G5585B PlateLoc Thermal Microplate Sealer

Quick Start

Original Instructions

Notices Agilent Technologies, Inc. 2018

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

User Guide Part Number

G5051-90001

Edition

Revision A, June 2018

Contact Information

Web: https://www.agilent.com

Contact page: https://www.agilent.com/en/contact-us/ page

Documentation feedback: documentation.automation@agilent.com

Acknowledgements

Microsoft and Windows are either registered trademarks or trademarks of the Microsoft Corporation in the United States and other countries.

Warranty

The material contained in this docu- ment is provided as is, and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor- mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend

If software is for use in the performance of a U.S. Government prime contract or sub- contract, Software is delivered and licensed as Commercial computer soft- ware as defined in DFAR 252.227-7014 (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Gov- ernment will receive no greater than Restricted Rights as defined in FAR 52.227- 19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14

(June1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.

Safety Notices

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

PlateLoc Thermal Microplate Sealer Quick Start

1

PlateLoc Thermal Microplate Sealer Quick Start

This quick start guide summarizes the operator instructions in the PlateLoc Thermal Microplate Sealer User Guide.

This guide contains the following topics:

About this guide on page 2

Safety information on page 2

Getting started on page 4

Sealing microplates on page 12

Touchscreen menus and commands on page 19

For user information about related products, you can search the product knowledge base or download the latest version of a PDF file from the Agilent website at:

http://www.agilent.com/en- us/products/automation- solutions

PlateLoc Thermal Microplate Sealer Quick Start About this guide

2 PlateLoc Thermal Microplate Sealer Quick Start

About this guide

This edition documents the PlateLoc color touchscreen. If your device has a black- and- white touchscreen, see the previous edition of the PlateLoc Thermal Microplate Sealer Quick Start (part number G5402- 90014A).

Safety information

Before using the PlateLoc Sealer

Before using the PlateLoc Sealer, make sure that you are properly trained in:

General laboratory safety

The correct and safe operation of the PlateLoc Sealer

The correct and safe operation of lab automation systems or components used in combination with the PlateLoc Sealer

If you are the person in your organization responsible for training others on the PlateLoc Sealer and you have a safety question, contact Agilent Automation Solutions Technical Support.

Safety standards

The PlateLoc Sealer complies with the applicable EU Directives and bears the CE mark. See the Declaration of Conformity for compliance details.

General precautions

For the general safety precautions, the intended product use statement, and the list of safety labels, see the Automation Solutions Products General Safety Guide.

3

PlateLoc Thermal Microplate Sealer Quick Start Safety information

PlateLoc Thermal Microplate Sealer Quick Start

Potential hazards specific to the PlateLoc Sealer

Chemical hazards

High-temperature hazard

Moving-parts hazards The PlateLoc Sealer contains moving parts that can cause injury. Under normal operating conditions, the PlateLoc Sealer is designed to protect you from the moving parts. The door sensor is designed so that the seal cycle cannot start unless the door is closed.

WARNING To avoid potential injury, do not use chemicals that have low flash points in the PlateLoc Sealer. Such chemicals include reagents that are used in radioactive scintillation proximity assays (SPAs) and solvents, including alcohols.

WARNING The hot surfaces in the PlateLoc Sealer can cause a burn injury. Do not touch the PlateLoc Sealer or pry open the door while it is in operation. Use care when accessing the PlateLoc Sealer interior after the device is warmed up. Do not reach into the PlateLoc Sealer when it is warmed up to a high temperature. Do not touch the hot plate with your bare hands.

WARNING A freshly sealed microplate might still be hot from the sealing process. To prevent potential injury, use caution when removing a sealed microplate from the PlateLoc Sealer.

PlateLoc

WARNING Do not disable the door sensor or attempt to access the interior of the device through any other openings. Exposure to the moving parts, such as the heated metal plate (hot plate) or the seal-cutting blade can cause severe injuries.

PlateLoc Thermal Microplate Sealer Quick Start Getting started

4 PlateLoc Thermal Microplate Sealer Quick Start

Getting started

Turning on and turning off the power

Before turning on the PlateLoc Sealer, make sure:

The lab requirements are met.

You have properly installed the PlateLoc Sealer.

You are trained in the proper operation of the PlateLoc Sealer.

The on/off switch is on the back of the PlateLoc Sealer.

