Diode Array Detector WR and Multiple Wavelength Detector Agilent InfinityLab LC Series

User Manual

Agilent InfinityLab LC Series DAD WR and MWD User Manual

Notices

Document Information Document No: SD-29000218 Rev. E Edition: 03/2021

Copyright Agilent Technologies, Inc. 2016-2021

No part of this manual may be repro- duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written con- sent from Agilent Technologies, Inc. as governed by United States and interna- tional copyright laws.

Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future edi- tions. Further, to the maximum extent permitted by applicable law, Agilent dis- claims all warranties, either express or implied, with regard to this manual and any information contained herein, includ- ing but not limited to the implied warran- ties of merchantability and fitness for a particular purpose. Agilent shall not be lia- ble for errors or for incidental or conse- quential damages in connection with the furnishing, use, or performance of this document or of any information con- tained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend U.S. Government Restricted Rights. Soft- ware and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Soft- ware and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Depart- ment of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commer- cial Computer Software or Computer Software Documentation).

Safety Notices

CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARN- ING notice until the indicated conditions are fully understood and met.

Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn, Germany

Agilent InfinityLab LC Series DAD WR and MWD User Manual

In This Guide...

In This Guide...

This manual covers the Agilent 1260 Infinity II Diode Array Detector WR and the Agilent 1260 Infinity II Multiple Wavelength Detector modules:

G7115A - 1260 Infinity II DAD WR

G7165A - 1260 Infinity II MWD

1 Introduction

This chapter gives an introduction to the module.

2 Site Requirements and Specifications

This chapter provides information on environmental requirements, physical and performance specifications.

3 Using the Module

This chapter explains the essential operational parameters of the module.

4 Preparing the Detector

This chapter provides information on how to set up the module for an analysis and explains the basic settings.

5 Optimizing the Detector

This chapter provides information on how to optimize the detector.

6 Troubleshooting and Diagnostics

This chapter gives an overview about the troubleshooting and diagnostic features.

7 Error Information

This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 3

In This Guide...

8 Test Functions and Calibration

This chapter describes the tests for the module.

9 Maintenance

This chapter describes the maintenance of the detector.

10 Parts for Maintenance

This chapter provides information on parts for maintenance and repair.

11 Identifying Cables

This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.

12 Hardware Information

This chapter describes the detector in more detail on hardware and electronics.

13 LAN Configuration

This chapter provides information on connecting the module to the Agilent ChemStation PC.

14 Appendix

This chapter provides addition information on safety, legal and web.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 4

Contents

1 Introduction 9

Introduction to the Detector 10 G7115A Diode Array Detector WR 11 G7165A Multiple Wavelength Detector 13 Optical System 15 Leak and Waste Handling 17 Bio-inert Materials 22

2 Site Requirements and Specifications 24

Site Requirements 25 Physical Specifications 28 Performance Specifications 29

3 Using the Module 36

Magnets 37 Turn on/off 38 Status Indicators 40 Instrument Configuration 41 Set up the Detector with Agilent Open Lab ChemStation 42 The Detector User Interface 43 Detector Control Settings 45 Method Parameter Settings 46 Agilent Local Control Modules 50

4 Preparing the Detector 51

Leak and Waste Handling 52 Setting up an Analysis 54 Solvent Information 61

Agilent InfinityLab LC Series DAD WR and MWD User Manual 5

5 Optimizing the Detector 67

Optimizing the Detector Performance 68 Optimization Overview 69 Optimizing for Sensitivity, Selectivity, Linearity and Dispersion 71 Optimizing Selectivity 81 Optimizing the Detector Regarding to the System 83 Warm up of the Detector 89

6 Troubleshooting and Diagnostics 91

Available Tests vs User Interfaces 92 Agilent Lab Advisor Software 93 Available Diagnostic Functions vs. Product Level 94

7 Error Information 97

What Are Error Messages 98 General Error Messages 99 Detector Error Messages 107

8 Test Functions and Calibration 115

Introduction 116 Conditions of Detector 120 Failing a Test 121 Self-Test 122 Intensity Test 124 Cell Test 127 Filter Test 129 Holmium Oxide Test 131 ASTM Drift and Noise Test 134 Slit Test 138 Wavelength Verification Test 141 Wavelength Calibration 143 D/A Converter (DAC) Test 146 Dark-Current Test 148 Spectral Scan 150 Other Lab Advisor Functions 151

Agilent InfinityLab LC Series DAD WR and MWD User Manual 6

9 Maintenance 152

Introduction to Maintenance 153 Cautions and Warnings 154 Overview of Maintenance 156 Cleaning the Module 157 Remove and Install Doors 158 Replace a Lamp 160 Remove and Install a Flow Cell 163 Maintenance of Standard, Semi-Micro or Micro Flow Cell 167 Maintenance of High Pressure Flow Cell 170 Replacing Capillaries on a Standard Flow Cell 172 Replacing Capillaries on a Semi-Micro and Micro Flow Cell 177 Nano Flow Cell - Replacing or Cleaning 181 Cleaning or Exchanging the Holmium Oxide Filter 185 Correcting Leaks 188 Replacing Leak Handling System Parts 190 Replacing the Modules Firmware 192 Information from Modules Assemblies 194

10 Parts for Maintenance 195

Overview of Maintenance Parts 196 Standard Flow Cell 198 Standard Flow Cell Bio-inert 200 Semi-Micro Flow Cell 202 Micro Flow Cell 204 High Pressure Flow Cell 206 Prep Flow Cell - SST 208 Prep Flow Cell - Quartz 210 Nano Flow Cells 212 Accessory Kits 215 Holmium Oxide Filter 216 Leak Handling Parts 217

Agilent InfinityLab LC Series DAD WR and MWD User Manual 7

11 Identifying Cables 218

Cable Overview 219 Analog Cables 221 Remote Cables 223 BCD Cables 226 CAN/LAN Cables 228 Agilent 1200 module to PC 229

12 Hardware Information 230

Firmware Description 231 Electrical Connections 234 Interfaces 236 Setting the 6-bit Configuration Switch 243 Instrument Layout 247 Early Maintenance Feedback (EMF) 248

13 LAN Configuration 250

What You Have to Do First 251 TCP/IP parameter configuration 252 Configuration Switches 253 Initialization Mode Selection 254 Dynamic Host Configuration Protocol (DHCP) 256 Manual Configuration 259

14 Appendix 264

General Safety Information 265 Waste Electrical and Electronic Equipment (WEEE) Directive 271 Radio Interference 272 Sound Emission 273 UV-Radiation 274 Solvent Information 275 Declaration of Conformity for HOX2 Filter 276 Installation of Stainless Steel Cladded PEEK Capillaries 277 Agilent Technologies on Internet 283

Agilent InfinityLab LC Series DAD WR and MWD User Manual 8

1 Introduction

Introduction to the Detector 10

G7115A Diode Array Detector WR 11 Product Description 11 Features 12

G7165A Multiple Wavelength Detector 13 Product Description 13 Features 14

Optical System 15

Leak and Waste Handling 17 Leak Sensor 21 Waste Concept 21

Bio-inert Materials 22

This chapter gives an introduction to the module.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 9

1 Introduction Introduction to the Detector

Introduction to the Detector

The detector is designed for highest optical performance, GLP compliance and easy maintenance. It includes the following features:

120 Hz data acquisition rate for (ultra-) fast LC applications,

RFID tags for all flow cells and UV-lamps provides traceable information about these assemblies,

long-life deuterium with RFID tag and tungsten lamps for highest intensity and lowest detection limit over a wavelength range of 190 950 nm,

no loss in sensitivity for up to eight wavelengths simultaneous,

programmable slit from 1 16 nm for complete optimization of sensitivity, linearity and spectral resolution,

optional flow-cell cartridges with RFID tag (standard 10 mm13 L, semi-micro 6 mm5 L, micro 3 mm2 L, 80 nL, 500 nL, 10 mm, high pressure 10 mm1.7 L and prep-cells) are available and can be used depending on the application needs,

easy front access to lamps and flow cell for fast replacement, and

built-in holmium oxide filter for fast wavelength accuracy verification,

built-in temperature control for improved baseline stability,

additional diagnostic signals for temperature and lamp voltage monitoring,

For specifications, see Specifications on page 29.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 10

1 Introduction G7115A Diode Array Detector WR

G7115A Diode Array Detector WR

Product Description The 1260 Infinity II DAD WR detector is designed for highest optical performance, GLP compliance, and easy maintenance. With its 120 Hz data acquisition rate the detector is perfectly suited for fast LC applications. The long life deuterium lamps allow highest intensity and lowest detection limits over a wavelength range of 190 950 nm. The use of RFID tags for all flow cells and UV-lamps provides traceable information about these assemblies.

The built-in holmium oxide filter features the fast wavelength accuracy verification, while the built-in temperature controls improves the baseline stability. Additional diagnostic signals for temperature and lamp voltage monitoring are available.

Figure 1 Overview of the detector

Status indicator

Flow cell

Power switch

Leak drain

Agilent InfinityLab LC Series DAD WR and MWD User Manual 11

1 Introduction G7115A Diode Array Detector WR

Features Higher sensitivity and selectivity - simultaneous detection of up to eight

compound-specific wavelengths.

Low detection limits - low noise front-end electronics and the patented flow cell design delivers very low detection limits thanks to the minimization of short-term noise (< 7 AU).

Up to 100 % resolution gain in fast LC - using an 120 Hz data acquisition rate.

Maximum baseline stability - electronic temperature control (ETC) reduces baseline drift under fluctuating ambient temperature and humidity conditions.

Wide linear range - for reliable, simultaneous quantification of primary compounds, by-products, and impurities.

Programmable slit (1 16 nm) for rapid optimization of sensitivity and linearity.

Excellent data traceability - radio frequency identification (RFID) tags on cells and source lamps improve traceability of data.

Automatic wavelength verification by built-in holmium oxide filter.

Nine analytical and preparative flow cells provide you with maximum application flexibility and choice.

Extensive diagnostics, error detection and display with Instant Pilot controller and Lab Advisor software.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 12

1 Introduction G7165A Multiple Wavelength Detector

G7165A Multiple Wavelength Detector

Product Description The 1260 Infinity II Multiple Wavelength Detector is designed for highest optical performance, GLP compliance, and easy maintenance. With its 120 Hz data acquisition rate, the detector is perfectly suited for fast LC applications. The long-life deuterium lamps allows highest intensity and lowest detection limits over a wavelength range of 190 950 nm. The use of RFID tags for all flow cells and UV-lamps provides traceable information about these assemblies.

The built-in holmium oxide filter features the fast wavelength accuracy verification, while the built-in temperature controls improves the baseline stability. Additional diagnostic signals for temperature and lamp voltage monitoring are available.

Figure 2 Overview of the detector

Status indicator

Flow cell

Power switch

Leak drain

Agilent InfinityLab LC Series DAD WR and MWD User Manual 13

1 Introduction G7165A Multiple Wavelength Detector

Features Higher sensitivity and selectivity - simultaneous detection of up to eight

compound-specific wavelengths.

Low detection limits - low noise front-end electronics and the patented flow cell design delivers very low detection limits thanks to the minimization of short-term noise (< 7 AU).

Up to 100 % resolution gain in fast LC - using a 120 Hz data acquisition rate.

Maximum baseline stability - electronic temperature control (ETC) reduces baseline drift under fluctuating ambient temperature and humidity conditions.

Wide linear range - for reliable, simultaneous quantification of primary compounds, by-products and impurities.

Programmable slit (1 16 nm) for rapid optimization of sensitivity and linearity.

Excellent data traceability - radio frequency identification (RFID) tags on cells and source lamps improve traceability of data.

Automatic wavelength verification by built-in holmium oxide filter.

Nine analytical and preparative flow cells provide you with maximum application flexibility and choice.

Extensive diagnostics, error detection and display with Instant Pilot controller and Lab Advisor software.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 14

1 Introduction Optical System

Optical System

The optical system of the detector is shown in Figure below. Its illumination source is a combination of a deuterium-arc-discharge lamp for the ultraviolet (UV) wavelength range and a tungsten lamp for the visible (VIS) and short-wave near-infrared (SWNIR) wavelength range. The image of the filament of the tungsten lamp is focused on the discharge aperture of the deuterium lamp by means of a special rear-access lamp design which allows both light sources to be optically combined and share a common axis to the source lens. The achromat (source lens) forms a single, focused beam of light through the flow cell. Each cell room and lamp are separated by a quartz window which can be cleaned or replaced. In the spectrograph, light is being dispersed onto the diode array by a holographic grating. This allows simultaneous access to all wavelength information.

Figure 3 Optical System of the Detector

Lamps The light source for the UV-wavelength range is a deuterium lamp with a shine-through aperture. As a result of plasma discharge in low-pressure deuterium gas, the lamp emits light over the 190 nm to approximately 800 nm wavelength range. The light source for the visible and SWNIR wavelength range is a low noise tungsten lamp. This lamp emits light over the wavelength range 470 950 nm.

Tungsten lamp

Deuterium lamp

Holmium oxide filter

Achromat (source lens)

Flow cell

Slit

Diode array Grating

Cell support window

Spectro lens

Coupling lens

Agilent InfinityLab LC Series DAD WR and MWD User Manual 15

1 Introduction Optical System

Achromat (Source Lens)

The achromat receives the light from both lamps and focuses it so that the beam passes through the flow cell.

Holmium Oxide Filter

The holmium oxide filter is electromechanically actuated. During the holmium filter test it moves into the light path.

Cell Support Window

The cell support window assembly separates the holmium filter area from the flow cell area.

Flow Cell Compartment

The optical unit has a flow cell compartment for easy access to flow cells. A variety of optional flow cells can be inserted using the same quick, simple mounting system. The flow cell can be removed to check the optical and electronic performance of the detector without having influences from the flow cell.

Spectrograph The spectrograph material is ceramic to reduce thermal effects to a minimum. The spectrograph consists of the spectrograph lens, the variable entrance slit, the grating and the photodiode array with front-end electronics. The spectrograph lens refocuses the light beam after it has passed through the flow cell. The sampling interval of the diode array is < 1 nm over the wavelength range 190 950 nm. Depending on the wavelength this varies from 1.0 to 1.25 diodes per nanometer (for example a diode every 0.8 to 1 nm).

For a small wavelength range, the small non-linearity could be neglected. With the wavelength range from 190 950 nm a new approach is required to achieve wavelength accuracy over the full range. Each spectrograph is calibrated individually. The calibration data is stored in the spectrograph on an EEPROM. Based on these data, the built-in processors calculate absorbance data with linear intervals (1.0, 2.0, ) between data points. This results in an excellent wavelength accuracy and instrument-to-instrument reproducibility.

Variable Entrance Slit

System

The micro-slit system makes use of the mechanical properties of silicon combined with the precise structuring capabilities of bulk micro-machining. It combines the required optical functions slit and shutter in a simple and compact component. The slit width is directly controlled by the micro-processor of the instrument and can be set as method parameter.

Grating The combination of dispersion and spectral imaging is accomplished by using a concave holographic grating. The grating separates the light beam into all its component wavelengths and reflects the light onto the photodiode array.

Diode Array The diode array is a series of 1024 individual photodiodes and control circuits located on a ceramic carrier. With a wavelength range from 190 950 nm the sampling interval is < 1 nm.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 16

1 Introduction Leak and Waste Handling

Leak and Waste Handling

The Agilent InfinityLab LC Series has been designed for safe leak and waste handling. It is important that all security concepts are understood and instructions are carefully followed.

The solvent cabinet is designed to store a maximum volume of 8 L solvent. The maximum volume for an individual bottle stored in the solvent cabinet should not exceed 2 L. For details, see the usage guideline for the Agilent Infinity II Solvent Cabinets (a printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available on the Internet).

All leak plane outlets are situated in a consistent position so that all Infinity and Infinity II modules can be stacked on top of each other. Waste tubes are guided through a channel on the right hand side of the instrument, keeping the front access clear from tubes.

The leak plane provides leak management by catching all internal liquid leaks, guiding them to the leak sensor for leak detection, and passing them on to the next module below, if the leak sensor fails. The leak sensor in the leak plane stops the running system as soon as the leak detection level is reached.

Solvent and condensate is guided through the waste channel into the waste container:

from the detector's flow cell outlet

from the Multisampler needle wash port

from the Sample Cooler or Sample Thermostat (condensate)

from the pump's Seal Wash Sensor (if applicable)

from the pump's Purge Valve or Multipurpose Valve

Agilent InfinityLab LC Series DAD WR and MWD User Manual 17

1 Introduction Leak and Waste Handling

Figure 4 Infinity II Leak Waste Concept (Flex Bench installation)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 18

1 Introduction Leak and Waste Handling

Figure 5 Infinity II Single Stack Leak Waste Concept (bench installation)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 19

1 Introduction Leak and Waste Handling

Figure 6 Infinity II Two Stack Leak Waste Concept (bench installation)

The waste tube connected to the leak plane outlet on each of the bottom instruments guides the solvent to a suitable waste container.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 20

1 Introduction Leak and Waste Handling

Leak Sensor

Waste Concept 1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with

4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.

CAUTION Solvent incompatibility

The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor. The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.

Do not use DMF as mobile phase.

Check the leak sensor regularly for corrosion.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 21

1 Introduction Bio-inert Materials

Bio-inert Materials

For the Bio-inert LC system, Agilent Technologies uses highest quality materials in the flow path (also referred to as wetted parts), which are widely accepted by life science scientists, as they are known for optimum inertness to biological samples and ensure best compatibility with common samples and solvents over a wide pH range. Explicitly, the complete flow path is free of stainless steel and free of other alloys containing metals such as iron, nickel, cobalt, chromium, molybdenum or copper, which can interfere with biological samples. The flow downstream of the sample introduction contains no metals whatsoever.

Table 1 Used Bio-inert materials

Module Materials

Agilent 1260 Infinity II Bio-inert Pump (G5654A)

Titanium, gold, platinum-iridium, ceramic, ruby, PTFE, PEEK

Agilent 1260 Infinity II Bio-inert Multisampler (G5668A)

Upstream of sample introduction: Titanium, gold, PTFE, PEEK, ceramic

Downstream of sample introduction: PEEK, ceramic

Agilent 1260 Infinity II Bio-inert Manual Injector (G5628A)

PEEK, ceramic

Agilent 1260 Infinity II Bio-inert Analytical Fraction Collector (G5664B)

PEEK, ceramic, PTFE

Bio-inert Flow Cells:

Standard flow cell bio-inert, 10 mm, 13 L, 120 bar ( 12 MPa) for MWD/DAD, includes 0890-1763 0.18 x 1500 mm PEEK capillary and 5063-6591 PEEK fittings (G5615-60022) (for Agilent 1260 Infinity II DAD G7115A, and MWD G7165A)

PEEK, ceramic, sapphire, PTFE

Bio-inert flow cell, 8 L, 20 bar (pH 112), includes 0890-1763 0.18 x 1500 mm PEEK capillary and 5063-6591 PEEK fittings (G5615-60005) (for Agilent 1260 Infinity II FLD G7121A/B)

PEEK, fused silica, PTFE

Agilent InfinityLab LC Series DAD WR and MWD User Manual 22

1 Introduction Bio-inert Materials

Bio-inert Heat Exchangers, Valves and Capillaries:

Quick Connect Heat Exchanger Bio-inert (G7116-60041) (for Agilent 1260 Infinity II Multicolumn Thermostat G7116A)

PEEK (steel-cladded)

Bio-inert Valve heads (G4235A, G5631A, G5632A, G5639A) PEEK, ceramic (Al2O3 based)

Bio-inert Connection capillaries Upstream of sample introduction: Titanium

Downstream of sample introduction: Agilent uses stainless-steel-cladded

PEEK capillaries, which keep the flow path free of steel and provide pressure stability up to 600 bar.

Table 1 Used Bio-inert materials

Module Materials

NOTE To ensure optimum bio-compatibility of your Agilent 1260 Infinity II Bio-inert LC system, do not include non-inert standard modules or parts to the flow path. Do not use any parts that are not labeled as Agilent Bio-inert. For solvent compatibility of these materials, see General Information about Solvent/Material Compatibility on page 61.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 23

2 Site Requirements and Specifications

Site Requirements 25

Physical Specifications 28

Performance Specifications 29 Specifications 29 Specification Conditions 35

This chapter provides information on environmental requirements, physical and performance specifications.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 24

2 Site Requirements and Specifications Site Requirements

Site Requirements

A suitable environment is important to ensure optimal performance of the instrument.

Power Considerations The module power supply has wide ranging capability. It accepts any line voltage in the range described in Table 2 on page 28. Consequently there is no voltage selector in the rear of the module. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.

WARNING Hazard of electrical shock or damage of your instrumentation

can result, if the devices are connected to a line voltage higher than specified.

Connect your instrument to the specified line voltage only.

WARNING Electrical shock hazard

The module is partially energized when switched off, as long as the power cord is plugged in.

The cover protects users from personal injuries, for example electrical shock.

Do not open the cover.

Do not operate the instrument and disconnect the power cable in case the cover has any signs of damage.

Contact Agilent for support and request an instrument repair service.

WARNING Inaccessible power plug.

In case of emergency it must be possible to disconnect the instrument from the power line at any time.

Make sure the power connector of the instrument can be easily reached and unplugged.

Provide sufficient space behind the power socket of the instrument to unplug the cable.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 25

2 Site Requirements and Specifications Site Requirements

Power Cords Country-specific power cords are available for the module. The female end of all power cords is identical. It plugs into the power-input socket at the rear. The male end of each power cord is different and designed to match the wall socket of a particular country or region.

Agilent makes sure that your instrument is shipped with the power cord that is suitable for your particular country or region.

WARNING Unintended use of power cords

Using power cords for unintended purposes can lead to personal injury or damage of electronic equipment.

Never use a power cord other than the one that Agilent shipped with this instrument.

Never use the power cords that Agilent Technologies supplies with this instrument for any other equipment.

Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.

WARNING Absence of ground connection

The absence of ground connection can lead to electric shock or short circuit.

Never operate your instrumentation from a power outlet that has no ground connection.

WARNING Electrical shock hazard

Solvents may damage electrical cables.

Prevent electrical cables from getting in contact with solvents.

Exchange electrical cables after contact with solvents.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 26

2 Site Requirements and Specifications Site Requirements

Bench Space The module dimensions and weight (see Table 2 on page 28) allow you to place the module on almost any desk or laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.

If the bench shall carry a complete HPLC system, make sure that the bench is designed to bear the weight of all modules.

The module should be operated in a horizontal position.

Environment Your detector will work within the specifications at ambient temperatures and relative humidity described in Table 2 on page 28.

ASTM drift tests require a temperature change below 2 C/hour (3.6 F/hour) over one hour period. Our published drift specification (refer also to Performance Specifications G7115A on page 29 and Performance Specifications G7165A on page 32) is based on these conditions. Larger ambient temperature changes will result in larger drift.

Better drift performance depends on better control of the temperature fluctuations. To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 C/hour (1.8 F/hour). Turbulences around one minute or less can be ignored.

NOTE Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench rack. This option helps to save bench space as all modules can be placed into one single stack. It also allows to easily relocate the instrument to another lab.

NOTE The module is designed to operate in a typical electromagnetic environment (EN61326-1) where RF transmitters, such as mobile phones, should not be used in close proximity.

CAUTION Condensation within the module

Condensation can damage the system electronics.

Do not store, ship or use your module under conditions where temperature fluctuations could cause condensation within the module.

If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 27

2 Site Requirements and Specifications Physical Specifications

Physical Specifications

Table 2 Physical Specifications

Type Specification Comments

Weight 12 kg (26.5 lbs)

Dimensions (height width depth)

140 x 396 x 436 mm (5.5 x 15.6 x 17 inches)

Line voltage 100 240 V~, 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, 5 %

Power consumption 110 VA / 100 W

Ambient operating temperature

455 C (39131 F)

Ambient non-operating temperature

-40 70 C (-40 158 F)

Humidity < 95 % r.h. at 40 C (104 F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Safety standards: IEC, EN, CSA, UL

Overvoltage category II, Pollution degree 2 For indoor use only.

ISM Classification ISM Group 1 Class B According to CISPR 11

Agilent InfinityLab LC Series DAD WR and MWD User Manual 28

2 Site Requirements and Specifications Performance Specifications

Performance Specifications

Specifications

Performance Specifications G7115A

Table 3 Performance Specifications G7115A

Type Specification

Detection type 1024-element photodiode array

Light source Deuterium and tungsten lamps

Number of signals 8

Maximum data rate 120 Hz

Short term signal noise (ASTM)

< 0.710-5 AU at 254 and 750 nm

Drift < 0.910-3 AU/h at 254 nm and 750 nm

Linear absorbance range > 2 AU (5 %) at 265 nm

Wavelength range 190 950 nm

Wavelength accuracy 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter

Wavelength bunching 1 400 nm, programmable in 1 nm steps

Slit width 1, 2, 4 , 8, 16 nm

Diode width ~1 nm

Time programmable Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength range, threshold, spectra storage mode

Agilent InfinityLab LC Series DAD WR and MWD User Manual 29

2 Site Requirements and Specifications Performance Specifications

Flow cells Standard: 13 L volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum Standard bio-inert: 13 L volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum Semi-micro: 5 L volume, 6 mm cell path length and 120 bar (1740 psi) pressure maximum Micro: 2 L volume, 3 mm cell path length, 120 bar (1740 psi) pressure maximum Semi-nano: 500 nL volume, 10 mm cell path length and 40 bar (580 psi) pressure maximum Nano: 80 nL volume, 6 mm cell path length and 40 bar (580 psi) pressure maximum High pressure: 1.7 L volume, 6 mm cell path length and 400 bar (5800 psi) pressure maximum Prep SST: 3 mm cell path length and 120 bar (1740 psi) pressure maximum Prep Quartz: 3 mm cell path length and 50 bar (1740 psi) pressure maximum Prep Quartz:0.3 mm cell path length and 50 or 20 bar (290 psi) pressure maximum Prep Quartz: 0.06 mm cell path length and 50 or 20 bar (290 psi) pressure maximum SFC Flow Cell: 13 L volume, 10 mm cell path length and 400 bar (5800 psi) pressure maximum SFC Flow Cell LD: 2 L volume, 3 mm cell path length and 400 bar (5800 psi) pressure maximum

Spectral tools Data analysis software for spectra evaluation, including spectral libraries and peak purity functions

Analog output Recorder/integrator: 100 mV or 1 V, output range 0.001 2 AU, one output

Instrument control Lab Advisor B.02.08 or above LC and CE Drivers A.02.14 or above For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers.

Local control Agilent Instant Pilot (G4208A) B.02.20 or above

Communication LAN, Controller-Area Network (CAN), USB Extended Remote Interface (ERI): ready, start, stop and shut-down signals

Table 3 Performance Specifications G7115A

Type Specification

Agilent InfinityLab LC Series DAD WR and MWD User Manual 30

2 Site Requirements and Specifications Performance Specifications

GLP RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage) Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-setable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.

Safety and maintenance Extensive diagnostics, error detection and display through Agilent Instant Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.

Housing All materials recyclable.

Others Second generation of Electronic temperature control (ETC) for the complete optical unit

Table 3 Performance Specifications G7115A

Type Specification

Agilent InfinityLab LC Series DAD WR and MWD User Manual 31

2 Site Requirements and Specifications Performance Specifications

Performance Specifications G7165A

Table 4 Performance Specifications G7165A

Type Specification

Detection type 1024-element photodiode array

Light source Deuterium and tungsten lamps

Number of signals 8

Maximum data rate 120 Hz

Short term signal noise (ASTM)

< 0.710-5 AU at 254 nm and 750 nm

Drift < 0.910-3 AU/h at 254 nm

Linear absorbance range > 2 AU (5 %) at 265 nm

Wavelength range 190 950 nm

Wavelength accuracy 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter

Wavelength bunching 1 400 nm

Slit width 1, 2, 4 , 8, 16 nm

Diode width < 1 nm

Time programmable Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength range, threshold, spectra storage mode

Agilent InfinityLab LC Series DAD WR and MWD User Manual 32

2 Site Requirements and Specifications Performance Specifications

Flow cells Standard: 13 L volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum Standard bio-inert: 13 L volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum Semi-micro: 5 L volume, 6 mm cell path length and 120 bar (1740 psi) pressure maximum Micro: 2 L volume, 3 mm cell path length, 120 bar (1740 psi) pressure maximum Semi-nano: 500 nL volume, 10 mm cell path length and 40 bar (580 psi) pressure maximum Nano: 80 nL volume, 6 mm cell path length and 40 bar (580 psi) pressure maximum High pressure: 1.7 L volume, 6 mm cell path length and 400 bar (5800 psi) pressure maximum Prep SST: 3 mm cell path length and 120 bar (1740 psi) pressure maximum Prep Quartz:0.3 mm cell path length and 50 or 20 bar (290 psi) pressure maximum Prep Quartz: 0.06 mm cell path length and 50 or 20 bar (290 psi) pressure maximum SFC Flow Cell: 13 L volume, 10 mm cell path length and 400 bar (5800 psi) pressure maximum SFC Flow Cell LD: 2 L volume, 3 mm cell path length and 400 bar (5800 psi) pressure maximum

Analog output Recorder/integrator: 100 mV or 1 V, output range 0.001 2 AU, one output

Instrument control Lab Advisor B.02.08 or above LC and CE Drivers A.02.14 or above For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers.

