- Manuals
- Brands
- Agilent
- Chromatograph
- 490 Micro GC Field Case
- User Manual
Agilent 490 Micro GC Field Case Chromatograph User Manual PDF
Summary of Content for Agilent 490 Micro GC Field Case Chromatograph User Manual PDF
Agilent 490 Micro GC Field Case
User Manual
Agilent Technologies
Notices Agilent Technologies, Inc. 2017
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla- tion into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
G3581-90005
Edition
Third edition, August 2017
Printed in China
Agilent Technologies, Inc. 412 Ying Lun Road Waigoaqiao Freed Trade Zone Shanghai 200131 P.R.China
Warranty
The material contained in this docu- ment is provided as is, and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor- mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or sub- contract, Software is delivered and licensed as Commercial computer soft- ware as defined in DFAR 252.227-7014 (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Govern- ment users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an oper- ating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi- tions are fully understood and met.
Contents
1 Getting Started
490 Micro GC Field Case User Manua
Safety Information 6
Initial Inspection 9
Unpacking 10 Packing list 10 Carrier gas 11
Instrument Overview 12 Inside view 13 Power source 15
2 Installation and Use
Minimum Requirements to Use the Field Case 18 Field Case Sample Inlet Requirements 19
Install Sample Inlet 20
Connecting Gas Supply for Single Carrier Gas 21
Connecting the Micro GC to the Field Case 22
Installing the Field Case Battery 30
Automatic Battery Selection 33
Charging Batteries 34
Field Case Display 35
Internal Buzzer 36
Carrier Gas Refill Instructions 37 Attach Carrier Gas Refill Assembly to Gas Source 37 Removing a Gas Cylinder From the Field Case 39 Reducing Gas Flow Module Pressure 41 Gas Cylinder Refill Procedure 41 Exchanging Gas Species or First Refill of the Gas Cylinder 45
Shipping 46
Cleaning 46
Disposal 46
3 Battery Pack and Charger
Battery Pack 48 Introduction 48 General precautions 48 Battery location 49 Battery handling 50 Charging 51 Discharging 52 Storage 52
l 3
4
Battery service life 53 Cleaning 53 Disposal 54 Technical specifications 54
Battery Charger 55 Introduction 55 Features 55 Charger status 56 Operation 56 Cleaning 56 Disposal instructions 57 Technical specifications 57
Error Message and Battery Status Indication 58 Error message 58 Battery status indicator 59
490 Micro GC Field Case User Manual
Agilent 490 Micro GC Field Case User Manual
1 Getting Started
Safety Information 6
Initial Inspection 9
Unpacking 10
Instrument Overview 12
5Agilent Technologies
1 Getting Started
Safety Information
6
In accordance with the Agilent commitment to customer service and safety, the:
Micro GC field case and its accompanying documentation (NEN 5509) complies with the CE specifications and the safety requirements for electrical equipment for measurement, control, and laboratory use (CEI/IEC 61010-1) and cCSAus.
Battery Pack and Charger and its accompanying documentation complies with the Safety, Battery Pack UL2054, UL1015 (cabling) and CE, Charger to CE, and documentation to NEN 5509.
This device has been tested and complied with CEC-400-2015-021, California Energy Commission appliance efficiency regulations on battery charger systems.
This device has been tested and found to comply with IEC 61326-1 and CISPR 11 limits. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to case harmful interference in which case the user will be required to correct the interference at his own expense.
To prevent any injury to the user or any damage to the instrument, it is essential that you read the information in this document.
This manual is provided to help you establish operating conditions, which will permit safe and efficient use of your equipment.
If you have any problems understanding the text in this manual, we advise you to contact your Agilent Technologies service office for assistance. Agilent Technologies, Inc. cannot accept responsibility for any damage or injury caused by misunderstanding of the information in this manual.
490 Micro GC Field Case User Manual
Getting Started 1
490 Micro GC Field Case User Manual
Special considerations and precautions appear in this manual in the form of NOTES, CAUTIONS, and WARNINGS. It is important that you operate your equipment in accordance with the instructions in this manual and any additional information, which may be provided by Agilent Technologies, Inc. Address any questions regarding the safe and proper use of your equipment to your local Agilent office.
Follow these safety practices to ensure safe equipment operation:
Perform periodic leak checks on all supply lines and pneumatic plumbing.
Do not allow gas lines to become kinked or punctured. Place lines away from foot traffic and extreme heat or cold.
Store organic solvents in fireproof, vented, and clearly labeled cabinets so they are easily identified as either toxic materials or flammable materials, or both.
Do not accumulate waste solvents. Dispose of such materials through a regulated disposal program and not through municipal sewage lines.
This instrument has been tested per applicable requirements of the EMC
WARNING This instrument is designed for chromatographic analysis of appropri- ately prepared samples. It must be operated using appropriate gases, solvents, or both, and within specified maximum ranges for pressure, flows, and temperatures as described in this manual. If the equipment is used in a manner not specified by the manufacturer, the protection pro- vided by the equipment may be impaired.
It is the responsibility of the Customer to inform Agilent Customer Sup-
Avoid back strain or injury by following all safety precautions when lift-
7
1 Getting Started
CAUTION Disconnect the instrument from all power sources before removing the protective panels to avoid exposure to potentially dangerous voltages.
