampler 2 Check which sample drawers are installed 3 Get the status of the EMF (Early Maintenance Feature) 4 Cooling Temperatur (if Sample Cooler/Sample Thermo-
stat is installed)
Current instrument information on: Injection volume Sample location
1
2
3
4
Agilent InfinityLab LC Series Vialsamplers User Manual 155
4 Preparing the Module Setting up the Vialsampler
A right-click into the Active Area will open a menu with the option to adjust the following parameters: Show the Control User Interface (special module set-
tings). See Control Settings on page 158. Show the Method User interface (same as via menu
Instrument> Set up Instrument Method> Setup G7129B). See Method Parameter Settings on page 160.
Injector Program
When you activate a pretreatment/injector program, it replaces the standard injection cycle. See Using an Injector Program on page 169
Identify Device Home Arm Reset Injector Release Vial Switch Valve to Bypass Switch on Tray Illumination Needle Wash: allows you to wash the needle without
modifying your current method. The needle wash spec- ified in this dialog box is carried out immediately.
Modify Flow Path: for the sample loops, needle seat, exten-
sion loops and metering devices External Tray Temperature Mode
Table 15 The Vialsampler User Interface
Agilent InfinityLab LC Series Vialsamplers User Manual 156
4 Preparing the Module Setting up the Vialsampler
Module Status shows Run / Ready / Error state and Not Ready text or Error text Error (Red) Not ready (yellow) Ready (green) Pre run, Post run (purple) Run (blue) Idle (green) Offline (dark gray) Standby (light gray)
EMF Status shows Run / Ready / Error state and Not Ready text or Error text Offline (gray) Ok
No Maintenance required (green) EMF warning. Maintenance might be required (yellow) EMF warning. Maintenance required (red)
Table 15 The Vialsampler User Interface
NOTE For additional help and support, highlight the desired area and press the F1 key. A help screen will open with additional information and documentation about the topic.
NOTE The Vialsampler configuration is done in the online module dashboard context menu, not in the instrument configuration view.
Agilent InfinityLab LC Series Vialsamplers User Manual 157
4 Preparing the Module Setting up the Vialsampler
Control Settings The control settings are available via right click on the active area of the grafical user interface, see Table 15 on page 155.
Agilent InfinityLab LC Series Vialsamplers User Manual 158
4 Preparing the Module Setting up the Vialsampler
Table 16 Control settings
Figure 30 Control settings
The Sampler control parameters are in the following sections: Missing Vial
Mark the Ignore missing vial check box to specify that, if a vial is missing, the injector ignores it and continues with a 6-second dummy run. The message "Missing vial
Illumination
Toggles the illumination of the sample area, On or Off. At Power On
The section is available when a cooler/thermostat is installed and configured. Mark the Turn on Thermostat check box to specify that the cooler/thermostat is switched on automatically when the instrument is switched on.
Thermostat
The section is available when a cooler/thermostat is installed and configured and the Constant temperature mode is selected.
Select On to switch on the cooler/thermostat. Specify the required temperature in the adjacent field. The specified temperature must be at least 5 C below ambient for proper temperature control.
Select Off to switch off the cooler/thermostat. Automatic Turn On
You can set a date and time at which the cooler/thermo- stat switches on automatically.
Pump connected to Sampler
Use this section to specify the pump that is used with the Vialsampler. If more than one pump is configured, display the drop-down list and select the appropriate pump from the list.
Enable Analysis
This feature requires LC & CE Drivers A.02.19 or newer and is only available for the Sample Thermostat. With this func- tion, you can specify if the analyses should start With any temperature or only when the Temperature is within 2 C range of the setpoint temperature.
NOTE The Enable Analysis section is disabled when Not con- trolled is selected in the Temperature section.
Agilent InfinityLab LC Series Vialsamplers User Manual 159
4 Preparing the Module Setting up the Vialsampler
Method Parameter Settings The method parameter settings are available via right click on the active area of the grafical user interface.
Figure 31 Method parameter settings
NOTE For additional help and support, highlight the desired area and press the F1 key. A help screen will open with additional information and documentation about the topic.
Agilent InfinityLab LC Series Vialsamplers User Manual 160
4 Preparing the Module Setting up the Vialsampler
Table 17 Method parameter settings
Injection Specify the injection volume (in L) in the Injection Volume field. The injection volume limits depend upon the configu- ration settings of the autosampler. Note that this setting is ignored when Pretreatment/Injector Program is defined.
Needle Wash Mark the Enable Needle Wash check box to specify an injec- tion with needle wash.
The following wash parameters can be specified: Mode: Select the wash location by clicking the down
arrow and selecting the needle wash mode from the list. You can select from Flush Port (default location) or Wash Vial.
Time: This field is enabled only if you have selected Flush Port. Specify a flush time in seconds between 1 and 100 in steps of 1.
Location: enabled only if you have selected a Wash Vial. Allows you to specify a wash location in the format Vial X, where X is the vial number. Specify well locations in the format P
Repeat: available only if you have selected Wash Vial. You can specify up to 10 repeat washes.
Stoptime: enables you to set the time that the analysis stops. Limits: 0.01 through 99999 min or As Pump/No Limit. Posttime: the Vialsampler remains in a post-run state during the Posttime to delay the start of the next analysis. A Post- time period can be used to allow the column to equilibrate after changes in solvent composition (for example after gradient elution). Limits: 0.01 through 99999 min or Off.
Agilent InfinityLab LC Series Vialsamplers User Manual 161
4 Preparing the Module Setting up the Vialsampler
Auxiliary Draw Speed: determines the rate at which the plunger
draws sample from the vial. Set the speed to an appro- priate value for your sample. For viscous samples, use a slow Draw Speed.
Eject Speed: determines the rate at which the plunger ejects sample from the metering device. If you are injecting large volumes of sample, setting a high Eject Speed will shorten the time needed for an injection cycle. For viscous samples in combination with multi- draw, use a slow Eject Speed.
Wait Time After Draw: ensures that the temporary vac- uum, which originates from the drawing of liquid from the sample vial, dissipates. The needle stays first on the seat for the specified time, then after drawing sample from the vial, remains there for the specified time.
Needle Height Offset: a vertical offset that enables you to position the needle a specific distance (in mm) away from its standard position. The Offset function is useful when analyzing very small sample volumes, or when only a specific part of the sample is required, for exam- ple, the top layer.
Table 17 Method parameter settings
Agilent InfinityLab LC Series Vialsamplers User Manual 162
4 Preparing the Module Setting up the Vialsampler
High Throughput Injection Valve to Bypass for Delay Volume Reduction: Mark this check box to switch the flow from the injector from mainpass (main path) to bypass after injection has taken place. This reduces the delay volume for low volume tech- niques. You can specify the point during the analyses when the valve switches to bypass by setting the Sample Flush-out Factor. Sample Flush-Out Factor: ensures that the sample is thor- oughly flushed out of the syringe and past the capillary seat and valve. The Sample Flush-Out Factor is preset to 5.0 at the factory. The preset Sample Flush-Out Factor is correct for most methods. However, for unusually viscous sam- ples, you should increase the Sample Flush-Out Factor to obtain the desired degree of flushing to prevent sample car- ryover.
Overlapped Injection Mode: provides faster throughput of samples by allowing the preparation of the next injection while the current injection is being completed. The follow- ing options can be selected from the drop-down list: Off: to switch off overlapped injection. Select When
sample is Flushed Out to specify that the sample can be prepared directly after the current sample has been flushed out.
Prefetch Vial: to specify that the next sample is fetched by the gripper while the current sample is running.
Overlap Injection Cycle: to specify that the next sample is fetched and loaded into the sample loop while the current sample is running. Specify a time (in minutes) after injection when this process should be started.
Table 17 Method parameter settings
NOTE For additional help and support, highlight the desired area in the user interface and press the F1 key. A help screen will open with additional information and documentation about the topic.
Agilent InfinityLab LC Series Vialsamplers User Manual 163
4 Preparing the Module Setting up the Vialsampler
Injector Programm The pretreatment/injector program comprises a series of numbered lines, each specifying an operation that the autosampler carries out sequentially. When you activate a pretreatment/injector program, it replaces the standard injection cycle. The following functions are available : Draw Eject Mix Inject Move Wait Valve Needle Wash Remote Wait For Repeat . End Repeat Comment
Agilent InfinityLab LC Series Vialsamplers User Manual 164
4 Preparing the Module Setting up the Vialsampler
Module Configuration View The module configuration settings are available via menu Instrument> Show ConfigUI
Figure 32 Configuration view
NOTE For additional help and support. Highlight the desired area and press the F1 key. A help screen will open with additional information and documentation about the topic.
NOTE The vialsampler configuration is done in the module dashboard context menu, not in the instrument configuration.
Agilent InfinityLab LC Series Vialsamplers User Manual 165
5 Optimizing Performance
Optimization for Lowest Carryover 167 Using the Automated Needle Wash (Wash Port) 167 Using an Injector Program 169 General Recommendation for Lowest Carryover 170
Fast Injection Cycle and Low Delay Volume 171 Overlapped Injection Mode 171 General Recommendations for Fast Injection Cycle Times 172
Precise Injection Volume 173 Draw and Eject Speed 174
Choice of Rotor Seal 175
This chapter provides information on how to optimize the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 166
5 Optimizing Performance Optimization for Lowest Carryover
Optimization for Lowest Carryover
Several parts of an injection system can contribute to carryover: needle outside needle inside needle seat sample loop seat capillary injection valve
The autosampler continuous flow-through design ensures that sample loop, needle inside, seat capillary, and the mainpass (main path) of the injection valve is always in the flowline. These parts are continuously flushed during an isocratic and also during a gradient analysis. The residual amount of sample remaining on the outside of the needle after injection may contribute to carryover in some instances. When using small injection volumes or when injecting samples of low concentration immediately after samples of high concentration, carryover may become noticeable. Using the automated needle wash enables the carryover to be minimized and also prevents contamination of the needle seat.
Using the Automated Needle Wash (Wash Port) The automated needle wash can be programmed either as method parameter "enable needle wash" or the needle wash can be included into the injector program. When the automated needle wash is used, the needle is moved into a wash vial before the sample is drawn. By washing the needle before drawing a sample, the sample is removed from the surface of the needle immediately.
For best results and to obtain minimum carryover, the wash port should contain solvent in which the sample components are soluble. The wash port flapper is located in the needle station and a peristaltic pump delivers the wash solvent. It has a volume of ca. 0.5 mL and the peristaltic pump delivers ca. 5 mL/min, which means the wash port volume is completely refilled with fresh solvent in ca. 6 s. If the wash port is selected, the user can set the length of time for washing the outside of the needle with fresh solvent. This may be as low as two or three seconds in routine situations where carryover is less of a problem and 10 to 20 s for more complete washing. It is recommended that washing the outside of the needle in the flush port should be standard procedure to avoid contaminating the needle seat.
Agilent InfinityLab LC Series Vialsamplers User Manual 167
5 Optimizing Performance Optimization for Lowest Carryover
Using the Needle Wash in the Wash Vial
For very critical applications where the outside of the needle cannot be cleaned sufficiently with one wash port solvent you can use an injector program and an additional wash vial with an appropriate and stronger solvent for cleaning. The wash vial should contain solvent in which the sample components are soluble, and the vial should not be capped. If the wash vial is capped, small amounts of sample remain on the surface of the septum, which may be carried on the needle to the next sample.
Injector Program with Needle Wash
The injector program includes the command WASH. When this command is included in the injector program, the needle is lowered once into the specified wash vial or flushport before injection. For example: 1 WASH > Location P2-A1 or flushport 2 Draw 5 L from Sample Location P1-A1 3 INJECT
Line 1 moves the needle to vial P2-A1 or the flushport. Line 2 draws 5 L from the current sample vial P1-A1. Line 3 injects the sample (valve switches to main pass).
Agilent InfinityLab LC Series Vialsamplers User Manual 168
5 Optimizing Performance Optimization for Lowest Carryover
Using an Injector Program The process is based on a program that switches the bypass grove of the injection valve into the flow line for cleaning. This switching event is performed at the end of the equilibration time to ensure that the bypass grove is filled with the start concentration of the mobile phase. Otherwise the separation could be influenced, especially if microbore columns are used.
Example
Outside wash of needle in vial p2-A1 before injection Injector program: Draw x.x (y) l from sample WASH vial location p2-A1 Inject
Wait (equilibration time - see text above) Valve bypass Wait 0.2 min Valve mainpass (main path) Valve bypass Valve mainpass (main path)
NOTE Overlapped injection together with additional injection valve switching is not possible.
Agilent InfinityLab LC Series Vialsamplers User Manual 169
5 Optimizing Performance Optimization for Lowest Carryover
General Recommendation for Lowest Carryover For samples where the outside of the needle cannot be cleaned sufficiently with the wash port use wash vials with an appropriate and stronger solvent. An injector program and an extra wash vial can be used for cleaning. In case the needle seat has got contaminated and carryover is significantly higher than expected, the following procedure can be used to clean the needle seat: In Lab Advisor go to Maintenance> Change Needle: this sets the needle into
home position. Remove the safety cover of the sampler. Be carefully not to risk an injury by
the uncovered needle. Pipette an appropriate solvent on to the needle seat. The solvent should be
able to dissolve the contamination. If the contamination is unknown, use two or three solvents of different polarity. Use several milliliters to clean the seat.
Clean the needle seat with a tissue and remove all liquid. Again be careful not to risk an injury by the uncovered needle.
Reinstall the safety cover of the sampler. In Lab Advisor, go to Maintenance> Change Needle and finish the procedure.
Agilent InfinityLab LC Series Vialsamplers User Manual 170
5 Optimizing Performance Fast Injection Cycle and Low Delay Volume
Fast Injection Cycle and Low Delay Volume
Short injection cycle times for high sample throughput is one of the most important requirements in laboratories. In order to shorten cycle times, you can: shorten the column length use high flow rates apply a steep gradient
Having optimized these parameters, further reduction of cycle times can be obtained using the overlapped injection mode.
Overlapped Injection Mode In this process, when the sample has reached the column, the injection valve is switched back to bypass, and the next injection cycle starts but waits with switching to mainpass (main path) until the actual run is finished. You gain the sample preparation time when using this process.
Switching the valve into the bypass position reduces the system delay volume. The mobile phase is directed to the column without passing sample loop, needle, and needle seat capillary. This can help to have faster cycle times especially if low flow rates have to be used like it is mandatory in narrow bore and micro bore HPLC.
The injection cycle times also depend on the injection volume. In identically standard condition, injecting 100 L instead of 1 L, increase the injection time by approximately 8 s. In this case and if the viscosity of the sample allows it, the draw and eject speed of the injection system has to be increased.
NOTE Having the valve in bypass position can increase the carryover in the system.
NOTE For the last injection of the sequence with overlapped injections, it has to be considered that for this run the injection valve is not switched as for the previous runs and consequently the injector delay volume is not bypassed. This means that the retention times are prolonged for the last run. Especially at low flow rates this can lead to retention time changes which are too big for the actual calibration table. To overcome this, it is recommended to add an extra blank injection as last injection to the sequence.
Agilent InfinityLab LC Series Vialsamplers User Manual 171
5 Optimizing Performance Fast Injection Cycle and Low Delay Volume
General Recommendations for Fast Injection Cycle Times As described in this section, the first step to provide short cycle times are optimizing the chromatographic conditions. If this is done the autosampler parameter should be set to: Overlapped injection mode Increase of draw and eject speed for large injection volumes Add at last run a blank, if overlapped injection is used
To reduce the injection time, the detector balance has to be set to OFF.
Agilent InfinityLab LC Series Vialsamplers User Manual 172
5 Optimizing Performance Precise Injection Volume
Precise Injection Volume
Injection Volumes Less Than 2 L
When the injection valve switches to the BYPASS position, the mobile phase in the sample loop is depressurized. When the syringe begins drawing sample, the pressure of the mobile phase is decreased further. If the mobile phase is not degassed adequately, small gas bubbles can form in the sample loop during the injection sequence. When using injection volumes < 2 L, these gas bubbles can affect the injection-volume precision. Use degassed mobile phases for best injection-volume precision with injection volumes < 2 L.
If the pump has no degasser, it is recommended to use an Agilent 1260 Infinity degasser.
Also, using the automated needle wash (see Optimization for Lowest Carryover on page 167) between injections reduces carryover to a minimum, further improving the injection volume precision.
Agilent InfinityLab LC Series Vialsamplers User Manual 173
5 Optimizing Performance Precise Injection Volume
Draw and Eject Speed
Draw Speed
The speed at which the metering unit draws sample out of the vial may have an influence on the injection volume precision when using viscous samples. If the draw speed is too high, air bubbles may form in the sample plug, affecting precision. The default draw speed is 100 L/min. This speed is suitable for the majority of applications, however, when using viscous samples, set the draw speed to lower speed for optimum results. A DRAW statement in an injector program also uses the draw speed setting which is configured for the autosampler.
Eject Speed
The default eject speed setting is 400 L/min. When using large injection volumes, setting the eject speed to a higher value speeds up the injection cycle by shortening the time the metering unit requires to eject solvent at the beginning of the injection cycle (when the piston returns to the home position).
An EJECT statement in an injector program also uses the eject speed setting which is configured for the autosampler. A faster eject speed shortens the time required to run the injector program. When using viscous samples, a high eject speed should be avoided.
Agilent InfinityLab LC Series Vialsamplers User Manual 174
5 Optimizing Performance Choice of Rotor Seal
Choice of Rotor Seal
Vespel Seal (for 1290 Infinity II Vialsampler Only)
The 1290 seal has sealing material made of Vespel. Vespel is suitable for applications using mobile phases within the pH range of 2.3 9.5, which is suitable for most applications. However, for applications using mobile phases with pH below 2.3 or above 9.5, the Vespel seal can degrade faster, leading to reduced seal lifetime.
PEEK Seal (for 1260 Infinity II Vialsamplers Only)
The 1260 injection valve has a sealing material made of PEEK. This material has high chemical resistance and versatility. It is suitable for application using mobile phases within a pH between 1 and 14.
NOTE Strong oxidizing acids such as concentrated nitric and sulfuric acids are not compatible with PEEK.
Agilent InfinityLab LC Series Vialsamplers User Manual 175
6 Troubleshooting and Diagnostics
Status Indicators 177 Overview of Tests and Tools 178 Maintenance and Troubleshooting Tools 179 Maintenance Positions 179 Gripper Verification 183 Injector Steps 184 Park Arm 186
Diagnostic Tests 187 Sample Cooler Function Test 187 Heater Test 188
Agilent Lab Advisor Software 189
This chapter gives an overview of the maintenance, troubleshooting, and diagnostic features available for the Agilent InfinityLab Series Vialsampler.
Agilent InfinityLab LC Series Vialsamplers User Manual 176
6 Troubleshooting and Diagnostics Status Indicators
Status Indicators
The Agilent InfinityLab Series LC modules are equipped with two status indicator LED lights to enable the user to get an immediate visual impression of the actual state of the instrument: The power indicator light is integrated into the main switch of the module and
provides information about whether or not the system is powered on. When the indicator illuminates in green, the module is ON.
The module status indicator light is situated in the upper right corner of the module and provides information on the actual operating state of the system (see Status Indicators on page 58 for more information).
Agilent InfinityLab LC Series Vialsamplers User Manual 177
6 Troubleshooting and Diagnostics Overview of Tests and Tools
Overview of Tests and Tools
Table 18 Overview of tests and tools
Name Type Lab Advisor Soft- ware
1200 Infinity Instant Pilot
InfinityLab Com- panion
Maintenance Posi- tions
Tool yes yes yes
Park Arm Tool yes yes yes
Injector Steps Tool yes yes yes
Gripper Verifica- tion
Tool yes no no
Automatic Refer- encing
Tool yes1
1 only available for customer engineers
no no
ALS Torque Verifi- cation
Tool yes1 no no
Sample Cooler Function Test
Test yes no no
Heater Test Test yes no no
Agilent InfinityLab LC Series Vialsamplers User Manual 178
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
Maintenance and Troubleshooting Tools
Maintenance Positions The replacement of certain components, such as the needle or the metering seal, requires the needle arm, the metering device plunger or the gripper arm to be moved into specific positions. With the Maintenance Positions function, these components can be moved into predefined service positions, enabling the easy and safe access to parts that would not be accessible otherwise.