Local control Agilent Instant Pilot (G4208A) B.02.20 or above

Communication LAN, Controller Area Network (CAN), USB Extended Remote Interface (ERI): ready, start, stop and shut-down signals

GLP RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage) Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-setable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.

Table 4 Performance Specifications G7165A

Type Specification

Agilent InfinityLab LC Series DAD WR and MWD User Manual 33

2 Site Requirements and Specifications Performance Specifications

Safety and maintenance Extensive diagnostics, error detection and display through Agilent Instant Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. Tracking of flow cells and lamps with RFID (radio frequency identification) tags.

Housing All materials recyclable.

Others Second generation of Electronic temperature control (ETC) for the complete optical unit

Table 4 Performance Specifications G7165A

Type Specification

Agilent InfinityLab LC Series DAD WR and MWD User Manual 34

2 Site Requirements and Specifications Performance Specifications

Specification Conditions ASTM: Standard Practice for Variable Wavelength Photometric Detectors Used in Liquid Chromatography.

Reference conditions: cell path length 10 mm, wavelength 254 and 750 nm with reference wavelength 360 nm/100 nm, slit width 4 nm, time constant 2 s (equal to response time 4 s), flow 1 mL/min LC-grade Methanol.

Linearity: Linearity is measured with caffeine at 265 nm/4 nm with slit width 4 nm and TC 2 s (or with RT 4 s) with 10 mm pathlength.

For environmental conditions refer to "Environment".

ASTM drift tests require a temperature change below 2 C/hour (3.6 F/hour) over one hour period. Our published drift specification is based on these conditions. Larger ambient temperature changes will result in larger drift. Better drift performance depends on better control of the temperature fluctuations. To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 C/hour (1.8 F/hour). Turbulences around one minute or less can be ignored.

Performance tests should be done with a completely warmed up optical unit (> two hours). ASTM measurements require that the detector should be turned on at least 24 h before start of testing.

Time Constant versus Response Time

According to ASTM E1657-98 Standard Practice of Testing Variable-Wavelength Photometric Detectors Used in Liquid Chromatography the time constant is converted to response time by multiplying by the factor 2.2.

NOTE The specifications are based on the standard RFID tag lamp (2140-0820) and may be not achieved when other lamp types or aged lamps are used.

NOTE Mobile devices used close to the intstrument could affect the detector's short term noise level.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 35

3 Using the Module

Magnets 37

Turn on/off 38

Status Indicators 40

Instrument Configuration 41

Set up the Detector with Agilent Open Lab ChemStation 42

The Detector User Interface 43

Detector Control Settings 45

Method Parameter Settings 46

Agilent Local Control Modules 50

This chapter explains the essential operational parameters of the module.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 36

3 Using the Module Magnets

Magnets

1 Magnets in doors of pumps, autosamplers, detectors, and fraction collectors.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 37

3 Using the Module Turn on/off

Turn on/off

This procedure exemplarily shows an arbitrary LC stack configuration.

1 2

Power switch: On

Agilent InfinityLab LC Series DAD WR and MWD User Manual 38

3 Using the Module Turn on/off

3 Turn instrument On/Off with the control software. 4

Power switch: Off

5

Agilent InfinityLab LC Series DAD WR and MWD User Manual 39

3 Using the Module Status Indicators

Status Indicators

This procedure exemplarily shows an arbitrary LC stack configuration.

1 The module status indicator indicates one of six possible module conditions:

Status indicators

1. Idle

2. Run mode

3. Not-ready. Waiting for a specific pre-run condition to be reached or completed.

4. Error mode - interrupts the analysis and requires attention (for example, a leak or defective internal components).

5. Resident mode (blinking) - for example, during update of main firmware.

6. Bootloader mode (fast blinking). Try to re-boot the module or try a cold-start. Then try a firmware update.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 40

3 Using the Module Instrument Configuration

Instrument Configuration

1 Set the switches of the Configuration switch at the rear of the module:

a All switches DOWN: module uses the default IP address 192.168.254.11.

b Switch 4 UP and others DOWN: module uses DHCP.

c Switch 5 UP and others DOWN: modules uses STORED address.

2 Enter the setup information (MAC / IP address and/or Instrument Name).

a Agilent OpenLab ChemStation (Configure Instrument):

b Lab Advisor (Instrument Overview - Add Instrument):

Agilent InfinityLab LC Series DAD WR and MWD User Manual 41

3 Using the Module Set up the Detector with Agilent Open Lab ChemStation

Set up the Detector with Agilent Open Lab ChemStation

The setup of the detector is shown with the Agilent OpenLab ChemStation C.01.07 and Driver A.02.14.

Figure 7 ChemStation Method and Run Control (just detector is shown)

After successful load of the OpenLab ChemStation, you should see the module as an active item in the graphical user interface (GUI).

NOTE This section describes the detector settings only. For information on the Agilent OpenLab ChemStation or other 1200 Infinity modules refer to the corresponding documentation.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 42

3 Using the Module The Detector User Interface

The Detector User Interface

Within the detector GUI, there are active areas. If you move the mouse cursor across the icons the cursor will change. 1 Lamp: turn on and off of UV-lamp 2 EMF status 3 Detector status 4 Detector balance status 5 Lamp status (on/off) 6 Lamp information (RFID tag) 7 Flow Cell information (RFID tag)

RFID tag information is displayed when moving with the mouse cursor on to the tag attached to the flow cell or lamp. The information provides flow cell and lamp related information like Part number Production date Serial number and other details.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 43

3 Using the Module The Detector User Interface

EMF Status shows Run / Ready / Error state and Not Ready text or Error text Offline (gray) Ok. No Maintenance required (green) EMF warning. Maintenance might be required (yellow) EMF warning. Maintenance required (red) Important: The EMF settings can be accessed via Agilent Lab Advisor. The limit(s) can be changed. Based on the limit, the User Interface displays the above status.

Module Status shows Run / Ready / Error state and Not Ready text or Error text Error (red) Not ready (yellow) Ready (green) Pre run, Post run (purple) Run (blue) Idle (green) Offline (dark gray) Standby (light gray)

A right-click into the Active Area will open a menu to Show the Control Interface (special module settings) Show the Method interface (similar as via menu

Instrument > Setup Instrument Method) Set Error Method Identify Module (Status LED will blink) Perform a Balance Switch the UV-lamp on/off (same as click on button

Make Device Ready/Turn device off (standby)) Switch the Vis-lamp on/off

Agilent InfinityLab LC Series DAD WR and MWD User Manual 44

3 Using the Module Detector Control Settings

Detector Control Settings

The figure shows the default settings. Lamps: can be turned ON/OFF. At Power On: automatic lamp-on at power on. Analog Output Range: can be set to either 100 mV or 1 V

full scale, for additional settings see Analog Output (under Method Parameter Settings on page 46).

Temperature Control: can be turned ON/OFF. The optical unit is kept on constant temperature and improves the baseline stability in unstable environments. See also note below. ON it will keep the optical unit stabilized.

UV Lamp Tag Use lamp only if lamp tag is available detects a lamp

with RFID tag. If no RFID tag lamp is used, UV lamp not ready is displayed and it cannot be ignited. A compatible mode has to be selected based on the used lamp; see Non-RFID-tag lamp information below.

Always use lamp: In case a non-RFID-tag lamp is used, the user interface will show this when selecting a compatible mode. You may operate the detector outside of the guaranteed specification. The correct selection is important for optimal performance and lifetime.

Cell Tag Use cell only if cell tag is available detects a cell with

RFID tag. If no RFID tag cell is used, cell tag not ready is displayed and it cannot be ignited and analysis is disabled.

Always use cell: In case a non-RFID-tag cell is used, the user interface will show this when selecting a compatible mode.

Automatic Turn On: automatic detector power on.

NOTE If the flow cell temperature is critical for your chromatography, you may set the Temperature Control to off. This will lower the optical unit and flow cell temperature by some degree C.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 45

3 Using the Module Method Parameter Settings

Method Parameter Settings

These settings are available via Menu > Instrument > Set up Instrument Method or via right click into the modules active area (does not show the Instrument Curves tab).

Figure 8 Method parameter settings

NOTE For additional help and support: Highlight the desired cell and press F1. A help screen will open with additional information and documentation about the topic.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 46

3 Using the Module Method Parameter Settings

Table 5 Method Parameter Settings

Signals Up to 8 individual signals can be set. For each of the signals, the wavelength and bandwidth can be set for sample and reference.

Limits: Wavelength: 190.0 to 950.0 nm in steps of 0.1 nm Bandwidth: 1.0 to 400.0 nm in steps of 0.1 nm Setting an appropriate reference wavelength could improve the baseline behavior.

Peakwidth (Responsetime, Data Rate) Peakwidth enables you to select the peak width (response time) for your analysis. The peak width is defined as the width of a peak, in minutes, at half the peak height. Set the peak width to the narrowest expected peak in your chromatogram. The peak width sets the optimum response time for your detector. The peak detector ignores any peaks that are considerably narrower, or wider, than the peak width setting. The response time is the time between 10 % and 90 % of the output signal in response to an input step function. When the All spectrum storage option is selected, then spectra are acquired continuously depending on the setting of the peak width. The time specified by the peak width is used as a factor in the acquisition of spectra. The acquisition time for one spectrum is slightly less than the peak width divided by 8, which is the acquisition time. Limits: When you set the peak width (in minutes), the corresponding response time is set automatically and the appropriate data rate for signal and spectra acquisition is selected. Do not use peak width shorter than necessary. For details see Peak width (response time) on page 73.

Stoptime/Posttime The stoptime is the time where either the complete system stops (As Pump/Injector) or the module (if different from system stop time). The data collection is stopped at this time. A posttime period can be used to allow modules items to equilibrate (e.g. after gradient change or temperature change).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 47

3 Using the Module Method Parameter Settings

Spectrum Settings (DAD only) Store: None / All

ALL spectra are taken continuously depending on the setting of the Peakwidth. Eight spectra are acquired per Peakwidth. The acquisition time for one spectrum is slightly less than the Peakwidth divided by 8, that is, greater than or equal to 0.01 s and less than or equal to 2.55 s.

Range defines the wavelength range for spectral storage.

Limits: 190 950 nm in steps of 1 nm for both low and high values. The high value must be greater than the low value by at least 2 nm.

Step defines the wavelength resolution for spectral storage.

Limits: 0.10 100.00 nm in steps of 0.1 nm.

Analog Output The range can be set to either 100 mV or 1 V full scale, see Detector Control Settings on page 45. Zero Offset: 1 99 % in steps of 1 % (5 % equal to 50 mV). Attenuation: 0.98 2000 mAU at discrete values for either

100 mV or 1 V full scale.

Margin for negative Absorbance Use this field to modify the detectors signal handling to increase the margin for negative absorbance. Use this option if, for example, your solvent gradient produces a decreasing baseline absorbance, and for GPC analyses. Limits: 100 to 4000 mAU. The higher the value the greater the baseline noise. Set this value only if you expect negative absorbance greater than -100 mAU. Slitwidth: You can select the optical bandwidth (1, 2, 4, 8 or 16 nm) of the detector; the narrower the slit, the smaller the optical bandwidth of the instrument, but the lower its sensitivity. The smaller the optical bandwidth the higher the spectral resolution.

Autobalance Defines, whether a balance is performed prior to a run and/or after a run has finished. Lamp on required for acquisition: If unchecked, the lamp will be turned off after the analysis has finished. Note that the lamp on requires at least one hour warm-up time, see Warm up of the Detector on page 89.

Table 5 Method Parameter Settings

Agilent InfinityLab LC Series DAD WR and MWD User Manual 48

3 Using the Module Method Parameter Settings

Timetable You may set up time events to change functions with their parameters over the run time. Add lines as required. Time Limits: 0.00 to 99999.00 min in steps of 0.01 min. Via the buttons in the bottom area, time table lines can be added, removed, cut copied, pasted or completely cleared. Based on the chosen function, a certain parameter can be selected.

Instrument Curves The detector has several signals (internal temperatures, voltages of lamps) that can be used for diagnosing problems. These can be baseline problems deriving from deuterium lamps wander / drift problems due to temperature changes. These signals can be used in addition to the normal baseline signal to determine whether correlation to temperature or voltage/current of the lamp. These signals are available via the Agilent ChemStation Online Plot/Data Signal and/or Agilent Lab Advisor Software.

Table 5 Method Parameter Settings

Agilent InfinityLab LC Series DAD WR and MWD User Manual 49

3 Using the Module Agilent Local Control Modules

Agilent Local Control Modules

Agilent InfinityLab Companion G7108AA

The Agilent InfinityLab Companion gives you complete control, system monitoring, signal plotting, and diagnostic capabilities for a wide range of LC system modules.

The instrument control solution is available as full package including all hardware and accessories, but can also be used on your own mobile devices like tablets, mobile phones and other electronic equipment.

Combining the conveniences of the Agilent Instant Pilot features with state-of-the-art mobile technology, the Agilent InfinityLab Companion gives you maximum flexibility and ease of use to control and monitor your LC system modules.

Features:

Complete local control and monitoring of Agilent Infinity II LC modules

Excellent usability and ease of use through a user interface specifically tailored for mobile devices - simple, intuitive touch-enabled, and visual controllable.

High flexibility through a modern Bring your own device approach. Connection between LC module and mobile device either wireless via Wi-Fi or wired over USB cable (with full package).

Convenient, ergonomic operation either handheld or attached to a module at the stack with newly developed, secure tablet holder (included in the full package).

Preconfigured tablet with all required software already installed (included in the full package).

Centerpiece of the solution is a USB dongle that activates the complete intelligence of the InfinityLab Companion on the instrument stack.

The InfinityLab Companion provides:

fast and direct control in front of the instrument

a clear overview of the system status

control functionalities

access to method parameters and sequences

a logbook showing events from the modules

diagnostic tests

Agilent InfinityLab LC Series DAD WR and MWD User Manual 50

4 Preparing the Detector

Leak and Waste Handling 52 Waste Concept 53

Setting up an Analysis 54 Before Using the System 55 Requirements and Conditions 56 Preparing the Detector 58 Preparing the HPLC System 58 Running the Sample and Verifying the Results 60

Solvent Information 61

This chapter provides information on how to set up the module for an analysis and explains the basic settings.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 51

4 Preparing the Detector Leak and Waste Handling

Leak and Waste Handling

For details on correct installation, see separate installation documentation.

WARNING Toxic, flammable and hazardous solvents, samples and reagents

The handling of solvents, samples and reagents can hold health and safety risks.

When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.

Do not use solvents with an auto-ignition temperature below 200 C (392 F). Do not use solvents with a boiling point below 56 C (133 F).

Avoid high vapor concentrations. Keep the solvent temperature at least 40 C (72 F) below the boiling point of the solvent used. This includes the solvent temperature in the sample compartment. For the solvents methanol and ethanol keep the solvent temperature at least 25 C (45 F) below the boiling point.

Do not operate the instrument in an explosive atmosphere. Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for

example, carbon disulfide). Reduce the volume of substances to the minimum required for the

analysis. Never exceed the maximum permissible volume of solvents (8 L) in the

solvent cabinet. Do not use bottles that exceed the maximum permissible volume as specified in the usage guideline for solvent cabinet.

Ground the waste container. Regularly check the filling level of the waste container. The residual free

volume in the waste container must be large enough to collect the waste liquid.

To achieve maximal safety, regularly check the tubing for correct installation.

NOTE For details, see the usage guideline for the solvent cabinet. A printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available in the Agilent Information Center or via the Internet.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 52

4 Preparing the Detector Leak and Waste Handling

Waste Concept 1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with

4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 53

4 Preparing the Detector Setting up an Analysis

Setting up an Analysis

This chapter may be used to

Prepare the system,

Get to know the set up of an HPLC analysis and

Use it as an instrument check to demonstrate that all modules of the system are correctly installed and connected. It is not a test of the instrument performance.

Learn about special settings

Agilent InfinityLab LC Series DAD WR and MWD User Manual 54

4 Preparing the Detector Setting up an Analysis

Before Using the System

Solvent Information

Observe recommendations on the use of solvents in chapter Solvents in the pumps reference manual.

Priming and Purging the System

When the solvents have been exchanged or the pumping system has been turned off for a certain time (for example, overnight) oxygen will re-diffuse into the solvent channel between the solvent reservoir, vacuum degasser (when available in the system) and the pump. Solvents containing volatile ingredients will slightly lose these. Therefore priming of the pumping system is required before starting an application.

1 Open the purge valve of your pump (by turning it counterclockwise) and set flow rate to 3 5 mL/min.

2 Flush all tubes with at least 30 mL of solvent. 3 Set flow to required value of your application and close the purge valve.

Table 6 Choice of Priming Solvents for Different Purposes

Activity Solvent Comments

After an installation

When switching between reverse phase and normal phase (both times)

Isopropanol

Isopropanol

Best solvent to flush air out of the system

Best solvent to flush air out of the system

After an installation Ethanol or Methanol Alternative to Isopropanol (second choice) if no Isopropanol is available

To clean the system when using buffers

After a solvent change

Bidistilled water

Bidistilled water

Best solvent to re-dissolve buffer crystals

Best solvent to re-dissolve buffer crystals

After the installation of normal phase seals (P/N 0905-1420)

Hexane + 5% Isopropanol Good wetting properties

NOTE The pump should never be used for priming empty tubings (never let the pump run dry). Use a syringe to draw enough solvent for completely filling the tubings to the pump inlet before continuing to prime with the pump.

NOTE Pump for approximately 10minutes before starting your application.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 55

4 Preparing the Detector Setting up an Analysis

Requirements and Conditions

What You Will Need

The table below lists the items you need to have for the set up of the analysis. Some of these are optional (not required for the basic system).

Conditions

A single injection of the isocratic test standard is made under the conditions given in Table 8 on page 56:

Table 7 What you will need

Agilent 1200 Infinity Series system Pump (plus degassing)

Autosampler

Detector, standard flow cell installed

Degasser (optional)

Column Compartment (optional)

Agilent CDS

System should be correctly set up for LAN communication with the Agilent ChemStation

Column: Zorbax Eclipse XDB-C18, 4.6 x 150 mm, 5 m (993967-902) or an equivalent column

Standard: Agilent isocratic checkout sample (01080-68704)

Table 8 Conditions

Flow 1.5 mL/min

Stoptime 8 min

Solvent 100% (30% water/70% Acetonitrile)

Temperature Ambient

Wavelength sample 254 nm

Injection Volume 1 L

Column Temperature (optional): 25 C or ambient

Agilent InfinityLab LC Series DAD WR and MWD User Manual 56

4 Preparing the Detector Setting up an Analysis

Typical Chromatogram

A typical chromatogram for this analysis is shown in Figure 9 on page 57. The exact profile of the chromatogram will depend on the chromatographic conditions. Variations in solvent quality, column packing, standard concentration and column temperature will all have a potential effect on peak retention and response.

Figure 9 Typical Chromatogram with UV-detector

Agilent InfinityLab LC Series DAD WR and MWD User Manual 57

4 Preparing the Detector Setting up an Analysis

Preparing the Detector For best performance of the detector

Let the lamp warm-up and stabilize for at least one hour (initial turn on of the module requires a longer time depending on the environment and the application needs); refer to Specification Conditions on page 35.

For high sensitivity measurements, a stable environment is required; refer to Environment on page 27. Prevent drafts from air condition systems.

Setting an appropriate reference wavelength could improve the baseline behavior.

Do not work with removed/open front panels/doors. When the system includes a G1316 TCC (typically located below the detector) and its front panel is removed while the TCC is set to high temperatures, the up-streaming air could influence the stability of the detector baseline.

Preparing the HPLC System 1 Turn on the control software and the monitor.

2 Turn on the modules.

3 Start the control software. The screen should show all modules and the system status is Not Ready.

4 Turn on the modules that require conditioning:

a Detector lamp (warm-up for at least 60 min to get a stable baseline).

b Column compartment (set temperature as required).

c Pump (purge).

d Sampler (prepare the standard isocratic sample into a vial).

e Solvents (fill water and Acetontrile into the solvent bottles).

5 Load the default method.

6 Pump the water/acetonitrile (30/70 %) mobile phase through the column for 10 min for equilibration.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 58

4 Preparing the Detector Setting up an Analysis

7 Select the menu item Run Control > Sample Info and enter information about this application. Click OK to leave this screen.

Figure 10 Sample Info

Agilent InfinityLab LC Series DAD WR and MWD User Manual 59

4 Preparing the Detector Setting up an Analysis

Running the Sample and Verifying the Results 1 To start a run select the menu item RunControl > Run Method.

2 This will start the modules and the online plot on the Agilent ChemStation will show the resulting chromatogram.

Figure 11 Chromatogram with Isocratic Test Sample

NOTE Information about using the Data Analysis functions can be obtained from the Using your ChemStation manual supplied with your system.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 60

4 Preparing the Detector Solvent Information

Solvent Information

Observe the following recommendations on the use of solvents.

Follow the recommendations for avoiding the growth of algae, see the pump manuals.

Small particles can permanently block capillaries and valves. Therefore, always filter solvents through 0.22 m filters.

Avoid or minimize the use of solvents that may corrode parts in the flow path. Consider specifications for the pH range given for different materials such as flow cells, valve materials etc. and recommendations in subsequent sections.

General Information about Solvent/Material Compatibility Materials in the flow path are carefully selected based on Agilents experiences in developing highest-quality instruments for HPLC analysis over several decades. These materials exhibit excellent robustness under typical HPLC conditions. For any special condition, please consult the material information section or contact Agilent.

Disclaimer

Subsequent data was collected from external resources and is meant as a reference. Agilent cannot guarantee the correctness and completeness of such information. Data is based on compatibility libraries, which are not specific for estimating the long-term life time under specific but highly variable conditions of UHPLC systems, solvents, solvent mixtures and samples. Information can also not be generalized due to catalytic effects of impurities like metal ions, complexing agents, oxygen etc. Apart from pure chemical corrosion, other effects like electro corrosion, electrostatic charging (especially for non-conductive organic solvents), swelling of polymer parts etc. need to be considered. Most data available refers to room temperature (typically 20 25 C, 68 77 F). If corrosion is possible, it usually accelerates at higher temperatures. If in doubt, please consult technical literature on chemical compatibility of materials.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 61

4 Preparing the Detector Solvent Information

MP35N

MP35N is a nonmagnetic, nickel-cobalt-chromium-molybdenum alloy demonstrating excellent corrosion resistance (for example, against nitric and sulfuric acids, sodium hydroxide, and seawater) over a wide range of concentrations and temperatures. In addition, this alloy shows exceptional resistance to high-temperature oxidation. Due to excellent chemical resistance and toughness, the alloy is used in diverse applications: dental products, medical devices, nonmagnetic electrical components, chemical and food processing equipment, marine equipment. Treatment of MP35N alloy samples with 10 % NaCl in HCl (pH 2.0) does not reveal any detectable corrosion. MP35N also demonstrates excellent corrosion resistance in a humid environment. Although the influence of a broad variety of solvents and conditions has been tested, users should keep in mind that multiple factors can affect corrosion rates, such as temperature, concentration, pH, impurities, stress, surface finish, and dissimilar metal contacts.

Polyphenylene Sulfide (PPS)

Polyphenylene sulfide has outstanding stability even at elevated temperatures. It is resistant to dilute solutions of most inorganic acids, but it can be attacked by some organic compounds and oxidizing reagents. Nonoxidizing inorganic acids, such as sulfuric acid and phosphoric acid, have little effect on polyphenylene sulfide, but at high concentrations and temperatures, they can still cause material damage. Nonoxidizing organic chemicals generally have little effect on polyphenylene sulfide stability, but amines, aromatic compounds, and halogenated compounds may cause some swelling and softening over extended periods of time at elevated temperatures. Strong oxidizing acids, such as nitric acid (> 0.1 %), hydrogen halides (> 0.1 %), peroxy acids (> 1 %), or chlorosulfuric acid degrade polyphenylene sulfide. It is not recommended to use polyphenylene sulfide with oxidizing material, such as sodium hypochlorite and hydrogen peroxide. However, under mild environmental conditions, at low concentrations and for short exposure times, polyphenylene sulfide can withstand these chemicals, for example, as ingredients of common disinfectant solutions.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 62

4 Preparing the Detector Solvent Information

PEEK

PEEK (Polyether-Ether Ketones) combines excellent properties regarding biocompatibility, chemical resistance, mechanical and thermal stability. PEEK is therefore the material of choice for UHPLC and biochemical instrumentation.

It is stable in the specified pH range (for the Bio-Inert LC system: pH 1 13, see bio-inert module manuals for details), and inert to many common solvents.

There is still a number of known incompatibilities with chemicals such as chloroform, methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %, sulfuric acid > 10 %, sulfonic acids, trichloroacetic acid), halogens or aqueous halogen solutions, phenol and derivatives (cresols, salicylic acid, and so on).

When used above room temperature, PEEK is sensitive to bases and various organic solvents, which can cause it to swell. Under such conditions, normal PEEK capillaries are sensitive to high pressure. Therefore, Agilent uses stainless steel cladded PEEK capillaries in bio-inert systems. The use of stainless steel cladded PEEK capillaries keeps the flow path free of steel and ensures pressure stability up to 600 bar. If in doubt, consult the available literature about the chemical compatibility of PEEK.

Polyimide

Agilent uses semi-crystalline polyimide for rotor seals in valves and needle seats in autosamplers. One supplier of polyimide is DuPont, which brands polyimide as Vespel, which is also used by Agilent.

Polyimide is stable in a pH range between 1 and 10 and in most organic solvents. It is incompatible with concentrated mineral acids (e.g. sulphuric acid), glacial acetic acid, DMSO and THF. It is also degraded by nucleophilic substances like ammonia (e.g. ammonium salts in basic conditions) or acetates.

Polyethylene (PE)

Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.

Polyethylene has a good stability for most common inorganic solvents including acids and bases in a pH range of 1 to 12.5. It is compatible with many organic solvents used in chromatographic systems like methanol, acetonitrile and isopropanol. It has limited stability with aliphatic, aromatic and halogenated hydrocarbons, THF, phenol and derivatives, concentrated acids and bases. For normal phase applications, the maximum pressure should be limited to 200 bar.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 63

4 Preparing the Detector Solvent Information

Tantalum (Ta)

Tantalum is inert to most common HPLC solvents and almost all acids except fluoric acid and acids with free sulfur trioxide. It can be corroded by strong bases (e.g. hydroxide solutions > 10 %, diethylamine). It is not recommended for the use with fluoric acid and fluorides.

Stainless Steel (SST)

Stainless steel is inert against many common solvents. It is stable in the presence of acids and bases in a pH range of 1 to 12.5. It can be corroded by acids below pH 2.3. It can also corrode in following solvents:

Solutions of alkali halides, their respective acids (for example, lithium iodide, potassium chloride, and so on) and aqueous solutions of halogens.

High concentrations of inorganic acids like nitric acid, sulfuric acid and organic solvents especially at higher temperatures (replace, if your chromatography method allows, by phosphoric acid or phosphate buffer which are less corrosive against stainless steel).

Halogenated solvents or mixtures which form radicals and/or acids, for example:

2 CHCl3 + O2 2 COCl2 + 2 HCl

This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol.

Chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, diisopropylether). Such ethers should be filtered through dry aluminium oxide which adsorbs the peroxides.

Solutions of organic acids (acetic acid, formic acid, and so on) in organic solvents. For example, a 1 % solution of acetic acid in methanol will attack steel.

