Agilent G4900DA SureScan Microarray Scanner System Microarray Scan Control Software 9.1 User Guide

For Research Use Only. Not for use in diagnostic procedures.

Revision C0, November 2022

Notices Agilent Technologies, Inc. 2011-2013, 2015, 2019, 2022

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

Manual Part Number G4900-90000

Edition Revision C0, November 2022

Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051 USA

Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation).

Safety Notices

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

SureScan Microarray Scanner System User Guide 2

This guide applies to the Microarray Scan Control Software 9.1.3 or higher until superseded.

In This Guide This user guide contains instructions to get started with, use, troubleshoot, and maintain your SureScan Microarray Scanner System. The system is comprised of the SureScan microarray scanner, computer workstation, and control and analysis programs.

1 Introduction

This chapter describes the microarray scanner and preparation for its operation.

2 Getting Started

In this chapter, learn how to operate the SureScan scanner safely and learn how to set up and then scan slides with the Agilent Microarray Scan Control program.

3 Using the Scan Control Program

This chapter describes how you use the Scan Control program to set up and run the scanner.

4 Using Scan Control Tools

In this chapter, learn how to use commands in the Scan Control Tools menu. This chapter includes how to set up scan protocols and scan regions, display log files, and select default settings for the program.

5 Maintaining and Troubleshooting Your System

This chapter describes maintenance procedures and tips, as well as error messages for troubleshooting your system.

6 Reference

This chapter includes descriptions of the Scan Control windows and dialog boxes. It also contains scanner and slide specifications and regulatory compliance information.

SureScan Microarray Scanner System User Guide 3

South Korean Class A EMC Declaration This equipment has been evaluated for its suitability for use in a commercial environment. When used in a domestic environment, there is a risk of radio interference.

.

SureScan Microarray Scanner System User Guide 4

Content

1 Introduction 9

A Powerful Tool for Microarray Analysis 10

System Description 11 Hardware and software features 11 Parts list 11 Computer system requirements 12 Scanner physical description 12 Site preparation 13 Safety guidelines 13

Principles of Operation 15 Slide positioning 15 Laser excitation 15 Scanning 15 Fluorescence detection 15

Programs Installed on the Computer Workstation 16

2 Getting Started 17

Operating the Scanner 18 Step 1. Turn on the SureScan scanner and start the Scan Control program 18 Step 2. Insert slides into slide holders 20 Step 3. Load the slide holders into the cassette 22 Step 4. Set or change protocol scan settings 24 Step 5. Add slides to the scan queue 24 Step 6. Scan your slides 25 Step 7. Remove the slides 25

About Scan Protocols 26

Offline Mode 27

Turning Off the SureScan Scanner 28

3 Using the Scan Control Program 29

Using the Slot Table 30 To change a Slide ID 30 To set or change the scan protocol 30 To change the output folder for a scan 31 To apply a selection to multiple slides 31 To add a slide to the scan queue 31

SureScan Microarray Scanner System User Guide 5

To add all slides to the queue 31 To move a slide in the queue 32 To remove a slide from the scan queue 32 To remove all slides from the scan queue 32 To open the scanner door 32 To close the scanner door 33 To start a scan 33

Changing Slide Scan Settings 34 To change settings for a single slide 34 To change settings for multiple slides 34 To add a description for the slide 35 To add a user name 35

4 Using Scan Control Tools 36

Creating and Changing Scan Protocols 37 To create a scan protocol 37 To change an existing scan protocol 37 To export a scan protocol 38 To import a scan protocol 38 To remove a scan protocol 39

Creating and Changing Scan Regions 40 To create a scan region 40 Change an existing scan region for non-Agilent slides 40 To export a scan region 41 To import a scan region 41 To remove a scan region 42

Adding a Barcode 43 To add a barcode 43

Turning on Lasers Manually 44 To turn on lasers 44

Troubleshooting Tools 45 To display recent errors 45 To display log files 45 To create a snapshot of the instrument state 45 To reset calibration warnings 45 To run a self test 46

Setting up Scanner Defaults 47 To set the default scan data folder 47 To set the laser saver delay 47 To set the location of the Feature Extraction program 47

SureScan Microarray Scanner System User Guide 6

To select to run Feature Extraction after scanning 48

Mapping Scan Protocols to Designs 49 To map a scan protocol to a slide design 49

5 Maintaining and Troubleshooting Your System 50

Maintaining Your System 51 Scheduled software maintenance 51 Scheduled hardware maintenance 51 Tips to prevent problems 52

Troubleshooting Your System 53 Technical Support 53 Frequently Asked Questions (FAQs) 54 Hardware Troubleshooting 55 Software Troubleshooting 56 File locations 57 System error messages 57

Using the Diagnostic Display 63

Updating the Scanner Firmware and Scan Control Program 64 To check if an update is available 64 To update the Scan Control program and firmware 64 To verify the software installation 65

6 Reference 66

Scan Control Program Window Reference 67 Scan Control main window 67 Tools menu 68 Help menu 69 Slot table 70 Function buttons 72 Settings pane 73 Log tabs 74

Scan Control Program Dialog Box Reference 75 Agilent Installation Qualification Tool 75 Export Scan Protocol dialog box 76 Export Scan Region dialog box 77 Input Barcode dialog box 77 Scan Protocol Editor dialog box 78 Scan Region Editor dialog box 83 Self Test dialog box 84 Settings dialog box General Settings 85

SureScan Microarray Scanner System User Guide 7

Settings dialog box Design To Protocol Mapping 86

About Adding Slides 88

Performing Feature Extraction on Scanned Images 89

SureScan Scanner Specifications 90

Slide Specifications 92 Scan dimensions 92 Glass specifications 92 Barcode and barcode label specifications 92

Regulatory Information 95

SureScan Microarray Scanner System User Guide 8

SureScan Microarray Scanner System User Guide

1 Introduction A Powerful Tool for Microarray Analysis 10 System Description 11 Principles of Operation 15 Programs Installed on the Computer Workstation 16

This chapter provides a general introduction to the SureScan Microarray Scanner System.

9

A Powerful Tool for Microarray Analysis

The SureScan scanner is part of the SureScan system solution from Agilent Technologies. The scanner is a sophisticated laser-induced fluorescence scanner designed to read microarrays printed on standard 1 in 3 in slides.

The scanner measures the fluorescence intensity of labeled sample nucleic acid (DNA and RNA) bound to microarrays. Its ability to measure fluorescence from two dyes simultaneously facilitates all two-color microarray studies. This technology provides for rapid, high-quality, automated scanning of microarrays.

Figure 1 SureScan scanner

Each slide is scanned in minutes, and the files are prepared for feature extraction analysis.

SureScan Microarray Scanner System User Guide 10

System Description

In this section you find listings of hardware and software features, parts, and computer requirements. A physical description of the SureScan scanner and information on site preparation and safety are also provided.

Hardware and software features The SureScan scanner provides the following features: Dynamic autofocus Single and dual color scanning Automatic PMT gain calibration before each scan 2-, 3-, 5-, or 10-micron pixel size Dynamic range of >104 for a single scan in 16-bit scan mode, >105 for a single scan in 20-bit

scan mode, and >106 for a dual scan in 16-bit scan mode (XDR) Uniformity specification of <5% CV (Coefficient of Variation) TIFF image file compression Flip and rotate images Internal and external barcode reading

The Agilent Microarray Scan Control program allows you to select the dye (fluorescence) channels, scan regions, resolution, dynamic range, PMT gain, and output folders for each of the slides in the cassette. You can load these settings automatically from saved application-specific protocols, or set them manually in the slot table.

Parts list The G4900DA SureScan microarray scanner system consists of the following components: G2600D SureScan scanner with integral 24-slide cassette 24 slide holders Computer workstation with recovery software on CD Power cords and network cable Agilent Microarray Scan Control Software installed Agilent Feature Extraction Software installed Agilent Installation Qualification Tool Software installed Declaration of Conformity

SureScan Microarray Scanner System User Guide 11

Computer system requirements The SureScan Microarray Scanner System comes with a computer that meets or exceeds the following configuration. Agilent Technologies supports only the computer provided with the system.

Software

Windows 10 64-bit Pro operating system For more details, see the release notes for your specific version of the Scan Control program or visit the Scan Control download site at: https://www.agilent.com/en/download-surescan-scan-control-software

Hardware

Intel Core 2 Duo E8500 3.16 GHz or equivalent 8 GB RAM Minimum 250-GB hard disk. (Proper disk maintenance is required to ensure that you always

have available disk space for data generation. See Table 11 on page 79 for estimated sizes of scanned images.)

Scanner physical description

Scanner front view

Figure 2 SureScan scanner, front view

Power On/Off

Door

Status indicator light

Removable front panel

SureScan Microarray Scanner System User Guide 12

Scanner rear view

Figure 3 SureScan scanner, rear view

Site preparation Make sure that the environment meets the SureScan Scanner Specifications on page 90 of Chapter 6. If you have any questions, contact your local Agilent sales and support center or www.agilent.com.

Safety guidelines The SureScan scanner is designed for safety and ease of use. Be sure that you understand and observe all the warnings and cautions before operating the SureScan scanner.

Power fuse Power plug

LAN connector

Laser cooling air outlet

Laser cooling air inlet

Main cooling air outlet

Main cooling air inlet

Power supply air outlet

WARNING Do not attempt to repair or gain access to SureScan scanner internal components. You risk exposure to high voltage and harmful laser radiation. Removing the main cover voids the warranty

WARNING Connect the SureScan scanner to a grounded power outlet. It relies on a protective earth ground for safety.

CAUTION In order to minimize vibration due to the rapid scanning of the laser excitation across the microarray, install the scanner on a sturdy lab bench or table. Do not install the scanner in proximity to other lab equipment that might cause vibration.

SureScan Microarray Scanner System User Guide 13

CAUTION The SureScan scanner is sensitive to condensing humidity conditions. Follow precautions stated in product documentation. See Tips to prevent problems on page 52 of Chapter 5 and the relative humidity specifications on page 90 of Chapter 6.

SureScan Microarray Scanner System User Guide 14

Principles of Operation

This section describes the operating features of the SureScan scanner.

Slide positioning The SureScan scanner holds up to 24 1 in 3 in slides in a nonremovable cassette. During scanning, the slides are sequentially transported into scanning position, scanned, and then returned to the cassette. Slides can be added to and removed from the scanner queue dynamically without stopping the scanner.

Laser excitation The SureScan scanner uses two lasers; a green diode-pumped solid-state laser (532 nm) and a red diode laser (640 nm). The lasers excite Cyanine-3 (Cy-3) and Cyanine-5 (Cy-5) labeled RNA or DNA to measure fluorescence after hybridization of the target nucleic acid to the microarray probes. Dyes similar to Cy-3 and Cy-5, such as Alexa 647, 555, and 660 dyes, can also be used.

The SureScan scanner is optimized for high signal-to-noise performance in the Cy-3 (550 610 nm) and Cy-5 (650 750 nm) emission bands. It has a wide dynamic range and low spectral crosstalk, allowing for measurement of a broad range of target concentrations and for higher data confidence at lower signal levels.

Scanning The laser excitation is scanned rapidly up and down across the microarray. The dynamic autofocus ensures that the microarray is always positioned in the focal plane of the scan lens, resulting in a uniform and calibrated-intensity scan.

Fluorescence detection Fluorescence from the labeled samples is converted to an electrical signal by a high-performance PMT. Very low noise amplifiers and digital integrators process the PMT signal into a digital measurement that is recorded in the TIFF file.

SureScan Microarray Scanner System User Guide 15

Programs Installed on the Computer Workstation

The computer that is included with your SureScan Microarray Scanner System has the following software programs preinstalled.

Agilent Microarray Scan Control program used to set up and operate the scanner.

Agilent Feature Extraction program automatically performs feature extraction using scanned image files created by the scanner. You can also run the Feature Extraction program manually to examine images and set up feature extraction options.

Agilent Installation Qualification Tool program verifies that the Scan Control program was installed correctly and was not corrupted after installation. Produces an Installation Qualification Report for your records.

SureScan Microarray Scanner System User Guide 16

SureScan Microarray Scanner System User Guide

2 Getting Started Operating the Scanner 18

Step 1. Turn on the SureScan scanner and start the Scan Control program 18 Step 2. Insert slides into slide holders 20 Step 3. Load the slide holders into the cassette 22 Step 4. Set or change protocol scan settings 24 Step 5. Add slides to the scan queue 24 Step 6. Scan your slides 25 Step 7. Remove the slides 25

About Scan Protocols 26 Offline Mode 27 Turning Off the SureScan Scanner 28

This chapter describes how to operate the scanner, including how to set up and scan your slides quickly and easily.

If you have any problems, see Chapter 5 for troubleshooting information.

17

Operating the Scanner

The following steps explain how to operate the scanner. For more information on how to use the Scan Control program, see Chapter 3, Using the Scan Control Program,.

