Agilent 7890 Series Gas Chromatograph

Advanced Operation Manual

Notices Agilent Technologies, Inc. 2007-2013

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla- tion into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

Manual Part Number

G3430-90015

Edition

Twelfth edition, January 2013 Eleventh edition, July 2012 Tenth edition, March 2012 Ninth edition, June 2011 Eighth edition, February 2011 Seventh edition, November 2010 Sixth edition, June 2010 Fifth edition, September 2009 Fourth edition, April 2009 Third edition, September 2008 Second edition, January 2008 First edition, April 2007

Printed in USA

Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA

412 800820 3278

Safety Notices

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Firmware Version

This manual is written for 7890 Series GCs using firmware version B.02.01.

Advanced Operation Manual 3

Contents

1 Programming

Run Time Programming 12

Using run time events 12 Programming run time events 13 The run table 13 Adding events to the run table 13 Editing events in the run table 14 Deleting run time events 14

Clock Time Programming 15

Using clock time events 15 Programming clock time events 15 Adding events to the clock table 16 Editing clock time events 16 Deleting clock time events 16

User-Key Programming 17

To program a User Key 17 To play back (execute) the stored keystrokes 17 To erase the stored keystrokes 17

Post Run Programming 18

To enable a post run program 18 To disable a post run program 18

2 Flow and Pressure Modules

About Flow and Pressure Control 20

Maximum operating pressure 20 PIDs 21

Inlet Modules 22

Detector Modules 23

Pressure Control Modules 24

Auxiliary Pressure Controllers 27

Restrictors 28

Examples 30

1. Using an Aux EPC channel to supply purge gas to a splitter 30 2. Using the PCM channels 30

4 Advanced Operation Manual

3 Inlets

Using Hydrogen 35

Inlet Overview 36

Carrier Gas Flow Rates 37

About Gas Saver 38

To use gas saver 39

Pre Run and Prep Run 40

The [Prep Run] key 40 Auto Prep Run 41

About Heaters 42

About the Split/Splitless Inlet 44

Septum tightening (S/SL) 44 Standard and high-pressure versions of the S/SL inlet 44 Split/Splitless inlet split mode overview 45 Split/Splitless inlet splitless mode overview 46 The S/SL inlet pulsed split and splitless modes 47 Split/Splitless inlet split mode minimum operating pressures 48 Selecting the correct S/SL inlet liner 49 Vapor Volume Calculator 51 Setting parameters for the S/SL split mode 51 Selecting parameters for the S/SL splitless mode 52 Setting parameters for the S/SL splitless mode 53 Setting parameters for the S/SL pulsed modes 54

About the Multimode Inlet 55

Septum tightening (MMI) 55 Heating the MMI 56 Cooling the MMI 56 MMI split mode minimum operating pressures 57 Selecting the correct MMI liner 58 Vapor Volume Calculator 60 MMI split and pulsed split modes 60 MMI splitless and pulsed splitless modes 64 MMI solvent vent mode 70 MMI Direct Mode 78 To develop a MMI method that uses large volume injection 79 Multiple injections with the MMI 82

About the Packed Column Inlet 88

Setting parameters 90

Advanced Operation Manual 5

About the Cool On-Column Inlet 92

Setup modes of the COC inlet 93 Retention gaps 93 COC inlet temperature control 93 Setting COC inlet flows/pressures 94 Setting COC inlet parameters 95

About the PTV Inlet 96

PTV sampling heads 96 Heating the PTV inlet 97 Cooling the PTV inlet 98 PTV inlet split and pulsed split modes 98 PTV inlet splitless and pulsed splitless modes 102 PTV inlet solvent vent mode 109 To develop a PTV method that uses large volume injection 117 Multiple injections with the PTV inlet 120

About the Volatiles Interface 126

VI operating modes 127 About the VI split mode 128 About the VI splitless mode 132 About the VI direct mode 137 Preparing the Interface for Direct Sample Introduction 140 VI direct mode setpoint dependencies 142 VI direct mode initial values 142 Setting parameters for the VI direct mode 143

4 Columns and Oven

About the Oven 146

Oven safety 146

Configuring the Oven 147

Cryogenic Operation 148

Cryogenic setpoints 148

About Oven Temperature Programming 150

Programming setpoints 150 Oven ramp rates 151 Setting the oven parameters for constant temperature 152 Setting the oven parameters for ramped temperature 152

About the Oven Insert 154

About Columns 155

Selecting the correct packed glass column type 155 About the column modes 155

6 Advanced Operation Manual

Select a column mode 156 Setting the column parameters for constant flow or constant

pressure 157 Enter a flow or pressure program (optional) 157 Programming column pressure or flow 158

Backflushing a Column 159

Backflushing when connected to an MSD 160 Backflushing using a capillary flow technology device 160

Nickel Catalyst Tube 165

About the nickel catalyst tube 165 Nickel catalyst gas flows 165 Setting temperatures for the nickel catalyst tube 166

5 Detectors

About Makeup Gas 168

About the FID 169

How FID units are displayed in Agilent data systems and on the GC 170 To light the FID flame 171 To extinguish the FID flame 171 FID automatic reignition (Lit offset) 171 Recommended starting conditions for new FID methods 172 Setting parameters for FID 173

About the TCD 174

TCD pneumatics 176 TCD carrier, reference, and makeup gas 176 TCD gas pressures 177 Selecting reference and makeup flows for the TCD 178 Chemically active compounds reduce TCD filament life 178 Changing the TCD polarity during a run 179 Detecting hydrogen with the TCD using helium carrier gas 179 Setting parameters for the TCD 180

About the uECD 182

uECD safety and regulatory information 182 uECD warnings 183 Safety precautions when handling uECDs 184 uECD gas flows 185 uECD linearity 185 uECD detector gas 185 uECD temperature 185 uECD analog output 186 Recommended starting conditions for new uECD methods 186

Advanced Operation Manual 7

uECD makeup gas notes 186 uECD temperature programming 187 Setting parameters for the uECD 187

About the NPD 188

New NPD features and changes 188 NPD software requirements 188 NPD flows and general information 188 NPD flow, temperature, and bead recommendations 189 NPD required gas purity 191 Setting parameters for the NPD 192 Selecting an NPD bead type 193 Changing from a ceramic bead to a Blos bead 194 Selecting an NPD jet 194 To configure the NPD 195 Automatically adjusting NPD bead voltage 196 Setting NPD adjust offset on the clock table 197 Aborting NPD adjust offset 197 Extending the NPD bead life 197 Setting the initial bead voltage for new beads 198 Setting NPD bead voltage manually (optional) 199

About the FPD+ 200

FPD+ linearity 201 FPD+ Lit Offset 201 FPD+ photomultiplier protection 201 FPD+ optical filters 201 Inlet liners for use with the FPD+ 201 FPD+ temperature considerations 202 FPD+ gas purity 202 FPD+ gas flows 202 Lighting the FPD+ flame 203 Setting parameters for the FPD+ 204

About the FPD 206

FPD linearity 207 FPD Lit Offset 207 Starting Up and Shutting Down the FPD 207 FPD photomultiplier protection 207 FPD optical filters 207 Inlet liners for use with the FPD 208 FPD temperature considerations 208 FPD gas purity 208 FPD gas flows 208

8 Advanced Operation Manual

Lighting the FPD flame 209 Setting parameters for the FPD 210

6 Valves

About Valves 214

The Valve Box 215

Heating the valves 215 Valve temperature programming 216 Configuring an Aux thermal zone 216

Valve Control 217

The valve drivers 217 The internal valve drivers 217 The external valve drivers 218

Valve Types 219

Configuring a Valve 220

Controlling a Valve 221

From the keyboard 221 From the run or clock time tables 221 Simple valve: column selection 221 Gas sampling valve 222 Multiposition stream selection valve with sampling valve 223

7 GC Output Signals

About Signals 226

Signal Types 227

Value 227

Analog Signals 229

Analog zero 229 Analog range 229 Analog data rates 230 Selecting fast peaks (analog output) 231

Digital Signals 232

Digital zero 232 Signal Freeze and Resume 232 Data rates with Agilent data systems 233 Zero Init Data Files 235

Advanced Operation Manual 9

Column Compensation 236

Creating a column compensation profile 237 Making a run using analog output column compensation 237 Making a run using digital output column compensation 237 Plotting a stored column compensation profile 238

Test Plot 239

8 Auxiliary Devices

Auxiliary Devices 242

About Auxiliary Pressure Control 242 About Aux Thermal Zone Control 243 About Cryo Trap Control 243 About Auxiliary Device Contacts 244 About the 24V Auxiliary Device Power Supply 244 About Auxiliary Columns 244 About Auxiliary Detectors 245

10 Advanced Operation Manual

11

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

1 Programming

Run Time Programming 12

Using run time events 12

Programming run time events 13

The run table 13

Adding events to the run table 13

Editing events in the run table 14

Deleting run time events 14

Clock Time Programming 15

Using clock time events 15

Programming clock time events 15

Adding events to the clock table 16

Editing clock time events 16

Deleting clock time events 16

Post Run Programming 18

User-Key Programming 17

To play back (execute) the stored keystrokes 17

To erase the stored keystrokes 17

To program a User Key 17

12 Advanced Operation Manual

1 Programming

Run Time Programming

Run time programming during a method allows certain setpoints to change automatically during a run as a function of the chromatographic run time. Thus an event that is programmed to occur at 2 minutes will occur 2 minutes after every injection.

Its uses include:

Controlling column switching or other valves

Changing analog signal definition, zero, or range

Controlling an auxiliary pressure channel

Changing polarity of a thermal conductivity detector (TCD)

Turning the hydrogen flow to a nitrogen- phosphorus detector (NPD) on or off

Switching digital signal output (requires an Agilent data system)

Pausing (freezing) and resuming digital signal output (requires an Agilent data system)

The changes are entered into a run table that specifies the setpoint to be changed, the time for the change, and the new value. At the end of the chromatographic run, most setpoints changed by a run time table are returned to their original values.

Valves can be run time programmed but are not restored to their starting position at the end of the run. You must program the reset operation in the run table if this action is desired. See From the run or clock time tables on page 221.

Using run time events

The [Run Table] key is used to program the following timed events.

Valves (1- 8)

Multiposition valve

Signal type (see Signal Types on page 227)

Analog signal definition, zero, and range

Auxiliary pressures (1 through 9)

TCD negative polarity (on/off)

Programming 1

Advanced Operation Manual 13

Detector gas flow (on/off), including NPD H2 fuel gas

Inlet septum purge flow

Programming run time events

1 Press [Run Table].

2 Press [Mode/Type] to see the available run time events.

3 Scroll to the event to be programmed. Press [Enter].

4 Enter values for the Time: and the otherparameter. Press [Enter] after each entry.

5 Press [Mode/Type] to add another event. Press [Status] to terminate entries.

The run table

The programmed events are arranged in order of execution time in the Run Table. This is a brief example:

RUN TABLE (1 of 3) Event 1 rotates a valve. Time: 0.10 Valve #2 On RUN TABLE (2 of 3) Event 2 adjusts the signal Time: 3 range. Analog signal 2 range 2 RUN TABLE (3 of 3) Event 3 resets Valve #2 to its Time: 4.20 original position in preparation Valve #2 Off for another run. Valves do not

reset automatically.

Adding events to the run table

1 To add new events to the run table, press [Mode/Type] while on the Time: line of any entry.

2 Select the event type.

3 Set appropriate Time: and other parameters. Some require numbers; others require [On/Yes] or [Off/No].

4 Repeat until all entries are added. Events are automatically placed in order by execution time.

14 Advanced Operation Manual

1 Programming

Editing events in the run table

1 Press [Run Table].

2 Move the cursor to the event you want to change.

3 To edit the time for an event, move the cursor to the line labeled Time:. Type the desired time and press [Enter].

4 To edit a setpoint value, scroll to the setpoint line. Press [On/Yes] or [Off/No] or enter a numeric value for the setpoint. Press [Enter].

Deleting run time events

1 Press [Run Table].

2 From within this table press [Delete] to delete events from the run time table. You will be asked to confirm the deletion.

3 Press [On/Yes] to delete the current timed event; press [Off/No] to cancel this operation.

4 To delete the entire table, press [Delete][Run Table].

Programming 1

Advanced Operation Manual 15

Clock Time Programming

Clock time programming allows certain setpoints to change automatically at a specified time during a 24- hour day. Thus, an event programmed to occur at 14:35 hours will occur at 2:35 in the afternoon. A running analysis or sequence has precedence over any clock table events occurring during this time. Such events are not executed.

Possible clock time events include:

Valve control

Method and sequence loading

Starting sequences

Initiating blank and prep runs

Column compensation changes

Adjustments of the detector offset

Using clock time events

The Clock Table function allows you to program events to occur during a day based on the 24- hour clock. Clock table events that would occur during a run or sequence are ignored.

For example, the clock table could be used to make a blank run before you even get to work in the morning.

Programming clock time events

1 Press [Clock Table].

2 Press [Mode/Type] to see the available clock time events.

3 Scroll to the parameter to be programmed.

4 Edit Time: and the setpoints for this event.

5 Press [Mode/Type] to add another event. Press [Status] to terminate entries.

When the clock event is executed, a confirming message appears.

16 Advanced Operation Manual

1 Programming

Adding events to the clock table

1 Press [Clock Table].

2 Press [Mode/Type]. When entries are added, they are automatically ordered chronologically.

3 Select the event type.

4 Set appropriate parameters.

5 Repeat this process until all entries are added.

Editing clock time events

1 Press [Clock Table] to view all events programmed.

2 Scroll to the event you want to change.

3 To edit the time for an event, move the cursor to the line labelled Time: and type the desired time.

4 To edit a setpoint value, scroll to the setpoint item. Press [On/Yes] or [Off/No], or enter a numerical value for the setpoint.

Deleting clock time events

1 Press [Clock Table].

2 Use [Delete] to remove events from the clock time table. You will be asked to confirm the deletion.

3 Press [On/Yes] to delete the current timed event; press [Off/No] to cancel this operation.

To delete the entire table, press [Delete][Clock Table].

Programming 1

Advanced Operation Manual 17

User-Key Programming

The two User Keys create macros (sets of frequently used keystrokes) and assign them to single keys. A macro is executed when the User Key is pressed.

The stored keystrokes may be any keys except [Start], [Prog], [User Key 1], or [User Key 2].

This discussion assumes that you wish to program [User Key 1]. The process is the same for [User Key 2].

To program a User Key

1 Press [Prog]. Press [User Key 1].

2 Press up to 31 keys, then press [User Key 1]. The keystrokes are stored.

To play back (execute) the stored keystrokes

Press [User Key 1].

To erase the stored keystrokes

Press [Prog][User Key 1][User Key 1]. This creates an empty macro.

18 Advanced Operation Manual

1 Programming

Post Run Programming

This function can be used with both isothermal and programmed methods. Post run is a period that begins at the end of the normal run. The parameters include:

TimeHow long is the post run period?

Oven TemperatureWhat is the oven temperature during the post run period?

Column n presFor a column controlled in a pressure mode, enter the pressure for this column during the post run period.

Column n flowFor a column controlled in a flow mode, enter the flow rate for this column during the post run period.

Enable Front inlet tempFor the Multimode inlet, set the post run inlet temperature. You can also press [On/Yes] and [Off/No] to turn this parameter on or off.

Enable Front inlet total flowFor the Multimode inlet, set the post run inlet total flow rate. You can also press [On/Yes] and [Off/No] to turn this parameter on or off.

Post run may be used to clean out a column in preparation for the next run, backflush a column to eliminate high- boilers, and other functions.

When the Post run Time elapses, the GC returns to the initial state defined in the current method. If it uses cryogenic cooling and you do not start another run quickly, you could waste considerable coolant while waiting for the next run. A solution is to use a sequence that loads a less wasteful method.

To enable a post run program

1 Press [Post Run].

2 Type a non- zero time for the post run duration and press [Enter]. The post run parameters available for the current GC configuration appear.

3 Scroll to each desired parameter, type the value for the post run period, and press [Enter].

To disable a post run program

1 Press [Post Run].

2 Type a 0 as the post run time and press [Enter].

19

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

2 Flow and Pressure Modules

About Flow and Pressure Control 20

Maximum operating pressure 20

PIDs 21

Inlet Modules 22

Detector Modules 23

Pressure Control Modules 24

Auxiliary Pressure Controllers 27

Restrictors 28

1. Using an Aux EPC channel to supply purge gas to a splitter 30

2. Using the PCM channels 30

20 Advanced Operation Manual

2 Flow and Pressure Modules

About Flow and Pressure Control

The GC uses four types of electronic flow or pressure controllers; inlet modules, detector modules, pressure control modules (PCMs), and auxiliary pressure controllers (Aux EPCs).

All of these modules mount in the slots at the top rear of the GC. The slots are identified by numbers, as shown here.

When a module is installed in a slot, it must be connected to the communications buss that runs underneath it. Each branch of the buss has an identifying label near the connector. The proper branch must be connected to the module for the firmware to recognize it.

Branch/slot assignments are:

If a detector is mounted in the left side carrier (as seen from the front side of the oven), it is controlled by EPC6. An extension cable connects to EPC6, runs across the top of the oven just in front of the top row of slots, and passes through an opening into the detector carrier.

Maximum operating pressure

The pneumatics modules of the GC will stand over 250 psi pressure, but may not function reliably. We recommend a maximum continuous operating pressure of 170 psi to avoid excessive wear and leaks.

Split vent traps and valves

Back of GC

Slot 1

Slot 2

Slot 3

Slot 4

Slot 5

Slot 6

Split vent traps and valves

Back of GC

EPC1

EPC2

EPC3

EPC4

EPC5

EPC6

Flow and Pressure Modules 2

Advanced Operation Manual 21

PIDs

The behavior of a pressure control module is governed by a set of three constants, called P (proportional), I (integral), and D (differential).

Certain gases or special applications (such as QuickSwap, headspace vial pressurization, or splitter and backflush applications) require different PIDs than those provided at the factory.

If you need to update or change the pneumatic PID values for an application, use the utility program on the documentation and utility DVD provided with the GC.

The table summarizes custom PID values required for selected applications. Note that if you are updating an AUX EPC module, you will need to change the frit for the channel used. See also Restrictors.

Table 1 PIDs and frits

Application Module AUX frit Select Available PID Values

QuickSwap AUX EPC 1 ring (or brown dot) Quickswap

Purged splitter and Deans Switch when using backflush

AUX EPC No color or rings Quickswap

Purged splitter and Deans Switch AUX EPC 1 ring (or brown dot) Standard

Headspace vial pressurization AUX EPC No color or rings AUX_EPC_Headspace

Headspace sampling loop PCM in backpressure control

PCM_Headspace

22 Advanced Operation Manual

2 Flow and Pressure Modules

Inlet Modules

These modules are used with specific inlets. They provide a controlled flow or pressure of carrier gas to the inlet and control the septum flow rate.

Module locations depend on the type of module and the length of the tubing connecting it to the inlet.

If you have a Front inlet, its flow module must go in Slot 1.

If you have a Back inlet, its flow module must go in Slot 2.

Flow and Pressure Modules 2

Advanced Operation Manual 23

Detector Modules

These are specific to the detector with which they are supplied, and differ according to the needs of that detector. For example, the FID module must supply controlled amounts of air, hydrogen, and makeup gas. The TCD module supplies the reference and makeup gases, but includes the reference switching valve that is essential to detector operation.

Module locations depend on the type of module and the length of the tubing connecting it to the detector.

If you have a Front detector, its flow module must go in Slot 3.

If you have a Back detector, its flow module must go in Slot 4.

If you have a TCD detector mounted on the left side of the GC, its flow module mounts in a special bracket near the detector.

24 Advanced Operation Manual

2 Flow and Pressure Modules

Pressure Control Modules

The PCM is a general purpose module with two independent control channels, designated 1 and 2. The general name of a PCM module is PCM #, where the # (actually, a letter) identifies the PCM (there can be up to 3 installed).

The two channels are not identical. Channel 1 is a simple forward- pressure regulated channel that maintains constant flow through a fixed restrictor.

Channel 2 may be used either as a forward- pressure regulator or as a back- pressure regulator, simply by reversing the input and output connections. The back

Flow and Pressure Modules 2

Advanced Operation Manual 25

pressure mode can be very useful with a gas sampling valve, where it ensures that the sample pressure in the loop remains constant.

For channel 1, gas input is via a threaded fitting. Gas output is via a coil of metal tubing with a Swagelok fitting on the end.

For channel 2, the connections are the same as for channel 1 if channel 2 is to be used as a forward- pressure regulator. They are reversedextra fittings/adapters will be neededif it is to be used as a back- pressure regulator.

PCMs can be installed in several locations:

In slot 1. The name is PCM A.

In slot 2. The name is PCM B.

26 Advanced Operation Manual

2 Flow and Pressure Modules

In slot 5. If there are no PCMs in slots 1 and 2, the name is PCM A. If another PCM is installed in either slot 1 or 2, the PCM in slot 5 takes the name that is not being used (A or B). If PCMs are installed in both slots 1 and 2, the name is PCM C.

In slot 6. The name is always PCM C.

Both channels of a PCM are controlled by the same parameter list. The first two lines refer to channel 1, the remaining lines refer to channel 2.

Split vent trap and valve

Slot 3

Slot 4

Back of GC

PCM C

PCM A

PCM B

PCM A or B

Flow and Pressure Modules 2

Advanced Operation Manual 27

Auxiliary Pressure Controllers

The Auxiliary Pressure Controller (Aux epc) is also a general purpose device. It has three independent forward- pressure regulated channels. Channels are designated by numbers 1 through 9 (there can be up to 3 Aux epcs), depending on where the module is installed.

If an auxiliary channel is specified as the Inlet during column configuration, that channel allows run time programming and three- ramp flow or pressure programming.

Gas input is via threaded fittings. Gas output is via coils of metal tubing with Swagelok fittings on the ends.

As you look at the module from the back of the GC, the channels are numbered from left to right according to this scheme:

In slot 6. The name is Aux epc #, where # is 1, 2, or 3 and identifies the channel. In this case, the Aux EPC is connected to the connector labeled EPC6.

In slot 5. The name is Aux epc #, where # is 4, 5, or 6 and identifies the channel. In this case, the Aux EPC is connected to the connector labeled EPC5.

In any open slot. The name is Aux epc #, where # is 7, 8, or 9 and identifies the channel. In this case, the Aux EPC is connected to the connector labeled AUX DET1 or AUX DET2.

In the third detector side of the GC. The name is Aux epc #, where # is 7, 8, or 9 and identifies the channel. In this case, the Aux EPC is connected to the connector labeled AUX DET1 or AUX DET2.

Split vent trap and valve

Slot 1

Slot 2

Slot 3

Back of GC

Slot 6

Slot 5

Slot 4

28 Advanced Operation Manual

2 Flow and Pressure Modules

Restrictors

Both PCMs and auxiliary channels are controlled by pressure setpoints. To work properly, there must be adequate flow resistance downstream of the pressure sensor. Each channel provides a frit- type restrictor. Four frits are available.

The one ring frit (low resistance, high flow) is installed in all channels in the AUX epc when the instrument (or accessory) is shipped. No frit ships in the PCM Aux channel.

Table 2 Auxiliary channel frits

Frit marking Flow resistance Flow characteristic Often used with

Three rings Blue

High 3.33 0.3 SCCM @ 15 PSIG NPD Hydrogen

Two rings Red

Medium 30 1.5 SCCM H2 @ 15 PSIG FID Hydrogen

One ring Brown

Low 400 30 SCCM AIR @ 40 PSIG FID Air, FPD Air, QuickSwap, Splitter, Deans Switch

None (brass tube) Zero No restriction Headspace vial pressurization

Flow and Pressure Modules 2

Advanced Operation Manual 29

When installing or replacing a frit, always use a new O- ring (5180- 4181, 12/pk).

Selecting a frit

The frits change the control range of the channels. The objective is to find a frit that allows the required range of flows at reasonable source pressures.

For an auxiliary channel ordered as an option (part of the GC order), use the frit supplied by the factory.

For an auxiliary channel ordered as an accessory (separate from the GC order), see the instruction information supplied with the accessory.

For a non- Agilent instrument, you must experiment to find the appropriate frit.

When you change a frit, you change the physical characteristics of the channel. It may be desirable (or necessary) to change the PID constants for that channel. See PIDs on page 21.

30 Advanced Operation Manual

2 Flow and Pressure Modules

Examples

1. Using an Aux EPC channel to supply purge gas to a splitter

An Aux EPC channel operates only in the forward- pressure mode; it provides a constant pressure at its outlet. It is used to provide gas to some other device, such as a splitter with a makeup gas input.

In this configuration, the gas eluting from the column is divided among three detectors. Because the resulting flows can be quite small, makeup gas is added at the splitter. An Aux EPC channel provides it.

Assume that we are using channel Aux epc 1 as the makeup source. The gas is plumbed to the leftmost (as seen from the back of the GC) fitting on the proper module. The outlet tubing for this channel goes to the splitter inside the ovenan extension tube may be needed to reach this far.

The medium frit is probably appropriate for this application.

Rather than have the makeup gas flowing all the time, consider turning it on and off with run time commands.

2. Using the PCM channels

The two channels in a PCM are different. Channel 1 is used to supply a pressure. Channel 2 may be used in the same way, but can also be used to maintain a pressure by reversing the input and output connections.

Split/splitless inlet

30 m x 0.25 mm x 0.25 m HP-MS5

0.507 m x 0.10 mm x 0 m 0.532 m x 0.18 mm x 0 m

1.444 m x 0.18 mm x 0 m

ECDAux EPC FPD MSD

Flow and Pressure Modules 2

Advanced Operation Manual 31

Channel 1: Forward-pressure only

This is identical to the carrier gas channel for the packed column inlet.

Channel 2: Two-way channel

If gas is supplied at the threaded connection and delivered by the tubing, this operates the same as channel 1. But the connections can be reversedrequiring some fittingsso that it will maintain the gas supplied to it at a fixed pressure. In this mode it behaves as a controlled leak.

This can be used to maintain the gas in a gas sampling valve at a fixed pressure, even if the supply varies somewhat. The result is improved reproducibility in sample amount.

The figure shows the connections with the valve in the Load position.

Valve PS

Sample in

Carrier in

To column

Back-pressure regulation

Sample out

Vent connectionThreaded

connection

32 Advanced Operation Manual

2 Flow and Pressure Modules

33

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

3 Inlets Using Hydrogen 35

Inlet Overview 36

Carrier Gas Flow Rates 37

About Gas Saver 38

Pre Run and Prep Run 40

Auto Prep Run 41

About Heaters 42

About the Split/Splitless Inlet 44

Split/Splitless inlet split mode overview 45

Split/Splitless inlet splitless mode overview 46

The S/SL inlet pulsed split and splitless modes 47

Split/Splitless inlet split mode minimum operating pressures 48

Selecting the correct S/SL inlet liner 49

Vapor Volume Calculator 51

Selecting parameters for the S/SL splitless mode 52

About the Multimode Inlet 55

Septum tightening (MMI) 55

Heating the MMI 56

Cooling the MMI 56

MMI split mode minimum operating pressures 57

Selecting the correct MMI liner 58

Vapor Volume Calculator 60

MMI split and pulsed split modes 60

MMI splitless and pulsed splitless modes 64

MMI solvent vent mode 70

MMI Direct Mode 78

To develop a MMI method that uses large volume injection 79

Multiple injections with the MMI 82

About the Cool On-Column Inlet 92

Retention gaps 93

COC inlet temperature control 93

Setting COC inlet flows/pressures 94

About the PTV Inlet 96

PTV sampling heads 96

Heating the PTV inlet 97

34 Advanced Operation Manual

3 Inlets

Cooling the PTV inlet 98

PTV inlet split and pulsed split modes 98

PTV inlet splitless and pulsed splitless modes 102

PTV inlet solvent vent mode 109

To develop a PTV method that uses large volume injection 117

Multiple injections with the PTV inlet 120

About the Volatiles Interface 126

About the VI split mode 128

About the VI splitless mode 132

About the VI direct mode 137

Preparing the Interface for Direct Sample Introduction 140

Setting parameters for the VI direct mode 143

Inlets 3

Advanced Operation Manual 35

Using Hydrogen

WARNING When using hydrogen (H2), as the carrier gas, be aware that hydrogen (H2) gas can flow into the oven and create an explosion hazard. Therefore, be sure that the supply is off until all connections are made, and ensure that the inlet and detector column fittings are either connected to a column or capped at all times when hydrogen (H2) gas is supplied to the instrument.

WARNING Hydrogen (H2) is flammable. Leaks, when confined in an enclosed space, may create a fire or explosion hazard. In any application using hydrogen (H2), leak test all connections, lines, and valves before operating the instrument. Always turn off the hydrogen (H2) supply at its source before working on the instrument.

36 Advanced Operation Manual

3 Inlets

Inlet Overview

Table 3 Comparing inlets

Inlet Column Mode Sample concentration

Comments

Sample to column

Split/splitless Capillary Split Pulsed split

Splitless Pulsed splitless

High High

Low Low

Useful with large injections

Useful with large injections

Very little Very little

All All

Multimode Capillary Split Pulsed split Splitless Pulsed splitless Solvent vent

Direct

High High Low Low Low Multiple injections

concentrate analytes and vent solvent

Very little Very little All All Most

All

Cool on-column Capillary n/a Low or labile Minimal discrimination and decomposition

All

Packed column Packed Large capillary

n/a n/a

Any Any OK if resolution

not critical

All All

Programmed temperature vaporization

Capillary Split Pulsed split Splitless Pulsed splitless Solvent vent

High High Low Low Low Multiple injections

concentrate analytes and vent solvent

Very little Very little All All Most

Volatiles interface (for use with external volatiles sampler)

Capillary Direct Split Splitless

Low High Low

Lowest dead volume Max flow = 100 mL/min

All Very little All

Inlets 3

Advanced Operation Manual 37

Carrier Gas Flow Rates

The flow rates in Table 4 are recommended for all column temperatures.