Solutions containing strong complexing agents (for example, EDTA, ethylene diamine tetra-acetic acid).

Mixtures of carbon tetrachloride with isopropanol or THF.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 64

4 Preparing the Detector Solvent Information

Titanium (Ti)

Titanium is highly resistant to oxidizing acids (for example, nitric, perchloric and hypochlorous acid) over a wide range of concentrations and temperatures. This is due to a thin oxide layer on the surface, which is stabilized by oxidizing compounds. Non-oxidizing acids (for example, hydrochloric, sulfuric and phosphoric acid) can cause slight corrosion, which increases with acid concentration and temperature. For example, the corrosion rate with 3 % HCl (about pH 0.1) at room temperature is about 13 m/year. At room temperature, titanium is resistant to concentrations of about 5 % sulfuric acid (about pH 0.3). Addition of nitric acid to hydrochloric or sulfuric acids significantly reduces corrosion rates. Titanium is sensitive to acidic metal chlorides like FeCl3 or CuCl2. Titanium is subject to corrosion in anhydrous methanol, which can be avoided by adding a small amount of water (about 3 %). Slight corrosion is possible with ammonia > 10 %.

Diamond-Like Carbon (DLC)

Diamond-Like Carbon is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.

Fused silica and Quartz (SiO2)

Fused silica is used in Max Light Cartridges. Quartz is used for classical flow cell windows. It is inert against all common solvents and acids except hydrofluoric acid and acidic solvents containing fluorides. It is corroded by strong bases and should not be used above pH 12 at room temperature. The corrosion of flow cell windows can negatively affect measurement results. For a pH greater than 12, the use of flow cells with sapphire windows is recommended.

Gold

Gold is inert to all common HPLC solvents, acids and bases within the specified pH range. It can be corroded by complexing cyanides and concentrated acids like aqua regia.

Zirconium Oxide (ZrO2)

Zirconium Oxide is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 65

4 Preparing the Detector Solvent Information

Platinum/Iridium

Platinum/Iridium is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.

Fluorinated polymers (PTFE, PFA, FEP, FFKM, PVDF)

Fluorinated polymers like PTFE (polytetrafluorethylene), PFA (perfluoroalkoxy), and FEP (fluorinated ethylene propylene) are inert to almost all common acids, bases, and solvents. FFKM is perfluorinated rubber, which is also resistant to most chemicals. As an elastomer, it may swell in some organic solvents like halogenated hydrocarbons.

TFE/PDD copolymer tubings, which are used in all Agilent degassers except G1322A/G7122A, are not compatible with fluorinated solvents like Freon, Fluorinert, or Vertrel. They have limited life time in the presence of hexafluoroisopropanol (HFIP). To ensure the longest possible life with HFIP, it is best to dedicate a particular chamber to this solvent, not to switch solvents, and not to let dry out the chamber. For optimizing the life of the pressure sensor, do not leave HFIP in the chamber when the unit is off.

The tubing of the leak sensor is made of PVDF (polyvinylidene fluoride), which is incompatible with the solvent DMF (dimethyl formamide).

Sapphire, Ruby and Al2O3-based ceramics

Sapphire, ruby and ceramics based on aluminum oxide Al2O3 are inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 66

5 Optimizing the Detector

Optimizing the Detector Performance 68

Optimization Overview 69

Optimizing for Sensitivity, Selectivity, Linearity and Dispersion 71 Flow Cell Path Length 71 Peak width (response time) 73 Sample and Reference Wavelength and Bandwidth 74 Slit Width 78 Optimizing Spectral Acquisition (DAD only) 79 Margin for Negative Absorbance 80

Optimizing Selectivity 81 Quantifying Coeluting Peaks by Peak Suppression 81

Optimizing the Detector Regarding to the System 83 Delay Volume and Extra-Column Volume 83 How to Configure the Optimum Delay Volume 83 How to Achieve Higher Sensitivity 84

Warm up of the Detector 89

This chapter provides information on how to optimize the detector.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 67

5 Optimizing the Detector Optimizing the Detector Performance

Optimizing the Detector Performance

The detector has a variety of parameters that can be used to optimize performance. Depending on whether signal or spectral data need to be optimized, different settings are recommended. The following sections describe optimization for:

signal sensitivity, selectivity and linearity,

spectral sensitivity and resolution (DAD only), and

disk space required for storing data.

How to Get the Best Detector Performance

The information below will guide you on how to get the best detector performance. Follow these rules as a start for new applications. It gives rules-of-thumb for optimizing detector parameters.

NOTE The information in this chapter should be seen as a basic introduction to diode array detector techniques. Some of these techniques may not be available in the instrument software controlling the detector.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 68

5 Optimizing the Detector Optimization Overview

Optimization Overview

Table 9 Optimization Overview

Parameter Impact

1 Selection of flow cell peak resolution versus sensitivity

Choose flow cell according to used column, see Figure 12 on page 70.

2 Connection of flow cell chromatographic resolution

For flow rates from 0.5 mL/min connect column using the zero-dead-volume fittings of the detector.

For small column i.d. (e.g 1 mm) the inlet capillary of the micro flow cell can be connected directly to the column.

3 Setting the peak width (response time) peak resolution versus sensitivity versus disk space

Use peak width according Figure 12 on page 70 as starting point. Set the peak-width close to the width of a narrow peak of interest in your

chromatogram.

4 Setting wavelength and bandwidth

Sample wavelength: Never miss a peak by the use of a browser wavelength like 250 nm with

100 nm bandwidth. Select specific wavelength with reduced bandwidth if you need

selectivity, e.g. 250,10 nm and 360,100 nm as reference wavelength. Set the sample wavelength to a peak or valley in the spectrum to get

best linearity for high concentrations.

sensitivity versus selectivity

sensitivity versus linearity

Reference wavelength: Select the reference wavelength with broad bandwidth (30...100 nm)

wavelength range where your analytes have little or no absorbance (e.g. sample at 254 nm, reference at 320 nm).

baseline drift due to RI effects.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 69

5 Optimizing the Detector Optimization Overview

Figure 12 Choosing a Flow Cell in HPLC

5 Setting the slit width

Use 4 nm slit for normal applications. Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance bands

and for high concentrations. Use a wide slit (e.g. 16 nm) to detect very low concentrations. Optimizing spectral acquisition (DAD only) Select spectra acquisition mode according to your needs. Set the spectral wavelength range (for colorless samples 190...400 nm is

sufficient). Set step to 4 nm for normal use; set small step (and slit width) if high

resolution of spectra with fine structure is wanted.

spectral resolution, sensitivity and linearity.

Table 9 Optimization Overview

Parameter Impact

10 cm

20 cm

0.025 min

0.05 min

0.1 min

0.2 min

Internal column diameter 0.5 ... 1 mm 2.1 mm 3.0 mm 4.6 mm

High pressure flow cell for pressures above 100bar

Typical column length

Typical peak width

Recommended flow cell

T <= 5 cm

>= 40 cm

Typical flow rate

0.01 ... 0.2 ml/min

0.2 ... 0.4 ml/min

0.4 ... 0.4 ml/min

1 ... 5 ml/min

0.01 ... 5 ml/min

Standard flow cell

Semi-micro flow cell

Micro or Semi-nano

Agilent InfinityLab LC Series DAD WR and MWD User Manual 70

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Flow Cell Path Length Lambert-Beers law shows a linear relationship between the flow cell path length and absorbance.

where

T is the transmission, defined as the quotient of the intensity of the transmitted light I divided by the intensity of the incident light, I0,

is the extinction coefficient, which is a characteristic of a given substance under a precisely-defined set of conditions of wavelength, solvent, temperature and other parameters,

C [mol/L] is the concentration of the absorbing species, and

d [m] is the path length of the cell used for the measurement.

Therefore, flow cells with longer path lengths yield higher signals. Although noise usually increases little with increasing path length, there is a gain in signal-to-noise ratio. For example, in Figure 13 on page 72 the noise increased by less than 10 % but a 70 % increase in signal intensity was achieved by increasing the path length from 6 10 mm.

When increasing the path length, the cell volume usually increases in our example from 5 13 L. Typically, this causes more peak dispersion. As Figure 13 on page 72 demonstrates, this did not affect the resolution in the gradient separation in our example.

As a rule-of-thumb the flow cell volume should be about 1/3 of the peak volume at half height. To determine the volume of your peaks, take the peak width as reported in the integration results multiply it by the flow rate and divide it by 3).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 71

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Figure 13 Influence of Cell Path Length on Signal Height

Traditionally LC analysis with UV detectors is based on comparing measurements with internal or external standards. To check photometric accuracy of the detector it is necessary to have more precise information on path lengths of the flow cells.

The correct response is:

expected response * correction factor

Please find below the details of the flow cells:

Ab so

rb an

ce

Time (min)

Analysis of pesticide standard

10-mm optical path length

6-mm optical path length

Table 10 Correction factors for flow cells

Flow cell Path length (actual) Correction factor

Standard flow cell, 10 mm, 13 L, 120 bar ( 12 MPa) (G1315-60022)

9.800.07 mm 10/9.8

Semi-micro flow cell, 6 mm, 5 L, 120 bar ( 12 MPa) (G1315-60025)

5.800.07 mm 6/5.8

Micro flow cell, 3 mm, 2 L, 120 bar ( 12 MPa) (G1315-60024) 3.00+0.05 mm/-0.07 mm 3/3

Semi-nano flow cell kit, 10 mm, 500 nL, 5 MPa (G1315-68724) 10.000.02 mm 10/10

Nano flow cell kit, 6 mm, 80 nL, 5 MPa ( G1315-68716) 6.000.02 mm 6/6

Standard flow cell bio-inert, 10 mm, 13 L, 120 bar ( 12 MPa) for MWD/DAD, includes 0890-1763 0.18 x 1500 mm PEEK capillary and 5063-6591 PEEK fittings (G5615-60022)

9.800.07 mm 10/9.8

Agilent InfinityLab LC Series DAD WR and MWD User Manual 72

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Peak width (response time) Response time describes how fast the detector signal follows a sudden change of absorbance in the flow cell. The detector uses digital filters to adapt response time to the width of the peaks in your chromatogram. These filters do not affect peak area nor peak symmetry. When set correctly, such filters reduce baseline noise significantly ( Figure 14 on page 73), but reduce peak height only slightly. In addition, these filters reduce the data rate to allow optimum integration and display of your peaks and to minimize disk space required to store chromatograms and spectra.

Figure 14 Influence of Response Time on Signal and Noise

Table 11 on page 74 list the filter choices of the detector. To get optimum results, set peak width as close as possible to a narrow peak of interest in your chromatogram. Response time will the be approximately 1/3 of the peak width, resulting in less than 5 % peak-height reduction and less than 5 % additional peak dispersion. Decreasing the peak width setting in the detector will result in less than 5 % gain in peak height but baseline noise will increase by a factor of 1.4 for a factor of 2 response-time reduction. Increasing peak width (response time) by factor of two from the recommended setting (over-filtering) will reduce peak height by about 20 % and reduce baseline noise by a factor of 1.4. This gives you the best possible signal-to-noise ratio, but may affect peak resolution.

Unfiltered

Response time 0.05 min

Response time 0.1 min

Agilent InfinityLab LC Series DAD WR and MWD User Manual 73

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Sample and Reference Wavelength and Bandwidth The detector measures absorbance simultaneously at wavelengths from 190 to 950 nm. Two lamps provide good sensitivity over the whole wavelength range. The deuterium discharge lamp provides the energy for the UV range (190 to 400 nm) and the tungsten lamp emits light from 400 to 950 nm for the visible and short wave near infrared.

If you know little about the analytes in your sample, use both lamps and store all spectra over the full wavelength range. This provides full information but fills up your disk space rather quickly. Spectra can be used to check a peaks purity and identity. Spectral information is also useful to optimize wavelength settings for your chromatographic signal.

Table 11 Peak Width Response Time Data Rate (G7115A/G7165A)

Peak width at half

height [min]1

1 Values in the User Interface may be rounded.

Response [s] Scan data rate[Hz] 251 pts/scan

Scan data rate[Hz] 501 pts/scan

Scan data rate[Hz] > 501 pts/scan

< 0.0015625 0.015625 120 120 40 20

> 0.0015625 0.03125 120 120 40 20

> 0.003125 0.0625 80 80 40 20

> 0.00625 0.125 40 40 40 20

> 0.0125 0.25 20 20 20 20

> 0.025 0.5 10 10 10 10

> 0.05 1 5 5 5 5

> 0.1 2 2.5 2.5 2.5 2.5

> 0.2 4 1.25 1.25 1.25 1.25

> 0.4 8 0.625 0.625 0.625 0.625

> 0.85 16 0.3125 0.3125 0.3125 0.3125

Agilent InfinityLab LC Series DAD WR and MWD User Manual 74

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

The detector can compute and store at run time up to 8 signals with these properties:

sample wavelength, the center of a wavelength band with the width of sample bandwidth (BW), and optionally

reference wavelength, the center of a wavelength band with the width of reference bandwidth.

The signals comprises a series of data points over time, with the average absorbance in the sample wavelength band minus the average absorbance of the reference wavelength band.

Signal A in the detector default method is set to sample 250,100, reference 360,100, that is, the average absorbance from 200 300 nm minus the average absorbance from 300 400 nm. As all analytes show higher absorbance at 200 300 nm than at 300 400 nm, this signal will show you virtually every compound which can be detected by UV absorbance.

Many compounds show absorbance bands in the spectrum. Figure 15 on page 76 shows the spectrum of anisic acid as an example.

To optimize for lowest possible detectable concentrations of anisic acid, set the sample wavelength to the peak of the absorbance band (that is, 252 nm) and the sample bandwidth to the width of the absorbance band (that is, 30 nm). A reference of 360,100 is adequate. Anisic acid does not absorb in this range.

If you work with high concentrations, you may get better linearity above 1.5 AU by setting the sample wavelength to a valley in the spectrum, like 225 nm for anisic acid.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 75

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Figure 15 Optimization of Wavelength Setting

A wide bandwidth has the advantage of reducing noise by averaging over a wavelength range compared to a 4 nm bandwidth, the baseline noise is reduced by a factor of approximately 2.5, whereas the signal is about 75 % of a 4 nm wide band. The signal-to-noise ratio for a 30 nm bandwidth is twice that for a 4 nm bandwidth in our example.

Figure 16 Influence of Bandwidth on Signal and Noise

Reference bandwidth 100 nm

Reference wavelength 360

Anisic acid

Wavelength (nm)

Ab so

rb an

ce (m

AU )

sample wavelength 252 nm

30 nm bandwidth

Bandwidth

30 nm

12 nm

4 nm

Agilent InfinityLab LC Series DAD WR and MWD User Manual 76

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Because the detector averages absorbance values that are calculated for each wavelength, using a wide bandwidth does not negatively impact linearity.

The use of a reference wavelength is highly recommended to further reduce baseline drift and wander induced by room temperature fluctuations or refractive index changes during a gradient.

An example of the reduction of baseline drifts is shown in Figure 17 on page 77 for PTH-amino acids. Without a reference wavelength, the chromatogram drifts downwards due to refractive index changes induced by the gradient. This is almost completely eliminated by using a reference wavelength. With this technique, PTH-amino acids can be quantified in the low picomole range even in a gradient analysis.

Figure 17 Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without Reference

Time (min)

Grad.: 0.02 m KH PO /ACN from 12% ACN to 45% ACN in 12 min

1pmol each

Reference 380nm

Wavelength 267 nm

Wavelength 267nm No reference

PT H-

AS N

PT H-

AR G

PT H-

AL A

PT H-

PH E

PT H-

PR O

2 4

Agilent InfinityLab LC Series DAD WR and MWD User Manual 77

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Slit Width The detector has a variable slit at the entrance of the spectrograph. This is an effective tool to adapt the detector to changing demand of different analytical problems.

A narrow slit provides spectral resolution for analytes with very fine structures in the absorbance spectrum. An example of such a spectrum is benzene. The five main absorbance bands (fingers) are only 2.5 nm wide and just 6 nm apart from each other.

Figure 18 Benzene at 1, 4 and 16 nm slit width (principle)

A wide slit uses more of the light shining through the flow cell. This gives lower baseline noise as shown in Figure 19 on page 78.

Figure 19 Influence of the Slit Width on Baseline Noise

4 nm

16 nm

1 nm

Slit width 1 nm

Slit width 4 nm

Slit width 16 nm

Agilent InfinityLab LC Series DAD WR and MWD User Manual 78

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

However, with a wider slit, the spectrographs optical resolution (its ability to distinguish between different wavelengths) diminishes. Any photodiode receives light within a range of wavelength determined by the slit width. This explains why the fine spectral structure of benzene disappears when using a 16-nm wide slit.

Furthermore, the absorbance is no longer strictly linear with concentration for wavelengths at a steep slope of a compounds spectrum.

Substances with fine structures and steep slopes like benzene are very rare.

In most cases the width of absorbance bands in the spectrum is more like 30 nm as with anisic acid (Figure 15 on page 76).

In most situations, a slit width of 4 nm will give the best results.

Use a narrow slit (1 or 2 nm) if you want to identify compounds with fine spectral structures or if you need to quantify at high concentrations (> 1000 mAU) with a wavelength at the slope of the spectrum. Signals with a wide bandwidth can be used to reduce baseline noise. Because (digital) bandwidth is computed as average of absorbance, there is no impact on linearity.

Use a wide (8 or 16 nm) slit when your sample contains very small concentrations. Always use signals with bandwidth at least as wide as the slit width.

Optimizing Spectral Acquisition (DAD only) Storage of all spectra consumes a lot of disk space. It is very useful to have all spectra available during optimization of a method or when analyzing unique samples. However when running many samples of the same type, the large size of data files with all spectra may become a burden. The detector provides functions to reduce the amount of data, yet retaining the relevant spectral information.

Range Only the wavelength range where the compounds in your sample absorb contains information that is useful for purity checks and library searches. Reducing the spectrum storage range saves disk space.

Step Most substances have broad absorbance bands. Display of spectra, peak purity and library search works best if a spectrum contains 5 to 10 data points per width of the absorbance bands. For anisic acid (the example used before) a step of 4 nm would be sufficient. However a step of 2 nm gives a more optimal display of the spectrum.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 79

5 Optimizing the Detector Optimizing for Sensitivity, Selectivity, Linearity and Dispersion

Margin for Negative Absorbance The detector adjusts its gain during balance such that the baseline may drift slightly negative (about -100 mAU). In some special case, for example, when gradient with absorbing solvents are used, the baseline may drift to more negative values.

Only for such cases, increase the margin for negative absorbance to avoid overflow of the analog-to-digital converter.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 80

5 Optimizing the Detector Optimizing Selectivity

Optimizing Selectivity

Quantifying Coeluting Peaks by Peak Suppression In chromatography, two compounds may often elute together. A conventional dual-signal detector can only detect and quantify both compounds independently from each other if their spectra do not overlap. However, in most cases this is highly unlikely.

With a dual-channel detector based on diode-array technology, quantifying two compounds is possible even when both compounds absorb over the whole wavelength range. The procedure is called peak suppression or signal subtraction. As an example, the analysis of hydrochlorothiazide in the presence of caffeine is described. If hydrochlorothiazide is analyzed in biological samples, there is always a risk that caffeine is present which might interfere chromatographically with hydrochlorothiazide. As the spectra in Figure 20 on page 81 shows, hydrochlorothiazide is best detected at 222 nm, where caffeine also shows significant absorbance. It would therefore be impossible, with a conventional variable wavelength detector, to detect hydrochlorothiazide quantitatively when caffeine is present.

Figure 20 Wavelength Selection for Peak Suppression

WL3 WL4

WL2WL1 Wavelength (nm)

hydrochlorothiazide)

WL1 (204 nm, caffeine)

WL4 (282 nm, reference to suppress caffeine)

WL3 (260 nm, reference to suppress) WL2 (222 nm, hydrochlorothiazide)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 81

5 Optimizing the Detector Optimizing Selectivity

With a UV-visible detector based on a diode array and the correct choice of a reference wavelength setting, quantitative detection is possible. To suppress caffeine, the reference wavelength must be set to 282 nm. At this wavelength, caffeine shows exactly the same absorbance as at 222 nm. When the absorbance values are subtracted from each another, any indication of the presence of caffeine is eliminated. In the same way, hydrochlorothiazide can be suppressed if caffeine is to be quantified. In this case the wavelength is set to 204 nm and the reference wavelength to 260 nm. Figure 21 on page 82 shows the chromatographic results of the peak suppression technique.

The trade-off for this procedure is a loss in sensitivity. The sample signal decreases by the absorbance at the reference wavelength relative to the signal wavelength. Sensitivity may be decreased by as much as 1030 %.

Figure 21 Peak Suppression Using Reference Wavelength

Time (min) Time (min)

Wavelength 204 nm No reference

Hydrochlorothiazide and caffeine

Wavelength 222 nm No reference

Hydrochlorothiazide and caffeine

Hydrochlorothiazides uppressed Wavelength 204 nm Reference 260 nm

Caffeine suppressed

Wavelength 222 nm Reference 282 nm

Agilent InfinityLab LC Series DAD WR and MWD User Manual 82

5 Optimizing the Detector Optimizing the Detector Regarding to the System

Optimizing the Detector Regarding to the System

Delay Volume and Extra-Column Volume The delay volume is defined as the system volume between the point of mixing in the pump and the front of the column.

The extra-column volume is defined as the volume between the injection point and the detection point, excluding the volume in the column.

Extra-Column Volume

Extra-column volume is a source of peak dispersion that will reduce the resolution of the separation and so should be minimized. Smaller diameter columns require proportionally smaller extra-column volumes to keep peak dispersion at a minimum.

In a liquid chromatograph the extra-column volume will depend on the connection tubing between the autosampler, column and detector; and on the volume of the flow cell in the detector. The extra-column volume is minimized with the Agilent InfinityLab LC Series system due to the narrow-bore (0.12 mm i.d.) tubing, the low-volume heat exchangers in the column compartment and the flow cell in the detector.

How to Configure the Optimum Delay Volume To maintain resolution in the Agilent 1290 Infinity II DAD/Agilent 1290 Infinity DAD FS the 10 mm Max-Light cartridge cell has a low dispersion volume ( volume 1.0 L) and no further volume optimization is required. In situations where the alternative 60 mm Max-Light high sensitivity cell is used to get higher sensitivity the cell volume is optimized for the use with 3 mm and 4.6 mm inner diameter columns.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 83

5 Optimizing the Detector Optimizing the Detector Regarding to the System

How to Achieve Higher Sensitivity The detector has a number of parameters that are used to optimize performance. The following sections describe how the detector parameters affect performance characteristics:

Flow cell affects sensitivity,

Wavelength and bandwidth affect sensitivity, selectivity and linearity,

Slit width affects sensitivity, spectral resolution and linearity,

Peak width affects sensitivity and resolution.

Flow Cell

The Max-Light cartridge flow cell has a standard 10 mm path length and is optimized for minimal volume and dispersion ( volume 1.0 L). It has high light transmission minimizing noise to reduce noise due to the optofluidic waveguide. It is suitable for use with a wide range of analytical columns from short narrow-bore columns to long standard diameter (4.6 mm) columns. Generally the peak dispersion volume (calculated from peak width x flow rate) should be greater than about 2 L for this cell (for example 0.02 min x 200 L/min = 4 L).

The Max-Light high sensitivity cell has a path length of 60 mm and this will give between three and five times increase in signal-to-noise values depending on the application conditions. The dispersion volume is fractionally increased compared to the standard cell.

Wavelength and Bandwidth

The detector measures absorbance simultaneously at wavelengths from 190 nm to 640 nm using diode-array detection. A UV-lamp provides good sensitivity over the whole wavelength range. The diode-array detector (DAD) can simultaneously compute and send to the data system up to eight chromatographic signals and the full-range spectra at every time point.

A UV chromatogram or signal is a plot of absorbance data versus time and is defined by its wavelength and bandwidth.

The wavelength indicates the center of the detection band.

The bandwidth defines the wavelength range over which the absorbance values are averaged to give the result at each time point.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 84

5 Optimizing the Detector Optimizing the Detector Regarding to the System

For example, a signal at wavelength 250 nm with a bandwidth of 16 nm will be an average of the absorbance data from 242 nm to 258 nm. Additionally, a reference wavelength and reference bandwidth can be defined for each signal. The average absorbance from the reference bandwidth centered on the reference wavelength will be subtracted from its equivalent value at the signal wavelength to produce the output chromatogram.

The signal wavelength and bandwidth can be chosen so that they are optimized for:

Broad band universal detection

Narrow band selective detection

Sensitivity for a specific analyte.

Broad band or universal detection works by having a wide bandwidth to detect any species with absorbance in that range. For example, to detect all absorbing molecules between 200 nm and 300 nm set a signal at 250 nm with a bandwidth of 100 nm. The disadvantage is that sensitivity will not be optimal for any one of those molecules. Narrow band or selective detection is used most often. The UV spectrum for a particular molecule is examined and an appropriate absorbance maximum is selected. If possible, the range where solvents absorb strongly should be avoided (below 220 nm for methanol, below 210 nm for acetonitrile). For example, in Figure 22 on page 86, anisic acid has a suitable absorbance maximum at 252 nm. A narrow bandwidth of 4 nm to 12 nm generally gives good sensitivity and is specific for absorbance in a narrow range.

The narrow band can be optimized for sensitivity for a specific molecule. As the bandwidth is increased the signal is reduced but so is the noise and there will be an optimum for best S/N. As an approximate guide, this optimum is often close to the natural bandwidth at half-height of the absorption band in the UV spectrum. In the anisic acid example this is 30 nm.

The analytical wavelength is usually set at a wavelength maximum to increase sensitivity to that molecule. The detector is linear up to 2 AU and beyond for many applications. This offers a wide linear range for concentration. For high concentration analysis the concentration linear range can be extended by setting the wavelength to one with a lower absorbance such as a wavelength minimum or by taking a wider bandwidth which usually includes lower absorbance values. The use of wavelength maxima and minima for quantitation dates back to conventional UV detectors which because of mechanical tolerances in moving gratings needed to avoid steeply sloping parts of the spectrum. Diode-array based detectors do not have this limitation but for reasons of convention maxima and minima are chosen in preference to other parts of the spectrum.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 85

5 Optimizing the Detector Optimizing the Detector Regarding to the System

The reference bandwidth is normally set on a region of the UV spectrum in which the analyte has no absorbance. This is shown in the spectrum for anisic acid in Figure 22 on page 86. This spectrum is typical of many small molecules containing a UV chromophore. For best results the reference has been set so that it is a wide band as close to the signal wavelength as possible but on a zero absorbance region. Reference bandwidths of 60 nm to 100 nm are commonly used. The default reference is 360 nm with a bandwidth of 100 nm. A wide bandwidth is used because this reduces the noise in the reference signal (from statistical theory, the error, i.e. noise in this case, is reduced by the square root of the number of determinations). It is important that the reference bandwidth does not extend to a part of the spectrum that has some absorbance as this would then reduce the resulting signal and sensitivity would be reduced. The use of a reference wavelength can help to reduce drift or wander in the chromatogram caused by refractive index changes due to room temperature fluctuation or gradient operation. The effect of a reference signal can be easily tested by setting two otherwise identical signals, one with and one without a reference signal. If there is no part of the spectrum with zero absorbance then it will be better to have the reference signal turned off.

Figure 22 Spectrum of Anisic Acid

Agilent InfinityLab LC Series DAD WR and MWD User Manual 86

5 Optimizing the Detector Optimizing the Detector Regarding to the System

Peak Width, Response Time and Data Collection Rate

The peak width setting, response time and data rate in the detector are all linked. The available settings are shown in Table 12 on page 88. It is important to set this correctly for optimum sensitivity and to preserve the resolution achieved in the separation.

The detector internally acquires data points faster than is needed for a chromatogram and processes them to produce the signal seen by the data system. Part of the processing reduces the data to an appropriate data rate which allows the chromatographic peaks to be accurately drawn. As with most analytical determinations groups of readings are effectively averaged to reduce error in the result. The detector bunches raw data points and produces the output signal data at the required data collection rate by an electronic filtering process. If the resulting data rate is too slow (over filtering) the peak heights will be reduced and the resolution between them reduced; too fast and the data is noisier than it need be to accurately profile narrow peaks.

