Contents

Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual PDF

1 of 52
1 of 52

Summary of Content for Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual PDF

Agilent Technologies

Agilent 1260 Infinity II High Temperature Viscometer

User Manual

Notices Agilent Technologies, Inc. 2017-2018 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla- tion into a foreign language) without prior agreement and written consent from Agi- lent Technologies, Inc. as governed by United States and international copyright laws.

Manual Part Number G7821-90000 Rev. B

Edition 05/2018 Printed in Germany Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn

Warranty The material contained in this docu- ment is provided as is, and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor- mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or sub- contract, Software is delivered and licensed as Commercial computer software as defined in DFAR 252.227-7014 (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regu- lation or contract clause. Use, duplication or disclosure of Software is subject to Agi- lent Technologies standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will

receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.

Safety Notices

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated condi- tions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indi- cated conditions are fully understood and met.

Agilent 1260 Infinity II High Temperature Viscometer User Manual

In This Guide...

In This Guide...

This manual covers the G7821B Agilent 1260 Infinity II High Temperature Viscometer

1 Introduction to the High Temperature Viscometer This chapter gives an introduction to the High Temperature Viscometer.

2 Site Requirements and Specifications This chapter provides information on environmental requirements, physical and performance specifications.

3 Installing the High Temperature Viscometer This chapter explains the installation of the High Temperature Viscometer.

4 Using the Module This chapter provides information on how to use the module.

5 Troubleshooting and Diagnostics This chapter gives an overview about the troubleshooting and diagnostic features and the different user interfaces.

6 Maintenance and Repair This chapter provides general information on maintenance and repair of the viscometer.

7 Appendix This chapter provides safetey and other general information.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 3

Contents

Contents

1 Introduction to the High Temperature Viscometer 7 Introduction to the High Temperature Viscometer 8 How the Detector Operates 9 Universal Calibration 11 Polymer Branching 13 Detection Principle 14

2 Site Requirements and Specifications 17 Site Requirements 18 Physical Specifications 21 Performance Specifications 22

3 Installing the High Temperature Viscometer 23 Installing the High Temperature Viscometer 24 Flow Split Ratio Calculations 26

4 Using the Module 27 Using the Module 28 Purging the Viscometer 29

5 Troubleshooting and Diagnostics 31 General Troubleshooting Guide 32

6 Maintenance and Repair 33 Introduction to Maintenance 34 Warnings and Cautions 35 Cleaning the Module 37 Storage of the Detector 37 Overview of Maintenance Parts 37

4 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Contents

7 Appendix 39 General Safety Information 40 Waste Electrical and Electronic Equipment Directive 46 Radio Interference 47 Sound Emission 48 Solvent Information 49 Agilent Technologies on Internet 50

Agilent 1260 Infinity II High Temperature Viscometer User Manual 5

Contents

6 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Agilent 1260 Infinity II High Temperature Viscometer User Manual

1 Introduction to the High Temperature Viscometer Introduction to the High Temperature Viscometer 8 How the Detector Operates 9 Universal Calibration 11 Polymer Branching 13 Detection Principle 14

This chapter gives an introduction to the High Temperature Viscometer.

7Agilent Technologies

1 Introduction to the High Temperature Viscometer Introduction to the High Temperature Viscometer

Introduction to the High Temperature Viscometer

The Agilent High Temperature viscometer is based on the proven four-capillary design and is intended for online measurement of viscosity when integrated into the 1260 Infinity II High Temperature GPC system. The combination of refractive index (RI) and viscosity detection provides accurate molecular weight determination for all polymer types based on the universal principle, as well as valuable branching information not otherwise obtainable from a concentration detector alone.