WARNING Sharp needle
Some of the maintenance positions require the removal of the safety cover, revealing the uncovered needle.
Risk of injury
Do not touch the tip of the needle.
Keep your fingers off the moving needle.
Wear safety goggles when handling an uncovered needle.
Agilent InfinityLab LC Series Vialsamplers User Manual 179
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
Using the Change Needle, Loop and Seat Function
The Change Needle, Loop, and Seat function grants easy access to the needle, the sample loop, and the needle seat by moving the needle and the gripper arm into predefined positions.
The execution of this function encompasses the following elementary steps when the Agilent Lab Advisor Software is used: 1 Select Maintenance Positions from the list under Service & Diagnostics. 2 Under Task Selection, select Change, Needle, Loop and Seat and click Start. 3 Specify the type of the seat and loop capillary being installed in the Verify
capillary configuration window and click Continue. 4 Wait until the needle arm moves into the upper position. 5 Remove the safety cover and move the wash port safety flap to the service
position by slightly lifting it up and then flapping it to the left. 6 Confirm that the wash port safety flap is in the service position and click Next
to continue.
7 Wait until the needle moves down and stops ca. 2 mm above the needle seat. 8 Remove the sample loop (see Exchange the Sample Loop Assembly on
page 241) and the needle (see Exchange the Needle Assembly on page 229) and click Next to continue.
9 Replace the needle seat (see Exchange the Needle Seat Assembly on page 235) and click Next to continue.
10 Install the new needle (see Exchange the Needle Assembly on page 229) and click Next to continue.
11 Reconnect the old sample loop or install a new one if needed (see Exchange the Sample Loop Assembly on page 241) and click Next to continue.
12 Ensure that the needle tip is concentric with the seat. Use the Move Up and Move Down buttons to move the needle closer to or further away from the seat.
WARNING Sharp needle
Risk of injury due to moving needle.
Do not touch the tip of the needle.
Keep your fingers off the moving needle.
Wear safety goggles when handling an uncovered needle.
Agilent InfinityLab LC Series Vialsamplers User Manual 180
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
13 Click Next and wait until the needle arm moves into the upper position. 14 Move the wash port safety flap back to its original position. 15 Confirm that the wash port safety flap is in the correct position and click Next. 16 Reinstall the safety cover back to the needle station and click Next to
continue. 17 Wait until the needle moves back to the needle seat. In the following pop-up
window, specify whether or not the EMF counters of the needle and the needle seat should be reset, then click OK.
18 Select Back in the bottom right corner to exit the Maintenance Positions function.
Using the Change Metering Seal and Piston Function
The Change Metering Seal and Piston function releases the tension on the metering spring by moving the plunger of the analytical head into the maintenance position.
The execution of this function encompasses the following elementary steps when the Agilent Lab Advisor Software is used: 1 Select Maintenance Positions from the list under Service & Diagnostics. 2 Under Task Selection, select Change Metering Seal and Piston and click Start.
The plunger is moved into the maintenance position, releasing the tension on the spring.
3 Remove the analytical head and the metering seal or replace any other component of the analytical head (see Exchange the Metering Seal and Piston on page 250).
4 Click Next to move the piston back to the home position. 5 In the following pop-up window, specify whether or not the EMF counter of the
metering seal should be reset, then click OK. 6 Select Back in the bottom right corner to exit the Maintenance Positions
function.
Agilent InfinityLab LC Series Vialsamplers User Manual 181
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
Using the Change Gripper Function
The Change Gripper function grants easy access to the release mechanism of the gripper assembly by bringing the gripper arm into the front position.
The execution of this function encompasses the following elementary steps when the Agilent Lab Advisor Software is used: 1 Select Maintenance Positions from the list under Service & Diagnostics. 2 Under Task Selection, select Change Gripper and click Start.
This gripper is moved into the maintenance position. 3 Replace the gripper arm (see Exchange the Gripper Arm on page 258). 4 Click Next to move the gripper arm back to the home position. 5 Select Back in the bottom right corner to exit the Maintenance Positions
function.
Agilent InfinityLab LC Series Vialsamplers User Manual 182
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
Gripper Verification The Gripper Verification is a diagnostic tool to verify that the gripper alignment is correct by using several predefined vial positions as reference points.
The execution of the Gripper Verification tool encompasses the following elementary steps when the Agilent Lab Advisor Software is used: 1 Select Gripper Verification in the list under Service & Diagnostics. 2 Select any of the available reference positions in the drop-down list and
ensure that there is a suitable vial in the selected position. 3 Click Pick & Put Vial.
The gripper arm fetches the vial, lifts it up and places it back to its original position.
4 Repeat the preceding steps with a different vial position if needed.
5 Select Back in the bottom right corner to exit the Gripper Verification function.
Table 19 Available reference positions for the various vial drawer types.
Vial Drawer Type Part Number Available Reference Positions
Drawer for 66 x 2 mL vials G7129-60010 A1, A11, D6, F1, F11
Drawer for 18 x 6 mL vials G7129-60110 A1, A6, B4. C1, C6
Classic drawer for 50 x 2 mL vials, left
G7129-60210 1, 10, 26, 41, 50
Classic drawer for 50 x 2 mL vials, right
G7129-60220 51, 60, 76, 91, 100
NOTE The Gripper Verification tool cannot be used to verify the positions of the external tray (G7129-60000).
NOTE In the case of an error, use the Reset button to bring the sampler back to its normal operating state.
Agilent InfinityLab LC Series Vialsamplers User Manual 183
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
Injector Steps Injector Steps is a diagnostic tool with which the user can walk through the whole injection process while having manual control over the elemental steps. This can be especially beneficial during troubleshooting when close observation of each individual step of the injection cycle can be crucial for success.
The execution of the Injector Steps tool encompasses the following elementary steps when the Agilent Lab Advisor Software is used: 1 Select Injector Steps from the list under Service & Diagnostics. 2 Enter a valid value for the vial location and the injection volume in the Location
and Volume fields, respectively. 3 Select and execute any of the step commands (see Table 20 on page 185). 4 Select Back in the bottom right corner to exit the Gripper Verification function.
The vial is returned, the metering plunger moves into the home position, and the injection valve switches into the mainpass position.
NOTE For some commands, there are prerequisite steps that are executed automatically if not already done (see Table 20 on page 185).
Agilent InfinityLab LC Series Vialsamplers User Manual 184
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
Table 20 List of injector step commands
Command Action Prerequisite Steps1
1 Prerequisite steps are executed automatically if not already done.
Valve Bypass The injection valve switches into the bypass posi- tion.
Plunger Home The metering piston moves into the home posi- tion.
Valve Bypass
Needle Up The needle arm moves into the upper position. Valve Bypass
Vial to Seat The gripper arm moves the specified vial to the needle seat.
Valve Bypass Needle Up
Needle into Sample The needle arm moves down and the needle pene- trates into the vial.
NOTE Always the default needle offset value (0.0 mm) is applied.
Valve Bypass Needle Up Vial to Seat
Draw The metering device draws the specified volume into the loop capillary.
NOTE This step can be performed multiple times, although the maximum volume of the analytical head will not be exceeded. Use Plunger Home to reset the metering device.
Valve Bypass Needle Up Vial to Seat Needle into Sam-
ple
Needle Up The needle arm moves into the upper position, lift- ing the needle out of the vial.
Valve Bypass
Vial to Tray The gripper arm returns the vial into its original position in the vial drawer.
Needle Up
Needle into Seat The needle moves down into the needle seat. Needle Up (only if a vial is present on the nee- dle seat)
Valve Mainpass The injection valve switches into the mainpass position.
Needle into Seat
Reset The sampler gets reset and will reinitialize
Agilent InfinityLab LC Series Vialsamplers User Manual 185
6 Troubleshooting and Diagnostics Maintenance and Troubleshooting Tools
Park Arm For transportation of the Vialsampler, the transport arm must be moved into the park position and secured with the transport foam ( Transport Protection Foam (G7129-40050)).
The Park Arm function can be found under Special Commands in the Instrument Control tab of the Agilent Lab Advisor software.
CAUTION Unsecured transportation of the module
The Vialsampler must be transported in the park position with the transport foam installed. Failing to do so might lead to damage to the internal parts.
Ensure that all vials are removed from the vial drawers.
Ensure that the transport assembly is in the park position.
Ensure that the Transport Protection Foam (G7129-40050) is correctly installed.
Agilent InfinityLab LC Series Vialsamplers User Manual 186
6 Troubleshooting and Diagnostics Diagnostic Tests
Diagnostic Tests
Sample Cooler Function Test The Sample Cooler Function Test is a diagnostic test to verify the correct functioning of the Sample Cooler/Sample Thermostat. The test takes up to 15 min to complete and returns a pass/fail type result. If the test failed or was aborted by the system, the final report will include some information on the possible root causes.
Before the test starts, the compressor is turned off to allow the system to reach the initial conditions. The test starts with acquiring data from the evaporator temperature sensor. If the reading is stable for at least 10 s (T < 0.5 C), the compressor turns on and the temperature inside the cooler/thermostat starts to drop. For the test to succeed, the system must pass three temperature checkpoints in a timely manner. These checkpoints are the following: Checkpoint 1: The temperature drops by 1/3 of the difference between the
starting temperature and 5 C. Checkpoint 2: The temperature drops below 5 C. Checkpoint 3: The temperature stabilizes at a value below 5 C and stays
stable for at least 60 s (T < 1.0 C).
For a Sample Thermostat, the heater resistance of the heating elements will also be tested and checked if the measured value is within the acceptance range (5 9 Ohm).
NOTE For testing the heater resistance of the Sample Thermostat, Lab Advisor version B.02.11 or higher is needed.
Agilent InfinityLab LC Series Vialsamplers User Manual 187
6 Troubleshooting and Diagnostics Diagnostic Tests
Heater Test The Heater Test is a diagnostic test to verify the correct functioning of the heating element of the Integrated Column Compartment (ICC). It measures the heating rate of the column oven and compares it against the threshold value. The test takes less than 1 min to complete and returns a pass/fail type result.
For the test to start, the actual temperature in the column oven must be equal to or below 60 C. If the temperature is above 60 C, the heater turns off and the system waits until the starting condition is reached.
NOTE To circumvent any negative effect on the test result, the Sample Cooler/Sample Thermostat is turned off for the duration of the test if present.
Agilent InfinityLab LC Series Vialsamplers User Manual 188
6 Troubleshooting and Diagnostics Agilent Lab Advisor Software
Agilent Lab Advisor Software
The Agilent Lab Advisor Software (basic license, shipped with an Agilent LC pump) is a standalone product that can be used with or without a chromatographic data system. Agilent Lab Advisor helps to manage the lab for high-quality chromatographic results by providing a detailed system overview of all connected analytical instruments with instrument status, Early Maintenance Feedback counters (EMF), instrument configuration information, and diagnostic tests. By the push of a button, a detailed diagnostic report can be generated. Upon request, the user can send this report to Agilent for a significantly improved troubleshooting and repair process. The Agilent Lab Advisor software is available in two versions: Lab Advisor Basic Lab Advisor Advanced
Lab Advisor Basic is included with every Agilent 1200 Infinity Series and Agilent InfinityLab LC Series instrument.
The Lab Advisor Advanced features can be unlocked by purchasing a license key, and include real-time monitoring of instrument actuals, all various instrument signals, and state machines. In addition, all diagnostic test results, calibration results, and acquired signal data can be uploaded to a shared network folder. The Review Client included in Lab Advisor Advanced allows to load and examine the uploaded data no matter on which instrument it was generated. This makes Data Sharing an ideal tool for internal support groups and users who want to track the instrument history of their analytical systems.
The optional Agilent Maintenance Wizard Add-on provides an easy-to-use, step-by-step multimedia guide for performing preventive maintenance on Agilent 1200 Infinity and Agilent InfinityLab LC Series instrument.
The tests and diagnostic features that are provided by the Agilent Lab Advisor software may differ from the descriptions in this manual. For details, refer to the Agilent Lab Advisor software help files.
Agilent InfinityLab LC Series Vialsamplers User Manual 189
7 Error Information
What are Error Messages 192 General Error Messages 193 Timeout 193 Shutdown 194 Remote Timeout 195 Lost CAN Partner 195 Leak Sensor 196 Leak Sensor Open 197 Leak Sensor Short 198 Compensation Sensor Open 199 Compensation Sensor Short 199 Fan Failed 200
Vialsampler Error Messages 201 Command failed errors 201 Metering home failed 202 Automatic Referencing procedure failed 202 Needle up/down failed errors 203 Missing vial errors 204 Motor overtemp 205 Motor alignment failed 206 Invalid vial position errors 207 Cleanup after abort failed 208 Wash procedure failed 208 Start of maintenance procedure failed errors 209 Injection valve position not found 209 Cannot move metering, the metering type definition is incorrect 210 Initialization failed errors 210 Unexpected block during axis movement 211 Timeout during axis movement 212
Agilent InfinityLab LC Series Vialsamplers User Manual 190
7 Error Information Agilent Lab Advisor Software
Sample Cooler/Sample Thermostat Error Messages 213 Sample temperature control voltage too low, check fuses and wires 213 Sample temperature control switched off due to condensate 213 Sample temperature control switched off due to overpressure 214 Sample temperature control sensor electronics calibration failed 214 Sample temperature control switched off due to supply voltage drop 215 Cooler condensate sensor defect 215 Cooler PCB is in error mode 215 Cooler condenser fan failed 216 Thermostat communication error 216 Heater defect 216 Heater in operating error 217 Thermostat sensor defect 217 Compressor has error 217 Sample Thermostat unknown 218 Thermostat fan defect 218
Integrated Column Compartment (ICC) Heater Error Messages 219 Wait for temperature function timed out 219 Heater over-current protection active 219 A heater sensor reading failed 220 Trend analysis for the heater temperature sensors failed 220 Maximum heater temperature exceeded 221 Column heater not connected 221
This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.
Agilent InfinityLab LC Series Vialsamplers User Manual 191
7 Error Information What are Error Messages
What are Error Messages
Error messages are displayed in the user interface when an electronic, mechanical, or hydraulic (flow path) failure occurs which requires attention before the analysis can be continued (for example, repair, or exchange of consumables is necessary). In the event of such a failure, the red status indicator at the front of the module is switched on, and an entry is written into the module logbook.
If an error occurs outside a method run, other modules will not be informed about this error. If it occurs within a method run, all connected modules will get a notification, all LEDs get red and the run will be stopped. Depending on the module type, this stop is implemented differently. For example, for a pump the flow will be stopped for safety reasons. For a detector, the lamp will stay on in order to avoid equilibration time. Depending on the error type, the next run can only be started, if the error has been resolved, for example liquid from a leak has been dried. Errors for presumably single time events can be recovered by switching on the system in the user interface.
Special handling is done in case of a leak. As a leak is a potential safety issue and may have occurred at a different module from where it has been observed, a leak always causes a shutdown of all modules, even outside a method run.
In all cases, error propagation is done via the CAN bus or via an APG/ERI remote cable (see documentation for the APG/ERI interface).
Agilent InfinityLab LC Series Vialsamplers User Manual 192
7 Error Information General Error Messages
General Error Messages
Timeout Error ID: 0062
The timeout threshold was exceeded.
Probable cause Suggested actions
1 The analysis was completed successfully, and the timeout function switched off the module as requested.
Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required.
2 A not-ready condition was present during a sequence or multiple-injection run for a period longer than the timeout threshold.
Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required.
Agilent InfinityLab LC Series Vialsamplers User Manual 193
7 Error Information General Error Messages
Shutdown Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals. A LOW signal input on pin 4 of the remote connector generates the error message.
Probable cause Suggested actions
1 Leak detected in another module with a CAN connection to the system.
Fix the leak in the external instrument before restarting the module.
2 Leak detected in an external instrument with a remote connection to the system.
Fix the leak in the external instrument before restarting the module.
3 Shut-down in an external instrument with a remote connection to the system.
Check external instruments for a shut-down con- dition.
4 The degasser failed to generate sufficient vacuum for solvent degassing.
Check the vacuum degasser for an error condi- tion. Refer to the Service Manual for the degasser or the pump that has the degasser built-in.
Agilent InfinityLab LC Series Vialsamplers User Manual 194
7 Error Information General Error Messages
Remote Timeout Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is started, the system expects all not-ready conditions (for example, a not-ready condition during detector balance) to switch to run conditions within one minute of starting the analysis. If a not-ready condition is still present on the remote line after one minute the error message is generated.
Lost CAN Partner Error ID: 0071
During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated.
Probable cause Suggested actions
1 Not-ready condition in one of the instru- ments connected to the remote line.
Ensure the instrument showing the not-ready condition is installed correctly, and is set up cor- rectly for analysis.
2 Defective remote cable. Exchange the remote cable.
3 Defective components in the instrument showing the not-ready condition.
Check the instrument for defects (refer to the instruments documentation).
Probable cause Suggested actions
1 CAN cable disconnected. Ensure all the CAN cables are connected cor- rectly.
Ensure all CAN cables are installed correctly.
2 Defective CAN cable. Exchange the CAN cable.
3 Defective mainboard in another module. Switch off the system. Restart the system, and determine which module or modules are not rec- ognized by the system.
Agilent InfinityLab LC Series Vialsamplers User Manual 195
7 Error Information General Error Messages
Leak Sensor Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted temperature-compensation sensor) are used by the leak algorithm to determine whether a leak is present. When a leak occurs, the leak sensor is cooled by the solvent. This changes the resistance of the leak sensor which is sensed by the leak sensor circuit on the main board.
Probable cause Suggested actions
1 Loose fittings. Ensure all fittings are tight.
2 Broken capillary. Exchange defective capillaries.
3 Leaking rotor seal or needle seat. Exchange the rotor seal or seat capillary.
4 Defective metering seal. Exchange the metering seal.
Make sure the leak sensor is thoroughly dry before restarting the autosampler.
5 Leaking peristaltic pump Exchange the peristaltic pump.
NOTE The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf respectively.
Agilent InfinityLab LC Series Vialsamplers User Manual 196
7 Error Information General Error Messages
Leak Sensor Open Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated.
Probable cause Suggested actions
1 Leak sensor not connected to the power switch board.
Please contact your Agilent service representa- tive.
2 Defective leak sensor. Please contact your Agilent service representa- tive.
3 Leak sensor incorrectly routed, being pinched by a metal component.
Please contact your Agilent service representa- tive.
4 Power switch assembly defective Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 197
7 Error Information General Error Messages
Leak Sensor Short Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current increases above the upper limit, the error message is generated.
Probable cause Suggested actions
1 Defective leak sensor. Please contact your Agilent service representa- tive.
2 Leak sensor incorrectly routed, being pinched by a metal component.
Please contact your Agilent service representa- tive.
3 Power switch assembly defective Please contact your Agilent service representa- tive.
4 Cable or contact problem. Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 198
7 Error Information General Error Messages
Compensation Sensor Open Error ID: 0081
The ambient-compensation sensor (NTC) on the power switch board in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the power switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
Compensation Sensor Short Error ID: 0080
The ambient-compensation sensor (NTC) on the power switch board in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the power switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor falls below the lower limit, the error message is generated.
Probable cause Suggested actions
1 Loose connection between the power switch board and the mainboard
Please contact your Agilent service representa- tive.
2 Defective power switch assembly Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Defective power switch assembly Please contact your Agilent service representa- tive.
