SureSelect Automated Strand-Specific RNA Library Prep Automated Poly-A Selection and Strand-Specific mRNA Library Preparation for the Illumina Platform

Protocol Version F0, June 2020

SureSelect platform manufactured with Agilent SurePrint Technology

For Research Use Only. Not for use in diagnostic procedures.

Agilent Technologies

Notices Agilent Technologies, Inc. 2013, 2014, 2017, 2020

No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

Manual Part Number G9691-90030

Edition Version F0, June 2020

Printed in USA

Agilent Technologies, Inc.

Warranty The material contained in this document is provided as is, and is subject to being changed, with- out notice, in future editions. Fur- ther, to the maximum extent permitted by applicable law, Agi- lent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchant- ability and fitness for a particular purpose. Agilent shall not be lia- ble for errors or for incidental or consequential damages in con- nection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this doc- ument that conflict with these terms, the warranty terms in the separate agreement shall control.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor- dance with the terms of such license.

Restricted Rights Legend U.S. Government Restricted Rights. Soft- ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus- tomers. Agilent provides this customary commercial license in Software and techni- cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com- puter Software Documentation).

Notice to Purchaser: Limited License Limited Use Label License: This product and its use are the subject of one or more issued and/or pending U.S. and foreign patent applications owned by Max Planck Gesellschaft exclusively licensed to New England Biolabs, Inc. and sublicensed to Agilent Technologies. The purchase of this product from Agilent Technologies, Inc., its affiliates, or its authorized resellers and dis- tributors conveys to the buyer the non-transferable right to use the purchased amount of the product and components of the product in research conducted by the buyer (whether the buyer is an academic or for profit entity). The purchase of this prod- uct does not convey a license under any claims in the foregoing patents or patent applications directed to producing the prod- uct. The buyer cannot sell or otherwise transfer this product or its components to a third party or otherwise use this product for the following COMMERCIAL PURPOSES: (1) use of the product or its components in manufacturing; or (2) use of the product or its components for therapeutic or prophy- lactic purposes in humans or animals.

5301 Stevens Creek Blvd Santa Clara, CA 95051 USA

Acknowledgment Oligonucleotide sequences 2006, 2008, and 2011 Illumina, Inc. All rights reserved. Only for use with the Illumina sequencer systems and associated assays.

Technical Support For US and Canada

Call (800) 227-9770 (option 3,4,4)

Or send an e-mail to: ngs.support@agilent.com

For all other regions Agilents world-wide Sales and Support Center contact details for your location can be obtained at www.agilent.com/en/contact-us/page.

2 SureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Safety Notices

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

SureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing 3

In this Guide...

4 SureSelect Autom

This guide describes an optimized protocol for Illumina paired- end multiplexed mRNA library preparation using the Agilent SureSelect Automated Strand- Specific Library Prep system.

This protocol is specifically developed and optimized to prepare mRNA sequencing libraries from total RNA samples. Sample processing steps are automated using the NGS Workstation.

1

Before You Begin

This chapter contains information (such as procedural notes, safety information, required reagents and equipment) that you should read and understand before you start an experiment.

2

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation

This chapter contains an orientation to the Agilent NGS Workstation, an overview of the SureSelect mRNA Library Prep protocol, and considerations for designing SureSelect experiments for automated processing using the Agilent NGS Workstation.

3

Sample Preparation

This chapter describes the steps to prepare strand- specific libraries from total RNA samples for mRNA sequencing on the Illumina platform.

4

Reference

This chapter contains reference information.

ated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Whats New in Version F0

SureSelect Automated Strand-Spec

Support for renamed Library Preparation Kit components. See Table 49 on page 88 and Table 50 on page 89 for a summary of component name changes. The new component names are utilized throughout the protocols in this document. No changes were made to the reagent formulations or methods of use.

Support for revised Poly- A Selection Module subkit. See Table 49 on page 88 and Table 51 on page 90 for a summary of component name and configuration changes. The new component names are utilized throughout the protocols in this document. No changes were made to the reagent formulations or methods of use. Water is no longer provided in this module (see Table 51 on page 90) and has been added to the list of required reagents in Table 1 on page 12.

Support for VWorks software version 13.1.0.1366 and Agilent NGS Workstation Option B p/n G5574AA (see Table 2 on page 13)

Updates to Agilent 2100 Bioanalyzer system ordering information (see page 14)

Support for Agilent 4150 TapeStation and addition of Agilent 4200/4150 TapeStation system- compatible plasticware ordering information (see page 14)

Updates to mixing instructions for sample qualification assays using an Agilent TapeStation instrument (see Caution on page 81)

Updates to ordering information for AMPure XP Kits (see Table 1 on page 12).

Updated recommendation for control reference RNA (see Table 3 on page 14).

Updates to sequencing support guidelines for RNA strandedness (see page 86) and recommended cycle number settings (see page 85).

Updated Notice to Purchaser (see page 2).

Updates to Technical Support contact information (see page 2)

ific RNA Library Prep for Illumina Platform Sequencing 5

6 SureSelect Autom

ated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Content

1 Before You Begin 9

Procedural Notes 10 Safety Notes 11 Required Reagents 12 Required Equipment 13 Optional Reagents and Equipment 14

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 15

About the Agilent NGS Workstation 16

About the Bravo Platform 16 VWorks Automation Control Software 20

Overview of the SureSelect RNA Library Prep Procedure 26

Experimental Setup Considerations for Automated Runs 28

Considerations for Placement of RNA Samples in 96-well Plates for Automation 29

Considerations for Equipment Setup 29 PCR Plate Type Considerations 30

3 Sample Preparation 31

Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA 32 Step 2. Purify first-strand cDNA using AMPure XP beads 53 Step 3. Prepare cDNA libraries for Illumina sequencing 57 Step 4. Index cDNA libraries by PCR amplification 65 Step 5. Purify amplified DNA using AMPure XP beads 73 Step 6. Remove adaptor-dimers using AMPure XP beads 77 Step 7. Assess Library DNA quantity and quality 80 Step 8. Pool samples for multiplexed sequencing 83 Step 9. Prepare and analyze sequencing samples 85

SureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing 7

Contents

4 Reference 87

Kit Contents 88

Nucleotide Sequences of SureSelectXT Indexes A01 to H12 91

Plasticware quantities for automation protocols 92

8 SureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform

SureSelect Automated Strand-Specific RNA Library Prep Protocol

1 Before You Begin

Procedural Notes 10

Safety Notes 11

Required Reagents 12

Required Equipment 13

Optional Reagents and Equipment 14

Make sure you read and understand the information in this chapter and have the necessary equipment and reagents listed before you start an experiment.

This protocol describes automated sample processing using the Agilent NGS Workstation. For non-automated sample processing procedures for Agilent's SureSelect Strand-Specific RNA Library Prep Kit for mRNA Sequencing on the Illumina platform, see publication G9691-90010.

NOTE

9Agilent Technologies

1 Before You Begin Procedural Notes

Procedural Notes

10 S

Certain protocol steps require the rapid transfer of sample plates between the Bravo deck and a thermal cycler. Locate your thermal cycler in close proximity to the Agilent NGS Workstation to allow rapid and efficient plate transfer.

Use of Agilents SureCycler 8800 thermal cycler and associated plasticware is recommended for optimal performance. The workflow is compatible with additional thermal cyclers, but performance should be validated before running a large number of samples. See page 30 for a list of supported PCR plate types and ensure that the thermal cycler to be used is compatible with one of the supported PCR plate types.

Prepare and load the Agilent NGS Workstation as detailed in each of the protocol steps before initiating each automated protocol run. When loading plates in the workstations Labware MiniHub, always place plates in the orientation shown in Figure 4 on page 37.

To prevent contamination of reagents by nucleases, always wear powder- free laboratory gloves and use dedicated solutions and pipettors with nuclease- free aerosol- resistant tips.

Avoid repeated freeze- thaw cycles of stock and diluted RNA and cDNA solutions. Possible stopping points, where samples may be stored at 20C, are marked in the protocol. Do not subject the samples to multiple freeze/thaw cycles.

When preparing master mix reagent stock solutions for use:

1 Thaw the reagent vial as rapidly as possible without heating above room temperature.

2 Mix thoroughly on a vortex mixer at high speed for 5 seconds, then briefly spin in a centrifuge to drive the contents off of walls and lid.

3 Store vials used during an experiment on ice or in a cold block.

4 Library Preparation Master Mixes should not be frozen and thawed more than five times. If you plan to use the reagents in more than five experiments, aliquot to multiple vials to minimize freeze/thaw cycles for each vial.

In general, follow Biosafety Level 1 (BL1) safety rules.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Before You Begin 1 Safety Notes

Safety Notes

Wear appropriate personal protective equipment (PPE) when working in the laboratory.

CAUTION

SureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing 11

1 Before You Begin Required Reagents

Required Reagents

12 S

Table 1 Required Reagents for SureSelect RNA Library Prep Automation

Description Vendor and part number

SureSelect Strand Specific RNA Reagent Kit

Illumina platforms (ILM), 96 Samples*

* Each 96-reaction kit contains sufficient reagents for 96 reactions used in runs that include at least 3 columns of samples per run.

Agilent

p/n G9691B

Actinomycin D

Actinomycin D should be obtained as a solid and prepared at 4 g/l concentration in DMSO then stored in single-use aliquots at 20C, protected from light. The aliquots may be stored for up to one year before use. See page 34 for additional information.

Sigma p/n A1410

DMSO Sigma p/n D8418

AMPure XP Kit 5 mL 60 mL 450 mL

Beckman Coulter Genomics p/n A63880 p/n A63881 p/n A63882

100% Ethanol, molecular biology grade Sigma-Aldrich p/n E7023

Nuclease-free Water (not DEPC-treated) Thermo Fisher Scientific p/n AM9930

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Before You Begin 1 Required Equipment

Required Equipment

SureSelect Automat

Table 2 Required Equipment for SureSelect RNA Library Prep Automation

Description Vendor and part number

Agilent NGS Workstation Option B

Contact Agilent Automation Solutions for more information: Customerservice.automation@agilent.com

Agilent p/n G5522A (VWorks software version 13.1.0.1366, 13.0.0.1360, or 11.3.0.1195)

OR

Agilent p/n G5574AA (VWorks software version 13.1.0.1366)

Bravo 96-well PCR plate insert (red) Agilent p/n G5498B#13

Robotic Pipetting Tips (Sterile, Filtered, 250 L) Agilent p/n 19477-022

Thermal cycler and accessories SureCycler 8800 Thermal Cycler (Agilent p/n G8810A), 96 well plate module (Agilent p/n G8810A) and compression mats (Agilent p/n 410187) or equivalent

PCR plates compatible with selected thermal cycler, e.g. Agilent semi-skirted PCR plate for the SureCycler 8800 Thermal Cycler

When selecting plates for another thermal cycler, see page 30 for the list of PCR plates supported in automation protocols

Agilent p/n 401334

Eppendorf twin.tec full-skirted 96-well PCR plates Eppendorf p/n 951020401 or 951020619

Thermo Scientific Reservoirs Thermo Scientific p/n 1064156

Nunc DeepWell Plates, sterile, 1.3-mL well volume Thermo Scientific p/n 260251

Axygen 96 Deep Well Plate, 2 mL, Square Well (waste reservoirs; working volume 2.2 mL)

Axygen p/n P-2ML-SQ-C E & K Scientific p/n EK-2440

Nucleic acid surface decontamination wipes DNA Away Surface Decontaminant Wipes, Thermo Scientific p/n 7008, or equivalent

Vacuum concentrator Savant SpeedVac, model DNA120, with 96-well plate rotor, model RD2MP, or equivalent

Low-Adhesion Tubes (RNase, DNase, and DNA-free) 1.5 mL 0.5 mL

USA Scientific p/n 1415-2600 p/n 1405-2600

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 13

1 Before You Begin Optional Reagents and Equipment

14 S

DNA Analysis Platform and Consumables

Agilent 2100 Bioanalyzer Instrument

Agilent 2100 Expert SW Laptop Bundle (optional)

DNA 1000 Kit

OR

Agilent 4200 TapeStation*

96-well sample plates

96-well plate foil seals

8-well tube strips

8-well tube strip caps

D1000 ScreenTape

D1000 Reagents

Agilent p/n G2939BA

Agilent p/n G2953CA

Agilent p/n 5067-1504

Agilent p/n G2991AA

Agilent p/n 5042-8502

Agilent p/n 5067-5154

Agilent p/n 401428

Agilent p/n 401425

Agilent p/n 5067-5582

Agilent p/n 5067-5583

Nuclease-free 0.2 mL PCR tubes, thin-walled Eppendorf p/n 951010006 or equivalent

P10, P20, P200 and P1000 pipettes Pipetman P10, P20, P200, P1000 or equivalent

Ice bucket

Powder-free gloves

Vortex mixer

Timer

* DNA samples may also be analyzed using the 4150 TapeStation, p/n G2992AA. ScreenTape devices and associated reagents listed in this table are compatible with both platforms.