Table 4 Column size and carrier flow rate

Column type Column size Carrier flow rate, mL/min

Hydrogen Helium Nitrogen

Packed 1/8-inch 30 20

1/4-inch 60 40

Capillary 0.05 mm id 0.5 0.4 n/a

0.10 mm id 1.0 0.8 n/a

0.20 mm id 2.0 1.6 0.25

0.25 mm id 2.5 2.0 0.5

0.32 mm id 3.2 2.6 0.75

0.53 mm id 5.3 4.2 1.5

38 Advanced Operation Manual

3 Inlets

About Gas Saver

Gas saver reduces carrier flow from the split vent after the sample is on the column. It applies to the Split/Splitless and PTV inlets (all modes) and to the split and splitless modes of the Volatiles Interface. It is most useful in split applications.

Column head pressure and flow rate are maintained, while purge and split vent flows decrease. Flowsexcept column flowremain at the reduced level until you press [Prep Run].

The pulsed modes of the split/splitless and PTV inlets are similar except for the pressure pulse starting at [Prep Run] and ending at Pulse time. The solvent vent mode of the PTV is more complex; see PTV inlet solvent vent mode.

0 1 2 3 4 5 6 7 8

10

20 30

40 50

-1-2

0 1 2 3 4 5 6 7 8

10

20 30

40 50

-1-2

Split mode

Splitless mode

Split vent flow mL/min

Split vent flow mL/min

Time, min

Time, min

Gas saver flow

Gas saver flow

Gas saver flow

Gas saver flow

Purge flow

Regular flow

[Prep Run]

[Prep Run]

[Start]

[Start]

Gas saver time, 3 min

Gas saver time, 5 min

Run ends

Run ends Purge time, 2 min

Inlets 3

Advanced Operation Manual 39

To use gas saver

1 Press [Front Inlet] or [Back Inlet].

2 Turn gas saver On.

3 Set Gas saver flow. It must be at least 15 mL/min greater than the column flow.

4 If in split mode, set Saver time after injection time. In all other modes, set after Purge time.

40 Advanced Operation Manual

3 Inlets

Pre Run and Prep Run

With some inlets and operating modes, certain instrument setpoints are different between runs than during an analysis. To restore the setpoints for injection, you must place the GC into the Pre Run state.

You must use the Pre Run state when:

Using gas saver with any inlet.

Using splitless mode with any inlet.

Using a pressure pulse mode with any inlet.

Using the solvent vent mode of the PTV inlet.

Using the direct or splitless mode of the Volatiles Interface.

There are three ways to begin Pre Runmanually (press [Prep Run] before each run), automatically (for Agilent samplers), or using Auto Prep Run (for non- Agilent samplers). The three methods are discussed below.

During the Pre Run state:

The Pre Run light blinks and Not Ready is on.

Setpoints change to the correct values for injection.

Inlet, detector, and oven equilibration times begin.

When all equilibration times expire, the Pre Run light is on steadily. When all criteria for a run are met, the Not Ready light turns off. The GC is now ready for sample injection.

The [Prep Run] key

Press [Prep Run] before you inject a sample manually. The GC enters the Pre Run state. When the Pre Run light is steady and the Not Ready light goes off, begin the analysis.

Agilent samplers

If you are using an Agilent automatic sampling system, the [Prep Run] function is automatic.

Start the sampler. It generates the [Prep Run] function, When all the setpoints are reached and the GC becomes Ready, sample injection begins.

Inlets 3

Advanced Operation Manual 41

Non-Agilent samplers

With most automatic injection systems, you do not need to use the [Prep Run] key. If your sampler or automation controller (for example, an integrator or workstation) does not support the [Prep Run] function, you must set the GC to Auto Prep Run.

Auto Prep Run

To set this parameter, usually for a non- Agilent integrator, workstation, or other controlling device:

1 Press [Config] to view a list of configurable parameters.

2 Scroll to Instrument and press [Enter].

3 Scroll to Auto prep run and press [On/Yes].

42 Advanced Operation Manual

3 Inlets

About Heaters

Inlets (and detectors, valve boxes, etc.) are heated. There are six heater connectors on the GC mainframe, located as shown here:

All heater connectors are square, 4- conductor receptacles mounted on brackets.

The next table describes the heater locations that are available for each module.

1

2

3

4

5

6

Front of GC

Near front inlet

Near back inlet

Near top right corner of front detector board

Near top right corner of back detector board

Left end of valve bracket

Right end of valve bracket

Table 5 Heater connection locations by module

Module Available heater connection location

Front inlet 1 or None

Back inlet 2 or None

Front detector 3 or 5

Back detector 4 or 6

Aux detector 1 5 or 6 or 2

Aux detector 2 None

PCM A 5 or 6 or 1 or 2

PCM B 5 or 6 or 1 or 2

PCM C 5 or 6 or 1 or 2

AUX 1,2,3 None

AUX 4,5,6 None

Inlets 3

Advanced Operation Manual 43

Front FPD uses heater connectors 3 and 5.

Back FPD uses heater connectors 4 and 6.

FPDs can be configured for one or two heater versions.

AUX 7,8,9 None

Valve box 5 or 6 or Both

Aux heater 1 5

Aux heater 2 6

Table 5 Heater connection locations by module (continued)

Module Available heater connection location

44 Advanced Operation Manual

3 Inlets

About the Split/Splitless Inlet

This inlet is used for split, splitless, pulsed splitless, or pulsed split analyses. You can choose the operating mode from the inlet parameter list. The split mode is generally used for major component analyses, while the splitless mode is used for trace analyses. The pulsed splitless and pulsed split modes are used for the same type of analyses as split or splitless, but allow you to inject larger samples.

Septum tightening (S/SL)

Septum retainer nuts must be tightened enough to obtain a good gas seal, but not so much as to compress the septum and make it difficult to push a syringe needle through it.

For the standard septum retainer nut, an internal spring in the septum retainer applies pressure to the septum. For inlet pressures up to 100 psi, tighten the retainer until the C- ring lifts about 1 mm above the top surface. This is adequate for most situations.

With higher inlet pressures, tighten the septum retainer until the C- ring stops turning, indicating that the retainer is in firm contact with the septum.

If using a Merlin Microseal septum, finger tighten the septum nut, until snug (not loose). The pressure capacity depends on the duckbill seal used.

Standard and high-pressure versions of the S/SL inlet

The standard split/splitless inlet is rated to 100 psi pressure at the inlet. It is appropriate for most columns. The high- pressure inlet is rated to 150 psi pressureit is useful with very small diameter capillary columns that offer considerable resistance to gas flow.

Recommended source pressures are 120 psi and 170 psi respectively.

1 mm

Inlets 3

Advanced Operation Manual 45

To determine the version that you have, press [Front Inlet] or [Back Inlet], scroll to the Pressure line, and press the [Info] key. The display will show the pressure range for the inleteither 1 to 100 psi (for the standard version) or 1 to 150 psi (for the high- pressure version).

Split/Splitless inlet split mode overview

During a split injection, a liquid sample is introduced into a hot inlet where it vaporizes rapidly. A small amount of the vapor enters the column while the major portion exits from the split/purge vent. The ratio of column flow to split vent flow is controlled by the user. Split injections are primarily used for high concentration samples when you can afford to lose most of the sample out the split/purge vent. It is also used for samples that cannot be diluted.

The split ratio is equal to the split vent flow divided by the column flow. If the column has been configured, the desired split ratio can be entered directly.

The pneumatics for this inlet in split mode operation are shown in the figure below.

46 Advanced Operation Manual

3 Inlets

Split/Splitless inlet splitless mode overview

In this mode, the split vent valve is closed during the injection and remains so while the sample is vaporized in the liner and transferred to the column. At a specified time after injection, the valve opens to sweep any vapors remaining in the liner out the split vent. This avoids solvent tailing due to the large inlet volume and small column flow rate. Specify the purge time and purge flow rate in the inlet parameter list.

The septum purge flow may be either on at all times (Standard mode) or on only between the purge time and the end of the run (Switched mode).

If you are using gas saver, the gas saver time should be after the purge time.

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Column

Inlets 3

Advanced Operation Manual 47

The S/SL inlet pulsed split and splitless modes

The pressure pulse modes increase inlet pressure just before the beginning of a run and return it to the normal value after a specified amount of time. The pressure pulse:

reduces the solvent vapor volume

reduces the risk of inlet overload

tightens the sample band

may allow use of a 2 mm liner, reducing the active glass area

If your chromatography is degraded by the pressure pulse, a retention gap may help restore peak shape.

You must press [Prep Run] before doing manual injections in the pressure pulse modes. See Pre Run and Prep Run on page 40 for details.

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Column

48 Advanced Operation Manual

3 Inlets

You can do column pressure and flow programming when in the pressure pulse mode. However, the pressure pulse will take precedence over the column pressure or flow ramp.

Split/Splitless inlet split mode minimum operating pressures

The minimum recommended inlet total flow is 20 mL/minute. When the split/splitless inlet is operated in Split mode, there will be a minimum pressure at which the inlet can operate. Typically, low inlet pressures may be required for shorter, wide bore columns. The minimum pressure is a function of carrier gas type, total inlet flow, liner design, and possible contamination of the split vent tube or trap.

A wide bore column requires a much lower inlet pressure than a typical capillary column to maintain a given flow. Setting the split ratio (total flow) too high when using a wide bore column can create an unstable control relationship between the pressure and flow control loops.

0 1 2 3 4 5 6 7 8

Pressure pulse

Pressure or flow program

Pressure or Flow

Inlets 3

Advanced Operation Manual 49

These numbers are based on the resistance to flow of new, clean inlet systems. Sample condensation in the split vent tube or a dirty filter can make these values non- attainable.

Selecting the correct S/SL inlet liner

Split liner

A good liner for split mode operation will offer very little restriction to the split flow path between the bottom of the liner and the inlet gold seal and between the outside of the liner and the inside of the injection port body. The preferred Agilent split liner, part number 5183- 4647, incorporates a glass positioning bead on the bottom to facilitate this. It will also incorporate glass wool or some other source of surface area inside the liner that provides for complete sample vaporization across the boiling point range of the sample. Select an appropriate liner from Table 7.

Table 6 Approximate minimum viable inlet pressures for split/splitless inlet in split mode, in psi (kPa)

Split vent flow (mL/min)

50100 100200 200400 400600

Helium and hydrogen carrier gases

Split liners - 5183-4647, 19251-60540 2.5 (17.2) 3.5 (24.1 4.5 (31) 6.0 (41.4)

Splitless liners - 5062-3587, 5181-8818 4.0 (27.6) 5.5 (37.9) 8.0 (55.2) 11.0 (75.4)

Nitrogen carrier gas

Split liners - 19251-60540, 5183-4647 3.0 (20.7) 4.0(27.6)

Splitless liners - 5062-3587, 5181-8818 4.0 (27.6) 6.0 (41.4)

Table 7 Split mode liners

Liner Description Volume Mode Deactivated Part Number

Low Pressure Drop Positioning Bead

870 L Split Fast Injection

Yes 5183-4647

4mm ID, Glass Wool 990 L Split Fast Injection

No 19251-60540

Empty Pin & Cup 800 L Split Manual Only

No 18740-80190

Packed Pin & Cup 800 L Split Manual Only

No 18740-60840

50 Advanced Operation Manual

3 Inlets

Splitless liner

The liner volume must contain the solvent vapor. The liner should be deactivated to minimize sample breakdown during the purge delay. Solvent vapor volume can be reduced by using Pulsed Splitless mode. Use the Vapor Volume Calculator to determine vapor volume requirements.

Vapor volume < 300 L Use 2 mm liner (250 L volume), 5181- 8818 or similar.

Vapor volume 225 300 L Consider pulsed splitless mode to reduce vapor volume.

Vapor volume > 300 L Use 4 mm liner, 5062- 3587 or similar.

Vapor volume > 800 L Consider pulsed splitless mode to reduce vapor volume.

For thermally labile or reactive samples, use G1544- 80700 (open top) or G1544- 80730 (top taper) liners.

Table 8 Splitless mode liners

Liner Description Volume Mode Deactivated Part Number

Single Taper Glass Wool

900 uL Splitless Yes 5062-3587

Single Taper 900 uL Splitless Yes 5181-3316

Dual Taper 800 uL Splitless Yes 5181-3315

2 mm Quartz 250 uL Splitless No 18740-80220

2 mm Quartz 250 uL Splitless Yes 5181-8818

1.5 mm 140 uL Direct Inject, Purge and Trap, Headspace

No 18740-80200

Single Taper Glass Wool

900 uL Splitless Yes 5062-3587

Single Taper 900 uL Splitless Yes 5181-3316

4 mm Single Taper Direct column connect Yes G1544-80730

4 mm Dual Taper Direct column connect Yes G1544-80700

Inlets 3

Advanced Operation Manual 51

Vapor Volume Calculator

Agilent provides a Vapor Volume Calculator to help you determine if a liner is suitable for a method. To use the calculator install the Agilent Instrument utility provided with the GC. The calculator is also provided with the Agilent Instrument Utilities software.

Setting parameters for the S/SL split mode

Mode The current operating modesplit

Temperature Actual and setpoint inlet temperatures

Pressure Actual and setpoint inlet pressure

Split ratio The ratio of split vent flow to column flow. Column flow is set at the Column parameter list. This line appears only if your columns in the flow path are defined.

Split flow Flow, in mL/min, from the split vent. This line appears only if your columns in the flow path are defined.

Total flow This is the total flow into the inlet, which is the sum of the split vent flow, column flow, and septum purge flow. When you change the total flow, the split ratio and split vent flow change while the column flow and pressure remain the same.

Septum Purge Flow, in mL/min, through the septum purge line. Recommended range is 1 to 5 mL/min.

If all columns in the flow path are defined

1 Press [Front Inlet] or [Back Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Split.

3 Set the inlet temperature.

4 If you want a specific split ratio, scroll to Split ratio and enter that number. Split flow will be calculated for you.

5 If you want a specific split flow, scroll to Split flow and enter that number. Split ratio will be calculated for you.

6 If desired, turn on Gas saver. Set Saver time after the injection time. Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting the sample.

52 Advanced Operation Manual

3 Inlets

If a column in the flow path is not defined

1 Press [Front Inlet] or [Back Inlet].

2 Set the inlet temperature.

3 Set Total flow into the inlet. It must exceed your intended Septum Purge flow. Measure the split vent flow using a flow meter.

4 Subtract split vent flow and septum purge flow (see Pre Run and Prep Run on page 40) from Total flow to get column flow.

5 Calculate the split ratio (split vent flow/column flow). Adjust as needed.

Selecting parameters for the S/SL splitless mode

A successful splitless injection consists of these steps:

1 Vaporize the sample and solvent in a heated inlet.

2 Use a low flow and low oven temperature to create a solvent- saturated zone at the head of the column.

3 Use this zone to trap and reconcentrate the sample at the head of the column.

4 Wait until all, or at least most, of the sample has transferred to the column. Then discard the remaining vapor in the inletwhich is mostly solventby opening a purge valve. This eliminates the long solvent tail that this vapor would otherwise cause.

5 Raise the oven temperature to release the solvent and then the sample from the head of the column.

Some experimentation is needed to refine the operating conditions. Table 9 provides starting values for the critical parameters.

Table 9 Splitless mode inlet parameters

Parameter Allowed setpoint range Suggested starting value

Oven temperature No cryo, 24 C to 450 C CO2 cryo, 60 C to 450 C N2 cryo, 80 C to 450 C

10 C below solvent boiling point

Oven initial time 0 to 999.9 minutes Inlet purge time

Inlet purge time 0 to 999.9 minutes 2 x Liner volume Column flow

Inlets 3

Advanced Operation Manual 53

Setting parameters for the S/SL splitless mode

Mode The current operating modesplitless

Oven temperature Below solvent boiling point

Temperature Actual and setpoint inlet temperatures

Pressure Actual and setpoint inlet pressure in psi, bar, or kPa

Purge time The time, after the beginning of the run, when you want the purge valve to open. This is the time in which the vaporized sample transfers from the liner to the column.

Purge flow The flow, in mL/min, from the purge vent, at Purge time. You will not be able to specify this value if any column in the flow path is not defined.

Total flow The actual flow to the inlet during a Pre- run (Pre- run light is on and not blinking) and during a run before purge time. You cannot enter a setpoint at these times. At all other times, Total flow will have both setpoint and actual values.

Septum Purge Flow, in mL/min, through the septum purge line.

Septum Purge Flow Mode Standard (septum purge flow is On at all times) or Switched (septum purge flow is Off during injection, turns On at Purge time).

If all columns in the flow path are defined

1 Press [Front Inlet] or [Back Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Splitless.

3 Set the inlet temperature.

Gas saver time 0 to 999.9 minutes After purge time

Gas saver flow 15 to 1000 mL/min 15 mL/min greater than maximum column flow

Table 9 Splitless mode inlet parameters

Parameter Allowed setpoint range Suggested starting value

54 Advanced Operation Manual

3 Inlets

4 Enter a Purge time and a Purge flow.

5 If desired, turn on Gas saver. Make certain the time is set after the Purge time.

6 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample (this is automatic for Agilent ALS).

If a column in the flow path is not defined

1 Press [Front Inlet] or [Back Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Splitless.

3 Set the inlet temperature.

4 Enter a Purge time.

5 Set your Total flow greater than the sum of column flow plus the septum purge flowsee Pre Run and Prep Run on page 40to guarantee adequate column flow.

6 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample.

Setting parameters for the S/SL pulsed modes

The pulsed mode parameters are the same as the non- pulsed parameters, but with two additional values.

Pulsed pressure The inlet pressure you want at the start of the run. The pressure rises to this value when [Prep Run] is pressed and remains constant until Pulse time elapses, when it returns to Pressure.

Pulse time This is the time after the start of the run when the inlet pressure returns to Pressure.

Inlets 3

Advanced Operation Manual 55

About the Multimode Inlet

The Agilent Multimode (MMI) Inlet System has five operating modes:

The split mode is generally used for major component analyses.

The pulsed split mode is like the split mode, but with a pressure pulse applied to the inlet during sample introduction to speed the transfer of material to the column.

The splitless mode is used for trace analyses.

The pulsed splitless mode allows for a pressure pulse during sample introduction.

The solvent vent mode is used for large volume injection. Either single or multiple injections can be made for each run.

The direct mode allows for a direct forward pressure of carrier gas through the column. The split vent valve is closed.

The MMI can be used with both manual and automatic injection.

Automatic multiple injections (large volume injections) is not available under GC control alone. See MMI solvent vent mode on page 70.

Septum tightening (MMI)

Septum retainer nuts must be tightened enough to obtain a good gas seal, but not so much as to compress the septum and make it difficult to push a syringe needle through it.

For the standard septum retainer nut, an internal spring in the septum retainer applies pressure to the septum. For inlet pressures up to 100 psi, tighten the retainer until the C- ring lifts about 1 mm above the top surface. This is adequate for most situations.

1 mm

56 Advanced Operation Manual

3 Inlets

With higher inlet pressures, tighten the septum retainer until the C- ring stops turning, indicating that the retainer is in firm contact with the septum.

If using a Merlin Microseal septum, finger tighten the septum nut, until snug (not loose). The pressure capacity depends on the duckbill seal used.

Heating the MMI

Programming the MMI temperature is similar to programming the column oven. Access the inlet parameters by pressing [Front Inlet] or [Back Inlet]. Temperature can be programmed with an initial temperature and up to 10 ramps (rates and plateaus). See MMI split and pulsed split modes for details.

At the end of the run and during post- run, the MMI is held at its final temperature. This permits backflushing without contaminating the inlet.

Additional temperature ramps

For most purposes, the MMI is designed to hold the sample in the inlet liner until the entire samplethere could be several injectionshas been injected. Then the MMI is heated rapidly to transfer the sample to the column. This can be accomplished with an initial hold, a single ramp, and a hold at the end to complete sample transfer.

Additional ramps are available and have several possible uses:

The inlet can be heated to a high temperature to thermally clean the liner for the next run.

The inlet can be programmed downwardjust set the Final temp below the previous temperatureto reduce thermal stress on the inlet.

Downward programming can be used to prepare the inlet for the next run. This can reduce cycle time for greater sample throughput.

Cooling the MMI

If using cryogen to cool the MMI inlet, both liquid carbon dioxide (LCO2) and liquid nitrogen (LN2) are supported. In addition to the cryogenic coolants, the MMI supports compressed air cooling on both the LCO2 and LN2 cooling options for legacy applications.

Inlets 3

Advanced Operation Manual 57

If using cryo as the coolant when configuring the initial inlet setpoint, set the Use cryo temperature equal to or higher than the inlet setpoint to cool the inlet and hold the setpoint until the inlet temperature program exceeds the Use cryo temperature. If the Use cryo temperature is less than the inlet setpoint, cryogen will cool the inlet to the initial setpoint and turn off.

If using compressed air as the coolant when configuring the initial inlet setpoint, Use cryo temperature behaves differently when in Compressed air mode than it does when in N2 cryo or CO2 cryo mode. In Compressed air mode, the air coolant is used to cool the inlet regardless of the Use cryo temperature setpoint during the cooling cycle. If the inlet reaches setpoint, the air coolant is turned off and stays off. If the oven temperature is high enough or the inlet was very hot previously, it is possible that the inlet temperature will rise and the GC will go not ready. For this reason, it is better to set the instrument cooling configuration as N2 cryo or CO2 cryo when using compressed air as the coolant. When using compressed air, the LN2 hardware cools the inlet faster than the LCO2 hardware. Never use LCO2 or LN2 when the instrument is configured in Compressed air mode.

If cryo is turned on, and if the inlet is cooled during a run, cryogen is used to achieve the setpoint. This may have a negative impact on the chromatographic performance of the oven and cause distorted peaks.

When a method ends, the cooling setpoint returns to the initial state of the method, unless you load another method. To conserve coolant when the GC is idle, load a method that does not use a cooling configuration.

MMI split mode minimum operating pressures

The minimum recommended inlet total flow is 20 mL/minute. When the inlet is operated in a Split mode, there will be a minimum pressure at which the inlet can operate. Typically, low inlet pressures may be required for shorter, wide bore columns. The minimum pressure is a function of carrier gas type, total inlet flow, liner design, and possible contamination of the split vent tube or trap.

A wide bore column requires a much lower inlet pressure than a typical capillary column to maintain a given flow. Setting the split ratio (total flow) too high when using a wide bore column can create an unstable control relationship between the pressure and flow control loops.

58 Advanced Operation Manual

3 Inlets

These numbers are based on the resistance to flow of new, clean inlet systems. Sample condensation in the split vent tube or a dirty filter can make these values non- attainable.

Selecting the correct MMI liner

Split liner

A good liner for split mode operation will offer very little restriction to the split flow path between the bottom of the liner and the inlet body and between the outside of the liner and the inside of the inlet body. The preferred Agilent split liner, part number 5183- 4647, incorporates a glass positioning bead on the bottom to facilitate this. It will also incorporate glass wool or some other source of surface area inside the liner that provides for complete sample vaporization across the boiling point range of the sample. Select an appropriate liner from Table 11.

Table 10 Approximate minimum viable inlet pressures for MMI in split mode, in psi (kPa)

Split vent flow (mL/min)

50100 100200 200400 400600

Helium and hydrogen carrier gases

Split liners - 5183-4647, 19251-60540 2.5 (17.2) 3.5 (24.1 4.5 (31) 6.0 (41.4)

Splitless liners - 5062-3587, 5181-8818 4.0 (27.6) 5.5 (37.9) 8.0 (55.2) 11.0 (75.4)

Nitrogen carrier gas

Split liners - 19251-60540, 5183-4647 3.0 (20.7) 4.0(27.6)

Splitless liners - 5062-3587, 5181-8818 4.0 (27.6) 6.0 (41.4)

Table 11 Split mode liners

Liner Description Volume Mode Deactivated Part Number

Low Pressure Drop Positioning Bead

870 L Split Fast Injection

Yes 5183-4647

4mm ID, Glass Wool 990 L Split Fast Injection

No 19251-60540

Empty Pin & Cup 800 L Split Manual Only

No 18740-80190

Packed Pin & Cup 800 L Split Manual Only

No 18740-60840

Inlets 3

Advanced Operation Manual 59

Splitless liner

The liner volume must contain the solvent vapor. The liner should be deactivated to minimize sample breakdown during the purge delay. Solvent vapor volume can be reduced by using Pulsed Splitless mode. Use the Vapor Volume Calculator to determine vapor volume requirements.

Vapor volume < 300 L Use 2 mm liner (250 L volume), 5181- 8818 or similar.

Vapor volume 225 300 L Consider pulsed splitless mode to reduce vapor volume.

Vapor volume > 300 L Use 4 mm liner, 5062- 3587 or similar.

Vapor volume > 800 L Consider pulsed splitless mode to reduce vapor volume.

For thermally labile or reactive samples, use G1544- 80700 (open top) or G1544- 80730 (top taper) liners.

Table 12 Splitless mode liners

Liner Description Volume Mode Deactivated Part Number

Single Taper Glass Wool

900 uL Splitless Yes 5062-3587

Single Taper 900 uL Splitless Yes 5181-3316

Dual Taper 800 uL Splitless Yes 5181-3315

2 mm Quartz 250 uL Splitless No 18740-80220

2 mm Quartz 250 uL Splitless Yes 5181-8818

1.5 mm 140 uL Direct Inject, Purge and Trap, Headspace

No 18740-80200

Single Taper Glass Wool

900 uL Splitless Yes 5062-3587

Single Taper 900 uL Splitless Yes 5181-3316

4 mm Single Taper Direct column connect Yes G1544-80730

60 Advanced Operation Manual

3 Inlets

Vapor Volume Calculator

Agilent provides a Vapor Volume Calculator to help you determine if a liner is suitable for a method. To use the calculator install the Agilent Instrument utility provided with the GC. The calculator is also provided with the Agilent Instrument Utilities software.

MMI split and pulsed split modes

The two split modeswith or without a pressure pulsedivide the gas stream entering the inlet between the column flow and the split vent flow through the solenoid valve. The ratio of the split vent flow to the column flow is called the split ratio.

The next figure shows the flows with split and pulsed split modes.

4 mm Dual Taper Direct column connect Yes G1544-80700

Table 12 Splitless mode liners (continued)

Liner Description Volume Mode Deactivated Part Number

Inlets 3

Advanced Operation Manual 61

Cold split introduction

For cold split sample introduction, use an initial inlet temperature below the normal boiling point of the solvent. If the liner volume is enough to hold all the vaporized solvent, start the first inlet temperature ramp at 0.1 minutes with a high heating rate (500 C/min or higher). The final temperature should be high enough to volatilize the heaviest analytes from the liner and should be held for at least 5 minutes. A final temperature of 350 C for 5 minutes has proven sufficient to quantitatively transfer C44.

For larger injection volumes or to eliminate the solvent, hold the initial temperature long enough to vent the solvent through the split vent and then begin the first ramp. Use a fast rate for thermally stable analytes. Slower rates may help minimize thermal degradation in the inlet.

Split Carrier supply 80 PSI

Septum Purge

Valve

Frit

EPC Module

PS

Inlet Weldment

Column

Split Vent Trap

FS = Flow Sensor PS = Pressure Sensor

Valve Valve

Frit

PS FS

62 Advanced Operation Manual

3 Inlets

A single temperature ramp is enough for the injection process. The remaining ramps may be used to clean the liner or to reduce the inlet temperature in preparation for the next injection.

Hot split introduction

For hot split introduction, set an initial temperature high enough to volatilize the analytes. No additional thermal parameters are required as the inlet will maintain the setpoint throughout the run.

Setting parameters for split mode operation

Mode: The current operating modesplit or pulsed split

Temperature Actual and setpoint inlet temperatures.

Initial temperature Starting temperature for the inlet.

Initial time Hold time at the inlet initial temperature.

Rate # Temperature program rates for inlet thermal ramps.

Final temp # Final inlet temperature for ramp 1- 10.

Final time # Hold time at Final temp 1- 10.

Pressure Actual and setpoint inlet pressure. Controls capillary column flow and linear velocity.

Split ratio The ratio of split flow to column flow. Column flow is set in the column parameter list. This line does not appear if a column in the flow path is not defined.

Split flow Flow, in mL/min, from the split/purge vent. This line does not appear if a column in the flow path is not defined.

Total flow These are the actual and setpoint values of the total flow into the inlet, which is the sum of the split flow, column flow, and septum purge flow. When you change the total flow, the split ratio and split flow change while the column flow and pressure remain the same.

Septum Purge Flow through the septum purge vent.

Gas saver On to reduce split vent flow at Saver time.

Inlets 3

Advanced Operation Manual 63

Saver flow Reduced split vent flow, at least 15 mL/min.

Saver time Time when flow is reduced to save gas.

If all columns in the flow path are defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Split or Pulsed split.

3 Set the inlet temperature (Initial temperature) and any desired ramps.

4 If you want a specific split ratio, scroll to Split ratio and enter that number. The split flow will be calculated and set for you.

5 If you want a specific split flow, scroll to Split flow and enter that number. The split ratio will be calculated and displayed for you.

6 If you selected Pulsed split, enter values for Pulsed pressure (pressure at injection) and Pulse time (minutes after injection to return to normal pressure).

7 If desired, turn on Gas saver. Set the Saver time after the injection time.

8 Press [Prep Run] before manually injecting the sample if the Gas Saver is on (see Pre Run and Prep Run on page 40 for details.).

If a column in the flow path is not defined

1 Press [Front Inlet].

2 Set the inlet temperature (Initial temperature) and any desired ramps.

3 Set other parameters as described for a defined column.

4 Set Total flow into the inlet. Measure flows out of the split vent and septum purge vent using a flow meter.

5 Subtract the septum purge flow and split vent flow from Total flow to get column flow.

6 Calculate the split ratio (split vent flow/column flow). Adjust as needed

64 Advanced Operation Manual

3 Inlets

MMI splitless and pulsed splitless modes

In these modeswith or without a pressure pulsethe split vent valve is closed during injection and vaporization of the sample and stays so while the sample transfers to the column.

At a specified time after injection, the valve opens to sweep vapors left in the liner out the split vent. This avoids solvent tailing due to the large inlet volume and small column flow rate.

Stage 1. Sample injection

With the split vent valve closed, the sample and solvent transfer to the column.

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Carrier Supply 80 PSI

Septum Purge

Column

Inlets 3

Advanced Operation Manual 65

Stage 2. Solvent purging

After the sample has transferred to the column, the split vent valve opens to purge remaining solvent vapor from the inlet.

