Agilent G2534A Microarray Hybridization Chamber User Guide

User Guide

Before You Begin 3

Safety Notes 3

Agilent Microarray Hybridization Chamber Kit (G2534A) Contents 4

Equipment 5

Procedures 6

To prepare to load the microarray hybridization chamber 6

To remove gasket slide from its packaging 8

To insert the hybridization gasket slide into the chamber base. 9

To load the sample 10

To add the microarray slide 11

To assemble the chamber 13

To transfer microarray slides to slide rack 16

To load assembled chambers into the hybridization oven 16

To clean the chamber assemblies 19

Agilent Technologies

This user guide is a supplement to Agilent microarray processing protocols (available online at: www.agilent.com) and is intended to demonstrate the assembly and disassembly procedures of the Agilent microarray chambers (G2534A).

Agilent lets you successfully run microarray experiments in your own lab, with enhanced ease-of-use and quality. The Agilent microarray hybridization chamber (G2534A) represents a second generation product advancement in engineering based on Agilents SureHyb technologycomprising an integrated set of hybridization reagents and other products.

Built to enable hybridizations anywhere, these chambers are made from stainless steel for precision fit and finish and long-term durability in most lab environments.

To ensure quality hybridizations from run to run, disposable glass gasket slides are used. The silicone gaskets are made in an industrial-scale setting and meet Agilents stringent quality standards. The gasket slides come in a variety of formats to accommodate most of Agilents microarray layouts. Because gasket slides are disposable, they are not meant for reuse, which reduces potential erroneous results.

When combined with Agilent microarray processing protocol and hybridization reagents, you benefit from a fully integrated and tested system that improves productivity while you reduce microarray experiment variability.

Agilent cannot guarantee microarray performance using non-Agilent protocols in microarray labeling and hybridization and it does not provide support to any non-Agilent protocols.

Updated protocols for all of Agilents microarray products are available at www.agilent.com.

2 Agilent Microarray Hybridization Chamber User Guide

Before You Begin

Make sure you read and understand the information in this chapter and have the necessary equipment and reagents listed before you start an experiment.

Safety Notes

WARNING Observe standard laboratory safety procedures when working in the laboratory.

The edges of the gasket slides may be sharp. Please handle carefully.

Agilent Microarray Hybridization Chamber User Guide 3

Agilent Microarray Hybridization Chamber Kit (G2534A) Contents

Table 1 Hybridization Chamber Kit Content

Component Description

Hybridization Chamber Assembly

Chamber base, chamber cover, clamp assembly and tweezers

User Guide Microarray Hybridization Chamber User Guide

CAUTION Store protected in a storage box or drawer. Do not stack or drop chamber components or irreparable harm can be done to the precision stainless steel hybridization surfaces.

TWEEZERS

CLAMP ASSEMBLY

CHAMBER COVER

CHAMBER BASE

Thumbscrew

Bridge

4 Agilent Microarray Hybridization Chamber User Guide

Equipment

Refer to the microarray-processing protocol for more required equipment, reagents, hardware and software.

Table 2 Disposable gasket slides for Agilent microarray hybridization chambers

Description Part Number. Description.

Microarray hybridization chamber, stainless steel G2534A

Gasket slide, 1 microarray/slide format G2534-60003 G2534-60008 G2534-60005

5 slides/kit 20slides/kit 100 slides/kit

Gasket slide, 2 microarrays/slide format G2534-60002 G2534-60009 G2534-60006

5 slides/kit 20slides/kit 100 slides/kit

Gasket slide, 4 microarrays/slide format G2534-60011 G2534-60012 G2534-60013

5 slides/kit 20slides/kit 100 slides/kit

Gasket slide, 8 microarrays/slide format G2534-60014 G2534-60015 G2534-60016

5 slides/kit 20slides/kit 100 slides/kit

Agilent Microarray Hybridization Chamber User Guide 5

Procedures

This section contains the instructions to use the hybridization chamber. Refer to your microarray protocol for details.

To prepare to load the microarray hybridization chamber

1 Check that the work area is clean and free from potential contaminants. 2 Get the needed quantity of labeled chamber assemblies and packaged

gasket slides to process samples. Disassemble chambers, if needed, and line up chamber bases, chamber covers and clamp assemblies.