2 Loose connection between the power switch board and the mainboard
Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 199
7 Error Information General Error Messages
Fan Failed Error ID: 0068
The fan in the autosampler module or in the Sample Cooler/Sample Thermostat has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan speed. If the fan speed falls below a certain limit for a certain length of time, the error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned off to assure that the module does not overheat inside.
Probable cause Suggested actions
1 Fan cable disconnected. Please contact your Agilent service representa- tive.
2 Defective fan. Please contact your Agilent service representa- tive.
3 Defective Sample Cooler/Sample Thermo- stat fan
Replace the Sample Cooler/Sample Thermostat.
4 Defective mainboard. Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 200
7 Error Information Vialsampler Error Messages
Vialsampler Error Messages
Command failed errors These are non-specific error messages and only indicate that something went wrong during the execution of the given action command. They are usually coupled with other error events, which provide more specific information on the nature of the issue. To identify these source error events, look for entries with the same timestamp as for the Command failed error message. Error ID: 34202 Metering command failed Error ID: 34205 Draw command failed Error ID: 34208 Eject command failed Error ID: 34211 Mix command failed Error ID: 34214 Needle command failed Error ID: 34217 Transport command failed Error ID: 34220 Valve command failed
Probable cause Suggested actions
1 Command failed Identify the source error message and follow the instructions provided for that.
Agilent InfinityLab LC Series Vialsamplers User Manual 201
7 Error Information Vialsampler Error Messages
Metering home failed Error ID: 34224
The home position of the metering plunger is not found or the plunger is unable to move.
Automatic Referencing procedure failed Error ID: 34225
The execution of the Automatic Referencing procedure was aborted by the system due to an unexpected error.
Probable cause Suggested actions
1 Blockage in the flow path. Check if there is a significant pressure difference (> 10 15 bar) between the mainpass and the bypass position of the injection valve. If yes, sys- tematically replace the components of the main- pass: Needle seat, Needle, Rotor seal, Sample loop
2 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
2 Physical blockage Look for potential sources of physical blockage in the path of the gripper arm and eliminate them. Restart the sampler and observe if the problem persists.
Agilent InfinityLab LC Series Vialsamplers User Manual 202
7 Error Information Vialsampler Error Messages
Needle up/down failed errors Error ID: 34303, 34304 Error ID: 34303 Needle up failed Error ID: 34304 Needle down failed
The upper or the bottom position of the needle arm is not found or the needle arm is unable to move.
Probable cause Suggested actions
1 Physical blockage Look for potential sources of physical block- age in the path of the needle arm and elimi- nate them.
Verify the correct functioning of the needle arm by using the Injector Steps Lab Advisor tool.
Restart the sampler and observe if the prob- lem persists.
2 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 203
7 Error Information Vialsampler Error Messages
Missing vial errors Error ID: 34305, 34306 Error ID: 34305 Missing vial Error ID: 34306 No wash vial
The gripper failed to detect any vial in the specified position of the vial tray.
Probable cause Suggested actions
1 Missing vial Verify the presence of a suitable vial in the position specified in the method or sequence.
If an injector program or purification method is used, verify that there is no controversial information specified there.
Rerun the method/sequence and observe if the problem persists.
2 Unsuitable vial type Verify that the vials and caps used by the method comply with the requirements from Agilent.
3 Broken gripper Verify the correct functioning of the gripper by using the Gripper Verification tool. Replace the gripper if needed.
4 Misaligned gripper arm Verify the correct functioning of the gripper arm by using the Gripper Verification and Injector Steps tools. If the gripper arm appears to be out of alignment, contact Agilent for more informa- tion.
5 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 204
7 Error Information Vialsampler Error Messages
Motor overtemp Error ID: 34307
The ongoing transport movement is aborted because the transport assembly stepper motors are overheated. 34307,0 corresponds to the X-axis stepper motor. 34307,1 corresponds to the Z-axis stepper motor. 34307,2 corresponds to the theta stepper motor. 34307,3 corresponds to the gripper stepper motor.
Probable cause Suggested actions
1 Physical blockage Look for potential sources of physical block- age in the path of the gripper arm and elimi- nate them.
Verify that the X-axis movement is not hin- dered by any obstacle, for example, by a vial that is stuck below the transport rods in the transport assembly body.
Restart the sampler and observe if the prob- lem still persists.
2 Broken gripper Verify the correct functioning of the gripper by using the Gripper Verification tool. Replace the gripper if needed.
3 Contaminated transport rods Inspect the x- and z-axis transport rods and clean them if you observe contamination or dust accu- mulation on them. Use isopropanol and lint-free cloth for this purpose.
4 Misaligned gripper arm Verify the correct functioning of the gripper arm by using the Gripper Verification and Injector Steps tools. If the gripper arm appears to be out of alignment, contact Agilent for more informa- tion.
5 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 205
7 Error Information Vialsampler Error Messages
Motor alignment failed Error ID: 34221
The alignment of the transport assembly stepper motors failed during the initialization process. 34221,0 corresponds to the X-axis stepper motor. 34221,1 corresponds to the Z-axis stepper motor. 34221,2 corresponds to the theta stepper motor. 34221,3 corresponds to the gripper stepper motor.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 206
7 Error Information Vialsampler Error Messages
Invalid vial position errors Error ID: 34309, 34313
The vial position defined in the method/sequence is invalid.
The vial drawers are automatically recognized by the reflection sensors on the transport assembly when they are installed in the Vialsampler. The vial drawers are distinguished on the basis of the tag information on their rear side (white markers). If the information from the optical sensors is not in agreement with the vial position defined in the method/sequence, the ongoing analysis will be aborted and the respective error ID reported. Error ID: 34309 Invalid vial position Error ID: 34313 Invalid wash vial position
Probable cause Suggested actions
1 Missing vial tray Verify that the correct vial drawer type is installed in the sampler and the drawer is pushed all the way in. Restart the sampler and observe if the problem still persists.
2 Missing tag information Verify that the white markers are still intact and they are not covered by dirt or dust. If necessary, clean the markers with isopropanol or replace the vial tray.
3 Misaligned gripper arm Verify the correct functioning of the gripper arm by using the Gripper Verification and Injector Steps tools. If the gripper arm appears to be out of alignment, contact Agilent for more informa- tion.
4 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 207
7 Error Information Vialsampler Error Messages
Cleanup after abort failed Error ID: 34310
This, similarly to the Command failed error events, is a non-specific error message and indicates only that something went wrong during the cleanup procedure. The cleanup procedure refers to the process, which is executed when an analysis is aborted (for example, due to a missing vial) and the components of the sampler (such as the needle, gripper arm, and the metering device) need to be reset to the initial conditions. As for the Command failed error events, this error message is usually also coupled with other error events, which provide more specific information on the nature of the issue. To identify these source error events, look for entries with the same timestamp as for the Cleanup after abort failed error message.
Wash procedure failed Error ID: 34314
This, similarly to the Command failed error events, is a non-specific error message and indicates only that something went wrong during the wash procedure. It is usually coupled with other error events, which provide more specific information on the nature of the issue. To identify these source error events, look for entries with the same timestamp as for the Wash procedure failed error message.
Probable cause Suggested actions
1 Cleanup procedure failed Identify the source error message and follow the instructions provided for that.
Probable cause Suggested actions
1 Wash procedure failed Identify the source error message and follow the instructions provided for that.
Agilent InfinityLab LC Series Vialsamplers User Manual 208
7 Error Information Vialsampler Error Messages
Start of maintenance procedure failed errors Error ID: 34315, 34316, 34317
These, similarly to the Command failed error events, are non-specific error messages and indicate only that something went wrong with initiating the given maintenance procedure. These error messages are usually coupled with other error events, which provide more specific information on the nature of the issue. To identify these source error events, look for entries with the same timestamp as for the Start of maintenance procedure failed error message. Error ID: 34315 Start of needle/seat change procedure failed Error ID: 34316 Start of piston change procedure failed Error ID: 34317 Start of gripper change procedure failed
Injection valve position not found Error ID: 34319
The injection valve position is not found or the valve is unable to switch.
Probable cause Suggested actions
1 Maintenance procedure failed Identify the source error message and follow the instructions provided for that.
Probable cause Suggested actions
1 Overtightened screws Verify that the stator screws are not overtight- ened as it can lead to hindered rotor movements. If necessary, loosen the stator screws and retighten them appropriately.
2 Broken rotor seal Check the rotor seal for signs of visible damage and if necessary, replace it.
3 Broken stator Check the stator for signs of visible damage and if necessary, replace it.
4 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 209
7 Error Information Vialsampler Error Messages
Cannot move metering, the metering type definition is incorrect Error ID: 34320
The analytical head is not operational because there is no valid information available for its type in the firmware.
Initialization failed errors These are non-specific error messages and indicate only that something went wrong during the initialization of the given component. They are usually coupled with other error events, which provide more specific information on the nature of the issue. To identify these source error events, look for entries with the same timestamp as for the Initialization failed error message.
Error ID: 34322 Metering device initialization failed Error ID: 34323 Injection valve initialization failed Error ID: 34324 Needle device initialization failed Error ID: 34325 Transport initialization failed Error ID: 34326 Sampler initialization failed
Probable cause Suggested actions
1 Missing configuration data Verify that the type of the analytical head is cor- rectly configured in the CDS or Lab Advisor.
2 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Initialization failed Identify the source error message and follow the instructions provided for that.
Agilent InfinityLab LC Series Vialsamplers User Manual 210
7 Error Information Vialsampler Error Messages
Unexpected block during axis movement Error ID: 34329
The ongoing transport movement is aborted due to a potential blockage in the pathway of the gripper arm. 34329,0 corresponds to the X-axis movement. 34329,1 corresponds to the Z-axis movement. 34329,2 corresponds to the theta movement. 34329,3 corresponds to the gripper finger movements.
Probable cause Suggested actions
1 Physical blockage Look for potential sources of physical blockage in the path of the gripper arm and eliminate them. Restart the sampler and observe if the problem persists.
2 Broken gripper Verify the correct functioning of the gripper by using the Gripper Verification tool. Replace the gripper if needed.
3 Contaminated transport rods Inspect the x- and z-axis transport rods and clean them if you observe contamination or dust accu- mulation on them. Use isopropanol and lint-free cloth for this purpose.
4 Misaligned gripper arm Verify the correct functioning of the gripper arm by using the Gripper Verification and Injector Steps tools. If the gripper arm appears to be out of alignment, contact Agilent for more informa- tion.
5 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 211
7 Error Information Vialsampler Error Messages
Timeout during axis movement Error ID: 34330
The system reported an error because the execution of the given transport command took longer than expected. 34330,0 corresponds to the X-axis movement. 34330,1 corresponds to the Z-axis movement. 34330,2 corresponds to the theta movement. 34330,3 corresponds to the gripper finger movements.
Probable cause Suggested actions
1 Contaminated transport rods Inspect the x- and z-axis transport rods and clean them if you observe contamination or dust accu- mulation on them. Use isopropanol and lint-free cloth for this purpose.
2 Broken gripper Verify the correct functioning of the gripper by using the Gripper Verification tool. Replace the gripper if needed.
3 Potential hardware error Please contact your Agilent service representa- tive.
4 Misaligned gripper arm Verify that the alignment parameters are in the correct range and perform the Automatic Refer- encing if needed.
Agilent InfinityLab LC Series Vialsamplers User Manual 212
7 Error Information Sample Cooler/Sample Thermostat Error Messages
Sample Cooler/Sample Thermostat Error Messages
Sample temperature control voltage too low, check fuses and wires Error ID: 30713
The compressor voltage is below the lower threshold value.
Sample temperature control switched off due to condensate Error ID: 30715
The cooler/thermostat was switched off due to a condensate event.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Overfilled container Empty the condensate container. Verify that the open end of the tubing doesnt immerse in the liq- uid.
2 Drainage issues Verify the correct plumbing of the conden- sate drainage system.
Make sure that no kinks or mechanical blocks are present in the drainage system.
Avoid the formation of the siphoning effect.
Make sure that the hosting sampler is level.
Agilent InfinityLab LC Series Vialsamplers User Manual 213
7 Error Information Sample Cooler/Sample Thermostat Error Messages
Sample temperature control switched off due to overpressure Error ID: 30716
The pressure in the refrigerant circuit exceeded the maximum allowed level. To prevent any damage to the system, the compressor was turned off.
Sample temperature control sensor electronics calibration failed Error ID: 30717
The system is in the error state because the calibration of the analog temperature sensor has failed.
Probable cause Suggested actions
1 Overheated condenser Turn off the cooler/thermostat and wait for 15 min to allow the system to cool down. Verify if there is enough space around the sampler for adequate ventilation and the cooler/thermostat is not exposed to direct sunlight.
2 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Sampler incompatibility If the hosting sampler is a Vialsampler, verify its compatibility with the Sample Cooler installed. Units with the serial number DEBAT02000 or below are equipped with an analog temperature sensor that is not compatible with the Vialsam- pler.
2 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 214
7 Error Information Sample Cooler/Sample Thermostat Error Messages
Sample temperature control switched off due to supply voltage drop Error ID: 30718
The compressor is turned off due to an unexpected drop in the supply voltage.
Cooler condensate sensor defect Error ID: 30719
The condensate sensor of the cooler/thermostat is not working properly.
Cooler PCB is in error mode Error ID: 30275
The system is in the error state because the compressor control board has encountered an unexpected error.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 215
7 Error Information Sample Cooler/Sample Thermostat Error Messages
Cooler condenser fan failed Error ID: 30726
The condenser fan of the cooler/thermostat is not working properly.
Thermostat communication error Error ID: 30738
The system is in the error state because the communication between the sampler and the thermostat has failed.
Heater defect Error ID: 30739
The heating function of the thermostat is not working properly.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 216
7 Error Information Sample Cooler/Sample Thermostat Error Messages
Heater in operating error Error ID: 30744
The system is in the error state because the thermostat heater has encountered an unexpected error.
Thermostat sensor defect Error ID: 30751
One of the digital temperature sensors of the cooler/thermostat is not working properly.
Compressor has error Error ID: 30756
The system is in the error state because the control board of the compressor has encountered an unexpected error.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 217
7 Error Information Sample Cooler/Sample Thermostat Error Messages
Sample Thermostat unknown Error ID: 30768
Thermostat fan defect Error ID: 30771
One of the cooling fans of the cooler/thermostat is not working properly.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 218
7 Error Information Integrated Column Compartment (ICC) Heater Error Messages
Integrated Column Compartment (ICC) Heater Error Messages
Wait for temperature function timed out Error ID: 30820
The system reported an error because reaching the set temperature took longer than expected. To verify the correct functioning of the heating, perform the Heater Test in the Agilent Lab Advisor software.
Heater over-current protection active Error ID: 30821
The heating function of the Integrated Column Compartment is switched due to overcurrent.
To verify the correct functioning of the heating element, perform the Heater Test in the Agilent Lab Advisor software.
Probable cause Suggested actions
1 Connection problems Power off the system and verify that the ICC is correctly mounted in the sampler.
Power on the system and check if the prob- lem persists.
2 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Connection problems Power off the system and verify that the ICC is correctly mounted in the sampler.
Power on the system and check if the prob- lem persists.
2 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 219
7 Error Information Integrated Column Compartment (ICC) Heater Error Messages
A heater sensor reading failed Error ID: 30822
The heating function of the Integrated Column Compartment is switched off due to the malfunctioning of the digital temperature sensor.
To verify the correct functioning of the heating element, perform the Heater Test in the Agilent Lab Advisor software.
Trend analysis for the heater temperature sensors failed Error ID: 30823
The heating function of the Integrated Column Compartment is switched off because the trend analysis of the reading from the digital temperature sensor has failed.
To verify the correct functioning of the heating element, perform the Heater Test in the Agilent Lab Advisor software.
Probable cause Suggested actions
1 Connection problems Power off the system and verify that the ICC is correctly mounted in the sampler.
Power on the system and check if the prob- lem persists.
2 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Connection problems Power off the system and verify that the ICC is correctly mounted in the sampler.
Power on the system and check if the prob- lem persists.
2 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 220
7 Error Information Integrated Column Compartment (ICC) Heater Error Messages
Maximum heater temperature exceeded Error ID: 30824
The heating function of the Integrated Column Compartment is switched off because the temperature registered by the temperature sensor exceeds the upper limit (100 C).
To verify the correct functioning of the heating element, perform the Heater Test in the Agilent Lab Advisor software
Column heater not connected Error ID: 30825
The Integrated Column Compartment is either not connected or cannot be recognized by the system.
Probable cause Suggested actions
1 Connection problems Power off the system and verify that the ICC is correctly mounted in the sampler.
Power on the system and check if the prob- lem persists.
2 Potential hardware error Please contact your Agilent service representa- tive.
Probable cause Suggested actions
1 Connection problems Power off the system and verify that the ICC is correctly mounted in the sampler.
Power on the system and check if the prob- lem persists.
2 Potential hardware error Please contact your Agilent service representa- tive.
Agilent InfinityLab LC Series Vialsamplers User Manual 221
8 Maintenance
Introduction to Maintenance 223 Warnings and Cautions 224 Overview of Maintenance 226 Cleaning the Module 227 Remove and Install Doors 228 Exchange the Needle Assembly 229 Exchange the Needle Seat Assembly 235 Exchange the Sample Loop Assembly 241 Exchange the Rotor Seal 245 Exchange the Metering Seal and Piston 250 Replace the Analytical Heads 255 Exchange the Gripper Arm 258 Replace the Peristaltic Pump Cartridge 261 Exchange the Wash Port Assembly 263 Replace the Module Firmware 269 Replace the Sample Cooler/Sample Thermostat 270
This chapter describes the maintenance of the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 222
8 Maintenance Introduction to Maintenance
Introduction to Maintenance
The module is designed for easy maintenance. Maintenance can be done from the front with module in place in the system stack.
Figure 33 Overview of the Vialsampler
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch
Agilent InfinityLab LC Series Vialsamplers User Manual 223
8 Maintenance Warnings and Cautions
Warnings and Cautions
WARNING Personal injury or damage to the product
Agilent is not responsible for any damages caused, in whole or in part, by improper use of the products, unauthorized alterations, adjustments or modifications to the products, failure to comply with procedures in Agilent product user guides, or use of the products in violation of applicable laws, rules or regulations.
Use your Agilent products only in the manner described in the Agilent product user guides.
WARNING Electrical shock
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the cover is opened.
Do not remove the cover of the module.
Only certified persons are authorized to carry out repairs inside the module.
WARNING Sharp metal edges
Sharp-edged parts of the equipment may cause injuries.
To prevent personal injury, be careful when getting in contact with sharp metal areas.
Agilent InfinityLab LC Series Vialsamplers User Manual 224
8 Maintenance Warnings and Cautions
WARNING Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
The volume of substances should be reduced to the minimum required for the analysis.
Do not operate the instrument in an explosive atmosphere.
CAUTION Safety standards for external equipment
If you connect external equipment to the instrument, make sure that you only use accessory units tested and approved according to the safety standards appropriate for the type of external equipment.
WARNING Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
Agilent InfinityLab LC Series Vialsamplers User Manual 225
8 Maintenance Overview of Maintenance
Overview of Maintenance
It is necessary to perform periodic inspection of this instrument to ensure its safe use. It is possible to have these periodic inspections performed by Agilent service representatives on a contractual basis. For information regarding the maintenance inspection contract, contact your Agilent representative.
The following pages describe the maintenance (simple repairs) of the module that can be carried out without opening the main cover.
Table 21 Overview of maintenance
Procedure Typical interval (minimum)
Change needle/needle seat 30000 needle into seat movements
Change peristaltic pump cartridge 3000 h on time
Change rotor seal 30000 injections
Change metering seal 30000 injections
Agilent InfinityLab LC Series Vialsamplers User Manual 226
8 Maintenance Cleaning the Module
Cleaning the Module
To keep the module case clean, use a soft cloth slightly dampened with water, or a solution of water and mild detergent. Avoid using organic solvents for cleaning purposes. They can cause damage to plastic parts.