Split Carrier supply 80 PSI

Septum Purge

Valve

Frit

EPC Module

PS

Inlet Weldment

Column

Split Vent Trap

FS = Flow Sensor PS = Pressure Sensor

Valve Valve

Frit

PS FS

66 Advanced Operation Manual

3 Inlets

Timelines

This figure summarizes the flow, pressure, and temperature changes during a splitless mode analysis.

Split vent flow Purge flow

Saver flow

Inlet pressure

Inlet is pressure controlled

Post pressure

Inlet pressure

Inlet temperature

Final temp 1

Initial temperature

Prep Run

Start Run

Purge Time

Saver Time

Stop Run

Post Time

Post Time

Stop Run

Purge Time

Purge Time

Prep Run

Prep Run

Start Run

Start Run

Column mode = Constant flow

Inlets 3

Advanced Operation Manual 67

Cold splitless introduction

For cold splitless introduction, use an initial inlet temperature below the normal boiling point of the solvent. For most solvents, starting the first inlet temperature ramp at 0.1 minutes provides good transfer and reproducibility. A program rate of 500 C/min or higher is appropriate for thermally stable analytes. A final temperature of 350 C, held for 5 minutes, has quantitatively transferred up to C44 alkane.

A main advantage of temperature programmability is that the inlet can be heated gently to transfer delicate analytes. If the oven temperature is initially low enough to refocus the analytes on the column, the inlet heating rate can be made slower (e.g., 120 C/min). This reduces thermal degradation from the inlet and can improve peak shape and quantitation.

For most applications of cold splitless, a single temperature ramp is enough. The remaining ramps can be used to clean the liner or to decrease the inlet temperature to prepare for the next injection.

Hot splitless introduction

For hot splitless introduction, select an initial temperature high enough to volatilize the analytes. No additional temperature parameters are required as the inlet will maintain the setpoint throughout the run.

Starting values

A successful splitless injection consists of these steps:

1 Inject the sample and temperature program the inlet to vaporize it.

2 Use a low column flow and low oven temperature to create a solvent- saturated zone at the head of the column.

3 Use this zone to trap and reconcentrate the sample at the head of the column.

4 Wait until all, or at least most, of the sample has transferred to the column. Then discard the remaining vapor in the inletwhich is mostly solventby opening a purge valve. This eliminates the long solvent tail that this vapor would otherwise cause.

5 Raise the oven temperature to analyze the sample.

68 Advanced Operation Manual

3 Inlets

Some experimentation is needed to refine the operating conditions. Table 13 provides starting values for the critical parameters.

Setting parameters for the splitless modes

Mode: The current operating modeSplitless or Pulsed splitless.

Temperature Actual and setpoint inlet temperatures.

Initial time Hold time at the initial inlet temperature.

Rate # Temperature program rates for inlet thermal ramps.

Final temp # Final inlet temperature for ramp 1- 10.

Final time # Hold time at Final temp 1- 10.

Pressure Actual and setpoint inlet pressure in psi, bar, or kPa

Pulsed pres The inlet pressure you desire at the beginning of a run. The pressure rises to this setpoint after [Prep Run] is pressed and remains constant until Pulse time elapses, when it returns to Pressure.

Table 13 Splitless mode inlet parameters

Parameter Allowed setpoint range Suggested starting value

Oven temperature No cryo, ambient+4 C to 450 C CO2 cryo, 40 C to 450 C N2 cryo, 80 C to 450 C

10 C below solvent boiling point

Oven initial time 0 to 999.9 minutes Inlet purge time

Inlet purge time 0 to 200.0 minutes 2 x Liner volume Column flow

Gas saver time 0 to 999.9 minutes After purge time

Gas saver flow 15 to 200 mL/min 15 mL/min greater than maximum column flow

Inlet temperature No cryo, oven temp + 10 C CO2 cryo, 70 C to 450 C N2 cryo, 160 C to 450 C

10 C below solvent boiling point for 0.1min, then ramp up

Inlets 3

Advanced Operation Manual 69

Pulse time Pressure returns to its normal setpoint at this time.

Purge time The time, after the beginning of the run, when you want the split vent valve to open.

Purge flow The flow, in mL/min, from the split vent, at Purge time. You will not be able to specify this value if operating with your column not defined.

Total flow The Total flow line displays the actual flow to the inlet during a Pre- run (Pre- run light is on and not blinking) and during a run before purge time. You cannot enter a setpoint at these times. At all other times, Total flow will have both setpoint and actual values.

Septum purge Flow through the septum purge vent

Gas saver On to reduce split vent flow at Saver time

Server flow Reduced split vent flow, at least 15 mL/min

Server time Time when flow is reduced to save gas

If the column is defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Splitless or Pulsed splitless.

3 Set the inlet temperature and any desired ramps.

4 Enter a Purge time and a Purge flow.

5 If desired, turn Gas saver on. Make certain the time is set after the Purge time.

6 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample. This is automatic if an Agilent sampler is used.

If the column is not defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Splitless or Pulsed splitless.

3 Set the inlet temperature and any desired ramps.

4 Enter a Purge flow.

70 Advanced Operation Manual

3 Inlets

5 Enter the Purge time when you wish the split valve to open.

6 Set Total flow greater than the column flow plus the septum purge flow to guarantee adequate column flow.

7 Turn Gas saver on, if desired. Set the time after Purge time.

8 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample. This is automatic if an Agilent sampler is used.

MMI solvent vent mode

This mode is typically used for large volume injections. For single injection use a 50 to 500 L syringe and the ALS variable plunger speed to slowly inject the sample.

The sample is injected into a cold inlet. If conditions are properly chosen and the sample is suitable, analytes deposit in the inlet liner while the solvent evaporates and is swept out. Large or multiple injections can be used to concentrate sample in the inlet before transferring to the column for analysis.

Inlets 3

Advanced Operation Manual 71

Stage 1. Sample and vent

During sampling and venting, the split valve is open. The inlet is at Initial temperature, which is at or below the solvent boiling point.

Solvent vapors are swept out the vent, while sample deposits on the liner walls or packing.

Split Carrier supply 80 PSI

Septum Purge

Valve

Frit

EPC Module

PS

Inlet Weldment

Column

Split Vent Trap

FS = Flow Sensor PS = Pressure Sensor

Valve Valve

Frit

PS FS

72 Advanced Operation Manual

3 Inlets

Stage 2. Sample transfer

When solvent venting ends, the split valve vent closes and the inlet heats to Final temperature 1. The sample transfers to the capillary column during Purge time. (Purge flow to split vent in a data system).

Stage 3. Purge and cleanup

The split valve opens again and the system returns to the Stage 1 configuration but with different setpoints. The MMI is flushed. Additional ramp rates are available to thermally clean the inlet or to reduce inlet temperature after sample transfer. This can extend the life of the liner.

Temperature, pressure, and flow considerations

The solvent vent mode goes through three distinct pneumatic states; venting, sample transfer, and purging. The vent portion allows the inlet pressure and the vent flow to be adjusted to optimize solvent elimination. The transfer state

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Carrier Supply 80 PSI

Septum Purge

Column

Inlets 3

Advanced Operation Manual 73

mimics traditional splitless operation and transports the analytes from the liner to the column. The purging mode allows the user to prepare the inlet for the next run.

A fundamental difficulty with solvent vent mode is the potential loss of volatile analytes with the solvent. Several solutions are possible for this situation:

The inlet liner can be packed with a more retentive material, such as Tenax. This greatly improves volatile analyte recovery but may impact recovery of higher boiling materials.

Some of the solvent can be left in the liner when sample transfer begins. The residual solvent acts like a stationary phase and retains volatile material, but at the expense of a larger solvent peak.

The inlet temperature can be reduced. This reduces the vapor pressure of the volatile analytes and permits higher recoveries.

Solvent removal can be speeded up by:

Reducing pressure in the inlet during sample introductionthe Vent pressure parameter

Increasing flow through the inletthe Vent flow parameter

While all these possibilities do complicate use of the PTV, they provide increased flexibility and new potential to solve difficult problems.

Sequence of operations

These are the steps in a typical analysis using the solvent vent mode:

Table 14 The solvent vent process

Step Parameter Value

1 Before injection Flow at split vent Either Purge flow or Saver flow

Inlet pressure Derived from column setpoint

The system is resting, with Purge flow (or Saver flow, if on) through the inlet.

2 Prep Run begins Flow at split vent Vent flow setpoint

Inlet pressure Vent pressure setpoint

Setpoints change to prepare for injection. When GC is ready, the sample is injected. Inlet and oven temperature program Init times begin. Solvent venting and analyte trapping begin.

3 At Vent end time Flow at split vent None, split valve closed

74 Advanced Operation Manual

3 Inlets

Some important points

The flow through the column is governed by the pressure in the inlet. This is controlled, during the analysis part of the process, by the flow or pressure setpoint or program entered for the column.

The controlling times must be in the order shown; Vent end time before Purge time, and Purge time before Saver time.

Vent end time must occur before the inlet starts to heat and release analytes.

Purge time must occur before the oven begins to heat and move sample through the column.

Timelines

Time increases downward; all other quantities increase to the right.

Inlet pressure Column pressure setpoint

Solvent venting ends, analyte transfer begins as inlet heats up.

4 At Purge time Flow at split vent Purge flow setpoint

Inlet pressure Column pressure setpoint

Analyte transfer ends, inlet is purged of residual vapor. Analysis begins.

5 At Saver time Flow at split vent Saver flow setpoint

Inlet pressure Column pressure setpoint

Analysis ends, carrier flow reduced to save gas (if Saver is on).

Table 14 The solvent vent process (continued)

Step Parameter Value

Inlets 3

Advanced Operation Manual 75

When is Start Run?

Both the inlet and oven temperature programs begin at Start Run. All timessuch as Purge timeare measured from Start Run. When does Start Run occur?

If the sample is injected manually, Start Run occurs when the user presses the [Start] key.

If a single injection per run is made using an autosampler, Start Run occurs when the syringe carrier moves down to make the injection.

If multiple injections per run are made using an autosampler, Start Run occurs when the syringe carrier moves down to make the first injection of the set. There are no Start Run signals for the rest of the injections in the set.

Time Oven temperature Inlet temperature Inlet pressure Split vent flow

Between runs

Prep Run

Start Run

Vent end time

Purge time

Saver time

Rate 1

Initial time

Final temperature 1 Final time 1

Rate 1

Final temperature 1 Final time 1

Other rates, temperatures, and times, if desired

Initial time

Other rates, temperatures, and times, if desired

Saver or Purge flow

Vent flow

Inlet is pressure controlled

Purge flow

Saver flow, if on

Controlled by column flow or pressure setpoint or program

Vent pressure

Controlled by column flow or pressure setpoint or program

76 Advanced Operation Manual

3 Inlets

These additional injections take time. The inlet and oven temperature programs, mainly the Initial time values, must be adjusted to allow for this. So must the various time values that control the inlet operation. This is discussed in more detail under To develop a MMI method that uses large volume injection on page 79.

Setting parameters for solvent vent operation

Mode: The current operating modesolvent vent.

Temperature Actual and setpoint initial inlet temperatures.

Initial time The time, measured from Start Run, when the initial inlet temperature hold ends. Must be greater than Vent end time.

Rate # Temperature program rate for inlet thermal ramps.

Final temperature # Final inlet temperature for ramp 1- 10.

Final time # Hold time at Final temp #. This time is a duration; it is not measured from Start Run.

Pressure Actual and setpoint inlet pressure before and after the vent period. It sets the starting point of column head pressure.

Vent pressure The inlet pressure during the vent period. By decreasing the inlet pressure while venting, solvent elimination proceeds faster. Also, the pressure reduction decreases the amount of carrier gasand solvent vaporthat enters the column during this time.

Users select from 0 to 100 psig. If 0 is chosen, the inlet uses the lowest pressure possible at the given vent flow. Table 15 shows approximate values for this minimum at various vent flows of helium. Pressures less than those in the table are not possible unless the flow is reduced.

Inlets 3

Advanced Operation Manual 77

Vent flow The flow of carrier gas out the split vent during the vent period. Higher flows sweep the liner more quickly and reduce the time for solvent elimination. For most columns, 100 mL/min vent flow eliminates solvent at an acceptable rate but puts minimal material on the column.

Vent end time The time, measured from Start Run, when solvent venting ends. For large volume injections, this time is normally greater than the time for the injection to complete.

Purge time The time, measured from Start Run, when sample transfer ends. It began at Vent end time.

Purge flow The flow of carrier gas to the inlet beginning at Purge time.

Total flow The actual flow into the inlet.

Septum Purge Flow through the septum purge vent

Gas saver On to reduce split vent flow at Saver time

Saver flow Reduced split vent flow, at least 15 mL/min

Saver time Time when flow is reduced to save gas

If the column is defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Solvent vent.

3 Enter a Vent pressure, a Vent flow, and a Vent end time.

4 Set the inlet temperature and ramps, as desired.

5 Enter a Purge time and a Purge flow.

Table 15 Minimum attainable pressures

Vent flow (mL/min) Actual vent pressure at 0 psig setpoint

Actual vent pressure at 0 kPa setpoint

50 0.7 5

100 1.3 10

200 2.6 18

500 6.4 44

1000 12.7 88

78 Advanced Operation Manual

3 Inlets

6 If desired, turn Gas saver on. Make certain the time is set after the Purge time.

7 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample.

If the column is not defined

1 Set up the parameters as described for the defined column case.

2 Set Total flow greater than the column flow plus the septum purge flow to guarantee adequate column flow.

MMI Direct Mode

MMI Direct Mode is a pneumatic configuration that allows on- column like behavior. In this mode, you still use a liner to trap involatile material but the sample can only enter the column.

Direct mode works best with direct connect liners. These liners form a seal with the column so that the sample cannot leak out of the liner. The inlet is held at a temperature below the solvent boiling point, just as in cool on- column (COC) during the injection. Pneumatically, the inlet is in forward pressure regulation, similar to the splitless portion of splitless mode. The inlet is then temperature programmed to transfer any remaining volatile material to the column.

The only control parameter for Direct Mode is inlet pressure. You can run the column in any of its normal control modes (constant pressure, constant flow, ramped pressure, ramped flow). Gas saver does not work in this mode as there is no purge state.

The next figure shows the flows with direct mode.

Inlets 3

Advanced Operation Manual 79

To develop a MMI method that uses large volume injection

This topic provides a recommended way to change from a splitless injection using a split/splitless inlet to a solvent vent mode injection using a Multimode inlet (MMI). It applies mainly to large volume injections (LVI) using a Multimode inlet, but the concepts can apply to general MMI use. This topic does not consider all items that can impact an analysis, for example the liner, solvent, analyte boiling points, or polarity. This topic also assumes knowledge of your data systemwhen to save a method, how to start a run, how to set up a single run, etc.

Split Carrier supply 80 PSI

Septum Purge

Valve

Frit

EPC Module

PS

Inlet Weldment

Column

Split Vent Trap

FS = Flow Sensor PS = Pressure Sensor

Valve Valve

Frit

PS FS

80 Advanced Operation Manual

3 Inlets

The main advantage of the MMI's solvent vent mode is that you can inject slowly into the inlet, allowing large amounts of solvent to evaporate in the liner (not in the split vent line), concentrating the analytes prior to injection. This requires an injector with variable speed injections, a "large" syringe, and knowledge of the sample and the solvent. When developing a solvent vent method, the goal for the injection is to determine the injection rates and temperatures needed to evaporate the solvent at the rate it enters the inlet. The development technique is to gradually scale up to an injection amount that produces a useful response. The most significant parameters are:

Inlet temperature. Hold the inlet temperature at or slightly below the solvent boiling point until after all the sample has been injected. This is important so that you do not boil away more volatile analytes, or boil away the solvent in the needle and trap your analytes there. An additional point to consider is that the boiling point of the earliest eluting analyte should be 100 C > the boiling point of the solvent.

Injection speed. Estimate the evaporation rate of solvent exiting the needle based on solvent type, inlet temperature, vent flow, and pressure. Start with about half that number. Note that you have to make sure that you configure the syringe properly. If not, you will over- or under- load the inlet.

Vent time. Make sure the vent time setpoint is greater than the time the needle spends in the inlet. If the vent time is too short, you will overload and contaminate the column and inlet. Change the vent time as you upscale the method.

If using a MMI with an MSD, another tip for method development is to scan for solvent ions. Detecting the solvent ions can be useful in troubleshooting residual solvent bleed onto the column.

To develop a MMI method for large volume injection, try the following:

1 Determine a small injection volume that works on a split/splitless inlet in splitless mode. Choose a volume that does not currently overload the inlet.

2 Start with a 510 uL syringe and make sure the syringe is properly configured in the instrument and data system.

3 Make sure the column is configured.

Inlets 3

Advanced Operation Manual 81

4 Set up the inlet to perform a 1 uL injection. Use the split/splitless inlet method conditions, except:

Start with the inlet temperature cold, near but slightly below the solvent boiling point. For example, if using methylene chloride (boiling point 39 C), start with a temperature of 3039 C.

Use Splitless mode.

Ramp to the normal split/splitless inlet temperature

5 Note the response achieved.

6 Next, change the inlet mode to Solvent Vent.

7 Check the injector timings.

a Install an empty sample vial in the injector turret or tray.

b Input the injection rate (Draw, Dispense, and Inject rates) for the injector and increase the injection volume to 5 uL.

c Enter draft Solvent Vent parameters:

Set a Vent Flow of 100 mL/min as a starting point.

Keep the inlet isothermal for now

Enter a Vent Pressure of 0 psi (0 kPa) until 0.1 min.

Make an injection using the empty vial. Use a stopwatch or your GC's timer feature to time how long the needle is in the inlet.

8 Enter revised Solvent Vent mode parameters.

Set the method's Vent Time to be about 0.05 min longer than the time the needle spends in the inlet.

Set the inlet temperature initial Hold Time to be about 0.05 min longer than the vent pressure until time.

Program the inlet to ramp quickly to the injection temperature. Make sure the ramp starts after the vent pressure until time.

Set the Purge Flow to Split Vent to 30 mL/min. Set the purge time to be the vent pressure until time + 1 minute.

9 Make a 5 uL injection of your standard. You should see 5 times the response.

If the response of all analytes is too low:

The dispense speed is too fast. Liquid was injected into the inlet and pushed out the vent.

82 Advanced Operation Manual

3 Inlets

The vent time is too long. The inlet started to heat while the vent was open.

If the response of early eluters is too low:

The inlet temperature is too high.

The Vent Flow is too high.

If the response of late eluters is too low:

The purge time is too short.

The final inlet temperature is too low.

10 If you need more injection volume and the 5 uL worked to give 5X response, change the syringe to a larger one, for example 50 uL.

11 Set up the data system to perform a 25 uL injection.

12 Configure the syringe. Make sure the plunger speeds on the injector are still set properly.

13 Recheck the injector timings. See step 12.

14 Set a new vent time and initial inlet temperature time based on the new time the needle spends in the syringe.

15 Make a 25 uL injection of your standard. Again, you should see 5 times the response. If not, see step 13.

16 If you need more response, repeat steps 10 15 to increase to larger volume. Also try performing 5 x 5 uL injections, then 5 x 50 uL injections.

Multiple injections with the MMI

The preferred technique for concentrating analytes in the inlet liner is to use a single, large volume injection. Using a high capacity syringe and one septum puncture reduces the possibility of contamination and generally improves results when compared against a multiple septum puncture technique. However, if needed, you can perform multiple septum punctures during the vent time. This technique requires an Agilent data system and automatic liquid sampler.

Data system requirements

An Agilent data system is necessary for multiple injection because the needed parameters are not available through the GC keyboard.

GC ChemStation Software revision B.04.01 SP1 or later.

Inlets 3

Advanced Operation Manual 83

MSD ChemStation Software revision E.02.00 SP2 or later

EZChrom Software revision 3.3.2 or later

Setting parameters for the inlet in solvent vent mode

Set or configure the following parameters in the data system's 7890A GC method editor.

Syringe size Verify the syringe size is configured correctly. The configured syringe size changes the available choices for injection volume.

Injection volume Select the injection volume, then enter a number of injections. The total injection volume will be displayed.

Multiple Injection Delay A pause time, in seconds, between injections. This is added to the minimum hardware cycle time.

Preinjection washes and pumps are performed only before the first injection of a multiple injection set.

Postinjection washes are performed only after the last injection in a multiple injection set.

An example

These values were used for a sample with a broad range of boiling points.

Table 16 General parameters

Name Value

Sample C10 to C44 hydrocarbons in hexane

Mode Solvent vent

MMI liner Glass wool packed

Injection volume One 10.0 L injection (25 L syringe)

Injection speed Fast

Column 30 m x 320 m x 0.25 m -5, part number 19091J-413

Column flow 4 mL/min constant flow

84 Advanced Operation Manual

3 Inlets

Table 17 Inlet parameters

Name Value Name Value

Initial temp 40 C Rate 2 (off)

Initial time 0.3 min Pressure 15.6 psig

Rate 1 720 C/min Vent pressure 0.0 psig

Final temp 1 450 C Vent flow 100 mL/min

Final time 1 5 min Vent end time 0.2 min

Rate 2 100 C/min Purge time 2.0 min

Final temp 2 250 C Purge flow 50 mL/min

Final time 2 0 min

Table 18 Oven parameters

Name Value

Initial temp 40 C

Initial time 2.5 min

Rate 1 25 C/min

Final temp 1 320 C

Final time 1 10.0 min

Rate 2 (off)

Table 19 Detector parameters

Name Value

Detector FID

Detector temp 400 C

Hydrogen flow 40 mL/min

Air flow 450 mL/min

Makeup (N2) 45 mL/min

Inlets 3

Advanced Operation Manual 85

These results were compared with a splitless analysis of the same sample, which should produce 100% recovery of all analytes. The data showed that, under these conditions, compounds above C20 were completely recovered and that the recovery was independent of injection size. Compounds lower than C20 were partially vented with the solvent.

Possible adjustments

Depending on what you are trying to accomplish, you have a number of possible adjustments available.

To eliminate more solvent

Increase the vent end time, inlet initial time, and purge time. This will not affect analytes that are quantitatively trapped but will eliminate more of the solvent peak.

Increase the vent flow to sweep the liner more rapidly with the same inlet timing. Increasing vent flow raises vent pressure if it is set to 0. This puts more solvent onto the column.

C20

86 Advanced Operation Manual

3 Inlets

Raise the inlet initial temperature to vaporize more solvent and allow more to be eliminated. This also increases the loss of volatile analytes since their vapor pressures also increase.

To improve recovery of low boiling analytes

Reduce inlet temperature to lower the vapor pressure of the analytes and trap them more effectively. This also reduces solvent vapor pressure and more time will be needed to eliminate it.

Use a retentive packing in the liner. Materials such as Tenax permit higher recovery of volatile analytes but may not release higher boiling compounds. This must be considered if quantitation on these high boiling peaks is desired.

Leave more solvent in the liner. The solvent acts as a pseudo stationary phase and helps retain volatile analytes. This must be balanced against the detectors tolerance for solvent.

An examplecontinued

The single injection example shown on the last few pages makes it clear that a 10 L injection does not overload the glass wool packed liner. This means that multiple 10 L injections are possible.

It was decided to make 10 injections per run, each of 10 L size. This would increase analytical sensitivity substantially. No adjustments were made to improve recovery of the low boilers since the purpose of this analysis was to detect and measure the high boiling components.

After timing a trial set of 10 injections, the total time for the multiple injection set was measured to be approximately 1.3 minutes. The following timing changes were made:

Table 20 Modifications

Parameter Increased from To

Inlet Init time 0.3 minutes 1.6 minutes

Vent end time 0.2 minutes 1.5 minutes

Purge time 2.0 minutes 3.0 minutes

Oven Init time 2.5 minutes 3.0 minutes

Inlets 3

Advanced Operation Manual 87

The result is shown in the next figure. Note the difference in the vertical scale (5000 versus 500).

C20

88 Advanced Operation Manual

3 Inlets

About the Packed Column Inlet

This inlet is also known as the purged packed inlet (PP). It is used with packed columns when high- efficiency separations are not required. It can also be used with wide- bore capillary columns, if flows greater than 10 mL/min are acceptable.

If the columns are not defined (packed columns and undefined capillary columns), the inlet is usually flow- controlled. If capillary columns are used and the columns in the flow path are defined, the inlet is normally pressure- controlled but can be put into the flow control mode.

The figure shows the inlet in the capillary column mode, with the column defined and the flow controlled by pressure.

Inlets 3

Advanced Operation Manual 89

The next figure shows the flow diagram for the packed column mode, with the column not defined the control is based on total carrier gas flow.

Valve

Frit Frit

Valve

EPC Module

Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

Carrier Supply 80 PSI

Septum Purge

Column

PS

90 Advanced Operation Manual

3 Inlets

Setting parameters

The inlet can operate in flow or pressure control mode. Flow is recommended for packed columns. Pressure is recommended for capillary columns.

Inlet in flow control mode

While in flow control mode, you cannot enter pressures here.

Mode Press [Mode/Type] to see the choices. You may select an inlet mode of either Pressure control or Flow control. When the Inlet control mode is Pressure control, the column control mode can be set to Constant pressure, Ramped pressure, Constant

Valve

Frit Frit

Valve

EPC Module

Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

Carrier Supply 80 PSI

Septum Purge

Packed Column

PS

Inlets 3

Advanced Operation Manual 91

flow, or Ramped flow. The column control mode is set from the Column parameters display. When the Inlet control mode is Flow control, only column flow can be set on the Column parameters display.

Temperature The setpoint and actual temperature values.

Pressure The actual pressure (in psi, bar, or kPa) supplied to the inlet. You cannot enter a setpoint here.

Total flow Enter your setpoint here, actual value is displayed. Inlet is mass flow controlled.

Septum purge The setpoint and actual flow, in mL/min, through the septum purge line, typically 1 to 5 mL/min

Inlet in pressure control mode

While in pressure control mode, you cannot enter flows here.

Mode Press [Mode/Type] to see the choices. You may select an inlet mode of either Pressure control or Flow control. When the Inlet control mode is Pressure control, the column control mode can be set to Constant pressure, Ramped pressure, Constant flow, or Ramped flow. The column control mode is set from the Column parameters display. When the Inlet control mode is Flow control, only column flow can be set on the Column parameters display.

Temperature The setpoint and actual temperature values of the inlet.

Pressure Enter your setpoint here (in psi, bar, or kPa). The actual value is displayed.

Total flow The actual total flow to the inlet. This is a reported value, not a setpoint.

Septum purge The setpoint and actual flow, in mL/min, through the septum purge line, typically 1 to 5 mL/min

92 Advanced Operation Manual

3 Inlets

About the Cool On-Column Inlet

This inlet introduces liquid sample directly onto a capillary column. To do this, both the inlet and the oven must be cool at injection, either at or below the boiling point of the solvent.

Because the sample does not vaporize immediately in the inlet, problems with sample discrimination and sample alteration are minimized. If done properly, cool- on column injection also provides accurate and precise results.

You can operate the inlet in track oven mode, where the inlet temperature follows the column oven, or you can program up to three temperature ramps. A cryogenic cooling option that uses liquid CO2 or N2 from the oven cryogenic system can reach sub- ambient temperatures.

Valve

Frit Frit

Valve

EPC Module

Cool on Column Inlet Weldment

PS = Pressure Sensor

PS

PS

Carrier Supply 80 PSI

Septum Purge

Column

Septum

Column AdapterFerrule

Column Nut

Column

Inlets 3

Advanced Operation Manual 93

Setup modes of the COC inlet

The COC inlet hardware must be set up for one of three usages, depending on the type of injection and column size.

0.25 mm or 0.32 mm automated on- column. Use predrilled septa.

0.53 mm automatic on- column or retention gap

0.2 mm manual

To select the correct hardware for a column and injection type, refer to Maintaining Your GC.

Retention gaps

Because the sample is injected directly onto the column, it is strongly suggested that a retention gapor guard columnbe used to protect your column. A retention gap is a deactivated column that is connected between the inlet and the analytical column. If you choose to use one, it is suggested that at least 1 m of retention gap be installed per 1 L of sample injected. Information on ordering retention gaps can be found in the Agilent catalog for consumables and supplies.

If you are using a retention gap and are operating with the column defined, the length of the retention gap could affect the calculations for flow and velocity through your column. If your retention gap is the same inside diameter as your column, it is a good idea to add the retention gap and column length before entering the number on the Configure Column parameter list. If the retention gap inside diameter is larger than your column, this step may not be necessary.

COC inlet temperature control

CryoBlast (optional)

CryoBlast shortens the cycle time between runs. If you have a CO2 or N2 cryogenic valve and the CryoBlast feature, you can cool the inlet to 37 C in either the track oven or temperature program modes.

The CryoBlast accessory uses coolant from the oven cryogenic system.

94 Advanced Operation Manual

3 Inlets

Track oven mode

In the Track oven mode, the inlet temperature stays 3 C higher than the oven temperature throughout the oven program. You cannot enter a temperature setpointit is set automatically. If you have CryoBlast, the inlet will track oven temperatures to 40C; without CryoBlast, the lower limit is set by room temperature.

Temperature programming mode

In this mode, you can enter up to three temperature ramps in the inlet parameter list so that the inlet and the oven operate independently. This is the recommended mode if operating below 20 C.

At these very low oven temperatures, the inlet temperature should be at least 20 C higher than the oven temperature. This will be more than adequate for solvent focusing.

At temperatures greater than ambient, the inlet should always be at least 3 C warmer than the oven for proper control of the inlet temperature.

The oven temperature program controls the run. If it is longer than the inlet temperature program, the inlet will remain at its final temperature until the oven program (and the run) ends.

Cryogenic considerations

When using track oven mode with a cryogenic oven, all other inlets must be off or in track oven mode.

If cryo is turned on, and if the inlet is cooled during a run, cryogen is used to achieve the setpoint. This may have a negative impact on the chromatographic performance of the oven and cause distorted peaks.

The inlet uses the same cryo coolant as configured for the oven.