Due to the four-capillary technology, the differential pressure (DP) signal is very stable against flow rate fluctuations. A continuous monitor of the pressure through the bridge network also provides a consistent reference of the flow system, the inlet pressure (IP). The instrument can be used in both a batch mode, which yields an average intrinsic viscosity, and a true gel permeation chromatography (GPC) mode, which yields a slice-by-slice intrinsic viscosity distribution. From the intrinsic viscosity and known retention volumes or known Mark-Houwink-Sakurada coefficients, the molecular weight as well as distribution of a polymer may be determined.

The Agilent high temperature viscometer consists of four components:

1 Precision-engineered 4-capillary bridge

2 High sensitivity pressure transducers

3 Electronic control module

4 Discrete 5 V power supply module (PSU)

The viscometer bridge is located in the Agilent column oven. The pressure transducer assembly is housed in the side panel of the Agilent High Temperature GPC system. The electronic control module, where the differential pressure (DP) and inlet pressure (IP) signals are obtained, is stand-alone and located alongside the GPC system.

8 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Introduction to the High Temperature Viscometer 1 How the Detector Operates

How the Detector Operates

When a polymer dissolves in a liquid, the interaction of the two components stimulates an increase in polymer dimensions over that of the unsolvated state. Due to the vast difference in size between solvent and solute, the frictional properties of the solvent in the mixture are drastically altered, and an increase in viscosity occurs which should reflect the size and shape of the dissolved solute, even in dilute solutions.

The dissolved polymer coil disturbs the linear flow of solvent, resulting in a resistance to flow, which is observed as an increase in viscosity.

This was first recognized in 1930 by Staudinger, who found an empirical relation between relative magnitude of increase in viscosity and the molar mass of the polymer.

The intrinsic viscosity [] is a measure of the viscosity of a polymer solution, as the concentration tends to zero, that is at very low concentrations. It is this parameter, therefore that is a characteristic of isolated polymer chains in solution and can be considered to be proportional to the density of the polymer coil. For a given polymer and solvent system at a specified temperature, [] can be related to molecular weight, M, through the Mark-Houwink-Sakurada equation:

where K and are coefficients specific to the polymer, solvent system and temperature. These can be determined by calibrating with polymer standards of known molecular weights. A plot of Log() versus Log(M), a Mark-Houwink-Sakurada plot, will be a straight line as long as the universal calibration is obeyed.

6 @ = KM

Agilent 1260 Infinity II High Temperature Viscometer User Manual 9

1 Introduction to the High Temperature Viscometer How the Detector Operates

Figure 1 Mark-Houwink-Sakurada plot

Values of reflect the size of the molecule in solution and for a random coil polymer, range between 0.5 for a polymer dissolved in a theta solvent to about 0.8 in a very good solvent.

10 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Introduction to the High Temperature Viscometer 1 Universal Calibration

Universal Calibration

Benoit (1967) showed that polymers of different structure fall on the same calibration curve if the intrinsic viscosity is included as a calibration parameter. A calibration plot of the product of intrinsic viscosity and molecular weight ([]*M) versus elution volume is a "universal calibration" as shown in Figure 2 on page 12. This allows the system to be calibrated with polymer standards of one type and then analyze polymers of another type and still obtain accurate molecular weight distributions.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 11

1 Introduction to the High Temperature Viscometer Universal Calibration

Figure 2 A universal calibration plot

Therefore for an unknown polymer sample the universal calibration is used to convert the intrinsic viscosity, determined for each slice across the distribution from the viscometer, to molecular weight.

12 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Introduction to the High Temperature Viscometer 1 Polymer Branching

Polymer Branching

The determination of the degree of long chain branching has long occupied polymer chemists. In polymer solutions, the branching factor, g, is the starting point for many branching calculations:

where

b Intrinsic viscosity of the branched polymer

l Intrinsic viscosity of the linear polymer of the same molecular weight

Structural model parameter, which is user defined

g = 1

b6 @c m 1

Agilent 1260 Infinity II High Temperature Viscometer User Manual 13

1 Introduction to the High Temperature Viscometer Detection Principle

Detection Principle

The Agilent 1260 High Temperature Viscometer employs the fluid flow equivalent of the analogous wheatstone bridge electrical circuit. Solvent travels down a bridge of four capillaries of equal resistance arranged as shown in Figure 3 on page 14.

