Agilent Microarray Analysis of Methylated DNA Immunoprecipitation

Protocol Version 3.0 August 2020

For Research Use Only. Not for use in diagnostic procedures.

Notices Manual Part Number G4170-90012 Version 3.0 August 2020

Copyright Agilent Technologies, Inc. 2008-2015, 2020

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051

Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation).

Safety Notices

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Technical Support For US and Canada

Call 800-227-9770 (option 3, 4, 2) Or send an email to genomics@agilent.com

For all other regions

Agilents world-wide Sales and Support Center contact details can be obtained at www.agilent.com/en/contact-us/page.

In This Guide... This protocol describes the use of methylation microarrays for the analysis of Methylated DNA Immunoprecipitation (MeDIP).

1 Before You Begin

This chapter contains information (such as procedural notes, safety information, required reagents and equipment) that you need to read and understand before you start an experiment.

2 Sample Preparation

This chapter describes the standard method to process DNA prior to labeling.

3 Sample Labeling

This chapter describes the steps to differentially label the DNA samples with fluorescent-labeled nucleotides.

4 Microarray Processing and Feature Extraction

This chapter describes the steps to hybridize, wash and scan Agilent DNA Methylation microarrays and to extract data using the Agilent Feature Extraction Software for use in Agilent Genomics Workbench.

5 Reference

This chapter contains reference information related to the amplification, labeling, hybridization and wash kits, and the protocol.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 3

Whats new in 3.0 Updated document look and feel. Instructions on handling the newly redesigned secure fit slide boxes in which

the microarray slides are shipped. Before opening the box for the first time, see Secure Fit Slide Box Opening Instructions" on page 63.

Expanded instructions and new images in the Microarray Processing and Feature Extraction procedures to help avoid common problems and optimize hybridization of your sample to the microarray.

Updated web addresses for Agilent materials. Updated Safety Notes. Removed microarray scanning instructions for the Agilent B scanner.

Whats new in 2.3 Corrected fluorescent labeling step to remove preparation of amplified

reference and IP gDNA. Updated temperature for denaturing step. Updated product labeling statement.

4 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

Content

1 Before You Begin 7

Procedural Notes 8 Safety Notes 9 Agilent Methylation Microarray Kit Contents 10 Required Equipment 11 Required Reagents 13 Required Hardware and Software 14

2 Sample Preparation 15

Step 1. Prepare the magnetic beads 17 Step 2. Prepare DNA for Immunoprecipitation 19 Step 3. Immunoprecipitate the methylated DNA 19 Step 4. Wash and elute methylated DNA from beads 20 Step 5. Extract immunoprecipitated and reference DNA with phenol-chloroform 21

3 Sample Labeling 23

Step 1. Fluorescent Labeling of DNA 24 Step 2. Purification of Labeled DNA 26 To determine yield and specific activity 27

4 Microarray Processing and Feature Extraction 29

Microarray Hybridization 30

Step 1. Prepare the 10 Blocking Agent 30 Step 2. Prepare labeled DNA for hybridization 30 Step 3. Prepare the hybridization assembly 34 Step 4. Hybridize 40

Microarray Wash 42

Step 1. Prewarm Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 (overnight) 43 Step 2. Wash with Milli-Q ultrapure water 43 Step 3. Clean with Acetonitrile (Wash Procedure B Only) 43 Step 4. Prewarm Stabilization and Drying Solution (Wash Procedure B

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 5

Content

Only) 44 Step 5. Wash microarrays 45 Step 6. Put slides in a slide holder 51

Microarray Scanning and Feature Extraction 54

Step 1. Scan the microarray slides 54 Step 2. Extract data using the Feature Extraction program 56

5 Reference 61

Reagent Kit Components 62 Secure Fit Slide Box Opening Instructions 63 Microarray Handling Tips 65 Agilent Microarray Layout and Orientation 66 Array/Sample tracking on microarray slides 68 Notes and Considerations 69

6 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

1 Before You Begin Procedural Notes 8 Safety Notes 9 Agilent Methylation Microarray Kit Contents 10 Required Equipment 11 Required Reagents 13 Required Hardware and Software 14

Make sure that you read and understand the information in this chapter and have the necessary equipment and reagents listed before you start an experiment.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 7

1 Before You Begin Procedural Notes

Procedural Notes

To prevent contamination of reagents by nucleases, always wear powder-free laboratory gloves, and use dedicated solutions and pipettors with nuclease-free aerosol-resistant tips.

Maintain a clean work area. Do not mix stock solutions and reactions containing gDNA or enzymes on a

vortex mixer. Instead, mix the solutions and reactions by gently tapping the tube with your finger.

Avoid repeated freeze-thaw cycles of solutions containing gDNA or enzymes. When preparing frozen reagent stock solutions for use:

1 Thaw the aliquot as quickly as possible without heating above room temperature.

2 Mix briefly on a vortex mixer, and then spin in a microcentrifuge for 5 to 10 seconds to drive the contents off the walls and lid.

3 Store on ice or in a cold block until use. In general, follow Biosafety Level 1 (BL1) safety rules.

8 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

1 Before You Begin Safety Notes

Safety Notes

CAUTION Wear appropriate personal protective equipment (PPE) when working in the laboratory.

WARNING Cyanine reagents are considered hazardous by the OSHA Hazard Communication Standard (29 CFR 1910.1200). Contains material that causes damage to the following organs: kidneys, liver, cardiovascular system, respiratory tract, skin, eye lens or cornea, stomach. May be harmful if swallowed. Avoid contact with eyes, skin and clothing.

2 HI-RPM Hybridization Buffer is considered hazardous by the OSHA Hazard Communication Standard (29 CFR 1910.1200). Contains material that causes damage to the following organs: skin, central nervous system. May be harmful if swallowed. Avoid contact with eyes, skin and clothing.

Triton is harmful if swallowed. Risk of serious damage to eyes. Wear suitable PPE. Triton is a component of the Agilent 2 HI-RPM Hybridization Buffer.

Acetonitrile is a flammable liquid and vapor. Harmful if inhaled, swallowed, or contacts skin. Target organs: liver, kidneys, cardiovascular system, and CNS.

Stabilization and Drying Solution is considered hazardous by the OSHA Hazard Communication Standard (29 CFR 1910.1200). Flammable liquid and vapor. Keep away from heat, sparks and flame. Keep container closed. Use only with adequate ventilation. This solution contains material which causes damage to the following organs: kidneys, liver, cardiovascular system, upper respiratory tract, skin, central nervous system (CNS), eye, lens or cornea.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 9

1 Before You Begin Agilent Methylation Microarray Kit Contents

Agilent Methylation Microarray Kit Contents

Store microarray kit at room temperature. After the microarray foil pouch is opened, store the microarray slides at room temperature (in the dark) under a vacuum desiccator or N2 purge box. Do not store microarray slides in open air after breaking foil.

Catalog SurePrint HD Methylation Microarray Kit

One or two microarrays printed on each 1-inch 3-inch glass slide Available as a 5-slide kit, as indicated in Table 1.

Design files can be downloaded from www.agilent.com/genomics/suredesign.

See Table 1 for available designs. For more information on Methylation designs, go to www.agilent.com.

Unrestricted SurePrint HD Methylation Microarrays

One microarray printed on each 1-inch 3-inch glass slide Number of microarray slides vary per kit and per order

Design files can be downloaded from www.agilent.com/genomics/suredesign.

See the tables that follow for available designs.

Custom SurePrint HD and G3 Microarrays

One, two, or four microarray(s) printed on each 1-inch 3-inch glass slide Number of microarrays varies per kit and per order

Table 1 Catalog SurePrint HD Methylation Microarray Kits

Part Number Description

G4492A Human CpG Island Microarray 5-slide Kit, 1244K (HD)

G4811A Mouse CpG Island Microarray, 2105K (HD)

Table 2 Unrestricted SurePrint Methylation Microarrays

Part Number Description

G4495A, AMAIDID 023795 Human DNA Methylation Microarray, 1244K (HD)

10 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

1 Before You Begin Required Equipment

See Table 3 for available formats.

Required Equipment

Table 3 Custom SurePrint HD and G3 Methylation Microarrays

Part Number Description

G4819A Custom SurePrint G3 ChIP-on-chip/DNA Methylation, 11M

G4820A Custom SurePrint G3 ChIP-on-chip/DNA Methylation, 2400K

G4821A Custom SurePrint G3 ChIP-on-chip/DNA Methylation, 4180K

G4496A Custom ChIP-on-chip/DNA Methylation, 1244K

G4498A Custom ChIP-on-chip/DNA Methylation, 2105K

G4497A Custom ChIP-on-chip/DNA Methylation, 444K

Table 4 Required equipment

Description Vendor and part number

200-L Thin-Wall Tube Agilent p/n 410091 or equivalent

Agilent Microarray Scanner Bundle for 1244K, 2105K, or 444K, or for 11M, 2400K, or 4180K

Agilent p/n G4900DA, G2565CA or G2565BA Agilent p/n G4900DA or G2565CA

Hybridization Chamber, stainless Agilent p/n G2534A

Hybridization gasket slides, 5-pack (20 and 100 packaging sizes are available) for 1-pack microarrays or for 2-pack microarrays or for 4-pack microarrays

Agilent p/n G2534-60003 Agilent p/n G2534-60002 Agilent p/n G2534-60011

Hybridization oven; temperature set at 67C Agilent p/n G2545A

Hybridization oven rotator for Agilent Microarray Hybridization Chambers

Agilent p/n G2530-60029

Ozone-barrier slide covers (box of 20)1 Agilent p/n G2505-60550

1.5-mL RNase-free Microfuge Tube (sustainable at 98C)