Table 2 Required Equipment (continued)for SureSelect RNA Library Prep Automation

Description Vendor and part number

Optional Reagents and Equipment

Table 3 Optional Reagents and Equipment

Description Vendor and part number

Labnet MPS1000 Mini Plate Spinner Labnet International p/n C1000

Agilent QPCR Human Reference Total RNA Agilent p/n 750500

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

SureSelect Automated Strand-Specific RNA Library Prep Protocol

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation

About the Agilent NGS Workstation 16

Overview of the SureSelect RNA Library Prep Procedure 26

Experimental Setup Considerations for Automated Runs 28

This chapter contains an orientation to the Agilent NGS Workstation, an overview of the SureSelect single- stranded RNA library preparation protocol, and considerations for designing SureSelect RNA experiments for automated processing using the Agilent NGS Workstation.

15Agilent Technologies

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation About the Agilent NGS Workstation

About the Agilent NGS Workstation

CAUTION Before you begin, make sure that you have read and understand operating, maintenance and safety instructions for using the Bravo platform and additional devices included with the workstation. Refer to the user guides listed in Table 4.

16 S

Review the user guides listed in Table 4 (available at Agilent.com) to become familiar with the general features and operation of the Agilent NGS Workstation Option B components. Instructions for using the Bravo platform and other workstation components for the SureSelect RNA Library Preparation workflow are detailed in this user guide.

Table 4 Agilent NGS Workstation components User Guide reference information

Device User Guide part number

Bravo Platform G5562-90000

VWorks Software G5415-90068 (VWorks versions 13.1.0.1366 and 13.0.0.1360), or G5415-90063 (VWorks version 11.3.0.1195)

BenchCel Microplate Handler G5400-90004

Labware MiniHub G5471-90002

PlateLoc Thermal Microplate Sealer G5402-90001

About the Bravo Platform

The Bravo platform is a versatile liquid handler with a nine plate- location platform deck, suitable for handling 96- well, 384- well, and 1536- well plates. The Bravo platform is controlled by the VWorks Automation Control software. Fitted with a choice of seven interchangeable fixed- tip or disposable- tip pipette heads, it accurately dispenses fluids from 0.1 L to 250 L.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 2 About the Bravo Platform

SureSelect Automat

Bravo Platform Deck

The protocols in the following sections include instructions for placing plates and reagent reservoirs on specific Bravo deck locations. Use Figure 1 to familiarize yourself with the location numbering convention on the Bravo platform deck.

Figure 1 Bravo platform deck

Setting the Temperature of Bravo Deck Heat Blocks

Bravo deck positions 4 and 6 are equipped with Inheco heat blocks, used to incubate sample plates at defined temperatures during the run. Runs that include high- (85C) or low- (4C) temperature incubation steps may be expedited by pre- setting the temperature of the affected block before starting the run.

Bravo deck heat block temperatures may be changed using the Inheco Multi TEC Control device touchscreen as described in the steps below. See Table 5 for designations of the heat block- containing Bravo deck positions on the Multi TEC control device.

Table 5 Inheco Multi TEC Control touchscreen designations

Bravo Deck Position Designation on Inheco Multi TEC Control Screen

4 CPAC 2 1

6 CPAC 2 2

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 17

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation About the Bravo Platform

18 S

1 Using the arrow buttons, select the appropriate block (CPAC 2 block 1 or CPAC 2 block 2).

2 To set the temperature of the selected block, press the SET button.

3 Using the numeral pad, enter the desired temperature. The entered temperature appears in the top, left rectangle. Once the correct temperature is displayed, press the rectangle to enter the temperature.

4 Press the Temp button until the new temperature is displayed on the SET button and until the Temp button is darkened, indicating that the selected heat block is heating or cooling to the new temperature setting. The current temperature of the block is indicated in the center of the display.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 2 About the Bravo Platform

SureSelect Automat

Setting the Temperature of Bravo Deck Position 9 Using the ThermoCube Device

Bravo deck position 9 is equipped with a ThermoCube thermoelectric temperature control system, used to incubate components at a defined temperature during the run. During protocols that require temperature control at position 9, you will be instructed to start and set the temperature of the ThermoCube device before starting the run.

ThermoCube temperature settings are modified using the control panel (LCD display screen and four input buttons) on the front panel of the device using the following steps.

1 Turn on the ThermoCube and wait for the LCD screen to display TEMP.

2 Press the UP or DOWN button to change SET TEMP 1 to the required set point.

3 Press the START button.

The ThermoCube will then initates temperature control of Bravo deck position 9 at the displayed set point.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 19

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation VWorks Automation Control Software

VWorks Automation Control Software

20 S

VWorks software, included with your Agilent NGS Workstation, allows you to control the robot and integrated devices using a PC. The Agilent NGS Workstation is preloaded with VWorks software containing all of the necessary SureSelect system liquid handling protocols. General instructions for starting up the VWorks software and the included protocols is provided below. Each time a specific VWorks protocol is used in the SureSelect procedure, any settings required for that protocol are included in the relevant section of this manual.

NOTE The instructions in this manual are compatible with VWorks software version 13.1.0.1366, 13.0.0.1360 or 11.3.0.1195.

If you have questions about VWorks version compatibility, please contact service.automation@agilent.com.

Logging in to the VWorks software

1 Double- click the VWorks icon or the SureSelect_RNA_ILM.VWForm shortcut on the Windows desktop to start the VWorks software.

2 If User Authentication dialog is not visible, click Log in on the VWorks window toolbar.

3 In the User Authentication dialog, type your VWorks user name and password, and click OK. (If no user account is set up, contact the administrator.)

VWorks protocol and runset files

VWorks software uses two file types for automation runs, .pro (protocol) files and .rst (runset) files. Runset files are used for automated procedures in which the workstation uses more than one automation protocol during the run.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 2 VWorks Automation Control Software

SureSelect Automat

Using the SureSelect_RNA_ILM.VWForm to setup and start a run

Use the VWorks form SureSelect_RNA_ILM.VWForm, shown below, to set up and start each SureSelect automation protocol or runset.

1 Open the form using the SureSelect_RNA_ILM.VWForm shortcut on your desktop.

2 Use the drop- down menus on the form to select the appropriate SureSelect workflow step and number of columns of samples for the run.

3 Once all run parameters have been specified on the form, click Display Initial Workstation Setup.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 21

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation VWorks Automation Control Software

22 S

4 The Workstation Setup region of the form will then display the required placement of reaction components and labware in the NGS Workstation for the specified run parameters.

5 After verifying that the NGS Workstation has been set up correctly, click Run Selected Protocol.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 2 VWorks Automation Control Software

SureSelect Automat

Error messages encountered at start of run

After starting the run, you may see the error messages displayed below. When encountered, make the indicated selections and proceed with the run. Encountering either or both of these error messages is not indicative of a problem with the NGS workstation or your run setup.

1 If you encounter the G- axis error message shown below, select Ignore and Continue, leaving device in current state.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 23

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation VWorks Automation Control Software

24 S

2 If you encounter the W- axis error message shown below, select Retry.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 2 VWorks Automation Control Software

SureSelect Automat

Verifying the Simulation setting

VWorks software may be run in simulation mode, during which commands entered on screen are not completed by the NGS workstation. If workstation devices do not respond when you start a run, verify the simulation mode status in VWorks using the following steps.

1 Verify that Simulation is off is displayed on the status indicator (accessible by clicking View > Control Toolbar).

2 If the indicator displays Simulation is on, click the status indicator button to turn off the simulation mode.

If you cannot see the toolbar above the SureSelect VWorks form, click the Full Screen buttonto exit full screen mode. If the toolbar is still not visible, right-click on the form and then select Control Toolbar from the menu.

NOTE

Finishing a protocol or runset

The window below appears when each run is complete. Click Yes to release the BenchCel racks to allow removal of components used in the current run in preparation for the next .pro or .rst run.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 25

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation Overview of the SureSelect RNA Library Prep Procedure

Overview of the SureSelect RNA Library Prep Procedure

26 S

Figure 2 summarizes the SureSelect workflow for RNA samples to be sequenced using the Illumina paired- read sequencing platform. For each sample to be sequenced, an individual cDNA library is prepared. The samples are then tagged by PCR with an index sequence. Depending on the capacity of the sequencing platform, up to 48 samples can be pooled and sequenced in a single lane using the multiplex index tags that are provided with the SureSelect Strand- Specific RNA Library Prep kit.

Figure 2 Overall sequencing sample preparation workflow

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 2 Overview of the SureSelect RNA Library Prep Procedure

SureSelect Automat

Table 6 summarizes how the VWorks protocols are integrated into the Strand- Specific RNA Library Prep workflow. See the Sample Preparation chapter for complete instructions for use of the VWorks protocols for sample processing.

Table 6 Overview of VWorks protocols and runsets used during the workflow

Workflow Step VWorks Protocols Used for Agilent NGS Workstation automation

Purify poly(A) RNA using oligo(dT) beads

Chemically fragment the poly(A) RNA

Synthesize first-strand cDNA

mRNA_Purification_v1.0.pro

Purify first-strand cDNA using AMPure XP beads AMPureXP_v1.1.pro:First Strand

Synthesize second-strand cDNA

Repair DNA ends

Purify end-repaired DNA

dA-tail DNA 3-ends

Ligate adaptors

Purify adaptor-ligated DNA

LibraryPrep_RNASeq_ILM_v1.1.rst

Amplify adaptor-ligated cDNA library using indexing primers

TranscriptomePCR_ILM_v1.0.pro

Purify indexed library amplicons using AMPure XP beads

AMPureXP_v1.1.pro:Transcriptome PCR

Remove adaptor-dimers using AMPure XP beads AMPureXP_v1.1.pro:Transcriptome Dimers

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 27

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation Experimental Setup Considerations for Automated Runs

Experimental Setup Considerations for Automated Runs

28 S

Agilent SureSelect Automated Strand- Specific RNA Library Prep runs may include 1, 2, 3, 4, 6, or 12 columns (equivalent to 8, 16, 24, 32, 48, or 96 wells) of RNA samples to be prepared for sequencing on the Illumina platform. Plan your experiments using complete columns of samples.

The number of columns or samples that may be processed using the supplied reagents will depend on the experimental design. For greatest efficiency of reagent use, plan experiments using at least 3 columns per run. Each 96- reaction kit contains sufficient reagents for 96 reactions configured as 4 runs of 3 columns of samples per run.

Table 7 Columns to Samples Equivalency

Number of Columns Processed Total Number of Samples Processed

1 8

2 16

3 24

4 32

6 48

12 96

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Using the Agilent NGS Workstation for SureSelect RNA Library Preparation 2 Considerations for Placement of RNA Samples in 96-well Plates for Automation

Considerations for Placement of RNA Samples in 96-well Plates for Automation

SureSelect Automat

The Agilent NGS Workstation processes samples column- wise beginning at column 1. RNA samples should be loaded into 96- well plates column- wise, in well order A1 to H1, then A2 to H2, ending with A12 to H12. When processing partial runs with <12 sample columns, do not leave empty columns between sample columns; always load the plate using the left- most column that is available.

