The G1289 Headspace Sampler requires an interface connection to the 6890 GCs carrier gas line and Auxiliary EPC control module.

Kit parts list Part Quantity

Tube, 520mm 1

Nut, Transfer Line 1

Ferrule, Transfer Line 1

ZDV Union 2

1/8-inch ferrule set, brass, Swagelok 1

PDD Self-seal Bag 3

Knife File 1

Installation Guide 1

Zero Tube Restictor 3

O-rings (6 pack) 1

Installing the Agilent 7694 Head- space Sampler Interface Kit for the Agilent 6890 Gas Chromatograph Accesory G1940A

2

Required tools Pliers

File (included in kit)

7-mm open-end wrench, 2 each

7/16-inch open-end wrench, 2 each

Installation This installation places the Headspace Sampler between the GC inlet flow control and the inlet itself. Since the GC is controlling the flows and pressures, these setpoints are part of the GC method and easy to reproduce.

The G1289 sampler does not contain a flow controller or a pressure regulator. Carrier gas is supplied by a GC inlet control channel. Vial pressurization gas is supplied by an Auxiliary gas channel in the GC.

1. If the GC does not have an Auxiliary EPC module, install the module provided.

2. On the Auxiliary module, locate the block that connects the three gas outlet tubes for the auxiliary channels to the pneumatics module.

3. Remove the screw that holds the block to the pneumatics module. Pull the block free of the module and rotate it so that the frits are on top.

Aux 5

Aux 4

Aux 3

Aux 3

Aux 4

Aux 5

Frits

Outputs

3

4. Decide which channel you will use for the vial pressurization control. Pull the frit for that channel out of the block. Remove the O-ring that seals it.

5. Place an O-ring on a zero-resistance brass tube frit (part no. G1570- 20540). Place the O-ring/frit combination in the block.

6. Reconnect the block to the pneumatics module. Tighten the screw firmly.

7. Use tubing and Swagelok fittings to connect the appropriate Auxiliary module output, the bleed weldment, and the VIAL PRESSURE fitting on the back of the Headspace Sampler as shown. Do not disturb the brass end on the bleed weldment.

8. Install a deactivated splitless liner (part no. 5181-8818) in the inlet.

9. Locate the carrier line to the split/splitless inlet.

Headspace Sampler 6890 GC

VIAL PRESSURE fitting To Auxiliary module output

Bleed weldment

Brass end

Carrier gas line

4

10. Cut the carrier line to the split/splitless inlet about 1 to 2 inches from the inlet body.

a. Carefully score the circumference of the tubing.

b. Hold the tubing with pliers on each side of the score and flex it until it breaks free.

c. File the edges until they are smooth.

11. Slide the nut onto the inlet side of the carrier line.

12. Attach a zero-dead-volume (ZDV) union to the short piece of carrier line attached to the inlet.

13. Attach the headspace sampler transfer line to the ZDV union.

14. Use the other ZDV union to connect the free end of the carrier gas line to the length of tubing provided. Route the tubing out the back of the GC to the rear panel of the sampler.

15. Connect the tubing to the CARRIER fitting on the sampler.

ZDV union

Carrier line Ferrule

Nut

ZDV union

Transfer line

5

Manualsnet FAQs

If you want to find out how the 7694 Agilent works, you can view and download the Agilent 7694 Interface Kit Installation Manual on the Manualsnet website.

Yes, we have the Installation Manual for Agilent 7694 as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Installation Manual should include all the details that are needed to use a Agilent 7694. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent 7694 Interface Kit Installation Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent 7694 Interface Kit Installation Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent 7694 Interface Kit Installation Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent 7694 Interface Kit Installation Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent 7694 Interface Kit Installation Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.