Figure 3 Plumbing Diagram of Agilent 1260 High Temperature Viscometer

The differential pressure transducer monitors the pressure drop across the bridge, P, and the inlet pressure transducer measures the pressure drop through the bridge, Pi. When the sample is introduced across capillary (C4) of the bridge, an imbalance is caused due to the hold-up columns. This change in pressure, P, is proportional to the specific viscosity, sp, according to the following equation:

where

P differential pressure across the bridge

Pi pressure of the flow through the capillary bridge

SP = Pi 2P 4P

14 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Introduction to the High Temperature Viscometer 1 Detection Principle

The instrument makes a true differential measurement directly, making it very sensitive to dilute solution viscosity measurements. The intrinsic viscosity, [] is ideally obtained by measuring the specific viscosity values at several finite concentrations and extrapolate to zero concentration.

However, the concentrations used in GPC are low enough for the equation to be valid over a typical chromatogram, and thus an extrapolation to the true intrinsic viscosity is not required, since the concentration is close enough to the limiting concentration.

All of the data acquisition and data processing are accomplished using the Agilent GPC-SEC software, which enables calculation of molecular weight and intrinsic viscosity distributions using algorithms for Universal Calibration.

Branching information can also be obtained.

6 @ = lim c " 0 c

SPa l

Agilent 1260 Infinity II High Temperature Viscometer User Manual 15

1 Introduction to the High Temperature Viscometer Detection Principle

16 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Agilent 1260 Infinity II High Temperature Viscometer User Manual

2 Site Requirements and Specifications Site Requirements 18 Physical Specifications 21 Performance Specifications 22

This chapter provides information on environmental requirements, physical and performance specifications.

17Agilent Technologies

2 Site Requirements and Specifications Site Requirements

Site Requirements

A suitable environment is important to ensure optimal performance of the instrument.

Power Considerations The module power supply has wide ranging capability. It accepts any line voltage in the range described in Table 1 on page 21. Consequently there is no voltage selector in the rear of the module. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.

WARNING Hazard of electrical shock or damage of your instrumentation can result, if the devices are connected to a line voltage higher than specified.

Connect your instrument to the specified line voltage only.

WARNING The module is partially energized when switched off, as long as the power cord is plugged in. Repair work at the module can lead to personal injuries, e.g. electrical shock, when the cover is opened and the module is connected to power.

Always unplug the power cable before opening the cover.

Do not connect the power cable to the instrument while the covers are removed.

WARNING Inaccessible power plug. In case of emergency it must be possible to disconnect the instrument from the power line at any time.

Make sure the power connector of the instrument can be easily reached and unplugged.

Provide sufficient space behind the power socket of the instrument to unplug the cable.

18 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Site Requirements and Specifications 2 Site Requirements

Power Cords Country-specific power cords are available for the module. The female end of all power cords is identical. It plugs into the power-input socket at the rear. The male end of each power cord is different and designed to match the wall socket of a particular country or region.

Agilent makes sure that your instrument is shipped with the power cord that is suitable for your particular country or region.

WARNING Absence of ground connection The absence of ground connection can lead to electric shock or short circuit.

Never operate your instrumentation from a power outlet that has no ground connection.

WARNING Unintended use of supplied power cords Using power cords for unintended purposes can lead to personal injury or damage of electronic equipment.

Never use a power cord other than the one that Agilent shipped with this instrument.

Never use the power cords that Agilent Technologies supplies with this instrument for any other equipment.

Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.

WARNING Power cords Solvents may damage electrical cables.

Prevent electrical cables from getting in contact with solvents.