Thermo Fisher Scientific p/n AM12400 or equivalent

Magnetic stir plate (1 or 3)2 Corning p/n 6795-410 or equivalent

Magnetic stir plate with heating element Corning p/n 6795-420 or equivalent

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 11

1 Before You Begin Required Equipment

Microcentrifuge Eppendorf p/n 5430 or equivalent

Tube rotator Labquake p/n 56264-306 or equivalent

DynaMag-2 Magnet Thermo Fisher Scientific p/n 12321D

Sterile storage bottle Nalgene 455-1000 or equivalent

UV-VIS spectrophotometer NanoDrop 8000 or 2000, or equivalent

P10, P20, P200 and P1000 pipettes Pipetman P10, P20, P200, P1000 or equivalent

1.5 L glass dish Pyrex p/n 213-R or equivalent

MaXtract High Density 2-mL tube Qiagen p/n 129056

Vacuum Concentrator Thermo Scientific Savant SpeedVac p/n DNA130-115 or equivalent

Magnetic stir bar, 7.9 38.1 mm (2 or 4)2 VWR p/n 58948-150 or equivalent

250 mL capacity slide-staining dish, with slide rack (3 or 5)2

Wheaton p/n 900200 or Thermo Fisher Scientific p/n 121

Circulating water baths or heat blocks set to 37C, 65C, and 95

Ice bucket

Clean forceps

Powder-free gloves

Sterile, nuclease-free aerosol barrier pipette tips

Timer

Vacuum desiccator or N2 purge box for slide storage

Vortex mixer

Sonicator machine

1 Optional. Recommended when processing arrays with a G2565CA scanner in environments in which ozone levels are 5 ppb or higher.

2 The number varies depending on if wash procedure A or B is selected.

Table 5 Optional. Recommended when using high-throughput method on 2-pack microarrays.

Description Vendor and part number

Tall Chimney PCR plate Fisher Scientific p/n 14-230-242

Table 4 Required equipment (continued)

Description Vendor and part number

12 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

1 Before You Begin Required Reagents

Required Reagents

Table 6 Required reagents for DNA Methylation sample preparation

Item Vendor and part number

PBS (Phosphate Buffered Saline) VWR p/n 97062-818

5-Methylcytidine, Monoclonal Antibody, purified Eurogentec p/n BI-MECY-1000

Dynabeads Pan Mouse IgG Thermo Fisher Scientific p/n 11041

Nuclease-free distilled water Thermo Fisher Scientific p/n 10977015

Yeast tRNA Thermo Fisher Scientific p/n 15401011

UltraPure 10% SDS Thermo Fisher Scientific p/n 15553027

Glycogen Sigma-Aldrich p/n 10901393001

Ethanol (95% to 100% molecular biology grade) Sigma-Aldrich p/n E7023-6500ML

BSA, powder Sigma-Aldrich p/n A7906

NaCl Sigma-Aldrich p/n S7653

Triton X-100 Sigma-Aldrich p/n T8787

Phenol-chloroform-isoamyl alcohol Sigma-Aldrich p/n 77617

Table 7 Required reagents for enzymatic sample prep and labeling with the SureTag DNA Labeling Kit

Description Vendor and part number

SureTag DNA Labeling Kit1

1 Kit content is listed in Reagent Kit Components on page 62.

Agilent p/n 5190-3400

1TE (pH 8.0), Molecular grade Thermo Fisher Scientific p/n AM9849

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 13

1 Before You Begin Required Hardware and Software

Required Hardware and Software

Refer to the Agilent Scanner or Feature Extraction manuals for minimum memory requirements and other specifications. Go to www.agilent.com.

Table 8 Required reagents for hybridization and wash

Description Vendor and part number

Oligo aCGH/ChIP-on-chip Wash Buffer Kit or Oligo aCGH/ChIP-on-chip Wash Buffer 1 and Oligo aCGH/ChIP-on-chip Wash Buffer 2

Agilent p/n 5188-5226 Agilent p/n 5188-5221 Agilent p/n 5188-5222

Stabilization and Drying Solution1

1 Optional components recommended if wash procedure B is selected.

Agilent p/n 5185-5979

Oligo aCGH/ChIP-on-chip Hybridization Kit Agilent p/n 5188-5220 (25) or p/n 5188-5380 (100)

Cot-1 DNA (1.0 mg/mL) Human Cot-1 DNA or Mouse Cot-1 DNA

Agilent p/n 5190-3393 Thermo Fisher Scientific p/n 18440016

DNase/RNase-free distilled water Thermo Fisher Scientific p/n 10977015

Milli-Q ultrapure water Millipore

Acetonitrile1 Sigma-Aldrich p/n 271004-1L

Deionized Formamide Sigma-Aldrich p/n F9037-100ML

14 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

2 Sample Preparation Step 1. Prepare the magnetic beads 17 Step 2. Prepare DNA for Immunoprecipitation 19 Step 3. Immunoprecipitate the methylated DNA 19 Step 4. Wash and elute methylated DNA from beads 20 Step 5. Extract immunoprecipitated and reference DNA with

phenol-chloroform 21

The steps in this protocol and the estimated amounts of time required are listed in Table 1.

Table 1 Overview and time requirements.

Step Time Requirement

Binding of antibody to magnetic beads 0.5 hour, then overnight

DNA sonication 1 hour

Methylated DNA immunoprecipitation 0.5 hour, then overnight

Wash, elution 2 hours

DNA Purification with phenol:chloroform:Isoamyl alcohol 2 hours

Cyanine 3/cyanine 5 labeling of immunoprecipitate and reference material

3 hours

Microarray hybridization 1 hour, then 40 hours

Microarray washing 1 hour

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 15

2 Sample Preparation

Figure 1. Methylated Genomic DNA overview

Hybridize to microarray

Purify and label DNA

IMMUNOPRECIPITATED SAMPLE GENOMIC DNA CONTROL

16 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

2 Sample Preparation Step 1. Prepare the magnetic beads

Step 1. Prepare the magnetic beads

The following steps to bind the antibody to the beads are to be done in a cold room or on ice. 1 Set up and label one 1.5-mL RNase-free Microfuge Tube for each

immunoprecipitation. 2 Vigorously mix bottle of Dynabeads Pan Mouse IgG in a vortex mixer to

resuspend. Dynal beads will have settled during storage. 3 Immediately add 50 L of Dynabeads Pan Mouse IgG to a 1.5-mL RNase-free

Microfuge Tube. 4 Put tubes on a magnetic separation stand, such as a DynaMag-2 Magnet. 5 Remove the supernatant with a pipette.

6 Add 750 L of Block Solution 1 and resuspend gently.

7 Gently mix the beads in Block Solution 1 for 5 minutes at 4C on a Tube rotator.

8 Remove the supernatant with a pipette and magnetic separation stand. 9 Repeat step 6 to step 8 for a second wash. 10 Spin in a centrifuge at 4C at 1500 g for 3 minutes. 11 Resuspend the beads in 230 L of Block Solution 2 (Table 3 on page 18) and

add 5 g of 5-Methylcytidine, Monoclonal Antibody, purified per immunoprecipitation.

NOTE Beads can also be processed as a batch in a 15 mL conical tube. Scale the volumes accordingly.

Table 2 Block Solution 1

Stock For 100 mL Final Concentration

PBS 100 mL 1

BSA, powder 500 mg 0.5% BSA (weight/volume)

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 17

2 Sample Preparation Step 1. Prepare the magnetic beads

The first time the antibody is used, divide the unused 5-Methylcytidine, Monoclonal Antibody, purified into 5 g aliquots and store at -80C to limit freeze/thaw cycles.

12 Place tubes on a Tube rotator at 4C for a minimum of 6 hours to bind the Antibody to the beads. You can also leave the tubes on the Tube rotator overnight.

13 The next day quick, spin to collect fluid at the bottom of the tube. Use a magnetic separation stand to collect the beads against the side of tube and remove the supernatant: a Add 750 L of Block Solution 2 to the beads. b Remove the tubes from the magnetic stand and gently resuspend beads. c Use a magnetic separation stand to collect the beads against the side of

tube and remove the supernatant with a pipette. d Repeat wash two more times.

14 Remove the last wash with a 1 mL pipette. 15 Resuspend the beads in 50 L of Block Solution 2.

The beads are now ready for the IP step.

Table 3 Block Solution 2*

* You can also make a 1:10 dilution of Block Solution 1 in PBS to create Block Solution 2. Prepare Blocking Solution fresh each time.

Stock For 100 mL Final Concentration

PBS 100 mL 1

BSA, powder 50 mg 0.05% BSA (weight/volume)

NOTE Beads for up to 6 immunoprecipitations can be combined into one 1.5 mL tube or 8 in a 2 mL tube.

NOTE If beads for multiple IPs were processed together, scale the wash volume accordingly and resuspend the washed beads in 50 L of Block Solution 2 per IP.

18 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

2 Sample Preparation Step 2. Prepare DNA for Immunoprecipitation

Step 2. Prepare DNA for Immunoprecipitation

1 For each sample, resuspend 5 g of purified genomic DNA in 250 L of PBS.

Do not use lower volumes and compensate with a whole genome amplification (WGA) as this will bias the data.

2 Sonicate the suspension with a microtip attached to sonicator. Keep samples in an ice water bath during sonication.

If you use a Branson Digital Sonifier Model 450, set output power to 57%. Sonicate 5 cycles of 5 seconds ON and 10 seconds OFF to decrease foaming.

3 Check volumes with a pipette and bring the volume of each of your samples to 250 L with PBS.

4 Transfer 50 L from each sheared sample into a new 1.5-mL RNase-free Microfuge Tube to use as the reference sample. (Store the sample at -20C until use.) With the remaining 200 L, go to the next step.

