Original Instructions

This guide contains the following topics: About this guide on page 2 Safety guidelines on page 3 Emergency stops on page 4 Hardware components and axes of motion on page 7 Labware considerations on page 11 Starting up and shutting down on page 12 Handling labware racks safely on page 14 Filling and emptying the labware racks on page 14 Installing and uninstalling labware racks on page 17 Using BenchCel Diagnostics on page 20 Cleaning up after use on page 29

G5580A BenchCel Microplate Handler Quick Guide

About this guide

2 G5580A BenchCel Microplate Handler Quick Guide

About this guide

This guide summarizes the operator instructions in the BenchCel Microplate Handler User Guide. This guide assumes the following: The BenchCel Microplate Handler is installed correctly. For details, see the

BenchCel Microplate Handler User Guide. The device profile for the specific BenchCel Microplate Handler configuration is

already created and the teachpoints are already set. For setup instructions, see the BenchCel Microplate Handler User Guide.

You are familiar with the VWorks Automation Control software. See the VWorks Automation Control Quick Start. For detailed instructions, see the VWorks Automation Control User Guide.

Where to find the user documentation

To access the user guides for Agilent Automation Solutions products, do one of the following: From within VWorks software, select Help > Knowledge Base or press F1. From the Windows desktop, select Start > Agilent Technologies >

VWorks Knowledge Base. Go to the online VWorks Knowledge Base at www.agilent.com/chem/askb.

Contacting Agilent Technologies

Web: https://www.agilent.com Contact page: https://www.agilent.com/en/contact-us/page Documentation feedback: documentation.automation@agilent.com

3

Safety guidelines

G5580A BenchCel Microplate Handler Quick Guide

Safety guidelines

WARNING Using controls, making adjustments, or performing procedures other than those specified in the user documentation can expose you to moving-parts hazards and hazardous voltage. Before using the BenchCel Microplate Handler, make sure you are aware of the potential hazards and understand how to avoid being exposed to them.

Ensure that you are trained in the safe operation of the device and that you have read the Agilent Automation Solutions Products General Safety Guide and the safety section of the BenchCel Microplate Handler User Guide.

Figure Safety label location on BenchCel Microplate Handler (front and side view)

For a description of the moving parts, see Hardware components and axes of motion on page 7. For the emergency-stop procedure, see Emergency stops on page 4.

WARNING If you touch any of the moving parts or attempt to move labware while the BenchCel Microplate Handler is in operation, the device could pinch, pierce, or bruise you. Keep your fingers, hair, clothing, and jewelry away from the device while it is in motion.

Emergency stops

4 G5580A BenchCel Microplate Handler Quick Guide

Emergency stops

The BenchCel Microplate Handler is equipped with an interlock circuit and emergency- stop pendant. When the BenchCel Microplate Handler is integrated with other devices in a third-party system, Agilent Technologies recommends that you install a main emergency-stop button to safely stop the robot and all devices simultaneously.

CAUTION You might not be able to resume a protocol after an emergency stop. Do not use an emergency stop to pause a run. To pause and continue a run, use the appropriate commands in the automation software.

Stopping in an emergency

WARNING After pressing the emergency-stop button, the robot arm might have momentum and continue to move until it comes to the end of its travel in the x-axis, z-axis, or theta- axis or until it bumps into an obstacle. Stay clear of the robot until it stops moving.

To perform an emergency stop: Press the red button on the pendant. The safety interlock circuit is interrupted, disabling the robot motors. The BenchCel Microplate Handler operation stops.

Recovering from an emergency stop

After you press the button on the emergency-stop pendant, the robot stops. One of the following occurs: If you stopped a protocol run, a dialog box opens in the VWorks software. If you stopped the robot while diagnosing problems in BenchCel Diagnostics, a

motor-disable message opens. Use the following procedure to recover in either case.