The peak width setting in the detector allows the user to correctly set these parameters without needing any knowledge other than sight of the chromatogram integration results to see how wide the peaks are. The peak width setting should be set for the narrowest peak width observed in the chromatogram. If it is set too wide it will make the peaks appear lower in height and wider (and potentially less resolved) and if it is set too narrow it will increase the baseline noise unnecessarily. Essentially the software uses this value to set the data collection rate such that it collects enough data points over the narrowest peaks and it is aiming for 15 to 25 points across a peak. The DAD can collect at a maximum up to 120 Hz if required which would allow enough data points to be collected over a peak that is only 0.1 s wide. The response time setting is another way of indicating how this filtering is set. It is measured in seconds and is about one-third of the peak width value (which is measured in minutes). It effectively shows how quickly the plotted signal responds to a step change in the input signal.

NOTE The full spectra is not available under all conditions.

Based on the data points, the scan data rate is reduced, see Table 12 on page 88.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 87

5 Optimizing the Detector Optimizing the Detector Regarding to the System

Table 12 Peak Width Response Time Data Rate (G7115A/G7165A)

Peak width at half

height [min]1

1 Values in the User Interface may be rounded.

Response [s] Scan data rate[Hz] 251 pts/scan

Scan data rate[Hz] 501 pts/scan

Scan data rate[Hz] > 501 pts/scan

< 0.0015625 0.015625 120 120 40 20

> 0.0015625 0.03125 120 120 40 20

> 0.003125 0.0625 80 80 40 20

> 0.00625 0.125 40 40 40 20

> 0.0125 0.25 20 20 20 20

> 0.025 0.5 10 10 10 10

> 0.05 1 5 5 5 5

> 0.1 2 2.5 2.5 2.5 2.5

> 0.2 4 1.25 1.25 1.25 1.25

> 0.4 8 0.625 0.625 0.625 0.625

> 0.85 16 0.3125 0.3125 0.3125 0.3125

NOTE The maximum spectra scan rate depends on the data points per scan, see Table 12 on page 88.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 88

5 Optimizing the Detector Warm up of the Detector

Warm up of the Detector

Give the optical unit enough time to warm-up and stabilize (> 60 minutes). The detector is temperature controlled. After turn-on of the detector, it goes through a cycle of different states:

0 to 0.5 minutes the heater control is OFF and the heater element runs at 0 % duty cycle.

0.5 to 1 minutes the heater control is OFF and the heater element runs at 66% duty cycle. This first minute is used as self-test of the heater functionality.

1 to 30 minutes the heater control is OFF and the heater element runs at 40% duty cycle.

After 30 minutes the heater control is ON and is working with optimized parameters to get the optical unit into the optimal temperature window stabilized.

This cycle starts

when the detector is turned off/on

when the lamp is turned off/on

to ensure that the temperature control operates in a defined control range.

The figures below show the first two hours of a detector warm-up phase. The lamp was turned on immediately after turn on of the detector.

NOTE The times to stabilize the baseline may vary from instrument to instrument and depends on the environment. The example below was done under stable environmental conditions.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 89

5 Optimizing the Detector Warm up of the Detector

Figure 23 Detector Warm-up 1st hour

Figure 24 Detector Warm-up 2nd hour

Agilent InfinityLab LC Series DAD WR and MWD User Manual 90

6 Troubleshooting and Diagnostics

Available Tests vs User Interfaces 92

Agilent Lab Advisor Software 93

Available Diagnostic Functions vs. Product Level 94

This chapter gives an overview about the troubleshooting and diagnostic features.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 91

6 Troubleshooting and Diagnostics Available Tests vs User Interfaces

Available Tests vs User Interfaces

Screenshots used within these procedures are based on the Agilent Lab Advisor software.

Figure 25 Tests in the Agilent Lab Advisor

NOTE Depending on the used interface, the available tests and the screens/reports may vary.

Preferred tool should be the Agilent Lab Advisor, see Agilent Lab Advisor Software on page 93.

Agilent Lab Advisor B.02.08 or later is required.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 92

6 Troubleshooting and Diagnostics Agilent Lab Advisor Software

Agilent Lab Advisor Software

The Agilent Lab Advisor Software (basic license, shipped with an Agilent LC pump) is a standalone product that can be used with or without a chromatographic data system. Agilent Lab Advisor helps to manage the lab for high-quality chromatographic results by providing a detailed system overview of all connected analytical instruments with instrument status, Early Maintenance Feedback counters (EMF), instrument configuration information, and diagnostic tests. With the push of a button, a detailed diagnostic report can be generated. Upon request, the user can send this report to Agilent for a significantly improved troubleshooting and repair process.

The Agilent Lab Advisor software is available in two versions:

Lab Advisor Basic

Lab Advisor Advanced

Lab Advisor Basic is included with every Agilent 1200 Infinity Series and Agilent InfinityLab LC Series instrument.

The Lab Advisor Advanced features can be unlocked by purchasing a license key, and include real-time monitoring of instrument actuals, all various instrument signals, and state machines. In addition, all diagnostic test results, calibration results, and acquired signal data can be uploaded to a shared network folder. The Review Client included in Lab Advisor Advanced allows to load and examine the uploaded data no matter on which instrument it was generated. This makes Data Sharing an ideal tool for internal support groups and users who want to track the instrument history of their analytical systems.

The optional Agilent Maintenance Wizard Add-on provides an easy-to-use, step-by-step multimedia guide for performing preventive maintenance on Agilent 1200 Infinity LC Series instrument.

The tests and diagnostic features that are provided by the Agilent Lab Advisor software may differ from the descriptions in this manual. For details, refer to the Agilent Lab Advisor software help files.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 93

6 Troubleshooting and Diagnostics Available Diagnostic Functions vs. Product Level

Available Diagnostic Functions vs. Product Level

Table 13 Available Diagnostic Functions vs. Product Level

Product Level

Tests

-ASTM Drift and Noise Test Basic Advanced

-Cell Test Basic Advanced

-D/A Converter Test Basic Advanced

-Dark Current Test Basic Advanced

-Filter Test Basic Advanced

-Holmium Oxide Test Basic Advanced

-Intensity Test Basic Advanced

-Quick Noise Test Basic Advanced

-Self Test Basic Advanced

-Slit Test Basic Advanced

-Wavelength Verification Test Basic Advanced

Calibrations

-Wavelength Calibration Basic Advanced

Tools

-Diagnostic Buffers Basic Advanced

-Module Info Basic Advanced

-Test Chromatogram Basic Advanced

-Spectral Scan Basic Advanced

Controls

-Advanced Method Parameters

-Vis lamp required Advanced

-Analog Output 1 Offset [% Full Scale]

Advanced

-D2 lamp required Advanced

-Analog Output 1 Attenuation Advanced

Agilent InfinityLab LC Series DAD WR and MWD User Manual 94

6 Troubleshooting and Diagnostics Available Diagnostic Functions vs. Product Level

-Configuration

-Analog Output 1 Range Advanced

-Remote Pulse Duration [s]* Basic Advanced

-Control

-Vis lamp Basic Advanced

-Balance Detector Advanced

-UV Lamp Basic Advanced

-Method Parameters

-Set Signal A Advanced

-Set Signal C Advanced

-Set Data Rate [HZ] Advanced

-Set Signal B Advanced

-Module Information

-Identify Module Basic Advanced

-Special Commands

-Lamp tag required Basic Advanced

-Cell tag required Basic Advanced

-Detector Reset Basic Advanced

-Clear Error Basic Advanced

Statemachines

-UV Lamp Basic Advanced

-Vis Lamp Advanced

Signals

-Signal A [mAU] Advanced

-Lamp Voltage [V] Advanced

-Board Temperature [C] Advanced

-Optical Temperature [C] Advanced

Table 13 Available Diagnostic Functions vs. Product Level

Product Level

Agilent InfinityLab LC Series DAD WR and MWD User Manual 95

6 Troubleshooting and Diagnostics Available Diagnostic Functions vs. Product Level

EMF Counters

-Accumulated UV Lamp On-Time

Basic Advanced

-Number of UV Lamp Ignitions Basic Advanced

-Accumulated Tungsten Lamp On-Time

Basic Advanced

-Number of Tungsten Lamp Ignitions

Basic Advanced

Table 13 Available Diagnostic Functions vs. Product Level

Product Level

Agilent InfinityLab LC Series DAD WR and MWD User Manual 96

7 Error Information

What Are Error Messages 98

General Error Messages 99 Timeout 99 Shutdown 100 Remote Timeout 101 Lost CAN Partner 101 Leak Sensor Short 102 Leak Sensor Open 102 Compensation Sensor Open 103 Compensation Sensor Short 103 Fan Failed 104 Leak 104 Open Cover 105 Cover Violation 105 ERI Messages 106

Detector Error Messages 107 Visible Lamp Current 107 Visible Lamp Voltage 107 Diode Current Leakage 108 UV Lamp Current 108 UV Lamp Voltage 109 UV Ignition Failed 109 UV Heater Current 110 Calibration Values Invalid 110 Holmium Oxide Test Failed 111 Illegal Temperature Value from Sensor on Main Board 111 Illegal Temperature Value from Sensor at Air Inlet 112 Wavelength Recalibration Lost 112 Heater at fan assembly failed 113 Heater Power At Limit 113 DSP Not Running 114

This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 97

7 Error Information What Are Error Messages

What Are Error Messages

Error messages are displayed in the user interface when an electronic, mechanical, or hydraulic (flow path) failure occurs which requires attention before the analysis can be continued (for example, repair, or exchange of consumables is necessary). In the event of such a failure, the red status indicator at the front of the module is switched on, and an entry is written into the module logbook.

If an error occurs outside a method run, other modules will not be informed about this error. If it occurs within a method run, all connected modules will get a notification, all LEDs get red and the run will be stopped. Depending on the module type, this stop is implemented differently. For example, for a pump the flow will be stopped for safety reasons. For a detector, the lamp will stay on in order to avoid equilibration time. Depending on the error type, the next run can only be started, if the error has been resolved, for example liquid from a leak has been dried. Errors for presumably single time events can be recovered by switching on the system in the user interface.

Special handling is done in case of a leak. As a leak is a potential safety issue and may have occurred at a different module from where it has been observed, a leak always causes a shutdown of all modules, even outside a method run.

In all cases, error propagation is done via the CAN bus or via an APG/ERI remote cable (see documentation for the APG/ERI interface).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 98

7 Error Information General Error Messages

General Error Messages

General error messages are generic to all Agilent series HPLC modules and may show up on other modules as well.

Timeout Error ID: 0062

The timeout threshold was exceeded.

Probable cause Suggested actions

1 The analysis was completed successfully,

and the timeout function switched off the module as requested.

Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required.

2 A not-ready condition was present during a sequence or multiple-injection run for a period longer than the timeout threshold.

Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 99

7 Error Information General Error Messages

Shutdown Error ID: 0063

An external instrument has generated a shutdown signal on the remote line.

The module continually monitors the remote input connectors for status signals. A LOW signal input on pin 4 of the remote connector generates the error message.

Probable cause Suggested actions

1 Leak detected in another module with a CAN connection to the system.

Fix the leak in the external instrument before restarting the module.

2 Leak detected in an external instrument with

a remote connection to the system.

Fix the leak in the external instrument before restarting the module.

3 Shut-down in an external instrument with a

remote connection to the system.

Check external instruments for a shut-down condition.

4 The degasser failed to generate sufficient vacuum for solvent degassing.

Check the vacuum degasser for an error condition. Refer to the Service Manual for the degasser or the pump that has the degasser built-in.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 100

7 Error Information General Error Messages

Remote Timeout Error ID: 0070

A not-ready condition is still present on the remote input. When an analysis is started, the system expects all not-ready conditions (for example, a not-ready condition during detector balance) to switch to run conditions within one minute of starting the analysis. If a not-ready condition is still present on the remote line after one minute the error message is generated.

Lost CAN Partner Error ID: 0071

During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed.

The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated.

Probable cause Suggested actions

1 Not-ready condition in one of the instruments connected to the remote line.

Ensure the instrument showing the not-ready condition is installed correctly, and is set up correctly for analysis.

2 Defective remote cable. Exchange the remote cable.

3 Defective components in the instrument

showing the not-ready condition.

Check the instrument for defects (refer to the instruments documentation).

Probable cause Suggested actions

1 CAN cable disconnected. Ensure all the CAN cables are connected correctly.

Ensure all CAN cables are installed correctly.

2 Defective CAN cable. Exchange the CAN cable.

3 Defective mainboard in another module. Switch off the system. Restart the system, and determine which module or modules are not recognized by the system.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 101

7 Error Information General Error Messages

Leak Sensor Short Error ID: 0082

The leak sensor in the module has failed (short circuit).

The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current increases above the upper limit, the error message is generated.

Leak Sensor Open Error ID: 0083

The leak sensor in the module has failed (open circuit).

The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated.

Probable cause Suggested actions

1 Defective leak sensor. Please contact your Agilent service representative.

2 Leak sensor incorrectly routed, being pinched by a metal component.

Please contact your Agilent service representative.

3 Power switch assembly defective Please contact your Agilent service representative.

4 Cable or contact problem. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Leak sensor not connected to the power switch board.

Please contact your Agilent service representative.

2 Defective leak sensor. Please contact your Agilent service representative.

3 Leak sensor incorrectly routed, being pinched by a metal component.

Please contact your Agilent service representative.

4 Power switch assembly defective Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 102

7 Error Information General Error Messages

Compensation Sensor Open Error ID: 0081

The ambient-compensation sensor (NTC) on the power switch board in the module has failed (open circuit).

The resistance across the temperature compensation sensor (NTC) on the power switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.

Compensation Sensor Short Error ID: 0080

The ambient-compensation sensor (NTC) on the power switch board in the module has failed (open circuit).

The resistance across the temperature compensation sensor (NTC) on the power switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor falls below the lower limit, the error message is generated.

Probable cause Suggested actions

1 Loose connection between the power switch board and the mainboard

Please contact your Agilent service representative.

2 Defective power switch assembly Please contact your Agilent service representative.

Probable cause Suggested actions

1 Defective power switch assembly Please contact your Agilent service representative.

2 Loose connection between the power switch board and the mainboard

Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 103

7 Error Information General Error Messages

Fan Failed Error ID: 0068

The cooling fan in the module has failed.

The hall sensor on the fan shaft is used by the mainboard to monitor the fan speed. If the fan speed falls below a certain limit for a certain length of time, the error message is generated.

This limit is given by 2 revolutions/second for longer than 5 seconds.

Depending on the module, assemblies (e.g. the lamp in the detector) are turned off to assure that the module does not overheat inside.

Leak Error ID: 0064

A leak was detected in the module.

The signals from the two temperature sensors (leak sensor and board-mounted temperature-compensation sensor) are used by the leak algorithm to determine whether a leak is present. When a leak occurs, the leak sensor is cooled by the solvent. This changes the resistance of the leak sensor which is sensed by the leak sensor circuit on the main board.

Probable cause Suggested actions

1 Fan cable disconnected. Please contact your Agilent service representative.

2 Defective fan. Please contact your Agilent service representative.

3 Defective mainboard. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Loose fittings. Ensure all fittings are tight.

2 Broken capillary. Exchange defective capillaries.

3 Leaking flow cell. Exchange flow cell components.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 104

7 Error Information General Error Messages

Open Cover Error ID: 0205

The top foam has been removed.

The sensor on the main board detects when the top foam is in place. If the foam is removed, the fan is switched off, and the error message is generated.

Cover Violation Error ID: 7461

The top foam has been removed.

The sensor on the main board detects when the top foam is in place. If the foam is removed while the lamps are on (or if an attempt is made to switch on for example the lamps with the foam removed), the lamps are switched off, and the error message is generated.

Probable cause Suggested actions

1 The top foam was removed during operation. Please contact your Agilent service representative.

2 Foam not activating the sensor. Please contact your Agilent service representative.

3 Defective sensor or main board. Please contact your Agilent service representative.

Probable cause Suggested actions

1 The top foam was removed during operation. Please contact your Agilent service representative.

2 Foam not activating the sensor. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 105

7 Error Information General Error Messages

ERI Messages Error ID: 11120 (+5 V) , 11121 (+25 V)

The ERI (Enhanced Remote Interface) provides two error events related to over current situations on the +5 V and +24 V lines.

Probable cause Suggested actions

1 The load on the ERI is too high. Reduce the load.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 106

7 Error Information Detector Error Messages

Detector Error Messages

These errors are detector specific.

Visible Lamp Current The visible lamp current is missing.

The processor continually monitors the lamp current during operation. If the current falls below the lower current limit, the error message is generated.

Visible Lamp Voltage The visible lamp voltage is missing.

The processor continually monitors the voltage across the lamp during operation. If the lamp voltage falls below the lower limit, the error message is generated.

Probable cause Suggested actions

1 Lamp disconnected. Ensure the visible lamp connector is seated firmly.

2 Defective visible lamp. Exchange the visible lamp.

3 Defective connector or cable. Please contact your Agilent service representative.

4 Defective power supply. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Defective connector or cable. Please contact your Agilent service representative.

2 Defective power supply. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 107

7 Error Information Detector Error Messages

Diode Current Leakage Error ID: 1041

When the detector is switched on, the processor checks the leakage current of each of the optical diodes. If the leakage current exceeds the upper limit, the error message is generated.

UV Lamp Current Error ID: 7450

The UV lamp current is missing.

The processor continually monitors the anode current drawn by the lamp during operation. If the anode current falls below the lower current limit, the error message is generated.

Probable cause Suggested actions

1 Defective PDA/optical unit. Please contact your Agilent service representative.

2 Defective connector or cable. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Lamp disconnected. Ensure the UV lamp connector is seated firmly.

2 Defective UV lamp or non-Agilent lamp. Exchange the UV lamp.

3 Defective detector main board. Please contact your Agilent service representative.

4 Defective power supply. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 108

7 Error Information Detector Error Messages

UV Lamp Voltage Error ID: 7451

The UV lamp anode voltage is missing.

The processor continually monitors the anode voltage across the lamp during operation. If the anode voltage falls below the lower limit, the error message is generated.

UV Ignition Failed Error ID: 7452

The UV lamp failed to ignite.

The processor monitors the UV lamp current during the ignition cycle. If the lamp current does not rise above the lower limit within 2 5 seconds, the error message is generated.

Probable cause Suggested actions

1 Defective UV lamp or non-Agilent lamp. Exchange the UV lamp.

2 Defective detector main board. Please contact your Agilent service representative.

3 Defective power supply. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Lamp too hot. Hot gas discharge lamps may

not ignite as easily as cold lamps.

Switch off the lamp and allow it to cool down for at least 15 minutes.

2 Lamp disconnected. Ensure the lamp is connected.

3 Defective UV lamp or non-Agilent lamp. Exchange the UV lamp.

4 Defective detector main board. Please contact your Agilent service representative.

5 Defective power supply. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 109

7 Error Information Detector Error Messages

UV Heater Current Error ID: 7453

The UV lamp heater current is missing.

During UV lamp ignition, the processor monitors the heater current. If the current does not rise above the lower limit within one second, the error message is generated.

Calibration Values Invalid Error ID: 1036

The calibration values read from the spectrometer ROM are invalid.

After recalibration, the calibration values are stored in ROM. The processor periodically checks if the calibration data are valid. If the data are invalid or cannot be read from the spectrometer ROM, the error message is generated.

Probable cause Suggested actions

1 Lamp disconnected. Ensure the UV lamp is connected.

2 Ignition started without the top foam in place.

Please contact your Agilent service representative.

3 Defective UV lamp or non-Agilent lamp. Exchange the UV lamp.

4 Defective detector main board. Please contact your Agilent service representative.

5 Defective power supply. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Defective connector or cable. Please contact your Agilent service representative.

2 Defective PDA/optical unit. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 110

7 Error Information Detector Error Messages

Holmium Oxide Test Failed

Illegal Temperature Value from Sensor on Main Board Error ID: 1071

This temperature sensor (located on the detector main board) delivered a value outside the allowed range. The parameter of this event equals the measured temperature in 1/100 centigrade. As a result the temperature control is switched off.

Probable cause Suggested actions

1 Lamps switched off. Ensure the lamps are switched on.

2 Defective or dirty flow cell. Ensure the flow cell is inserted correctly, and is free from contamination (cell windows, buffers etc.).

3 Defective filter assembly. Please contact your Agilent service representative.

4 Defective achromat assembly. Please contact your Agilent service representative.

5 Defective PDA/optical unit. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Defective sensor or main board. Please contact your Agilent service representative.

2 Detector is exposed to illegal ambient conditions.

Verify that the ambient conditions are within the allowed range.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 111

7 Error Information Detector Error Messages

Illegal Temperature Value from Sensor at Air Inlet Error ID: 1072

This temperature sensor delivered a value outside the allowed range. The parameter of this event equals the measured temperature in 1/100 centigrade. As a result the temperature control is switched off.

Wavelength Recalibration Lost Error ID: 1037

The calibration information needed for your detector to operate correctly has been lost.

During calibration of the detector the calibration values are stored in ROM. If no data is available in the spectrometer ROM, the error message is generated.

Probable cause Suggested actions

1 The temperature sensor is defect. Replace the cable to the main board.

Please contact your Agilent service representative.

2 Detector is exposed to illegal ambient

conditions.

Verify that the ambient conditions are within the allowed range.

Probable cause Suggested actions

1 The detector is new. Recalibrate the detector.

2 The detector has been repaired. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 112

7 Error Information Detector Error Messages

Heater at fan assembly failed Error ID: 1073

Every time the deuterium lamp or the tungsten lamp (DAD only) is switched on or off a heater self-test is performed. If the test fails an error event is created. As a result the temperature control is switched off.

Heater Power At Limit Error ID: 1074

The available power of the heater reached either the upper or lower limit. This event is sent only once per run. The parameter determines which limit has been hit:

0 means upper power limit hit (excessive ambient temperature drop).

1 means lower power limit hit (excessive ambient temperature increase).

Probable cause Suggested actions

1 Defective connector or cable. Please contact your Agilent service representative.

2 Defective heater. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Excessive ambient temperature change. Wait until temperature control equilibrates.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 113

7 Error Information Detector Error Messages

DSP Not Running This error message comes up when the communication between the optical unit and the main board has a problem.

Probable cause Suggested actions

1 Random communication error. Switch the detector off and on again at the power switch. If the error reoccurs:

Please contact your Agilent service representative.

2 Defective detector main board. Please contact your Agilent service representative.

3 Defective PDA/optical unit. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 114

8 Test Functions and Calibration

Introduction 116

Conditions of Detector 120

Failing a Test 121

Self-Test 122

Intensity Test 124 Test Failed 126

Cell Test 127 Test Failed (low ratio value) 128

Filter Test 129 Test Evaluation 130

Holmium Oxide Test 131 Test Evaluation 131

ASTM Drift and Noise Test 134 Test Failed 137

Slit Test 138 Test Failed 140

Wavelength Verification Test 141

Wavelength Calibration 143 Wavelength Recalibration Fails 145

D/A Converter (DAC) Test 146 Test Evaluation 147

Dark-Current Test 148 Test Failed 149

Spectral Scan 150

Other Lab Advisor Functions 151 EMFs - Early Maintenance Feature 151

This chapter describes the tests for the module.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 115

8 Test Functions and Calibration Introduction

Introduction

All tests are described based on the Agilent Lab Advisor Software B.02.08. Other user interfaces may not provide any test or just a few.

For details on the use of the interface refer to the interface documentation.

The Lab Advisor shows the available test under Service & Diagnostics.

Table 14 Available Diagnostic Functions vs. Product Level

Product Level

Tests

-ASTM Drift and Noise Test Basic Advanced

-Cell Test Basic Advanced

-D/A Converter Test Basic Advanced

-Dark Current Test Basic Advanced

-Filter Test Basic Advanced

-Holmium Oxide Test Basic Advanced

-Intensity Test Basic Advanced

-Quick Noise Test Basic Advanced

-Self Test Basic Advanced

-Slit Test Basic Advanced

-Wavelength Verification Test Basic Advanced

Calibrations

-Wavelength Calibration Basic Advanced

Tools

-Diagnostic Buffers Basic Advanced

-Module Info Basic Advanced

-Test Chromatogram Basic Advanced

-Spectral Scan Basic Advanced

Agilent InfinityLab LC Series DAD WR and MWD User Manual 116

8 Test Functions and Calibration Introduction

Controls

-Advanced Method Parameters

-Vis lamp required Advanced

-Analog Output 1 Offset [% Full Scale]

Advanced

-D2 lamp required Advanced

-Analog Output 1 Attenuation Advanced

-Configuration

-Analog Output 1 Range Advanced

-Remote Pulse Duration [s]* Basic Advanced

-Control

-Vis lamp Basic Advanced

-Balance Detector Advanced

-UV Lamp Basic Advanced

-Method Parameters

-Set Signal A Advanced

-Set Signal C Advanced

-Set Data Rate [HZ] Advanced

-Set Signal B Advanced

-Module Information

-Identify Module Basic Advanced

-Special Commands

-Lamp tag required Basic Advanced

-Cell tag required Basic Advanced

-Detector Reset Basic Advanced

-Clear Error Basic Advanced

Table 14 Available Diagnostic Functions vs. Product Level

Product Level

Agilent InfinityLab LC Series DAD WR and MWD User Manual 117

8 Test Functions and Calibration Introduction

Statemachines

-UV Lamp Basic Advanced

-Vis Lamp Advanced

Signals

-Signal A [mAU] Advanced

-Lamp Voltage [V] Advanced

-Board Temperature [C] Advanced

-Optical Temperature [C] Advanced

EMF Counters

-Accumulated UV Lamp On-Time

Basic Advanced

-Number of UV Lamp Ignitions Basic Advanced

-Accumulated Tungsten Lamp On-Time

Basic Advanced

-Number of Tungsten Lamp Ignitions

Basic Advanced

Table 14 Available Diagnostic Functions vs. Product Level

Product Level

Agilent InfinityLab LC Series DAD WR and MWD User Manual 118

8 Test Functions and Calibration Introduction

Figure 26 The Lab Advisor shows the available tests

Agilent InfinityLab LC Series DAD WR and MWD User Manual 119

8 Test Functions and Calibration Conditions of Detector

Conditions of Detector

The test usually should be performed with a detector turned on for at least one hour, so that the temperature regulation of the optical unit is working (not active during the first 30 minutes after turn on). If the detector is on, tests can be performed usually 10 minutes after the UV-lamp has been turned on.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 120

8 Test Functions and Calibration Failing a Test

Failing a Test

If a test fails with the flow cell installed, repeat the test with removed flow cell and compare. If the test fails also, then start with proposed actions mentioned in the details of the tests.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 121

8 Test Functions and Calibration Self-Test

Self-Test

The self-test runs a series of individual tests (described on the next pages), and evaluates the results automatically. The following tests are run:

Filter Test

Slit Test

Dark Current Test

Intensity Test

Wavelength Verification Test

Holmium Oxide Test

Spectral Flatness Test

ASTM Noise Test (without testing the Drift)

When For complete detector check.

Parts required Description

Removed Flow Cell or Flow Cell (filled with water)

Preparations Lamp must be on for at least 10 min. For noise test a longer warm-up time may be required (> 2 h).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 122

8 Test Functions and Calibration Self-Test

1 Run the Self Test with Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 27 Self Test Results

Under the tab Signals you can find the detailed signals from the tests.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 123

8 Test Functions and Calibration Intensity Test

Intensity Test

The intensity test measures the intensity of the UV-lamp over the full wavelength range (190 - 950 nm). Four spectral ranges are used to evaluate the intensity spectrum. The test is used to determine the performance of the lamp and optics (see also Cell Test on page 127). When the test is started, the 1 nm slit is moved into the light path automatically. To eliminate effects due to absorbing solvents, the test should be done with water in the flow cell or with the flow cell removed. The shape of the intensity spectrum is primarily dependent on the lamp, grating, and diode array characteristics. Therefore, intensity spectra will differ slightly between instruments.

When In case of UV-lamp problem (drift, noise).

Parts required Description

Removed Flow Cell or Flow Cell (filled with water)

Preparations Lamp must be on for at least 10 min. When using a flow cell a flow rate of 1 mL/min with water is required.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 124

8 Test Functions and Calibration Intensity Test

1 Run the Intensity Test with Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 28 Intensity Test Results (w/o flow cell)

Figure 29 Intensity Test Signals (w/o flow cell)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 125

8 Test Functions and Calibration Intensity Test

Test Failed Intensity Test Evaluation

Probable cause Suggested actions

1 Absorbing solvent or air bubble in flow cell. Ensure the flow cell is filled with water, and free from air bubbles.

Repeat test without flow cell and compare results.