To turn on the PlateLoc Sealer: Press the on/off switch to the on position (I). If you have not yet turned on the air, the touchscreen displays the Insufficient Air Pressure error message. See Turning on and turning off the air on page 4. If you have not yet connected the air supply to the PlateLoc Sealer, see the PlateLoc Thermal Microplate Sealer User Guide.

To turn off the PlateLoc Sealer:

Press the on/off switch to the off position (o). The touchscreen turns off.

Turning on and turning off the air

Before turning on the air, make sure:

You have properly installed the PlateLoc Sealer.

The air is turned on at the source (house, cylinder, or pump) and the air pressure and flow rate meet the PlateLoc Sealer requirements.

The PlateLoc Sealer is turned on.

00191 PlateLoc connections

Lower gripper to load seal

Press to close gripper

5

PlateLoc Thermal Microplate Sealer Quick Start Getting started

PlateLoc Thermal Microplate Sealer Quick Start

The AIR ON/OFF switch is on the back of the PlateLoc Sealer.

To turn on the air:

Push the AIR ON/OFF switch to the ON position.

When you turn on the air, the PlateLoc Sealer door opens. If this is the first time you are turning on the air, you must remove the packing foam inside the sealing chamber before you can use the device.

To check that air is flowing into the PlateLoc Sealer:

Check the pressure gauge on the back of the PlateLoc Sealer. If air is flowing into the device, the gauge should show a non- zero pressure value.

If the gauge shows 0 psi, check the air tubing connections and the connections at the source (house, cylinder, or pump).

IMPORTANT The pressure reading on the back of the PlateLoc Sealer might be different from the value shown on the touchscreen. The pressure gauge on the back of the PlateLoc Sealer should be used to determine the presence of air flow only. To check the air pressure inside of the device, view the air pressure information displayed on the touchscreen.

00191 PlateLoc connections

Lower gripper to load seal

Press to close gripper

00191 PlateLoc connections

PlateLoc Thermal Microplate Sealer Quick Start Getting started

6 PlateLoc Thermal Microplate Sealer Quick Start

To check the internal regulated air pressure:

On the touchscreen Home page, locate the Air Pressure readout. The regulated air pressure should be approximately 0.5990.62 MPa (8790 psi).

To turn off the air: Push the AIR ON/OFF switch to the OFF position.

Loading and unloading a roll of seal

Make sure you have the supplied seal- loading card.

To warm up or cool down the PlateLoc Sealer:

1 On the touchscreen, press to display the Settings pages.

2 Set the Seal Temp to a value from 30 to 40 C.

Note: Press or to adjust the value down or up in 0.5 C increments.

3 Press Save to save the new Seal Temp value.

4 Press to return to the Home page. A status message appears at the

top of the page, indicating that the PlateLoc Sealer is warming up or cooling down to the new set temperature.

7

PlateLoc Thermal Microplate Sealer Quick Start Getting started

PlateLoc Thermal Microplate Sealer Quick Start

5 Wait for the PlateLoc Sealer to warm up or cool down so that the Seal Temp is within 30 to 40 C as displayed on the Home page.

To assemble a roll of seal on the axle and hubs:

1 Remove the seal roll (1), seal- roll hubs (2), and axle (3) from their packaging. See the following figure.

2 Screw one of the hubs onto the axle.

3 Place the axle through the hole at the center of the seal roll. Make sure the hub fits snugly against the roll.

4 Screw the second hub onto the axle so that the roll is firmly secured between both hubs.

00195 PlateLoc seal roll

1

2

3

PlateLoc Thermal Microplate Sealer Quick Start Getting started

8 PlateLoc Thermal Microplate Sealer Quick Start

5 Using scissors, cut off any wrinkled or torn material from the end of the seal so that the end of the seal has a clean, straight edge.

To mount the seal roll on the seal-roll supports:

1 Orient the roll, and set the assembled seal roll onto the seal- roll supports (1) as shown in the following figure. Notice that the end of the seal material rolls out from under the roll.

2 Pull the seal from under the cross- beam (2), as the following figure shows.

To insert the seal into the PlateLoc Sealer:

1 On the back of the PlateLoc Sealer, push down the red recessed open- gripper button until it clicks.

Pressing the button opens the seal gripper inside the device.