Step 1. Turn on the SureScan scanner and start the Scan Control program

1 Turn on the scanner using the power switch on the front of the instrument. After you turn on the scanner, the scanner loads and initializes its firmware. When the scanner is ready (about two minutes), the status indicator light turns off.

2 Turn on the computer workstation and wait for it to boot up. 3 Double-click the Agilent Microarray Scan Control icon to start the Scan Control program.

Or Click Start > All Programs > Agilent > Agilent Microarray Scan Control.

Figure 4 Agilent Microarray Scan Control icon

When the program starts, the Agilent Microarray Scan Control program main window opens and the scanner performs its initialization sequence. See Scan Control main window on page 67. The Scan Control program communicates with the scanner via the LAN cable, sending

commands and parameters, and receiving status and data. The lasers turn on and start to warm up. The autoloader initializes and performs a slide eject cycle (to make sure that no slide is

loaded).

The scanning system is initialized and the data acquisition system is calibrated. After the initialization sequence finishes, the Open Door button is enabled and you can load slides.

NOTE To learn how to set up the lasers to turn off automatically when not in use, see To set the laser saver delay on page 47.

NOTE If the scanner has 24 slides loaded when you turn it on, the initialization will fail because it cannot perform the slide eject cycle.

SureScan Microarray Scanner System User Guide 18

For more information on the Scan Control program window, see Scan Control Program Window Reference on page 67.

Figure 5 Agilent Microarray Scan Control program window ready to add slides

The status of the scanner is indicated at the lower right corner of the Scan Control window, in the status bar.

Initialize The scanner is initializing. When the initialization is finished, the Open Door button is enabled, and the lasers continue to warm up.

WarmUp The lasers typically take less than 30 seconds to warm up. During warm-up, you can load slides, set protocols, and add slides to the queue. Once the lasers are warmed up, you can start scanning.

LasersOff Indicates that the lasers are turned off.

Ready You can load slides or begin a scan.

Error Indicates that the scanner has encountered an error. Close the Scan Control program then shut off the scanner. Restart the scanner and the Scan Control program. If the error continues, contact Agilent technical support.

SureScan Microarray Scanner System User Guide 19

Step 2. Insert slides into slide holders

A slide is inserted into a slide holder before loading it into the scanner. 1 Before you insert the slide, place the slide holder on a flat surface, with the clear cover facing

up, and the tab on the right. This helps to ensure that you have the slide aligned properly when you insert it into the slide holder.

2 Gently push in and pull up on the tabbed end of the clear plastic cover to open it.

Figure 6 Opening the slide holder

Figure 7 Inserting slide into the slide holder

3 Insert the slide into the holder, as follows: a Hold the slide at the barcode end. b Make sure that the active microarray surface faces up, toward the slide cover, with the

barcode on the left.

NOTE You cannot start scanning until both lasers are warmed up, at least one slide is in the queue, and the scanner status is Ready.

NOTE Fingerprints cause errors in the fluorescence detection. For accurate readings, touch only the edges of the slide and always use gloves when handling slides.

Also, do not write on the slides with markers or place any labels on the slide other than an appropriate barcode.

Slide ledge

SureScan Microarray Scanner System User Guide 20

c Carefully place the end of the slide without the barcode label onto the slide ledge. See Figure 7.

d Gently lower the slide into the slide holder. See Figure 8. e Close the plastic slide cover, pushing on the tab end until you hear it click. This moves the

slide into position in the holder. f Gently push in and pull up on the tabbed end of the clear plastic cover to open it again and

verify that the slide is correctly positioned.

Once inserted, the slide lies flat and matches up with the alignment points on the slide holder.

g Close the plastic slide cover, pushing on the tab end until you hear it click. See Figure 9.

Figure 8 Slide inserted in slide holder

Figure 9 Slide holder closed with slide

For instructions on removing the slides, see Step 7. Remove the slides on page 25.

Agilent slides have two barcodes, one on each side of the glass. See Figure 10. Typically, non-Agilent slides have only one barcode. For both Agilent and non-Agilent slides, place the active microarray side of the slide facing toward the slide holder cover.

If you have a slide whose active surface is on the side opposite to the barcode, the scanner cannot read the barcode.

NOTE See Barcode and barcode label specifications on page 92 to apply a second readable barcode.

CAUTION An improperly inserted slide can damage the scanner.

SureScan Microarray Scanner System User Guide 21

Figure 10 Slide orientation

Step 3. Load the slide holders into the cassette When the slides are properly inserted in the slide holders, you can load the slide holders into the cassette. The cassette and slide holders are designed to ensure that the slide holders are inserted correctly.

Figure 11 Slide holder helps you to insert slides correctly

NOTE Do not load slide holders that do not contain slides into the scanner.

Arrow helps identify scanner orientation

Finger-hold

SureScan Microarray Scanner System User Guide 22

1 In the Scan Control program window, click Open Door to open the scanner door.

2 Pick up the slide holder using the finger hold. The arrow on top of the slide holder points to the left when you pick up the slide holder correctly. See Figure 11. Insert a slide holder into any open slot. The slot numbers are clearly labeled on the slide cassette. Do not force the slide holder into the cassette; it inserts easily if properly aligned with the finger-hold on top and the arrow facing to the left.

Figure 12 Inserting slide holder into cassette

3 Make sure that the slide holder is seated in the bottom of the cassette slot. The slot number for the loaded slide blinks blue.

4 Repeat steps 2 and 3 until all slide holders are loaded in the cassette.

The slide numbers next to the cassette and in the slot table of the Scan Control program window change color to indicate the state of the slot. For more information, see Table 8 on page 70.

5 In the Scan Control program, click Close Door. The following events happen: The scanner door closes. The scanner reads the barcode for each slide. The barcode is displayed under Slide ID in the Scan Control software slot table. Default output folder is applied.

CAUTION The correct way to open the scanner door is using the Open Door button in the Scan Control program. Do not attempt to open the door manually.

NOTE The scanner scans slides in the order set in the scan queue. The scanner skips over any empty slots. See To add a slide to the scan queue on page 31.

CAUTION Improper placement of the slide holder in the cassette can result in severe damage to the scanner.

SureScan Microarray Scanner System User Guide 23

For slides that have a scan protocol mapped to their design, the scan protocol is assigned in the Slot Table, and the slot State changes to Ready for queue.

For slides that do not have a scan protocol mapped to their design, the scan protocol remains empty and the slot State remains Present. Assign a scan protocol, as described in Step 4. Set or change protocol scan settings. For more information on the Scan Control program main window, see Scan Control Program Window Reference on page 67.

For information on how to map scan protocols to slide designs, see To map a scan protocol to a slide design on page 49.

You can add slides to the cassette while a scan is in process. See About Adding Slides on page 88.

Step 4. Set or change protocol scan settings The first time you set up to scan a slide, select a scan protocol to use. See About Scan Protocols on page 26. Once the slide is scanned, the program remembers that scan protocol and assigns it to all slides with the same microarray design. You can change these assignments later. You can also manually set scan settings for a selected slide. 1 For each slide in the slot table, click the Scan Protocol and select a scan protocol to use for

scanning the slide. See About Scan Protocols on page 26 and To set or change the scan protocol on page 30.

2 (Optional) For a selected slide, in the scan settings pane, change one or more scan settings to use for scanning only that slide. See Changing Slide Scan Settings on page 34.

The current scan protocol settings are displayed for each selected slide in the right pane of the Scan Control software main window. For more information on these settings, see Scan Control Program Window Reference on page 67.

Step 5. Add slides to the scan queue

1 In the Scan Control main window, click All to Queue to add all slides in the slot table with a State of Ready for queue to the scan queue. A confirmation dialog box appears. Click Yes to add the slides to the queue. Or In the Scan Control slot table, click the State cell for the first slide to scan and click Add to Queue.

2 For each additional slide you want to scan, Click the State cell and select Add to queue first to add the slide to the top of the scan

queue. Or

NOTE Once you add a slide to the scan queue, you cannot change its scan settings. To change the scan settings, remove the slide from the queue.

To add a slide to the scan queue, its State must be Ready for queue.

SureScan Microarray Scanner System User Guide 24

Click the State cell and select Add to queue last to add the slide to the bottom of the scan queue.

As each slide is added to the queue, its State indicates that it is in the queue and the order in which the slide is scanned. (In queue 1, In queue 2, for example.) The status indicator light changes to solid blue.

Step 6. Scan your slides A scan refers to creating an image of the probes on a single slide. 1 If necessary, in the Scan Control main window, click Close Door.

Wait until the door closes and the Start Scan button is enabled. 2 In the Scan Control main window, click Start Scan to begin scanning the slides that were

added to the queue. The scanner scans the slides in their order in the scan queue. See Step 5. Add slides to the scan queue on page 24. During a scan, you see the following: The slot status indicator light for the current slide blinks green during the scan process, and

the scan progress (for example, Scanning 50%) is displayed in the slot State. The remaining scan time and required disk space are displayed at the bottom of the Scan

Control main window. See Scan Control Program Window Reference on page 67. Events during the scan are logged in the Scan Log and Status Log. See Log tabs on

page 74.

Step 7. Remove the slides When the Open Door button is enabled, you can unload the slide holders from the cassette and then remove the slides from the slide holders. If the Open Door button is not available, you cannot open the door. Check to make sure that the scanning process is finished. 1 In the Scan Control main window, click Open Door to open the scanner door. 2 Remove the slide holders from the cassette. 3 Remove the slides from the slide holders, as follows:

a Hold the slide holder on the sides with the Agilent logo facing up. b Gently push in and pull up on the tabbed end of the clear plastic cover to open it. c Push up on the barcode end of the slide from underneath the slide holder to avoid

fingerprints on the sample area. d Grasp the slide from the sides and remove from the slide holder.

SureScan Microarray Scanner System User Guide 25

About Scan Protocols

A scan protocol is a collection of scan and image settings that, when selected, is applied to the slide as it is scanned.

Agilent supplies eight preloaded protocols for your selection and use with Agilent high density (HD) microarrays, Agilent G3 microarrays, and non-Agilent microarrays.

AgilentHD_GX_2Color Agilent HD 2-color gene expression microarrays

AgilentHD_GX_1Color Agilent HD 1-color gene expression microarrays

AgilentG3_GX_2Color Agilent GX 2-color gene expression microarrays

AgilentG3_GX_1Color Agilent GX 1-color gene expression microarrays

AgilentHD_CGH Agilent HD CGH/CNV/ChIP microarrays

AgilentG3_CGH Agilent G3 CGH/CNV/ChIP microarrays

AgilentHD_miRNA Agilent HD miRNA microarrays

AgilentG3_miRNA Agilent G3 miRNA microarrays

For Agilent slides, you select the predefined protocol that applies to your type of Agilent microarray.

For non-Agilent slides it is necessary to change individual scan settings. Or, you can create a protocol from an existing one. For information on how to create a scan protocol or change an existing protocol, see To create a scan protocol on page 37 or To change an existing scan protocol on page 37.

SureScan Microarray Scanner System User Guide 26

Offline Mode

If no instrument is available, the Scan Control program runs in offline mode. In this mode, you can create, import, and export scan protocols and scan regions. You can also open log files, display recent errors, set general settings, and map scan protocols to design IDs.

SureScan Microarray Scanner System User Guide 27

Turning Off the SureScan Scanner

1 In the Scan Control program window, make sure that the scanner is not scanning, ejecting, or loading a slide.

2 Click Open Door to open the scanner door. 3 Remove the slide holders from the scanner cassette. 4 Remove the slides from the slide holders. 5 Click Close Door.

6 In the Scan Control main window, click the red X at the upper right corner to close the program. The lasers are turned off automatically when you close the program.

7 Turn off the power switch on the front of the scanner.

CAUTION You cannot open the scanner door manually. Use the Open Door/Close Door button in the Scan Control program to open and close the door.

SureScan Microarray Scanner System User Guide 28

SureScan Microarray Scanner System User Guide

3 Using the Scan Control Program Using the Slot Table 30

To change a Slide ID 30 To set or change the scan protocol 30 To change the output folder for a scan 31 To apply a selection to multiple slides 31 To add a slide to the scan queue 31 To add all slides to the queue 31 To move a slide in the queue 32 To remove a slide from the scan queue 32 To remove all slides from the scan queue 32 To open the scanner door 32 To close the scanner door 33 To start a scan 33

Changing Slide Scan Settings 34 To change settings for a single slide 34 To change settings for multiple slides 34 To add a description for the slide 35 To add a user name 35

The Scan Control program is used to control all features of the SureScan scanner, including setting and changing scan settings and protocols, starting and stopping scans, reviewing scan status, and troubleshooting.

This chapter describes how you use the Scan Control program to set up and run the SureScan scanner. For information on how to use the Scan Control Tools to change protocols and scan regions, and to turn on and off the lasers, see Chapter 4, Using Scan Control Tools.

