Agilent 1260 Infinity Manual Injector

Using the Manual Injector

Using the Manual Injector

Warnings and Cautions

NOTE This guide just briefly describes how to use the manual injector. Please refer to the User Manual for further details and important information.

WARNING Ejection of mobile phase

When using sample loops larger than 100 L, mobile phase may be ejected from the needle port as the mobile phase in the sample loop decompresses.

Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.

Agilent Technologies

Injecting Sample

LOAD Position

In the LOAD position (see Figure 1 on page 3), the pump is connected directly to the column (ports 2 and 3 connected), and the needle port is connected to the sample loop. At least 2 to 3 sample-loop volumes (more if better precision is required) of sample should be injected through the needle port to provide good precision. The sample fills the loop, and excess sample is expelled through the vent tube connected to port 6.

WARNING Splashing of solvent

When using the Needle Port Cleaner, empty the syringe slowly to prevent solvent from splashing back at you.

Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.

CAUTION Potential damage to the valve

Rinse the valve with water after using buffer solutions to prevent crystals from forming, which can cause scratches on the rotor seal.

2

Figure 1 LOAD Position

INJECT Position

In the INJECT position (see Figure 2 on page 3), the pump is connected to the sample loop (ports 1 and 2 connected). All of the sample is washed out of the loop onto the column. The needle port is connected to the vent tube (port 5).

Figure 2 INJECT Position

3

Complete Loop Filling

In complete-filling, the volume of sample injected is set by the volume of the loop (this includes the valve passages). This method produces the highest precision.

At least 2 to 3 sample-loop volumes (more if better precision is required) of sample should be injected through the needle port to provide good precision.

The sample fills the loop, and excess sample is expelled through the vent tube connected to port 6.

An excess of sample is needed because mobile phase near the wall of the loop is displaced slowly due to the laminar flow effect.

1 Turn the handle to the LOAD position

2 Insert the syringe into the needle port. You should feel slight resistance as the needle passes through the needle seal before it stops against the stator face.

3 Load the sample slowly onto the loop. Repeat this step for higher precision.

4 Leave the syringe in and turn the handle to INJECT.

Partial Loop Filling

If you only have small quantities of sample, this is the method of choice. In the partialfilling method the volume of sample injected is set by the syringe. In this method, no more than half a loop volume of sample should be loaded into the loop. For example, load no more than 10 L into a 20 L loop. With larger than half the loop volume, some of the sample is lost out Vent Line 6. This is because sample flows down the center of the loop at twice the average velocity due to the laminar flow effect.

1 In INJECT, use the Needle Port Cleaner to flush out the needle port with about 1 mL of mobile phase to flush out contamination from the earlier injection. This liquid will exit out Vent Line 5.

2 Insert the syringe into the needle port. You should feel slight resistance as the needle passes through the needle seal before it stops against the stator face.

3 Load the sample slowly onto the loop.

4 Leave the syringe in and turn the handle to INJECT.

Manualsnet FAQs

If you want to find out how the 1260 Infinity Manual Injector Agilent works, you can view and download the Agilent 1260 Infinity Manual Injector Chromatograph Instructions on the Manualsnet website.

Yes, we have the Instructions for Agilent 1260 Infinity Manual Injector as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Instructions should include all the details that are needed to use a Agilent 1260 Infinity Manual Injector. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent 1260 Infinity Manual Injector Chromatograph Instructions is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent 1260 Infinity Manual Injector Chromatograph Instructions consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent 1260 Infinity Manual Injector Chromatograph Instructions free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent 1260 Infinity Manual Injector Chromatograph Instructions, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent 1260 Infinity Manual Injector Chromatograph Instructions as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.