Agilent InfinityLab LC Series Refractive Index Detector
User Manual
Agilent InfinityLab LC Series RID User Manual
Notices
Document Information Document No: SD-29000641 Rev. C Edition: 10/2021
Copyright Agilent Technologies, Inc. 2015-2021
No part of this manual may be repro- duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written con- sent from Agilent Technologies, Inc. as governed by United States and interna- tional copyright laws.
Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future edi- tions. Further, to the maximum extent permitted by applicable law, Agilent dis- claims all warranties, either express or implied, with regard to this manual and any information contained herein, includ- ing but not limited to the implied warran- ties of merchantability and fitness for a particular purpose. Agilent shall not be lia- ble for errors or for incidental or conse- quential damages in connection with the furnishing, use, or performance of this document or of any information con- tained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend U.S. Government Restricted Rights. Soft- ware and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Soft- ware and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Depart- ment of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commer- cial Computer Software or Computer Software Documentation).
Safety Notices
CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARN- ING notice until the indicated conditions are fully understood and met.
Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn, Germany
Agilent InfinityLab LC Series RID User Manual
In This Guide...
This manual covers the
Agilent 1290 Infinity II Refractive Index Detector (G7162B)
Agilent 1260 Infinity II Refractive Index Detector (G7162A)
1 Introduction to the Refractive Index Detector
This chapter gives an introduction to the Refractive Index Detector.
2 Site Requirements and Specifications
This chapter provides information on environmental requirements, physical and performance specifications.
3 Using the Module
This chapter explains the essential operational parameters of the module.
4 Preparing the Module
This chapter provides information on how to set up the module for an analysis and explains the basic settings.
5 Optimizing the Detector
This chapter provides information on how to optimize the detector.
6 Troubleshooting and Diagnostics
This chapter gives an overview about the troubleshooting and diagnostic features and the different user interfaces.
7 Error Information
This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.
Agilent InfinityLab LC Series RID User Manual 3
8 Test Functions
This chapter describes the detectors built in test functions.
9 Maintenance
This chapter provides general information on maintenance of the detector.
10 Parts for Maintenance
This chapter provides information on parts for maintenance.
11 Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.
12 Hardware Information
This chapter describes the module in more detail on hardware and electronics.
13 LAN Configuration
This chapter provides information on connecting the module to the Agilent ChemStation PC.
14 Appendix
This chapter provides safetey and other general information.
Agilent InfinityLab LC Series RID User Manual 4
Contents
1 Introduction to the Refractive Index Detector 9
Introduction to the Refractive Index Detector 10 G7162A Refractive Index Detector 12 G7162B Refractive Index Detector 14 How the Detector Operates 16 Detection Principle 18 Flow Path 20 Leak and Waste Handling 27
2 Site Requirements and Specifications 33
Site Requirements 34 Physical Specifications 37 Performance Specifications 38
3 Using the Module 42
Magnets 43 Turn on/off 44 Status Indicators 46 Instrument Configuration 47 Set up the Detector with Agilent OpenLab ChemStation 49 The Detector User Interface 50 Detector Control Settings 53 Method Parameter Settings 55 Agilent Local Control Modules 59
4 Preparing the Module 60
Leak and Waste Handling 61 Setting up an Analysis 63 Solvent Information 71
Agilent InfinityLab LC Series RID User Manual 5
5 Optimizing the Detector 77
Refractive Index Detector Optimization 78
6 Troubleshooting and Diagnostics 82
Available Tests vs User Interfaces 83 Agilent Lab Advisor Software 84
7 Error Information 85
What Are Error Messages 86 General Error Messages 87 Refractive Index Detector Specific Error Messages 94 Not-Ready Messages 101
8 Test Functions 103
Introduction 104 Refractive Index Calibration 108 Optical Balance 114 Using the Build-in Test Chromatogram 117 ASTM Drift and Noise Test 119 D/A Converter (DAC) Test 123 Other Lab Advisor Functions 125
9 Maintenance 126
Introduction to Maintenance 127 Warnings and Cautions 128 Overview of Maintenance 130 Cleaning the Module 131 Remove and Install Doors 132 Flow Cell Flushing 134 Correct Leaks 135 Replace Leak Handling System Parts 138 Replace the Module Firmware 140
Agilent InfinityLab LC Series RID User Manual 6
10 Parts for Maintenance 141
Overview of Maintenance Parts 142 Accessory Kit 144 Leak Handling Parts 147
11 Identifying Cables 148
Cable Overview 149 Analog Cables 151 Remote Cables 153 CAN/LAN Cables 157 Agilent Module to PC 158 USB Cables 159
12 Hardware Information 160
Data Flow for Chromatographic Output 161 Firmware Description 162 Electrical Connections 165 Interfaces 168 Setting the 6-bit Configuration Switch 176 Instrument Layout 180 Early Maintenance Feedback 181
13 LAN Configuration 182
What You Have to Do First 183 TCP/IP parameter configuration 184 Configuration Switches 185 Initialization Mode Selection 186 Dynamic Host Configuration Protocol (DHCP) 188 Manual Configuration 191 PC and Agilent ChemStation Setup 195
Agilent InfinityLab LC Series RID User Manual 7
14 Appendix 204
General Safety Information 205 Waste Electrical and Electronic Equipment (WEEE) Directive 211 Radio Interference 212 Sound Emission 213 Solvent Information 214 Agilent Technologies on Internet 216
Agilent InfinityLab LC Series RID User Manual 8
1 Introduction to the Refractive Index Detector
Introduction to the Refractive Index Detector 10
G7162A Refractive Index Detector 12 Product Description 12 Features 13
G7162B Refractive Index Detector 14 Product Description 14 Features 15
How the Detector Operates 16
Detection Principle 18
Flow Path 20
Leak and Waste Handling 27 Leak Sensor 31 Waste Guidance 31 Waste Concept 32
This chapter gives an introduction to the Refractive Index Detector.
Agilent InfinityLab LC Series RID User Manual 9
1 Introduction to the Refractive Index Detector Introduction to the Refractive Index Detector
Introduction to the Refractive Index Detector
The Standard Optical (G7162A) and Micro Optical (G7162B) detectors are designed for highest optical performance, GLP compliance, and easy maintenance. They include the following features:
advanced temperature controlled detector optics ready to use within two hours of installation,
automatic zero and automatic purge combined with a recycle valve for automatic solvent recycling allow uninterrupted operation,
durable tungsten lamp with a life expectancy of 40000 hours,
automatic light intensity control circuit to ensure the optimum performance of the optics,
integrated diagnostics for efficient troubleshooting,
built-in refractive index calibration,
front access to valves and capillaries for easy maintenance.
For specifications, see Performance Specifications on page 38.
Figure 1 The Agilent 1200 Infinity Series Refractive Index Detector
Agilent InfinityLab LC Series RID User Manual 10
1 Introduction to the Refractive Index Detector Introduction to the Refractive Index Detector
G7162A Refractive Index Detector
This detector is equipped with the standard optical unit and is used for all standard applications.
It comprises the following features:
inlet port to sample cell 62 L,
inlet port to outlet port 590 L,
maximum data rate up to 74 Hz.
G7162B Refractive Index Detector Micro
This detector is equipped with the micro optical unit and is used for all low volume applications.
It comprises the following features:
inlet port to sample cell 2.5 L,
inlet port to outlet port 265 L,
maximum data rate up to 148 Hz,
LED illumination.
Agilent InfinityLab LC Series RID User Manual 11
1 Introduction to the Refractive Index Detector G7162A Refractive Index Detector
G7162A Refractive Index Detector
Product Description The Agilent 1260 Infinity II Refractive Index Detector (RID) is the ideal detector for fast and reliable LC results when routinely analyzing non-UV absorbing substances, such as carbohydrates, lipids, and polymers. The 1260 Infinity II RID is also the detector of choice for gel permeation chromatography (GPC) or size exclusion chromatography (SEC).
Figure 2 Overview of the detector
Status indicator
Interface port area
Service door
Power switch
Leak drain
Agilent InfinityLab LC Series RID User Manual 12
1 Introduction to the Refractive Index Detector G7162A Refractive Index Detector
Features High sensitivity and robust detection with an excellent reproducibility of only
0.1 % RSD
Precise temperature control from ambient +5 C to 55 C to minimize baseline drift and optimize compound response
Automatic light level adjustment compensates for lamp degradation and flow cell staining, ensuring optimum performance and lowest maintenance
Integrated solvent recycle valve for solvent recycling between runs
Programmable reference cell purging (autopurge) and auto-zeroing of the detector baseline before each run for unattended operation
Convenient front access to the flow connections and valves next to the optical unit
Early Maintenance Feedback (EMF) for prior notification of detector maintenance and continuous tracking of reference flow cell purging with user settable limits
Fast and simple troubleshooting via diagnostic signal data collection
Automatic offset adjustment for easy detection and quantification of negative peaks, and a time-programmable polarity switch for simple inversion of peaks for ease of integration
Excellent linear response for compounds with little or no UV chromaphore in a wide concentration range
Agilent InfinityLab LC Series RID User Manual 13
1 Introduction to the Refractive Index Detector G7162B Refractive Index Detector
G7162B Refractive Index Detector
Product Description The 1290 Infinity II Refractive Index Detector is equipped with an ultralow dispersion microflow cell, which significantly reduces run times for higher sample throughput and improved resolution. Lower solvent consumption means much lower cost of analysis. A high-performance detector of choice for accurate, reproducible, routine analysis of polymers and other compounds that arent detectable by UV.
Figure 3 Overview of the detector
Status indicator
Interface port area
Service door
Power switch
Leak drain
Agilent InfinityLab LC Series RID User Manual 14
1 Introduction to the Refractive Index Detector G7162B Refractive Index Detector
Features Shorter run times for considerably higher sample throughput.
Ultralow dispersion for improved sample definition and resolution.
Reduced solvent consumption for significant savings in analysis cost.
Ideal tool for polymer analysis consistent molecular weights, micro or analytical scale.
Excellent sensitivity achieve low limits of detection.
148 Hz data rate even the narrowest distribution peaks can be detected and accurately quantified.
Minimized band broadening high-definition sample profiling.
Fast startup advanced low thermal mass design means that initial setup is typically less than two hours.
Further time and solvent saving a recycle valve enables sample to be recirculated when the sample is not passing through the flow cell.
Automatic purging of reference flow cell using programmable purge and wait times.
Easy maintenance convenient front access for maintaining instrument uptime.
Early maintenance feedback (EMF) indicates that preventative maintenance is due.
Extensive diagnostics includes error detection and display with Instant Pilot controller and Lab Advisor software.
RoHS compliant aligned with European Union directives regarding specific hazardous materials.
Agilent InfinityLab LC Series RID User Manual 15
1 Introduction to the Refractive Index Detector How the Detector Operates
How the Detector Operates
Refractive index
When a beam of light passes from one medium into another, the wave velocity and direction changes. The change in direction is called refraction. The relationship between the angle of incidence and the angle of refraction is expressed in Snell's Law of refraction.
Where:
n = Refractive index of medium 1 relative to medium 2
n2 = Refractive index of medium 2
n1 = Refractive index of medium 1
1 = angle of incident light in medium 1
2 = angle of refraction in medium 2
Figure 4 Light Refraction
Medium 1
Medium 2
Agilent InfinityLab LC Series RID User Manual 16
1 Introduction to the Refractive Index Detector How the Detector Operates
According to the formula below small angles of external deflection are proportional to the difference between the refractive indices of medium 1 and medium 2.
Where:
= angle of external deflection
n2 = Refractive index of medium 2
n1 = Refractive index of medium 1
Factors that Affect Refractive Index
The refractive index of a medium is affected by a number of factors;
1 Wavelength
The refractive index varies with changes in the wavelength of the incident light beam.
2 Density
As the density of the medium changes the refractive index changes. At a fixed wavelength of incident light the changes in refractive index are generally linear in relation to the changes in medium density.
The density of a medium will be affected by the following factors:
Composition (if not a pure substance)
Temperature
Pressure
Agilent InfinityLab LC Series RID User Manual 17
1 Introduction to the Refractive Index Detector Detection Principle
Detection Principle
Detector Design
The Agilent 1200 Infinity II Refractive Index Detector is a differential refractometer that measures the deflection of a light beam due to the difference in refractive index between the liquids in the sample and reference cells of a single flow cell.
A beam of light from the lamp passes through a flow cell which is separated diagonally into sample and reference cells. At the rear of the flow cell a mirror reflects the light back through the flow cell and via a zero glass, which affects the path of the light beam, to the light receiver. The light receiver has two diodes each of which produces an electrical current proportional to the amount of light that falls upon it (see Figure 5 on page 18).
Figure 5 Light path
Agilent InfinityLab LC Series RID User Manual 18
1 Introduction to the Refractive Index Detector Detection Principle
Measurements
Initially both sample and reference cell are flushed with mobile phase. The reference cell is then closed and solvent flows only through the sample cell. The refractive index of the mobile phase in both cells is the same and the position of the zero glass can be adjusted so that the detector is in optical balance with an equal amount of light falls on each diode.
When sample elutes from the column into the sample cell the refractive index of the cell contents changes. The change in refractive index deflects the light beam as it passes through the flow cell resulting in an unequal amount of light falling on each diode. The change in current from the diodes that this causes is amplified and used to produce the calibrated detector signal. This signal expressed, as nano Refractive Index Units (nRIU), corresponds to the difference between the refractive index of sample in the sample cell and the mobile phase in the reference cell.
Figure 6 Optical Path
Zero glass
Deflected light
Collimator lens
Reference cell
Mirror
Condenser lens
Incident light
Sample cell
Lamp
Diodes
Agilent InfinityLab LC Series RID User Manual 19
1 Introduction to the Refractive Index Detector Flow Path
Flow Path
The column eluent enters the optical unit through the in port and passes through a heat exchanger. The combination of the heat exchanger and control of the optical unit temperature in the range of 5 C above ambient to 55 CC minimizes changes in refractive index due to temperature variations. The eluent flows through the sample cell and via the same heat exchanger to the purge valve. With the purge valve in the OFF position the eluent passes to the recycle valve. If the recycle valve is also in the OFF/WASTE position the eluent will flow via the waste port into the waste container.
If the recycle valve is in the ON/BOTTLE position the eluent will flow via the recycle port back to the solvent bottle. The recycle valve can be manually set to the ON or OFF position or the Automatic recycling after analysis mode can be enabled. In this mode the recycle valve will automatically switch to the ON position after each analysis has been completed and return to the OFF position before the next analysis starts. Using this mode provides the benefits of uninterrupted flow through the detector without the problems of excessive solvent usage or the contamination of mobile phase with recycled sample compounds.
If the purge valve is in the on position the eluent cannot pass immediately to the recycle valve but will instead flow via a second heat exchanger through the reference cell and then into the recycle valve. Periodically switching the purge valve to the on position while only mobile phase is flowing will ensure that the liquid in the reference cell is as similar as possible to the flowing solvent. The purge valve can be manually set to the on position for a defined time or the Automatic purge mode can be enabled. In this mode the purge valve will automatically switch to the ON position for a defined purgetime prior to the start of each analysis. If a purgetime is set then a waittime must also be set to allow the detector baseline to stabilize after the switching of the purge valve position.
After both the purgetime and waittime have been completed the analysis will start. If the Automatic zero before analysis mode is enabled the detector output will be set to zero immediately before the analysis begins.
The different flow paths depend on the status of the Purge Valve and Recycle Valve.
Agilent InfinityLab LC Series RID User Manual 20
1 Introduction to the Refractive Index Detector Flow Path
Normal Flow Mode
Purge Valve = OFF, Recycle Valve = OFF
Mirror
Sample cell
Reference cell
Heater
Collimator lens
Zero glass Diodes
Slit
lens
Lamp
Purge valve
valve Recycle
Safety cap
Interconnection
Heat exchanger
Condenser
Agilent InfinityLab LC Series RID User Manual 21
1 Introduction to the Refractive Index Detector Flow Path
Purge Mode
This mode is used for flushing the complete detector inclusive the reference side of the flow cell.
Purge Valve = ON, Recycle Valve = OFF
Mirror
Sample cell
Reference cell
Heater
Collimator lens
Zero glass Diodes
Slit
lens
Lamp
Purge valve
valve Recycle
Safety cap
Interconnection
Heat exchanger
Condenser
Agilent InfinityLab LC Series RID User Manual 22
1 Introduction to the Refractive Index Detector Flow Path
Recycle Mode
This mode is used for saving solvent during periods where no sample is running through.
Purge Valve = OFF, Recycle Valve = ON
Mirror
Sample cell
Reference cell
Heater
Collimator lens
Zero glass Diodes
Slit
lens
Lamp
Purge valve
valve Recycle
Safety cap
Interconnection
Heat exchanger
Condenser
Agilent InfinityLab LC Series RID User Manual 23
1 Introduction to the Refractive Index Detector Flow Path
Physical Plumbing
Figure 7 Physical Plumbing Connections
from optical unit sample cell (top right)
bottom left
(1)
Purge valve
top left
optical unit (from and to reference cell)
(3)
(2)
(4)
to optical unit sample cell (bottom right)
Metal union block
Recycle valveNC
NO NC
NO
COM = COMMON NO = NORMALLY OPEN NC = NORMALLY CLOSED
C ON
N E
C T
I N
G P
O R
T S
COM WASTE
IN
COM
RECYCLE
Capillaries (1) to (4) are part of the optical unit assembly. They are made of SST with an ID of 1.0 mm, except for (2), which has an ID of 0.2 mm. All other tubings (to and from the purge and the reference valve) are made of PTFE (available as Tubing kit (G1362-68709)).
Agilent InfinityLab LC Series RID User Manual 24
1 Introduction to the Refractive Index Detector Flow Path
Figure 8 Flow path with the Purge- and Recycle-Valves = OFF
* COM COM
NONC NONC
Metal union block
Sample cell
Reference cell
WASTERECYCLE
Purge valve Recycle valve
COM = COMMON NO = NORMALLY OPEN NC = NORMALLY CLOSED
IN
Grey lines = flowing path Black lines = immobilized mobile phase
*The T-connection in the metal union block results in both sides of the flow cell (sample and reference) always being exposed to the same pressure
Agilent InfinityLab LC Series RID User Manual 25
1 Introduction to the Refractive Index Detector Flow Path
Figure 9 Flow path with the Purge- and Recycle-Valves = ON
NO NC
* COM
Purge valve Recycle valve
COM
NC NO
Metal union block
Sample cell
Reference cell
WASTERECYCLE
IN
COM = COMMON NO = NORMALLY OPEN COM = NORMALLY CLOSED
Grey lines = flowing path Black lines = immobilized mobile phase
*The T-connection in the metal union block results in both sides of the flow cell (sample and reference) always being exposed to the same pressure
Agilent InfinityLab LC Series RID User Manual 26
1 Introduction to the Refractive Index Detector Leak and Waste Handling
Leak and Waste Handling
The Agilent InfinityLab LC Series has been designed for safe leak and waste handling. It is important that all security concepts are understood and instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The maximum volume for an individual bottle stored in the solvent cabinet should not exceed 2 L. For details, see the usage guideline for the Agilent Infinity II Solvent Cabinets (a printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and Infinity II modules can be stacked on top of each other. Waste tubes are guided through a channel on the right hand side of the instrument, keeping the front access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks, guiding them to the leak sensor for leak detection, and passing them on to the next module below, if the leak sensor fails. The leak sensor in the leak plane stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste container:
from the detector's flow cell outlet
from the Multisampler needle wash port
from the Sample Cooler or Sample Thermostat (condensate)
from the pump's Seal Wash Sensor (if applicable)
from the pump's Purge Valve or Multipurpose Valve
Agilent InfinityLab LC Series RID User Manual 27
1 Introduction to the Refractive Index Detector Leak and Waste Handling
Figure 10 Infinity II Leak Waste Concept (Flex Bench installation)
Agilent InfinityLab LC Series RID User Manual 28
1 Introduction to the Refractive Index Detector Leak and Waste Handling
Figure 11 Infinity II Single Stack Leak Waste Concept (bench installation)
Agilent InfinityLab LC Series RID User Manual 29
1 Introduction to the Refractive Index Detector Leak and Waste Handling
Figure 12 Infinity II Two Stack Leak Waste Concept (bench installation)
The waste tube connected to the leak plane outlet on each of the bottom instruments guides the solvent to a suitable waste container.