29

Using the Slot Table

The slot table provides a display of the cassette and its contents. Once you load slides into the cassette and close the door, the Scan Control software reads the barcode for each slide and shows it as the Slide ID in the slot table. The numbers to the left of the table correspond to the slots in the scanner cassette. The color of the number indicates the status of the slot. For details on the Scan Control main window, see Scan Control Program Window Reference on page 67.

The topics in this section describe how to use the Scan Control slot table to prepare for scanning slides.

To change a Slide ID The Slide ID is used in the image file name. By default, it is the slide barcode read by the scanner. See Barcode and barcode label specifications on page 92 for information on barcodes.

Typically, the scanner automatically reads the barcode for a slide and displays it as the Slide ID in the slot table. To add or change the Slide ID in the table, follow the instructions below. 1 In the slot table, click the Slide ID cell for the slide you want to add or change. Cell editing is

enabled only when the slot has a slide loaded and is not yet in the queue. 2 Type the new slide ID.

To set or change the scan protocol A scan protocol is a predefined set of scan settings. Several default scan protocols are provided with the software. See About Scan Protocols on page 26.

If no scan protocol is selected for a slide, or if you want to change the scan protocol, follow the instructions below. 1 In the slot table, click the Scan Protocol cell for the selected slide and then click again to show

the list of available scan protocols. 2 Click a scan protocol from the drop-down list.

NOTE After you change the Slide ID, move your mouse cursor over the Slide ID cell in the scan table to see the barcode for the slide. The barcode appears in a tooltip.

NOTE You cannot change scan protocols for slides in the queue.

SureScan Microarray Scanner System User Guide 30

To change the output folder for a scan The output folder is where the scanned image files for a slide are saved. By default, the output folder is D:\ScanData. 1 In the slot table, click the Output Folder cell for a slide. 2 Click the browse icon.

The Browse For Folder dialog box opens. 3 Browse to the location where you want to save the scanned image files for this slide, and click

OK.

To apply a selection to multiple slides Within the slot table, you can select more than one slide and then make a selection for Scan Protocol and Output Folder. 1 In the slot table, click to highlight the first slide. 2 To select multiple contiguous slides, hold down the Shift key and then select a second slide.

All slides between and including the selected slides are highlighted in the slot table. Or To select a series of noncontiguous slides, hold down the Ctrl key and then click additional slides you want to select. Selected slides are highlighted in the slot table.

3 Within the last selected slide, select the Scan Protocol or Output Folder cell. The selection is applied to all highlighted slides.

To add a slide to the scan queue 1 In the slot table, click the State cell for the slide you want to add to the queue. 2 Click Add to queue (if no other slides are in the queue).

Or Click Add to queue first to add the slide to the beginning of the queue. Or Click Add to queue last to add the slide to the end of the queue.

To add all slides to the queue 1 In the Scan Control main window, click All to Queue.

A confirmation dialog box opens. 2 Click Yes.

NOTE You cannot make changes to slides in the queue.

SureScan Microarray Scanner System User Guide 31

All slides in the slot table with a State of Ready for queue are added to the queue, in the order they appear in the slot table.

To move a slide in the queue 1 In the slot table, click the State cell for the slide you want to move. 2 Click again to show selections for the slide. 3 Click one of the following possible options to move the slide position in the queue:

Move to first Move the slide to the first position Move to last Move the slide to the last position Move up Move the slide up one position Move down Move the slide down one position

To remove a slide from the scan queue 1 In the slot table, click the State cell for the slide you want to remove from the queue. 2 Click again to show selections for the slide. 3 Click Remove from queue.

The slide is removed from the queue and its State changes to Ready for queue.

To remove all slides from the scan queue 1 In the Scan Control main window, click Empty Queue.

A confirmation dialog box opens. 2 Click Yes.

All queued slides are removed from the queue, and the State changes to Ready for queue.

To open the scanner door In the Scan Control main window, click Open Door.

NOTE You must use the Scan Control program to open the scanner door.

You cannot open the door while the scanner is loading or ejecting a slide.

SureScan Microarray Scanner System User Guide 32

To close the scanner door In the Scan Control main window, click Close Door.

To start a scan In the Scan Control main window, click Start Scan.

The slot status indicator light blinks green during the scan process. The scan progress (for example, Scanning 50%) is displayed in the slot State cell.

NOTE You must close the door before you can start a scan. After the scan begins, you can open the door and add or remove slides.

You must use the Scan Control program to close the scanner door.

SureScan Microarray Scanner System User Guide 33

Changing Slide Scan Settings

When you select a slide in the slot table of the Scan Control main window, the scan settings for that slide are shown in the Settings Pane on the right side of the window. See Scan Control Program Window Reference on page 67. The values displayed are defined in the selected Scan Protocol for that slide.

There are two ways to change scan settings: Make one-time changes to the scan settings for a slide before it is added to the queue. These

instructions are shown in the following sections. Select a scan protocol or create a new one. See To create a scan protocol on page 37.

You can also apply setting changes to multiple slides. For information, see To apply a selection to multiple slides on page 31.

To change settings for a single slide You can only change the scan settings for a slide when it is not in the scan queue. See To remove a slide from the scan queue on page 32. 1 In the scan table, select the slide whose settings you want to change. 2 In the Scan Settings area, click the arrow next to the setting you want to change, and then

select the new setting. When you manually change a setting, the Scan Protocol for the slide changes to .

3 When you are finished changing the settings, click the State cell and add the slide to the queue. See To add a slide to the scan queue on page 31.

For more information on all the settings, see Chapter 6, Reference.

To change settings for multiple slides You can only change the scan settings for a slide when it is not in the scan queue. See To remove a slide from the scan queue on page 32. 1 In the scan table, click to select the first slide whose settings you want to change. 2 Hold down the Ctrl key and then click to select other slides.

Or To select a contiguous block of slides, click to select the first slide, and then hold down the Shift key and then click the last slide.

3 In the Scan Settings area, click the arrow next to the setting you want to change, and then select the new setting. When you manually change a setting, the Scan Protocol for the selected slides changes to .

SureScan Microarray Scanner System User Guide 34

4 When you are finished changing the settings, add the slides to the queue. See To add a slide to the scan queue on page 31 or To add all slides to the queue on page 31.

To add a description for the slide

1 In the slot table, select the slide whose settings you want to change. 2 In the Scan Description area, type the desired information about the slide and scan. 3 When you are finished typing the description, click the State cell and add the slide to the queue.

(See To add a slide to the scan queue on page 31.)

To add a user name You can only change the user name for a slide when it is not in the scan queue. See To remove a slide from the scan queue on page 32. 1 In the slot table, select the slide whose settings you want to change. 2 In the User area, type user information. 3 When you are finished typing the user, click the State cell and add the slide to the queue. (See

To add a slide to the scan queue on page 31.)

NOTE You can only change the slide scan description for a slide when it is not in the scan queue. See To remove a slide from the scan queue on page 32.

SureScan Microarray Scanner System User Guide 35

SureScan Microarray Scanner System User Guide

4 Using Scan Control Tools Creating and Changing Scan Protocols 37 Creating and Changing Scan Regions 40 Adding a Barcode 43 Turning on Lasers Manually 44 Troubleshooting Tools 45 Setting up Scanner Defaults 47 Mapping Scan Protocols to Designs 49

The Tools menu in the Scan Control program provides general settings and functions that help with troubleshooting. Scan control tools let you Create or change scanner protocols Change the scanning region for slides Input barcodes Switch on and off lasers Display recent errors and log files Create a snapshot file of the current scanner state Reset calibration warnings Perform a self test Set general scanner settings Map protocols to slide designs

This chapter describes how to use the tools available in the Scan Control program.

36

Creating and Changing Scan Protocols

The Scan Control program comes with a default set of protocols that are designed to work with typical Agilent microarray slide designs. See About Scan Protocols on page 26.

To create a scan protocol You can create a protocol from any existing protocol by saving the existing protocol with a new name.

1 Click Tools > Scan Protocol Editor. 2 Select an existing protocol that is similar to the protocol you want to create. 3 Click Save As to save the existing protocol with a new name.

The Save As New Name dialog box opens. 4 Type a new name for the protocol, and then click Save.

The scan protocol settings become active. 5 Change the scan and image settings as desired. For information on the available settings, see

Scan Protocol Editor dialog box on page 78. 6 When you are finished, click Save.

To change an existing scan protocol

1 Click Tools > Scan Protocol Editor. 2 Select an existing protocol you want to change. 3 In the Scan Protocol Editor dialog box, change one or more settings. 4 Click Save.

NOTE You cannot modify a protocol that is currently assigned to a slide in the slot table.

NOTE You cannot change the default protocols provided with the scanner. To change one of these protocols, save it with a different name first.

SureScan Microarray Scanner System User Guide 37

To export a scan protocol You can export one or more protocols to a file on your hard disk, as a backup, or to import on another SureScan Microarray Scanner System. 1 Click Tools > Scan Protocol Editor.

The Scan Protocol Editor dialog box opens. 2 Click Export.

The Export Scan Protocol dialog box opens. A list of protocols in the program is displayed. 3 Click to select a protocol to export.

Or To select a series of contiguous protocols to export, click to select a protocol, and then hold down the Shift key and click another protocol. Or Click to select a protocol, and then hold down the Ctrl key and select additional noncontiguous protocols to export.

4 Click Export. The Save As dialog box appears.

5 Browse to the location where you want to save the exported protocols file. 6 Type a name for the exported protocols file, and click Save.

To import a scan protocol 1 Click Tools > Scan Protocol Editor.

The Scan Protocol Editor dialog box opens. 2 Click Import.

The Open dialog box opens. 3 Browse to where the exported protocols file you want to import is located. Exported protocols

files have .exp extensions. 4 Click to select the protocols file, and click Open.

The protocols in the file are imported.

NOTE If a protocol in the file has the same name as an existing protocol, the program does not import it.

SureScan Microarray Scanner System User Guide 38

To remove a scan protocol

1 Click Tools > Scan Protocol Editor. The Scan Protocol Editor dialog box opens.

2 In the Scan Protocol list, select a protocol to remove. 3 Click Remove.

NOTE You cannot remove any of the default protocols or any protocols that are currently assigned to a slide in the slot table.

SureScan Microarray Scanner System User Guide 39

Creating and Changing Scan Regions

The scan region determines the area of the slide that is scanned. The larger the region, the longer the scan time. If you use non-Agilent slides, you can create your own scan region or change a scan region, up to the maximum scan region of 71 mm x 21.6 mm. The new region appears as a selection in the slot table and in the Protocol Editor.

To create a scan region Make sure that the scan region is at least 4 mm away from the barcode label and does not overlap any other opaque or translucent areas of the slide. 1 In the Scan Control program menu bar, click Tools > Scan Region Editor.

The Scan Region Editor opens. 2 In the list next to Scan Region, select New Scan Region.

Or To use an existing scan region as a template, a In the list next to Scan Region, select one of the available scan regions. b Select Save As.

The Save As New Name dialog box opens. c Type the name for the new scan region, and then click OK. The scan region settings become active.

3 Under Scan Region, type the measurements (in mm) for the region. If you type an invalid value, a red box appears around the measurement. The red box at the top of the dialog box shows the scan region currently defined.

4 Click Save. If no errors are found, the Scan Region Editor appears with the new region listed in the Scan Region Editor.

Change an existing scan region for non-Agilent slides You can only change the scan regions that you created. You cannot change or remove the regions provided by Agilent.

When creating a scan region or using existing scan regions, make sure that the scan region is at least 4 mm away from the barcode label. 1 In the Scan Control program menu bar, click Tools > Scan Region Editor.

The Scan Region Editor opens. 2 In the list next to Scan Region, select the scan region you want to modify. 3 Change the measurements for the region, as desired. For more information on the settings

available, see Scan Region Editor dialog box on page 83.

SureScan Microarray Scanner System User Guide 40

4 Click Save to save the changes for the selected scan region.

To export a scan region 1 In the Scan Control program menu bar, click Tools > Scan Region Editor.

The Scan Region Editor opens. 2 Click Export.

The Export Scan Region dialog box opens. 3 Click to select the scan region you want to export.

Or To select noncontiguous scan regions to export, hold down the Ctrl key and then click additional scan regions. Or To select a contiguous set of scan regions to export, click to select the first scan region, and then hold down the Shift key and then click to select the last scan region to export.

4 Click Export. 5 The Save As dialog opens. 6 Browse to the location where you want to save the exported scan region file. 7 In File name, type the name for the exported scan region file. 8 Click Save.

To import a scan region 1 In the Scan Control program menu bar, click Tools > Scan Region Editor.

The Scan Region Editor opens. 2 Click Import.

The Open dialog box opens. 3 Browse to where the exported scan regions file you want to import is located. Exported scan

regions files have .exp extensions. 4 Click to select the scan regions file, and click Open.

NOTE If a scan region in the file has the same name as an existing scan region, the program does not import it.