When it is necessary to use a non-original power cord plug, ensure the replacement cord adheres to the color-coding and polarity described in the manual and all local building safety codes.
Replace faulty or frayed power cords immediately with the same type and rating. Place this instrument in a suitable location with sufficient ventilation to remove
gases and vapors. Space around the instrument must be sufficient to enable cooling of the instrument.
Before plugging the instrument in or turning the power on, always make sure that the voltage and fuses are set appropriately for your local power source.
Do not turn the instrument on if there is a possibility of any kind of electrical damage. Instead, disconnect the power cord and battery packs and contact your Agilent office.
Insert the supplied power cord into a power outlet with a protective earth ground connection. When using an extension cord, ensure that the cord is also properly grounded.
Do not change the external or internal grounding connections as this could endanger you, or damage the instrument, or both.
The instrument is properly grounded when shipped. You do not need to make any changes to the electrical connections or to the instrument chassis to ensure safe operation.
When working with this instrument, follow the regulations for Good Laboratory Practice (GLP). Take care to wear safety glasses and appropriate clothing.
Do not place containers with flammable liquids on this instrument. Spilling the liquid over hot parts may cause fire.
This instrument may use flammable or explosive gases. Be sure to be familiar with and to accurately follow the operation procedures prescribed for those gases before operating the instrument.
Never try to repair or replace any component that is not described in this manual without the assistance of an Agilent service engineer. Unauthorized repairs or modifications will result in rejection of warranty claims.
Always disconnect the AC power cord and battery packs before attempting any type of maintenance.
Use proper tools when working on the instrument to prevent danger for you, damage to the instrument, or both.
Do not attempt to replace fuses in this instrument. Damage can result if the instrument is stored under unfavorable conditions for
prolonged periods (for example, subject to heat, or water). Do not shut off column flow when the oven temperature is high, it may damage
the column. This unit has been designed and tested in accordance with recognized safety
standards and designed for use indoors. If the instrument is used in a manner not specified by Agilent Technologies, the
protection provided by the instrument may be impaired. Substituting parts or performing any unauthorized modification to the
instrument may result in a safety hazard. Changes or modifications not expressly approved by the responsible party for
compliance could void the user's authority to operate the equipment.
8 490 Micro GC Field Case User Manual
Getting Started 1
Initial Inspection
490 Micro GC Field Case User Manual
The Micro GC field case will arrive packed in one large box and one or more smaller cartons. Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to your local Agilent Technologies, Inc. office.
Unpack the Micro GC field case and accessories carefully and transfer to the work area, using proper handling techniques. Inspect the Micro GC field case and accessories carefully for damage or signs of rough handling. Report damage to the carrier and to your local Agilent Technologies, Inc. office.
The Micro GC field case will be supplied with an integrated trolley for easier transportation.
Figure 1 Field case with trolley
9
1 Getting Started
Unpacking
10
Check the packing list to ensure you have received all that you require.
Packing list
Micro GC field case
Accessories (optional)
Ship kit
Table 1 Accessories
Car cigarette lighter adapter CP740291
Charger for 12 V battery pack NiMH CP740427
Battery pack NiMH CP740328
Gas cylinder for 2-channel field case 75mL G3581-63011
Gas cylinder for 4-channel field case 300mL G3581-63012
Field Case kit for manual injection G3581-63006
Carrier gas refill assembly DIN10 G3581-60701
Carrier gas refill assembly BS3 G3581-60702
Carrier gas refill assembly FRANCE G3581-60703
Carrier gas refill assembly DIN6 G3581-60704
Carrier gas refill assembly CGA580 G3581-60705
Table 2 Field case ship kit G3581-63010
Part Number Description Quantity
0100-0090 Tee, 1/8 inch Brass 1
G1530-21020 Field case Gas Supply Tubing 1
G3581-60612 Back Sample in Kit 1
G3581-00524 FC Back Sample in Holder 1
1390-1023 Captive screw M4X0.7X12MM 1
CP86757 Screw M4X8 Torx 8.8 SV 3
490 Micro GC Field Case User Manual
Getting Started 1
Carrier gas
490 Micro GC Field Case User Manual
The carrier gas supply cylinders are empty (transport regulations) and should be refilled using the procedure Carrier Gas Refill Instructions on page 37.
The common carrier gas for the Micro GC is He, N2, or Ar. The recommended purity for carrier gas is 99.999 % minimum.
NOTE Never use hydrogen carrier gas in the gas cylinder.
11
1 Getting Started
Instrument Overview
12
Congratulations and thank you for purchasing the Agilent Technologies, Inc. Micro GC field case (Figure 2). The Micro GC field case is used to bring the Micro GC (2 and 4-Channel) to the sample source. Operation is continuous with interchangeable, rechargeable battery packs, and an internal gas supply.
For problems or questions about your Micro GC field case, please contact your nearest Agilent Technologies, Inc. representative.