Exchange electrical cables after contact with solvents.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 19

2 Site Requirements and Specifications Site Requirements

Bench Space The module dimensions and weight (see Table 1 on page 21) allow you to place the module on a bench next to the 1260 Infinity II HT GPC System. It needs an additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.

If the bench shall carry a complete GPC system, make sure that the bench is designed to bear the weight of all modules.

The module should be operated in a horizontal position.

Environment Your module will work within the specifications at ambient temperatures and relative humidity described in Table 1 on page 21.

ASTM drift tests require a temperature less than 2 C/h (3.6 F/h) over one hour period. Our published drift specification (refer also to Table 2 on page 22) is based on these conditions. Larger ambient temperature changes will result in larger drift.

Drift performance depends on control of the temperature environment fluctuations. To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 C/h (1.8 F/h).

CAUTION Condensation within the module Condensation can damage the system electronics.

Do not store, ship or use your module under conditions where temperature fluctuations could cause condensation within the module.

If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.

NOTE This module is designed to operate in a typical electromagnetic environment, that is where RF transmitters such as mobile telephones may not be used in close proximity.

20 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Site Requirements and Specifications 2 Physical Specifications

Physical Specifications

Table 1 Physical Specifications

Type Specification Comments

Weight 7 kg (15.5 lbs)

Dimensions (height width depth)

135 x 235 x 130 mm (5.3 x 9.3 x 5.2 inches)

Line voltage 100 240 V~, 10 Control unit

Fuses 5 V (2.4 A) PSU

Line frequency 47 or 63 Hz, 5 %

Power consumption 70 W / 80 VA

Ambient operating temperature

10 40 C (50 104 F)

Ambient non-operating temperature

-40 70 C (-40 158 F)

Humidity < 80 % r.h. at 40 C (104 F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, CSA, UL

Installation category II, Pollution degree 2

For indoor use only.

ISM Classification ISM Group 1 Class A According to CISPR 11

Agilent 1260 Infinity II High Temperature Viscometer User Manual 21

2 Site Requirements and Specifications Performance Specifications

Performance Specifications

Table 2 Agilent 1260 Infinity II High Temperature Viscometer (G7821B) Performance Specifications

Type Specification Comments

Detection type Viscometer

Linearity 0.5 % FS

Shear rate 110-5 Pa*s

Baseline Noise <0.25 mV DP output

Baseline Drift <5 mV

Capillary Dimensions 0.01 in (id) x 24 in (l)

Output Differential Pressure 1 mV/Pa

Output Inlet Pressure 10 mV/Pa

22 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Agilent 1260 Infinity II High Temperature Viscometer User Manual

3 Installing the High Temperature Viscometer Installing the High Temperature Viscometer 24 Flow Split Ratio Calculations 26

This chapter explains the installation of the High Temperature Viscometer.

23Agilent Technologies

3 Installing the High Temperature Viscometer Installing the High Temperature Viscometer

Installing the High Temperature Viscometer

Figure 4 Front View of the Electronic Control Module for the Agilent GPC220 Viscometer

1 Viscometer analog outputs; 1V FSD The viscometer analog outputs, DP, IP and Specific Viscosity (Visc).

2 Transducers cable connection port Connects the DP and IP pressure transducers to the control module.

3 DP and IP autozero buttons Autozeros the DP and IP outputs. For the IP autozero button, the button needs to be pressed and held for a few seconds before the IP output is autozeroed.

4 DP/IP output display Displays the DP/IP output.

5 DP/IP output display indicators A red LED indicates which output is being displayed.

6 Toggle display button Toggles the output display between the DP and IP outputs.

7 DP offset adjustment screw Adjusts the DP zero offset by 20 Pa. To adjust the zero offset turn the screw.

24 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Installing the High Temperature Viscometer 3 Installing the High Temperature Viscometer

Connect the electronic control module for the Agilent High Temperature viscometer as follows:

1 Using the supplied transducer cable connect the 6 pin MIL-C connector end to the port on the right hand side panel of the Agilent GPC 220 instrument and the 9 pin D connector end to the port labelled Transducers on the control module.