Agilent InfinityLab LC Series RID User Manual 30
1 Introduction to the Refractive Index Detector Leak and Waste Handling
Leak Sensor
Waste Guidance
CAUTION Solvent incompatibility
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor. The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
Do not use DMF as mobile phase.
Check the leak sensor regularly for corrosion.
NOTE The waste drainage must go straight into the waste containers. The waste flow must not be restricted at bends or joints.
Agilent InfinityLab LC Series RID User Manual 31
1 Introduction to the Refractive Index Detector Leak and Waste Handling
Waste Concept 1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.
NOTE To optimize detector performance the waste container and solvent bottle should be positioned above the level of the refractive index detector and solvent pump (e.g. in the solvent compartment). This will maintain a slight pressure in the sample cell.
Agilent InfinityLab LC Series RID User Manual 32
2 Site Requirements and Specifications
Site Requirements 34
Physical Specifications 37
Performance Specifications 38
This chapter provides information on environmental requirements, physical and performance specifications.
Agilent InfinityLab LC Series RID User Manual 33
2 Site Requirements and Specifications Site Requirements
Site Requirements
A suitable environment is important to ensure optimal performance of the instrument.
Power Considerations The module power supply has wide ranging capability. It accepts any line voltage in the range described in Table 1 on page 37. Consequently there is no voltage selector in the rear of the module. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.
WARNING Incorrect line voltage at the module
Shock hazard or damage of your instrument can result if the devices are connected to line voltage higher than specified.
Connect your module to the specified line voltage.
WARNING Module is partially energized when switched off, as long as the power cord is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the cover is opened and the module is connected to power.
Make sure that it is always possible to access the power plug.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the Instrument while the covers are removed.
WARNING Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power line at any time.
Make sure the power connector of the instrument can be easily reached and unplugged.
Provide sufficient space behind the power socket of the instrument to unplug the cable.
Agilent InfinityLab LC Series RID User Manual 34
2 Site Requirements and Specifications Site Requirements
Power Cords Country-specific power cords are available for the module. The female end of all power cords is identical. It plugs into the power-input socket at the rear. The male end of each power cord is different and designed to match the wall socket of a particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is suitable for your particular country or region.
WARNING Unintended use of power cords
Using power cords for unintended purposes can lead to personal injury or damage of electronic equipment.
Never use a power cord other than the one that Agilent shipped with this instrument.
Never use the power cords that Agilent Technologies supplies with this instrument for any other equipment.
Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
WARNING Absence of ground connection
The absence of ground connection can lead to electric shock or short circuit.
Never operate your instrumentation from a power outlet that has no ground connection.
WARNING Electrical shock hazard
Solvents may damage electrical cables.
Prevent electrical cables from getting in contact with solvents.
Exchange electrical cables after contact with solvents.
Agilent InfinityLab LC Series RID User Manual 35
2 Site Requirements and Specifications Site Requirements
Bench Space The module dimensions and weight (see Table 1 on page 37) allow you to place the module on almost any desk or laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Environment Your module will work within the specifications at ambient temperatures and relative humidity described in Table 1 on page 37.
ASTM drift tests require a temperature change below 2 C/hour (3.6 F/hour) over one hour period. Our published drift specification (refer also to Performance Specifications on page 38) is based on these conditions. Larger ambient temperature changes will result in larger drift.
Better drift performance depends on better control of the temperature fluctuations. To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 C/hour (1.8 F/hour). Turbulences around one minute or less can be ignored.
NOTE Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench rack. This option helps to save bench space as all modules can be placed into one single stack. It also allows to easily relocate the instrument to another lab.
CAUTION Condensation within the module
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.
NOTE This module is designed to operate in a typical electromagnetic environment, i.e. where RF transmitters such as mobile telephones may not be used in close proximity.
Agilent InfinityLab LC Series RID User Manual 36
2 Site Requirements and Specifications Physical Specifications
Physical Specifications
Table 1 Physical Specifications
Type Specification Comments
Weight 14.7 kg (32.4 lbs)
Dimensions (height width depth)
180 x 396 x 436 mm (7.1 x 15.6 x 17.2 inches)
Line voltage 100 240 V~, 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, 5 %
Power consumption 80 VA / 70 W
Ambient operating temperature
455 C (39131 F)
Ambient non-operating temperature
-40 70 C (-40 158 F)
Humidity < 95 % r.h. at 40 C (104 F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN, CSA, UL
Overvoltage category II, Pollution degree 2 For indoor use only
ISM Classification ISM Group 1 Class B According to CISPR 11
Agilent InfinityLab LC Series RID User Manual 37
2 Site Requirements and Specifications Performance Specifications
Performance Specifications
Table 2 Performance Specifications G7162A
Type Specification
Detection type Refractive Index
Refractive index range 1.00 1.75 RIU, calibrated
Measurement range 60010-6 RIU
Optical zeroing Digitally via software, manually via set screw
Optics temperature control 5 C above ambient to 55 C
Sample cell Volume: 8 L Maximum pressure: 5 bar (0.5 MPa) Maximum flow rate: 5 mL/min
Valves Automatic purge and automatic solvent recycle
Volumes Inlet port to sample cell 62 L, inlet port to outlet port 590 L
Liquid contact materials 316 stainless steel, PTFE and quartz glass
pH range 2.3 9.5
Performance specifications Short term noise: <1.2510-9 RIU
Drift: <20010-9 RIU/hr
Time programmable parameters
Polarity, peak width
Maximum data rate 74 Hz
Detector zero Automatic zero before analysis
Instrument Control Lab Advisor B.02.07 or above LC and CE Drivers A.02.12 or above
Local control Agilent Instant Pilot (G4208A)
Analog outputs Recorder/integrator: 100 mV or 1 V, output range selectable, one output
Communication LAN, Controller Area Network (CAN), Extended Remote Interface (ERI): ready, start, stop and shut-down signals
Agilent InfinityLab LC Series RID User Manual 38
2 Site Requirements and Specifications Performance Specifications
Safety and maintenance Extensive diagnostics, error detection and display, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
GLP Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-selectable limits and feedback messages. Electronic records of maintenance and errors. Automated operational qualification/performance verification (OQ/PV).
Housing All materials recyclable
Table 2 Performance Specifications G7162A
Type Specification
NOTE Based on ASTM method E-1303-95 Practice for Refractive Index Detectors used in Liquid Chromatography. Reference conditions; optics temperature 35 C, response time 4 s, flow 1.0 mL/min LC-grade Water, restriction capillary, column compartment temperature 35 C, Agilent on-line degasser (e.g. G4225A), pump and thermostatted column compartment. Instrument equilibrated for 2 hours.
Agilent InfinityLab LC Series RID User Manual 39
2 Site Requirements and Specifications Performance Specifications
Performance Specifications
Table 3 Performance Specifications G7162B
Type Specification
Detection type Refractive Index
Refractive index range 1.00 1.75 RIU, calibrated
Measurement range 60010-6 RIU
Optical zeroing Digitally via software, manually via set screw
Optics temperature control 5 C above ambient to 55 C
Sample cell Volume: 2.5 L Maximum pressure: 5 bar (0.5 MPa) Maximum flow rate: 1 mL/min (100 % water)
Valves Automatic purge and automatic solvent recycle
Volumes Inlet port to sample cell 2.5 L, inlet port to outlet port 265 L
Liquid contact materials 316 stainless steel, PTFE and quartz glass
pH range 2.3 9.5
Performance specifications Short term noise: <1.7510-9 RIU
Drift: <20010-9 RIU/hr
Time programmable parameters
Polarity, peak width
Maximum data rate 148 Hz
Detector zero Automatic zero before analysis
Instrument Control Lab Advisor B.02.07 or above LC and CE Drivers A.02.12 or above
Local control Agilent Instant Pilot (G4208A)
Analog outputs Recorder/integrator: 100 mV or 1 V, output range selectable, one output
Communication LAN, Controller Area Network (CAN), Extended Remote Interface (ERI): ready, start, stop and shut-down signals
Safety and maintenance Extensive diagnostics, error detection and display, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
Agilent InfinityLab LC Series RID User Manual 40
2 Site Requirements and Specifications Performance Specifications
GLP Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-selectable limits and feedback messages. Electronic records of maintenance and errors. Automated operational qualification/performance verification (OQ/PV).
Housing All materials recyclable.
Table 3 Performance Specifications G7162B
Type Specification
NOTE Based on ASTM method E-1303-95 Practice for Refractive Index Detectors used in Liquid Chromatography. Reference conditions; optics temperature 35 C, response time 4 s, flow 1.0 mL/min LC-grade Water, restriction capillary, column compartment temperature 35 C, Agilent on-line degasser (e.g. G4225A), pump and thermostatted column compartment. Instrument equilibrated for 2 hours.
Agilent InfinityLab LC Series RID User Manual 41
3 Using the Module
Magnets 43
Turn on/off 44
Status Indicators 46
Instrument Configuration 47
Set up the Detector with Agilent OpenLab ChemStation 49
The Detector User Interface 50
Detector Control Settings 53
Method Parameter Settings 55
Agilent Local Control Modules 59
This chapter explains the essential operational parameters of the module.
Agilent InfinityLab LC Series RID User Manual 42
3 Using the Module Magnets
Magnets
1 Magnets in doors of pumps, autosamplers, detectors, and fraction collectors.
Agilent InfinityLab LC Series RID User Manual 43
3 Using the Module Turn on/off
Turn on/off
This procedure exemplarily shows an arbitrary LC stack configuration.
1 2
Power switch: On
Agilent InfinityLab LC Series RID User Manual 44
3 Using the Module Turn on/off
3 Turn instrument On/Off with the control software. 4
Power switch: Off
5
Agilent InfinityLab LC Series RID User Manual 45
3 Using the Module Status Indicators
Status Indicators
This procedure exemplarily shows an arbitrary LC stack configuration.
1 The module status indicator indicates one of six possible module conditions:
Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or completed.
4. Error mode - interrupts the analysis and requires attention (for example, a leak or defective internal components).
5. Resident mode (blinking) - for example, during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a cold-start. Then try a firmware update.
Agilent InfinityLab LC Series RID User Manual 46
3 Using the Module Instrument Configuration
Instrument Configuration
1 Set the switches of the Configuration switch at the rear of the module:
a All switches DOWN: module uses the default IP address 192.168.254.11.
b Switch 4 UP and others DOWN: module uses DHCP.
c Switch 5 UP and others DOWN: modules uses STORED address.
Agilent InfinityLab LC Series RID User Manual 47
3 Using the Module Instrument Configuration
2 Enter the setup information (MAC / IP address and/or Instrument Name).
a Agilent OpenLab ChemStation (Configure Instrument):
b Lab Advisor (Instrument Overview - Add Instrument):
Agilent InfinityLab LC Series RID User Manual 48
3 Using the Module Set up the Detector with Agilent OpenLab ChemStation
Set up the Detector with Agilent OpenLab ChemStation
The setup of the detector is shown with the Agilent OpenLab ChemStation C.01.07 and LC and CE Drivers A.02.13.
Figure 13 ChemStation Method and Run Control (just detector is shown)
After successful load of the OpenLab ChemStation, you should see the module as an active item in the graphical user interface (GUI).
NOTE This section describes the detector settings only. For information on the Agilent OpenLab ChemStation or other 1200 Infinity modules refer to the corresponding documentation.
Agilent InfinityLab LC Series RID User Manual 49
3 Using the Module The Detector User Interface
The Detector User Interface
Within the detector GUI, there are active areas. If you move the mouse cursor across the icons the cursor will change. 1 Make device ready or turn off 2 EMF status 3 Detector status 4 Balance 5 Heater status (on/off) 6 Heater temperature
The Purge Valve status (closed/open) and the Recycle Valve status (off/on) is shown when the cursor moves across the flow cell icon.
Agilent InfinityLab LC Series RID User Manual 50
3 Using the Module The Detector User Interface
EMF Status shows Run / Ready / Error state and Not Ready text or Error text Offline (gray) Ok. No Maintenance required (green) EMF warning. Maintenance might be required (yellow) EMF warning. Maintenance required (red) Important: The EMF settings can be accessed via Agilent Lab Advisor. The limit(s) can be changed. Based on the limit, the User Interface displays the above status.
Module Status shows Run / Ready / Error state and Not Ready text or Error text Error (red) Not ready (yellow) Ready (green) Pre run, Post run (purple) Run (blue) Idle (green) Offline (dark gray) Standby (light gray)
Agilent InfinityLab LC Series RID User Manual 51
3 Using the Module The Detector User Interface
A right-click into the Active Area will open a menu to Show the Control Interface (special module settings) Show the Method interface (similar as via menu
Instrument >Setup Instrument Method) Set Error Method Identify Module (Status LED will blink) Perform a Balance Switch the Heater on/off (same as click on button Make
Device Ready/Turn device off (standby)) Open/close Purge Valve Recycle Valve on/off
If the Purge Valve or Recycle Valve has been turned on, the on-time is displayed below the flow cell icon.
Agilent InfinityLab LC Series RID User Manual 52
3 Using the Module Detector Control Settings
Detector Control Settings
Figure 14 Control settings
The figure shows the default settings.
Heater: Select the On option to switch the RID heater on. This parameter requires setting the Optical Unit Temperature. Select the Off option to switch the optical unit heater off.
Analog Output Range: can be set to either 100 mV or 1 V full scale, for additional settings see Analog Output (under Method Parameter Settings on page 55).
At Power On: When activated, the optical unit heater is turned on automatically when the RID is switched on. For shortest equilibration times, Agilent recommends to leave this function always on.
Agilent InfinityLab LC Series RID User Manual 53
3 Using the Module Detector Control Settings
Purge Reference Cell: This parameter is used to exchange the content of the reference cell in the case of solvent change or reference cell contamination. Enter a time interval (minutes) to purge the reference cell. This will be started immediately if you click OK on this window. Allow additional time for baseline stabilization after purging.
Recycling Valve: Select the On option to switch the recycling of the eluent on. The Off option diverts the flow of the RID to the waste bottle.
Automatic Turn On: This function allows you to turn on the optical unit heater at a specified date and time. It requires that the At Power On function is turned off. Select Turn Heater on at: to activate the date and time fields, and enter the date and time in the appropriate fields in the specified format.
Agilent InfinityLab LC Series RID User Manual 54
3 Using the Module Method Parameter Settings
Method Parameter Settings
These settings are available via Menu >Instrument >Set up Instrument Method or via right click into the modules active area (does not show the Instrument Curves tab).
Figure 15 Method parameter settings
NOTE For additional help and support: Highlight the desired cell and press F1. A help screen will open with additional information and documentation about the topic.
Agilent InfinityLab LC Series RID User Manual 55
3 Using the Module Method Parameter Settings
Table 4 Method parameter settings
Optical Unit Temperature: This item sets the temperature of the optical unit. The optical unit of the Agilent 1260 Infinity RID can be operated between 5 C above ambient and 55 C. The recommended setting is 5 C above ambient. This will improve baseline stability.
Peakwidth (Response time, Data Rate): Peakwidth enables you to select the peak width (response time) for your analysis. The peak width is defined as the width of a peak, in minutes, at half the peak height. Set the peak width to the narrowest expected peak in your chromatogram. The peak width sets the optimum response time for your RID. Limits: When you set the peak width (in minutes), the corresponding response time is set automatically and the appropriate data rate for signal acquisition is selected. If Acquire Signal check box is not marked, signals are not stored.
NOTE The 1260 Infinity II RID (G7162A) has a data rate of up to 74 Hz. The 1290 Infinity II RID (G7162B) has a data rate of up to 148 Hz.
Stoptime/Posttime The Stoptime is the time where either the complete system stops (As Pump/Injector) or the module (if different from system stop time). The data collection is stopped at this time. You can set the Posttime so that your module remains in the not ready state during the Posttime to delay the start of the next analysis. A Posttime period can be used to allow your column to equilibrate after changes in solvent composition or temperature change.
Analog Output The range can be set to either 100 mV or 1 V full scale, see Detector Control Settings on page 53. Zero Offset: 1 99 % in steps of 1 % (5 % equal to 50 mV). Attenuation: 488 1000000 nRIU at discrete values for
either 100 mV or 1 V full scale.
Agilent InfinityLab LC Series RID User Manual 56
3 Using the Module Method Parameter Settings
Signal Polarity Use this item to set the polarity of the RID signal. Because of the nature of analytes and eluents, refractive index detectors can show negative and positive peaks, even within a run.
Automatic Zero Before Analysis Use this setting to activate an automatic zeroing of the signal before the run is started. If automatic purge is selected, the purge will be performed before the automatic zero. Automatic Recycling After Analysis: This parameter can be used to select between automatic recycling of the effluent (On) or directing the effluent to the waste outlet of the RID (Off) after the run.
Automatic Purge This parameter can be used to purge the reference cell and wait additional time for baseline stabilization. It is initiated each time the run is started. This should be used only if the content of the reference cell is expected to degrade during a run. The automatic purge is completed before the autozero is performed and before the injection is done.
Table 4 Method parameter settings
Agilent InfinityLab LC Series RID User Manual 57
3 Using the Module Method Parameter Settings
Timetable You may set up time events to change functions with their parameters over the run time. Add lines as required. Time Limits: 0.00 to 99999.00 min in steps of 0.01 min. Via the buttons in the bottom area time table lines can be added, removed, cut copied, pasted or completely cleared. Based on the chosen function, a certain parameter can be selected.
Instrument Curves The detector has several signals that can be used for diagnosing problems. These signals can be used in addition to the normal baseline signal to determine whether correlation to other signals exists. These signals are available via the Agilent ChemStation Online Plot/Data Signal and/or Agilent Lab Advisor Software. Optical unit temperature Diode Balance to diagnose peaks that exceed the
dynamic range of the RID, for example in the case of extremely high concentrations/signals.