5

Emergency stops

G5580A BenchCel Microplate Handler Quick Guide

To recover the BenchCel device after an emergency stop: 1 If the robot dropped labware before or during the emergency stop, remove the

labware that was dropped. Also remove labware at teachpoints or other locations. 2 If the BenchCel robot attempted to place labware at a location that was not free, a

collision might have occurred resulting in misalignment of the robot gripper. Check the robot gripper alignment: a Move the robot arms so that they are perpendicular to the x-axis. b Make sure the bottom of the robot gripper pads are perpendicular to the robot

arms. If they are not, contact Agilent Technical Support.

Figure Gripper alignment: (A) Correct, (B) Incorrect

3 At the emergency-stop pendant, turn the button clockwise to restore power to the motors.

4 If you stopped a protocol run in an emergency, select one of the following in the VWorks message dialog box:

A B

Selection Description Diagnostics Opens the BenchCel Diagnostics dialog box. See step 5.

Note: This selection is available only when you are in the middle of a protocol run and not while you are in BenchCel Diagnostics.

Retry Attempts to restart the current command or task in the run.

Ignore and continue Ignores the current command or task and continues to the next command or task in the protocol sequence.

Abort Aborts the current command or task in the run. Select Abort if you have determined that the protocol run is not recoverable.

Emergency stops

6 G5580A BenchCel Microplate Handler Quick Guide

For a full description of the selections, see the VWorks Automation Control User Guide.

5 Optional. In BenchCel Diagnostics, use the available commands to manually move the robot or other components, including: Release the microplate that the robot is currently holding. Upstack the microplate that the robot is currently holding. Replace the lid on the microplate. Home the robot. Verify teachpoints.

If a physical crash occurred, always start BenchCel Diagnostics to home the robot and verify teachpoints.

7

Hardware components and axes of motion

G5580A BenchCel Microplate Handler Quick Guide

Hardware components and axes of motion

Front view

The following figure and table describe the primary hardware components and axes of motion.

Figure BenchCel Microplate Handler components (front view)

1

2 3

4

5

6

7

8

9

Item Feature Description

1 Labware rack (with labware)

The accessory that stores labware to be processed in a run.

2 Stacker head

The structure at which: A labware rack is loaded. Two sensors inside of each stacker head detect the

presence of the racks. A microplate is checked for type and orientation using a plate-presence sensor and

four plate-orientation sensors. A microplate is lowered into the stacker grippers to begin a run.

Hardware components and axes of motion

8 G5580A BenchCel Microplate Handler Quick Guide

3 Air pressure regulator

The knob that adjusts the air pressure inside the device. Compressed air is used to move the components inside the stacker heads. Each regulator controls the air pressure to the two adjacent stacker heads. For details, see the BenchCel Microplate Handler User Guide.

4 Safety shield

The clear panel that is installed on the front of the BenchCel device to prevent access while it is in operation.

5 Pendant The component that is part of the safety interlock circuit, which must be closed for the BenchCel device to operate. Pressing the raised button on the pendant interrupts the interlock circuit and stops the robot motors. Use this method of stopping the robot for emergencies only.

6 Rack release button

The button that unlocks the rack for removal. The rack-release button (1) at the top of each stacker head displays different colors to indicate the state of the stacker head.

Green. The labware rack is installed correctly on the BenchCel device and the microplates are unloaded. The stack of microplates are ready for processing or you can unlock and remove the labware rack.

Flashing green. The labware rack is unlocked and can be removed. Blue. The stack of microplates is loaded. You cannot unlock and remove the

labware rack. Red. The clamps are open without a rack installed. Do not install a rack until the

clamps are closed.

7 Robot head The component that moves (1) horizontally along the x-axis and (2) vertically along the z-axis.

Item Feature Description

9

Hardware components and axes of motion

G5580A BenchCel Microplate Handler Quick Guide

8 Robot arms The two parallel structures that are attached to and rotate about the robot head along the theta-axis.

9 Robot grippers

The structures inside the robot arms that close and open to hold and release a microplate. Using the provided software, you can adjust the distance between the grippers (1) to hold a microplate loosely or tightly.