2 Remove the strip of paper that covers the adhesive on the end of the seal- loading card.

3 Follow the instructions on the card to:

a Attach the end of the seal to the card.

b Insert the card into the seal entry slot on the back of the PlateLoc Sealer.

c Push the card into the PlateLoc Sealer until it protrudes beyond the chamber doorway at the front of the device.

CAUTION The PlateLoc Sealer will not seal microplates properly if the seal is wrinkled or torn.

1

2

00191 PlateLoc connections

9

PlateLoc Thermal Microplate Sealer Quick Start Getting started

PlateLoc Thermal Microplate Sealer Quick Start

d Pull the card through the PlateLoc Sealer so that the end of the seal is past the plate stage.

4 While maintaining the alignment of the seal with the plate stage, press the silver close- gripper button on the back of the PlateLoc Sealer until it clicks. The red open- gripper button pops back up.

Pressing the silver button closes the gripper and holds the seal in place inside the device.

00203 PlateLoc remove loading card

PlateLoc

00210 PlateLoc Loading card removal

PlateLoc

PlateLoc Thermal Microplate Sealer Quick Start Getting started

10 PlateLoc Thermal Microplate Sealer Quick Start

5 Turn the seal roll to remove excess slack in the seal.

6 Load a plate stage on the plate- stage support.

7 Load a spare microplate on the plate stage.

8 Hold the seal- loading card at the touchscreen level, keeping the tension on the card.

9 On the touchscreen Home page , press Run. Continue to keep the

tension on the card as the seal feeds through the PlateLoc Sealer.

The purpose of this seal cycle is to cut the seal and prepare it for use. The seal is not applied to the spare microplate.

10 If an error displays on the touchscreen, press Clear to clear the error. Press Run again to make sure that the seal will apply to the spare microplate.

00191 PlateLoc connections

00 19

3

11

PlateLoc Thermal Microplate Sealer Quick Start Getting started

PlateLoc Thermal Microplate Sealer Quick Start

Unloading procedure

To unload a roll of seal:

1 Power cycle the PlateLoc Sealer by turning off and turning on the device.

Power cycling resets the components inside the device and removes the hold on the seal.

2 On the back of the PlateLoc Sealer, push down the red recessed gripper- release button until it clicks.

Pressing the button opens the seal gripper inside the device.

3 Pull the seal out of the slot.

4 Lift the seal roll off of the seal- roll supports.

5 Unscrew one hub from the axle.

6 Remove the roll from the axle.

00191 PlateLoc connections

PlateLoc Thermal Microplate Sealer Quick Start Sealing microplates

12 PlateLoc Thermal Microplate Sealer Quick Start

Sealing microplates

Loading an insert and a microplate

Before you load an insert or microplate, make sure:

The PlateLoc Sealer is turned on.

The air is turned on and the door is open.

The seal is properly loaded.

Lab automation systems only. Communication between the controlling computer and the PlateLoc Sealer is established.

The sealing parameters are set correctly.

The PlateLoc Sealer is warmed up or cooled down to the sealing temperature.

Standalone device procedure

To load a microplate and an insert in the standalone device:

1 Place the removable plate stage (1a) on the plate- stage support (1b), as the following figure shows.

2 Optional. Place the insert (2) on the plate stage.

3 Place the microplate (3) on the insert or plate stage. Make sure the microplate is level and the microplate skirt is within the raised tabs on the plate stage.

00198 PlateLoc plate insert

0 0 1 9 8

PlateLoc

3

2

1a

1b

13

PlateLoc Thermal Microplate Sealer Quick Start Sealing microplates

PlateLoc Thermal Microplate Sealer Quick Start

Lab automation system procedure If the PlateLoc Sealer is installed in a lab automation system, you need to load the insert before you start the protocol run. During the run, the robot loads the microplates automatically. You do not need to manually load each microplate.

Setting the sealing parameters

Before you set the sealing parameters, make sure:

The PlateLoc Sealer is turned on.

The air is turned on.

The seal is properly loaded.

Lab automation systems only. Communication between the controlling computer and the PlateLoc Sealer is established.

Agilent Technologies recommends that you perform seal optimization tests to determine the optimal seal time and temperature. For general optimization guidelines, see the PlateLoc Thermal Microplate Sealer User Guide.