SureScan Microarray Scanner System User Guide 41

To remove a scan region

1 Click Tools > Scan Region Editor. The Scan Region Editor dialog box opens.

2 In the Scan Region list, select a scan region to remove. 3 Click Remove.

NOTE You cannot remove any of the default scan regions or any scan region that is currently used in a protocol.

SureScan Microarray Scanner System User Guide 42

Adding a Barcode

Barcodes are the means by which microarray slides are identified, both physically and within the Scan Control program. In addition, the barcode is saved in the metadata of the TIFF image, and is displayed in Feature Extraction reports.

To add a barcode If the barcode of a microarray slide is missing or is not readable by the scanner, you can add it manually by either typing it in or using an external barcode reader. To add a barcode manually, at least one slot of the cassette must be available. 1 If the scanner door is not open, in the Scan Control program main window, click Open Door,

and wait for the door to open. 2 If the barcode for a slide already in a slot is unreadable, remove the slide holder that contains

the slide from the cassette. 3 In the Scan Control main window, click Tools > Input Barcode.

The Input Barcode dialog box opens. 4 Enter the barcode using an external barcode reader or type the barcode into the Barcode text

box. 5 Insert the slide holder that contains the slide into the designated slot of the cassette. 6 Click Set.

In the Scan Table, the barcode is displayed in the Slide ID for that slot. The slot State changes to Present.

7 If desired, follow step 3 through step 6 to add another barcode. 8 When finished, click Close.

NOTE By default, the scanner reads the barcode of a slide and displays it as the Slide ID in the Scan Control program Scan Table. If you change the Slide ID, you can still see the barcode of the slide by moving the mouse cursor over its Slide ID. The barcode appears in a tooltip.

SureScan Microarray Scanner System User Guide 43

Turning on Lasers Manually

The lasers are turned on automatically when you start the Scan Control program, or when you add slides to the queue. They turn off automatically, based on the Laser Saver Delay settings. See To set the laser saver delay on page 47. This section describes how to turn on the lasers manually.

To turn on lasers In the Scan Control program window, click Tools > Switch on Lasers.

The lasers are turned on. The Status Log displays Warming up lasers and the status bar displays Warming up. When the lasers are warmed up, the status bar displays Ready.

NOTE Once the lasers are turned on, it takes up to 5 minutes for them to warm up before the instrument is ready to scan

SureScan Microarray Scanner System User Guide 44

Troubleshooting Tools

To help with troubleshooting, you can display recent errors, or open log files that were generated for the scanner. You can also create a file that contains a snapshot of the current state of the instrument.

To display recent errors In the Scan Control program menu bar, click Tools > Show Recent Errors.

The LogMessages.txt file opens in Notepad (or your default text editor program).

To display log files 1 In the Scan Control program menu bar, click Tools > Log Files.

The C:\ProgramData\Agilent\MicroArrayScanner\Logs folder opens with a list of log files: ScanLog.csv contains information about scan activity SysLog- - .csv contains information about system activity

ExceptionLog.txt contains information about special conditions that affect the software execution

2 Double-click to select and open a log file. Files with the .csv (comma-separated variable) extension are opened by default with an available spreadsheet program. Files with the .txt extension are opened by default with an available text editor.

To create a snapshot of the instrument state In the Scan Control program menu bar, click Tools > State Snapshot.

A file is created in the C:\ProgramData\Agilent\MicroArrayScanner\Snapshots\ StateSnapshots folder.

To reset calibration warnings During typical operation, the scanner calibrates the PMTs before every scan. It also calibrates the lasers during system initialization. If the calibration is unsuccessful, or if the calibration changes significantly since the previous time it was performed, the scanner software records this information, and generates warnings in the Scan Log.

NOTE The troubleshooting tools described in this section are typically used when you are working with an Agilent technical support specialist.

SureScan Microarray Scanner System User Guide 45

The PMT calibration warning is set when the PMT gain changes by more than 20% from the previous calibration value.

The laser calibration warnings are set when the lasers cannot achieve their specified power within the warm-up period. If this problem occurs, the system sets the warning, and recalibrates the lasers at 80% of their specified power. The Scan Control program scales the TIFF file to compensate for the lower laser power.

If the next calibration is again unsuccessful, and the warnings reappear, contact your local Agilent sales and support center.

To reset calibration warnings, In the Scan Control menu bar, click Tools > Reset Calibration Warnings.

To run a self test

1 In the Scan Control menu bar, click Tools > Self Test. The Self Test dialog box opens.

2 Click Start. The self test examines various scanner subsystems to check for out-of-specification behavior. After the self test is finished, a summary of the results is opened in your internet browser.

NOTE To run a self test, remove all slide holders from the scanner. The Self Test command is enabled when the scanner is Ready and the door is closed.

The self-test does not test all subsystems or specifications. For a full retest, contact Agilent service for a preventative maintenance and scanner check.

SureScan Microarray Scanner System User Guide 46

Setting up Scanner Defaults

To set the default scan data folder This location is the default output folder where the image files generated by the scanner are deposited. This file is shown by default as the Output Folder in the slot table. You can change the output folder manually for a scan before it is added to the queue. By default, the scan data output folder is D:\ScanData. 1 In the Scan Control program menu bar, click Tools > Settings.

The Settings dialog box opens. 2 Next to Default Scan Data Folder, type the path for the folder where you want to save the scan

images. Or Click Browse and browse to the location where you want to save the scan images, and then click OK.

3 Click Save.

Changes to the scan data folder setting are not applied to slides with barcodes already read by the scanner. To change the default setting for slides already in the slot table, open and then close the scanner door so that the scanner reads the barcodes again.

To set the laser saver delay The laser saver delay is designed to turn off the lasers automatically when not in use, to maximize the lifetime of the lasers.

1 In the Scan Control program menu bar, click Tools > Settings. The Settings dialog box opens.

2 Next to Laser Saver Delay, select a value for the number of minutes the scanner waits after the last scan before it turns off the lasers.

3 Click Save.

To set the location of the Feature Extraction program If you want to run the Feature Extraction program automatically at the end of a scan, you must set the location where the Feature Extraction program is installed. 1 In the Scan Control program menu bar, click Tools > Settings.

The Settings dialog box appears.

NOTE The lasers turn on automatically when you start the Scan Control program, and after you add scans to a queue. You can also turn them on manually. See Turning on Lasers Manually on page 44.

SureScan Microarray Scanner System User Guide 47

2 Click Browse and browse to the folder where the Feature Extraction program is installed, select FENoWindows.exe, and then click OK. By default, this location is C:\Program Files (x86)\Agilent\MicroArray\FeatureExtraction\ FeNoWindows.exe.

3 Click Save.

To select to run Feature Extraction after scanning You can select to run the Feature Extraction program automatically at the end of a scan. For more information, see Settings dialog box General Settings on page 85 and Performing Feature Extraction on Scanned Images on page 89. 1 In the Scan Control program menu bar, click Tools > Settings.

The Settings dialog box opens. 2 Next to Run Feature Extraction after scan is finished, select Yes. 3 Click Save.

SureScan Microarray Scanner System User Guide 48

Mapping Scan Protocols to Designs

The first time you scan a microarray of a given design, the Scan Control program assigns, or maps the selected protocol to that design. After that, any time the Scan Control program recognizes a slide with the same design, the Scan Control program automatically fills in that protocol for the slide in the slot table. You can also assign protocols to slide designs manually.

To map a scan protocol to a slide design 1 In the Scan Control program menu bar, click Tools > Settings.

The Settings dialog box opens. 2 Click the Design to Protocol Mapping tab. 3 Under Design ID, type the Design ID number for the design you want to assign to the protocol.

The Design ID is determined from an Agilent slide barcode. All barcodes start with 25. The following five digits are the Design ID. For example, the Design ID for barcode 251727810298 is 17278.

4 Under Scan Protocol, select a protocol to use for slides for the selected design. 5 (Optional) Under Description, type information about the protocol or design. 6 Click Save.

Whenever you add a slide with the same ID, the program automatically uses the selected protocol in the slot table.

SureScan Microarray Scanner System User Guide 49

SureScan Microarray Scanner System User Guide

5 Maintaining and Troubleshooting Your System Maintaining Your System 51

Scheduled software maintenance 51 Scheduled hardware maintenance 51 Tips to prevent problems 52

Troubleshooting Your System 53 Technical Support 53 Frequently Asked Questions (FAQs) 54 Hardware Troubleshooting 55 Software Troubleshooting 56 System error messages 57

Updating the Scanner Firmware and Scan Control Program 64 To check if an update is available 64 To update the Scan Control program and firmware 64 To verify the software installation 65

This chapter provides maintenance and troubleshooting information for the SureScan Microarray Scanner System.

50

Maintaining Your System

Use proper care and maintenance to keep your SureScan Microarray Scanner System running trouble-free. When in doubt about the use or care of the system, contact your local Agilent sales and support center to answer your questions.

Scheduled software maintenance Perform these tasks to help maintain the performance of your computer workstation and operating system: At least once a week, check disk space and archive data as needed. Once a week delete any temporary files (*.mp, *.tmp files) from the C:\Temp folder.

Scheduled hardware maintenance The SureScan Microarray Scanner System contains no user-serviceable parts.

Inspecting the vents

Good air flow through the scanner is essential to proper operation. Blocked vents can adversely affect performance.

Check all vents weekly to ensure that they are not blocked by dust, debris, furniture, or other instrumentation.

SureScan Microarray Scanner System User Guide 51

Tips to prevent problems Follow these tips to help you maintain the SureScan Microarray Scanner System and its performance.

Tips to avoid data loss

Avoid running software programs that cause high CPU workload, that can affect the acquisition of data during scanning.

Tips to avoid damage to the scanner

Keep liquids and vapors away from the scanner. Never place anything on the scanner or on the scanner door. Minimize and control temperature fluctuations.

Do not place the scanner in direct sunlight. Do not locate the scanner near windows even if they have blinds or window coverings. The hot sun can heat up the scanner housing in a nonuniform fashion, which can cause problems with the alignment of the optics. Scan only when the laboratory temperature is consistent with the operating temperature specifications for the scanner. To assure optimal scanner performance, operate the scanner only in the specified temperature ranges. (See SureScan Scanner Specifications on page 90.)

Control the humidity. The scanner is sensitive to condensing humidity conditions. To ensure optimal performance, operate the scanner only in the specified humidity ranges. (See SureScan Scanner Specifications on page 90.) Always allow 12 hours thermal equilibration time on site before opening the shipping box.

If the power cord needs to be replaced, use a power cord that is appropriately rated.

Tips to maintain hardware performance Avoid moving the scanner.

If you must move the scanner, there is a chance for adverse affects on performance. Call your local Agilent sales and support center for assistance in moving the scanner.

Place the scanner on a sturdy lab bench or table. Avoid leaning on the scanner. To extend the life of the lasers, set up your lasers to turn on and off automatically. See To set

the laser saver delay on page 47. After turn-on, allow time for laser warm-up and stabilization. Typical warm-up time is less than

five minutes. To clean the exterior surfaces of the scanner, wipe down with laboratory wipes moistened

with 70% isopropanol. Do not use acetone or other solvents for cleaning. Do not allow the electronic components of the scanner to become contaminated. For maintenance of the slide holders, see Agilent publication G5761-90004.

WARNING Do not remove the main cover. Do not attempt to repair or gain access to internal components. You risk exposure to high voltage and harmful laser radiation.

SureScan Microarray Scanner System User Guide 52

Troubleshooting Your System

The scanner was designed for low maintenance and high ease of use. If you cannot resolve a problem with the system, read this chapter. If the problem still exists, contact your local Agilent sales and support center.

Technical Support Technical support is available for the SureScan Microarray Scanner System. Read the rest of this chapter before calling your local Agilent sales and support center.

Use the web to reach technical support

The Help menu of the Scan Control program provides a link to Agilent technical support for the scanner. Click Help > Online Support.

Computer workstation support

If you have a problem with your computer workstation, see the documentation that came with the computer.

If you are still unable to resolve the problem, contact your local Agilent sales and support center or click Help > Online Support.

SureScan Microarray Scanner System support

If you have a problem with your SureScan Microarray Scanner System that requires assistance from your local Agilent sales and support center, be prepared to provide the latest diagnostic log files created by the Scan Control program. To open the folder that contains the log files, in the Scan Control program main window, click Tools > Log Files. Log files are located in the folder C:\ ProgramData\Agilent\MicroArrayScanner\Logs.

Each scanner has a unique 10-character serial number. The serial number is located on the front of the instrument at the lower right and on the rear of the instrument.

When corresponding with your local Agilent sales and support center about your SureScan Microarray Scanner System, be sure to include the model number and 10-character serial number of the scanner.

Make a note of the serial number of your scanner, the software version # and the installation date in the spaces shown (if you print the following page) or on a sheet of paper that you keep close to your scanner.