2 Using the Detector output cables supplied, attach one end of the cables to the Ports labeled "DP out" and "IP out" on the front of the control module and the other ends to the Detector Input Channels (for example, Channels 2 and 3) on the front of the 1260 Infinity II DataStream.

3 Connect the supplied 5 V DC power supply unit to the rear of the control module and connect the power supply unit to an earthed mains supply using the power cable provided.

NOTE The Agilent Viscometer also has a specific viscosity output that is obtained from the VISC outport. The specific viscosity output cannot be used with Cirrus Multi.

NOTE Powering the control module off and on will reset the output offsets.

NOTE The lengths of the inlet tubes are such as to obtain a flow split between the viscometer NOTE and RI of between 50 and 70 %. For flow split ratio calculations, see Flow Split Ratio Calculations on page 26.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 25

3 Installing the High Temperature Viscometer Flow Split Ratio Calculations

Flow Split Ratio Calculations

The flow split ratio is the ratio between the recorded inlet pressures (IP) from the viscometer with and without the RI detector connected in the system.

Therefore, if the flow split ratio is 60 %, then 60 % of the flow are going through the viscometer, that is, if running the system at 1 mL/min then the flow through the viscometer is 0.6 mL/min.

FlowSplitRatio = IP(Visc Only)

IP(RI+Visc) * 100

26 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Agilent 1260 Infinity II High Temperature Viscometer User Manual

4 Using the Module Using the Module 28 Purging the Viscometer 29

This chapter provides information on how to use the module.

27Agilent Technologies

4 Using the Module Using the Module

Using the Module

The 1260 Infinity II High Temperature Viscometer module is powered on and off using a mains power socket with the supplied 5 V DC power supply unit and mains power cable. Switching this on provides power to the electronic control module. Confirmation of power on is provided when the output display on the module displays a value and one of the output display indicators is illuminated.

NOTE Powering the control module off and on will reset the output offsets.

28 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Using the Module 4 Purging the Viscometer

Purging the Viscometer

This procedure describes how to purge the inlet pressure (IP) and differential pressure (DP) transducers.

1 Ensure the solvent is flowing through the system.

2 Turn the knob labelled IP on the front RHS panel to PURGE.

The IP transducer and its connecting tubes are flushed through with solvent.

3 Turn the IP knob to RUN and the knob labelled DP to PURGE.

The solvent flow is switched and the DP transducer and its connecting tubes are flushed with solvent.

4 Turn the DP knob to RUN.

The solvent flow is switched back to normal operation, that is through the viscometer bridge.

5 Purge the IP and DP transducers for at least 5 minutes, at a flow rate of 1.0 mL/min, or until the IP/DP meter shows a steady signal.

NOTE The waste line for the DP transducer is a permanent waste line.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 29

4 Using the Module Purging the Viscometer

30 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Agilent 1260 Infinity II High Temperature Viscometer User Manual

5 Troubleshooting and Diagnostics General Troubleshooting Guide 32

DP/IP Baseline Drift 32 DP/IP Baseline Noise 32 Low Sensitivity 32 Long Retention Times 32

This chapter gives an overview about the troubleshooting and diagnostic features and the different user interfaces.

31Agilent Technologies

5 Troubleshooting and Diagnostics General Troubleshooting Guide

General Troubleshooting Guide

DP/IP Baseline Drift Allow sufficient time for system to equilibrate.

Ensure the instrument is located in a thermally stable environment.

Degas solvent.

Purge DP/IP transducers.

Ensure laboratory temperature not changing dramatically.

DP/IP Baseline Noise Purge DP/IP transducers thoroughly to remove any air bubbles.

Purge pump head to remove any air bubbles.

Degas solvent.

Low Sensitivity Purge DP/IP transducers.

Check flow split.

Long Retention Times Check pump flow rate.

