How to take care of your Agilent Max-Light Cartridge Cells

Technical Note

Thank you for purchasing a Max-Light Cartridge Cell for your Agilent G4212A/B or G7117A/B/C DAD detectors. These flow cells have been designed for maximum sensitivity and simplicity. To ensure maximum lifetime, we recommend following the Best Practices for these products as provided below.

General Information

Included part numbers:

p/n Description

G4212-60008 Max-Light Cartridge Cell (10 mm, V() 1.0 L)

G4212-60007 Max-Light Cartridge Cell (60 mm, V() 4.0 L)

G4212-60032 HDR Max-Light Cartridge Cell (3.7 mm, V() 0.9 L)

G4212-60038 ULD Max-Light Cartridge Cell (10 mm, V() 0.6 L)

G7117-60020 Max-Light Cartridge Cell LSS (10 mm, V() 1.0 L)

Table 1 Maximum pressure specification

Maximum operating pressure 70 bar

Maximum incidental pressure 150 bar

Best Practices

Best Practices

Avoid pressure increase related to sources behind the flow cell

Use a standard (0.7 mm ID) waste tubing on the Max-Light Cartridge Cell and ensure that there are no kinks or blockages.

Agilent strongly recommends installing a pressure relief valve (Inline Pressure Relief Valve Kit (G4212-68001)) when another module is installed in the flow path behind the flow cell (for example, LC/MSD, FLD, ELSD).

The pressure relief valve protects the flow cell from overpressure and can be cleaned and re-activated.

Do not touch the light inlet and outlet of the cell with your fingers. This will add a layer of contamination on the flow cell window and reduce the light throughput.

Do not let buffers stay for long times in the Max-Light Cartridge Cell.

Flush the Max-Light Cartridge Cell when finished with the application.

Use fresh, clean water, followed by isopropanol, if the cell is not in use for more than 1-2 days.

Flush out the isopropanol first with water, before switching to buffers again.

Avoid blockages within the flow cell

Dont use a VWD outlet tubing on a Max-Light Cartridge Cell, as it contains a restrictor.

Using PEEK-FS capillaries is not recommended.

In combination with the SST zero dead volume fitting (for example, at the inlet) the capillary could break, and the glass particles could block or damage the flow cell.

Be aware of possible precipitation. Buffers, such as KH2PO4, can precipitate crystals once the percentage of the organic solvent is too high.

From a chromatographic point of view, theres no need to always go up to, for example, 100 % ACN.

Consider installing an inline filter (1290 Infinity II Inline Filter Kit (5067-6189)) in front of the flow cell.

NOTE Pay attention to the pressure limit of the cell.

2

Best Practices

General handling recommendations

Do not expose the Max-Light Cartridge Cells to shocks.

Avoid storage in a drawer.

Storage in a dry cabinet is recommended.

Install the white plugs and black caps before storing.

Annually perform a preventive maintenance on the instrument.

Always use HPLC-grade or higher-grade solvents. Do not use technical grade isopropanol for flushing.

Replace water-based mobile phases daily.

Consider filtering solvents (also water) through 0.2 m filters.

Replace organic mobile phases at least every second day.

Always use fresh solvents and solvent bottles:

Do not refill the bottles straightaway.

Rinse the bottle first with fresh solvent, empty it, then fill it with fresh solvent.

Inspect solvent bottles and inlet filters for damage or discoloration.

If not in use for several days, flush the entire system with water and finally with isopropanol.

At installation, flush the entire system thoroughly with isopropanol and then with water before connecting the flow to the Max-Light Cartridge Cell.

3

Light Sensitive Samples (LSS) Aperture (G7117-60101) for Max-Light Cartridge Cell LSS 10 mm

Figure 1 Aperture G7117-60101

The Aperture should be installed for light-sensitive samples, which are likely to undergo photodegradation.

The Aperture reduces the intensity of the light entering the cell and thus limits the photodegradation effects.

The use of the Aperture increases the detector noise; for this reason, the use of the aperture is recommended for the light-sensitive samples only.

The flow cell performance without the installed Aperture is the same as of the standard Max-Light Cartridge Cell 10 mm.

To install the Aperture, simply screw it on the light inlet of the Max-Light Cartridge Cell LSS ( 10 mm, V() 1.0 L) (G7117-60020). Do not apply any tools to affix the Aperture. Finger-tight fixation is enough. Make sure that the Aperture is aligned with the light path and does not obstruct the light path (not tilted in relation to the light axis). Check that after installation, the light intensity is approximately 10 % of the original.

Store the Aperture in a clean environment protected from dust. During installation and deinstallation wear powder-free gloves to protect the Aperture from possible contamination.

When performing detector tests in Lab Advisor, remove the Aperture.

Aperture is not compatible with other Max-Light Cartridges. It can only be installed on the Max-Light Cartridge Cell LSS ( 10 mm, V() 1.0 L) (G7117-60020).

To cartridge To detector

Manualsnet FAQs

If you want to find out how the Max-Light Agilent works, you can view and download the Agilent Max-Light Cartridge Cells Technical Note on the Manualsnet website.

Yes, we have the Technical Note for Agilent Max-Light as well as other Agilent manuals. All you need to do is to use our search bar and find the user manual that you are looking for.

The Technical Note should include all the details that are needed to use a Agilent Max-Light. Full manuals and user guide PDFs can be downloaded from Manualsnet.com.

The best way to navigate the Agilent Max-Light Cartridge Cells Technical Note is by checking the Table of Contents at the top of the page where available. This allows you to navigate a manual by jumping to the section you are looking for.

This Agilent Max-Light Cartridge Cells Technical Note consists of sections like Table of Contents, to name a few. For easier navigation, use the Table of Contents in the upper left corner.

You can download Agilent Max-Light Cartridge Cells Technical Note free of charge simply by clicking the “download” button in the upper right corner of any manuals page. This feature allows you to download any manual in a couple of seconds and is generally in PDF format. You can also save a manual for later by adding it to your saved documents in the user profile.

To be able to print Agilent Max-Light Cartridge Cells Technical Note, simply download the document to your computer. Once downloaded, open the PDF file and print the Agilent Max-Light Cartridge Cells Technical Note as you would any other document. This can usually be achieved by clicking on “File” and then “Print” from the menu bar.