SureScan Microarray Scanner System User Guide 53

Scanner information

Model #:

Serial #:

Installation Date:

Software Version#:

Software Update Version#/Date:

Software Update Version#/Date:

Find the version information for scanner software

1 Click Help > About in the menu bar to find version information. 2 To close the program, click OK.

Frequently Asked Questions (FAQs) The following are frequently asked questions (FAQs) that can help you operate and maintain the SureScan Microarray Scanner System and troubleshoot issues that occur.

Table 1 FAQs

FAQ Answer

I want to move the scanner to another area.

The move can adversely affect scanner performance. Call your local Agilent sales and support center for assistance in moving the scanner and assuring proper operation afterward.

Can I save files over the network while scanning?

Agilent recommends that you save your data files directly to the local hard disk. You can also save data files to a network folder. If a network access problem is experienced during the scan, data is saved to a temporary local folder, and a warning is included in the scan log.

Where do I find support information, such as drivers, guides, and troubleshooting solutions, for my computer workstation?

If you have a problem with your computer workstation, see the documentation that came with the computer. If you are still unable to resolve the problem, contact your local Agilent sales and support center.

Can I open the door to the scanner manually?

No. You must use the Open Door/Close Door button in the Scan Control program to open or close the scanner door.

The scanner is turned on and the Scan Control program is open, but the scanner does not scan.

1 Close and then restart the Scan Control program. You must turn on the scanner before starting the Scan Control program. If you started the Scan Control program first, the connection is not made when the scanner is turned on.

2 Contact your local Agilent sales and support center.

I want to remove a slide from the scanner, but the Scan Control program will not let me open the door.

The Scan Control program prevents you from opening the door while it is loading or ejecting a slide. Wait until the Open Door button is available to open the door. If you continue to have problems, check the Status Log and contact Agilent technical support.

SureScan Microarray Scanner System User Guide 54

Hardware Troubleshooting Except for the power fuse, the SureScan Microarray Scanner Systemhas no user-serviceable parts. The status indicator light on the front of the scanner indicates possible problems. You can also replace the fuses that protect the system. For any other problems, including jams, contact your local Agilent sales and support center.

If you are having a problem with your computer workstation, see Computer workstation support on page 53 for support information.

Troubleshooting with the status indicator light

The front panel has an indicator light that shows the status of the SureScan scanner.

Figure 13 Location of the status indicator light

Table 2 describes the possible states of the indicator light.

Checking and replacing scanner fuses

The scanner has two fuses for the power supply, on the rear of the scanner. The power supply fuses are directly above the power cord plug.

The fuses are ordered directly from Agilent Technologies.

Table 2 Scanner status indicator light states

Light state Meaning/action required

Yellow Scanner is initializing. The yellow light appears after you turn on the scanner, and also when the Scan Control program connects and initializes communication with the scanner.

Off Scanner is fully initialized

Green Scanning is in process.

Red An error has occurred. Check the Scan Control Status log, Scan log, and list of recent errors, and then contact Agilent technical support.

Status indicator light

SureScan Microarray Scanner System User Guide 55

Checking and replacing the power supply fuses If you cannot turn on the scanner, even though the power outlet is active when tested, check, and replace the fuses if needed. 1 Disconnect the power cord. 2 Use a small flat-edge screwdriver to pry up the small plastic tab on the bottom edge of the fuse

holder until it releases. 3 Pull out the fuse holder, and check the fuse integrity. 4 If a fuse is blown, replace the fuse with a T4A, 250 VAC rated fuse (part number 2110-1491). 5 Push the fuse holder back in until it clicks into place. 6 Plug in the power cord.

Software Troubleshooting In case you experience a computer failure or you want to reload the hard drive image that Agilent ships with the computer, Agilent supplies a recovery CD that lets you re-image the hard drive.

WARNING Always disconnect the power cord before checking or replacing the fuses.

CAUTION Replace the fuses with only the same or equivalent rated fuses. If you are unsure about the fuses, contact your local Agilent sales and support center before installing.

Power supply fuse holder

SureScan Microarray Scanner System User Guide 56

File locations The SureScan Microarray Scanner System uses the following folders:

Scan Control program files (installation folder)

C:\Program Files (x86)\Agilent\ScanControl

Log files

C:\ProgramData\Agilent\MicroarrayScanner\Logs

Scanned image files (default otherwise as set in Scan Control Settings)

D:\ScanData

Scanned image files (in case of failure to find a network storage location)

If the Default Scan Data Folder set in Tools > Settings is not available, the program performs the following actions: Posts an error message in the Status Log Clears the Default Scan Data Folder set in Tools > Settings Sets the Output Folder in the Slot Table to C:\ProgramData\Agilent\MicroArrayScanner\Temp

System error messages This section explains how to use error messages and error logs generated by the SureScan Microarray Scanner System.

Hardware monitoring

The scanner continuously monitors internal temperatures and fan speeds, as well as fault conditions on many subsystems. If the monitored parameters reach warning levels, a message is displayed in the status log, and

the instrument goes into a scanning suspended state. In this state, the currently running scan finishes, but no new scan starts.

If the monitored parameters reach alarm levels, the instrument immediately stops scanning, and reduces its power consumption as much as possible.

To display details of the fault that shut down the instrument, in the Scan Control program, click Tools > Show Recent Errors.

SureScan Microarray Scanner System User Guide 57

Where scanner error messages can appear

Error messages appear in the following places: Error messages appear in the Status Log in the Scan Control program main window. If the

error can result in compromised data, it also appears in the Scan Log. Errors are also captured in greater detail in the system log file. System log files are saved as

comma separated value files, with extension .csv. They can be opened with a text editor program (such as Notepad) or with Microsoft Excel. To open the folder where system log files are located, click Tools > Log Files.

To display the most recent errors, click Tools > Show Recent Errors. The recent errors are opened in Notepad.

Troubleshooting with error messages

The SureScan Microarray Scanner System creates error messages to help you solve issues that arise. Many of the error messages include a solution within the text box; follow those instructions.

The following table contains some of the error messages that appear in popup dialog boxes, along with descriptions and suggested actions. If a problem continues after you try the suggested action, contact Agilent technical support.

Table 3 Error messages in popup dialog boxes

Error message Description and suggested action

A slide has been placed in the active slot (Slot number {*}). Remove the slide to allow scanning to continue.

A slide is currently in the active slot and the scanner cannot return the scanned slide to its slot in the cassette. No additional scans can take place, and Close Door is disabled. Remove the slide from the active slot.

Cannot connect to instrument: Firmware version is more recent than host software version. Firmware version: {*} Host software version: {*} Contact Agilent product support.

Instrument firmware does not match the Scan Control program version. Contact Agilent technical support.

Cannot connect to instrument: Verify firmware failed. {*} Contact Agilent product support.

Scan Control cannot communicate with the scanner due to a firmware problem and goes into offline mode. Contact Agilent technical support.

Check disk space failed: {*} Occurs when you try to add a scan to the queue but there is not enough disk space for the scan to complete. Clean up your hard disk to provide adequate space for storage of scan files.

Configuration items are missing from config files. Cannot save.

Reinstall the Scan Control program.

Configuration items missing. Reinstall the Scan Control program.

Error during shutdown: {*} An error occurred when closing the Scan Control program. Restart the Scan Control program and try again.

Find log files failed: {*} Close the Scan Control program and restart it. If the problem continues, reinstall the Scan Control program.

SureScan Microarray Scanner System User Guide 58

* Detail added when the message is generated

Find recent errors failed: {*} Close the Scan Control program and restart it. If the problem continues, reinstall the Scan Control program.

Get 'About' information failed: {*} Close the Scan Control program and restart it. If the problem continues, reinstall the Scan Control program.

Initialize logger failed: {*} Log file is open in another application. For the Exception logger, locate the log file (named ExceptionLog.txt). Rename the file (e.g., ExceptionLog_old.txt) but do not delete it. Then, restart the Scan Control program. A new ExceptionLog.txt file is created.

To avoid re-occurance of this error message, always exit the Scan Control program and allow it to close completely before powering off the scanner.

Instrument is busy: Please wait for the instrument to become idle.

Occurs when you attempt a command that cannot be completed while the instrument is busy. Wait and try the action again.

Instrument self test failed: {*} Reinstall the Scan Control program.

Load application configuration failed: {*}

Scan Control program installation is corrupt. Reinstall the Scan Control program.

Load instrument configuration failed: {*}

Scan Control program installation is corrupt. Reinstall the Scan Control program.

Load scan configurations failed: {*} Scan Control program installation is corrupt. Reinstall the Scan Control program.

Load test script set failed: {*} Reinstall the Scan Control program.

Open Online Support website failed: {*}

Web page is currently unavailable. Check your internet connection. Try again later.

Open Scanner Home Page website failed: {*}

Web page is currently unavailable. Check your internet connection. Try again later.

Open Users Guide '{*}' failed: {*} Close the Scan Control program and restart it. If the problem continues, reinstall the Scan Control program.

Recovering door jam failed: {*} An error occurred when the instrument tried to recover from a door jam. Close the Scan Control software, restart the scanner, and then restart the Scan Control program.

Save instrument state failed: {*} Close the Scan Control program and restart it. If the problem continues, reinstall the Scan Control program.

Show 'About' information failed: {*} Close the Scan Control program and restart it. If the problem continues, reinstall the Scan Control program.

Update firmware failed: {*} Update of the instrument firmware failed. Contact Agilent technical support.

Table 3 Error messages in popup dialog boxes (continued)

Error message Description and suggested action

SureScan Microarray Scanner System User Guide 59

The following table shows error messages that appear in the Status Log or Scan Log. When an error occurs, to display additional information, click Tools > Show Recent Errors.

Table 4 Error message in logs

Error message Description and suggested action

{*} State Machine unknown error during state '{*}': {*}.

Low-level error. Contact Agilent technical support.

Activity EjectSlide failed. Multiple possible causes. Try restarting the scanner and Scan Control program. If the problem persists, contact Agilent technical support.

Activity InitLoader failed Multiple possible causes. Try restarting the scanner and Scan Control program. If the problem persists, contact Agilent technical support.

Activity InitStages failed Multiple possible causes. Try restarting the scanner and Scan Control program. If the problem persists, contact Agilent technical support.

Activity LaserWarmup completed with warnings

Lasers failed to warm up at calibrated power.

Activity LaserWarmup failed. Lasers did not stabilize within a certain time.

Activity LoadSlide failed Multiple possible causes. Try restarting the scanner and Scan Control program. If the problem persists, contact Agilent technical support.

AutoFocus hold percentage is more than warning limit

Dirt or obstructions on the slide; slide not installed correctly in slide holder.

Data system calibration completed with warnings

Indicates data system calibration failure. If the problem persists, call service.

Data system calibration failed Indicates data system calibration failure. If the problem persists call service.

Default scan data folder verification '{*}' failed: {*}

The default folder has been removed or cannot be accessed. Recreate it or fix the network connection.

Eject failed: unable to move slide into cassette.

Restart the scanner to attempt to clear fault.

Failed to create FE project file This message indicates FE project file creation failed. Check to make sure the Feature Extraction program is set up properly.

Failed to start FE application This message indicates the program failed to launch the Feature Extraction application. Make sure the Feature Extraction program is installed properly.

Fanspeed error detected: Instrument operation halted

Hardware problem contact Agilent technical support.

Fanspeed warning cleared: Instrument operation resuming

Hardware problem contact Agilent technical support.

Fanspeed warning detected: Instrument operation suspended.

Hardware problem contact Agilent technical support.

Find focus failed Multiple possible causes. Try restarting the scanner and Scan Control program. If the problem persists, contact Agilent technical support.

General communication failure. Check the LAN cable. Restart the scanner and the Scan Control program.

Green Laser power is not set to calibrated value

Laser power has been reduced. Compensation was applied to the output TIFF. The green laser may fail soon.

SureScan Microarray Scanner System User Guide 60

* Detail added when the message is generated

If an error message does not appear in the table

This table does not list all the possible error messages. If you have an error message that is not listed and you are unable to resolve the problem, do the following: 1 Write down the error message. 2 Restart the Scan Control program.

Green PMT calibration completed with warnings

Indicates PMT calibration completed successfully, but some of the values it calculated are not good.

Hardware error detected in subsystem '{*}':Instrument operation halted

Low-level error. Contact Agilent Technical support.

Hardware warning detected in subsystem '{*}':Instrument operation suspended

Low-level error. Check recent errors and contact Agilent technical support.

High temperature error detected: Instrument operation halted

Check ventilation slots.

High temperature warning cleared: Instrument operation resuming

Check ventilation slots.

High temperature warning detected: Instrument operation suspended

Check ventilation slots.

PMT calibration failed Indicates PMT calibration failure. If the problem persists, call service.

Red Laser power is not set to calibrated value

Laser power has been reduced. Compensation was applied to the output TIFF. The red laser may fail soon.

Red PMT calibration completed with warnings

Indicates PMT calibration issues. If the problem persists, call service.

Scan slide failed Multiple possible causes. Click Tools > Show Recent Errors for more information. Try restarting the scanner and the Scan Control program. If the problem persists, contact Agilent technical support.