Figure 2 Micro GC field case
490 Micro GC Field Case User Manual
Getting Started 1
Inside view
490 Micro GC Field Case User Manual
Figure 3 Inside view
Battery compartment
Power cable Power to GC
Communication cable (Data cable) connects to GC
Micro GC compartment
13
14
1 Getting Started
Figure 4 Inside view detail
Power connector Power IN male connector
Carrier gas supply Connects to GC
Carrier gas cylinders
Display panel
Carrier gas selector Select between gas cylinder and outer carrier gas
Outer carrier gas connector Connects to outer gas supply
For more information on the Battery status see Battery Pack on page 48.
For more information on the Carrier 1 and Fill 1 see Carrier Gas Refill Instructions on page 37.
4
Getting Started 1
Power source
490 Micro GC Field Case User Manual
The Micro GC field case can be powered by batteries or the power supply.
Power supply
Voltage of 12 Vdc, 150 W
Only use the power supply (p/n G3581-60080) provided with the Micro GC. See the Micro GC User Manual for more information.
Battery pack
The Micro GC field case is equipped with a 12 V 10 Ah (ampere-hours) Ni-MH (Nickel metal hydride) battery. See Chapter 3 Battery Pack and Charger on page 47 for more details.
Depending on the condition in which the Micro GC is used, the battery will provide power up to 8 hours before recharging becomes necessary.
Maximum temperatures when using battery as power source: Column temp 150C, Injector temp 90C, Sample line temp 90C.
When reconditioning columns or stabilizing the system it is advisable to plug in the power supply during this period.
Both the 2-channel system and 4-channel system support the use of up to two batteries simultaneously.
All ambient requirements for the Micro GC, except for the maximum operating temperature, are applicable to the field case. Please read the Micro GC user manual for more details.
When operating the Micro GC in the field case, while using the batteries, the maximum allowable ambient temperature is 50C.
WARNING This battery (p/n CP740328) is designed for use in combination with the Micro GC field case. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
15
16
1 Getting Started
490 Micro GC Field Case User Manual
Agilent 490 Micro GC Field Case User Manual
2 Installation and Use
Minimum Requirements to Use the Field Case 18
Install Sample Inlet 20
Connecting Gas Supply for Single Carrier Gas 21
Connecting the Micro GC to the Field Case 22
Installing the Field Case Battery 30
Automatic Battery Selection 33
Charging Batteries 34
Field Case Display 35
Internal Buzzer 36
Carrier Gas Refill Instructions 37
Shipping 46
Cleaning 46
Disposal 46
17Agilent Technologies
2 Installation and Use
Minimum Requirements to Use the Field Case
18
To use the field case, the Micro GC must equipped with main board G3581-65000 version F0 and Firmware version 4.01
Table 3 Firmware compatibility
Micro GC main board with FW CP490240/41/ 42/50/51/52 (legacy)
G3581-63001/ 002/003
CP740010 with FW 2.44 yes no
G3581-65000 (D0) with FW 3.xx yes no
G3581-65000 (F0) with FW 3.32 yes no
G3581-65000 (F0) with FW 4.0x yes yes
Table 4 Field Cases
Part Number Description
CP490240 Legacy Field case 2-channel standard
CP490241 Legacy Field case 2-channel double volume
CP490242 Legacy Field case 2-channel two carrier gases
CP490250 Legacy Field case 4-channel standard
CP490251 Legacy Field case 4-channel double volume
CP490252 Legacy Field case 4-channel two carrier gases
G3581-63001 Fieldcase 2-channel 2 Carrier Gases
G3581-63002 Fieldcase 4-channel Std 1 Carrier Gas
G3581-63003 Fieldcase 4-channel 2 Carrier Gases
490 Micro GC Field Case User Manual
Installation and Use 2
Field Case Sample Inlet Requirements
490 Micro GC Field Case User Manual
Table 5 Field Case Sample Inlet Requirements
Sample and sample inlet type
Remarks and Requirements
Micro GC Front inlet Not compatible with field case. Marketing will block ordering front inlet & field case in the configurator (SAP and sales).
Single or dual rear inlet Items required for install are provided in field case shipping kit Table 2 on page 10. See Install Sample Inlet on page 20 for installation instructions.
Heat -traced sample line Install on internal inlet. See Micro GC user manual G3581-90001 for instructions.
Manual syringe injection Install optional Field Case kit for manual injection PN G3581-63006.
Micro GC front inlet is not compatible with the field case.
Table 6 Manual injection kit for field case G3581-63006
Part Number Description Quantity
G3581-63006 Sample-in connection kit 1
G3581-00524 FC Back Sample-in Holder 1
G3581-60610 Septum Nut IN Kit 1
G3581-60611 Luer Lock IN Kit 1
CP742983 Septum injection port Assembly 1
VLZLA1 Female Luer adapter to 1/16 fittings 1
CP914757 Fitting, ext. Luer lock,10-32 thrd,1/pk 1
CP86757 Screw M4X8 Torx 8.8 SV 1
1390-1023 Captive Screw M4X0.7X12mm 1
19
2 Installation and Use
Install Sample Inlet
20
Install the back sample-in connector.
1 Connect the G3581-60612 to the back sample in connector on micro the GC, using a wrench to tighten the nut on the adapter.