Step 3. Immunoprecipitate the methylated DNA

1 Add 50 L antibody/magnetic bead mixture from Step 1. Prepare the magnetic beads on page 17 to the 1.5-mL RNase-free Microfuge Tube that contains the 200 L of sheared genomic DNA from Step 2. Prepare DNA for Immunoprecipitation on page 19.

2 Add 250 L of 2 IP Buffer.

NOTE You may need to optimize sonication conditions. Use the lowest settings that result in sheared DNA that ranges from 200 to 1000 bp in size. Shearing varies greatly depending on quantity, volume, and equipment. Depending on the specific experiment, and using power settings as high as 70%, you can use anywhere from 3 to 8 cycles and variable ratios of time ON and time OFF. You can determine the degree with the use of the Agilent 2100 Bioanalyzer. See QC Metrics on page 66.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 19

2 Sample Preparation Step 4. Wash and elute methylated DNA from beads

3 Gently mix tubes overnight on a Tube rotator at 4C.

Step 4. Wash and elute methylated DNA from beads

Do these steps in a 4C cold room or on ice.

1 Place tubes in a magnetic separation stand to collect the beads. Remove supernatant with a pipette. Keep as many magnetic beads as you can.

2 Add 1 mL of IP Wash Buffer to each tube. 3 Remove tubes from magnetic device and shake or agitate tube gently to

resuspend beads. 4 Rotate beads for 3 minutes at 4C. 5 Replace tubes in magnetic device to collect beads. 6 Remove supernatant. Repeat this wash 2 more times.

7 Remove any residual IP Wash Buffer with a pipette.

Table 4 2 IP Buffer

Final concentration in 2X buffer

Volume

1% Triton X-100 in PBS 0.05% 200 L

Yeast tRNA 50 g/ml 8 L

PBS 3.8 mL

Total 4 mL

Table 5 IP Wash Buffer

Stock Volume

PBS 48.75 mL

1% Triton X-100 in PBS 1.25 mL

20 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

2 Sample Preparation Step 5. Extract immunoprecipitated and reference DNA with phenol-chloroform

Elution

1 Add 150 L of Elution Buffer and resuspend beads.

2 Incubate in water bath at 65C for 2.5 minutes. 3 Remove tubes and quickly mix on a vortex mixer to resuspend beads. 4 Return to 65C water bath and incubate for an additional 2.5 minutes. 5 Place tubes on magnetic separation stand. 6 Transfer 150 L of supernatant to a new labeled 1.5-mL RNase-free

Microfuge Tube.

The supernatant contains your IP DNA. Do not discard. 7 Repeat elution step 1 through step 6 with additional 150 L of Elution Buffer.

Combine the two elutions into one tube.

Step 5. Extract immunoprecipitated and reference DNA with phenol-chloroform

1 Add 250 L of elution buffer to each of your reference samples, which you set aside from Step 2. Prepare DNA for Immunoprecipitation on page 19, for a total volume of 300 L.

2 Add 300 L of Phenol-chloroform-isoamyl alcohol to each immunoprecipitate and reference tube.

3 Mix the samples on a vortex mixer for 20 seconds. 4 On a centrifuge, spin one MaXtract High Density 2-mL tube for each

immunoprecipitated and reference DNA sample at 14,000 g for 30 seconds at room temperature.

5 Transfer all 600 L of the sample to the MaXtract High Density 2-mL tube. 6 Spin the sample in a centrifuge at 14,000 g for 5 minutes at room

temperature.

Table 6 Elution Buffer

Component Volume

1TE (pH 8.0) 45 mL

10% SDS 5 mL

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 21

2 Sample Preparation Step 5. Extract immunoprecipitated and reference DNA with phenol-chloroform

If any samples remains cloudy, repeat the Phenol-chloroform-isoamyl alcohol extraction one more time.

7 Transfer the aqueous layer (top) to a new 1.5-mL RNase-free Microfuge Tube and dispose the MaXtract High Density 2-mL tube.

8 To each tube that contains the aqueous layer, add: 16 L of 5 M NaCl (200 mM final concentration) 1.5 L of 20 g/L Glycogen 880 L Ethanol

9 Mix on a vortex mixer. 10 Cool the mixture for 30 minutes at -80C. 11 Spin the mixture in a centrifuge at 20,000 g for 10 minutes at 4C to create

DNA pellets. 12 Carefully remove the supernatant and retain the pellets. 13 Wash the pellets with 500 L of 70% ice-cold Ethanol. 14 Spin the mixture in a centrifuge at 12,000 g for 3 minutes. 15 Discard the supernatant while you retain the pellets. 16 Dry the pellets for 10 minutes with a vacuum desiccator, such as a Savant

Speed Vac. 17 If you want to keep a part of this sample for more analysis, such as

quantification, resuspend the DNA in 31 L of nuclease-free water. Otherwise, resuspend in 26 L of nuclease-free water.

18 Use the Nanodrop spectrophotometer to measure the yield of the IP and reference samples using the reserved 5 L of the sample.

Yield from a typical tissue ranges from 500 to 700 ng. Hypomethylated or hypermethylated abnormal tissues can produce slightly different yields.

22 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

3 Sample Labeling Step 1. Fluorescent Labeling of DNA 24 Step 2. Purification of Labeled DNA 26 To determine yield and specific activity 27

The SureTag DNA Labeling Kit contains sufficient two-color labeling reaction reagents for 25 microarrays (25 reactions of each color).

It also contain clean-up columns for 25 reactions of each color.

The kit uses random primers and the Exo (-) Klenow fragment to differentially label IP and reference DNA samples with fluorescent-labeled nucleotides. For the Agilent DNA Methylation application, the experimental sample is labeled with one dye while the reference sample is labeled with the other dye. The polarity of the sample labeling is a matter of experimental choice. Typically, the reference sample is labeled with Cy3 and the IP is labeled with Cy5.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 23

3 Sample Labeling Step 1. Fluorescent Labeling of DNA

Step 1. Fluorescent Labeling of DNA

1 Equilibrate heat blocks or water baths to 98C, 37C and 65C, or use a thermal cycler.

2 Spin the samples in a centrifuge for 1 minute at 6,000 g to drive the contents off the walls and lid.

3 Add Random Primer: For 1-pack, 2-pack, and 4-pack microarrays, add 5 L of Random Primer to

each reaction tube containing 26 L of reference or IP DNA to make a total volume of 31 L. Mix well by pipetting up and down gently.

4 Transfer sample tubes to a circulating water bath or heat block at 98C. Incubate at 98C for 3 minutes, then move to ice and incubate on ice for 5 minutes.

or

Transfer sample tubes to a thermal cycler. Program the thermal cycler according to Table 7 and run the program.

5 Spin the samples in a centrifuge for 1 minute at 6,000 g to drive the contents off the walls and lid.

NOTE Cyanine 3-dUTP and cyanine 5-dUTP are light sensitive and are subject to degradation by multiple freeze thaw cycles. Minimize light exposure throughout the labeling procedure.

CAUTION The test/reference sample pairs must be treated identically when they are processed, or else the quality of your data can be adversely affected. The best way to ensure that the sample pairs are exposed to the same temperature during the denaturation step is to use a water bath.

Table 7 DNA denaturation using a thermal cycler

Step Temperature Time

Step 1 95 C 3 minutes

Step 2 4 C hold

24 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

3 Sample Labeling Step 1. Fluorescent Labeling of DNA

6 Prepare Labeling Master Mix: a Mix the components in Table 8 on ice in the order indicated to prepare one

cyanine-3 and one cyanine-5 Labeling Master Mix.

b Add 19 L of Labeling Master Mix to each reaction tube containing the reference or IP DNA to make a total volume of 50 L. Mix well by gently pipetting up and down.

7 Incubate the samples: a Transfer sample tubes to a circulating water bath or heat block at 37C.

Incubate at 37C for 2 hours. b Transfer sample tubes to a circulating water bath or heat block at 65C.

Incubate at 65C for 10 minutes to inactivate the enzyme. c Move the sample tubes to ice.

or

Transfer sample tubes to a thermal cycler. Program the thermal cycler according to Table 9 and run the program.

Reactions can be stored up to a month at -20C in the dark.

Table 8 Labeling Master Mix (for 1-pack, 2-pack and 4-pack microarrays)

Component Per reaction (L)

8 rxns (L) (including excess)

24 rxns (L) (including excess)

48 rxns (L) (including excess)

5 Reaction Buffer 10.0 85 250 500

10 dNTPs 5.0 42.5 125 250

Cyanine 3-dUTP or Cyanine 5-dUTP

3.0 25.5 75 150

Exo (-) Klenow 1.0 8.5 25 50

Final volume of Labeling Master Mix

19.0 161.5 475 950

Table 9 DNA labeling using a thermal cycler

Step Temperature Time

Step 1 37C 2 hours

Step 2 65C 10 minutes

Step 3 4C hold

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 25

3 Sample Labeling Step 2. Purification of Labeled DNA

Step 2. Purification of Labeled DNA

Labeled DNA is purified using the reaction Purification Column provided with the SureTag DNA Labeling Kit. The Purification Column includes:

column 2-mL collection tube

1 Spin the labeled reference or IP DNA samples in a centrifuge for 1 minute at 6,000 g to drive the contents off the walls and lid.

2 Add 430 L of 1TE (pH 8.0) to each reaction tube. 3 For each reference or IP DNA sample to be purified, place a column into a

2-mL collection tube and label the column appropriately. Load each labeled reference or IP DNA onto a column.

4 Cover the column with a cap and spin for 10 minutes at 14,000 g in a microcentrifuge at room temperature. Discard the flow-through and place the column back in the 2-mL collection tube.

5 Add 480 L of 1TE (pH 8.0) to each column. Spin for 10 minutes at 14,000 g in a microcentrifuge at room temperature. Discard the flow-through.