Polarity monitors of polarity switching during the run. Diode 1 and Diode 2: The RID signal is based on the ratio
of the light level that is measured by two photodiodes. The RID signal is zero if the two diodes show the same light level. This parameter allows you to store individually the signal measured by the diode 1 and diode 2.
Board Temperature
The temperature around the Leak Sensor (at the power switch electronics) can be used for looking at ambient temperature changes.
Table 4 Method parameter settings
Agilent InfinityLab LC Series RID User Manual 58
3 Using the Module Agilent Local Control Modules
Agilent Local Control Modules
Agilent InfinityLab Companion G7108AA
The Agilent InfinityLab Companion gives you complete control, system monitoring, signal plotting, and diagnostic capabilities for a wide range of LC system modules.
The instrument control solution is available as full package including all hardware and accessories, but can also be used on your own mobile devices like tablets, mobile phones and other electronic equipment.
Combining the conveniences of the Agilent Instant Pilot features with state-of-the-art mobile technology, the Agilent InfinityLab Companion gives you maximum flexibility and ease of use to control and monitor your LC system modules.
Features:
Complete local control and monitoring of Agilent Infinity II LC modules
Excellent usability and ease of use through a user interface specifically tailored for mobile devices - simple, intuitive touch-enabled, and visual controllable.
High flexibility through a modern Bring your own device approach. Connection between LC module and mobile device either wireless via Wi-Fi or wired over USB cable (with full package).
Convenient, ergonomic operation either handheld or attached to a module at the stack with newly developed, secure tablet holder (included in the full package).
Preconfigured tablet with all required software already installed (included in the full package).
Centerpiece of the solution is a USB dongle that activates the complete intelligence of the InfinityLab Companion on the instrument stack.
The InfinityLab Companion provides:
fast and direct control in front of the instrument
a clear overview of the system status
control functionalities
access to method parameters and sequences
a logbook showing events from the modules
diagnostic tests
Agilent InfinityLab LC Series RID User Manual 59
4 Preparing the Module
Leak and Waste Handling 61 Waste Concept 62
Setting up an Analysis 63 Before Using the System 64 Requirements and Conditions 65 Preparing the Detector 68 Preparing the HPLC System 68 Running the Sample and Verifying the Results 70
Solvent Information 71
This chapter provides information on how to set up the module for an analysis and explains the basic settings.
Agilent InfinityLab LC Series RID User Manual 60
4 Preparing the Module Leak and Waste Handling
Leak and Waste Handling
For details on correct installation, see separate installation documentation.
WARNING Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
Do not use solvents with an auto-ignition temperature below 200 C (392 F). Do not use solvents with a boiling point below 56 C (133 F).
Avoid high vapor concentrations. Keep the solvent temperature at least 40 C (72 F) below the boiling point of the solvent used. This includes the solvent temperature in the sample compartment. For the solvents methanol and ethanol keep the solvent temperature at least 25 C (45 F) below the boiling point.
Do not operate the instrument in an explosive atmosphere.
Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for example, carbon disulfide).
Reduce the volume of substances to the minimum required for the analysis.
Never exceed the maximum permissible volume of solvents (8 L) in the solvent cabinet. Do not use bottles that exceed the maximum permissible volume as specified in the usage guideline for solvent cabinet.
Ground the waste container.
Regularly check the filling level of the waste container. The residual free volume in the waste container must be large enough to collect the waste liquid.
To achieve maximal safety, regularly check the tubing for correct installation.
NOTE For details, see the usage guideline for the solvent cabinet. A printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available in the Agilent Information Center or via the Internet.
Agilent InfinityLab LC Series RID User Manual 61
4 Preparing the Module Leak and Waste Handling
Waste Concept 1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.
NOTE To optimize detector performance the waste container and solvent bottle should be positioned above the level of the refractive index detector and solvent pump (e.g. in the solvent compartment). This will maintain a slight pressure in the sample cell.
Agilent InfinityLab LC Series RID User Manual 62
4 Preparing the Module Setting up an Analysis
Setting up an Analysis
This chapter can be used for
preparing the system,
to learn the set up of an HPLC analysis and
to use it as an instrument check to demonstrate that all modules of the system are correctly installed and connected. It is not a test of the instrument performance.
Learn about special settings
Agilent InfinityLab LC Series RID User Manual 63
4 Preparing the Module Setting up an Analysis
Before Using the System
Solvent Information
Observe recommendations on the use of solvents in chapter Solvents in the pumps reference manual.
Priming and Purging the System
When the solvents have been exchanged or the pumping system has been turned off for a certain time (for example, overnight) oxygen will re-diffuse into the solvent channel between the solvent reservoir, vacuum degasser (when available in the system) and the pump. Solvents containing volatile ingredients will slightly lose these. Therefore priming of the pumping system is required before starting an application.
Table 5 Choice of Priming Solvents for Different Purposes
Activity Solvent Comments
After an installation
When switching between reverse phase and normal phase (both times)
Isopropanol
Isopropanol
Best solvent to flush air out of the system
Best solvent to flush air out of the system
After an installation Ethanol or Methanol Alternative to Isopropanol (second choice) if no Isopropanol is available
To clean the system when using buffers
After a solvent change
Bidistilled water
Bidistilled water
Best solvent to re-dissolve buffer crystals
Best solvent to re-dissolve buffer crystals
After the installation of normal phase seals (P/N 0905-1420)
Hexane + 5% Isopropanol Good wetting properties
NOTE The pump should never be used for priming empty tubings (never let the pump run dry). Use a syringe to draw enough solvent for completely filling the tubings to the pump inlet before continuing to prime with the pump.
Agilent InfinityLab LC Series RID User Manual 64
4 Preparing the Module Setting up an Analysis
1 Open the purge valve of your pump (by turning it counterclockwise) and set flow rate to 3 5 mL/min.
2 Flush all tubes with at least 30 mL of solvent.
3 Set flow to required value of your application and close the purge valve.
Requirements and Conditions
What You Will Need
The table below lists the items you need to have for the set up of the analysis. Some of these are optional (not required for the basic system).
NOTE Pump for approximately 10minutes before starting your application.
NOTE Do NOT exceed the flow rate of the 1290 Infinity II RID (Micro) whilst purging pump. Bypass detector when purging >1 mL/min.
Table 6 What you will need
Agilent 1200 Infinity Series system Pump (plus degassing)
Autosampler
Detector
Degasser (optional)
Column Compartment (optional)
Agilent CDS
System should be correctly set up for LAN communication with the Agilent ChemStation
Column: Zorbax Eclipse XDB-C18, 4.6 x 150 mm, 5 m (993967-902) or an equivalent column
Standard: Agilent isocratic checkout sample (01080-68704)
Agilent InfinityLab LC Series RID User Manual 65
4 Preparing the Module Setting up an Analysis
Conditions
A single injection of the isocratic test standard is made under the conditions given in Table 7 on page 66:
Table 7 Chromatographic conditions
Mobile phases 30 % water, 70 % acetonitrile
Flow rate 1.5 mL/min
Stroke A 20 L
Stop time 10 min
Injection volume 20 L
Column compartment temperature 25 C
Optical unit Temperature 35 C
Polarity Positive
Peak Width (Response time) 0.2 min (4 s, standard)
Agilent InfinityLab LC Series RID User Manual 66
4 Preparing the Module Setting up an Analysis
Typical Chromatogram
A typical chromatogram for this analysis is shown in Figure 16 on page 67. The exact profile of the chromatogram will depend on the chromatographic conditions. Variations in solvent quality, column packing, standard concentration and column temperature will all have a potential effect on peak retention and response.
Figure 16 Isocratic standard sample chromatogram
NOTE The resulting chromatogram should only be seen as a qualitative example, the checkout procedure is not meant as a quantitative procedure. Its intent is only to verify the presence of the four peaks from the checkout sample.
Please be aware of the large negative air / solvent peak from the injection (cut out from the bottom of the following figure) prior to the first peak of interest. This is to be expected in a regular chromatogram, especially if a non-degassed sample is injected into degassed solvent and the sample solvent properties dont perfectly match with the mobile phase properties. Only similar zooming factors in the display of a chromatogram will lead to similar looking results.
Agilent InfinityLab LC Series RID User Manual 67
4 Preparing the Module Setting up an Analysis
Preparing the Detector For best performance of the detector
Let the lamp warm-up and stabilize for at least one hour (initial turn on of the module requires a longer time depending on the environment and the application needs).
For high sensitivity measurements, a stable environment is required; refer to Environment on page 36. Prevent drafts from air condition systems.
Do not work with removed/open front panels/doors. When the system includes a G1316 TCC (typically located below the detector) and its front panel is removed while the TCC is set to high temperatures, the up-streaming air could influence the stability of the detector baseline.
Preparing the HPLC System 1 Turn on the control software and the monitor.
2 Turn on the modules.
3 Start the control software. The screen should show all modules and the system status is Not Ready.
4 Turn on the modules that require conditioning:
a Detector heater (warm-up for at least 60 min to get a stable baseline).
b Column compartment (set temperature as required).
c Pump (purge).
d Purge the detector reference cell for 20 min.
e Sampler (prepare the standard isocratic sample into a vial).
f Solvents (fill water and Acetontrile into the solvent bottles).
5 Load the default method.
6 Pump the water/acetonitrile (30/70 %) mobile phase through the column for 10 min for equilibration.
Agilent InfinityLab LC Series RID User Manual 68
4 Preparing the Module Setting up an Analysis
7 Select the menu item Run Control >Sample Info and enter information about this application. Click OK to leave this screen.
Figure 17 Sample Info
Agilent InfinityLab LC Series RID User Manual 69
4 Preparing the Module Setting up an Analysis
Running the Sample and Verifying the Results 1 To start a run select the menu item RunControl > Run Method.
2 This will start the modules and the online plot on the Agilent ChemStation will show the resulting chromatogram.
Figure 18 Isocratic standard sample chromatogram
Agilent InfinityLab LC Series RID User Manual 70
4 Preparing the Module Solvent Information
Solvent Information
Observe the following recommendations on the use of solvents.
Follow the recommendations for avoiding the growth of algae, see the pump manuals.
Small particles can permanently block capillaries and valves. Therefore, always filter solvents through 0.22 m filters.
Avoid or minimize the use of solvents that may corrode parts in the flow path. Consider specifications for the pH range given for different materials such as flow cells, valve materials etc. and recommendations in subsequent sections.
General Information about Solvent/Material Compatibility Materials in the flow path are carefully selected based on Agilents experiences in developing highest-quality instruments for HPLC analysis over several decades. These materials exhibit excellent robustness under typical HPLC conditions. For any special condition, please consult the material information section or contact Agilent.
Disclaimer
Subsequent data was collected from external resources and is meant as a reference. Agilent cannot guarantee the correctness and completeness of such information. Data is based on compatibility libraries, which are not specific for estimating the long-term life time under specific but highly variable conditions of UHPLC systems, solvents, solvent mixtures and samples. Information can also not be generalized due to catalytic effects of impurities like metal ions, complexing agents, oxygen etc. Apart from pure chemical corrosion, other effects like electro corrosion, electrostatic charging (especially for non-conductive organic solvents), swelling of polymer parts etc. need to be considered. Most data available refers to room temperature (typically 20 25 C, 68 77 F). If corrosion is possible, it usually accelerates at higher temperatures. If in doubt, please consult technical literature on chemical compatibility of materials.
Agilent InfinityLab LC Series RID User Manual 71
4 Preparing the Module Solvent Information
MP35N
MP35N is a nonmagnetic, nickel-cobalt-chromium-molybdenum alloy demonstrating excellent corrosion resistance (for example, against nitric and sulfuric acids, sodium hydroxide, and seawater) over a wide range of concentrations and temperatures. In addition, this alloy shows exceptional resistance to high-temperature oxidation. Due to excellent chemical resistance and toughness, the alloy is used in diverse applications: dental products, medical devices, nonmagnetic electrical components, chemical and food processing equipment, marine equipment. Treatment of MP35N alloy samples with 10 % NaCl in HCl (pH 2.0) does not reveal any detectable corrosion. MP35N also demonstrates excellent corrosion resistance in a humid environment. Although the influence of a broad variety of solvents and conditions has been tested, users should keep in mind that multiple factors can affect corrosion rates, such as temperature, concentration, pH, impurities, stress, surface finish, and dissimilar metal contacts.
Polyphenylene Sulfide (PPS)
Polyphenylene sulfide has outstanding stability even at elevated temperatures. It is resistant to dilute solutions of most inorganic acids, but it can be attacked by some organic compounds and oxidizing reagents. Nonoxidizing inorganic acids, such as sulfuric acid and phosphoric acid, have little effect on polyphenylene sulfide, but at high concentrations and temperatures, they can still cause material damage. Nonoxidizing organic chemicals generally have little effect on polyphenylene sulfide stability, but amines, aromatic compounds, and halogenated compounds may cause some swelling and softening over extended periods of time at elevated temperatures. Strong oxidizing acids, such as nitric acid (> 0.1 %), hydrogen halides (> 0.1 %), peroxy acids (> 1 %), or chlorosulfuric acid degrade polyphenylene sulfide. It is not recommended to use polyphenylene sulfide with oxidizing material, such as sodium hypochlorite and hydrogen peroxide. However, under mild environmental conditions, at low concentrations and for short exposure times, polyphenylene sulfide can withstand these chemicals, for example, as ingredients of common disinfectant solutions.
Agilent InfinityLab LC Series RID User Manual 72
4 Preparing the Module Solvent Information
PEEK
PEEK (Polyether-Ether Ketones) combines excellent properties regarding biocompatibility, chemical resistance, mechanical and thermal stability. PEEK is therefore the material of choice for UHPLC and biochemical instrumentation.
It is stable in the specified pH range (for the Bio-Inert LC system: pH 1 13, see bio-inert module manuals for details), and inert to many common solvents.
There is still a number of known incompatibilities with chemicals such as chloroform, methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %, sulfuric acid > 10 %, sulfonic acids, trichloroacetic acid), halogens or aqueous halogen solutions, phenol and derivatives (cresols, salicylic acid, and so on).
When used above room temperature, PEEK is sensitive to bases and various organic solvents, which can cause it to swell. Under such conditions, normal PEEK capillaries are sensitive to high pressure. Therefore, Agilent uses stainless steel cladded PEEK capillaries in bio-inert systems. The use of stainless steel cladded PEEK capillaries keeps the flow path free of steel and ensures pressure stability up to 600 bar. If in doubt, consult the available literature about the chemical compatibility of PEEK.
Polyimide
Agilent uses semi-crystalline polyimide for rotor seals in valves and needle seats in autosamplers. One supplier of polyimide is DuPont, which brands polyimide as Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic solvents. It is incompatible with concentrated mineral acids (e.g. sulphuric acid), glacial acetic acid, DMSO and THF. It is also degraded by nucleophilic substances like ammonia (e.g. ammonium salts in basic conditions) or acetates.
Polyethylene (PE)
Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents including acids and bases in a pH range of 1 to 12.5. It is compatible with many organic solvents used in chromatographic systems like methanol, acetonitrile and isopropanol. It has limited stability with aliphatic, aromatic and halogenated hydrocarbons, THF, phenol and derivatives, concentrated acids and bases. For normal phase applications, the maximum pressure should be limited to 200 bar.
Agilent InfinityLab LC Series RID User Manual 73
4 Preparing the Module Solvent Information
Tantalum (Ta)
Tantalum is inert to most common HPLC solvents and almost all acids except fluoric acid and acids with free sulfur trioxide. It can be corroded by strong bases (e.g. hydroxide solutions > 10 %, diethylamine). It is not recommended for the use with fluoric acid and fluorides.
Stainless Steel (SST)
Stainless steel is inert against many common solvents. It is stable in the presence of acids and bases in a pH range of 1 to 12.5. It can be corroded by acids below pH 2.3. It can also corrode in following solvents:
Solutions of alkali halides, their respective acids (for example, lithium iodide, potassium chloride, and so on) and aqueous solutions of halogens.
High concentrations of inorganic acids like nitric acid, sulfuric acid and organic solvents especially at higher temperatures (replace, if your chromatography method allows, by phosphoric acid or phosphate buffer which are less corrosive against stainless steel).
Halogenated solvents or mixtures which form radicals and/or acids, for example:
2 CHCl3 + O2 2 COCl2 + 2 HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol.
Chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, diisopropylether). Such ethers should be filtered through dry aluminium oxide which adsorbs the peroxides.
Solutions of organic acids (acetic acid, formic acid, and so on) in organic solvents. For example, a 1 % solution of acetic acid in methanol will attack steel.
Solutions containing strong complexing agents (for example, EDTA, ethylene diamine tetra-acetic acid).
Mixtures of carbon tetrachloride with isopropanol or THF.
Agilent InfinityLab LC Series RID User Manual 74
4 Preparing the Module Solvent Information
Titanium (Ti)
Titanium is highly resistant to oxidizing acids (for example, nitric, perchloric and hypochlorous acid) over a wide range of concentrations and temperatures. This is due to a thin oxide layer on the surface, which is stabilized by oxidizing compounds. Non-oxidizing acids (for example, hydrochloric, sulfuric and phosphoric acid) can cause slight corrosion, which increases with acid concentration and temperature. For example, the corrosion rate with 3 % HCl (about pH 0.1) at room temperature is about 13 m/year. At room temperature, titanium is resistant to concentrations of about 5 % sulfuric acid (about pH 0.3). Addition of nitric acid to hydrochloric or sulfuric acids significantly reduces corrosion rates. Titanium is sensitive to acidic metal chlorides like FeCl3 or CuCl2. Titanium is subject to corrosion in anhydrous methanol, which can be avoided by adding a small amount of water (about 3 %). Slight corrosion is possible with ammonia > 10 %.
Diamond-Like Carbon (DLC)
Diamond-Like Carbon is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Fused silica and Quartz (SiO2)
Fused silica is used in Max Light Cartridges. Quartz is used for classical flow cell windows. It is inert against all common solvents and acids except hydrofluoric acid and acidic solvents containing fluorides. It is corroded by strong bases and should not be used above pH 12 at room temperature. The corrosion of flow cell windows can negatively affect measurement results. For a pH greater than 12, the use of flow cells with sapphire windows is recommended.
Gold
Gold is inert to all common HPLC solvents, acids and bases within the specified pH range. It can be corroded by complexing cyanides and concentrated acids like aqua regia.
Agilent InfinityLab LC Series RID User Manual 75
4 Preparing the Module Solvent Information
Zirconium Oxide (ZrO2)
Zirconium Oxide is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Fluorinated polymers (PTFE, PFA, FEP, FFKM, PVDF)
Fluorinated polymers like PTFE (polytetrafluorethylene), PFA (perfluoroalkoxy), and FEP (fluorinated ethylene propylene) are inert to almost all common acids, bases, and solvents. FFKM is perfluorinated rubber, which is also resistant to most chemicals. As an elastomer, it may swell in some organic solvents like halogenated hydrocarbons.