2 With one CP86757 screw, mount bracket G3581-00524 to adapter.
3 Remove the rear panel screw
4 Mount the assembly to the rear panel of the micro GC with one 1390-1023 screw.
Figure 5 Remove rear panel screw
Figure 6 Install the back sample-in connector
Rear panel screw
490 Micro GC Field Case User Manual
Installation and Use 2
Connecting Gas Supply for Single Carrier Gas
490 Micro GC Field Case User Manual
1 Connect carrier gas supply to Tee union (0100-0090).
2 Connect field case gas supply tube (G1530-21020) to Tee union output.
21
2 Installation and Use
Connecting the Micro GC to the Field Case
22
NOTE The GC must have a sample-in connector installed BEFORE the GC is installed into the field case. See the section Field Case Sample Inlet Requirements on page 19.
1 Open the case and remove the mounting bracket.
2 Place the GC on the bracket as shown.
Figure 7 Remove bracket
490 Micro GC Field Case User Manual
Installation and Use 2
490 Micro GC Field Case User Manual
Figure 8 Attach bracket to GC with mounting screw
Mounting screw location Attach GC with screw at this location
3 Attach GC to bracket using screw (CP86757) at bottom. There is 1 screw for the 2-channel GC and 2 for the 4-channel GC.
4 Fasten straps around GC.
Figure 9 Straps fastened
WARNING Do not install the batteries or charge the carrier gas tank until all parts are fully assembled and reinstalled inside the field case.
23
24
2 Installation and Use
5 To gain access to the power cable, data cable, and carrier gas supply, remove the protective cover by loosening the captive screw with a Torx T-20 driver.
NOTE With the four channel field case, it is necessary to remove the gas cylinders before removing the protective cover. See Removing a Gas Cylinder From the Field Case on page 39
Figure 10 Remove protective cover
490 Micro GC Field Case User Manual
Installation and Use 2
490 Micro GC Field Case User Manual
6 Place the mounted GC on top of the open field case and attach the power cable.
Figure 11 Protective cover removed
Figure 12 Attach the power cable
25
26
2 Installation and Use
7 Lower the GC into the field case. The straps assist in installing and removing the GC.
When placing GC into the field case, do not damage or obstruct the data
Figure 13 Placing the Micro GC in the field case
490 Micro GC Field Case User Manual
Installation and Use 2
490 Micro GC Field Case User Manual
8 Connect the carrier gas supply to the GC. The connector can only be fastened after the GC is mounted in field case.
Figure 14 Connect carrier gas supply
27
28
2 Installation and Use
9 Open the GC to access the data connection. You will need to unfasten the straps.
10 Connect the RS232 communication cable to the COM 2 connector.
11 Close the GC and refasten the straps.
12 Replace the protective cover.
Figure 15 Micro GC Field case communication cable connected
490 Micro GC Field Case User Manual
Installation and Use 2
490 Micro GC Field Case User Manual
13 Connect the Micro GC to the sample source. For details, see the Micro GC user manual.
14 Install the field case battery (Installing the Field Case Battery on page 30).
15 Connect the Workstation PC to the Micro GC. For details, see the Micro GC user manual.
16 Turn the carrier gas on by selecting the gas source and opening the appropriate gas cylinder valve.
17 Switch on the field case.
18 After a few minutes, the system should be ready for use (see the Micro GC user manual).
Figure 16 GC installed in field case
29
2 Installation and Use
Installing the Field Case Battery
30
The field case is shipped with battery disconnected. Connect battery
1 Inside near the hinge of the field case, open the battery compartment by unscrewing the captive Torx T-20 screws.
Figure 17 Removing the battery compartment cover
The field case may only be used with the supplied battery (CP740328) or
2 When installing one battery, place the battery in the compartment and connect the battery to the cable labeled battery 1 (Figure 18 and Figure 19).
Figure 18 Labeled battery connectors
490 Micro GC Field Case User Manual
Installation and Use 2
490 Micro GC Field Case User Manual
3 To install more than one battery in the 2-channel field case, remove the metal battery spacer and foam insert (Figure 20). Connect a battery to one of the connectors and install it in place of the foam insert. Replace the metal bracket and install the second battery on top of the bracket and connect to the remaining connector (Figure 19). To install more than one battery in the 4-channel field case, insert the batteries side by side (Figure 21). Connect each battery to a connector.
Figure 19 Installing the battery in the 2-Channel field case
Figure 20 Metal battery spacer removed, foam insert shown
31
32
2 Installation and Use
4 Ensure the cables are correctly positioned in the cable guides to prevent damage, then replace the battery compartment door.
Figure 21 4-Channel field case with two batteries installed
It is advisable to charge the batteries before operation.
CAUTION Place the field case in such a way that both the mains appliance inlet or adapter and the battery supply cables are easy to reach for the operator. These connections must be disconnected in case of emergency.
CAUTION If the field case will not be used for a long period of time, please discon- nect batteries inside the battery compartment to avoid over-discharging.
490 Micro GC Field Case User Manual
Installation and Use 2
Automatic Battery Selection
490 Micro GC Field Case User Manual
During startup, the system will check how many batteries are available and will first use the one connected to the battery 1 connector (Bat1) to power the Micro GC.