Item Feature Description

Hardware components and axes of motion

10 G5580A BenchCel Microplate Handler Quick Guide

Back view

Figure BenchCel connection panel (Back view)

Item Part Description

1 Air-input fitting Connects the air tubing to the BenchCel device. Compressed air is used to actuate components inside the stacker head.

2 Power switch Turns on or off the power to the BenchCel device.

3 AC power entry Connects the power cord to the BenchCel device.

4 Ethernet port Connects the Ethernet cable from the controlling computer to the BenchCel device to allow communication between the computer and the device. Use this port as an alternative to the serial connection.

5 Serial port Connects the serial cable from the controlling computer to the BenchCel device to allow communication between the computer and the device. Use this port as an alternative to the Ethernet connection.

6 Pendant port Connects the pendant to the safety interlock circuit.

1

2 3

4

5 6

11

Labware considerations

G5580A BenchCel Microplate Handler Quick Guide

Labware considerations

Acceptable microplates

The BenchCel device is designed to handle labware that comply with the standards ANSI/SLAS 1-2004 (R2012) through ANSI/SLAS 4-2004 (R2012). For use of nonconforming labware, contact Agilent Technical Support. The BenchCel device uses gripping mechanisms to hold microplates securely and repeatably in the labware rack and in the robot arms. The BenchCel device typically holds the microplates halfway between the top of the microplate and the top of the microplate skirt (5 to 10 mm above the bottom of the microplate). In the following figure, notice the gripper-microplate contact point

Figure Labware rack detail with microplate held by the stacker grippers (front view)

Lidded microplates

Microplates that do not have lids or have shallow lids (lids that do not reach the microplate skirt) provide enough clearance to allow secure and repeatable gripping. Microplates with deeper lids can be more challenging, because the microplate must be held by the skirt. If the skirt is too flexible, the stacker grippers will bend the skirt. The bent skirt can grip the microplate lid stacked beneath, inadvertently removing the lid. Some labware vendors might offer alternative lids that are shallower. Contact the vendor for details.

Figure Lidded microplate examples

Item Description

1 No lid: excellent gripper clearance

2 Shallow lid: good gripper clearance

3 Deep lid: no gripper clearance, must be held by skirt

1

2

3

Starting up and shutting down

12 G5580A BenchCel Microplate Handler Quick Guide

Challenging microplate characteristics

Microplates that have the following characteristics might require additional setup time to ensure repeatable performance for the BenchCel device: Microplate material. Although you can adjust the robot grip distance to compensate

for microplate flexibility, some microplates are too soft and tend to bend in the robot grippers or become warped after thermal cycling. (for example, low-profile polypropylene PCR microplates).

Manufacturing variance. Gross variations in microplate dimensions can reduce repeatability of secure gripping. In addition, because the BenchCel device uses reflected light to sense microplate presence and orientation, variations in the reflective properties of the microplates can affect optimal operation.

Microplate design. Some microplates have special features specifically designed for particular instruments but are not optimized for the BenchCel device.

Thermal cycling effects. Microplates that have been through thermal cycling might become warped.

Tall labware. Especially tall tube racks and tip boxes that are taller than 65 mm might pose challenges in the BenchCel device. Contact Agilent Technical Support about acceptable tall labware.

Extra long lid. Some microplates that have lids that extend past the microplate skirt tend to pose challenges for the BenchCel device. Contact Technical Support for guidance.

Starting up and shutting down

The following procedures describe how to start up and shut down the BenchCel Microplate Handler when you are operating it as a standalone device. For instructions on how to turn on and turn off the BenchCel Microplate Handler when it is integrated into a workstation or system, see the workstation or system user documentation.

Starting up the device

WARNING Do not touch the BenchCel Microplate Handler as you start the software. The robot head moves when the device moves to home position.