Setting COC inlet flows/pressures

1 Configure capillary column to inletselect constant flow or pressure.

2 Set column flow, linear velocity, or inlet pressure.

3 Set Septum Purge, typically 3 to 10 mL/min.

Inlets 3

Advanced Operation Manual 95

Setting COC inlet parameters

Track oven mode

1 Press [Front Inlet] or [Back Inlet].

2 Press [Mode/Type] and select Track oven.

There is no setpoint for Track oven mode.

Ramped temperature mode

1 Press [Front Inlet] or [Back Inlet].

2 Press [Mode/Type] and select Ramped temp.

3 Enter a value for Temp. This is the starting temperature.

4 Enter an Init time. This is the length of time the inlet will stay at the starting temperature after a run has begun.

5 Enter Rate 1. This is the rate at which the inlet will be heated or cooled. A Rate of 0 halts further programming.

6 Enter Final temp 1. This is the inlet temperature at the end of the first ramp.

7 Enter Final time 1. This is the number of minutes the inlet holds Final temp 1.

8 To enter a second (or third) ramp, scroll to the appropriate Rate line and repeat steps 5 through 7.

96 Advanced Operation Manual

3 Inlets

About the PTV Inlet

The Agilent Programmed Temperature Vaporization (PTV) Inlet System has five operating modes:

The split mode is generally used for major component analyses.

The pulsed split mode is like the split mode, but with a pressure pulse applied to the inlet during sample introduction to speed the transfer of material to the column.

The splitless mode is used for trace analyses.

The pulsed splitless mode allows for a pressure pulse during sample introduction.

The solvent vent mode is used for large volume injection. Either single or multiple injections can be made for each run.

The PTV inlet can be used with both manual and automatic injection.

For automatic multiple injections (large volume injections), an Agilent GC or MSD ChemStation is required. This function is not available under GC control alone. See PTV inlet solvent vent mode on page 109.

PTV sampling heads

Two heads are available for the PTV inlet.

The septum head uses either a regular septum or a Merlin Microseal to seal the syringe passage. A stream of gas sweeps the inner side of the septum and exits through the septum purge vent on the pneumatics module. It may be used with either automatic or manual injection.

The septum head uses either standard 11 mm septa or (with a different cap) a Merlin Microseal.

The septumless head uses a check valve instead of a septum to seal the syringe entrance passage. It may be used with either automatic or manual injection. This head is recommended for subambient inlet operation.

Inlets 3

Advanced Operation Manual 97

The flow diagrams in the rest of this document show the septum head in place with a separate drawing for the septumless head.

Heating the PTV inlet

The control parameters for PTV temperature programming are the same as for the column oven, but are reached by pressing [Front Inlet]. Temperature can be programmed with an initial temperature and up to 3 rates and plateaus. Rates between 0.1 and 720 C/min can be selected. See Setting COC inlet parameters on page 95 for details.

At the end of the run and during post- run, the PTV is held at its final temperature. This permits backflushing without contaminating the inlet.

Additional temperature ramps

For most purposes, the PTV is designed to hold the sample in the inlet liner until the entire samplethere could be several injectionshas been injected. Then the PTV is heated rapidly to transfer the sample to the column. This can be accomplished with an initial hold, a single ramp, and a hold at the end to complete sample transfer.

Two additional ramps are available and have several possible uses:

The inlet can be heated to a high temperature to thermally clean the liner for the next run.

Septum head Septumless head

CAUTION If the initial inlet temperature and the oven initial temperature are too close, the inlet may be unable to maintain its setpoint. We recommend a difference of at least 6 C, either higher or lower.

98 Advanced Operation Manual

3 Inlets

The inlet can be programmed downwardjust set the Final temp below the previous temperatureto reduce thermal stress on the inlet.

Downward programming can be used to prepare the inlet for the next run. This can reduce cycle time for greater sample throughput.

Cooling the PTV inlet

If cryo is turned on, and if the inlet is cooled during a run, cryogen is used to achieve the setpoint. This may have a negative impact on the chromatographic performance of the oven and cause distorted peaks.

The sample may be injected into either a cooled or heated inlet. The initial inlet temperature can be reduced to 60 C (with CO2 cooling) or to 160 C (with liquid N2 cooling).

The inlet uses the same coolant as configured for the oven.

PTV inlet split and pulsed split modes

The two split modeswith or without a pressure pulsedivide the gas stream entering the inlet between the column flow, the split vent flow through the solenoid valve, and the septum purge flow. The ratio of the split vent flow to the column flow is called the split ratio.

The next figure shows the flows with the septum head. Flows with the septumless head are the same except that the septum purge flow bypasses the head.

Inlets 3

Advanced Operation Manual 99

Cold split introduction

For cold split sample introduction, use an initial inlet temperature below the normal boiling point of the solvent. If the liner volume is enough to hold all the vaporized solvent, start the first inlet temperature ramp at 0.1 minutes with a high heating rate (500 C/min or higher). The final temperature should be high enough to volatilize the heaviest analytes from the liner and should be held for at least 5 minutes. A final temperature of 350 C for 5 minutes has proven sufficient to quantitatively transfer C44.

For larger injection volumes or to eliminate the solvent, hold the initial temperature long enough to vent the solvent through the split vent and then begin the first ramp. Use a fast rate for thermally stable analytes. Slower rates may help minimize thermal degradation in the inlet.

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

PTV Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Column

100 Advanced Operation Manual

3 Inlets

A single temperature ramp is enough for the injection process. The remaining ramps may be used to clean the liner or to reduce the inlet temperature in preparation for the next injection.

Hot split introduction

For hot split introduction, set an initial temperature high enough to volatilize the analytes. No additional thermal parameters are required as the inlet will maintain the setpoint throughout the run.

Because of the small liner volume (about 120 microliters), the PTV has a limited injection capacity with hot split introduction. Injection volumes exceeding 1 L in the hot split mode may overflow the inlet causing analytical problems. Cold split introduction avoids this potential problem.

Setting parameters for split mode operation

Mode: The current operating modesplit or pulsed split

Temperature Actual and setpoint inlet temperatures.

Initial temperature Starting temperature for the inlet.

Initial time Hold time at the inlet initial temperature.

Rate # Temperature program rates for inlet thermal ramps.

Final temp # Final inlet temperature for ramps 1, 2, and 3.

Final time # Hold time at Final temp 1, 2, and 3.

Pressure Actual and setpoint inlet pressure. Controls capillary column flow and linear velocity.

Split ratio The ratio of split flow to column flow. Column flow is set in the column parameter list. This line does not appear if a column in the flow path is not defined.

Split flow Flow, in mL/min, from the split/purge vent. This line does not appear if a column in the flow path is not defined.

Inlets 3

Advanced Operation Manual 101

Total flow These are the actual and setpoint values of the total flow into the inlet, which is the sum of the split flow, column flow, and septum purge flow. When you change the total flow, the split ratio and split flow change while the column flow and pressure remain the same.

Septum Purge Flow through the septum purge vent.

Gas saver On to reduce split vent flow at Saver time.

Saver flow Reduced split vent flow, at least 15 mL/min.

Saver time Time when flow is reduced to save gas.

If all columns in the flow path are defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Split or Pulsed split.

3 Set the inlet temperature (Initial temperature) and any desired ramps.

4 If you want a specific split ratio, scroll to Split ratio and enter that number. The split flow will be calculated and set for you.

5 If you want a specific split flow, scroll to Split flow and enter that number. The split ratio will be calculated and displayed for you.

6 If you selected Pulsed split, enter values for Pulsed pressure (pressure at injection) and Pulse time (minutes after injection to return to normal pressure).

7 If desired, turn on Gas saver. Set the Saver time after the injection time.

8 Press [Prep Run] before manually injecting the sample if the Gas Saver is on (see Pre Run and Prep Run on page 40).

If a column in the flow path is not defined

1 Press [Front Inlet].

2 Set the inlet temperature (Initial temperature) and any desired ramps.

3 Set other parameters as described for a defined column.

4 Set Total flow into the inlet. Measure flows out of the split vent and septum purge vent using a flow meter.

102 Advanced Operation Manual

3 Inlets

5 Subtract the septum purge flow and split vent flow from Total flow to get column flow.

6 Calculate the split ratio (split vent flow/column flow). Adjust as needed

PTV inlet splitless and pulsed splitless modes

In these modeswith or without a pressure pulsethe split vent valve is closed during injection and vaporization of the sample and stays so while the sample transfers to the column.

At a specified time after injection, the valve opens to sweep vapors left in the liner out the split vent. This avoids solvent tailing due to the large inlet volume and small column flow rate.

The figures show the flows with the septum head. Flows with the septumless head are the same except that the septum purge flow bypasses the head.

Inlets 3

Advanced Operation Manual 103

Stage 1. Sample injection

With the split vent valve closed, the sample and solvent transfer to the column.

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

PTV Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Column

104 Advanced Operation Manual

3 Inlets

Stage 2. Solvent purging

After the sample has transferred to the column, the split vent valve opens to purge remaining solvent vapor from the inlet.

Timelines

This figure summarizes the flow, pressure, and temperature changes during a splitless mode analysis.

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

PTV Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Column

Inlets 3

Advanced Operation Manual 105

Cold splitless introduction

For cold splitless introduction, use an initial inlet temperature below the normal boiling point of the solvent. For most solvents, starting the first inlet temperature ramp at 0.1 minutes provides good transfer and reproducibility. A program rate of 500 C/min or higher is appropriate for thermally stable analytes. A final temperature of 350 C, held for 5 minutes, has quantitatively transferred up to C44 alkane.

A main advantage of temperature programmability is that the inlet can be heated gently to transfer delicate analytes. If the oven temperature is initially low enough to refocus the analytes on the column, the inlet heating rate can be made slower (e.g., 120 C/min). This reduces thermal degradation from the inlet and can improve peak shape and quantitation.

Split vent flow Purge flow

Saver flow

Inlet pressure

Inlet is pressure controlled

Post pressure

Inlet pressure

Inlet temperature

Final temp 1

Initial temperature

Prep Run

Start Run

Purge Time

Saver Time

Stop Run

Post Time

Post Time

Stop Run

Purge Time

Purge Time

Prep Run

Prep Run

Start Run

Start Run

Column mode = Constant flow

106 Advanced Operation Manual

3 Inlets

For most applications of cold splitless, a single temperature ramp is enough. The remaining ramps can be used to clean the liner or to decrease the inlet temperature to prepare for the next injection.

Hot splitless introduction

For hot splitless introduction, select an initial temperature high enough to volatilize the analytes. No additional temperature parameters are required as the inlet will maintain the setpoint throughout the run.

Because of the small liner volume (about 120 L), the PTV cannot contain vapor resulting from large liquid injection volumes. Injection volumes greater than 1 L may overflow vapor from the inlet, causing analysis variations. Cold splitless introduction avoids this problem.

Starting values

A successful splitless injection consists of these steps:

1 Inject the sample and temperature program the inlet to vaporize it.

2 Use a low column flow and low oven temperature to create a solvent- saturated zone at the head of the column.

3 Use this zone to trap and reconcentrate the sample at the head of the column.

4 Wait until all, or at least most, of the sample has transferred to the column. Then discard the remaining vapor in the inletwhich is mostly solventby opening a purge valve. This eliminates the long solvent tail that this vapor would otherwise cause.

5 Raise the oven temperature to analyze the sample.

Some experimentation is needed to refine the operating conditions. Table 21 provides starting values for the critical parameters.

Table 21 Splitless mode inlet parameters

Parameter Allowed setpoint range Suggested starting value

Oven temperature No cryo, ambient+10 C to 450 C CO2 cryo, 60 C to 450 C N2 cryo, 80 C to 450 C

10 C below solvent boiling point

Inlets 3

Advanced Operation Manual 107

Setting parameters for the splitless modes

Mode: The current operating modeSplitless or Pulsed splitless.

Temperature Actual and setpoint inlet temperatures.

Initial time Hold time at the initial inlet temperature.

Rate # Temperature program rates for inlet thermal ramps.

Final temp # Final inlet temperature for ramps 1, 2, and 3.

Final time # Hold time at Final temp 1, 2, and 3.

Pressure Actual and setpoint inlet pressure in psi, bar, or kPa

Pulsed pres The inlet pressure you desire at the beginning of a run. The pressure rises to this setpoint after [Prep Run] is pressed and remains constant until Pulse time elapses, when it returns to Pressure.

Pulse time Pressure returns to its normal setpoint at this time.

Purge time The time, after the beginning of the run, when you want the split vent valve to open.

Oven initial time 0 to 999.9 minutes Inlet purge time

Inlet purge time 0 to 999.9 minutes 2 x Liner volume Column flow

Gas saver time 0 to 999.9 minutes After purge time

Gas saver flow 15 to 1000 mL/min 15 mL/min greater than maximum column flow

Inlet temperature No cryo, oven temp + 10 C CO2 cryo, 50 C to 450 C N2 cryo, 60 C to 450 C

10 C below solvent boiling point for 0.1min, then ramp up

Table 21 Splitless mode inlet parameters (continued)

Parameter Allowed setpoint range Suggested starting value

108 Advanced Operation Manual

3 Inlets

Purge flow The flow, in mL/min, from the split vent, at Purge time. You will not be able to specify this value if operating with your column not defined.

Total flow The Total flow line displays the actual flow to the inlet during a Pre- run (Pre- run light is on and not blinking) and during a run before purge time. You cannot enter a setpoint at these times. At all other times, Total flow will have both setpoint and actual values.

Septum purge Flow through the septum purge vent

Gas saver On to reduce split vent flow at Saver time

Server flow Reduced split vent flow, at least 15 mL/min

Server time Time when flow is reduced to save gas

If the column is defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Splitless or Pulsed splitless.

3 Set the inlet temperature and any desired ramps.

4 Enter a Purge time and a Purge flow.

5 If desired, turn Gas saver on. Make certain the time is set after the Purge time.

6 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample. This is automatic if an Agilent sampler is used.

If the column is not defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Splitless or Pulsed splitless.

3 Set the inlet temperature and any desired ramps.

4 Enter a Purge flow.

5 Enter the Purge time when you wish the split valve to open.

6 Set Total flow greater than the column flow plus the septum purge flow to guarantee adequate column flow.

7 Turn Gas saver on, if desired. Set the time after Purge time.

Inlets 3

Advanced Operation Manual 109

8 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample. This is automatic if an Agilent sampler is used.

PTV inlet solvent vent mode

This mode is typically used for large volume injections. For single injection use a 50 or 100 L syringe with variable plunger speedslowly, 5 to 30 seconds.

The sample is injected into a cold inlet. If conditions are properly chosen and the sample is suitable, analytes deposit in the inlet liner while the solvent evaporates and is swept out. Large or multiple injections can be used to concentrate sample in the inlet before transferring to the column for analysis.

The figure shows the flows with the septum head. Flows with the septumless head are the same except that the septum purge flow bypasses the head.

110 Advanced Operation Manual

3 Inlets

Stage 1. Sample and vent

During sampling and venting, the split valve is open. The inlet is at Initial temperature, which is at or below the solvent boiling point.

Solvent vapors are swept out the vent, while sample deposits on the liner walls or packing.

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

PTV Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Column

Inlets 3

Advanced Operation Manual 111

Stage 2. Sample transfer

When solvent venting ends, the split valve vent closes and the inlet heats to Final temperature 1. The sample transfers to the capillary column during Purge delay time.

Stage 3. Purge and cleanup

The split valve opens again and the system returns to the Stage 1 configuration but with different setpoints. The PTV inlet is flushed. Additional ramp rates are available to thermally clean the inlet or to reduce inlet temperature after sample transfer. This can extend the life of the liner.

Temperature, pressure, and flow considerations

The solvent vent mode goes through three distinct pneumatic states; venting, sample transfer, and purging. The vent portion allows the inlet pressure and the vent flow to be adjusted to optimize solvent elimination. The transfer state

Valve

Frit Frit

Valve

EPC Module

Valve

Split Vent Trap

PTV Inlet Weldment

FS = Flow Sensor PS = Pressure Sensor

FS PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Column

112 Advanced Operation Manual

3 Inlets

mimics traditional splitless operation and transports the analytes from the liner to the column. The purging mode allows the user to prepare the inlet for the next run.

A fundamental difficulty with solvent vent mode is the potential loss of volatile analytes with the solvent. Several solutions are possible for this situation:

The inlet liner can be packed with a more retentive material, such as Tenax. This greatly improves volatile analyte recovery but may impact recovery of higher boiling materials.

Some of the solvent can be left in the liner when sample transfer begins. The residual solvent acts like a stationary phase and retains volatile material, but at the expense of a larger solvent peak.

The inlet temperature can be reduced. This reduces the vapor pressure of the volatile analytes and permits higher recoveries.

Solvent removal can be speeded up by:

Reducing pressure in the inlet during sample introductionthe Vent pressure parameter

Increasing flow through the inletthe Vent flow parameter

While all these possibilities do complicate use of the PTV, they provide increased flexibility and new potential to solve difficult problems.

Sequence of operations

These are the steps in a typical analysis using the solvent vent mode:

Table 22 The solvent vent process

Step Parameter Value

1 Before injection Flow at split vent Either Purge flow or Saver flow

Inlet pressure Derived from column setpoint

The system is resting, with Purge flow (or Saver flow, if on) through the inlet.

2 Prep Run begins Flow at split vent Vent flow setpoint

Inlet pressure Vent pressure setpoint

Setpoints change to prepare for injection. When GC is ready, the sample is injected. Inlet and oven temperature program Init times begin. Solvent venting and analyte trapping begin.

3 At Vent end time Flow at split vent None, split valve closed

Inlets 3

Advanced Operation Manual 113

Some important points

The flow through the column is governed by the pressure in the inlet. This is controlled, during the analysis part of the process, by the flow or pressure setpoint or program entered for the column.

The controlling times must be in the order shown; Vent end time before Purge time and PTV temperature Initial time before Saver time.

Vent end time must occur before the inlet starts to heat and release analytes.

Purge time must occur before the oven begins to heat and move sample through the column.

Timelines

Time increases downward; all other quantities increase to the right.

Inlet pressure Column pressure setpoint

Solvent venting ends, analyte transfer begins as inlet heats up.

4 At Purge time Flow at split vent Purge flow setpoint

Inlet pressure Column pressure setpoint

Analyte transfer ends, inlet is purged of residual vapor. Analysis begins.

5 At Saver time Flow at split vent Saver flow setpoint

Inlet pressure Column pressure setpoint

Analysis ends, carrier flow reduced to save gas (if Saver is on).

Table 22 The solvent vent process (continued)

Step Parameter Value

114 Advanced Operation Manual

3 Inlets

When is Start Run?

Both the inlet and oven temperature programs begin at Start Run. All timessuch as Purge timeare measured from Start Run. When does Start Run occur?

If the sample is injected manually, Start Run occurs when the user presses the [Start] key.

If a single injection per run is made using an autosampler, Start Run occurs when the syringe carrier moves down to make the injection.

If multiple injections per run are made using an autosampler, Start Run occurs when the syringe carrier moves down to make the first injection of the set. There are no Start Run signals for the rest of the injections in the set.

Time Oven temperature Inlet temperature Inlet pressure Split vent flow

Between runs

Prep Run

Start Run

Vent end time

Purge time

Saver time

Rate 1

Initial time

Final temperature 1 Final time 1

Rate 1

Final temperature 1 Final time 1

Other rates, temperatures, and times, if desired

Initial time

Other rates, temperatures, and times, if desired

Saver or Purge flow

Vent flow

Inlet is pressure controlled

Purge flow

Saver flow, if on

Controlled by column flow or pressure setpoint or program

Vent pressure

Controlled by column flow or pressure setpoint or program

Inlets 3

Advanced Operation Manual 115

These additional injections take time. The inlet and oven temperature programs, mainly the Initial time values, must be adjusted to allow for this. So must the various time values that control the inlet operation. This is discussed in more detail under To develop a PTV method that uses large volume injection on page 117.

Setting parameters for solvent vent operation

Mode: The current operating modesolvent vent.

Temperature Actual and setpoint initial inlet temperatures.

Initial time The time, measured from Start Run, when the initial inlet temperature hold ends. Must be greater than Vent end time.

Rate # Temperature program rate for inlet thermal ramps.

Final temperature # Final inlet temperature for ramps 1, 2, and 3.

Final time # Hold time at Final temp #. This time is a duration; it is not measured from Start Run.

Pressure Actual and setpoint inlet pressure before and after the vent period. It sets the starting point of column head pressure.

Vent pressure The inlet pressure during the vent period. By decreasing the inlet pressure while venting, solvent elimination proceeds faster. Also, the pressure reduction decreases the amount of carrier gasand solvent vaporthat enters the column during this time.

Users select from 0 to 100 psig. If 0 is chosen, the inlet uses the lowest pressure possible at the given vent flow. Table 23 shows approximate values for this minimum at various vent flows of helium. Pressures less than those in the table are not possible unless the flow is reduced.

Table 23 Minimum attainable pressures

Vent flow (mL/min) Actual vent pressure at 0 psig setpoint

Actual vent pressure at 0 kPa setpoint

50 0.7 5

100 1.3 10

116 Advanced Operation Manual

3 Inlets

Vent flow The flow of carrier gas out the split vent during the vent period. Higher flows sweep the liner more quickly and reduce the time for solvent elimination. For most columns, 100 mL/min vent flow eliminates solvent at an acceptable rate but puts minimal material on the column.

Vent end time The time, measured from Start Run, when solvent venting ends. For large volume injections, this time is normally greater than the time for the injection to complete.

Purge time The time, measured from Start Run, when sample transfer ends. It began at Vent end time.

Purge flow The flow of carrier gas to the inlet beginning at Purge time.

Total flow The actual flow into the inlet.

Septum Purge Flow through the septum purge vent

Gas saver On to reduce split vent flow at Saver time

Saver flow Reduced split vent flow, at least 15 mL/min

Saver time Time when flow is reduced to save gas

If the column is defined

1 Press [Front Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Solvent vent.

3 Enter a Vent pressure, a Vent flow, and a Vent end time.

4 Set the inlet temperature and ramps, as desired.

5 Enter a Purge time and a Purge flow.

6 If desired, turn Gas saver on. Make certain the time is set after the Purge time.

200 2.6 18

500 6.4 44

1000 12.7 88

Table 23 Minimum attainable pressures (continued)

Vent flow (mL/min) Actual vent pressure at 0 psig setpoint

Actual vent pressure at 0 kPa setpoint

Inlets 3

Advanced Operation Manual 117

7 Press [Prep Run] (see Pre Run and Prep Run on page 40) before manually injecting a sample.

If the column is not defined

1 Set up the parameters as described for the defined column case.

2 Set Total flow greater than the column flow plus the septum purge flow to guarantee adequate column flow.

To develop a PTV method that uses large volume injection

This topic provides a recommended way to change from a splitless injection using a split/splitless inlet to a solvent vent mode injection using a programmable temperature vaporization inlet (PTV). It applies mainly to large volume injections (LVI) using a PTV, but the concepts can apply to general PTV use. This topic does not consider all items that can impact an analysis, for example the liner, solvent, analyte boiling points, or polarity. This topic also assumes knowledge of your data systemwhen to save a method, how to start a run, how to set up a single run, etc.

The main advantage of the PTV inlet's solvent vent mode is that you can inject slowly into the inlet, allowing large amounts of solvent to evaporate in the liner (not in the split vent line). This concentrates the analytes prior to injection. It requires an injector with variable speed injections, a large syringe, and knowledge of the sample and the solvent.

When developing a solvent vent method, the goal is to determine the injection rates and temperatures needed to evaporate the solvent at the rate it enters the inlet. The development technique is to gradually scale up to an injection amount that produces a useful response. The most significant parameters are:

Temperature Hold the inlet temperature at or slightly below the solvent boiling point until after all the sample has been injected. This is important so that you do not boil away more volatile analytes, or boil away the solvent in the needle and trap your analytes there. An additional point to consider is that the boiling point of the earliest eluting analyte should be 100 C less than the boiling point of the solvent.

118 Advanced Operation Manual

3 Inlets

Injection speed Estimate the evaporation rate of solvent exiting the needle based on solvent type, inlet temperature, vent flow, and pressure. Start with about half that number. Note that you have to make sure that you configure the syringe properly. If not, you will over- or under- load the inlet.

Vent end time The vent end time setpoint must be greater than the time the needle spends in the inlet. If the vent time is too short, you will overload and contaminate the column and inlet. Adjust the vent time as you scale up the method.

If using a PTV with an MSD, another tip for method development is to scan for solvent ions. Detecting the solvent ions can be useful in troubleshooting residual solvent bleed onto the column.

To develop a PTV method for large volume injection, try the following:

1 Determine a small injection volume that works on a split/splitless inlet in splitless mode or the PTV in splitless mode. Choose a volume that does not currently overload the inlet.

2 Start with a 510 uL syringe and make sure the syringe is properly configured in the instrument and data system.

3 Make sure the column is configured.

4 Set up the inlet to perform a 1 uL injection.

Use the splitless method conditions, except:

Start with the inlet temperature cold, near but slightly below the solvent boiling point. For example, if using methylene chloride (boiling point 39 C), start with a temperature of 3039 C.

Use Splitless mode.

Ramp to the normal splitless inlet temperature.

5 Note the response achieved.

6 Next, change the inlet mode to Solvent vent.

7 Check the injector timings.

a Install an empty sample vial in the injector turret or tray.

b Input the injection rate (Sample Draw Speed, Sample Disp Speed, and Inject Dispense Speed rates) for the injector and increase the injection volume to 5 uL.

c Enter draft Solvent vent parameters:

Inlets 3

Advanced Operation Manual 119

Set a Vent flow of 100 mL/min as a starting point.

Keep the inlet isothermal for now.

Enter a Vent pressure of 0 psi (0 kPa) and Vent end time of 0.1 minutes.

d Make an injection using the empty vial. Use a stopwatch or your GC's timer feature to time how long the needle is in the inlet.

8 Enter revised Solvent vent mode parameters.

Set the method's Vent end time to be about 0.05 min longer than the time the needle spends in the inlet.

Set the inlet temperature Initial time to be about 0.05 min longer than the Vent pressure time.

Program the inlet to ramp quickly to the injection temperature. Make sure the ramp starts after the Vent end time.

Set the Purge Flow to 30 mL/min. Set the purge time to be the vent pressure Vent end time + 1 minute.

9 Make a 5 uL injection of your standard. You should see 5 times the response.

If the response of all analytes is too low:

The dispense speed is too fast. Liquid was injected into the inlet and pushed out the vent.

The vent time is too long. The inlet started to heat while the vent was open.

If the response of early eluters is too low:

The inlet temperature is too high.

The Vent flow is too high.

If the response of late eluters is too low:

The purge time is too short.

The final inlet temperature is too low.

10 If you need more injection volume and the 5 uL worked to give 5 times the response, change the syringe to a larger one, for example, 50 uL.

11 Set up the data system to perform a 25 uL injection.

12 Configure the syringe. Make sure the plunger speeds on the injector are still set properly.

13 Recheck the injector timings. See the step above.

120 Advanced Operation Manual

3 Inlets

14 Set a new vent time and initial inlet temperature time based on the new time the needle spends in the syringe.

15 Make a 25 uL injection of your standard. Again, you should see 5 times the response. If not, see step 13.

16 If you need more response, repeat steps 10 through 15 to increase to larger volume. See the next section. Try performing 5 x 5 uL injections, then 5 x 50 uL injections.

Multiple injections with the PTV inlet

The preferred technique for concentrating analytes in the inlet liner is to use a single, large volume injection. Using a high capacity syringe and one septum puncture reduces the possibility of contamination and generally improves results when compared against a multiple septum puncture technique. However, if needed, you can perform multiple septum punctures during the vent time. This technique requires an Agilent data system and automatic liquid sampler.

Data system requirements

An Agilent data system is necessary for multiple injection because the needed parameters are not available through the GC keyboard.

GC ChemStation Software revision B.04.01 SP1 or later.

MSD ChemStation Software revision E.02.00 SP2 or later

EZChrom Software revision 3.3.2 or later

Setting parameters for the inlet in solvent vent mode

Set or configure the following parameters in the data system's 7890A GC method editor.

Syringe size Verify the syringe size is configured correctly. The configured syringe size changes the available choices for injection volume.

Injection volume Select the injection volume, then enter a number of injections. The total injection volume will be displayed.

Multiple Injection Delay A pause time, in seconds, between injections. This is added to the minimum hardware cycle time.

Inlets 3

Advanced Operation Manual 121

Preinjection washes and pumps are performed only before the first injection of a multiple injection set.

Postinjection washes are performed only after the last injection in a multiple injection set.

An example

These values were used for a sample with a broad range of boiling points.

Table 24 General parameters

Name Value

Sample C10 to C44 hydrocarbons in hexane

Mode Solvent vent

MMI liner Glass wool packed

Injection volume One 10.0 L injection (25 L syringe)

Injection speed Fast

Column 30 m x 320 m x 0.25 m -5, part number 19091J-413

Column flow 4 mL/min constant flow

Table 25 Inlet parameters

Name Value Name Value

Initial temp 40 C Rate 2 (off)

Initial time 0.3 min Pressure 15.6 psig

Rate 1 720 C/min Vent pressure 0.0 psig

Final temp 1 450 C Vent flow 100 mL/min

Final time 1 5 min Vent end time 0.2 min

Rate 2 100 C/min Purge time 2.0 min

Final temp 2 250 C Purge flow 50 mL/min

Final time 2 0 min

Table 26 Oven parameters

Name Value

Initial temp 40 C

Initial time 2.5 min

122 Advanced Operation Manual

3 Inlets

Rate 1 25 C/min

Final temp 1 320 C

Final time 1 10.0 min

Rate 2 (off)

Table 27 Detector parameters

Name Value

Detector FID

Detector temp 400 C

Hydrogen flow 40 mL/min

Air flow 450 mL/min

Makeup (N2) 45 mL/min

Table 26 Oven parameters

Name Value

Inlets 3

Advanced Operation Manual 123

These results were compared with a splitless analysis of the same sample, which should produce 100% recovery of all analytes. The data showed that, under these conditions, compounds above C20 were completely recovered and that the recovery was independent of injection size. Compounds lower than C20 were partially vented with the solvent.

Possible adjustments

Depending on what you are trying to accomplish, you have a number of possible adjustments available.

