Agilent 490 Micro GC

Manual Injection Port Field Kit

Installation Instructions

This document describes the installation of the field kits for the manual injection ports (septum nut and luerlock) at the front of the Agilent 490 Micro GC.

NOTE When the Agilent 490 Micro GC is equipped with a standard unheated front inlet, an additional field upgrade kit for the sample inlet box is required. The part number for a dual cabinet Agilent Micro GC is CP490202, for the quad cabinet Agilent Micro GC use part number CP190203.

Agilent Technologies

2

Agilent 490 Micro GC

Table 1 Parts list for the four kit types

Kit types available

Description and Part number

CP490204 Septum

CP490205 Luerlock

CP490206 Septum + Rear

CP490207 Luerlock + Rear

Bracket with pressure relief valve CP742849/CP742987

X X

Bracket with pressure relief valve + ball-valve CP742849/CP742987/CP958942

X X

Sample in bracket CP739952

X X X X

Torx screw M4 8 CP86757

X X X X

Bulkhead reducing union 1/8 1/16-inch CP72472

X X

Septum injection port Assy CP742983

X X

Septum low bleed 10/pk (Spare part) CP724384

X X

ZF2SI Support 1/pk (Spare part) CP742990

X X

Unheated sample inlet CP740570

X X

Fitting ext. luerlock 10-32 thrd 1/pk VLZLA1

X X

Female luerlock adapter to 1/16-inch ftgs CP914757

X X

CP742983 VLZLA1 CP914757

Inst

allation Instructions

Agilent 490 Micro GC

Mounting the Sample-in at the Front of the Agilent 490 Micro GC

Installation Instructio

1 Remove the side and front panels of the Agilent 490 Micro GC.

2 Make a hole in the preferred place on the front panel next to Sample 1. See Figure 1.

3 Install the sample bracket. See Figure 2.

Figure 1 Sample 1 front panel hole

Figure 2 Sample bracket

ns 3

4

Agilent 490 Micro GC

If using the Injection-Port CP490204 or CP490206 use the parts listed in Figure 3.

If using the Injection-Port CP490205 or CP490207 use the parts listed in Figure 4.

4 Reinstall the front panel.

5 In case of heated or unheated Sample-Inlet on the rear remove the Sample-Inlet-Box.

Place the preassembled bracket above the sample parts. To fix the bracket use three M4 Torx screws, p/n CP86757. See Figure 5 for placement of the screws.

Figure 3 Sample bracket parts for CP490204 or CP490206

Figure 4 Sample bracket parts for CP490205 or CP490207

1. CP742983

2. CP72472

3. CP86757

4. CP739952

1. VLZLA1

2. CP914757

3. CP740570

4. CP86757

5. CP739952

Installation Instructions

Agilent 490 Micro GC

Installation Instructio

The diagrams below show the differences for the different field kits (Figure 6 and Figure 7). Plumb the tubing according to these diagrams and Figure 8.

Figure 5 Sample bracket parts for CP490205 or CP490207

Figure 6 Diagram of unheated sample-in-Agilent 490 Micro GC for CP490204 or CP490205

ns 5

6

Agilent 490 Micro GC

Figure 7 Diagram of unheated sample-in-Agilent 490 Micro GC for CP490206 or CP490207

Figure 8 Diagram of unheated sample-in-Agilent 490 Micro GC

Installation Instructions

Agilent 490 Micro GC

Installation Instructio

6 Make an 18 mm incision in the upper insulation plate as seen in Figure 9 and install it in the Sample-Inlet-Box, see Figure 10 and Figure 11.

7 Reinstall the side panel.

Figure 9 Upper insulation plate

Figure 10 Sample-Inlet-Box without insulation

Figure 11 Sample-Inlet-Box with insulation

18 mm

ns

7

Agilent Technologies

Warranty

The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warran- ties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms

Manualsnet FAQs

If you want to find out how the 490 Micro GC Agilent works, you can view and download the Agilent 490 Micro GC Port Field Kit Installation Instructions on the Manualsnet website.

Yes, we have the Installation Instructions for Agilent 490 Micro GC as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Installation Instructions should include all the details that are needed to use a Agilent 490 Micro GC. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent 490 Micro GC Port Field Kit Installation Instructions is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent 490 Micro GC Port Field Kit Installation Instructions consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent 490 Micro GC Port Field Kit Installation Instructions free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent 490 Micro GC Port Field Kit Installation Instructions, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent 490 Micro GC Port Field Kit Installation Instructions as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.