To start up the BenchCel Microplate Handler: 1 On the back side of the BenchCel Microplate Handler, press the power switch to the

on (I) position. Every time you turn on the BenchCel Microplate Handler, the robot homes (the robot is sent to the factory-defined home position for each axis of motion).

13

Starting up and shutting down

G5580A BenchCel Microplate Handler Quick Guide

2 Turn on the compressed air supply to the BenchCel Microplate Handler. 3 Turn on the controlling computer. See the user documentation from the computer

manufacturer. 4 Start the VWorks software. To do this, on the Windows desktop, double-click the

VWorks shortcut icon. For instructions on using BenchCel Diagnostics to operate the BenchCel Microplate Handler, see Using BenchCel Diagnostics on page 20. For instructions on running protocols in BenchCel workstations, see the VWorks Automation Control User Guide.

Shutting down the BenchCel Microplate Handler

Shut down the BenchCel Microplate Handler if you intend to: Leave it unused for a long period of time. Service the device. Move it to another location.

To shut down the BenchCel Microplate Handler: 1 Exit the VWorks software. 2 Turn off the compressed air to the BenchCel Microplate Handler. 3 On the back of the device, press the power switch to the off (o) position.

Handling labware racks safely

14 G5580A BenchCel Microplate Handler Quick Guide

Handling labware racks safely

Carrying the labware racks

The labware racks store the stacks of labware (microplates, tip boxes, and tube racks) that are processed during a protocol run. This topic describes how to carry labware racks safely.

WARNING Do not hold a rack by the interior edges. The interior edges can have sharp surfaces that can cause cuts if handled improperly.

CAUTION A rack that is fully loaded with labware can be heavy. Grasp the rack handle firmly to prevent the rack from slipping or tilting.

To carry a rack, firmly grasp the rack by the handle as shown in the following figure.

Figure Carrying a front-load rack

Filling and emptying the labware racks

Before you start

WARNING Make sure you understand how to handle the racks safely.

CAUTION Before you place labware into a rack that is installed on the BenchCel device, the clamps in the BenchCel stacker head must be closed (extended). If the clamps are open the stacker grippers will not prevent the stack of labware from dropping.

Before filling or emptying a rack: Position the rack so that the opening is facing you. Determine how the microplates should be oriented in the rack.

15

Filling and emptying the labware racks

G5580A BenchCel Microplate Handler Quick Guide

For example, if the BenchCel orientation-sensing feature is enabled, make sure the A1 wells are oriented in the rack as specified.

Depending on the rack model, the procedure can vary: Front-load racks. The rack can be installed on the BenchCel device or placed on

a flat, level surface. Standard or top-load racks. Place the rack on a flat, level surface. For details on how to install or uninstall the racks, see Installing and uninstalling labware racks on page 17.

Filling a front-load rack

The doors on the front-load rack provide easy access for placing labware into the front of a rack that is installed on a device.

Before you start: If the rack is already installed on the BenchCel device, verify that the stack is in the unloaded state, as follows:

To verify the stack is in the unloaded state: 1 In BenchCel Diagnostics, initialize the profile. See Initializing the BenchCel

Microplate Handler in Diagnostics on page 21. 2 In the Controls tab, click Unload All to unload all the stackers.

To place labware into a front-load rack: 1 On each side of the rack, slide the (1) door-release (black) buttons forward, while

pushing outward on the (2) thumb tabs. The (3) rack doors open.

Figure Door mechanism on the front-load rack and opening the front-load rack

Filling and emptying the labware racks

16 G5580A BenchCel Microplate Handler Quick Guide

2 Place the labware directly through the open rack doors so that the bottom labware rests on the rack stacker grippers (4). Ensure the labware is level in the rack.

3 Close the doors, press the thumb tabs inward until the doors snap shut.

To remove labware from a front-load rack: 1 If the rack is installed on the BenchCel device, verify that the stack is in the

unloaded state. See To verify the stack is in the unloaded state: on page 15. 2 To open the rack doors, slide the black door-release buttons forward on each side

of the rack, while pressing outward on the thumb tabs. 3 Carefully, lift the labware out through the front of the rack.