To eliminate more solvent

Increase the vent end time, inlet initial time, and purge time. This will not affect analytes that are quantitatively trapped but will eliminate more of the solvent peak.

Increase the vent flow to sweep the liner more rapidly with the same inlet timing. Increasing vent flow raises vent pressure if it is set to 0. This puts more solvent onto the column.

C20

124 Advanced Operation Manual

3 Inlets

Raise the inlet initial temperature to vaporize more solvent and allow more to be eliminated. This also increases the loss of volatile analytes since their vapor pressures also increase.

To improve recovery of low boiling analytes

Reduce inlet temperature to lower the vapor pressure of the analytes and trap them more effectively. This also reduces solvent vapor pressure and more time will be needed to eliminate it.

Use a retentive packing in the liner. Materials such as Tenax permit higher recovery of volatile analytes but may not release higher boiling compounds. This must be considered if quantitation on these high boiling peaks is desired.

Leave more solvent in the liner. The solvent acts as a pseudo stationary phase and helps retain volatile analytes. This must be balanced against the detectors tolerance for solvent.

An examplecontinued

The single injection example shown on the last few pages makes it clear that a 10 L injection does not overload the glass wool packed liner. This means that multiple 10 L injections are possible.

It was decided to make 10 injections per run, each of 10 L size. This would increase analytical sensitivity substantially. No adjustments were made to improve recovery of the low boilers since the purpose of this analysis was to detect and measure the high boiling components.

After timing a trial set of 10 injections, the total time for the multiple injection set was measured to be approximately 1.3 minutes. The following timing changes were made:

Table 28 Modifications

Parameter Increased from To

Inlet Init time 0.3 minutes 1.6 minutes

Vent end time 0.2 minutes 1.5 minutes

Purge time 2.0 minutes 3.0 minutes

Oven Init time 2.5 minutes 3.0 minutes

Inlets 3

Advanced Operation Manual 125

The result is shown in the next figure. Note the difference in the vertical scale (5000 versus 500).

C20

126 Advanced Operation Manual

3 Inlets

About the Volatiles Interface

The volatiles interface provides a simple, reliable way to introduce a gas sample into your GC from an external device such as a headspace, purge and trap, or air toxic sampler. Manual syringe injections cannot be made with this interface. The interface has a small volume and is highly inert, ensuring high sensitivity and resolution for applications requiring trace level detection.

The figure shows the pneumatics in the split mode.

Total flow to the interface is measured by a flow sensor and is divided into two streams. One stream connects to the septum purge regulator; the other connects to a frit block. At the frit block, the flow is further divided. The first stream

Valve

Frit Frit

Frit

Valve

EPC Module

Valve

Split Vent Trap

Gas Sampling Valve Headspace Purge and Trap

FS = Flow Sensor PS = Pressure Sensor Column

FS

PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Inlets 3

Advanced Operation Manual 127

goes to the gas- phase sampler and from there is introduced into the interface. The second stream, called the pressure sensing line, passes through the frit block and is measured by a pressure sensor. This stream also provides a trickle flow to the interface.

VI operating modes

There are three modes of operationsplit, splitless, and direct. The pneumatics differ for each operating mode and are discussed in detail in the rest of this document.

Table 29 summarize some issues to consider when choosing an operating mode. Specifications for the interface are also listed.

Table 29 Overview of the volatiles interface

Mode Sample concentration

Sample to column

Comments

Split High Very little, most is vented

Splitless Low All Can switch to split mode electronically.

Direct Low All Must physically disconnect split vent, plug the interface, and reconfigure the GC. Maximizes sample recovery and eliminates possibility of contamination to pneumatic system.

Table 30 Specifications of the volatiles interface

Specification Value/Comment

Deactivated flow path

Volume 32 L

Internal dimensions 2 mm by 10 mm

Maximum flow to interface 100 mL/min

Split range Depends on column flow Typically no split to 100:1

128 Advanced Operation Manual

3 Inlets

About the VI split mode

When you introduce a sample in the split mode, a small amount of the sample enters the column while the major portion exits from the split vent. The ratio of split flow to column flow is controlled by the user. The split mode is primarily used for high concentration samples when you can afford to lose most of the sample out the split vent and for samples that cannot be diluted.

Split ratio

Because of the interfaces small internal volume, the maximum total flow to the interface is 100 mL/min. This maximum flow puts some restriction on the split ratio you can set.

Split mode pneumatics

During Pre Run, during sampling, and after sampling, total flow to the interface is measured by a flow sensor and controlled by a proportional valve. Flow at the head of the column is back- pressure regulated. Pressure is sensed upstream from the proportional valve.

Temperature range 10 C above ambient (with oven at ambient) to 400 C

Recommended temperature: transfer line temperature of the external sampling device

Table 30 Specifications of the volatiles interface (continued)

Specification Value/Comment

Table 31 Maximum split ratios

Column diameter (mm)

Column flow (mL/min)

Maximum split ratio

Total flow (mL/min)

0.20 1 100:1 100

0.53 5 20:1 100

Inlets 3

Advanced Operation Manual 129

Setpoint dependencies

Some setpoints are interdependent. If you change one setpoint, other setpoints may change to compensate. With a defined capillary column, setting column flow or linear velocity will set the inlet pressure.

Valve

Frit Frit

Frit

Valve

EPC Module

Valve

Split Vent Trap

Gas Sampling Valve Headspace Purge and Trap

FS = Flow Sensor PS = Pressure Sensor Column

FS

PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Table 32 Setpoint dependencies

When you change These setpoints change

Column defined Column not defined

Pressure Column flow* Split flow Total flow

No changes

130 Advanced Operation Manual

3 Inlets

Initial values

Use the information in Table 33 to help you set up the operating conditions for your interface.

Setting parameters for the split mode

Mode: The current operating modesplit

Temperature Actual and setpoint interface temperatures

Pressure Actual and setpoint interface pressure. Controls capillary column flow.

Column flow* Pressure Split flow Total flow

not available

Split flow Split ratio Total flow

not available

Split ratio Split flow Total flow

not available

Total flow Split flow Split ratio

No changes

* This setpoint appears in [Col 1] or [Col 2].

Table 33 Suggested starting values

Parameter Allowed setpoint range Suggested starting value

Oven initial time 0 to 999.9 minutes After sample on column

Interface temperature Ambient + 10 C to 400 C Transfer line temperature

Gas saver time 0 to 999.9 minutes After sample on column

Gas saver flow 15 to 100 mL/min 15 mL/min greater than maximum column flow

Table 32 Setpoint dependencies (continued)

When you change These setpoints change

Column defined Column not defined

Inlets 3

Advanced Operation Manual 131

Split ratio The ratio of split flow to column flow. Column flow is set using [Col 1] or [Col 2]. This parameter is not available if your column is not defined.

Split flow Flow, in mL/min, from the split vent. This parameter is not available if your column is not defined.

Total flow The total flow into the interface, both setpoint and actual

Septum Purge Flow through the septum purge vent

Gas saver On to reduce split vent flow at Saver time

Saver flow Reduced split vent flow, at least 15 mL/min.

Saver time Time when flow is reduced to save gas

If the column is defined

1 Press [Front Inlet] or [Back Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Split.

3 Set the interface temperature.

4 If you want a specific split ratio, scroll to Split ratio and enter that number. The split flow will be calculated and set for you.

5 If you want a specific split flow, scroll to Split flow and enter that number. The split ratio will be calculated and set for you.

6 If desired, turn Gas saver on. Set Saver time after the sample has been introduced.

7 If Gas saver is on, be certain Auto prep run is On (see Pre Run and Prep Run on page 40) or press [Prep Run] before introducing the sample.

If the column is not defined

1 Press [Front Inlet] or [Back Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Split.

3 Set the interface temperature.

4 Set Total flow into the interface. Measure flow out of the split vent using a flow meter.

5 Subtract the split vent flow from Total flow. Subtract the septum purge flow from the result to get column flow.

132 Advanced Operation Manual

3 Inlets

6 Calculate the split ratio (split vent flow/column flow). Adjust as needed.

7 If desired, turn Gas saver on. Set Saver time after the sample has been introduced.

8 If Gas saver is on, be certain Auto prep run is On (see Pre Run and Prep Run on page 40) or press [Prep Run] before introducing the sample.

About the VI splitless mode

This mode is used to concentrate sample at the head of the GC column during desorb. Purge timing delay must take into consideration the volume of the loop or trap in the external sampler plus the transfer line versus desorb/total flow rate. Cryo focussing is required for very volatile samples in splitless mode.

When you introduce a sample, the split valve remains closed while the sample enters the interface and is transferred to the column. At a specified time after the sample is introduced, the split valve opens.

Splitless mode pneumatics

Before Pre Run When the GC is preparing for sample introduction, total flow to the interface is measured by a flow sensor and controlled by a proportional valve. Column flow is controlled via back- pressure regulation. The split valve is open.

Inlets 3

Advanced Operation Manual 133

During sampling Pressure upsets caused by switching valves and trap restrictions in the external sampling device can cause fluctuations in column flow rates. To compensate for this, the interface is flow controlled during sampling time. The sampling flow rate is calculated from the pressure setpoint that is active when sample introduction begins. This flow control starts when the GC goes into the Pre Run state (when your system is automated and the Pre Run light is on or during manual operation when you press [Prep Run]) and ends after the interfaces Sampling end setpoint expires.

Sampling end is required for purge and trap or thermal desorption systems and should be set the sampler desorb time.

Valve

Frit Frit

Frit

Valve

EPC Module

Valve

Split Vent Trap

Gas Sampling Valve Headspace Purge and Trap

FS = Flow Sensor PS = Pressure Sensor Column

FS

PS

PS

Split Carrier Supply 80 PSI

Septum Purge

134 Advanced Operation Manual

3 Inlets

During this user- specified sampling period, the solenoid valve is closed. Flow to the interface is measured by a flow sensor and controlled by a proportional valve.

After sampling end The solenoid valve opens. The system returns to the Before Prep Run state. Flow to the interface is again measured by a flow sensor and controlled by a proportional valve while column flow is controlled via back- pressure regulation. The purge flow is controlled by the user. If desired, gas saver can be turned on at the end of the run.

Valve

Frit Frit

Frit

Valve

EPC Module

Valve

Split Vent Trap

Gas Sampling Valve Headspace Purge and Trap

FS = Flow Sensor PS = Pressure Sensor Column

FS

PS

PS

Split Carrier Supply 80 PSI

Septum Purge

Inlets 3

Advanced Operation Manual 135

Setpoint dependencies

Some setpoints in the flow system are interdependent. If you change one setpoint, other setpoints may change to compensate.

Initial values

The table shows recommended starting values for selected parameters.

Table 34 Setpoint dependencies

When you change These setpoints change

Column defined Column not defined

Purging You can change the Pressure and Total flow setpoints; other setpoints are not affected.

Purge flow Total flow**

Pressure Total flow** Column flow*

Column flow*

* This setpoint appears in the column parameters.

**This value is actual only.

Pressure Total flow**

Before and after sampling, not purging You can change the Pressure setpoint; other setpoints are not affected.

Pressure Column flow* Total flow**

Column flow* Pressure Total flow**

During sampling: You cannot change pressure and flow setpoints during sampling time.

136 Advanced Operation Manual

3 Inlets

Setting parameters for the VI splitless mode

Mode: The current operating modesplitless

Temperature Actual and setpoint interface temperatures. Column temperature must be low enough to cold trap the volatile sample. Cryo focussing is recommended.

Sampling end The sample introduction interval, in minutes. The flow rate is calculated from the pressure setpoint that is active at the start of sample introduction.

Set the sampling end setpoint 0.2 minutes longer than the time the sampler needs to introduce the sample. For example, the 7694 headspace sampler has an Inject time parameter which controls how long the valve remains in the inject position. If Inject time is 1 minute, the sampling end setpoint should be set to 1.2 minutes. If youre using an 7695 Purge and Trap Concentrator, set Sampling end 0.2 minutes longer than the Desorb time parameter.

If your column is defined and you specify a flow or pressure program for your column, the ramp does not begin until after the sampling end setpoint expires.

Pressure Actual and setpoint interface pressure in psi, bar, or kPa.

Purge time The time, after the beginning of the run, when purging resumes. Purge time must be greater than Samplg end.

Table 35 Suggested starting values

Parameter Allowed setpoint range Suggested starting value

Oven initial time 0 to 999.9 minutes Interface purge time

Interface temperature Ambient + 10 C to 400 C Transfer line temperature

Interface sampling end 0 to 999.9 minutes 0.2 minutes longer than introduction time

Interface purge time 0 to 999.9 minutes

Gas saver time 0 to 999.9 minutes Must be after purge time

Gas saver flow 15 to 100 mL/min 15 mL/min greater than maximum column flow

Inlets 3

Advanced Operation Manual 137

Purge flow The flow, in mL/min, from the split vent at Purge time. You will not be able to access or specify this value if your column is not defined.

Total flow When your column is defined, Total flow displays the actual flow to the interface. You cannot enter a setpoint. If your column is not defined, Total flow will have both setpoint and actual values during purge time. All other times, the actual flow to the interface is displayed.

Septum Purge Flow through the septum purge vent, at least 15 mL/min.

Gas saver On to reduce split vent flow at Saver time.

Saver flow Flow through the split vent after Saver time.

These instructions apply to both column defined and not defined.

1 Press [Front Inlet] or [Back Inlet].

2 Scroll to Mode: and press [Mode/Type]. Select Splitless.

3 Set the interface temperature and a sampling end time.

4 If your column is defined, enter a Purge time and Purge flow. Turn Gas saver on if desired. Set the Gas saver time after the purge time and enter a Gas saver flow.

5 If your column is not defined, enter a Purge time (Purge flow is not available). Set Total flow greater than column flow plus septum purge flow to guarantee adequate column flow.

6 Make certain Auto Prep Run is On (see Pre Run and Prep Run on page 40) or press [Prep Run] before introducing a sample.

About the VI direct mode

Direct sample introduction permits a quantitative transfer of analyte without risking contamination to the pneumatic system. It provides the sensitivity required for air toxic analyses. The interfaces minimal dead volume also eliminates the potential interaction of solutes with poorly swept, active surfaces.

To operate in the direct mode, you must physically disconnect the split vent and reconfigure the GC.

138 Advanced Operation Manual

3 Inlets

Before Pre Run

The interface is forward pressure controlled; pressure is sensed downstream from the flow proportional valve.

During sampling

Pressure upsets caused by switching valves in the external sampler can cause fluctuations in column flow rates. To compensate for this, the interface is flow controlled during sampling time. The sampling flow rate is calculated from the pressure setpoint that is active when sample introduction begins. This flow control starts when the GC goes into the Pre Run state (when your system is automated and the Pre Run light is on or during manual operation when you press [Prep Run]) and ends after the interfaces Sampling end setpoint expires.

Valve

Frit Frit

Frit

Valve

EPC Module

FS = Flow Sensor PS = Pressure Sensor Column

FS

PS

PS

Carrier Supply 80 PSI

Septum Purge

Gas Sampling Valve Headspace Purge and Trap

Inlets 3

Advanced Operation Manual 139

Flow to the interface is measured by a flow sensor and controlled by a proportional valve.

After sampling end

The interface is forward pressure controlled; pressure is sensed downstream from the proportional valve. The system returns to the idle state.

Valve

Frit Frit

Frit

Valve

EPC Module

Gas Sampling Valve Headspace Purge and Trap

FS = Flow Sensor PS = Pressure SensorColumn

FS

PS

PS

Carrier Supply 80 PSI

Septum Purge

140 Advanced Operation Manual

3 Inlets

Preparing the Interface for Direct Sample Introduction

Before you can operate your interface in direct mode, you must:

Disconnect the split vent line

Configure the GC for a direct injection

Disconnecting the split vent line

1 Press [Front Inlet] or [Back Inlet]. Turn off the interface temperature and pressure and allow the interface to cool.

2 If desired, remove the transfer line by loosening the hex nut with a wrench.

3 Remove the clamping plate from the interface by loosening the captive screw with a screwdriver. Put the plate in a safe place.

4 Carefully lift the interface out of the heater block.

WARNING Be careful! The interface may be hot enough to cause burns.

Inlets 3

Advanced Operation Manual 141

5 Loosen the hex nut connecting the split vent line to the interface until you can remove the line. Put the line aside. You do not need to plug it.

6 Install a blanking nut into the split line port and finger- tighten the nut. Tighten the nut an additional 1/4- turn using two wrenches in opposition.

7 Place the interface in the heater block. Replace the clamping plate you removed earlier and tighten the screw until snug. Do not overtighten. If you removed the transfer line, replace it.

142 Advanced Operation Manual

3 Inlets

8 Restore the GC to normal operating conditions. Perform a leak test on the interface fittings.

Configuring the GC for a direct injection

The GC cannot sense the presence of the split vent. When you disconnect or reconnect the vent, you must configure the GC so that the pneumatics work properly.

1 Press [Config][Back Inlet] or [Config][Front Inlet].

2 Scroll to Mode and press [Mode/Type].

3 Select Split vent removed. Press [Enter].

4 Press [Back Inlet] or [Front Inlet]. If the inlet is correctly configured, you will see the Direct injection display.

VI direct mode setpoint dependencies

Some setpoints in the flow system are interdependent. If you change one setpoint, other setpoints may change to compensate.

VI direct mode initial values

Use the information in Table 37 to help you set up the operating conditions for your interface.

Table 36 Setpoint changes

When you change These setpoints change

Column defined Column not defined

Before and after sampling

The Column flow* setpoint is not available.

Pressure Column flow*

Total flow**

* This setpoint appears in the column parameters

.**This value is actual only

You can change the pressure setpoint; other setpoints are not affected.

Column flow* Pressure Total flow**

During sampling You cannot change pressure and flow setpoints during sampling time.

Inlets 3

Advanced Operation Manual 143

Setting parameters for the VI direct mode

Temperature Actual and setpoint interface temperatures

Sampling end The sample introduction interval, in minutes. The flow rate is calculated from the pressure setpoint that is active at the start of sample introduction.

Set Sampling end 0.2 minutes longer than the time the sampler needs to introduce the sample. For example, the 7694 headspace sampler has an Inject time parameter which controls how long the valve remains in the inject position. If Inject time is 1 minute, Sampling end should be set to 1.2 minutes. If youre using a 7695 Purge and Trap Concentrator, set Sampling end 0.2 minutes longer than the Desorb time parameter.

If your column is defined and you specify a flow or pressure program for your column, the ramp does not begin until after Sampling end expires.

Pressure Actual and setpoint interface pressure before a run and after sampling time.

Total flow The actual flow to the interface. This is a reported value, not a setpoint.

Septum Purge Flow through the septum purge vent, range 0 to 30 mL/min.

These instructions apply to both column defined and column not defined.

1 Press [Front Inlet] or [Back Inlet].

2 Confirm that your GC is configured for direct injection.

3 Set the interface temperature.

Table 37 Suggested starting values

Parameter Allowed setpoint range Suggested starting value

Oven initial time 0 to 999.9 minutes interface sampling end

Interface temperature Ambient + 10 C to 400 C

transfer line temperature

Interface sampling end 0 to 999.9 minutes 0.2 minutes longer than actual sampling time

144 Advanced Operation Manual

3 Inlets

4 Set Sampling end at 0.2 minutes longer than the sample introduction time.

5 Make certain Auto Prep Run is On (see Pre Run and Prep Run on page 40) or press [Prep Run] before introducing a sample.

145

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

4 Columns and Oven

About the Oven 146

Oven safety 146

Configuring the Oven 147

Cryogenic Operation 148

Cryogenic setpoints 148

About Oven Temperature Programming 150

Programming setpoints 150

Oven ramp rates 151

Setting the oven parameters for constant temperature 152

Setting the oven parameters for ramped temperature 152

About the Oven Insert 154

Selecting the correct packed glass column type 155

About the column modes 155

Select a column mode 156

Setting the column parameters for constant flow or constant pressure 157

Enter a flow or pressure program (optional) 157

Programming column pressure or flow 158

About Columns 155

Backflushing a Column 159

Backflushing when connected to an MSD 160

Nickel Catalyst Tube 165

About the nickel catalyst tube 165

Nickel catalyst gas flows 165

Setting temperatures for the nickel catalyst tube 166

146 Advanced Operation Manual

4 Columns and Oven

About the Oven

Oven safety

For safety, opening the oven door turns off power to the oven heater, fan, and cryogenic valve (if installed) but maintains the setpoints in memory.

Closing the oven door returns the oven to normal operation.

If the oven cannot attain or maintain an entered setpoint temperature during normal above- ambient operation, a problem is assumed and the oven is switched off.

Possible problems include:

The oven vent flaps not working

The oven fan, heater, or temperature sensor not working properly

An electronic problem

When a shutdown occurs, the Off line in the oven parameter list blinks and the oven remains off until switched on again by pressing [Oven][On] or by editing the Temperature setpoint.

Table 38 Oven capabilities

Capability Range

Temperature range 80 C (liquid N2) or 40 C (CO2) to the configured limit

Maximum temperature 450 C

Temperature programming Up to six ramps

Maximum run time 999.99 minutes

Temperature ramp rates 0 to 120 C/min, depending on instrument configuration

Columns and Oven 4

Advanced Operation Manual 147

Configuring the Oven

Oven configuration sets maximum temperature, equilibration time the cool down mode, and the cryogenic setpoints, if cryo is installed.

Maximum temperature Maximum allowable oven temperature setpoint. Some accessories, such as the valve box, valves and columns have specific temperature limits. When configuring Maximum temperature, these limits should be considered so that the accessories are not damaged. Oven setpoints are verified as they are entered; a message is displayed when an entered setpoint is inconsistent with a previously defined maximum.

Equilibration time The time required for the oven temperature to equilibrate after temperature is modified. Equilibration time begins when the actual oven temperature comes within 1 C of the oven temperature setting. The Equilibration time setpoint can be 0 to 999.99 minutes.

Cryo See Cryogenic Operation.

External oven mode Press the GC [Info] key for details.

Slow oven cool down mode During the oven cooling period, the fan can operate at full speed or at a reduced speed. This parameter makes the choice.

148 Advanced Operation Manual

4 Columns and Oven

Cryogenic Operation

The cryogenic valve lets you operate the oven below ambient temperature. Minimum attainable oven temperature depends on the type of valve installed.

The GC senses the presence and type of cryogenic valve and disallows setpoints if no valve is installed. When cryogenic cooling is not needed or cryogenic coolant is not available, the cryogenic operation should be turned off. If this is not done, proper oven temperature control may not be possible, particularly at temperatures near ambient.

Cryogenic setpoints

Equilibration time Set from 0 to 999.999 minutes.

Cryo [ON] enables cryogenic cooling, [OFF] disables it.

Quick cryo cool This is separate from Cryo. Quick cryo cool cools the oven faster after a run than it would without assistance. This is useful when maximum sample throughput is necessary, however it does use more coolant. Quick cryo cool turns off soon after the oven reaches its setpoint and Cryo takes over, if needed.

Ambient temp The temperature in the laboratory. This setpoint determines the temperature at which cryogenic cooling is enabled.

Ambient temp + 25 C, for regular cryo operation

Ambient temp + 45 C, for Quick Cryo Cool.

Cryo timeout Cryo timeout occurs, and the oven shuts off, when a run does not start within a specified time (10 to 120 minutes) after the oven equilibrates. Turning cryo timeout Off disables this feature. We recommend that it be turned On because cryo timeout conserves coolant at the end of a sequence or if automation fails.

Cryo fault Shuts the oven down if it does not reach setpoint temperature after 16minutes of continuous cryo operation. Note that this is the time to reach the setpoint, not the time to stabilize and become ready at the setpoint. For example, with a cool on- column inlet and cryo control in the track oven mode, it may take the oven 20 to 30 minutes to achieve readiness.

Columns and Oven 4

Advanced Operation Manual 149

If the temperature goes below the minimum allowed temperature (90 C for liquid nitrogen, 70 C for liquid CO2), the oven will shut down.

External oven mode Isothermal internal mode and programmed external oven used to calculate column flow.

Slow oven cool down mode On to run oven fan at slow speed during cool down.

150 Advanced Operation Manual

4 Columns and Oven

About Oven Temperature Programming

You can program the oven temperature from an initial temperature to a final temperature using up to 20 ramps during a run.

A single ramp temperature program raises the initial oven temperature to a specified final temperature at a specified rate and holds at the final temperature for a specified period of time.

The multiple- ramp temperature program is similar. You can program the oven from an initial temperature to a final temperature, but with various rates, times, and temperatures in between. Multiple ramps can also be programmed for temperature decreases as well as increases.

Programming setpoints

Temperature Starting temperature of a temperature programmed run. When the program begins, this value is copied into a temporary setpoint called Init temp. At the end of the run, Temperature is reset to the value in Init temp and the oven returns to its starting temperature.

Initial time Time in minutes that the oven will stay at the starting temperature after a programmed run has begun.

Rate The rate in C/min at which the oven will be heated or cooled.

Initial temperature Initial time 1

Rate 1

Final temperature 1 Final time 1

Rate 2 = 0

Initial temperature Initial time 1

Rate 1

Final temperature 1 Final time 1

Rate 2

Final temperature 2 Final time 2

Rate 3 = 0

Columns and Oven 4

Advanced Operation Manual 151

Final temperature Temperature of the oven at the end of a heating or cooling rate.

Final time Time in minutes that the oven will be held at the final temperature of a temperature- programmed rate.

Total length of a run is determined by its oven temperature program. The maximum allowable time for a run is 999.99 minutes. If the program is still running at that time, the run terminates.

See also Post Run Programming.

Oven ramp rates

To use the fast oven ramp rates (a 240 V power option is required), your electric service must be able to supply 200 V at 15 Amp.

The highest rate that you can achieve depends on many factors, including the room temperature, temperatures of the inlets and detectors, the amount of material inside the oven (columns, valves, etc.), and whether or not this is the first run of the day.

The optional oven insert for fast chromatography (see About the Oven Insert on page 154), increases oven ramp rates for the back column. Table 39 lists typical oven ramp rates.

Table 39 Oven ramp rates

100/120 V oven ramp rate (C/minute)

200/220/230/240 V oven ramp rate (C/minute)

Temperature range (C)

Without insert

With optional insert

Without insert

With optional insert

50 to 70 75 120 120 120

70 to 115 45 95 95 120

115 to 175 40 65 65 110

175 to 300 30 45 45 80

300 to 450 20 35 35 65

152 Advanced Operation Manual

4 Columns and Oven

Setting the oven parameters for constant temperature

An isothermal run is one in which the oven is maintained at a constant temperature. For an isothermal run, set Rate 1 to zero.

1 Press [Oven] to open the oven parameter list.

2 Enter the oven temperature for the isothermal run.

3 Enter the number of minutes (Initial time) that you want the oven to stay at this temperature. This time is the duration of the run.

4 If Rate 1 is not already 0, enter zero for an isothermal run.

Setting the oven parameters for ramped temperature

Single ramp

1 Press [Oven] to open the oven parameter list.

2 Enter a starting temperature (Temperature).

3 Enter the time (Initial time) that you want the oven to stay at Temperature.

4 Enter the rate (Rate 1) at which the oven temperature is to change.

5 Enter the final temperature (Final temperature 1).

6 Enter the time (Final time 1) the oven is to hold Final temperature 1.

7 To end the oven ramp program after Ramp 1, set Rate 2 to zero.

Multiple ramps

In a multiple- ramp program, Final time for one ramp is also Initial time for the next ramp. Thus, there is only one Initial time.

1 Set up the first oven ramp as described in Single ramp.

2 Enter the rate (Rate 2) at which you want the oven temperature to increase for the second oven ramp.

3 Enter the final temperature (Final temperature 2).

4 Enter the number of minutes (Final time 2) that you want the oven to hold the final temperature.

5 To end the temperature program after the second ramp, set Rate 3 to zero.

Columns and Oven 4

Advanced Operation Manual 153

6 To add additional oven ramps, repeat the steps described.

154 Advanced Operation Manual

4 Columns and Oven

About the Oven Insert

The Oven Insert for Fast Chromatography reduces the oven volume so that the column and sample heat more quickly, yielding faster separation and faster chromatography. Furthermore, the smaller volume oven cools faster than a full- sized oven, reducing the overall analytical cycle time.

The oven insert is used with any inlet, column, and detector mounted in the back position. It is not compatible with any accessory which obstructs access to the front of the oven or which requires the use of either the front inlet or the front part of the oven.

Carrying strap

WARNING! Metal parts may be hot enough to burn.

Columns and Oven 4

Advanced Operation Manual 155

About Columns

In all GCs, a samplewhich is a mixture of several componentsis vaporized in an inlet, separated in a column, and examined in a detector.

The column separates components in time because:

When a vaporized component is presented with a gas phase and a coating phase, it divides between the two phases according to its relative attraction to the two phases.

The attraction can be solubility, volatility, polarity, specific chemical interaction, or any other property that differs from one component to another.

If one phase is stationary (the coating) and the other is moving (the carrier gas), the component will travel at a speed less than that of the moving phase. How much less depends on the strength of the attraction.

If different components have different attractions, they will separate in time.

Selecting the correct packed glass column type

This topic is covered in the Maintenance manual. See To attach a packed column to the purged packed inlet for details.

About the column modes

The flow modes available are determined by the GC inlets control mode. When the inlets control mode is set to Pressure control, all of the flow modes and pressure modes below are available for the column. When the inlets control mode is set to Flow control, the columns mode is not selectable. For an inlets mode of Flow control, only column flow can be entered.

The flow modes

Flow rates are corrected to NTP (normal temperature and pressure, 25 C and 1 atmosphere.

156 Advanced Operation Manual

4 Columns and Oven

Constant flowMaintains a constant mass flow rate of carrier gas in the column throughout the run. If the column resistance changes due to a temperature program, the column head pressure is adjusted to keep the flow rate constant. This can shorten runs significantly.

Ramped flowIncreases the mass flow rate in the column during the run according to a program you enter. A column flow profile can have up to three ramps, each consisting of a programmed increase followed by a hold period.

The pressure modes

The pressure modes are not available if the column is not defined or the inlets mode is set to Flow control.