The seal time and temperature you specify depends on the type of microplate and the type of seal you are using. For the most up- to- date starting point settings, see the PlateLoc Thermal Plate Sealer Seal Selection Guide. You can locate the guide in the Literature section of the product page at https://www.agilent.com/en/products/automation- solutions/microplate- management- robotics/plateloc- thermal- microplate- sealer

Standalone device procedure

To set the sealing parameters on the touchscreen:

1 Press to display the Settings pages.

2 In the Settings 1/3 page, set the following values:

Seal Time. The duration (seconds) that the hot plate will be in contact with the seal material and microplate.

Seal Temp. The hot plate temperature (C) when the seal cycle starts.

Gas Purge Time. Gas- Purging PlateLocs only. The amount of time (seconds) the microplate will be exposed to argon before the microplate is sealed.

Press or to change the value for the corresponding setting.

PlateLoc Thermal Microplate Sealer Quick Start Sealing microplates

14 PlateLoc Thermal Microplate Sealer Quick Start

3 Press Save to save the new settings.

The touchscreen automatically returns to the Home page. Status

messages appear at the top of the Home page, such as, Warming up, Cooling down, or Ready.

You can also monitor the following:

Seal Time displays the time remaining in the seal cycle.

Seal Temp displays the current temperature. The text turns blue when the hot plate temperature is within Set Temp run temperature range.

Note: The touchscreen readout displays the temperature as measured on the opposite side of the hot plate sealing surface. The actual temperature of the sealing surface may vary slightly.

15

PlateLoc Thermal Microplate Sealer Quick Start Sealing microplates

PlateLoc Thermal Microplate Sealer Quick Start

Selecting and editing a saved method (touchscreen)

To select and edit a saved plate sealing method:

1 On the touchscreen, press to display the Preset page.

2 Press or to move the highlight to the desired Preset row, and then do one of the following:

To use the highlighted Preset without changing it, press Select. The selected values are loaded to settings on the Home page.

Status messages appear at the top of the Home page, such as, Warming up, Cooling down, or Ready.

To edit the highlighted Preset, press Edit, and then proceed to step 3.

3 In the Preset edit page, set the following:

Seal Time. The duration (seconds) that the hot plate will be in contact with the seal material and microplate.

Seal Temp. The hot plate temperature (C) when the seal cycle starts.

Gas Purge Time. Gas- Purging PlateLocs only. The amount of time (seconds) the microplate will be exposed to argon before the microplate is sealed.

Press or to change the value for the corresponding setting.

PlateLoc Thermal Microplate Sealer Quick Start Sealing microplates

16 PlateLoc Thermal Microplate Sealer Quick Start

4 Click Save to save the edits for the Preset.

5 To load the values for the edited Preset, highlight the corresponding Preset row on the Preset page, and then press Select.

6 Press to return to the Home page.

The selected values are loaded to settings on the Home page. Status messages appear at the top of the Home page, such as, Warming up, Cooling down, or Ready.

Lab automation system procedure In the VWorks software, you set the sealing parameters when you set the Seal (PlateLoc) task properties. For details, see the VWorks Automation Control User Guide.

If the PlateLoc Sealer is integrated with third- party software, see the PlateLoc Sealer ActiveX version 13.0 User Guide.

Starting the seal cycle

Before you set the seal cycle, make sure:

The PlateLoc Sealer is turned on.

The air is turned on and the internal air pressure is 8790 psi.

The seal is properly loaded.

Lab automation systems only. Communication between the controlling computer and the PlateLoc Sealer is established.

You have set the correct sealing parameters.

Standalone PlateLoc Sealer only. The microplate is loaded.

17

PlateLoc Thermal Microplate Sealer Quick Start Sealing microplates

PlateLoc Thermal Microplate Sealer Quick Start

Standalone device procedure

To use the touchscreen to start the seal cycle:

1 If not already displayed, press to go to the Home page.

2 Verify the following:

The correct values are specified for Set Time and Set Temp.

At the top of the page, the status bar displays Ready.

3 Press Run. The microplate moves into the sealing chamber and the door closes. The seal cycle begins. The status bar at the top of the touchscreen displays In Progress.

Gas- Purging PlateLoc only. If the Gas Purge Time is greater than 0.0 seconds, the argon fills the sealing chamber for the specified amount of time before the seal is applied. The argon automatically turns off just before the seal is applied.