WARNING Liquid dripping into the electronic compartment of your module can cause shock hazard and damage the module
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any connections in the flow path.
NOTE A solution of 70 % isopropanol and 30 % water might be used if the surface of the module needs to be disinfected.
Agilent InfinityLab LC Series Vialsamplers User Manual 227
8 Maintenance Remove and Install Doors
Remove and Install Doors
When The instrument doors or the hinges are broken.
Tools required Description Flat screwdriver
Parts required p/n Description G7129-68702 Door Kit for Infinity II Vialsampler/Preparative Autosampler
Preparations Finish any pending acquisition job.
1 Press the release buttons and pull the front door out.
2 For the Installation of the front door, insert the hinges into their guides and push the door in until the release buttons click into their final position.
Agilent InfinityLab LC Series Vialsamplers User Manual 228
8 Maintenance Exchange the Needle Assembly
Exchange the Needle Assembly
When The needle is visibly damaged. Leaks or blockages are observed. The limit for the needle interaction EMF counter is exceeded. The needle needs to be replaced as part of the yearly maintenance.
Tools required p/n Description 8710-1924 Open-end wrench 14 mm 8710-2140 Screwdriver Torx TX-10
Parts required # p/n Description 1 G7129-87200 Needle assembly, 1260 Vialsampler
for G7129-87017 needle seat OR 1 G7129-87201 Needle assembly, 1290 Vialsampler
for G7129-87012 needle seat OR 1 G7129-87202 Needle assembly, slotted, for high injection volumes
Preparations Finish any pending acquisition job. Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent. Close the shutoff valves at the pump if available.
WARNING Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
Agilent InfinityLab LC Series Vialsamplers User Manual 229
8 Maintenance Exchange the Needle Assembly
WARNING Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
Do not open the safety cover of the needle station during normal operation.
Wear safety goggles and safety gloves when removing the needle assembly.
NOTE It is recommended to always exchange the needle assembly and the needle seat at the same time to prevent premature leakage.
NOTE When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See Table 15 on page 155.
1 In the Agilent Lab Advisor software select Service & Diagnostics> Maintenance Positions> Change Needle, Loop and Seat, click Start and wait until the needle assembly is in maintenance position.
OR
In the Local Controller start the maintenance mode and select Change Needle, Loop and Seat function.
2 Open the doors of the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 230
8 Maintenance Exchange the Needle Assembly
3 Loosen the safety cover screw. WARNING Sharp needle Uncovered needles may cause injuries
Do not touch the tip of the needle.
4 Remove the safety cover.
WARNING Sharp needle Uncovered needles may cause injuries.
Do not change the needle seat at this point.
Do not touch the tip of the needle.
5 Move the wash port into the service position by slightly lifting it up and then flapping to the left.
6 In Lab Advisor use Next to move the needle in the down position (ca. 2 mm above the seat).
OR
In the Local Controller, move the needle down until the needle tip is ca. 2 4 mm above the seat.
Agilent InfinityLab LC Series Vialsamplers User Manual 231
8 Maintenance Exchange the Needle Assembly
7 Disconnect the sample loop from the needle, and, if needed, also from the metering device.
8 Loosen the fixing screw.
9 Lift out the needle. 10 Mount and align the needle on its fixture, then tighten the screw firmly.
Agilent InfinityLab LC Series Vialsamplers User Manual 232
8 Maintenance Exchange the Needle Assembly
11 Align the needle tip with the seat and ensure that they are concentric. If needed, carefully bend the needle into the right position with your fingers.
12 Reconnect the sample loop to the needle. Make sure of the correct positioning of the loop capillary, the uncoated part of the capillary must be horizontal.
NOTE Do not overtighten the fitting!
NOTE Incorrect positioning and installation of the needle/loop connection can result in damaging or breaking the sample loop.
13 In Lab Advisor use Next to lift the needle slowly into the up position.
OR
In the Local Controller, move the needle up to the upper- most position.
14 Move the wash port back into its normal position by turning it to the right.
Agilent InfinityLab LC Series Vialsamplers User Manual 233
8 Maintenance Exchange the Needle Assembly
15 Install the safety cover (1.) and fix the screw (2.). 16 In Lab Advisor, select Next to move the needle into the needle seat and then Back to finish the maintenance procedure.
OR
In the Local Controller, select Continue to move the nee- dle into the needle seat and then Exit to finish the main- tenance procedure.
17 Close the doors. 18 Perform a System Pressure Test.
(1.) (2.)
Agilent InfinityLab LC Series Vialsamplers User Manual 234
8 Maintenance Exchange the Needle Seat Assembly
Exchange the Needle Seat Assembly
When The seat is visibly damaged. Leaks or blockages are observed. The limit for the seat interaction EMF counter is exceeded. The needle seat needs to be replaced as part of the yearly maintenance.
Tools required p/n Description 8710-1924 Open-end wrench 14 mm 8710-2140 Screwdriver Torx TX-10
Flat screwdriver
Parts required # p/n Description 1 G7129-87017 Seat Assembly, 1260 Vialsampler, PEEK, 0.17 mm
OR 1 G7129-87012 Seat Assembly, 1290 Vialsampler, PEEK, 0.12 mm OR 1 G7129-87117 Seat Assembly, 1260 Vialsampler, Vespel, 0.17 mm OR 1 G7129-87112 Seat Assembly, 1290 Vialsampler, Vespel, 0.12 mm
Preparations Finish any pending acquisition job. Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent. Close the shutoff valves at the pump if available.
WARNING Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
Do not open the safety cover of the needle station during normal operation.
Wear safety goggles and safety gloves when removing the needle assembly.
NOTE When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See Table 15 on page 155.
NOTE It is recommended to always exchange the needle assembly and the needle seat at the same time to prevent premature leakage.
Agilent InfinityLab LC Series Vialsamplers User Manual 235
8 Maintenance Exchange the Needle Seat Assembly
NOTE The use of a Vespel needle seat (G7129-87112 and G7129-87117) limits the maximum system pressure to 600 bar.
1 In the Agilent Lab Advisor software select Service & Diagnostics> Maintenance Positions> Change Needle, Loop and Seat, click Start and wait until the needle assembly is in maintenance position.
OR
In the Local Controller start the maintenance mode and select Change Needle, Loop and Seat function.
2 Open the doors of the module.
3 Loosen the safety cover screw. WARNING Sharp needle Uncovered needles may cause injuries
Do not touch the tip of the needle.
4 Remove the safety cover.
Agilent InfinityLab LC Series Vialsamplers User Manual 236
8 Maintenance Exchange the Needle Seat Assembly
WARNING Sharp needle Uncovered needles may cause injuries.
Do not change the needle seat at this point.
Do not touch the tip of the needle.
5 Move the wash port into the service position by slightly lifting it up and then flapping to the left.
6 In Lab Advisor use Next to move the needle in the down position (ca. 2 mm above the seat).
OR
In the Local Controller, move the needle down until the needle tip is ca. 2 4 mm above the seat.
7 Disconnect the sample loop from the needle, and, if needed, also from the metering device.
8 Loosen the fixing screw.
Agilent InfinityLab LC Series Vialsamplers User Manual 237
8 Maintenance Exchange the Needle Seat Assembly
9 Lift out the needle.
10 Disconnect the seat capillary fitting from the injection valve (port 5) and ease out the needle seat from the leak tray with the help of a flathead screwdriver.
11 Insert the new needle seat and press it firmly into posi- tion. Then, connect the seat capillary fitting to port 5 of the injection valve.
12 Mount and align the needle on its fixture, then tighten the screw firmly.
Agilent InfinityLab LC Series Vialsamplers User Manual 238
8 Maintenance Exchange the Needle Seat Assembly
13 Reconnect the sample loop to the needle. Make sure of the correct positioning of the loop capillary, the uncoated part of the capillary must be horizontal.
NOTE Do not overtighten the fitting!
NOTE Incorrect positioning and installation of the needle/loop connection can result in damaging or breaking the sample loop.
14 In Lab Advisor use Next to lift the needle slowly into the up position.
OR
In the Local Controller, move the needle up to the upper- most position.
15 Move the wash port back into its normal position by turning it to the right.
16 Install the safety cover (1.) and fix the screw (2.).
(1.) (2.)
Agilent InfinityLab LC Series Vialsamplers User Manual 239
8 Maintenance Exchange the Needle Seat Assembly
17 In Lab Advisor, select Next to move the needle into the needle seat and then Back to finish the maintenance procedure.
OR
In the Local Controller, select Continue to move the nee- dle into the needle seat and then Exit to finish the main- tenance procedure.
18 Close the doors.
19 Perform a System Pressure Test.
Agilent InfinityLab LC Series Vialsamplers User Manual 240
8 Maintenance Exchange the Sample Loop Assembly
Exchange the Sample Loop Assembly
When The sample loop is visibly damaged. Leaks or blockages are observed.
Tools required p/n Description 8710-1924 Open-end wrench 14 mm
Parts required # p/n Description 1 G7129-60500 Sample Loop 100 L
OR 1 G7129-60400 Sample Loop 40 L OR 1 G7129-60300 Sample Loop 20 L
Preparations Finish any pending acquisition job. Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent. Close the shutoff valves at the pump if available.
WARNING Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
Do not open the safety cover of the needle station during normal operation.
Wear safety goggles and safety gloves when removing the needle assembly.
NOTE When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See Table 15 on page 155.
Agilent InfinityLab LC Series Vialsamplers User Manual 241
8 Maintenance Exchange the Sample Loop Assembly
1 In the Agilent Lab Advisor software select Service & Diagnostics> Maintenance Positions> Change Needle, Loop and Seat, click Start and wait until the needle assembly is in maintenance position.
OR
In the Local Controller start the maintenance mode and select Change Needle, Loop and Seat function.
2 Open the doors of the module.
3 Loosen the safety cover screw. WARNING Sharp needle Uncovered needles may cause injuries
Do not touch the tip of the needle.
4 Remove the safety cover.
Agilent InfinityLab LC Series Vialsamplers User Manual 242
8 Maintenance Exchange the Sample Loop Assembly
5 In Lab Advisor use Next to move the needle in the down position (ca. 2 mm above the seat).
OR
In the Local Controller, move the needle down until the needle tip is ca. 2 4 mm above the seat.
6 Disconnect the loop capillary from the metering device and from the needle.
7 Install the new sample loop. Make sure of the correct positioning of the loop capillary, the uncoated part of the capillary must be horizontal.
NOTE Incorrect positioning and installation of the needle/loop connection can result in damaging and breaking the sample loop.
8 After installing the new sample loop, ensure that the needle tip is concentric with the seat.
Agilent InfinityLab LC Series Vialsamplers User Manual 243
8 Maintenance Exchange the Sample Loop Assembly
9 Click on Next in the LabAdvisor to lift the Needle up again.
OR
In the Local Controller, move the needle up to the upper- most position.
10 Move the wash port back into its normal position by turning it to the right.
11 Install the safety cover (1.) and fix the screw (2.). 12 In Lab Advisor, select Next to move the needle into the needle seat and then Back to finish the maintenance procedure.
OR
In the Local Controller, select Continue to move the nee- dle into the needle seat and then Exit to finish the main- tenance procedure.
13 Close the doors. 14 Perform a System Pressure Test.
(1.) (2.)
Agilent InfinityLab LC Series Vialsamplers User Manual 244
8 Maintenance Exchange the Rotor Seal
Exchange the Rotor Seal
When Injection volume reproducibility problems are observed. Leaks or blockages are observed. The limit for the rotor seal EMF counter is exceeded. The rotor seal needs to be replaced as part of the yearly maintenance.
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch 8710-2394 Hex key 9/64 inch 15 cm long T-handle
Cleaning tissue and appropriate solvent like isopropanol or methanol
Parts required # p/n Description 1 5068-0007 Rotor Seal, Vespel, 1300 bar
for 1290 Infinity II Injection Valve OR 1 0101-1416 Rotor Seal, PEEK, 600/800 bar
for 1260 Infinity Injection Valve
Preparations Finish any pending acquisition job. Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent. Close the shutoff valves at the pump if available.
CAUTION Reduced life time of the injection valve
Component cleanliness is crucial for the life time of the injection valve.
Replace the rotor seal in a clean environment.
CAUTION Removing the stator head
The stator face is held in place by the stator head. When you remove the stator head, the stator face can fall out of the valve.
Carefully handle the valve to prevent damage to the stator face.
Carefully handle the stator face during sonication.
Agilent InfinityLab LC Series Vialsamplers User Manual 245
8 Maintenance Exchange the Rotor Seal
1 Open the doors of the module. 2 Remove all capillaries from the injection valve with a 1/4 inch wrench.
3 Loosen each fixing bolt two turns at a time. Remove the bolts from the head.
4 Remove the stator head and stator ring.
Agilent InfinityLab LC Series Vialsamplers User Manual 246
8 Maintenance Exchange the Rotor Seal
5 Remove the rotor seal and isolation seal. CAUTION Damage to the rotor seal and cross-port leaks
Before you replace the rotor seal, clean the stator.
Inspect the stator head and swab it with the appropri- ate solvent. If more stringent cleaning is required, use a sonicator. Inspect the remaining valve components for contamination. Clean them as necessary.
If the stator head is scratched, replace it.
6 Install the new rotor seal and isolation seal. Ensure the metal spring inside the isolation seal faces towards the valve body.
NOTE Make sure that the rotor sealing surface with its engraved flow passages is facing out. The pattern is asymmetrical to prevent improper placement.
Agilent InfinityLab LC Series Vialsamplers User Manual 247
8 Maintenance Exchange the Rotor Seal
7 Install the stator ring with the short of the two pins fac- ing towards you at the 12 o'clock position. Ensure the ring sits flat on the valve body.
8 Install the stator head. Tighten the bolts alternately two turns at a time until the stator head is secure.
NOTE Do not over-tighten the screws. The screws hold the assem- bly together and do not affect the sealing force. The sealing force is automatically set as the screws close the stator head against the valve body.
Agilent InfinityLab LC Series Vialsamplers User Manual 248
8 Maintenance Exchange the Rotor Seal
9 Reconnect the capillaries and tubes to the valve ports.
The correct plumbing is shown below and on the safety cover of the module.
10 Slide the waste tube into the waste holder in the leak tray.
11 Close the doors. 12 Perform a System Pressure Test.
Waste
Column
Pump
Agilent InfinityLab LC Series Vialsamplers User Manual 249
8 Maintenance Exchange the Metering Seal and Piston
Exchange the Metering Seal and Piston
When Injection volume reproducibility problems are observed Leaks or blockages are observed. The limit for the metering seal EMF counter is exceeded. The metering seal needs to be replaced as part of the yearly maintenance.
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch 8710-2392 4 mm Hex key 8710-2411 3 mm Hex key, 12 cm long 01018-23702 Insert tool
Cleaning tissue and appropriate solvent like isopropanol or methanol
Parts required p/n Description 0905-1717 Metering Seal, 40 L
OR 0905-1503 Metering Seal, 100 L OR 0905-1294 Metering Seal, 900 L
5067-5920 Piston, 40 L, Zirconia for 40 L analytical head
OR 5067-5678 Piston, 100 L, Zirconia for 100 L analytical head
OR G4267-60462 Piston, 900 L, Sapphire
Preparations Finish any pending acquisition job. Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent. Close the shutoff valves at the pump if available.
NOTE When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See Table 15 on page 155.
Agilent InfinityLab LC Series Vialsamplers User Manual 250
8 Maintenance Exchange the Metering Seal and Piston
1 In the Agilent Lab Advisor software, select Service & Diagnostics> Maintenance Positions> Change Metering Seal and Piston, click Start and wait until the metering plunger is in maintenance position.
OR
In the Local Controller start the maintenance mode and select Change Metering Device function.
2 Open the doors of the module.
3 Disconnect the two capillaries from the metering device.
4 Remove the two fixing bolts.
5 Remove the metering head assembly from the sampler. 6 Remove the two fixing bolts from the base of the meter- ing head assembly.
Agilent InfinityLab LC Series Vialsamplers User Manual 251
8 Maintenance Exchange the Metering Seal and Piston
7 Disassemble the metering head assembly.
NOTE The internal construction slightly differs for the 900 L analyitical head (refer to Figure 46 on page 296).
8 Use a seal insert tool to carefully remove the seal.
9 Clean the chamber with lint-free cloth and an appropri- ate solvent. Ensure all particulate matter is removed.
10 Use the plastic side of the insert tool to install the new seal. Press the seal firmly into position.
Agilent InfinityLab LC Series Vialsamplers User Manual 252
8 Maintenance Exchange the Metering Seal and Piston
11 Place the seal support ring on top of the seal and fix the screws.
NOTE There are no seal support screws for the 900 L analytical head.
12 Reassemble the metering head assembly. Carefully insert the piston into the base.
13 Install the fixing bolts and tighten them alternately two turns at a time until the stator head is secure. The side without the drain hole must be matching with the capil- lary port sitting on the non-truncated side of the analyti- cal head.
14 Install the metering head assembly in the autosampler. Ensure that the drain hole of the metering body is facing downwards. Tighten the bolts alternately two turns at a time until the metering device is secure.
15 Reinstall the capillaries. 16 In Lab Advisor, select Next to move the piston back into its normal position and then Back to finish the mainte- nance procedure.
OR
In the Local Controller, select Continue to move the pis- ton back into its normal position and then Exit to finish the maintenance procedure.
1.
2. 2.
Agilent InfinityLab LC Series Vialsamplers User Manual 253
8 Maintenance Exchange the Metering Seal and Piston
17 Close the doors. 18 Perform a System Pressure Test.
Agilent InfinityLab LC Series Vialsamplers User Manual 254
8 Maintenance Replace the Analytical Heads
Replace the Analytical Heads
When The maximum injection volume needs to be changed.
Tools required p/n Description 8710-0510 Open-end wrench 1/4 5/16 inch
Parts required # p/n Description 1 G7129-60084 Analytical Head Assembly 40 L
OR 1 G7129-60082 Analytical Head Assembly 100 L OR 1 G7129-60083 Analytical Head Assembly 900 L
Preparations Finish any pending acquisition job. Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent. Close the shutoff valves at the pump if available.
NOTE If the sampler with 900 L metering device is linked to a 1260 pump, such as the G7112B Binary Pump, the 400 bar pressure limit must be set manually in the method parameter settings of the pump (see Setup of Basic Pump Parameters chapter in the pump user manual). For all 1290 pumps, the pressure limit will be set automatically.
NOTE When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See Table 15 on page 155.
1 In the Agilent Lab Advisor software, select Service & Diagnostics> Maintenance Positions> Change Metering Seal and Piston, click Start and wait until the metering plunger is in maintenance position.
OR
In the Local Controller start the maintenance mode and select Change Metering Device function.
2 Open the doors of the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 255
8 Maintenance Replace the Analytical Heads
3 Disconnect the two capillaries from the metering device.
4 Remove the two fixing bolts.
5 Remove the metering head assembly from the sampler. 6 Install the metering head assembly in the autosampler. Ensure that the drain hole of the metering body is facing downwards. Tighten the bolts alternately two turns at a time until the metering device is secure.
7 Reinstall the capillaries. 8 In Lab Advisor, select Next to move the piston back into its normal position and then Back to finish the mainte- nance procedure.
OR
In the Local Controller, select Continue to move the pis- ton back into its normal position and then Exit to finish the maintenance procedure.
9 Select Back to finish the procedure.
1.
2. 2.
Agilent InfinityLab LC Series Vialsamplers User Manual 256
8 Maintenance Replace the Analytical Heads
10 Close the doors. Next Steps:
11 Perform a System Pressure Test.
12 If needed, install a sample loop matching the volume of the new analytical head (see Exchange the Sample Loop Assembly on page 241).
Agilent InfinityLab LC Series Vialsamplers User Manual 257
8 Maintenance Exchange the Gripper Arm
Exchange the Gripper Arm
When The gripper arm needs to be replaced due to malfunctioning.