6 Invert the column into a fresh 2-mL collection tube that has been appropriately labeled. Spin for 1 minute at 1,000 g in a microcentrifuge at room temperature to collect purified sample.

7 Add 1TE (pH 8.0), or use a concentrator to bring the sample volume to that listed in Table 10. Do not excessively dry the DNA because the pellets will become difficult to resuspend.

8 Mix thoroughly. If the sample has dried or precipitated after concentration, incubate the tube that contains DNA sample on ice for 5 minutes, and then pipette the solution up and down 10 times.

9 Take 1.5 L of each sample to determine yield and specific activity. See To determine yield and specific activity on page 27. Refer to Table 11 on page 28 for expected yield of labeled DNA and specific activity after labeling and clean-up.

10 In a fresh 1.5-mL RNase-free Microfuge Tube or 200-L Thin-Wall Tube, combine test and reference sample using the appropriate cyanine-5-labeled

NOTE Keep cyanine-3 and cyanine-5 labeled DNA samples separated throughout this clean-up step.

26 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

3 Sample Labeling To determine yield and specific activity

sample and cyanine-3-labeled sample for a total mixture volume listed in Table 10. Use the appropriate container listed in Table 10.

Labeled DNA can be stored up to one month at -20C in the dark.

To determine yield and specific activity

Use the NanoDrop 8000 or 2000 UV-VIS Spectrophotometer to measure yield and specific activity.

1 From the main menu, select MicroArray Measurement, then from the Sample Type menu, select DNA-50.

2 Use 1.5 L of 1TE (pH 8.0) to blank the instrument. 3 Use 1.5 L of purified labeled DNA for quantitation. Measure the absorbance

at A260nm (DNA), A550nm (cyanine 3), and A650nm (cyanine 5). 4 Calculate the Specific Activity of the labeled DNA:

*pmol dyes per g DNA

The Specific Activity is Degree of Labeling divided by 0.034. 5 Record the DNA concentration (ng/L) for each sample. Calculate the yield as

Table 10 Sample volume and total mixture volumes

Microarray Cy3 or Cy5 sample volume after purification

Total mixture volume after Nanodrop and combining

Container

1-pack 42 L 80 L 1.5-mL RNase-free Microfuge Tube

2-pack 22 L 40 L 1.5-mL RNase-free Microfuge Tube or Tall Chimney PCR plate

4-pack 13.25 L 22.5 L 1.5-mL RNase-free Microfuge Tube or Tall Chimney PCR plate

Specific Activity* = pmol per L of dye g per L DNA-------------------------------------------------------

Yield (g) DNA Concentration (ng/L) Sample Volume (L) 1000 ng/g-----------------------------------------------------------------------------------------------------------------------------------------------------=

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 27

3 Sample Labeling To determine yield and specific activity

Refer to Table 11 for expected yield of labeled DNA and specific activity after labeling and purification.

The cyanine 3 and cyanine 5 yield after labeling should be the same.

Table 11 Expected Yield and Specific Activity after Labeling and Purification

Yield (g) Specific activity of cyanine 3 labeled sample (pmol/g)

Specific activity of cyanine 5 labeled sample (pmol/g)

> 2.5 18 to 25 7 to 20

28 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Microarray Hybridization 30 Microarray Wash 42 Microarray Scanning and Feature Extraction 54

Microarray processing consists of hybridization, washing, and scanning.

Feature Extraction is the process by which data is extracted from the scanned microarray image (.tif) and translated into log ratios, allowing researchers to measure DNA copy number changes in their experiments in conjunction with Agilent Genomic Workbench Software.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 29

4 Microarray Processing and Feature Extraction Microarray Hybridization

Microarray Hybridization

Before you begin, make sure you read and understand Secure Fit Slide Box Opening Instructions on page 63 and Microarray Handling Tips on page 65.

To practice hybridization, prepare a 1:1 2 HI-RPM Hybridization Buffer and water mix and use a microscope slide or used microarray slide, and a gasket slide. You can use the same slide to practice wash and placement of slide in the slide holder.

Step 1. Prepare the 10 Blocking Agent

1 Add 1,350 L of DNase/RNase-free distilled water to the vial containing lyophilized 10 aCGH Blocking Agent (included in the Oligo aCGH/ChIP-on-chip Hybridization Kit).

2 Leave at room temperature for 60 minutes and mix on a vortex mixer to reconstitute sample before use or storage.

Step 2. Prepare labeled DNA for hybridization

1 Equilibrate water baths or heat blocks to 95C and 37C or use a thermal cycler.

2 Mix the components according to the microarray format to prepare the Hybridization Master Mix. Refer to Table 12 through Table 14.

NOTE Reconstituted 10 Blocking Agent is stable for 120 days when stored at -20C and can be frozen and thawed up to three times.

30 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 2. Prepare labeled DNA for hybridization

Table 12 Hybridization Master Mix for 1-pack microarray

Component Volume (L) per hybridization

8 rxns (L) (including excess)

24 rxns (L) (including excess)

48 rxns (L) (including excess)

Cot-1 DNA (1.0 mg/mL)*

* Use Cot-1 DNA (1.0 mg/mL) from the appropriate species.

50 425 1,250 2,500

10 aCGH Blocking Agent

Included in the Oligo aCGH/ChIP-on-chip Hybridization Kit

52 442 1,300 2,600

2 HI-RPM Hybridization Buffer 260 2,210 6,500 13,000

Deionized Formamide 78 663 1950 3900

Final Volume of Hybridization Master Mix

440 3,740 11,000 22,000

Table 13 Hybridization Master Mix for 2-pack microarray

Component Volume (L) per hybridization

8 rxns (L) (including excess)

24 rxns (L) (including excess)

48 rxns (L) (including excess)

Cot-1 DNA (1.0 mg/mL)*

* Use Cot-1 DNA (1.0 mg/mL) from the appropriate species.

25 212.5 625 1,250

10 aCGH Blocking Agent

Included in the Oligo aCGH/ChIP-on-chip Hybridization Kit

26 221 650 1,300

2 HI-RPM Hybridization Buffer 130 1,105 3,250 6,500

Deionized Formamide 39 331.5 975 1950

Final Volume of Hybridization Master Mix

220 1,870 5,500 11,000

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 31

4 Microarray Processing and Feature Extraction Step 2. Prepare labeled DNA for hybridization

3 Add the appropriate volume of the Hybridization Master Mix to the 1.5-mL RNase-free Microfuge TubeTall Chimney PCR plate well that contains the labeled reference or IP DNA to make the total volume listed in Table 15.

4 Mix the sample by pipetting up and down, then quickly spin in a centrifuge to drive contents to the bottom of the reaction tube.

5 Transfer sample tubes to a circulating water bath or heat block at 95C. Incubate at 95C for 3 minutes, then immediately transfer sample tubes to a circulating water bath or heat block at 37C. Incubate at 37C for 30 minutes.

or

Transfer sample tubes to a thermal cycler. Program the thermal cycler according to the following table and run the program:

Table 14 Hybridization Master Mix for 4-pack microarray

Component Volume (L) per hybridization

8 rxns (L) (including excess)

24 rxns (L) (including excess)

48 rxns (L) (including excess)

Cot-1 DNA (1.0 mg/mL)*

* Use Cot-1 DNA (1.0 mg/mL) from the appropriate species.

5 42.5 125 250

10 aCGH Blocking Agent

Included in the Oligo aCGH/ChIP-on-chip Hybridization Kit

11 93.5 275 550

2 HI-RPM Hybridization Buffer 55 467.5 1,375 2,750

Deionized Formamide 16.5 140.25 412.25 824.5

Final Volume of Hybridization Master Mix

87.5 743.75 2187.25 4374.5

Table 15 Volume of Hybridization Master Mix per hybridization

Microarray format Volume of Hybridization Master Mix

Total volume

1-pack 440 L 520 L

2-pack 220 L 260 L

4-pack 87.5 L 110 L

32 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 2. Prepare labeled DNA for hybridization

6 Remove sample tubes from the water bath, heat block, or thermal cycler. Spin 1 minute at 6000 g in a centrifuge to collect the sample at the bottom of the tube.

The samples are ready to be hybridized.

Table 16 Thermal cycler program

Step Temperature Time

Step 1 95C 3 minutes exactly

Step 2 37C 30 minutes

CAUTION The samples must be hybridized immediately. If not, keep the temperature of hybridization sample mixtures as close to 37C as possible on a heat block, thermal cycler or in an oven.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 33

4 Microarray Processing and Feature Extraction Step 3. Prepare the hybridization assembly

Step 3. Prepare the hybridization assembly

Refer to the Agilent Microarray Hybridization Chamber User Guide (publication G2534-90004) for in-depth instructions on how to load samples, assemble and disassemble chambers, as well as other helpful tips. This user guide can be downloaded from the Agilent Web site at www.agilent.com.

Before you begin, make sure you read and understand Secure Fit Slide Box Opening Instructions on page 63 and Microarray Handling Tips on page 65.

Remove gasket slide from its packaging

1 With tweezers, carefully lift up the corner of the clear plastic covering and slowly pull back the protective film.

Figure 2. Removal of clear plastic covering

1 With clean, powder-free gloved fingers, remove the gasket slide from its package. Handle the slide only on its edges.

To avoid any potential contamination from surrounding surface materials, immediately insert the gasket slide in the chamber base using the instructions below.

NOTE Do not remove gasket slide from protective sleeve until ready for use. Do not slice or cut open the gasket slide protective packaging. Handle only the edges of the gasket slide. Prior to use, inspect gasket slides for visible gaps or cuts through the gaskets

or any debris within the hybridization areas as these are indications of instability. Do not use gasket slides that have these features.