TFE/PDD copolymer tubings, which are used in all Agilent degassers except G1322A/G7122A, are not compatible with fluorinated solvents like Freon, Fluorinert, or Vertrel. They have limited life time in the presence of hexafluoroisopropanol (HFIP). To ensure the longest possible life with HFIP, it is best to dedicate a particular chamber to this solvent, not to switch solvents, and not to let dry out the chamber. For optimizing the life of the pressure sensor, do not leave HFIP in the chamber when the unit is off.
The tubing of the leak sensor is made of PVDF (polyvinylidene fluoride), which is incompatible with the solvent DMF (dimethyl formamide).
Sapphire, Ruby and Al2O3-based ceramics
Sapphire, ruby and ceramics based on aluminum oxide Al2O3 are inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications.
Agilent InfinityLab LC Series RID User Manual 76
5 Optimizing the Detector
Refractive Index Detector Optimization 78 Potential Causes for Baseline Problems 80 Detector Equilibration 81
This chapter provides information on how to optimize the detector.
Agilent InfinityLab LC Series RID User Manual 77
5 Optimizing the Detector Refractive Index Detector Optimization
Refractive Index Detector Optimization
Follow these thirteen points to optimize the performance of your refractive index detector.
1 Position the solvent and waste reservoirs correctly
Position the solvent and waste reservoirs above the level of the refractive index detector and solvent pump. This maintains a slight pressure in the sample cell and will improve the performance of the detector.
2 Do not overpressurize the flow cell
Be aware to not exceed a 5 bar pressure drop after the flow cell when hooking up additional devices like other detectors or a fraction collector. If an additional detector is installed place upstream in the flow path it before the refractive index detector.
3 Use the correct solvents
To minimize baseline noise and drift solvents must be LC grade and filtered prior to use.
4 Check for leaks
Leaks within the LC instrument that the refractive index detector is connected to will cause problems with baseline long term noise or drift. Confirm that the instrument is free from leaks by performing the diagnostic pressure test (for the high pressure parts of the system between pump and column). Ensure that the connections from the on-line vacuum degasser to the pump and the detector inlet, waste and recycle connections are air tight.
5 Verify frit, filter and fitting quality
Partially blocked frits, filters and fittings can cause baseline long term noise. Verify that the pressure drop across all such parts is within expected limits.
NOTE Depending on the detector different inlet tubings are used (from the Accessory Kit):
G7162A:
Interfacing capillary (G1362-87300)
G7162B:
Capillary ST 0.075 mm x 220 mm (5067-4784) (Sampler to TCC/MCT)
Capillary ST 0.075 mm x 340 mm (5067-4783) (Column to RID)
Agilent InfinityLab LC Series RID User Manual 78
5 Optimizing the Detector Refractive Index Detector Optimization
6 Control the optical unit temperature
Always control the optical unit temperature (heater = ON) for maximum detector sensitivity or with samples that could precipitate in the sample cell at room temperature and set an elevated optical unit temperature at least 5 C above ambient conditions.
7 Use an appropriate response time
For most applications a setting of 4 seconds is adequate. Only for high speed analyses (short columns at high flow rates) a lower setting is recommended. Bear in mind that even if the response time setting is too high fast peaks will appear a little smaller and broader but retention time and peak areas are still correct and reproducible.
8 Recycle mobile phase
Use the recycle valve to allow automatic recycling of mobile phase delivered when no analysis is running. The pump flow can therefore continue uninterrupted until the next analysis without wasting mobile phase solvents. In addition the refractive index detector is always stabilized and ready for immediate use.
9 Consider using a degasser
For many solvents you can achieve better baseline stability, when using a degasser. For some solvents a degasser might not lead to a better baseline quality.
10 Flush the degasser
If flow is stopped and mobile phase remains inside the on-line vacuum degasser the solvent composition will change. When re-starting the flow or when using new mobile phase flush each degasser channel used for 10 minutes at the maximum flow rate of the pump (with the purge valve of the pump open to avoid a potential over-pressure in the RI detectors flow cell).
11 Use pre-mixed solvents, only
Dont use a pump for mixing solvents. When operating the RI detector together with a quaternary pump, bypass the MCGV in the quaternary pump. You have to virtually convert the quaternary pump into an isocratic one, by directly connecting the solvent inlet tubing from degasser or solvent bottle to the active inlet valve of the pump (use PEEK adapter 1/4-28 to 10-32 (0100-1847), which is delivered with the accessory kit of the detector).
Agilent InfinityLab LC Series RID User Manual 79
5 Optimizing the Detector Refractive Index Detector Optimization
12 Consider solvent changes with time
Baseline drift can be caused by the tendency of certain solvents to change over time. For example the acetonitrile content of acetonitrile/water mixtures will decrease, tetrahydrofuran will form peroxides, the amount of water in hygroscopic organic solvents will increase and solvents such as tetrahydrofuran held in the reference cell may begin to regas.
13 Eliminate mobile phase/column combination problems
Certain mobile phases in combination with specific columns can generate long term baseline noise. For example acetonitrile/water mobile phases with certain aminopropyl bonded phase columns. To eliminate the combination of mobile phase and column as a cause of long term noise replace the column with Restriction capillary (G1362-87301) and re-evaluate the detector performance.
Potential Causes for Baseline Problems
Noise (short term)
Typically the sources for short term noise are either electronic (check the settings for the peak widths, check for ambient sources of electronic noise) or they are related to the solvents, their composition and flow (in order to verify this, turn off the pump, consider degassing your solvents, use only premixed solvents).
Wander (long term noise)
Excessive wander is an indication for a general system or environmental instability (system or laboratory might not be thermally stable, control instrument and laboratory temperature). Verify that the solvent properties are constant over time (flush out contamination, use only stabilized and premixed solvents). Clean the parts in the flow path and allow the system to be flushed out and equilibrated.
Drift
Excessive drift is an indication for a general system or environmental instability (system or laboratory might not be thermally stable, control instrument and laboratory temperature). Verify that the solvent properties are constant over time (flush out of contamination, use only stabilized solvents). Clean parts in the flow path and allow the system to be flushed out and equilibrated.
Agilent InfinityLab LC Series RID User Manual 80
5 Optimizing the Detector Refractive Index Detector Optimization
Detector Equilibration The Refractive Index (RI) is a function of temperature, pressure and a property of the used solvent (it changes with solvent composition, degassing level and due to any trace of contamination). Therefore the Refractive Index Detector will detect any change in any of these parameters as a change in its signal and a variation of its baseline. Therefore the detector will trace down any instabilities in the system and the environment as well. It may sometimes appear, as if the detector itself was unstable or generating an unstable baseline, where in fact, the detector is simply displaying the instabilities of the environment and the rest of the system. By this the detector is often - without justification - blamed for instabilities, which it does not generate itself, but only detect. The fact that this detector is a universal detector makes it also sensitive to instabilities introduced to it from outside the detector.
This makes it very important to have a very stable environment and system for achieving best possible baseline stability. The baseline will get the better, the longer the system is used under identical and stable conditions. Keep the temperature in your laboratory and system constant and controlled. Ideally a system with an RID should be used always with the same type of analysis (stable solvent composition, temperature, flow rates, dont switch the pump off after analysis, instead just recycle solvents or at least reduce only the flow. Switch valves and settings only when needed. Dont expose the detector to draft of air or to vibrations). A change of any of these parameters may require a considerable amount of time for re-equilibration.
Agilent InfinityLab LC Series RID User Manual 81
6 Troubleshooting and Diagnostics
Available Tests vs User Interfaces 83
Agilent Lab Advisor Software 84
This chapter gives an overview about the troubleshooting and diagnostic features and the different user interfaces.
Agilent InfinityLab LC Series RID User Manual 82
6 Troubleshooting and Diagnostics Available Tests vs User Interfaces
Available Tests vs User Interfaces
NOTE Depending on the used interface, the available tests and the screens/reports may vary.
Preferred tool should be the Agilent Lab Advisor, see Agilent Lab Advisor Software on page 84.
Agilent Lab Advisor B.02.07 SP1 or later is required.
The Instant Pilot supports the G7162A/B with B.02.19 or later. It does not provide any test functions!
Agilent InfinityLab LC Series RID User Manual 83
6 Troubleshooting and Diagnostics Agilent Lab Advisor Software
Agilent Lab Advisor Software
The Agilent Lab Advisor Software (basic license, shipped with an Agilent LC pump) is a standalone product that can be used with or without a chromatographic data system. Agilent Lab Advisor helps to manage the lab for high-quality chromatographic results by providing a detailed system overview of all connected analytical instruments with instrument status, Early Maintenance Feedback counters (EMF), instrument configuration information, and diagnostic tests. With the push of a button, a detailed diagnostic report can be generated. Upon request, the user can send this report to Agilent for a significantly improved troubleshooting and repair process.
The Agilent Lab Advisor software is available in two versions:
Lab Advisor Basic
Lab Advisor Advanced
Lab Advisor Basic is included with every Agilent 1200 Infinity Series and Agilent InfinityLab LC Series instrument.
The Lab Advisor Advanced features can be unlocked by purchasing a license key, and include real-time monitoring of instrument actuals, all various instrument signals, and state machines. In addition, all diagnostic test results, calibration results, and acquired signal data can be uploaded to a shared network folder. The Review Client included in Lab Advisor Advanced allows to load and examine the uploaded data no matter on which instrument it was generated. This makes Data Sharing an ideal tool for internal support groups and users who want to track the instrument history of their analytical systems.
The optional Agilent Maintenance Wizard Add-on provides an easy-to-use, step-by-step multimedia guide for performing preventive maintenance on Agilent 1200 Infinity LC Series instrument.
The tests and diagnostic features that are provided by the Agilent Lab Advisor software may differ from the descriptions in this manual. For details, refer to the Agilent Lab Advisor software help files.
Agilent InfinityLab LC Series RID User Manual 84
7 Error Information
What Are Error Messages 86
General Error Messages 87 Timeout 87 Shutdown 88 Remote Timeout 88 Lost CAN Partner 89 Leak Sensor Open 89 Leak Sensor Short 90 Compensation Sensor Open 90 Compensation Sensor Short 91 Fan Failed 91 Leak 92 Open Cover 92 Cover Violation 93 ERI Messages 93
Refractive Index Detector Specific Error Messages 94 Thermal Fuse Open 94 Heater Resistance Too High 94 Heater Fuse 95 Wrong Temperature Profile 95 Undecipherable Temperature Signal 96 Maximum Temperature Exceeded 96 Purge Valve Fuse Blown 97 Recycle Valve Fuse Blown 97 Purge Valve Not Connected 98 Recycle Valve Not Connected 98 Lamp Voltage too Low 99 Lamp Voltage too High 99 Lamp Current too High 99 Lamp Current too Low 100 Wait Function Timed Out 100
Not-Ready Messages 101 Purge Time Running 101 Wait for Purge 101 Unbalanced Diodes 102 Not Enough Light 102 Too Much Light 102
This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.
Agilent InfinityLab LC Series RID User Manual 85
7 Error Information What Are Error Messages
What Are Error Messages
Error messages are displayed in the user interface when an electronic, mechanical, or hydraulic (flow path) failure occurs which requires attention before the analysis can be continued (for example, repair, or exchange of consumables is necessary). In the event of such a failure, the red status indicator at the front of the module is switched on, and an entry is written into the module logbook.
If an error occurs outside a method run, other modules will not be informed about this error. If it occurs within a method run, all connected modules will get a notification, all LEDs get red and the run will be stopped. Depending on the module type, this stop is implemented differently. For example, for a pump the flow will be stopped for safety reasons. For a detector, the lamp will stay on in order to avoid equilibration time. Depending on the error type, the next run can only be started, if the error has been resolved, for example liquid from a leak has been dried. Errors for presumably single time events can be recovered by switching on the system in the user interface.
Special handling is done in case of a leak. As a leak is a potential safety issue and may have occurred at a different module from where it has been observed, a leak always causes a shutdown of all modules, even outside a method run.
In all cases, error propagation is done via the CAN bus or via an APG/ERI remote cable (see documentation for the APG/ERI interface).
Agilent InfinityLab LC Series RID User Manual 86
7 Error Information General Error Messages
General Error Messages
General error messages are generic to all Agilent series HPLC modules and may show up on other modules as well.
Timeout Error ID: 0062
The timeout threshold was exceeded.
Probable cause Suggested actions
1 The analysis was completed successfully,
and the timeout function switched off the module as requested.
Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required.
2 A not-ready condition was present during a sequence or multiple-injection run for a period longer than the timeout threshold.
Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required.
Agilent InfinityLab LC Series RID User Manual 87
7 Error Information General Error Messages
Shutdown Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals. A LOW signal input on pin 4 of the remote connector generates the error message.
Remote Timeout Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is started, the system expects all not-ready conditions (for example, a not-ready condition during detector balance) to switch to run conditions within one minute of starting the analysis. If a not-ready condition is still present on the remote line after one minute the error message is generated.
Probable cause Suggested actions
1 Leak detected in an external instrument with a remote connection to the system.
Fix the leak in the external instrument before restarting the module.
2 Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down condition.
3 The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error condition. Refer to the Service Manual for the degasser or the pump that has the degasser built-in.
Probable cause Suggested actions
1 Not-ready condition in one of the instruments connected to the remote line.
Ensure the instrument showing the not-ready condition is installed correctly, and is set up correctly for analysis.
2 Defective remote cable. Exchange the remote cable.
3 Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the instruments documentation).
Agilent InfinityLab LC Series RID User Manual 88
7 Error Information General Error Messages
Lost CAN Partner Error ID: 0071
During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated.
Leak Sensor Open Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated.
Probable cause Suggested actions
1 CAN cable disconnected. Ensure all the CAN cables are connected correctly.
Ensure all CAN cables are installed correctly.
2 Defective CAN cable. Exchange the CAN cable.
3 Defective mainboard in another module. Switch off the system. Restart the system, and determine which module or modules are not recognized by the system.
Probable cause Suggested actions
1 Leak sensor not connected to the power switch board.
Please contact your Agilent service representative.
2 Defective leak sensor. Please contact your Agilent service representative.
3 Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service representative.
4 Power switch assembly defective Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 89
7 Error Information General Error Messages
Leak Sensor Short Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current increases above the upper limit, the error message is generated.
Compensation Sensor Open Error ID: 0081
The ambient-compensation sensor (NTC) on the power switch board in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the power switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
Probable cause Suggested actions
1 Defective leak sensor. Please contact your Agilent service representative.
2 Leak sensor incorrectly routed, being pinched by a metal component.
Please contact your Agilent service representative.
3 Power switch assembly defective Please contact your Agilent service representative.
4 Cable or contact problem. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Loose connection between the power switch board and the mainboard
Please contact your Agilent service representative.
2 Defective power switch assembly Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 90
7 Error Information General Error Messages
Compensation Sensor Short Error ID: 0080
The ambient-compensation sensor (NTC) on the power switch board in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the power switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor falls below the lower limit, the error message is generated.
Fan Failed Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan speed. If the fan speed falls below a certain limit for a certain length of time, the error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned off to assure that the module does not overheat inside.
Probable cause Suggested actions
1 Defective power switch assembly Please contact your Agilent service representative.
2 Loose connection between the power switch board and the mainboard
Please contact your Agilent service representative.
Probable cause Suggested actions
1 Fan cable disconnected. Please contact your Agilent service representative.
2 Defective fan. Please contact your Agilent service representative.
3 Defective mainboard. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 91
7 Error Information General Error Messages
Leak Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted temperature-compensation sensor) are used by the leak algorithm to determine whether a leak is present. When a leak occurs, the leak sensor is cooled by the solvent. This changes the resistance of the leak sensor which is sensed by the leak sensor circuit on the main board.
Open Cover Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam is removed, the fan is switched off, and the error message is generated.
Probable cause Suggested actions
1 Loose fittings. Ensure all fittings are tight.
2 Broken capillary. Exchange defective capillaries.
3 Leaking valve. Exchange valve.
4 Leaking flow cell. Exchange optical unit.
Probable cause Suggested actions
1 The top foam was removed during
operation.
Please contact your Agilent service representative.
2 Foam not activating the sensor. Please contact your Agilent service representative.
3 Defective sensor or main board. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 92
7 Error Information General Error Messages
Cover Violation Error ID: 7461
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam is removed while the lamps are on (or if an attempt is made to switch on for example the lamps with the foam removed), the lamps are switched off, and the error message is generated.
ERI Messages Error ID: 11120 (+5 V) , 11121 (+25 V)
The ERI (Enhanced Remote Interface) provides two error events related to over current situations on the +5 V and +24 V lines.
Probable cause Suggested actions
1 The top foam was removed during operation. Please contact your Agilent service representative.
2 Foam not activating the sensor. Please contact your Agilent service representative.
Probable cause Suggested actions
1 The load on the ERI is too high. Reduce the load.
Agilent InfinityLab LC Series RID User Manual 93
7 Error Information Refractive Index Detector Specific Error Messages
Refractive Index Detector Specific Error Messages
Thermal Fuse Open Error ID: 2680
The thermal fuse of the optical unit heater has failed.
Heater Resistance Too High Error ID: 2681
The resistance of the heater foil is above the set limit.
Probable cause Suggested actions
1 Heater cable disconnected. Please contact your Agilent service representative.
2 Defective mainboard. Please contact your Agilent service representative.
3 Defective thermal fuse. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Heater cable disconnected. Please contact your Agilent service representative.
2 Defective mainboard. Please contact your Agilent service representative.
3 Defective heater. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 94
7 Error Information Refractive Index Detector Specific Error Messages
Heater Fuse Error ID: 2682
The electronic fuse of the heater has been activated.
Wrong Temperature Profile Error ID: 2683
After turning ON the optical unit heat control, the temperature does not increase at a sufficiently fast rate to reach the set point.
Probable cause Suggested actions
1 Short in heater circuit. Powercycle the module.
2 Defective mainboard. Please contact your Agilent service representative.
3 Defective heater. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Defective mainboard. Please contact your Agilent service representative.
2 Defective heater. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 95
7 Error Information Refractive Index Detector Specific Error Messages
Undecipherable Temperature Signal Error ID: 2684
Maximum Temperature Exceeded Error ID: 2685
The maximum heater temperature has been exceeded.
Probable cause Suggested actions
1 Heater cable disconnected. Please contact your Agilent service representative.
2 Defective mainboard. Please contact your Agilent service representative.
3 Defective heater. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Defective mainboard. Please contact your Agilent service representative.
2 Defective heater. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 96
7 Error Information Refractive Index Detector Specific Error Messages
Purge Valve Fuse Blown Error ID: 2686
The electronic fuse on the purge valve has been activated.
Recycle Valve Fuse Blown Error ID: 2687
The electronic fuse on the recycle valve has been activated.
Probable cause Suggested actions
1 Short in purge valve circuit. Power cycle the module.
2 Defective purge valve. Please contact your Agilent service representative.
3 Defective mainboard. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Short in recycle valve circuit. Power cycle the module.
2 Defective recycle valve. Please contact your Agilent service representative.
3 Defective mainboard. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 97
7 Error Information Refractive Index Detector Specific Error Messages
Purge Valve Not Connected Error ID: 2688
When activated no response was received from the purge valve.
Recycle Valve Not Connected Error ID: 2689
When activated no response was received from the recycle valve.