Tools required p/n Description 5023-2499 Hex Key Set
(The 1.5 mm hex key is needed)
Parts required p/n Description G1313-60010 Gripper assembly
Preparations Finish any pending acquisition job
NOTE Alternatively, any other tool or item can be used that can be fitted into the block position slot of the transport arm (for example, a straightened paper clip, or a small-sized screwdriver).
1 In the Agilent Lab Advisor software, select Service & Diagnostics> Maintenance Positions> Change Gripper and wait until the gripper is in maintenance position (see also Maintenance Positions on page 179).
OR
In the Local Controller start the maintenance mode and select Change Gripper function.
2 Open the doors of the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 258
8 Maintenance Exchange the Gripper Arm
3 Identify the slot for blocking the gripper arm rotation on the left side of the transport arm (1) and the gripper arm release button (2).
4 Place the hex key (1.5 mm) into the slot and rotate the gripper arm slightly to the left or right until you find the block position.
5 Press in the gripper arm release button and carefully rotate the gripper arm approximately 40 50 to the right, while firmly holding the hex key in place.
You should feel that the lock mechanism releases the gripper arm.
6 Remove the gripper arm and the hex key.
NOTE With firmware version D.07.28 or older, it is recommended to turn off the sampler to complete the rest of the proce- dure.
1
2
1.
2.
Agilent InfinityLab LC Series Vialsamplers User Manual 259
8 Maintenance Exchange the Gripper Arm
7 Fit the new gripper arm into the socket of the transport arm (1.) and rotate it by 360 while firmly holding it against the transport arm (2.) to find the matching posi- tion for the lock mechanism.
8 Place the hex key back into the slot (1.) and rotate the gripper arm slightly to the left or right until you find the block position (2.).
NOTE The gripper arm orientation is slightly different compared to before taking it off.
9 Rotate the gripper arm approximately 40 50 to the left while firmly holding the hex key in place to snap it into the lock position.
10 Remove the hex key from the block position slot and close the instrument doors.
NOTE With firmware version D.07.28 or older, turn on the sampler and wait until it initializes.
11 Complete the remaining steps prompted in the user interface of the Local Controller or the Agilent Lab Advisor soft- ware.
1.
2.
1.
2.
Agilent InfinityLab LC Series Vialsamplers User Manual 260
8 Maintenance Replace the Peristaltic Pump Cartridge
Replace the Peristaltic Pump Cartridge
When The tubing is blocked or damaged The peristaltic cartridge needs to be replaced as part of the yearly maintenance.
Parts required p/n Description 5065-4445 Peristaltic pump with Pharmed tubing
Preparations Finish any pending acquisition job. Remove the solvent line from the wash bottle to avoid spilling solvent.
WARNING When opening capillary or tube fittings solvents may leak out.
The handling of toxic and hazardous solvents and reagents can hold health risks.
Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
NOTE The peristaltic pump cartridge must be replaced as a single entity. The Pharmed tubing is not replaceable separately.
1 Open the doors of the module. 2 Press the two clips on the front of the peristaltic pump cartridge.
Agilent InfinityLab LC Series Vialsamplers User Manual 261
8 Maintenance Replace the Peristaltic Pump Cartridge
3 Pull the cartridge forward off the motor shaft. 4 Disconnect the tubing leading to the wash port and the tubing coming from the solvent bottle.
5 Push the new cartridge onto the motor shaft until the clips click into place.
6 Connect the wash port tubing to the left side of the car- tridge tubing (use sand paper to get a good grip on the tubing).
7 Connect the wash solvent line with the plastic fitting to the right side of the cartridge tubing.
8 Close the doors.
Agilent InfinityLab LC Series Vialsamplers User Manual 262
8 Maintenance Exchange the Wash Port Assembly
Exchange the Wash Port Assembly
When The safety wash port is broken or damaged. Carryover problems are observed while needle wash is used.
Tools required p/n Description 5182-3466 Torx screwdriver T10
OR 5023-3089 Torx key set
Parts required p/n Description G7129-60033 Safety Wash Port for Needle Station
Preparations Finish any pending acquisition job. Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent. Close the shutoff valves at the pump if available.
1 In the Agilent Lab Advisor software select Service & Diagnostics> Maintenance Positions> Change Needle, Loop and Seat, click Start and wait until the needle assembly is in maintenance position.
OR
In the Local Controller start the maintenance mode and select Change Needle, Loop and Seat function.
2 Open the doors of the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 263
8 Maintenance Exchange the Wash Port Assembly
3 Verify that the needle in the uppermost position before continuing.
4 Loosen the safety cover screw.
WARNING Sharp needle Uncovered needles may cause injuries
Do not touch the tip of the needle.
5 Remove the safety cover.
WARNING Sharp needle Uncovered needles may cause injuries.
Do not change the needle seat at this point.
Do not touch the tip of the needle.
6 Move the wash port into the service position by slightly lifting it up and then flapping to the left.
Agilent InfinityLab LC Series Vialsamplers User Manual 264
8 Maintenance Exchange the Wash Port Assembly
7 In Lab Advisor use Next to move the needle in the down position (ca. 2 mm above the seat).
OR
In the Local Controller, move the needle down until the needle tip is ca. 2 4 mm above the seat.
8 Disconnect the sample loop from the needle, and, if needed, also from the metering device.
9 Loosen the fixing screw. 10 Lift out the needle.
11 Release the spring loaded pin. 12 Flap the wash port to the right and lift it out of the bracket. Be careful not to squeeze the wash tubing.
NOTE Move the needle holder arm up by a couple of mm if you are facing difficulties with moving the wash port to the right.
Agilent InfinityLab LC Series Vialsamplers User Manual 265
8 Maintenance Exchange the Wash Port Assembly
13 Disconnect the wash tube fitting and then move the old wash port completely out of the sampler.
14 Connect the wash tubing fitting to the new wash port assembly.
15 Install the new wash port by mounting it in the respec- tive brackets of the needle station housing. The pin must be latched on the housing. Check the movement of the flapping mechanism. Check if the tension on the spring is high enough to move the wash port back into position in time. When everything seems to be in order, move the wash port into the service position (see step 6).
16 Mount and align the needle on its fixture, then tighten the screw firmly.
Agilent InfinityLab LC Series Vialsamplers User Manual 266
8 Maintenance Exchange the Wash Port Assembly
17 Align the needle tip with the seat and ensure that they are concentric. If needed, carefully bend the needle into the right position with your fingers.
18 Reconnect the sample loop to the needle. Make sure of the correct positioning of the loop capillary, the uncoated part of the capillary must be horizontal.
NOTE Do not overtighten the fitting!
NOTE Incorrect positioning and installation of the needle/loop connection can result in damaging or breaking the sample loop.
19 In Lab Advisor use Next to lift the needle slowly into the up position.
OR
In the Local Controller, move the needle up to the upper- most position.
20 Move the wash port back into its normal position by turning it to the right.
Agilent InfinityLab LC Series Vialsamplers User Manual 267
8 Maintenance Exchange the Wash Port Assembly
21 Install the safety cover (1.) and fix the screw (2.). 22 In Lab Advisor, select Next to move the needle into the needle seat and then Back to finish the maintenance procedure.
OR
In the Local Controller, select Continue to move the nee- dle into the needle seat and then Exit to finish the main- tenance procedure.
23 Select Back to finish the procedure.
24 Close the doors. Next Steps:
25 Purge the wash port. Check if solvent is delivered from the peristaltic pump.
26 Perform a System Pressure Test.
(1.) (2.)
Agilent InfinityLab LC Series Vialsamplers User Manual 268
8 Maintenance Replace the Module Firmware
Replace the Module Firmware
To upgrade/downgrade the modules firmware carry out the following steps: 1 Download the required module firmware, the latest FW Update Tool and the
documentation from the Agilent web. http://www.agilent.com/en-us/firmwareDownload?whid=69761
2 For loading the firmware into the module follow the instructions in the documentation.
Module Specific Information
There is no specific information for this module.
When The installation of newer firmware might be necessary if a newer version solves problems of older versions or to keep all systems on the same (validated) revision. The installation of older firmware might be necessary to keep all systems on the same (validated) revision or if a new module with newer firmware is added to a system or if third party control software requires a special version.
Tools required Description Agilent Lab Advisor software
Parts required # Description 1 Firmware, tools and documentation from Agilent web site
Preparations Read update documentation provided with the Firmware Update Tool.
Agilent InfinityLab LC Series Vialsamplers User Manual 269
8 Maintenance Replace the Sample Cooler/Sample Thermostat
Replace the Sample Cooler/Sample Thermostat
For the Sample Cooler installation in a Vialsampler the serial number of the Sample Cooler must be DEBAT02001 or higher.
When If the Sample Cooler/Sample Thermostat is damaged or defective.
Tools required p/n Description 8710-0899 Screwdriver Pozidrive Shaft
(for the Sample Cooler) 5182-3466 Torx screwdriver T10
(for the Sample Thermostat) OR 5023-3089 Torx key set
Parts required # p/n Description 1 G7167-60005 Sample Cooler
OR 1 G7167-60101 Sample Thermostat
WARNING Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample Thermostat and laboratory.
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m3 (1 m3 for every 8 g of R600a refrigerant inside of the Sample Thermostat).
Ensure adequate ventilation: typical air exchange of 25 m3/h per m2 of laboratory floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do not block the openings on the circumference of the Sample Thermostat.
WARNING Flammable refrigerant used
When handling, installing and operating the Sample Thermostat, care should be taken to avoid damage to the refrigerant tubing or any part of the Sample Thermostat.
Agilent InfinityLab LC Series Vialsamplers User Manual 270
8 Maintenance Replace the Sample Cooler/Sample Thermostat
WARNING In the event of a damage
Keep open fire or sources of ignition away from the device.
Ventilate the room for several minutes.
Do not use the Sample Thermostat any more.
WARNING Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
CAUTION Routing of the condensation tubing
Proper routing of the condensation tubing is critical for correct condensate drainage.
Do not place the sampler directly on the bench.
CAUTION Condensate inside the cooler or thermostat
Damage to the electronics
Unplug the power cords.
Drain off all condensate before dismounting the sample cooler or thermostat.
Make sure that there is no condensate left.
Agilent InfinityLab LC Series Vialsamplers User Manual 271
8 Maintenance Replace the Sample Cooler/Sample Thermostat
1 Ensure that the power switch on the front of the module is OFF (switch stands out).
2 Disconnect the power cable from the sampler.
3 Ensure that no condensate remains inside the cooler/thermostat before proceeding forward.
NOTE Gently tapping on the sides of the sampler can help to remove the last traces of condensate from the system.
4 Remove the condensate tubing.
NOTE If there is still some condensate inside the cooler/thermo- stat, place a suitable container underneath the outlet pipe and keep tapping on the sides of the sampler until no water is coming out.
5 Remove the fixation screws on the back of Sample Cooler/Sample Thermostat.
6 Pull the cooler/thermostat halfway out, disconnect the power and the data cable and then remove the unit completely from the sampler.
Agilent InfinityLab LC Series Vialsamplers User Manual 272
8 Maintenance Replace the Sample Cooler/Sample Thermostat
7 Slide the new cooler/thermostat halfway into the sam- pler and connect the power and the data cable. CAUTION
Damage to the cables
Do not bend or pinch the cables.
Make sure that the Sample Cooler/Sample Thermostat fits perfectly in the sampler.
8 Slide the cooler/thermostat all the way into the sampler, making sure that the cables dont get jammed between the metal parts.
9 Fix the unit with the four screws. 10 Reconnect the condensate tubing.
NOTE For information on proper condensate handling, see Install the Sample Cooler/Sample Thermostat on page 117.
Agilent InfinityLab LC Series Vialsamplers User Manual 273
8 Maintenance Replace the Sample Cooler/Sample Thermostat
11 Connect the power cable to the power connector at the rear of the module. CAUTION
Damage to the Sample Cooler/Sample Thermostat
Wait at least 30 min before switching on the compres- sor of the cooler/thermostat.
This allows the refrigerant and system lubrication to reach equilibrium.
12 Switch on the sampler and perform the Sample Cooler Function Test to verify the correct functioning of the new cooler/thermostat (see Sample Cooler Function Test on page 187).
Agilent InfinityLab LC Series Vialsamplers User Manual 274
9 Parts and Materials for Maintenance
Standard Parts For Maintenance 276 PM Kits 276 Needle Assemblies 276 Needle Seats 276 Sample Loops 277 Rotor Seals for Injection Valves 277 Metering Seals for Analytical Heads 277 Capillary Connections 278 Other Parts 278
Accessory Kits 279 Accessory Kit for 1290 Infinity II Vialsampler 279 Accessory Kit for 1260 Infinity II Vialsampler 281
Vial Drawers and Trays 283 Cartesian Vial Drawers 283 Classic Vial Drawers 284 External Tray 285
Multidraw Kits 286 Multidraw Kit 286 Large Volume Injection Kit 286
Sample Thermostat Upgrade 287 Integrated Column Compartment 288 Column ID Upgrade Kit 289 Parts for 900 L Injection Upgrade 290 Cabinet and Door Kits 291 Cabinet Kit 291 Door Kit 293 Door Insulation Kit 294
Analytical Head Assembly (100 L) 295 Analytical-Head Assembly (900 L) 296 Analytical Head Assembly (40 L) 297 2 Position/6 Port Injection Valve, 600 bar 298 2 Position/6 Port Injection Valve, 800 bar 299 2 Position/6 Port Injection Valve, 1300 bar 300
This chapter provides information on parts for maintenance.
Agilent InfinityLab LC Series Vialsamplers User Manual 275
9 Parts and Materials for Maintenance Standard Parts For Maintenance
Standard Parts For Maintenance
PM Kits
Needle Assemblies
Needle Seats
p/n Description
G7129-68740 PM Kit 1260 Inf. II Vialsampler includes needle assembly, needle seat assembly, peristaltic pump cartridge, rotor seal and finger caps.
G7129-68730 PM Kit 1290 Inf. II Vialsampler includes needle assembly, needle seat assembly, peristaltic pump cartridge, rotor seal and finger caps.
p/n Description
G7129-87200 Needle assembly, 1260 Vialsampler for G7129-87017 needle seat
G7129-87201 Needle assembly, 1290 Vialsampler for G7129-87012 needle seat
G7129-87202 Needle assembly, slotted, for high injection volumes
p/n Description
G7129-87017 Seat Assembly, 1260 Vialsampler, PEEK, 0.17 mm
G7129-87012 Seat Assembly, 1290 Vialsampler, PEEK, 0.12 mm
G7129-87117 Seat Assembly, 1260 Vialsampler, Vespel, 0.17 mm
G7129-87112 Seat Assembly, 1290 Vialsampler, Vespel, 0.12 mm
NOTE The use of a Vespel needle seat (G7129-87112 and G7129-87117) limits the maximum system pressure to 600 bar.
Agilent InfinityLab LC Series Vialsamplers User Manual 276
9 Parts and Materials for Maintenance Standard Parts For Maintenance
Sample Loops
Rotor Seals for Injection Valves
Metering Seals for Analytical Heads
p/n Description
G7129-60500 Sample Loop 100 L (white coded)
G7129-60400 Sample Loop 40 L (green coded)
G7129-60300 Sample Loop 20 L (red coded)
p/n Description
0101-1416 Rotor Seal, PEEK, 600/800 bar
5068-0007 Rotor Seal, Vespel, 1300 bar
p/n Description
0905-1503 Metering Seal, 100 L
0905-1717 Metering Seal, 40 L
0905-1294 Metering Seal, 900 L
Agilent InfinityLab LC Series Vialsamplers User Manual 277
9 Parts and Materials for Maintenance Standard Parts For Maintenance
Capillary Connections
Other Parts
p/n Description
5500-1246 Capillary ST 0.17 mm x 500 mm SI/SI (pump to Vialsampler without ICC)
5500-1217 Capillary, ST, 0.17 mm x 900 mm SI/SX (pump to Vialsampler with ICC)
5500-1252 Capillary, ST, 0.17 mm x 400 mm SL/SL (1260 Vialsampler to MCT)
5500-1240 Capillary ST 0.17 mm x 105 mm SL/SL (ICC, 6 L to column)
5500-1251 Capillary ST 0.12 mmX 400 mm SL/SL (1290 Vialsampler to MCT)
5500-1238 Capillary ST 0.12 mm x 105 mm SL/SL (ICC, 3 L to column)
5500-1250 Capillary, ST, 0.17 mm x 120 mm SL/SL, long socket (1260 Vialsampler to ICC, 6 L)
5500-1249 Capillary ST 0.12 mm x 120 mm SL/SL, long socket (1290 Vialsampler to ICC, 3 L)
5067-5966 InfinityLab Quick Turn Fitting
5067-6166 InfinityLab Quick Connect Assy ST 0.17 mm x 105 mm (alternative, ICC, 6 L to column)
5067-5957 InfinityLab Quick Connect Assy ST 0.12 mm x 105 mm (alternative, ICC, 3 L to column)
5500-1228 Capillary ST 0.3 mm x 80 mm SL-SL (metering head to injection valve)
p/n Description
G1313-60010 Gripper assembly
5063-6506 Finger Caps
5065-4445 Peristaltic pump with Pharmed tubing
Agilent InfinityLab LC Series Vialsamplers User Manual 278
9 Parts and Materials for Maintenance Accessory Kits
Accessory Kits
Accessory Kit for 1290 Infinity II Vialsampler Accessory Kit for 1290 Infinity II Vialsampler (G7129-68705) contains the following parts:
Item # p/n Description
1 1 5500-1251 Capillary ST 0.12 mmX 400 mm SL/SL
2 3 5063-6506 Finger Caps (15/pk, if ordered separately)
3 2 5043-1013 Tubing Clip
4 1 5500-1411 Tubing Connector, 180 , ID 6.4 mm (5/pk, if ordered separately)
5 2 5500-1223 Tubing Connector, 90, ID 6.4 mm (5/pk, if ordered separately)
6 1 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm
7 1 5181-1519 CAN cable, Agilent module to module, 1 m
1 5182-0716 Screw Cap Vial, 2 mL, amber glass, write-on spot, 100/Pack (not shown)
1 5190-7024 Screw Cap, PTFE/silicone, 100/pk (not shown)
Agilent InfinityLab LC Series Vialsamplers User Manual 279
9 Parts and Materials for Maintenance Accessory Kits
Figure 34 Accessory Kit for Vialsamplers
1
2
3
4
5
6 7
Agilent InfinityLab LC Series Vialsamplers User Manual 280
9 Parts and Materials for Maintenance Accessory Kits
Accessory Kit for 1260 Infinity II Vialsampler Accessory Kit for 1260 Infinity II Vialsampler (G7129-68715) contains the following parts:
Item # p/n Description
1 1 5500-1252 Capillary, ST, 0.17 mm x 400 mm SL/SL
2 3 5063-6506 Finger Caps (15/pk, if ordered separately)
3 2 5043-1013 Tubing Clip
4 1 5500-1411 Tubing Connector, 180 , ID 6.4 mm (5/pk, if ordered separately)
5 2 5500-1223 Tubing Connector, 90, ID 6.4 mm (5/pk, if ordered separately)
6 1 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm
7 1 5181-1519 CAN cable, Agilent module to module, 1 m
1 5182-0716 Screw Cap Vial, 2 mL, amber glass, write-on spot, 100/Pack (not shown)
1 5190-7024 Screw Cap, PTFE/silicone, 100/pk (not shown)
Agilent InfinityLab LC Series Vialsamplers User Manual 281
9 Parts and Materials for Maintenance Accessory Kits
Figure 35 Accessory Kit for Vialsamplers
1
2
3
4
5
6 7
Agilent InfinityLab LC Series Vialsamplers User Manual 282
9 Parts and Materials for Maintenance Vial Drawers and Trays
Vial Drawers and Trays
Cartesian Vial Drawers
Figure 36 Cartesian Vial Drawers
Item p/n Description
1 G7129-60010 Drawer for 66 x 2 mL Vials
2 G7129-60110 Drawer for 18 x 6 mL Vials
1
2
NOTE Do not use cartesian vial drawers in combination with classic vial drawers.