34 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 3. Prepare the hybridization assembly

Insert the gasket slide into the chamber base

1 Hold the gasket slide so that the barcode label is facing towards you. This side of the slide is the gasket side.

Figure 3. Gasket slide, gasket side

1 Locate the four chamber base guideposts and rectangular barcode guide in the chamber base.

2 Position the gasket slide between the 4 chamber base guide posts (see Figure 4) with the barcode label resting over the bases rectangular barcode guide.

Figure 4. Chamber base, guide posts denoted with arrows

3 Gently place the gasket slide into the chamber base.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 35

4 Microarray Processing and Feature Extraction Step 3. Prepare the hybridization assembly

4 Make sure the gasket slide rests flush against the chamber base. Re-adjust to a flush position against the chamber base if needed.

Figure 5. Correct positioning of gasket slide in chamber base

Load the sample

1 Slowly dispense the appropriate volume of hybridization sample mixture onto the gasket well in a drag and dispense manner (described below). For multi-pack microarray formats (2-pack or 4-pack microarray), load all gasket wells before you add the microarray slide. 490 L (for 1-pack microarray) 245 L (for 2-pack microarray) 100 L (for 4-pack microarray)

The drag and dispense method helps to distribute the sample evenly across the surface of the well and avoids spillover of sample over the gasket edge. Start with the pipette tip near the top edge of the well. Do not directly touch the gasket or the glass with the pipette tip. Then, dispense the mixture while you move your pipette tip to the opposite end of the well so that the sample is distributed across the well space. Avoid creating large air bubbles as you dispense the mixture as they could lead to spillover.

Figure 6. Drag and dispense method Start dispensing when the pipette tip is near the top of the well. Finish dispensing when the pipette tip is near the bottom of the well.

Slide and gasket are flush

36 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 3. Prepare the hybridization assembly

Add the microarray slide

1 Remove a microarray slide from the slide storage box between your thumb and index finger, numeric barcode side facing up and Agilent label facing down.

2 Use the four chamber base guideposts and rectangular end of the base to position the microarray slide as you lower it to within 3 mm (1/8) above the gasket slide, making sure the microarray slide is not tilted with respect to the gasket slide. Barcode ends of both the gasket slide and the microarray slide must line up at the corners of the chamber base. Once positioned, gently rest the microarray slide on the lower gasket slide. Refer to Figure 7 for proper technique on holding the microarray slide with both hands.

Figure 7. Chamber base with gasket and microarray slide applied, guide posts denoted with arrows

CAUTION Keep the temperature of hybridization sample mixtures as close to 37C as possible. To do this, process them in small batches and/or put them on a heat block, thermal cycler or in an oven.

CAUTION Do not drop the microarray slide onto the gasket slide as this increases the chances of sample mixing between gasket wells.

Once placed, do not attempt to move the chamber and sandwiched slides as this can cause leakage of the hybridization solution.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 37

4 Microarray Processing and Feature Extraction Step 3. Prepare the hybridization assembly

Assemble the chamber

1 Place the chamber cover, correct side facing up, onto the chamber base which contains the sandwiched slides.

Figure 8. Chamber cover in correct (left) and incorrect (right) orientations

2 From the rounded corner of the chamber base, slip the clamp onto the chamber base and cover until it stops firmly in place, resting at the center of the two pieces.

Keep the chamber assembly flat on the lab bench to avoid spilling the hybridization solution.

Figure 9. Slipping the clamp onto the chamber base

3 Firmly tighten the thumbscrew fully.

The slides will not be harmed by hand-tightening.

Correct Wrong

38 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 3. Prepare the hybridization assembly

Figure 10. Tightening of the thumbscrew on the clamp

4 Rotate the final assembled chamber in a vertical orientation, clockwise, 2 to 3 times to wet the gaskets (see Figure 11).

Rotation helps ensure that the hybridization solution will coat the entire surface of the microarray during the incubation process.

Figure 11. Rotation of the final assembled chamber

5 Inspect for good bubble formation. Hold the chamber vertically and inspect for stray or small bubbles that do

not move as you rotate the chamber. Use the large mixing bubble to dislodge small stray or stationary bubbles. If the small stray or stationary bubbles persist, gently tap the assembled

chamber on a firm surface. Rotate the chamber on its sides as you tap. Inspect again and repeat if needed until the small stray or stationary bubbles dissipate.

CAUTION If you do not completely tighten the thumbscrew, hybridization solution can leak out during hybridization.

Do not use tools to tighten the thumbscrew. The use of pliers or other tools can damage the parts and will void the warranty.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 39

4 Microarray Processing and Feature Extraction Step 4. Hybridize

Figure 12. The slide on the left shows a stray, stationary bubble (denoted with arrow), which must be removed before hybridization. The slide on the right shows only large mixing bubbles, which move freely around the chamber when rotated. Bubbles are acceptable, as long as they move freely when you rotate the chamber.

Step 4. Hybridize

1 Load the assembled chamber into the oven rotator rack, starting from the center of the rack (position 3 or 4 when counting from left to right). Refer to the figure below for correct and incorrect orientations.

Figure 13. Assembled chambers in correct (left) and incorrect (middle and right) orientations

2 Close the door and set the rotator speed to 20 rpm. 3 Hybridize at 67C for:

24 hours (4-pack microarrays) 40 hours (1-pack and 2-pack microarrays)

Wrong Correct

CAUTION If you are not loading all the available positions on the hybridization rotator rack, be sure to balance the loaded hybridization chambers on the rack similar to a centrifuge to prevent unnecessary strain on the oven motor.

40 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 4. Hybridize

CAUTION You must calibrate the hybridization oven regularly for accuracy of the collected data. Refer to Agilent G2545A Hybridization Calibration Procedure (publication G2545-90002) for more information.

NOTE The Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 that is used in the microarray wash procedure needs to be warmed overnight. While you are waiting for the microarray slides to hybridize, do the steps in Step 1. Prewarm Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 (overnight) on page 43.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 41

4 Microarray Processing and Feature Extraction Microarray Wash

Microarray Wash

Before you begin, determine which wash procedure to use:

Always use clean equipment when conducting the wash procedures. Use only dishes that are designated and dedicated for use in Agilent

Methylation experiments.

NOTE The microarray wash procedure must be done in environments where ozone levels are 5 ppb or less. For Scanner C and Scanner B, if ozone levels are between 5 to 10 ppb in your laboratory, use the Agilent Ozone Barrier Slide Cover. SureScan microarray scanner uses a slide holder with a built-in ozone barrier. If ozone levels exceed 10 ppb, use the Stabilization and Drying Solution together with the ozone barrier.

You can also use Carbon Loaded Non-woven Filters to remove ozone from the air. These filters can be installed in either your HVAC system, or as part of small Ozone Controlled Enclosures. These free-standing enclosures can be installed either on a lab bench or as a walk-in room within your lab. These products are available through filter suppliers.

Table 17 Wash procedure to follow

Ozone level in your lab

Wash Procedure Ozone-Barrier Slide Cover

< 5 ppb Wash Procedure A (without Stabilization and Drying Solution) on page 45

No

> 5 ppb < 10 ppb Wash Procedure A (without Stabilization and Drying Solution) on page 45

Yes

> 10 ppb Wash Procedure B (with Stabilization and Drying Solution) on page 49

Yes

CAUTION Do not use detergent to wash the staining dishes as some detergents may leave fluorescent residue on the dishes. If you do, you must ensure that all traces are removed by thoroughly rinsing with Milli-Q ultrapure water.

42 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 1. Prewarm Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 (overnight)

Step 1. Prewarm Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 (overnight)

The temperature of Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 must be at 37C for optimal performance. 1 Add the volume of buffer required to a Sterile storage bottle and warm

overnight in an incubator or circulating water bath set to 37C. 2 Put a slide-staining dish with a lid, a 1.5 L glass dish, and one to two liters of

Milli-Q ultrapure water in an incubator or water bath set at 37C to warm overnight.

Step 2. Wash with Milli-Q ultrapure water

Rinse slide-staining dishes, slide racks and stir bars thoroughly with high-quality Milli-Q ultrapure water before use and in between washing groups. 1 Run copious amounts of Milli-Q ultrapure water through the slide-staining

dishes, slide racks and stir bars. 2 Empty out the water collected in the dishes at least five times. 3 Repeat step 1 and step 2 until all traces of contaminating material are

removed.

Step 3. Clean with Acetonitrile (Wash Procedure B Only)

Acetonitrile wash removes any remaining residue of Stabilization and Drying Solution from slide-staining dishes, slide racks and stir bars that were used in previous experiments with Wash Procedure B (with Stabilization and Drying Solution) on page 49.

WARNING Do Acetonitrile washes in a vented fume hood. Acetonitrile is highly flammable and toxic.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 43

4 Microarray Processing and Feature Extraction Step 4. Prewarm Stabilization and Drying Solution (Wash Procedure B Only)

1 Add the slide rack and stir bar to the slide-staining dish, and transfer to a magnetic stir plate.

2 Fill the slide-staining dish with 100% Acetonitrile. 3 Turn on the magnetic stir plate and adjust the speed to a setting of 4 (medium

speed). 4 Wash for 5 minutes at room temperature. 5 Discard the Acetonitrile as is appropriate for your site. 6 Repeat step 1 through step 5. 7 Air dry everything in the vented fume hood. 8 Continue with the Milli-Q ultrapure water wash as previously instructed.

Step 4. Prewarm Stabilization and Drying Solution (Wash Procedure B Only)

The Stabilization and Drying Solution contains an ozone scavenging compound dissolved in Acetonitrile. The compound in solution is present in saturating amounts and may precipitate from the solution under normal storage conditions. If the solution shows visible precipitation, warming of the solution will be necessary to redissolve the compound. Washing slides using Stabilization and Drying Solution showing visible precipitation will have profound adverse affects on microarray performance.