When the battery voltage becomes lower than 11 Volts, the system will (in case of two batteries) automatically switch to the second battery (Bat2).
If both batteries' voltage drop below 10.5 Volts, both batteries will power the system simultaneously
If both batteries' voltage drop below 9.5 Volts, the micro GC will be shut down.
33
2 Installation and Use
Charging Batteries
34
Plug the micro GC power supply (G3581-60080) into the power connector at the upper right corner of main panel. The system will start to charge the battery if it is below 90% power.
If the micro GC is powered off, the field case will charge both batteries simultaneously.
If the micro GC is powered on, the field case will charge one battery at a time. Note that it may take longer time to have battery fully charged than usual.
The percentage of battery power will be displayed. See Field Case Display on page 35
490 Micro GC Field Case User Manual
Installation and Use 2
Field Case Display
490 Micro GC Field Case User Manual
The field case display shows the battery status, GC status, and gas pressure.
Percentage of charge is shown for each battery, followed by a letter indicating battery status.
When the GC is connected, GC status is shown on the right of the display. The possible states are Not Ready, Ready, Run, and Error. In addition to the display text, color LED will show at the far right of the display.
Table 7 Battery status descriptions
Battery Status Description
R Ready
U Using
C Charging
Figure 22 Field case display
Table 8 GC status color descriptions
GC Status LED color
Not Ready Yellow
Ready Off
Run Green
Error Red
35
2 Installation and Use
Internal Buzzer
36
The buzzer is located in the field case and gives information about the battery status.
Table 9 Buzzer action
Duration Action
Interval beep (10 seconds) Battery voltage below 10.5 Volts; recharge battery
Continuous beep Battery voltage below 10 Volts; recharge battery
490 Micro GC Field Case User Manual
Installation and Use 2
Carrier Gas Refill Instructions
490 Micro GC Field Case User Manual
Your Micro GC field case is equipped with refillable, high pressure carrier supply cylinders which have been approved to 12,400 kPa. The 2-channel field case is equipped with two 75mL gas cylinders. The 4-channel field case is quipped with either one or two 300mL gas cylinders.
When the instrument is in use, the Micro GC needs at least 550 10 kPa (80 2 psi) to work properly.
Refilling is done by means of a carrier gas refill assembly.
This special tool is connected directly to the valve on the gas supply cylinder. Because this connection differs from country to country, Agilent Technologies, Inc. offers a range of Refill Assemblys to meet all major standards. If the connector of this device, despite a careful choice, does not match your supply unit, ask your local gas supplier for the right part. It should have 1/4 inch NPT male thread at one end. Exchanging is easy.
Figure 23 Carrier gas refill assembly
Pressure gauge
Relief valve
Connection to gas supply cylinder
3-way ball valve check valve
Fast connection to carrier fill port
Copper tubing
Attach Carrier Gas Refill Assembly to Gas Source
1 Connect refill gas assembly and fast connector (0100-2908) with copper tubing (G1530-20620).
Figure 24 Fast connector
37
38
2 Installation and Use
2 Ensure the 3-way ball valve is in the check valve position.
3 Connect the refill assembly to a gas source as shown in Figure 25.
4 Open the gas source valve.
Figure 25 Gas refill assembly connected to a gas supply
High-pressure gas stores an incredible amount of energy and is danger-
490 Micro GC Field Case User Manual
Installation and Use 2
Removing a Gas Cylinder From the Field Case
490 Micro GC Field Case User Manual
1 Close gas cylinder valve by turning valve knob clockwise.
Figure 26 Close gas cylinder
39
40
2 Installation and Use
2 Push the release lever at the right side of the gas cylinder.
3 Pull vertically on cylinder assembly to remove.
Figure 27 Push release lever
Figure 28 Removing the gas cylinder
490 Micro GC Field Case User Manual
Installation and Use 2
490 Micro GC Field Case User Manual
If pressure is too high, for example above 50bar, it will not be easy to
Reducing Gas Flow Module Pressure
NOTE release the gas cylinder from field case. Follow below step to reduce gas flow module pressure
1 Close the gas cylinder valve.
2 Remove the external gas supply connector plug.
3 Switch between internal and external gas source several times to reduce pressure inside flow module. Every switch will reduce pressure approximately 4-5bar.
Figure 29 Gas source switch
Gas source switch
Gas Cylinder Refill Procedure
1 Install the Refill Assembly onto the gas supply cylinder (see page 37).
2 Remove gas cylinder from field case (see page 39).
3 Ensure that on/off valve on gas cylinder is closed.
41
42
2 Installation and Use
4 Ensure 3-way ball valve on refill assembly is set to the check valve position.
5 Connect the gas cylinder to the fast connector at the end of refill assembly.
Figure 30 Gas cylinder
Figure 31 Ball valve in check valve position
Figure 32 Fast connector connected to gas cylinder
On/Off valve
3-way ball valve
Fast connector
On/Off valve
490 Micro GC Fi
Installation and Use 2
490 Micro GC Field Case User Manual
6 Open the on/off valve on the gas cylinder. If some carrier gas is still in the cylinder, you will hear the sound of gas relief from check valve.