Agilent InfinityLab LC Series Vialsamplers User Manual 283
9 Parts and Materials for Maintenance Vial Drawers and Trays
Classic Vial Drawers Classic Vial Drawer Kit (G7129-68210) contains:
Figure 37 Classic Vial Drawer Kit
p/n Description
G7129-60210 Classic Drawer for 50 x 2 mL Vials, Left
G7129-60220 Classic Drawer for 50 x 2 mL Vials, Right
1
2
NOTE Do not use classic vial drawers in combination with cartesian vial drawers.
NOTE Make sure of the correct position of the left (vials 1 50) and the right (vials 51 100) drawer of the Classic Vial Drawer Kit.
Agilent InfinityLab LC Series Vialsamplers User Manual 284
9 Parts and Materials for Maintenance Vial Drawers and Trays
External Tray
Figure 38 External tray
p/n Description
G7129-60000 External Tray for 5 x 2 mL Vials
G1313-27302 Disposal tube (not shown)
Agilent InfinityLab LC Series Vialsamplers User Manual 285
9 Parts and Materials for Maintenance Multidraw Kits
Multidraw Kits
Multidraw Kit Multidraw Kit for InfinityLab Sampler (G7167-68711) contains the following items:
Large Volume Injection Kit Large Volume Injection Kit for 1290 Infinity II Samplers (G4216-68711) contains:
p/n Description
G1313-87307 Seat Capillary, ST, 500 L, ID 0.5 mm
G1313-87308 Seat Capillary, ST, 1500 L, ID 0.94 mm
0100-0900 Zero Dead Volume Union, ST
NOTE With the Multidraw Kit, the maximum injection volume can be extended to 500 or 1500 L for the 100 L analytical head and to 1800 L for the 900 L analytical head. The nominal volume of the seat capillaries is 656 and 2415 L, respectively.
p/n Description
Extension Seat Capillary, 80 L, 0.5 mm ID (0.9 mm OD)
NOTE With the Large Volume Injection Kit, the maximum injection volume can be extended to 100 or 120 L, depending on the sample loop installed in the system. The nominal volume of the seat capillary is 157 L.
Agilent InfinityLab LC Series Vialsamplers User Manual 286
9 Parts and Materials for Maintenance Sample Thermostat Upgrade
Sample Thermostat Upgrade
InfinityLab Sample Thermostat Upgrade Kit (G4761A) contains:
Figure 39 Sample Thermostat
p/n Description
G7167-60101 Sample Thermostat
5067-6208 Condensate Drainage Kit (not shown)
NOTE The Sample Thermostat contains flammable refrigerant R600a. Please check further details for installation.
Agilent InfinityLab LC Series Vialsamplers User Manual 287
9 Parts and Materials for Maintenance Integrated Column Compartment
Integrated Column Compartment
Figure 40 Integrated Column Compartment
p/n Description
G7130-60030 Integrated Column Compartment, 3 L Heater
G7130-60060 Integrated Column Compartment, 6 L Heater
G7116-68003 Column Holder Lamella, 2/pk (not shown)
NOTE The Integrated Column Compartment is always shipped with an accessory kit, including a pair of column-holder lamellas and two stainless steel capillaries (ID 0.12 mm or 0.17 mm) with a length of 105 mm and 120 mm, respectively (see Capillary Connections on page 278).
Agilent InfinityLab LC Series Vialsamplers User Manual 288
9 Parts and Materials for Maintenance Column ID Upgrade Kit
Column ID Upgrade Kit
Column ID Upgrade Kit for the Integrated Column Compartment (G4751A) contains:
Figure 41 Column Identifier Kit
p/n Description
5067-5964 Column ID Tag Reader
5067-5917 InfinityLab Column Identification Tag
NOTE The InfinityLab Column Identification Tags used in G7116A/B and G7129A/B/C are not compatible with the former Column Identification Tags used in G1316x.
NOTE The Column Identifier Kit can only be used in combination with the Integrated Column Compartment (ICC).
Agilent InfinityLab LC Series Vialsamplers User Manual 289
9 Parts and Materials for Maintenance Parts for 900 L Injection Upgrade
Parts for 900 L Injection Upgrade
Item p/n Description
1 G7129-60083 Analytical Head Assembly 900 L
G7129-87202 Needle assembly, slotted, for high injection volumes
3 G1313-87303 Capillary ST 0.17 mm x 400 mm S/S (Loop extension capillary (900 L))
5022-2133 High Flow Union, ST
1
2
3
4
NOTE With the 900 L analytical head, the maximum system pressure is limited to 400 bar.
Agilent InfinityLab LC Series Vialsamplers User Manual 290
9 Parts and Materials for Maintenance Cabinet and Door Kits
Cabinet and Door Kits
Cabinet Kit Cabinet Kit for Infinity II Vialsampler/Preparative Autosampler (G7129-68713) contains:
Related parts:
Item p/n Description
1 Top cover
2 5043-1804 Name Plate, Infinity II 1290
2 5043-1805 Name Plate, Infinity II 1260
3 Side Cover left
4 Side Cover right
5 5043-0286 Base Cover Infinity II
NOTE The top cover and the two side covers are not separately operable.
Item p/n Description
6 G7129-60023 Column Shelf
7 5043-1013 Tubing Clip
8 5043-0874 Tubing Holder Insert (included in Side Cover, Right)
9 G1313-22406 Fixation Screw (included in Side Cover, Right)
Agilent InfinityLab LC Series Vialsamplers User Manual 291
9 Parts and Materials for Maintenance Cabinet and Door Kits
Figure 42 Cabinet Kit for Infinity II Vialsampler
3
6
5
1
2
4
7
8
9
Agilent InfinityLab LC Series Vialsamplers User Manual 292
9 Parts and Materials for Maintenance Cabinet and Door Kits
Door Kit Door Kit for Infinity II Vialsampler/Preparative Autosampler (G7129-68702) contains:
Figure 43 Door Kit for Infinity II Vialsampler
Item p/n Description
1 Door panel, left
2 Door panel, right
3 Side window
4 5067-5412 Hinge for Infinity II Instrument Doors (preinstalled in the doors)
1
2
3
4
Agilent InfinityLab LC Series Vialsamplers User Manual 293
9 Parts and Materials for Maintenance Cabinet and Door Kits
Door Insulation Kit Door Insulation Kit (G7129-67003) contains:
Figure 44 Door Insulation Kit
Item p/n Description
1 Foam for Left Panel
2 Foam for Right Panel
3 Foam for Side Window
1
2
3
Agilent InfinityLab LC Series Vialsamplers User Manual 294
9 Parts and Materials for Maintenance Analytical Head Assembly (100 L)
Analytical Head Assembly (100 L)
Figure 45 Analytical head assembly, 100 L
Item p/n Description
G7129-60082 Analytical Head Assembly 100 L
1 G7129-27710 Analytical-Head 100 L
2 0905-1503 Metering Seal, 100 L
3 G7129-60182 Seal Support, 100 L
4 0515-1052 Screw, ST, M3x0.5, Hex 2.5 mm
5 G7129-60006 Analytical Head Adapter
6 0515-0850 Screw, ST, M4x0.7, Hex 4 mm
7 5067-5678 Piston, 100 L, Zirconia
1 2
3 4
5
6 7
NOTE With the 100 L analytical head, the maximum system pressure is limited to 800 bar.
Agilent InfinityLab LC Series Vialsamplers User Manual 295
9 Parts and Materials for Maintenance Analytical-Head Assembly (900 L)
Analytical-Head Assembly (900 L)
Figure 46 Analytical head assembly (900 L)
Item p/n Description
G7129-60083 Analytical Head Assembly 900 L
1 G7129-27790 Analytical-Head 900 L
2 0905-1294 Metering Seal, 900 L
3 5001-3764 Seal Support, 900 L
4 G7129-60006 Analytical Head Adapter
5 0515-0850 Screw, ST, M4x0.7, Hex 4 mm
6 G4267-60462 Piston, 900 L, Sapphire
1
2
3
4
5 6
NOTE With the 900 L analytical head, the maximum system pressure is limited to 400 bar.
Agilent InfinityLab LC Series Vialsamplers User Manual 296
9 Parts and Materials for Maintenance Analytical Head Assembly (40 L)
Analytical Head Assembly (40 L)
Figure 47 Analytical head assembly (40 L)
Item p/n Description
G7129-60084 Analytical Head Assembly 40 L
1 G7129-27704 Analytical-Head 40 L
2 0905-1717 Metering Seal, 40 L
3 G7129-60184 Seal Support, 40 L
4 0515-1052 Screw, ST, M3x0.5, Hex 2.5 mm
5 G7129-60006 Analytical Head Adapter
6 0515-0850 Screw, ST, M4x0.7, Hex 4 mm
7 5067-5920 Piston, 40 L, Zirconia
1
2 3
4
5
6 7
Agilent InfinityLab LC Series Vialsamplers User Manual 297
9 Parts and Materials for Maintenance 2 Position/6 Port Injection Valve, 600 bar
2 Position/6 Port Injection Valve, 600 bar
Figure 48 2 Position/6 Port Injection Valve 600 bar/800 bar
Item p/n Description
5067-4245 2 Position/6 Port Injection Valve, 600 bar
1 5068-0018 Screws, ST, 8-32, Hex 9/64, 10/pk
2 5068-0215 Stator, Injection Valve, 600 bar/800 bar
3 5068-0118 Stator ring
4 0101-1416 Rotor Seal, PEEK, 600/800 bar
5 1535-4045 Bearing ring
1
2
3 4 5
Agilent InfinityLab LC Series Vialsamplers User Manual 298
9 Parts and Materials for Maintenance 2 Position/6 Port Injection Valve, 800 bar
2 Position/6 Port Injection Valve, 800 bar
Figure 49 2 Position/6 Port Injection Valve 600 bar/800 bar
Item p/n Description
5067-6689 2 Position/6 Port Injection Valve, 800 bar
1 5068-0018 Screws, ST, 8-32, Hex 9/64, 10/pk
2 5068-0215 Stator, Injection Valve, 600 bar/800 bar
3 5068-0118 Stator ring
4 0101-1416 Rotor Seal, PEEK, 600/800 bar
5 1535-4045 Bearing ring
1
2
3 4 5
Agilent InfinityLab LC Series Vialsamplers User Manual 299
9 Parts and Materials for Maintenance 2 Position/6 Port Injection Valve, 1300 bar
2 Position/6 Port Injection Valve, 1300 bar
Figure 50 2 Position/6 Port Injection Valve 1300 bar
Item p/n Description
5067-4238 2 Position/6 Port Injection Valve, 1300 bar
1 5068-0018 Screws, ST, 8-32, Hex 9/64, 10/pk
2 5068-0216 Stator, Injection Valve, 1300 bar
3 5068-0118 Stator ring
4 5068-0007 Rotor Seal, Vespel, 1300 bar
5 1535-4045 Bearing ring
1
2
3 4
5
Agilent InfinityLab LC Series Vialsamplers User Manual 300
10 Identifying Cables
Cable Overview 302 Analog Cables 304 Remote Cables 306 CAN/LAN Cables 310 RS-232 Cables 311 USB 312
This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.
Agilent InfinityLab LC Series Vialsamplers User Manual 301
10 Identifying Cables Cable Overview
Cable Overview
Analog cables
Remote cables
CAN cables
NOTE Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
p/n Description
35900-60750 Agilent 35900A A/D converter
01046-60105 Analog cable (BNC to general purpose, spade lugs)
p/n Description
5188-8029 ERI to general purpose
5188-8044 Remote Cable ERI ERI
5188-8045 Remote Cable APG ERI
5188-8059 ERI-Extension-Cable 1.2 m
5061-3378 Remote Cable to 35900 A/D converter
01046-60201 Agilent module to general purpose
5188-8057 Fraction Collection ERI remote Y-cable
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
5181-1519 CAN cable, Agilent module to module, 1 m
Agilent InfinityLab LC Series Vialsamplers User Manual 302
10 Identifying Cables Cable Overview
LAN cables
RS-232 cables (not for
FUSION board)
USB cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connec- tion)
p/n Description
RS232-61601 RS-232 cable, 2.5 m Instrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It is also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
5181-1561 RS-232 cable, 8 m
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
5188-8049 USB A F-USB Mini B M OTG (Module to Flash Drive)
Agilent InfinityLab LC Series Vialsamplers User Manual 303
10 Identifying Cables Analog Cables
Analog Cables
One end of these cables provides a BNC connector to be connected to Agilent modules. The other end depends on the instrument to which connection is being made.
Agilent Module to 35900 A/D converters
p/n 35900-60750 35900 Pin Agilent module
Signal Name
1 Not connected
2 Shield Analog -
3 Center Analog +
Agilent InfinityLab LC Series Vialsamplers User Manual 304
10 Identifying Cables Analog Cables
Agilent Module to BNC Connector
Agilent Module to General Purpose
p/n 8120-1840 Pin BNC Pin Agilent module
Signal Name
Shield Shield Analog -
Center Center Analog +
p/n 01046-60105 Pin Pin Agilent module
Signal Name
1 Not connected
2 Black Analog -
3 Red Analog +
Agilent InfinityLab LC Series Vialsamplers User Manual 305
10 Identifying Cables Remote Cables
Remote Cables
ERI (Enhanced Remote Interface) 5188-8029 ERI to general purpose (D-Sub 15 pin male - open end) 5188-8044 ERI to ERI (D_Sub 15 pin male - male) 5188-8059 ERI-Extension-Cable 1.2 m (D-Sub15 pin male / female)
p/n 5188-8029 pin Color code Enhanced Remote
Classic Remote
Active (TTL)
1 white IO1 START REQUEST
Low
2 brown IO2 STOP Low
3 green IO3 READY High
4 yellow IO4 POWER ON High
5 grey IO5 NOT USED
6 pink IO6 SHUT DOWN Low
7 blue IO7 START Low
8 red IO8 PREPARE Low
9 black 1wire DATA
10 violet DGND
11 grey-pink +5V ERI out
12 red-blue PGND
13 white-green PGND
14 brown-green +24V ERI out
15 white-yellow +24V ERI out
NC yellow-brown
Agilent InfinityLab LC Series Vialsamplers User Manual 306
10 Identifying Cables Remote Cables
5188-8045 ERI to APG (Connector D_Subminiature 15 pin (ERI), Connector D_Subminiature 9 pin (APG))
p/n 5188-8045 Pin (ERI) Signal Pin (APG) Active (TTL)
10 GND 1
1 Start Request 9 Low
2 Stop 8 Low
3 Ready 7 High
5 Power on 6 High
4 Future 5
6 Shut Down 4 Low
7 Start 3 Low
8 Prepare 2 Low
Ground Cable Shielding NC
Agilent InfinityLab LC Series Vialsamplers User Manual 307
10 Identifying Cables Remote Cables
5188-8057 ERI to APG and RJ45 (Connector D_Subminiature 15 pin (ERI), Connector D_Subminiature 9 pin (APG), Connector plug Cat5e (RJ45))
One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent modules. The other end depends on the instrument to be connected to.
Table 22 5188-8057 ERI to APG and RJ45
p/n 5188-8057 Pin (ERI) Signal Pin (APG) Active (TTL) Pin (RJ45)
10 GND 1 5
1 Start Request
9 High
2 Stop 8 High
3 Ready 7 High
4 Fraction Trig- ger
5 High 4
5 Power on 6 High
6 Shut Down 4 High
7 Start 3 High
8 Prepare 2 High
Ground Cable Shield- ing
NC
Agilent InfinityLab LC Series Vialsamplers User Manual 308
10 Identifying Cables Remote Cables
Agilent Module to Agilent 35900 A/D Converters
Agilent Module to General Purpose
p/n 5061-3378 Pin 35900 A/D Pin Agilent module
Signal Name Active (TTL)
1 - White 1 - White Digital ground
2 - Brown 2 - Brown Prepare run Low
3 - Gray 3 - Gray Start Low
4 - Blue 4 - Blue Shut down Low
5 - Pink 5 - Pink Not connected
6 - Yellow 6 - Yellow Power on High
7 - Red 7 - Red Ready High
8 - Green 8 - Green Stop Low
9 - Black 9 - Black Start request Low
p/n 01046-60201 Wire Color Pin Agilent module
Signal Name Active (TTL)
White 1 Digital ground
Brown 2 Prepare run Low
Gray 3 Start Low
Blue 4 Shut down Low
Pink 5 Not connected
Yellow 6 Power on High
Red 7 Ready High
Green 8 Stop Low
Black 9 Start request Low
Agilent InfinityLab LC Series Vialsamplers User Manual 309
10 Identifying Cables CAN/LAN Cables
CAN/LAN Cables
Both ends of this cable provide a modular plug to be connected to Agilent modules CAN or LAN connectors.
CAN Cables
LAN Cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
5181-1519 CAN cable, Agilent module to module, 1 m
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connec- tion)
Agilent InfinityLab LC Series Vialsamplers User Manual 310
10 Identifying Cables RS-232 Cables
RS-232 Cables
p/n Description
RS232-61601 RS-232 cable, 2.5 m Instrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It is also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
5181-1561 RS-232 cable, 8 m
Agilent InfinityLab LC Series Vialsamplers User Manual 311
10 Identifying Cables USB
USB
To connect a USB Flash Drive use a USB OTG cable with Mini-B plug and A socket.
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
5188-8049 USB A F-USB Mini B M OTG (Module to Flash Drive)
Agilent InfinityLab LC Series Vialsamplers User Manual 312
11 Hardware Information
Firmware Description 314 Electrical Connections 317 Rear View of the Module 318 Serial Number Information 318
Interfaces 319 Overview Interfaces 321 ERI (Enhanced Remote Interface) 323 USB (Universal Serial Bus) 325
Setting the 6-bit Configuration Switch 326 Special Settings 328
Instrument Layout 330 Early Maintenance Feedback (EMF) 331
This chapter describes the detector in more detail on hardware and electronics.
Agilent InfinityLab LC Series Vialsamplers User Manual 313
11 Hardware Information Firmware Description
Firmware Description
The firmware of the instrument consists of two independent sections: a non-instrument specific section, called resident system an instrument specific section, called main system
Resident System
This resident section of the firmware is identical for all Agilent 1100/1200/1220/1260/1290 series modules. Its properties are: the complete communication capabilities (CAN, LAN, USB and RS- 232) memory management ability to update the firmware of the 'main system'
Main System
Its properties are: the complete communication capabilities (CAN, LAN, USB and RS- 232) memory management ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are divided into common functions like run synchronization through APG/ERI remote, error handling, diagnostic functions, or module specific functions like
internal events such as lamp control, filter movements, raw data collection and conversion to absorbance.
Agilent InfinityLab LC Series Vialsamplers User Manual 314
11 Hardware Information Firmware Description
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web: http://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are: PPPP_RVVV_XXX.dlb, where PPPP is the product number, for example, 1315B for the G1315B DAD, R the firmware revision, for example, A for G1315B or B for the G1315C DAD, VVV is the revision number, for example 650 is revision 6.50, XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in chapter "Maintenance" or use the documentation provided with the Firmware Update Tools.
Figure 51 Firmware Update Mechanism
NOTE Update of main system can be done in the resident system only. Update of the resident system can be done in the main system only.
Main and resident firmware must be from the same set.
Resident System Main FW update
Main System
Resident FW Update
Agilent InfinityLab LC Series Vialsamplers User Manual 315
11 Hardware Information Firmware Description
The firmware update tools, firmware and documentation are available from the Agilent web. http://www.agilent.com/en-us/firmwareDownload?whid=69761
NOTE Some modules are limited in downgrading due to their mainboard version or their initial firmware revision. For example, a G1315C DAD SL cannot be downgraded below firmware revision B.01.02 or to a A.xx.xx.