2 Incorrect calibration Recalibrate and repeat the test.

3 Dirty or contaminated flow cell. Run the cell test. If the test fails, flush the flow cell. See also Maintenance of Standard, Semi-Micro or Micro Flow Cell on page 167.

4 Dirty or contaminated optical components. Please contact your Agilent service representative.

5 Old UV-lamp. Exchange the UV-lamp.

6 Defect optical unit. If the test fails with flow cell and new UV-lamp, please contact your Agilent service representative.

NOTE If only one range fails and the application does not require this range, the lamp may not be changed.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 126

8 Test Functions and Calibration Cell Test

Cell Test

The cell test measures the intensity of the UV- and tungsten lamps over the full wavelength range (190 - 950 nm), once with the flow cell installed, and once with the flow cell removed. The resulting intensity ratio is a measure of the amount of light absorbed by the flow cell. The test can be used to check for dirty or contaminated flow cell windows. When the test is started, the 1 nm slit is moved into the light path automatically.

This test should be performed initially with a new detector/flow cell. The values should be kept for later reference/comparison.

1 Run the Cell-Test with Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 30 Cell Test Results

When In case of low intensity or noise and drift problem.

Parts required Description

Flow Cell (filled with water)

Preparations Lamp must be on for at least 10 min. When using a flow cell a flow rate of 1 mL/min with water is required.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 127

8 Test Functions and Calibration Cell Test

Figure 31 Cell Test Signals (example shows low intensity for flow cell)

Test Failed (low ratio value) Cell Test Evaluation

Probable cause Suggested actions

1 Absorbing solvent or air bubble in flow cell. Ensure the flow cell is filled with water, and free from air bubbles.

2 Dirty or contaminated flow cell. Clean the flow cell as described in Maintenance of Standard, Semi-Micro or Micro Flow Cell on page 167.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 128

8 Test Functions and Calibration Filter Test

Filter Test

The filter test checks the correct operation of the filter assembly. When the test is started, the holmium oxide filter is moved into position. During filter movement, the absorbance signal is monitored. As the edge of the filter passes through the light path, an absorbance maximum is seen. Once the filter is in position, the absorbance maximum (of holmium oxide) is determined. Finally, the filter is moved out of the light path. During movement, an additional absorbance maximum is expected as the edge of the filter passes through the light path. The test passes successfully, if the two maxima resulting from the edge of the filter assembly (during filter movement) are seen, and the absorbance maximum of holmium oxide is within the limits.

Figure 32 Filter Test

Agilent InfinityLab LC Series DAD WR and MWD User Manual 129

8 Test Functions and Calibration Filter Test

Test Evaluation

Filter Test Failed

Test Failed

Holmium Oxide Maximum out of Limits

Probable cause Suggested actions

1 Filter assembly (lever and filter) not installed. Install the filter assembly.

2 Defective filter motor. Please contact your Agilent service representative.

Probable cause Suggested actions

1 Holmium oxide filter not installed. Install the holmium oxide filter.

2 Dirty or contaminated filter. Exchange the holmium oxide filter.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 130

8 Test Functions and Calibration Holmium Oxide Test

Holmium Oxide Test

The holmium oxide test uses characteristic absorbance maxima of the built-in holmium oxide filter to verify wavelength accuracy (see also Wavelength Verification Test on page 141). When the test is started, the 1-nm slit is moved into the light path automatically. To eliminate effects due to absorbing solvents, the test should be done with water in the flow cell or with removed flow cell.

Test Evaluation

Holmium Oxide Test Evaluation

Limits:

The test is evaluated by the instrument, and the measured maxima are displayed automatically. The test fails if one or more of the maxima lies outside of the limits (see Figure 33 on page 132).

NOTE See also Declaration of Conformity for HOX2 Filter on page 276.

361.0 nm 360.0 - 362.0 nm ( 1nm)

418.9 nm 417.9 - 419.9 nm ( 1nm) (not with ChemStation)

453.7 nm 452.7 - 454.7 nm ( 1nm)

536.7 nm 535.7 - 537.7 nm ( 1nm)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 131

8 Test Functions and Calibration Holmium Oxide Test

Figure 33 Holmium Oxide Test

Figure 34 Holmium Oxide Test (Signal)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 132

8 Test Functions and Calibration Holmium Oxide Test

Test Failed

Probable cause Suggested actions

1 Absorbing solvent or air bubble in flow cell. Ensure the flow cell is filled with water.

2 Incorrect calibration Recalibrate (see Wavelength Calibration on page 143) and repeat the test.

3 Dirty or contaminated flow cell. Run the cell test (see Cell Test on page 127). If the test fails, exchange the flow cell windows.

4 Dirty or contaminated optical components (achromat, windows).

Clean optical components with alcohol and lint-free cloth or replace the parts (see Intensity Test on page 124).

5 Old or non-Agilent lamp. Exchange the UV lamp.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 133

8 Test Functions and Calibration ASTM Drift and Noise Test

ASTM Drift and Noise Test

The ASTM noise test determines the detector noise over a period of 20 minutes. The test is done with installed flow cell or flow cell removed.

This test also checks for the drift. It is also part of the Self Test (without checking for the drift).

If the test is performed with the flow cell removed, the test results are not influenced by solvent or pump effects.

When In case of noise and drift problem.

Parts required Description

Removed Flow Cell or Flow Cell (filled with water)

Preparations Detector and UV-lamp must be on for at least 2 hours. ASTM measurements based on specifications may require longer stabilization times. When using a flow cell a flow rate of 1 mL/min with water is required.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 134

8 Test Functions and Calibration ASTM Drift and Noise Test

1 Run the ASTM Drift and Noise Test with Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 35 ASTM Drift and Noise Test Results (with Flow Cell removed)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 135

8 Test Functions and Calibration ASTM Drift and Noise Test

Figure 36 ASTM Drift and Noise Test Signal (with Flow Cell removed)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 136

8 Test Functions and Calibration ASTM Drift and Noise Test

Test Failed ASTM Noise Test Evaluation

Probable cause Suggested actions

1 Insufficient lamp warm-up time. Allow detector and UV-lamp turned on for at least 2 hours.

2 Absorbing solvent or air bubble in flow cell. Ensure the flow cell is filled with water, and free from air bubbles.

3 Dirty or contaminated flow cell. Flush flow cell.

Clean the flow cell as described in Maintenance of Standard, Semi-Micro or Micro Flow Cell on page 167.

4 Old UV-lamp. Exchange the UV-lamp.

5 Old Vis-lamp. Exchange the Vis-lamp.

6 Environment not according to specifications. Improve environment.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 137

8 Test Functions and Calibration Slit Test

Slit Test

The slit test verifies correct operation of the micromechanical slit.

During the test, the slit is moved through all slit positions while the detector monitors the lamp intensity change. When the slit position is changed, the intensity drop (move to smaller slit) or intensity increase (move to larger slit) must be within a defined range.

If the test is performed with the flow cell removed, the test results are not influenced by solvent or pump effects.

If the intensity changes are outside the expected range, the test fails.

1 Run the Slit Test with the Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 37 Slit Test Results

When In case of problems.

Parts required Description

Removed Flow Cell or Flow Cell (filled with water)

Preparations Lamp must be on for at least 10 min. When using a flow cell a flow rate of 1 mL/min with water is required.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 138

8 Test Functions and Calibration Slit Test

Figure 38 Slit Test Signal

Agilent InfinityLab LC Series DAD WR and MWD User Manual 139

8 Test Functions and Calibration Slit Test

Test Failed Slit Test Evaluation

Probable cause Suggested actions

1 Air bubble in flow cell. Flush the flow cell or remove the flow cell.

2 Old lamp. Run the Intensity Test. Exchange the lamp if old or defective.

3 Defective slit assembly. Please contact your Agilent service representative.

4 Defective detector main board. Please contact your Agilent service representative.

5 Defective optical unit. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 140

8 Test Functions and Calibration Wavelength Verification Test

Wavelength Verification Test

The detector uses the alpha (656.1 nm) and beta (486 nm) emission lines of the UV-lamp for wavelength calibration. The sharp emission lines enable accurate calibration. When verification is started, the 1-nm slit is moved into the light path automatically. The test is run with Flow Cell removed or with Flow Cell installed.

If the test is performed with the Flow Cell removed, the test results are not influenced by solvent or pump effects.

When The detector is calibrated at the factory, and under normal operating conditions should not require recalibration. However, it is advisable to recalibrate: after repair of components in the optical unit, after exchange of the optical unit or main board, after replacing the Flow Cell or UV-lamp, after significant environmental condition changes (temperature, humidity), at a regular interval, at least once per year (for example, prior to an Operational

Qualification/Performance Verification procedure), and when chromatographic results indicate the detector may require recalibration.

Parts required Description

Removed Flow Cell or Flow Cell (filled with water)

Preparations Lamp must be on for at least 10 min. When using a Flow Cell a flow rate of 0.5 mL/min with water is required.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 141

8 Test Functions and Calibration Wavelength Verification Test

1 Run the Wavelength Verification Test with the Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 39 Wavelength Verification Results

Agilent InfinityLab LC Series DAD WR and MWD User Manual 142

8 Test Functions and Calibration Wavelength Calibration

Wavelength Calibration

The detector uses the alpha (656.1 nm) and beta (486 nm) emission lines of the deuterium lamp for wavelength calibration. The sharp emission lines enable more accurate calibration than is possible with holmium oxide. When recalibration is started, the 1 nm slit is moved into the light path automatically. The gain is set to zero.

On completion of the scan, the alpha- and beta-line deviations (in nm) are displayed. These values indicate how far the detector calibration deviates from the actual positions of the alpha and beta emission lines. After calibration, the deviation is zero.

To eliminate effects due to absorbing solvents, remove the flow cell before starting the test.

When The detector is calibrated at the factory, and under normal operating conditions should not require recalibration. However, it is advisable to recalibrate: after maintenance (flow cell or UV-lamp), after repair of components in the optical unit, after exchange of the optical unit or main board, after significant environmental condition changes (temperature, humidity), at a regular interval, at least once per year (for example, prior to an Operational

Qualification/Performance Verification procedure), and when chromatographic results indicate the detector may require recalibration.

Parts required Description

Removed Flow Cell or Flow Cell (filled with water)

Preparations Detector/lamp must be on for more than 1 h. When using a Flow Cell a flow rate of 0.5 mL/min with water is required.

NOTE If the detector is operated in a lab environment that differs at average from the final test environment (25 C) then the detector should be recalibrated for this temperature.

NOTE If the detector was repaired (opened covers), the wavelength calibration can be done 10 minutes after lamp on. A final wavelength calibration should be repeated after complete warm-up of the detector.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 143

8 Test Functions and Calibration Wavelength Calibration

1 Run the Wavelength Calibration with the Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 40 Wavelength Calibration - Results

If you select No, the test is aborted.

If you select Yes, the re-calibration is performed (the offset is corrected).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 144

8 Test Functions and Calibration Wavelength Calibration

Wavelength Recalibration Fails

Probable cause Suggested actions

1 Absorbing solvent or air bubble in Flow Cell. Repeat calibration with Flow Cell and compare results.

2 Dirty or contaminated Flow Cell. Ensure the Flow Cell is filled with water.

Recalibrate.

3 Old UV-lamp. Exchange the UV-lamp.

4 Dirty or contaminated optical components. Run the Cell Test. If the test fails, flush the flow cell.

NOTE If the test fails with Flow Cell and new UV-lamp, the optical unit must be checked/replaced.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 145

8 Test Functions and Calibration D/A Converter (DAC) Test

D/A Converter (DAC) Test

The detector provides analog output of chromatographic signals for use with integrators, chart recorders or data systems. The analog signal is converted from the digital format by the digital-analog-converter (DAC).

The DAC test is used to verify correct operation of the digital-analog-converter by applying a digital test signal to the DAC.

The DAC outputs an analog signal of approximately 50 mV (if the zero offset of the analog output is set to the default value of 5 %) which can be plotted on an integrator. A continuous square wave with an amplitude of 10 V and a frequency of approximately 1 cycle/24 seconds is applied to the signal.

The amplitude of the square wave and the peak-to-peak noise are used to evaluate the DAC test.

1 Run the D/A Converter (DAC) Test with the Agilent Lab Advisor (for further information see Online-Help of user interface).

Figure 41 D/A Converter (DAC) Test Results

When If the analog detector signal is noisy or missing.

Preparations Lamp must be on for at least 10 minutes. Connect integrator, chart recorder or data system to the detector analog output.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 146

8 Test Functions and Calibration D/A Converter (DAC) Test

Test Evaluation The noise on the step should be less than 3 V.

Probable cause Suggested actions

1 Bad cable or grounding problem between

detector and external device.

Check or replace the cable.

2 Defective detector main board. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 147

8 Test Functions and Calibration Dark-Current Test

Dark-Current Test

The dark-current test measures the leakage current from each diode. The test is used to check for leaking diodes which may cause non-linearity at specific wavelengths. During the test, the slit assembly moves to the dark position, cutting off all light falling onto the diode array. Next, the leakage current from each diode is measured, and displayed graphically. The leakage current (represented in counts) for each diode should fall within the limits.

1 Run the Dark Current Test with the recommended user interface (for further information see Online-Help of user interface).

Figure 42 Dark Current Test Results

When In case of problem.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 148

8 Test Functions and Calibration Dark-Current Test

Figure 43 Dark Current Test Signals

Test Failed

Probable cause Suggested actions

1 Defective slit assembly (stray light). Run the Slit Test on page 138 (part of the Self-Test on page 122).

2 Defective detector main board. Please contact your Agilent service representative.

3 Defective PDA/optical unit. Please contact your Agilent service representative.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 149

8 Test Functions and Calibration Spectral Scan

Spectral Scan

The Spectral Scan tool is available for diode-array and variable wavelength detectors (DAD/MWD and VWD). It allows you to scan a spectrum over a specified wavelength range and export the data to a csv (comma-separated values) file that can be used in other applications (for example, Microsoft Excel).

Figure 44 Spectral Scan

Agilent InfinityLab LC Series DAD WR and MWD User Manual 150

8 Test Functions and Calibration Other Lab Advisor Functions

Other Lab Advisor Functions

EMFs - Early Maintenance Feature The EMFs screen allows you to view and manage the EMF counters for all modules in all systems.

Figure 45 Early Maintenance Feature

Agilent InfinityLab LC Series DAD WR and MWD User Manual 151

9 Maintenance

Introduction to Maintenance 153

Cautions and Warnings 154

Overview of Maintenance 156

Cleaning the Module 157

Remove and Install Doors 158

Replace a Lamp 160

Remove and Install a Flow Cell 163

Maintenance of Standard, Semi-Micro or Micro Flow Cell 167

Maintenance of High Pressure Flow Cell 170

Replacing Capillaries on a Standard Flow Cell 172

Replacing Capillaries on a Semi-Micro and Micro Flow Cell 177

Nano Flow Cell - Replacing or Cleaning 181 Wrench plus Torque 184

Cleaning or Exchanging the Holmium Oxide Filter 185

Correcting Leaks 188

Replacing Leak Handling System Parts 190

Replacing the Modules Firmware 192

Information from Modules Assemblies 194

This chapter describes the maintenance of the detector.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 152

9 Maintenance Introduction to Maintenance

Introduction to Maintenance

The module is designed for easy maintenance. Maintenance can be done from the front with module in place in the system.

NOTE There are no serviceable parts inside.

Do not open the module.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 153

9 Maintenance Cautions and Warnings

Cautions and Warnings

WARNING Module is partially energized when switched off, as long as the power cord is plugged in.

Risk of stroke and other personal injury. Repair work at the module can lead to personal injuries, e. g. shock hazard, when the module cover is opened and the instrument is connected to power.

Never perform any adjustment, maintenance or repair of the module with the top cover removed and with the power cord plugged in.

The security lever at the power input socket prevents that the module cover is taken off when line power is still connected. Never plug the power line back in when cover is removed.

WARNING Sharp metal edges

Sharp-edged parts of the equipment may cause injuries.

To prevent personal injury, be careful when getting in contact with sharp metal areas.

WARNING Toxic, flammable and hazardous solvents, samples and reagents

The handling of solvents, samples and reagents can hold health and safety risks.

When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.

The volume of substances should be reduced to the minimum required for the analysis.

Do not operate the instrument in an explosive atmosphere.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 154

9 Maintenance Cautions and Warnings

CAUTION Electronic boards and components are sensitive to electrostatic discharge (ESD).

ESD can damage electronic boards and components.

Be sure to hold the board by the edges, and do not touch the electrical components. Always use ESD protection (for example, an ESD wrist strap) when handling electronic boards and components.

WARNING Eye damage by detector light

Eye damage may result from directly viewing the UV-light produced by the lamp of the optical system used in this product.

Always turn the lamp of the optical system off before removing it.

CAUTION Safety standards for external equipment

If you connect external equipment to the instrument, make sure that you only use accessory units tested and approved according to the safety standards appropriate for the type of external equipment.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 155

9 Maintenance Overview of Maintenance

Overview of Maintenance

The following pages describe maintenance (simple repairs) of the detector that can be carried out without opening the main cover.

Table 15 Overview of Maintenance

Procedure Typical Frequency Notes

Cleaning of module If required.

Deuterium lamp or tungsten lamp exchange

If noise and/or drift exceeds your application limits or lamp does not ignite.

An intensity test should be performed after replacement.

Flow cell exchange If application requires a different flow cell type. A holmium or wavelength calibration test should be performed after replacement.

Flow cell parts Cleaning or exchange

If leaking or if intensity drops due to contaminated flow cell windows.

A pressure tightness test should be done after repair.

Holmium oxide filter Cleaning or exchange

If contaminated. A holmium or wavelength calibration test should be performed after replacement.

Leak sensor drying If leak has occurred. Check for leaks.

Leak handling System replacement

If broken or corroded. Check for leaks.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 156

9 Maintenance Cleaning the Module

Cleaning the Module

The module case should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and mild detergent. Do not use an excessively damp cloth allowing liquid to drip into the module.

WARNING Liquid dripping into the electronic compartment of your module can cause shock hazard and damage the module

Do not use an excessively damp cloth during cleaning.

Drain all solvent lines before opening any connections in the flow path.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 157

9 Maintenance Remove and Install Doors

Remove and Install Doors

Parts required p/n Description

5067-5737 Door left

5067-5736 Door right

NOTE The figures shown in this procedure exemplarily show the Infinity II Multisampler module.

The principle of how to remove and/or install doors works in the same way for all Infinity II modules.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 158

9 Maintenance Remove and Install Doors

1 Press the release buttons and pull the front door out. 2 For the Installation of the front door. Insert the hinges into their guides and move the door in until the release buttons click into their final position.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 159

9 Maintenance Replace a Lamp

Replace a Lamp

When If noise or drift exceeds application limits or lamp does not ignite

Tools required Description

Screwdriver, Pozidriv #1 PT3

Parts required # p/n Description

1 2140-0820 Longlife Deuterium lamp C (with black cover and RFID tag)

1 G1103-60001 Tungsten lamp

Preparations Turn the lamp(s) off.

WARNING Eye damage by detector light

Eye damage may result from directly viewing the light produced by the deuterium lamp used in this product.

Always turn the deuterium lamp off before removing it.

WARNING Injury by touching hot lamp

If the detector has been in use, the lamp may be hot.

If so, wait for lamp to cool down.

NOTE The lamp house cover includes a magnet.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 160

9 Maintenance Replace a Lamp

1 Open the doors. 2 Grab the lamp cover and pull it off (it is fixed by two magnets in the center of the cover).

3 Unplug the lamp connector as required .

4 Unscrew (do not remove) the two lamp screws (Pozidriv) as required.

5 Remove the deuterium lamp and place it on a clean place.

NOTE Do not touch the glass bulb with your fingers. It may reduce the light output.

6 Remove the Vis-lamp and place it on a clean place.

NOTE Do not touch the glass bulb with your fingers. It may reduce the light output.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 161

9 Maintenance Replace a Lamp

7 Insert the deuterium lamp (RFID tag on the top side). 8 Insert the Vis-lamp (flat side to the right).

9 Reconnect the lamp connector as required. 10 Place the lamp cables in the lamp cover.

11 Slide the lamp cover into the top position of the metal front and press the lamp cover completely in until it clicks.

12 Close the doors.

Next Steps:

13 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp (Wavelength Verification Test on page 141, Holmium Oxide Test on page 131).

14 Perform an Intensity Test.

1.

2.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 162

9 Maintenance Remove and Install a Flow Cell

Remove and Install a Flow Cell

For bio-inert modules use bio-inert parts only!

When If an application needs a different type of flow cell or the flow cell needs repair.

Tools required p/n Description

Wrench, 1/4 inch for capillary connections

5043-0915 Fitting mounting tool

Parts required p/n Description

G1315-60022 Standard flow cell, 10 mm, 13 L, 120 bar (12 MPa)

G1315-60025 Semi-micro flow cell, 6 mm, 5 L, 120 bar (12 MPa)

G1315-60024 Micro flow cell, 3 mm, 2 L, 120 bar (12 MPa)

G1315-60015 High pressure flow cell, 6 mm, 1.7 L, 400 bar (40 MPa)

Nano flow cell

G5615-60022 Standard flow cell bio-inert, 10 mm, 13 L, 120 bar (12 MPa) for MWD/DAD, includes 0890-1763 0.18 x 1500 mm PEEK capillary and 5063-6591 PEEK fittings

Preparations Turn the lamp(s) off.

CAUTION Sample degradation and contamination of the instrument

Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination.

For bio applications, always use dedicated bio parts, which can be identified by the bio-inert symbol or other markers described in this manual.

Do not mix bio, and non-bio modules or parts in a bio system.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 163

9 Maintenance Remove and Install a Flow Cell

NOTE The lamp house cover includes a magnet.

1 Open the doors. 2 Grab the lamp cover and pull it off (it is fixed by two magnets in the center of the cover).

3 Open the flow cell door. 4 Disconnect the flow cell inlet capillary (top) from the union.

5 Disconnect the waste tubing (bottom) from the union. 6 Loosen the thumb screw (1.) and remove the flow cell outlet capillary (bottom) with the union (2.).

1. 2.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 164

9 Maintenance Remove and Install a Flow Cell

7 Remove the flow cell while pressing the flow cell holder.

NOTE The label attached to the flow cell provides information on part number, path length, and maximum pressure.

If you want to replace flow cell parts, see Maintenance of Standard, Semi-Micro or Micro Flow Cell on page 167 or Maintenance of High Pressure Flow Cell on page 170.

8 Insert the flow cell while pressing the flow cell holder.

9 Insert the flow cell capillaries into the union holder (top is inlet, bottom is outlet).

10 Tighten the thumb screw.

1.

2. 3.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 165

9 Maintenance Remove and Install a Flow Cell

11 Reconnect the waste tubing (bottom) to the union. Establish a flow and check for leaks.

NOTE To check for leaks, establish a flow and observe the flow cell (outside of the cell compartment) and all capillary connections.

12 Close the flow cell door.

13 Slide the lamp cover into the top position of the metal front and press the lamp cover completely in until it clicks.

14 Close the doors.

15 Perform a Wavelength Verification and Calibration or a Holmium Oxide Test to check the correct positioning of the flow cell.

1.

2.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 166

9 Maintenance Maintenance of Standard, Semi-Micro or Micro Flow Cell

Maintenance of Standard, Semi-Micro or Micro Flow Cell

When If the flow cell needs repair due to leaks or contaminations (reduced light throughput)

Tools required p/n Description

Wrench, 1/4 inch for capillary connections

5043-0915 Fitting mounting tool

Hexagonal key, 4 mm

Toothpick

Parts required Description

For parts, see Standard Flow Cell on page 198, Semi-Micro Flow Cell on page 202, Micro Flow Cell on page 204.

Preparations Turn the flow off. Open the doors of the module. Remove the flow cell, see Remove and Install a Flow Cell on page 163

NOTE The gaskets used in the standard and semi-micro/micro flow cell are different.

CAUTION Sample degradation and contamination of the instrument

Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination.

For bio applications, always use dedicated bio parts, which can be identified by the bio-inert symbol or other markers described in this manual.

Do not mix bio, and non-bio modules or parts in a bio system.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 167

9 Maintenance Maintenance of Standard, Semi-Micro or Micro Flow Cell

1 Use a 4 mm hex key to unscrew the window assembly (1) and remove the gasket (6,7) from the cell body.

NOTE Carefully take one of the gaskets (#6 back or # 7 front) and insert it into the cell body.

Do not mix the gasket #6 and # 7.

Gasket # 7 has the smaller hole and must be on the light entrance side.

Verify that the gasket is positioned flat on the bottom and the light path is not blocked.

If you removed all individual parts from the window assembly refer to Standard Flow Cell on page 198 for the correct orientation of the parts.

2 Use a tooth pick to remove the quartz window from the window assembly.

NOTE If the washers fall out of the window assembly, they must be inserted in the correct order with the PTFE ring to prevent any leaks from the flow cell window.

3 Orientation of Flow Cell Parts.

NOTE Gaskets # 6 and #7 have different hole diameters.

4 Assemble the washers and the window assembly in correct order.

1

1

7

6

1

2 3 4 5 6

7 8

9

10

11

Agilent InfinityLab LC Series DAD WR and MWD User Manual 168

9 Maintenance Maintenance of Standard, Semi-Micro or Micro Flow Cell

5 Correct orientation of spring washers (2) is required. 6 Press the PTFE ring into the window assembly.

7 Press the window assembly onto the new or cleaned quartz window.

8 Insert the window assembly (1) into the cell body.

NOTE Do not mix the gasket #6 and #7 (different hole diameter).

Next Steps:

9 Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.

10 Reconnect the capillaries, see Remove and Install a Flow Cell on page 163.

11 Perform a leak test.

12 Insert the flow cell.

13 Replace the front cover

14 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp (Wavelength Verification Test on page 141, Holmium Oxide Test on page 131).

1 - Window screw 2 - Spring washers 3 - Compression washer 4 - Window holder 5 - Quartz window 6 - Gasket

1 2 3 4 5 6

1

1

7

6

Agilent InfinityLab LC Series DAD WR and MWD User Manual 169

9 Maintenance Maintenance of High Pressure Flow Cell

Maintenance of High Pressure Flow Cell

1 Use a 4 mm hex key to unscrew the window assembly (1) and remove the gasket (2) from the cell body.

If you want to replace the gasket only, reinsert the window assembly into the cell body.

When If the flow cell needs repair due to leaks or contaminations (reduced light throughput)

Tools required p/n Description

Wrench, 1/4 inch for capillary connections

OR 5043-0915 Fitting mounting tool

Hexagonal key, 4 mm

Toothpick

Parts required Description

For parts see High Pressure Flow Cell on page 206

Preparations Turn the flow off. Open the doors of the module. Remove the flow cell, see Remove and Install a Flow Cell on page 163

NOTE All descriptions in this procedure are based on the default orientation of the cell (as it is manufactured). The heat exchanger/capillary and the cell body can be fixed mirror symmetrically to have both capillaries routed to the bottom or to the top (depending on the routing of the capillaries to the column).

1 2

2 1

Agilent InfinityLab LC Series DAD WR and MWD User Manual 170

9 Maintenance Maintenance of High Pressure Flow Cell

2 Use a tooth pick to remove the quartz window from the window assembly.

3 Follow the procedure Maintenance of Standard, Semi-Micro or Micro Flow Cell on page 167 for reassembling.

NOTE If the washers fall out of the window assembly, they must be inserted in the correct order with the PTFE ring to prevent any leaks from the flow cell window.

1 Window assembly (contains items 2, 3, 4, 5 and 6

2 Seal ring

3 Quartz window

4 Compression washer

5 Spring washers

6 Window screw

7 Inlet capillary

8 Outlet capillary

6 5

4 2 3 2

7

8

9

1

Agilent InfinityLab LC Series DAD WR and MWD User Manual 171

9 Maintenance Replacing Capillaries on a Standard Flow Cell

Replacing Capillaries on a Standard Flow Cell

For bio-inert modules use bio-inert parts only!

When If the capillary is blocked

Tools required p/n Description

Wrench, 1/4 inch for capillary connections

5043-0915 Fitting mounting tool

Wrench, 4 mm (for capillary connections)

Screwdriver, Pozidriv #1 PT3

Parts required Description

For parts see Standard Flow Cell on page 198

Preparations Turn the lamp(s) off. Open the doors of the module. Remove the flow cell, see Remove and Install a Flow Cell on page 163.