For sample indexing by PCR (see Figure 2), you will need to prepare a separate plate containing the indexing primers. Assign the wells to be indexed with their respective indexing primers during experimental design.

Considerations for Equipment Setup

Some workflow steps require the rapid transfer of sample plates between the Bravo deck and a thermal cycler. Locate your thermal cycler in close proximity to the Agilent NGS Workstation to allow rapid and efficient plate transfer.

Several workflow steps require that the sample plate be sealed using the PlateLoc thermal microplate sealer included with the Agilent NGS Workstation, and then centrifuged to collect any dispersed liquid. To maximize efficiency, locate the centrifuge in close proximity to the Agilent NGS Workstation.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 29

2 Using the Agilent NGS Workstation for SureSelect RNA Library Preparation PCR Plate Type Considerations

PCR Plate Type Considerations

30 S

Automation protocols include several liquid- handling steps in which reagents are dispensed to PCR plates in preparation for transfer to a thermal cycler. For these steps you must specify the PCR plate type to be used on the SureSelect_RNA_ILM.VWForm to allow correct configuration of the liquid handling components for the PCR plate type. Before you begin the automation protocol, make sure that you are using a supported PCR plate type. The PCR plate type to be used in the protocol is specified using the menu below. Vendor and part number information is provided for the supported plate types in Table 8.

Table 8 Ordering information for supported PCR plates

Description in VWorks menu Vendor and part number

96 ABI PCR half-skirted plates (MicroAmp Optical plates)

Thermo Fisher Scientific p/n N8010560

96 Agilent semi-skirted PCR plate Agilent p/n 401334

96 Eppendorf Twin.tec half-skirted PCR plates Eppendorf p/n 951020303

96 Eppendorf Twin.tec PCR plates (full-skirted) Eppendorf p/n 951020401 or 951020619

CAUTION The plates listed in Table 8 are compatible with the Agilent NGS Bravo and associated VWorks automation protocols, designed to support use of various thermal cyclers.

Accordingly, some plates listed in Table 8 are not compatible with the recommended SureCycler 8800 Thermal Cycler. When using the SureCycler 8800 Thermal Cycler in the SureSelect automation workflow, use 96 Agilent semi-skirted PCR plates.

When using a different thermal cycler in the workflow, be sure to select a PCR plate that is compatible with your thermal cycler and that is listed in Table 8.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

SureSelect Automated Strand-Specific RNA Library Prep Protocol

3 Sample Preparation

Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA 32

Step 2. Purify first-strand cDNA using AMPure XP beads 53

Step 3. Prepare cDNA libraries for Illumina sequencing 57

Step 4. Index cDNA libraries by PCR amplification 65

Step 5. Purify amplified DNA using AMPure XP beads 73

Step 6. Remove adaptor-dimers using AMPure XP beads 77

Step 7. Assess Library DNA quantity and quality 80

Step 8. Pool samples for multiplexed sequencing 83

Step 9. Prepare and analyze sequencing samples 85

This section contains instructions for RNA library preparation specific to the Illumina paired- read sequencing platform and to automated processing using the Agilent NGS Workstation.

For each sample to be sequenced, individual library preparations are performed in separate wells of a 96- well plate. The samples are then indexed by PCR amplification allowing multiplexing of up to 48 samples for sequencing on Illumina platforms.

Refer to Illuminas protocol Preparing Samples for Paired- End Sequencing (p/n 1005361), or the appropriate Illumina protocol for more information.

31Agilent Technologies

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

32 S

In this step, automation protocol mRNA_Purification_v1.0.pro is used to complete multiple steps of the RNA Library Preparation workflow. First, poly(A) RNA is purified from total RNA using two serial rounds of binding to oligo(dT) magnetic particles. After purification, the poly(A) RNA is chemically- fragmented to the appropriate size and then is converted to first- strand cDNA.

Total RNA samples containing 50 ng to 4 g RNA are suitable for the mRNA library preparation automation protocol. Each total RNA sample must be prepared for the run in 25 L of nuclease- free water.

NOTE For optimal performance, total RNA samples should have an RNA Integrity Number (RIN) of 8 or more, based on analysis using Agilents 2100 Bioanalyzer.

Consider preparing an additional sequencing library in parallel, using a high- quality control RNA sample, such as Agilent QPCR Human Reference Total RNA (p/n 750500). Use of this control is especially recommended during the first run of the protocol, to verify that all protocol steps are being successfully performed. Routine use of this control is helpful for any required troubleshooting, in order to differentiate any performance issues related to RNA input from other factors.

A workstation operator must be present during this automation protocol to transfer plates between the workstation, which completes most liquid handling steps, and the thermal cycler, which is used for several incubation steps. In addition, the operator must prepare and dispense a master mix immediately before it is used in the automation protocol (see step 30 on page 48).

Prepare the workstation

1 Open the SureSelect setup form using the SureSelect_RNA_ILM.VWForm shortcut on your desktop.

2 Log in to the VWorks software.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

3 Turn on the ThermoCube, set to 0C, at position 9 of the Bravo deck. Be sure that the chiller reservoir contains at least 300 mL of 25% ethanol.

4 Clear the Labware MiniHub and BenchCel of all plates and tip boxes.

Prepare reagents for the run

5 Bring the reagents listed in Table 9 to room temperature and thaw the reagents listed in Table 10 on ice.

Table 9 Reagents brought to room temperature before use in protocol

Table 10 Reagents thawed and held on ice before use in protocol

Storage Location Kit Component Where Used in Protocol

SureSelect Poly-A Selection Module*, 4C

* May also be labeled as SureSelect Strand Specific RNA Library Prep, ILM, Box 2.

Oligo(dT) Microparticles (bottle) page 35

Bead Washing Buffer (bottle)

May also be labeled as RNA Seq Bead Washing Buffer.

page 35

Bead Elution Buffer (bottle)

May also be labeled as RNA Seq Bead Elution Buffer.

page 35

Bead Binding Buffer** (bottle)

** May also be labeled as RNA Seq Bead Binding Buffer.

page 35

Storage Location Kit Component Where Used in Protocol

SureSelect RNA Library Prep, ILM (Pre PCR)*, 20C

* May also be labeled as SureSelect Strand Specific RNA Library Prep, ILM, Box 1.

Fragmentation Mix (bottle)

May also be labeled as RNA Seq Fragmentation Mix.

page 36

First Strand Master Mix (tube with orange cap)

May also be labeled as RNA Seq First Strand Master Mix.

page 49

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 33

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

34 S

6 Locate or prepare a stock solution of 4 g/L Actinomycin D in DMSO. When first prepared, aliquot the stock solution into single- use volumes (typically 3 L) and store the aliquots at 20C, protected from light. Do not subject the aliquots to multiple freeze- thaw cycles. The aliquots may be stored for up to one year before use in the library preparation protocol.

A 3- L aliquot of this DMSO stock solution will be used on page 48 to prepare a fresh dilution of 120 ng/L Actinomycin D in water for the run.

CAUTION To ensure strand specificity, take care to follow the storage instructions above for the 4 g/L Actinomycin D in DMSO stock solution, including the one-year maximum storage duration.

Prepare the RNA samples source plate

7 Place 25 L of each RNA sample (0.054 g RNA in nuclease- free water) into the wells of a 96- well Eppendorf twin.tec plate. Load samples into the plate column- wise, in well order A1 to H1, then A2 to H2, ending with A12 to H12 for processing on the Agilent NGS Workstation.

SureSelect Strand- Specific RNA Library Prep runs may include 1, 2, 3, 4, 6, or 12 columns of the plate. See Using the Agilent NGS Workstation for SureSelect RNA Library Preparation for additional sample placement considerations.

NOTE

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

Prepare the oligo(dT) beads and mRNA purification source plates

When preparing each of the source plates below, add the indicated amount of reagent to wells of the source plate corresponding to the total RNA sample wells in step 7 above. For example, for 3- column runs, fill source well plate wells A1 to H3, but leave wells A4 to H12 empty.

8 Prepare the oligo(dT) beads source plate.

a Vortex the Oligo(dT) Microparticles until the suspension appears homogeneous and consistent in color. If bead aggregates are still present after vortexing, mix thoroughly by pipetting up and down until the suspension appears homogeneous.

b In a PCR plate that is compatible with the thermal cycler to be used in the run, place 25 L of the homogeneous Oligo(dT) bead suspension into each well to be used for sample purification.

9 Prepare the Bead Binding Buffer source plate. Place 30 L of Bead Binding Buffer into wells of a 96- well Eppendorf twin.tec plate. Fill each well that corresponds to an RNA sample well.

10 Prepare the Bead Elution Buffer source plate. Place 30 L of Bead Elution Buffer into wells of a 96- well Eppendorf twin.tec plate. Fill each well that corresponds to an RNA sample well.

11 Prepare the Bead Wash Buffer source plate. Place 410 L of Bead Washing Buffer into wells of a Nunc DeepWell plate. Fill each well that corresponds to an RNA sample well.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 35

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

36 S

Prepare the master mix source plate

12 Prepare the master mix source plate by adding the appropriate volume of Fragmentation Mix (see Table 11) to all wells of Column 1 of a Nunc DeepWell plate. The configuration of the source plate is shown in Figure 3.

Table 11 Preparation of the Master Mix Source Plate for mRNA_Purification_v1.0.pro

Master Mix Solution Position on Source Plate

Volume of Master Mix added per Well of Nunc Deep Well Source Plate

1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Fragmentation Mix Column 1

(A1-H1)

28.5 L 47.5 L 66.5 L 85.5 L 123.5 L 247.0 L

Figure 3 Initial configuration of master mix source plate for mRNA_Purification_v1.0.pro

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

Load the Agilent NGS Workstation

13 Load the Labware MiniHub according to Table 12, using the plate orientations shown in Figure 4.

Table 12 Initial MiniHub configuration for mRNA_Purification_v1.0.pro

Figure 4 Agilent Labware MiniHub plate orientation. For Thermo Scientific reservoirs, place the notched corner facing the center of the hub.

Vertical Shelf Position

Cassette 1 Cassette 2 Cassette 3 Cassette 4

Shelf 5 (Top) Empty Nunc DeepWell plate

Empty Nunc DeepWell plate

Empty Nunc DeepWell plate

Empty

Shelf 4 Empty Bead Binding Buffer in twin.tec plate

Empty Empty

Shelf 3 Empty Bead Elution Buffer in twin.tec plate

Empty Empty

Shelf 2 Empty tip box Empty Bead Wash Buffer in Nunc DeepWell plate

Empty

Shelf 1 (Bottom) New tip box Empty Empty Empty tip box

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 37

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

38 S

14 Load the Bravo deck according to Table 13.

Table 13 Initial Bravo deck configuration for mRNA_Purification_v1.0.pro

15 Load the BenchCel Microplate Handling Workstation according to Table 14.

Table 14 Initial BenchCel configuration for mRNA_Purification_v1.0.pro

Location Content

1 Empty waste reservoir (Axygen 96 Deep Well Plate, square wells)

4 Oligo(dT) beads in PCR plate seated on red insert (PCR plate type must be specified on setup form under step 2)

6 Empty PCR plate seated on red insert (PCR plate type must be specified on setup form under step 2)

7 Total RNA samples in twin.tec plate

9 Master Mix Source Plate seated on silver insert (Nunc DeepWell; see Figure 3 on page 36 for column content)

No. of Columns Processed

Rack 1 Rack 2 Rack 3 Rack 4

1 1 Tip box Empty Empty Empty

2 2 Tip boxes Empty Empty Empty

3 3 Tip boxes Empty Empty Empty

4 3 Tip boxes Empty Empty Empty

6 5 Tip boxes Empty Empty Empty

12 9 Tip boxes Empty Empty Empty

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

Run VWorks protocol mRNA_Purification_v1.0.pro

16 On the SureSelect setup form, under Select Protocol to Run, select mRNA_Purification_v1.0.pro.

17 Under Select PCR plate labware for Thermal Cycling, select the specific type of PCR plate that was loaded on Bravo deck positions 4 and 6. The plate type selected must be compatible with the thermal cycler to be used for incubation steps during the protocol.