Check for leak on system.

32 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Agilent 1260 Infinity II High Temperature Viscometer User Manual

6 Maintenance and Repair Introduction to Maintenance 34 Warnings and Cautions 35 Cleaning the Module 37 Storage of the Detector 37 Overview of Maintenance Parts 37

This chapter provides general information on maintenance and repair of the viscometer.

33Agilent Technologies

6 Maintenance and Repair Introduction to Maintenance

Introduction to Maintenance

The module is designed for easy maintenance. Maintenance can be done from the front with the module in place at the back of the Agilent 1260 Infinity II High Temperature System.

NOTE There are no serviceable parts inside.

Do not open the module.

34 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Maintenance and Repair 6 Warnings and Cautions

Warnings and Cautions

WARNING Module is partially energized when switched off, as long as the power cord is plugged in. Risk of stroke and other personal injury. Repair work at the module can lead to personal injuries, e. g. shock hazard, when the module cover is opened and the instrument is connected to power.

Never perform any adjustment, maintenance or repair of the module with the cover removed and with the power cord plugged in.

WARNING Sharp metal edges Sharp-edged parts of the equipment may cause injuries.

To prevent personal injury, be careful when getting in contact with sharp metal areas.

WARNING Toxic, flammable and hazardous solvents, samples and reagents The handling of solvents, samples and reagents can hold health and safety risks.

When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.

The volume of substances should be reduced to the minimum required for the analysis.

Do not operate the instrument in an explosive atmosphere.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 35

6 Maintenance and Repair Warnings and Cautions

CAUTION Electronic boards and components are sensitive to electrostatic discharge (ESD). ESD can damage electronic boards and components.

Be sure to hold the board by the edges, and do not touch the electrical components. Always use ESD protection (for example, an ESD wrist strap) when handling electronic boards and components.

CAUTION Safety standards for external equipment

If you connect external equipment to the instrument, make sure that you only use accessory units tested and approved according to the safety standards appropriate for the type of external equipment.

36 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Maintenance and Repair 6 Cleaning the Module

Cleaning the Module

To keep the module case clean, use a soft cloth slightly dampened with water, or a solution of water and mild detergent.

Storage of the Detector

In case the detector is not used for some time (stored), fill the bridge (be sure to purge both IP and DP) with Tetrahydrofuran (THF).

Overview of Maintenance Parts

WARNING Liquid dripping into the electronic compartment of your module can cause shock hazard and damage the module

Do not use an excessively damp cloth during cleaning.

Drain all solvent lines before opening any connections in the flow path.

Item p/n Description

1 PL0810-3076 PL-GPC220 Delay Columns (x2)

Agilent 1260 Infinity II High Temperature Viscometer User Manual 37

6 Maintenance and Repair Overview of Maintenance Parts

38 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Agilent 1260 Infinity II High Temperature Viscometer User Manual

7 Appendix General Safety Information 40

General Safety Information 40 Safety Standards 40 General 40 Before Applying Power 41 Ground the Instrument 41 Do Not Operate in an Explosive Atmosphere 42 Do Not Remove the Instrument Cover 42 Do Not Modify the Instrument 42 In Case of Damage 42 Solvents 43 Symbols 44

Waste Electrical and Electronic Equipment Directive 46 Radio Interference 47 Sound Emission 48 Solvent Information 49 Agilent Technologies on Internet 50

This chapter provides safetey and other general information.

39Agilent Technologies

7 Appendix General Safety Information

General Safety Information

General Safety Information The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customers failure to comply with these requirements.

Safety Standards This is a Safety Class I instrument (provided with terminal for protective earthing) and has been manufactured and tested according to international safety standards.

General Do not use this product in any manner not specified by the manufacturer. The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.

WARNING Ensure the proper usage of the equipment. The protection provided by the equipment may be impaired.

The operator of this instrument is advised to use the equipment in a manner as specified in this manual.

