Agilent InfinityLab LC Series 1260 Infinity II Delay Coil Organizer Technical Note

This Technical Note describes the installation of the G9324A Delay Coil Organizer to an Agilent Infinity II system.

Contents

Delay Coil Organizer 2 Overview 2 Delivery Checklist 3 Parts 4

Installation 5 Assemble the Delay Coil Organizer 5 Assemble and Install the Leak Tray and Leak Sensor 7 Install the Delay Coil Organizer to another Module 8 Install the Delay Coil Organizer into the 1290 Infinity II Preparative Column Compartment 10

Agilent Technologies

Delay Coil Organizer Overview

Delay Coil Organizer

Overview

2

The Delay Coil Organizer allows the storage of up to three Delay Coil combinations to achieve the required delay volume, and to easily detect and dispose of any possible leaks.

The Delay Coil Organizer comes with a Pressure Relief Valve (PRV, as part of the accessory kit) for protection of the Fraction Collector inlet line.

Figure 1 Delay Coil Organizer

Delay Coil Organizer Delivery Checklist

Delivery Checklist

p/n Description

G7163-60010 Delay Coil Box Assembly

G7163-68750 Accessory Kit Delay Coil Organizer

5067-6181 Knitted Delay Kit 1/2/5 m x 1.0 mm ESD Black (OPTIONAL)

5067-6800 Delay Coil Kit 4 8 mL/min ESD (OPTIONAL)

5067-6801 Delay Coil Kit 15 40 mL/min ESD (OPTIONAL)

5067-6802 Delay Coil Kit 40 80 mL/min ESD 1 (OPTIONAL)

5067-6804 Delay Coil Kit 80 200 mL/min ESD 1 (OPTIONAL)

5188-8053 ERI Remote / Leak Module (OPTIONAL)

1 requires Adapter Nut FB 1/4-28-M 5/16-24-F ST (5023-3093), which comes as part of the Delay Coil Kit

Accessory Kit

NOTE Delay coils are specified to 25 bar.

p/n Description

5063-6527 Tubing assembly, i.d. 6 mm, o.d. 9 mm, 1.2 m (to waste)

5067-6655 Leak Sensor Assembly 3 m

5067-6857 PREP Pressure Relief Valve Kit

5067-5685 Clamp Guide Kit-IF-II

5043-0270 Leak plane

5043-0271 Holder leak plane

5043-0275 Clamp guide

3

Delay Coil Organizer Parts

Parts

4

Figure 2 Delay Coil Organizer parts

Item p/n Description

1 5043-0270 Leak plane

2 5067-6857 PREP Pressure Relief Valve Kit

3 G7163-60010 Delay Coil Box Assembly

4 5067-6181 Knitted Delay Kit 1/2/5 m x 1.0 mm ESD Black (OPTIONAL)

5067-6800 Delay Coil Kit 4 8 mL/min ESD (OPTIONAL)

5067-6801 Delay Coil Kit 15 40 mL/min ESD (OPTIONAL)

5067-6802 Delay Coil Kit 40 80 mL/min ESD (OPTIONAL)

5067-6804 Delay Coil Kit 80 200 mL/min ESD (OPTIONAL)

Installation Assemble the Delay Coil Organizer

Installation

You can install the Delay Coil Organizer either to the side of a particular module (see Install the Delay Coil Organizer to another Module on page 8), or in the 1290 Infinity II Preparative Column Compartment (see Install the Delay Coil Organizer into the 1290 Infinity II Preparative Column Compartment on page 10).

Assemble the Delay Coil Organizer

1 Remove the delay coil organizer from its housing by pulling it to the front.

2 Remove the distributor from the delay coil organizer.

3 Install the Adapter Nut FB 1/4-28-M 5/16-24-F ST (5023-3093) for the Delay Coil Kit mL/min ESD (5067-6802) and the Delay Coil Kit mL/min ESD (5067-6804).

4 Tigthen the adapter nut with a torque of 2.3 Nm (about 10 after fingertight).

5 Mount the required delay coils to the available ports of the distributor, and fix them fingertight. Do not use any wrench or fitting handle.

NOTE Use two vertical aligned ports for one coil (combination).

You can install up to three combinations.