During run setup, be sure to use the plate type selected from this menu at positions 4 and 6 of the Bravo deck. In addition, when the workstation issues prompts to add plates to postion 4 or 6 during the run, use only the same PCR plate type specified here.

NOTE

18 Select the number of columns of samples to be processed. Runs must include 1, 2, 3, 4, 6, or 12 columns.

19 Click Display Initial Workstation Setup.

20 Verify that the NGS workstation has been set up as displayed in the Workstation Setup region of the form.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 39

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

40 S

21 When verification is complete, click Run Selected Protocol.

If workstation devices do not respond when you start the run, but activity is recorded in the Log, verify that VWorks is not running in Simulation mode. See page 25 for more information.

NOTE

Running the mRNA_Purification_v1.0.pro protocol takes approximately 90 minutes, including four incubation periods on the thermal cycler.

During the automation protocol run, a workstation operator must be present to transfer plates between the workstation and thermal cycler when prompted, as detailed on the following pages.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

22 When the workstation has finished combining the RNA samples with the oligo(dT) beads, you will be prompted by VWorks as shown below.

a Remove the plate from position 4 of the Bravo deck and seal the plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 seconds.

b Briefly spin the plate in a centrifuge or mini- plate spinner to collect the liquid without pelleting the beads.

c Transfer the PCR plate to a thermal cycler (with the heated lid ON) and run the RNA denaturation + bead binding program shown in Table 15. After transferring the plate, click Continue on the VWorks screen.

Table 15 Thermal cycler program for RNA denaturation and RNA-bead binding

Step Temperature Time

Step 1 65C 5 minutes

Step 2 4C 1 minute

Step 3 20C 5 minutes

Step 4 20C Hold

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 41

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

42 S

23 After the thermal cycler reaches the 20C Hold step, and when prompted by the dialog below, transfer the RNA sample plate to position 4 of the Bravo deck, seated in the red insert. Carefully unseal the plate, then click Continue.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

24 When the workstation has finished collecting and washing the bead- bound RNA samples, you will be prompted by VWorks as shown below.

a Remove and discard the PCR plate from position 4 of the Bravo deck.

b Place a fresh PCR plate at position 4, seated in the red insert. The PCR plate type added here must be the same plate type as the one removed and as was specified during the run setup.

c After positioning the plate, click Continue on the VWorks screen.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 43

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

44 S

25 When the workstation has finished collecting and washing the bead- bound RNA samples, you will be prompted to transfer the plate to the thermal cycler for the RNA Elution step as shown below.

a Remove the plate from position 4 of the Bravo deck and seal the plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 seconds.

b Briefly spin the plate in a centrifuge or mini- plate spinner to collect the liquid without pelleting the beads.

c Transfer the PCR plate to a thermal cycler (with the heated lid ON) and run the RNA elution program shown in Table 16. After transferring the plate, click Continue on the VWorks screen.

Table 16 Thermal cycler program for RNA elution

Step Temperature Time

Step 1 80C 2 minutes

Step 2 4C 1 minute

Step 3 4C Hold

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

26 After the thermal cycler reaches the 4C Hold step and when prompted by the dialog below, transfer the RNA sample plate to position 4 of the Bravo deck, seated in the red insert. Carefully unseal the plate, then click Continue.

27 The workstation adds Bead Binding Buffer to the eluted RNA samples and then holds the samples at room temperature for 5 minutes to allow the poly(A) RNA to re- bind the beads.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 45

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

46 S

28 When the workstation has finished the collecting and washing the bound RNA samples in this second round of purification, you will be prompted by VWorks as shown below. Remove and discard the PCR plate from position 4, then click Continue.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

29 The workstation adds Fragmentation Mix to the bead- bound RNA samples in preparation for the RNA fragmentation step. When the workstation has finished, you will be prompted by VWorks as shown below.

a Remove the plate from position 6 of the Bravo deck and seal the plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 seconds.

b Briefly spin the plate in a centrifuge or mini- plate spinner to collect the liquid.

c Transfer the PCR plate to a thermal cycler (with the heated lid ON) and run the RNA fragmentation program shown in Table 17. After transferring the plate, click Continue on the VWorks screen.

Table 17 Thermal cycler program for RNA fragmentation

Step Temperature Time

Step 1 94C 8 minutes

Step 2 4C Hold

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 47

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

48 S

30 During the 8- minute incubation step, prepare the reagents and workstation for first- strand cDNA synthesis:

a When prompted by the dialog below, place a fresh PCR plate (use the plate type specified during the run setup) at position 6, seated in the red insert. Proceed immediately to step b, below.

b Prepare a fresh 120 ng/L Actinomycin D dilution in water from a stock solution of 4 g/L Actinomycin D in DMSO, according to Table 18.

CAUTION To ensure strand-specificity, you must prepare the 120 ng/L Actinomycin D dilution in water immediately before use in step d, below.

Table 18 Preparation of 120 ng/l Actinomycin D

Reagent Volume for up to 12-column run (includes excess)

Actinomycin D (4 g/l in DMSO) 3 L

Nuclease-free water 97 L

Total 100 L

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

c Vortex the thawed vial of First Strand Master Mix 5 seconds at high speed to ensure homogeneity.

CAUTION SureSelect RNA Library Prep master mixes are viscous. Mix thoroughly by vortexing at high speed for 5 seconds before removing an aliquot for use and after combining the master mix with other solutions. Pipetting up and down is not sufficient to mix this reagent.

d Prepare the appropriate amount of First Strand Master Mix + Actinomycin D mixture, according to the table below. Mix by vortexing, then spin briefly and keep on ice.

Table 19 Preparation of First Strand Master Mix/Actinomycin D mixture

Reagent Volume for 1 Library

Volume for 1 Column

Volume for 2 Columns

Volume for 3 Columns

Volume for 4 Columns

Volume for 6 Columns

Volume for 12 Columns

First Strand Master Mix

8.0 L 98.4 L 196.8 L 262.4 L 360.8 L 492.0 L 918.4 L

Actinomycin D (120 ng/l in H2O)

0.5 L 6.2 L 12.3 L 16.4 L 22.6 L 30.8 L 57.4 L

Total Volume 8.5 L 104.6 L 209.1 L 278.8 L 383.4 L 522.8 L 975.8 L

e Add the volume listed in Table 20 of the First Strand Master Mix + Actinomycin D mixture to column 2 of the Master Mix source plate at position 9 of the Bravo deck. The final configuration of the source plate is shown in Figure 13. After adding the master mix to the source plate, click Continue on the VWorks screen.

Table 20 Preparation of the Master Mix Source Plate for mRNA_Purification_v1.0.pro

Master Mix Solution Position on Source Plate

Volume of Master Mix added per Well of Nunc Deep Well Source Plate

1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

First Strand Master Mix + Actinomycin D mixture

Column 2

(A2-H2)

12.0 L 25.1 L 33.8 L 46.9 L 64.3 L 120.9 L

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 49

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

50 S

Figure 5 Final configuration of master mix source plate for mRNA_Purification_v1.0.pro

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

SureSelect Automat

31 With the RNA sample plate still on the thermal cycler, the workstation prepares the remaining components for first- strand cDNA synthesis. When the workstation has finished, you will be prompted by VWorks as shown below.

a After the thermal cycler reaches the 4C Hold step for the RNA fragmentation program (Table 17), remove the plate from the thermal cycler and briefly spin in a centrifuge or mini- plate spinner to collect the liquid.

b Place the RNA sample plate on position 4 of the Bravo deck, seated in the red insert.

c Carefully unseal the plate, then click Continue.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 51

3 Sample Preparation Step 1. Purify poly(A) RNA/fragment RNA/synthesize first strand cDNA

52 S

32 The workstation removes the fragmented RNA samples from the bead- containing wells and combines the samples with First Strand Master Mix + Actinomycin D. When the workstation has finished, you will be prompted by VWorks as shown below.

a Remove the plate from position 6 of the Bravo deck and seal the plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 seconds.

b Briefly spin the plate in a centrifuge or mini- plate spinner to collect the liquid.

c Transfer the PCR plate to a thermal cycler (with the heated lid ON) and run the first- strand cDNA synthesis program shown in Table 21. After transferring the plate, click Continue on the VWorks screen.

Table 21 Thermal cycler program for first-strand cDNA synthesis

Step Temperature Time

Step 1 25C 10 minutes

Step 2 37C 40 minutes

Step 3 4C Hold

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 2. Purify first-strand cDNA using AMPure XP beads

Step 2. Purify first-strand cDNA using AMPure XP beads

SureSelect Automat

In this step, the Agilent NGS Workstation transfers AMPure XP beads and first- strand cDNA samples to a Nunc DeepWell plate and then collects and washes the bead- bound DNA.

Prepare the workstation and reagents

1 Leave tip boxes on shelves 1 and 2 in casette 1 of the Labware MiniHub from the previous mRNA_Purification_v1.0.pro run. Otherwise, clear the remaining positions of the MiniHub and BenchCel of plates and tip boxes.

2 Gently wipe down the Labware MiniHub, Bravo deck, and BenchCel with a DNA Away decontamination wipe.

3 Turn on the ThermoCube, set to 0C, at position 9 of the Bravo deck. Be sure that the chiller reservoir contains at least 300 mL of 25% ethanol.

4 Let the AMPure XP beads come to room temperature for at least 30 minutes. Do not freeze the beads at any time.

5 Mix the bead suspension well so that the reagent appears homogeneous and consistent in color.

6 Prepare a Nunc DeepWell source plate for the beads by adding 51 L of homogeneous AMPure XP beads per well, for each well to be processed.

7 Prepare a Thermo Scientific reservoir containing 15 mL of nuclease- free water.

8 Prepare a separate Thermo Scientific reservoir containing 45 mL of freshly- prepared 70% ethanol.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 53

3 Sample Preparation Step 2. Purify first-strand cDNA using AMPure XP beads

54 S

9 Load the Labware MiniHub according to Table 22, using the plate orientations shown in Figure 4.

Table 22 Initial MiniHub configuration for AMPureXP_v1.1.pro:First Strand

10 Load the Bravo deck according to Table 23.

Table 23 Initial Bravo deck configuration for AMPureXP_v1.1.pro:First Strand

Vertical Shelf Position

Cassette 1 Cassette 2 Cassette 3 Cassette 4

Shelf 5 (Top) Empty Nunc DeepWell plate

Empty Empty Empty

Shelf 4 Empty Empty Empty Empty

Shelf 3 Empty Empty Eppendorf twin.tec Plate

Empty Empty

Shelf 2 Waste tip box (retained from mRNA_Purification protocol)*

* The tip boxes retained in Cassette 1 are not shown on the VWorks Workstation Setup table. These tip boxes are not used in AMPureXP_v1.1.pro:First Strand but are used in a later protocol. This labware should be retained in the MiniHub to ensure that empty and full tip positions are properly defined for the subsequent protocol.

Nuclease-free water reservoir from step 7

AMPure XP beads in Nunc DeepWell plate from step 6

Empty

Shelf 1 (Bottom) Clean tip box (retained from mRNA_Purification protocol)*

70% ethanol reservoir from step 8

Empty Empty tip box

Location Content

1 Empty waste reservoir (Axygen 96 Deep Well Plate, square wells)

9 First-strand cDNA samples in PCR plate seated on red insert (PCR plate type must be specified on setup form under step 2)

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 2. Purify first-strand cDNA using AMPure XP beads

SureSelect Automat

11 Load the BenchCel Microplate Handling Workstation according to Table 24.

Table 24 Initial BenchCel configuration for AMPureXP_v1.1.pro:First Strand

Run VWorks protocol AMPureXP_v1.1.pro:First Strand

12 On the SureSelect setup form, under Select Protocol to Run, select AMPureXP_v1.1.pro:First Strand.

13 Under Select PCR plate labware for Thermal Cycling, select the specific type of PCR plate containing the cDNA samples at position 9.