NOTE All descriptions in this procedure are based on the default orientation of the cell (as it is manufactured). The heat exchanger/capillary and the cell body can be fixed mirror symmetrically to have both capillaries routed to the bottom or to the top (depending on the routing of the capillaries to the column).

NOTE The fittings at the flow cell body are special types for low dead volumes and not compatible with other fittings.

When retightening the fittings, make sure that they are carefully tightened (handtight plus 1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 172

9 Maintenance Replacing Capillaries on a Standard Flow Cell

CAUTION Sample degradation and contamination of the instrument

Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination.

For bio applications, always use dedicated bio parts, which can be identified by the bio-inert symbol or other markers described in this manual.

Do not mix bio, and non-bio modules or parts in a bio system.

1 Identify the inlet and outlet capillaries. To replace the inlet capillary, continue with step "To replace the inlet capillary, use a 4 mm wrench for the fitting."

2 Remove the outlet capillary.

3 To replace the inlet capillary, use a 4 mm wrench to open the fitting.

4 Unscrew the cell body from the heat exchanger.

Inlet

Outlet

Agilent InfinityLab LC Series DAD WR and MWD User Manual 173

9 Maintenance Replacing Capillaries on a Standard Flow Cell

5 Remove the heat exchanger from the clamp unit. 6 Use a small flat screw driver to carefully lift off the I.D. tag. Shown is the default orientation. See Note at the beginning of this section.

7 Unscrew the fixing screw and unwrap the inlet capillary from the grove in the flow cell body.

8 Take the new inlet capillary and bend it 90 about 35 mm from its end.

9 Bend the capillary again by 90 as shown below. 10 Insert the capillary into the hole between fixing screw and the inlet fitting.

3 5

cm

90

90

Agilent InfinityLab LC Series DAD WR and MWD User Manual 174

9 Maintenance Replacing Capillaries on a Standard Flow Cell

11 The capillary lays in the grove and should be tied around the body (in the grove) 5 times.

12 Insert the fixing screw, so that the capillary cannot leave the grove.

13 Carefully insert the I.D. tag into the new heat exchanger. Shown is the default orientation. See Note at the beginning of this section.

14 Fix the heat exchanger to the clamp unit.

15 Fix the flow cell body to the heat exchanger. 16 Fix the inlet capillary to the flow cell body handtight first. Then do a 1/4 turn with a 4 mm wrench.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 175

9 Maintenance Replacing Capillaries on a Standard Flow Cell

17 After replacing the outlet capillary, fix it handtight first. Then do a 1/4 turn with a 4 mm wrench.

18 Check for a centered holder vs. hole. If required adjust with the holder screws.

Next Steps:

19 Reconnect the capillaries, see Remove and Install a Flow Cell on page 163.

20 Perform a leak test.

21 Insert the flow cell.

22 Close the doors of the module.

23 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp (Wavelength Verification Test on page 141, Holmium Oxide Test on page 131).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 176

9 Maintenance Replacing Capillaries on a Semi-Micro and Micro Flow Cell

Replacing Capillaries on a Semi-Micro and Micro Flow Cell

When If the capillary is blocked

Tools required p/n Description

Wrench, 1/4 inch for capillary connections

OR 5043-0915 Fitting mounting tool

Wrench, 4 mm (for capillary connections)

Screwdriver, Pozidriv #1 PT3

Parts required Description

For parts see Semi-Micro Flow Cell on page 202 or Micro Flow Cell on page 204.

Preparations Turn the lamp(s) off. Open the doors of the module. Remove the flow cell, see Remove and Install a Flow Cell on page 163.

NOTE All descriptions in this procedure are based on the default orientation of the cell (as it is manufactured). The heat exchanger/capillary and the cell body can be fixed mirror symmetrically to have both capillaries routed to the bottom or to the top (depending on the routing of the capillaries to the column).

NOTE The fittings at the flow cell body are special types for low dead volumes and not compatible with other fittings. When retightening the fittings, make sure that they are carefully tightened (handtight plus 1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 177

9 Maintenance Replacing Capillaries on a Semi-Micro and Micro Flow Cell

1 Identify the inlet and outlet capillaries. 2 Remove the outlet capillary.

3 To replace the inlet capillary, use a 4 mm wrench to open the fitting.

4 Unscrew the cell body from the heat exchanger.

5 Remove the heat exchanger from the clamp unit. 6 Use a small flat screw driver to carefully lift off the I.D. tag. Shown is the default orientation. See Note at the beginning of this section.

Inlet

Outlet

Agilent InfinityLab LC Series DAD WR and MWD User Manual 178

9 Maintenance Replacing Capillaries on a Semi-Micro and Micro Flow Cell

7 Carefully insert the I.D. tag into the new heat exchanger. Shown is the default orientation. See Note at the beginning of this section.

8 Fix the heat exchanger to the clamp unit.

9 Fix the flow cell body to the heat exchanger. 10 Fix the inlet capillary to the flow cell body handtight first. Then do a 1/4 turn with a 4 mm wrench.

11 After replacing the outlet capillary, fix it handtight first. Then do a 1/4 turn with a 4 mm wrench.

12 Check for a centered holder vs. hole. If required adjust with the holder screws.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 179

9 Maintenance Replacing Capillaries on a Semi-Micro and Micro Flow Cell

Next Steps:

13 Reconnect the capillaries, see Remove and Install a Flow Cell on page 163.

14 Perform a leak test.

15 Insert the flow cell.

16 Close the doors.

17 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp (Wavelength Verification Test on page 141, Holmium Oxide Test on page 131).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 180

9 Maintenance Nano Flow Cell - Replacing or Cleaning

Nano Flow Cell - Replacing or Cleaning

When If parts are contaminated or leaky.

Tools required Description

Screwdriver, Pozidriv #1 PT3

Wrench, 1/4 inch for capillary connections

Parts required Description

For parts identification refer to Nano Flow Cells on page 212 (80 nL and 500 nL).

Preparations Turn the lamp(s) off. Open the doors of the module. Remove the flow cell, see Remove and Install a Flow Cell on page 163.

NOTE For details refer to the technical note that comes with the nano-flow cell kit.

NOTE The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet windows at the quartz block.

1 Disconnect the capillaries from the capillary holder and remove the flow cell, see Remove and Install a Flow Cell on page 163.

2 Unscrew the cell body from the holder.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 181

9 Maintenance Nano Flow Cell - Replacing or Cleaning

3 Unscrew the capillaries from the flow cell. DO NOT use the adapter at this time!

4 Using for example a toothpick, press on the plastic part and slide the quartz body out of the cell housing.

5 The quartz body and the cell seal assembly can be separated for cleaning purpose.

6 This figure shows the correct holding of the quartz body and the cell seal assembly.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 182

9 Maintenance Nano Flow Cell - Replacing or Cleaning

7 Replace the cell seal assembly onto the quartz body. Always use a new seal assembly to exclude damage during disassembling.

8 Slide the quartz body completely into the cell body to the front stop (use for example a toothpick).

9 Insert the flow cell capillaries and tighten them fingertight. Use the wrench and torque adapter as described on Figure 46 on page 184 and tighten the fittings alternately.

10 Reassemble the flow cell body to the holder.

Next Steps:

11 Install the flow cell, see Remove and Install a Flow Cell on page 163.

12 Perform a leak test with the flow cell outside of the detector.

13 If no leak is observed, install the flow cell and you are ready to work.

14 Make sure that the flow cell assembly is inserted correctly and fits perfectly in the optical unit (especially when PEEK capillaries are used).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 183

9 Maintenance Nano Flow Cell - Replacing or Cleaning

Wrench plus Torque

Figure 46 Wrench plus Torque Adapter

NOTE The cell body can be fitted in two positions to allow the capillaries routed upwards or downwards (depending on where the column is located). Route the capillaries directly column (inlet) and waste assembly (outlet).

NOTE With the instrument accessory kit comes a 4-mm wrench and with the Sealing Kit a special adapter. Both together work as a torque wrench with pre-defined torque (maximum allowed torque for the cell fittings is 0.7 Nm). It can be used to tight the capillary fittings at the flow cell body. The wrench has to be plugged into the adapter as shown in Figure 46 on page 184.

Wrench

Adapter

DO NOT press down more than shown here

max 0.7 Nm

Agilent InfinityLab LC Series DAD WR and MWD User Manual 184

9 Maintenance Cleaning or Exchanging the Holmium Oxide Filter

Cleaning or Exchanging the Holmium Oxide Filter

When If holmium oxide filter is contaminated

Tools required Description

Screwdriver, Pozidriv #1 PT3

Screwdriver, flat blade

Wrench, 1/4 inch for capillary connections

Pair of tweezers

Parts required # p/n Description

1 79880-22711 Holmium oxide filter

Preparations Turn the lamp(s) off. Open the doors of the module. Remove the flow cell, see Remove and Install a Flow Cell on page 163.

NOTE See also Declaration of Conformity for HOX2 Filter on page 276.

The glass tends to build a film on its surface even under normal environmental conditions. This is a phenomenon, which can be found also on the surface of several other glasses and has something to do with the composition of the glass. There is no indication, that the film has an influence on the measurement. Even in the case of a thick film, which scatters the light remarkably, no shift of the peak positions is to be expected. A slight change in the absorbance might be possible. Other components within the light path (lenses, windows, ...) are also changing their behavior over the time.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 185

9 Maintenance Cleaning or Exchanging the Holmium Oxide Filter

1 Unscrew the six screws and remove the flow cell cover. 2 If not already in this position, move the filter up.

3 Release the holder with a screw driver (at the top).

NOTE Do not scratch the holmium oxide filter.

4 Remove the holmium oxide filter.

NOTE The holmium oxide filter can be cleaned with alcohol and a lint-free cloth.

5 Insert the holmium oxide filter. 6 Replace the flow cell cover and fix the six screws.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 186

9 Maintenance Cleaning or Exchanging the Holmium Oxide Filter

7 Perform a holmium oxide test (see Holmium Oxide Test on page 131) to check the proper function of the holmium oxide filter.

8 Install the flow cell, see Remove and Install a Flow Cell on page 163.

9 Close the doors.

10 Turn on the flow.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 187

9 Maintenance Correcting Leaks

Correcting Leaks

When If a leakage has occurred in the flow cell area or at the heat exchanger or at the capillary connections

Tools required p/n Description

Tissue

Wrench, 1/4 inch for capillary connections

5043-0915 Fitting mounting tool

NOTE Use tissue to dry the leak sensor area and the leak pan.

1 Open the doors. 2 Grab the lamp cover and pull it off (it is fixed by two magnets in the center of the cover).

Agilent InfinityLab LC Series DAD WR and MWD User Manual 188

9 Maintenance Correcting Leaks

3 Open the flow cell door. 4 Observe the capillary connections and the flow cell area for leaks and correct, if required.

NOTE To check for leaks, establish a flow and observe the flow cell (outside of the cell compartment) and all capillary connections.

5 Check the Leak Sensor area for leaks and correct, if required.

6 Install the flow cell and close the flow cell door.

7 Slide the lamp cover into the top position of the metal front and press the lamp cover completely in until it clicks.

8 Close the doors.

9 Perform a Wavelength Verification and Calibration or a Holmium Oxide Test to check the correct positioning of the flow cell.

1.

2.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 189

9 Maintenance Replacing Leak Handling System Parts

Replacing Leak Handling System Parts

When If the parts are corroded or broken

Parts required p/n Description

5043-0856 Leak Adapter

5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm

1 Open the doors. 2 Press the Leak Adapter down (1.) and remove it together with the tubing (2.).

3 Install the Leak Adapter by pressing it into the Main Cover.

4 Insert the Tubing (approximately 85 mm required for replacement) between Leak Adapter outlet and Leak Panel.

1. 2.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 190

9 Maintenance Replacing Leak Handling System Parts

5 Close the doors.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 191

9 Maintenance Replacing the Modules Firmware

Replacing the Modules Firmware

To upgrade/downgrade the modules firmware carry out the following steps:

1 Download the required module firmware, the latest FW Update Tool and the documentation from the Agilent web. http://www.agilent.com/en-us/firmwareDownload?whid=69761

2 For loading the firmware into the module follow the instructions in the documentation.

When The installation of newer firmware might be necessary if a newer version solves problems of older versions or to keep all systems on the same (validated) revision.

The installation of older firmware might be necessary to keep all systems on the same (validated) revision or if a new module with newer firmware is added to a system or if third party control software requires a special version.

Tools required Description

Agilent Lab Advisor software

Parts required # Description

1 Firmware, tools and documentation from Agilent web site

Preparations Read update documentation provided with the Firmware Update Tool.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 192

9 Maintenance Replacing the Modules Firmware

Module Specific Information

Table 16 Module Specific Information (G7115A DAD/G7165A MWD)

G7115A DAD G7165A MWD

Initial firmware (main and resident)

D.07.01 D.07.01

Compatibility with 1260/1290 Infinity modules

When using the G7115A or G7165A in a system, all other modules must have firmware revision from firmware set 7.00 or above (main and resident).

Compatibility with 1100/1200 Series modules

When using the G7115A or G7165A in a system, all other modules must have firmware revision from firmware set 7.00 or above (main and resident). Otherwise the communication will not work.

Conversion to / emulation of G1315D or G1365D

G7115A allows conversion to G1315C/D DAD, G1365C/D MWD G7165A allows conversion to G1365C/D MWD

Agilent InfinityLab LC Series DAD WR and MWD User Manual 193

9 Maintenance Information from Modules Assemblies

Information from Modules Assemblies

Lamp and Flow Cell RFID Tag The detector is equipped with a UV lamp and flow cell identification system using RFID (radio frequency identification) tags attached to the assemblies and RFID tag readers at the optical unit. The table below lists all parameters stored in the RFID tag.

Serial Number and Firmware Revision The user interface provides module specific information that is stored in the main board. These are for example the serial number, firmware revision.

Table 17 RFID Tag Data

Lamp information Flow cell information

product number product number

serial number serial number

production date production date

accumulated UV on time (in hours) nominal path length of the cell (in mm)

actual UV lamp on time (in hours) cell volume () in L

number of ignitions maximum pressure (in bar)

date of last intensity test date of last cell test

NOTE The pressure value is always displayed in bar, even if the user interface uses other units, e.g. PSI.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 194

10 Parts for Maintenance

Overview of Maintenance Parts 196

Standard Flow Cell 198

Standard Flow Cell Bio-inert 200

Semi-Micro Flow Cell 202

Micro Flow Cell 204

High Pressure Flow Cell 206

Prep Flow Cell - SST 208

Prep Flow Cell - Quartz 210

Nano Flow Cells 212

Accessory Kits 215

Holmium Oxide Filter 216

Leak Handling Parts 217

This chapter provides information on parts for maintenance and repair.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 195

10 Parts for Maintenance Overview of Maintenance Parts

Overview of Maintenance Parts

Item p/n Description

1 5067-5737 Door left

2 5067-5736 Door right

3 G1103-60001 Tungsten lamp

4 2140-0820 Longlife Deuterium lamp C (with black cover and RFID tag)

5 5043-1013 Tubing Clip

6 5043-0856 Leak Adapter

7 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm for Waste and Leak Adapter (ca. 85 mm required)

8 Flow cell with RFID tag for details refer to specific flow cell

G1315-60022 Standard flow cell, 10 mm, 13 L, 120 bar (12 MPa)

G5615-60022 Standard flow cell bio-inert, 10 mm, 13 L, 120 bar (12 MPa) for MWD/DAD, includes 0890-1763 0.18 x 1500 mm PEEK capillary and 5063-6591 PEEK fittings

G1315-60025 Semi-micro flow cell, 6 mm, 5 L, 120 bar (12 MPa)

G1315-60024 Micro flow cell, 3 mm, 2 L, 120 bar (12 MPa)

G1315-60015 High pressure flow cell, 6 mm, 1.7 L, 400 bar (40 MPa)

G1315-60016 Prep flow cell SST - 3 mm, 120 bar (12 MPa)

G1315-60017 Prep flow cell quartz, 0.3 mm, 20 bar (2 MPa)

G1315-60018 Prep flow cell quartz, 0.06 mm (2 MPa)

G1315-68724 Semi-nano flow cell kit, 10 mm, 500 nL, 5 MPa

G1315-68716 Nano flow cell kit, 6 mm, 80 nL, 5 MPa

9 5022-6515 Union ZDV

10 Tube PTFE 0.8 mm x 2 m, re-order 5 m

11 G1315-87311 Capillary ST 0.17 mm x 380 mm S/S

79880-22711 Holmium oxide filter (not shown)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 196

10 Parts for Maintenance Overview of Maintenance Parts

Figure 47 Overview of maintenance parts

1

2

3

4

5

6

7 8

9

10

11

Agilent InfinityLab LC Series DAD WR and MWD User Manual 197

10 Parts for Maintenance Standard Flow Cell

Standard Flow Cell

Item p/n Description

G1315-60022 Standard flow cell, 10 mm, 13 L, 120 bar (12 MPa)

1 79883-22402 Window screw

2 5062-8553 Washer kit (10/pk)

3 79883-28801 Compression washer

4 79883-22301 Window holder

5 1000-0488 Quartz window

6 G1315-68711 Gasket BACK (PTFE), 2.3 mm hole, outlet side (12/pk)

7 G1315-68710 Gasket FRONT (PTFE), 1.3 mm hole, inlet side (12/pk)

8 Window assembly (comprises window screw, spring washers, compression washer, window holder and quartz window)

9 G1315-87321 Capillary IN (0.17 mm, 590 mm lg) including heat exchanger

10 G1315-87302 Capillary OUT (0.17 mm, 200 mm lg)

11 G1315-84910 Clamp unit

0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

5022-2184 Union ZDV

G1315-68712 Cell repair kit STD includes window screw kit, 4 mm hexagonal wrench and seal kit

79883-68703 Window screw kit, includes 2 quartz windows, 2 compression washers, 2 window holders, 2 window screws and 10 washers

Agilent InfinityLab LC Series DAD WR and MWD User Manual 198

10 Parts for Maintenance Standard Flow Cell

Figure 48 Standard Flow Cell Parts

Figure 49 Orientation of Spring Washers

1

2 3 4

5 6

7 8

9

10

11

NOTE Gaskets # 6 and #7 have different hole diameters.

1 - Window screw 2 - Spring washers 3 - Compression washer 4 - Window holder 5 - Quartz window 6 - Gasket

1 2 3 4 5 6

Agilent InfinityLab LC Series DAD WR and MWD User Manual 199

10 Parts for Maintenance Standard Flow Cell Bio-inert

Standard Flow Cell Bio-inert

Item p/n Description

G5615-60022 Standard flow cell bio-inert, 10 mm, 13 L, 120 bar (12 MPa) for MWD/DAD, includes 0890-1763 0.18 x 1500 mm PEEK capillary and 5063-6591 PEEK fittings

1 79883-22402 Window screw

2 5062-8553 Washer kit (10/pk)

3 79883-28801 Compression washer

4 79883-22301 Window holder

5 5190-0921 Sapphire window

6 G1315-68711 Gasket BACK (PTFE), 2.3 mm hole, outlet side (12/pk)

7 G1315-68710 Gasket FRONT (PTFE), 1.3 mm hole, inlet side (12/pk)

8 Window assembly (comprises window screw, spring washers, compression washer, window holder and sapphire window)

9 G5615-87331 Capillary In (0.17 mm, 590 mm lg), including heat exchanger)

10 G5615-87302 Capillary Out (0.17 mm, 200 mm lg)

11 G1315-84910 Clamp unit

0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

5022-2184 Union ZDV

G1315-68712 Cell repair kit STD includes window screw kit, 4 mm hexagonal wrench and seal kit

G5615-68712 Cell repair kit for bio-inert cell Contains: 1 x HEX KEY 4MM, 1 x Front seals 12/PK 1 x Back seals-std cell 12/PK, 1 x Air cushion envelope A5, 1 x ST Cell Screw Kit bio-inert

79883-68704 Cell Screw Kit bio-inert Contains: 2 x Sapphire Window, 2 x Cell screw assembly, 1 x Container

5067-5695 UHP-FF Fitting

Agilent InfinityLab LC Series DAD WR and MWD User Manual 200

10 Parts for Maintenance Standard Flow Cell Bio-inert

Figure 50 Standard Flow Cell Bio-inert

Figure 51 Orientation of Spring Washers

1

2

3 4 5

6

7 8

9

10

11

1 - Window screw 2 - Spring washers 3 - Compression washer 4 - Window holder 5 - Sapphire window 6 - Gasket

1 2 3 4 5 6

Agilent InfinityLab LC Series DAD WR and MWD User Manual 201

10 Parts for Maintenance Semi-Micro Flow Cell

Semi-Micro Flow Cell

Item p/n Description

G1315-60025 Semi-micro flow cell, 6 mm, 5 L, 120 bar (12 MPa)

1 79883-22402 Window screw

2 5062-8553 Washer kit (10/pk)

3 79883-28801 Compression washer

4 79883-22301 Window holder

5 1000-0488 Quartz window

6 79883-68702 Gasket BACK (PTFE), 1.8 mm hole, outlet side (12/pk)

7 G1315-68710 Gasket FRONT (PTFE), 1.3 mm hole, inlet side (12/pk)

8 Window assembly (comprises window screw, spring washers, compression washer, window holder and quartz window)

9 G1315-87319 Capillary IN (0.17 mm, 310 mm lg) including heat exchanger

10 G1315-87306 Capillary OUT (0.12 mm, 200 mm lg)

10 G1315-87302 Capillary OUT (0.17 mm, 200 mm lg)

11 G1315-84910 Clamp unit

0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

5022-2184 Union ZDV

G1315-68713 Cell repair kit semi-micro, includes window screw kit, Gasket Kit BACK, Gasket Kit FRONT and 4 mm hexagonal wrench

79883-68703 Window screw kit, includes 2 quartz windows, 2 compression washers, 2 window holders, 2 window screws and 10 washers

Agilent InfinityLab LC Series DAD WR and MWD User Manual 202

10 Parts for Maintenance Semi-Micro Flow Cell

Figure 52 Semi-Micro Flow Cell Parts

Figure 53 Orientation of Spring Washers

1

2 3 4

5 6

7 8

9

10

11

NOTE Gaskets # 6 and #7 have different hole diameters.

1 - Window screw 2 - Spring washers 3 - Compression washer 4 - Window holder 5 - Quartz window 6 - Gasket

1 2 3 4 5 6

Agilent InfinityLab LC Series DAD WR and MWD User Manual 203

10 Parts for Maintenance Micro Flow Cell

Micro Flow Cell

Item p/n Description

G1315-60024 Micro flow cell, 3 mm, 2 L, 120 bar (12 MPa)

1 79883-22402 Window screw

2 5062-8553 Washer kit (10/pk)

3 79883-28801 Compression washer

4 79883-22301 Window holder

5 1000-0488 Quartz window

6 79883-68702 Gasket BACK (PTFE), 1.8 mm hole, outlet side (12/pk)

7 G1315-68710 Gasket FRONT (PTFE), 1.3 mm hole, inlet side (12/pk)

8 Window assembly (comprises window screw, spring washers, compression washer, window holder and quartz window)

9 G1315-87339 DAD Heat Exchanger Capillary 310 mm, 0.12 mm i.d.

10 G1315-87306 Capillary OUT (0.12 mm, 200 mm lg)

10 G1315-87302 Capillary OUT (0.17 mm, 200 mm lg)

11 G1315-84910 Clamp unit

0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

5022-2184 Union ZDV

G1315-68713 Cell repair kit semi-micro, includes window screw kit, Gasket Kit BACK, Gasket Kit FRONT and 4 mm hexagonal wrench

79883-68703 Window screw kit, includes 2 quartz windows, 2 compression washers, 2 window holders, 2 window screws and 10 washers

Agilent InfinityLab LC Series DAD WR and MWD User Manual 204

10 Parts for Maintenance Micro Flow Cell

Figure 54 Micro Flow Cell Parts

Figure 55 Orientation of Spring Washers

1

2 3 4

5 6

7 8

9

10

11

NOTE Gaskets # 6 and #7 have different hole diameters.

1 - Window screw 2 - Spring washers 3 - Compression washer 4 - Window holder 5 - Quartz window 6 - Gasket

1 2 3 4 5 6

Agilent InfinityLab LC Series DAD WR and MWD User Manual 205

10 Parts for Maintenance High Pressure Flow Cell

High Pressure Flow Cell

Item p/n Description

G1315-60015 High pressure flow cell, 6 mm, 1.7 L, 400 bar (40 MPa)

1 Window assembly, comprises items 2, 3, 4, 5 and 6

2 79883-27101 Seal ring

3 1000-0953 Quartz window

4 79883-28802 Compression washer

5 5062-8553 Washer kit (10/pk)

6 79883-22404 Window screw

7 G1315-87325 Capillary IN (0.12 mm, 290 mm lg) including heat exchanger

8 G1315-87306 Capillary OUT (0.12 mm, 200 mm lg)

9 G1315-84901 Clamp unit

0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

G1315-87312 Capillary ST 0.12 mm x 150 mm S/S

G1315-87311 Capillary ST 0.17 mm x 380 mm S/S

79883-68700 High pressure cell repair kit (includes 1 quartz window, 1 compression washer, 5 spring washers, 2 seal rings)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 206

10 Parts for Maintenance High Pressure Flow Cell

Figure 56 High pressure flow cell - parts

6 5

4 2 3 2

7

8

9

1

Agilent InfinityLab LC Series DAD WR and MWD User Manual 207

10 Parts for Maintenance Prep Flow Cell - SST

Prep Flow Cell - SST

NOTE For more details on the Preparative Flow Cells refer to the technical note that comes with the flow cells.

Item p/n Description

G1315-60016 Prep flow cell SST - 3 mm, 120 bar (12 MPa) recommended flow rate: 50 mL/min

1 G1315-60021 Cell screw assembly (comprises window screw, spring washers, compression washer, window holder and quartz window)

G1315-68712 Cell repair kit STD includes window screw kit, 4 mm hexagonal wrench and seal kit

2 G1315-68711 Gasket BACK (PTFE), 2.3 mm hole, outlet side (12/pk)

3 G1315-68710 Gasket FRONT (PTFE), 1.3 mm hole, inlet side (12/pk)

4 G1315-87305 Capillary SST, 250 mm length, 0.5 mm i.d., o.D. 0.9 mm with fittings for flow cell assembled

4a 5062-2418 1/16 fittings and ferrules 10/pk

5 G1315-27706 Cell body

6 G1315-84901 Clamp unit

7 G1315-84902 Handle for Clamp unit

0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

Agilent InfinityLab LC Series DAD WR and MWD User Manual 208

10 Parts for Maintenance Prep Flow Cell - SST

Figure 57 Prep Flow Cell - SST Parts

Figure 58 Orientation of Spring Washers

1

1 2

3

4 - inlet 4 - outlet

4a

5

6

7

NOTE Gaskets #2 and #3 have different hole diameters.

1 - Window screw 2 - Spring washers 3 - Compression washer 4 - Window holder 5 - Quartz window 6 - Gasket

1 2 3 4 5 6

Agilent InfinityLab LC Series DAD WR and MWD User Manual 209

10 Parts for Maintenance Prep Flow Cell - Quartz

Prep Flow Cell - Quartz

NOTE For more details on the Preparative Flow Cells refer to the technical note that comes with the flow cells.

Item p/n Description

G1315-60017 Prep flow cell quartz, 0.3 mm, 20 bar (2 MPa)

G1315-60018 Prep flow cell quartz, 0.06 mm (2 MPa)

1 0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

2 G1315-84902 Handle for Clamp unit

3 G1315-84901 Clamp unit

4 G1315-80004 Quartz body - Prep Cell 0.3 mm

4 G1315-80003 Quartz body - Prep Cell 0.06 mm

5 G1315-67302 PTFE tubing 80 cm length, 0.5 mm i.d., o.D. 1.6 mm

6 G1315-27705 Cell housing

7 G1315-67301 PTFE tubing 2 m length, 0.8 mm i.d., o.D. 1.6 mm

0100-1516 Finger-tight fitting PEEK, 2/pk

NOTE The flow cell comes with two tubings 0.8 mm i.d. and one 0.5 mm i.d. so that the combination at the flow cell could be either 0.8/0.8 or 0.5/0.8 (inlet/outlet).