Note: If the Run button is unavailable, the PlateLoc Sealer is warming up or cooling down. You cannot start the seal cycle until the PlateLoc Sealer is within the run temperature range of the specified Set Temp.

Note: If an error occurs during the seal cycle, the cycle is aborted.

After the seal cycle is finished: When the seal cycle is finished, the Status message changes to Ready. The door opens. The microplate moves out of the sealing chamber.

PlateLoc Thermal Microplate Sealer Quick Start Sealing microplates

18 PlateLoc Thermal Microplate Sealer Quick Start

Wait a few seconds, and then remove the sealed microplate and insert from the plate stage.

If you have finished using the PlateLoc Sealer:

1 Turn off the PlateLoc Sealer.

2 Turn off the air at the back of the PlateLoc Sealer or at the source (house, cylinder, or pump).

Lab automation system procedure You can start one or more seal cycles when you start the protocol run in the VWorks software. For detailed instructions, see the VWorks Automation Control User Guide.

Stopping a seal cycle

Standalone device procedure

To use the touchscreen to stop a seal cycle:

On the Home page, press Stop.

WARNING Do not handle the microplate immediately after the seal cycle is finished. The sealed microplate and the insert might be hot.

19

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

PlateLoc Thermal Microplate Sealer Quick Start

To stop a seal cycle in an emergency: If the Stop button on the touchscreen is not responding fast enough, turn off the PlateLoc Sealer (1). The on/off switch is on the back of the device. Alternatively, you can disconnect the power to the PlateLoc Sealer (2).

Lab automation system procedure For instructions on how to stop a seal cycle from the VWorks software, see the VWorks Automation Control User Guide.

Touchscreen menus and commands

Navigation pane

00191 PlateLoc connections

1 2

Icon Description

Displays the Home page.

Displays the Settings pages, which enable you to adjust sealing parameters and view information about the device.

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

20 PlateLoc Thermal Microplate Sealer Quick Start

Home page

Displays the Preset page, which enables you to select from a list of saved plate- sealing methods.

Agilent factory and service personnel only.

Displays a keypad that authorized Agilent personnel use to display the PlateLoc service pages to set up the device and troubleshoot problems.

Icon Description

WARNING Using the commands in the Service pages can expose you to potential injury and cause equipment damage. The Service pages are for use by Agilent factory and service personnel only.

Control or indicator Description

Status bar Spans the top of the page and displays messages about the PlateLoc status. Possible values are:

Ready. The PlateLoc is within range of the specified Set Temp and is ready for a run.

Warming Up. The PlateLoc is in the process of warming up to the specified Set Temp.

Cooling Down. The PlateLoc is in the process of cooling down to the specified Set Temp.

In progress. The PlateLoc is in the process of sealing a microplate.

Error. The PlateLoc has encountered an error during the seal cycle.

21

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

PlateLoc Thermal Microplate Sealer Quick Start

Seal Time (s) Displays the time remaining (seconds) in the current seal cycle.

Note: The seal time counts down when the seal cycle starts.

Set Time (s) Displays the duration (seconds) that the hot plate will be in contact with the seal material and the microplate, as specified in the Settings pages.

Seal Temp (C) Displays the current temperature (C) of the hot plate.

The text turns blue when the hot plate temperature is within the Set Temp run temperature range.

The text is grayed when the hot plate temperature is above or below the Set Temp.

Note: The actual temperature of the sealing surface does not match the touchscreen Seal Temp readout. The touchscreen readout displays the temperature as measured on the opposite side of the hot plate sealing surface.

Set Temp (C) Displays the temperature (C) setting specified in the Settings pages.

Air Pressure Displays the device regulated air pressure (psi).

The regulated air pressure should be approximately 0.5990.62 MPa (8790 psi).

RUN/STOP Starts the seal cycle or stops the seal cycle in progress.

Note: When a seal cycle is in progress, the button changes to STOP.

Control or indicator Description

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

22 PlateLoc Thermal Microplate Sealer Quick Start

Settings pages

Settings 1/3

Control or indicator Description

Seal Time (s) Sets the amount of time (seconds) the hot plate will be in contact with the sealing material and the microplate.

Press or to adjust the value down or up in 0.5- second increments.

Default: 0.5 s

Range: 0.5 to 12.0 s

Seal Temp (C) Sets the temperature (C) of the hot plate when the seal cycle starts.