WARNING The Stabilization and Drying Solution is a flammable liquid. Warming the solution will increase the generation of ignitable vapors. Use gloves and eye/face protection in every step of the warming procedures.

WARNING Do not use a hot plate, oven, an open flame or a microwave. Do not increase temperature rapidly. Warm and mix the material away from ignition sources.

WARNING Failure to follow the outlined process will increase the potential for fire, explosion, and possible personal injury.

44 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 5. Wash microarrays

1 Put a clean magnetic stir bar into the Stabilization and Drying Solution bottle and recap.

2 Partially fill a plastic bucket with hot water at approximately 40C to 45C (for example from a hot water tap).

3 Put the Stabilization and Drying Solution bottle into the hot water in the plastic bucket.

4 Put the plastic bucket on a magnetic stirrer (not a hot-plate) and stir. 5 The hot water cools to room temperature. If the precipitate has not all

dissolved replenish the cold water with hot water. 6 Repeat step 5 until the solution is clear. 7 After the precipitate is completely dissolved, allow the solution to equilibrate

to room temperature prior to use.

Step 5. Wash microarrays

Wash Procedure A (without Stabilization and Drying Solution)

Always use fresh Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 for each wash group (up to five slides).

Table 18 lists the wash conditions for the Wash Procedure A without Stabilization and Drying Solution.

CAUTION Do not filter the Stabilization and Drying Solution, or the concentration of the ozone scavenger may vary.

Table 18 Wash conditions

Dish Wash buffer Temperature Time

Disassembly #1 Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1

Room temperature

1st wash #2 Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1

Room temperature 5 minutes

2nd wash #3 Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2

37C 1 minute

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 45

4 Microarray Processing and Feature Extraction Step 5. Wash microarrays

1 Completely fill slide-staining dish #1 with Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 at room temperature.

2 Prepare dish #2: a Put a slide rack into slide-staining dish #2. b Add a magnetic stir bar. Fill slide-staining dish #2 with enough Agilent

Oligo aCGH/ChIP-on-Chip Wash Buffer 1 at room temperature to cover the slide rack.

c Put this dish on a magnetic stir plate. 3 Prepare dish #3:

a Put the prewarmed 1.5 L glass dish on a magnetic stir plate with heating element.

b Put the slide-staining dish #3 into the 1.5 L glass dish. c Fill the 1.5 L glass dish with pre-warmed Milli-Q ultrapure water. d Fill the slide-staining dish #3 approximately three-fourths full with Agilent

Oligo aCGH/ChIP-on-Chip Wash Buffer 2 (warmed to 37C). e Add a magnetic stir bar. f Turn on the heating element and maintain temperature of Agilent Oligo

aCGH/ChIP-on-Chip Wash Buffer 2 at 37C. Monitor with a thermometer. 4 Remove one hybridization chamber from the incubator and resume rotation

of the others. Record whether bubbles formed during hybridization and if all bubbles are rotating freely.

5 Prepare the hybridization chamber disassembly. a Put the hybridization chamber assembly on a flat surface and loosen the

thumbscrew, turning counter-clockwise.

Figure 14. Loosening of the thumbscrew

b Slide off the clamp assembly and remove the chamber cover.

46 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 5. Wash microarrays

Figure 15. Removal of the clamp

c With gloved fingers, remove the microarray-gasket sandwich from the chamber base by lifting one end and then grasping in the middle of the long sides. Keep the microarray slide numeric barcode facing up as you quickly transfer the sandwich to slide-staining dish #1.

d Without letting go of the slides, submerge the microarray-gasket sandwich into slide-staining dish #1 containing Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1.

6 With the sandwich completely submerged in Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1, pry the sandwich open from the barcode end only: a Slip one of the blunt ends of the forceps between the slides. b Gently twist the forceps to separate the slides. c Let the gasket slide drop to the bottom of the staining dish. d Remove the microarray slide, grasp it from the upper corners with thumb

and forefinger, and quickly put into slide rack in the slide-staining dish #2 containing Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 at room temperature. Minimize exposure of the slide to air. Touch only the barcode portion of the microarray slide or its edges!

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 47

4 Microarray Processing and Feature Extraction Step 5. Wash microarrays

\

Figure 16. Removal of the microarray slide from the staining dish

7 Repeat step 4 through step 6 for up to four additional slides in the group. A maximum of five disassembly procedures yielding five microarray slides is advised at one time in order to facilitate uniform washing.

8 When all slides in the group are put into the slide rack in slide-staining dish #2, stir at 350 rpm for 5 minutes. Adjust the setting to get good but not vigorous mixing.

9 Wash the slides in Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2: a Transfer slide rack to slide-staining dish #3, which contains Agilent Oligo

aCGH/ChIP-on-Chip Wash Buffer 2 at 37C: a Activate the magnetic stirrer. b Wash microarray slides for at least 1 minute and no more than 2 minutes.

Adjust the setting to get thorough mixing without disturbing the microarray slides.

10 Slowly remove the slide rack trying to minimize droplets on the slides. It should take 5 to 10 seconds to remove the slide rack.

11 Discard used Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2.

12 Repeat step 1 through step 11 for the next group of five slides using fresh Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 warmed to 37C.

48 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 5. Wash microarrays

13 Scan slides immediately to minimize the impact of environmental oxidants on signal intensities. If necessary, store slides in orange slide boxes in a N2 purge box, in the dark.

Wash Procedure B (with Stabilization and Drying Solution)

Cyanine reagents are susceptible to degradation by ozone. Use this wash procedure if the ozone level exceeds 10 ppb in your laboratory.

Always use fresh Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 for each wash group (up to five slides).

The Acetonitrile (dish #4) and Stabilization and Drying Solution (dish #5) below may be reused for washing up to 4 batches of 5 slides (total 20 slides) in one experiment. Do not pour the Stabilization and Drying Solution back in the bottle.

Table 19 lists the wash conditions for the Wash Procedure B with Stabilization and Drying Solution.

WARNING The Stabilization and Drying Solution must be set-up in a fume hood. Put the Wash Buffer 1 and Wash Buffer 2 set-up areas close to, or preferably in, the same fume hood. Use gloves and eye/face protection in every step of the washing procedure.

Table 19 Wash conditions

Dish Wash Buffer Temperature Time

Disassembly #1 Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1

Room temperature

1st wash #2 Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1

Room temperature 5 minutes

2nd wash #3 Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2

37C 1 minute

Acetonitrile wash #4 Acetonitrile Room temperature 10 seconds

3rd wash #5 Stabilization and Drying Solution Room temperature 30 seconds

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 49

4 Microarray Processing and Feature Extraction Step 5. Wash microarrays

1 In the fume hood, fill slide-staining dish #4 approximately three-fourths full with Acetonitrile. Add a magnetic stir bar and put this dish on a magnetic stir plate.

2 In the fume hood, fill slide-staining dish #5 approximately three-fourths full with Stabilization and Drying Solution. Add a magnetic stir bar and put this dish on a magnetic stir plate.

3 Do step 1 through step 9 in Wash Procedure A (without Stabilization and Drying Solution) on page 45.

4 Remove the slide rack from Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 and tilt the rack slightly to minimize wash buffer carry-over. Quickly transfer the slide rack to slide-staining dish #4 containing Acetonitrile, and stir at 350 rpm for 10 seconds.

5 Transfer slide rack to slide-staining dish #5 filled with Stabilization and Drying Solution, and stir at 350 rpm for 30 seconds.

6 Slowly remove the slide rack trying to minimize droplets on the slides. It should take 5 to 10 seconds to remove the slide rack.

7 Discard used Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2.

8 Repeat step 1 through step 7 for the next group of five slides using fresh Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2 prewarmed to 37C.

9 Dispose of Acetonitrile and Stabilization and Drying Solution as flammable solvents.

NOTE The Acetonitrile and the Stabilization and Drying Solution may be reused for washing of up to four batches of five slides (that is, total 20 microarray slides) in one experiment. Pour the Stabilization and Drying Solution to a different marked bottle, and protect from light with other flammables. After each use, rinse the slide rack and the slide-staining dish that were in contact with the Stabilization and Drying Solution with Acetonitrile followed by a rinse in Milli-Q ultrapure water.

50 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 6. Put slides in a slide holder

Step 6. Put slides in a slide holder

Scan slides immediately to minimize impact of environmental oxidants on signal intensities. If necessary, store slides in the original slide boxes in a N2 purge box, in the dark.

For SureScan microarray scanner

1 Before you insert the slide into a slide holder, inspect the slide holder for any dust or fingerprints. If found, remove the dust or fingerprints with compressed air or a soft, dust-free cloth. If the slide holder is scratched, worn, or damaged, has a lid that does not close tightly, or has a hinge that does not move freely, discard the slide holder and select a different one.

2 Place the slide holder on a flat surface, with the clear cover facing up, and the tab on the right. This helps to ensure that you have the slide aligned properly when you insert it into the slide holder.

3 Gently push in and pull up on the tabbed end of the clear plastic cover to open the slide holder.

Figure 17. Opening the slide holder

4 Insert the slide into the holder. a Hold the slide at the barcode end and position the slide over the open slide

holder. Make sure that the active microarray surface faces up with the barcode on the left, as shown in Figure 18.

b Carefully place the end of the slide without the barcode label onto the slide ledge. See Figure 18.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 51

4 Microarray Processing and Feature Extraction Step 6. Put slides in a slide holder

Figure 18. Inserting slide into the slide holder

c Gently lower the slide into the slide holder. See Figure 19. d Close the plastic slide cover, pushing on the tab end until you hear it click.