State machine failure {*} Low-level software error. Contact Agilent technical support.

Status communication failure Check connection to instrument.

Suspending autoloader operation due to autoloader errors

Click Tools > Show Recent Errors for more information.

Unable to access folder \"{*}\": Saving output file in folder \"{*}\"

Destination folder was not available during the scan.

Watchdog communication failure: {*}. Low-level error. If the problem recurs, call Agilent technical support.

Table 4 Error message in logs (continued)

Error message Description and suggested action

SureScan Microarray Scanner System User Guide 61

3 If step 2 does not solve the problem, do the following: a Close the Scan Control program. a Restart the computer workstation. b Turn off the scanner, and then back on. c Restart the Scan Control program.

4 If step 3 does not solve the problem, contact your local Agilent sales and support center.

SureScan Microarray Scanner System User Guide 62

Using the Diagnostic Display

The scanner diagnostic display is used for advanced troubleshooting tasks. This display is located on the front of the instrument, behind the upper front cover. To open the cover, grasp the finger holds on the sides of the cover, and pull forward. You see the diagnostic display and a 4-way control switch.

Figure 14 Diagnostic display location

The diagnostic display control switch has the following capabilities: Before Scan Control connects to the instrument, the display shows the IP address of the

scanner. If the IP address is not displayed, the firmware is not running. Toggle the switch Up to cycle through a menu, with the following choices:

The first item lets you reset the IP address to factory default (10.0.0.2). The second item lets you reboot the firmware. (The firmware is also reset by power cycling

the instrument.) To perform the selected operation, move the switch to the right (to the Select position). Once the Scan Control program connects to the instrument, the display reads Client

Connected.

NOTE Use the diagnostic display switch only when requested by Agilent technical support.

Upper front cover

Finger holds

4-way control switch

Diagnostic display

SureScan Microarray Scanner System User Guide 63

Updating the Scanner Firmware and Scan Control Program

Agilent Technologies occasionally makes software updates available. Firmware updates (if necessary) are included with the Scan Control program update. This section describes how to update the scanner program and firmware.

To check if an update is available If an update for the Scan Control program is available, an Information icon appears at the lower right of the Scan Control main window: Click the icon to display information about your installed version and the most recent version available.

Figure 15 Update notification dialog box

To update the Scan Control program and firmware 1 From the Scan Control main window, click Help > Scanner Home Page.

The Agilent Technologies Genomics High Resolution Scanner Overview web page opens. 2 In the web page, click Download Software. 3 Follow the instructions to read the Release Notes and Installation Notes 4 Click Download Software to download the software installer and save it to your computer. 5 Start the software installer and follow the prompts to install the software. Accept the defaults.

It is not necessary to remove the previous version of the software.

6 When the software installation is finished, start the Scan Control program. 7 If a firmware update is needed, a message appears, and the Scan Control program changes to

Offline Mode. 8 Close the Scan Control program.

NOTE Software updates do not overwrite scan regions, protocols, or the calibration of the instrument

SureScan Microarray Scanner System User Guide 64

9 Turn off the power to the scanner. 10 Wait 10 seconds, and then turn on the scanner power. 11 Start the Scan Control program.

The Scan Control program and scanner firmware are now updated. If you have problems, contact Agilent technical support.

To verify the software installation The SureScan Microarray Scanner System workstation includes an installation qualification tool (IQT). Use this program after updating your software to verify that the update installed correctly. 1 Click Start > All Programs > Agilent Technologies > Installation Qualification Tool.

Figure 16 Agilent Installation Qualification Tool dialog box

2 Mark the check box next to Agilent Microarray Scan Control. 3 Under Products to be qualified, click Qualify.

The installation is verified, and an installation qualification report is generated. If you selected Open report after qualification, the installation qualification report opens in your internet browser.

4 When finished, click the close button in the upper right corner of the program dialog box.

SureScan Microarray Scanner System User Guide 65

SureScan Microarray Scanner System User Guide

6 Reference Scan Control Program Window Reference 67 Scan Control Program Dialog Box Reference 75 About Adding Slides 88 Performing Feature Extraction on Scanned Images 89 SureScan Scanner Specifications 90 Slide Specifications 92 Regulatory Information 95

This chapter includes descriptions of the Microarray Scan Control program windows and dialog boxes. It also contains specifications and regulatory information.

66

Scan Control Program Window Reference

This section describes the main window of the Microarray Scan Control program and its contents.

Scan Control main window

Figure 17 Agilent Microarray Scan Control main window

The Microarray Scan Control window appears when you start the Microarray Scan Control Program. It has the following features:

Menu bar

Slot table

Function buttons

Settings pane

Logs

Status bar

Slot status indicators

Image processing status

Table 5 Scan Control window features

Feature Description

Menu bar Open tools menu and help.

Slot table Display status, scan protocols, and output folders for microarray slides currently in the scanner.

SureScan Microarray Scanner System User Guide 67

Tools menu

Figure 18 Tools menu

The following functions are available on the Tools menu:

Settings pane Display settings for selected slide. For slides not in the queue, you can change settings from here.

Slot status indicators Indicates the status of the slot. Matches the slot indicator lights on the cassette.

Function buttons Add and remove slides from the scan queue Open and close the scanner door Start and stop a scan

Logs Display instrument and scan status logs.

Status bar Show status of scanner, remaining scan time, and disk space required for the scan.

Table 5 Scan Control window features

Feature Description

Table 6 Scan Control Tools commands

Tool Description

Scan Protocol Editor Opens the Scan Protocol Editor dialog box, where you can create, change, or remove scan protocols.

Scan Region Editor Opens the Scan Region Editor dialog box, where you can create, change, or remove custom slide scan regions.

Input Barcode Lets you use the keyboard or a keyboard emulation barcode reader to enter a barcode for a slide that does not have a barcode, or a when the scanner cannot read a barcode.

Switch on Lasers If the lasers are off, use this command to turn on the scanner lasers.

Show Recent Errors Opens notepad (or your default text editor) and displays details of the most recent errors.

SureScan Microarray Scanner System User Guide 68

Help menu

Figure 19 Scan Control Help menu

The following commands are available in the Help menu:

* The first time you open the Scan Control software, you are prompted to select your preferred version of the User Guide. Until the preferred version is selected, the User Guide option on the Help menu is unavailable for selection. You can changethe selected User Guide at any time from the Settings dialog box (see Settings dialog box General Settings on page 85).

Log Files Opens the Logs folder, where you can open any of the logs created by the program.

State Snapshot Creates a file that contains the status of the scanner at the time the snapshot was created. This file is helpful in troubleshooting.

Reset Calibration Warnings Laser calibration warnings are set when the lasers cannot achieve their specified power within the warm-up period. If this problem occurs, the system sets the warning, and recalibrates the lasers at 80% of their specified power. Use this function to reset the warnings to default settings.

Self Test The self test examines various scanner subsystems to check for out-of-specification behavior. After the self test is finished, a summary of the results is opened in your internet browser.

Settings Opens the Settings dialog box, where you can set defaults and map scan protocols to microarray designs.

Table 6 Scan Control Tools commands (continued)

Tool Description

Table 7 Scan Control Help commands

Item Description

Users Guide* Opens this guide in Adobe Reader.

Scanner Home Page Opens the Agilent Technologies website for the SureScan Microarray Scanner System in your internet browser.

Online support Opens the Agilent Technologies Technical Support web page, where you can find support information for your scanner.

About Agilent Scan Control Displays version information for the Scan Control program and the serial number and model of your scanner.

SureScan Microarray Scanner System User Guide 69

Slot table

Figure 20 Scan Control program slot table

The slot table is a virtual representation of the contents of your scanner cassette. It is used to set up, start/stop, and monitor the progress of scans.

Slot Status Indicator

Numbers to the left of the slot table are identical to the slot status indicator lights on the cassette. The color of the number changes to indicate the status of the slot and scan, as described in Table 8 on page 70.

Slide ID The Slide ID used to name the scanned image file. After you load slides into the scanner cassette and close the scanner door, the program reads the barcode label for each slide in the cassette and displays it in this column. You can change the Slide ID to any text acceptable in a file name. The barcode for the slide is retained, and is shown when you move the mouse over its Slide ID in the Scan Table.

State Displays the current state of the slot. Possible states and the color of the slot status indicator are shown in the following table.

Table 8 Slot states and indicators

Slot status indicator State Meaning

Off Empty No slide is present in the cassette.

Blinks blue Present Slide is present in the cassette slot. When barcode is read successfully, it appears in Slide ID. Slide is not ready to add to a queue because it has no scan protocol assigned yet.

Blinks blue Ready for queue The slide is ready to add to a scan queue.

SureScan Microarray Scanner System User Guide 70

Menu selections for State

For slides that are not scanning, commands are available that let you add or remove a slide from the queue. The selections available for a particular slide vary depending on the location of the slide in the queue, or if the slide is ready to add to the queue.

Scan Protocol Displays the scan protocol to use for scanning the selected slide. Available scan protocols include the default Agilent-supplied scan protocols and any scan protocols that were created or imported. See About Scan Protocols on page 26.

Solid blue In queue x Slide is in the scan queue, in position x, where x indicates the order in which the slides are scanned.

Blinks green Scanning (x%) Slide is in the process of scanning, where x% indicates the percent of completion for the scan.

Solid green Complete Scan finished successfully.

Yellow Warning A warning was generated during the scan.

Red Error An error occurred during the scan.

Blinks yellow Removed A slide that was Ready or In Queue was removed from the cassette.

Blinks yellow Replaced A slide was placed into a slot whose state was Removed. If the barcode matches the original slide that was removed, the State is

changed back to Ready or In Queue. If the barcode does not match the slide that was removed, the Status

changes to Ready if a protocol is mapped to the slide Design ID. It changes to Present if no protocol is mapped to the slide Design ID.

Table 8 Slot states and indicators (continued)

Slot status indicator State Meaning

Table 9 Menu selections for State

Selection Description

Move to First Moves the slide to the first position in the queue.

Move to Last Moves the slide to the last position in the queue.

Move up Moves the slide one place up in the scan queue.

Move down Moves the slide one place down in the slide queue.

Remove from queue Removes the slide from the scan queue and sets the State to Ready for Queue.

Add to queue Available if no slides are in the scan queue. Adds the selected slide to the scan queue.

Add to queue first Adds slide to the first position in the queue. If a scan is already in process, the slide becomes the first slide to scan after completion of the current scan.

Add to queue last Adds the slide to the last position in the queue.

SureScan Microarray Scanner System User Guide 71

Output Folder Displays the folder where image files created by the scanner are saved. By default, this location is D:\ScanData. You can change the default output folder in Tools > Settings. You can change the output folder for a slide before it is added to the queue. The Browse button lets you select a folder to store the data from each scan. Agilent recommends that the data be acquired to a local folder on a secondary hard drive. You can also select a network folder. If a network access problem is experienced during the scan, data is saved to a temporary local folder, and a warning is included in the scan log.

Function buttons Buttons next to the slot table are available depending on the instrument status.

All to Queue Adds all slides not currently in the slide queue to the slide queue. Slides are added to the queue in the order they appear in the slot table.

Empty Queue Removes all slides from the queue, except those currently scanning.

Open Door/ Close Door

Opens or closes the door on the scanner.

Start Scan/ Stop Scan

Starts or stops the scan. Slides are scanned in the order they appear in the scan queue.

SureScan Microarray Scanner System User Guide 72

Settings pane

Figure 21 Scan Control settings pane

The settings pane lets you change individual settings for a selected slide. To change settings, the slide must not be in the scan queue. For more information on the settings available in this table, see Scan Protocol Editor dialog box on page 78.

Scan Settings Displays scan settings from the assigned scan protocol. To change a setting, click next to the setting name and select a new value from the list. For more information, see Scan Settings on page 79.

Image Settings Some data analysis programs have specific requirements for the images. This section lets you change how the image is created from the scan. For more information, see Image Settings on page 81.

File Naming Settings

Shows selections for how the program names scan files. For more information, see File Naming Settings on page 81.

Scan Description

In this area, you can type information about the microarray slide or scan. This information becomes part of the file information that can be displayed in the Feature Extraction program.

User In this area, you can type information about the operator who set up and performed the microarray scans. This information becomes part of the file information that can be displayed in the Feature Extraction program.

SureScan Microarray Scanner System User Guide 73

Log tabs The software documents instrument and scan status in log files that are saved in the C:\ ProgramData\Agilent\MicroArrayScanner\Logs folder. These logs are also displayed in the Log tabs at the bottom of the Scan Control program window.

Status Log tab

Figure 22 Status Log tab

Displays information about the status of the instrument.

Scan Log tab

Figure 23 Scan Log tab

Displays information about the scans. When you start the Scan Control program, the Scan Log displays the scans from the previous 30 days.