Filling the standard and top-load racks

WARNING Use care to avoid sliding your hand on the interior edges in the rack. The edges can have sharp surfaces.

The following figure shows how to slide a stack of labware into a standard rack.

Figure Filling a standard rack: (1) Sliding the stack down through the top, (2) Supporting the stack through the bottom slot, and (3) Ensuring stack rests on the rack stacker grippers

To put labware into a standard or top-load rack: 1 Place the rack on a flat, level surface. 2 Using both hands, carefully slide a small stack of labware down through the top of

the rack. You can use one hand to support underneath the labware stack, while the other hand holds the top of the labware to keep it level. See figure.

17

Installing and uninstalling labware racks

G5580A BenchCel Microplate Handler Quick Guide

3 Standard racks only. When you reach the bottom of the open slot, transfer your hand positions so that you continue supporting the labware through the bottom slot.

4 Ensure that the bottom labware in the stack rests on the rack stacker grippers.

To remove the labware from a standard or top-load rack: 1 If possible, remove the rack from the device, and place the rack on a flat, level

surface.

IMPORTANT If you are removing labware from an installed rack, ensure that the stack is unloaded. See To verify the stack is in the unloaded state: on page 15.

2 Carefully slide the labware in small stacks, up and out of the top of the rack.

Installing and uninstalling labware racks

Before you start

IMPORTANT Make sure the BenchCel Microplate Handler power and compressed air are turned on before you install or uninstall a labware rack.

When lifting the labware rack onto and off of the stacker head, use both hands to grasp the rack securely around the four corners near the base.

WARNING Avoid touching the interior edges of a rack when lifting the rack. The interior edges can have sharp surfaces.

Installing and uninstalling labware racks

18 G5580A BenchCel Microplate Handler Quick Guide

Installing labware racks on the BenchCel Microplate Handler

To install a labware rack: 1 At the BenchCel Microplate Handler, verify that the clamps are closed (extended) in

the stacker head. Note: To close the clamps, open BenchCel Diagnostics. In the Controls tab, click Stacker at the top of the rack that you want to remove, and then click Close Stacker grippers. When you install the rack, the prongs in the labware rack tabs will be inserted into the slots in the clamps. If the clamps are open, the slots will be hidden, and you cannot install the rack.

Figure Stacker head with closed (extended) clamps

2 With the racks open side facing the front, lower the rack onto the stacker head. The rack is automatically locked into position.

Figure Installing a labware rack on the stacker head (closeup view)

IMPORTANT Make sure the prongs at the bottom of the rack tabs insert into the slots in the extended clamps.

19

Installing and uninstalling labware racks

G5580A BenchCel Microplate Handler Quick Guide

Uninstalling labware racks

To uninstall a rack when rack-release button is green: 1 Verify that the rack-release button on the stacker head is lit green, and then press it.

A click sounds as the locks retract. 2 While the green light is flashing, lift the labware rack up and out of the stacker

head.

Note: If the green light stops flashing, press the rack-release button again.

To uninstall a rack when rack-release button is blue: 1 In BenchCel Diagnostics, initialize the profile. 2 In the Controls tab, click Unload All to unload all the stackers. See figure, item 1.

In the Controls tab image: (2) A blue rack-release button indicates that the stack is loaded. (3) A rack image appears on the stacker if the sensors detect a rack. (4) A green labware image indicates that the sensors detect a plate.

3 To unlock the rack, press the green rack-release button at the top of the stacker head. Wait for the click to sound as the locks retract. While the green light is flashing, lift the labware rack up and out of the stacker head.

Using BenchCel Diagnostics

20 G5580A BenchCel Microplate Handler Quick Guide

Using BenchCel Diagnostics

BenchCel Diagnostics has three tabbed pages: Controls, General Settings, and Profiles. You use the commands in the Profiles tab to establish communication with the BenchCel Microplate Handler. You use the commands and parameters in the Controls tab to control or move the device components.