This moves the slide into position in the holder.

e Gently push in and pull up on the tabbed end of the clear plastic cover to open it again and verify that the slide is correctly positioned.

f Close the plastic slide cover, gently pushing on the tab end until you hear it click. See Figure 20. Make sure that the slide holder is completely closed.

Figure 19. Slide inserted in slide holder cover open

Slide ledge

CAUTION An improperly inserted slide can damage the scanner.

52 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 6. Put slides in a slide holder

Figure 20. Slide inserted in slide holder cover closed

For Agilent Scanner C

In environments in which the ozone level exceeds 5 ppb, immediately put the slides with active microarray surface (Agilent-labeled barcode) facing up in a slide holder. Make sure that the slide is not caught up on any corner. Put an ozone-barrier slide cover on top of the array as shown in Figure 21. Refer to the Agilent Ozone-Barrier Slide Cover User Guide (publication G2505-90550), included with the slide cover, for more information.

Figure 21. Inserting the ozone-barrier slide cover

In environments in which the ozone level is below 5 ppb, put the slides with Agilent barcode facing up in a slide holder.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 53

4 Microarray Processing and Feature Extraction Microarray Scanning and Feature Extraction

Microarray Scanning and Feature Extraction

Step 1. Scan the microarray slides

A SureScan or Agilent C microarray scanner is required for G3 microarrays.

Agilent provides support for Agilent microarrays scanned on select non-Agilent scanners. Please see Feature Extraction Compatibility Matrix for Non Agilent scanners for scanner compatibility and settings (www.agilent.com/cs/library/usermanuals/Public/G1662-90043_ScannerCom patibilityMatrix.pdf).

However, Agilent can guarantee the quality of data only if the data comes from Agilent microarrays scanned on Agilent scanners.

Agilent SureScan Microarray Scanner

1 Put assembled slide holders into the scanner cassette. Refer to Figure 22 and Figure 23.

Figure 22. Slide holder helps you to insert slides correctly

Finger-hold Arrow helps identify scanner orientation

54 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 1. Scan the microarray slides

Figure 23. Inserting slide holder into cassette

2 Select Protocol AgilentG3_CGH for G3 microarrays. Select Protocol AgilentHD_CGH for HD microarrays.

3 Verify that the Scanner status in the main window says Scanner Ready. 4 Click Start Scan.

Agilent C Scanner Settings

1 Put assembled slide holders with or without the ozone-barrier slide cover into scanner carousel.

2 Select Start Slot m End Slot n where the letter m represents the Start slot where the first slide is located and the letter n represents the End slot where the last slide is located.

3 Select Profile AgilentG3_CGH for G3 microarrays. Select Profile AgilentHD_CGH for HD microarrays.

4 Verify scan settings. See Table 20.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 55

4 Microarray Processing and Feature Extraction Step 2. Extract data using the Feature Extraction program

5 Check that Output Path Browse is set for desired location. 6 Verify that the Scanner status in the main window says Scanner Ready. 7 Click Scan Slot m-n on the Scan Control main window where the letter m

represents the Start slot where the first slide is located and the letter n represents the End slot where the last slide is located.

Step 2. Extract data using the Feature Extraction program

The Feature Extraction software v10.5 or higher supports extraction of microarray TIFF images (.tif) of Agilent Methylation microarrays scanned on the Agilent SureScan or C Scanner.

The Feature Extraction software v9.5 supports extraction of microarray TIFF images (.tif) of Agilent Methylation microarrays scanned on the Agilent B Scanner.

Feature Extraction version 10.7 or higher can automatically download Grid Templates, protocols and QC metrics (QCMT) from SureDesign if configured appropriately. See Automatic Download from SureDesign on page 60 for configuration.

Figure 24 shows an example of an Agilent SurePrint G3 11M microarray image opened in the Feature Extraction software.

Table 20 C Scanner Scan Settings

For HD Microarray Formats For G3 Microarray Formats

Dye channel R+G (red and green) R+G (red and green)

Scan region Agilent HD (61 x 21.6 mm) Agilent HD (61 x 21.6 mm)

Scan resolution 5 m 3 m

Tiff file dynamic range 16 bit 16 bit

Red PMT gain 100% 100%

Green PMT gain 100% 100%

XDR

56 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 2. Extract data using the Feature Extraction program

Figure 24. Agilent SurePrint G3 11M microarray shown in red and green channels, full and zoomed view

1 Open the Agilent Feature Extraction program. 2 Add the images (.tif) to be extracted to the Feature Extraction Project.

a Click Add New Extraction Set(s) icon on the toolbar or right-click the Project Explorer and select Add Extraction...

You can also drag the image (.tif) from the desktop to the Feature Extraction project pane.

b Browse to the location of the .tif files, select the .tif file(s) and click Open. To select multiple files, use the Shift or Ctrl key when selecting.

The Feature Extraction program automatically assigns a default grid template and protocol for each extraction set, if the following conditions are met:

As of v10.5, the Feature Extraction program automatically associates the protocol for a given microarray based on the application specified in the design file and the number of channels present in the image. If you need to use a protocol other than the Agilent default protocol, specify it in the Grid Template properties.

For auto assignment of the ChIP Feature Extraction protocol, the default ChIP protocol must be specified in the Feature Extraction Grid Template properties.

To access the Feature Extraction Grid Template properties, double-click on the grid template in the Grid Template Browser.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 57

4 Microarray Processing and Feature Extraction Step 2. Extract data using the Feature Extraction program

3 Set Feature Extraction Project Properties. a Select the Project Properties tab. b In the General section, enter your name in the Operator field. c In all other sections, verify that at least the following default settings as

shown in Figure 25 below are selected. d For Feature Extraction 9.5, in the Other section, select CGH_QCMT_Feb08.

For Feature Extraction 10.5 or higher, the metric sets are part of the protocol, and there is no need to set them.

QC metrics updates are available automatically from SureDesign if configured appropriately. See Automatic Download from SureDesign on page 60 for configuration.

Figure 25. Default settings in Feature Extraction 10.5

4 Check the Extraction Set Configuration. a Select the Extraction Set Configuration tab. b Verify that the correct grid template is assigned to each extraction set in

the Grid Name column. To assign a different grid template to an extraction set, select one from the pull down menu.

58 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

4 Microarray Processing and Feature Extraction Step 2. Extract data using the Feature Extraction program

If a grid template is not available to select from the pull down menu, you must add it to the Grid Template Browser. To add, right-click inside the Grid Template Browser, select Add. Look for the design file (.xml) and click Open to load grid template into the Feature Extraction database.

To update to the latest grid templates via Online Update, right-click Grid Template Browser and select Online Update. You can also download the latest grid templates from Agilent Web site at www.agilent.com/genomics/SureDesign. After downloading, add the grid template to the Grid Template Browser.

c Verify that the most recent protocol is assigned to each extraction set in the Protocol Name column.

If a protocol is not available to select from the pull down menu, you must import it to the Feature Extraction Protocol Browser. To import, right-click the Feature Extraction Protocol Browser, select Import. Browse for the Feature Extraction protocol (.xml) and click Open to load the protocol into the Feature Extraction database. Visit Agilent Web site at www.genomics.agilent.com/article.jsp?pageId=2058 to download the latest protocols.

Protocols are also available automatically from SureDesign if configured appropriately. See Automatic Download from SureDesign on page 60 for configuration.

5 Save the Feature Extraction Project (.fep) by selecting File > Save As and browse for desired location.

6 Verify that the icons for the image files in the Feature Extraction Project Window no longer have a red X through them. A red X through the icon indicates that an extraction protocol was not selected or that the Grid Template is not in the database. If needed, reselect the extraction protocol for that image file.

7 Select Project > Start Extracting. 8 After the extraction is completed successfully, view the QC report for each

extraction set by double-clicking the QC Report link in the Summary Report tab. Determine whether the grid has been properly placed by inspecting Spot Finding of the Four Corners of the Array.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 59

4 Microarray Processing and Feature Extraction Step 2. Extract data using the Feature Extraction program

Automatic Download from SureDesign

Feature Extraction version 10.7 or higher can automatically download Grid Templates, protocols and QC metrics (QCM or QCMT). To set this up, in the eArray Login Setting dialog box, under Advanced Options, click Use eArray server during extraction. See Figure 26.

Figure 26. eArray Login Setting. You can mark the other two check boxes under Advanced Options if you want to get update of grid templates already in the database or to get protocol updates. See the Feature Extraction user guide for more information.

60 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

5 Reference Reagent Kit Components 62 Secure Fit Slide Box Opening Instructions 63 Microarray Handling Tips 65 Agilent Microarray Layout and Orientation 66 Array/Sample tracking on microarray slides 68 Notes and Considerations 69

This chapter contains reference information that pertains to this protocol.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 61

5 Reference Reagent Kit Components

Reagent Kit Components

The contents of the reagent kits used in this protocol are listed here.

SureTag DNA Labeling Kit 10 Restriction Enzyme Buffer*

BSA* Alu I* Rsa I* Purification Column Nuclease-Free Water Exo (-) Klenow 5 Reaction Buffer Cyanine 5-dUTP Cyanine 3-dUTP 10 dNTPs Random Primer

Oligo aCGH/ChIP-on-chip Hybridization Kit 2 HI-RPM Hybridization Buffer 10 aCGH Blocking Agent

Oligo aCGH/ChIP-on-chip Wash Buffer Kit Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2

* Not used in this protocol.

62 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

5 Reference Secure Fit Slide Box Opening Instructions

Secure Fit Slide Box Opening Instructions

Agilent now ships all microarray slides in a newly designed secure fit slide box. The instructions below describe how to remove the slide box from the shipping pouch, how to open the slide box, and how to properly remove a microarray slide. 1 Use scissors to cut below the seal and remove box from its foil pouch.