7 Read the pressure gauge on the refill assembly and ensure the pressure is zero.
8 Switch the 3-way ball valve to the gas source position. If the pressure of the pressure source is higher than 124bar, you will hear the pressure released from the relief valve. switch the 3-way ball valve back to middle position.
9 If the pressure of gas source is lower than 124bar, switch the 3-way ball valve back to the middle position when the needle in gauge stabilizes.
10 Close the on/off valve on the gas cylinder.
Figure 33 Ball valve in gas source position
Figure 34 Ball valve in middle position
3-way ball valve
3-way ball valve
43
44
2 Installation and Use
11 Switch the 3-way ball valve on refill assembly to the check valve position to release the remaining gas in transfer tubing. You will hear the relief sound.
12 Disconnect the gas cylinder.
Figure 35 Gas cylinder On/Off valve
Figure 36 Ball valve in check valve position
On/Off valve
3-way ball valve
The carrier gas cylinder can only be filled to the maximum when the
NOTE Refilling at high pressure may make the surface of cylinder warm, while fast flushing may make it colder.
490 Micro GC Fi
Installation and Use 2
Exchanging Gas Species or First Refill of the Gas Cylinder
490 Micro GC Field Case User Manual
1 Open the on/off valve on the gas cylinder. If some carrier gas is still in the cylinder, you will hear the sound of gas relief from check valve.
2 Read the pressure gauge and ensure the pressure is zero.
3 If the pressure of gas source is lower than 124bar, switch the 3-way ball valve back to middle position when the needle in gauge stabilizes. If the pressure of the gas source is higher than 124bar, switch the 3-way ball valve back to middle position when the needle in gauge shows higher than 124bar (12400kpa).
4 Switch the 3-way ball valve on the refill assembly to the check valve position. Release the carrier gas in the gas cylinder. You will hear the relief sound.
5 Repeat steps 2 through 4 three times.
6 Close the on/off valve on the gas cylinder.
7 Switch the 3-way ball valve on refill assembly to the check valve position. Release the remaining gas in the transfer tubing. You will hear the relief sound.
NOTE Refilling at high pressure may make the surface of cylinder warm, while fast flushing may make it colder.
45
2 Installation and Use
Shipping
46
If the Micro GC field case must be transported by shipping, it is very important to follow these shipping instructions:
Relieve the pressure completely from the internal carrier gas cylinders
Avoid shipping the battery connected with the instrument.
Cleaning
CAUTION
To keep the Micro GC field case surface clean:
Clean only when the Micro GC field case is disconnected from the charger, power supply, and battery packs.
Use a soft (not hard or abrasive) brush to carefully brush away all dust and dirt.
Use a soft, clean cloth, dampened with mild detergent to clean the outside of the case.
Never clean the inside of the case!
Be careful not to get water on the electronics components.
Do not use compressed air to clean the case.
Disposal
Disposal must be carried out in accordance with all (environmental) regulations applicable in your country.
490 Micro GC Field Case User Manual
Agilent 490 Micro GC Field Case User Manual
3 Battery Pack and Charger
Battery Pack 48
Battery Charger 55
Error Message and Battery Status Indication 58
47Agilent Technologies
3 Battery Pack and Charger
Battery Pack
Introduction
48
This Battery Pack (CP740328), Figure 37, may be charged inside field case or with the Charger (CP740427), see Figure 41 on page 55. It may only be used to power the field case. This Battery Pack is tailored to meet the power needs of the Micro GC.
Figure 37 Battery Pack (p/n CP740328)
General precautions
WARNING It is the responsibility of the Customer to inform Agilent Technologies Customer Support Representatives if this battery has been used in com- bination with the Micro GC for the analysis of hazardous samples prior to any instrument service being performed or when an instrument is being returned to the Agilent Service Center for repair.
WARNING This battery (p/n CP740328) is designed for use in combination with the Micro GC field case. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
490 Micro GC Field Case User Manual
Battery Pack and Charger 3
Battery location
490 Micro GC Field Case User Manual
The Ni-MH rechargeable battery is located inside the field case along the hinge in a specially designed compartment.
Figure 38 The battery compartment
Battery Compartment
To access the battery, loosen the captive Torx T-20 screws and lift off the battery compartment cover.
The battery will be visible.
Figure 39 Removing the battery compartment cover
49
50
3 Battery Pack and Charger
The battery can be removed by unplugging the connector and lifting the battery out of the compartment.
Figure 40 Removing the battery
Battery handling
Never disassemble the battery as the electrolyte inside is strong alkaline and can damage skin and clothes.
Never attempt to short-circuit the battery. Doing so can damage the product and generated heat can cause burns.
Do not dispose of the battery in a fire. Disposing of the battery in fire can cause the battery to rupture. Also avoid placing batteries in water as this causes batteries to cease to function.
Never solder anything directly to the battery. This can destroy the safety features of the battery by damaging the safety vent inside the cap.
Do not use the battery in an appliance or for purposes for which it was not intended.
Do not place containers with flammable liquids near the battery. Spilling liquid over hot parts may cause fire.
Never use alcohol or thinners to clean the Battery Pack. These chemicals can damage the case.