3 Prepare pipettes and tips for sample loading.

CAUTION The precision-engineered stainless steel parts are vulnerable to damage. Use caution when you assemble, disassemble, wash and store the chamber assembly,

6 Agilent Microarray Hybridization Chamber User Guide

4 To better track samples, label your chamber bases. Mark them with a black waterproof marking pen.

Agilent Microarray Hybridization Chamber User Guide 7

To remove gasket slide from its packaging

1 With tweezers, carefully lift up the corner of the clear plastic covering and slowly pull back the protective sheath.

2 With clean, powder-free gloved fingers, remove the gasket slide from its package. Handle the slide only from its edges.

3 Immediately place in the chamber base to avoid any potential contamination from surrounding surface materials.

CAUTION Do not remove gasket slide from protective sleeve until ready for use.

Do not slice or cut open the gasket slide protective packaging.

Handle only the edges of the slide.

8 Agilent Microarray Hybridization Chamber User Guide

To insert the hybridization gasket slide into the chamber base.

1 Hold the slide with the label reading Agilent face up, which puts the gasket side of the slide up.

2 Locate the four chamber base guideposts and rectangular barcode guide in the chamber base.

3 Position the gasket slide between the 4 chamber base guide posts (see figure below) with the Agilent Label resting over the bases rectangular barcode guide.

4 Gently place the gasket slide into the chamber base.

5 Make sure the gasket slide rests flush against the chamber base. Re-adjust to a flush position against the chamber base if needed.

Slide and gasket are not flush

Agilent Microarray Hybridization Chamber User Guide 9

To load the sample

1 Select the correct volume of hybridization mixture to load onto the gasket slide surface. Refer to the microarray processing protocol for hybridization solution preparation steps and volume.

The labeled target and hybridization solution will be loaded onto the gasket slide, not the microarray slide.

2 Start close to one end of space contained within the gaskets. 3 Slowly dispense your sample while you move your pipette to the

opposite end of the well so that the sample is distributed across the space within the gaskets.

Do not touch the slide.

This action is called drag and dispense. This method helps to distribute the sample evenly across the surface and avoids the spillover of sample over the gasket edge.

These images are for demonstration purposes only. Always put the backing slide in the chamber base before you dispense the hybridization solution.

10 Agilent Microarray Hybridization Chamber User Guide

To add the microarray slide

1 With gloved fingers, grab a microarray from its slide holder between your thumb and index finger, numeric barcode side facing up and Agilent label facing down, and carefully lower it on top of the gasket slide.

Barcode ends of both slides must line up at rectangular end of the chamber base.

2 Use the four chamber base guideposts and rectangular end of the base to position the oligo microarray slide as you lower it to within 3 mm (1/8) above the gasket slide. Once positioned, gently rest it on the lower gasket slide.

Agilent Microarray Hybridization Chamber User Guide 11

CAUTION Do not drop the array slide onto the gasket slide as this increases the chances of sample mixing between gasket wells. Make sure that the oligo microarray is not exposed.

Quickly reposition the slide if needed. Once correctly placed, do not attempt to move the chamber and sandwiched slides as this can cause leakage of the hybridization solution.

12 Agilent Microarray Hybridization Chamber User Guide

To assemble the chamber

1 Place the chamber cover, correct side facing up, onto the chamber base which contains both sandwiched slides.

2 From the rounded corner of the chamber base, slip the clamp assembly onto the chamber base and cover until it stops firmly in place, resting at the center of the two pieces.

3 Keep the chamber assembly flat on the lab bench to avoid spillage of the hybridization solution.

Correct

Wrong

Agilent Microarray Hybridization Chamber User Guide 13

4 Firmly tighten the thumbscrew fully.

The slides will not be harmed by hand-tightening.

5 Rotate the final assembled chamber in a vertical orientation, clockwise, 2 to 3 times to wet the gaskets.

This helps ensure that the hybridization solution will coat the entire surface of the microarray during the incubation process.

6 Inspect for good bubble formation.

Hold the chamber vertically and inspect for stray or small bubbles that do not move as you rotate the chamber.

Use the large mixing bubble to dislodge small stray or stationary bubbles.

CAUTION If you do not completely tighten the thumbscrew, hybridization solution can leak in during hybridization.

Do not use tools to tighten the thumbscrew. The use of pliers or other tools can damage the parts and will void the warranty.

14 Agilent Microarray Hybridization Chamber User Guide

If the small stray or stationary bubbles persist, gently tap the assembled chamber on a firm surface. Rotate the chamber on its sides as you tap. Inspect again and repeat if needed until the small stray or stationary bubbles dissipate.