IMPORTANT Some of the BenchCel Diagnostics features are available only if you have VWorks administrator- or technician-level privileges. For details, see your lab administrator.

This topic describes the following: Opening BenchCel Diagnostics on page 20 Initializing the BenchCel Microplate Handler in Diagnostics on page 21 Sending the robot to the home position on page 22 Homing the robot on page 23 Disabling and enabling the robot motors on page 23 Jogging the robot on page 24 Changing the robot speed on page 25 Moving plates between teachpoints on page 26 Opening and closing clamps on page 27 Extending and retracting shelves on page 27 For more details, see the BenchCel Microplate Handler User Guide.

Opening BenchCel Diagnostics

To open BenchCel Diagnostics: 1 In the VWorks window, ensure that simulation is off and the correct device file

(*.dev) is open. 2 In the Devices area, select the BenchCel Microplate Handler device, and then click

Device Diagnostics.

21

Using BenchCel Diagnostics

G5580A BenchCel Microplate Handler Quick Guide

The BenchCel Diagnostics dialog box opens. By default, the Profiles tab is displayed.

Initializing the BenchCel Microplate Handler in Diagnostics

Before you use BenchCel Diagnostics to operate the BenchCel Microplate Handler, you must initialize an appropriate BenchCel profile to: Establish communication with the BenchCel Microplate Handler. Load the profile information.

WARNING When you initialize the BenchCel device, the robot head can move. To prevent potential injury, keep clear of the device while it is in motion.

CAUTION To prevent potential equipment damage, remove objects in the path of the robot head, arms, and grippers before operating the BenchCel Microplate Handler.

CAUTION Using an incorrect device profile can cause an error or damage the BenchCel Microplate Handler. Ensure that the correct device profile is initialized before operating the BenchCel Microplate Handler.

To initialize a profile in BenchCel Diagnostics: 1 In the Profiles tab, select the desired profile in the Profile name list.

Using BenchCel Diagnostics

22 G5580A BenchCel Microplate Handler Quick Guide

2 Click Initialize this profile to establish communication with the BenchCel Microplate Handler and load the profile information.

Sending the robot to the home position

The home position is where the robot head is at the center of the BenchCel Microplate Handler and the robot arms are perpendicular to the x-axis. You send the robot to the home position if you want the robot out of the way in a safe position.

To send the robot to the home position: 1 In the Controls tab, click the Jog/Teach tab. 2 Click Go Home.

23

Using BenchCel Diagnostics

G5580A BenchCel Microplate Handler Quick Guide

Homing the robot

Homing the robot sends the robot to the factory-defined home position for each axis of motion. The homing process recalibrates the robot position along each axis. Home the robot if you notice that the robot is not accurately picking up or placing plates. You might also want to home the robot after recovering from an emergency stop. Note: If a labware definition is selected when you click Home motors, the robot grippers return to the Gripper open position defined for the selected labware.

To home the robot: 1 In the Controls tab, click the Jog/Teach tab. 2 Click Home Motors.

Disabling and enabling the robot motors

Disabling the robot motors allows you to move the robot by hand, making it easier to set and edit teachpoints.

To disable or enable the robot motors: 1 In the Controls tab, click the Jog/Teach tab. 2 In the Motors Enabled area, select or click the following:

Option or command Description X Select the option to enable the x-axis motor. Clear

the check box to disable the x-axis motor. Z Select the option to enable the z-axis motor. Clear

the check box to disable the z-axis motor.

Using BenchCel Diagnostics

24 G5580A BenchCel Microplate Handler Quick Guide

Jogging the robot

Jogging the robot moves the robot and robot grippers in small, precise increments along one of the axes. You can jog the robot to fine-tune its position when creating and editing teachpoints.