40 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Appendix 7 General Safety Information

Before Applying Power

Ground the Instrument

WARNING Wrong voltage range, frequency or cabling Personal injury or damage to the instrument

Verify that the voltage range and frequency of your power distribution matches to the power specification of the individual instrument.

Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.

Make all connections to the unit before applying power.

NOTE Note the instrument's external markings described under Symbols on page 44.

WARNING Missing electrical ground Electrical shock

If your product is provided with a grounding type power plug, the instrument chassis and cover must be connected to an electrical ground to minimize shock hazard.

The ground pin must be firmly connected to an electrical ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 41

7 Appendix General Safety Information

Do Not Operate in an Explosive Atmosphere

Do Not Remove the Instrument Cover

Do Not Modify the Instrument Do not install substitute parts or perform any unauthorized modification to the product. Return the product to an Agilent Sales and Service Office for service and repair to ensure that safety features are maintained.

In Case of Damage

WARNING Presence of flammable gases or fumes Explosion hazard

Do not operate the instrument in the presence of flammable gases or fumes.

WARNING Instrument covers removed Electrical shock

Do Not Remove the Instrument Cover

Only Agilent authorized personnel are allowed to remove instrument covers. Always disconnect the power cables and any external circuits before removing the instrument cover.

WARNING Damage to the module Personal injury (for example electrical shock, intoxication)

Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.

42 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Appendix 7 General Safety Information

Solvents

WARNING Toxic, flammable and hazardous solvents, samples and reagents The handling of solvents, samples and reagents can hold health and safety risks.

When working with these substances observe appropriate safety procedures (for example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.

Do not use solvents with an auto-ignition temperature below 200 C (392 F). Do not use solvents with a boiling point below 56 C (133 F).

Avoid high vapor concentrations. Always keep the temperature in the sample compartment at least 25 K below the boiling point of the solvent used.

Do not operate the instrument in an explosive atmosphere.

Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for example, carbon disulfide).

Reduce the volume of substances to the minimum required for the analysis.

Never exceed the maximum permissible volume of solvents (8 L) in the solvent cabinet. Do not use bottles that exceed the maximum permissible volume as specified in the usage guideline for solvent cabinet.

Ground the waste container.

Regularly check the filling level of the waste container. The residual free volume in the waste container must be large enough to collect the waste liquid.

To achieve maximal safety, regularly check the tubing for correct installation.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 43

7 Appendix General Safety Information

Symbols Table 3 Symbols

The apparatus is marked with this symbol when the user should refer to the instruction manual in order to protect risk of harm to the operator and to protect the apparatus against damage.

Indicates dangerous voltages.

Indicates a protected ground terminal.

The apparatus is marked with this symbol when hot surfaces are available and the user should not touch it when heated up.

Sample Cooler unit is designed as vapor-compression refrigeration system. Contains fluorinated greenhouse gas (refrigerant) according to the Kyoto protocol. For specifications of refrigerant, charge capacity, carbon dioxide equivalent (CDE), and global warming potential (GWP) see instrument label.

Flammable Material For Sample Thermostat which uses flammable refrigerant consult Agilent Information Center / User Manual before attempting to install or service this equipment. All safety precautions must be followed.

Confirms that a manufactured product complies with all applicable European Community directives. The European Declaration of Conformity is available at: http://regulations.corporate.agilent.com/DoC/search.htm

Manufacturing date.

Power symbol indicates On/Off. The apparatus is not completely disconnected from the mains supply when the power switch is in the Off position

44 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Appendix 7 General Safety Information

Pacemaker Magnets could affect the functioning of pacemakers and implanted heart defibrillators. A pacemaker could switch into test mode and cause illness. A heart defibrillator may stop working. If you wear these devices keep at least 55 mm distance to magnets. Warn others who wear these devices from getting too close to magnets.