Some modules can be re-branded (e.g. G1314C to G1314B) to allow operation in specific control software environments. In this case, the feature set of the target type is used and the feature set of the original one is lost. After re-branding (e.g. from G1314B to G1314C), the original feature set is available again.
All this specific information is described in the documentation provided with the firmware update tools.
Agilent InfinityLab LC Series Vialsamplers User Manual 316
11 Hardware Information Electrical Connections
Electrical Connections
The CAN bus is a serial bus with high-speed data transfer. The two connectors for the CAN bus are used for internal module data transfer and synchronization.
The ERI/REMOTE connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features such as start, stop, common shutdown, prepare, and so on.
With the appropriate software, the LAN connector may be used to control the module from a computer through a LAN connection. This connector is activated and can be configured with the configuration switch.
With the appropriate software, the USB connector may be used to control the module from a computer through a USB connection.
The power input socket accepts a line voltage of 100 240 VAC 10 % with a line frequency of 50 or 60 Hz. Maximum power consumption varies by module. There is no voltage selector on your module because the power supply has wide-ranging capability. There are no externally accessible fuses because automatic electronic fuses are implemented in the power supply.
NOTE Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
Agilent InfinityLab LC Series Vialsamplers User Manual 317
11 Hardware Information Electrical Connections
Rear View of the Module
Figure 52 Rear view of Vialsampler - electrical connections and label
Serial Number Information The serial number information on the instrument labels provide the following information:
Configuration switch
Power socket
CAN ERI LAN
USB-Mini-Port
CCXZZ00000 Format
CC Country of manufacturing DE = Germany JP = Japan CN = China
X Alphabetic character A-Z (used by manufacturing)
ZZ Alpha-numeric code 0-9, A-Z, where each combination unambig- uously denotes a module (there can be more than one code for the same module)
00000 Serial number
Agilent InfinityLab LC Series Vialsamplers User Manual 318
11 Hardware Information Interfaces
Interfaces
The Agilent InfinityLab LC Series modules provide the following interfaces:
Table 23 Agilent InfinityLab LC Series Interfaces
Module CAN USB LAN (on-board)
RS-232 Analog APG (A) / ERI (E)
Special
Pumps
G7104A/C 2 No Yes Yes 1 A
G7110B 2 Yes Yes No No E
G7111A/B, G5654A 2 Yes Yes No No E
G7112B 2 Yes Yes No No E
G7120A 2 No Yes Yes 1 A
G7161A/B 2 Yes Yes No No E
Samplers
G7129A/B/C 2 Yes Yes No No E
G7167B/C, G5667A 2 Yes Yes No No E
G7157A 2 Yes Yes No No E
Detectors
G7114A/B 2 Yes Yes No 1 E
G7115A 2 Yes Yes No 1 E
G7117A/B/C 2 Yes Yes No 1 E
G7121A/B 2 Yes Yes No 1 E
G7162A/B 2 Yes Yes No 1 E
G7165A 2 Yes Yes No 1 E
Agilent InfinityLab LC Series Vialsamplers User Manual 319
11 Hardware Information Interfaces
CAN connectors as interface to other modules LAN connector as interface to the control software RS-232C as interface to a computer USB (Universal Series Bus) as interface to a computer REMOTE connector as interface to other Agilent products Analog output connector(s) for signal output
Fraction Collectors
G7158B 2 Yes Yes No No E
G7159B 2 Yes Yes No No E
G7166A 2 No No No No No Requires a host module with on-board LAN with minimum FW B.06.40 or C.06.40, or with addi- tional G1369C LAN Card
G1364E/F, G5664B 2 Yes Yes No No E THERMOSTAT for G1330B
Others
G7116A/B 2 No No No No No Requires a HOST mod- ule via CAN
G7122A No No No Yes No A
G7170B 2 No No No No No Requires a host module with on-board LAN with minimum FW B.06.40 or C.06.40, or with addi- tional G1369C LAN Card
Table 23 Agilent InfinityLab LC Series Interfaces
Module CAN USB LAN (on-board)
RS-232 Analog APG (A) / ERI (E)
Special
NOTE The detector (DAD/MWD/FLD/VWD/RID) is the preferred access point for control via LAN. The inter-module communication is done via CAN.
Agilent InfinityLab LC Series Vialsamplers User Manual 320
11 Hardware Information Interfaces
Overview Interfaces
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus system supporting high speed data communication and real-time requirement.
LAN
The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D DAD and G1365C/D MWD). This interface allows the control of the module/system via a PC with the appropriate control software. Some modules have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A Valve Drive or G4227A Flexible Cube). These are hosted modules and require a Host module with firmware B.06.40 or later or with additional G1369C LAN Card.
USB
The USB interface replaces the RS-232 Serial interface in new FUSION generation modules. For details on USB refer to USB (Universal Serial Bus) on page 325.
Analog Signal Output
The analog signal output can be distributed to a recording device. For details refer to the description of the modules mainboard.
Remote (ERI)
The ERI (Enhanced Remote Interface) connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features as common shut down, prepare, and so on.
It allows easy connection between single instruments or systems to ensure coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote connector which is inputs/outputs (wired- or technique).
NOTE If an Agilent detector (DAD/MWD/FLD/VWD/RID) is in the system, the LAN should be connected to the DAD/MWD/FLD/VWD/RID (due to higher data load). If no Agilent detector is part of the system, the LAN interface should be installed in the pump or autosampler.
Agilent InfinityLab LC Series Vialsamplers User Manual 321
11 Hardware Information Interfaces
To provide maximum safety within a distributed analysis system, one line is dedicated to SHUT DOWN the systems critical parts in case any module detects a serious problem. To detect whether all participating modules are switched on or properly powered, one line is defined to summarize the POWER ON state of all connected modules. Control of analysis is maintained by signal readiness READY for next analysis, followed by START of run and optional STOP of run triggered on the respective lines. In addition PREPARE and START REQUEST may be issued. The signal levels are defined as: standard TTL levels (0 V is logic true, + 5.0 V is false), fan-out is 10, input load is 2.2 kOhm against + 5.0 V, and output are open collector type, inputs/outputs (wired- or technique).
Special Interfaces
There is no special interface for this module.
NOTE All common TTL circuits operate with a 5 V power supply. A TTL signal is defined as "low" or L when between 0 V and 0.8 V and "high" or H when between 2.0 V and 5.0 V (with respect to the ground terminal).
Table 24 ERI signal distribution
Pin Signal Description
1 START REQUEST (L) Request to start injection cycle (for example, by start key on any module). Receiver is the autosampler.
2 STOP (L) Request to reach system ready state as soon as possible (for example, stop run, abort or finish and stop injection). Receiver is any module performing run-time controlled activities.
3 READY (H) System is ready for next analysis. Receiver is any sequence con- troller.
4 POWER ON (H) All modules connected to system are switched on. Receiver is any module relying on operation of others.
5 Not used
6 SHUT DOWN (L) System has serious problem (for example, leak: stops pump). Receiver is any module capable to reduce safety risk.
7 START (L) Request to start run / timetable. Receiver is any module perform- ing run-time controlled activities.
8 PREPARE (L) Request to prepare for analysis (for example, calibration, detec- tor lamp on). Receiver is any module performing pre-analysis activi- ties.
Agilent InfinityLab LC Series Vialsamplers User Manual 322
11 Hardware Information Interfaces
ERI (Enhanced Remote Interface) ERI replaces the AGP Remote Interface that is used in the HP 1090/1040/1050/1100 HPLC systems and Agilent 1100/1200/1200 Infinity HPLC modules. All new InfinityLab LC Series products using the FUSION core electronics use ERI. This interface is already used in the Agilent Universal Interface Box 2 (UIB2)
ERI Description
The ERI interface contains eight individual programmable input/output pins. In addition, it provides 24 V power and 5 V power and a serial data line to detect and recognize further add-ons that could be connected to this interface. This way the interface can support various additional devices like sensors, triggers (in and out) and small controllers, etc.
Figure 53 Location of the ERI interface (example shows a G7114A/B VWD)
ERI
Agilent InfinityLab LC Series Vialsamplers User Manual 323
11 Hardware Information Interfaces
IO (Input/Output) Lines Eight generic bi-directional channels (input or output). Same as the APG Remote. Devices like valves, relays, ADCs, DACs, controllers can be
supported/controlled.
1-Wire Data (Future Use)
This serial line can be used to read out an EPROM or write into an EPROM of a connected ERI-device. The firmware can detect the connected type of device automatically and update information in the device (if required).
Pin Enhanced Remote
1 IO 1 (START REQUEST)
2 IO 2 (STOP)
3 IO 3 (READY)
4 IO 4 (POWER ON)
5 IO 5 (NOT USED)
6 IO 6 (SHUT DOWN)
7 IO 7 (START)
8 IO 8 (PREPARE)
9 1 wire DATA
10 DGND
11 +5 V ERI out
12 PGND
13 PGND
14 +24 V ERI out
15 +24 V ERI out
Agilent InfinityLab LC Series Vialsamplers User Manual 324
11 Hardware Information Interfaces
5V Distribution (Future Use) Available directly after turning on the hosting module (assures that the
firmware can detect certain basic functionality of the device). For digital circuits or similar. Provides 500 mA maximum. Short-circuit proof with automatic switch off (by firmware).
24V Distribution (Future Use) Available by firmware command (defined turn on/off). For devices that need higher power
Class 0: 0.5 A maximum (12 W) Class 1: 1.0 A maximum (24 W) Class 2: 2.0 A maximum (48 W)
Class depends on hosting modules internal power overhead. If a connected device requires more power the firmware detects this
(overcurrent detection) and provides the information to the user interface. Fuse used for safety protection (on board). Short circuit will be detected through hardware.
USB (Universal Serial Bus) USB (Universal Serial Bus) - replaces RS232, supports: a PC with control software (for example Agilent Lab Advisor) USB Flash Disk
Agilent InfinityLab LC Series Vialsamplers User Manual 325
11 Hardware Information Setting the 6-bit Configuration Switch
Setting the 6-bit Configuration Switch
The 6-bit configuration switch is located at the rear of the module with FUSION electronics. Switch settings provide configuration parameters for LAN and instrument specific initialization procedures. All modules with FUSION electronics: Default is ALL switches DOWN (best settings).
Default IP address for LAN 192.168.254.11 For specific LAN modes switches 4-5 must be set as required. For boot resident/cold start modes switches 1+2 or 6 must be UP.
Figure 54 Location of Configuration switch (example shows a G7114A/B VWD)
Configuration switch
Agilent InfinityLab LC Series Vialsamplers User Manual 326
11 Hardware Information Setting the 6-bit Configuration Switch
Table 25 6-bit Configuration Switch
Mode Function/Setting
Switch 1 Switch 2 Switch 3 Switch 4 Switch 5 Switch 6
COM1
1 When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are taken from non-volatile memory.
0 n.a.2
2 not assigned - Always keep these switches on position 0 (off)
n.a. LAN Init Mode n.a.
Use Default IP Address3
3 Default IP Address is 192.168.254.11
0 0 0 0 0
Use Stored IP Address 0 0 0 1 0
Use DHCP to request IP Address4
4 Host Name will be the MAC address.
0 0 1 0 0
Test 1 System n.a. n.a. n.a. ColdStart
Boot Main System / Keep Data 0 0 0 0 0
Boot Resident System / Keep Data 1 0 0 0 0
Boot Main System / Revert to Default Data
0 0 0 0 1
Boot Resident System / Revert to Default Data
1 0 0 0 1
Agilent InfinityLab LC Series Vialsamplers User Manual 327
11 Hardware Information Setting the 6-bit Configuration Switch
Special Settings
Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware loading errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again, the instrument firmware stays in the resident/main mode. In resident mode, it is not operable as a module. It only uses basic functions of the operating system for example, for communication. In this mode the main firmware can be loaded (using update utilities).
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with default parameter settings. Boot Main System / Revert to Default Data
The instrument will boot to main mode and changes to the modules default parameter. May be also required to load resident firmware into the module.
Boot Resident System / Revert to Default Data The instrument will boot to resident mode and changes to the modules default parameter. May be also required to load main firmware into the module.
If you use the following switch settings and power the instrument up again, it will start as described above.
CAUTION Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory. Exceptions are calibration settings, diagnosis and repair log books which will not be erased.
Save your methods and data before executing a forced cold start.
Agilent InfinityLab LC Series Vialsamplers User Manual 328
11 Hardware Information Setting the 6-bit Configuration Switch
Table 26 Boot Resident / Forced Coldstart
SW1 SW2 SW3 SW4 SW5 SW6 Init Mode
1 0 0 0 0 0 Boot Main System / Keep Data
1 1 0 0 0 0 Boot Resident System / Keep Data
1 0 0 0 0 1 Boot Main System / Revert to Default Data
1 1 0 0 0 1 Boot Resident System / Revert to Default Data
Note: The setting '0' (down) is essential.
Agilent InfinityLab LC Series Vialsamplers User Manual 329
11 Hardware Information Instrument Layout
Instrument Layout
The industrial design of the module incorporates several innovative features. It uses Agilents E-PAC concept for the packaging of electronics and mechanical assemblies. This concept is based upon the use of expanded polypropylene (EPP) layers of foam plastic spacers in which the mechanical and electronic boards components of the module are placed. This pack is then housed in a metal inner cabinet which is enclosed by a plastic external cabinet. The advantages of this packaging technology are: virtual elimination of fixing screws, bolts or ties, reducing the number of
components and increasing the speed of assembly/disassembly, the plastic layers have air channels molded into them so that cooling air can
be guided exactly to the required locations, the plastic layers help cushion the electronic and mechanical parts from
physical shock, and the metal inner cabinet shields the internal electronics from electromagnetic
interference and also helps to reduce or eliminate radio frequency emissions from the instrument itself.
Agilent InfinityLab LC Series Vialsamplers User Manual 330
11 Hardware Information Early Maintenance Feedback (EMF)
Early Maintenance Feedback (EMF)
Maintenance requires the exchange of components which are subject to wear or stress. Ideally, the frequency at which components are exchanged should be based on the intensity of usage of the module and the analytical conditions, and not on a predefined time interval. The early maintenance feedback (EMF) feature monitors the usage of specific components in the instrument, and provides feedback when the user-selectable limits have been exceeded. The visual feedback in the user interface provides an indication that maintenance procedures should be scheduled.
EMF Counters
EMF counters increment with use and can be assigned a maximum limit which provides visual feedback in the user interface when the limit is exceeded. Some counters can be reset to zero after the required maintenance procedure.
Using the EMF Counters
The user-settable EMF limits for the EMF Counters enable the early maintenance feedback to be adapted to specific user requirements. The useful maintenance cycle is dependent on the requirements for use. Therefore, the definition of the maximum limits need to be determined based on the specific operating conditions of the instrument.
Setting the EMF Limits
The setting of the EMF limits must be optimized over one or two maintenance cycles. Initially the default EMF limits should be set. When instrument performance indicates maintenance is necessary, take note of the values displayed by the EMF counters. Enter these values (or values slightly less than the displayed values) as EMF limits, and then reset the EMF counters to zero. The next time the EMF counters exceed the new EMF limits, the EMF flag will be displayed, providing a reminder that maintenance needs to be scheduled.
Agilent InfinityLab LC Series Vialsamplers User Manual 331
12 LAN Configuration
What You Have to Do First 333 TCP/IP parameter configuration 334 Configuration Switches 335 Initialization Mode Selection 336 Dynamic Host Configuration Protocol (DHCP) 338 General Information (DHCP) 338 Setup (DHCP) 339
Manual Configuration 341 With Telnet 342
PC and Agilent ChemStation Setup 345 PC Setup for Local Configuration 345 Agilent ChemStation Setup 348
This chapter provides information on connecting the module to the Agilent ChemStation PC.
Agilent InfinityLab LC Series Vialsamplers User Manual 332
12 LAN Configuration What You Have to Do First
What You Have to Do First
The module has an on- board LAN communication interface.
1 Note the MAC (Media Access Control) address for further reference. The MAC or hardware address of the LAN interfaces is a world wide unique identifier. No other network device will have the same hardware address. The MAC address can be found on a label at the rear of the module underneath the configuration switch (see Figure 56 on page 333).
Figure 55 MAC-Label
2 Connect the instrument's LAN interface (see Figure 56 on page 333) to the PC network card using a crossover network cable (point-to-point) or a hub or switch using a standard LAN cable.
Figure 56 Location of LAN interfaces and MAC label
NOTE This chapter is generic and may show figures that differ from your module. The functionality is the same.
Part number of the detector main board Revision Code, Vendor, Year and Week of assembly MAC address Country of Origin
LAN interface
MAC label
Agilent InfinityLab LC Series Vialsamplers User Manual 333
12 LAN Configuration TCP/IP parameter configuration
TCP/IP parameter configuration
To operate properly in a network environment, the LAN interface must be configured with valid TCP/IP network parameters. These parameters are: IP address Subnet Mask Default Gateway The TCP/IP parameters can be configured by the following methods: by automatically requesting the parameters from a network-based DHCP
Server (using the so-called Dynamic Host Configuration Protocol). This mode requires a LAN-onboard Module or a G1369C LAN Interface card, see Setup (DHCP) on page 339
by manually setting the parameters using Telnet by manually setting the parameters using the Local Controller
The LAN interface differentiates between several initialization modes. The initialization mode (short form init mode) defines how to determine the active TCP/IP parameters after power-on. The parameters may be derived non-volatile memory or initialized with known default values. The initialization mode is selected by the configuration switch, see Table 27 on page 336.
Agilent InfinityLab LC Series Vialsamplers User Manual 334
12 LAN Configuration Configuration Switches
Configuration Switches
The configuration switch can be accessed at the rear of the module.
Figure 57 Location of Configuration switch (example shows a G7114A/B VWD)
The module is shipped with all switches set to OFF, as shown above.
Configuration switch
NOTE To perform any LAN configuration, SW1 and SW2 must be set to OFF.
Agilent InfinityLab LC Series Vialsamplers User Manual 335
12 LAN Configuration Initialization Mode Selection
Initialization Mode Selection
The following initialization (init) modes are selectable:
Default IP address for LAN is 192.168.254.11.
DHCP address is the modules LAN MAC address.
Using Stored
When initialization mode Using Stored is selected, the parameters are taken from the non-volatile memory of the module. The TCP/IP connection will be established using these parameters. The parameters were configured previously by one of the described methods.
Figure 58 Using Stored (Principle)
Table 27 Initialization Mode Switches
SW1 SW2 SW3 SW4 SW5 SW6 Init Mode
0 0 0 0 0 0 Use Default IP Address
0 0 0 0 1 0 Use Stored IP Address
0 0 0 1 0 0 Use DHCP
Note: The setting 0 (down) is essential.
Non-Volatile RAM
Stored Parameter
Active Parameter
Agilent InfinityLab LC Series Vialsamplers User Manual 336
12 LAN Configuration Initialization Mode Selection
Using Default
When Using Default is selected, the factory default parameters are taken instead. These parameters enable a TCP/IP connection to the LAN interface without further configuration, see Table 28 on page 337.
Figure 59 Using Default (Principle)
Since the default IP address is a so-called local address, it will not be routed by any network device. Thus, the PC and the module must reside in the same subnet.
The user may open a Telnet session using the default IP address and change the parameters stored in the non-volatile memory of the module. He may then close the session, select the initialization mode Using Stored, power-on again and establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a local hub), separated from the local area network, the user may simply keep the default parameters to establish the TCP/IP connection.
Active Parameter
Default Parameter
NOTE Using the default address in your local area network may result in network problems. Take care and change it to a valid address immediately.
Table 28 Using Default Parameters
IP address: 192.168.254.11
Subnet Mask: 255.255.255.0
Default Gateway not specified
NOTE In the Using Default mode, the parameters stored in the memory of the module are not cleared automatically. If not changed by the user, they are still available, when switching back to the mode Using Stored.