To jog the robot: 1 In the Controls tab, click the Jog/Teach tab. 2 Enable the robot motors. See Disabling and enabling the robot motors on

page 23.

3 In the robot movement area, click to select the jog distance from the corresponding increment list, if applicable.

CAUTION Use smaller jog increments than you think you need to ensure that the robot does not bump into obstacles in its path (such as the stacker heads and plate stages).

4 Click the directional button:

Theta Select the option to enable the theta-axis motor. Clear the check box to disable the theta-axis motor.

Enable All Click to turn on all the motors. Disable All Click to turn off all the motors.

Option or command Description

Command Description Jogs the robot arm counterclockwise from the current position by the specified theta-axis increment.

Jogs the robot arm clockwise from the current position by the specified theta-axis increment.

Jogs the robot head left from the current position by the specified x-axis increment.

Jogs the robot head right from the current position by the specified x-axis increment.

Jogs the robot head up from the current position by the specified z- axis increment.

Jogs the robot head down from the current position by the specified z-axis increment.

Opens the robot grippers by the specified grip increment.

Closes the robot grippers by the specified grip increment.

25

Using BenchCel Diagnostics

G5580A BenchCel Microplate Handler Quick Guide

Changing the robot speed

The speed you select in Diagnostics applies only to the robot commands in Diagnostics (Jog, Move, Transfer, and so on). If the robot is holding a microplate, the slower of the following will be applied: the speed you selected in the Labware Editor or the speed you selected in Diagnostics. You can select the robot speed to accommodate the task you are performing. For example, you can select the Slow speed when you are creating new teachpoints, creating and testing protocols, or diagnosing problems with the system. You can adjust the speed setting for the x-axis, z-axis, and theta-axis for each speed (high, medium, or slow) as a percentage of the factory-set maximum speed.

To select the robot speed: 1 In Diagnostics, click the Controls tab. 2 In the Speed list, select Fast, Medium, or Slow.

Note: During a protocol run, the robot uses the speed selection in the VWorks Tools > Options dialog box. If the robot is holding a microplate, the slower of the following will be applied: the speed in the Labware Editor or the speed in the Tools > Options dialog box. For more information, see the VWorks Automation Control User Guide.

Full Open Opens the robot grippers to the Robot Gripper Open Position value set in the Labware tab.

Full Close Closes the robot grippers to the Robot Gripper Holding Stack value set in the Labware tab.

Command Description

Using BenchCel Diagnostics

26 G5580A BenchCel Microplate Handler Quick Guide

Moving plates between teachpoints

You can move a microplate between teachpoints when you are verifying a teachpoint or to determine whether to home the motors.

CAUTION To prevent collision, remove obstacles in the path of the robot.

To move plates between teachpoints: 1 Manually place a microplate at one of the two teachpoints. 2 In Diagnostics, click the Controls tab, and then select the Slow speed.

3 In the graphical display area, click the plus sign ( ) at either of the following:

The teachpoint you want to move to Two teachpoints between which you want to move a plate

The selected teachpoints should be highlighted in red circles ( ).

4 Rest the pointer on a selected teachpoint. In the command menu that appears, select one of the following:

Command Description Move to

Moves the robot from its current position to the selected teachpoint. The robot stays at the teachpoint.

Pick from

Picks up the microplate from the selected teachpoint and moves the plate to the ready-for-upstack position under the stacker head.

Place at

Moves the robot from its current position and places the microplate at the selected teachpoint. After placing the microplate, the robot backs away from the teachpoint into the safe zone.

Transfer to

Moves a microplate from the currently selected teachpoint to the other selected teachpoint. To transfer a microplate from a stack, you must first click Load Stacker. During the transfer, the robot downloads the plate, places it at the other selected teachpoint, and then backs into the safe zone. If the robot is transferring a plate from one teachpoint to another, the robot picks up the microplate from the currently selected teachpoint, places the microplate at the other selected teachpoint, and then backs into the safe zone.