Standard is 0.8/0.8. Depending on the system pressure (<30 mL/min) or bandbroadening, the inlet tubing might be changed to 0.5 mm.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 210

10 Parts for Maintenance Prep Flow Cell - Quartz

Figure 59 Prep Flow Cell - Quartz Parts

1

2 3

4

5

6

7

Agilent InfinityLab LC Series DAD WR and MWD User Manual 211

10 Parts for Maintenance Nano Flow Cells

Nano Flow Cells

The following kits are available:

Figure 60 on page 213 shows all parts delivered with the nano-flow cell kits.

Generic parts for both nano-flow cells:

Table 18 Nano-flow cell kits

Part number Comments

Semi-nano flow cell kit, 10 mm, 500 nL, 5 MPa (G1315-68724)

completely assembled (includes items 1, 2, 3, 4, 10, 11, 12, 13, 14, 15, and 16)

Nano flow cell kit, 6 mm, 80 nL, 5 MPa ( G1315-68716)

completely assembled (includes items 1, 2, 3, 4, 10, 11, 12, 13, 14, 15, and 16)

Item p/n Description

3 5063-6593 Fitting Screw (for 4 mm wrench)

4 Cell ferrules are factory installed

5 5065-4422 PEEK fitting 1/32"

7 5063-6592 Litetouch ferrules LT-100, (1/32" Ferrule and SS lock ring)

8 5022-2146 Union Adjustment Tool

9 5022-2184 Union ZDV

10 G1315-45003 Torque adapter

14 G1315-84902 Handle for Clamp unit

15 G1315-84910 Clamp unit

16 0515-1056 Screw M 2.5, 4 mm lg for cell body/clamp

17 8710-1534 Wrench, 4 mm both ends, open end

Agilent InfinityLab LC Series DAD WR and MWD User Manual 212

10 Parts for Maintenance Nano Flow Cells

Figure 60 Content of kits

Specific parts for the semi-nano flow cell

Item p/n Description

G1315-68724 Semi-nano flow cell kit, 10 mm, 500 nL, 5 MPa

1 G1315-87333 PEEK coated fused silica capillary Inlet (100 m) pre-mounted to cell, includes Inlet capillary, 300 mm long, 100 m i.d. with pre-fixed ferrules (#4) and fittings (#3), plus one PEEK Fitting FT (#5)

2 G1315-87338 PEEK coated fused silica capillary Outlet (100 m) pre-mounted to cell, includes Outlet capillary, 120 mm long, 100 m i.d. with pre-fixed ferrules (#4) and fitting (#3), plus one PEEK Fitting FT (#5)

1 G1315-87323 PEEK coated fused silica capillary Inlet (50 m) alternative, includes Inlet capillary, 400 mm long, 50 m i.d. with pre-fixed ferrules (#4) and fittings (#3), plus one PEEK Fitting FT (#5)

2 G1315-87328 PEEK coated fused silica capillary Outlet (50 m), alternative, includes Outlet capillary, 120 mm long, 50 m i.d. with pre-fixed ferrules (#4) and fitting (#3), plus one PEEK Fitting FT (#5)

11 G1315-27703 Cell Housing (500 nL)

12 G1315-87101 Cell Seal Assembly (500 nL)

13 G1315-80001 Quartz Body (500 nL)

G1315-68715 Sealing Kit

17

2 3, 43, 4

7

8 9

10

1112

13

14

16 15

5

1

Agilent InfinityLab LC Series DAD WR and MWD User Manual 213

10 Parts for Maintenance Nano Flow Cells

Specific parts for the nano flow cell

Item p/n Description

G1315-68716 Nano flow cell kit, 6 mm, 80 nL, 5 MPa

1 G1315-87323 PEEK coated fused silica capillary Inlet (50 m) alternative, includes Inlet capillary, 400 mm long, 50 m i.d. with pre-fixed ferrules (#4) and fittings (#3), plus one PEEK Fitting FT (#5)

2 G1315-87328 PEEK coated fused silica capillary Outlet (50 m), alternative, includes Outlet capillary, 120 mm long, 50 m i.d. with pre-fixed ferrules (#4) and fitting (#3), plus one PEEK Fitting FT (#5)

1 G1315-87313 PEEK coated fused silica capillary Inlet (25 m) alternative, includes Inlet capillary, 200 mm long, 25 m i.d. with pre-fixed ferrules (#4) and fittings (#3), plus one PEEK Fitting FT (#5)

2 G1315-87318 PEEK coated fused silica capillary Outlet (25 m) alternative, includes Outlet capillary, 600 mm long, 25 m i.d. with pre-fixed ferrules (#4) and fitting (#3), plus one PEEK Fitting FT (#5)

G1315-27704 Cell Housing (80 nL)

G1315-87102 Cell Seal Assembly (80 nL)

G1315-80002 Quartz Body (80 nL)

G1315-68725 Sealing Kit 80 nL cell

Agilent InfinityLab LC Series DAD WR and MWD User Manual 214

10 Parts for Maintenance Accessory Kits

Accessory Kits

Detector Accessory Kit (G7115-68755) contains the following items:

p/n Description

5062-8535 Waste accessory kit

5500-1155 Tube Connector, 90 degree, ID 6.4

0100-1516 Finger-tight fitting PEEK, 2/pk

5181-1516 CAN cable, Agilent module to module, 0.5 m

5500-1191 InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket

Agilent InfinityLab LC Series DAD WR and MWD User Manual 215

10 Parts for Maintenance Holmium Oxide Filter

Holmium Oxide Filter

Figure 61 Holmium Oxide Filter Parts

Item p/n Description

1 G7115-68700 Filter motor assembly (includes filter lever G1315-45001 and spring 1460-1510)

2 1460-1510 Spring

3 G1315-45001 Filter lever

4 79880-22711 Holmium oxide filter

NOTE When the filter motor has been removed, the filter lever should not be reused. Use always a new filter lever to assure correct fit on the filter motor shaft.

1

2

3 4

Agilent InfinityLab LC Series DAD WR and MWD User Manual 216

10 Parts for Maintenance Leak Handling Parts

Leak Handling Parts

Figure 62 Leak Parts

Item p/n Description

1 5043-0856 Leak Adapter

2 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm

3 5061-3356 Leak Sensor Assembly

4 G7115-45000 Leak Plane 1260 DAD Infinity II

0515-2529 Screw Tapping PAN-HD-TORX T10 3x8 ST-ZN (not shown)

5043-1013 Tubing Clip

1

2

3

4

Agilent InfinityLab LC Series DAD WR and MWD User Manual 217

11 Identifying Cables

Cable Overview 219

Analog Cables 221

Remote Cables 223

BCD Cables 226

CAN/LAN Cables 228

Agilent 1200 module to PC 229

This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 218

11 Identifying Cables Cable Overview

Cable Overview

Analog cables

Remote cables

BCD cables

NOTE Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.

p/n Description

35900-60750 Agilent module to 3394/6 integrators

35900-60750 Agilent 35900A A/D converter

01046-60105 Analog cable (BNC to general purpose, spade lugs)

p/n Description

03394-60600 Agilent module to 3396A Series I integrators

3396 Series II / 3395A integrator, see details in section Remote Cables on page 223

03396-61010 Agilent module to 3396 Series III / 3395B integrators

5061-3378 Remote Cable

01046-60201 Agilent module to general purpose

p/n Description

03396-60560 Agilent module to 3396 integrators

G1351-81600 Agilent module to general purpose

Agilent InfinityLab LC Series DAD WR and MWD User Manual 219

11 Identifying Cables Cable Overview

CAN cables

LAN cables

RS-232 cables

p/n Description

5181-1516 CAN cable, Agilent module to module, 0.5 m

5181-1519 CAN cable, Agilent module to module, 1 m

p/n Description

5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)

5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)

p/n Description

RS232-61601 RS-232 cable, 2.5 m Instrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It is also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.

5181-1561 RS-232 cable, 8 m

Agilent InfinityLab LC Series DAD WR and MWD User Manual 220

11 Identifying Cables Analog Cables

Analog Cables

One end of these cables provides a BNC connector to be connected to Agilent modules. The other end depends on the instrument to which connection is being made.

Agilent Module to 3394/6 Integrators

p/n 35900-60750 Pin 3394/6 Pin Agilent module

Signal Name

1 Not connected

2 Shield Analog -

3 Center Analog +

Agilent InfinityLab LC Series DAD WR and MWD User Manual 221

11 Identifying Cables Analog Cables

Agilent Module to BNC Connector

Agilent Module to General Purpose

p/n 8120-1840 Pin BNC Pin Agilent module

Signal Name

Shield Shield Analog -

Center Center Analog +

p/n 01046-60105 Pin Pin Agilent module

Signal Name

1 Not connected

2 Black Analog -

3 Red Analog +

Agilent InfinityLab LC Series DAD WR and MWD User Manual 222

11 Identifying Cables Remote Cables

Remote Cables

One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent modules. The other end depends on the instrument to be connected to.

Agilent Module to 3396A Integrators

Agilent Module to 3396 Series II / 3395A Integrators

Use the cable Agilent module to 3396A Series I integrators (03394-60600) and cut pin #5 on the integrator side. Otherwise the integrator prints START; not ready.

p/n 03394-60600 Pin 3396A Pin Agilent module

Signal Name Active (TTL)

9 1 - White Digital ground

NC 2 - Brown Prepare run Low

3 3 - Gray Start Low

NC 4 - Blue Shut down Low

NC 5 - Pink Not connected

NC 6 - Yellow Power on High

5,14 7 - Red Ready High

1 8 - Green Stop Low

NC 9 - Black Start request Low

13, 15 Not connected

Agilent InfinityLab LC Series DAD WR and MWD User Manual 223

11 Identifying Cables Remote Cables

Agilent Module to 3396 Series III / 3395B Integrators

Agilent Module to Agilent 35900 A/D Converters

p/n 03396-61010 Pin 33XX Pin Agilent module

Signal Name Active (TTL)

9 1 - White Digital ground

NC 2 - Brown Prepare run Low

3 3 - Gray Start Low

NC 4 - Blue Shut down Low

NC 5 - Pink Not connected

NC 6 - Yellow Power on High

14 7 - Red Ready High

4 8 - Green Stop Low

NC 9 - Black Start request Low

13, 15 Not connected

p/n 5061-3378 Pin 35900 A/D Pin Agilent module

Signal Name Active (TTL)

1 - White 1 - White Digital ground

2 - Brown 2 - Brown Prepare run Low

3 - Gray 3 - Gray Start Low

4 - Blue 4 - Blue Shut down Low

5 - Pink 5 - Pink Not connected

6 - Yellow 6 - Yellow Power on High

7 - Red 7 - Red Ready High

8 - Green 8 - Green Stop Low

9 - Black 9 - Black Start request Low

Agilent InfinityLab LC Series DAD WR and MWD User Manual 224

11 Identifying Cables Remote Cables

Agilent Module to General Purpose

p/n 01046-60201 Wire Color Pin Agilent module

Signal Name Active (TTL)

White 1 Digital ground

Brown 2 Prepare run Low

Gray 3 Start Low

Blue 4 Shut down Low

Pink 5 Not connected

Yellow 6 Power on High

Red 7 Ready High

Green 8 Stop Low

Black 9 Start request Low

Agilent InfinityLab LC Series DAD WR and MWD User Manual 225

11 Identifying Cables BCD Cables

BCD Cables

One end of these cables provides a 15-pin BCD connector to be connected to the Agilent modules. The other end depends on the instrument to be connected to

Agilent Module to General Purpose

p/n G1351-81600 Wire Color Pin Agilent module

Signal Name BCD Digit

Green 1 BCD 5 20

Violet 2 BCD 7 80

Blue 3 BCD 6 40

Yellow 4 BCD 4 10

Black 5 BCD 0 1

Orange 6 BCD 3 8

Red 7 BCD 2 4

Brown 8 BCD 1 2

Gray 9 Digital ground Gray

Gray/pink 10 BCD 11 800

Red/blue 11 BCD 10 400

White/green 12 BCD 9 200

Brown/green 13 BCD 8 100

not connected 14

not connected 15 + 5 V Low

Agilent InfinityLab LC Series DAD WR and MWD User Manual 226

11 Identifying Cables BCD Cables

Agilent Module to 3396 Integrators

p/n 03396-60560 Pin 3396 Pin Agilent module

Signal Name BCD Digit

1 1 BCD 5 20

2 2 BCD 7 80

3 3 BCD 6 40

4 4 BCD 4 10

5 5 BCD0 1

6 6 BCD 3 8

7 7 BCD 2 4

8 8 BCD 1 2

9 9 Digital ground

NC 15 + 5 V Low

Agilent InfinityLab LC Series DAD WR and MWD User Manual 227

11 Identifying Cables CAN/LAN Cables

CAN/LAN Cables

Both ends of this cable provide a modular plug to be connected to Agilent modules CAN or LAN connectors.

CAN Cables

LAN Cables

p/n Description

5181-1516 CAN cable, Agilent module to module, 0.5 m

5181-1519 CAN cable, Agilent module to module, 1 m

p/n Description

5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)

5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)

Agilent InfinityLab LC Series DAD WR and MWD User Manual 228

11 Identifying Cables Agilent 1200 module to PC

Agilent 1200 module to PC

p/n Description

RS232-61601 RS-232 cable, 2.5 m Instrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It is also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.

5181-1561 RS-232 cable, 8 m

Agilent InfinityLab LC Series DAD WR and MWD User Manual 229

12 Hardware Information

Firmware Description 231

Electrical Connections 234 Rear View of the Module 235 Serial Number Information (ALL) 235

Interfaces 236 Interfaces Overview 238 ERI (Enhanced Remote Interface) 240 USB (Universal Serial Bus) 242

Setting the 6-bit Configuration Switch 243 Special Settings 245

Instrument Layout 247

Early Maintenance Feedback (EMF) 248

This chapter describes the detector in more detail on hardware and electronics.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 230

12 Hardware Information Firmware Description

Firmware Description

The firmware of the instrument consists of two independent sections:

a non-instrument specific section, called resident system

an instrument specific section, called main system

Resident System

This resident section of the firmware is identical for all Agilent 1100/1200/1220/1260/1290 series modules. Its properties are:

the complete communication capabilities (CAN, LAN, USB and RS- 232)

memory management

ability to update the firmware of the 'main system'

Main System

Its properties are:

the complete communication capabilities (CAN, LAN, USB and RS- 232)

memory management

ability to update the firmware of the 'resident system'

In addition the main system comprises the instrument functions that are divided into common functions like

run synchronization through APG/ERI remote,

error handling,

diagnostic functions,

or module specific functions like

internal events such as lamp control, filter movements,

raw data collection and conversion to absorbance.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 231

12 Hardware Information Firmware Description

Firmware Updates

Firmware updates can be done with the Agilent Lab Advisor software with files on the hard disk (latest version should be used).

Required tools, firmware and documentation are available from the Agilent web: http://www.agilent.com/en-us/firmwareDownload?whid=69761

The file naming conventions are:

PPPP_RVVV_XXX.dlb, where

PPPP is the product number, for example, 1315B for the G1315B DAD,

R the firmware revision, for example, A for G1315B or B for the G1315C DAD,

VVV is the revision number, for example 650 is revision 6.50,

XXX is the build number of the firmware.

For instructions on firmware updates refer to section Replacing Firmware in chapter "Maintenance" or use the documentation provided with the Firmware Update Tools.

Figure 63 Firmware Update Mechanism

NOTE Update of main system can be done in the resident system only. Update of the resident system can be done in the main system only.

Main and resident firmware must be from the same set.

Resident System Main FW update

Main System

Resident FW Update

Agilent InfinityLab LC Series DAD WR and MWD User Manual 232

12 Hardware Information Firmware Description

The firmware update tools, firmware and documentation are available from the Agilent web.

http://www.agilent.com/en-us/firmwareDownload?whid=69761

NOTE Some modules are limited in downgrading due to their mainboard version or their initial firmware revision. For example, a G1315C DAD SL cannot be downgraded below firmware revision B.01.02 or to a A.xx.xx.

Some modules can be re-branded (e.g. G1314C to G1314B) to allow operation in specific control software environments. In this case, the feature set of the target type is used and the feature set of the original one is lost. After re-branding (e.g. from G1314B to G1314C), the original feature set is available again.

All this specific information is described in the documentation provided with the firmware update tools.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 233

12 Hardware Information Electrical Connections

Electrical Connections

The CAN bus is a serial bus with high-speed data transfer. The two connectors for the CAN bus are used for internal module data transfer and synchronization.

One analog output provides signals for integrators or data handling systems.

The ERI/REMOTE connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features such as start, stop, common shutdown, prepare, and so on.

With the appropriate software, the LAN connector may be used to control the module from a computer through a LAN connection. This connector is activated and can be configured with the configuration switch.

With the appropriate software, the USB connector may be used to control the module from a computer through a USB connection.

The power input socket accepts a line voltage of 100 240 VAC 10 % with a line frequency of 50 or 60 Hz. Maximum power consumption varies by module. There is no voltage selector on your module because the power supply has wide-ranging capability. There are no externally accessible fuses because automatic electronic fuses are implemented in the power supply.

NOTE Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 234

12 Hardware Information Electrical Connections

Rear View of the Module

Figure 64 Rear view of the detector

Serial Number Information (ALL) The serial number information on the instrument labels provide the following information:

CCXZZ00000 Format

CC Country of manufacturing DE = Germany JP = Japan CN = China

X Alphabetic character A-Z (used by manufacturing)

ZZ Alpha-numeric code 0-9, A-Z, where each combination unambiguously denotes a module (there can be more than one code for the same module)

00000 Serial number

Agilent InfinityLab LC Series DAD WR and MWD User Manual 235

12 Hardware Information Interfaces

Interfaces

The Agilent InfinityLab LC Series modules provide the following interfaces:

Table 19 Agilent InfinityLab LC Series Interfaces

Module CAN USB LAN (on-board)

RS-232 Analog APG (A) / ERI (E)

Special

Pumps

G7104A/C 2 No Yes Yes 1 A

G7110B 2 Yes Yes No No E

G7111A/B, G5654A 2 Yes Yes No No E

G7112B 2 Yes Yes No No E

G7120A, G7132A 2 No Yes Yes 1 A

G7161A/B 2 Yes Yes No No E

Samplers

G7129A/B/C 2 Yes Yes No No E

G7167A/B, G7137A, G5668A 2 Yes Yes No No E

G7157A 2 Yes Yes No No E

Detectors

G7114A/B 2 Yes Yes No 1 E

G7115A 2 Yes Yes No 1 E

G7117A/B/C 2 Yes Yes No 1 E

G7121A/B 2 Yes Yes No 1 E

G7162A/B 2 Yes Yes No 1 E

G7165A 2 Yes Yes No 1 E

Agilent InfinityLab LC Series DAD WR and MWD User Manual 236

12 Hardware Information Interfaces

CAN connectors as interface to other modules

LAN connector as interface to the control software

RS-232C as interface to a computer

USB (Universal Series Bus) as interface to a computer

REMOTE connector as interface to other Agilent products

Analog output connector(s) for signal output

Fraction Collectors

G7158B 2 Yes Yes No No E

G7159B 2 Yes Yes No No E

G7166A 2 No No No No No Requires a host module with on-board LAN with minimum FW B.06.40 or C.06.40, or with additional G1369C LAN Card

G1364E/F, G5664B 2 Yes Yes No No E THERMOSTAT for G1330B

Others

G7116A/B 2 No No No No No Requires a host module with on-board LAN or with additional G1369C LAN Card.

G7122A No No No Yes No A

G7170B 2 No No No No No Requires a host module with on-board LAN with minimum FW B.06.40 or C.06.40, or with additional G1369C LAN Card

Table 19 Agilent InfinityLab LC Series Interfaces

Module CAN USB LAN (on-board)

RS-232 Analog APG (A) / ERI (E)

Special

NOTE The detector (DAD/MWD/FLD/VWD/RID) is the preferred access point for control via LAN. The inter-module communication is done via CAN.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 237

12 Hardware Information Interfaces

Interfaces Overview

CAN

The CAN is inter-module communication interface. It is a 2-wire serial bus system supporting high speed data communication and real-time requirement.

LAN

The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D DAD and G1365C/D MWD). This interface allows the control of the module/system via a PC with the appropriate control software. Some modules have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A Valve Drive or G4227A Flexible Cube). These are hosted modules and require a Host module with firmware B.06.40 or later or with additional G1369C LAN Card.

USB

The USB interface replaces the RS-232 Serial interface in new FUSION generation modules. For details on USB refer to USB (Universal Serial Bus) on page 242.

Analog Signal Output

The analog signal output can be distributed to a recording device. For details refer to the description of the modules mainboard.

Remote (ERI)

The ERI (Enhanced Remote Interface) connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features as common shut down, prepare, and so on.

It allows easy connection between single instruments or systems to ensure coordinated analysis with simple coupling requirements.

The subminiature D connector is used. The module provides one remote connector which is inputs/outputs (wired- or technique).

NOTE If an Agilent detector (DAD/MWD/FLD/VWD/RID) is in the system, the LAN should be connected to the DAD/MWD/FLD/VWD/RID (due to higher data load). If no Agilent detector is part of the system, the LAN interface should be installed in the pump or autosampler.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 238

12 Hardware Information Interfaces

To provide maximum safety within a distributed analysis system, one line is dedicated to SHUT DOWN the systems critical parts in case any module detects a serious problem. To detect whether all participating modules are switched on or properly powered, one line is defined to summarize the POWER ON state of all connected modules. Control of analysis is maintained by signal readiness READY for next analysis, followed by START of run and optional STOP of run triggered on the respective lines. In addition PREPARE and START REQUEST may be issued. The signal levels are defined as:

standard TTL levels (0 V is logic true, + 5.0 V is false),

fan-out is 10,

input load is 2.2 kOhm against + 5.0 V, and

output are open collector type, inputs/outputs (wired- or technique).

NOTE All common TTL circuits operate with a 5 V power supply. A TTL signal is defined as "low" or L when between 0 V and 0.8 V and "high" or H when between 2.0 V and 5.0 V (with respect to the ground terminal).

Table 20 ERI signal distribution

Pin Signal Description

1 START REQUEST (L) Request to start injection cycle (for example, by start key on any module). Receiver is the autosampler.

2 STOP (L) Request to reach system ready state as soon as possible (for example, stop run, abort or finish and stop injection). Receiver is any module performing run-time controlled activities.

3 READY (H) System is ready for next analysis. Receiver is any sequence controller.

4 POWER ON (H) All modules connected to system are switched on. Receiver is any module relying on operation of others.

5 Not used

6 SHUT DOWN (L) System has serious problem (for example, leak: stops pump). Receiver is any module capable to reduce safety risk.

7 START (L) Request to start run / timetable. Receiver is any module performing run-time controlled activities.

8 PREPARE (L) Request to prepare for analysis (for example, calibration, detector lamp on). Receiver is any module performing pre-analysis activities.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 239

12 Hardware Information Interfaces

Special Interfaces

There is no special interface for this module.

ERI (Enhanced Remote Interface) ERI replaces the AGP Remote Interface that is used in the HP 1090/1040/1050/1100 HPLC systems and Agilent 1100/1200/1200 Infinity HPLC modules. All new InfinityLab LC Series products using the FUSION core electronics use ERI. This interface is already used in the Agilent Universal Interface Box 2 (UIB2)

ERI Description

The ERI interface contains eight individual programmable input/output pins. In addition, it provides 24 V power and 5 V power and a serial data line to detect and recognize further add-ons that could be connected to this interface. This way the interface can support various additional devices like sensors, triggers (in and out) and small controllers, etc.

Figure 65 Location of the ERI interface (example shows a G7114A/B VWD)

ERI

Agilent InfinityLab LC Series DAD WR and MWD User Manual 240

12 Hardware Information Interfaces

IO (Input/Output) Lines

Eight generic bi-directional channels (input or output).

Same as the APG Remote.

Devices like valves, relays, ADCs, DACs, controllers can be supported/controlled.

1-Wire Data (Future Use)

This serial line can be used to read out an EPROM or write into an EPROM of a connected ERI-device. The firmware can detect the connected type of device automatically and update information in the device (if required).

Pin Enhanced Remote

1 IO 1 (START REQUEST)

2 IO 2 (STOP)

3 IO 3 (READY)

4 IO 4 (POWER ON)

5 IO 5 (NOT USED)

6 IO 6 (SHUT DOWN)

7 IO 7 (START)

8 IO 8 (PREPARE)

9 1 wire DATA

10 DGND

11 +5 V ERI out

12 PGND

13 PGND

14 +24 V ERI out

15 +24 V ERI out

Agilent InfinityLab LC Series DAD WR and MWD User Manual 241

12 Hardware Information Interfaces

5V Distribution (Future Use)

Available directly after turning on the hosting module (assures that the firmware can detect certain basic functionality of the device).

For digital circuits or similar.

Provides 500 mA maximum.

Short-circuit proof with automatic switch off (by firmware).

24V Distribution (Future Use)

Available by firmware command (defined turn on/off).

For devices that need higher power

Class 0: 0.5 A maximum (12 W)

Class 1: 1.0 A maximum (24 W)

Class 2: 2.0 A maximum (48 W)

Class depends on hosting modules internal power overhead.

If a connected device requires more power the firmware detects this (overcurrent detection) and provides the information to the user interface.

Fuse used for safety protection (on board).

Short circuit will be detected through hardware.

USB (Universal Serial Bus) USB (Universal Serial Bus) - replaces RS232, supports:

a PC with control software (for example Agilent Lab Advisor)

USB Flash Disk

Agilent InfinityLab LC Series DAD WR and MWD User Manual 242

12 Hardware Information Setting the 6-bit Configuration Switch

Setting the 6-bit Configuration Switch

The 6-bit configuration switch is located at the rear of the module with FUSION electronics. Switch settings provide configuration parameters for LAN and instrument specific initialization procedures.

All modules with FUSION electronics:

Default is ALL switches DOWN (best settings).

Default IP address for LAN 192.168.254.11

For specific LAN modes switches 4-5 must be set as required.

For boot resident/cold start modes switches 1+2 or 6 must be UP.

Figure 66 Location of Configuration switch (example shows a G7114A/B VWD)

Configuration switch

Agilent InfinityLab LC Series DAD WR and MWD User Manual 243

12 Hardware Information Setting the 6-bit Configuration Switch

Table 21 6-bit Configuration Switch

Mode Function/Setting

Switch 1 Switch 2 Switch 3 Switch 4 Switch 5 Switch 6

COM1

1 When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are taken from non-volatile memory.

0 n.a.2

2 not assigned - Always keep these switches on position 0 (off)

n.a. LAN Init Mode n.a.

Use Default IP Address3

3 Default IP Address is 192.168.254.11

0 0 0 0 0

Use Stored IP Address 0 0 0 1 0

Use DHCP to request IP Address4

4 Host Name will be the MAC address.

0 0 1 0 0

Test 1 System n.a. n.a. n.a. ColdStart

Boot Main System / Keep Data 0 0 0 0 0

Boot Resident System / Keep Data 1 0 0 0 0

Boot Main System / Revert to Default Data

0 0 0 0 1

Boot Resident System / Revert to Default Data

1 0 0 0 1

Agilent InfinityLab LC Series DAD WR and MWD User Manual 244

12 Hardware Information Setting the 6-bit Configuration Switch

Special Settings

Boot-Resident/Main

Firmware update procedures may require this mode in case of firmware loading errors (main/resident firmware part).

If you use the following switch settings and power the instrument up again, the instrument firmware stays in the resident/main mode. In resident mode, it is not operable as a module. It only uses basic functions of the operating system for example, for communication. In this mode the main firmware can be loaded (using update utilities).

Forced Cold Start

A forced cold start can be used to bring the module into a defined mode with default parameter settings.

Boot Main System / Revert to Default Data

The instrument will boot to main mode and changes to the modules default parameter. May be also required to load resident firmware into the module.

Boot Resident System / Revert to Default Data

The instrument will boot to resident mode and changes to the modules default parameter. May be also required to load main firmware into the module.

If you use the following switch settings and power the instrument up again, it will start as described above.

CAUTION Loss of data

Forced cold start erases all methods and data stored in the non-volatile memory. Exceptions are calibration settings, diagnosis and repair log books which will not be erased.