Probable cause Suggested actions
1 Purge valve disconnected. Please contact your Agilent service representative.
2 Defective purge valve. Please contact your Agilent service representative.
3 Defective mainboard. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Recycle valve disconnected. Please contact your Agilent service representative.
2 Defective recycle valve. Please contact your Agilent service representative.
3 Defective mainboard. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 98
7 Error Information Refractive Index Detector Specific Error Messages
Lamp Voltage too Low Error ID: 2693
Lamp Voltage too High Error ID: 2690
Lamp Current too High Error ID: 2691
Probable cause Suggested actions
1 Defective mainboard. Please contact your Agilent service representative.
2 Defective lamp or optics. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Contaminated flow cell. Flush the flow cell.
2 Defective mainboard. Please contact your Agilent service representative.
3 Defective lamp or optics. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Defective mainboard. Please contact your Agilent service representative.
2 Defective lamp or optics. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 99
7 Error Information Refractive Index Detector Specific Error Messages
Lamp Current too Low Error ID: 2694
Wait Function Timed Out Error ID: 2692
Wait for temperature or wait for defined signal has not been fulfilled within the specified time frame.
Probable cause Suggested actions
1 Optical unit cable disconnected. Please contact your Agilent service representative.
2 Defective mainboard. Please contact your Agilent service representative.
3 Defective lamp or optics. Please contact your Agilent service representative.
Probable cause Suggested actions
1 Time too short. Increase time.
Agilent InfinityLab LC Series RID User Manual 100
7 Error Information Not-Ready Messages
Not-Ready Messages
Not-ready messages are displayed during the wait for a specific condition to be reached or completed or while a self-test procedure is running. In the event of such a failure, the yellow status indicator at the front of the detector is switched ON.
This section describes the meaning of detector not-ready messages.
Purge Time Running Event ID 2600
Wait for Purge Event ID 2692
Probable cause Suggested actions
1 The purge valve is open, liquid is flowing
through both sample and reference cell.
Allow the reference purge time to elapse.
Probable cause Suggested actions
1 The detector is waiting after the automatic purge of the reference cell.
Allow the wait time to elapse.
Agilent InfinityLab LC Series RID User Manual 101
7 Error Information Not-Ready Messages
Unbalanced Diodes
Not Enough Light
Too Much Light The amount of light falling on the light receiving diodes is too high to generate a refractive index signal.
Probable cause Suggested actions
1 The diode balance value is outside the
pre-set range -0.5 to +0.5, an unequal amount of light is falling on the two light receiving diodes.
Flush the reference cell with the mobile phase being used.
Perform the RID Optical Balance procedure (see The Optical Balance Procedure on page 115).
Probable cause Suggested actions
1 There is insufficient light falling on the light
receiving diodes to generate a refractive index signal.
Flush the flow cell with the mobile phase being used to ensure that it is free of air bubbles or other contamination.
Probable cause Suggested actions
1 The sample cell content is varying too much from the reference cell.
Purge reference and sample cell.
Agilent InfinityLab LC Series RID User Manual 102
8 Test Functions
Introduction 104
Refractive Index Calibration 108
Optical Balance 114 The Optical Balance Procedure 115
Using the Build-in Test Chromatogram 117 Procedure Using the Agilent LabAdvisor 117
ASTM Drift and Noise Test 119 Setting the Test Conditions 119 Evaluation 122
D/A Converter (DAC) Test 123 Test Evaluation 124
Other Lab Advisor Functions 125 EMFs - Early Maintenance Feature 125
This chapter describes the detectors built in test functions.
Agilent InfinityLab LC Series RID User Manual 103
8 Test Functions Introduction
Introduction
All tests are described based on the Agilent Lab Advisor Software B.02.07 SP1 [147].
Table 8 Interfaces and available test functions
Interface Comment Available Function
Agilent Lab Advisor For functions see Table 9 on page 105
Available functions depend on Product Level (Basic Advanced FSE)
Agilent ChemStation No tests available Adding of temperature/lamp signals to chromatographic signals possible
Diagnostic Signals Temperature optical unit Diode 1 & 2 Balance Polarity
Agilent Instant Pilot G4208A No tests available FW Update EMF
Agilent InfinityLab LC Series RID User Manual 104
8 Test Functions Introduction
Table 9 Function Overview Lab Advisor Basic/Advanced
Functions Product Level
Tests
- ASTM Drift and Noise Test Basic Advanced
- D/A Converter Test Basic Advanced
Calibrations
- RID Calibration Basic Advanced
Tools
- Diagnostic Buffers Basic Advanced
- Module Info Basic Advanced
- Test Chromatogram Basic Advanced
- RID Tool Screen Basic Advanced
Controls
- Advanced Method Parameters
- Attenuation Analog Output Advanced
- Analog Output 1 Offset [% Full Scale] Advanced
- Polarity Advanced
- Configuration
- Remote Pulse Duration [s] * Basic Advanced
- Analog Output 1 Range Advanced
- Control
- Recycle Basic Advanced
- Heater On/Off Basic Advanced
- Method Parameters
- Set Peakwidth Advanced
- Set Optical temperature [C] Advanced
- Module Information
- Identify Module Basic Advanced
Agilent InfinityLab LC Series RID User Manual 105
8 Test Functions Introduction
- Purge + Prime
- Start Purge Advanced
- Set Purge Time [minutes] Advanced
- Special Commands
- Detector Reset Basic Advanced
- Clear Error Basic Advanced
Signals
- RI Signal [nRIU] Advanced
- Optical Temperature [C] Advanced
- Board Temperature [C] Advanced
- Signal Diode 1 [counts] Advanced
- Signal Diode 2 [counts] Advanced
- Diode Balance Advanced
- Polarity Advanced
EMF Counters
- Time since last Purge Basic Advanced
Table 9 Function Overview Lab Advisor Basic/Advanced
Functions Product Level
Agilent InfinityLab LC Series RID User Manual 106
8 Test Functions Introduction
Figure 19 The Lab Advisor shows the available tests
Agilent InfinityLab LC Series RID User Manual 107
8 Test Functions Refractive Index Calibration
Refractive Index Calibration
Refractive Index Calibration The refractive index calibration is based on a sucrose calibration solution, which has a known refractive index compared to LC grade water. After both the sample and reference cells have been purged with LC grade water the sucrose solution is introduced into the flow cell and then the built-in refractive index calibration functionality is used.
Filling the sample cell with the sucrose calibration solution will give a theoretical detector response of 512,000 nRIU +/- 5,000 nRIU. The calibration algorithm will allow the actual detector response, if different, to be changed to the theoretical value.
NOTE Refractive index calibration is only required after exchange of the optical unit or the main (RIM) - board.
Agilent InfinityLab LC Series RID User Manual 108
8 Test Functions Refractive Index Calibration
Preparing the Calibration
1 Preparation of the sucrose calibration solution.
a To prepare 25 ml of the calibration solution 87.5 mg of the Sucrose sample is required.
b Add the weighed amount of sample into a suitable volumetric flask.
c Dispense 10 ml of LC grade water into the flask and shake or stir to dissolve.
d Dilute the contents of the flask to volume with LC grade water.
Wait five minutes and shake again. The solution is now ready for use.
2 Preparing the pump.
a Fill a suitable solvent bottle with LC grade water.
b Connect this bottle to Channel A of the pump, A1 if a binary pump.
3 Flush the degasser and pump.
When Recommended after exchange of the optical unit or RIM board.
Tools required Description
Laboratory balance
Parts required # p/n Description
1 DAB/Ph Eur/BP/JP/NF/USP Grade Sucrose
1 9301-1446A Syringe
1 9301-0407A Needle
1 5190-5111A Syringe filter, 0.45 m, 100/pk
1 0100-1516A Finger-tight fitting PEEK, 2/pk
Agilent InfinityLab LC Series RID User Manual 109
8 Test Functions Refractive Index Calibration
4 Run the RID Tools Screen with the Agilent Lab Advisor (for further information see Online-Help of user interface). (If an Agilent pump is part of the system, the pump section is active.)
5 Purging the sample and reference cells.
a The purge valve will automatically switch to the ON position.
b Using a syringe or LC pump flush the sample and reference cell with about 20 mL of LC grade water. (If an Agilent pump is part of the system, the pump section is active).
c The purge valve will automatically switch to the OFF position when you click continue.
Agilent InfinityLab LC Series RID User Manual 110
8 Test Functions Refractive Index Calibration
6 Fill the sample cell with calibration solution.
a Remove the inlet capillary or flushing syringe from the in port.
b Take the syringe and fix the needle to the syringe adapter.
c Suck about 1.5 mL of the calibration sample into the syringe.
d Keep the syringe in a horizontal position.
e Remove the needle.
f Add the filter to the syringe and fit the needle to filter.
Figure 20 Syringe with Sample Filter
g Lift the needle tip and carefully eject approximately 0.5 mL to remove air out of the syringe and to flush the needle.
h Add the PEEK fitting to the needle tip and fix both at the flow cell inlet.
i Slowly inject about 1.0 ml and wait for about 10 s to inject another 0.1 mL. This will assure that the cell is filled properly.
7 Leave the RID Tools Screen.
Sample filter
NOTE Do not inject the calibration solution without the sample filter.
Agilent InfinityLab LC Series RID User Manual 111
8 Test Functions Refractive Index Calibration
Performing the Calibration 1 Run the RID Calibration Screen with the Agilent Lab Advisor (for further
information see Online-Help of user interface).
2 If the purge of the sample and reference side has been done, start the calibration process with OK.
3 Inject the calibration sample.
Agilent InfinityLab LC Series RID User Manual 112
8 Test Functions Refractive Index Calibration
4 If the detector response differs from the theoretical response of 512000 nRIU5000 nRIU enter the theoretical value (512000) in the dialog box. If the detector response is within the theoretical response click OK.
5 The calibration was successful.
NOTE Rinse the sample cell with pure water at a minimum of 1.5 mL/min to flush the sucrose from the cell and the capillaries. When organic solvent is sequentially applied (without rinsing), a blockage of capillaries may occur.
Agilent InfinityLab LC Series RID User Manual 113
8 Test Functions Optical Balance
Optical Balance
When the sample and reference cells both contain the same liquids an equal amount of light should fall on each light receiving diode, the diode balance will equal 0. If this balance of light needs to be corrected the optical balance procedure can be used.
Diode balance is calculated as follows:
Where:
diode1 = signal proportional to the amount of light falling on diode1
diode2 = signal proportional to the amount of light falling on diode2
Optical balance adjustment is a manual procedure where the position of the light beam falling on the light receiving diode is adjusted using the zero glass adjustment screw.
NOTE The detector will become not-ready when the diode balance value falls outside the range - 0.5 to + 0.5.
NOTE Both sample and reference cell must be purged with the same solvent before optical balance is performed. Prior to performing this procedure, the system must be well equilibrated.
Agilent InfinityLab LC Series RID User Manual 114
8 Test Functions Optical Balance
The Optical Balance Procedure
1 Run the RID Tools Screen with the Agilent Lab Advisor (for further information see Online-Help of user interface). (If an Agilent pump is part of the system, the pump section is active.)
2 Purging the sample and reference cells.
a Switch the purge valve to the ON position.
b Purge the sample and reference cells for around 10 min with the solvents to be used.
c Switch the purge valve to the OFF position
3 Select Optical Balance.
When When light falling on light receiving diodes is out of balance.
Tools required Description
Flat head screwdriver
NOTE This procedure should only be performed to correct a permanent misalignment of the light beam that cannot be eliminated by flushing the sample and the reference cell with the same solvent and by equilibrating the system.
Agilent InfinityLab LC Series RID User Manual 115
8 Test Functions Optical Balance
4 Adjust Optical Balance.
a While monitoring the optical balance use the flat-headed screwdriver to turn the zero glass adjustment screw slowly.
b When the diode balance value reaches 0.00 (green area), optical balance is restored.
Figure 21 Turning the zero glass adjustment screw
Zero glass adjustment (screw left)
Agilent InfinityLab LC Series RID User Manual 116
8 Test Functions Using the Build-in Test Chromatogram
Using the Build-in Test Chromatogram
This function is available from the Agilent ChemStation, Lab Advisor and Instant Pilot.
The built-in Test Chromatogram can be used to check the signal path from the detector to the data system and the data analysis or via the analog output to the integrator or data system. The chromatogram is continuously repeated until a stop is executed either by means of a stop time or manually.
Procedure Using the Agilent LabAdvisor This procedure works for all Agilent 1200 Infinity detectors (DAD, MWD, VWD, FLD and RID). The example figure is from the RID detector.
1 Assure that the default LC method is loaded via the control software.
2 Start the Agilent Lab Advisor software (B.01.03 SP4 or later) and open the detector's Tools selection.
3 Open the test chromatogram screen
4 Turn the Test Chromatogram on.
5 Change to the detector's Module Service Center and add the detector signal to the Signal Plot window.
NOTE The peak height is always the same but the area and the retention time depend on the set peakwidth, see example below.
Agilent InfinityLab LC Series RID User Manual 117
8 Test Functions Using the Build-in Test Chromatogram
6 To start a test chromatogram enter in the command line: STRT
Figure 22 Test Chromatogram with Agilent Lab Advisor
7 To stop the test chromatogram enter in the command line: STOP
NOTE The test chromatogram is switched off automatically at the end of a run.
Agilent InfinityLab LC Series RID User Manual 118
8 Test Functions ASTM Drift and Noise Test
ASTM Drift and Noise Test
Setting the Test Conditions
1 Turn ON the detector.
You are now ready to change the settings of your detector.
2 Connect the restriction capillary directly between the column compartment heat exchanger outlet and the in port of the detector.
3 Set up the instrument with the following test conditions.
When If you want to checkout the detector
Tools required Description
LC system with G7162A RID
Parts required # p/n Description
1 G1362-87301A Restriction capillary
Table 10 Chromatographic Conditions
Mobile phases LC grade water
Column Restriction capillary 2.7 m x 1.7 mm i.d.
Flow rate 1.0 mL/min
Compressibility 46
Stroke 20 L
Stop time 20 min
Column compartment temperature 35 C
Optical unit temperature 35 C
Polarity Positive
Peak width (response time) 0.2 min (4 s, standard)
Agilent InfinityLab LC Series RID User Manual 119
8 Test Functions ASTM Drift and Noise Test
4 Set the RID setpoints.
5 Turn the heater ON and purge the detector reference cell for 20 min.
6 When purging has finished allow the baseline to stabilize.
7 Run the ASTM Drift and Noise Test Screen with the Agilent Lab Advisor (for further information see Online-Help of user interface).
NOTE The optical unit temperature must be set at least 5 C above ambient conditions. Therefore if ambient temperature is above 30 C higher values for optical unit temperature and column compartment temperature must be set.
NOTE For optimum performance a stabilization time of 2 hours minimum is recommend.
Agilent InfinityLab LC Series RID User Manual 120
8 Test Functions ASTM Drift and Noise Test
8 Click OK.
The measurement starts.
The test has completed and shows either PASSED of FAILED.
Agilent InfinityLab LC Series RID User Manual 121
8 Test Functions ASTM Drift and Noise Test
Evaluation For the Instant Pilot Rescale the plot and measure the baseline noise and drift on the screen. If a printer is configured for the instrument the plot can be printed by pressing the m key and selecting Print Plot.
The following values are calculated automatically by the Agilent ChemStation.
Noise (ASTM): The short term noise in nRIU based on ASTM method E-1303-95 Practice for Refractive Index Detectors used in Liquid Chromatography using 0.5 minute segments.
Wander:The long term noise in nRIU based on ASTM method E-1303-95 Practice for Refractive Index Detectors used in Liquid Chromatography using 0.5 minute segments.
Drift: The drift in nRIU/hour based on ASTM method E-1303-95 Practice for Refractive Index Detectors used in Liquid Chromatography measured over 20 minutes.
Factors that will affect the baseline stability include:
Variations in the optics or eluent temperature
Pressure fluctuations in the sample cell
The quality of the water used
Air bubbles in the flow cell
For diagnosing signal problems refer to chapter "Troubleshooting and Diagnostics".
Agilent InfinityLab LC Series RID User Manual 122
8 Test Functions D/A Converter (DAC) Test
D/A Converter (DAC) Test
The detector provides analog output of chromatographic signals for use with integrators, chart recorders or data systems. The analog signal is converted from the digital format by the digital-analog-converter (DAC).
The DAC test is used to verify correct operation of the digital-analog-converter by applying a digital test signal to the DAC.
The DAC outputs an analog signal of approximately 50 mV (if the zero offset of the analog output is set to the default value of 5 %) which can be plotted on an integrator. A continuous square wave with an amplitude of 10 V and a frequency of approximately 1 cycle/24 seconds is applied to the signal.
The amplitude of the square wave and the peak-to-peak noise are used to evaluate the DAC test.
1 Run the D/A Converter (DAC) Test with the Agilent Lab Advisor (for further information see Online-Help of user interface).
When If the analog detector signal is noisy or missing.
Preparations Lamp must be on for at least 10 minutes. Connect integrator, chart recorder or data system to the detector analog output.
Agilent InfinityLab LC Series RID User Manual 123
8 Test Functions D/A Converter (DAC) Test
Test Evaluation The noise on the step should be less than 3 V.
Probable cause Suggested actions
1 Bad cable or grounding problem between
detector and external device.
Check or replace the cable.
2 Defective detector main board. Please contact your Agilent service representative.
Agilent InfinityLab LC Series RID User Manual 124
8 Test Functions Other Lab Advisor Functions
Other Lab Advisor Functions
EMFs - Early Maintenance Feature The EMFs screen allows you to view and manage the EMF counters for all modules in all systems.
Figure 23 EMFs - Early Maintenance Feature
Agilent InfinityLab LC Series RID User Manual 125
9 Maintenance
Introduction to Maintenance 127
Warnings and Cautions 128
Overview of Maintenance 130
Cleaning the Module 131 Storage of the Detector 131
Remove and Install Doors 132
Flow Cell Flushing 134
Correct Leaks 135
Replace Leak Handling System Parts 138
Replace the Module Firmware 140
This chapter provides general information on maintenance of the detector.
Agilent InfinityLab LC Series RID User Manual 126
9 Maintenance Introduction to Maintenance
Introduction to Maintenance
The module is designed for easy maintenance. Maintenance can be done from the front with module in place in the system.
NOTE There are no serviceable parts inside.
Do not open the module.
Agilent InfinityLab LC Series RID User Manual 127
9 Maintenance Warnings and Cautions
Warnings and Cautions
WARNING Module is partially energized when switched off, as long as the power cord is plugged in.
Risk of stroke and other personal injury. Repair work at the module can lead to personal injuries, e. g. shock hazard, when the module cover is opened and the instrument is connected to power.
Never perform any adjustment, maintenance or repair of the module with the top cover removed and with the power cord plugged in.
The security lever at the power input socket prevents that the module cover is taken off when line power is still connected. Never plug the power line back in when cover is removed.
WARNING Sharp metal edges
Sharp-edged parts of the equipment may cause injuries.
To prevent personal injury, be careful when getting in contact with sharp metal areas.
WARNING Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
The volume of substances should be reduced to the minimum required for the analysis.