Pressures are gauge pressuresthe difference between the absolute pressure and the local atmospheric pressure. Because most detectors present little resistance to the column flow, the gauge pressure at the column head is usually the same as the pressure difference between column inlet and exit. The mass selective detector and the atomic emission detector are the exceptions.

Constant pressureMaintains a constant gauge pressure at the head of the column throughout the run. If the column resistance and gas density changes during a run, the gauge pressure does not change but the mass flow rate does.

Ramped pressureIncreases the column head gauge pressure during the run according to a program you enter. A column pressure profile can have up to three ramps, each consisting of a programmed increase followed by a hold period.

Select a column mode

The columns mode parameter is not available if the inlets mode parameter is set to Flow control.

When a splitter controlled by a PCM or AUX pressure is used to divide flow between multiple columns, only the mode setting of the lowest column number configured determines the mode used. All other column mode settings exiting the splitter are ignored by the GC.

1 Press [Col 1] or [Col 2], or press [Aux col #] and enter the column number.

2 Scroll to the Mode line.

Columns and Oven 4

Advanced Operation Manual 157

3 Press [Mode/Type] to see the column mode list.

4 Scroll to the column mode you want. Press [Enter].

This completes column mode selection. Next you must specify the inlet conditions either during the entire run (if you selected either of the constant modes) or at the beginning of the run (if you selected either of the ramped modes).

Setting the column parameters for constant flow or constant pressure

If the column is defined, you can enter any one of these quantitiesthe GC will calculate and display the other two.

For example, you may have selected Constant pressure as the column mode. You decide to specify, as a starting condition, the column flow. The GC will compute the pressure necessary to achieve this flow (as well as the average linear velocity) and hold this pressure constant during the run.

If you select Constant flow as the mode and specify column flow as the initial condition, the GC will still calculate the pressure necessary to achieve this flow, but it will adjust the pressure as necessary to maintain constant flow.

If the column is not defined, you can enter only pressure. Constant flow can still be specified, but the GC cannot know what the flow is.

1 Press [Col 1] or [Col 2], or press [Aux col #] and enter the column number.

2 Scroll to the Pressure or Flow or Velocity line.

3 Type the desired initial value, followed by [Enter]. The GC will compute and display the other two values. Adjust them, if you choose to, by repeating steps 2 and 3 but note that changing any one changes all three.

This completes setting the initial carrier gas condition.

Enter a flow or pressure program (optional)

If you selected either the ramped pressure or ramped flow column mode, the column parameter list contains entries for setting up a ramp program.

You begin with an initial value, either Initial Pressure or Initial Flow, and an Initial time. At the end of that time, Rate 1 begins and runs until it reaches Final pressure (or Final flow). It

158 Advanced Operation Manual

4 Columns and Oven

remains at that value for Final time 1. You can then add a second and third ramp, each consisting of a Rate, a Final value (pressure or flow), and a Final time.

The program ends when it reaches a Rate that is set to 0 (Off).

When a flow or pressure program is running, the Pressure, Flow, and Velocity lines that you used to set constant conditions show the progress of the program.

The oven program determines the length of the run. If a flow or pressure program ends before the analytical run does, the flow (or pressure) remains at the last final value.

Programming column pressure or flow

1 Press [Col 1] or [Col 2], or press [Aux col #] and enter the column number.

2 Scroll to Initial pressure (or Initial flow). Type the desired value and press [Enter].

3 Similarly, enter a value for Initial time. This completes the initial part of the program.

4 To begin a ramp, enter a positive value for Rate 1. It does not matter whether you are programming up or downthe rate is always positive.

5 If Rate 1 is zero, the program ends here. If you enter any other value, the Final value lines for the first ramp appear and the cursor moves to the line.

6 Enter values for Final pressure 1 (or Final flow 1) and Final time 1. This completes the first ramp.

7 To enter a second ramp, scroll to the appropriate Rate line and repeat steps 5 and 6. A maximum of 20 ramps can be entered.

Columns and Oven 4

Advanced Operation Manual 159

Backflushing a Column

Backflush is a means of discarding high- boilers from a column after the peaks of interest have eluted. It saves analysis time and has these additional benefits:

Longer column life

less high temperature exposure

removal of high- boilers

protection from air and water at high temperatures

Less chemical background

ghost peaks

wrap- around of late eluters from previous runs

stationary phase decomposition peaks

Less contamination of the mass selective detector (MSD) source, if using MSD

longer time between source cleanings

higher stability of calibrations

Agilent now provides two Backflush Wizard software utilities. One is available with the Instrument Utilities software, and the other is an add- on for your Agilent data systems. These utilities complement each other, and automate the process of updating the GC and method to implement and validate successful backflushing. Contact your Agilent sales representative for more information

If not using an Agilent Backflush Wizard, you will need to experimentally determine the correct flow/pressure settings and the backflush time appropriate for your analysis. Determine the correct values by testing the backflush method, then running a blank to check for residual late eluters.

Refer to the Agilent web site at http://www.agilent.com/chem for example backflush applications, especially if using an MSD.

CAUTION Only backflush an inlet that has a split vent line with a chemical trap, such as the split/splitless, PTV, and VI. Attempts to backflush using other inlet types will most likely damage the pneumatics flow modules.

160 Advanced Operation Manual

4 Columns and Oven

Backflushing when connected to an MSD

If using an MSD with the GC, backflushing becomes more complex. Set up backflush as a post run event by using the Backflush Wizard present in the MSD Productivity ChemStation.

Backflushing using a capillary flow technology device

Because the following capillary flow technology (CFT) devices connect to a controlled carrier gas supply, they can be used to backflush a column:

G2855B Capillary Flow Technology Deans Switching System

G3180B Capillary Flow Technology Two- Way Splitter (with makeup gas)

G3183B Capillary Flow Technology Three- Way Splitter

G3185B QuickSwap Accessory

A conceptual diagram for a simple setup is shown in the figure.

After the last analyte of interest elutes, you can program a decrease in inlet column flow with an increase in the pressure/flow from the Aux EPC or PCM connected to the CFT device, thereby reversing column flow and forcing the remaining analytes off the column and out of the inlet split vent. You can do this using a post run program or as a ramped pressure or flow program.

Split Vent Trap

Inlet

Column

CFT Device

Auxiliary EPC/PCM

Detector

Columns and Oven 4

Advanced Operation Manual 161

To set up a post run backflush

A main advantage of the post- run backflush method is that the GC turns off detection during this time. If using an MSD, this helps prevent damage to the detector and is the recommended backflush method.

If using an Agilent data system, a Backflush Wizard provides a straightforward interface for making these settings.

1 Verify that all columns are properly configured.

2 Since the backflush will run as a post- run program, set the method so that the oven program ends after the last peak of interest returns to baseline, or after reaching the last temperature of interest.

3 Press [Post Run] and enter the backflush duration as the Time.

The oven temperature and column pressure/flow setpoints for installed columns appear.

Whether you can enter a pressure or flow for a column depends on its mode (pressure or flow).

Some inlets, such as the multimode inlet, also provide a post run inlet temperature and total flow rate.

4 Enter the oven temperature for the backflush.

5 Enter the backflush flow(s) or pressure(s).

If in flow mode:

Enter a negative flow for the column connected between the inlet and the CFT device. The GC will automatically establish the corresponding pressures needed in the inlet and the CFT device.

Increase the flow rates in the column(s) connected between the CFT device and the detector(s).

If in pressure mode:

Set the pressure of the column connected between the inlet and the CFT device to 0.000. This turns off flow from the inlet.

Increase the pressure of the primary column connected between the CFT device and the detector(s). This creates the backward flow through the column and increases the flow through any connected detectors.

6 If using a multimode inlet, enter a total flow and inlet temperature for the post run (backflush) period.

162 Advanced Operation Manual

4 Columns and Oven

When developing the backflush portion of your method, consider the following:

Ensure that the split vent flow setpoint is at least 25 mL/min and at least 50% more than the column backflush flow rate.

If using gas saver, ensure that the gas saver flow setpoint is at least 25 mL/min and at least 50% more than the column backflush flow rate.

To backflush using a ramped pressure program

In this case, the backflush occurs as part of the run, so the detectors continue to collect data. During the backflush, you may wish to turn off data collection in the data system.

1 Verify that all columns are properly configured.

2 Enter all method parameters for the analysis: sampler parameters, inlet parameters, oven temperature profile, detector flows and temperatures, and so forth.

3 Program the oven for the backflush.

Include any temperature profile needed for backflush.

Set the total run time to include sufficient time for backflush.

4 Program the pressure ramp for the column installed between the inlet and the CFT device. After the last analyte elutes or after reaching the last temperature of interest, program a fast ramp (for example, 30 psi/min) with a final pressure of 0.

5 Program the pressure ramp for the primary column installed between the CFT device and the detector. The pressure should increase slightly during the backflush duration so that the flow into the detectors remains relatively stable.

If you turned off data acquisition in a data system during backflush, remember to turn it on again at the end of the run.

CAUTION To avoid damage to an MSD, Agilent strongly recommends setting up backflush as a post run event, not as part of a ramped column program. If you still choose to backflush as part of a run, be very careful that the flow into the MSD does not exceed the limits of the vacuum pump.

Columns and Oven 4

Advanced Operation Manual 163

To backflush using a ramped flow program

In this case, the backflush occurs as part of the run, so the detectors continue to collect data. During the backflush, you may wish to turn off data collection in the data system.

1 Verify that all columns are properly configured.

2 Enter all method parameters for the analysis: sampler parameters, inlet parameters, oven temperature profile, detector flows and temperatures, and so forth.

3 Program the oven for the backflush.

Include any temperature profile needed for backflush.

Set the total run time to include sufficient time for backflush.

4 Program the flow ramp for the column installed between the inlet and the CFT device. After the last analyte elutes or after reaching the last temperature of interest, program a fast ramp with a final flow that is negative.

5 Program the flow ramp for the primary column installed between the CFT device and the detector. Typically hold at the methods final value for the backflush duration.

If you turned off data acquisition in a data system during backflush, remember to turn it on again at the end of the run.

Backflushing using a switching valve

Backflushing is done using a column switching valve controlled by the Run Table. See Run Time Programming on page 12.

The valve is plumbed as follows:

Position 1 Carrier gas flows through the column to the detector. This is the normal flow path.

Position 2 Carrier gas flows through the column toward the inlet, removing components on the column through the inlet vent line.

CAUTION To avoid damage to an MSD, Agilent strongly recommends setting up backflush as a post run event, not as part of a ramped column program. If you still choose to backflush as part of a run, be very careful that the flow into the MSD does not exceed the limits of the vacuum pump.

164 Advanced Operation Manual

4 Columns and Oven

The Run Table contains commands to perform these actions:

After the last peak of interest appears, switch the valve to Position 2. Higher boiling peaks are discarded through the inlet vent.

At the same time, turn data acquisition off.

At the end of the backflush period (determined experimentally), switch the valve to Position 1 and turn data acquisition on. The system is now ready for the next run.

Columns and Oven 4

Advanced Operation Manual 165

Nickel Catalyst Tube

About the nickel catalyst tube

The Nickel Catalyst Tube accessory, G2747A, is used for trace analysis of CO and CO2 with a flame ionization detector. The gas sample is separated on the column and passed over a heated catalyst in the presence of hydrogen, which converts the CO and CO2 peaks to CH4.

Nickel catalyst gas flows

For a standard FID installation:

Carrier gas

Sample Hydrogen Air

Gas sample valve Column Nickel catalyst FID

Table 40 Gas flows for a standard FID

Gas Flow rate, mL/min

Carrier (helium) 30

FID hydrogen 30 (see Caution)

FID air 400

166 Advanced Operation Manual

4 Columns and Oven

Setting temperatures for the nickel catalyst tube

The nickel catalyst tube is usually mounted in the back inlet position and controlled by the back inlet temperature setpoint. For most analyses, set these temperatures:

Nickel catalyst tube375 C

FID400 C

Table 41 Gas flows for a TCD/FID series installation

Gas Flow rate, mL/min

Carrier (helium) 30

TCD switching flow 25

FID hydrogen 45 (see Caution)

FID air 500

CAUTION Hydrogen flow is pressure-controlled, where an FID provides a known resistance. The nickel catalyst tube increases flow resistance, so that the calibration is no longer valid. You must measure hydrogen flow with a bubble or similar meter.

The nickel catalyst can be damaged by exposure to air.

167

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

5 Detectors

About Makeup Gas 168

About the FID 169

About the TCD 174

About the uECD 182

About the NPD 188

About the FPD+ 200

About the FPD 206

168 Advanced Operation Manual

5 Detectors

About Makeup Gas

Most detectors use a makeup gas to increase the flow rate through the detector body. This sweeps peaks out of the detector quickly, avoiding mixing of components and loss of resolution. This is particularly important with capillary columns because the column flow rates are so small.

The makeup gas line of your detector parameter list changes depending on your instrument configuration.

If you have an inlet with the column not defined, the makeup flow is constant. If you are operating with column defined, you have a choice of two makeup gas modes.

Constant makeup This mode provides a constant flow of makeup gas to the detector.

Column + makeup = constant This mode provides a variable flow of makeup gas to the detector. As column flow increases or decreases, the makeup flow changes to provide a constant combined flow to the detector. If you choose this option, enter a value under Combined flow. The Combined flow line always displays the same value, while the Makeup line changes as the actual makeup flow changes.

To change the makeup gas flow mode

1 Press [Front Det] or [Back Det].

2 Scroll to Mode. Press [Mode/Type].

3 Scroll to the correct mode and press [Enter].

Detectors 5

Advanced Operation Manual 169

About the FID

The FID passes sample and carrier gas from the column through a hydrogen- air flame. The hydrogen- air flame alone creates few ions, but burning an organic compound increases the number of ions produced. A polarizing voltage attracts these ions to a collector located near the flame. The current produced is proportional to the amount of sample being burned. This current is sensed by an electrometer, converted to digital form, and sent to an output device.

The FID uses three supply gases (hydrogen, makeup gas, and air) and two supply lines. Air flows through one supply line, while hydrogen mixed with the makeup gas flows through the other. All three supply gases use:

A filter frit to protect the flow path and limit the flow rate

A proportional valve to control the pressure

A pressure sensor and restrictor to control the valve

The hydrogen and makeup mix outside the flow module and enter the detector at the base of the jet. Air enters above the jet.

170 Advanced Operation Manual

5 Detectors

How FID units are displayed in Agilent data systems and on the GC

The GC displays the FID signal in picoamperes (pA). The following table lists how different data systems convert the display units to reporting units.

Valve

Frit Frit

Valve

PS = Pressure Sensor Column

PS

PSPS

Valve

Frit

Restrictor Restrictor

Restrictor

EPC Module

H2 Air

Vent

Makeup

Table 42 Properties of the FID

Property Value/Comment

Dynamic range 1 x 107

Sensitivity 10 to 100 picograms of an organic compound dependent on the molecular structure.

Selectivity Responds to all organic compounds (C-H bonds) except those which do not burn or ionize in the hydrogen-air flame. There is little or no response for H2O, CO2, CO, N2, O2 CS2 or inert gases. Formaldehyde and heavily halogenated compounds give minimal response.

Detectors 5

Advanced Operation Manual 171

To light the FID flame

Press [Front Det] or [Back Det], scroll to Flame, then press [On/Yes].

To extinguish the FID flame

Press [Front Det] or [Back Det], scroll to Flame, then press [Off/No].

FID automatic reignition (Lit offset)

Lit offset is the expected minimum difference between the FID output with the flame lit and the output with the flame off. The GC checks this value during runs and when loading a method.

During a run, if the output falls below the Lit offset value, the FID will attempt to reignite three times. If after the third attempt the output does not increase by at least this value, the detector shuts down all functions except temperature and makeup gas flow.

When loading a method that includes a Flame On setting, the GC performs a similar check. If the detector output is less than the Lit offset, it will attempt reignition after reaching method setpoints.

The default setting for Lit offset is 2.0 picoamps. This is a good working value for all but very clean gases and systems. You may want to lower this setpoint if the detector attempts to reignite when the flame is still on, thus producing a shutdown.

Table 43 Unit conversions

Data system Height units LSV (height units) Area units Noise (ASTM)*

Agilent data system 1 pA 1.3 x10-4 pA 1 pA-sec 0.038 pA

SIGRange 0 1 x 10-4 pA 1.3 x 10-4 pA 1 x 10-4 pA 0.038 pA

SIGRange 5 3.2 x 10-3 pA 4.2 x 10-3 pA 3.2 x 10-3 pA 0.038 pA

Analog 1V 1.25 x 10-4 pA device dependent 1.25 x 10-4 pA 0.038 pA

* Noise is recommended maximum when determining MDL.

SIGRange used with 3393 and 3396 integrators.

Analog 1V is an approximate value.

172 Advanced Operation Manual

5 Detectors

To change Lit offset:

1 Press [Config][Front Det] or [Config][Back Det].

2 Scroll to Lit offset.

3 Enter the new value and press [Enter].

Recommended starting conditions for new FID methods

See Table 44 for guidelines and rules to select initial detector settings for new methods.

Table 44 Recommended starting conditions

Combustible gas mix Make sure that the final hydrogen-to-air ratio is between 8% and 12%

Detector temperature Set to 20 C above the highest oven temperature, depending on the column type.

A temperature of 300 C provides a good starting point and easier ignition, and minimizes water condensation.

The GC will not attempt to ignite the flame at a temperature <150 C.

Carrier gas flow (hydrogen, helium, nitrogen)

Packed columns Suggest 10 to 60 mL/min

Capillary columns Suggest 1 to 5 mL/min

Detector gases Flow range mL/min Suggested flow mL/min

Standard installation

Column plus capillary makeup 10 to 60 30

Hydrogen 24 to 60 30

Air 200 to 600 400

With Nickel Catalyst Accessory: Standard installation

Column plus capillary makeup 10 to 60 30

Hydrogen 24 to 60 30*

Air 200 to 600 400

With Nickel Catalyst Accessory: TCD to FID series installation

Column plus capillary makeup 10 to 60 30

TCD switching flow 25

Hydrogen 24 to 60 45*

Detectors 5

Advanced Operation Manual 173

Use nitrogen makeup gas (instead of helium) for greater sensitivity.

Setting parameters for FID

To set the FID parameters:

1 Verify:

Makeup gas is configured

Installed jet type is correct for column type

2 Press [Front Det] or [Back Det].

3 Set the detector temperature. The temperature must be greater than 150 C for the flame to light.

4 Set the hydrogen flow rate, if desired, and press [Off/No].

5 Change the air flow rate, if desired, and press [Off/No].

6 If using a packed column, set the FID makeup gas to 0.0/Off.

7 If using a defined capillary column, set the makeup gas flow or combined column plus makeup gas flow.

8 Scroll to Flame and press [On/Yes]. This turns on the air and hydrogen and initiates the ignition sequence. The signal typically increases to 5 to 20 pA after ignition.

Verify that the flame is lit by holding a cold, shiny surface, such as a mirror or chrome- plated wrench, over the collector exit. Steady condensation indicates that the flame is lit.

Air 200 to 600 450

* Detector hydrogen is pressure controlled, where the detector provides a known resistance. If using a nickel catalyst tube, the resistance changes, and the flow rates displayed by the GC will not be accurate. Measure the actual hydrogen flow using a flow meter at the detector vent, with all other flows turned off.

Table 44 Recommended starting conditions (continued)

WARNING Verify that a column is installed or the FID column fitting is plugged before turning on the air or hydrogen. An explosion may occur if air and hydrogen are allowed to leak into the oven.

174 Advanced Operation Manual

5 Detectors

About the TCD

The TCD compares the thermal conductivities of two gas flowspure carrier gas (the reference gas) and carrier gas plus sample components (the column effluent).

This detector contains a filament that is heated electrically so that it is hotter than the detector body. The filament temperature is held constant while alternate streams of reference gas and column effluent pass over it. When a sample component appears in the effluent, the power required to keep the filament temperature constant changes. The two gas streams are switched over the filament five times per second (hence the ticking sound) and the power differences are measured and recorded.

When helium (or hydrogen) is used as carrier gas, the sample causes the thermal conductivity to fall. If nitrogen is used, the thermal conductivity usually goes up because most things are more conductive than nitrogen.

Because the TCD does not destroy the sample during the detection process, this detector can be connected in series to a flame ionization detector or other detector.

Detectors 5

Advanced Operation Manual 175

Column effluent is forced away from the filament. TCD measures reference gas.

Column effluent is forced toward the filament. TCD measures peaks (if present).

176 Advanced Operation Manual

5 Detectors

TCD pneumatics

This is the pneumatics design of the TCD.

TCD carrier, reference, and makeup gas

Reference and makeup gas must be the same as the carrier gas, and the gas type must be specified in both the inlet and detector parameter lists.

When using packed columns, we recommend a small makeup gas flow (2 to 3 mL/min) to get the best peak shapes.

Use the next figure to select a value for reference gas flow for either capillary or packed columns. Any ratio within 0.25 of that in the figure is suitable.

PS = Pressure Sensor

Vent

PS

Valve

FritFrit

Reference switching valve

Restrictor

EPC Module

Reference GasMakeup

PS

Restrictor

Valve

Column

Detectors 5

Advanced Operation Manual 177

TCD gas pressures

Choose a flow, find a pressure, set source pressure 10 psi (70 kPa) higher.

Ratio of reference flow to column + makeup flow

4.0

3.0

2.0

1.0

0

Column + makeup flow, mL/min

605040302010

Hydrogen Helium

NitrogenReference gas flow, mL/min

Pressure (psig) (kPa)

4

3

2

1

10 69

60 414

50 345

40 276

30 207

20 138

178 Advanced Operation Manual

5 Detectors

Selecting reference and makeup flows for the TCD

Sample components with higher thermal conductivities than the carrier gas produce negative peaks. For example, helium or hydrogen form a negative peak with nitrogen or argon- methane as the carrier gas.

Chemically active compounds reduce TCD filament life

The tungsten- rhenium TCD filament has been chemically passivated to protect against oxygen damage. However, chemically active compounds such as acids and halogenated

Makeup gas flow, mL/min

Hydrogen

Helium

Nitrogen

Pressure (psig) (kPa)

10 69

60 414

50 345

40 276

30 207

20 138

4

8

12

16

20

Table 45 Recommended flow rates and temperatures

Gas type Flow range

Carrier gas (hydrogen, helium, nitrogen)

Packed, 10 to 60 mL/min Capillary, 1 to 5 mL/min

Reference (same gas type as carrier)

15 to 60 mL/min See the figures to select a value.

Capillary makeup (same gas type as carrier)

5 to 15 mL/mincapillary columns 2 to 3 mL/minpacked columns

Detector temperature

<150 C, cannot turn on filament Detector temperature should be 30 C to 50 C greater than highest oven ramp temperature.

Detectors 5

Advanced Operation Manual 179

compounds may attack the filament. The immediate symptom is a permanent change in detector sensitivity due to a change in filament resistance.

If possible, such compounds should be avoided. If this is not possible, the filament may have to be replaced frequently.

Changing the TCD polarity during a run

Negative polarity On inverts the peak so the integrator or ChemStation can measure it. Negative polarity can be a run table entry; see Run Time Programming on page 12.

Detecting hydrogen with the TCD using helium carrier gas

Hydrogen is the only element with thermal conductivity greater than helium, and mixtures of small amounts of hydrogen (<20%) in helium at moderate temperatures exhibit thermal conductivities less than either component alone. If you are analyzing for hydrogen with helium carrier gas, a hydrogen peak may appear as positive, negative, or as a split peak.

There are two solutions to this problem:

Use nitrogen or argon- methane as carrier gas. This eliminates problems inherent with using helium as carrier, but causes reduced sensitivity to components other than hydrogen.

Operate the detector at higher temperaturesfrom 200 C to 300 C.

You can find the correct detector operating temperature by analyzing a known range of hydrogen concentrations, increasing the operating temperature until the hydrogen peak exhibits normal shape and is always in the same direction (negative relative to normal response to air or propane) regardless of concentration. This temperature also ensures high sensitivity and linear dynamic range.

Because hydrogen peaks are negative, you must turn negative polarity on at appropriate times so the peak appears positive.

180 Advanced Operation Manual

5 Detectors

Setting parameters for the TCD

1 Press [Front Det] or [Back Det].

2 Set the detector temperature. Do not set higher than the maximum temperature allowed for the column because part of the column passes through the heated block and into the cell.

3 Verify that makeup gas type is the same as that plumbed to your instrument (next to Makeup line in the parameter list). Change the gas type, if necessary.

4 Set the reference gas flow rate.

5 If you are using packed columns, turn off the makeup gas (or proceed to step 6 and enter 2 to 3 mL/min, see TCD carrier, reference, and makeup gas on page 176) and proceed to step 7

6 If you are using capillary columns:, choose a flow mode and set the makeup gas flow or combined flow.

7 Turn on the filament. Allow about 30 minutes for thermal stabilization. A longer period may be needed for the highest sensitivity.

8 If necessary, turn Negative polarity [On/Yes] to invert negative- going peaks. When a sample contains components giving both positive- and negative- going peaks, Negative polarity can be switched on and off during a run as a timetable event.

Example: Packed mode (packed and large capillary columns)

Column flow is 15 to 60 mL/min. Set the reference flow to 1.5 times the sum of column flow + makeup flow.

Makeup gas is recommended with all capillary columns. It allows the column to be inserted all the way into the detector and withdrawn 1 mm. If makeup is not used, the column must be no more than 3 mm above the ferrule. Minimum makeup flow is 1 mL/min.

1/8-inch stainless steel column If column flow is 30 mL/min, set the reference flow to 30 1.5 = 45 mL/min. Total detector flow is 30 + 45 = 75 mL/min.

10 m 0.53 mm column If column flow is 15 mL/min and makeup flow = 2 mL/min, set the reference flow to 1.5 (15 + 2) = 25.5 mL/min. Total detector flow is 17 + 25.5 = 42.5 mL/min.

Detectors 5

Advanced Operation Manual 181

Example: Capillary mode (small capillary columns)

If combined column plus makeup flow is between 5 and 10 mL/min, set the reference flow at 3 the combined flow. For a combined flow between 10 and 15 mL/min, use a multiplier of 2. This will bring the TCD within 25% of the maximum response. For further optimization, adjust the reference flow.

2 m 0.2 mm capillary column If column flow is 0.75 mL/min, the makeup must be at least 4.25 mL/min. Set it = 5. Reference flow will then be 3 5.75 = 17.25 mL/min. Total detector flow = 5.75 + 17.25 = 22.5 mL/min.

25 m 0.32 mm capillary column If column flow = 10 mL/min, set makeup low to minimize sample dilution. Set it = 2 mL/min. Reference flow will then be 12 2 = 24 mL/min. Total detector flow = 12 + 24 = 36 mL/min.

182 Advanced Operation Manual

5 Detectors

About the uECD

The micro- cell detector (uECD) contains a cell plated with 63Ni, a radioactive isotope. The 63Ni releases particles that collide with carrier gas molecules to produce low- energy electronseach particle produces approximately 100 electrons. The free electrons produce a small currentcalled the reference or standing currentthat is collected and measured in a pulsed circuit.

When a sample component molecule comes into contact with the free electrons, the electrons may be captured by the sample molecules to create negatively charged ions. The voltage across the cell electrodes is pulsed to collect the remaining free electrons while the heavier ions are relatively unaffected and swept out the vent with the carrier gas flow.

Cell current is measured and compared to a reference current. The pulse rate is adjusted to maintain a constant cell current. The more uncaptured electrons, the lower the pulse frequency required to match the reference current. When a component that captures electrons passes through the cell, the pulse rate rises. This pulse rate is converted to a voltage and recorded.

uECD safety and regulatory information

The 63Ni isotope

The radioactive isotope used in the cell is 63Ni. It is plated onto the inner surface of the cell body and is solid at temperatures used in chromatography. Some other properties are listed below.

Table 46 Properties of 63Ni

Property Value

Halflife: 101.1 years

Emission: 65.87 keV max., beta radiation

Melting point: 1453 C

Dimensions of the active part of the uECD:

Inside diameter: 6 mm Height: 4.2 mm

Total activity (uECD cell): 555 MBq (15 millicuries) maximum

Detectors 5

Advanced Operation Manual 183

uECD licenses

Customers in the United states can purchase an exempt model uECD. Customers outside the United States should contact their local Agilent sales office for information.

uECD warnings

Although beta particles at this energy level have little penetrating power the surface layer of the skin or a few sheets of paper will stop most of themthey may be hazardous if the isotope is ingested or inhaled. For this reason the cell must be handled with care: Radioactive leak tests must be performed at the required intervals (not applicable to exempt models), the inlet and outlet fittings must be capped when the detector is not in use, corrosive chemicals must not be introduced into the detector, and the effluent from the detector must be vented outside the laboratory environment.

After turning off the main power and allowing the instrument to cool, cap the detector inlet and exhaust vent openings. Wear disposable plastic gloves and observe normal laboratory safety precautions.

Return the cell for exchange, following directions included with the License Verification Form (part no. 19233- 90750).

WARNING Materials that may react with the 63Ni source, either to form volatile products or to cause physical degradation of the plated film, must be avoided. These materials include oxidizing compounds, acids, wet halogens, wet nitric acid, ammonium hydroxide, hydrogen sulfide, PCBs, and carbon monoxide. This list is not exhaustive but indicates the kinds of compounds that may cause damage to 63Ni detectors.

WARNING In the extremely unlikely event that both the oven and the detector heated zone should go into thermal runaway (maximum, uncontrolled heating in excess of 400 C) at the same time, and that the detector remains exposed to this condition for more than 12 hours, take the following steps:

184 Advanced Operation Manual

5 Detectors

Include a letter stating the condition of abuse.

It is unlikely, even in this very unusual situation, that radioactive material will escape the cell. However, permanent damage to the 63Ni plating within the cell is possible, and therefore, the cell must be returned for exchange.

Safety precautions when handling uECDs

Never eat, drink, or smoke when handling uECDs.

Always wear safety glasses when working with or near open uECDs.

Wear protective clothing such as laboratory jackets, safety glasses, and gloves, and follow good laboratory practices. Wash hands thoroughly with a mild non- abrasive cleaner after handling uECDs.

Cap the inlet and outlet fittings when the uECD is not in use.