5

Installation Assemble the Delay Coil Organizer

6 Place the distributor with the installed coil(s) back into the delay coil organizer.

7 Install the distributor by pushing it through the holes in the module frame.

8 Slide the delay coil organizer into its housing to close it.

9 Install the Pressure Relief Valve (PRV) to the top port (OUT) of the Delay Coil (combination) to be used. This will indicate which port is in use.

6

Installation Assemble and Install the Leak Tray and Leak Sensor

1 Attach the leak tray clamp to the leak tray. 2 Push the leak sensor into its holder.

3 Slide the leak tray onto the bottom of the Delay Coil Organizer.

Assemble and Install the Leak Tray and Leak Sensor

7

Installation Install the Delay Coil Organizer to another Module

Install the Delay Coil Organizer to another Module

1 Assemble and install the leak tray and leak sensor to the delay coil organizer, see Assemble and Install the Leak Tray and Leak Sensor on page 7.

2 Ensure the power switch at the rear of the module is OFF and the power connector is unplugged.

3 Identify the marks for the screws on the side panel of the cabinet assy and screw the clamp guides to the side panel of the cabinet.

4 Slide the delay coil organizer into the clamp guides from front to back.

5 Attach the waste tubing to the leak plane and guide it to a proper waste container.

6 Connect the leak sensor cable to a module with ERI Remote connector (particularly G7159B, or G7161A/B), or to an ERI Remote / Leak Module (5188-8053).

7 Install the capillary from the detector to the bottom port (IN) of the Delay Coil (combination) to be used.

8

Installation Install the Delay Coil Organizer to another Module

8 Install the tubing that goes to the fraction collector to the extended top front port of the PRV. Use the Fitting Handle to apply the correct torque.

9 Fix the waste tubing (1.6 mm x 650 mm ESD) to the bottom of the PRV.

10 Twist the PRV, such that the bottom does not get in the way of the bottom Delay Coil connection, but still ends inside the Leak Tray.

NOTE All fittings used in combination with PRV, or inside the Delay Coil Organizer are -28 FB. Fittings to the bottom port of the Delay Coil Organizer are 10-32 coned.

9

Installation Install the Delay Coil Organizer into the 1290 Infinity II Preparative Column Compartment

Install the Delay Coil Organizer into the 1290 Infinity II Preparative Column Compartment

1 Install the Delay Coil Organizer into the top valve position of the 1290 Infinity II Preparative Column Compartment according to the Purification Capillary Kits Technical Note.

2 Install the capillary from the detector to the bottom port (IN) of the Delay Coil (combination) to be used.

3 Install the tubing that goes to the fraction collector to the extended top front port of the PRV. Use the Fitting Handle to apply the correct torque.

4 Fix the waste tubing (1.6 mm x 650 mm ESD) to the bottom of the PRV.

10

Installation Install the Delay Coil Organizer into the 1290 Infinity II Preparative Column Compartment

5 Twist the PRV, such that the bottom doesn't get in the way of the bottom Delay Coil connection, while the 650 mm waste line is guided alongside the small mounting plate inside the Column Compartment, to the Leak Plate and the tubing is fixed directly under the leak sensor assembly.

NOTE All fittings used in combination with PRV, or inside the Delay Coil Organizer are -28 FB. Fittings to the bottom port of the Delay Coil Organizer are 10-32 coned.

11

Installation Install the Delay Coil Organizer into the 1290 Infinity II Preparative Column Compartment

*G9324-90 *G9324-90 G9324-90000 Rev. B SD-29000217 Rev.B Edition: 10/2019

Manualsnet FAQs

If you want to find out how the 1260 Infinity II Agilent works, you can view and download the Agilent InfinityLab LC Series 1260 Infinity II Coil Organizer Technical Note on the Manualsnet website.

Yes, we have the Technical Note for Agilent 1260 Infinity II as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Technical Note should include all the details that are needed to use a Agilent 1260 Infinity II. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent InfinityLab LC Series 1260 Infinity II Coil Organizer Technical Note is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent InfinityLab LC Series 1260 Infinity II Coil Organizer Technical Note consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent InfinityLab LC Series 1260 Infinity II Coil Organizer Technical Note free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent InfinityLab LC Series 1260 Infinity II Coil Organizer Technical Note, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent InfinityLab LC Series 1260 Infinity II Coil Organizer Technical Note as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.