Do not operate the instrument in an explosive atmosphere.
Agilent InfinityLab LC Series RID User Manual 128
9 Maintenance Warnings and Cautions
CAUTION Electronic boards and components are sensitive to electrostatic discharge (ESD).
ESD can damage electronic boards and components.
Be sure to hold the board by the edges, and do not touch the electrical components. Always use ESD protection (for example, an ESD wrist strap) when handling electronic boards and components.
CAUTION Safety standards for external equipment
If you connect external equipment to the instrument, make sure that you only use accessory units tested and approved according to the safety standards appropriate for the type of external equipment.
Agilent InfinityLab LC Series RID User Manual 129
9 Maintenance Overview of Maintenance
Overview of Maintenance
On the following pages maintenance procedures are described that can be carried out without opening the main cover.
Table 11 Maintenance Procedures
Procedure Typical Frequency Notes
Flow cell flushing If flow cell is contaminated.
Leak sensor drying If leak has occurred. Check for leaks.
Leak handling System replacement If broken or corroded. Check for leaks.
Replacing the detectors Firmware If not up to date or corrupted.
Agilent InfinityLab LC Series RID User Manual 130
9 Maintenance Cleaning the Module
Cleaning the Module
To keep the module case clean, use a soft cloth slightly dampened with water, or a solution of water and mild detergent. Avoid using organic solvents for cleaning purposes. They can cause damage to plastic parts.
Storage of the Detector
WARNING Liquid dripping into the electronic compartment of your module can cause shock hazard and damage the module
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any connections in the flow path.
NOTE A solution of 70 % isopropanol and 30 % water might be used if the surface of the module needs to be disinfected.
NOTE In case the detector is not used for some time (stored), then fill the flow cell (sample and reference side) with isopropanol.
Agilent InfinityLab LC Series RID User Manual 131
9 Maintenance Remove and Install Doors
Remove and Install Doors
Parts required p/n Description
5067-5745A Door Assembly Infinity 180 Left
5067-5746A Door Assembly Infinity 180 Right
NOTE The figures in this procedure show the Infinity II Multisampler module as an example.
The principle of how to remove and/or install doors works in the same way for all Infinity II modules.
Agilent InfinityLab LC Series RID User Manual 132
9 Maintenance Remove and Install Doors
1 Press the release buttons and pull the front door out. 2 For the Installation of the front door. Insert the hinges into their guides and move the door in until the release buttons click into their final position.
Agilent InfinityLab LC Series RID User Manual 133
9 Maintenance Flow Cell Flushing
Flow Cell Flushing
In case the cell is contaminated, follow the procedure below.
1 Use the purge mode and flush with the strong solvent.
2 Leave this solution in the cell for about one hour.
3 Flush with mobile phase.
When If flow cell is contaminated
Tools required Description
Glass syringe
Adapter
Parts required # Description
1 Strong solvent
1 Tubings to waste
WARNING Dangerous solvents
The strong solvents used in this procedure are toxic and flammable and proper precautions are necessary.
Wear protective gloves and goggles.
Don't expose yourself to the vapors.
NOTE Aqueous solvents can build up algae and therefore should do not be left in the flow cell for long periods. Add a small percentage of organic solvents (e.g. acetonitrile or methanol ~ 5 %).
NOTE The strong solvent should dissolve any potential contaminants in the flow cell. For example water for aqueous mobile phase buffers, chloroform or tetrahydrofuran for not water soluble contaminants.
NOTE Do not exceed the flow cell pressure limit of 5 bar (0.5 MPa).
Agilent InfinityLab LC Series RID User Manual 134
9 Maintenance Correct Leaks
Correct Leaks
When If a leakage has occurred in the valve area or at the capillary connections
Tools required Description
Tissue
Wrench, 1/4 inch for capillary connections
1 Open the doors. 2 Observe the leak sensor area for leaks and correct, if required.
3 Observe the interface ports for leaks and correct, if required.
4 Unscrew the screws of the Service Door.
Agilent InfinityLab LC Series RID User Manual 135
9 Maintenance Correct Leaks
5 Open the Service Door. 6 Remove the Service Door.
7 Observe the valve area for leaks and correct, if required. 8 Install the Service Door.
9 Close the Service Door. 10 Fasten the screws of the Service Door.
Agilent InfinityLab LC Series RID User Manual 136
9 Maintenance Correct Leaks
11 Close the doors.
Agilent InfinityLab LC Series RID User Manual 137
9 Maintenance Replace Leak Handling System Parts
Replace Leak Handling System Parts
Parts required p/n Description
5043-0856A Leak Adapter
5063-6527A Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm approximately 85 mm required
1 Open the doors. 2 Locate the Leak Adapter (1) and Tubing (2).
3 Press the Leak Adapter down and remove it together with the tubing.
4 Install the Leak Adapter by pressing it into the Main Cover.
(1)
(2)
1. 2.
1. 2.
Agilent InfinityLab LC Series RID User Manual 138
9 Maintenance Replace Leak Handling System Parts
5 Insert the Tubing (approximately 115 mm required for replacement) between Leak Adapter outlet and Leak Panel.
6 Close the doors.
Agilent InfinityLab LC Series RID User Manual 139
9 Maintenance Replace the Module Firmware
Replace the Module Firmware
To upgrade/downgrade the modules firmware carry out the following steps:
1 Download the required module firmware, the latest FW Update Tool and the documentation from the Agilent web. http://www.agilent.com/en-us/firmwareDownload?whid=69761
2 For loading the firmware into the module follow the instructions in the documentation.
Module Specific Information
When The installation of newer firmware might be necessary if a newer version solves problems of older versions or to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary to keep all systems on the same (validated) revision or if a new module with newer firmware is added to a system or if third party control software requires a special version.
Tools required Description
Agilent Lab Advisor software
Parts required # Description
1 Firmware, tools and documentation from Agilent web site
Preparations Read update documentation provided with the Firmware Update Tool.
G7162A/B
Initial firmware D.06.76
Compatibility with 1100 / 1200 series modules When using the G7162A/B in a system, all other modules must have firmware from set 6.50 or above (main and resident). Otherwise the communication will not work.
Conversion to / emulation Not available
Agilent InfinityLab LC Series RID User Manual 140
10 Parts for Maintenance
Overview of Maintenance Parts 142
Accessory Kit 144
Leak Handling Parts 147
This chapter provides information on parts for maintenance.
Agilent InfinityLab LC Series RID User Manual 141
10 Parts for Maintenance Overview of Maintenance Parts
Overview of Maintenance Parts
For cables, see Cable Overview on page 149.
Item p/n Description
1 5067-5745 Door Assembly Infinity 180 Left
2 5067-5746 Door Assembly Infinity 180 Right
3 5043-0856 Leak Adapter
4 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm for Waste and Leak Adapter (ca. 85 mm required)
5 5043-1013 Tubing Clip
G1362-68706 Interface tubing kit
G1362-87300 Interfacing capillary
G1362-87301 Restriction capillary
G7162-87300 Waste Tube Kit (recycle/waste)
Agilent InfinityLab LC Series RID User Manual 142
10 Parts for Maintenance Overview of Maintenance Parts
Figure 24 Maintenance Parts
1
2
3
4
5
5
Agilent InfinityLab LC Series RID User Manual 143
10 Parts for Maintenance Accessory Kit
Accessory Kit
Figure 25 Accessory Kit
Agilent InfinityLab LC Series RID User Manual 144
10 Parts for Maintenance Accessory Kit
Accessory Kit for G7162A Accessory kit (G7162-68755) contains some accessories needed for the installation of the G7162A RID.
Item # p/n Description
1 1 G1362-87300 Interfacing capillary
2 1 G1362-87301 Restriction capillary
3 2 0100-1700 FERRULE-AY-18IN
4 2 0100-1708 Nut 1/8 PPS
5 1 G7162-87300 Waste Tube Kit (recycle/waste) includes: Fitting, PK (2x), Tubing 2 m (2x)
1 5042-6449 Flex-Tube OD 1.6 mm
1 0100-1516 Fitting, PK, 1/16'' x 10-32, SH, male nut, 2/pk
6 1 0890-1760 Tubing Flexible 1 ea / 1 meter
1 5181-1519 CAN cable, Agilent module to module, 1 m
8 2 5043-1013 Tubing Clip
9 1 5500-1155 Tube Connector, 90 degree, ID 6.4
10 1 0100-1847 PEEK adapter 1/4-28 to 10-32 (Adapter AIV to solvent inlet tubes)
Agilent InfinityLab LC Series RID User Manual 145
10 Parts for Maintenance Accessory Kit
Accessory Kit for G7162B Accessory kit (G7162-68765) contains some accessories needed for the installation of the G7162B RID.
Item # p/n Description
1 1 5067-4784 Capillary ST 0.075 mm x 220 mm (Sampler to TCC/MCT)
1 1 5067-4783 Capillary ST 0.075 mm x 340 mm (Column to RID)
2 1 G1362-87301 Restriction capillary
3 2 0100-1700 FERRULE-AY-18IN
4 2 0100-1708 Nut 1/8 PPS
5 1 G7162-87300 Waste Tube Kit (recycle/waste) includes: Fitting, PK (2x), Tubing 2 m (2x)
1 5042-6449 Flex-Tube OD 1.6 mm
1 0100-1516 Fitting, PK, 1/16'' x 10-32, SH, male nut, 2/pk
6 1 0890-1760 Tubing Flexible 1 ea / 1 meter
1 5181-1519 CAN cable, Agilent module to module, 1 m
8 2 5043-1013 Tubing Clip
9 1 5500-1155 Tube Connector, 90 degree, ID 6.4
10 1 0100-1847 PEEK adapter 1/4-28 to 10-32 (Adapter AIV to solvent inlet tubes)
Agilent InfinityLab LC Series RID User Manual 146
10 Parts for Maintenance Leak Handling Parts
Leak Handling Parts
Item p/n Description
1 5043-0856 Leak Adapter
2 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm
3 5061-3356 Leak Sensor Assembly
4 G7162-44111 Leak Plane
5 0515-2529 Screw Tapping PAN-HD-TORX T10 3x8 ST-ZN
5043-1013 Tubing Clip (not shown)
6 G1362-44110 Leak Pan
7 0535-0030 Nut M 14
8 3050-0900 Washer
1
2
3
4
5 6
7, 8
Agilent InfinityLab LC Series RID User Manual 147
11 Identifying Cables
Cable Overview 149
Analog Cables 151
Remote Cables 153
CAN/LAN Cables 157
Agilent Module to PC 158
USB Cables 159
This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.
Agilent InfinityLab LC Series RID User Manual 148
11 Identifying Cables Cable Overview
Cable Overview
Analog cables
Remote cables
CAN cables
NOTE Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
p/n Description
35900-60750 Agilent 35900A A/D converter
01046-60105 Analog cable (BNC to general purpose, spade lugs)
p/n Description
5188-8029 ERI to general purpose
5188-8044 Remote Cable ERI ERI
5188-8045 Remote Cable APG ERI
5188-8059 ERI-Extension-Cable 1.2 m
5061-3378 Remote Cable to 35900 A/D converter
01046-60201 Agilent module to general purpose
5188-8057 Fraction Collection ERI remote Y-cable
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
5181-1519 CAN cable, Agilent module to module, 1 m
Agilent InfinityLab LC Series RID User Manual 149
11 Identifying Cables Cable Overview
LAN cables
RS-232 cables (not for
FUSION board)
USB cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)
p/n Description
RS232-61601 RS-232 cable, 2.5 m Instrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It is also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
5181-1561 RS-232 cable, 8 m
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
5188-8049 USB A F-USB Mini B M OTG (Module to Flash Drive)
Agilent InfinityLab LC Series RID User Manual 150
11 Identifying Cables Analog Cables
Analog Cables
One end of these cables provides a BNC connector to be connected to Agilent modules. The other end depends on the instrument to which connection is being made.
Agilent Module to 35900 A/D converters
p/n 35900-60750 35900 Pin Agilent module
Signal Name
1 Not connected
2 Shield Analog -
3 Center Analog +
Agilent InfinityLab LC Series RID User Manual 151
11 Identifying Cables Analog Cables
Agilent Module to BNC Connector
Agilent Module to General Purpose
p/n 8120-1840 Pin BNC Pin Agilent module
Signal Name
Shield Shield Analog -
Center Center Analog +
p/n 01046-60105 Pin Pin Agilent module
Signal Name
1 Not connected
2 Black Analog -
3 Red Analog +
Agilent InfinityLab LC Series RID User Manual 152
11 Identifying Cables Remote Cables
Remote Cables
ERI (Enhanced Remote Interface)
5188-8029 ERI to general purpose (D-Sub 15 pin male - open end)
5188-8044 ERI to ERI (D_Sub 15 pin male - male)
5188-8059 ERI-Extension-Cable 1.2 m (D-Sub15 pin male / female)
p/n 5188-8029 pin Color code Enhanced Remote
Classic Remote
Active (TTL)
1 white IO1 START REQUEST
Low
2 brown IO2 STOP Low
3 green IO3 READY High
4 yellow IO4 POWER ON High
5 grey IO5 NOT USED
6 pink IO6 SHUT DOWN Low
7 blue IO7 START Low
8 red IO8 PREPARE Low
9 black 1wire DATA
10 violet DGND
11 grey-pink +5V ERI out
12 red-blue PGND
13 white-green PGND
14 brown-green +24V ERI out
15 white-yellow +24V ERI out
NC yellow-brown
Agilent InfinityLab LC Series RID User Manual 153
11 Identifying Cables Remote Cables
5188-8045 ERI to APG (Connector D_Subminiature 15 pin (ERI), Connector D_Subminiature 9 pin (APG))
p/n 5188-8045 Pin (ERI) Signal Pin (APG) Active (TTL)
10 GND 1
1 Start Request 9 Low
2 Stop 8 Low
3 Ready 7 High
5 Power on 6 High
4 Future 5
6 Shut Down 4 Low
7 Start 3 Low
8 Prepare 2 Low
Ground Cable Shielding NC
Agilent InfinityLab LC Series RID User Manual 154
11 Identifying Cables Remote Cables
5188-8057 ERI to APG and RJ45 (Connector D_Subminiature 15 pin (ERI), Connector D_Subminiature 9 pin (APG), Connector plug Cat5e (RJ45))
One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent modules. The other end depends on the instrument to be connected to.
Table 12 5188-8057 ERI to APG and RJ45
p/n 5188-8057 Pin (ERI) Signal Pin (APG) Active (TTL) Pin (RJ45)
10 GND 1 5
1 Start Request
9 High
2 Stop 8 High
3 Ready 7 High
4 Fraction Trigger
5 High 4
5 Power on 6 High
6 Shut Down 4 High
7 Start 3 High
8 Prepare 2 High
Ground Cable Shielding
NC
Agilent InfinityLab LC Series RID User Manual 155
11 Identifying Cables Remote Cables
Agilent Module to Agilent 35900 A/D Converters
Agilent Module to General Purpose
p/n 5061-3378 Pin 35900 A/D Pin Agilent module
Signal Name Active (TTL)
1 - White 1 - White Digital ground
2 - Brown 2 - Brown Prepare run Low
3 - Gray 3 - Gray Start Low
4 - Blue 4 - Blue Shut down Low
5 - Pink 5 - Pink Not connected
6 - Yellow 6 - Yellow Power on High
7 - Red 7 - Red Ready High
8 - Green 8 - Green Stop Low
9 - Black 9 - Black Start request Low
p/n 01046-60201 Wire Color Pin Agilent module
Signal Name Active (TTL)
White 1 Digital ground
Brown 2 Prepare run Low
Gray 3 Start Low
Blue 4 Shut down Low
Pink 5 Not connected
Yellow 6 Power on High
Red 7 Ready High
Green 8 Stop Low
Black 9 Start request Low
Agilent InfinityLab LC Series RID User Manual 156
11 Identifying Cables CAN/LAN Cables
CAN/LAN Cables
Both ends of this cable provide a modular plug to be connected to Agilent modules CAN or LAN connectors.
CAN Cables
LAN Cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
5181-1519 CAN cable, Agilent module to module, 1 m
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)
Agilent InfinityLab LC Series RID User Manual 157
11 Identifying Cables Agilent Module to PC
Agilent Module to PC
p/n Description
RS232-61601 RS-232 cable, 2.5 m Instrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It is also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
5181-1561 RS-232 cable, 8 m
Agilent InfinityLab LC Series RID User Manual 158
11 Identifying Cables USB Cables
USB Cables
To connect a USB Flash Drive use a USB OTG cable with Mini-B plug and A socket.
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
5188-8049 USB A F-USB Mini B M OTG (Module to Flash Drive)
Agilent InfinityLab LC Series RID User Manual 159
12 Hardware Information
Data Flow for Chromatographic Output 161
Firmware Description 162
Electrical Connections 165 Rear View of the Module 166 Information on Instrument Serial Number 167
Interfaces 168 Overview Interfaces 170 ERI (Enhanced Remote Interface) 173 USB (Universal Serial Bus) 175
Setting the 6-bit Configuration Switch 176 Special Settings 178
Instrument Layout 180
Early Maintenance Feedback 181
This chapter describes the module in more detail on hardware and electronics.
Agilent InfinityLab LC Series RID User Manual 160
12 Hardware Information Data Flow for Chromatographic Output
Data Flow for Chromatographic Output
The data flow is shown below.
Figure 26 Data flow for chromatographic output
Peak width
nRIU calculation
A/D electronics diode difference
signal
Diode 1 signal Diode 2 signal
Selected signals
Raw data Analog outputs
Digital interface Analog interface
Pre-Filter
Filter
calculation zero Polarity
Temperature signal Scale offset
Agilent InfinityLab LC Series RID User Manual 161
12 Hardware Information Firmware Description
Firmware Description
The firmware of the instrument consists of two independent sections:
a non-instrument specific section, called resident system
an instrument specific section, called main system
Resident System
This resident section of the firmware is identical for all Agilent 1100/1200/1220/1260/1290 series modules. Its properties are:
the complete communication capabilities (CAN, LAN, USB and RS- 232)
memory management
ability to update the firmware of the 'main system'
Main System
Its properties are:
the complete communication capabilities (CAN, LAN, USB and RS- 232)
memory management
ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are divided into common functions like
run synchronization through APG/ERI remote,
error handling,
diagnostic functions,
or module specific functions like
internal events such as lamp control, filter movements,
raw data collection and conversion to absorbance.
Agilent InfinityLab LC Series RID User Manual 162
12 Hardware Information Firmware Description
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web: http://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315B for the G1315B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
VVV is the revision number, for example 650 is revision 6.50,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in chapter "Maintenance" or use the documentation provided with the Firmware Update Tools.
Figure 27 Firmware Update Mechanism
NOTE Update of main system can be done in the resident system only. Update of the resident system can be done in the main system only.
Main and resident firmware must be from the same set.
Resident System Main FW update
Main System
Resident FW Update
Agilent InfinityLab LC Series RID User Manual 163
12 Hardware Information Firmware Description
The firmware update tools, firmware and documentation are available from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
NOTE Some modules are limited in downgrading due to their mainboard version or their initial firmware revision. For example, a G1315C DAD SL cannot be downgraded below firmware revision B.01.02 or to a A.xx.xx.