14 Select the number of columns of samples to be processed. Runs must include 1, 2, 3, 4, 6, or 12 columns.

15 Click Display Initial Workstation Setup.

16 Verify that the NGS workstation has been set up as displayed in the Workstation Setup region of the form.

No. of Columns Processed

Rack 1 Rack 2 Rack 3 Rack 4

1 1 Tip box Empty Empty Empty

2 1 Tip box Empty Empty Empty

3 2 Tip boxes Empty Empty Empty

4 2 Tip boxes Empty Empty Empty

6 3 Tip boxes Empty Empty Empty

12 6 Tip boxes Empty Empty Empty

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 55

3 Sample Preparation Step 2. Purify first-strand cDNA using AMPure XP beads

56 S

17 When verification is complete, click Run Selected Protocol.

Running the AMPureXP_v1.1.pro:First Strand protocol takes approximately 45 minutes. During this time, you can prepare the purification reagents for the Library Prep automation protocol, as described on page 58.

Once the AMPureXP_v1.1.pro:First Strand protocol is complete, the purified cDNA samples are located in the Eppendorf plate at position 7 of the Bravo deck. Proceed immediately to Step 3. Prepare cDNA libraries for Illumina sequencing on page 57.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 3. Prepare cDNA libraries for Illumina sequencing

Step 3. Prepare cDNA libraries for Illumina sequencing

SureSelect Automat

This step is automated using the LibraryPrep_RNASeq_ILM_v1.1.rst runset. During the runset, the Agilent NGS Workstation completes second- strand cDNA library synthesis and the end modification steps required for sequencing, including end- repair, A- tailing, and adaptor ligation. After certain modification steps, the Agilent NGS Workstation purifies the prepared cDNA using AMPure XP beads.

This step uses the SureSelect RNA Reagent Kit components listed in Table 25 in addition to the purification reagents prepared for use on page 58. Thaw each reagent vial and keep on ice. Vortex each vial for 5 seconds at high speed to mix before use.

Prepare the workstation

1 Turn on the ThermoCube, set to 0C, at position 9 of the Bravo deck. Be sure that the chiller reservoir contains at least 300 mL of 25% ethanol.

2 Leave tip boxes on shelves 1 and 2 in cassette 1 of the Labware MiniHub from the previous mRNA_Purification_v1.0.pro run. Otherwise, clear the remaining MiniHub and BenchCel positions of plates and tip boxes.

Table 25 Reagents for automation runset LibraryPrep_RNASeq_ILM_v1.1.rst

Storage Location Kit Component Where Used in Protocol

SureSelect RNA Library Prep, ILM (Pre PCR)*, 20C

* May also be labeled as SureSelect Strand Specific RNA Library Prep, ILM, Box 1.

Second Strand Enzyme Mix (bottle)

May also be labeled as RNA Seq Second Strand + End Repair Enzyme Mix.

page 59

Second Strand Oligo Mix (tube with yellow cap)

May also be labeled as RNA Seq Second Strand + End Repair Oligo Mix.

page 59

SureSelect Ligation Master Mix (tube with purple cap) page 59

SureSelect Oligo Adaptor Mix (tube with blue cap) page 59

dA Tailing Master Mix** (bottle)

** May also be labeled as RNA Seq dA Tailing Master Mix.

page 59

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 57

3 Sample Preparation Step 3. Prepare cDNA libraries for Illumina sequencing

58 S

3 Pre- set the temperature of Bravo deck position 4 to 14C and of position 6 to 4C using the Inheco Multi TEC control touchscreen, as described in Setting the Temperature of Bravo Deck Heat Blocks. On the control touchscreen, Bravo deck positions 4 corresponds to CPAC 2, position 1, while deck position 6 corresponds to CPAC 2, position 2.

Prepare the purification reagents

4 Verify that the AMPure XP bead suspension is at room temperature. Do not freeze the beads at any time.

5 Mix the bead suspension well so that the reagent appears homogeneous and consistent in color.

6 Prepare a Nunc DeepWell source plate for the beads by adding 160 L of homogeneous AMPure XP beads per well, for each well to be processed.

7 Prepare a Thermo Scientific reservoir containing 20 mL of nuclease- free water.

8 Prepare a separate Thermo Scientific reservoir containing 150 mL of freshly- prepared 70% ethanol.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 3. Prepare cDNA libraries for Illumina sequencing

SureSelect Automat

Prepare the master mix source plate

9 Vortex the thawed vials of Second Strand Enzyme Mix, SureSelect Ligation Master Mix, and dA Tailing Master Mix for 5 seconds at high speed to ensure homogeneity.

CAUTION SureSelect RNA Library Prep master mixes are viscous. Mix thoroughly by vortexing at high speed for 5 seconds before removing an aliquot for use and after combining the master mix with other solutions. Pipetting up and down is not sufficient to mix these reagents.

10 Prepare the appropriate amount of Second Strand + End Repair Master Mix according to Table 26 below.

Table 26 Preparation of Second Strand + End Repair Master Mix for LibraryPrep_RNASeq_ILM_v1.1.rst

Reagent Volume for 1 Library

Volume for 1 Column

Volume for 2 Columns

Volume for 3 Columns

Volume for 4 Columns

Volume for 6 Columns

Volume for 12 Columns

Second Strand Enzyme Mix

25.0 L 307.5 L 615 L 820 L 1127.5 L 1640 L 3075 L

Second Strand Oligo Mix

5.0 L 61.5 L 123 L 164 L 225.5 L 328 L 615 L

Total Volume 30 L 369 L 738 L 984 L 1353 L 1968 L 3690 L

11 Prepare the appropriate amount of Adaptor Ligation Master Mix, containing the SureSelect Ligation Master Mix and the adaptors, according to Table 27 below.

Table 27 Preparation of Adaptor Ligation Master Mix for LibraryPrep_RNASeq_ILM_v1.1.rst

Reagent Volume for 1 Library

Volume for 1 Column

Volume for 2 Columns

Volume for 3 Columns

Volume for 4 Columns

Volume for 6 Columns

Volume for 12 Columns

Nuclease-free water 2.5 L 30.8 L 61.5 L 82.0 L 112.8 L 164.0 L 307.5 L

SureSelect Ligation Master Mix

5.0 L 61.5 L 123.0 L 164.0 L 225.5 L 328.0 L 615.0 L

SureSelect Oligo Adaptor Mix

5.0 L 61.5 L 123.0 L 164.0 L 225.5 L 328.0 L 615.0 L

Total Volume 12.5 L 153.8 L 307.5 L 410.0 L 563.8 L 820.0 L 1537.5 L

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 59

3 Sample Preparation Step 3. Prepare cDNA libraries for Illumina sequencing

60 S

12 Using the same Nunc DeepWell master mix source plate that was used for the mRNA_Purification_v1.0.pro run, prepare the Library Prep master mix source plate. Add the volumes indicated in Table 28 of each master mix to all wells of the indicated column of the plate. Keep the master mixes on ice during the aliquoting steps. The final configuration of the master mix source plate is shown in Figure 13.

Table 28 Preparation of the Master Mix Source Plate for LibraryPrep_RNASeq_ILM_v1.1.rst

Master Mix Solution

Position on Source Plate

Volume of Master Mix added per Well of Nunc Deep Well Source Plate

1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Second Strand + End Repair Master Mix (Table 26)

Column 3

(A3-H3)

42.4 L 88.5 L 119.3 L 165.4 L 242.3 L 457.5 L

dA Tailing Master Mix

Column 4

(A4-H4)

30.0 L 50.0 L 70.0 L 90.0 L 130.0 L 260.0 L

Adaptor Ligation Master Mix (from Table 27)

Column 5

(A5-H5)

17.7 L 36.9 L 49.7 L 68.9 L 100.9 L 190.6 L

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 3. Prepare cDNA libraries for Illumina sequencing

SureSelect Automat

Figure 6 Configuration of the master mix source plate for LibraryPrep_RNASeq_ ILM_v1.1.rst

If you are using a new DeepWell plate for the Library Prep Master Mix source plate, leave columns 1 and 2 empty and add the PCR Master Mix to columns 3 to 5 of the new plate.

NOTE

13 Seal the master mix source plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 sec.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 61

3 Sample Preparation Step 3. Prepare cDNA libraries for Illumina sequencing

62 S

14 Centrifuge the plate for 30 seconds to drive the well contents off the walls and plate seal and to eliminate any bubbles. Keep the master mix source plate on ice.

The presence of bubbles in source plate solutions may cause inaccurate volume transfer by the Bravo liquid handling platform. Ensure that the source plate is sealed and centrifuged prior to use in a run.

NOTE

Load the Agilent NGS Workstation

15 Load the Labware MiniHub according to Table 29, using the plate orientations shown in Figure 4.

Table 29 Initial MiniHub configuration for LibraryPrep_RNASeq_ILM_v1.1.rst

Vertical Shelf Position

Cassette 1 Cassette 2 Cassette 3 Cassette 4

Shelf 5 (Top) Empty Empty Nunc DeepWell plate

Empty Nunc DeepWell plate

Empty

Shelf 4 Empty Empty Eppendorf twin.tec plate

Empty Eppendorf twin.tec plate

Empty

Shelf 3 Empty Empty Empty Eppendorf twin.tec plate

Empty

Shelf 2 Waste tip box (retained from mRNA_Purification protocol)

Nuclease-free water reservoir from step 7

AMPure XP beads in Nunc DeepWell plate from step 6

Empty

Shelf 1 (Bottom) Clean tip box (retained from mRNA_Purification protocol)

70% ethanol reservoir from step 8

Empty Empty tip box

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 3. Prepare cDNA libraries for Illumina sequencing

SureSelect Automat

16 Load the Bravo deck according to Table 30.

Table 30 Initial Bravo deck configuration for LibraryPrep_RNASeq_ILM_v1.1.rst

17 Load the BenchCel Microplate Handling Workstation according to Table 31.

Table 31 Initial BenchCel configuration for LibraryPrep_RNASeq_ILM_v1.1.rst

Run VWorks runset LibraryPrep_RNASeq_ILM_v1.1.rst

18 On the SureSelect setup form, under Select Protocol to Run, select LibraryPrep_RNASeq_ILM_v1.1.rst.

19 Select the number of columns of samples to be processed. Runs must include 1, 2, 3, 4, 6, or 12 columns.

Location Content

1 Empty waste reservoir (Axygen 96 Deep Well Plate, square wells)

6 Empty Eppendorf twin.tec plate, oriented with well A1 in the upper-left

7 Purified first-strand cDNA samples in Eppendorf twin.tec plate, oriented with well A1 in the upper-left

9 Library Prep Master Mix Source Plate (Nunc DeepWell), unsealed and seated on silver insert

No. of Columns Processed

Rack 1 Rack 2 Rack 3 Rack 4

1 2 Tip boxes Empty Empty Empty

2 3 Tip boxes Empty Empty Empty

3 4 Tip boxes Empty Empty Empty

4 5 Tip boxes Empty Empty Empty

6 7 Tip boxes Empty Empty Empty

12 11 Tip boxes 3 Tip boxes Empty Empty

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 63

3 Sample Preparation Step 3. Prepare cDNA libraries for Illumina sequencing

64 S

20 Click Display Initial Workstation Setup.

21 Verify that the NGS workstation has been set up as displayed in the Workstation Setup region of the form.

22 When verification is complete, click Run Selected Protocol.

23 When ready to begin the run, click OK in the following window.

Running the LibraryPrep_RNASeq_ILM_v1.1.rst runset takes approximately 3 hours. Once complete, the purified, adaptor- ligated cDNA samples are located in the Eppendorf plate at position 7 of the Bravo deck.