Right-click in one of the log tabs to open a shortcut menu with the following options:

Table 10 Shortcut menu options for log tabs

Menu command Description

Clear Clears the contents of the tab. The contents of the log file is not affected.

Copy Active after you hold down the mouse button and drag to select a portion of the log. Copies the selected region to the Clipboard. You can paste the selection into a text editor or program of your choice.

Select all Selects all of contents of the log tab.

Auto Scroll Turns auto scrolling within the log on or off. If autoscroll is on, when a new message appears, the program automatically scrolls to the bottom of the log so you can see it easily. If autoscroll is off, the pane does not scroll when new messages appear. Turning off autoscroll is useful if you want to review the log while the scanner is active.

SureScan Microarray Scanner System User Guide 74

Scan Control Program Dialog Box Reference

This section contains descriptions of the parameters available in the dialog boxes that appear when you use the Scan Control program. The dialog box descriptions appear in alphabetical order.

Agilent Installation Qualification Tool

Figure 24 Installation Qualification Tool dialog box

Purpose: Verifies that the Scan Control program was installed correctly and generates an Installation Qualification report.

To open: In the Windows Start menu, click All Programs > Agilent Technologies > Installation Qualification Tool.

Show OK files in report

When selected, the qualification report includes a list of all files verified as OK. (Default is not selected. Invalid files are always shown.)

Open report after

qualification

When selected, the qualification report opens in your web browser, after the installation qualification is finished.

Products to be qualified

Displays a list of Agilent software products that you can qualify with the tool.

Qualify Starts the installation qualification for the selected product.

Re-Qualify Appears after installation qualification. Lets you requalify the installation. Requalify after you correct any problems, to generate a new installation qualification report.

Report saved at

Appears after installation qualification is finished. Displays a link to the location of the qualification report. Click the link to open the report in your web browser.

Opens the IQT reports folder.

SureScan Microarray Scanner System User Guide 75

Export Scan Protocol dialog box

Figure 25 Export Scan Protocol dialog box

Purpose: Lets you select available scan protocols to export.

To open: In the Scan Protocol Editor dialog box, click Export.

Export When one or more scan protocols are selected, this command opens the Save As dialog box, where you select a location and file name for the exported protocols file.

Close Closes the dialog box.

SureScan Microarray Scanner System User Guide 76

Export Scan Region dialog box

Figure 26 Export Scan Region dialog box

Purpose: Displays available scan regions that you can select to export.

To open: In the Scan Regions Editor dialog box, click Export.

Export When one or more scan regions are selected, this command opens the Save As dialog box, where you select a location and file name for the exported scan regions file.

Close Closes the dialog box.

Input Barcode dialog box

Figure 27 Input Barcode dialog box

Purpose: Used to type or enter a barcode for a slide without a barcode or whose barcode is unreadable by the scanner.

To Open: In the Scan Control program, click Tools > Input Barcode.

SureScan Microarray Scanner System User Guide 77

Barcode The barcode you enter using an external barcode reader or your keyboard.

Set After the barcode is entered and the slide is loaded into slot 1 of the scanner, this button is used to assign the barcode to the slide in slot 1.

Close Used to close the dialog box.

Scan Protocol Editor dialog box

Figure 28 Scan Protocol Editor dialog box

Purpose: Used to create or change scan protocols.

To open: In the Scan Control program window, click Tools > Scan Protocol Editor.

Scan Protocol Displays a list of available scan protocols. The settings for the selected scan protocol are displayed in the Scan Protocol Editor dialog box.

General

Agilent Defined

A protocol provided by Agilent.

SureScan Microarray Scanner System User Guide 78

Locked When a protocol is locked, it cannot be changed.

Scan Settings

Dye channel Determines whether only red (for example, Cy-5 dye), only green (for example, Cy-3 dye), or both dye channel information is gathered. The selection has no effect on the scan time, but selecting only one dye channel does reduce file size up to a factor of 2.

Scan region The scan region determines the area of the slide that is scanned. It must be large enough to capture the entire print region of the microarray. It must be small enough to avoid scanning too close to the barcode or other nontransparent border areas of the slide and affecting the ability of the scanner to auto focus properly. Minimizing the scan region also reduces scan time and save storage space.

Agilent provides a scan region suitable for all Agilent High Density and G3 slides, and another suitable for scanning full 1 x 3 slides without barcode labels.

For non-Agilent slides, you may need to use a custom scan region that better matches the printed region. See Creating and Changing Scan Regions on page 40.

Resolution Sets the scan resolution (pixel size) to 2, 3, 5 or 10 microns.

For 10-micron scans, each row of pixels in the TIFF image represents the average of two scan lines, one acquired in each direction. For 5-, 3- and 2-micron scans, you can select double-pass scanning to perform this averaging, or acquire only a single scan line for each image row.

A high-sensitivity scan mode is also available at all scan resolutions. The high-sensitivity mode provides sensitivity comparable to the double-pass mode but with scan times reduced by 25%. The high-sensitivity mode can be selected in the Resolution pull-down menu of the Scan Protocol Editor.

TIFF Dynamic Range

Sets the dynamic range to 16-bit or 20-bit. The G4900DA scanner has extended the dynamic range of the PMT and signal processing electronics. When you select the 20-bit TIFF file option, you can access this extended range, enabling quantitation of high and low signal features in a single scan.

The Scan Control software still enables conventional XDR 16-bit scanning, as described below.

Table 11 shows the storage space and scan time for single and double pass scans for each resolution selected, for either a 16-bit TIFF dynamic range or a 20-bit range. The Scan Region is 61 X 21.6 mm.

The status bar at the bottom of the Scan Control main window shows an estimate of the storage space and run time required for the current queue of slides.

NOTE If 20-bit option is selected, no XDR option is allowed; if an XDR option is selected already, it is reset automatically.

Table 11 Storage space and scan time

Resolution Storage Space, Mb, 16-bit

Storage Space, Mb, 20-bit

Scan Time, min.

2-micron single pass 1300 1600 24

3-micron single pass 620 760 16

SureScan Microarray Scanner System User Guide 79

Double pass scans do not require more storage space than single pass scans, but they take twice as long to finish.

Red PMT Sensitivity (%)

and Green PMT

Sensitivity (%)

Sets the sensitivity level of the red channel and green channel PMTs. The PMT detects fluorescence emitted by the microarray.

The default output level (100%) sets the gain to the factory-calibrated level; this setting is recommended for Agilent microarrays. You can reduce each color channel setting independently to as low as 1%.

If a microarray is so bright that the upper end of the output signal is saturated, the PMT sensitivity level can be lowered to a sensitivity range that allows all the information to be read.

XDR Ratio Before 20-bit TIFF file dynamic range was available, the eXtended Dynamic Range function was used on both the scanner and Feature Extraction program to capture all the data scanned between very low signal features and very high signal features. Now, instead of using XDR, you can select the 20-bit TIFF file dynamic range to capture the wide dynamic range of data.

With the XDR function turned on, the scanner automatically scans the same slide twice, at two different PMT sensitivity levels. It generates two linked images that Feature Extraction coprocesses to yield a single unified set of extracted intensity data. This data covers up to 6+ orders of magnitude.

The Red or Green PMT Gain sets the high sensitivity level. The XDR function is automatically turned off at values lower than 10%.

The XDR ratio setting instructs the scanner to perform a second scan at lower PMT settings. For example, if the Red and Green PMTs are set to 100%, and the XDR ratio is set to 0.1, then the second scan is performed with the PMTs set to 10%. Available values are: , 0.5, 0.2, 0.1, and 0.05.

5-micron single pass 200 300 10

10-micron single pass 52 115 10

2-micron double pass 1300 1600 46

3-micron double pass 620 760 31

5-micron double pass 200 300 19

2-micron high-sensitivity 1300 1600 36

3-micron high-sensitivity 620 760 24

5-micron high-sensitivity 200 300 15

10-micron high-sensitivity 52 115 15

Table 11 Storage space and scan time (continued)

Resolution Storage Space, Mb, 16-bit

Storage Space, Mb, 20-bit

Scan Time, min.

NOTE If 20-bit option is selected, no XDR option is allowed.

SureScan Microarray Scanner System User Guide 80

Image Settings

Transform Image

Some analysis programs require data from one-color images rotated 90 degrees. If you select Flip/Rotate, the image is transformed as shown in Figure 29.

Figure 29 Image with Flip/Rotate option set

Split When you select Yes, the color file is split into two color files. The split files now have the names of FileName_green and FileName_red, where FileName is the name that was automatically assigned to the file before it was split.

This option is available only for 16-bit TIFF dynamic range scans.

Compress When you select Yes, the program reduces the final amount of storage space occupied by scan images by compressing the TIFF files. The compression algorithm used, LZW, reduces the storage space by 20 to 70 percent.

File Naming Settings

Field 1, Field 2, and Field 3

Displays the settings used for naming TIFF image files created by the scanner. Choices are shown in Table 12.

Scan files are named using the following rules.

For standard scans, the Scan Control program uses up to three user-defined name prefixes to compose the file name. These prefixes are defined in the scan protocol.

Field1_Field2_Field3_ScanNumber.tif

For XDR scans, an additional segment (either _H or _L) is added to the file name to distinguish the XDR Hi image from the XDR Lo image:

Field1_Field2_Field3_ScanNumber_H.tif Field1_Field2_Field3_ScanNumber_L.tif

NOTE If a single dye channel, 20-bit scan, or any XDR option is selected, no Split option is allowed; if the Split option is selected already, it is reset automatically.

Table 12 Choices for file naming fields

Choice Description

The field is not included in the image file name.

Instrument SN Includes the serial number of the scanner in the image file name.

Slide ID Includes the slide identification number (barcode) in the image file name.

Scan DateTime Includes the date and time of the scan in the image file name.

Customize Lets you type custom information to include in the image file name. After you select Customize, double-click Field X. (Where X = the field number 1, 2, or 3.) In the adjacent box, type the custom information to include in the file name.

SureScan Microarray Scanner System User Guide 81

The Scan Control program automatically assigns the Scan Number. The program compares the file name of a new scan with file names in the selected data folder. If Field1_Field2_Field3 is unique, the scan number is set to S01. If a match is found, the scan number is increased until the file name is unique.

Example

US4510PP02_251485023883_S03.tif Instrument Serial # = US4510PP02 Slide ID = 251485023883 ScanNumber = S03. Indicates the third scan file in the folder with the same Instrument Serial #

and Slide ID.

Image File Info

This section is a read-only area that displays information about the image file name, geometry of the slide, disk space required for the file, and estimated time to finish the scan.

SureScan Microarray Scanner System User Guide 82

Scan Region Editor dialog box

Figure 30 Scan Region Editor dialog box

Purpose: Lets you adjust or define the area of the slide that is scanned.

To open: In the Scan Control program menu bar, click Tools > Scan Region Editor.

General

Agilent Defined

A protocol provided by Agilent.

Locked When a protocol is locked, it cannot be changed.

Scan Region

TopLeft X X-axis measurement for the upper left corner. Type this measurement and that of the Y-axis in mm to position the region on the slide.

TopLeft Y Y-axis measurement for the upper left corner.

Width Width of the scan region measured from the end of the x-axis measurement in the upper left corner.

SureScan Microarray Scanner System User Guide 83

Height Height of the scan region measured from the end of the Y-axis measurement in the upper left corner.

Chip Package

A chip package describes the maximum size of the scan region for slides of a designated type. This read-only area displays the default scan region for the selected chip package. There are two types of chip packages; Full Agilent for Agilent slides with barcode labels, and Full Standard for slides without barcode labels.

Save Saves the current scan region values in the current slide scan region. This button is unavailable if the scan region is an Agilent defined scan region or if it is used in an existing scan protocol.

Save As Opens the Save As New Name dialog box, where you can save the current scan region with a new name.

Remove Removes the selected scan region. This button is unavailable if the scan region is an Agilent defined scan region or if it is used in an existing scan protocol.

Import Opens the Open dialog box, where you select an exported scan region file to import to the program.

Export Opens the Export Scan Region dialog box, where you select one or more scan regions to export.

Close Closes the dialog box.

Self Test dialog box

Figure 31 Self Test dialog box

Purpose: Examines various scanner subsystems to check for out-of-specification behavior. After the self test is finished, a summary of the results is opened in your internet browser. The test results are also saved in the C:\ProgramData\Agilent\MicroArrayScanner\SelfTestReport folder.

To open: In the Scan Control program menu bar, click Tools > Self Test.

SureScan Microarray Scanner System User Guide 84

Start/Stop Starts or stops the self-test.

Show Last Results

If more than one self test was run without closing this dialog box, this command opens your internet browser with the results.

Close Closes the Self Test dialog box and reinitializes the scanner.

Settings dialog box General Settings

Figure 32 Settings dialog box General Settings tab

Browse For settings that require a folder location, lets you browse to the folder and select it, rather than typing the path.

Save Saves the settings. If you change settings and want to save them, select this button before you close the dialog box.

Close Closes the dialog box without saving changes.