Some modules can be re-branded (e.g. G1314C to G1314B) to allow operation in specific control software environments. In this case, the feature set of the target type is used and the feature set of the original one is lost. After re-branding (e.g. from G1314B to G1314C), the original feature set is available again.
All this specific information is described in the documentation provided with the firmware update tools.
Agilent InfinityLab LC Series RID User Manual 164
12 Hardware Information Electrical Connections
Electrical Connections
The CAN bus is a serial bus with high-speed data transfer. The two connectors for the CAN bus are used for internal module data transfer and synchronization.
One analog output provides signals for integrators or data handling systems.
The ERI/REMOTE connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features such as start, stop, common shutdown, prepare, and so on.
With the appropriate software, the LAN connector may be used to control the module from a computer through a LAN connection. This connector is activated and can be configured with the configuration switch.
With the appropriate software, the USB connector may be used to control the module from a computer through a USB connection.
The power input socket accepts a line voltage of 100 240 VAC 10 % with a line frequency of 50 or 60 Hz. Maximum power consumption varies by module. There is no voltage selector on your module because the power supply has wide-ranging capability. There are no externally accessible fuses because automatic electronic fuses are implemented in the power supply.
NOTE Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
Agilent InfinityLab LC Series RID User Manual 165
12 Hardware Information Electrical Connections
Rear View of the Module
Figure 28 Rear view of the Refractive Index Detector
Configuration switch
Analog output
USB-Mini-Port
LAN
ERI
CAN
Power socket
Agilent InfinityLab LC Series RID User Manual 166
12 Hardware Information Electrical Connections
Information on Instrument Serial Number
Serial Number Information 1200 Series and 1290 Infinity
The serial number information on the instrument labels provide the following information:
Serial Number Information 1260/1290 Infinity
The serial number information on the instrument labels provide the following information:
CCYWWSSSSS Format
CC country of manufacturing DE = Germany JP = Japan CN = China
YWW year and week of last major manufacturing change, e.g. 820 could be week 20 of 1998 or 2008
SSSSS real serial number
CCXZZ00000 Format
CC Country of manufacturing DE = Germany JP = Japan CN = China
X Alphabetic character A-Z (used by manufacturing)
ZZ Alpha-numeric code 0-9, A-Z, where each combination unambiguously denotes a module (there can be more than one code for the same module)
00000 Serial number
Agilent InfinityLab LC Series RID User Manual 167
12 Hardware Information Interfaces
Interfaces
The Agilent InfinityLab LC Series modules provide the following interfaces:
Table 13 Agilent InfinityLab LC Series Interfaces
Module CAN USB LAN (on-board)
RS-232 Analog APG (A) / ERI (E)
Special
Pumps
G7104A/C 2 No Yes Yes 1 A
G7110B 2 Yes Yes No No E
G7111A/B, G5654A 2 Yes Yes No No E
G7112B 2 Yes Yes No No E
G7120A, G7132A 2 No Yes Yes 1 A
G7161A/B 2 Yes Yes No No E
Samplers
G7129A/B/C 2 Yes Yes No No E
G7167A/B, G7137A, G5668A, G3167A
2 Yes Yes No No E
G7157A 2 Yes Yes No No E
Detectors
G7114A/B 2 Yes Yes No 1 E
G7115A 2 Yes Yes No 1 E
G7117A/B/C 2 Yes Yes No 1 E
G7121A/B 2 Yes Yes No 1 E
G7162A/B 2 Yes Yes No 1 E
G7165A 2 Yes Yes No 1 E
Agilent InfinityLab LC Series RID User Manual 168
12 Hardware Information Interfaces
CAN connectors as interface to other modules
LAN connector as interface to the control software
RS-232C as interface to a computer
USB (Universal Series Bus) as interface to a computer
REMOTE connector as interface to other Agilent products
Analog output connector(s) for signal output
Fraction Collectors
G7158B 2 Yes Yes No No E
G7159B 2 Yes Yes No No E
G7166A 2 No No No No No Requires a host module with on-board LAN with minimum FW B.06.40 or C.06.40, or with additional G1369C LAN Card
G1364E/F, G5664B 2 Yes Yes No No E THERMOSTAT for G1330B
Others
G1170A 2 No No No No No
G7116A/B 2 No No No No No Requires a host module with on-board LAN or with additional G1369C LAN Card.
G7122A No No No Yes No A
G7170B 2 No No No No No Requires a host module with on-board LAN with minimum FW B.06.40 or C.06.40, or with additional G1369C LAN Card
Table 13 Agilent InfinityLab LC Series Interfaces
Module CAN USB LAN (on-board)
RS-232 Analog APG (A) / ERI (E)
Special
NOTE The detector (DAD/MWD/FLD/VWD/RID) is the preferred access point for control via LAN. The inter-module communication is done via CAN.
Agilent InfinityLab LC Series RID User Manual 169
12 Hardware Information Interfaces
Overview Interfaces
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus system supporting high speed data communication and real-time requirement.
LAN
The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D DAD and G1365C/D MWD). This interface allows the control of the module/system via a PC with the appropriate control software. Some modules have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A Valve Drive or G4227A Flexible Cube). These are hosted modules and require a Host module with firmware B.06.40 or later or with additional G1369C LAN Card.
USB
The USB interface replaces the RS-232 Serial interface in new FUSION generation modules. For details on USB refer to USB (Universal Serial Bus) on page 175.
Analog Signal Output
The analog signal output can be distributed to a recording device. For details refer to the description of the modules mainboard.
NOTE If an Agilent detector (DAD/MWD/FLD/VWD/RID) is in the system, the LAN should be connected to the DAD/MWD/FLD/VWD/RID (due to higher data load). If no Agilent detector is part of the system, the LAN interface should be installed in the pump or autosampler.
Agilent InfinityLab LC Series RID User Manual 170
12 Hardware Information Interfaces
Remote (ERI)
The ERI (Enhanced Remote Interface) connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features as common shut down, prepare, and so on.
It allows easy connection between single instruments or systems to ensure coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote connector which is inputs/outputs (wired- or technique).
To provide maximum safety within a distributed analysis system, one line is dedicated to SHUT DOWN the systems critical parts in case any module detects a serious problem. To detect whether all participating modules are switched on or properly powered, one line is defined to summarize the POWER ON state of all connected modules. Control of analysis is maintained by signal readiness READY for next analysis, followed by START of run and optional STOP of run triggered on the respective lines. In addition PREPARE and START REQUEST may be issued. The signal levels are defined as:
standard TTL levels (0 V is logic true, + 5.0 V is false),
fan-out is 10,
input load is 2.2 kOhm against + 5.0 V, and
output are open collector type, inputs/outputs (wired- or technique).
NOTE All common TTL circuits operate with a 5 V power supply. A TTL signal is defined as "low" or L when between 0 V and 0.8 V and "high" or H when between 2.0 V and 5.0 V (with respect to the ground terminal).
Agilent InfinityLab LC Series RID User Manual 171
12 Hardware Information Interfaces
Special Interfaces
There is no special interface for this module.
Table 14 ERI signal distribution
Pin Signal Description
1 START REQUEST (L) Request to start injection cycle (for example, by start key on any module). Receiver is the autosampler.
2 STOP (L) Request to reach system ready state as soon as possible (for example, stop run, abort or finish and stop injection). Receiver is any module performing run-time controlled activities.
3 READY (H) System is ready for next analysis. Receiver is any sequence controller.
4 POWER ON (H) All modules connected to system are switched on. Receiver is any module relying on operation of others.
5 Not used
6 SHUT DOWN (L) System has serious problem (for example, leak: stops pump). Receiver is any module capable to reduce safety risk.
7 START (L) Request to start run / timetable. Receiver is any module performing run-time controlled activities.
8 PREPARE (L) Request to prepare for analysis (for example, calibration, detector lamp on). Receiver is any module performing pre-analysis activities.
Agilent InfinityLab LC Series RID User Manual 172
12 Hardware Information Interfaces
ERI (Enhanced Remote Interface) ERI replaces the AGP Remote Interface that is used in the HP 1090/1040/1050/1100 HPLC systems and Agilent 1100/1200/1200 Infinity HPLC modules. All new InfinityLab LC Series products using the FUSION core electronics use ERI. This interface is already used in the Agilent Universal Interface Box 2 (UIB2)
ERI Description
The ERI interface contains eight individual programmable input/output pins. In addition, it provides 24 V power and 5 V power and a serial data line to detect and recognize further add-ons that could be connected to this interface. This way the interface can support various additional devices like sensors, triggers (in and out) and small controllers, etc.
Figure 29 Location of the ERI interface (example shows a G7114A/B VWD)
ERI
Agilent InfinityLab LC Series RID User Manual 173
12 Hardware Information Interfaces
IO (Input/Output) Lines
Eight generic bi-directional channels (input or output).
Same as the APG Remote.
Devices like valves, relays, ADCs, DACs, controllers can be supported/controlled.
1-Wire Data (Future Use)
This serial line can be used to read out an EPROM or write into an EPROM of a connected ERI-device. The firmware can detect the connected type of device automatically and update information in the device (if required).
Pin Enhanced Remote
1 IO 1 (START REQUEST)
2 IO 2 (STOP)
3 IO 3 (READY)
4 IO 4 (POWER ON)
5 IO 5 (NOT USED)
6 IO 6 (SHUT DOWN)
7 IO 7 (START)
8 IO 8 (PREPARE)
9 1 wire DATA
10 DGND
11 +5 V ERI out
12 PGND
13 PGND
14 +24 V ERI out
15 +24 V ERI out
Agilent InfinityLab LC Series RID User Manual 174
12 Hardware Information Interfaces
5V Distribution (Future Use)
Available directly after turning on the hosting module (assures that the firmware can detect certain basic functionality of the device).
For digital circuits or similar.
Provides 500 mA maximum.
Short-circuit proof with automatic switch off (by firmware).
24V Distribution (Future Use)
Available by firmware command (defined turn on/off).
For devices that need higher power
Class 0: 0.5 A maximum (12 W)
Class 1: 1.0 A maximum (24 W)
Class 2: 2.0 A maximum (48 W)
Class depends on hosting modules internal power overhead.
If a connected device requires more power the firmware detects this (overcurrent detection) and provides the information to the user interface.
Fuse used for safety protection (on board).
Short circuit will be detected through hardware.
USB (Universal Serial Bus) USB (Universal Serial Bus) - replaces RS232, supports:
a PC with control software (for example Agilent Lab Advisor)
USB Flash Disk
Agilent InfinityLab LC Series RID User Manual 175
12 Hardware Information Setting the 6-bit Configuration Switch
Setting the 6-bit Configuration Switch
The 6-bit configuration switch is located at the rear of the module with FUSION electronics. Switch settings provide configuration parameters for LAN and instrument specific initialization procedures.
All modules with FUSION electronics:
Default is ALL switches DOWN (best settings).
Default IP address for LAN 192.168.254.11
For specific LAN modes switches 4-5 must be set as required.
For boot resident/cold start modes switches 1+2 or 6 must be UP.
Figure 30 Location of Configuration switch (example shows a G7114A/B VWD)
Configuration switch
Agilent InfinityLab LC Series RID User Manual 176
12 Hardware Information Setting the 6-bit Configuration Switch
Table 15 6-bit Configuration Switch
Mode Function/Setting
Switch 1 Switch 2 Switch 3 Switch 4 Switch 5 Switch 6
COM1
1 When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are taken from non-volatile memory.
0 n.a.2
2 not assigned - Always keep these switches on position 0 (off)
n.a. LAN Init Mode n.a.
Use Default IP Address3
3 Default IP Address is 192.168.254.11
0 0 0 0 0
Use Stored IP Address 0 0 0 1 0
Use DHCP to request IP Address4
4 Host Name will be the MAC address.
0 0 1 0 0
Test 1 System n.a. n.a. n.a. ColdStart
Boot Main System / Keep Data 0 0 0 0 0
Boot Resident System / Keep Data 1 0 0 0 0
Boot Main System / Revert to Default Data
0 0 0 0 1
Boot Resident System / Revert to Default Data
1 0 0 0 1
Agilent InfinityLab LC Series RID User Manual 177
12 Hardware Information Setting the 6-bit Configuration Switch
Special Settings
Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware loading errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again, the instrument firmware stays in the resident/main mode. In resident mode, it is not operable as a module. It only uses basic functions of the operating system for example, for communication. In this mode the main firmware can be loaded (using update utilities).
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with default parameter settings.
Boot Main System / Revert to Default Data
The instrument will boot to main mode and changes to the modules default parameter. May be also required to load resident firmware into the module.
Boot Resident System / Revert to Default Data
The instrument will boot to resident mode and changes to the modules default parameter. May be also required to load main firmware into the module.
If you use the following switch settings and power the instrument up again, it will start as described above.
CAUTION Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory. Exceptions are calibration settings, diagnosis and repair log books which will not be erased.
Save your methods and data before executing a forced cold start.
Agilent InfinityLab LC Series RID User Manual 178
12 Hardware Information Setting the 6-bit Configuration Switch
Table 16 Boot Resident / Forced Coldstart
SW1 SW2 SW3 SW4 SW5 SW6 Init Mode
1 0 0 0 0 0 Boot Main System / Keep Data
1 1 0 0 0 0 Boot Resident System / Keep Data
1 0 0 0 0 1 Boot Main System / Revert to Default Data
1 1 0 0 0 1 Boot Resident System / Revert to Default Data
Note: The setting '0' (down) is essential.
Agilent InfinityLab LC Series RID User Manual 179
12 Hardware Information Instrument Layout
Instrument Layout
The industrial design of the module incorporates several innovative features. It uses Agilents E-PAC concept for the packaging of electronics and mechanical assemblies. This concept is based upon the use of expanded polypropylene (EPP) layers of foam plastic spacers in which the mechanical and electronic boards components of the module are placed. This pack is then housed in a metal inner cabinet which is enclosed by a plastic external cabinet. The advantages of this packaging technology are:
virtual elimination of fixing screws, bolts or ties, reducing the number of components and increasing the speed of assembly/disassembly,
the plastic layers have air channels molded into them so that cooling air can be guided exactly to the required locations,
the plastic layers help cushion the electronic and mechanical parts from physical shock, and
the metal inner cabinet shields the internal electronics from electromagnetic interference and also helps to reduce or eliminate radio frequency emissions from the instrument itself.
Agilent InfinityLab LC Series RID User Manual 180
12 Hardware Information Early Maintenance Feedback
Early Maintenance Feedback
Maintenance requires the exchange of components that are subject to wear or stress. Ideally, the frequency at which components are exchanged should be based on the intensity of use of the module and the analytical conditions, and not on a predefined time interval. The early maintenance feedback (EMF) feature monitors the use of specific components in the instrument, and provides feedback when the user-selectable limits have been exceeded. The visual feedback in the user interface provides an indication that maintenance procedures should be scheduled.
EMF Counters
EMF counters increment with use and can be assigned a maximum limit which provides visual feedback in the user interface when the limit is exceeded. Some counters can be reset to zero after the required maintenance procedure.
Using the EMF Counters
The user-settable EMF limits for the EMF Counters enable the early maintenance feedback to be adapted to specific user requirements. The useful maintenance cycle is dependent on the requirements for use. Therefore, the definition of the maximum limits need to be determined based on the specific operating conditions of the instrument.
Setting the EMF Limits
The setting of the EMF limits must be optimized over one or two maintenance cycles. Initially the default EMF limits should be set. When instrument performance indicates maintenance is necessary, take note of the values displayed by the EMF counters. Enter these values (or values slightly less than the displayed values) as EMF limits, and then reset the EMF counters to zero. The next time the EMF counters exceed the new EMF limits, the EMF flag will be displayed, providing a reminder that maintenance needs to be scheduled.
Agilent InfinityLab LC Series RID User Manual 181
13 LAN Configuration
What You Have to Do First 183
TCP/IP parameter configuration 184
Configuration Switches 185
Initialization Mode Selection 186
Dynamic Host Configuration Protocol (DHCP) 188 General Information (DHCP) 188 Setup (DHCP) 189
Manual Configuration 191 With Telnet 192
PC and Agilent ChemStation Setup 195 PC Setup for Local Configuration 195 Agilent ChemStation Setup 198
This chapter provides information on connecting the module to the Agilent ChemStation PC.
Agilent InfinityLab LC Series RID User Manual 182
13 LAN Configuration What You Have to Do First
What You Have to Do First
The module has an on- board LAN communication interface.
1 Note the MAC (Media Access Control) address for further reference. The MAC or hardware address of the LAN interfaces is a world wide unique identifier. No other network device will have the same hardware address. The MAC address can be found on a label at the rear of the module underneath the configuration switch (see Figure 32 on page 183).
Figure 31 MAC-Label
2 Connect the instrument's LAN interface (see Figure 32 on page 183) to
the PC network card using a crossover network cable (point-to-point) or
a hub or switch using a standard LAN cable.
Figure 32 Location of LAN interfaces and MAC label
NOTE This chapter is generic and may show figures that differ from your module. The functionality is the same.
Part number of the detector main board Revision Code, Vendor, Year and Week of assembly MAC address Country of Origin
LAN interface
MAC label
Agilent InfinityLab LC Series RID User Manual 183
13 LAN Configuration TCP/IP parameter configuration
TCP/IP parameter configuration
To operate properly in a network environment, the LAN interface must be configured with valid TCP/IP network parameters. These parameters are:
IP address
Subnet Mask
Default Gateway
The TCP/IP parameters can be configured by the following methods:
by automatically requesting the parameters from a network-based DHCP Server (using the so-called Dynamic Host Configuration Protocol). This mode requires a LAN-onboard Module or a G1369C LAN Interface card, see Setup (DHCP) on page 189
by manually setting the parameters using Telnet
by manually setting the parameters using the Local Controller
The LAN interface differentiates between several initialization modes. The initialization mode (short form init mode) defines how to determine the active TCP/IP parameters after power-on. The parameters may be derived non-volatile memory or initialized with known default values. The initialization mode is selected by the configuration switch, see Table 17 on page 186.
Agilent InfinityLab LC Series RID User Manual 184
13 LAN Configuration Configuration Switches
Configuration Switches
The configuration switch can be accessed at the rear of the module.
Figure 33 Location of Configuration switch (example shows a G7114A/B VWD)
The module is shipped with all switches set to OFF, as shown above.
Configuration switch
NOTE To perform any LAN configuration, SW1 and SW2 must be set to OFF.
Agilent InfinityLab LC Series RID User Manual 185
13 LAN Configuration Initialization Mode Selection
Initialization Mode Selection
The following initialization (init) modes are selectable:
Default IP address for LAN is 192.168.254.11.
DHCP address is the modules LAN MAC address.
Using Stored
When initialization mode Using Stored is selected, the parameters are taken from the non-volatile memory of the module. The TCP/IP connection will be established using these parameters. The parameters were configured previously by one of the described methods.