Press or to adjust the value down or up in 0.5 C increments.

Default: 20 C

Range: 20.0 to 200.0 C

Gas Purge Time (s) Available for Gas- Purging PlateLoc only. Sets the amount of time (seconds) the microplate is exposed to argon before the microplate is sealed.

Press or to adjust the value down or up in 0.5- second increments.

Default: 0.5 s

Range: 0.0 to 25.0 s

Seal Counter Displays the number of sealing cycles that have occurred since the last time the Reset button was pressed.

Reset Resets the Seal Counter to 0.

Save Saves changes to the Seal Time, Seal Temp, and, if applicable, the Gas Purge Time values.

23

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

PlateLoc Thermal Microplate Sealer Quick Start

Settings 2/3

Firmware version Displays the version of the installed firmware for the device.

Control or indicator Description

Control or indicator Description

Screen Brightness Sets the brightness of the touchscreen display. As you reduce the brightness, the text in the touchscreen dims.

To adjust the brightness, press or .

Sleep Mode Turns on or off the Sleep Mode. When Sleep Mode is on, sets the time (minutes) to wait after a seal cycle completes before cooling down the device when not in use.

Press or to select one of the following:

20 minutes

60 minutes

300 minutes

Off. Turns off the Sleep Mode.

Cycle Mode (100) Turns on or off the Cycle Mode. When set to On, the Cycle Mode places 100 seals on the current microplate.

Note: This feature is typically used only by Agilent factory and service personnel.

Press or to turn On or Off this feature.

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

24 PlateLoc Thermal Microplate Sealer Quick Start

Settings 3/3

Indicator Description

Gas Purge Displays No or Yes to indicate if the device is a Gas- Purging model.

Small Hot Plate Displays No or Yes to indicate if the device is fitted with a small hot plate.

Voltage Displays the AC voltage model of the device, which is one of the following:

120 V. The 100120 V model.

230 V. The 200240 V model.

Cycle Count Displays a running total of all the seal cycles that this device has run.

25

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

PlateLoc Thermal Microplate Sealer Quick Start

Preset page

Figure Example of Preset edit page

Control or indicator Description

Preset 01 10 Displays the list of saved plate sealing methods that you can use.

Edit Displays the Preset edit page where you can modify the seal time and temperature values for a highlighted preset.

Press or to highlight the desired row in the table.

Select Loads the settings of the highlighted Preset row to the Home page.

PlateLoc Thermal Microplate Sealer Quick Start Touchscreen menus and commands

26 PlateLoc Thermal Microplate Sealer Quick Start

Control or indicator Description

Seal Time (s) Sets the amount of time (seconds) the hot plate will be in contact with the sealing material and the microplate.

Press or to adjust the value down or up in 0.5- second increments.

Default: 0.5 s

Range: 0.5 to 12.0 s

Seal Temp (C) Sets the temperature (C) of the hot plate when the seal cycle starts.

Press or to adjust the value down or up in 0.5 C increments.

Default: 20 C

Range: 20.0 to 200.0 (C)

Gas Purge Time Gas- Purging PlateLoc only. Sets the amount of time (seconds) the microplate is exposed to argon before the microplate is sealed.

Click the left or right arrow to adjust the value down or up in 0.5- second increments.

Default: 0.5 s

Range: 0.0 to 25.0 s

Reset Resets the Preset settings to the default values.

Save Saves the changes to the specified Preset.

Manualsnet FAQs

If you want to find out how the G5585B PlateLoc Thermal Microplate Sealer Agilent works, you can view and download the Agilent G5585B PlateLoc Thermal Microplate Sealer Handler Quick Start on the Manualsnet website.

Yes, we have the Quick Start for Agilent G5585B PlateLoc Thermal Microplate Sealer as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Quick Start should include all the details that are needed to use a Agilent G5585B PlateLoc Thermal Microplate Sealer. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent G5585B PlateLoc Thermal Microplate Sealer Handler Quick Start is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent G5585B PlateLoc Thermal Microplate Sealer Handler Quick Start consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent G5585B PlateLoc Thermal Microplate Sealer Handler Quick Start free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent G5585B PlateLoc Thermal Microplate Sealer Handler Quick Start, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent G5585B PlateLoc Thermal Microplate Sealer Handler Quick Start as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.