Connect the uECD exhaust vent to a fume hood or vent it to the outside. See the latest revision of title 10, Code of Federal Regulations, part 20, (including appendix B) or the applicable State regulation. For other countries, consult with the appropriate agency for equivalent requirements.

Agilent Technologies recommends a vent line inside diameter of 6 mm (1/4 inch) or greater. With a line of this diameter, the length is not critical.

WARNING Do not use solvents to clean the uECD.

WARNING You may not open the uECD cell unless authorized to do so by your local nuclear regulatory agency. Do not disturb the four socket-head bolts. These hold the cell halves together. United States customers removing or disturbing them is a violation of the terms of the exemption and could create a safety hazard.

Detectors 5

Advanced Operation Manual 185

uECD gas flows

uECD linearity

The uECD response factor versus concentration curve is linear for four orders of magnitude or more (linear dynamic range = 104 or higher) for a broad range of compounds. You should still run a calibration curve on your samples to find the limits of the linear range for your materials.

uECD detector gas

The uECD operates with either nitrogen or argon/methane as the makeup and anode gas. Purity is critical; gases must exceed 99.9995% purity.

Because of the high detector sensitivity, carrier and makeup gas must be dry and oxygen- free. Moisture, chemical, and oxygen traps in good condition should be installed in carrier and makeup gas supply lines. Do not use plastic (including PTFE) tubing, plastic- bodied traps, or O- ring seals.

uECD temperature

To prevent peak tailing and to keep the cell clean, the detector temperature should be set higher than the highest oven temperature usedthe setpoint should be based on the

PS = Pressure Sensor

PS

Valve

Frit

Restrictor

Restrictor

EPC Module

Anode purge and makeup gas

Vent 63Ni

Column

Capillary adapter

186 Advanced Operation Manual

5 Detectors

elution temperature of the last compound. If you operate at excessively high temperatures, your results will not necessarily improve and you may increase sample and column decomposition.

uECD analog output

If you intend to use the analog output from the uECD, you must set the output Range to 10.

1 Press [Analog Out 1] or [Analog Out 2].

2 Scroll to Range.

3 Type 10 and press [Enter].

Recommended starting conditions for new uECD methods

Use the following information when selecting temperatures and flows. Maximum source pressure must not exceed 100 psi. Use the maximum source pressure to achieve maximum makeup flow rate.

uECD makeup gas notes

If the carrier gas type is different from the makeup gas type, the makeup gas flow rate must be at least three times the carrier gas flow rate.

uECD sensitivity can be increased by reducing the makeup gas flow rate.

Table 47 Starting values

Gas Recommended flow range

Carrier gas Packed columns (nitrogen or argon-methane) Capillary columns (hydrogen, nitrogen, or argon-methane)

30 to 60 mL/min 0.1 to 20 mL/min, depending on diameter

Capillary makeup (nitrogen or argon-methane)

10 to 150 mL/min (30 to 60 mL/min typical)

Temperature

250 C to 400 C Detector temperature is typically set 25 C greater than the highest oven ramp temperature.

Detectors 5

Advanced Operation Manual 187

uECD chromatographic speed (for fast peaks) can be increased by increasing the makeup gas flow rate.

uECD temperature programming

The uECD is flow sensitive. If you are using temperature programming, in which the column flow resistance changes with temperature, set up the instrument as follows:

Set the carrier gas in the Constant flow mode. Set detector makeup gas to Constant makeup.

If you choose to work in the constant pressure mode, the makeup gas should be set in the Column +makeup=constant mode.

Setting parameters for the uECD

Verify that your detector gases are connected, a column is properly installed, and the system is free of leaks. Set the oven temperature and the inlet temperature and flow. Make sure your carrier gas type is the same as that plumbed to your GC.

1 Press [Front Det] or [Back Det].

2 Set the detector temperature. To keep the uECD cell clean, this temperature must be higher than the oven temperature.

3 Verify that the makeup gas type is the same as that plumbed to your instrument. The gas type is in parentheses next to the Makeup line on the parameter list. Change the gas type, if necessary.

4 Enter a value for the makeup gas flow.

If you are using packed columns, turn off the makeup gas.

If your capillary column is defined, choose a flow mode and set the makeup or combined gas flow.

If your capillary column is not defined, only constant makeup flow is available. Enter a makeup gas flow.

188 Advanced Operation Manual

5 Detectors

About the NPD

New NPD features and changes

The NPD firmware in this GC (A.01.08 or higher) is considerably different from that in earlier versions and from the 6890 firmware. Major changes are:

Equilibration time parameter has been removed.

Adjust Offset feature has been changed.

Auto Adjust on/off has been added.

Dry Mode has been added.

Blos/ceramic bead selector has been added.

We strongly recommend that you allow the firmware to perform Auto Adjust and set the Bead Voltage.

NPD software requirements

This discussion assumes that the following firmware/software is installed:

Software/firmware with numbers less than shown in the table can cause reduced bead lifetime. See the Agilent web site (www.agilent.com) for firmware and software updates.

NPD flows and general information

The NPD passes sample and carrier through a hydrogen/air plasma. A heated ceramic or glass source, called the bead, is just above the jet. The low hydrogen/air ratio cannot sustain a flame,

Table 48 Software requirements

Product Software/firmware revision

7890B or 7890A+ GC All

7890A GC A.01.08 or higher

Agilent OpenLAB CDS All

Agilent MassHunter WorkStation All

Agilent GC ChemStation B.04.01 or higher

Agilent MSD ChemStation G1701EA or higher

Agilent EZ Chrom Elite 3.3.2 or higher

Detectors 5

Advanced Operation Manual 189

minimizing hydrocarbon ionization, while the alkali ions on the bead surface facilitate ionization of nitrogen- or phosphorus-organic compounds. The output current is proportional to the number of ions collected. It is sensed by an electrometer, converted to digital form, and sent to an output device.

NPD flow, temperature, and bead recommendations

Valve

Frit Frit

Valve

PS = Pressure Sensor Column

PS

PS PS

Valve

Frit

Restrictor Restrictor

Restrictor

EPC Module

H2 Air

Vent

Electrically heated bead

Makeup

Table 49 General operating values

Gas or Setting Recommendation

Carrier gas (helium, hydrogen, nitrogen) Capillary, choose optimum flow based on column dimensions.

Detector gases

Hydrogen Ceramic bead 2 to 5 mL/min

Blos bead 1 to 3 mL/min

Air 60 mL/min

190 Advanced Operation Manual

5 Detectors

Source gas pressures

Choose a flow, find a pressure, and set source pressure 10 psi (70 kPa) higher.

Capillary makeup (helium, nitrogen) Ceramic bead Nitrogen: 5 to 10 mL/min Helium: less than 5 mL/min

Blos bead 1 to 20 mL/min

Temperature Default is 250 C; operating range is 150 C to 400 C. <150 C, the Adjust offset process will not start. 325 to 335 C is recommended. Detector temperature should be greater than the highest oven temperature.

With higher detector temperatures, less bead heating voltage is required.

Adjust offset Default is 30 pA, suggested operating range is 20 to 40 pA, and allowable range is 0 to 99.9 pA. 50 pA increases sensitivity but reduces bead life. Lower settings reduce sensitivity and increase bead life, but settings too low

will result in solvent quenching. The time required for Adjust offset depends on the bead type and condition.

Bead voltage Ceramic bead. Range is 0 to 4.095 V. Blos bead. Range is 0.5 to 1.1 V. Use Auto Adjust On, Dry Bead, and let the GC set the Bead Voltage for you.

Table 49 General operating values (continued)

Gas or Setting Recommendation

Detectors 5

Advanced Operation Manual 191

Temperature programming

The NPD is flow sensitive. If you are using temperature programming, in which the column flow resistance changes with temperature, set up the instrument as follows:

Set the carrier gas in the Constant flow mode. Set detector makeup gas to Constant makeup.

If you choose to work in the constant pressure mode, the makeup gas should be set in the Column +makeup=constant mode.

NPD required gas purity

Because of its high sensitivity, the NPD requires very pure (at least 99.9995%) gases. We strongly recommend that moisture and organics traps be used on the carrier gas and

Flow, mL/min

Flow, mL/min

Pressure (psig) (kPa)

Pressure (psig) (kPa)

50

100

150

5

4

3

2

1

Air

Helium

Nitrogen

Hydrogen

10 69

20 138

30 207

40 276

50 345

60 414

70 483

20 138

16 110

12 83

8 55

4 28

192 Advanced Operation Manual

5 Detectors

all detector gases, including the detector hydrogen, air, and makeup gases. Do not use plastic (including PTFE) tubing, plastic- bodied traps, or O- ring seals.

Setting parameters for the NPD

Before operating the NPD, make sure that detector gases are connected, a column is installed, and the system is free of leaks. Set the oven temperature, inlet temperature, and column flow.

1 Select the bead (white ceramic, black ceramic, Blos).

2 Select the jet.

3 Install bead and jet as required. (See the Maintenance Manual for details.)

4 Press [Config][Front Det] or [Config][Back Det].

5 If you are using makeup gas, verify that the configured makeup gas type is the same as that plumbed to your instrument. Change the gas type, if necessary. Nitrogen is recommended.

6 If the displayed Bead Type is incorrect, set the Bead Type using the [Mode/Type] key.

7 Set Auto Adjust (On recommended).

8 Set Dry Bead (On recommended).

9 Press [Front Det] or [Back Det].

10 Set the detector temperature. The recommended range is 325 to 335 C.

11 Enter a hydrogen flow (3.0 mL/min is recommended). Turn the flow On.

12 Enter an air flow (60 is recommended for ceramic beads, 120 for Blos beads). Turn the flow On.

a If you are using packed columns, turn off makeup gas and proceed to step 13.

b If your capillary column is defined, choose a flow mode and set the makeup gas flow. For a column in the constant flow mode, choose Constant makeup. For a

WARNING Make sure that a column is installed or the NPD column fitting is plugged before turning on the air or hydrogen. An explosion may occur if air and hydrogen are allowed to leak into the oven.

Detectors 5

Advanced Operation Manual 193

column in the constant pressure mode, choose Column +makeup=constant.

c If your column is not defined, enter a makeup gas flow. Only constant flow is available.

13 Monitor the offset adjustment process.

a If Auto Adjust is On, the adjust offset process starts automatically when the detector reaches setpoint. If Auto Adjust is Off, the Bead Voltage will gradually go to the last setpoint after the bead reaches setpoint temperature and the Dry Bead time has elapsed.

b If you need to set a new target offset, enter an Adjust offset value. Adjust offset starts when the detector reaches setpoint.

c If Auto Adjust is Off, you can manually start the Adjust offset process by scrolling to Adjust offset, then pressing [On/Yes].

d If your standard operating procedures require that you set the bead voltage directly, see Setting NPD bead voltage manually (optional) on page 199.

Selecting an NPD bead type

Three beads are available:

Table 50 NPD beads

Bead type Part number Advantages Disadvantages

White ceramic G1534-60570 Standard Phosphorus tails

Black ceramic 5183-2007 Durable, no phosphorus tailing

Lower nitrogen sensitivity, about 40%

Blos bead G3434-60806 Moisture resistant, long life

Ceramic beads Blos bead

194 Advanced Operation Manual

5 Detectors

Changing from a ceramic bead to a Blos bead

When you turn the NPD off, the GC remembers the bead voltage used and applies that voltage when the detector is turned back on. If you have changed from a ceramic bead to a Blos bead, this voltage will be too high and may damage the new bead.

To avoid this, reduce the voltage to the ceramic bead to 0.5 V before turning the detector off. When it restarts, the lower voltage will be applied.

Selecting an NPD jet

Open the oven door and locate the column connection fitting at the base of the detector. It will look like either a capillary optimized fitting or an adaptable fitting.

If you have an application that tends to clog the jet, select a jet with a wider tip id.

When using packed columns in high column- bleed applications, the jet tends to clog with silicon dioxide.

For capillary optimized fittings, select one of the following from Table 51.

CAUTION The Blos bead is more delicate than the ceramic beads, and may be distorted during shipping. Before installing a Blos bead, verify that it is centered and adjust it if necessary.

Capillary optimized fitting Adaptable fitting

Detector fitting

Adapter

Table 51 Jets for capillary optimized fittings

Figure 3 ID Jet type Part number Jet tip id Length

1 Capillary with extended jet (recommended)

G1534-80580 0.29 mm (0.011 inch) 51.5 mm

2 Capillary G1531-80560 0.29 mm (0.011 inch) 43 mm

3 High-temperature G1531-80620 0.47 mm (0.018 inch) 43 mm

Detectors 5

Advanced Operation Manual 195

For the adjustable NPD, select one of the following from Table 52.

To configure the NPD

In addition to the Ignore Ready and Makeup gas type, the NPD requires the following configuration settings. Scroll to each and enable/disable using [On/Yes] or [Off/No].

Figure 1 Capillary optimized NPD jets

1

2

3 1 2 3

43 mm

51.5 mm

43 mm

Table 52 Jets for adaptable fittings

Figure 4 ID Jet type Part number Jet tip id Length

1 Capillary with extended jet (recommended)

G1534-80590 0.29 mm (0.11 inch) 70.5 mm

2 Capillary 19244-80560 0.29 mm (0.011 inch) 61.5 mm

3 Capillary, high-temperature 19244-80620 0.47 mm (0.018 inch) 61.5 mm

4 Packed 18710-20119 0.46 mm (0.018 inch) 63.6 mm

Figure 2 Adaptable NPD jets

1

2

3

4 1 2 3 4

70.5 mm

61.5 mm

61.5 mm

63.6 mm

196 Advanced Operation Manual

5 Detectors

Auto Adjust Bead Recommended On. When On, the automatic adjust offset process starts when the bead reaches the temperature setpoint after having been turned off or cooled below 150 C. Auto adjust starts after Dry Bead hold time, if enabled. Auto Adjust Bead uses the adjust offset feature to protect the beadespecially new beadsby making sure that the desired offset is obtained with the lowest possible bead voltage. When Off, the bead voltage will rise as soon as the Dry Bead time elapses, or as soon as the temperature setpoint is reached if Dry Bead is off.

Dry Bead Recommended On. When On, the bead temperature holds at 150 C for 5 minutes before continuing to the setpoint. This allows any condensation to evaporate and be swept out of the detector.

Blos Bead If using a Blos bead, set to On. For ceramic beads, set to Off. The Blos Bead On setting restricts the allowable bead voltage to the range appropriate for this bead.

Maximum Bead Voltage Display only. Shows the current maximum bead voltage for the configured bead type (4.095 V for ceramic beads, 1.1 V for the Blos bead).

Automatically adjusting NPD bead voltage

Agilent recommends using the Adjust offset feature to automatically determine the lowest bead voltage needed to give the desired response.

When the detector is turned on, the temperature rises at a controlled rate.

If Dry Bead is On, temperature holds at 150 C for 5 minutes to drive off moisture, then continues to the setpoint.

If Dry Bead is Off, the temperature rises directly to the setpoint.

When the temperature reaches the setpoint, the Bead Voltage gradually rises until it produces the desired output.

Adjust offset

When you enter a value here, or press [On/Yes] to use the stored value, detector gas flows turn on, the bead heats, and the bead voltage adjusts until Output is stable and equal to the entered value. There are five stages of Adjust offset.

Detectors 5

Advanced Operation Manual 197

Detector off When the detector is off, Adjust offset and Bead voltage are Off and initial Output is displayed.

Detector ondetector temperature less than 150 C. When you enter an Adjust offset value or press [On], detector gases turn on and the display blinks the Temp not ready message.

Detector onwaiting for oven and/or detector to reach temperature setpoint and equilibrium. If the oven or detector is not at setpoint, the display continues to blink the Temp not ready message.

Detector onDry Bead On If Dry Bead is On, the temperature rise holds at 150 C for 5 minutes to remove moisture, then continues to the setpoint.

Detector onduring adjust offset. When the detector and oven temperatures reach setpoint and equilibrate, the Adjust offset process begins. The bead voltage is slowly increased until the output is close to the Adjust offset value. The display blinks Detector Slewing.

Detector on and ready. When the Adjust offset value is reached, the Adjust offset line reads Done and displays the offset target setpoint. Your detector is on and ready. The display shows the actual Bead voltage.

Setting NPD adjust offset on the clock table

You can use the Clock table feature to begin Adjust offset at a specified time.

Aborting NPD adjust offset

Press [Delete] with the cursor on the Adjust offset line. This cancels the adjustment without turning off the detector gases and bead voltage.

Extending the NPD bead life

These actions, together with the automated heatup and adjust procedures, can extend ceramic bead life considerably.

Use the lowest practical Adjust offset value. This will result in a lower Bead Voltage during operation.

Run clean samples.

Turn the bead off when not in use.

198 Advanced Operation Manual

5 Detectors

Keep the detector temperature high (320 to 335 C).

Turn the hydrogen flow off during solvent peaks and between runs.

Turning hydrogen off during a solvent peak

When using the NPD, the baseline shifts after a solvent peak and can take some time to stabilize, especially with chlorinated solvents. To minimize this effect, turn off the hydrogen flow during the solvent peak and turn it back on after the solvent elutes. With this technique, the baseline recovers to its original value in less than 30 seconds. This also extends the life of the bead. The hydrogen can be turned on and off automatically as part of a Run Table. See Run Time Programming on page 12.

Turning hydrogen off between runs

To extend bead life, turn off the hydrogen flow between runs. Leave all other flows and the detector temperature on. Turn on the hydrogen flow for the next run; the bead will ignite almost immediately. The process can be automated with Run Table entries.

Turning off the detector

Setting the initial bead voltage for new beads

Before you turn on the bead for the first time, manually set its voltage to a safe value so that the new bead is not destroyed.

1 Make sure Adjust Offset is turned Off.

2 After the temperature stabilizes at setpoint, set the initial Bead Voltage, depending on bead type:

Blos bead: 0.0 V to 0.5 V

Ceramic bead (white or black): 0.0 V to 2.0 V

CAUTION If you turn Adjust offset [Off] at any time, the bead voltage, hydrogen, and air flows all turn off.

Detectors 5

Advanced Operation Manual 199

Setting NPD bead voltage manually (optional)

Bead voltage shows the voltage used to heat the bead. It can be a value derived from the Adjust offset value, or can be entered as a setpoint. Entering a setpoint causes the voltage to change at 13 mV/second until it reaches the setpoint provided that

the detector is at the temperature setpoint

temperature is at least 150 C

gas flows are on

Dry Bead time, if On, has elapsed

Bead voltage is also useful for small adjustments between runs. If you observe a baseline drift, you can enter a small, one- time change to compensate for the drift.

If you are not using the recommended Adjust offset process, note that large voltage jumps reduce bead life. Use increments no greater than 0.05 V, spaced 10 seconds apart, until you reach the desired offset.

New beads

After a new bead reaches the initial voltage, begin to increase the voltage value in 0.05 V increments until the bead ignites. Wait about 10 seconds between each voltage adjustment. Monitor the detector output. When the bead ignites, the output will rise suddenly, then decrease towards a more stable value. It is best to allow the NPD to remain in this state without further adjustment for about 24 hours. Then you may adjust the bead voltage in small increments (0.05 to 0.1 V) until reaching the desired offset. With a clean environment, clean gas supplies, and low bleed column, a typical offset may decrease 6- 12 pA during a 24 hour period.

Typical voltages for new ceramic beads range from 2.5 to 3.7 volts. Higher values reduce bead life.

Typical voltages for new Blos beads range from 0.5 to 1.0 V.

200 Advanced Operation Manual

5 Detectors

About the FPD+

The sample burns in a hydrogen- rich flame, where some species are reduced and excited. The gas flow moves the excited species to a cooler emission zone above the flame where they decay and emit light. A narrow bandpass filter selects light unique to one species, while a shield prevents intense carbon emission from reaching the photomultiplier tube (PMT).

The light strikes a photosensitive surface in the PMT where a light photon knocks loose an electron. The electron is amplified inside the PMT for an overall gain of up to a million.

The current from the PMT is amplified and digitized by the FPD+ electronics board. The signal is available either as a digital signal on the communications output or as a voltage signal on the analog output.

The FPD+ should not be stored at temperatures above 50 C, based on the original manufacturers specifications for the PMT.

Valve

Frit Frit

Valve

PS = Pressure Sensor Column

PS

PS PS

Valve

Frit

Restrictor Restrictor

Restrictor

EPC Module

H2 Air

Wavelength filter

Window

Vent

Makeup

Emission Zone

Transfer line Restrictor

Detectors 5

Advanced Operation Manual 201

FPD+ linearity

Several mechanisms produce sulfur emission. The excited species is diatomic, so that emission intensity is approximately proportional to the square of the sulfur atom concentration.

The excited species in the phosphorus mode is monatomic, leading to a linear relationship between emission intensity and atom concentration.

FPD+ Lit Offset

The default Lit Offset is 2.0 pA.

FPD+ photomultiplier protection

The PMT is extremely sensitive to light. Always turn the PMT voltage off (which turns off the high voltage to the PMT) before removing the PMT housing or opening the emissions chamber. Failing to do this can destroy the PMT.

Even with the PMT voltage off, protect the PMT from room light. Cap the housing when removed, place it end down to exclude light, reduce room light level before exposing the PMT, and so on. A brief exposure (always with the PMT voltage turned off) will not damage it but prolonged exposure will cause a gradual loss of sensitivity.

FPD+ optical filters

The filters are marked on the edge with the transmission wavelength. Each filter has a small arrow or triangle on its side which must point toward the PMT when installed.

The sulfur filter is silvery on both sides and transmits at 393 nanometers.

The phosphorus filter is yellow/green and transmits at 525 nanometers.

Inlet liners for use with the FPD+

Compounds containing sulfur may adsorb on an inlet liner and degrade the GCs performance. Use deactivated, clean liners or a cool on- column inlet, which injects directly onto the column.

For best results with splitless injection, use liner 5181- 3316.

202 Advanced Operation Manual

5 Detectors

FPD+ temperature considerations

The FPD+ provides two temperatures zones, one for the transfer line (the main detector temperature) and one for the emission block. For the transfer line temperature, we recommend a temperature that is 25 C higher than the highest column temperature.

The emission block temperature ranges from 125175 C. Typically, the default temperature of 150 C is sufficient for most applications. However, when setting the emission block temperature, consider the following:

If using the GC oven at high temperature (>325 C) with the transfer line set to 400 C, set the emission block temperature to 165 C to avoid a system Not Ready if the emission block temperature cannot be maintained.

If using the transfer line at 400 C, set the emission block temperature to at least 150 C to avoid a system Not Ready.

For sulfur analyses, the highest area response will be realized with the lowest possible emission block temperature.

For phosphorus analyses, the area response is independent of the emission block temperature.

FPD+ gas purity

High- purity gases have a lower sulfur content. Standard purity gases have a higher sulfur content which impairs sulfur detection in the compound being studied. Instrument or Chromatographic grades work well.

Agilent recommends using helium carrier, nitrogen makeup gas, and air with 99.9995% purity or better. Use carbon, oxygen, and moisture traps. Select traps to remove sulfur compounds from detector air and nitrogen gases. A helium getter is also recommended. See the GC, GC/MS, and ALS Site Preparation Guide for and the Agilent web site at http://www.agilent.com more information.

FPD+ gas flows

Table 53 gives the flows for the maximum sensitivity FPD+ flame, which is hydrogen- rich and oxygen- poor.

Detectors 5

Advanced Operation Manual 203

Helium, either as carrier or makeup gas, may cool the detector gases below the ignition temperature. We recommend using nitrogen rather than helium.

Lighting the FPD+ flame

Before trying to light the flame, heat the detector to operating temperature and let it stabilize.

During the ignition sequence, the detector does the following:

1 Until the detector temperature zones reach setpoint, the GC purges the detector with all gas flows.

2 Once the detector stabilizes at its thermal setpoints, the GC turns off hydrogen and makeup gas flows. Carrier and air flows remain on.

3 Sets air flow to 500 mL/min.

4 Turns the glow plug ignitor on.

5 Ramps the hydrogen flow from 10 to 70 mL/min.

6 Resets the air flow to the air flow setpoint.

7 Resets the hydrogen flow to the hydrogen flow setpoint.

8 Turns the makeup gas on.

9 Compares the signal change with the Lit offset value. If the change is greater than Lit offset, declares the flame on (lit). If it is less, declares the flame off (not lit).

For this process to work, there must be enough air pressure to the pneumatics module to provide up to 500 mL/min flow. We recommend a supply pressure of at least 60 psi. See the Installation and First Startup manual.

Table 53 Recommended flows

mL/min

Carrier (hydrogen, helium, nitrogen, argon)

Packed columns 10 to 60

Capillary columns 1 to 5

Detector gases

Hydrogen 60

Air 60

Carrier + makeup 60

204 Advanced Operation Manual

5 Detectors

Manual ignition

1 Press [Front Det] or [Back Det].

2 Scroll to Flame. Press [On/Yes]. The flame ignition sequence begins.

Automatic ignition

If the FPD+ output with the flame on falls below the flame- off output plus the Lit offset value, this is interpreted as a flame- out condition. The FPD+ runs the flame ignition sequence to relight the flame. If this fails, it runs the sequence again. If the second attempt also fails, the detector shuts down all functions except temperature and makeup gas flow.

Setting parameters for the FPD+

1 Press [Front det] or [Back det].

2 Set the detector temperature. Scroll to Temperature and input the desired value, and press [Enter]. It must be greater than 120 C for the flame to light.

3 Scroll to H2 flow and change the hydrogen flow rate, if desired. Press [Enter]. If the flame is not lit, press [Off/No] to turn off the flow. (It is not good practice to pass uncombusted hydrogen into the air. The hydrogen flow will be turned on automatically when igniting the flame.)

4 Scroll to Air flow and change the air flow rate, if desired. Press [Off/No] to turn off the flow for now.

5 If you are using packed columns, turn off the makeup gas and proceed to step 9.

6 If you are using capillary columns, scroll to the Makeup parameter:

a Verify that makeup gas type is the same as that plumbed to your instrument. The configured gas type is

WARNING Verify that a column is installed or the FPD+ fitting is plugged before turning on the air or hydrogen. An explosion may occur if air and hydrogen are allowed to leak into the oven.

Detectors 5

Advanced Operation Manual 205

shown next to Makeup in the parameter list. Change the gas type, if necessary.

b If your capillary column is defined, choose a flow mode and set the makeup gas flow or combined flow.

c If your capillary column is not defined, enter a makeup gas flow. Only constant flow is available.

7 Scroll to PMT voltage and make sure it is turned on.

8 Optionally, set the emission block temperature. (Normally, the default temperature of 150 C is sufficient for most applications.) Scroll to Emission Block, input the desired temperature, and press [Enter]. See FPD+ temperature considerations on page 202 for more information.

9 Scroll to Flame and press [On/Yes]. This starts the automatic ignition sequence (see Lighting the FPD+ flame on page 203).

On successful ignition, the signal increases. Typical signal output increases are 4 to 40 pA in sulfur mode, and 10 to 70 pA in phosphorus mode. You can also verify that the flame is lit by holding a cold, shiny surface, such as a mirror or chrome- plated wrench, over the vent exit. Steady condensation indicates that the flame is lit.

206 Advanced Operation Manual

5 Detectors

About the FPD

The sample burns in a hydrogen- rich flame, where some species are reduced and excited. The gas flow moves the excited species to a cooler emission zone above the flame where they decay and emit light. A narrow bandpass filter selects light unique to one species, while a shield prevents intense carbon emission from reaching the photomultiplier tube (PMT).

The light strikes a photosensitive surface in the PMT where a light photon knocks loose an electron. The electron is amplified inside the PMT for an overall gain of up to a million.

The current from the PMT is amplified and digitized by the FPD electronics board. The signal is available either as a digital signal on the communications output or as a voltage signal on the analog output.

The FPD should not be stored at temperatures above 50 C, based on the original manufacturers specifications for the PMT.

Valve

Frit Frit

Valve

PS = Pressure Sensor Column

PS

PS PS

Valve

Frit

Restrictor Restrictor

Restrictor

EPC Module

H2 Air

Wavelength filter

Window

Vent

Makeup

Emission Zone

Detectors 5

Advanced Operation Manual 207

FPD linearity

Several mechanisms produce sulfur emission. The excited species is diatomic, so that emission intensity is approximately proportional to the square of the sulfur atom concentration.

The excited species in the phosphorus mode is monatomic, leading to a linear relationship between emission intensity and atom concentration.

FPD Lit Offset

The default Lit Offset is 2.0 pA.

Starting Up and Shutting Down the FPD

The FPD creates a great deal of water vapor when the flame is on. This could condense in the vent tube on top of the detector and drop onto the flame, possibly extinguishing it. To avoid this, turn the heaters on, wait 20 minutes for the vent to heat up, and then ignite the flame. Water vapor will now make it over the top of the vent tube before condensing.

For similar reasons, extinguish the flame before turning the heaters off.

FPD photomultiplier protection

The PMT is extremely sensitive to light. Always turn the PMT voltage off (which turns off the high voltage to the PMT) before removing the PMT housing or opening the emissions chamber. Failing to do this can destroy the PMT.

Even with the PMT voltage off, protect the PMT from room light. Cap the housing when removed, place it end down to exclude light, reduce room light level before exposing the PMT, and so on. A brief exposure (always with the PMT voltage turned off) will not damage it but prolonged exposure will cause a gradual loss of sensitivity.

FPD optical filters

The filters are marked on the edge with the transmission wavelength. Each filter has a small arrow on its side which must point toward the PMT when installed.

The sulfur filter is silvery on both sides and transmits at 393 nanometers.

208 Advanced Operation Manual

5 Detectors

The phosphorus filter is yellow/green and transmits at 525 nanometers.

Inlet liners for use with the FPD

Compounds containing sulfur may adsorb on an inlet liner and degrade the GCs performance. Use deactivated, clean liners or a cool on- column inlet, which injects directly onto the column.

For best results with splitless injection, use liner 5181- 3316.

FPD temperature considerations

The minimum detector temperature to prevent water condensation is 120 C. We recommend a temperature that is 25 C higher than the highest column temperature, but no higher than 250 C.

FPD gas purity

High- purity gases have a lower sulfur content. Standard purity gases have a higher sulfur content which impairs sulfur detection in the compound being studied. Instrument or Chromatographic grades work well.