Stopping Point

If you do not continue to the next step, seal the plate and store at 20C.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 4. Index cDNA libraries by PCR amplification

Step 4. Index cDNA libraries by PCR amplification

SureSelect Automat

In this step, the Agilent NGS Workstation completes the liquid handling steps for PCR amplification of the adaptor- ligated cDNA samples in a 3- primer PCR that includes the appropriate indexing primer. After the reactions are set up by the workstation, you transfer the PCR plate to a thermal cycler for amplification. The amplification cycle number is based on the initial amount of total RNA sample used for library preparation.

Before you begin, thaw and mix the reagents listed in Table 32 below and keep on ice.

Prepare the workstation

1 Turn on the ThermoCube, set to 0C, at position 9 of the Bravo deck. Be sure that the chiller reservoir contains at least 300 mL of 25% ethanol.

2 Clear the Labware MiniHub and BenchCel of all plates and tip boxes.

3 Pre- set the temperature of Bravo deck position 6 to 4C using the Inheco Multi TEC control touchscreen, as described in Setting the Temperature of Bravo Deck Heat Blocks. Bravo deck position 6 corresponds to CPAC 2, position 2 on the Multi TEC control touchscreen.

Table 32 Reagents for use in pre-capture PCR protocol

Storage Location Kit Component Where Used in Protocol

SureSelect RNA Library Prep, ILM (Pre PCR)*, 20C

* May also be labeled as SureSelect Strand Specific RNA Library Prep, ILM, Box 1.

PCR Master Mix (bottle)

May also be labeled as RNA Seq PCR Master Mix.

page 67

Uracil DNA Glycosylase (UDG) (tube with yellow cap)

page 67

SureSelect Primer (tube with brown cap) page 67

ILM Reverse PCR Primer (tube with black cap)

May also be labeled as RNA Seq ILM Reverse PCR Primer.

page 66

SureSelectXT Indexes, 8 bp (blue 96-well plate) page 66

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 65

3 Sample Preparation Step 4. Index cDNA libraries by PCR amplification

66 S

Prepare the index source plate

4 Add 5 L of the appropriate indexing primer (Index A01H06)to the appropriate wells of a PCR plate.

The well position for each index should correspond to the position of the RNA sample assigned to that index in the original total RNA sample plate.

CAUTION For 96-reaction kits, only use the indexing primers provided in columns 1 through 6 (Indexes A01H06) in step 4. Provided plates also contain indexing primers in columns 7 through 12, which are used in other SureSelect protocols.

See page 90 for a plate map.

Prepare the PCR reaction mix and the master mix source plate

5 Prepare a 1:20 dilution of the ILM Reverse PCR Primer, according to Table 33.

Table 33 Preparation of reverse primer dilution

Reagent Volume for up to 96 reactions (includes excess)

Nuclease-free water 114 L

ILM Reverse PCR Primer 6 L

Total 120 L

CAUTION The PCR Master Mix used at this step is highly viscous and thorough mixing prior to use is critical for optimal kit performance. Mix by vortexing at high speed for 5 seconds before removing an aliquot for use and after combining with other solutions as directed in Table 34. Pipetting up and down is not sufficient to mix this reagent.

6 Vortex the thawed vial of PCR Master Mix 5 seconds at high speed to ensure homogeneity.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 4. Index cDNA libraries by PCR amplification

SureSelect Automat

7 Prepare the appropriate volume of PCR Reaction Mix, according to Table 34. Mix well by vortexing at high speed and keep on ice.

Table 34 Preparation of PCR Reaction Mix

SureSelect Reagent

Volume for 1 Library

Volume for 1 Column

Volume for 2 Columns

Volume for 3 Columns

Volume for 4 Columns

Volume for 6 Columns

Volume for 12 Columns

PCR Master Mix 25.0 L 307.5 L 512.5 L 717.5 L 922.5 L 1332.5 L 2665 L

Uracil DNA Glycosylase (UDG)

1.0 L 12.3 L 20.5 L 28.7 L 36.9 L 53.3 L 106.6 L

SureSelect Primer

(Forward primer)

1.0 L 12.3 L 20.5 L 28.7 L 36.9 L 53.3 L 106.6 L

ILM Reverse PCR Primer (1:20 from Table 33)

1.0 L 12.3 L 20.5 L 28.7 L 36.9 L 53.3 L 106.6 L

Total Volume 28 L 344.4 L 574.0 L 803.6 L 1033.2 L 1492.4 L 2984.8 L

8 Using the same Nunc DeepWell master mix source plate that was used for the LibraryPrep_RNASeq_ILM_v1.1.rst run, add the volume of PCR Master Mix indicated in Table 35 to all wells of column 6 of the master mix source plate. The final configuration of the master mix source plate is shown in Figure 7.

Table 35 Preparation of the Master Mix Source Plate for TranscriptomePCR_ILM_v1.0.pro

If you are using a new DeepWell plate for the PCR source plate, leave columns 1 to 5 empty and add the PCR Master Mix to column 6 of the new plate.

Master Mix Solution

Position on Source Plate

Volume of Master Mix added per Well of Nunc Deep Well Source Plate

1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

PCR Reaction Mix Column 6

(A6-H6)

39.6 L 68.3 L 97.0 L 125.7 L 183.1 L 369.6 L

NOTE

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 67

3 Sample Preparation Step 4. Index cDNA libraries by PCR amplification

68 S

Figure 7 Configuration of the master mix source plate for Transcriptome- PCR_ILM_v1.0.pro. Columns 1-5 were used to dispense master mixes during previous protocols.

9 Seal the master mix source plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 sec.

10 Centrifuge the plate for 30 seconds to drive the well contents off the walls and plate seal and to eliminate any bubbles.

The presence of bubbles in source plate solutions may cause inaccurate volume transfer by the Bravo liquid handling platform. Ensure that the source plate is sealed and centrifuged prior to use in a run.

NOTE

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 4. Index cDNA libraries by PCR amplification

SureSelect Automat

Load the Agilent NGS Workstation

11 Load the Labware MiniHub according to Table 36, using the plate orientations shown in Figure 4.

Table 36 Initial MiniHub configuration for TranscriptomePCR_ILM_v1.0.pro

12 Load the Bravo deck according to Table 37.

Table 37 Initial Bravo deck configuration for TranscriptomePCR_ILM_v1.0.pro

Vertical Shelf Position

Cassette 1 Cassette 2 Cassette 3 Cassette 4

Shelf 5 (Top)

Empty Empty Empty Empty

Shelf 4 Empty Empty Empty Empty

Shelf 3 Empty Empty Empty Empty

Shelf 2 Empty tip box Empty Empty Empty

Shelf 1 (Bottom)

New tip box Empty Empty Empty tip box

Location Content

6 Indexing primers in PCR plate seated on red insert (PCR plate type must be specified on setup form under step 2)

7 Prepped cDNA samples in Eppendorf twin.tec plate, oriented with well A1 in the upper-left

9 Master mix source plate (Nunc DeepWell), unsealed and seated on silver insert

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 69

3 Sample Preparation Step 4. Index cDNA libraries by PCR amplification

70 S

13 Load the BenchCel Microplate Handling Workstation according to Table 38.

Table 38 Initial BenchCel configuration for TranscriptomePCR_ILM_v1.0.pro

Run VWorks protocol TranscriptomePCR_ILM_v1.0.pro

14 On the SureSelect setup form, under Select Protocol to Run, select TranscriptomePCR_ILM_v1.0.pro.

15 Under Select PCR plate labware for Thermal Cycling, select the specific type of PCR plate that was loaded on Bravo deck position 6. The plate type selected must be compatible with the thermal cycler to be used for amplification.

16 Select the number of columns of samples to be processed. Runs must include 1, 2, 3, 4, 6, or 12 columns.

17 Click Display Initial Workstation Setup.

18 Verify that the NGS workstation has been set up as displayed in the Workstation Setup region of the form.

No. of Columns Processed

Rack 1 Rack 2 Rack 3 Rack 4

1 1 Tip box Empty Empty Empty

2 1 Tip box Empty Empty Empty

3 1 Tip box Empty Empty Empty

4 1 Tip box Empty Empty Empty

6 1 Tip box Empty Empty Empty

12 1 Tip box Empty Empty Empty

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 4. Index cDNA libraries by PCR amplification

SureSelect Automat

19 When verification is complete, click Run Selected Protocol.

20 Running the TranscriptomePCR_ILM_v1.0.pro protocol takes approximately 15 minutes. Once complete, the PCR- ready samples, containing prepped DNA, PCR master mix, and indexing primer are located in the PCR plate at position 6 of the Bravo deck and you will see the following prompt:

a Remove the PCR plate from position 6 of the Bravo deck and seal the plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 3.0 seconds.

b Centrifuge the plate for 30 seconds to drive the well contents off the walls and plate seal and to eliminate air bubbles.

c Transfer the plate to the thermal cycler (with the heated lid ON) and run the program in Table 39.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 71

3 Sample Preparation Step 4. Index cDNA libraries by PCR amplification

72 S

Table 39 Thermal cycler program for mRNA Library PCR indexing

Segment Number of Cycles Temperature Time

1 1 37C 15 minutes

2 1 95C 2 minutes

3 1016 cycles (see Table 40)

95C 30 seconds

65C 30 seconds

72C 1 minute

4 1 72C 5 minutes

5 1 4C Hold

Table 40 mRNA Library PCR indexing cycle number recommendations

Amount of total RNA used for library prep Cycle Number

50 ng200 ng 1416

201 ng2 g 1214

2.1 g4 g 1012

NOTE If you started with the minimum total RNA input amount of 50 ng, use 16 amplification cycles in the PCR indexing amplification program (Table 39).

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 5. Purify amplified DNA using AMPure XP beads

Step 5. Purify amplified DNA using AMPure XP beads

SureSelect Automat

In this step, the Agilent NGS Workstation transfers AMPure XP beads and amplified cDNA libraries to a Nunc DeepWell plate and then collects and washes the bead- bound DNA.

Prepare the workstation and reagents

1 Clear the Labware MiniHub and BenchCel of all plates and tip boxes.

2 Gently wipe down the Labware MiniHub, Bravo deck, and BenchCel with a DNA Away decontamination wipe.

3 Turn on the ThermoCube, set to 0C, at position 9 of the Bravo deck. Be sure that the chiller reservoir contains at least 300 mL of 25% ethanol.

4 Let the AMPure XP beads come to room temperature for at least 30 minutes. Do not freeze the beads at any time.

5 Mix the bead suspension well so that the reagent appears homogeneous and consistent in color.

6 Prepare a Nunc DeepWell source plate for the beads by adding 125 L of homogeneous AMPure XP beads per well, for each well to be processed.

7 Prepare a Thermo Scientific reservoir containing 30 mL of nuclease- free water.

8 Prepare a separate Thermo Scientific reservoir containing 90 mL of freshly- prepared 70% ethanol.

NOTE The reagents prepared in step 6 to step 8 above will be used for two sequential purification protocols and should be retained on the workstation at the conclusion of this protocol.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 73

3 Sample Preparation Step 5. Purify amplified DNA using AMPure XP beads

74 S

9 Load the Labware MiniHub according to Table 41, using the plate orientations shown in Figure 4.

Table 41 Initial MiniHub configuration for AMPureXP_v1.1.pro:Transcriptome PCR

10 Load the Bravo deck according to Table 42.

Table 42 Initial Bravo deck configuration for AMPureXP_v1.1.pro:Transcriptome PCR

Vertical Shelf Position

Cassette 1 Cassette 2 Cassette 3 Cassette 4

Shelf 5 (Top) Empty Nunc DeepWell plate

Empty Empty Empty

Shelf 4 Empty Empty Empty Empty

Shelf 3 Empty Empty Eppendorf twin.tec Plate

Empty Empty

Shelf 2 Empty Nuclease-free water reservoir from step 7

AMPure XP beads in Nunc DeepWell plate from step 6

Empty

Shelf 1 (Bottom) Empty 70% ethanol reservoir from step 8

Empty Empty tip box

Location Content

1 Empty waste reservoir (Axygen 96 Deep Well Plate, square wells)

9 Amplified cDNA samples in PCR plate, unsealed and seated on red insert (PCR plate type must be specified on setup form under step 2)

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 5. Purify amplified DNA using AMPure XP beads

SureSelect Automat

11 Load the BenchCel Microplate Handling Workstation according to Table 43.

Table 43 Initial BenchCel configuration for AMPureXP_v1.1.pro:Transcriptome PCR

Run VWorks protocol AMPureXP_v1.1.pro:Transcriptome PCR

12 On the SureSelect setup form, under Select Protocol to Run, select AMPureXP_v1.1.pro:Transcriptome PCR.

13 Under Select PCR plate labware for Thermal Cycling, select the specific type of PCR plate that was loaded on Bravo deck position 9.