490 Micro GC Field Case User Manual
Battery Pack and Charger 3
Charging
490 Micro GC Field Case User Manual
In order to take full advantage of the properties of the Ni-MH Battery Pack and to prevent problems due to improper use, note the following points when charging.
WARNING Charge the battery within an ambient temperature range of 0 C to 45 C.
WARNING Never attempt reverse charging. Charging with polarity reversed can cause a gas pressure inside the batter to rise, which can activate the safety vent, lead to alkaline electrolyte leakage, rapid deterioration ini battery performance, battery swelling, or battery rupture.
Batteries should always be charged prior to use. Be sure to charge correctly, and use only the battery charger that has been supplied by Agilent Technologies, Inc.
The batteries can be charged with the external charger or from in the field case. It is recommended that the batteries be charged in the field case.
Ambient temperature affects charging efficiency. As charging efficiency is best within a temperature range of 10 C to 45 C, when charging, be sure the Battery Pack and charger are within this temperature range.
At temperatures below 0 C the gas absorption reaction is not adequate, causing gas pressure inside the battery to rise, which can activate the safety vent and lead to leakage of alkaline gas and deterioration in performance and battery leakage.
Do not parallel charge batteries.
Avoid overcharging when using the external charger. Repeated overcharging can lead to deterioration in performance. Overcharging means charging a battery when it is already fully charged.
To avoid overcharging, use a timer to measure the total charge time.
51
52
3 Battery Pack and Charger
When charging for the first time after long-term storage, deactivation of reactants may lead to increased battery voltage and decreased battery capacity. Restore such batteries to original performance by using the external charger and repeating several cycles of charging and discharging.
When recharging, place the battery in a suitable location with sufficient ventilation. Space around the battery must be sufficient to enable cooling.
Never reverse charge or overcharge with high currents. Doing so causes rapid gas generation and increased gas pressure, thus causing batteries to swell or rupture.
Never insert the battery with the positive and negative poles reversed as this can cause the battery to swell or rupture.
Do not charge the battery if there is a possibility of any kind of electrical damage. Instead, disconnect the battery power cords and contact your Agilent Technologies office.
Discharging
Since over-discharging (deep discharge) damages the battery characteris-
Storage
CAUTION tics do not leave the battery connected to the instrument for long periods of time.
When storing batteries for more than 1 year, disconnect batteries from the
Store the battery in a dry location with low humidity, no corrosive gases, and at a temperature range of -20 C to +50 C.
Storing the battery in a location where humidity is extremely high or where temperatures fall below -20 C or above +50 C can lead to rusting of metallic parts and battery leakage due to expansion or contraction of parts composed of organic materials.
Long-term storage can accelerate battery self-discharging and lead to the deactivation of reactants; locations where the temperature ranges between +10 C and +30 C are suitable for long-term storage.
490 Micro GC Field Case User Manual
Battery Pack and Charger 3
490 Micro GC Field Case User Manual
When charging for the first time after long-term storage, deactivation of reactants may lead to increased battery voltage and decreased battery capacity. Restore such batteries to their original performance by using the external charger and repeating several cycles of charging and discharging.
Battery service life
Batteries used under proper conditions of charging and discharging can be used 500 cycles or more.
Batteries are chemical products involving internal chemical reactions. Performance deteriorates not only with use but also during prolonged storage.
Normally, a battery will last two years (or 500 cycles) if used under proper conditions and not overcharged or overdischarged.
However, failure to satisfy conditions concerning charging, temperature, and other factors during actual use can lead to shortened life (or cycle life), damage to products, and deterioration in performance due to leakage and shortened service life.
Significantly reduced service time in spite of proper charging means that the life of the battery has been exceeded. At the end of service life, an increase in internal resistance or an internal short-circuit failure may occur.
Cleaning
To keep the Battery Pack surface clean:
Clean only when the Battery Pack is disconnected from the charger.
Use a soft (not hard or abrasive) brush to carefully brush away all dust and dirt.
If the outer case is dirty (never clean the inside!), clean it with a soft, clean cloth dampened with mild detergent.
Never use alcohol or thinners to clean the Battery Pack, these chemicals can damage the case.
Be careful not to get water on the electronics components.
Do not use compressed air to clean the battery pack.
53
3 Battery Pack and Charger
Disposal
54
When the Battery Pack or parts of it has reached the end of its useful life, disposal must be carried out in accordance with all (environment) regulations applicable in your country.
Technical specifications
Table 10 Battery Pack technical specifications
Technology Ni-MH (Nickel Metal Hydride)
Number of cells 10
Cell-orientation 1 10 in row
Nominal output voltage 12 Vdc
Capacity 10000 mAh
Protection Built-in thermo fuse resettable (75 C) and current (14 A)
Output cable Two wires, UL1015 approved
Output connector 2-circuit JST connector Pin 1 (red) +12 V Pin 2 (black) ground
Operation temperature 0 C to +50 C
Charge temperature 0 C to +45 C
Storage temperature 0 C to +70 C
Humidity 0 % to 95 % (non-condensing)
Safety UL2054, UL1015 (cabling) and CE
Size 348 71 37 mm
Weight 2,000 grams
490 Micro GC Field Case User Manual
Battery Pack and Charger 3
Battery Charger
Introduction
490 Micro GC Field Case User Manual
This Charger (p/n CP740427), Figure 41, must only be used in combination with the Micro GC Battery Pack (CP740328), see Figure 37 on page 48. This Charger is tailored to meet the power needs of the Battery Pack.