Save your methods and data before executing a forced cold start.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 245

12 Hardware Information Setting the 6-bit Configuration Switch

Table 22 Boot Resident / Forced Coldstart

SW1 SW2 SW3 SW4 SW5 SW6 Init Mode

1 0 0 0 0 0 Boot Main System / Keep Data

1 1 0 0 0 0 Boot Resident System / Keep Data

1 0 0 0 0 1 Boot Main System / Revert to Default Data

1 1 0 0 0 1 Boot Resident System / Revert to Default Data

Note: The setting '0' (down) is essential.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 246

12 Hardware Information Instrument Layout

Instrument Layout

The industrial design of the module incorporates several innovative features. It uses Agilents E-PAC concept for the packaging of electronics and mechanical assemblies. This concept is based upon the use of expanded polypropylene (EPP) layers of foam plastic spacers in which the mechanical and electronic boards components of the module are placed. This pack is then housed in a metal inner cabinet which is enclosed by a plastic external cabinet. The advantages of this packaging technology are:

virtual elimination of fixing screws, bolts or ties, reducing the number of components and increasing the speed of assembly/disassembly,

the plastic layers have air channels molded into them so that cooling air can be guided exactly to the required locations,

the plastic layers help cushion the electronic and mechanical parts from physical shock, and

the metal inner cabinet shields the internal electronics from electromagnetic interference and also helps to reduce or eliminate radio frequency emissions from the instrument itself.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 247

12 Hardware Information Early Maintenance Feedback (EMF)

Early Maintenance Feedback (EMF)

Maintenance requires the exchange of components which are subject to wear or stress. Ideally, the frequency at which components are exchanged should be based on the intensity of usage of the module and the analytical conditions, and not on a predefined time interval. The early maintenance feedback (EMF) feature monitors the usage of specific components in the instrument, and provides feedback when the user-selectable limits have been exceeded. The visual feedback in the user interface provides an indication that maintenance procedures should be scheduled.

EMF Counters

EMF counters increment with use and can be assigned a maximum limit which provides visual feedback in the user interface when the limit is exceeded. Some counters can be reset to zero after the required maintenance procedure.

The detector provides the following EMF counters:

Deuterium Lamp On-Time

Number of UV lamp ignitions

Using the EMF Counters

The user-settable EMF limits for the EMF Counters enable the early maintenance feedback to be adapted to specific user requirements. The useful maintenance cycle is dependent on the requirements for use. Therefore, the definition of the maximum limits need to be determined based on the specific operating conditions of the instrument.

Lamp Type Counter Reset Comment

lamp with RFID tag NO

lamp without RFID tag YES via Lab Advisor or Instant Pilot

Agilent InfinityLab LC Series DAD WR and MWD User Manual 248

12 Hardware Information Early Maintenance Feedback (EMF)

Setting the EMF Limits

The setting of the EMF limits must be optimized over one or two maintenance cycles. Initially the default EMF limits should be set. When instrument performance indicates maintenance is necessary, take note of the values displayed by the EMF counters. Enter these values (or values slightly less than the displayed values) as EMF limits, and then reset the EMF counters to zero. The next time the EMF counters exceed the new EMF limits, the EMF flag will be displayed, providing a reminder that maintenance needs to be scheduled.

NOTE This function is only available via Agilent Lab Advisor or Instant Pilot.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 249

13 LAN Configuration

What You Have to Do First 251

TCP/IP parameter configuration 252

Configuration Switches 253

Initialization Mode Selection 254

Dynamic Host Configuration Protocol (DHCP) 256 General Information (DHCP) 256 Setup (DHCP) 257

Manual Configuration 259 With Telnet 260 With the Instant Pilot (G4208A) 263

This chapter provides information on connecting the module to the Agilent ChemStation PC.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 250

13 LAN Configuration What You Have to Do First

What You Have to Do First

The module has an on- board LAN communication interface.

1 Note the MAC (Media Access Control) address for further reference. The MAC or hardware address of the LAN interfaces is a world wide unique identifier. No other network device will have the same hardware address. The MAC address can be found on a label at the rear of the module underneath the configuration switch (see Figure 68 on page 251).

Figure 67 MAC-Label

2 Connect the instrument's LAN interface (see Figure 68 on page 251) to

the PC network card using a crossover network cable (point-to-point) or

a hub or switch using a standard LAN cable.

Figure 68 Location of LAN interfaces and MAC label

NOTE This chapter is generic and may show figures that differ from your module. The functionality is the same.

Part number of the detector main board Revision Code, Vendor, Year and Week of assembly MAC address Country of Origin

LAN interface

MAC label

Agilent InfinityLab LC Series DAD WR and MWD User Manual 251

13 LAN Configuration TCP/IP parameter configuration

TCP/IP parameter configuration

To operate properly in a network environment, the LAN interface must be configured with valid TCP/IP network parameters. These parameters are:

IP address

Subnet Mask

Default Gateway

The TCP/IP parameters can be configured by the following methods:

by automatically requesting the parameters from a network-based DHCP Server (using the so-called Dynamic Host Configuration Protocol). This mode requires a LAN-onboard Module or a G1369C LAN Interface card, see Setup (DHCP) on page 257

by manually setting the parameters using Telnet

by manually setting the parameters using the Local Controller

The LAN interface differentiates between several initialization modes. The initialization mode (short form init mode) defines how to determine the active TCP/IP parameters after power-on. The parameters may be derived non-volatile memory or initialized with known default values. The initialization mode is selected by the configuration switch, see Table 23 on page 254.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 252

13 LAN Configuration Configuration Switches

Configuration Switches

The configuration switch can be accessed at the rear of the module.

Figure 69 Location of Configuration switch (example shows a G7114A/B VWD)

The module is shipped with all switches set to OFF, as shown above.

Configuration switch

NOTE To perform any LAN configuration, SW1 and SW2 must be set to OFF.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 253

13 LAN Configuration Initialization Mode Selection

Initialization Mode Selection

The following initialization (init) modes are selectable:

Default IP address for LAN is 192.168.254.11.

DHCP address is the modules LAN MAC address.

Using Stored

When initialization mode Using Stored is selected, the parameters are taken from the non-volatile memory of the module. The TCP/IP connection will be established using these parameters. The parameters were configured previously by one of the described methods.

Figure 70 Using Stored (Principle)

Table 23 Initialization Mode Switches

SW1 SW2 SW3 SW4 SW5 SW6 Init Mode

0 0 0 0 0 0 Use Default IP Address

0 0 0 0 1 0 Use Stored IP Address

0 0 0 1 0 0 Use DHCP

Note: The setting 0 (down) is essential.

Non-Volatile RAM

Stored Parameter

Active Parameter

Agilent InfinityLab LC Series DAD WR and MWD User Manual 254

13 LAN Configuration Initialization Mode Selection

Using Default

When Using Default is selected, the factory default parameters are taken instead. These parameters enable a TCP/IP connection to the LAN interface without further configuration, see Table 24 on page 255.

Figure 71 Using Default (Principle)

Since the default IP address is a so-called local address, it will not be routed by any network device. Thus, the PC and the module must reside in the same subnet.

The user may open a Telnet session using the default IP address and change the parameters stored in the non-volatile memory of the module. He may then close the session, select the initialization mode Using Stored, power-on again and establish the TCP/IP connection using the new parameters.

When the module is wired to the PC directly (e.g. using a cross-over cable or a local hub), separated from the local area network, the user may simply keep the default parameters to establish the TCP/IP connection.

Active Parameter

Default Parameter

NOTE Using the default address in your local area network may result in network problems. Take care and change it to a valid address immediately.

Table 24 Using Default Parameters

IP address: 192.168.254.11

Subnet Mask: 255.255.255.0

Default Gateway not specified

NOTE In the Using Default mode, the parameters stored in the memory of the module are not cleared automatically. If not changed by the user, they are still available, when switching back to the mode Using Stored.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 255

13 LAN Configuration Dynamic Host Configuration Protocol (DHCP)

Dynamic Host Configuration Protocol (DHCP)

General Information (DHCP) The Dynamic Host Configuration Protocol (DHCP) is an auto configuration protocol used on IP networks. The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and B-firmware (B.06.40 or above) or modules with "D"-firmware. All modules should use latest firmware from the same set.

When the initialization mode DHCP is selected, the card tries to download the parameters from a DHCP Server. The parameters obtained become the active parameters immediately. They are not stored to the non-volatile memory of the card.

Besides requesting the network parameters, the card also submits its hostname to the DHCP Server. The hostname equals the MAC address of the card, e.g. 0030d3177321. It is the DHCP server's responsibility to forward the hostname/address information to the Domain Name Server. The card does not offer any services for hostname resolution (e.g. NetBIOS).

Figure 72 DHCP (Principle)

DHCP Server

Active Parameter

NOTE 1 It may take some time until the DHCP server has updated the DNS server with the hostname information.

2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g. 0030d3177321.country.company.com.

3 The DHCP server may reject the hostname proposed by the card and assign a name following local naming conventions.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 256

13 LAN Configuration Dynamic Host Configuration Protocol (DHCP)

Setup (DHCP)

1 Note the MAC address of the LAN interface (provided with G1369C LAN Interface Card or mainboard). This MAC address is on a label on the card or at the rear of the mainboard, for example, 0030d3177321.

On the Local Controller the MAC address can be found under Details in the LAN section.

Figure 73 LAN Setting on Instant Pilot

The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and B-firmware (B.06.40 or above) or modules with "D"-firmware. All modules should use latest firmware from the same set.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 257

13 LAN Configuration Dynamic Host Configuration Protocol (DHCP)

2 Set the configuration switch to DHCP either on the G1369C LAN Interface Card or the mainboard of above mentioned modules.

3 Turn on the module that hosts the LAN interface.

4 Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition, Lab Advisor, Firmware Update Tool) and use MAC address as host name, e.g. 0030d3177321.

The LC system should become visible in the control software (see Note in section General Information (DHCP) on page 256).

Table 25 G1369C LAN Interface Card (configuration switch on the card)

SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode

ON OFF OFF OFF OFF DHCP

Table 26 LC Modules with 8-bit configuration switch (B-firmware) (configuration switch at rear

of the instrument)

SW 6 SW 7 SW 8 Initialization Mode

ON OFF OFF DHCP

Agilent InfinityLab LC Series DAD WR and MWD User Manual 258

13 LAN Configuration Manual Configuration

Manual Configuration

Manual configuration only alters the set of parameters stored in the non-volatile memory of the module. It never affects the currently active parameters. Therefore, manual configuration can be done at any time. A power cycle is mandatory to make the stored parameters become the active parameters, given that the initialization mode selection switches are allowing it.

Figure 74 Manual Configuration (Principle)

TELNET Session

Control Module

Stored Parameter

Non-Volatile RAM

Agilent InfinityLab LC Series DAD WR and MWD User Manual 259

13 LAN Configuration Manual Configuration

With Telnet Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set by any method), the parameters may be altered by opening a Telnet session.

1 Open the system (DOS) prompt window by clicking on Windows START button and select Run.... Type cmd and press OK.

2 Type the following at the system (DOS) prompt:

c:\>telnet or

c:\>telnet

Figure 75 Telnet - Starting a session

where may be the assigned address from a Bootp cycle, a configuration session with the Handheld Controller, or the default IP address (see Configuration Switches on page 253).

When the connection was established successfully, the module responds with the following:

Figure 76 A connection to the module is made

Agilent InfinityLab LC Series DAD WR and MWD User Manual 260

13 LAN Configuration Manual Configuration

3 Type ? and press enter to see the available commands.

Figure 77 Telnet Commands

4 To change a parameter follows the style:

parameter value, for example: ip 134.40.28.56

Then press [Enter], where parameter refers to the configuration parameter you are defining, and value refers to the definitions you are assigning to that parameter. Each parameter entry is followed by a carriage return.

Table 27 Telnet Commands

Value Description

? displays syntax and descriptions of commands

/ displays current LAN settings

ip sets new ip address

sm sets new subnet mask

gw sets new default gateway

exit exits shell and saves all changes

Agilent InfinityLab LC Series DAD WR and MWD User Manual 261

13 LAN Configuration Manual Configuration

5 Use the / and press Enter to list the current settings.

6 Change the IP address (in this example 192.168.254.12) and type / to list current settings.

7 When you have finished typing the configuration parameters, type exit and press Enter to exit with storing parameters.

Figure 80 Closing the Telnet Session

Figure 78 Telnet - Current settings in "Using Stored" mode

information about the LAN interface MAC address, initialization mode Initialization mode is Using Stored active TCP/IP settings

TCP/IP status - here ready connected to PC with controller software (e.g. Agilent ChemStation), here not connected

Figure 79 Telnet - Change IP settings

change of IP setting to Initialization mode is Using Stored

active TCP/IP settings

stored TCP/IP settings in non-volatile memory

connected to PC with controller software (e.g. Agilent ChemStation), here not connected

NOTE If the Initialization Mode Switch is changed now to Using Stored mode, the instrument will take the stored settings when the module is re-booted. In the example above it would be 192.168.254.12.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 262

13 LAN Configuration Manual Configuration

With the Instant Pilot (G4208A) To configure the TCP/IP parameters before connecting the module to the network, the Instant Pilot (G4208A) can be used.

1 From the Welcome screen press the More button.

2 Select Configure.

3 Press the module button of the module that hosts the LAN interface (usually the detector).

4 Scroll down to the LAN settings.

Figure 81 Instant Pilot - LAN Configuration (Edit mode)

5 Press the Edit button (only visible if not in Edit mode), perform the required changes and press the Done button.

6 Leave the screen by clicking Exit.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 263

14 Appendix

General Safety Information 265 General Safety Information 265 Safety Standards 265 General 265 Before Applying Power 266 Ground the Instrument 266 Do Not Operate in an Explosive Atmosphere 267 Do Not Remove the Instrument Cover 267 Do Not Modify the Instrument 267 In Case of Damage 267 Solvents 268 Safety Symbols 269

Waste Electrical and Electronic Equipment (WEEE) Directive 271

Radio Interference 272

Sound Emission 273

UV-Radiation 274

Solvent Information 275

Declaration of Conformity for HOX2 Filter 276

Installation of Stainless Steel Cladded PEEK Capillaries 277 First Step: Finger-tight Fitting 277 Second Step: Installation to Connector 278 Removing Capillaries 282

Agilent Technologies on Internet 283

This chapter provides addition information on safety, legal and web.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 264

14 Appendix General Safety Information

General Safety Information

General Safety Information The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customers failure to comply with these requirements.

Safety Standards This is a Safety Class I instrument (provided with terminal for protective earthing) and has been manufactured and tested according to international safety standards.

General Do not use this product in any manner not specified by the manufacturer. The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.

WARNING Ensure the proper usage of the equipment.

The protection provided by the equipment may be impaired.

The operator of this instrument is advised to use the equipment in a manner as specified in this manual.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 265

14 Appendix General Safety Information

Before Applying Power

Ground the Instrument

WARNING Wrong voltage range, frequency or cabling

Personal injury or damage to the instrument

Verify that the voltage range and frequency of your power distribution matches to the power specification of the individual instrument.

Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.

Make all connections to the unit before applying power.

NOTE Note the instrument's external markings described under Safety Symbols on page 269.

WARNING Missing electrical ground

Electrical shock

If your product is provided with a grounding type power plug, the instrument chassis and cover must be connected to an electrical ground to minimize shock hazard.

The ground pin must be firmly connected to an electrical ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 266

14 Appendix General Safety Information

Do Not Operate in an Explosive Atmosphere

Do Not Remove the Instrument Cover

Do Not Modify the Instrument Do not install substitute parts or perform any unauthorized modification to the product. Return the product to an Agilent Sales and Service Office for service and repair to ensure that safety features are maintained.

In Case of Damage

WARNING Presence of flammable gases or fumes

Explosion hazard

Do not operate the instrument in the presence of flammable gases or fumes.

WARNING Instrument covers removed

Electrical shock

Do Not Remove the Instrument Cover

Only Agilent authorized personnel are allowed to remove instrument covers. Always disconnect the power cables and any external circuits before removing the instrument cover.

WARNING Damage to the module

Personal injury (for example electrical shock, intoxication)

Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 267

14 Appendix General Safety Information

Solvents

WARNING Toxic, flammable and hazardous solvents, samples and reagents

The handling of solvents, samples and reagents can hold health and safety risks.

When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.

Do not use solvents with an auto-ignition temperature below 200 C (392 F). Do not use solvents with a boiling point below 56 C (133 F).

Avoid high vapor concentrations. Keep the solvent temperature at least 40 C (72 F) below the boiling point of the solvent used. This includes the solvent temperature in the sample compartment. For the solvents methanol and ethanol keep the solvent temperature at least 25 C (45 F) below the boiling point.

Do not operate the instrument in an explosive atmosphere.

Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for example, carbon disulfide).

Reduce the volume of substances to the minimum required for the analysis.

Never exceed the maximum permissible volume of solvents (8 L) in the solvent cabinet. Do not use bottles that exceed the maximum permissible volume as specified in the usage guideline for solvent cabinet.

Ground the waste container.

Regularly check the filling level of the waste container. The residual free volume in the waste container must be large enough to collect the waste liquid.

To achieve maximal safety, regularly check the tubing for correct installation.

NOTE For details, see the usage guideline for the solvent cabinet. A printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available in the Agilent Information Center or via the Internet.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 268

14 Appendix General Safety Information

Safety Symbols

Table 28 Symbols

The apparatus is marked with this symbol when the user shall refer to the instruction manual in order to protect risk of harm to the operator and to protect the apparatus against damage.

Indicates dangerous voltages.

Indicates a protected ground terminal.

The apparatus is marked with this symbol when hot surfaces are available and the user should not touch it when heated up.

Sample Cooler unit is designed as vapor-compression refrigeration system. Contains fluorinated greenhouse gas (refrigerant) according to the Kyoto protocol. For specifications of refrigerant, charge capacity, carbon dioxide equivalent (CDE), and global warming potential (GWP) see instrument label.

Flammable Material For Sample Thermostat which uses flammable refrigerant consult Agilent Information Center / User Manual before attempting to install or service this equipment. All safety precautions must be followed.

Confirms that a manufactured product complies with all applicable European Community directives. The European Declaration of Conformity is available at: http://regulations.corporate.agilent.com/DoC/search.htm

Manufacturing date.

Power symbol indicates On/Off. The apparatus is not completely disconnected from the mains supply when the power switch is in the Off position

Pacemaker Magnets could affect the functioning of pacemakers and implanted heart defibrillators. A pacemaker could switch into test mode and cause illness. A heart defibrillator may stop working. If you wear these devices keep at least 55 mm distance to magnets. Warn others who wear these devices from getting too close to magnets.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 269

14 Appendix General Safety Information

Magnetic field Magnets produce a far-reaching, strong magnetic field. They could damage TVs and laptops, computer hard drives, credit and ATM cards, data storage media, mechanical watches, hearing aids and speakers. Keep magnets at least 25 mm away from devices and objects that could be damaged by strong magnetic fields.

Indicates a pinching or crushing hazard

Indicates a piercing or cutting hazard.

Table 28 Symbols

WARNING A WARNING

alerts you to situations that could cause physical injury or death.

Do not proceed beyond a warning until you have fully understood and met the indicated conditions.

CAUTION A CAUTION

alerts you to situations that could cause loss of data, or damage of equipment.

Do not proceed beyond a caution until you have fully understood and met the indicated conditions.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 270

14 Appendix Waste Electrical and Electronic Equipment (WEEE) Directive

Waste Electrical and Electronic Equipment (WEEE) Directive

This product complies with the European WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.

NOTE Do not dispose of in domestic household waste

To return unwanted products, contact your local Agilent office, or see http://www.agilent.com for more information.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 271

14 Appendix Radio Interference

Radio Interference

Cables supplied by Agilent Technologies are screened to provide optimized protection against radio interference. All cables are in compliance with safety or EMC regulations.

Test and Measurement

If test and measurement equipment is operated with unscreened cables, or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 272

14 Appendix Sound Emission

Sound Emission

Sound pressure

Sound pressure Lp <70 db(A) accroding to DIN-EN 27779

Schalldruckpegel

Schalldruckpegel Lp <70 db(A) nach DIN-EN 27779

Agilent InfinityLab LC Series DAD WR and MWD User Manual 273

14 Appendix UV-Radiation

UV-Radiation

Emissions of ultraviolet radiation (200-315 nm) from this product is limited such that radiant exposure incident upon the unprotected skin or eye of operator or service personnel is limited to the following TLVs (Threshold Limit Values) according to the American Conference of Governmental Industrial Hygienists:

Typically the radiation values are much smaller than these limits:

NOTE This information is only valid for UV-lamps without cover (e.g. 2140-0590 and 2140-0813).

Table 29 UV-Radiation Limits

Exposure/day Effective Irradiance

8 hours 0.1 W/cm2

10 minutes 5.0 W/cm2

Table 30 UV-Radiation Typical Values

Position Effective Irradiance

Lamp installed, 50 cm distance Average 0.016 W/cm2

Lamp installed, 50 cm distance Maximum 0.14 W/cm2

Agilent InfinityLab LC Series DAD WR and MWD User Manual 274

14 Appendix Solvent Information

Solvent Information

Flow Cell To protect optimal functionality of your flow-cell:

Avoid the use of alkaline solutions (pH > 9.5) which can attack quartz and thus impair the optical properties of the flow cell.

Use of Solvents Observe the following recommendations on the use of solvents.

Brown glass ware can avoid growth of algae.

Avoid the use of the following steel-corrosive solvents:

solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on),

high concentrations of inorganic acids like sulfuric acid and nitric acid, especially at higher temperatures (if your chromatography method allows, replace by phosphoric acid or phosphate buffer which are less corrosive against stainless steel),

halogenated solvents or mixtures which form radicals and/or acids, for example:

2CHCl3 + O2 2COCl2 + 2HCl

This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol,

chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, diisopropyl ether) should be filtered through dry aluminium oxide which adsorbs the peroxides,

solvents containing strong complexing agents (e.g. EDTA),

mixtures of carbon tetrachloride with 2-propanol or THF.

Avoid the use of dimethyl formamide (DMF). Polyvinylidene fluoride (PVDF), which is used in leak sensors, is not resistant to DMF.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 275

14 Appendix Declaration of Conformity for HOX2 Filter

Declaration of Conformity for HOX2 Filter

Agilent InfinityLab LC Series DAD WR and MWD User Manual 276

14 Appendix Installation of Stainless Steel Cladded PEEK Capillaries

Installation of Stainless Steel Cladded PEEK Capillaries

The 1260 Infinity Bio-inert LC system uses PEEK capillaries that are cladded with stainless steel. These capillaries combine the high pressure stability of steel with the inertness of PEEK. They are used in the high pressure flow path after sample introduction (loop/needle seat capillary) through the thermostatted column compartment/heat exchangers to the column. Such capillaries need to be installed carefully in order to keep them tight without damaging them by over-tightening.

The installation consists of two steps. In the first step, the fitting is installed finger-tight without using tools. Finger-tight means that the fitting will grip and hold the capillary. This brings the fitting to the appropriate start position (marked as 0 below) for the second step.

First Step: Finger-tight Fitting 1 Tighten the fitting using your fingers.

NOTE This installation procedure applies for capillaries and corresponding fittings used in modules delivered before January 2013.

Agilent InfinityLab LC Series DAD WR and MWD User Manual 277

14 Appendix Installation of Stainless Steel Cladded PEEK Capillaries

Second Step: Installation to Connector In the second step (Second Step: Installation to Hard Connectors on page 278 or Second Step: Installation to Soft Connectors on page 279), a wrench is used to rotate the fitting relative to the finger-tight position by a defined angle. For each of the cases mentioned above, there is a recommended range in which the fitting is tight.

Staying below this range could create a leak, either a visible one or a micro-leak, potentially biasing measurement results. Exceeding the recommended range could damage the capillary.

Alternatively, a torque wrench may be used. The target torque for all connections is about 0.7 Nm. When using a torque wrench, read instructions for that tool carefully, as wrong handling may easily miss the correct torque.

Second Step: Installation to Hard Connectors

Use this procedure for hard connectors made from metal (titanium) or ceramics. In the system, these are connections to and from the analytical head of the autosampler (connections to injection valve and needle), and to a metal column.

First installation of a capillary to a hard connector

1 When tightening a fitting for the first time, start from the finger-tight position (which is not necessarily a vertical wrench position) and rotate the wrench by 135 180 . Staying below 135 (grey arrow) will be insufficiently tight, more than 180 (red arrow) could damage the capillary.

possibly leaky

recommended range

Agilent InfinityLab LC Series DAD WR and MWD User Manual 278

14 Appendix Installation of Stainless Steel Cladded PEEK Capillaries

Second and subsequent installations of a capillary to a hard connector

1 When tightening the fitting for the second and subsequent times, again start from the finger-tight position (which is not necessarily a vertical wrench position) and rotate the wrench by 90 135 . Staying below 90 (grey arrow) could be insufficiently tight, more than 135 (red arrow) could damage the capillary.

Second Step: Installation to Soft Connectors

Use this procedure for soft connectors, which are typically made from PEEK. These are the following connections:

to and from all bio-inert valves (injection valve in the autosampler and valves in the thermostatted column compartment and 1290 Infinity Valve Drive),

bio-inert ZDV unions (detector flow cells, multi-draw upgrade kit, capillary to capillary connections, for example, for heat exchangers),

to the autosampler needle and

to PEEK columns (like many bio-inert columns).

For the installation of bio-inert ZDV unions, refert to the Technical Note "Installation of stainless steel cladded PEEK capillaries" (p/n G5611-90120).

possibly leaky

recommended range

Agilent InfinityLab LC Series DAD WR and MWD User Manual 279

14 Appendix Installation of Stainless Steel Cladded PEEK Capillaries

First installation of a capillary to a soft connector

1 When tightening a fitting for the first time, start from the finger-tight position (which does not necessarily need to be a vertical wrench position) and rotate the wrench by 180 210 . Staying below 180 (grey arrow) will not be sufficiently tight, more than 210 (red arrow) could damage the capillary.

Second and subsequent installations of a capillary to a soft connector

1 When tightening the fitting for the second and subsequent times, again start from the finger-tight position (which is not necessarily a vertical wrench position) and rotate the wrench by 135 180 . Staying below 135 (grey arrow) could be insufficiently tight enough, more than 180 (red arrow) could damage the capillary.

possibly leaky

recommended range

possibly leaky

recommended range

Agilent InfinityLab LC Series DAD WR and MWD User Manual 280

14 Appendix Installation of Stainless Steel Cladded PEEK Capillaries

Summary for Second Step

Table 31 Summary for second step

2ndStep First installation Subsequent installations

Hard connectors

Soft connectors

possibly leaky

recommended range

possibly leaky

recommended range

recommended range

possibly leaky possibly

leaky

recommended range

Agilent InfinityLab LC Series DAD WR and MWD User Manual 281

14 Appendix Installation of Stainless Steel Cladded PEEK Capillaries

Removing Capillaries

To keep the flow path free of stainless steel, the front end of the capillary is made of PEEK. Under high pressure, or when in contact with some solvents, PEEK can expand to the shape of the connector where the capillary is installed. If the capillary is removed, this may become visible as a small step. In such cases, do not try to pull the fitting from the capillary, as this can destroy the front part of the capillary. Instead, carefully pull it to the rear. During installation of the capillary, the fitting will end up in the correct position.

Figure 82 Capillary fitting

CAUTION Potential damage of capillaries

Do not remove fittings from used capillaries.

Step

Rear Front

Agilent InfinityLab LC Series DAD WR and MWD User Manual 282

14 Appendix Agilent Technologies on Internet

Agilent Technologies on Internet

For the latest information on products and services visit our worldwide web site on the Internet at:

http://www.agilent.com

Agilent InfinityLab LC Series DAD WR and MWD User Manual 283

In This Book

This manual contains technical reference information about the Agilent 1260 Infinity II Diode Array and Multiple Wavelength Detectors G7115A WR and G7165A.

The manual describes the following:

introduction and specifications,

installation,

using and optimizing,

troubleshooting and diagnose,

maintenance,

parts identification,

hardware information,

Manualsnet FAQs

If you want to find out how the InfinityLab LC Series Agilent works, you can view and download the Agilent InfinityLab LC Series Diode Array Detector User Manual on the Manualsnet website.

Yes, we have the User Manual for Agilent InfinityLab LC Series as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Manual should include all the details that are needed to use a Agilent InfinityLab LC Series. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent InfinityLab LC Series Diode Array Detector User Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent InfinityLab LC Series Diode Array Detector User Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent InfinityLab LC Series Diode Array Detector User Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent InfinityLab LC Series Diode Array Detector User Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent InfinityLab LC Series Diode Array Detector User Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.