14 Select the number of columns of samples to be processed. Runs must include 1, 2, 3, 4, 6, or 12 columns.

15 Click Display Initial Workstation Setup.

16 Verify that the NGS workstation has been set up as displayed in the Workstation Setup region of the form.

No. of Columns Processed

Rack 1 Rack 2 Rack 3 Rack 4

1 1 Tip box Empty Empty Empty

2 1 Tip box Empty Empty Empty

3 2 Tip boxes Empty Empty Empty

4 2 Tip boxes Empty Empty Empty

6 3 Tip boxes Empty Empty Empty

12 6 Tip boxes Empty Empty Empty

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 75

3 Sample Preparation Step 5. Purify amplified DNA using AMPure XP beads

76 S

17 When verification is complete, click Run Selected Protocol.

The purification protocol takes approximately 45 minutes. When complete, the purified DNA samples are in the Eppendorf plate located on Bravo deck position 7.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 6. Remove adaptor-dimers using AMPure XP beads

Step 6. Remove adaptor-dimers using AMPure XP beads

SureSelect Automat

In this step, the Agilent NGS Workstation completes a second round of purification of the amplified cDNA libraries to remove any adaptor- dimers prior to sequencing.

Prepare the workstation and reagents

1 Leave the purification reagents (Nunc plate containing beads and reservoirs containing water and ethanol) in cassettes 2 and 3 of the Labware MiniHub from the previous AMPureXP_v1.1.pro:Transcriptome PCR run. Otherwise, clear the remaining positions of the MiniHub and BenchCel of plates and tip boxes.

2 Leave the waste reservior at position 1 of the Bravo deck. Clear the remaining positions of the Bravo deck of labware.

3 Verify that the ThermoCube at position 9 of the Bravo deck remains set to 0C.

4 Load the Labware MiniHub according to Table 44, using the plate orientations shown in Figure 4.

Table 44 Initial MiniHub configuration for AMPureXP_v1.1.pro:Transcriptome Dimers

Vertical Shelf Position

Cassette 1 Cassette 2 Cassette 3 Cassette 4

Shelf 5 (Top) Empty Nunc DeepWell plate

Empty Empty Empty

Shelf 4 Empty Empty Empty Empty

Shelf 3 Empty Empty Eppendorf twin.tec Plate

Empty Empty

Shelf 2 Empty Nuclease-free water reservoir (retained from previous protocol)

AMPure XP beads in Nunc DeepWell plate (retained from previous protocol)

Empty

Shelf 1 (Bottom) Empty 70% ethanol reservoir (retained from previous protocol)

Empty Empty tip box

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 77

3 Sample Preparation Step 6. Remove adaptor-dimers using AMPure XP beads

78 S

5 Load the Bravo deck according to Table 45.

Table 45 Initial Bravo deck configuration for AMPureXP_v1.1.pro:Transcriptome Dimers

6 Load the BenchCel Microplate Handling Workstation according to Table 46.

Table 46 Initial BenchCel configuration for AMPureXP_v1.1.pro:Transcriptome Dimers

Location Content

1 Waste reservoir (Axygen 96 Deep Well Plate, square wells, retained from previous protocol)

9 Once-purified cDNA samples in Eppendorf twin.tec plate seated on red insert (full-skirted twin.tec plate type must be specified on setup form under step 2)

No. of Columns Processed

Rack 1 Rack 2 Rack 3 Rack 4

1 1 Tip box Empty Empty Empty

2 1 Tip box Empty Empty Empty

3 2 Tip boxes Empty Empty Empty

4 2 Tip boxes Empty Empty Empty

6 3 Tip boxes Empty Empty Empty

12 6 Tip boxes Empty Empty Empty

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 6. Remove adaptor-dimers using AMPure XP beads

SureSelect Automat

Run VWorks protocol AMPureXP_v1.1.pro:Transcriptome Dimers

7 On the SureSelect setup form, under Select Protocol to Run, select AMPureXP_v1.1.pro:Transcriptome Dimers.

8 Under Select PCR plate labware for Thermal Cycling, select 96 Eppendorf Twin.tec PCR in Red Alum Insert.

9 Select the number of columns of samples to be processed. Runs must include 1, 2, 3, 4, 6, or 12 columns.

10 Click Display Initial Workstation Setup.

11 Verify that the NGS workstation has been set up as displayed in the Workstation Setup region of the form.

12 When verification is complete, click Run Selected Protocol.

The purification protocol takes approximately 45 minutes. When complete, the purified DNA samples are in the Eppendorf plate located on Bravo deck position 7.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 79

3 Sample Preparation Step 7. Assess Library DNA quantity and quality

Step 7. Assess Library DNA quantity and quality

80 S

Option 1: Analysis using the Agilent 2100 Bioanalyzer and DNA 1000 Assay

Use a Bioanalyzer DNA 1000 chip and reagent kit. Perform the assay according to the Agilent DNA 1000 Kit Guide.

1 Set up the 2100 Bioanalyzer as instructed in the reagent kit guide.

2 Seal the sample plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 sec.

3 Vortex the plate to mix samples in each well, then centrifuge the plate for 30 seconds to drive the well contents off the walls and plate seal.

4 Prepare the chip, samples and ladder as instructed in the reagent kit guide, using 1 L of each sample for the analysis.

5 Load the prepared chip into the 2100 Bioanalyzer and start the run within five minutes after preparation.

6 Measure the concentration of the library (ng/L) by integrating under the peak at approximately 200 to 600 bp. For accurate quantification, make sure that the concentration falls within the linear range of the assay. A sample electropherogram is shown in Figure 8.

Stopping Point

Figure 8 Analysis of amplified library DNA using a DNA 1000 assay.

If you do not continue to the next step, seal the plate and store at 20C.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 7. Assess Library DNA quantity and quality

SureSelect Automat

Option 2: Analysis using Agilent 4200 TapeStation and D1000 ScreenTape

Use a D1000 ScreenTape and associated reagent kit to analyze the amplified libraries. For more information to do this step, see the Agilent D1000 Assay Quick Guide.

1 Seal the DNA sample plate using the PlateLoc Thermal Microplate Sealer, with sealing settings of 165C and 1.0 sec.

2 Vortex the plate to mix samples in each well, then centrifuge the plate for 30 seconds to drive the well contents off the walls and plate seal.

3 Prepare the TapeStation samples as instructed in the instrument user manual. Use 1 L of each amplified library DNA sample diluted with 3 L of D1000 sample buffer for the analysis.

CAUTION For accurate quantitation, make sure to thoroughly mix the combined DNA and sample buffer by vortexing the assay plate or tube strip for 1 minute on the IKA MS3 vortex mixer provided with the 4200 TapeStation system before loading the samples.

If no IKA MS3 vortex mixer is available, ensure that the samples are mixed thoroughly by vortexing for 10 seconds at high speed on a manual vortex mixer.

4 Load the sample plate or tube strips from step 3, the D1000 ScreenTape, and loading tips into the 4200 TapeStation as instructed in the Assay Quick Guide. Start the run.

5 For each sample, measure the concentration of the library (ng/L) by integrating under the peak at approximately 200 to 600 bp. A sample electropherogram is shown in Figure 9.

Stopping Point

If you do not continue to the next step, seal the plate and store at 20C

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 81

3 Sample Preparation Step 7. Assess Library DNA quantity and quality

82 S

.

Figure 9 Analysis of amplified library DNA using a D1000 ScreenTape.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 8. Pool samples for multiplexed sequencing

Step 8. Pool samples for multiplexed sequencing

SureSelect Automat

The number of indexed libraries that may be multiplexed in a single sequencing lane is determined by the output specifications of the platform used, together with the amount of sequencing data required for your research design. Calculate the number of indexes that can be combined per lane, according to the capacity of your platform and the amount of sequencing data required per sample.

1 Combine the libraries such that each index- tagged sample is present in equimolar amounts in the pool. For each library, use the formula below to determine the amount of indexed sample to use.

Volume of Index V f C f # C i

--------------------------------=

where V(f) is the final desired volume of the pool,

C(f) is the desired final concentration of all the DNA in the pool

# is the number of indexes, and

C(i) is the initial concentration of each indexed sample.

Table 47 shows an example of the amount of 4 index- tagged samples (of different concentrations) and Low TE needed for a final volume of 20 L at 10 nM.

2 Adjust the final volume of the pooled library to the desired final concentration.

Table 47 Example of indexed sample volume calculation for total volume of 20 L

Component V(f) C(i) C(f) # Volume to use (L)

Sample 1 20 L 20 nM 10 nM 4 2.5

Sample 2 20 L 10 nM 10 nM 4 5

Sample 3 20 L 17 nM 10 nM 4 2.9

Sample 4 20 L 25 nM 10 nM 4 2

Low TE 7.6

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 83

3 Sample Preparation Step 8. Pool samples for multiplexed sequencing

84 S

If the final volume of the combined index- tagged samples is less than the desired final volume, V(f), add Low TE to bring the volume to the desired level.

If the final volume of the combined index- tagged samples is greater than the final desired volume, V(f), lyophilize and reconstitute to the desired volume.

3 If you store the library before sequencing, add Tween 20 to 0.1% v/v and store at - 20C short term.

Exact library pool dilution and processing can vary based on the flow cell capacity and analysis pipeline versions being used. Refer to the appropriate Illumina user guide for instructions.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Sample Preparation 3 Step 9. Prepare and analyze sequencing samples

Step 9. Prepare and analyze sequencing samples

SureSelect Automat

Proceed to cluster amplification using the Illumina Paired- End Cluster Generation Kit; refer to the manufacturers instructions for this step. The optimal seeding concentration for cluster amplification from SureSelect mRNA libraries is approximately 10- 12 pM.

NOTE The optimal seeding concentration may vary, depending on the method used for library quantification and fragment size distribution.

This protocol has been validated with 2 x 100- base paired- end reads. However, read length can be adjusted to achieve the desired research goals.

Sequencing run setup guidelines for 8-bp indexes

Sequencing runs must be set up to perform an 8- nt index read. For the HiSeq platform, use the Cycles settings shown in Table 48. Cycle number settings can be specified on the Run Configuration screen of the instrument control software interface after choosing Custom from the index type selection buttons.

For complete index sequence information, see the Reference chapter starting on page 87.

Table 48 HiSeq platform Run Configuration screen Cycle Number settings

Run Segment Cycle Number

Read 1 100

Index 1 (i7) 8

Index 2 (i5) 0

Read 2 100

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 85

3 Sample Preparation Step 9. Prepare and analyze sequencing samples

86 S

Sequence analysis guidelines

The SureSelect RNA sequencing library preparation method preserves RNA strandedness using dUTP second- strand marking. The sequence of read 1, which starts at the P5 end, matches the reverse complement of the poly- A RNA transcript strand. Read 2, which starts at the P7 end, matches the poly- A RNA transcript strand. When running analysis of this data to determine strandedness, it is important to include this information. For example, when using the Picard tools (https://broadinstitute.github.io/picard) to calculate RNA sequencing metrics, it is important to include the parameter STRAND_SPECIFICITY= SECOND_READ_TRANSCRIPTION_STRAND to correctly calculate the strand specificity metrics.

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

SureSelect Automated Strand-Specific RNA Library Prep Protocol

4 Reference

Kit Contents 88

Nucleotide Sequences of SureSelectXT Indexes A01 to H12 91

Plasticware quantities for automation protocols 92

This chapter contains reference information, including component kit contents, index sequences, and plasticware requirements.