Agilent InfinityLab LC Series Vialsamplers User Manual 337
12 LAN Configuration Dynamic Host Configuration Protocol (DHCP)
Dynamic Host Configuration Protocol (DHCP)
General Information (DHCP) The Dynamic Host Configuration Protocol (DHCP) is an auto configuration protocol used on IP networks. The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and B-firmware (B.06.40 or above) or modules with "D"-firmware. All modules should use latest firmware from the same set.
When the initialization mode DHCP is selected, the card tries to download the parameters from a DHCP Server. The parameters obtained become the active parameters immediately. They are not stored to the non-volatile memory of the card.
Besides requesting the network parameters, the card also submits its hostname to the DHCP Server. The hostname equals the MAC address of the card, e.g. 0030d3177321. It is the DHCP server's responsibility to forward the hostname/address information to the Domain Name Server. The card does not offer any services for hostname resolution (e.g. NetBIOS).
Figure 60 DHCP (Principle)
DHCP Server
Active Parameter
NOTE 1 It may take some time until the DHCP server has updated the DNS server with the hostname information.
2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g. 0030d3177321.country.company.com.
3 The DHCP server may reject the hostname proposed by the card and assign a name following local naming conventions.
Agilent InfinityLab LC Series Vialsamplers User Manual 338
12 LAN Configuration Dynamic Host Configuration Protocol (DHCP)
Setup (DHCP)
1 Note the MAC address of the LAN interface (provided with G1369C LAN Interface Card or mainboard). This MAC address is on a label on the card or at the rear of the mainboard, for example, 0030d3177321. On the Local Controller the MAC address can be found under Details in the LAN section.
Figure 61 LAN Setting on Instant Pilot
The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and B-firmware (B.06.40 or above) or modules with "D"-firmware. All mod- ules should use latest firmware from the same set.
Agilent InfinityLab LC Series Vialsamplers User Manual 339
12 LAN Configuration Dynamic Host Configuration Protocol (DHCP)
2 Set the configuration switch to DHCP either on the G1369C LAN Interface Card or the mainboard of above mentioned modules.
3 Turn on the module that hosts the LAN interface. 4 Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition, Lab
Advisor, Firmware Update Tool) and use MAC address as host name, e.g. 0030d3177321. The LC system should become visible in the control software (see Note in section General Information (DHCP) on page 338).
Table 29 G1369C LAN Interface Card (configuration switch on the card)
SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode
ON OFF OFF OFF OFF DHCP
Table 30 LC Modules with 8-bit configuration switch (B-firmware) (configuration switch at rear of the instrument)
SW 6 SW 7 SW 8 Initialization Mode
ON OFF OFF DHCP
Agilent InfinityLab LC Series Vialsamplers User Manual 340
12 LAN Configuration Manual Configuration
Manual Configuration
Manual configuration only alters the set of parameters stored in the non-volatile memory of the module. It never affects the currently active parameters. Therefore, manual configuration can be done at any time. A power cycle is mandatory to make the stored parameters become the active parameters, given that the initialization mode selection switches are allowing it.
Figure 62 Manual Configuration (Principle)
TELNET Session
Control Module
Stored Parameter
Non-Volatile RAM
Agilent InfinityLab LC Series Vialsamplers User Manual 341
12 LAN Configuration Manual Configuration
With Telnet Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set by any method), the parameters may be altered by opening a Telnet session. 1 Open the system (DOS) prompt window by clicking on Windows START
button and select Run.... Type cmd and press OK. 2 Type the following at the system (DOS) prompt:
c:\>telnet
Figure 63 Telnet - Starting a session
where
Figure 64 A connection to the module is made
Agilent InfinityLab LC Series Vialsamplers User Manual 342
12 LAN Configuration Manual Configuration
3 Type ? and press enter to see the available commands.
Figure 65 Telnet Commands
4 To change a parameter follows the style: parameter value, for example:
ip 134.40.28.56 Then press [Enter], where parameter refers to the configuration parameter you are defining, and value refers to the definitions you are assigning to that parameter. Each parameter entry is followed by a carriage return.
Table 31 Telnet Commands
Value Description
? displays syntax and descriptions of commands
/ displays current LAN settings
ip
sm
gw
exit exits shell and saves all changes
Agilent InfinityLab LC Series Vialsamplers User Manual 343
12 LAN Configuration Manual Configuration
5 Use the / and press Enter to list the current settings.
6 Change the IP address (in this example 192.168.254.12) and type / to list current settings.
7 When you have finished typing the configuration parameters, type exit and press Enter to exit with storing parameters.
Figure 68 Closing the Telnet Session
Figure 66 Telnet - Current settings in "Using Stored" mode
information about the LAN interface MAC address, initialization mode Initialization mode is Using Stored active TCP/IP settings
TCP/IP status - here ready connected to PC with controller software (e.g. Agilent ChemStation), here not connected
Figure 67 Telnet - Change IP settings
change of IP setting to Initialization mode is Using Stored
active TCP/IP settings
stored TCP/IP settings in non-volatile memory
connected to PC with controller software (e.g. Agilent ChemStation), here not connected
NOTE If the Initialization Mode Switch is changed now to Using Stored mode, the instrument will take the stored settings when the module is re-booted. In the example above it would be 192.168.254.12.
Agilent InfinityLab LC Series Vialsamplers User Manual 344
12 LAN Configuration PC and Agilent ChemStation Setup
PC and Agilent ChemStation Setup
PC Setup for Local Configuration This procedure describes the change of the TCP/IP settings on your PC to match the modules default parameters in a local configuration (see Table 28 on page 337). 1 Open the Local Area Connection Properties and select Internet Protocol
(TCP/IP). Then click on Properties.
Agilent InfinityLab LC Series Vialsamplers User Manual 345
12 LAN Configuration PC and Agilent ChemStation Setup
2 You may enter here the fixed IP address of the module or use the Alternative Configuration.
Agilent InfinityLab LC Series Vialsamplers User Manual 346
12 LAN Configuration PC and Agilent ChemStation Setup
3 We will use the direct LAN access via Cross-over LAN cable with the modules IP address.
4 Click on OK to save the configuration.
Agilent InfinityLab LC Series Vialsamplers User Manual 347
12 LAN Configuration PC and Agilent ChemStation Setup
Agilent ChemStation Setup This procedure describes the Agilent ChemStation B.04.02 setup for the 1290 Infinity system using the 1290 Infinity DAD (G4212A) as the interfacing module. The setup works in the same way for all other systems.
1 Open the ChemStation Configuration Editor.
2 Select from the menu Configure - Instruments. 3 Select Modular 3D LC System. 4 Give the Instrument a name.
NOTE The LAN must be connected to detector due to high data load on communication to Control Software.
Agilent InfinityLab LC Series Vialsamplers User Manual 348
12 LAN Configuration PC and Agilent ChemStation Setup
5 Click on OK.
6 Select LC System Access Access Point and click on Add.
7 Click on OK. The Configuration Editor shows now the new instrument.
Agilent InfinityLab LC Series Vialsamplers User Manual 349
12 LAN Configuration PC and Agilent ChemStation Setup
8 If required, change under Configure Path the folder locations. 9 Save the current configuration via File Save.
10 Exit the Configuration Editor. 11 Start the Agilent ChemStation.
During first startup or when the system configuration has changed, a notification shows up.
12 The left column shows the modules that could be configured. You may select the module manually from the list. We use the Auto Configuration mode. Click on Yes.
Agilent InfinityLab LC Series Vialsamplers User Manual 350
12 LAN Configuration PC and Agilent ChemStation Setup
13 Enter the IP address or the Hostname of the module with the LAN-access.
14 Click on OK. The selected module is shown now in the right window (with serial number). In addition all other modules connected via CAN to the detector are shown as well.
15 Click on OK to continue the ChemStation loading.
Agilent InfinityLab LC Series Vialsamplers User Manual 351
12 LAN Configuration PC and Agilent ChemStation Setup
16 You may see the details of the module by selecting the module and clicking on Configure.
Under Connection Settings you may change the IP/Hostname of the module (may require a re-start of the ChemStation).
After successful load of the ChemStation, you should see the module(s) as active item in the graphical user interface (GUI).
Agilent InfinityLab LC Series Vialsamplers User Manual 352
12 LAN Configuration PC and Agilent ChemStation Setup
Figure 69 Screen After Successful Load of ChemStation
Agilent InfinityLab LC Series Vialsamplers User Manual 353
13 Appendix
General Safety Information 355 General Safety Information 355 Safety Standards 355 General 355 Before Applying Power 356 Ground the Instrument 356 Do Not Operate in an Explosive Atmosphere 357 Do Not Remove the Instrument Cover 357 Do Not Modify the Instrument 357 In Case of Damage 357 Solvents 358 Symbols 359
Waste Electrical and Electronic Equipment (WEEE) Directive 361 Refrigerant 362 Radio Interference 365 Sound Emission 366 Solvent Information 367 Agilent Technologies on Internet 368
This chapter provides additional information on safety, legal and web.
Agilent InfinityLab LC Series Vialsamplers User Manual 354
13 Appendix General Safety Information
General Safety Information
General Safety Information The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customers failure to comply with these requirements.
Safety Standards This is a Safety Class I instrument (provided with terminal for protective earthing) and has been manufactured and tested according to international safety standards.
General Do not use this product in any manner not specified by the manufacturer. The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
WARNING Ensure the proper usage of the equipment.
The protection provided by the equipment may be impaired.
The operator of this instrument is advised to use the equipment in a manner as specified in this manual.
Agilent InfinityLab LC Series Vialsamplers User Manual 355
13 Appendix General Safety Information
Before Applying Power
Ground the Instrument
WARNING Wrong voltage range, frequency or cabling
Personal injury or damage to the instrument
Verify that the voltage range and frequency of your power distribution matches to the power specification of the individual instrument.
Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
Make all connections to the unit before applying power.
NOTE Note the instrument's external markings described under Symbols on page 359.
WARNING Missing electrical ground
Electrical shock
If your product is provided with a grounding type power plug, the instrument chassis and cover must be connected to an electrical ground to minimize shock hazard.
The ground pin must be firmly connected to an electrical ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Agilent InfinityLab LC Series Vialsamplers User Manual 356
13 Appendix General Safety Information
Do Not Operate in an Explosive Atmosphere
Do Not Remove the Instrument Cover
Do Not Modify the Instrument Do not install substitute parts or perform any unauthorized modification to the product. Return the product to an Agilent Sales and Service Office for service and repair to ensure that safety features are maintained.
In Case of Damage
WARNING Presence of flammable gases or fumes
Explosion hazard
Do not operate the instrument in the presence of flammable gases or fumes.
WARNING Instrument covers removed
Electrical shock
Do Not Remove the Instrument Cover
Only Agilent authorized personnel are allowed to remove instrument covers. Always disconnect the power cables and any external circuits before removing the instrument cover.
WARNING Damage to the module
Personal injury (for example electrical shock, intoxication)
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
Agilent InfinityLab LC Series Vialsamplers User Manual 357
13 Appendix General Safety Information
Solvents
WARNING Toxic, flammable and hazardous solvents, samples and reagents The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
Do not use solvents with an auto-ignition temperature below 200 C (392 F). Do not use solvents with a boiling point below 56 C (133 F).
Avoid high vapor concentrations. Keep the solvent temperature at least 40 K below the boiling point of the solvent used. This includes the solvent temperature in the sample compartment. For the solvents methanol and ethanol keep the solvent temperature at least 25 K below the boiling point.
Do not operate the instrument in an explosive atmosphere.
Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for example, carbon disulfide).
Reduce the volume of substances to the minimum required for the analysis.
Never exceed the maximum permissible volume of solvents (8 L) in the solvent cabinet. Do not use bottles that exceed the maximum permissible volume as specified in the usage guideline for solvent cabinet.
Ground the waste container.
Regularly check the filling level of the waste container. The residual free volume in the waste container must be large enough to collect the waste liquid.
To achieve maximal safety, regularly check the tubing for correct installation.
NOTE For details, see the usage guideline for the solvent cabinet. A printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available in the Agilent Information Center or via the Internet.
Agilent InfinityLab LC Series Vialsamplers User Manual 358
13 Appendix General Safety Information
Symbols Table 32 Symbols
The apparatus is marked with this symbol when the user shall refer to the instruction manual in order to protect risk of harm to the operator and to protect the apparatus against damage.
Indicates dangerous voltages.
Indicates a protected ground terminal.
The apparatus is marked with this symbol when hot surfaces are available and the user should not touch it when heated up.
Sample Cooler unit is designed as vapor-compression refrigeration system. Contains fluorinated greenhouse gas (refrigerant) according to the Kyoto protocol. For specifications of refrigerant, charge capacity, carbon dioxide equivalent (CDE), and global warming potential (GWP) see instrument label.
Flammable Material For Sample Thermostat which uses flammable refrigerant consult Agilent Information Center / User Manual before attempting to install or service this equipment. All safety precautions must be followed.
Confirms that a manufactured product complies with all applicable Euro- pean Community directives. The European Declaration of Conformity is available at: http://regulations.corporate.agilent.com/DoC/search.htm
Manufacturing date.
Power symbol indicates On/Off. The apparatus is not completely disconnected from the mains supply when the power switch is in the Off position
Pacemaker Magnets could affect the functioning of pacemakers and implanted heart defibrillators. A pacemaker could switch into test mode and cause illness. A heart defibril- lator may stop working. If you wear these devices keep at least 55 mm dis- tance to magnets. Warn others who wear these devices from getting too close to magnets.
Agilent InfinityLab LC Series Vialsamplers User Manual 359
13 Appendix General Safety Information
Magnetic field Magnets produce a far-reaching, strong magnetic field. They could damage TVs and laptops, computer hard drives, credit and ATM cards, data storage media, mechanical watches, hearing aids and speakers. Keep magnets at least 25 mm away from devices and objects that could be damaged by strong magnetic fields.
Indicates a pinching or crushing hazard
Indicates a piercing or cutting hazard.
Table 32 Symbols
WARNING A WARNING
alerts you to situations that could cause physical injury or death.
Do not proceed beyond a warning until you have fully understood and met the indicated conditions.
CAUTION A CAUTION
alerts you to situations that could cause loss of data, or damage of equipment.
Do not proceed beyond a caution until you have fully understood and met the indicated conditions.
Agilent InfinityLab LC Series Vialsamplers User Manual 360
13 Appendix Waste Electrical and Electronic Equipment (WEEE) Directive
Waste Electrical and Electronic Equipment (WEEE) Directive
This product complies with the European WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.
NOTE Do not dispose of in domestic household waste
To return unwanted products, contact your local Agilent office, or see http://www.agilent.com for more information.
Agilent InfinityLab LC Series Vialsamplers User Manual 361
13 Appendix Refrigerant
Refrigerant
The refrigerant HFC-134a is used only in the Agilent Infinity II Sample Cooler.
Table 33 Physical properties of refrigerant HFC-134a
Molecular weight 102
Critical temperature 101.1 C
Critical pressure 40.6 bar
Boiling point -26.5 C
Table 34 Physical properties of refrigerant R600a (isobutane)
Molecular weight 58.12
Critical temperature 134.98 C
Critical pressure 36.6 bar
Boiling point -11.7 C
Agilent InfinityLab LC Series Vialsamplers User Manual 362
13 Appendix Refrigerant
WARNING Refrigerant
Refrigerant HFC-134a is known as a safe refrigerant, however accidents can occur if it is handled incorrectly. For this reason, the following instructions must be observed:
Avoid contact with liquid refrigerant HFC-134a. At atmospheric pressure HFC-134a evaporates at approximately -26 C and causes frost bite.
After skin contact, rinse the affected area with water.
After eye contact, rinse the eye(s) with plenty of water for at least 15 minutes and consult a doctor.
HFC-134a must not be allowed to escape in enclosed areas. Although HFC-134a is not toxic, there is a danger of suffocation as gaseous refrigerant is heavier than air.
Please observe the following first aid instructions. After inhalation, move the affected person to fresh air, keep him warm and allow him to rest. If necessary, he should be supplied with oxygen. If he has stopped breathing or is breathing erratically, he should be given artificial respiration. In the case of cardiac arrest, carry out heart massage. Send for a doctor immediately.
Moreover, it must be noted that HFC-134a must always be extracted from the system and collected. It must never be discharged into the atmosphere on environmental grounds (greenhouse effect).
Agilent InfinityLab LC Series Vialsamplers User Manual 363
13 Appendix Refrigerant
CAUTION General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
The disposal or scrapping of the Sample Cooler or the Sample Thermostat must be carried out by a qualified disposal company.
All media must be disposed of in accordance with national and local regulations.
Please contact your local Agilent Service Center in regard to safe environmental disposal of the appliance or check www.agilent.com for more info.
CAUTION Risk of fire or explosion
Dispose of properly in accordance with federal or local regulations. Flammable Refrigerant Used.
Do not dispose of in domestic household waste.
To return unwanted products, contact your local Agilent office, or see http://www.agilent.com for more information.
Agilent InfinityLab LC Series Vialsamplers User Manual 364
13 Appendix Radio Interference
Radio Interference
Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
Test and Measurement
If test and measurement equipment is operated with equipment unscreened cables and/or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.
Agilent InfinityLab LC Series Vialsamplers User Manual 365
13 Appendix Sound Emission
Sound Emission
Manufacturers Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive of 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB. Sound Pressure Lp < 70 dB (A) At Operator Position Normal Operation According to ISO 7779:1988/EN 27779/1991 (Type Test)
Agilent InfinityLab LC Series Vialsamplers User Manual 366
13 Appendix Solvent Information
Solvent Information
Observe the following recommendations on the use of solvents. Brown glass ware can avoid growth of algae. Avoid the use of the following steel-corrosive solvents:
solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on),
high concentrations of inorganic acids like sulfuric acid and nitric acid, especially at higher temperatures (if your chromatography method allows, replace by phosphoric acid or phosphate buffer which are less corrosive against stainless steel),
halogenated solvents or mixtures which form radicals and/or acids, for example: 2CHCl3 + O2 2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol,
chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, diisopropyl ether) should be filtered through dry aluminium oxide which adsorbs the peroxides,
solvents containing strong complexing agents (e.g. EDTA), mixtures of carbon tetrachloride with 2-propanol or THF.
Avoid the use of dimethyl formamide (DMF). Polyvinylidene fluoride (PVDF), which is used in leak sensors, is not resistant to DMF.
Agilent InfinityLab LC Series Vialsamplers User Manual 367
13 Appendix Agilent Technologies on Internet
Agilent Technologies on Internet
For the latest information on products and services visit our worldwide web site on the Internet at:
http://www.agilent.com
Agilent InfinityLab LC Series Vialsamplers User Manual 368
In This Book
This manual contains technical reference information about the Agilent 1260 Infinity II Vialsampler (G7129A), the Agilent 1290 Infinity II Vialsampler (G7129B), and the Agilent 1260 Infinity II Vialsampler (G7129C). The manual describes the following: Introduction, site requirements and specifications, using the module, optimizing performance, troubleshooting and diagnostics, error information, maintenance, parts and materials, hardware information, LAN configuration, safety and related information.
www.agilent.com Agilent Technologies Inc. 2015-2020
Published in Germany 04/2020
G7129-90010*
Part No: G7129-90
Related manuals for Agilent InfinityLab LC Series Sampler User Manual
Manualsnet FAQs
If you want to find out how the InfinityLab LC Series Agilent works, you can view and download the Agilent InfinityLab LC Series Sampler User Manual on the Manualsnet website.
Yes, we have the User Manual for Agilent InfinityLab LC Series as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.
The User Manual should include all the details that are needed to use a Agilent InfinityLab LC Series. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.
The best way to navigate the Agilent InfinityLab LC Series Sampler User Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.
This Agilent InfinityLab LC Series Sampler User Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.
You can download Agilent InfinityLab LC Series Sampler User Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.
To be able to print Agilent InfinityLab LC Series Sampler User Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent InfinityLab LC Series Sampler User Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.