Delid from

Lidded labware only. Removes the lid from the labware that is at the selected teachpoint.

Relid to

Lidded labware only. Replaces the lid on the labware that is at the selected teachpoint.

27

Using BenchCel Diagnostics

G5580A BenchCel Microplate Handler Quick Guide

Opening and closing clamps

The clamps in the stacker head close and open the grippers at the bottom of the labware rack to hold and release the first plate in position for the robot grippers. Compressed air is used to move the clamps. The clamps close and open the stacker grippers automatically during the loading, unloading, downstacking, and stacking procedures. When diagnosing problems or after an aborted run, you can use the commands in Diagnostics to open or close the clamps (stacker grippers). For example, you might want to open the clamps (stacker grippers) to remove a microplate.

CAUTION Opening the clamps (stacker grippers) might cause the microplate or stack of microplates to drop.

To open or close the clamps (stacker grippers): 1 In Diagnostics, click the Controls tab. 2 In the graphical display area at the top of the rack, click Stacker, and then choose

Open Stacker gripper or Close stacker gripper. A click sounds as the clamps (stacker grippers) open or close.

Extending and retracting shelves

The shelves in the stacker head are used to hold the stack of labware temporarily during the downstacking and upstacking processes. Resting the microplates on the shelves levels the microplates, allowing the robot grippers to accurately hold the microplate at the specified offset position. Compressed air is used to move the shelves.

Using BenchCel Diagnostics

28 G5580A BenchCel Microplate Handler Quick Guide

The shelves extend and retract automatically during the downstacking and upstacking processes. When diagnosing problems or after an aborted run, you can use the commands in Diagnostics to extend or retract the shelves.

CAUTION Retracting the shelves might cause the microplate or stack of microplates to drop.

To extend or retract the shelves: 1 In BenchCel Diagnostics, click the Controls tab. 2 In the graphical display area at the top of the desired rack, click Stacker, and then

choose Extend Shelf or Retract Shelf. A click sounds as the shelves extend or retract.

29

Cleaning up after use

G5580A BenchCel Microplate Handler Quick Guide

Cleaning up after use

CAUTION Make sure you clean up spilled liquids immediately. Use only the recommended cleaning materials. Using other cleaning solutions and materials can cause damage to the device. Do not use abrasive, corrosive cleaning agents. Do not use metal brushes.

To clean up the BenchCel Microplate Handler after use: 1 If applicable, remove the labware that remains in the robot grippers. You might

need to use BenchCel Diagnostics to move the robot head to a convenient position, and then open the robot grippers. For instructions see Jogging the robot on page 24.

2 Unload used labware from the labware racks. See Filling and emptying the labware racks on page 14.

3 Uninstall the labware racks from the BenchCel Microplate Handler. See Installing and uninstalling labware racks on page 17.

IMPORTANT Make sure the power and compressed air are turned on when removing the racks.

4 Make sure the x-axis tracks (1) are free of debris.

5 Use standard laboratory wipes and a mild detergent or ethanol alcohol to clean the exterior painted white surfaces and the metal surfaces of dust, grime, chemical deposits, and other debris on the BenchCel Microplate Handler.

www.a

Manualsnet FAQs

If you want to find out how the G5580A BenchCel Microplate Handler Agilent works, you can view and download the Agilent G5580A BenchCel Microplate Handler Handler Quick Guide on the Manualsnet website.

Yes, we have the Quick Guide for Agilent G5580A BenchCel Microplate Handler as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Quick Guide should include all the details that are needed to use a Agilent G5580A BenchCel Microplate Handler. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent G5580A BenchCel Microplate Handler Handler Quick Guide is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent G5580A BenchCel Microplate Handler Handler Quick Guide consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent G5580A BenchCel Microplate Handler Handler Quick Guide free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent G5580A BenchCel Microplate Handler Handler Quick Guide, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent G5580A BenchCel Microplate Handler Handler Quick Guide as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.