Figure 34 Using Stored (Principle)
Table 17 Initialization Mode Switches
SW1 SW2 SW3 SW4 SW5 SW6 Init Mode
0 0 0 0 0 0 Use Default IP Address
0 0 0 0 1 0 Use Stored IP Address
0 0 0 1 0 0 Use DHCP
Note: The setting 0 (down) is essential.
Non-Volatile RAM
Stored Parameter
Active Parameter
Agilent InfinityLab LC Series RID User Manual 186
13 LAN Configuration Initialization Mode Selection
Using Default
When Using Default is selected, the factory default parameters are taken instead. These parameters enable a TCP/IP connection to the LAN interface without further configuration, see Table 18 on page 187.
Figure 35 Using Default (Principle)
Since the default IP address is a so-called local address, it will not be routed by any network device. Thus, the PC and the module must reside in the same subnet.
The user may open a Telnet session using the default IP address and change the parameters stored in the non-volatile memory of the module. He may then close the session, select the initialization mode Using Stored, power-on again and establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a local hub), separated from the local area network, the user may simply keep the default parameters to establish the TCP/IP connection.
Active Parameter
Default Parameter
NOTE Using the default address in your local area network may result in network problems. Take care and change it to a valid address immediately.
Table 18 Using Default Parameters
IP address: 192.168.254.11
Subnet Mask: 255.255.255.0
Default Gateway not specified
NOTE In the Using Default mode, the parameters stored in the memory of the module are not cleared automatically. If not changed by the user, they are still available, when switching back to the mode Using Stored.
Agilent InfinityLab LC Series RID User Manual 187
13 LAN Configuration Dynamic Host Configuration Protocol (DHCP)
Dynamic Host Configuration Protocol (DHCP)
General Information (DHCP) The Dynamic Host Configuration Protocol (DHCP) is an auto configuration protocol used on IP networks. The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and B-firmware (B.06.40 or above) or modules with "D"-firmware. All modules should use latest firmware from the same set.
When the initialization mode DHCP is selected, the card tries to download the parameters from a DHCP Server. The parameters obtained become the active parameters immediately. They are not stored to the non-volatile memory of the card.
Besides requesting the network parameters, the card also submits its hostname to the DHCP Server. The hostname equals the MAC address of the card, e.g. 0030d3177321. It is the DHCP server's responsibility to forward the hostname/address information to the Domain Name Server. The card does not offer any services for hostname resolution (e.g. NetBIOS).
Figure 36 DHCP (Principle)
DHCP Server
Active Parameter
NOTE 1 It may take some time until the DHCP server has updated the DNS server with the hostname information.
2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g. 0030d3177321.country.company.com.
3 The DHCP server may reject the hostname proposed by the card and assign a name following local naming conventions.
Agilent InfinityLab LC Series RID User Manual 188
13 LAN Configuration Dynamic Host Configuration Protocol (DHCP)
Setup (DHCP)
1 Note the MAC address of the LAN interface (provided with G1369C LAN Interface Card or mainboard). This MAC address is on a label on the card or at the rear of the mainboard, for example, 0030d3177321.
On the Local Controller the MAC address can be found under Details in the LAN section.
Figure 37 LAN Setting on Instant Pilot
The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and B-firmware (B.06.40 or above) or modules with "D"-firmware. All modules should use latest firmware from the same set.
Agilent InfinityLab LC Series RID User Manual 189
13 LAN Configuration Dynamic Host Configuration Protocol (DHCP)
2 Set the configuration switch to DHCP either on the G1369C LAN Interface Card or the mainboard of above mentioned modules.
3 Turn on the module that hosts the LAN interface.
4 Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition, Lab Advisor, Firmware Update Tool) and use MAC address as host name, e.g. 0030d3177321.
The LC system should become visible in the control software (see Note in section General Information (DHCP) on page 188).
Table 19 G1369C LAN Interface Card (configuration switch on the card)
SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode
ON OFF OFF OFF OFF DHCP
Table 20 LC Modules with 8-bit configuration switch (B-firmware) (configuration switch at rear of the instrument)
SW 6 SW 7 SW 8 Initialization Mode
ON OFF OFF DHCP
Agilent InfinityLab LC Series RID User Manual 190
13 LAN Configuration Manual Configuration
Manual Configuration
Manual configuration only alters the set of parameters stored in the non-volatile memory of the module. It never affects the currently active parameters. Therefore, manual configuration can be done at any time. A power cycle is mandatory to make the stored parameters become the active parameters, given that the initialization mode selection switches are allowing it.
Figure 38 Manual Configuration (Principle)
TELNET Session
Control Module
Stored Parameter
Non-Volatile RAM
Agilent InfinityLab LC Series RID User Manual 191
13 LAN Configuration Manual Configuration
With Telnet Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START button and select Run.... Type cmd and press OK.
2 Type the following at the system (DOS) prompt:
c:\>telnet
c:\>telnet
Figure 39 Telnet - Starting a session
where
When the connection was established successfully, the module responds with the following:
Figure 40 A connection to the module is made
3 Type ? and press enter to see the available commands.
Figure 41 Telnet Commands
Agilent InfinityLab LC Series RID User Manual 192
13 LAN Configuration Manual Configuration
4 To change a parameter follows the style:
parameter value, for example: ip 134.40.28.56
Then press [Enter], where parameter refers to the configuration parameter you are defining, and value refers to the definitions you are assigning to that parameter. Each parameter entry is followed by a carriage return.
5 Use the / and press Enter to list the current settings.
Table 21 Telnet Commands
Value Description
? displays syntax and descriptions of commands
/ displays current LAN settings
ip
sm
gw
exit exits shell and saves all changes
Figure 42 Telnet - Current settings in "Using Stored" mode
information about the LAN interface MAC address, initialization mode Initialization mode is Using Stored active TCP/IP settings
TCP/IP status - here ready connected to PC with controller software (e.g. Agilent ChemStation), here not connected
Agilent InfinityLab LC Series RID User Manual 193
13 LAN Configuration Manual Configuration
6 Change the IP address (in this example 192.168.254.12) and type / to list current settings.
7 When you have finished typing the configuration parameters, type exit and press Enter to exit with storing parameters.
Figure 44 Closing the Telnet Session
Figure 43 Telnet - Change IP settings
change of IP setting to Initialization mode is Using Stored
active TCP/IP settings
stored TCP/IP settings in non-volatile memory
connected to PC with controller software (e.g. Agilent ChemStation), here not connected
NOTE If the Initialization Mode Switch is changed now to Using Stored mode, the instrument will take the stored settings when the module is re-booted. In the example above it would be 192.168.254.12.
Agilent InfinityLab LC Series RID User Manual 194
13 LAN Configuration PC and Agilent ChemStation Setup
PC and Agilent ChemStation Setup
PC Setup for Local Configuration This procedure describes the change of the TCP/IP settings on your PC to match the modules default parameters in a local configuration (see Table 18 on page 187).
1 Open the Local Area Connection Properties and select Internet Protocol (TCP/IP). Then click on Properties.
Agilent InfinityLab LC Series RID User Manual 195
13 LAN Configuration PC and Agilent ChemStation Setup
2 You may enter here the fixed IP address of the module or use the Alternative Configuration.
Agilent InfinityLab LC Series RID User Manual 196
13 LAN Configuration PC and Agilent ChemStation Setup
3 We will use the direct LAN access via Cross-over LAN cable with the modules IP address.
4 Click on OK to save the configuration.
Agilent InfinityLab LC Series RID User Manual 197
13 LAN Configuration PC and Agilent ChemStation Setup
Agilent ChemStation Setup This procedure describes the Agilent ChemStation B.04.02 setup for the 1290 Infinity system using the 1290 Infinity DAD (G4212A) as the interfacing module. The setup works in the same way for all other systems.
1 Open the ChemStation Configuration Editor.
2 Select from the menu Configure - Instruments.
3 Select Modular 3D LC System.
4 Give the Instrument a name.
NOTE The LAN must be connected to detector due to high data load on communication to Control Software.
Agilent InfinityLab LC Series RID User Manual 198
13 LAN Configuration PC and Agilent ChemStation Setup
5 Click on OK.
6 Select LC System Access Access Point and click on Add.
7 Click on OK.
The Configuration Editor shows now the new instrument.
Agilent InfinityLab LC Series RID User Manual 199
13 LAN Configuration PC and Agilent ChemStation Setup
8 If required, change under Configure Path the folder locations.
9 Save the current configuration via File Save.
10 Exit the Configuration Editor.
11 Start the Agilent ChemStation.
During first startup or when the system configuration has changed, a notification shows up.
12 The left column shows the modules that could be configured. You may select the module manually from the list. We use the Auto Configuration mode. Click on Yes.
Agilent InfinityLab LC Series RID User Manual 200
13 LAN Configuration PC and Agilent ChemStation Setup
13 Enter the IP address or the Hostname of the module with the LAN-access.
14 Click on OK.
The selected module is shown now in the right window (with serial number). In addition all other modules connected via CAN to the detector are shown as well.
15 Click on OK to continue the ChemStation loading.
Agilent InfinityLab LC Series RID User Manual 201
13 LAN Configuration PC and Agilent ChemStation Setup
16 You may see the details of the module by selecting the module and clicking on Configure.
Under Connection Settings you may change the IP/Hostname of the module (may require a re-start of the ChemStation).
Agilent InfinityLab LC Series RID User Manual 202
13 LAN Configuration PC and Agilent ChemStation Setup
After successful load of the ChemStation, you should see the module(s) as active item in the graphical user interface (GUI).
Figure 45 Screen After Successful Load of ChemStation
Agilent InfinityLab LC Series RID User Manual 203
14 Appendix
General Safety Information 205 General Safety Information 205 Safety Standards 205 General 205 Before Applying Power 206 Ground the Instrument 206 Do Not Operate in an Explosive Atmosphere 207 Do Not Remove the Instrument Cover 207 Do Not Modify the Instrument 207 In Case of Damage 207 Solvents 208 Safety Symbols 209
Waste Electrical and Electronic Equipment (WEEE) Directive 211
Radio Interference 212
Sound Emission 213
Solvent Information 214
Agilent Technologies on Internet 216
This chapter provides safetey and other general information.
Agilent InfinityLab LC Series RID User Manual 204
14 Appendix General Safety Information
General Safety Information
General Safety Information The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customers failure to comply with these requirements.
Safety Standards This is a Safety Class I instrument (provided with terminal for protective earthing) and has been manufactured and tested according to international safety standards.
General Do not use this product in any manner not specified by the manufacturer. The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
WARNING Ensure the proper usage of the equipment.
The protection provided by the equipment may be impaired.
The operator of this instrument is advised to use the equipment in a manner as specified in this manual.
Agilent InfinityLab LC Series RID User Manual 205
14 Appendix General Safety Information
Before Applying Power
Ground the Instrument
WARNING Wrong voltage range, frequency or cabling
Personal injury or damage to the instrument
Verify that the voltage range and frequency of your power distribution matches to the power specification of the individual instrument.
Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
Make all connections to the unit before applying power.
NOTE Note the instrument's external markings described under Safety Symbols on page 209.
WARNING Missing electrical ground
Electrical shock
If your product is provided with a grounding type power plug, the instrument chassis and cover must be connected to an electrical ground to minimize shock hazard.
The ground pin must be firmly connected to an electrical ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Agilent InfinityLab LC Series RID User Manual 206
14 Appendix General Safety Information
Do Not Operate in an Explosive Atmosphere
Do Not Remove the Instrument Cover
Do Not Modify the Instrument Do not install substitute parts or perform any unauthorized modification to the product. Return the product to an Agilent Sales and Service Office for service and repair to ensure that safety features are maintained.
In Case of Damage
WARNING Presence of flammable gases or fumes
Explosion hazard
Do not operate the instrument in the presence of flammable gases or fumes.
WARNING Instrument covers removed
Electrical shock
Do Not Remove the Instrument Cover
Only Agilent authorized personnel are allowed to remove instrument covers. Always disconnect the power cables and any external circuits before removing the instrument cover.
WARNING Damage to the module
Personal injury (for example electrical shock, intoxication)
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
Agilent InfinityLab LC Series RID User Manual 207
14 Appendix General Safety Information
Solvents
WARNING Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
Do not use solvents with an auto-ignition temperature below 200 C (392 F). Do not use solvents with a boiling point below 56 C (133 F).
Avoid high vapor concentrations. Keep the solvent temperature at least 40 C (72 F) below the boiling point of the solvent used. This includes the solvent temperature in the sample compartment. For the solvents methanol and ethanol keep the solvent temperature at least 25 C (45 F) below the boiling point.
Do not operate the instrument in an explosive atmosphere.
Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for example, carbon disulfide).
Reduce the volume of substances to the minimum required for the analysis.
Never exceed the maximum permissible volume of solvents (8 L) in the solvent cabinet. Do not use bottles that exceed the maximum permissible volume as specified in the usage guideline for solvent cabinet.
Ground the waste container.
Regularly check the filling level of the waste container. The residual free volume in the waste container must be large enough to collect the waste liquid.
To achieve maximal safety, regularly check the tubing for correct installation.
NOTE For details, see the usage guideline for the solvent cabinet. A printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available in the Agilent Information Center or via the Internet.
Agilent InfinityLab LC Series RID User Manual 208
14 Appendix General Safety Information
Safety Symbols
Table 22 Symbols
The apparatus is marked with this symbol when the user shall refer to the instruction manual in order to protect risk of harm to the operator and to protect the apparatus against damage.
Indicates dangerous voltages.
Indicates a protected ground terminal.
The apparatus is marked with this symbol when hot surfaces are available and the user should not touch it when heated up.
Sample Cooler unit is designed as vapor-compression refrigeration system. Contains fluorinated greenhouse gas (refrigerant) according to the Kyoto protocol. For specifications of refrigerant, charge capacity, carbon dioxide equivalent (CDE), and global warming potential (GWP) see instrument label.
Flammable Material For Sample Thermostat which uses flammable refrigerant consult Agilent Information Center / User Manual before attempting to install or service this equipment. All safety precautions must be followed.
Confirms that a manufactured product complies with all applicable European Community directives. The European Declaration of Conformity is available at: http://regulations.corporate.agilent.com/DoC/search.htm
Manufacturing date.
Power symbol indicates On/Off. The apparatus is not completely disconnected from the mains supply when the power switch is in the Off position
Pacemaker Magnets could affect the functioning of pacemakers and implanted heart defibrillators. A pacemaker could switch into test mode and cause illness. A heart defibrillator may stop working. If you wear these devices keep at least 55 mm distance to magnets. Warn others who wear these devices from getting too close to magnets.
Agilent InfinityLab LC Series RID User Manual 209
14 Appendix General Safety Information
Magnetic field Magnets produce a far-reaching, strong magnetic field. They could damage TVs and laptops, computer hard drives, credit and ATM cards, data storage media, mechanical watches, hearing aids and speakers. Keep magnets at least 25 mm away from devices and objects that could be damaged by strong magnetic fields.
Indicates a pinching or crushing hazard
Indicates a piercing or cutting hazard.
Table 22 Symbols
WARNING A WARNING
alerts you to situations that could cause physical injury or death.
Do not proceed beyond a warning until you have fully understood and met the indicated conditions.
CAUTION A CAUTION
alerts you to situations that could cause loss of data, or damage of equipment.
Do not proceed beyond a caution until you have fully understood and met the indicated conditions.
Agilent InfinityLab LC Series RID User Manual 210
14 Appendix Waste Electrical and Electronic Equipment (WEEE) Directive
Waste Electrical and Electronic Equipment (WEEE) Directive
This product complies with the European WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.
NOTE Do not dispose of in domestic household waste
To return unwanted products, contact your local Agilent office, or see http://www.agilent.com for more information.
Agilent InfinityLab LC Series RID User Manual 211
14 Appendix Radio Interference
Radio Interference
Cables supplied by Agilent Technologies are screened to provide optimized protection against radio interference. All cables are in compliance with safety or EMC regulations.
Test and Measurement
If test and measurement equipment is operated with unscreened cables, or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.
Agilent InfinityLab LC Series RID User Manual 212
14 Appendix Sound Emission
Sound Emission
Sound pressure
Sound pressure Lp <70 db(A) according to DIN EN ISO 7779
Schalldruckpegel
Schalldruckpegel Lp <70 db(A) nach DIN EN ISO 7779
Agilent InfinityLab LC Series RID User Manual 213
14 Appendix Solvent Information
Solvent Information
Flow Cell To protect optimal functionality of your flow-cell:
Avoid the use of alkaline solutions (pH > 9.5) which can attack quartz and thus impair the optical properties of the flow cell.
If the flow cell is transported while temperatures are below 5 degree C, it must be assured that the cell is filled with alcohol.
Aqueous solvents in the flow cell can built up algae. Therefore do not leave aqueous solvents sitting in the flow cell. Add a small % of organic solvents (e.g. acetonitrile or methanol ~5%).
Agilent InfinityLab LC Series RID User Manual 214
14 Appendix Solvent Information
Use of Solvents Observe the following recommendations on the use of solvents.
Brown glass ware can avoid growth of algae.
Small particles can permanently block capillaries and valves. Therefore always filter solvents through 0.4 m filters.
Avoid the use of the following steel-corrosive solvents:
solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on),
high concentrations of inorganic acids like sulfuric acid and nitric acid, especially at higher temperatures (if your chromatography method allows, replace by phosphoric acid or phosphate buffer which are less corrosive against stainless steel),
halogenated solvents or mixtures which form radicals and/or acids, for example:
2CHCl3 + O2 2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol,
chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, diisopropyl ether) should be filtered through dry aluminium oxide which adsorbs the peroxides,
solvents containing strong complexing agents (e.g. EDTA),
mixtures of carbon tetrachloride with 2-propanol or THF.
Avoid the use of dimethyl formamide (DMF). Polyvinylidene fluoride (PVDF), which is used in leak sensors, is not resistant to DMF.
Agilent InfinityLab LC Series RID User Manual 215
14 Appendix Agilent Technologies on Internet
Agilent Technologies on Internet
For the latest information on products and services visit our worldwide web site on the Internet at:
http://www.agilent.com
Agilent InfinityLab LC Series RID User Manual 216
In This Book
This manual contains technical reference information about the Agilent 1290 Infinity II Refractive Index Detector (G7162B) and the Agilent 1260 Infinity II Refractive Index Detector (G7162A):
introduction,
specifications,
configuration,
optimizing,
Related manuals for Agilent InfinityLab LC Series Detector User Manual V2
Manualsnet FAQs
If you want to find out how the InfinityLab LC Series Agilent works, you can view and download the Agilent InfinityLab LC Series Detector User Manual V2 on the Manualsnet website.
Yes, we have the User Manual for Agilent InfinityLab LC Series as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.
The User Manual should include all the details that are needed to use a Agilent InfinityLab LC Series. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.
The best way to navigate the Agilent InfinityLab LC Series Detector User Manual V2 is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.
This Agilent InfinityLab LC Series Detector User Manual V2 consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.
You can download Agilent InfinityLab LC Series Detector User Manual V2 free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.
To be able to print Agilent InfinityLab LC Series Detector User Manual V2, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent InfinityLab LC Series Detector User Manual V2 as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.