Agilent recommends using helium carrier, nitrogen makeup gas, and air with 99.9995% purity or better. Use carbon, oxygen, and moisture traps. Select traps to remove sulfur compounds from detector air and nitrogen gases. A helium getter is also recommended. See the GC, GC/MS, and ALS Site Preparation Guide for and the Agilent web site at http://www.agilent.com more information.

FPD gas flows

Table 53 gives the flows for the maximum sensitivity FPD flame, which is hydrogen- rich and oxygen- poor.

Table 54 Recommended flows

Sulfur mode flows, mL/min

Phosphorus mode flows, mL/min

Carrier (hydrogen, helium, nitrogen, argon)

Packed columns 10 to 60 10 to 60

Capillary columns 1 to 5 1 to 5

Detectors 5

Advanced Operation Manual 209

Helium, either as carrier or makeup gas, may cool the detector gases below the ignition temperature. We recommend using nitrogen rather than helium.

Lighting the FPD flame

Before trying to light the flame, have the detector at operating temperature. Removing the condensate tubing may help, but be sure to replace it before making runs.

It is difficult to light the flame with the flows shown in Table 53, particularly in the sulfur mode. If the flame will not light with the sulfur mode flows shown, change to the phosphorus mode flows. After ignition, gradually reduce the flows to the sulfur values. Some experimentation will be needed.

When either of the flame ignition methods in this section is used, the FPD automatically performs this sequence:

1 Turns all detector gasesair, hydrogen, makeupoff. Carrier remains on.

2 Sets air flow to 200 mL/min.

3 Turns the glow plug ignitor on.

4 Ramps the hydrogen flow from 10 to 70 mL/min.

5 Resets the air flow to the air flow setpoint.

6 Resets the hydrogen flow to the hydrogen flow setpoint.

7 Turns the makeup gas on.

8 Compares the signal change with the Lit offset value. If the change is greater than Lit offset, declares the flame on (lit). If it is less, declares the flame off (not lit).

For this process to work, there must be enough air pressure to the pneumatics module to provide 200 mL/min flow. We recommend a supply pressure of 90 psi.

Detector gases

Hydrogen 50 75

Air 60 100

Carrier + makeup 60 60

Table 54 Recommended flows (continued)

Sulfur mode flows, mL/min

Phosphorus mode flows, mL/min

210 Advanced Operation Manual

5 Detectors

Manual ignition

1 Press [Front Det] or [Back Det].

2 Scroll to Flame. Press [On/Yes]. The flame ignition sequence begins.

Automatic ignition

If the FPD output with the flame on falls below the flame- off output plus the Lit offset value, this is interpreted as a flame- out condition. The FPD runs the flame ignition sequence to relight the flame. If this fails, it runs the sequence again. If the second attempt also fails, the detector shuts down all functions except temperature and makeup gas flow.

Setting parameters for the FPD

1 Press [Front det] or [Back det].

2 Set the detector temperature. It must be greater than 120 C for the flame to light.

3 Change the hydrogen flow rate, if desired. Press [Off/No].

4 Change the air flow rate, if desired. Press [Off/No].

5 If you are using packed columns, turn off the makeup gas and proceed to step 9.

6 If you are using capillary columns:

a Verify that makeup gas type is the same as that plumbed to your instrument (next to Makeup in the parameter list). Change the gas type, if necessary.

b If your capillary column is defined, choose a flow mode and set the makeup gas flow or combined flow.

c If your capillary column is not defined, enter a makeup gas flow. Only constant flow is available.

7 Scroll to Flame and press [On/Yes]. This turns on the air and hydrogen and initiates the ignition sequence.

On ignition, the signal increases. Typical levels are 4 to 40 pA in sulfur mode, 10 to 70 pA in phosphorus mode. Verify that the flame is lit by holding a cold, shiny

WARNING Verify that a column is installed or the FPD fitting is plugged before turning on the air or hydrogen. An explosion may occur if air and hydrogen are allowed to leak into the oven.

Detectors 5

Advanced Operation Manual 211

surface, such as a mirror or chrome- plated wrench, over the vent exit. Steady condensation indicates that the flame is lit.

212 Advanced Operation Manual

5 Detectors

213

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

6 Valves

About Valves 214

The Valve Box 215

Heating the valves 215

Valve temperature programming 216

Configuring an Aux thermal zone 216

Valve Control 217

The valve drivers 217

The internal valve drivers 217

The external valve drivers 218

Valve Types 219

Configuring a Valve 220

Controlling a Valve 221

From the keyboard 221

From the run or clock time tables 221

Simple valve: column selection 221

Gas sampling valve 222

Multiposition stream selection valve with sampling valve 223

214 Advanced Operation Manual

6 Valves

About Valves

Valves may be used to alter the usual inlet/column/detector flow path in the GC. Sampling valves can replace the inlet, switching valves can select columns, multiposition valves, used in conjunction with sampling valves, can perform the same functions for sample streams that an ALS performs for liquid samples.

Valves 6

Advanced Operation Manual 215

The Valve Box

The GC holds up to four valves in a heated valve box on top of the oven.

The valve box is the preferred location for valves because it is a stable temperature zone, isolated from the column oven.

Valves may also be mounted inside the column oven.

Heating the valves

The valve box contains two heated blocks, each with two valve mounting locations (shaded in Figure 3). The middle hole on each block is used to pass tubing into the column oven.

If two valves are used, mount them on the same block. This allows them to be heated using only one control channel (Aux Temp 1 or Aux Temp 2, depending on how the heaters are wired). With more than two valves, both channels must be used for heating the two blocks. Set them at the same temperature.

Figure 3 Diagram of valve locations on GC

Valve heater blocks

Valve box, cover removed

Back of chromatograph

216 Advanced Operation Manual

6 Valves

Valve temperature programming

Most valve applications are isothermal; however, you can define three temperature ramps if desired. Press [Aux Temp #], then [1] or [2]. Program this ramp the same as an oven ramp. Refer to Setting the oven parameters for ramped temperature on page 152 for more information.

Configuring an Aux thermal zone

To configure a thermal Aux zone (1 or 2), press [Config][Aux Temp #], then [1] or [2]. Press [Mode/Type], then select the type of device to be controlled. Press [Enter].

Valves 6

Advanced Operation Manual 217

Valve Control

Valves can be controlled manually from the keyboard or as part of a clock or run time program. Note that only sampling valves automatically reset at the end of a run. Other valve types remain at the new position until activated again. For other valve types, you must include any desired resets in the program.

The valve drivers

A valve driver is the software and circuitry in the GC that controls a valve or related function. There are eight drivers known as Valve 1 through Valve 8.

The internal valve drivers

Valve drivers 1 through 4 are usually used to control pneumatically operated valves mounted in the valve box. The wiring for these appears at a set of connectors inside the right cover of the GC.

Pneumatically driven valves are controlled by solenoids mounted near the connectors that control the flow of air to the valve actuators.

Table 55 Valve drivers

Valve number Type Volts Power or current Use

1, 2, 3, and 4 Current source 24 VDC 13 watts Pneumatic valve control

5 and 6 Current source 24 VDC 100 mA Relays and low-power devices

7 and 8 Contact closure 48 VDC or 48 VAC RMS

Control an external current source

218 Advanced Operation Manual

6 Valves

There is no direct relationship between the location of a valve in the valve box and the driver that controls it. This depends on how the solenoids are wired and the actuators are plumbed.

Manual valves must be switch by hand, and are heated or unheated.

The external valve drivers

Valve drivers 5 and 6 control a current that may be used to drive a relay or other low- power device. Valve drivers 7 and 8 switch a current from an external source. Electrical details are in Table 55 on page 217.

These drivers, particularly Valve 7 and 8, may be used to control a motor driven multiposition valve for stream selection.

All four of these drivers appear on the External Event connector on the back of the GC.

Internal valve drivers (1 through 4)

Keyboard

Run table

Clock table

or

or

Connector V1

Connector V2

Connector V3

Connector V4

Keyboard

Run table

Clock table

or

or

External valve drivers (5 through 8)

Valve 5 (pin 1) and ground (pin 3 or 4)

Valve 6 (pin 2) and ground (pin 3 or 4)

Valve 7 (pin 5 and pin 6)

Valve 8 (pin 7 and pin 8)

Valves 6

Advanced Operation Manual 219

Valve Types

The possible valve types are:

Sampling A two- position (load and inject) valve. In load position, an external sample stream flows through an attached (gas sampling) or internal (liquid sampling) loop and out to waste. In inject position, the filled sampling loop is inserted into the carrier gas stream. When the valve switches from Load to Inject, it starts a run if one is not already in progress. See the example on page 222.

Switching A two- position valve with four, six, or more ports. These are general- purpose valves used for such tasks as column selection, column isolation, and many others. For an example of valve control, see page 223.

Multiposition Also called a stream selection valve. It is usually used to select one from a number of gas streams and feed it to a sampling valve for analysis. It has a special actuator that advances the valve one position each time it is activated, or it may be motor driven. An example that combines a stream selection valve with a gas sampling valve is on page 223.

Other Could be anything.

Not installed Self- explanatory.

220 Advanced Operation Manual

6 Valves

Configuring a Valve

1 Press [Config]. Scroll to Valve #.

2 Enter the valve number and press [Enter]. The current valve type is displayed.

3 To change the valve type, press [Mode/Type], select the new valve type, and press [Enter].

Valves 6

Advanced Operation Manual 221

Controlling a Valve

From the keyboard

Valves (except multiposition valves) have two positions controlled by the [On] and [Off] keys. The keyboard commands for two- position valves are:

[Valve #] [On] Rotates valve to one stop

and

[Valve #] [Off] Rotates valve to the other stop

From the run or clock time tables

The Valve On and Valve Off commands can be run time or clock time programmed. See Run Time Programming on page 12 and Clock Time Programming on page 15.

If a valve is rotated by a run time program, it is not automatically returned to its initial position at the end of the run. You must program this reset operation yourself.

Simple valve: column selection

This is the plumbing for a single valve, configured as a switching valve, that selects one of two columns for analysis. It has no configuration parameters.

The column is selected by pressing:

[Valve #] [On] (for the front column) or [Off] (for the back column).

Front column

Back column

To detector

From inlet or sampling valve

ON OFF

222 Advanced Operation Manual

6 Valves

Use a run table entry to ensure that the valve is in the Off state between runs.

Gas sampling valve

If a valve is configured as a gas sampling valve, it starts a run automatically when it is switched to the Inject position. This can be done with a keyboard command or by a subsequence or clock table entry. You may have two gas sampling valves installed.

Load position The loop is flushed with a stream of the sample gas. The column is flushed with carrier gas.

Inject position The filled loop is inserted into the carrier gas stream. The sample is flushed onto the column. The run starts automatically.

Carrier gas may be provided by an (optional) auxiliary gas channel. To do this, configure the column and specify an Aux # channel as the inlet. The channel then becomes programmable with four operating modes.

Sampling valves have two positions:

Load position The loop (external for gas sampling, internal for liquid sampling) is flushed with a stream of the sample. The column is flushed with carrier gas.

Inject position The filled loop is inserted into the carrier gas stream. The sample is flushed onto the column. The run starts automatically.

The sampling valve control parameters are:

Carrier in

To column

Sample in

Sample out

LOAD position INJECT position

Loop

Valves 6

Advanced Operation Manual 223

Load time Time in minutes that the valve remains in the Load position before becoming ready.

Inject time Time in minutes that the valve remains in the Inject position before returning to the Load position.

The sampling valve cycle is:

1 The sampling valve rotates to the Load position. Load time begins. Valve is not ready.

2 Load time ends. The valve becomes ready.

3 If everything else is ready, the GC becomes ready. If anything is not ready:

If you are using Clock Table or sequence control, the GC waits until everything is ready, then executes the valve inject command.

If you are not using Clock Table or sequence control, the valve injection can be made at any time from the keyboard.

4 The sampling valve rotates (keyboard command or sequence control) to the Inject position. Inject time begins. The run begins.

5 Inject time ends. Return to step 1.

Multiposition stream selection valve with sampling valve

Several manufacturers provide multiposition stream selection valves that can be driven by valve drivers 1 through 4. Only one multiposition valve can be configured.

If a valve is configured as a multiposition valve and has a BCD position output connected to the GC, the valve position can be selected directly.

224 Advanced Operation Manual

6 Valves

If the GC has one valve configured as a multiposition valve and another configured as a gas or liquid sampling valve, it assumes that they are to be used in series. This double configuration can be used to replace an automatic liquid sampler and sample tray in an analytical sequence. The multiposition valve becomes the sample tray; the sampling valve becomes the injector.

Two configuration parameters provide mechanical and electrical compatibility with most multiposition valve actuators.

Switching time In seconds, is a delay between successive actuator movements. It allows time for the actuator mechanism to prepare for the next movement.

Invert BCD Complements the BCD input; 1s become 0s and 0s become 1s. This accommodates coding convention differences among manufacturers.

Sample streams in

Selected stream out

Multiposition stream selection valve

Sampling valve

225

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

7 GC Output Signals

About Signals 226

Signal Types 227

Value 227

Analog Signals 229

Analog zero 229

Analog range 229

Analog data rates 230

Selecting fast peaks (analog output) 231

Digital Signals 232

Digital zero 232

Signal Freeze and Resume 232

Data rates with Agilent data systems 233

Column Compensation 236

Creating a column compensation profile 237

Making a run using analog output column compensation 237

Plotting a stored column compensation profile 238

Test Plot 239

226 Advanced Operation Manual

7 GC Output Signals

About Signals

Signal is the GC output to a data handling device, analog or digital. It can be a detector output or the output from flow, temperature, or pressure sensors. Two signal output channels are provided.

Signal output can be either analog or digital, depending on your data handling device. Analog output is available at either of two speeds, suitable to peaks with minimum widths of 0.004 minutes (fast data rate) or 0.01 minutes (normal rate). Analog output ranges are 0 to 1 V, 0 to 10 V, and 0 to 1 mV.

Digital output rates are set by your Agilent data system, such as ChemStation or EZChrom.

GC Output Signals 7

Advanced Operation Manual 227

Signal Types

When assigning detector signals, use the [Mode/Type] key and choose from the Signal Type parameter list, or press a key or combination of keys.

[Front], [Back], [], and [Column Comp] will work, alone or in combination. For example, press [Back] for back detector or [Back][][Front] for back detector minus front detector. The menu choices for signal subtraction (Front - Back and Back - Front) only appear if the front and back detectors are of the same type.

The nondetector signals are test plot, thermal, pneumatic, and diagnostic. Access them by pressing [Mode/Type]. Diagnostic signals are for use by your service representative and are not described in detail here.

Signal type can be programmed as a run time event.

Value

Value on the signal parameter list is the same as Output on the detector parameter list if your signal type is Front or Back. If you are subtracting one signal from another (as in Front Back), the signal Value will be the difference. You cannot enter a setpoint for Value.

A conversion factor may be involved when interpreting Valuefor example, one FID unit is one picoamp; one uECD unit is 1 Hz. The units for detector and other signals are listed below.

Table 56 Signal conversions

Signal type 1 display unit is equivalent to:

Detector:

FID, NPD 1.0 pA (1.0 10-12 A)

FPD 150 pA (150 10-12 A)

TCD 25 uV (2.5 10-5 V)

ECD 1 Hz

Analog input board (use to connect the GC to non-Agilent detector)

15 V

Nondetector: Thermal 1 C

228 Advanced Operation Manual

7 GC Output Signals

Pneumatic: Flow Pressure

Diagnostic

1 mL/min 1 pressure unit (psi, bar, or kPa) Mixed, some unscaled

Table 56 Signal conversions (continued)

Signal type 1 display unit is equivalent to:

GC Output Signals 7

Advanced Operation Manual 229

Analog Signals

If you use an analog recorder, you may need to adjust the signal to make it more usable. Zero and Range in the Signal parameter list do this.

Analog zero

Zero Subtracts value entered from baseline. Press [On/Yes] to set to current Value or [Off/No] to cancel.

This is used to correct baseline elevation or offsets. A common application is to correct a baseline shift that occurs as the result of a valve operation. After zeroing, the analog output signal is equal to the Value line of the parameter list minus the Zero setpoint.

Zero can be programmed as a run time event. For details, see Run Time Programming on page 12.

1 Verify that the detector is on and in a ready state.

2 Press [Analog Out 1] or [Analog Out 2].

3 Scroll to Zero.

4 Press [On/Yes] to set Zero at the current signal value, or Enter a number between - 500000 and +500000. A value smaller than the current Zero shifts baseline up.

Analog range

Range Scales data coming from the detector

Range is also referred to as gain, scaling, or sizing. It sizes the data coming from the detector to the analog signal circuits to avoid overloading the circuits (clamping). Range scales all analog signals.

If a chromatogram looks like A or B in the next figure, the data needs to be scaled (as in C) so that all peaks are visible on the paper.

Valid setpoints are from 0 to 13 and represent 20 (=1) to 213 (=8192). Changing a setpoint by 1 changes the height of the chromatogram by a factor of 2. The following chromatograms illustrate this. Use the smallest possible value to minimize integration error.

230 Advanced Operation Manual

7 GC Output Signals

There are limits to usable range settings for some detectors. The table lists the valid range setpoints by detector.

Range may be run time programmed. See Run Time Programming on page 12 for details.

Analog data rates

Your integrator or recorder must be fast enough to process data coming from the GC. If it cannot keep up with the GC, the data may be damaged. This usually shows up as broadened peaks and loss of resolution.

Speed is measured in terms of bandwidth. Your recorder or integrator should have a bandwidth twice that of the signal you are measuring.

The GC allows you to operate at two speeds. The faster speed allows minimum peak widths of 0.004 minutes (8 Hz bandwidth), while the standard speed allows minimum peak widths of 0.01 minutes (1.6 Hz bandwidth).

If you use the fast peaks feature, your integrator should operate at around 15 Hz.

C: Range = 1B: Range = 3A: Range = 0

Table 57 Range limits

Detector Usable range settings (2x)

FID 0 to 13

NPD 0 to 13

FPD 0 to 13

TCD 0 to 6

ECD 0 to 6

Analog input 0 to 7

GC Output Signals 7

Advanced Operation Manual 231

Selecting fast peaks (analog output)

1 Press [Config][Analog 1] or [Config][Analog 2].

2 Scroll to Fast peaks and press [On].

Agilent does not recommend using Fast peaks with a thermal conductivity detector. Since the gas streams switch at 5 Hz, the gain in peak width is offset by increased noise.

232 Advanced Operation Manual

7 GC Output Signals

Digital Signals

The GC outputs digital signals only to an Agilent data system. The following discussions describe features that impact the data sent to data systems, not the analog data available to integrators. Access these features from the data system. These features are not accessible from the GC keypad.

Digital zero

Available only from an Agilent data system.

Digital signal outputs respond to a zero command by subtracting the signal level at the time of the command from all future values.

Signal Freeze and Resume

Available only from an Agilent data system.

Some run time operations, such as changing signal assignments or switching a valve, can cause baseline upsets. Other factors can cause baseline upsets also. The GC can compensate for this by pausing (freezing) the signal at a particular value, using that signal value for a specified duration, and then resuming normal signal output.

Consider a system that uses a switching valve. When the valve switches, an anomaly occurs in the baseline. By freezing and resuming the signal, the anomaly can be removed so that the peak identification and integration software operates more smoothly.

GC Output Signals 7

Advanced Operation Manual 233

Data rates with Agilent data systems

The GC can process data at various data rates, each corresponding to a minimum peak width. The table shows the effect of data rate selection.

Baseline upset due to valve switch

Pause signal here

Resume signal here

Table 58 EZChrom/ChemStation data processing

Data rate, Hz Minimum peak width, minutes

Relative noise

Detector Column type

500 0.0001 5 FID Narrow-bore, 0.05 mm

200 0.001 3.1 FID Narrow-bore, 0.05 mm

100 0.002 2.2 FID/FPD/NPD only

Capillary

50 0.004 1.6

20 0.01 1

234 Advanced Operation Manual

7 GC Output Signals

You cannot change the data rate during a run.

You will see higher relative noise at the faster sampling rates. Doubling the data rate can double peak height while the relative noise increases by 40%. Although noise increases, the signal- to- noise ratio is better at the faster rates.

This benefit only occurs if the original rate was too low, leading to peak broadening and reduced resolution. We suggest that rates be chosen so that the product of data rate and peak width in seconds is about 10 to 20.

The figure shows the relationship between relative noise and data rates. Noise decreases as the data rate decreases until you get to data rates of around 5 Hz. As the sampling rate slows, other factors such as thermal noise increase noise levels.

10 0.02 0.7

5 0.04 0.5 to

2 0.1 0.3 All types

1 0.2 0.22

0.5 0.4 0.16

0.2 1.0 0.10

0.1 2.0 0.07 Slow packed

Table 58 EZChrom/ChemStation data processing (continued)

Data rate, Hz Minimum peak width, minutes

Relative noise

Detector Column type

GC Output Signals 7

Advanced Operation Manual 235

Zero Init Data Files

This feature applies to digital output only, and is mainly intended for non- Agilent data systems. It may help systems that have trouble with non- zero baseline output.

When you turn it On, the GC immediately begins to subtract the current detector output value(s) from any future values. For example, if you turn it on when the output is 20 pA, the GC subtracts 20 pA from the digital output until you turn Zero Init Data Files Off.

You will not see any change in the GC display, but you will see the change in the online plot available in the data system.

To change this setting, press [Config], scroll to Instrument, then scroll to Zero Init Data Files.

Relative noise level

Slower data ratesFaster data rates

Excess noise (due to flow, oven temperature, detector block temperatures, etc.)

236 Advanced Operation Manual

7 GC Output Signals

Column Compensation

In temperature programmed analysis, bleed from the column increases as the oven temperature rises. This causes a rising baseline which makes peak detection and integration more difficult. Column compensation corrects for this baseline rise.

A column compensation run is made with no sample injected. The GC collects an array of data points from all 4 detectors, whether installed, off, or working. If a detector is not installed or is turned off, that part of the array is filled with zeros.

One array (Column compensation 1) can be created for analog signals. Two independent arrays (Column compensation 1 and 2) can be created for digital signals.

Each array defines a set of curves, one for each detector, that can be subtracted from the real run to produce a flat baseline. The next figure illustrates the concept.

All conditions must be identical in the column compensation run and the real run. The same detector and column must be used, operating under the same temperature and gas flow conditions.

Chromatogram with rising baseline

Chromatogram with column compensation

Blank column compensation profile

GC Output Signals 7

Advanced Operation Manual 237

Creating a column compensation profile

1 Set up the instrument for a run.

2 Make a blank run to verify that the baseline is clean. This is particularly important for new conditions or if the GC has been idle for several hours.

3 Press [Column Comp].

4 Select Col comp 1 or Col comp 2 (these are the two arrays).

5 Select Start compensation run and press [Enter].

6 If the run is successful, the first line of the parameter list will say Data ok, and a time and date will appear at the bottom.

Making a run using analog output column compensation

1 Set the chromatographic conditions. They must be identical to those in the stored column compensation run except that Final time in the last ramp of the oven program can be longer or shorter.

2 Press [Analog Out 1] or [Analog Out 2].

3 Scroll to Type and press [Mode/Type].

4 The choices for an analog signal are:

Front detector Back detector Front - column comp 1 Front column compensation 1 Back column compensation 1 Aux 1 column compensation 1 Aux 2 column compensation 1 Test plot

5 Choose an option from the list.

6 Enter setpoints for Zero and Range, if applicable.

7 Start your run.

Making a run using digital output column compensation

1 Set each detector output separately.

2 Press [Config][Detector name]. Scroll to Signal and press [Mode/Type]. Select from:

No Column Compensation Detector - ColComp 1 Detector - ColComp 2

238 Advanced Operation Manual

7 GC Output Signals

This changes the digital output. You cannot get both compensated and uncompensated digital data from the same detector at the same time. However, it does not affect the analog output.

Plotting a stored column compensation profile

1 Press [Analog Out 1] or [Analog Out 2].

2 Scroll to Type: and press [Mode/Type].

3 Select the profile to be plotted.

4 Press [Start].

GC Output Signals 7

Advanced Operation Manual 239

Test Plot

Test plot is an internally generated chromatogram that can be assigned to a signal output channel. It consists of three baseline- resolved, repeating peaks. The area of the largest is approximately 1 Volt- sec, the middle one is 0.1 times the largest, and the smallest is 0.01 times the largest.

Test plot can be used to verify the operation of external data processing devices without having to perform repeated chromatographic runs. It may also be used as a stable signal to compare the results from different data processing devices.

To use the Test Plot:

1 Press [Analog Out 1] or [Analog Out 2].

2 Scroll to Type: and press [Mode/Type].

3 Choose Test Plot.

4 Press [Start]. The plot will repeat until you press [Stop].

Test Plot is the default choice for the analog outputs.

240 Advanced Operation Manual

7 GC Output Signals

241

Agilent 7890 Series Gas Chromatograph Advanced Operation Manual

Agilent Technologies

8 Auxiliary Devices

Auxiliary Devices 242

About Auxiliary Pressure Control 242

About Aux Thermal Zone Control 243

About Cryo Trap Control 243

About Auxiliary Device Contacts 244

About the 24V Auxiliary Device Power Supply 244

About Auxiliary Columns 244

About Auxiliary Detectors 245

242 Advanced Operation Manual

8 Auxiliary Devices

Auxiliary Devices

About Auxiliary Pressure Control

Pressure units

There are two common ways of expressing gas pressures:

psia Absolute pressure, measured relative to vacuum.

psig Gauge pressure, measured relative to atmospheric pressure. This name is used because most pressure gauges have one side of the sensing element exposed to the atmosphere.

The two measurements are related by:

psia = psig + atmospheric pressure

Two-channel pressure control module (PCM)

The PCM has two different channels. Channel 1 may be either flow or pressure controlled and may be flow/pressure programmed. It is essentially identical to the packed column inlet flow module (see About the Multimode Inlet).

Channel 2 is pressure only, but may be used in either a forward- or back- pressure mode by changing connections.

Both channels are controlled by the same parameter list. The first two lines are for channel 1; the remaining lines are for channel 2.

The forward- pressure mode requires the user to supply a downstream flow resistance, possibly a frit.

The back- pressure mode is most useful with gas sampling valves. By connecting the sample exit line from the valve to the vent fitting of the PCM, pressure in the sample loop can be controlled.

Three-channel auxiliary pressure controller (Aux PCM)

All channels are 3- ramp programmable. Up to three modules may be installed, for a total of 9 pressure- regulated channels.

Auxiliary Devices 8

Advanced Operation Manual 243

Setting parameters for auxiliary pressure control

1 Press [Aux EPC #] and scroll to the channel you wish to control. Press [Enter].

2 Scroll to Initial pressure. Enter a value and press [Enter].

3 If desired, enter a pressure program using the time and rate functions.

4 Press [On/Yes] to apply Initial pressure and start the program.

About Aux Thermal Zone Control

There can be up to 3 aux thermal controlled zones.

These zones are 3- ramp programmable.

Run time events may be used to schedule specific temperatures during the run.

Setting parameters for the aux thermal zone control

1 Press [Aux Temp #] and scroll to the zone you wish to control. Press [Enter].

2 Scroll to Temperature. Enter a value and press [Enter].

3 Scroll to Initial time. Enter a value and press [Enter].

4 Scroll to Rate 1. Enter 0 to end the program here or a positive value to create a temperature program.

About Cryo Trap Control

When configured as a cryo trap, an aux thermal zone has the following parameters:

Temperature Displays the current trap temperature and setpoint. Press [On/Off] to turn the zone on or off.

If On, the cryo trap temperature will remain 10 C above the current oven temperature, following the oven program.

If Off, the GC does not control the trap temperature.

Cold Trap Temperature Enter the target temperature for cryo trap operation and press [Enter]. Enable this temperature using run time On and Off events for the aux thermal zone.

244 Advanced Operation Manual

8 Auxiliary Devices

To use the cryo trap during a run:

1 Press [Aux Temp #] and scroll to the zone you wish to control. Press [Enter].

2 Scroll to Temperature. Press [On/Off] to turn the zone on or off, as desired (typically, On).

3 Scroll to Cold Trap Temperature. Enter a value and press [Enter].

4 Press [Run Table].

5 Press [Mode/Type].

6 Scroll to Cryo trap. Press [Enter] to add a cryo trap event to the run table.

7 Enter a value for the Time: and press [Enter]. Scroll to Cryo trap cooling and press [On/Yes] to turn it on.

This event sets the cryo trap temperature to the Cold Trap Temperature setpoint. The trap remains at this temperature until another event turns it off, or until the run ends.

8 Enter another Cryo Trap event to turn the trap temperature Off at the desired run time.

This event sets the cryo trap to track the oven temperature + 10 C for the remainder of the run.

About Auxiliary Device Contacts

These contacts are controlled by the external valve drivers. See The external valve drivers on page 218.

About the 24V Auxiliary Device Power Supply

This is controlled by the external valve drivers. See The external valve drivers on page 218.

About Auxiliary Columns

Defines up to 6 columns (includes Col 1 and Col 2).

1 Press [Config][Aux Col #] and enter a column number. Press [Enter].

2 Define/configure the column.

Auxiliary Devices 8

Advanced Operation Manual 245

About Auxiliary Detectors

The GC supports up to two auxiliary detectors in addition to the Front and Back detectors that mount on the top of the oven.

Aux Det # 1 This can only be a TCD, and mounts in a carrier on the left side of the oven together with its flow module.

Aux Det # 2 This can only be an analog input board (AIB). It is used to receive and process data from a non- Agilent detector or other source.

Manualsnet FAQs

If you want to find out how the 7890 Series Agilent works, you can view and download the Agilent 7890 Series Chromatograph Operation Manual on the Manualsnet website.

Yes, we have the Operation Manual for Agilent 7890 Series as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Operation Manual should include all the details that are needed to use a Agilent 7890 Series. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent 7890 Series Chromatograph Operation Manual is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent 7890 Series Chromatograph Operation Manual consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent 7890 Series Chromatograph Operation Manual free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent 7890 Series Chromatograph Operation Manual, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent 7890 Series Chromatograph Operation Manual as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.