Magnetic field Magnets produce a far-reaching, strong magnetic field. They could damage TVs and laptops, computer hard drives, credit and ATM cards, data storage media, mechanical watches, hearing aids and speakers. Keep magnets at least 25 mm away from devices and objects that could be damaged by strong magnetic fields.

Indicates a pinching or crushing hazard

Indicates a piercing or cutting hazard.

Table 3 Symbols

WARNING A WARNING alerts you to situations that could cause physical injury or death.

Do not proceed beyond a warning until you have fully understood and met the indicated conditions.

CAUTION A CAUTION alerts you to situations that could cause loss of data, or damage of equipment.

Do not proceed beyond a caution until you have fully understood and met the indicated conditions.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 45

7 Appendix Waste Electrical and Electronic Equipment Directive

Waste Electrical and Electronic Equipment Directive

Abstract The Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC), adopted by EU Commission on 13 February 2003, is introducing producer responsibility on all electric and electronic appliances starting with 13 August 2005.

NOTE This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.

Product Category:

With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a Monitoring and Control Instrumentation product.

NOTE Do not dispose of in domestic household waste

To return unwanted products, contact your local Agilent office, or see http://www.agilent.com for more information.

46 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Appendix 7 Radio Interference

Radio Interference

Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.

Test and Measurement If test and measurement equipment is operated with equipment unscreened cables and/or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 47

7 Appendix Sound Emission

Sound Emission

Manufacturers Declaration This statement is provided to comply with the requirements of the German Sound Emission Directive of 18 January 1991.

This product has a sound pressure emission (at the operator position) < 70 dB.

Sound Pressure Lp < 70 dB (A)

At Operator Position

Normal Operation

According to ISO 7779:1988/EN 27779/1991 (Type Test)

48 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Appendix 7 Solvent Information

Solvent Information

Observe the following recommendations on the use of solvents.

Brown glass ware can avoid growth of algae.

Avoid the use of the following steel-corrosive solvents:

solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on),

high concentrations of inorganic acids like sulfuric acid and nitric acid, especially at higher temperatures (if your chromatography method allows, replace by phosphoric acid or phosphate buffer which are less corrosive against stainless steel),

halogenated solvents or mixtures which form radicals and/or acids, for example:

2CHCl3 + O2 2COCl2 + 2HCl

This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol,

chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, di-isopropyl ether) should be filtered through dry aluminium oxide which adsorbs the peroxides,

solvents containing strong complexing agents (e.g. EDTA),

mixtures of carbon tetrachloride with 2-propanol or THF.

Avoid the use of dimethyl formamide (DMF). Polyvinylidene fluoride (PVDF), which is used in leak sensors, is not resistant to DMF.

Agilent 1260 Infinity II High Temperature Viscometer User Manual 49

7 Appendix Agilent Technologies on Internet

Agilent Technologies on Internet

For the latest information on products and services visit our worldwide web site on the Internet at:

http://www.agilent.com

50 Agilent 1260 Infinity II High Temperature Viscometer User Manual

Appendix 7 Agilent Technologies on Internet

Agilent 1260 Infinity II High Temperature Viscometer User Manual 51

www.agilent.com

In This Book

This manual contains information on the Agilent 1260 Infinity II High-Temperature Viscometer (G7821B).

The manual describes the following:

Introduction

Site requirements

Using the module

Troubleshooting and diagnostics

Maintenance

Safety information

Agilent Technologies 2017-2018

Printed in Germany 05/2018

*G7821-90000* *G7821-90000* G7821-900

Manualsnet FAQs

If you want to find out how the 1260 Infinity 2 High Temperature Viscometer Agilent works, you can view and download the Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual on the Manualsnet website.

Yes, we have the User Manual for Agilent 1260 Infinity 2 High Temperature Viscometer as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The User Manual should include all the details that are needed to use a Agilent 1260 Infinity 2 High Temperature Viscometer. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent 1260 Infinity 2 High Temperature Viscometer Chromatograph User Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.