Figure 41 Battery Charger (p/n CP740427)
Features
Micro controller controlled charging
Short circuit detection and electronic protection against reversed battery
Battery condition at the beginning of the charge is of no importance for the Battery Packs
Safety stages like voltage gradient supervision and -dV switch off as well as a safety timer are integrated
55
3 Battery Pack and Charger
Charger status
56
LED Status
Yellow Battery not connected*
* There is power to the charging unit, but the battery is not connected.
Yellow Battery initialization and analysis
After the battery is connected to the charging unit.
Orange Fast charge
Green with intermittent Yellow flash Top-off charge
Green Ready to use and trickle charge
Alternating Orange-Green Error
Operation
CAUTION Keep this charger in a dry place (indoor use only). The charger should be disconnected from the power supply when not in use. Do not plug in the charger if there is a damaged cabinet or power plug.
1 The charger will start automatically as soon as a Battery Pack is installed and the charger is plugged into the power supply. (The LED is yellow.)
2 The charger starts the fast charging procedure (LED is orange).
3 After the fast charging procedure, the charger switches automatically over to top-off charge (LED is green with a yellow flash).
4 After the top-off charging procedure, the charger switches automatically over to trickle charge (LED is green).
5 Trickle charge will last until either the battery or the power supply cable is disconnected.
Cleaning
To keep the charger surface clean:
Remove the power cable.
Use a soft brush (not hard or abrasive) to carefully brush away all dust and dirt.
490 Micro GC Field Case User Manual
Battery Pack and Charger 3
490 Micro GC Field Case User Manual
If the outer case is dirty (never clean the inside!), clean it with a soft, clean cloth dampened with mild detergent.
Never use alcohol or thinners to clean the Battery Charger, these chem icals can damage the case.
Be careful not to get water on the electronics components.
Do not use compressed air to clean the Battery Charger.
Disposal instructions
Disposal must be carried out in accordance with all environment regulations applicable in your country.
Technical specifications
Table 11 Battery Charger technical specifications
Model Mascot MC-2215
Main voltage 90 to 264 Vac
Frequency 50 to 60 Hz
Power consumption 45 W
Output voltage 6.2 - 18 Vdc
Maximum charge current
1.8 A
Output cable Two wires, 1.7 meter
Output connector 2-circuit Mate-N-lok connector Pin 1 (red) +12 V Pin 2 (black) ground
Humidity 0 % to 90 % (non-condensing)
Conformations 2006/95/EC (low voltage)
2004/108/EC (EMC)
93/42/EEC (Medical)
Size 107 67 36.5 mm
Weight 250 grams
57
3 Battery Pack and Charger
Error Message and Battery Status Indication
Error message
58
While charging battery inside field case, the system will monitor and display battery status on screen. When an error message appears, contact a service engineer for support.
During battery charging mode, battery voltage over 15.2 V or below 10.0 V, or a voltage drop over 0.5 V indicates a defective battery pack.
During battery charging mode, a charging current over 2.5 A or below 0.4 A may be caused by a defective charging circuit.
If an error is detected, the system will stop operation and show an error message on the display panel, see Figure 42.
Follow the instructions to power off micro GC:
1 Unplug the power adapter from field case power connector.
2 Disconnect the batteries from the field case.
Figure 42 Error message
Table 12 Error codes
Error message Battery 1 Battery 2
Over voltage (>15.2 V) 101 201
Under voltage (<10.0 V) 102 202
Over current (>2.5 A) 103 203
Under current (<0.4 A) 104 204
Over voltage drop (>0.5 V) 105 205
490 Micro GC Field Case User Manual
Battery Pack and Charger 3
Battery status indicator
490 Micro GC Field Case User Manual
The Ni-MH rechargeable battery, used under proper conditions of charging and discharging, can be used 500 cycles or more. The system monitors the battery status, and will indicate when it is reaching its service life. At that time, it is recommended to replace it with a new battery pack, CP740328. See Figure 43.
Figure 43 Battery status indicator
59
60
3 Battery Pack and Charger
490 Micro GC Field Case User Manual
Agilent Technologies
Related manuals for Agilent 490 Micro GC Field Case Chromatograph User Manual
Manualsnet FAQs
If you want to find out how the 490 Micro GC Field Case Agilent works, you can view and download the Agilent 490 Micro GC Field Case Chromatograph User Manual on the Manualsnet website.
Yes, we have the User Manual for Agilent 490 Micro GC Field Case as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.
The User Manual should include all the details that are needed to use a Agilent 490 Micro GC Field Case. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.
The best way to navigate the Agilent 490 Micro GC Field Case Chromatograph User Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.
This Agilent 490 Micro GC Field Case Chromatograph User Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.
You can download Agilent 490 Micro GC Field Case Chromatograph User Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.
To be able to print Agilent 490 Micro GC Field Case Chromatograph User Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent 490 Micro GC Field Case Chromatograph User Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.