87Agilent Technologies

4 Reference Kit Contents

Kit Contents

88 S

The SureSelect Strand Specific RNA Reagent Kit includes the component kits listed in Table 49. The contents of each component kit are detailed in Table 50 and Table 51.

Table 49 Kit Content

Component Kits Storage Condition Part Number

SureSelect RNA Library Prep, ILM (Pre PCR)*

* May also be labeled as SureSelect Strand Specific RNA Library Prep, ILM, Box 1.

20C 5500-0135

SureSelect Poly-A Selection Module (Pre PCR)

May also be labeled as SureSelect Strand Specific RNA Library Prep, ILM, Box 2.

4C 5190-6411

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Reference 4 Kit Contents

SureSelect Automat

Table 50 SureSelect RNA Library Prep, ILM (Pre PCR) Content

Equivalent RNA Library Preparation Kit Component Names* Format (96 Reaction Kit, p/n 5500-0135)

Current name Retired name

Fragmentation Mix RNA Seq Fragmentation Mix bottle

First Strand Master Mix RNA Seq First Strand Master Mix tube with orange cap

Second Strand Enzyme Mix RNA Seq Second-Strand + End Repair Enzyme Mix

bottle

Second Strand Oligo Mix RNA Seq Second-Strand + End Repair Oligo Mix

tube with yellow cap

dA Tailing Master Mix RNA Seq dA Tailing Master Mix bottle

SureSelect Ligation Master Mix (no change) tube with purple cap

SureSelect Oligo Adaptor Mix (no change) tube with blue cap

PCR Master Mix RNA Seq PCR Master Mix bottle

Uracil DNA Glycosylase (UDG) (no change) tube with yellow cap

SureSelect Primer (no change) tube with brown cap

ILM Reverse PCR Primer RNA Seq ILM Reverse PCR Primer tube with black cap

ILM Post-Capture PCR Primer RNA Seq ILM Post-Capture PCR Primer tube with green cap

SureSelectXT Indexes, 8 bp** (no change) SureSelect 8 bp Indexes A01 through H12, provided in blue 96-well plate

* Some component names were updated in June, 2020. Formulations of the reagents supplied and protocols for use of the re- agents are unchanged. All components are supported through the expiration date listed on the Certificate of Analysis.

The Second Strand Enzyme Mix and Second Strand Oligo Mix also supply the reagents used for cDNA end repair; formula- tions are equivalent to vials labeled as RNA Seq Second Strand + End Repair Enzyme Mix and Second Strand + End Repair Oligo Mix, respectively.

The provided ILM Post-capture PCR Primer is not used in the workflow described in this manual.

** See Table 53 on page 91 for index sequences.

See Table 52 on page 90 for a plate map. Although the provided plate contains 96 indexing primers, only indexes A01H06 should be used for the mRNA library preparation workflow. Wells contain enough volume for two mRNA library preparation reactions per index, using the protocol on page 66.

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 89

4 Reference Kit Contents

90 S

Table 51 SureSelect Poly-A Selection Module (Pre PCR) Content

Kit Component Format (96 Reaction Kit, p/n 5190-6411)*

* Component kit part number 5190-6411 labeled as SureSelect Strand Specific RNA Library Prep, ILM, Box 2 also contains a vial of Nuclease Free Water and contains reagent tubes labeled as RNA Seq Bead Binding Buffer, RNA Seq Bead Washing Buffer, and RNA Seq Bead Elution Buffer. Formula- tions of the reagents supplied and protocols for use of these reagents are unchanged. Ordering in- formation for nuclease-free water is provided in Table 1 on page 12.

Oligo(dT) Microparticles bottle

Bead Binding Buffer bottle

Bead Washing Buffer bottle

Bead Elution Buffer bottle

for SSEL 8bp Indexes A01 through H12 provided in blue plate in Library Prep kit p/n 5500-0135. Use

Table 52 Plate map only indexes A01H06 (Columns 1 to 6) for the mRNA library preparation workflow.

1 2 3 4 5 6 7 8 9 10 11 12

A A01 A02 A03 A04 A05 A06 A07 A08 A09 A10 A11 A12

B B01 B02 B03 B04 B05 B06 B07 B08 B09 B10 B11 B12

C C01 C02 C03 C04 C05 C06 C07 C08 C09 C10 C11 C12

D D01 D02 D03 D04 D05 D06 D07 D08 D09 D10 D11 D12

E E01 E02 E03 E04 E05 E06 E07 E08 E09 E10 E11 E12

F F01 F02 F03 F04 F05 F06 F07 F08 F09 F10 F11 F12

G G01 G02 G03 G04 G05 G06 G07 G08 G09 G10 G11 G12

H H01 H02 H03 H04 H05 H06 H07 H08 H09 H10 H11 H12

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Reference 4 Nucleotide Sequences of SureSelectXT Indexes A01 to H12

Nucleotide Sequences of SureSelectXT Indexes A01 to H12

SureSelect Automat

Each index is 8 nt in length. See page 85 for sequencing run setup requirements for sequencing libraries using 8- bp indexes.

Table 53 SureSelect Indexes, for indexing primers in blue 96-well plates

Index Sequence Index Sequence Index Sequence Index Sequence

A01 ATGCCTAA A04 AACTCACC A07 ACGTATCA A10 AATGTTGC

B01 GAATCTGA B04 GCTAACGA B07 GTCTGTCA B10 TGAAGAGA

C01 AACGTGAT C04 CAGATCTG C07 CTAAGGTC C10 AGATCGCA

D01 CACTTCGA D04 ATCCTGTA D07 CGACACAC D10 AAGAGATC

E01 GCCAAGAC E04 CTGTAGCC E07 CCGTGAGA E10 CAACCACA

F01 GACTAGTA F04 GCTCGGTA F07 GTGTTCTA F10 TGGAACAA

G01 ATTGGCTC G04 ACACGACC G07 CAATGGAA G10 CCTCTATC

H01 GATGAATC H04 AGTCACTA H07 AGCACCTC H10 ACAGATTC

A02 AGCAGGAA A05 AACGCTTA A08 CAGCGTTA A11 CCAGTTCA

B02 GAGCTGAA B05 GGAGAACA B08 TAGGATGA B11 TGGCTTCA

C02 AAACATCG C05 CATCAAGT C08 AGTGGTCA C11 CGACTGGA

D02 GAGTTAGC D05 AAGGTACA D08 ACAGCAGA D11 CAAGACTA

E02 CGAACTTA E05 CGCTGATC E08 CATACCAA E11 CCTCCTGA

F02 GATAGACA F05 GGTGCGAA F08 TATCAGCA F11 TGGTGGTA

G02 AAGGACAC G05 CCTAATCC G08 ATAGCGAC G11 AACAACCA

H02 GACAGTGC H05 CTGAGCCA H08 ACGCTCGA H11 AATCCGTC

A03 ATCATTCC A06 AGCCATGC A09 CTCAATGA A12 CAAGGAGC

B03 GCCACATA B06 GTACGCAA B09 TCCGTCTA B12 TTCACGCA

C03 ACCACTGT C06 AGTACAAG C09 AGGCTAAC C12 CACCTTAC

D03 CTGGCATA D06 ACATTGGC D09 CCATCCTC D12 AAGACGGA

E03 ACCTCCAA E06 ATTGAGGA E09 AGATGTAC E12 ACACAGAA

F03 GCGAGTAA F06 GTCGTAGA F09 TCTTCACA F12 GAACAGGC

G03 ACTATGCA G06 AGAGTCAA G09 CCGAAGTA G12 AACCGAGA

H03 CGGATTGC H06 CCGACAAC H09 CGCATACA H12 ACAAGCTA

ed Strand-Specific RNA Library Prep for Illumina Platform Sequencing 91

4 Reference Plasticware quantities for automation protocols

Plasticware quantities for automation protocols

92 S

The tables below show the quantity of each plasticware type used in each automation protocol in the workflow. Quantities listed in the tables only include unique labware that was not used in other protocols or runsets. For example, Nunc DeepWell master mix plates may be reused in multiple protocols but are counted below only where first used.

mRNA_Purification_v1.0.pro

AMPureXP_v1.1.pro:First Strand

Labware 1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Tip boxes (filled) 2 3 4 4 6 10

Empty tip boxes (for waste tips) 2 2 2 2 2 2

Nunc DeepWell Plates 5 5 5 5 5 5

96 Eppendorf twin.tec full-skirt plates

3 3 3 3 3 3

PCR plates (compatible with thermal cycler)

4 4 4 4 4 4

Axygen square-well plate (waste) 1 1 1 1 1 1

Labware 1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Tip boxes (filled) 1 1 2 2 3 6

Empty tip boxes (for waste tips) 1 1 1 1 1 1

Nunc DeepWell Plates 2 2 2 2 2 2

96 Eppendorf twin.tec full-skirt plates

1 1 1 1 1 1

Thermo Scientific Reservoirs 2 2 2 2 2 2

Axygen square-well plate (waste) 1 1 1 1 1 1

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

Reference 4 Plasticware quantities for automation protocols

LibraryPrep_RNASeq_ILM_v1.1.rst

TranscriptomePCR_ILM_v1.0.pro

AMPureXP_v1.1.pro:Transcriptome PCR

Labware 1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Tip boxes (filled) 2 3 4 5 7 14

Empty tip boxes (for waste tips) 1 1 1 1 1 1

Nunc DeepWell Plates 3 3 3 3 3 3

96 Eppendorf twin.tec full-skirt plates

4 4 4 4 4 4

Thermo Scientific Reservoirs 2 2 2 2 2 2

Axygen square-well plate (waste) 1 1 1 1 1 1

Labware 1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Tip boxes (filled) 2 2 2 2 2 2

Empty tip boxes (for waste tips) 2 2 2 2 2 2

PCR plates (compatible with thermal cycler)

1 1 1 1 1 1

Labware 1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Tip boxes (filled) 1 1 2 2 3 6

Empty tip boxes (for waste tips) 1 1 1 1 1 1

Nunc DeepWell Plates 2 2 2 2 2 2

96 Eppendorf twin.tec full-skirt plates

1 1 1 1 1 1

Thermo Scientific Reservoirs 2 2 2 2 2 2

Axygen square-well plate (waste) 1 1 1 1 1 1

SureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing 93

4 Reference Plasticware quantities for automation protocols

AMPureXP_v1.1.pro:Transcriptome Dimers

Labware 1-Column Runs

2-Column Runs

3-Column Runs

4-Column Runs

6-Column Runs

12-Column Runs

Tip boxes (filled) 1 1 2 2 3 6

Empty tip boxes (for waste tips) 1 1 1 1 1 1

Nunc DeepWell Plates 1 1 1 1 1 1

96 Eppendorf twin.tec full-skirt plates

1 1 1 1 1 1

94 S

ureSelect Automated Strand-Specific RNA Library Prep for Illumina Platform Sequencing

www.agilent.com

In This Book

This guide contains information to run the SureSelect Automated Strand- Specific RNA Library Prep protocol using the Agilent NGS Workstation.

Agilent Technologies, Inc. 2013, 2014, 2017, 2020

Version F0, June 2020

*G9691-90030* G9691-90030

Manualsnet FAQs

If you want to find out how the SureSelect Automated Strand-Specific RNA Library Prep Agilent works, you can view and download the Agilent SureSelect Automated Strand-Specific RNA Library Prep Handler Protocol on the Manualsnet website.

Yes, we have the Protocol for Agilent SureSelect Automated Strand-Specific RNA Library Prep as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Protocol should include all the details that are needed to use a Agilent SureSelect Automated Strand-Specific RNA Library Prep. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent SureSelect Automated Strand-Specific RNA Library Prep Handler Protocol is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent SureSelect Automated Strand-Specific RNA Library Prep Handler Protocol consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent SureSelect Automated Strand-Specific RNA Library Prep Handler Protocol free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent SureSelect Automated Strand-Specific RNA Library Prep Handler Protocol, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent SureSelect Automated Strand-Specific RNA Library Prep Handler Protocol as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.