Figure 1 The slide on the left shows stray or stationary bubbles, which must be removed before hybridization. The slide on the right shows a single large mixing bubble, which moves freely around the chamber when rotated. Multiple bubbles are acceptable, as long as they move freely when you rotate the chamber.

Wrong Correct

Agilent Microarray Hybridization Chamber User Guide 15

To load assembled chambers into the hybridization oven

1 Load each assembled chamber into the oven rotator rack. Refer to the figures below for correct orientation.

2 Close the door and adjust the rotator speed to match what is specified in your microarray processing protocol.

To transfer microarray slides to slide rack

1 Prepare the wash staining dishes as described in microarray processing protocol.

2 Fill dishes with wash solutions. 3 Check that the slide racks, stir bars and magnetic stirrer are ready to

go.

4 Put a pair of blunt end tweezers next to the first wash staining dish for quick disassembly of slides.

5 Remove 1 microarray hybridization chamber assembly from the oven. Leave the remaining assemblies in the oven to minimize the loss of heat in the chambers during the disassembly process.

6 Put the hybridization chamber assembly on a flat surface and loosen the thumbscrew, turning counterclockwise.

Correct Incorrect

16 Agilent Microarray Hybridization Chamber User Guide

7 Slide off the clamp assembly and remove the chamber cover. 8 With gloved fingers, grab the slides from their ends and remove the

sandwiched slides from the chamber base.

9 Keep the oligo microarray slide numeric barcode face up as you quickly transfer and submerge the sandwiched slides into the FIRST wash staining dish.

10 With the sandwiched slides completely submerged in the wash solution, pry the two slides apart from the barcode end only.

To do so, slip one of the blunt ends of the tweezers between the slides and then gently turn the tweezers upward or downward to separate the slides.

Agilent Microarray Hybridization Chamber User Guide 17

11 Let the gasket slide drop to the bottom of the wash staining dish.

12 Remove the microarray slide quickly and put it into the slide rack in the second wash staining dish. Minimize exposure of the slide to air.

13 Repeat disassembly procedure until all slides are in the rack. 14 When all gasket slides have been removed, dispose of them in a sharp

objects disposal container.

18 Agilent Microarray Hybridization Chamber User Guide

To clean the chamber assemblies

1 Line up each chamber assembly part side by side in a wash basin.

Refer to the figures below for the correct and incorrect ways to handle the chamber assembly parts. Do not stack, pile, throw or otherwise mishandle the chamber assembly parts.

2 Use tap water to clean the chamber assembly parts.

3 Thoroughly rinse with copious amounts of water.

4 Dry each part completely.

CAUTION The chamber assemblies are precision-engineered and are vulnerable to damage when mishandled. Do not stack, pile, or throw the assemblies.

Correct Wrong

CAUTION Do not use deionized water or you will corrode the stainless steel surface.

CAUTION Make sure cleaning solution is thoroughly rinsed. Any remaining solution can cause damage to the stainless steel surface.

Agilent Microarray Hybridization Chamber User Guide 19

5 Assemble and store appropriately.

Line up neatly in a drawer, or store in a protective plastic container.

Store gasket slides in original shipping container or in a protected container.

Store the chamber assemblies and gasket slides in a nuclease-free environment to avoid contamination.

20 Agilent Microarray Hybridization Chamber User Guide

This page intentionally left blank.

Agilent Technologies, Inc. 2010-2016

Printed in USA Revision A1, June 2016

*G2534-90004* G2534-90004

www.agilent.com

Agilent Technologies

In This Guide

This guide contains in

Manualsnet FAQs

If you want to find out how the G2534A Microarray Hybridization Chamber Agilent works, you can view and download the Agilent G2534A Microarray Hybridization Chamber Platform User Guide on the Manualsnet website.

Yes, we have the User Guide for Agilent G2534A Microarray Hybridization Chamber as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Guide should include all the details that are needed to use a Agilent G2534A Microarray Hybridization Chamber. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent G2534A Microarray Hybridization Chamber Platform User Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent G2534A Microarray Hybridization Chamber Platform User Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent G2534A Microarray Hybridization Chamber Platform User Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent G2534A Microarray Hybridization Chamber Platform User Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent G2534A Microarray Hybridization Chamber Platform User Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.