Timeouts

Laser Saver Delay

When no scan or scan queue is running, the lasers automatically turn off after this amount of time (in minutes).

Data

Default Scan Data Folder

Displays the folder where images created by the scanner are stored by default. You can change this folder for a scan in the Scan Table.

SureScan Microarray Scanner System User Guide 85

Feature Extraction

Location of Feature

Extraction

The path to the FENoWindows.exe program file. By default, this location is C:\Program Files (x86)\Agilent\MicroArray\FeatureExtraction\FeNoWindows.exe.

Run Feature Extraction

when scan is complete

When selected, slides are feature extracted automatically after they are scanned. To run Feature Extraction automatically, the design files for all images must already be present in the Feature Extraction database.

Settings dialog box Design To Protocol Mapping

Figure 33 Settings dialog box - Design To Protocol Mapping tab

Purpose: Used to assign default scan protocols to microarray slide Design IDs. Whenever the scanner recognizes a slide that has a scan protocol mapped to its design, the program automatically fills in the mapped scan protocol in the Slot Table.

To open: In the Scan Control program menu bar, click Tools > Settings and then click Design To Protocol Mapping.

Design ID For Agilent microarray slides, you can determine the Design ID from the barcode. All barcodes start with 25; the next five digits represent the Design ID. For example, the Design ID for barcode 251727810298 is 17278.

NOTE To use Agilent CytoGenomics in auto-processing mode, do not select Run Feature Extraction when scan is complete. (The Agilent CytoGenomics program is not included with the SureScan Microarray Scanner System.)

SureScan Microarray Scanner System User Guide 86

Scan Protocol The scan protocol that is assigned to the Design ID. This scan protocol is automatically assigned in the Slot Table whenever the scanner recognizes a slide with the associated Design ID.

Description An area where you can type information about the mapped scan protocol.

SureScan Microarray Scanner System User Guide 87

About Adding Slides

You can add slides to the scanner even when it is scanning. Use the following guidelines when adding slides to the scanner. You can add slides (in slide holders) to the cassette even when scanning is in process. If a slide

is actively loading or unloading, you cannot open the door. Wait approximately 30 seconds for the loading or unloading process to finish.

When the door is open, the scanner waits to eject a slide that is currently scanning. If the door is left open with no activity for 5 minutes, a message appears to warn you that the

door is about to close, and then the door closes automatically. If you place a slide into the slot for a slide that is currently scanning, the slot status indicator

turns red, and a message appears instructing you to remove the slide. If the door jams while closing (due to an incorrectly inserted slide holder, for example) a dialog

is displayed that asks you to clear the jam and then click OK to try again.

SureScan Microarray Scanner System User Guide 88

Performing Feature Extraction on Scanned Images

There are a number of ways you can do feature extraction on the scanned image files created by your scanner: Set up the Scan Control program to run Feature Extraction automatically as the slides are

scanned. See To set the location of the Feature Extraction program on page 47 and To select to run Feature Extraction after scanning on page 48. The Scan Control program automatically creates a project file that is used to feature extract the image after it is scanned. The Feature Extraction database must contain the design (grid) for the image, or you must have the Feature Extraction program configured to use the eArray server during extraction.

Run the Feature Extraction program manually after the scanned images are created. For information on using the Feature Extraction program, see the documentation provided with the program.

To use Agilent CytoGenomics to feature extract the images, set the output folder to the auto-processing TIFF image folder configured in the Agilent CytoGenomics program. The CytoGenomics program picks up the images automatically and then feature extracts and analyzes them. For more information, see the Agilent CytoGenomics program documentation.

NOTE The Feature Extraction program is preinstalled on your scanner workstation.

The Agilent CytoGenomics program is not included with the SureScan Microarray Scanner System.

SureScan Microarray Scanner System User Guide 89

SureScan Scanner Specifications

The scanner operates within the following specifications:

Approximate dimensions

Height: 42 cm (16.5 in)

Width: 43 cm (17 in)

Depth: 67 cm (26 in)

Weight 56.8 kg (125 lbs)

Power input 100 240 Vac, 50 - 60 Hz, 250-VA max.

Fuses Two power supply fuses: T4A, 250 VAC (part# 2110-1491)

Temperature range

Operating: 15 to 30 C

Storage: 40 to +50 C

Humidity Operating: 15% to 85% RH at 30 C

Potentially sensitive to condensing humidity conditions. Follow precautions stated in Tips to avoid damage to the scanner on page 52. Always allow 12 hours thermal equilibration time on site before opening the shipping box.

Altitude Operating maximum: 3000 m

Storage maximum: 9200 m (30,000 ft) at 40 C

Usage Indoor use

Laser information

Wavelengths: Green solid-state laser: 532 nm Red solid-state laser: 640 nm

Power: both controlled to 13 mW

Maximum scan region

71 mm x 21.6 mm

Suggested microarray

print region

1 mm smaller than scan region on the right, 2 mm on the left, and 0.6 mm on the top and bottom. For non-Agilent slides, the print region is smaller than this by an amount equal to the array printing tolerance.

Dyes supported

Cyanine-3 (Cy-3) and cyanine-5 (Cy-5) and dyes similar to Cy-3 and Cy-5 and Alexa 647, 555, and 660 dyes

Resolution (pixel size)

2, 3, 5 or 10 microns

Pixel placement

error

< 1 pixel at 5-micron resolution

Uniformity Average Global non-uniformity: 5% using 100 m2 features Average Local non-uniformity: 2% using 100 m2 features*

* Typical average Local non-uniformity is 1% using 100 m2 features

SureScan Microarray Scanner System User Guide 90

Scan time

Dynamic range Single scan 16-bit dynamic range >104

Single scan 20-bit dynamic range >105

Dual scan extended dynamic range (XDR) >106

Table 13 Scan Time for Single and Double Pass Scans for Agilent HD Scan Region 61 x 21.6 mm

Resolution Scan Time, min.

2-micron single pass 24

3-micron single pass 16

5-micron single pass 10

10-micron single pass 10

2-micron double pass 46

3-micron double pass 31

5-micron double pass 19

2-micron high-sensitivity 36

3-micron high-sensitivity 24

5-micron high-sensitivity 15

10-micron high-sensitivity 15

SureScan Microarray Scanner System User Guide 91

Slide Specifications

Scan dimensions The scan region for a standard Agilent microarray is specified in Figure 34. All dimensions are in millimeters and the reference point is the lower right side of the glass.

Figure 34 Default scan region for G4900DA scanner

Glass specifications The scanner uses slide holders to move the microarrays in and out of the cassette. These slide holders are designed to accept a 1 in 3 in nominal piece of glass.

The detailed specifications of the glass are as follows: 1 inch wide (25.4 mm, 0.45 mm, or +0.7 mm) 3 inches long (76.2 mm, +0.25 mm, or 1.4 mm) 1 mm thick (+/-0.1 mm) No mirrored slides High quality with low intrinsic fluorescence Index of refraction from 1.510 to 1.515

Barcode and barcode label specifications

Barcode specifications for Agilent slides

The G4900DA scanner reads barcodes placed on the active side of the slide.

For backwards compatibility with the G2565AA model scanners, Agilent microarrays continue to have barcodes on both sides. The label with the text Agilent denotes the active side; the label with the numeric value is the inactive side.

Ba rc

od e

La be

l

Scan Region

4.0

61.0

1.9

21.6

4.2

25.4 + 0.7 or -0.45

76.2 +0.25 -1.4

1.0

SureScan Microarray Scanner System User Guide 92

Figure 35 Agilent slide barcode orientation vs. microarray surface

Barcode specifications for non-Agilent slides

The barcode label maximum thickness is 0.2 mm. Make sure that there are no bubbles in the label. Do not add labels on top of old labels or replace labels that are beginning to peel off.

Do not place barcode labels on slides whose scan region exceeds 61 mm. Enter the slide identifier or barcode manually for slides whose scan regions exceed 61 mm.

For a barcoded slide, the distance between the edge of the microarray scan region and the edge of the barcode label can be no less than 4 mm.

Figure 36 Barcode label location on non-Agilent slide

The barcode label width should be about 6 mm; the recommended barcode label width is for use with the 61-mm scan region. Larger or longer barcode labels can be used by decreasing the scan region proportionally.

Barcode labels should be narrow enough so as not to interfere with the slide holder tabs. Maximum allowable tilt in the placement of the barcode label on the slide is 3 .

The barcode must meet the following requirements: The minimum quiet zone (the region between the edge of the label and the start or end of the

barcode) is ten times the X-dimension on each side of the barcode. Barcode lines must be parallel to the long edge of the slide within tilt tolerance of 3 . Agilent supports the following barcode formats:

>4 mm

Ba rc

od e

la be

l a re

a

Nominal scan region

SureScan Microarray Scanner System User Guide 93

Code 128: Non-Agilent barcodes with 12 digits must not start with 2 CODABAR: Maximum of eight digits; minimum line thickness = 6 mils Code 39: Maximum of eight digits; minimum line thickness = 6 mils Code 93: Maximum of eight digits; minimum line thickness = 6 mils

Minimum height of barcode must be Code 128: 15% of barcode length for Code 128 Code 39, Code 93, CODABAR: 5 mm or 15% of barcode length, whichever is greater

SureScan Microarray Scanner System User Guide 94

Regulatory Information

This section lists regulatory information for the SureScan Microarray Scanner System, which includes the G2600D SureScan microarray scanner, computer workstation, and control software.

For Research Use Only. Not for use in diagnostic procedures.

Acoustic noise information

Manufacturers Declaration:

English This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991. Sound Pressure Lp < 70 dB(A), at operators position, normal operation, according to EN 27779/ISO 7779 (Type Test).

Deutsch Die folgende Information wird in bereinstimmung mit den Anforderungen der Maschinenlrminformationsverordnung vom 18. Januar 1991 erteilt. Schalldruckpegel am Arbeisplatz bei normalem Betrieb, Lp < 70 dB(A), nach EN 27779/ISO 7779 (Typprfung).

Recycling and disposal

This device is designed to accommodate recycling at the end of its useful life. Please dispose of this device in accordance with local regulations.

Contact Agilent Technologies for more information on recycling and disposal.

Electromagnetic interference

The scanner is intended for use with shielded cables only.

Emissions Complies with the emissions limits for Class A, Group 1 equipment specified in CISPR 11/EN5011 as required in IEC 61326-1 for Class A equipment. This equipment is not intended for use in residential areas.

Immunity This device complies with the immunity levels required in IEC 61326-2-6 for a non-controlled, electromagnetic environment. This equipment is not intended for use in residential or industrial environment. See accompanying Declaration of Conformity for specific levels.

Canada This ISM (Industrial-Scientific-Medical) device complies with Canadian ICES-001.

Cet appareil ISM est conforme a la norme NMB-001 du Canada.

Safety information

This scanner complies with the following safety standards:

SureScan Microarray Scanner System User Guide 95

Pollution Degree: 2 Installation Category: II Class I Equipment; requires a grounding system Class 1 Laser Product CSA and NRTL Certified Product

CAN/CSA-. C22.2 No. 61010-1 - 04

Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements

UL Std No. 61010-1 (2nd Edition)

Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use: Part 1: General Requirements

IEC 61010-1:2001 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements

EN 61010-1:2001 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use Part 1: General Requirements

IEC 60825-1:2007 Safety of laser products Part 1: Equipment classification and requirements

EN60825-1:2007 Safety of Laser Product Part 1: Equipment Classification and Requirements

SureScan Microarray Scanner System User Guide 96

Revision C0, November 2022

*G4900-90000* G4900-90000

www.agilent.com

Manufacturer Agilent Technologies Singapore (International) Pte Ltd. No. 1 Yishun Avenue 7, Singapore 768923 Manufactured at: Agilent Technologies Singapore Pte Ltd. No. 1 Yishun Avenue 7, Singapore 768923 SINGAPORE www.agilent.com

Agilent Worldwide Technical Support Telephone support for US and Canada Call 800-227-9770 Telephone and email support for all regions Agilents world-wide Sales and Support Center contact details for your location can be obtained at www.agilent.com/en/contact-us/page.

Agilent Technologies, Inc. 2011-2013, 2015, 2019, 2022

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign

Manualsnet FAQs

If you want to find out how the G4900DA SureScan MicroArray Scanner System Agilent works, you can view and download the Agilent G4900DA SureScan MicroArray Scanner System Platform User Guide on the Manualsnet website.

Yes, we have the User Guide for Agilent G4900DA SureScan MicroArray Scanner System as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Guide should include all the details that are needed to use a Agilent G4900DA SureScan MicroArray Scanner System. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent G4900DA SureScan MicroArray Scanner System Platform User Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent G4900DA SureScan MicroArray Scanner System Platform User Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent G4900DA SureScan MicroArray Scanner System Platform User Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent G4900DA SureScan MicroArray Scanner System Platform User Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent G4900DA SureScan MicroArray Scanner System Platform User Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.