After breaking foil on microarray pouch, store microarray slides in the slide box at room temperature (in the dark) under a vacuum desiccator or nitrogen purge box.

Figure 27. Opening foil pouch (left) and removing the slide box (right)

2 Place the slide box on a flat surface. While stabilizing the box from the top with one hand, use a sharp edge to cut the sealing tape on both sides of the box before opening.

Figure 28. Cutting the sealing tape

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 63

5 Reference Secure Fit Slide Box Opening Instructions

3 With one hand, firmly hold the base of the box on the sides with the indentations (or dimples) for added grip.

Figure 29. Gripping the base at the indentations (top) and close-up of the indentations (bottom)

4 Using your free hand, grasp the lid and gently lift it away from the base as if it is hinged. Set the lid aside.

Depending on your preference, you can reverse the hand positions so that the left hand holds the base while the right hand grasps the lid.

Figure 30. Grasping the lid (left) and lifting the lid from the base (right)

64 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

5 Reference Microarray Handling Tips

Microarray Handling Tips

Each microarray is printed on the side of the glass slide containing the Agilent-labeled barcode. This side is called the active side. The numeric barcode is on the inactive side of the slide.

In this processing and hybridization procedure, the hybridization mixture is applied directly to the gasket slide, and not to the active side of the oligo microarray. Instead, the active side of the oligo microarray is placed on top of the gasket slide to form a sandwich slide pair.

To avoid damaging the microarray, always handle glass slides carefully by their edges. Wear powder-free gloves. Never touch the surfaces of the slides. If you do, you may cause irreparable damage to the microarray.

Never allow the microarray surface to dry out during the hybridization process and washing steps.

CAUTION You must familiarize yourself with the assembly and disassembly instructions for use with the Agilent Microarray Hybridization Chamber (G2534A) and gasket slides. Practice slide kits are available.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 65

5 Reference Agilent Microarray Layout and Orientation

Agilent Microarray Layout and Orientation

Agilent oligo microarray (1 microarray/slide format) as imaged on the Agilent microarray scanner

Figure 31. Agilent microarray slide and slide holder

Agilent oligo microarrays formats and the resulting microarray design files are based on how the Agilent microarray scanner images 1-inch x 3-inch glass slides. Agilent designed its microarray scanner to scan through the glass slide (back side scanning). The glass slide is securely placed in an Agilent microarray slide holder with the Agilent labeled barcode facing the opening of the slide holder (on SureScan Microarray Scanner) or facing the inside of the slide holder (Scanner C or Scanner B). In this orientation, the active side containing the microarrays is protected from potential damage by fingerprints and other elements. Once securely placed, the numeric barcode, non-active side of the slide, is visible from the outside of the slide holder.

Figure 31 depicts how the Agilent microarray scanner reads the microarrays and how this relates to the microarray design files that Agilent generates during the manufacturing process of its in situ-synthesized oligonucleotide microarrays. Thus, if you have a scanner that reads microarrays from the front side of the glass slide, the collection of microarray data points will be different in relation to the microarray design files. Therefore, please take a moment to become familiar with the microarray layouts for each of the Agilent oligo microarrays and the layout information as it pertains to scanning using a front side scanner.

Microarrays are printed on the side of the glass with the Agilent-labeled barcode (also referenced as "active side" or "front side").

Agilent Microarray Scanner scans through the glass. (Back side scanning.)

00116789

Agilent microarray slide holder for SureScan (above) and Scanner C (below) microarray scanners.

66 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

5 Reference Agilent Microarray Layout and Orientation

Non-Agilent Front Side Microarray Scanners

When scanning Agilent oligo microarray slides, the user must determine: If the scanner images the microarrays by reading them on the front side of

the glass slide (Agilent-labeled barcode side of the slide) and If the microarray image produced by the non-Agilent scanner is oriented in a

portrait or landscape mode, and Agilent-labeled barcode is on the left-side, right-side, up or down, as viewed as an image in the imaging software (see Figure 32).

This changes the feature numbering and location as it relates to the microarray design files.

Figure 32. Microarray slide orientation

Front side barcode left (landscape)

Front side barcode right

(landscape)

Front side barcode up (portrait)

Front side barcode down

(portrait)

Agilent

A gi

le nt

Agilent

A gilent

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 67

5 Reference Array/Sample tracking on microarray slides

Array/Sample tracking on microarray slides

Use the forms below to make notes to track your samples on microarray slides.

Figure 33. 2-pack microarray slides

Figure 34. 4-pack microarray slides

Array 1_ Array 1_2

Sample: Sample:

Arrays

B A R C O D E

Barcode Number ______________________________________________________________

Array 1_1 Array 1_2 Array 1_3 Array 1_4

Sample: Sample: Sample ample:

Arrays

B A R C O D E

Barcode Number __________________________________________________________

68 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

5 ReferenceReference Notes and Considerations

Notes and Considerations

Methylated DNA IP lets investigators capture DNA sequences and study their relative methylation levels across an entire genome. The protocol requires an antibody to 5-methyl cytosine that will immunoprecipitate methylated DNA from a DNA sample.

The Methylated DNA Immunoprecipitation protocol consists of five general steps: DNA shearing Methylated DNA immunoprecipitation DNA labeling Microarray hybridization and washing Microarray scanning and storage

This reference summarizes the goals and steps for this protocol. Other considerations outside of this protocol include initial probe and microarray design and the design and implementation of robust quantitative metrics that validate success at multiple steps of the protocol.

1. Chromatin immunoprecipitation (ChIP)

Goal Use selective antibody bound to magnetic beads to specifically capture the Methylated DNA.

SOP 1 Mix antibody bound to magnetic beads (Dynal) with DNA sample. 2 Place at 4C overnight on a rotating platform. 3 Isolate the beads containing the antibody bound to the methylated DNA. 4 Wash 2 times with buffer to remove non-specific contaminants. 5 After the wash, heat the complexes for a few minutes with detergent to elute

the methylated DNA from the antibody and beads.

Key variables Beads, type, and quantity Time Temperature Immunoprecipitation buffer, volume, and composition Wash buffer composition Number of washes

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 69

5 ReferenceReference Notes and Considerations

QC Metrics After the DNA is isolated (step 2), you can run the Bioanalyzer or gel equivalent to assess fragment size after shearing. Use 1 L for the Bioanalyzer, or 5 L on a 1.5% agarose gel.

Notes Magnetic beads coated with protein G are routinely used due to their ease-of-use and ability to bind a variety of antibodies. Other coatings (e.g. protein A) and bead types (e.g. agarose) are available but have not been validated by Agilent.

Figure 35. Human liver DNA (Biochain p/n D1234149) was sonicated following the procedure outlined in Step 2. Prepare DNA for Immunoprecipitation on page 19. 1 L (~20 ng) of sheared DNA was analyzed using the Agilent 2100 Bioanalyzer in conjunction with the DNA High Sensitivity Kit (p/n 5067-4626). The expected size distribution is between 200 and 1000 bp.

2. DNA isolation

Goal Purify DNA from associated proteins and RNA and protein contaminants.

SOP 1 Purify the DNA via organic extraction and ethanol precipitation.

Key variables Temperature Time SDS concentration

QC Metrics None

3. DNA labeling

Goal Incorporate fluorescent-tagged nucleotides into the IP and reference DNA for hybridization.

SOP 1 Use the Agilent SureTag DNA Labeling Kit. 2 Do 1 Labeling reaction for each sample. Approximately 1 g input per reaction

for the reference channel cyanine 3, and all IP DNA for the cyanine 5. 3 Anneal random primers to the DNA. 4 Extend primers using high concentration exo- Klenow enzyme and

fluorescent-labeled nucleotides. 5 Purify labeled DNA using the Amicon 30 kDa columns.

70 Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol

5 ReferenceReference Notes and Considerations

Key variables Reaction size Reagent quantity (input DNA material, Cy dye, enzyme) per reaction

QC Metrics Nanodrop measurement of total DNA yield (expect >2.5 g per reaction); Nanodrop measurement of pmol/L dye (expect >2 pmol/L with Cy5-dUTP and >3 pmol/L with Cy3-dUTP).

4. Microarray hybridization and washing

Goal Hybridize material to and wash excess/nonspecific material from Agilent 60-mer oligo arrays to yield low background and high signal (flat background with high peaks)

SOP 1 Hybridize for 40 hours at 67C in hybridization oven rotating at 20 rpm. Hybridization buffer contains a proprietary wetting agent (that keeps bubbles moving freely), approximately 5 g labeled DNA per channel (10 g total) and competitor nucleic acids.

2 Wash slides in a series of two buffers with an optional 3rd wash that contains ozone-scavenging reagents to help prevent premature dye degradation.

Key variables Hybridization duration Quantity of labeled material Temperature Type and quality of detergent Type and quantity of nucleic acid competitors

Notes These conditions are identical to those developed for Agilent aCGH hybridizations. Refer to the Bioreagent Wash/Dry Solution application note for more information. The wash conditions are specific for Agilent's ChIP-on-chip application.

5. Microarray scanning and storage

Goal Extract data from microarray; store microarray for possible future analysis

SOP 1 Use default settings on Agilent scanner.

2 Store used slides in N2 box.

Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Protocol 71

Manualsnet FAQs

If you want to find out how the Microarray Analysis of Methylated DNA Immunoprecipitation Agilent works, you can view and download the Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Platform Protocol on the Manualsnet website.

Yes, we have the Protocol for Agilent Microarray Analysis of Methylated DNA Immunoprecipitation as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Protocol should include all the details that are needed to use a Agilent Microarray Analysis of Methylated DNA Immunoprecipitation. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Platform Protocol is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Platform Protocol consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Platform Protocol free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Platform Protocol, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent Microarray Analysis of Methylated DNA Immunoprecipitation Platform Protocol as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.